Professional Documents
Culture Documents
Matrix Effect
Matrix Effect
%ME= 28.0%
c (mg/Kg) Signal
0.01 2.52E+06
0.04 8.40E+06
0.1 3.53E+07
0.4 9.08E+07
1 2.93E+08
4 1.08E+09
c (mg/Kg) Signal
0.01 2.52E+06 slope 2.71E+08 0.00E+00 intercept
0.04 8.40E+06 3.21E+06 #N/A
0.1 3.53E+07 9.99E-01 1.33E+07
0.4 9.08E+07
1 2.93E+08
4 1.08E+09
tomato
1.60E+09
Matrix calibration function
1.40E+09
slope 3.65E+08 0.00E+00 intercept 1.20E+09
5.34E+06 #N/A
1.00E+09
9.99E-01 2.21E+07
8.00E+08
señal
6.00E+08
Calibration function in solvent
4.00E+08
%ME= 123.56%
*mejora ionizacion
correlating compound taht has exactly the same ms/ms transition como aldicarb
correlating matrix compound that increases our ionization efficiency of our compound
para tener unn metodo robusto necesitamos tener un efecto matriz cercano a 100%
(𝑝𝑜𝑠𝑡−𝑒𝑥𝑡𝑟𝑎𝑐𝑡𝑖𝑜𝑛 𝑠𝑝𝑖𝑘𝑒𝑑 𝑚𝑎𝑡𝑟𝑖𝑥)/ 〖𝐴𝑛𝑎𝑙𝑦𝑡𝑒 𝑠𝑖𝑔𝑛𝑎𝑙〗 _𝑠𝑜𝑙𝑣𝑒𝑛𝑡 ∗100%
1.60E+09
1.40E+09
1.20E+09
solvente
1.00E+09
8.00E+08 Linear
señal
(solvente)
6.00E+08
matrix
4.00E+08
efecto matriz