Water Equivalence of Some Plastic-Water Phantom Materials For Clinical Proton Beam Dosimetry
Water Equivalence of Some Plastic-Water Phantom Materials For Clinical Proton Beam Dosimetry
a r t i c l e i n f o a b s t r a c t
Available online 13 February 2012 Plastic-water phantom materials are not exactly water equivalent since they have a different elemental
Keywords: composition and different interaction cross sections for protons than water. Several studies of the water
Proton dosimetry equivalence of plastic-water phantom materials have been reported for photon and electron beams, but
Fluence correction factor none for clinical proton beams. In proton beams, the difference between non-elastic nuclear interac-
Plastic water tions in plastic-water phantom materials compared to those in water should be considered. In this
Monte Carlo work, the water equivalence of Plastic Waters (PW)1, Plastic Waters Diagnostic Therapy (PWDT)1 and
Water equivalent solid water (WT1)2 phantoms was studied for clinical proton energies of 60 MeV and 200 MeV. This was
done by evaluating the fluence correction factor at equivalent depths; first with respect to water and
then with respect to graphite by experiment and Monte Carlo (MC) simulations using FLUKA. MC
simulations showed that the fluence correction with respect to water was less than 0.5% up to the
entire penetration depth of the protons at 60 MeV and less than 1% at 200 MeV up to 20 cm depth for
PWDT, PW and WT1. With respect to graphite the fluence correction was about 0.5% for 60 MeV and
about 4% for 200 MeV. The experimental results for modulated and un-modulated 60 MeV proton
beams showed good agreement with the MC simulated fluence correction factors with respect to
graphite deviating less than 1% from unity for the three plastic-water phantoms.
& 2012 Elsevier Ltd. All rights reserved.
0969-8043/$ - see front matter & 2012 Elsevier Ltd. All rights reserved.
doi:10.1016/j.apradiso.2012.02.002
L. Al-Sulaiti et al. / Applied Radiation and Isotopes 70 (2012) 1052–1057 1053
by Monte Carlo codes (Palmans et al., 2002; Palmans and alpha particles for water, graphite and plastic-water materials at
Verhaegen, 2005; AL-Sulaiti et al., 2010; Palmans et al., 2010). equivalent depths were simulated. The contributions of other
The correction on the dose conversion due to the difference in the secondary particles to the dose is negligible (Paganetti, 2002).
fluence in a graphite and a water phantom was found to be unity P hR Emax,i i
with an uncertainty of about 0.3% for 60 MeV protons (Palmans i 0 FE,w,i ðEÞ Sel,irðEÞ dE
kf l ¼ P hR E w i ð3Þ
et al., 2002). max,i
FE,pl,i ðEÞ Sel,irðEÞ dE
i 0
In this study, the water equivalence for three plastic-water w
materials; Plastic Waters (PW), Plastic Waters Diagnostic (Sel,i/r)w in Eq. (3) is the mass collision stopping power for water.
Therapy (PWDT) and WT1 was studied. This was done by
evaluating the fluence correction factor kfl at water equivalent
depths; first with respect to water and then with respect to 2.2. Experiments
graphite since water-to-graphite fluence correction factor were
studied for 60 MeV protons in a previous study concerning the Two sets of measurements (Fig. 1) were performed at the
application of a graphite calorimeter to proton beam dosimetry 60 MeV clinical proton cyclotron in CCO: one to measure kfl with
(Palmans et al., 2002). In the present work Monte Carlo (MC) respect to graphite and another with respect to water. The beam
simulations were performed for 60 and 200 MeV beams and was collimated to a narrow field using a 4 mm diameter circular
experiments at the Clatterbridge Center for Oncology (CCO) brass collimator. Both experiments were performed using the
60 MeV un-modulated and full modulated beams (the latter refers un-modulated beam and the full modulated beam. The latter was
to a modulation leading to an SOBP which extends from the surface realized by using a PMMA modulator wheel to spread out the
to the distal edge). Bragg peak, i.e. with the plateau of the SOBP extending from the
surface until the deepest Bragg peak. For the first experiment,
a plane-parallel NACP 02 ionization chamber was placed at the
2. Methodology beam isocentre which was about 7 cm away from the collimator
‘fixed source detector distance (SDD)’, while another plane-parallel
2.1. Monte Carlo simulation Roos ionization chamber was placed just after the collimator for
monitoring purposes. Plates of different thicknesses of graphite and
FLUKA 2008.3c simulations (Fasso et al., 2005; Battistoni et al., plastic-water materials (PW, PWDT and WT1) were positioned
2007) of dose distribution and fluence spectra were performed for up-stream of the NACP 02 chamber. The second experiment was
both protons and secondary alpha particles in water, graphite and performed for PWDT and water only. A similar setup was used but
in three plastic-water phantom materials, PWDT, PW and WT1, in the NACP 02 chamber was placed at 10 cm away from the collimator
a cylindrical phantom of 4 cm radius. The simulations were done instead of 7 cm due to the water phantom size. The water phantom
for 60 MeV and 200 MeV mono-energetic proton pencil beams with a thin MICA widow (0.1 mm or 0.24 mm water equivalent) was
and 106 protons histories for each material. Dose distributions in based on a plate which was adjusted to be moved forward and
depth were scored by the USRBIN detector card in 0.1 mm bins. backward in 0.1 mm steps while the NACP 02 chamber remained in
The fluence spectra were scored by the USRTRACK detector card the same position. In this way the thickness of water upstream the
in 200 energy bins. In order to calculate the fluence spectra as a chamber was accurately controlled.
function of depth, the phantom was split into several separated The ratio of ionization chamber readings, Mw, in water phan-
regions of 0.1 mm water-equivalent thickness and with each of these tom and, Mpl, in plastic phantom at equivalent depths, which is
regions a USRTRACK detector was associated. For physics setting, defined as the fluence scaling factor, hpl, in the IAEA TRS-398 code
the DEFAULTS card set to HADROTHErapy was used in which of practice (IAEA-TRS-398, 2000), is equal to the fluence correc-
the charged particle transport threshold was 100 keV. In addition, tion factor as defined above if the perturbation correction factors
the PHYSICS card was used and set to SDUM¼ EVAPORATion to use for the ionization chamber in both water and plastic phantoms
the new hadronic evaporation model with heavy fragment evapora- are equal. Accepting this assumption, the fluence correction factor
tion and inclusion of heavy ion nuclear interactions. with respect to water at water equivalent depths, wkfl, is given by
The fluence correction factor was calculated by two methods
(Palmans et al., 2002): one was based on a dose ratio calculation
using Eq. (1).
Dw ðdweq Þ
kf l ¼ ð1Þ
Dpl ðdpl Þsw,pl
where Dw(dw eq) is the dose in the water phantom and Dpl(dpl) is the
dose in plastic phantom. sw,pl is the water to plastic-water stopping
power ratio at equivalent depths which was extracted from the
FLUKA calculation.. The depth in the phantom, dpl, and the corre-
sponding equivalent depth in water, dw eq, are related by Eq. (2).
dweq ¼ dpl cpl ð2Þ
The range scaling factor, cpl ¼ r80 w/r80 pl, is the ratio of r80 w
and r80 pl which are the ranges in water and plastic-water,
respectively, where the dose at the distal edge of Bragg-peak
dropped to 80% of the dose maximum.
In the second method, the fluence differential in energy (E) for
protons and for secondary particles as a function of depth was
simulated. The fluence correction factor in Eq. (3) is based on the Fig. 1. Schematic diagram of the experimental set up for depth dose distribution
calculated fluence spectra for all charged particle types, i. In in water (top) and graphite, PW, PWDT and WT1 plastic water phantoms (bottom)
this study, only the fluence spectra for protons and secondary using 4 mm narrow proton beam at CCO. (Palmans et al., 2010).
1054 L. Al-Sulaiti et al. / Applied Radiation and Isotopes 70 (2012) 1052–1057
w Mw
k f l ¼ hpl ¼ ð4Þ
M pl
g g Mg
k f l ¼ h pl ¼ ð5Þ
Mpl
1.02 modulated
un-modulated The results of FLUKA simulations for 60 and 200 MeV mono-
energetic beams in plastic-water phantoms were obtained via
1.01
two methods; by calculating the dose ratio using Eq. (1) and by
k fl
1.00
mass collision stopping powers using Eq. (4). The second method
included two cases. In the first case the ratio was calculated for
0.99 the proton fluence only while in the second case the calculation
0.0 0.5 1.0 1.5 2.0 2.5 3.0
was extended to include the fluence of secondary alpha particles.
dw-eq (g/cm2)
From the results for 60 MeV protons plotted in Fig. 4, the
Experiment - 60 MeV observation was made that the wkfl considering only the proton
fluence was almost unity up to the entire depth for all plastic-
1.02
k fl normalized
0.99 the wkfl value below unity can be explained by the presence of
0.0 0.5 1.0 1.5 2.0 2.5 3.0
carbon in plastic-water materials, as presented in Table 1, know-
dw-eq (g/cm2) ing that the alpha production cross-section is higher for carbon
than for water especially at energies less than 60 MeV as shown
Fig. 2. Fluence correction factor for PWDT with respect to water in 60 MeV
modulated and un-modulated proton beams at the CCO cyclotron, (a) before
in Fig. 6. Despite this difference, the wkfl was almost unity for the
normalization and (b) after normalization at the surface. The lines linking data three plastic-water phantoms over the entire depth in a 60 MeV
points are only provided as a guide for the eye. mono-energetic proton beam.
L. Al-Sulaiti et al. / Applied Radiation and Isotopes 70 (2012) 1052–1057 1055
Fig. 5 shows the fluence correction factor with respect to had a value of about 0.5% above unity at the surface and (ii) the
graphite, gkfl, for 60 MeV protons. Similarly, it was observed that g
kfl decreased with penetration depth by about 0.5% for all plastic-
both calculation methods (the one which based on the dose ratio water materials in both methods.
and the one based on the fluences of both proton and alpha
particles) were matching. The differences from wkfl are that: (i) gkfl
PWDT - 60 MeV mono-energetic
PWDT - 60 MeV mono-energetic 1.02
Fluence calculation (P)
1.02 Fluence calculation (P) Fluence calculation (P+α)
Fluence calculation (P+α) 1.01 Dose calculation
k fl
1.01 Dose calculation
k fl
g
1.00
w
1.00
0.99
0.99 0.0 0.5 1.0 1.5 2.0 2.5 3.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0
d w-eq (cm)
dw-eq (cm)
PW - 60 MeV mono-energetic
PW - 60 MeV momo-energetic
1.02 Fluence calculation (P)
1.02
Fluence calculation (P) Fluence calculation (P+α)
Fluence calculation (P+α) 1.01 Dose calculation
1.01
k fl
Dose calculation
k fl
g
w
1.00 1.00
0.99 0.99
0.0 0.5 1.0 1.5 2.0 2.5 3.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0
dw-eq (cm) d w-eq (cm)
WT1 - 60 MeV mono-energetic WT1 - 60 MeV momo-energetic
1.02 1.02
Fluence calculation (P) Fluence calculation (P)
Fluence calculation (P+α) Fluence calculation (P+α)
1.01 Dose calculation 1.01 Dose calculation
k fl
k fl
w
1.00 1.00
0.99 0.99
0.0 0.5 1.0 1.5 2.0 2.5 3.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0
dw-eq (cm) d w-eq (cm)
Fig. 4. FLUKA calculated fluence correction factors with respect to water for Fig. 5. FLUKA calculated fluence correction factors with respect to graphite for
plastic-water phantoms. (a) PWDT, (b) PW and (c) WT1 in a 60 MeV mono- plastic-water phantoms. (a) PWDT, (b) PW and (c) WT1 in a 60 MeV mono-
energetic proton beam as a function of water equivalent depth. energetic proton beam as a function of water equivalent depth.
Table 1
Elemental composition in terms of fraction by weight, mass density r, electron density, ranges and range scaling factor of plastic-water materials considered in this study.
a
Elemental composition as given by (Tello et al., 1995).
b
Elemental composition as given by (Araki et al., 2009).
c P
Electron density is given by (NA Z/A) ¼NA wi Zi/Ai.
d P
I-value is given by lnI¼ [ wi Zi/Ai] lnIi/(Z/A) (ICRU-37, 1984).
e
Calculated range in FLUKA.
1056 L. Al-Sulaiti et al. / Applied Radiation and Isotopes 70 (2012) 1052–1057
k fl
0.08 1.00
(barn/amu)
g
water
0.06 0.99 Fluence calculation (P)
0.98 Fluence calculation (P+α)
0.04 Dose calculation
0.97
0.02 0 5 10 15 20 25
0.00 d w-eq (cm)
0 50 100 150 200 250 300
PW - 200 MeV mono-energetic
E (MeV)
1.02
Fig. 6. Secondary alpha production cross-section per nucleon for different 1.01
incident proton energy, from ICRU-63 (ICRU-63, 2000).
1.00
k fl
g
0.99 Fluence calculation (P)
0.98 Fluence calculation (P+α)
Dose calculation
PWDT - 200 mono-energetic 0.97
1.02 0 5 10 15 20 25
1.01
d w-eq (cm)
1.02
w
0.99
Fluence calculation (P)
dw-eq (cm)
0.98 Fluence calculation (P+α)
PW - 200 mono-energetic 0.97
Dose calculation
1.02 0 5 10 15 20 25
1.01 d w-eq (cm)
1.00 Fig. 8. FLUKA calculated fluence correction factors with respect to graphite for
kfl
plastic-water phantoms. (a) PWDT, (b) PW and (c) WT1 in a 200 MeV mono-
w
0.99 Fluence calculation (P) energetic proton beam as a function of water equivalent depth.
0.98 Fluence calculation (P+α)
Dose calculation
0.97 penetration depth in different target materials (AL-Sulaiti et al.,
0 5 10 15 20 25
2010). Thus, for high proton energies the increase in gkfl at
dw-eq (cm) shallow depths of less than 10 cm indicates that the secondary
WT1 - 200 mono-energetic proton production is higher in graphite than in plastic-water.
1.02
1.01 4. Conclusion
1.00
k fl
Acknowledgment ICRU-37, 1984. Stopping Powers and Ranges for Electrons and Positrons. Report
37. International Commissions on Radiation Units and Measurements,
Bethesda, MD, USA.
The authors are extremely thankful to the staff at the Douglas Matsuzaki, Y., Date, H., Sutherland, K., Kiyanagi, Y., 2010. Nuclear collision
Cyclotron in the Clatterbridge Center for Oncology for their processes around the Bragg peak in proton therapy. Radiol. Phys. Technol. 3,
helpful assistance and beam delivery and would like to express 84–92.
McEwen, M.R., Niven, D., 2006. Characterization of the phantom material Virtual
their gratitude to the Ministry of Environment, Qatar, for its support WaterTM in high-energy photon and electron beams. Med. Phys. 33, 876–887.
of Leena Al-Sulaiti. Paganetti, H., 2002. Nuclear interactions in proton therapy: dose and relative
biological effect distributions originating from primary and secondary parti-
cles. Phys. Med. Biol. 47, 747.
References Palmans, H., Verhaegen, F., 2005. Assigning nonelastic nuclear interaction cross
sections to Hounsfield units for Monte Carlo treatment planning of proton
beams. Phys. Med. Biol. 50, 991.
AL-Sulaiti, L., Shipley, D., Thomas, R., Kacperek, A., Regan, P.H., Palmans, H., 2010.
Palmans, H., Symons, J., Denis, J., DE Kock, E., Jones, D., Vynckier, S., 2002. Fluence
Water equivalence of various materials for clinical proton dosimetry by
correction factors in plastic phantoms for clinical proton beams. Phys. Med.
experiment and Monte Carlo simulation. Nucl. Instrum. Methods Phys. Res.
Biol. 47, 3055.
Sect. A 619, 344–347.
Palmans, H., AL-Sualiti, L., Andreo, P., Thomas, R., Shipley, D., Martinkovic, J. & A., K.
Araki, F., H. Y., Fukuoka, M., Matsumoto, K., Okumura, M., Oguchi, H., 2009. Monte
2010. Conversion of dose-to-graphite to dose-to-water in a clinical proton
Carlo calculations of correction factors for plastic phantoms in clinical photon
beam. Standards, Applications and Quality Assurance in Medical Radiation
and electron beam dosimetry. Med. Phys. 36, 2992–3001.
Dosimetry, Vienna, Austria, IAEA-CN-182/277.
Battistoni, G., S. M., Sala, P.R., Cerutti, F., Ferrari, A., & Roesler, S., A. F., Ranft, J.,
Schneider, U., Pemler, P., Besserer, J., Dellert, M., Moosburger, M., DE Boer,
2007. The FLUKA code: Description and benchmarking. Albrow M., Raja R.
J., Pedroni, E., Boehringer, T., 2002. The water equivalence of solid materials
(Eds.), In: Proceedings of the Hadronic Shower Simulation Workshop 2006,
used for dosimetry with small proton beams. Med. Phys. 29, 2946–2951.
Fermilab 6–8 September 2006, AIP Conference Proceeding 896, pp. 31–49.
F., Ranft, J. and Sala, P.R., 2005. FLUKA: a multi-particle transport code. CERN Seuntjens, J., Olivares, M., Evans, M., Podgorsak, E., 2005. Absorbed dose to water
Fasso,
reference dosimetry using solid phantoms in the context of absorbed-dose
2005-10 (2005), INFN/TC_05/11, SLAC-R-773.
protocols. Med. Phys. 32, 2945–2953.
IAEA-TRS-398, 2000. Absorbed Dose Determination in External Beam Radio-
Tello, V.M., Tailor, R.C., Hanson, W.F., 1995. How water equivalent are water-
therapy: An International Code of Practice for Dosimetry based on Standards of
equivalent solid materials for output calibration of photon and electron-
Absorbed Dose to Water. International atomic energy agency, Veinna TRS-398.
beams. Med. Phys. 22, 1177–1189.
ICRU-63 2000. Nuclear data for neutron and proton radiotherapy and for radiation
protection.: International Commission on Radiation Units and Measurements,
Bethesda, MD, USA, Report 63.