MP 120
KEY POINTS
- Easy to use - Selection of units
- Adjustable backlight - Manual autozero
- Hold-min-max function - Air velocity measurement
TECHNICAL FEATURES
compensation
VELOCITY SPECIFICATIONS ●
Selection of units
● Autozero Manual
Measuring units Measuring range Accuracy1 Resolution
● Hold function
m/s, fpm, km/h De 0 à 40 m/s From 5 to 40 m/s : ±0.5% of 0.1 m/s values
reading ±0,3 m/s
● Configurable auto shut-off
1
All the accuracies indicated in this technical datasheet were stated in laboratory conditions, and can be guaranteed for measurements carried out in the same conditions, or carried out
with calibration compensation ● Backlight
8.5 mm
● Transport case (ref : ST 110)
FTang_portable –MP120 – 11/08/16 – RCS (24) Périgueux 349 282 095 Document non contractuel – Nous nous réservons la possibilité de faire évoluer les caractéristiques de nos produits sans préavis.
147.9 mm
ACCESSORIES
CQ 15 : Magnetic protective
housing
Voltage
+ -
+ - MT 51 : ABS transport
+ - case
Pressure + -
Vacuum
+ -
+ -
+ - MAINTENANCE
Silicon layer
We carry out calibration, adjustment and
maintenance of your instruments to
guarantee a constant level of quality of
Pitot tube your measurements. As part of Quality
Dynamic pressure is measured by Pitot tube: Assurance Standards, we recommend you
Pd = Total pressure (Pt) – Static pressure (Ps) Total pressure to carry out a yearly checking.
Static pressure
Air velocity is calculated according to Bernoulli
simplified formula. Pressure sensor
GUARANTEE
Formula with temperature correction :
Po = Barometric pressure in Pa Instruments have 1-year guarantee for any
Vm/s = K x = Temperature in °C manufacturing defect (return to our After-
574,2 + 156842,77 x P en Pa K = Pitot tube coefficient Sales Service required for appraisal).
P0