Professional Documents
Culture Documents
ARt EFActoos
ARt EFActoos
Doppler Artifacts1
Myron A. Pozniak, MD
James A. Zagzebski, PhD
KatbleenA. Scanlan, MD
U INTRODUCTION
Doppler evaluation of hemodynamics provides useful clinical information. The pres-
ence or absence of flow in a vessel, flow direction, pulsatility, and velocity can all be
studied with this modality. In addition, color Doppler provides a visual image of the
flow characteristics, profiling regions of turbulence. With both modalities, correct
interpretation of tracings and flow images requires knowledge of physical and tech-
nical factors that influence Doppler signals. Artifacts caused by physical limitations
of the modality or inappropriate equipment settings can result in displayed flow
conditions that may differ considerably from the actual physiologic situation. Under-
standing limiting factors and appropriately modifying instrument settings should
minimize misdiagnosis.
In this article, we review artifacts in spectral and color Doppler imaging. We have
grouped them into three categories: those resulting from inappropriate equipment
I From the Department of Radiology, University ofWisconsin Hospital and Clinics, 600 Highland Ave. Madison, WI
53792 (M.A.P.,J.A.Z., K.A.S.), and the Department ofMedical Physics, University ofWisconsin, Madison (J.A.Z.). Recipi-
ent ofa Certificate ofMerit for a scientific exhibit at the 1990 RSNA scientific assembly. Receivedjune 10. 1991; revision
requested August 23 and received September 24; accepted September 26. Supported in part by a grant from Radiation
Measurements, Middleton, Wisconsin. Address reprint requests to M.A.P.
‘ RSNA, 1992
35
a. b.
Figure 1. Images were generated of a common
sonographic gray-scale phantom. Transducer was
moved back and forth across the phantom to gener-
ate a Doppler shift. (a) With proper gain settings, a
clear distinct velocity envelope is displayed. Flow
manifests in a uniform direction with a window un-
der the systolic peak indicating a narrow range of
velocities at any point in time. (b) With a moderate
increase in gain, the velocity tracing is thickened,
that is, the spectrum broadens (*). Systolic window
becomes filled in (**). Additionally, a mirror image
tracing that suggests reversed flow is also displayed
(arrow). (c) Further increase in gain produces
spikes over the tracing where the detection circuitry
is overwhelmed by the higher velocity shifts. rn/s =
meters per second.
settings, those arising from anatomic factors, tive to the transducer, flow velocity, and im-
and artifacts related to physical and technical portant flow conditions such as turbulence
limitations of the modality. Phantom and clini- are inferred from this display.
cal examples are shown in both color flow Most color flow instruments derive a flow
imaging and spectral Doppler. image by applying Doppler signal processing
and detection to the echo signal waveform
U DOPPLER SIGNAL and then estimating the mean Doppler fre-
Doppler processing detects phase or fre- quency at multiple locations along each beam
quency shifts in returning echo signals when line (1). Different methods are used by manu-
there is relative motion between scatterers facturers for this estimation. All methods re-
and the ultrasound
transducer. The Doppler quire a series ofperhaps eight to 15 pulse-
equation,f,) 2f)Vcos(O)/c,
= gives the magni- echo sequences along each beam line to
tude of the frequency shiftJ) when reflectors obtain sufficient data for estimating the mean
are moving with velocity V and angle 0 relative Doppler frequency within each pixel. This
to the ultrasound beam.f, is the ultrasound prolonged data acquisition time reduces the
frequency, and c is the speed of sound in the frame rate below that of conventional B-mode
medium. In addition to an audio presenta- imaging and introduces trade-offs between
tion, frequency (or spectral) analysis is pre- color image quality, image size, and temporal
sented as a tracing of Doppler frequency dis- resolution. The color image portrays mean
tribution versus time. This display may also be flow velocity and direction relative to the
presented in terms of velocity by solving the transducer for pixels throughout the targeted
above equation for V. Direction of flow rela- field.
#{149}
Incorrect Wail-Filter Setting U ANATOMICALLY RELATED
Filtration is designed to remove unwanted ARTIFACTS
low-frequency Doppler signals originating
from slowly moving soft-tissue reflectors. The #{149}
Mirror Image Artifact
cutoff frequency is operator selectable. With Mirror image artifacts are commonly observed
filtration set too high, diagnostically signifi- in conventional B-mode imaging (3). A similar
cant velocity information can be lost (2). This artifact can occur with color Doppler imaging
is especially true in low-flow states and in the of any vessel adjacent to a highly reflective
measurement of resistive index (Fig 5). surface, such as the lung (4). The subdia-
phragmatic region of the liver and the supra-
clavicular region are most notorious for this
artifact (Fig 6) (5,6).
Figure 6. (a) Image of the supraclavicular region shows two apparent subclavian veins. The more anterior
vessel is the true vein. The deeper vessel below the pleura (arrows) is actually an artifactual mirror image
reflecting off the lung apex. (b) Longitudinal image of the inferior vena cava. A second vessel is perceived
deep to the true inferior vena cava (arrows). Doppler shift is more prominent in the mirrored vessel. As the
sound beam reflects from the diaphragm, it interrogates the inferior vena cava flow at less of an angle,
thereby undergoing a greater Doppler shift than when it interrogates the true vessel, which is more perpen-
dicular to the ultrasound beam. (c) Schematic shows the course of the true signal (large arrows) and the mir-
rored signal (small arrows). Increased distance that the mirrored signal travels causes the phantom vessel to
project deeper in the image.
#{149}
Vascular-Motion Artifact velocity peripheral flow and back. This can
When a vessel that is moving relative to the introduce or increase periodicity of the portal
transducer is interrogated, artifactual pulsatil- venous flow pattern, an appearance sugges-
ity can be introduced into the spectral tracing. tive of, for example, tricuspid regurgitation.
This artifact can be perceived in the portal Increasing the size of the Doppler gate to in-
vein and its branches. Transmitted cardiac elude the entire vessel, imaging other portal
contraction pushes on the liver and tugs on branches, or varying the angle may decrease
the inferior vena cava. This rocking action of this artifact (Fig 7).
the liver causes slight motion of the vessels
with respect to the Doppler sample volume. #{149}
Color in Nonvascular Structures
Even a small degree of motion is sufficient to An area of low echogenicity such as a cyst or a
displace the sample volume from the higher- duct is susceptible to color flash artifact. Any
velocity central laminar flow to the slower- motion of a reflector relative to the trans-
ducer produces a Doppler shift. Most color
flow processors incorporate motion discrimi-
nators that separate true flow from random
motion of soft-tissue reflectors. However, the they usually manifest as a tracing both above
lower level signals arising from hypoechoic and below the spectral baseline. At higher
soft-tissue regions less effectively trigger the gain settings, directional ambiguity is worse
motion discriminator and do not suppress the and the velocity envelope on the spectral trac-
resultant color flash. Even minute motion may ing becomes more indistinct. The ambiguity is
produce artifactual color signals within di- easily corrected when the interrogating beam
lated bile ducts, cysts, or the gallbladder (Fig direction is shifted to an angle at either side
8). This can be erroneously perceived as flow, of 90#{176}
(Fig 9). When the sample volume is po-
especially if the color sensitivity settings are sitioned on a sector image, care should be
high (7). taken to avoid that portion of a vessel that
intercepts the interrogating beam at 90#{176}.
With
U INSTRUMENT- AND a linear transducer, electronic angulation or
PROCESSOR-RELATED ARTIFACTS wedge standoffs can improve the angle of
spectral Doppler interrogation (8).
#{149}
Directional Ambiguity
Directional ambiguity can result when the
interrogating beam intercepts the vessel at a
90#{176}
angle. If Doppler signals are detected,
When a color Doppler image is produced (Fig 1 1). Side lobes occur in proximity to the
with a sector transducer, flow perpendicular primary beam. Grating lobes, however, can be
to the beam usually occurs over a small seg- quite far removed from the central beam (9).
ment of a vessel parallel to the transducer sur- The exact configuration of these lobes de-
face (Fig 10). With a linear transducer, this pends on the construction of the transducer,
becomes a more significant problem, but that is, crystal element size and spacing.
again, electronic beam steering or wedge Tightly curved, convex-array transducers are
standoffs can compensate. most susceptible, as are higher frequency lin-
ear transducers. If these off-axis lobes of
#{149}
Grating- or Side-Lobe Artifact sound strike a highly reflective surface, they
Electronically focused, phased-array transduc- can return to the transducer and cause mis-
ers concentrate the primary interrogating registration of an object on the display of the
beam toward the Doppler sample volume. primary beam. This artifact commonly con-
Because of the spacing of the array elements fuses needle biopsy guidance (10, 1 1).
and the sequence of firing, however, weak Off-axis lobes can interrogate vessels unre-
secondary lobes of focused sound may inter- lated to the Doppler sample volume. If a
rogate areas unrelated to the primary beam strong enough signal is encountered off axis,
it may appear on the spectral Doppler tracing,
displaying a Doppler shift where none is ex-
pected (Fig 12).
#{149}
Spectral-broadening Artifact target traverses the sample volume, the Dop-
Spectral broadening is one of the characteris- pier angle varies. This introduces a slightly
tics identified in stenotic flow patterns. There- different Doppler frequency shift for each po-
fore, it is an important clinical sign. However, sition of the target and hence a broadened
spectral broadening may be introduced arti- spectrum.
factually and can cause errors in interpreting Spectral broadening is dependent on the
peak velocities in spectral waveforms. The angle of insonation (Fig 1 3) and the velocity
artifact occurs in pulsed and continuous-wave of the blood. Spectral broadening generally
Doppler because of the finite dimensions of becomes more severe for angles closer to 90#{176}.
the sample volume. A target traveling through Thus, errors can be minimized by using
the sample volume at a given speed intro- sound-beam orientation that results in a Dop-
duces a range of Doppler frequencies cen- pler angle of less than 60#{176}.
tered about the “true” Doppler frequency (ie,
the Doppler frequencyf0 calculated with the
equation mentioned previously) (12). As the