Tracheal Tube Cuff Care
Tracheal Tube Cuff Care
PURPOSE:
The tracheal tube cuff helps stabilize the endotracheal or tracheal tube and maintains an adequate airway seal so
that air moves through the tube into the lungs. The cuff also may decrease the risk of aspiration of large food
particles, but it does not protect against aspiration of liquid.
• High-volume low-pressure cuffs allow a large surface area to come into contact with the tracheal wall, distributing the
pressure over a much greater area. The older cuff design (low-volume high-pressure) may require 40 mm Hg (54.4
cm H2 O) to obtain an effective seal and is undesirable.
• The amount of pressure and volume necessary to obtain a seal and prevent mucosal damage depends on tube size and
design, cuff configuration, mode of ventilation, and the patient’s arterial blood pressure.
• A variety of devices are available to measure cuff pressures, including bedside sphygmomanometers, special aneroid
cuff manometers, and electronic cuff pressure devices. Ideally, the cuff pressures should be between 20 and 25 mm
Hg and still meet the goals of cuff use. Tracheal capillary perfusion pressure is 25 to 35 mm Hg for patients with
normotensive conditions. Lower cuff pressures are associated with less mucosal damage but also are associated with
silent aspiration, which has been shown to be more prevalent at cuff pressures less than 20 mm Hg.1,6,8
• Two techniques, minimal leak technique (MLT) and minimal occlusion volume (MOV), are used to inflate and
monitor air in the cuff.
The MLT involves air inflation of the tube cuff until any leak stops; then, a small amount of air is removed slowly
until a small leak is heard on inspiration. Problems with this technique include difficulty maintaining positive end
expiratory pressure (PEEP), aspiration around the cuff, and increased movement of the tube in the trachea during
cuff deflation.2,4,5,7,8 Aspiration may be prevented with deep pharyngeal suctioning before use of the MLT.
The MOV consists of injection of air into the cuff until no leak is heard, then withdrawal of the air until a small leak
is heard on inspiration, and then addition of more air until no leak is heard on inspiration.2,4,5,7,8
• Each technique has distinct advantages. MLT decreases tracheal mucosal injury and assists in mobilizing secretions
forward into the pharynx. MOV is used if the patient needs a seal to provide adequate ventilation or is at risk for
aspiration.4,8
• Although rare since the use of high-volume low-pressure devices became common, the adverse effects of tracheal tube
cuff inflation include tracheal stenosis, necrosis, tracheoesophageal fistulas, and tracheomalacia. These complications
may be more likely to occur in conditions that adversely affect tissue response to mucosal injury, such as hypotension.
Two major mechanisms are mainly responsible for airway damage: tube movement and pressure. Duration of
intubation also plays a significant role.4,8
• Routine cuff deflation is unnecessary and is no longer recommended.4
• Unintentional extubation and tube manipulation can occur with ineffective patient restraint or sedation, inadequate
securing of the tube, incorrect tube size and length, improper support or respiratory underinflation of endotracheal
cuff, and prolonged intubation.4
EQUIPMENT
• 10-mL syringe
• Pressure manometer with extension line or specially designed manometer to measure cuff pressures
• Three-way stopcock
• Stethoscope
• Self-inflating manual resuscitation bag-valve device
• Oxygen source and tubing
Additional equipment (for cuff inflation with faulty inflating device) includes the following:
• Scissors
• Padded hemostats
• Short 18-gauge or 23-gauge blunt needle
• Tongue depressor
• Tape (1 inch wide)
• Reintubation equipment, in case of accidental extubation
• Suction supplies (see Procedure 12)
Patient Preparation
• Ensure that the patient understands preprocedural teachings. Answer questions as they arise, and reinforce
information as needed. Rationale: This communication evaluates and reinforces understanding of previously
taught information.
• Place patient in semi-Fowler’s position. Rationale: This positioning promotes general relaxation, oxygenation, and
ventilation. It also reduces stimulation of the gag reflex and risk of aspiration.
Procedure for Tracheal Tube Cuff Care
FIGURE 13-2 Measuring cuff pressure w ith a homemade pressure monitor. (From Eubanks DH, Bone RC: Comprehensive respiratory care, ed 2, St Louis, 1990,
Mosby.)
FIGURE 13-3 Attachments for emergency cuff inflation for faulty inflation line. (From Sills J: An emergency cuff inflation technique, Respir Care 31:200, 1986.)
References
1. Hess, D. Tracheostomy tubes and related appliances. Respir Care. 2005; 50(4):497–509.
2. MacIntyre, N, Branson, R. Mechanical ventilation,, ed 2. Philadelphia: Saunders; 2009.
3. Morris, L, Zoumalan, R, Roccaforte, D, et al, Monitoring -tracheal tube cuff pressures in the intensive care unit. a
comparison of digital palpation and manometry. Ann Otol Rhinol Laryngol. 2007; 116(9):639–642.
4. Pierce, L, Airway maintenance. Management of the mechanically ventilated patient. ed 2. Saunders, St Louis, 2007.
5. Plambeck, A, Adult ventilation management. Corexcel. www.corexcel.com/courses/vent.htm, 2004 [retrieved
April 22, 2004].
6. Roman, M. Tracheostomy tubes. Medsurg Nurs. 2005; 14(2):143–145.
7. St John R, Protocols for practice. airway management. Crit Care Nurse. 2004; 24(2):93–96.
8. Urden, L, Stacy, K, Lough, M, et al, Thelan’s critical -care nursing. diagnosis and management. Mosby, St Louis,
2005.
Additional Reading
Winn, M, Right, K, Trac heostomy. a guide to nursing c are. Austr Nurs J. 2005; 13(5):1–4.