Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

Acta Biomaterialia 6 (2010) 2264–2273

Contents lists available at ScienceDirect

Acta Biomaterialia
journal homepage: www.elsevier.com/locate/actabiomat

Functionally graded hydroxyapatite coatings doped with antibacterial components


Xiao Bai a, Karren More b, Christopher M. Rouleau c, Afsaneh Rabiei a,*
a
Mechanical and Aerospace Engineering, North Carolina State University, Raleigh, NC 27695, USA
b
Materials Science and Technology Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA
c
Center for Nanophase Materials Sciences, Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA

a r t i c l e i n f o a b s t r a c t

Article history: A series of functionally graded hydroxyapatite (FGHA) coatings incorporated with various percentages of
Received 13 July 2009 silver were deposited on titanium substrates using ion beam-assisted deposition. The analysis of the coat-
Received in revised form 26 November 2009 ing’s cross-section using transmission electron microscopy (TEM) and scanning transmission electron
Accepted 1 December 2009
microscopy equipped with energy dispersive X-ray spectroscopy has shown a decreased crystallinity
Available online 5 December 2009
as well as a distribution of nanoscale (10–50 nm) silver particles from the coating/substrate interface
to top surface. Both X-ray diffraction and fast Fourier transforms on high-resolution TEM images revealed
Keywords:
the presence of hydroxyapatite within the coatings. The amount of Ag (wt.%) on the outer surface of the
Ion beam processing
Hydroxyapatite coating
FGHA, as determined from X-ray photoelectron spectroscopy, ranged from 1.09 to 6.59, which was about
Adhesion strength half of the average Ag wt.% incorporated in the entire coating. Average adhesion strengths evaluated by
Antimicrobial pull-off tests were in the range of 83 ± 6 to 88 ± 3 MPa, which is comparable to 85 MPa for FGHA without
silver. Further optical observations of failed areas illustrated that the dominant failure mechanism was
epoxy failure, and FGHA coating delamination was not observed.
Ó 2009 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.

1. Introduction Among those, Ag incorporation has gained more attention due to


its broad-spectrum antibacterial properties [4,8]. The most com-
Infection, which results in morbidity and implantation failure, is mon technique to incorporate Ag into HA coatings is via an ion ex-
a serious problem associated with prosthetic implantation, and change method, in which the Ca ions in HA are replaced by Ag ions
usually leads to subsequent implant removal and debridement while dipping the HA coatings into AgNO3 for a period of time
[1–3]. The infection mechanism has been postulated as bacterial [2,4,24]. The limitation of the ion exchange method is that Ag will
colonization of the prosthesis at the time of implantation by direct reside mostly on the outer surface of the coating and will be
inoculation during manipulation of tissue and implantation, or as a quickly depleted in vivo/in vitro without long-term antibacterial
result of contaminated sterile implants [3]. In 1033 cases of total effect. In order to achieve the continuous release of Ag, HA coatings
hip and total knee prosthetic arthroplasty infections, the major doped with Ag through the entire thickness has been developed
ones have been found to be caused by aerobic cocci, most com- using sol–gel [10,18], co-sputtering [5,9], and thermal or cold
monly Staphylococcus aureus (23%), and coagulase-negative staph- spraying [25,26]. Although Ag in small percentages can have an
ylococci (25%) [3]. The overall rate of prosthetic joint infection is antibacterial effect, larger amounts can be toxic [9], and therefore
highest in the first 6 months postoperatively, and declines contin- optimization of the Ag concentration in the coating is critical to
uously thereafter [3]. It has also been observed that the combined guarantee an optimum antibacterial effect without cytotoxicity.
incidence of rates of total hip and knee arthroplasty infections dur- It has been reported that the presence of 2 wt.% Ag in HA coating
ing the first and second postoperative years, and after 2 years, were processed by co-sputtering can have significant antibacterial effect
approximately 6.5, 3.2 and 1.4 per 1000 joint-years, respectively to Staphylococcus aureus and Staphylococcus epidermidis without
[3]. osteoblast-precursor cell cytotoxicity [9].
Hydroxyapatite (HA)-coated implants incorporated with anti- In this study, we investigated processing and characterization of
microbial agents are able to prevent or cure infections by releasing functionally graded hydroxyapatite (FGHA) coatings incorporated
directly the antimicrobial agents to local regions. HA incorporated with Ag as an antibacterial component. The processing technique
with Ag/Ag+ [2,4–22], Cu2+/Zn2+ [18,21], ampicillin [23] and doxy- was ion beam-assisted deposition (IBAD) with in situ heat treat-
cycline [12] have been studied to evaluate their antibacterial effect. ment. While the amorphous top layer of the coating allows a high-
er release rate of silver due to its higher dissolution immediately
after implantation, the crystalline layer will maintain the silver
* Corresponding author. Tel.: +1 919 513 2674; fax: +1 919 515 7968.
as a reservoir, resulting in long term infection protection. It is
E-mail address: arabiei@eos.ncsu.edu (A. Rabiei).

1742-7061/$ - see front matter Ó 2009 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.actbio.2009.12.002
X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273 2265

anticipated that this technique will be more efficient in infection SEM–EDS) in FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3, respectively.
control than conventional oral/intravenous antibiotic intake with For comparison of coating properties, FGHA coatings without silver
fewer side effects. In this study, the processing of FGHA coatings were deposited using the second HA target using the same deposi-
with three different silver contents is reported along with their tion parameters.
mechanical and microstructural properties. The biological re-
sponses of these coatings will be reported in a future article. 2.2. Materials characterization

2.2.1. Transmission electron microscopy and scanning transmission


2. Materials and methods electron microscopy
A combined transmission electron microscope/scanning trans-
2.1. Material processing mission electron microscope (TEM/STEM, Hitachi model HF3300)
equipped with an energy dispersive X-ray spectrometer (EDS,
Titanium disc substrates of 7.62 mm thickness were machined Thermo Scientific) and operated at 300 kV was used to examine
from a 12.7 mm diameter, 99.5% commercially pure titanium rod the cross-sectional microstructure of each of the coatings. The
(Alfa Asear). Prior to deposition, the Ti substrate surfaces were pre- cross-section samples for TEM and STEM analysis were prepared
pared by wet grinding with 240-, 400- and 600-grit silicon carbide in a dual-beam focused ion beam (FIB) instrument (Hitachi
paper (Buehler) and subsequent polishing with 9, 3 and 1 lm poly- NB5000) using an in situ lift-out technique.
crystalline diamond suspension (Buehler). In between each of the
grinding and polishing steps the discs were washed and ultrason- 2.2.2. Scanning electron microscopy–energy dispersive spectroscopy
ically cleaned to prevent cross contamination of abrasive particles. A scanning electron microscope (SEM, Hitachi S3200) equipped
After polishing and grinding the discs were ultrasonically cleaned with an EDS was used to examine the composition of the different
in acetone and isopropyl alcohol for 10 min each and passivated coatings from the top surface down to the FGHA/Ti interface. A
by exposing the discs to a 40% (by volume) nitric acid solution at minimum of three random locations on the coating surface were
room temperature for 30 min. The discs were then rinsed with selected for analysis. The instrument was calibrated by measuring
water and dried by compressed air (ASTM F86-91). and comparing a pure Mn standard to the ideal for Mn.
Two sintered HA targets (Cerac Inc.) of 152.4 mm diameter,
6.6 mm thickness and 99% purity were used to deposit either pure
2.2.3. X-ray diffraction
FGHA or FGHA-Ag coatings in order to avoid cross contamination.
X-ray diffraction (XRD, Xpert-Pro Powder Diffractometer) was
5 N purity silver wires with various diameters (ESPI) were affixed
used to identify various phases in the FGHA coatings along with
to the top of the target to dope the FGHA coatings with various
the blank Ti and HA target. The coatings were analyzed with CuKa1
concentrations of silver.
(k = 0.1540598 nm) radiation with 2h between 10° and 70° (with a
The IBAD system (Ion Tech.) employed to synthesize the coat-
step size of 0.017°). Results were then compared to JCPDS powder
ings consisted of a primary 5 cm Kaufman ion source for sputtering
diffraction standards to confirm the presence of crystalline HA, Ag
the target and a secondary 15 cm Kaufman ion source for bom-
or any other calcium phosphate compounds that might be present
barding the substrate/coating surface; both primary and secondary
in the coating.
beams were aligned to focus on the center of target and substrate
holder. The incident angle from the target and secondary beam was
45°. Substrates were heated by a radiant heater (Momentative Per- 2.2.4. X-ray photoelectron spectroscopy
formance Mater.) located behind the substrate holder, but without X-ray photoelectron spectroscopy (XPS, Thermo Scientific K-Al-
direct contact. A base pressure of 5  10 7 Torr was reached before pha) Ar+ depth-profiling was employed to quantify the chemical
starting each deposition run. composition of the coating throughout the thickness. Prior to the
The FGHA coatings were deposited over a total period of 6 h; the XPS measurements, the XPS was calibrated with the pure HA target
substrate temperature was 550 °C for the first 4 h and 450 °C for having a Ca/P ratio of 1.67.
the last 2 h in order to grow the functionally graded microstruc-
ture. Beam voltage and current density of the primary source were 2.2.5. Fourier transform infrared spectroscopy
1200 V and 8.92 mA cm 2, respectively, during deposition. Beam Fourier transform infrared spectroscopy (FTIR, Nicolet Nexus
voltage and current density of the secondary ion source, however, 470) with a smart orbit diamond attenuated total reflectance
were gradually decreased in order to facilitate a well-bonded inter- (ATR) insert was used to evaluate the structural and molecular
face and appropriate composition at the top surface of the coatings. composition of the coatings on the titanium substrates as well as
Manual rotation by 90° of the substrate holder was performed the HA target in order to identify the presence of the chemical
every 5 min in the first hour of deposition and every 15 min during groups OH and PO43 . Each sample was scanned with a resolution
the remainder of the deposition in order to improve the uniformity of 2 cm 1 with a scan number of 64.
of the coating. Water vapor was introduced into the chamber by
passing a 20 sccm argon gas flow through a water bubbler to pro- 2.2.6. Adhesion strength
vide more OH at the surface of the substrate and prevent dehydr- The adhesion strength of the coating to the titanium substrate
oxylation of HA. For the duration of the depositions, 10 and was evaluated using a Romulus materials tester (Pull off tester,
20 sccm of argon gas flow was provided to the primary and second- Quad Group). Epoxy coated studs with a cured adhesive strength
ary ion source, respectively, and 5 sccm of argon gas flow was pro- of 70–100 MPa were secured perpendicular to the coating surface.
vided to the plasma bridge neutralizer of each ion source. Using the The samples were cured for 1 h at 150 °C in an oven, then removed
parameters above, three sets of Ag-containing FGHA coatings and allowed to cool to room temperature in an ambient atmo-
(FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3) were deposited onto a total sphere. Using the Romulus Materials Tester IV and Romulus auto-
of 21 Ti substrates secured into a custom-made substrate holder mated program for coating adhesion strength, each sample was
for each run. One line of Ag wire with diameter 1 mm, two lines loaded at a rate of 26.7 N s–1 until failure, with a maximum load
of diameter 1 mm and two lines of diameter 1.5 mm were secured of 445 N. The adhesion strength was averaged from a minimum
on the surface of the target to provide an average Ag content of of four measurements for each set of coatings. Failure areas under
2.34 ± 0.37, 6.29 ± 0.24 and 13.53 ± 0.22 wt.% silver (measured by the stud were observed via an optical microscope (Zeiss) and fail-
2266 X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273

ure modes were then classified as epoxy failure (EF), coating cohe- talline bottom layers in all three coatings, with their sum
sive failure (CF) or coating delamination (CD). thickness around 2/3 of the total coating thickness, were formed
at a substrate temperature of 550 °C during the first 4 h of deposi-
tion; the initial interface layer was formed during the first hour of
2.2.7. Nanoindentation test
deposition with a beam voltage and current of 200 V and
Nanohardness and Young’s modulus values for all the coatings
0.28 mA cm 2, respectively, as well as a substrate rotation of 90°/
were determined using an MTS Nanoindenter IIÒ instrument (MTS
5 min. The top (surface) amorphous layer as shown in a-3, b-3
Instruments). The samples were indented with a Standard Berkovich
and c-3, on the other hand, was formed during the final 2 h of
indenter, with a strain rate set to 0.05 s 1 and an allowable drift rate
deposition with a substrate temperature of 450 °C and secondary
of 0.05 nm s 1. The maximum penetration depth was 1200 nm. The
beam parameters of 50 V/0.17 mA cm 2. Fast Fourier transform
tip was calibrated with fused silica (SiO2) and was cleaned with iso-
(FFT) on a high-resolution image of coating FGHA-Ag2 shows three
propanol between indentations. The Young’s modulus and hardness,
rings, with d-spacing of 2.8, 2.4 and 1.7Å, corresponding to (1 1 2),
measured as a function of indentation depth (only 100–200 nm in
(2 2 0) and (0 0 4) of HA. The selective area diffraction was used to
order to exclude the substrate effect), were determined using the
verify the transmission electron microscopy observation of the
Oliver–Pharr model [27] with the Poisson’s ratio set to 0.3 [28]. A
mostly amorphous top layer and crystalline bottom layer in the
minimum of 10 measurements from each coating were used to eval-
coatings. However, the high density of crystalline silver particles
uate the average nanohardness and Young’s modulus.
in the coating precludes us from having a meaningful selected area
diffraction (SAD) result. In addition, converging the beam (CBED) to
3. Results collect data only from the areas between Ag particles resulted in
rapid degradation of the sample. Therefore, a set of samples depos-
Fig. 1 shows TEM images of the cross-sections of samples FGHA- ited under the same exact substrate temperatures, but without
Ag1, FGHA-Ag2 and FGHA-Ag3 and Fig. 2 shows the high-resolu- including Ag, was used to identify the crystalline and amorphous
tion TEM image of FGHA-Ag2 and its fast Fourier transform. All regions in the coatings. TEM, SAD and XRD observations confirmed
the coatings consist of three distinct FGHA layers, as labeled in the crystalline nature of the coating deposited at 550 °C and mostly
images (a-1), (b-1) and (c-1) shown in Fig. 1: an equiaxed polycrys- amorphous nature of coating deposited at 450 °C (Figs. 3 and 4).
talline interface layer adjacent to the Ti, a crystalline bottom layer More detailed information on such experiments can be found else-
and a highly-amorphous top layer. Higher magnification images of where [29].
the coating interface and bottom layers are shown in images (a-2), The isolated dark particles shown in Fig. 1 were identified as
(b-2) and (c-2) of Fig. 1. The thicknesses of the coatings are consis- metallic Ag by STEM-EDS. Ag particles were found throughout
tently in the range of 700 to 800 nm. Both the interface and crys- the coating thickness in all three layers of the coating, for

Fig. 1. TEM images of cross-section of coatings FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3. (a-1, b-1 and c-1) Showing the entire thickness of coatings; (a-2, b-2 and c-2) close-up
of the interface areas; (a-3, b-3 and c-3) close-up of the top layer.
X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273 2267

Fig. 2. High-resolution TEM image of coating FGHA-Ag2 (left) and its FFT (right). The FFT shows three rings labeled by 1, 2 and 3 at the top; their measured d-spacing and
planes are also labeled.

Fig. 3. TEM and SAD analysis of coating without silver deposited at 450 and 550 °C.

FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3. There were, however, clear fusion resulted from a lower substrate temperature during the final
variations in their distribution and size. In coating FGHA-Ag1, the stage of deposition. It can be concluded that, in order to achieve a
Ag was mostly distributed in the transitional region between the more uniform distribution of silver throughout the thickness of the
top amorphous HA layer and the bottom crystalline HA layer. On coating, a minimum amount of silver (more than the amount of sil-
the other hand, coating FGHA-Ag1, which had the lowest percent- ver in FGHA-Ag1) is needed to facilitate the formation of an equi-
age of Ag, exhibited a columnar grain structure within the bottom axed grain structure in the crystalline layer.
layer. Coatings FGHA-Ag2 and FGHA-Ag3, however, with higher Fig. 5 shows the XRD data from the HA target, blank Ti, FGHA-
percentage of Ag, had a more equiaxed grain structure in their Ag1, FGHA-Ag2, FGHA-Ag3 and FGHA coatings. As can be seen in
crystalline layer. This can be attributed to the fact that silver can Fig. 5, the most intense peaks are HA(2 1 1), HA(3 0 0), HA(1 1 2)
act as nucleation sites for smaller, equiaxed HA grains in samples and HA(0 0 2) in the HA target. The HA peak (2 1 1) at 32.11°, ob-
with higher percentages of Ag at higher substrate temperatures served in all coatings, is broad enough to cover both HA(2 1 1)
(550 °C) during the first 4 h of deposition. An equiaxed grain struc- and HA(1 1 2), which are the two most intense peaks in the target.
ture will accommodate more silver atoms/particles compared to a Broad peaks of HA(3 2 3), CaO (2 2 0) and (2 2 2) were also ob-
columnar grain structure because the silver can be pushed away by served in all coatings. In general, all coatings show broader peaks,
the tip of the growing columnar grains. In the amorphous layer, which is due to the nanoscale structure of the coating. An Ag dif-
however, the Ag particle size is much smaller due to the lower dif- fraction peak was not clearly observed in coating FGHA-Ag1, which
is likely due to the low concentration of Ag in this coating. How-
ever, a broad Ag peak (2 0 0) has been clearly observed at 44.13°
in coatings FGHA-Ag2 and FGHA-Ag3. In general, the higher the
concentration of Ag in the coating, the stronger the diffraction
peaks in the XRD data.
Fig. 6 shows the average Ag wt.% evaluated from the coating top
surface by SEM–EDS and XPS. The average Ag wt.% for the entire
thickness of coatings FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3 mea-
sured by SEM–EDS were 2.34 ± 0.37, 6.29 ± 0.24, and 13.53 ± 0.22,
respectively, whereas the Ag wt.% measured on the coating top sur-
face by XPS were 1.09, 3.16 and 6.59, respectively. The average Ag
concentration for the entire coating is about twice that of the out-
Fig. 4. XRD analysis of coating without silver deposited at 450 and 550 °C. ermost coating surface, which confirmed that Ag was not uni-
2268 X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273

Measured by XPS Measured by SEM-EDS

4.0
3.5

Atomic Ca/P ratio


3.0
2.5
2.0
1.5
1.0
0.5
0.0
FGHA FGHA-Ag1 FGHA-Ag2 FGHA-Ag3

Fig. 7. Atomic Ca/P ratio evaluated by XPS and SEM–EDS on the top surface of
coatings.

For the FGHA coating, the Ca/P is 3.5 within the amorphous re-
gion and 4 within the crystalline region. Similar Ca/P trends were
observed in the coatings FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3.
However, atomic Ca/P ratios in the top layer of coatings FGHA-
Ag1, FGHA-Ag2 and FGHA-Ag3 were 3.5, 3.5 and 3.7, respectively,
whereas Ca/P was 5, 4.5 and 5 in the crystalline region, respec-
Fig. 5. XRD results of samples FGHA-Ag1, FGHA-Ag2, FGHA-Ag3 and FGHA.
tively. The Ag concentration steadily increased from the coating
top surface inward, and approached a maximum concentration at
the transition region between the top layer and bottom layer for
formly distributed within the coatings, as shown in Fig. 1. In addi-
all coatings. The Ag concentration steadily decreased within the
tion to the variation of Ag distributions on the cross section of coat-
bottom layer towards the interface with the Ti substrate. Another
ings, the high sputtering rate of Ag compared to HA could lower the
high concentration of Ag was observed at the interface region with
Ag concentrations at the coating surface because of ion beam etch-
the substrate, and this could be formed during the first half hour of
ing during XPS characterization.
deposition. It is well known that Ag has a higher sputtering rate
Fig. 7 shows the Ca/P measured from the coating top surfaces by
when compared to HA [30,31]. As a result, the density of Ag parti-
SEM–EDS and XPS. The average Ca/P ratios of coating FGHA, FGHA-
cles sputtered from the target and deposited on the substrate are
Ag1, FGHA-Ag2 and FGHA-Ag3 measured by SEM–EDS were
higher than HA. However, the re-sputtering from the heated sub-
2.33 ± 0.01, 2.95 ± 0.03, 2.19 ± 0.10 and 3.09 ± 0.03, respectively,
strate surface is greater for Ag particles than Ca and P during the
whereas the Ca/P ratios measured by XPS were 3.25, 3.33, 3.75
first few minutes, lowering the concentration of Ag in the coating.
and 2.96, respectively. The higher Ca/P ratio measured by XPS
Fig. 9 shows the FTIR spectra of the coatings and the HA target.
may also be related to the higher sputtering rate of P compared
The peaks are labeled with their corresponding structures. As can
to Ca, which may sputter more P from the sample during XPS
be seen in the HA target spectrum in Fig. 9, OH peaks were ob-
characterization.
served in stretching mode and OH liberation mode, while the sharp
In order to evaluate the variation of the Ca/P ratio and the Ag
peaks at 564 and 599 belong to v4 of PO43 . The tiny peak at
concentration throughout the coating thickness, XPS depth-profil-
879 cm 1 is v2 of CO32 whereas the peak at 963 cm 1 is v1 of
ing was performed on coatings FGHA, FGHA-Ag1, FGHA-Ag2 and
PO43 . The peaks at 1027 and 1091 cm 1 are v3 of PO43 . The peaks
FGHA-Ag3. These results are presented in Fig. 8. Fig. 8a, b–1, c-1
within 1300–1650 cm 1 are due to v3 vibration mode of surface
and d-1 showing the depth-profile for the primary elements ob-
carbonate ions instead of carbonate ions in the lattice of the mate-
served in each coating (P, Ca, O), while Fig. 8b–2, c-2 and d-2 show
rial, which have been observed in commercial pure HA [33]. The
the corresponding observation of the Ag concentration throughout
peaks within 1700–2500 cm 1 are noise of ATR–FTIR. Compared
the coating thickness. The profiles are segmented and labeled in
to the target material, all the coatings show the presence of OH-
Fig. 8 as interface, bottom and top layer, consistent with layer
at 3640 cm 1 [32]. However, the OH- peaks of stretching and libra-
thicknesses measured from images in Fig. 1. Ca, P and O were the
tional modes were not observed at 3570 and 633 cm 1, which is
primary elements in coating FGHA (Fig. 8a), and Ag, Ca, P and O
the same as the HA coating processed by ion beam mixing deposi-
were detected in coatings FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3.
tion [34]. In addition, all the coatings show a broad peak of PO43
v4 mode at 564 cm 1 instead of the two sharp peaks of PO43 v4
mode at 564 cm 1 and PO43 v3 mode at 599 cm 1 presented in
Measured by XPS Measured by SEM-EDS the target material. Moreover, a broad peak of PO43 v3 mode
16 was observed at 1004 cm 1 for all the coatings. It is concluded that
14 all coatings include OH and PO43 . These results, in combination
12 with the XRD and SAD results presented earlier, confirm the pres-
ence of functionally graded hydroxyapatite in the coating. It is
Ag wt %

10
notable that the presence of broader peaks in this study is attrib-
8
uted to the small thickness of the coating together with its partially
6
amorphous structure.
4
Table 1 lists the average adhesion strength results from pull-off
2
testing of the coatings. The average adhesion strengths for all coat-
0 ings fall within the range of 83–88 MPa. The fracture surfaces after
FGHA-Ag1 FGHA-Ag2 FGHA-Ag3
pull-off test were also observed using optical microscopy in order
Fig. 6. Ag wt.% evaluated by XPS and SEM–EDS on the top surface of coatings. to further evaluate the failure modes. The optical images are
X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273 2269

Fig. 8. Results of depth-profiling measured by XPS: (a) FGHA, (b) coating FGHA-Ag1, (c) FGHA-Ag2 and (d) FGHA-Ag3.

shown in Fig. 10. The associated failure modes are classified as Table 2 summarizes the results of the Young’s modulus and
coating cohesive failure, coating delamination and epoxy failure, hardness measured by nanoindentation. The coatings FGHA and
and are also summarized in Table 1. The epoxy failures given in Ta- FGHA-Ag1 have similar elastic moduli in the range of 77 to
ble 1 are the sum of both epoxy failures (EF) that occurred within 78 GPa, respectively, and hardness values ranging from 2.59 to
the epoxy and the epoxy delaminations (ED), as shown in Fig. 10. 2.97 GPa. However, the elastic modulus and hardness of coatings
There were no coating delaminations (CD) observed in any of coat- FGHA-Ag2 and FGHA-Ag3 are higher than those of FGHA and
ings, which indicates a well-bonded interface between the coating FGHA-Ag1, ranging from 101.85 to 106.92 GPa and from 4.17 to
and the substrate. 4.59 GPa.
2270 X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273

Fig. 9. FTIR results of FGHA, FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3 coatings along with the HA target.

Table 1 of deposition; with decreasing values of the secondary beam


The adhesion strength measured by the pull-off test. parameters to 100 V/50 mA in the second hour and 60 V/30 mA
in the third and fourth hours of deposition, the Ca/P ratio steadily
Coatings Average adhesion Failure modes
strength (MPa) deceased. During the first 4 h of deposition, the decreasing Ca/P
Coating Cohesive Epoxy was controlled primarily by the gradual reduction of beam volt-
delamination (%) failure (%) failure (%) age/current because the re-sputtering rate reduction for Ca is sig-
FGHA 85 ± 6 0 40 60 nificant when compared to that of P, although Ca and P both
FGHA-Ag1 83 ± 6 0 18 82 have reduced re-sputtering rates at a reduced beam voltage/cur-
FGHA-Ag2 86 ± 3 0 33 67 rent. In the final 2 h, the voltage was decreased to 50 V and the
FGHA-Ag3 88 ± 3 0 35 65
temperature decreased to 450 °C, which resulted in decreased Ca/
P ratios of 3.25, 3.33, 3.75 and 2.96 for the FGHA, FGHA-Ag1,
FGHA-Ag2 and FGHA-Ag3, respectively. The other primary factor
4. Discussion controlling the Ca/P ratio for this stage was the substrate temper-
ature that is significantly affecting the re-sputtering rate of Ca
The coatings’ microstructures, as well as their mechanical prop- when compared to that of P.
erties, resulted from manipulating deposition parameters and sub- The Ag concentration in the coating is critical to its antibacterial
strate temperature during deposition. The well-bonded interfaces effects and cytotoxicity. In particular, the concentration of Ag in
shown in Fig. 1 as well as the compositional transition interface the top amorphous region of the coating has more severe impact
shown in Fig. 8 illustrate that higher values for secondary beam on its antibacterial effect and cytotoxicity than the average amount
parameters of 200 V/50 mA at the beginning of deposition facili- of Ag within the entire coating. This is due to the fact that the top
tated the formation of a dense and compositionally graded struc- amorphous region is directly in contact with the body tissue, and is
ture. The high bond strength (greater than 83 MPa) without any the first layer to be gradually dissolved and thus enhance new bone
coating delamination for all the coatings verified the formation of growth by providing Ca and P, as well as inhibiting possible infec-
this well-bonded interface. In addition, standard deviations less tions by releasing Ag. The amorphous top surface, with the lowest
than 6.42 MPa (7.6%) of the bonding strength resulted from a high Ag concentration and highest dissolution rate, is able to quench
reproducibility of coatings due to continuous substrate rotation of infection immediately after implantation. Ag release then reaches
90o/5 min at the beginning of deposition. Although the in situ sub- a steady-state level since the high concentration of the Ag in the
strate temperature has been found to be the dominant factor con- crystalline layer will compensate for its lower release rate. Based
trolling the degree of crystallinity of HA coatings [29,35–39], a on the current observations, a silver concentration between the
difference in grain structures in coatings deposited at the same measured in coating FGHA-Ag2 and that in coating FGHA-Ag3
temperature has been observed. The columnar grain structure in should provide the best cell attachment/cell growth as well as anti-
the bottom layer of FGHA-Ag1 and the equiaxed grain structure bacterial effect. Further studies are needed to pinpoint the exact
in bottom layer of both FGHA-Ag2 and FGHA-Ag3, which were all optimum Ag concentration in FGHA coatings to achieve the best
deposited at 550 °C, suggest that Ag concentrations exceeding biological response.
6.29 wt.% may enhance the crystallization of the calcium phos- XRD revealed the broad peaks of HA along with calcium oxide
phate coating. In addition, dense coating structures throughout and silver in the coatings; particularly, the HA peaks (2 1 1) at
the coating thickness for all the coatings were resulted from ion 32.11° in all the coatings is broad enough to cover the most intense
bombardment by the secondary beam. peaks of HA(211) at 31.74° and HA(112) at 32.16° that appeared in
Similarly, the variation of coating composition throughout the both the HA target and JCPDS card (9–432). The broad peaks to-
coating thickness (as shown in Fig. 8) was a direct result of varying gether with the absence of some other HA peaks in these coatings
the deposition parameters of the secondary beam and the sub- compared to the HA target or JCPDS card (9–432) are due to the
strate temperature. The highest Ca/P ratio near the interface re- nature of the coating, which is partially amorphous, with some
sulted from the highest secondary beam parameters of 200 V/ nanoscale crystals much smaller than 100 nm in size. In a study
50 mA at a substrate temperature of 550 °C during the first hours of size effect on X-ray diffraction patterns of HA crystals ranging
X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273 2271

Fig. 10. Optical images of the pull-off area for coatings FGHA, FGHA-Ag1, FGHA-Ag2 and FGHA-Ag3.

Table 2
32.11° and HA(3 2 3) at 64.10° combined with FFT results indicated
Nanoindentation results.
the presence of HA phase. Compared to the fact that various cal-
Coatings Elastic modulus (GPa) Hardness (GPa) cium phosphate compounds instead of HA were observed in func-
FGHA 78.01 ± 8.47 2.97 ± 0.57 tionally graded coatings processed without the presence of water
FGHA-Ag1 77.04 ± 8.05 2.59 ± 0.45 vapor during deposition in our previous work [39], it can be con-
FGHA-Ag2 106.92 ± 9.47 4.59 ± 0.73 cluded that the presence of water vapor with a pressure of
FGHA-Ag3 101.85 ± 9.90 4.17 ± 0.55
2  10 4 Torr during deposition was favorable to the formation of
HA according to both XRD and FFT data. A significantly higher
from 0.5 lm to 50 nm, the well-defined characteristic HA peaks water vapor pressure of 0.25 Torr has been observed to better en-
were observed in HA crystals with an average length of about hance the formation of HA using pulsed laser deposition with
0.5 lm, while only two broad peaks ranging from 25° to 35° and water vapor [41]. It is suggested that a higher pressure of water va-
from 45° to 55° were observed in HA crystals with an average size por might be better to incorporate OH into coatings during future
of 50 nm [40]. In our coatings, both broad peaks of HA(2 1 1) at depositions.
2272 X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273

No significant difference of adhesion strength was observed be- Appendix A. Figures with essential color discrimination
tween the coatings with or without silver, suggesting that incorpo-
ration of Ag (2.34–13.53 wt.%) into the coating does not have any Certain figures in this article, particularly Figures 4–10, are dif-
significant impact on the coating’s adhesion strength. In addition, ficult to interpret in black and white. The full color images can be
the high adhesion strength (greater than 83 MPa) for all the coat- found in the on-line version, at doi: 10.1016/j.actbio.2009.12.002.
ings illustrates a well-bonded interface and overall good mechan-
ical properties of the coatings. The optical observations showed
that epoxy failure was the dominant failure mode during the References
pull-off test, with 60–80% of the failed area falling under this mode,
[1] Vaudaux PE, Daniel PL, Waldvgel FA. Host factors predisposing to and
indicating that the bond strength of the coating is superior to that
influencing therapy of foreign body infections. In: Bisno AL, Waldvogel FA,
of the epoxy. The superior bond strength of the coating can be editors. Infections associated with indwelling medical devices. Washington,
attributed to the presence of the intermixed interface layer and DC: ASM Press; 1994. p. 1.
[2] Shirkhanzadeh M, Azadegan M, Liu GQ. Bioactive delivery systems for the
dense coatings produced by IBAD.
slow-release of antibiotics – incorporation of Ag+ ions into micro-porous
Nanoindentation results from the coatings suggest that incor- hydroxyapatite coatings. Mater Lett 1995;24:7.
poration of Ag into the FGHA coatings generally resulted in an in- [3] Steckelberg JM, Osmon DR. Prothetic joint infections. In: Bisno AL, Waldvogel
crease in both the elastic modulus and nanohardness, with the FA, editors. Infections associated with indwelling medical
devices. Washington, DC: ASM Press; 1994. p. 259.
exception being coating FGHA-Ag1 as compared with coating [4] Feng QL, Kim TN, Wu J, Park ES, Kim JO, Lim DY, et al. Antibacterial effects of
FGHA. Apparently, the very low Ag content (2.34 ± 0.37 wt.%) in Ag–HAp thin films on alumina substrates. Thin Solid Films 1998;335:214.
coating FGHA-Ag1 rendered it similar to coating FGHA, whereas [5] Park SS, Lee HJ, Oh IH, Lee BT. Effects of Ag-doping on microstructure and
mechanical properties of hydroxyapatite films. Key Eng Mat 2005;277–
for coatings having higher Ag concentrations, of 6.29 ± 0.24 and 279:113.
13.53 ± 0.22 wt.% (less than 5 vol.%) for FGHA-Ag2 and FGHA- [6] Alt V, Bechert T, Steinrucke P, Wagener M, Seidel P, Dingeldein E, et al. An
Ag3, the Young’s modulus increased sharply by 30% compared in vitro assessment of the antibacterial properties and cytotoxicity of
nanoparticulate silver bone cement. Biomaterials 2004;25:4383.
to coating FGHA-Ag1, indicating that the Ag acts as a reinforcement [7] Rameshbabu N, Kumar TSS, Prabhakar TG, Sastry VS, Murty KVGK, Rao KP.
in the HA matrix [42]. Antibacterial nanosized silver substituted hydroxyapatite: synthesis and
characterization. J Biomed Mater Res A 2007;80A:581.
[8] Feng QL, Wu J, Chen GQ, Cui FZ, Kim TN, Kim JO. A mechanistic study of the
5. Conclusions antibacterial effect of silver ions on Escherichia coli and Staphylococcus aureus. J
Biomed Mater Res 2000;52:662.
[9] Chen W, Liu Y, Courtney HS, Bettenga M, Agrawal CM, Bumgardner JD, et al. In
A series of functionally graded hydroxyapatite coatings incor- vitro anti-bacterial and biological properties of magnetron co-sputtered silver-
porated with different concentrations of Ag were deposited using containing hydroxyapatite coating. Biomaterials 2006;27:5512.
[10] Chen W, Oh S, Ong AP, Oh N, Liu Y, Courtney HS, et al. Antibacterial and
IBAD. TEM/STEM observations on FIB-prepared coating cross-sec- osteogenic properties hydroxyapatite coatings produced using of silver-
tions show three distinct layers within the coatings: a ‘‘mixed” containing a sol gel process. J Biomed Mater Res A 2007;82A:899.
crystalline interface layer adjacent to the Ti substrate, a crystal- [11] Alt V, Bechert T, Steinrucke P, Wagener M, Seidel P, Dingeldein E, et al. In vitro
testing of antimicrobial activity of bone cement. Antimicrob Agents Ch
line bottom layer and a mostly amorphous layer on the top sur-
2004;48:4084.
face. TEM/STEM-EDS confirmed the presence of 10–50 nm silver [12] Ramesh N, Kumar KR, Kumar TSS, Sunder M, Victor SP. Biphasic calcium
particles distributed throughout the coating thickness, with lar- phosphates for antibiotic release. Trends Biomater Artif Organs 2005;18:6.
ger particles observed in the crystalline layer(s) and smaller Ag [13] Feng QL, Cui FZ, Kim TN, Kim JW. Ag-substituted hydroxyapatite coatings with
both antimicrobial effects and biocompatibility. J Mater Sci Lett 1999;18:559.
particles in the amorphous layer. XRD results showed that the [14] Badrour L, Sadel A, Zahir M, Kimakh L, El Hajbi A. Synthesis and physical and
coatings are primarily HA phase, with small amounts of CaO. chemical characterization of Ca10 xAgx(PO4)(6)(OH)(2 x)square(x)
SEM–EDS analysis on the coating top surfaces indicated that apatites. Ann Chim-Sci Mat 1998;23:61.
[15] Aimanova OJ, LeGeros RZ, Sinyayev VA. Antimicrobiologic property hydrated
the average Ag wt.% in coatings FGHA-Ag1, FGHA-Ag2 and amorphous calcium phosphates containing silver. Key Eng Mat 2005;17:439.
FGHA-Ag3 were 2.34, 6.29 and 13.53, respectively, whereas [16] Karlov AV, Shakhov VP, Kolobov JR. Definition of silver concentration in
XPS analysis of the top surfaces of the coatings indicated an calcium phosphate coatings on titanium implants ensuring balancing of
bactericidity and cytotoxicity. Key Eng Mat 2000;192-1:207.
average Ag wt.% of 1.09, 3.16 and 6.59, respectively. Adhesion [17] Oh KS, Kim KJ, Jeong YK, Park EK, Kim SY, Kwon JH, et al. Cytotoxicity and
strength, evaluated by pull-off tests, showed that the average antimicrobial effect of Ag doped hydroxyapatite. Key Eng Mat 2004;264–
bonding strengths were in the range of 83–88 MPa, which is 268:2107.
[18] Chung RJ, Hsieh MF, Huang KC, Perng LH, Chou FI, Chin TS. Anti-microbial
comparable to 85 MPa for an FGHA coating without incorporated hydroxyapatite particles synthesized by a sol–gel route. J Sol-Gel Sci Techn
silver. The optical observation of failure areas indicated that the 2005;33:229.
dominant failure mode was epoxy failure and no coating delam- [19] Lee BT, Shin NY, Han JK, Song HY. Microstructures and fracture characteristics
of spark plasma-sintered HAp–5 vol.% Ag composites. Mat Sci Eng A-Struct
inations were observed. The biological responses of these coat-
2006;429:348.
ings are being prepared for future publication. [20] Montali A. Antibacterial coating systems. Injury 2006;37:81.
[21] Kim TN, Feng QL, Kim JO, Wu J, Wang H, Chen GC, et al. Antimicrobial effects of
metal ions (Ag+, Cu2+, Zn2+) in hydroxyapatite. J Mater Sci-Mater M
Acknowledgements 1998;9:129.
[22] Choi JW, Cho HM, Kwak EK, Kwon TG, Ryoo HM, Jeong YK, et al. Effect of Ag-
doped hydroxyapatite as a bone filler for inflamed bone defects. Key Eng Mat
The authors would like to acknowledge the financial support 2004;254-2:47.
from the National Science Foundation (Grant No. 0600596) and ac- [23] Paul W, Sharma CP. Antibiotic loaded hydroxyapatite osteoconductive implant
cess to the research facilities at the Center for Nanophase Materials material – in vitro release studies. J Mater Sci Lett 1995;14:1792.
[24] Lee IS, Whang CN, Oh KS, Park JC, Lee KY, Lee GH, et al. Formation of silver
Sciences (CNMS) and the Shared Research Equipment (SHaRE) User incorporated calcium phosphate film for medical applications. Nucl Instrum
Facilities at Oak Ridge National Laboratory, both of which are sup- Meth B 2006;242:45.
ported by the Scientific User Facilities Division, Office of Basic En- [25] Noda I, Miyaji F, Ando Y, Miyamoto H, Shimazaki T, Yonekura Y, et al.
Development of novel thermal sprayed antibacterial coating and evaluation of
ergy Sciences, the U.S. Department of Energy (Award No.
release properties of silver ions. J Biomed Mater Res B: Appl Biomater
CNMS2005-070). The authors would also like to acknowledge Mr. 2008;89B.
Stefan Sandukas from NCSU for his assistance with SEM–EDS, [26] Sanpo N, Tan ML, Cheang P, Khor KA. Antibacterial property of cold-sprayed
and Drs. Harry M. Meyer, Sang Hoon Shim, Adam J. Rondinone HA–Ag/PEEK coating. J Therm Spray Techn 2009;18:10.
[27] Oliver WC, Pharr GM. Measurement of hardness and elastic modulus by
and Andrew Payzant from ORNL for their training and assistance instrumented indentation: advances in understanding and refinements to
with XPS, nanoindentation and X-ray diffraction analysis. methodology. J Mater Res 2004;19:3.
X. Bai et al. / Acta Biomaterialia 6 (2010) 2264–2273 2273

[28] Cofino B, Fogarassy P, Millet P, Lodini A. Thermal residual stresses near the [35] Blalock TL, Bai X, Narayan R, Rabiei A. Effect of substrate temperature on
interface between plasma-sprayed hydroxyapatite coating and titanium mechanical properties of calcium phosphate coatings. J Biomed Mater Res B
substrate: finite element analysis and synchrotron radiation measurements. 2008;85B:60.
J Biomed Mater Res A 2004;70A:20. [36] Rabiei A, Thomas B, Neville B, Lee JW, Cuomo J. A novel technique for
[29] Blalock T, Bai X, Rabiei A. A study on microstructure and properties of calcium processing functionally graded HA coatings. Mat Sci Eng C-Bio S 2007;27:523.
phosphate coatings processed using ion beam assisted deposition on heated [37] Rabiei A, Thomas B, Jin C, Narayan R, Cuomo J, Yang Y, et al. A study on
substrates. Surf Coat Tech 2007;201:5850. functionally graded HA coatings processed using ion beam assisted deposition
[30] Ding SJ, Ju CP, Lin JHC. Characterization of hydroxyapatite and titanium with in situ heat treatment. Surf Coat Tech 2006;200:6111.
coatings sputtered on Ti–6Al–4V substrate. J Biomed Mater Res 1999;44: [38] Rabiei A, Blalock B, Thomas B, Cuomo J, Yang Y, Ong J. Microstructure,
266. mechanical properties, and biological response to functionally graded HA
[31] Ohring M, NetLibrary Inc.. Materials science of thin films: deposition and coatings. Mat Sci Eng C-Bio S 2007;27:529.
structure. San Diego, CA: Academic Press; 2002. [39] Xiao Bai, Stefan Sandukas, Appleford MR, Ong Joo L, Rabiei A. Deposition and
[32] Hou XH, Choy KL, Leach SE. Processing and in vitro behavior of hydroxyapatite investigation of functionally graded calcium phosphate coatings on titanium.
coatings prepared by electrostatic spray assisted vapor deposition method. J Acta Biomater 2009;5:3563.
Biomed Mater Res A 2007;83A:683. [40] Suvorova EI, Buffat PA. Size effect in X-ray and electron diffraction patterns
[33] Rehman I, Bonfield W. Characterization of hydroxyapatite and carbonated from hydroxyapatite particles. Crystallogr Rep 2001;46:722.
apatite by photo acoustic FTIR spectroscopy. J Mater Sci-Mater M [41] Lee WJ, Lee SW, Kim HL, Kim DJ, Han JS. Characteristics of calcium phosphate
1997;8:1. films prepared by pulsed laser deposition under various water vapor
[34] Wang CX, Chen ZQ, Wang M, Liu ZY, Wang PL. Ion-beam-sputtering/mixing pressures. J Korean Phys Soc 2005;47:152.
deposition of calcium phosphate coatings. I. Effects of ion-mixing beams. J [42] Zhang X, Gubbels GHM, Terpstra RA, Metselaar R. Toughening of calcium
Biomed Mater Res 2001;55:587. hydroxyapatite with silver particles. J Mater Sci 1997;32:235.

You might also like