Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

Journal of Tropical Pediatrics, 2020, 00, 1–10

doi: 10.1093/tropej/fmaa111
Original Paper

Point-of-Care Testing Using Invasive and Non-


Invasive Hemoglobinometers: Reliable and
Valid Method for Estimation of Hemoglobin

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


among Children 6–59 Months
Gomathi Ramaswamy MD, MBBS, Kashish Vohra MSc, BSc, Kapil
Yadav MD, MBBS, Ravneet Kaur MD, MBBS, Tripti Rai MSc, BSc,
Abhishek Jaiswal, MD, MBBS, and Shashi Kant, MAE, MBA, DNB,
MD, MBBS
Centre for Community Medicine, All India Institute of Medical Sciences, New Delhi 110029, India
Correspondence: Kapil Yadav, Centre for Community Medicine, All India Institute of Medical Sciences, Room No 37,
New Delhi, India. Tel: 011-265-93848. E-mail: <dr.kapilyadav@gmail.com>.

ABSTRACT

Introduction: Globally around 47.4% of children and in India, 58% of children aged 6–59 months
are anemic. Diagnosis of anemia in children using accurate technologies and providing adequate
treatment is essential to reduce the burden of anemia. Point-of-care testing (POCT) devices is a po-
tential option for estimation of hemoglobin in peripheral and field settings were the hematology ana-
lyzer and laboratory services are not available.
Objectives: To access the validity of the POCTs (invasive and non-invasive devices) for estimation
of hemoglobin among children aged 6–59 months compared with hematology analyzer.
Methods: The study participants were enrolled from the pediatric outpatient department in
Haryana, India, from November 2019 to January 2020. Hemoglobin levels of the study participants
were estimated in Sahli’s hemoglobinometer and invasive digital hemoglobinometers (DHs) using
capillary blood samples. Hemoglobin levels in non-invasive DH were assessed from the finger/toe
of the children. Hemoglobin levels measured in POCTs were compared against the venous blood
hemoglobin estimated in the hematology analyzer.
Results: A total of 120 children were enrolled. The mean (SD) of hemoglobin (g/dl) estimated in
auto-analyzer was 9.4 (1.8), Sahli’s hemoglobinometer was 9.2 (1.9), invasive DH was 9.7 (1.9), and
non-invasive DH was 11.9 (1.5). Sahli’s hemoglobinometer (95.5%) and invasive DH (92.2%) had
high sensitivity for the diagnosis of anemia compared with non-invasive DH (24.4%). In contrast,
non-invasive DH had higher specificity (96.7%) compared with invasive DH (83.3%) and Sahli’s
hemoglobinometer (70%). Invasive DH took the least time (2–3 min) for estimation of hemoglobin
per participant, followed by Sahli’s (4–5 min) and non-invasive DH (5–7 min).
Conclusion: All three POCT devices used in this study are reasonable and feasible for estimating
hemoglobin in under-5 children. Invasive DHs are potential POCT devices for diagnosis of anemia

C The Author(s) [2020]. Published by Oxford University Press. All rights reserved.
V
For permissions, please email: journals.permissions@oup.com  1
2  Validation of Hemoglobinometers

among under-5 children, while Sahli’s can be considered as a possible option, where trained and
skilled technicians are available. Further research and development are required in non-invasive DH
to improve accuracy.

LAY SUMMARY
In India, anemia is a serious public health problem, where 58% of the children aged 6–59 months

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


are anemic. Point-of-care testing (POCT) using digital hemoglobinometers (DHs) has been recom-
mended as one of the key interventions by the Anemia Mukt Bharat program since 2018 in India.
These POCT devices are easy to use, less invasive, can be carried to field, require minimal training
and results are available immediately. Therefore this study assessed the validity of POCT devices—
invasive DH, non-invasive DH and Sahli’s hemoglobinometer among 6–59 months children in facil-
ity setting compared with the gold standard hematology analyzer. A total of 120 children under 6–
59 months of age were enrolled from the pediatric outpatient department in Haryana, India, from
November 2019 to January 2020. The (mean hemoglobin in g/dl) invasive (9.7) and non-invasive
DH (11.9) overestimated hemoglobin value, while Sahli’s (9.2) underestimated hemoglobin com-
pared with hematology analyzer (9.4). Invasive DH (92.2%) and Sahli’s hemoglobinometer (95.5%)
reported high ability to correctly identify those with anemia compared with non-invasive DH
(24.4%). In contrast, non-invasive DH (96.73%) had higher ability to correctly identify those with-
out the anemia compared with invasive DH (83.3%) and Sahli’s (70%).

INTRODUCTION hospital are few of the challenges associated with this


Children are one of the most vulnerable groups for method. Venous puncture itself is a stressful proced-
anemia. Globally, 47.4% of children under 5 years of ure among the children, parents, and the healthcare
age are anemic [1]. In India, where anemia is a ser- worker. A study conducted in Brazil reported that a
ious public health problem, around 6.8 crores (58%) failure rate of 10% was observed in the peripheral
of children aged 6–59 months are anemic [2]. venous sampling, and it was higher among malnour-
Childhood anemia impairs cognitive, motor, and lan- ished children [7]. Hence, alternate less invasive or
guage development, and academic achievement. It non-invasive technique is imperative for hemoglobin
also increases the risk of infection, overall morbidity, estimation among under-5 children.
and mortality [3–5]. In the long run, anemia under- POCT for hemoglobin is recently becoming a
mines work productivity in adulthood with substan- popular strategy to overcome the challenges associ-
tial economic consequences to the individual as well ated with laboratory-based investigations. POCT
as the country [4, 5]. Given the detrimental long- devices are easy to use, less invasive, can be carried
term effects, prevention in early childhood is crucial. to the field, requires minimal training, and results are
Prophylactic iron–folic acid supplementation, available immediately. The commonly available
deworming, food-based approaches like dietary di- POCT methods for detection of anemia are Sahli’s
versity, food fortification, and behavior change com- method, World Health Organization’s Hemoglobin
munication interventions are implemented by the Color Scale, invasive digital hemoglobinometers
governments to address anemia. (DHs) using reagent-containing microcuvettes
Hemoglobin is the most commonly used param- (HemoCue 201, TrueHb), newer invasive DH using
eter for the diagnosis of anemia. The standard reagent-free microcuvettes (HemoCue 301,
method for hemoglobin estimation is using a hema- DiaSpect), non-invasive DH such as multiwavelength
tology analyzer in clinical laboratories [6]. However, co-oximetry [Radical-7(TM) and Pronto-7(TM)];
the prerequisite of venipuncture, clinical laboratory, Masimo Corp. All invasive POCT requires less than
trained laboratory technicians, the requirement of at one drop of blood and a capillary blood sample can
least 1-ml blood and more than one visit to the be used to estimate hemoglobin.
Validation of Hemoglobinometers  3

Sahli’s hemoglobinometer is a widely used POCT and respiratory diseases, children with any illness for
for more than two decades globally [8]. It is relative- more than 2 weeks, and severely ill children assessed
ly inexpensive, simple to use with no higher recur- by the investigator.
ring costs for testing materials and electricity, and Ethics approval was obtained from the AIIMS
requires only a small amount of blood taken by a Institute Ethics Committee. Written informed con-
skilled technician [9, 10]. The invasive DH is being sent was obtained from Mother/Guardian.
used in health camps, blood donation camps, nation-
al surveys, and field-level testing. The invasive DH Description of devices

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


has reported acceptable diagnostic accuracy, and Hematology analyzer (Sysmex XS 1000i, Kobe,
Sahli’s reported a diverse range of diagnostic accur-
Japan)
acy [11–13]. However, these studies have majorly
This autoanalyzer was selected as the reference test.
conducted among adolescents and adult age group.
It uses Sodium Lauryl Sulfate method, which is a
Under-5 children are a challenging group for obtain-
non-cyanide method using an automated blood cell
ing the capillary sample, and studies have reported a
counter machine for hemoglobin estimation, and its
significant change in the diagnostic accuracy of
principle is similar to the gold standard, i.e. cyanme-
hemoglobin due to variation in the execution of ca-
themoglobin method.
pillary blood sampling procedure [14, 15].
The newer non-invasive DH can be a potential
Sahli’s hemoglobinometer
option for estimating hemoglobin since these is no
The principle in Sahli’s method involves the conver-
pain associated with this procedure. It may be more
sion of hemoglobin to acid hematin and comparing
acceptable among the children and parents and con-
visually the color developed with that of standard
venient for the healthcare workers. However, there
tinted glass. This method requires 20 ml of blood,
are limited studies published on the validity of both
and results can be read by diffused daylight, 3 min
invasive and non-invasive DH for the estimation of
after adding the blood sample to hydrochloric acid.
hemoglobin among the under-5 age group. Thus,
It does not run on electricity nor requires cuvettes or
this study was aimed to assess the validity of POCT
strips for every patient; thus, it is relatively
devices Sahli’s hemoglobinometer, invasive DH, and
inexpensive.
non-invasive DH among 6–59 months children in fa-
cility setting compared with the gold standard hema- HemoCue Hb 301 System (HemoCue AB,
tology analyzer.
Ängelholm, Sweden)
It is a pre-calibrated, portable, battery-operated, re-
METHODOLOGY
agent-free microcuvette device that estimates hemo-
Study design and setting globin by measuring the absorbance of whole blood
The present validation study was conducted from at the Hb/HbO2 isosbestic points at the wavelength
November 2019 to January 2020 at the Sub District of 506 and 880 nm for compensation of turbidity. It
Hospital (SDH), Comprehensive Rural Health requires 10 ml amount of blood, and provides results
Services Project, Ballabgarh in Haryana, India. within 3 s. It can operate in temperature ranging
from 10 to 40 C and measures hemoglobin from 0
Participants and ethics approval to 25.6 g/dl.
A total of 120 children from 6 to 59 months of age
attending pediatric outpatient department clinics at Rad 67 Spot check pulse co oximeter (Rad-67TM
R R R
SDH Ballabgarh, Haryana, were enrolled in the Pulse CO-OximeterV and rainbowV DCIV-mini
study. Children with minor ailments such as cough Sensor, Masimo, USA)
or fever for <1-week, acute otitis media, mild impe- It measures SpO2, total hemoglobin- SpHb, and
tigo, viral warts, tenia versicolor etc., were included. SpCO with the easy-to-use finger gauge for patients
We excluded children with a history of renal, cardiac,  3 kg. The sensor collects the patient’s vital signs
4  Validation of Hemoglobinometers

and sends them to the device, which, in turn, displays thumb or great toe was used to select the appropriate
the calculated SpHb value in g/dl. It displays results finger or toe for measuring hemoglobin. The ring
within 30 s as claimed by the manufacturing com- sensor model was used on the selected finger or toe,
pany and weighs around 0.37 kg. The total hemoglo- and the reading was obtained according to the manu-
bin (SpHb) accuracy range of the device is 8–17 g/ facturer instructions (as shown in Fig. 2). The probe
dl. The device has three parameters viz. perfusion was covered by a black sheet (provided by the manu-
index, signal quality, and interference to indicate the facturer) to avoid light interference. Various cartoon
quality of the test while measuring hemoglobin. characters were showed to calm the child while test-

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


Accurate results will be obtained when all three ing using non-invasive DH.
parameters fall in the green zone indicating stable Under strict aseptic precautions, the second drop
condition (Fig. 1). of capillary blood was used for Sahli’s, and the third
drop was used for invasive DH for estimation of
Sample size and sampling method hemoglobin. The sequence of the blood sample
All children who attended the pediatric clinic at the taken was alternated between Sahli’s and invasive
SDH, Haryana, during the study period, and pre- DH to overcome any bias that may arise due to the
scribed for any blood investigation by pediatricians difference in the second and third drop of blood.
were included. Considering bias (SD of difference) Similarly, 2 ml of venous blood drawn from antecu-
as 0.1 (1.1) g/dl alpha error of 5%, power 80%, and bital fossa was used for hemoglobin estimation in a
maximum allowed difference 2.8 g/dl in hemoglobin, hematology analyzer. The blood samples for all tests
the minimum required sample size was 105 [16, 17]. were collected in the same setting on the same day,
The sample size of 120 children was finalized, con- and there was no treatment administered in
sidering a 10% non-response rate. between.

Data collection Quality assurance


A standard protocol was followed for performing all Hemoglobin values displayed from Sahli’s hemoglo-
the tests. As non-invasive DH involves prick free binometer, invasive DH, and non-invasive DH were
method, it was used first for testing of hemoglobin not disclosed among technicians to avoid inter-
to keep the child calm for further invasive proce- observer bias, especially with Sahli’s. Internal quality
dures. Non-invasive DH displays spot-check results control (IQC) was run daily for invasive DH and
correctly when its signal quality indicators for mo-
tion, perfusion index quality, and ambient light inter-
ference are stable. Each child was sampled at sitting
position, with the non-invasive DH positioned on a
flat surface at the level of the child’s hand. The finger
gauge, which measures the size of the finger or

Fig. 1. Non-invasive DH displaying hemoglobin value, Fig. 2. Estimation of hemoglobin using non-invasive
signal quality indicators and SpO2 values. digital hemoglobinometer.
Validation of Hemoglobinometers  5

hematology analyzer. All three levels of IQC samples anemia ¼ 8.3% (n ¼ 10)]. Prevalence of anemia
(normal, low, and high) were checked and plotted in using Sahli’s was 79.1% (n ¼ 95), invasive DH was
the Levy Jennings (LJ) plot. All values were found to 73.3% (n ¼ 88) and non-invasive DH was 19.2%
be within the two standard deviation range in the LJ (n ¼ 23). (Table 1 and Fig. 3) There was no signifi-
plot. The end-user for non-invasive DH requires no cant difference in the prevalence of anemia between
calibration. The non-invasive DH sensor uses a valid- invasive DH (p ¼ 0.768) and Sahli’s (p ¼ 0.442)
ation process different from the conventional quality compared with hematology analyzer. There was a
control materials of the automated hematology ana- significant difference in prevalence of anemia in esti-

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


lyzer and the control cuvette of the HemoCue [18]. mated in non-invasive DH (p < 0.001) compared
with hematology analyzer.
Data management and analysis
The data were entered in an android based mobile
application tool Epicollect5 and exported to Stata 14 Bland Altman analysis
software (Stata Corp. 2015, College Station, TX: The mean (SD) of hemoglobin (g/dl) values esti-
Stata Corp LP) for data analysis. The mean differ- mated in hematology analyzer was 9.4 (1.8) vs. mean
ence (SD of difference) and limits of agreement (SD) observed in Sahli’s was 9.2 (1.9), p < 0.001, in-
(LOA) were calculated based on the Bland Atman vasive DH was 9.7 (1.9), p < 0.001 and non-invasive
plot. Lin’s concordance coefficient and intra class DH was 11.9 (1.5), p < 0.001. When compared with
correlation coefficient were calculated with 95% CI. the hematology analyzer, bias and LOA for Sahli’s
Weighted Kappa was calculated to assess the agree- was 0.27 g/dl (2.0 to 1.4), invasive DH was
ment between the equipment on grades of anemia. 0.27 g/dl (1.4 to 1.9), and non-invasive DH was
Diagnostic accuracy for each test was examined 2.49 g/dl (0.3 to 5.2; Fig. 4 and Table 2).
based on sensitivity, specificity, positive and negative
predictive values and ROC curve to detect anemia. A
p-value of <0.05 was considered for ascertaining Lin’s concordance and agreement between DHs
statistical significance. and hematology analyzer
Lin’s concordance coefficient for non-invasive DH
RESULTS was 0.3 (p < 0.001), invasive DH was 0.9
In total, 120 children aged 6–59 months were (p < 0.001) and Sahli’s was 0.9 (p < 0.001) com-
included in this study. Of the total, 77 (64.2%) were pared with hematology analyzer for estimation of
male and 43 (35.8%) were female children. The hemoglobin. The weighted Cohen’s Kappa coeffi-
mean age of the children was 33.7 6 15.7 months, cient for different grades (no/mild/moderate/se-
ranging from 6 to 59 months. The overall prevalence vere) of anemia for Sahli’s was 0.70, invasive DH
of anemia was 75.0% (n ¼ 90) estimated using hema- was 0.73, and non-invasive DH was 0.12, indicating a
tology analyzer [mild anemia ¼ 18.3% (n ¼ 22); good degree of agreement for invasive DH and
moderate anemia ¼ 48.3%, (n ¼ 58); and severe Sahli’s compared with hematology analyzer.

Table 1. Prevalence of anemia and mean (SD) of hemoglobin determined by hematology analyzer,
Device Sahli’s, invasive DH and non-invasive DH among 6–59 months children in India (n ¼ 120)
Hemoglobin concentration, g/dl Hematology analyzer Device Sahli’s Invasive DH Device non-invasive DH

Mean 6 SD 9.4 6 1.8 9.2 6 1.9* 9.7 6 1.9* 11.9 6 1.5*


Range 4.0–13.6 3.5–12.6 4.4–13.5 5.1–15.2
Prevalence of anemia, n (%) 90 (75) 95 (79.1)** 88 (73.3)** 23 (19.2)*

*p < 0.001;
**p > 0.05.
6  Validation of Hemoglobinometers

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


Fig. 3. Prevalence of anemia by severity (%) determined from the non-invasive DH, invasive DH, Device Sahli’s and
hematology analyzer among 6–59 months children in India (n ¼ 120).

Fig. 4. Bland–Altman plot showing agreement in hemoglobin concentration assessed by hematology analyzer, Sahli’s
hemoglobinometer, invasive DH and non-invasive DH.

Table 2. Bias (SD) of difference, LOA and concordance determined by hematology analyzer,
Device Sahli’s, invasive DH and non-invasive DH among 6 to 59 months children in India
(N ¼ 120)
Sahli’s hemoglobinometer Invasive DH Non-invasive
(n ¼ 120) (n ¼ 120) DH (n ¼ 120)

Bias (SD) of difference in hemoglobin 0.27 6 1.9 0.27 6 1.9 2.49 6 0.1
95% LOA in hemoglobin 2.0, 1.4 1.4, 1.9 0.3, 5.2
Sensitivity (%) for detecting anemia 95.5 92.2 24.4
Specificity (%) for detecting anemia 70 83.3 96.7
Positive predicted value (%) for detecting anemia 79.2 73.3 19.2
Negative predicted value (%) for detecting anemia 20.8 26.7 80.8
Lin’s concordance correlation coefficient (qc) 0.9 0.9 0.3
Cohen’s Kappa coefficient 0.70 0.73 0.12
Validation of Hemoglobinometers  7

not moving during the procedure) whereas other


indicators such as perfusion index quality and ambi-
ent light interference did not affect the sensitivity of
the device. Specificity of the device was above 95%
only when all three indicators or even any of them
were in green zone. When the indicator for motion
alone, or motion with perfusion index quality, or all
three indicators were in green zone, device showed

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


100% specificity.
Invasive DH required 2–3 min to complete whole
procure for estimation of hemoglobin, whereas
Sahli’s took 4–5 min and non-invasive DH took 5–
7 min per participant for estimation of hemoglobin.
Fig. 5. ROC of hemoglobin values measured using
hemoglobinometers compared with hematology analyzers.
DISCUSSION
The area under the curve for Sahli’s was 0.90, in- This study assessed the diagnostic accuracy of three
vasive DH was 0.94 and non-invasive DH was 0.78 POCT hemoglobinometers compared with the
(p < 0.001) and depicted in Fig. 5. hematology analyzer to estimate hemoglobin among
children from 6 to 59 months of age. The mean dif-
Sensitivity and specificity for detecting anemia ference (LOA) in the hemoglobin values of the
Device Sahli’s had 95.5% sensitivity, 70% specificity, hematology analyzer compared with Sahli’s was
79.2% positive predictive value, and 20.8% negative 0.3 g/dl (2.0 to 1.4), invasive DH was 0.3 g/dl
predictive value. For invasive DH, the sensitivity, (1.4 to 1.9), and non-invasive DH was 2.5 g/dl
specificity, positive predictive value and negative pre- (0.3 to 5.2). Both invasive DH and Sahli showed
dictive value was 92.2%, 83.3%, 73.3% and 26.7% re- similar excellent concordance, whereas non-invasive
spectively and for device non-invasive DH its DH showed poor concordance with hematology ana-
sensitivity, specificity, positive predictive value and lyzer. The present study reported high sensitivity of
negative predictive value was 24.4%, 96.7%, 19.2% Sahli’s (95.5%) and invasive DH (92.2%) compared
and 80.8%, respectively for detection of anemia as with the gold standard hematology analyzer for
compared with the hematology analyzer. detecting anemia, exhibiting good validity between
Sahli’s device had a sensitivity of 100% and speci- the two methods. In contrast, non-invasive DH
ficity of 90.8% to detect severe anemia (hemoglobin (24.4%) had reported poor sensitivity with but with
< 7 g/dl) compared with auto-analyzer. Invasive DH higher specificity (96.73%) compared with other
had sensitivity of 90% and specificity of 90.8% and POCTs.
non-invasive DH had sensitivity of 10% and specifi- A study carried out among school-going children
city of 100% to detect severe anemia. (6–8 years) stated higher sensitivity and specificity
Non-invasive DH’s performance or it’s sensitivity for similar invasive DH (HemoCue 201) compared
and specificity is dependent on signal quality indica- with the hematology analyzer for detecting anemia
tors such as motion, perfusion index quality, and am- [12]. Studies reported that sensitivity and specificity
bient light interference. If all these indicators are of invasive DH varies with the type of blood sample
stable, then quick results with better diagnostic ac- (venous or capillary) used for estimation of hemo-
curacy can be achieved. Supplementary Table S1 globin [18–24]. Another study conducted in Mexico
depicts how all these indicators combined or indi- among 1–5 years children; and other studies con-
vidually affects the sensitivity and specificity of the ducted in California and India among adult patients
device. Sensitivity of the device improves when indi- and healthy volunteers reported that non-invasive
cator for motion was in green zone (when child is device under reports hemoglobin values. In contrast,
8  Validation of Hemoglobinometers

the current study reported overestimated hemoglo- system for the diagnosis and management of anemia.
bin values [25–27]. Few invasive devices also have a data storage facility.
Several factors affect the accuracy of hemoglobin- However, only hemoglobin values and in a few, date
ometers. Geographical variations (altitude and tem- and time are stored.
perature), procedural differences in capillary blood The approximate device cost for Sahli’s was 13
sample collection (Intra- and inter-observer varia- USD, invasive DH was 286 USD and non-invasive
tions), study participants related factors (age groups, DH was 1770 USD. The approximate operating cost
sex and biological determinants including Fahraeus per test is 0.20 USD for Sahli’s hemoglobinometer,

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


effect [11, 12, 18] lower hematocrit in capillary 0.41 USD for invasive DH and 0.68 USD for invasive
blood vessels compared with venous vessels) might DH (for 1000 test package). The operating cost of
be the reasons for the observed deviations in the non-invasive DH is based on the number of tests
hemoglobin measurement. Non-invasive DH dis- included in the package. Higher the number of tests,
plays spot-check results correctly only when it’s sig- lower the operating cost. A study conducted by
nal quality indicators are stable. Figure 1 shows the Ahankari et al. [9], reported similar cost for the
error when indicators are not stable. Due to this POCT devices for hemoglobin estimation. Though
characteristic, amongst children with continuous the operating cost of non-invasive DH is relatively
body movement, it was difficult to obtain stable indi- high, the intangible costs associated with invasive
cators (all three indicators in green zone). DH such as handling of biomedical waste, pain and
All three devices had certain advantages and dis- discomfort associated with the procedure for the
advantages to be used in a facility setting. Invasive patients are avoided in non-invasive DH.
DH was quick in displaying results and was easy to Certain limitations were associated with the
handle. Sahli’s requires a trained technician. study. The study was conducted in the winter season
However, it has an advantage of low cost without no and temperature can affect the findings of hemoglo-
expensive recurrent logistics is involved, and it can binometers, primarily non-invasive DH. Second, we
be considered as an alternative POCT when a skilled have conducted this study among the children who
technician is available. Non-invasive DH was a pain- were attending healthcare centers for management
less procedure and favorable for children. However, of minor health ailments. Since these results are
the issues with signal quality indicators among chil- derived from children attending health care centers
dren have to be dealt to obtain accurate results. thus these results should be cautiously used with
Invasive devices show overall better performance general population having low prevalence of anemia.
than the non-invasive device in healthcare settings. We have purposefully selected this group to avoid
However, non-invasive hemoglobin measurement any unwarranted invasive procedures among the
could be significantly advantages as it avoids pricking under-5 children. However, the results have to The
and can be executed as a pain-free procedure which cost of the devices was based on quotations from the
is favorable especially among children. Non-invasive manufacturers. We have not conducted any cost-
DH also has an added advantage as there is no risk analysis.
of exposure to bloodborne infections or discomforts One strength of the study is adherence to the
associated with invasive procedure. It is also environ- quality control process for both the hematology ana-
ment friendly as it does not produce biomedical lyzer and DHs throughout the study. To our know-
waste for every patient test. ledge, this is the first study from India to assess the
Non-invasive DH also has inbuilt storage capacity, validity of non-invasive DHs among under-5 chil-
gives an option for error rectification, and also check- dren. Last, a comparison of various hemoglobinome-
ing previous visit hemoglobin at small healthcare ters aids the policymakers and programmers to
centers with minimal patient load ensuring follow- identify the best possible devices for the estimation
up. This feature of non-invasive devices provides us of hemoglobin. In India, POCT using DHs has been
an opportunity to incorporate artificial intelligence recommended as one of the key interventions by the
techniques and health management information Anemia Mukt Bharat program since 2018.
Validation of Hemoglobinometers  9

CONCLUSION 6. H15-A3 Reference and Selected Procedures for the


Accurate diagnosis of using point of testing devices Quantitative Determination of Hemoglobin in Blood;
is essential for ensuring the quality of the anemia Approved Standard-Third Edition. 2000. https://commu
nity.clsi.org/media/1401/h15a3_sample.pdf (24 May
control program. The validity of both invasive, non- 2020, date last accessed).
invasive, non-digital hemoglobinometers was 7. de Negri DC, Avelar AFM, Andreoni S, et al. Predisposing
explored in this study. All three POCT devices were factors for failure of peripheral intravenous puncture in
found to be potential to use for estimation of hemo- children. Rev Lat Am Enfermagem 2012;20:1072–80.
globin in under-5 children. Invasive DHs have high 8. Kulkarni MV, Durge PM, Kasturwar NB. Prevalence of an-

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


emia among adolescent girls in an urban slum. Natl J
diagnostic accuracy compared with other devices.
Community Med 2012;3:108–11.
Sahli’s hemoglobinometer can be considered as a po- 9. Ahankari AS, Dixit JV, Fogarty AW, et al. Comparison of
tential alternative when trained technicians are avail- the NBM 200 non-invasive haemoglobin sensor with
able. Non-invasive DHs may require further research Sahli’s haemometer among adolescent girls in rural India.
in improving the diagnostic accuracy for the estima- BMJ Innov 2016;2:144–8.
tion of hemoglobin. 10. Patil P, Thakare G, Patil S, et al. Variability and accuracy of
Sahli’s method in estimation of haemoglobin concentra-
tion. Natl J Integr Res Med 2013;4:38–44.
SUPPLEMENTARY DATA 11. Neufeld LM, Larson LM, Kurpad A, et al. Hemoglobin
Supplementary data are available at Journal of concentration and anemia diagnosis in venous and capil-
Tropical Pediatrics online. lary blood: biological basis and policy implications. Ann
NY Acad Sci 2019;1450:172–89.
ACKNOWLEDGEMENTS 12. Gwetu TP, Chhagan MK. Evaluation of the diagnostic ac-
curacy of the HemoCue device for detecting anaemia in
We would like to thank Dr Neelesh Kapoor and Masimo US
healthy school-aged children in KwaZulu-natal. S Afr Med
Team for providing the non-invasive digital hemoglobinome-
J 2015;105:596–9.
ter ‘Rad 67 Spot check pulse co oximeter’. We also thank Dr
13. Srinivasan N, Kasturba MP. Intra-operative point of care
Pawan Gupta and HemoCue Team for providing ‘HemoCue haemoglobin estimation: a comparison of three methods.
Hb 301’ device. Sri Lankan J Anaesthesiol 2010;18:15–9.
14. Bhaskaram P, Balakrishna N, Radhakrishna KV, et al.
FUNDING Validation of hemoglobin estimation using Hemocue.
The work was not supported by Masimo or HemoCue. Indian J Pediatr 2003;70:25–8.
15. Whitehead RD, Jr, Mei Z, Mapango C, et al. Methods and
REFERENCES analyzers for hemoglobin measurement in clinical labora-
tories and field settings. Ann NY Acad Sci 2019;1450:
1. Mclean E, Cogswell M, Egli I, et al. Worldwide preva-
147–71.
lence of anaemia, WHO Vitamin and Mineral Nutrition
16. Hinnouho G-M, Barffour MA, Wessells R, et al.
Information System. Public Health Nutr 2009;12:
Comparison of haemoglobin assessments by HemoCue
444–54. and two automated haematology analysers in young
2. International Institute for Population Sciences. National Laotian children. J Clin Pathol 2018;71:532–8.
Family Health Survey (NFHS-4) 2015-16 India. Deonar, 17. Sample Size Calculation: Bland-Altman Plot. https://
Mumbai: International Institute for Population Sciences, www.medcalc.org/manual/sampling_blandaltman.php
2017, 1–192. (30 May 2020, date last accessed).
3. Srivastava T, Negandhi H, Neogi SB, et al. Methods for 18. Kim MJ, Park Q, Kim MH, et al. Comparison of the accur-
hemoglobin estimation: a review of “what works”. J acy of noninvasive hemoglobin sensor (NBM-200) and
Hematol Transfus 2014;2:1028. portable hemoglobinometer (HemoCue) with an auto-
4. WHAT’S AT STAKE TARGET: 50% reduction of an- mated hematology analyzer (LH500) in blood donor
aemia in women of reproductive age BOX 1: WHAT IS screening. Ann Lab Med 2013;33:261–7.
ANAEMIA? https://www.who.int/nutrition/topics/glob 19. Akhtar K, Sherwani RK, Rahman K, et al. HemoCue pho-
altargets_anaemia_policybrief.pdf (24 May 2020, date last tometer: a better alternative of hemoglobin estimation in
accessed). blood donors. Indian J Hematol Blood Transfus 2008;24:
5. Subramanian SV, Balarajan Y, Ramakrishnan U, et al. 170–2.
Anaemia in low-income and middle-income countries.
Lancet 2011;378:2123–35.
10  Validation of Hemoglobinometers

20. Sari M, de Pee S, Martini E, et al. Estimating the preva- haemoglobin levels in whole blood donors. Blood
lence of anaemia: a comparison of three methods. Bull Transfus 2013;11:385–90.
World Health Organ 2001;79:506. 25. Levy TS, Méndez-Gómez-Humarán I, Ruán MDCM, et al.
21. van Den Broek NR, Ntonya C, Mhango E, et al. Validation of masimo pronto 7 and HemoCue 201 for
Diagnosing anaemia in pregnancy in rural clinics: assessing hemoglobin determination in children from 1 to 5 years of
the potential of the Haemoglobin Colour Scale. Bull age. PLoS One 2017;12:e0170990.
World Health Organ 1999;77:15–21. 26. Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive
22. Munoz M, Romero A, gomez JF, et al. Utility of point-of- hemoglobin and invasive point-of-care hemoglobin testing
care haemoglobin measurement in the HemoCue-B compared with a laboratory analyzer. Int J Lab Hematol

Downloaded from https://academic.oup.com/tropej/article/67/1/fmaa111/6047276 by guest on 27 October 2022


haemoglobin for the initial diagnosis of anaemia. Clin Lab 2014;36:56–61.
Haematol 2005;27:99–104. 27. Neogi SB, Negandhi H, Sharma J, et al. Diagnostic efficacy
23. Mendrone A, Jr, Sabino EC, Sampaio L, et al. Anemia of digital hemoglobinometer (TrueHb), HemoCue and
screening in potential female blood donors: comparison of non invasive devices for screening patients for anemia in
two different quantitative methods. Transfusion 2009;49: the field settings- a proposal. Indian J Community Heal
662–8. 2018;30:86–8.
24. Shahshahani HJ, At Mansouri MN. F. Evaluation of the
validity of a rapid method for measuring high and low

You might also like