Professional Documents
Culture Documents
HB Sahli
HB Sahli
doi: 10.1093/tropej/fmaa111
Original Paper
ABSTRACT
Introduction: Globally around 47.4% of children and in India, 58% of children aged 6–59 months
are anemic. Diagnosis of anemia in children using accurate technologies and providing adequate
treatment is essential to reduce the burden of anemia. Point-of-care testing (POCT) devices is a po-
tential option for estimation of hemoglobin in peripheral and field settings were the hematology ana-
lyzer and laboratory services are not available.
Objectives: To access the validity of the POCTs (invasive and non-invasive devices) for estimation
of hemoglobin among children aged 6–59 months compared with hematology analyzer.
Methods: The study participants were enrolled from the pediatric outpatient department in
Haryana, India, from November 2019 to January 2020. Hemoglobin levels of the study participants
were estimated in Sahli’s hemoglobinometer and invasive digital hemoglobinometers (DHs) using
capillary blood samples. Hemoglobin levels in non-invasive DH were assessed from the finger/toe
of the children. Hemoglobin levels measured in POCTs were compared against the venous blood
hemoglobin estimated in the hematology analyzer.
Results: A total of 120 children were enrolled. The mean (SD) of hemoglobin (g/dl) estimated in
auto-analyzer was 9.4 (1.8), Sahli’s hemoglobinometer was 9.2 (1.9), invasive DH was 9.7 (1.9), and
non-invasive DH was 11.9 (1.5). Sahli’s hemoglobinometer (95.5%) and invasive DH (92.2%) had
high sensitivity for the diagnosis of anemia compared with non-invasive DH (24.4%). In contrast,
non-invasive DH had higher specificity (96.7%) compared with invasive DH (83.3%) and Sahli’s
hemoglobinometer (70%). Invasive DH took the least time (2–3 min) for estimation of hemoglobin
per participant, followed by Sahli’s (4–5 min) and non-invasive DH (5–7 min).
Conclusion: All three POCT devices used in this study are reasonable and feasible for estimating
hemoglobin in under-5 children. Invasive DHs are potential POCT devices for diagnosis of anemia
C The Author(s) [2020]. Published by Oxford University Press. All rights reserved.
V
For permissions, please email: journals.permissions@oup.com 1
2 Validation of Hemoglobinometers
among under-5 children, while Sahli’s can be considered as a possible option, where trained and
skilled technicians are available. Further research and development are required in non-invasive DH
to improve accuracy.
LAY SUMMARY
In India, anemia is a serious public health problem, where 58% of the children aged 6–59 months
Sahli’s hemoglobinometer is a widely used POCT and respiratory diseases, children with any illness for
for more than two decades globally [8]. It is relative- more than 2 weeks, and severely ill children assessed
ly inexpensive, simple to use with no higher recur- by the investigator.
ring costs for testing materials and electricity, and Ethics approval was obtained from the AIIMS
requires only a small amount of blood taken by a Institute Ethics Committee. Written informed con-
skilled technician [9, 10]. The invasive DH is being sent was obtained from Mother/Guardian.
used in health camps, blood donation camps, nation-
al surveys, and field-level testing. The invasive DH Description of devices
and sends them to the device, which, in turn, displays thumb or great toe was used to select the appropriate
the calculated SpHb value in g/dl. It displays results finger or toe for measuring hemoglobin. The ring
within 30 s as claimed by the manufacturing com- sensor model was used on the selected finger or toe,
pany and weighs around 0.37 kg. The total hemoglo- and the reading was obtained according to the manu-
bin (SpHb) accuracy range of the device is 8–17 g/ facturer instructions (as shown in Fig. 2). The probe
dl. The device has three parameters viz. perfusion was covered by a black sheet (provided by the manu-
index, signal quality, and interference to indicate the facturer) to avoid light interference. Various cartoon
quality of the test while measuring hemoglobin. characters were showed to calm the child while test-
Fig. 1. Non-invasive DH displaying hemoglobin value, Fig. 2. Estimation of hemoglobin using non-invasive
signal quality indicators and SpO2 values. digital hemoglobinometer.
Validation of Hemoglobinometers 5
hematology analyzer. All three levels of IQC samples anemia ¼ 8.3% (n ¼ 10)]. Prevalence of anemia
(normal, low, and high) were checked and plotted in using Sahli’s was 79.1% (n ¼ 95), invasive DH was
the Levy Jennings (LJ) plot. All values were found to 73.3% (n ¼ 88) and non-invasive DH was 19.2%
be within the two standard deviation range in the LJ (n ¼ 23). (Table 1 and Fig. 3) There was no signifi-
plot. The end-user for non-invasive DH requires no cant difference in the prevalence of anemia between
calibration. The non-invasive DH sensor uses a valid- invasive DH (p ¼ 0.768) and Sahli’s (p ¼ 0.442)
ation process different from the conventional quality compared with hematology analyzer. There was a
control materials of the automated hematology ana- significant difference in prevalence of anemia in esti-
Table 1. Prevalence of anemia and mean (SD) of hemoglobin determined by hematology analyzer,
Device Sahli’s, invasive DH and non-invasive DH among 6–59 months children in India (n ¼ 120)
Hemoglobin concentration, g/dl Hematology analyzer Device Sahli’s Invasive DH Device non-invasive DH
*p < 0.001;
**p > 0.05.
6 Validation of Hemoglobinometers
Fig. 4. Bland–Altman plot showing agreement in hemoglobin concentration assessed by hematology analyzer, Sahli’s
hemoglobinometer, invasive DH and non-invasive DH.
Table 2. Bias (SD) of difference, LOA and concordance determined by hematology analyzer,
Device Sahli’s, invasive DH and non-invasive DH among 6 to 59 months children in India
(N ¼ 120)
Sahli’s hemoglobinometer Invasive DH Non-invasive
(n ¼ 120) (n ¼ 120) DH (n ¼ 120)
Bias (SD) of difference in hemoglobin 0.27 6 1.9 0.27 6 1.9 2.49 6 0.1
95% LOA in hemoglobin 2.0, 1.4 1.4, 1.9 0.3, 5.2
Sensitivity (%) for detecting anemia 95.5 92.2 24.4
Specificity (%) for detecting anemia 70 83.3 96.7
Positive predicted value (%) for detecting anemia 79.2 73.3 19.2
Negative predicted value (%) for detecting anemia 20.8 26.7 80.8
Lin’s concordance correlation coefficient (qc) 0.9 0.9 0.3
Cohen’s Kappa coefficient 0.70 0.73 0.12
Validation of Hemoglobinometers 7
the current study reported overestimated hemoglo- system for the diagnosis and management of anemia.
bin values [25–27]. Few invasive devices also have a data storage facility.
Several factors affect the accuracy of hemoglobin- However, only hemoglobin values and in a few, date
ometers. Geographical variations (altitude and tem- and time are stored.
perature), procedural differences in capillary blood The approximate device cost for Sahli’s was 13
sample collection (Intra- and inter-observer varia- USD, invasive DH was 286 USD and non-invasive
tions), study participants related factors (age groups, DH was 1770 USD. The approximate operating cost
sex and biological determinants including Fahraeus per test is 0.20 USD for Sahli’s hemoglobinometer,
20. Sari M, de Pee S, Martini E, et al. Estimating the preva- haemoglobin levels in whole blood donors. Blood
lence of anaemia: a comparison of three methods. Bull Transfus 2013;11:385–90.
World Health Organ 2001;79:506. 25. Levy TS, Méndez-Gómez-Humarán I, Ruán MDCM, et al.
21. van Den Broek NR, Ntonya C, Mhango E, et al. Validation of masimo pronto 7 and HemoCue 201 for
Diagnosing anaemia in pregnancy in rural clinics: assessing hemoglobin determination in children from 1 to 5 years of
the potential of the Haemoglobin Colour Scale. Bull age. PLoS One 2017;12:e0170990.
World Health Organ 1999;77:15–21. 26. Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive
22. Munoz M, Romero A, gomez JF, et al. Utility of point-of- hemoglobin and invasive point-of-care hemoglobin testing
care haemoglobin measurement in the HemoCue-B compared with a laboratory analyzer. Int J Lab Hematol