Professional Documents
Culture Documents
Biomedical Short Learning Article Complete
Biomedical Short Learning Article Complete
Biomedical Short Learning Article Complete
Author
Akhil. S
June 2021
AKHIL 1
FORCE Biomedical
Table of Contents
Introduction to Biomedical Engineering ............................................................................. 4
What Is Biomedical Engineering? ......................................................................................... 4
What Biomedical Engineers do? ....................................................................................... 5
Introduction to IEC Standards – 60601 ................................................................................ 6
Introduction to Biomedical Equipment’s ............................................................................ 9
Aspirator ............................................................................................................................... 9
Clinical Thermometer ....................................................................................................... 10
Sphygmomanometer......................................................................................................... 11
Weighing scale ................................................................................................................... 12
Glucometer ......................................................................................................................... 13
Flowmeter ........................................................................................................................... 14
Pulse oximeter.................................................................................................................... 16
Nebulizer ............................................................................................................................ 17
Breast pump ....................................................................................................................... 20
Electrical wheelchair ......................................................................................................... 20
Oxygen concentrator ......................................................................................................... 21
ABG Machine ..................................................................................................................... 23
ACT Machine ..................................................................................................................... 24
Biothesiometer ................................................................................................................... 25
Bronchoscope ..................................................................................................................... 26
Biosafety cabinet ................................................................................................................ 28
C Arm .................................................................................................................................. 30
Capnography ..................................................................................................................... 32
CT Machine ........................................................................................................................ 34
Anaesthesia machine ........................................................................................................ 38
Defibrillator ........................................................................................................................ 41
Dental chair ........................................................................................................................ 43
AKHIL 2
FORCE Biomedical
Mammography .................................................................................................................. 45
ECG machine ..................................................................................................................... 47
Electro surgical unit .......................................................................................................... 49
ECT machine ...................................................................................................................... 52
Vein detector ...................................................................................................................... 54
Fetal Monitor ...................................................................................................................... 54
Haemodialysis Unit .......................................................................................................... 57
IABP..................................................................................................................................... 64
Baby Incubators ................................................................................................................. 66
Infusion pump ................................................................................................................... 68
PCA pump .......................................................................................................................... 71
Enteral Feeding pump ...................................................................................................... 72
Multipara Monitor ............................................................................................................ 73
NIBP .................................................................................................................................... 75
Surgical light ...................................................................................................................... 77
OT table ............................................................................................................................... 80
Radiant warmer ................................................................................................................. 86
Phototherapy unit ............................................................................................................. 88
Stethoscope ......................................................................................................................... 90
Syringe pump..................................................................................................................... 91
Neuronavigation system .................................................................................................. 92
Ultrasound machine.......................................................................................................... 93
Ventilator ............................................................................................................................ 96
AKHIL 3
FORCE Biomedical
There are many subdisciplines within biomedical engineering, including the design
and development of active and passive medical devices, orthopaedic implants,
medical imaging, biomedical signal processing, tissue and stem cell engineering, and
clinical engineering
AKHIL 4
FORCE Biomedical
Biomedical engineers work in a wide variety of settings and disciplines. There are
opportunities in industry for innovating, designing, and developing new
technologies; in academia furthering research and pushing the frontiers of what is
medically possible as well as testing, implementing, and developing new diagnostic
tools and medical equipment; and in government for establishing safety standards
for medical devices. Many biomedical engineers find employment in cutting-edge
start-up companies or as entrepreneurs themselves. Tissue and stem cell engineers
are working towards artificial recreation of human organs, aiding in transplants and
helping millions around the world live better lives. Experts in medical devices
develop new implantable and external devices such as pacemakers, coronary stents,
orthopedic implants, prosthetics, dental products, and ambulatory devices. Clinical
engineers work to ensure that medical equipment is safe and reliable for use in
clinical settings. Biomedical engineering is an extremely broad field with many
opportunities for specialization.
*****
AKHIL 5
FORCE Biomedical
It can be assumed that not all people will understand the dangers associated with
the exposure to electricity. It is this danger that has triggered several discussions
relating to the safety of all members of the public.
Applied Part
Part of the medical equipment which is designed to come into physical contact with
the patient or parts that are likely to be brought into contact with the patient.
Patient Connection
Individual physical connections and / or metal parts intended for connection with
the patient which form (part of) an Applied Part.
AKHIL 6
FORCE Biomedical
Patient Environment
Volumetric area in which a patient can come into contact with medical equipment or
contact can occur between other persons touching medical equipment and the
patient, both intentional and unintentional.
Applied Part which is electrically isolated from Earth and other parts of the medical
equipment i.e., floating F-type Applied Parts are either type BF or type CF Applied
Parts.
Applied Part complying with specified requirements for protection against electric
shock. Type B Applied Parts are those parts, which are usually Earth referenced.
Type B are those parts not suitable for direct cardiac application.
F-Type Applied Part complying with a higher degree of protection against electric
shock than type B Applied Parts. Type BF Applied Parts are those parts not suitable
for direct cardiac application.
F-Type Applied Part complying with the highest degree of protection against electric
shock. Type CF Applied Parts are those parts suitable for direct cardiac application.
AKHIL 7
FORCE Biomedical
Class I
Class II
*****
AKHIL 8
FORCE Biomedical
Aspirator
Surgical aspirators consist of a line-powered vacuum pump, a vacuum regulator and
gauge, a collection canister, and an optional bacterial filter. Plastic tubing connects
these components, completing an open-ended system that continuously draws
tissue debris and fluid from the surgical fi eld to the collection canister. The gauge
allows the user to set a safe limit for suctioning, to assess the performance of the
vacuum pump, and to detect leaks or blockages. Units are either portable or
Principles of operation
AKHIL 9
FORCE Biomedical
Clinical Thermometer
Clinical thermometers are meant for clinical purposes. It is developed for measuring
the human body temperature. It is a long narrow glass tube with a bulb containing
mercury at the end. The normal human body temperature is 37˚C; which can
fluctuate between the ranges 35˚C to 42˚C. Hence, the clinical thermometers have the
range 35˚C to 42˚C. The level of mercury tells our body temperature in ˚C. Since
mercury is a toxic element, thus these thermometers have been replaced by digital
thermometers nowadays.
Digital Thermometer
These thermometers are used for measuring the temperature by the means of an
electronic circuit. The information captured is sent to a microchip that processes it
and gets displayed numerically on the digital screen. They are easy to use,
inexpensive, and accurate. Digital thermometers can be considered to be advanced
thermometers that are used for measuring body temperature.
Digital Thermometer
AKHIL 10
FORCE Biomedical
Sphygmomanometer
A sphygmomanometer is a device that measures blood pressure. It is composes of an
inflatable rubber cuff, which is wrapped around the arm. A measuring device
indicates the cuff's pressure. A bulb inflates the cuff and a valve releases pressure. A
stethoscope is used to listen to arterial blood flow sounds. As the heart beats, blood
forced through the arteries cause a rise in pressure, called systolic pressure, followed
by a decrease in pressure as the heart's ventricles prepare for another beat. This low
pressure is called the diastolic pressure.
Systolic and diastolic pressures are commonly stated as systolic 'over' diastolic. For
example, 120 over 80. Blood flow sounds are called Korotkoff sounds.
Types
• ANEROID SPHYGMOMANOMETERS
• DIGITAL SPYGMOMANOMETERS
ANEROID DIGITAL
AKHIL 11
FORCE Biomedical
Weighing scale
Weighting Scales are used to measure the weight of an item. To use a scale, the item
which needs to be weighed is put on one side of the scale. Then, usually stones are
put on the other side to compare the weight of the stone with the weight of the object
you have chosen.
Basic components
• Pressure sensor
• Microprocessor
• Analog to digital converter
• LCD
OPERATIONAL ISSUES
➢ Error reading
AKHIL 12
FORCE Biomedical
Glucometer
AKHIL 13
FORCE Biomedical
Flowmeter
The medical gas flowmeter is a medical device for oxygen or medical air inhalation
of first-aid and hypoxic patients in the hospital, including oxygen flowmeter, and
medical air flowmeter. It mainly consists of gas pressure gauge, pressure reducer,
safety valve, flow tube, flow control valve, and humidification bottle, and other
components.
AKHIL 14
FORCE Biomedical
Oxygen regulators are intended for the administration of oxygen to patients that are
deemed by a physician to need increased oxygen levels to improve or stabilize their
breathing conditions. It is a pressure-reducing device that lowers the pressure of the
oxygen from a cylinder to a level that can safely be used.
The Diameter Index Safety System, or DISS, was designed by the Compressed Gas
Association specifically for medical gases at 200 psig or less. It uses unique, gas-
specific threaded connections to fit equipment to station outlets.
AKHIL 15
FORCE Biomedical
Pulse oximeter
A pulse oximeter is a small, lightweight device used to monitor the amount of
oxygen carried in the body. This noninvasive tool attaches painlessly to your
fingertip, sending two wavelengths of light through the finger to measure
your pulse rate and how much oxygen is in your system.
• Transmittance
• Reflectance
AKHIL 16
FORCE Biomedical
Nebulizer
A nebulizer turns liquid medicine into a mist to help treat your asthma. They come
in electric or battery-run versions. They come in both a portable size you can carry
with you and a larger size that’s meant to sit on a table and plug into a wall. Both are
made up of a base that holds an air compressor, a small container for liquid
medicine, and a tube that connects the air compressor to the medicine container.
Above the medicine container is a mouthpiece or mask you use to inhale the mist.
TYPES
➢ Jet nebulizer
Vibrating mesh nebulizers use mesh deformation or vibration to push the liquid
drug through the mesh. An annular piezo element, which is in contact with the
mesh, is used to produce vibration around the mesh, and the liquid drug is in direct
contact with the mesh. Holes in the mesh have a conical structure, with the largest
cross-section of the cone in contact with the liquid drug [6]. The mesh deforms into
the liquid side, thus pumping and loading the holes with liquid. This deformation
on the other side of the liquid-drug reservoir ejects droplets through the holes,
which can be inhaled by the patient
AKHIL 17
FORCE Biomedical
Jet nebulizer
Jet nebulization was the first technical operation developed for aerosol production. It
uses gas flow either from a compressor or a central air supply. The gas passes
through a small aperture in the nebulizer in order to pick up and atomize the liquid
drug. The aerosol that is generated by atomization contains large and small droplets,
and is driven to a baffle. Large droplets are impacted by the baffle and forced onto
the side of the nebulizer to be recycled in liquid form in a reservoir. More than 90%
of the droplets produced by atomization are selected and recirculated in the
nebulizer to be recycled in the liquid-drug reservoir. Small droplets are transported
out of the nebulizer by the gas to be inhaled by the patient.
AKHIL 18
FORCE Biomedical
Ultrasonic nebulizers
AKHIL 19
FORCE Biomedical
Breast pump
Breast pumps are medical devices
regulated by the U.S. Food and Drug
Administration. They can be used to
maintain or increase a
woman's milk supply, relieve
engorged breasts and
plugged milk ducts, or pull out flat or
inverted nipples so a nursing baby can
latch on more easily.
Breast pump
Electrical wheelchair
AKHIL 20
FORCE Biomedical
Oxygen concentrator
Oxygen concentrator definition: An oxygen concentrator is a type of medical device
used for delivering oxygen to individuals with breathing-related disorders.
Individuals whose oxygen concentration in their blood is lower than normal often
require an oxygen concentrator to replace that oxygen.
BASIC PRINCIPLES
The most common oxygen concentrators molecular sieves that operate in a two-part
cycle
AKHIL 21
FORCE Biomedical
o In the second step, the function of the cylinders is reversed in a timed cycle,
providing a continuous flow of oxygen to the patient
o The oxygen concentration produced by molecular sieve concentrators varies
inversely with the flow of gas through the cylinders: the lower the flow, the
higher the oxygen concentration in the end- product gas
AKHIL 22
FORCE Biomedical
ABG Machine
Blood gas analyzers are used to measure combinations of pH, blood gas (i.e. pCO2
and pO2), electrolytes, and metabolites parameters from whole blood samples.
Blood conservation is an important initiative in every clinical diagnostic lab or
critical care facility. Therefore, a highly accurate blood gas analyzer can streamline
work processes and reduce errors with minimum maintenance. Most blood gas
analyzers offer automatic sample mixing and easy sample aspiration with intuitive
user interface built-in and automatic QC for accuracy and regulatory compliance.
Also available in a lightweight, small footprint design, most models offer simple
replacement of solutions for analysis such as, sensor cassettes and solution packs. A
particularly unique feature to certain blood gas analyzers is liquid calibration,
thereby eliminating previously required gas tanks, regulators, tubing lines, and
humidifiers.
Principle
AKHIL 23
FORCE Biomedical
● Urea: Urea is broken into ammonia and carbon dioxide through urease.
Ammonia and carbon dioxide react through hydrolysis with physiological pH
to form ammonia or bicarbonate ions. The ammonia ions can be determined
using a potentiometrical ammonia ion-selective electrode. This measurement
requires a reference electrode such as those used in ion-selective electrodes
● tHb/SO2: Light absorption in whole blood is measured at four different
wavelengths, the sample is subjected to light radiation and the dispersed light
is also evaluated.
● COOX: The hemoglobin derivatives and the total bilirubin (= neonatal) are
determined spectrophotometrically based on the Lambert-Beer law.
ACT Machine
The activated clotting time (ACT) is commonly used to monitor treatment with high-
dose heparin before, during, and for a short time after medical or surgical
procedures that require that blood be prevented from clotting, such as heart bypass
surgery, coronary angioplasty, and dialysis. 70 to 120 seconds is the usual amount
of time for blood to clot without heparin. 180 to 240 seconds is the usual amount
of time for blood to clot with heparin. This is called the therapeutic range.
AKHIL 24
FORCE Biomedical
Biothesiometer
The biothesiometer is an instrument which measures the threshold of appreciation of
vibration sense. The amplitude of the stimulus (measured in volts) is gradually
increased until the threshold of vibratory sensation is reached and the stimulus is
appreciated by the patient.
AKHIL 25
FORCE Biomedical
Bronchoscope
Devices that are introduced at the nose or mouth to observe distal branches of the
bronchi. Through working channels in the bronchoscope, the physician can sample
lung tissue (e.g., when pulmonary malignancies are suspected), instill radiographic
media for bronchographic studies, perform laser therapy, remove foreign objects,
suction sputum for microbiological culturing, insert catheters, and perform difficult
intubations
These devices consist of a proximal housing, a flexible insertion tube ranging from
0.5 to 7.0 mm in diameter, and an “umbilical cord” connecting the light source and
the proximal housing. The proximal housing, which is designed to be held in one
hand, typically includes the eyepiece (fiberoptic models only), controls for distal tip
(bending section) angulation and suction, and the working channel port.
Principles
The bronchoscope (either flexible or rigid) is inserted into the airways, usually
through the mouth or nose. Sometimes the bronchoscope is inserted via a
tracheostomy. Rigid bronchoscopes are used for the removal of foreign bodies while
flexible video bronchoscopes are intended to provide images of a patient’s airways
and lungs. Images provided by the bronchoscope can be focused by adjusting the
ocular on the scope’s proximal housing. A video bronchoscope uses a charge-
coupled device (CCD) located at the distal tip of the scope to sense and transmit
images, replacing the image guide and eyepiece. These images can then be recorded,
printed, stored on digital media, or transmitted to another location for simultaneous
viewing
AKHIL 26
FORCE Biomedical
Flexible bronchoscope
Rigid bronchoscope
AKHIL 27
FORCE Biomedical
Biosafety cabinet
A biological safety cabinet (BSC) is a primary engineering control used to protect
personnel against biohazardous or infectious agents and to help maintain quality
control of the material being worked with as it filters both the inflow and exhaust
air. It is sometimes referred to as a laminar flow or tissue culture hood.
Classification
o Class I: BSCs should be limited to basic microbiological work with low- and
moderate risk agents
o Class II: BSCs also provide product protection through high-efficiency
particulate air filtration (HEPA filtration) of the intake air.
o Class III: cabinets offer the maximum protection to laboratory personnel, the
community, and the environment because all hazardous materials are
contained in a completely enclosed, ventilated cabinet.
AKHIL 28
FORCE Biomedical
CLASS I CLASS II
CLASS III
AKHIL 29
FORCE Biomedical
C Arm
A mobile C-arm is a medical imaging device that is based on X-ray technology
and can be used flexibly in various ORs within a clinic. The name is derived from the
C-shaped arm used to connect the X-ray source and X-ray detector to one another.
Since the introduction of the first C-arm in 1955 the technology has advanced
rapidly. Today, mobile imaging systems are an essential part of everyday hospital
life: Specialists in fields such as surgery, orthopedics, traumatology, vascular surgery
and cardiology use C-arms for intraoperative imaging. The devices provide high-
resolution X-ray images in real time, thus allowing the physician to monitor progress
at any point during the operation and immediately make any corrections that may
be required. Consequently, the treatment results are better and patients recover
more quickly. Hospitals benefit from cost savings through fewer follow-up
operations and from minimized installation efforts
AKHIL 30
FORCE Biomedical
C ARM PARTS
AKHIL 31
FORCE Biomedical
Capnography
Capnography, the measurement of exhaled carbon dioxide (CO2), has been gaining
popularity in hospital critical care environments and more recently in the
prehospital setting as well. Capnography was first used in the OR to continuously
and noninvasively estimate arterial carbon dioxide (PaCO2) levels on a breath-to-
breath basis.1 Measuring carbon dioxide in the patient’s exhaled breath allowed
anaesthesiologists to determine the adequacy and effectiveness of ventilation. The
amount of carbon dioxide exhaled at the end of each breath (EtCO2) is measured
through a sensor located between the patient’s airway and ventilator and is then
numerically and graphically displayed as a waveform.
Measurement methods
AKHIL 32
FORCE Biomedical
non-intubated patients. Heavy and bulky adapter and sensor assemblies may make
this method uncomfortable for non-intubated patients.
In side stream capnographs the exhaled CO2 is aspirated via ETT, cannula, or mask
through a 5–10-foot-long sampling tube connected to the instrument for analysis;
this method is intended for the nonincubated patient. Both mainstream and side
stream technologies calculate the CO2 value and waveform.
Sidestream Mainstream
AKHIL 33
FORCE Biomedical
CT Machine
These devices are also called as Axial CT scanners, CAT scanners, cine CT scanners,
EBT scanners, helical CT scanners, multislice CT scanners, spiral CT scanners,
ultrafast CT scanners. CT scanners produce thin cross-sectional images of the human
body for a wide variety of diagnostic procedures.
AKHIL 34
FORCE Biomedical
Basic components
o X-ray subsystem
o Gantry
o Patient table
o Controlling computer
o High-voltage x-ray generator
o X-ray tube
o Detector system
o Collimators
o Rotational frame
o Solid-state detectors
o Silicon photodiodes
BASIC PRINCIPLES
AKHIL 35
FORCE Biomedical
o Slip-ring scanners can perform helical CT scanning, in which the x-ray tube
and detector rotate around the patient’s body, continuously acquiring data
while the patient moves through the gantry
o The acquired volume of data can be reconstructed at any point during the
scan
o All modern CT scanners are multislice
o In addition to the gantry, a multislice CT scanner has a powerful computer
processor
o Inside the gantry, an x-ray tube projects a fan-shaped x-ray beam through the
patient to the detector array.
o As the x-ray tube and detector rotate, x-rays are detected continuously
through the patient.
o The computer mathematically reconstructs data from each full rotation to
produce an image of one slice.
o The second component is a detector design that incorporates approximately
1,000 elements per row along the length of the arc (x/y axes) and up to 320
rows across the width (z-axis) of the detector.
o When using a multislice CT scanner, the slice width is chosen by combining
data from adjacent rows across the detector in the z-axis.
IMAGE MANIPULATION
o The quantitative nature of the CT image allows the reviewer to easily perform
a large number of image manipulations
o Although the numerical range of pixels in the image is rather large, the
numerical range spanned by most soft tissues is relatively narrow
o To adequately display the values for soft tissue and still maintain the ability
to discriminate density differences, CT scanners are designed to display user-
selected CT numerical ranges (also called Hounsfield units) over the entire
greyscale
o The range to be displayed (window width) and the central value (level) are
also user selectable
o Regions of interest in the image can be selected to obtain average CT values
within the region or to calculate total lesion volume
AKHIL 36
FORCE Biomedical
CT IMAGE RECONSTRUCTION
RADIATION DOSE
o CT uses some of the highest doses of any diagnostic imaging method, and the
fact that multislice CT has the potential to increase these doses adds to the
need for some form of automatic dose control
o The simplest dose-control system uses just one scout view, although some
systems can use two views.
o A more advanced dose-control method uses real-time information about the
patient’s anatomy derived from the beam signal received by the detectors as
the scan is progressing
******
AKHIL 37
FORCE Biomedical
Anaesthesia machine
An anaesthesia system comprises of a gas delivery platform, a data analysis and
distribution system, and physiologic and multigas monitors (optional in most units),
which indicate levels and variations of several physiologic variables and parameters
associated with cardiopulmonary function and/or gas and agent concentrations in
breathed-gas mixtures. Manufacturers typically offer a minimum combination of
monitors, alarms, and other features that customers must purchase to meet
standards and ensure patient safety.
BASIC PRINCIPLE
AKHIL 38
FORCE Biomedical
CIRCUIT DIAGRAM
o Gases used in anaesthesia are usually supplied under high pressure either in
cylinders or as a piped gas supply
o The cylinders are made from molybdenum steel, in which gases and vapours
are stored under pressure
o H - free-standing, attached to the anaesthesia machine by a flexible hose.
These are most economical, but reduce the mobility of the anaesthesia
machine
o The hose to the anaesthetic machine should be at a regulated pressure of 50
psi
o E - attached directly to the anaesthesia machine via a yoke
AKHIL 39
FORCE Biomedical
VENTILATION
AKHIL 40
FORCE Biomedical
Defibrillator
Defibrillators are devices that restore a normal heartbeat by sending an electric pulse
or shock to the heart. They are used to prevent or correct an arrhythmia, a heartbeat
that is uneven or that is too slow or too fast. Defibrillators can also restore the heart’s
beating is suddenly stops.
Other defibrillators can prevent sudden death among people who have a high risk of
a life-threatening arrhythmia. They include implantable cardioverter defibrillators
(ICDs), which are surgically placed inside your body, and wearable cardioverter
defibrillators (WCDs), which rest on the body. It can take time and effort to get used
to living with a defibrillator, and it is important to be aware of possible risks and
complications.
AKHIL 41
FORCE Biomedical
Types of Defibrillators
• Monophasic Defibrillator
• Biphasic Defibrillator
AKHIL 42
FORCE Biomedical
Dental chair
The Dental Chair is a specially designed medical device intended to support a
patient's whole body, like a recliner, but articulated, so that the dentist can recline
patients to virtually any position. Dental chairs can feature a staggering variety of
attachments, either on the chair itself, or on the ever-present Dental Engine: spit
bowls, suction tubes, pneumatic tubes to power various pieces of equipment used in
cleanings and surgeries, and so forth. Modern dental chairs are crafted from a
combination of metal and plastic, are often infused with antibacterial properties to
minimize the risk of infection, and notably have thin chair backs, which allow the
dentist to sit on a stool behind the patient to facilitate ease of access.
Ceiling Mounted Design – None of the attachments are on the chair proper, but
built into the ceiling, with the chair positioned beneath them.
Dental Chair Mounted Design – The Dental Engine and all necessary attachments
are built into the chair itself.
AKHIL 43
FORCE Biomedical
AKHIL 44
FORCE Biomedical
Mammography
Mammographic radiographic units use x-rays to produce images of the breast—a
and gross pathology. Mammography is used primarily to detect and diagnose breast
ray tube and gantry, and a recording medium. The x-ray generator modifies
incoming voltage to provide the x-ray tube with the power necessary to produce an
x-ray beam. They also include a “paddle” for compression and placement of the
phosphorescent screen with phosphor crystals that emit light when exposed to x-
replace the traditional film cassette and digital cassette reader, producing a digital
image from the cassette instead of developing film through a film processor.
Principle
Low energy X-rays are produced by the x-ray tube (an evacuated tube with an anode
voltage supply from the generator, collides with the tube’s target anode. The cathode
contains a wire filament that, when heated, provides the electron source. The target
anode is struck by the impinging electrons. X-rays exit the tube through a port
window of beryllium. Additional filters are placed in the path of the x-ray beam to
modify the x-ray spectrum. The x-rays that pass through the filter are shaped by
either a collimator or cone apertures and then directed through the breast.
AKHIL 45
FORCE Biomedical
AKHIL 46
FORCE Biomedical
ECG machine
Electrocardiographs detect the electrical signals associated with cardiac activity and
produce an ECG, a graphic record of the voltage versus time. They are used to
diagnose and assist in treating some types of heart disease and arrhythmias,
leads simultaneously and are frequently used in place of single-channel units. Some
ECG units consist of the ECG unit, electrodes, and cables. The 12-lead system
heart’s electrical activity; producing ECG waveforms in which the P waves, QRS
complex, and T waves vary in amplitude and polarity. Single-channel ECGs record
the electric signals from only one lead confi guration at a time, although they may
electrocardiographs only record the electric signals from the electrodes (leads) and
do not use any internal procedure for their interpretation. Interpretive multichannel
Principles
Electrocardiographs record small voltages of about one millivolt (mV) that appear
on the skin as a result of cardiac activity. The voltage differences between electrodes
are measured; these differences directly correspond to the heart’s electrical activity.
Each of the 12 standard leads presents a different perspective of the heart’s electrical
activity; producing ECG waveforms in which the P waves, QRS complex, and T
AKHIL 47
FORCE Biomedical
waves vary in amplitude and polarity. Other lead configurations include those of the
Frank system and Cabrera leads. The Frank configuration measures voltages from
electrodes applied to seven locations—the forehead or neck, the center spine, the
midsternum, the left and right midaxillary lines, a position halfway between the
AKHIL 48
FORCE Biomedical
Principles
denaturation of intracellular
resulting in coagulation or
desiccation effects
a cutting effect
AKHIL 49
FORCE Biomedical
o Electro-desiccation
o Fulguration
o Electrocoagulation
o Electro section
▪ In Fulguration, the electrode is held away from the skin surface and more
superficial surgery
▪ The blend mode combines both cutting and coagulation characteristics with
▪ These various modes allow the physician to select the most efficient method
AKHIL 50
FORCE Biomedical
▪ COAGULATION MODES
I. soft coagulation
II. Forced coagulation
TECHNIQUES
MONOPOLAR SURGERY
BIPOLAR TECHNIQUE
AKHIL 51
FORCE Biomedical
ECT machine
Electroconvulsive therapy (ECT) is a procedure, done under general anesthesia, in
which small electric currents are passed through the brain, intentionally triggering a
brief seizure. ECT seems to cause changes in brain chemistry that can quickly reverse
symptoms of certain mental health conditions.
BASIC COMPONENTS
▪ Control module
▪ Electrodes
▪ EEG Cable
TYPES OF ECT
The bitemporal position (right and left temporal areas) is used to administer
bilateral ECT
The bifrontal electrodes placed on right and left forehead over the frontal lobes
o Unilateral ECT
AKHIL 52
FORCE Biomedical
BASIC PRINCIPLES
I. response is poor
II. generalised seizures of sufficient duration are not achieved
III. reduction in the length of the seizure
AKHIL 53
FORCE Biomedical
Vein detector
The device works by using proprietary vein visualization technology that shines an
infrared light on the patients’ skin. The hemoglobin (oxygen-carrying protein)
within the patient’s blood absorbs the light, creating a red pattern that can be seen
on the surface of the skin. The oxygen-depleted veins appear darker in the pattern,
indicating the presence of a vein. These dark “roads” along the skin guide our
specialists to their target as they find a suitable vein to use for treatment.
Fetal Monitor
Electronic fetal monitoring (EFM) provides graphic and numeric information on fetal
heart rate (FHR) and maternal uterine activity (UA) to help clinicians assess fetal
well-being before and during labor. FHR often exhibits decelerations and
accelerations in response to uterine contractions or fetal movements; certain patterns
are indicative of hypoxia. Examination of these patterns, the baseline level, and
variability characteristics can indicate the need to alter the course of labor with drugs
or perform an operative delivery
Fetal monitors are bedside units that consist of a monitoring unit, cables, and
electrodes. They are designed to measure, record, and display FHR, uterine
contractions, and/or maternal blood pressure and heart rate before and during
childbirth. These monitors may sense FHR and uterine contraction indirectly
AKHIL 54
FORCE Biomedical
through the mother’s abdomen and/or directly by placing an electrode on the fetal
scalp (or other exposed skin surface) and measuring the change in pressure within
the uterus. Antepartum fetal monitors are typically used in physician’s offi ces and
clinics long before the beginning of labor. Most hospital-based monitors have
additional capabilities, including fetal and maternal ECG recording.
Principles
This method uses a device to listen to and record your baby’s heartbeat through
your belly (abdomen). One type of monitor is a Doppler ultrasound device. It’s often
used during prenatal visits to count the baby’s heart rate. It may also be used to
check the fetal heart rate during labor.
The healthcare provider may also check
your baby’s heart rate continuously
during labor and birth. To do this, the
ultrasound probe (transducer) is fastened
to your belly. It sends the sounds of your
baby’s heart to a computer. The rate and
pattern of your baby’s heart rate are
shown on a screen and printed on paper.
AKHIL 55
FORCE Biomedical
This method uses a thin wire (electrode) put on your baby’s scalp. The wire runs from the
baby through your cervix. It is connected to the monitor. This method gives better readings
because things like movement don’t affect it. But it can only be done if the fluid-filled sac
that surrounds the baby during pregnancy (amniotic sac) has broken and the cervix is
opened. Your provider may use internal monitoring when external monitoring is not giving
a good reading. Or your provider may use this method to watch your baby more closely
during labor.
During labor, your healthcare provider will watch your uterine contractions and your baby’s
heart rate. Your provider will note how often you are having contractions and how long
each lasts. Because the fetal heart rate and contractions are recorded at the same time, these
results can be looked at together and compared.
Your provider may check the pressure inside your uterus while doing internal fetal heart
monitoring. To do this, he or she will put a thin tube (catheter) through your cervix and into
your uterus. The catheter will send uterine pressure readings to a monitor
AKHIL 56
FORCE Biomedical
Haemodialysis Unit
These devices perform extracorporeal dialysis to replace the main activity of the
kidneys in patients with impaired renal function, such as those with end-stage renal
Principles
the dialysate delivery system, the extracorporeal blood-delivery circuit, and the
dialyzer. Blood is taken via the extracorporeal circuit, passed through a dialyzer for
solute and fl uid removal, and returned to the patient. Each system has its own
monitoring and control circuits. The delivery system prepares dialysate—a solution
circuit) circulates a portion of the patient’s blood through the dialyzer and returns it
BASIC COMPONENTS
AKHIL 57
FORCE Biomedical
AKHIL 58
FORCE Biomedical
BASIC PRINCIPLES
dialyzer
o Dialysate acts to remove metabolic wastes from the blood and also acts as a
buffering agent
o To prevent short- and long-term toxic effects, incoming water must be treated
o Treated water enters the dialysis machine and usually passes through a heater
and a deaerator before being mixed with the concentrate to form dialysate
o Two types of proportioning systems are used to mix the water and
concentrate
AKHIL 59
FORCE Biomedical
o The temperature of the dialysate is kept in the 34° to 42°C range to prevent
o The temperature and conductivity sensors can initiate alarms and divert the
portion of the patient’s blood through the dialyzer and returns it to the
patient.
o Usually, an artery and a vein in the patient’s arm are surgically joined for
decreased, the blood entering the fistula maintains high pressure, causing the
o One or two large-bore needles can then be inserted into the enlarged vessel
venous catheter that has separate connectors and lumens for the venous and
arterial lines
o A blood pump moves blood through the external tubing and dialyzer
o As the pump draws blood into the extracorporeal circuit, it creates a partial
vacuum that will draw air into the tubing if connections are not absolutely
tight
o As a safety feature, air detectors are employed to detect air in the blood line
AKHIL 60
FORCE Biomedical
o Pressures are monitored on both venous and arterial blood lines; high- and
low-pressure alarms turn off the blood pump if alarm limits are exceeded
o Because blood tends to clot when it comes into contact with foreign surfaces
o A drip chamber on the venous side of the blood circuit contains a clot-
trapping filter to help prevent upstream clots and other debris from reaching
the patient
DIALYZER
o There are three basic design configurations: coil, parallel plate, and hollow
fiber
o Red and white blood cells and proteins are too large to pass through the pores
in the membrane
AKHIL 61
FORCE Biomedical
o This type of therapy allows tighter control of volume transfer and more
o CRRT is especially useful in patients who cannot tolerate the rapid volume
patients, such as those in the intensive care unit; and in patients undergoing
cardiac surgery
modalities as an option.
AKHIL 62
FORCE Biomedical
AKHIL 63
FORCE Biomedical
IABP
An intra-aortic balloon pump (IABP) is a mechanical device that helps the heart
pump blood.
This device is inserted into the aorta, the body's largest artery. It is a long, thin tube
called a catheter with a balloon on the end of it. If you are hospitalized, your doctor
may insert an IABP. Your doctor will numb an area of your leg and thread the IABP
through the femoral artery in your leg into your aorta. He or she then positions the
IABP at the center of your aorta, below your heart. The doctor will use an X-ray
The IABP reduces the workload on your heart, allowing your heart to pump more
blood. The IABP is placed inside your aorta, the artery that takes blood from the
heart to the rest of the body. The balloon on the end of the catheter inflates and
deflates with the rhythm of your heart. This helps your heart pump blood to the
body
BASIC PRINCIPLES
o The coronary arteries, which originate at the base of the ascending aorta,
coronary perfusion
o As a result, 70% to 90% of coronary perfusion occurs during the heart’s filling
stage (diastole)
developed
AKHIL 64
FORCE Biomedical
in early systole. Balloon inflation causes ‘volume displacement’ of blood within the
aorta, both proximally and distally. This leads to a potential increase in coronary
the intrinsic ‘Windkessel effect’, whereby potential energy stored in the aortic root
during systole is converted to kinetic energy with the elastic recoil of the aortic root.
AKHIL 65
FORCE Biomedical
Baby Incubators
Types of Incubators
There are different types of incubators that can accommodate the changing needs of
the preemie. Among the five types commonly found in the NICU:
o Closed box incubators have a fresh air filtration system that minimizes the
risk of infection and prevents the loss of moisture from the air.
o Double-walled incubators have two walls that can further prevent heat and
air moisture loss.
o Open box incubators, also known as Armstrong incubators, provide radiant
heat below the baby but are otherwise open to the air, allowing for easy
access.
o Portable incubators, also known as transport incubators, are used to move the
new born from one part of the hospital to another.
o Servo-control incubators are automatically programmed to adjust
temperature and humidity levels based on skin sensors attached to the baby.
AKHIL 66
FORCE Biomedical
Incubator
Parts of incubator
AKHIL 67
FORCE Biomedical
Infusion pump
An external infusion pump is a medical device used to deliver fluids into a patient’s
body in a controlled manner. There are many different types of infusion pumps,
Infusion pumps may be capable of delivering fluids in large or small amounts, and
Some infusion pumps are designed mainly for stationary use at a patient’s bedside.
A number of commonly used infusion pumps are designed for specialized purposes.
These include:
BASIC COMPONENTS
AKHIL 68
FORCE Biomedical
BASIC PRINCIPLES
Peristaltic mechanism: Rippling, wavelike motion of finger like discs to occlude the
IV tubing successively . The finger like disc is forcing the fluid to flow through the
tubings
Pumping mechanism: using cassette fitted either syringe/piston like device and
tubing running from two sides. Motor-driven plunger moves inward for giving
patient
Alarms includes
• Air in line
• Upstream/downstream occlusion
• Set disengagement
• Flow error
Sensors includes
• Optical sensors
AKHIL 69
FORCE Biomedical
AKHIL 70
FORCE Biomedical
PCA pump
you medicine for pain when you press a button. In most cases, PCA pumps supply
The pump is attached to a thin, flexible tube (intravenous or IV line) that is placed in
your vein. This medicine is usually delivered only when you press the button
(bolus), but a continuous rate may be added by your doctor if needed (basal rate).
which are programmed for the pain-relieving drug that the doctor orders based on
your age, weight and type of surgery. The PCA pump is safe to use because you
receive medication by pressing the button when you feel pain, but the pump won’t
give you the drug if it’s not time to receive another dose yet. Remember, no one
should press the button on the PCA pump except you. When the pump is empty, an
PCA pump
AKHIL 71
FORCE Biomedical
An enteral feeding pump is an electronic medical device that controls the timing and
procedure in which the doctor inserts a tube into the patient’s digestive tract to
deliver liquid nutrients and medicines to the body. Tube feeding is administered to
patients who cannot eat normally due to oral cancer, surgery, injury, or another
condition that affects the normal ingestion and digestion process in the
gastrointestinal tract. The enteral feeding pump ensures that the right amount of
AKHIL 72
FORCE Biomedical
Multipara Monitor
It is mainly used in intensive care, hospitalization or ER. In general, basic models are
used to monitor cardiac activity (ECG), blood pressure (NIBP), respiration (RESP),
Monitoring Parameters
o ECG
o Respiration
o Temperature
AKHIL 73
FORCE Biomedical
Block diagram
AKHIL 74
FORCE Biomedical
NIBP
Blood pressure is the pressure exerted by circulating blood upon the walls of blood
vessels.
Principle
AKHIL 75
FORCE Biomedical
AKHIL 76
FORCE Biomedical
Surgical light
Surgical lights, also known as surgical lighting or operating lights, are mainly used
in hospital operating rooms and ambulatory surgery centers, but can also be used in
various locations throughout the facility to provide high quality lighting for
procedures. Examples include emergency rooms, labor and delivery, examination
rooms, and anywhere where procedures are completed. They are used by clinicians,
surgeons and proceduralists. A surgical light illuminates the operative site on a
patient for optimal visualization during a procedure.1 Surgical lights can provide
hours of bright light without excessively heating the patient or staff. A variety of
lights are available to meet the needs of providing optimal visualization during
surgery and procedures. An examination light is used during medical exams, while
operating room lights are used during surgical procedures.
• Lux (lx)
Unit for the amount of visible light measured by a luxmeter at a certain point.
Illuminance (lx) at 1m distance from the light emitting surface in the light field
centre.
Point in the light field (lighted area) where illuminance reaches maximum lux
intensity. It is the reference point for most measurements.
• Depth of illumination
The distance under the light emitting area where the illumination reaches 20% of the
central illuminance
• Shadow dilution
Diameter of light field around the light field centre, ending where the illuminance
reaches 10% of Ec. The average of four different cross sections through the light field
centre.
• D50
Diameter of light field around the light field centre, ending where the illuminance
reaches 50% of Ec. The average of four different cross sections through the light field
centre
Scientific principles
AKHIL 78
FORCE Biomedical
AKHIL 79
FORCE Biomedical
OT table
An operating table, sometimes called operating room table, is the table on which the
patient lies during a surgical operation. This surgical equipment is usually found
inside the surgery room of a hospital.
Requirements
o Height
o Tilting
o Adjustable segments
o Radiolucent surface
o Table top
o Operating table column
o The transporter
Types
o Fixed/Stationary
o Mobile unit
o Table column for a stationary operating table system is fixed to the floor.
o Medical devices like X-ray, C ARM can easily be brought to the operating area
and positioned.
AKHIL 80
FORCE Biomedical
o Accessibility
o Stability
o Imaging Support
o weight support
AKHIL 81
FORCE Biomedical
❖ Imaging Table
Imaging Table
AKHIL 82
FORCE Biomedical
❖ Neurology Table
o Back section,Lateral
slide, Tilt
o Longitudinal slide
o Trendelenburg and
elevations
o Electro Hydraulic
operation
Neurology Table
AKHIL 83
FORCE Biomedical
❖ Orthopaedics Table
o Femur treatment
o Arm treatment
o Shoulder treatment
o Hip treatment
o Bi-lateral hip treatment
o Spinal column treatment
Orthopaedics Table
AKHIL 84
FORCE Biomedical
OT Table positions
AKHIL 85
FORCE Biomedical
Radiant warmer
Radiant Warmer, is a body warming device to provide heat to the body. This device
helps to maintain the body temperature of the baby and limit the metabolism rate.
Heat has a tendency to flow in the heat gradient direction that is from high
temperature to low temperature.
Principles
Function
The heat output of radiant warmers is usually regulated by servo-control to keep the
skin temperature constant at a site on the abdomen where a thermistor probe is
attached. Compared with incubators, the partition of body heat loss is quite different
under radiant warmers. Radiant warmers increase convective and evaporative heat
loss and insensible water loss but eliminate radiant heat loss or change it to net gain.
The major advantage of the radiant warmer is the easy access it provides to
critically-ill infants without disturbing the thermal environment.
Its major disadvantage is the increase in insensible water loss produced by the
radiant warmer. ‘Insensible' water loss (‘insensible’ because we are not aware of it)
refers to water loss due to:
AKHIL 86
FORCE Biomedical
Modes
• Servo Mode
• Manual Mode
Servo Mode
Manual Mode
• Once connected to mains heater output regulated by knob on front panel. The
output is displayed as % or bars or bulbs.
• Use maximum (100% output) for rapid warming of bassinet in labor room 10
minutes before delivery. Reduce output to 25-75% after 10 minutes depending
on ambient temperature. If left on with heater output >80% alarm is activated
within 15 or 20 minutes later and there after the heater output goes to 40%; if
alarm is silenced the heater will kept on for another 15 to 20 minutes as per
manufacturers recommendation.
• For low birth weight or sick neonate adjust heater output depending on baby
temperature.
AKHIL 87
FORCE Biomedical
Phototherapy unit
Devices used to treat hyperbilirubinemia, characterized by high bilirubin
concentrations in the blood. Bilirubin, a product of hemoglobin breakdown, remains
in the body until the liver can convert it to a form that can be excreted. Jaundice, a
yellowish discoloration of the skin, eyes, and mucous membranes, results when
bilirubin levels in the blood are too high. High bilirubin levels can be caused by the
inability of an immature liver to process high levels of bilirubin, particularly in
neonates.
Phototherapy units consist of a light source and a means of allowing the light to
radiate the infant. Devices using overhead lamps can be freestanding on casters,
ceiling or wall mounted, or attached to infant radiant warmers or infant incubators;
some units have height and hood angle adjustments. Bassinet style units, in which
the infant is placed in a
plastic bassinet containing a
bank of lights in an
overhead case, are also
available. Fiberoptic
phototherapy pad systems
use a tungsten halogen bulb
in a metal case, a flexible
fiberoptic cable, and a light-
emitting plastic pad.
Filtered blue light is
delivered from the source
through the fiberoptic cable
and emitted from the sides
and ends of the fibers inside
the pad, which is wrapped
around the infant
AKHIL 88
FORCE Biomedical
Principles
• It can use fluorescent tubes, tungsten halogen bulbs, light emitting diodes or
quartz bulbs as light sources
• Fluorescent tubes should emit wide spectrum white light, wide spectrum blue
light, or narrow spectrum blue light
• It has shields to filter any ultraviolet radiation
• Tungsten-halogen bulbs are filtered for maximum light output within the
blue spectrum
• It has internal filters and dichroic reflectors to reduce harmful radiant energy
• It consists of a light source in a metal case, a flexible fiber optic cable, and a
light emitting plastic pad
• Filtered blue light is delivered from the source through the fiber optic cable
and emitted from the sides and ends of the fibers inside the pad
AKHIL 89
FORCE Biomedical
Stethoscope
A stethoscope is used to hear the heart beat sounds, sound due to inhalation and
exhalation of air in the lungs and the respiratory pathways and also the stomach
movement. It is a very useful diagnostic tool to help localize problems and to
diagnose disease. Stethoscopes are also used along with the sphygmomanometer.
The first usable binaural stethoscope was invented in 1855. The modern electronic
stethoscopes are high precisioned instruments. These can be used to hear a patient's
heart and lung clearly even in high noisy environments and even through layers of
clothing. The electronic stethoscopes also make it possible to hear the foetal sounds
in mother's womb.
Uses
AKHIL 90
FORCE Biomedical
Syringe pump
A syringe pump is a motor-driven precision pump that uses one or more syringes to
deliver precise and accurate amounts of fluid in high-impact research environments.
Types
There are two broad types of pumps: Laboratory syringe pump and medical infusion
pump.
o Medical Infusion Pump: They are devices used to deliver controlled quantities
of fluids such as nutrients, drugs, and blood to patients. This type of infusion
pump can be used for in vivo diagnosis, treatment, and research.
AKHIL 91
FORCE Biomedical
Neuronavigation system
Neuronavigation is the set of computer-assisted technologies used by neurosurgeons
to guide or "navigate” within the confines of the skull or vertebral column during
surgery, and used by psychiatrists to accurately target rTMS (Transcranial Magnetic
Stimulation).
BASIC PRINCIPLE
o Any point within the brain can be specified by measuring its distance along
these three intersecting planes
o It has become the standard of care for resection of many intra-axial tumours
AKHIL 92
FORCE Biomedical
Ultrasound machine
An ultrasound machine makes images so that organs inside the body can be
examined. The machine sends out high-frequency sound waves, which reflect off
body structures. A computer receives the waves and uses them to create a picture
BASIC PRINCIPLES
human hearing
used
AKHIL 93
FORCE Biomedical
and reflected
materials
o When the ultrasonic energy emitted from the probe is reflected from the
tissue, the transducer receives some of these reflections and reconverts them
Types of transducer
AKHIL 94
FORCE Biomedical
MODES
o A MODE
o B MODE
o M MODE
transducer scans a line through the body with the echoes plotted on
BASIC COMPONENTS
• Transducer
AKHIL 95
FORCE Biomedical
Ventilator
patients who cannot breathe on their own or who require assistance to maintain
monitors, and alarms. The gas is delivered using a double-limb breathing circuit.
The gas may be heated or humidified using appropriate devices. The exhalation limb
releases the gas to the ambient air. Intensive care ventilators are usually connected to
a wall gas supply. Most ventilators are microprocessor controlled and regulate the
pressure, volume, and FiO2. Power is supplied from either an electrical wall outlet
or a battery.
AKHIL 96
FORCE Biomedical
BASIC COMPONENTS
o Expiratory valve
o Patient circuit
o Flow sensor
o Blender
o Control circuits
o Humidifier
BASIC PRINCIPLES
❖ SPONTANEOUS BREATHING
Inspiration: -
o Air comes passive from the environment through the upper airways
AKHIL 97
FORCE Biomedical
SPONTANEOUS BREATHING
Inspiration
Expiration: -
AKHIL 98
FORCE Biomedical
Expiration
VOCABULARY OF VENTILATION
o MV minute volume, calculated from the tidal volume and the frequency
MV = f * Vt
mmHg
o AaDO2 difference between the partial pressure of oxygen in the alveoli and
AKHIL 99
FORCE Biomedical
MANDATORY VENTILATION
o Gas delivered by the ventilator goes through the tube, through the airways
o Thorax will be extended and the diaphragm will be pressed down positive
Modes of ventilation
❖ Volume modes
(CMV). Each breath is either an assist or control breath, but they are all of the same
volume. The larger the volume, the more expiratory time required. If the I:E ratio is
less than 1:2, progressive hyperinflation may result. ACV is particularly undesirable
for patients who breathe rapidly – they may induce both hyperinflation and
respiratory alkalosis. Note that mechanical ventilation does not eliminate the work
Guarantees a certain number of breaths, but unlike ACV, patient breaths are
AKHIL 100
FORCE Biomedical
of SIMV are increased work of breathing and a tendency to reduce cardiac output,
which may prolong ventilator dependency. The addition of pressure support on top
of spontaneous breaths can reduce some of the work of breathing. SIMV has been
breathe rapidly on ACV should switch to SIMV 2. Patients who have respiratory
Pressure Modes
Less risk of barotrauma as compared to ACV and SIMV. Does not allow for patient-
AKHIL 101
FORCE Biomedical
peak pressures and improving gas exchange [Chest 122: 2096, 2002]. The major
disadvantage is that there are no guarantees for volume, especially when lung
mechanics are changing. Thus, PCV has traditionally been preferred for patients
Allows the patient to determine inflation volume and respiratory frequency (but not
Pressure controlled ventilatory mode in which the majority of time is spent at the
higher (inspiratory) pressure. Early trials were promising, however the risks of auto
PEEP and hemodynamic deterioration due to the decreased expiratory time and
increased mean airway pressure generally outweight the small potential for
improved oxygenation
of PCV in which the I:E ratio is reversed, APRV is a variation of CPAP that releases
higher average airway pressures. Patients are able to spontaneously ventilate at both
low and high pressures, although typically most (or all) ventilation occurs at the
high pressure. In the absence of attempted breaths, APRV and PCIRV are identical.
AKHIL 102
FORCE Biomedical
Block diagram
AKHIL 103
FORCE Biomedical
About Author
My name is Akhil S, born in Kollam.
Thanking you
Akhil S
Mob: +91-9747144464
AKHIL 104
SHORT LEARNING
BIOMEDICAL ENGINEERING
Author
Akhil. S
December 2021
FORCE BIOMEDICAL
1
SHORT LEARNING ARTICLES
Table of Contents
FORCE BIOMEDICAL
2
SHORT LEARNING ARTICLES
Apnea monitors detect the cessation of breathing (apnea) in infants and adults who
are at risk of respiratory failure and alert the parent or attendant to the condition.
Some prolonged respiratory pauses result in low oxygen concentration levels in the
body, which can lead to irreversible brain damage and, if prolonged, death. The
components of apnoea monitors depend specifically on the type. However, in general
they are composed of a set of sensors which obtain the information of different
physiological parameters. This information is passed to a micro-computer system,
which analyses the sensors’ information and determines if apnea is occurring
Principles
FORCE BIOMEDICAL
3
SHORT LEARNING ARTICLES
FORCE BIOMEDICAL
4
SHORT LEARNING ARTICLES
Principle
Categories
(i) Multi wavelength Spectral Reflectance meters
(ii) Two-wavelength (460 nm, 540 nm) Spectral Reflectance meters
The major components which impart the spectral reflectance in neonates are:
melanin, dermal maturity, haemoglobin and bilirubin. Most bilirubinometers analyse
only a few wavelengths; as a result, the dermal maturity and melanin content would
interfere with the result. With these meters, separate analyses for each patient
FORCE BIOMEDICAL
5
SHORT LEARNING ARTICLES
Blood samples are required for spectrophotometric analysis. The analysis technique
depends on both the type or types of bilirubin being measured and the age of the
patient (neonate versus child or adult). A light-emitting sensor is placed on the
infant’s skin (optimally on the forehead or sternum). The reflected light is split into
two beams by a dichroic mirror, and wavelengths of 455 nm and 575 nm are
measured by optical detectors.
*****
FORCE BIOMEDICAL
6
SHORT LEARNING ARTICLES
o Hollow probe
o Irrigation sleeve
o Foot pedal
o Remote
Principle
o A very small " No Stitch " incision is made in the side of the cornea. Such an
incision promotes fast and more comfortable recovery.
o The bag of the lens is opened and the ultrasonic probe (phacoemulsifier)
removes the cloudy lens.
o A small artificial lens is inserted through the small incision to replace the
cataract lens (IOL)
o In case of opacification of the capsule (secondary cataract) , YAG laser can be
used for treatment
Technique
There are two types.
o Coaxial: a single probe is used for irrigation, aspiration and emulsification
o Bimanual: probes are smaller and can make smaller incisions .
FORCE BIOMEDICAL
7
SHORT LEARNING ARTICLES
04: ECMO
FORCE BIOMEDICAL
8
SHORT LEARNING ARTICLES
Basic principles
➢ VENO-Arterial (VA)
o Venous cannula is usually placed in the right common femoral vein for
extraction and an arterial cannula is usually placed into the right
femoral artery for infusion
o The tip of the femoral venous cannula should be maintained near the
junction of the inferior vena cava and right atrium, while the tip of the
femoral arterial cannula is maintained in the iliac artery
➢ VENO-VENOUS (VV)
o In Veno-venous ECMO – venous cannulae are usually placed in the
right common femoral vein for drainage and right internal jugular vein
for infusion
o Alternatively, a dual lumen catheter is inserted into the right internal
jugular vein, draining blood from the superior and inferior vena cavae
and returning it to the right atrium
o Patient is anticoagulated with intravenous heparin and then the
cannulae are inserted
o ECMO support is initiated once the cannulae are connected to the
appropriate limbs of the ECMO circuit
o Cannula are usually placed percutaneously by the Seldinger technique,
the largest cannulas that can be placed in the vessels are used in order
to maximise flow and minimise pressures bernoulli equation
o ECMO required for complications of cardiac surgery can be placed
directly into the appropriate chambers of the heart or great vessels
FORCE BIOMEDICAL
9
SHORT LEARNING ARTICLES
*****
FORCE BIOMEDICAL
10
SHORT LEARNING ARTICLES
It also called a scintillation camera or Anger camera and it is used to image gamma
radiation emitting radioisotopes, a technique known as scintigraphy. The
applications of scintigraphy include early drug development and nuclear medical
imaging to view and analyse images of the human body or the distribution of
medically injected, inhaled, or ingested radionuclides emitting gamma rays
Basic components
Construction
A gamma camera consists of one or more flat crystal planes (or detectors) optically
coupled to an array of photomultiplier tubes, the assembly is known as a "head",
mounted on a gantry. The gantry is connected to a computer system that both
controls the operation of the camera as well as acquisition and storage of acquired
images. The construction of a gamma camera is sometimes known as a
compartmental radiation construction. The system accumulates events, or counts, of
gamma photons that are absorbed by the crystal in the camera. Usually, a large flat
crystal of sodium iodide with thallium doping in a light-sealed housing is used. The
FORCE BIOMEDICAL
11
SHORT LEARNING ARTICLES
o Photomultiplier tubes (PMTs) behind the crystal detect the fluorescent flashes
and a computer sums the counts
o Generally, each tube has an exposed face of about 7.6 cm in diameter and the
tubes are arranged in hexagon configurations, behind the absorbing crystal
o All the PMTs simultaneously detect the (presumed) same flash of light to
varying degrees, depending on their position from the actual individual event
o The location of the interaction between the gamma ray and the crystal can be
determined by processing the voltage signals from the photomultipliers; in
FORCE BIOMEDICAL
12
SHORT LEARNING ARTICLES
simple terms, the location can be found by weighting the position of each
photomultiplier tube by the strength of its signal, and then calculating a mean
position from the weighted positions
o The total sum of the voltages from each photomultiplier is proportional to the
energy of the gamma ray interaction, thus allowing discrimination between
different isotopes or between scattered and direct photons
Spatial resolution
o In order to obtain spatial information about the gamma-ray emissions from an
imaging subject, a method of correlating the detected photons with their point
of origin is required
o The conventional method is to place a collimator over the detection
crystal/PMT array
o The collimator consists of a thick sheet of lead, typically 1-3 inches thick, with
thousands of adjacent holes through it
o The collimator attenuates most of incident photons and thus greatly limits the
sensitivity of the camera system
Imaging techniques
*****
FORCE BIOMEDICAL
13
SHORT LEARNING ARTICLES
Advantages
o Similar to digital photography
o It does not require dark room procedure and cassette reading
o It reduces X-ray printing time
o X-ray directly converted into electronic signal, convert into digital values and
into images
Basic components
o A digital image receptor
o A digital image processing unit
FORCE BIOMEDICAL
14
SHORT LEARNING ARTICLES
➢ Digital Receptor
The digital receptor is the device that intercepts the x-ray beam after it has passed
through the patient’s body and produces an image in digital form, that is, a matrix of
pixels, each with a numerical value.
This replaces the cassette containing intensifying screens and film that is used in non-
digital, film-screen radiography.
FORCE BIOMEDICAL
15
SHORT LEARNING ARTICLES
➢ Imaging Processing
One of the major advantages of digital radiography is the ability to process the
images after they are recorded. Various forms of digital processing can be used to
change the characteristics of the digital images. For digital radiographs the ability to
change and optimize the contrast is of great value.
It is also possible to use digital processing to enhance visibility of detail in some
radiographs. The various processing methods are explored in much more detail in
another module.
➢ Communications Network
Another advantage of digital images is the ability to transfer them from one location
to another very rapidly.
This can be:
Within the imaging facility to the storage and display devices and To other locations
(Teleradiology) Anywhere in the world (by means of the internet)
➢ Digital Image Display and Display Control
Compared to radiographs recorded and displayed on film, i.e. "softcopy", there are
advantages of "softcopy" displays.
Basic components of Radiography
o X-ray tube
o X-ray detector
o Collimator
o HV generator
o Filters
o Console computer
FORCE BIOMEDICAL
16
SHORT LEARNING ARTICLES
➢ X-ray tube
o Includes cathode, anode and casing
o Cathode filament is supplied with 20-150 KV
o X-ray produced by mainly two method electron ejection and electron
deceleration
o Electron ejection (Characteristic X-ray radiation): Fast moving electron
hit electrons in innermost shell, hence electron vacancy is created, to
occupy this vacancy by electron from another shell by emitting X-ray
radiation
o Electron deceleration (Bremsstrahlung radiation): When fast moving
electrons coming to anode, anode atoms look’s as Positive nucleus and
negative electrons thus it moves any of region emitting X-ray radiation
o Heating curve: time taken to achieve particular MA and KV
o Cooling curve: time taken to cool X-ray tube
o Energy at x-ray tube= Kv*MA*Exposure time
o No of electrons in X-ray depends current, voltage which improves the
intensity of x-ray beam
➢ H V generator
o Used for giving supply to the X-ray tube
o Voltages ranges
o This is mainly including single phase or three phases
o Supply specification mainly represented by Peak kilo voltage (Kvp)
o H V: Transformer Step up the input voltage
o Rectifier: It convert High voltage Ac supply into Dc supply
o Chopper: Increases the frequencies up to 200KHZ
FORCE BIOMEDICAL
17
SHORT LEARNING ARTICLES
FORCE BIOMEDICAL
18
SHORT LEARNING ARTICLES
Basic principle
❖ BEER-LAMBERT LAW
When a monochromatic light source pass through a medium,
attenuation of light is directly proportional to concentration of
substance present in the solution
PRINCIPLE OF OPERATION
When an x-ray radiation from an x-ray tube passed through the body,
then it falls to the x-ray detector. Detector converts x-ray into electrical
signal and then it is digitized by A-D Convertor
*****
FORCE BIOMEDICAL
19
SHORT LEARNING ARTICLES
FORCE BIOMEDICAL
20
SHORT LEARNING ARTICLES
Basic principle
o To conduct the scan, a short-lived radioactive tracer isotope is injected into the
living subject
o During the scan a record of tissue concentration is made as the tracer decays
o The emitted positron travels in tissue for a short distance, during which time it
loses kinetic energy, until it decelerates to a point where it can interact with an
electron
FORCE BIOMEDICAL
21
SHORT LEARNING ARTICLES
o These are detected when they reach a scintillator in the scanning device,
creating a burst of light which is detected by photomultiplier tubes or silicon
avalanche photodiodes (Si APD)
o The technique depends on simultaneous or coincident detection of the pair of
photons moving in approximately opposite directions
o Photons that do not arrive in temporal "pairs" (i.e. within a timing-window of
a few nanoseconds) are ignored
RADIONUCLIDES & RADIOTRACERS
o Radionuclides used in PET scanning are typically isotopes with short half-
lives such as carbon-11 (~20 min), nitrogen-13 (~10 min), oxygen-15 (~2 min),
fluorine-18 (~110 min)., or rubidium-82(~1.27 min)
FORCE BIOMEDICAL
22
SHORT LEARNING ARTICLES
Alternate’s inflation and deflation of cells to constantly change pressure points and
promote circulation. A single cell slowly deflates, re-inflates, and then the adjacent
cell does the same, slowly, up and down the mattress. The head area of the mattress
does not alternate so the patient remains undisturbed during sleep. In fact, the
patient doesn't really feel the cells alternating at all since the process is slow. The
speed at which the alternating process cycles from one end of the mattress to the
other end is called cycle time and it can be adjusted or set on most mattress systems.
Usually, 10-minute cycle time is sufficient.
Basic features
FORCE BIOMEDICAL
23
SHORT LEARNING ARTICLES
Basic components
o Arterial pump
o Backup arterial pump
o One or two suction pumps
o Cardioplegia pump
Basic principles
o During heart-lung bypass, blood that would normally return to the heart
through the venae cavae is diverted by means of plastic cannulae through a
presterilized disposable tubing set to an arterial pump
FORCE BIOMEDICAL
24
SHORT LEARNING ARTICLES
o Blood that pools in the surgical site is suctioned and collected in a cardiotomy
reservoir, where it is filtered, defoamed, and either pumped or drained to the
oxygenator
o The oxygenated blood is then returned to the patient, typically through the
aorta, bypassing the heart and lungs
Oxygenation
o The type of oxygenator used is the membrane oxygenator.
o Have an integral heat exchanger to control the temperature of the blood,
although heat exchangers are also available as separate disposable units
o Must be primed before use, usually with a priming solution such as lactated
Ringer’s solution with 5% dextrose, to displace air pockets and bubbles from
the bypass circuit
FORCE BIOMEDICAL
25
SHORT LEARNING ARTICLES
*****
FORCE BIOMEDICAL
26
SHORT LEARNING ARTICLES
Autotransfusion units are designed to collect and process blood lost by a patient to
extract red blood cells (RBCs) for subsequent transfusion into the same patient. They
are particularly intended for use in surgeries that involve substantial volume blood
loss.
Basic components
o Collection reservoir
o Roller pump
o Centrifuge bowl
o Air detector
o Microfilter
Basic principles
o Use of auto transfusion devices requires aspiration of extravascular blood
from the surgical site into the collection reservoir of the unit, where it
undergoes gross filtration
FORCE BIOMEDICAL
27
SHORT LEARNING ARTICLES
o To keep the blood from clotting while in the unit, an anticoagulant (heparin) is
added to the aspiration line at a rate proportional to the rate of blood
collection
o From the reservoir, the unit’s roller pump sends the blood into a centrifuge
bowl
o Once the bowl is full, the pump shuts off and the RBCs are spun and washed
with normal saline to remove anticoagulant
o The saline wash is monitored by an air detector
o When a saline bag is depleted during the wash cycle, residual air in the saline
bag is detected by the air detector, triggering an alarm to alert the operator to
replace the bag
o During the wash cycle, the saline flows into a waste bag, leaving the packed
RBCs in the centrifuge bowl
o The RBCs are then pumped into a holding bag for reinfusion and delivered to
the patient through a microfilter
o Operate in an automatic or a manual mode
o The automatic mode is most often used for routine procedures; the operator
simply selects a programmed cycle choice with preset wash volumes and
pumping rates and initiates the cycle
o The system detects when the bowl is full, automatically triggering the
separation and wash cycle
o High-speed processing units can process one unit of packed red cells (250 ±50
mL) in about three minutes; faster cycle times typically yield less efficient
washes
o A manual mode allows operators to control the start of each processing stage
and the saline volume in wash cycles for maximum flexibility in exceptional
cases
o Faster cycle times can be achieved in this mode because wash cycles can be
eliminated if rapid reinfusion is necessary; however, the operator should try
increasing pump speeds before choosing to eliminate wash cycles to avoid
giving patients potentially contaminated blood
FORCE BIOMEDICAL
28
SHORT LEARNING ARTICLES
FORCE BIOMEDICAL
29
SHORT LEARNING ARTICLES
11: Humidifiers
Basic principle
o Inhaling dry gases can cause damage to the cells lining the respiratory tract,
impairing ciliary function
o Within a short period of just 10 mint of ventilation with dry gases, cilia
function will be disrupted
o This increases the patient’s susceptibility to respiratory tract infection
o A decrease in body temperature occurs at the respiratory tract humidifies the
dry gases
o Air fully saturated with water vapour has an absolute humidity of about 44
mg/L at 37◦C
o During nasal breathing at rest, inspired gases become heated to 36◦C with a
relativve humidity of about 80-90% by the time they reach the carina, largely
because of heat transfer in the nose
FORCE BIOMEDICAL
30
SHORT LEARNING ARTICLES
Basic components
FORCE BIOMEDICAL
31
SHORT LEARNING ARTICLES
• Water vapor content and temperature of the inspired and exhaled gases
• Inspiratory and expiratory flow rates affecting the time the gas is in contact
with the HME medium hence the heat and moisture exchange
• The volume and efficiency of the HME medium- the larger the medium,
the greater the performance
*****
FORCE BIOMEDICAL
32
SHORT LEARNING ARTICLES
12: CUSA
Intended use
o Innovative tool for dissecting through the liver parenchyma, which can
potentially reduce intraoperative blood loss and perioperative morbidity
o Powerful ultrasonic aspirator and dissector with a wide application not only
in liver surgery but also in other surgical specialties as well
o Used in multiple surgical subspecialties, including neurosurgery,
gastrointestinal, hepatobiliary surgery, gynecology, and urology
o Ultrasonic surgical aspirator, where fragmentation, suction, and irrigation
occur simultaneously, allowing the surgeon to remove tissue with accurate
control
o In liver surgery, it is an invaluable tool, particularly in a situation where the
tumour is closely adjacent to a vital structure that needs to be saved
FORCE BIOMEDICAL
33
SHORT LEARNING ARTICLES
o CUSA will enable dissection around any structure that needs to be preserved
Basic components
o Transducer: -
• A device that converts electromagnetic energy into mechanical
vibration
• The transducer is composed of a stack of nickel alloy plates
• A magnetic field is produced by a coil plced around the plates and
causes mechanical motion of approximately 300 microns
o Connecting body: -
• Mechanically conveys the motions of the transducer to the surgical tip
• It also amplifies the vibration motion of the transducer
o Surgical tip: -
• Completes the amplifications of the motion and also contacts the tissue.
• Hence tip is relatively long compared to its diameter and this provides
adequate motion amplification
Basic principle
o Cavitation
• Cavitation is defined as the process of formation of the vapour phase of
a liquid when it is subjected to reduced pressures at constant ambient
temperature
• Thus, it is the process of boiling in a liquid as a result of pressure
reduction rather than heat addition
• Cavitation occurs when, on the negative side of a pressure cycle, such
as when the probe-tip is retracting with sufficient amplitude and
frequency, suspended gas bubbles either within fluid, tissue or trapped
at solid interfaces expand and collapse resuling in the generation of
shockwaves
FORCE BIOMEDICAL
34
SHORT LEARNING ARTICLES
o Aspiration
o Ultrasound
▪ Audible Range:-20Hz---------20,000Hz.
▪ Infrasonic waves:-Sound waves with frequencies
<20Hz
▪ Ultrasonic waves:-Sound waves with frequencies
>20,000Hz
FORCE BIOMEDICAL
35
SHORT LEARNING ARTICLES
• This mechanical energy is delivered through a hollow 3mm tip that vibrates at
23,000 cycles per second
Piezoelectric transducer
FORCE BIOMEDICAL
36
SHORT LEARNING ARTICLES
*****
FORCE BIOMEDICAL
37
SHORT LEARNING ARTICLES
13: LINAC
A medical linear accelerator (LINAC) is the device most commonly used for external
beam radiation treatments for patients with cancer. It delivers high-energy x-rays or
electrons to the region of the patient's tumour. These treatments can be designed in
such a way that they destroy the cancer cells while sparing the surrounding normal
tissue. The LINAC is used to treat all body sites, using conventional techniques,
Intensity-Modulated Radiation Therapy (IMRT), Volumetric Modulated Arc Therapy
(VMAT), Image Guided Radiation Therapy (IGRT), Stereotactic Radiosurgery (SRS)
and Stereotactic Body Radio Therapy (SBRT).
Component
Equipment working
The linear accelerator uses microwave technology (similar to that used for radar) to
accelerate electrons in a part of the accelerator called the "wave guide," then allows
these electrons to collide with a heavy metal target to produce high-energy x-rays.
FORCE BIOMEDICAL
38
SHORT LEARNING ARTICLES
These high energy x-rays are shaped as they exit the machine to conform to the shape
of the patient's tumor and the customized beam is directed to the patient's tumor.
The beam is usually shaped by a multileaf collimator that is incorporated into the
head of the machine. The patient lies on a moveable treatment couch and lasers are
used to make sure the patient is in the proper position. The treatment couch can
move in many directions including up, down, right, left, in and out. The beam comes
out of a part of the accelerator called a gantry, which can be rotated around the
patient. Radiation can be delivered to the tumor from many angles by rotating the
gantry and moving the treatment couch.
Six classes
• Injection system;
• RF power generation system;
• Accelerating waveguide;
• Auxiliary system;
• Beam transport system;
• Beam collimation and beam monitoring system.
Magnetron
Device that produces microwaves. It has a cylindrical construction, having a central
cathode and outer cathode. Space between cathode and anode is evacuated. Cathode
is heated by an inner filament and the electrons are generated by thermionic
emission.
FORCE BIOMEDICAL
39
SHORT LEARNING ARTICLES
Klysteron
FORCE BIOMEDICAL
40
SHORT LEARNING ARTICLES
cavity they induce charges on the ends of the cavity and there by generate a retarding
electric field.
The electrons suffer deceleration and by the principle of conservation of energy the
K.E of electrons is converted into high power microwaves.
Gantry
o Source of radiation can rotate 360 degree
o As the gantry rotates collimator axis moves in a vertical plane. The point of
intersection of collimator axis and the axis of rotation of the gantry is known
as iso center.
o The gantry consists of: Electron gun, Accelerator Structure, Treatment Head
FORCE BIOMEDICAL
41
SHORT LEARNING ARTICLES
FORCE BIOMEDICAL
42
SHORT LEARNING ARTICLES
Beam collimation
In a typical modern medical linac, the photon beam collimation is achieved with two
or three collimator devices:
• A primary collimator
• Secondary movable beam defining collimators
With the current trend using linear accelerator a variety of treatment techniques are
possible. Some of the techniques as follow,
o 3D conformal radiotherapy
o Intensity Modulated Radiotherapy
o Stereotactic Radiotherapy (SRT)
o Stereotactic Radiosurgery (SRS)
o Dynamic Adoptive Radiotherapy (DART)
o Image guided Radiotherapy (IGRT)
*****
FORCE BIOMEDICAL
43
SHORT LEARNING ARTICLES
Principle
A basic principle is Gay Loussac’s law i.e., when the pressure of a gas increases, the
temperature of the gas increase proportionally. i.e., about 15 pounds of pressure per
square inch (Psi), the temperature rises to 121oC. Increasing the pressure to 20 psi
raises the temperature to 134oC.
Working
o Steam flows through the sterilizer beginning the process of displacing the air.
o Exhaust valve will be closed, causing the interior temperature and pressure to
ramp up to the desired setpoint.
o Desired temperature is maintained until desired time is reached.
o The pressure is released from the chamber through an exhaust valve
FORCE BIOMEDICAL
44
SHORT LEARNING ARTICLES
Parts
• Chamber and jacket
• Thermostatic traps
• Vacuum system
• Steam generator
Thermostatic traps
A device designed to allow air and water (condensate) to escape from the chamber.
Steam traps are temperature sensitive valves that close when heated past a certain set
point.
Vacuum system
Forcibly remove air by pulling a vacuum on the chamber before a cycle (also known
as pre-vacuum), it also helps by pulling a vacuum after the cycle (also known as post-
vacuum) to remove the steam remaining in the chamber and dry off the goods inside
the autoclave.
Steam generator
Boilers are typically i.e., underneath the chamber to the steriliser and utilize electric
heating elements to heat water and generate steam.
FORCE BIOMEDICAL
45
SHORT LEARNING ARTICLES
➢ 5 Phases of Sterilization
1. Phase conditioning
• Chamber closes with jacket
• Steam loaded
• Jacket pressure and chamber pressure displayed
2. Heating conditioning
• Steam generated
3. Sterilization phase
• 121 C at 10 psi in 20 min
• 134 C at 20 psi in 5 min
4. Dry phase
• Removal of water
5. Aeration/Vent
• Releasing high pressure
• Use vacuum
FORCE BIOMEDICAL
46
SHORT LEARNING ARTICLES
*****
FORCE BIOMEDICAL
47
SHORT LEARNING ARTICLES
Principles
Spirometry instruments measure the volume of gases exhaled by the patient (i.e.,
volume changes of the lungs) either by volume displacement or flow sensing
methods. Spirometers measure the volume directly; these devices include water-seal
bellows and rolling-seal spirometers, or the flow of gas that is integrated to yield
volume. Such flow sensing instruments can employ a pneumotachometer, a hot-wire
anemometer, or a turbinometer. Some analysers incorporate computers with software
that permits customized reports or the inclusion of specialized predictive equations
for normal function.
FORCE BIOMEDICAL
48
SHORT LEARNING ARTICLES
It is Used primarily to locate and identify irregularities within the heart and its
vasculature, the aorta, or the vena cava, as well as to define the size or severity of
lesions and Complete evaluation of a patient with known or suspected heart
disease. It also used in diagnosing disorders such as reduced left ventricular
function, valve incompetence, pulmonary vascular disease, and congenital
anomalies and used for pacemaker lead placement.
Basic components
• C-Arm
• Pressure injector
• Dye injector
• Hemodynamic monitor
• Review station
• Diagnostic catheter
• Guide catheter
• Guide wires
• Stent
• Angioplasty balloon
FORCE BIOMEDICAL
49
SHORT LEARNING ARTICLES
Basic principle
➢ The basic cathlab recording system consists of a computer console and a chart
recorder, which typically reside in a control room
➢ A slave scope and patient interface modules reside in the cathlab
➢ All the systems are multichannel and display and record up to 32 traces of
information from the patient
➢ Linking the cathlab monitor/recorder with a computer allows most of the
hemodynamic parameters to be calculated automatically
➢ The parameters that are automatically derived typically include cardiac
output, pressure gradients, valve areas, shunt flows, vascular resistance,
diastolic filling period, systolic ejection period, work, and preejection period
➢ These values are stored chronologically, along with other important events
occurring during catheterization, and can be printed in report form at the
conclusion of the procedure
➢ It is possible to interface with the cathlab imaging system to provide x-ray
image visualization on the physiologic monito
➢ Before catheterization, a full 12-lead electrocardiogram (ECG) is usually
performed to obtain baseline values
➢ This preliminary study can also disclose unsuspected abnormalities. During
the procedure, continuous arterial pressure can be monitored directly
➢ For right-heart catheterization (RHC), the catheter is introduced into a vein in
the arm, the external jugular vein, or the femoral vein in the groin and then
manipulated into the right atrium of the heart under direct observation using
fluoroscopy
➢ The catheter is then passed through the right ventricle and lodged in the
pulmonary artery
➢ In left-heart catheterization, which is less common, the catheter is introduced
into an artery in the arm or into the femoral artery and maneuvered through
the aortic valve and into the left ventricle
FORCE BIOMEDICAL
50
SHORT LEARNING ARTICLES
➢ The fluid-filled lumen of the catheter transfers pressure fluctuations from the
open distal end of the catheter inside the patient to a transducer at the
proximal end outside the patient, which converts the mechanical force of the
applied pressure into electrical signals that are proportional to that pressure
➢ These signals are then transmitted to a pressure amplifier in the recording
system
➢ Generally, a larger lumen allows for a better pressure reading
FORCE BIOMEDICAL
51
SHORT LEARNING ARTICLES
➢ The second method uses a transducer in the catheter tip to measure pressure
fluctuations
➢ The status of the myocardium can be evaluated by measuring pulmonary and
systemic blood flows to determine cardiac output
➢ The most widely used method is the thermal dilution technique in which a
chilled or room-temperature saline solution is injected through the catheter
into the right atrium; its temperature is measured by a thermistor at the distal
end of the catheter, located in the pulmonary artery
➢ The temperature change correlates with the blood flow through the heart
➢ ECG/His amplifiers receive signals from electrode catheters placed inside the
various chambers of the heart and from electrodes on the chest
➢ Studies of the bundle of His provide an assessment of the heart’s electrical
conduction system
FORCE BIOMEDICAL
52
SHORT LEARNING ARTICLES
17: Ophthalmoscope
It is used for performing ophthalmoscopy and used to see inside the fundus of an eye
and other structures using an ophthalmoscope. Usually used to check the retina and
vitreous humour.
Basic Principle
If patient and observer are both
emmetropic, the rays emitting from a
point in the patient’s fundus will emerge
as a parallel beam and will be focused
on the observer’s retina.
Basic components
• Red button
• Power base
• Number settings
• Lens
• Eye rest
• Vertical dial
• Red button
• Power base
• Number settings
FORCE BIOMEDICAL
53
SHORT LEARNING ARTICLES
Types
➢ Direct ophthalmoscope
➢ Indirect ophthalmoscope
Direct ophthalmoscope
Indirect ophthalmoscope
FORCE BIOMEDICAL
54
SHORT LEARNING ARTICLES
Features
FORCE BIOMEDICAL
55
SHORT LEARNING ARTICLES
18: Centrifuge
A centrifuge is a laboratory device that is used for the separation of fluids, gas or
liquid, based on density. These are high- or low-speed centrifuges capable of
spinning a large volume of specimen.
Basic components
➢ Basic centrifuge components include
• electric motor
• Shaft
• rotor heads (often interchangeable) on which the
centrifuge head turns
• motor-drive assembly.
➢ The entire system is housed within a chamber
➢ The centrifuge head contains the cups or shields that cover the rotor and turns
on a spindle
FORCE BIOMEDICAL
56
SHORT LEARNING ARTICLES
Basic Principle
FORCE BIOMEDICAL
57
SHORT LEARNING ARTICLES
19: Lithotripters
It is used to locate, view, fragment and remove urinary tract or renal calculi found in
the kidney, ureter and bladder and also disintegrate difficult-to-treat bile direct
stones and salivary stones.
Basic Principles
➢ Based on the principal of operation lithotripters can be classified into
• Electro hydraulic lithotripsy
• Laser lithotripsy
• Ultrasonic Impact lithotripsy
• Electromechanical and pneumatic lithotripsy
• Percutaneous lithotripsy
• Transurethral lithotripsy
FORCE BIOMEDICAL
58
SHORT LEARNING ARTICLES
• With the stone in view, the operator fires these pulses at the stone while
the probe is irrigated with saline solution
• Firing the probe produces a plasma bubble causes a hydraulic shock
that fragments the stone, that can subsequently be removed using a
grasper or stone basket
Laser lithotripsy
FORCE BIOMEDICAL
59
SHORT LEARNING ARTICLES
Percutaneous lithotripsy
➢ In this, a needle is inserted into the patients flank and guided into the
renal pelvis with the help of fluoroscopic visualization, a guide wire is
inserted through the needle and the needle is removed
FORCE BIOMEDICAL
60
SHORT LEARNING ARTICLES
➢ A second guide wire is introduced down the ureter in the same manner
to maintain access throughout the procedure in case the nephroscope
becomes dislodged
➢ Dilation of the tract can then be performed using a balloon catheter
➢ Following dilation, the nephroscopic sheath is inserted over the guide
wire and positioned near the calculus
➢ The nephroscope and lithotripter probe or laser fiber are placed inside
the sheath and activated to disintegrate the stone
➢ The resulting fragments are evacuated by suction through the probe
Transurethral lithotripsy
➢ Transurethral ureteroscopic lithotripsy approaches the calculus in an
ascending direction through the urinary tract
➢ A sheath and then a ureterorenoscope are passed into the ureter usually
following ureteral dilation where the lithotripsy probe fragments the
stone
FORCE BIOMEDICAL
61
SHORT LEARNING ARTICLES
Basic components
FORCE BIOMEDICAL
62
SHORT LEARNING ARTICLES
Basic principles
➢ All CSUs employ either a closed or an open system
➢ In a closed-system CSU, the cryogen flows through an insulated shaft in the
hollow probe, cools the tip, and is exhausted back through the probe
➢ For e.g. in ophthalmic surgery, closed systems are used for cryonecrosis,
cryoadhesion, and cryoinflammation because they afford greater control over
both the rate of cooling and the area of freezing than do open systems
➢ Open-system CSUs apply cryogen directly to the target tissue; they are
generally limited to cryonecrosis and are rarely used for ophthalmic surgery
➢ CSUs using N2O or CO2 are not usually suitable for use as open systems
because cryogen “snow” would build up on the target tissue and insulate the
lesion from the cryogen spray
➢ The mechanisms of cooling differ among cryogens
➢ Liquid nitrogen CSUs deliver the cryogen to the tip as a liquid, where its rapid
vaporization cools the probe
➢ In closed-system N2O and CO2 units, cooling occurs through the Joule-
Thomson effect, in which a compressed gas is allowed to expand suddenly
through a small aperture inside the probe tip, causing a considerable drop in
gas temperature and liquefaction of some of the cryogen
➢ The vaporization of the liquefied cryogen from
the interior of the tip, combined with the drop in
gas temperature caused by expansion, lowers the
tip temperature to near the boiling point of the
cryogen
➢ When a cryoprobe or spray of liquid nitrogen is
placed on the target tissue, an adhesive bond of
ice crystals forms between the probe and the
moist tissue, and a cryolesion forms within the
tissue surface
FORCE BIOMEDICAL
63
SHORT LEARNING ARTICLES
➢ Within the cryolesion, the lowest temperature is closest to the cryoprobe; the
temperature increases with distance
➢ Internal cryoprobes, such as those used for prostate cryosurgery, develop an
iceball that surrounds the entire probe tip
➢ Regardless of the freezing method, the cryolesion must extend beyond the
target tissue to ensure that the cryobiologically lethal temperature (-20° to -
40°C) is reached in the desired area
FORCE BIOMEDICAL
64
SHORT LEARNING ARTICLES
About Author
Name: Akhil S
Working Status:
➢ Biomedical Technician under Sur MOH
Hospital. Oman
Previously worked institution:
➢ KIMS Health. Trivandrum, Kerala
➢ Aster DM WIMS. Wayanad, Kerala
➢ Sree Gokulam Medical. Trivandrum, Kerala
Akhil S
FORCE BIOMEDICAL
65