ACON Laboratories, Inc.

U500 Urine Analyzer

Service Manual

Page 1 of 67
CONFIDENTIAL
for Distributor Reference Only
ACON Laboratories, Inc.

Content

Part 1 General Information of U500


Part 2 Operating Instruction
Part 3 Check and Repair
Part 4 Spare Parts
Part 5 Error messages, possible reasons and troubleshooting
Part 6 Cleaning and Maintainance
Part 7 Language Chip Replacement Package Insert
Part 8 Detection Optimization

Page 2 of 67
CONFIDENTIAL
for Distributor Reference Only
ACON Laboratories, Inc.

1. General Information
1.1 Instruction
The U500 Urine Analyzer is a reflectance photometer that analyzes the intensity and color of light reflected
from the reagent areas of a urinalysis reagent strip. The analyzer also prints testing results through the
internal or external printer.
This service manual describes the control and repair of the U500. It is written for use by a well-trained
service engineer, who is experienced in repairing and maintaining instruments for -Diagnostic
investigations.
The U500 is a precise, calibrated optical measurement system. The listed operations below must be
accomplished with special attention and precision. Please take special care to disconnect the device from
the power supply if this is required by the instruction. If spare parts or accessories are needed, it is
absolutely necessary to use only original spare parts for this instrument.

Page 3 of 67
ACON Laboratories, Inc.

1.2 Components

Page 4 of 67
ACON Laboratories, Inc.

Page 5 of 67
ACON Laboratories, Inc.

2. Operating instruction

Turn on the Power Switch

Possible reason:
Yes Yes
No response from No response of Power 1. No electrical supply
Analyzer Switch indicator light 2. AC power cord disconnected or
damaged
No

Yes
Fuse is broken. Replace the fuse. Refer to section3.1.1

No
Possible reason:
The input voltage of Yes 1. The wire from the power socket
Switching Power to Switching Power Supply is
Supply is abnormal. loose or damaged.
2. Replace the Power Switch Unit.
No Refer to section 3.1.7
No
The output voltage of Yes Replace the Switching Power Supply.
Switching Power
Refer to section 3.1.5.
Supply is abnormal.

No
The wire harness connecting
The input voltage of main Yes Switching Power Supply to main
board is abnormal. board is loose or damaged. Refer to
section 3.1.6

No

Replace main board.


Refer to section 3.1.4

Analyzer keeps beeping

Yes
Replace main board.
Refer to section 3.1.4

Page 6 of 67
ACON Laboratories, Inc.

No
Yes The FFC connecting
Yes 1. Reseat the FFC connectors;
No LCD display LCD to Printer board is
2. Replace the FFC. Refer to section
loose or damaged.
3.2.1.

No

Yes
The 50-pin flat cable from 1. Reseat the cable connectors;
main board to Printer
2. Replace the 50-pin flat cable
board is loose or
damaged.

No No
Yes
LCD is damaged. Replace the LCD.
Refer to section 3.2.3

No
Yes
Replace the Printer board.
Printer board is damaged.
Refer to section 3.2.4

No
1. Potentiometer adjustment on
Replace main board. Main board is not proper.
Refer to section 3.1.4 Refer to section 3.2.5.
2. LCD damaged. Refer to section
The brightness of the Yes 3.2.3.
Yes
LCD display is abnormal characters on LCD is too 3. Printer board damaged. Refer
dark or contrast is low. to section 3.2.4

1. Reseat the connectors of cable


connecting LCD to Printer
board
2. Replace the FFC connecting
No LCD to Printer board. Refer to
section 3.2.1
3. Reseat the connectors of 50-pin
Yes
Other abnormal flat cable
appearance 4. Replace 50-pin flat cable. Refer
to section 3.2.2
5. Replace the LCD. Refer to
section 3.2.3
6. Replace the Printer board. Refer
to section 3.2.4
Page 7 of 67
7. Replace main board.
Section3.1.4
ACON Laboratories, Inc.

No
1. Reseat the Strip sensor board
Yes Yes connector. Refer to section
The strip sensor The sensor can’t respond
can’t sense the strip the strip at all 3.3.1
properly 2. Replace the Detection Unit.
No Refer to section 3.1.3

Yes 1. Clean the Strip sensor board.


The sensor can’t sense the Refer to section 3.3.2
strip in partly area. 2. Replace the strip sensor
board. Refer to section 3.3.1

No

No
Replace main board.
Refer to section 3.1.4

1. Reseat the motor connector.


Yes Yes
The strip catch The strip catch motor Refer to section 3.3.5;
motor can’t move can’t move at all. 2. Replace the Detection Unit.
properly. Refer to section 3.1.3;

No

1. Clean the Strip Carry


Furcation. Refer to section
Yes
The strip catch motor 3.3.3
can’t move smoothly. 2. Replace the Strip Carry
Furcation, refer to section
No 3.3.13

No
1. Clean the optical coupler
Yes Refer to 3.3.8
The strip catch motor
2. Replace Detection unit. Refer
can’t drive the belt to
to section 3.1.3
desired position.
3. Replace the Synchro-Strap;
No refer to section 3.3.12

Replace main board.


Refer to section 3.1.4

Page 8 of 67
ACON Laboratories, Inc.

No

Yes Yes 1. Reseat the motor connector.


The strip transfer The strip transfer motor
Refer to section 3.3.6
motor can’t move can’t move at all.
2. Replace the Detection Unit.
properly
Refer to section 3.1.3;

No

Yes
The strip transfer motor Replace the Detection Unit. Refer
can’t move smoothly. to section 3.1.3;

No

1. Clean the optical coupler


Refer to Section 3.3.9;
Yes 2. Replace the Detection Unit.
The strip transfer motor
can’t drive the belt to Refer to section 3.1.3;
desired position. 3. Replace the Strip Carry
Furcation, refer to
section3.3.13;

No
No

Replace main board.


Refer to section 3.1.4

Page 9 of 67
ACON Laboratories, Inc.

Yes 1. Reseat the motor connector.


The strip scanning Yes The strip scanning motor
motor does not does not move at all. Refer to section 3.3.7
move properly. 2. Replace Detection unit,
Refer to section 3.1.3,
3. Replace the strip scanning
No motor,refer to section 3.3.14;

1. Replace the MXL


Yes Synchro-Strap, refer to
The strip scanning motor
section 3.3.11
does not move smoothly.
2. Replace the Following
Pulley, refer to section3.3.15
No 3. Replace Detection unit,
Refer to section 3.1.3;
No
Yes
The test results behave
Replace Detection unit, Refer to
obviously error.
section 3.1.3

No

Replace main board.


Refer to section3.1.4.

Yes Yes Replace the main board. Refer to


Self testing failed Main control unit failed section 3.1.4.

1. White calibration target is not


No clean, Refer to Section 3.3.4;
Yes 2. The Strip bed and waste box;
Optical sensor failed is positioned incorrectly on
Detection unit.
3. Replace the Detection Unit.
Refer to section 3.1.3
Yes
No Detection Unit failed. 1. Replace the Detection Unit.
Refer to section 3.1.3
2. Replace main board. Refer to
No
section 3.1.4

Yes 1. Replace the Detection Unit.


“Excessive light” failed Refer to section 3.1.3
2. Replace the main board. Refer
No to section 3.1.4

Page 10 of 67
ACON Laboratories, Inc.

Touch screen Yes The cable connecting Yes 1. Reseat the cable
abnormal LCD to Printer board is connectors;
loose or damaged. 2. Replace the cable. Refer
to section 3.2.1;

No

The 50-pin flat cable 1. Reseat the cable


Yes
connecting the main connectors;
board to Printer board is 2. Replace the cable. Refer
loose or damaged. No to section 3.2.2;
No

No
Yes
Replace the LCD. Refer to
LCD is damaged.
section 3.2.3;

No

Yes
Replace the printer board.
Printer board is damaged.
Refer to section 3.2.4;

Yes
Menu display failure Replace the main board.
Refer to section 3.1.4;

No

Yes Yes
Replace the battery. Refer to
Wrong configuration Battery is out of power.
section 3.4.1
of Date/Time

No
Yes
Bad contact between 1. Install the battery again.
battery and battery 2. Replace the main board.
contacts. Refer to section 3.1.4
No

No

Replace the main board.


Refer to section 3.1.4

Page 11 of 67
ACON Laboratories, Inc.

Yes Wrong External Data Yes


Data Transfer Use correct cable.
Transfer cable.
incorrect or failed

No
Yes
Wrong communication Setup the parameters as
parameters. defined in the user’s manual.

No
1. Reseat the cable
Yes
connectors
FFC connecting RS232 to
2. Replace the FFC. Refer to
main board is loose or
section 3.5.1
damaged.
No

No

Yes Replace the Data Transfer


The Data Transfer Cable
Unit is damaged. Cable Unit. Refer to section

No Possible Reason:
1. The printer is not set as
Replace the main board. “Internal”;
Refer to section 3.1.4. 2. The cable connecting
printer to Printer board is
loose or damaged; Refer
Yes
Internal See 1,2,3,5,6,7 to section 3.6.1
Printer has no response
printer 3. The 50-pin flat cable
abnormal connecting main board to
Printer board is loose or
Printer result is illegible See 2,4,5,10
damaged; Refer to section
3.2.2
4. The printer roller and
Printer does not feed paper See 2,3,5,6,7
priter head is
contaminated; Refer to
Printer feeds paper, but section 3.6.3
See 2,3,7,8,9,10 5. Printer is damaged. Refer
the printout is blank
to section 3.6.2
No
6. Printer board is damaged.
Printerout is incomplete See 2,4,5
Refer to section 3.2.4
7. Main board is damaged.
Characters in printedout Refer to section 3.1.4
See 2,5,6
are narrow 8. The printer paper isn’t
placed properly
Does not feed paper 9. The pinter paper isn’t
See 2,3,5,6,7
automatically thermal paper;
10. The green printer release
Page 12 of 67 lever is not seated.
ACON Laboratories, Inc.

Yes Yes
External Printer Printer is not set as Set printer “External” in
abnormal “External”. menu.
No
Yes
The External Printer is Use proper printer.
not the proper type.

No
No 1. Reseat the cable
The 50-pin flat cable Yes
connecting the main connectors
No 2. Replace the cable. Refer
board to Printer board is
loose or damaged. to section 3.2.2.

No
Yes
Replace the Data Transfer
RS232 board is damaged.
Cable Unit. Refer to section
No

Replace the main board.


Refer to section 3.1.4.

The cable connecting the 1. Reseat the cable


Barcode Reader Yes Yes
main board to USB board connectors
abnormal is loose or damaged. 2. Replace the cable.

No
Yes
Replace the Data Transfer
RS232 board is damaged.
Cable Unit. Refer to section
No 3.5.2;
Yes
Barcode Reader is damaged. Replace Barcode Reader.

No
No

Replace main board.


Refer to section 3.1.4.

The Strip bed and waste


Yes Yes Install the Strip bed and waste
Test data is incorrect box is not in proper
location. box as in the user’s manual.

No
Yes
The Strip bed and waste Clean the Strip bed and waste
box is not clean. box as in the user’s manual.

No

Page 13 of 67
ACON Laboratories, Inc.

Yes
The white calibration is Clean the white calibration as
not clean. in the user’s manual.

No
Yes
Strip type error Use proper strip.

No
Yes
High and low frequency Replace Detection Unit. Refer
No is abnornal. to section 3.1.3.

No
Yes Replace Detection Unit. Refer
The Optical AMP board is
damaged. to section 3.1.3.

No

Replace the main board.


Refer to section 3.1.4.

Displayed and
Yes Replace the main board.
stored data is
Refer to section 3.1.4.
abnormal

No

For other abnormal


cases please
contact technical
support.

Page 14 of 67
ACON Laboratories, Inc.

3. Check and Repair


Section Content Operation Figure
3.1 3.1.1 1. Pull out the fuse Figure 1.
Switch Replace the box at the back of
Unit Fuse analyzer, Figure 1.
2. Take out the
damaged fuse,
Figure 2.
3. Replace a new fuse
into the box.
4. Install the fuse box
into the analyzer.

Figure 2.

Figure 3.

Page 15 of 67
ACON Laboratories, Inc.

3.1.2 3.1.2.1 Open the Upper



Figure 1.
Open and Housing
Close the 1. Take the strip bed
Upper and waste box out
Housing according to Figure
1.
2. Remove 3 screws at
the back of
analyzer, Figure 2.

Figure 2.

Page 16 of 67
ACON Laboratories, Inc.

3. Remove 3 screws at
the bottom of Figure 3.
analyzer, Figure 3.
4. Open the Upper
Housing according
to the Figure 4.

Figure 4.

Page 17 of 67
ACON Laboratories, Inc.

3.1.2.2 Close the Upper Figure 1.


Housing
1. Close the Upper
Housing and
Bottom Housing,
make sure they are
in proper location,
especially the screw
location marked in
Figure 1 and Figure
2;
2. Tighten all 6
screws ; Figure 2.
3. Insert the strip bed
and waste box,
Figure 3.


Figure 3.

Page 18 of 67
ACON Laboratories, Inc.

3.1.3 1. Open the Upper Figure 1.


Replace the Housing, refer to
Detection Section3.1.2.1;
Unit 2. Unplug all the
connectors marked
in Figure 1 and
remove 2 screws
marked in Figure 1;
3. Remove 6 screws
marked in Figure 1
and take off the
Protection Board,
Figure 2;
4. Unplug all the Figure 2.
connectors marked
in Figure 1 and
remove 6 screws
marked in Figure 1;

Page 19 of 67
ACON Laboratories, Inc.

5. Take out the Figure 3.


Detection Unit
carefully, Figure 4;
especially note the
Serrated Lock
Washers marked in
Figure 3;
6. Replace the
Detection Unit,
Install all screws and
connectors Figure 1.

Figure 4.

Page 20 of 67
ACON Laboratories, Inc.

3.1.4 1. Open the Upper Figure 1.


Replace the Housing, refer to
Main Section3.1.2.1;
Board 2. Unplug all the
connectors marked
in Figure 1;
3. Remove 5 screws in
Figure 2;
4. Turn back the main
board and remove
the FFC, take out
the damaged main
board, Figure 3.
5. Replace the main Figure 2.
board, tighten all
screws, and plug in
all connectors as in
Figure 1. The FFC
should be fastened
in its original
position.

Figure 3.

Page 21 of 67
ACON Laboratories, Inc.

3.1.5 1. Open the Upper Figure 1.


Replace the Housing, refer to
Switching Section 3.1.2.1;
Power 2. Remove the 2
Supply screws marked in
Figure 1 and take off
the Protection
Board, Figure 2;
3. Take off the
Detection Unit, refer
to Section 3.1.3 and
remove the 3 screws
marked in Figure 2;
4. Take off the main Figure 2.
board, refer to
Section 3.1.4;

Page 22 of 67
ACON Laboratories, Inc.

5. Remove the 4 Figure 3.


screws marked in
Figure 3;
6. Turn the analyzer to
its back and remove
the 2 screws marked
in Figure 4;
7. Take off the
damaged Switching
Power Supply;
8. Place a new
Switching Power
Supply onto the base Figure 4.
and reinstall all
screws.

Page 23 of 67
ACON Laboratories, Inc.

3.1.6 1. Open the Upper Figure 1.


Replace the Housing, refer to
wire Section3.1.2.1;
harness 2. Unplug the
connecting connector and
Switching remove the 3
Power screws marked in
Supply to Figure 1;
Main 3. Replace the new
Board wire harness, plug
the connector and
fasten 3 screws,
espically attention
the color of wire
harness.

Page 24 of 67
ACON Laboratories, Inc.

3.1.7 1. Open the Upper Figure 1.


Replace the Housing, refer to
Power Section3.1.2.1;
Switch 2. Take off the
Unit Detection Unit,
refer to Section
3.1.3;
3. Take off the main
board, refer to
Section 3.1.6;
4. Loose the wire
harness by
removing 3 Figure 2.
screws marked in
Figure 1, and
removing 2
screws marked in
Figure 1;
5. Turn back the
Grounding Bar
and loose the nut
with the
hexagonal
wrench, Figure 2;
6. Turn back the Figure 3.
Grounding Bar
and remove the
screw marked in
Figure 3;

Page 25 of 67
ACON Laboratories, Inc.

7. Press the snap at Figure 4.


the top of Power
Switch Unit using
flathead
screwdriver and
push it out,Figure
4;
8. Push out both of
the two snaps at
the top of Power
Switch Unit,
Figure 5;
Figure 5.

Page 26 of 67
ACON Laboratories, Inc.

9. For there are also Figure 6.


two snaps at the
bottom of the
Power Switch
Unit. Pull out the
Power Switch by
using the Allen
Hex plus Multi
Angle Long Arm
Key like Figure 6
and Figure 7;
10. Remove the
Power Switch Figure 7.
Unit, Figure 8;
11. Replace a new
Power Switch.
Screwing all the
wires down to the
power supply
terminals.

Figure 8.

Page 27 of 67
ACON Laboratories, Inc.

3.2 3.2.1 1. Open the Upper Figure 1.


LCD Replace the Housing, refer to
FFC Section 3.1.2.1;
connecting 2. Remove 4 screws
LCD to marked in Figure 1
Printer and take off the
Board Board shield
LCD,Figure 2;

Figure 2.

Page 28 of 67
ACON Laboratories, Inc.

3. Unplug the both Figure 3.


connectors of
FFC,Figure 3, and
take off this
damaged FFC;
4. Plug all the
connectors and
fasten all the
screws, Figure 4.

Figure 4.

Page 29 of 67
ACON Laboratories, Inc.

3.2.2 1. Open the Upper Figure 1.


Replace the Housing, refer to
50-pin flat Section3.1.2.1;
cable from 2. Unplug the
Main connectors marked
Board to in Figure 1;
Printer 3. Replace the new
Board 50-pin flat cable
from Main Board to
Printer Board and
plug the connectors;

Page 30 of 67
ACON Laboratories, Inc.

3.2.3 1. Open the Upper Figure 1.


Repace the Housing, refer to
LCD Section3.1.2.1;
2. Remove the screws
and take off the
Board shield
LCD,refer to
section 3.2.1.2;
3. Unplug the
connector of FFC
marked in Figure 1;
4. Remove 4 screws
marked in Figure 2; Figure 2.
5. Take off the
damaged LCD and
replace a new one.

Page 31 of 67
ACON Laboratories, Inc.

3.2.4 1. Open the Upper Figure 1.


Replace the Housing, refer to
Printer Section 3.1.2.1;
Board 2. Unplug the 3
Printer board
connectors and
remove the 4
screws marked in
Figure 1;
3. Replace the Printer
board, plug all the
connectors and
tighten all 4 screws.

3.2.5 1. Open the Upper Figure 1.


Potentiome Housing, refer to
ter Section3.1.2.1;
adjustment 2. Adjust the
potentiometer on the
Main board using
ceramic screwdriver
slotted tip as shown
in Figure 1;
3. Observe the
brightness of the
LCD, adjusting until
the proper
brightness is
observed.
3.3 3.3.1 1. Open the Upper Figure 1.
Detection Reseat and Housing, refer to
Unit Replace the Section3.1.2.1;
Strip 2. Reseat the connector
Sensor marked in Figure 1;
Board 3. Unplug the
connector and
remove the 2 screws
marked in Figure 1;
4. Replace the Strip
Sensor Board,
tighten all 2 screws
and plug the
connector.

Page 32 of 67
ACON Laboratories, Inc.

3.3.2 1. Open the Upper Figure 1.


Clean the Housing, refer to
Strip
Section3.1.2.1;
Sensor
2. Take off the Strip
bed and waste box;
3. Clean the Strip
Sensor using cotton
swab dipping in
75% alcohol

solution, Figure 1;

4. After a while, the


cleaned area is dry.

Page 33 of 67
ACON Laboratories, Inc.

3.3.3 1. Turn off the Figure 1.


Clean the analyzer and open
Strip Carry up the LCD
Furcation
Housing;

2. Take out the strip


bed and waste box,
Figure 1;
3. Turn the LCD
Housing, push the
Detection Unit to
the benmost position
slowly (marked in Figure 2.
Figure 2), stretch
hand and seize the
corner of Strip
CarryFurcation,
Figure 2;
4. Take out the Strip
CarryFurcation from
the meter, Figure 3;

Figure 3.

Page 34 of 67
ACON Laboratories, Inc.

5. Wipe the dirt on the Figure 4.


dentate rack
(especially on the
right side of upper
dentate) with a clean
and dry cotton ball
or soft paper, Figure

4;

6. Take the strip bed,


Figure 5 and waste
box back and close
the LCD housing, Figure 5.
Figure 6.

Figure 6.

Page 35 of 67
ACON Laboratories, Inc.

3.3.4 1. Open the Upper Figure 1.


Clean the Housing, refer to
White Section 3.1.2.1;
Calibration 2. Take off the Strip
bed and waste box;
3. Clean the White
Calibration using
cotton swab dipping
in 75% alcohol
solution, Figure 1;
4. After a while, the
cleaned area is dry.

3.3.5 1. Open the Upper Figure 1.


Reseat the Housing, refer to
connector Section 3.1.2.1;
of Strip 2. Reseat the No. “2”
Catch connector of Strip
Motor catch motor,
marked in Figure 1;

3.3.6 1. Open the Upper Figure 1.


Reseat Housing, refer to
the Section 3.1.2.1;
connector 2. Reseat the No. “4”
of Strip connector of Strip
Transfer Transfer motor,
Motor marked in Figure 1;

Page 36 of 67
ACON Laboratories, Inc.

3.3.7 1. Open the Upper Figure 1.


Reseat Housing, refer to
the Section3.1.2.1;
connector 2. Reseat the No. “7”
of Strip connector of Strip
Scanning Transfer motor,
Motor marked in Figure 1;

3.3.8 1. Open the Upper Figure 1.


Clean the Housing, refer to

No. “ 1 ” Section 3.1.2.1;

Optical 2. Clean the No. “1”


Coupler optical coupler using
cotton swab dipping
in 75% alcohol

solution, Figure 1;

3. After a while, the


cleaned area is dry.

3.3.9 1. Open the Upper Figure 1.


Clean the Housing, refer to

No. “ 2 ” Section3.1.2.1;

Optical 2. Clean the No. “2”


Coupler optical coupler using
cotton swab dipping
in 75% alcohol

solution, Figure 1;

3. After a while, the


cleaned area is dry.

Page 37 of 67
ACON Laboratories, Inc.

3.3.10 1. Open the Upper Figure 1.


Clean the Housing, refer to

No. “ 3 ” Section3.1.2.1;

Optical 2. Clean the No. “3”


Coupler optical coupler using
cotton swab dipping
in 75% alcohol

solution, Figure 1;

3. After a while, the


cleaned area is dry.

3.3.11 1. Open the Upper Figure 1.


Replace the Housing, refer to
MXL
Section 3.1.2.1;
Synchro-
Strap 2. Take off the
Protection Board,
refer to Section
3.1.5.2, remove 2
screws marked in
Figure 1 and take
off the Optical PCB
Cover;
3. Press the motor unit Figure 2.
with your left hand,
you should poke the
MXL
Synchro-Strap
using your other
hand ,like Figure 2;

Page 38 of 67
ACON Laboratories, Inc.

4. Once the MXL Figure 3.


Synchro-Strap has
loose for one side ,
take off it from the
clamp, Figure 3;
5. Take off the MXL
Synchro-Strap from
the hole of
Slide-rail
Subsidiary Holder,
make sure the
Following Pully is
OK, Figure 4; Figure 4.
6. Replace the new
MXL
Synchro-Strap and
install it carefully.

3.3.12 1. Open the Upper Figure 1.


Replace the Housing, refer to
Synchro-St
Section3.1.2.1;
rap (Take
one side for 2. Take off the
example) Synchro-Strap
slowly like Figure
1;

Page 39 of 67
ACON Laboratories, Inc.

3. Replace a new Figure 2.


Synchro-Strap,
install one side first,
make sure the stick
rubber is in the
proper postion
using a straight
object like Figure 2;
4. Install the other
side of
Synchro-Strap like
Figure 3;
5. You can use the Figure 3.
same way to
replace the other
Synchro-Strap.

Figure 4.

Page 40 of 67
ACON Laboratories, Inc.

3.3.13 1. Turn off the Figure 1.


Replace the analyzer and open
Strip Carry up the LCD
Furcation
Housing;

2. Take out the strip


bed and waste box,
Figure 1;
3. Turn the LCD
Housing, push the
Detection Unit to
the benmost
position slowly Figure 2.
(marked in Figure
2), stretch hand and
seize the corner of
Strip
CarryFurcation,
Figure 2;

Page 41 of 67
ACON Laboratories, Inc.

4. Take out the Strip Figure 3.


CarryFurcation from
the meter, Figure 3;
5. Replace a new
arry Strip Furcation
and install it , make
sure the direction is
right; Figure 4;
6. Take the strip bed
and waste box back
and close the LCD
housing, Figure 5.
Figure 4.

Figure 5.

Page 42 of 67
ACON Laboratories, Inc.

3.3.14 1. Take out the Figure 1.


Replace the Detection Unit, refer
Strip to section 3.1.3.5;
Scanning 2. Take out the MXL
Motor Synchro-Strap, refer
to section 3.3.11.3;
3. Remove 4 screws
marked in Figure 3
and take out the
Strip Scanning
Motor;

Figure 2.

Figure 3.

Page 43 of 67
ACON Laboratories, Inc.

4. Replace a new Figure 4.


Strip Scanning
Motor, Figure 4;
5. Install the Strip
Scanning Motor,
make sure the
position of motor
cable is correct,
Figure 5;
6. Poke the Following
Pulley and tighten
all 4 screws , Figure
6; Figure 5.

Figure 6.

Page 44 of 67
ACON Laboratories, Inc.

3.3.15 1. Loose the MXL Figure 1.


Replace the Synchro-Strap,
Following refer to section
Pully 3.3.11.4;
2. Replace a new
Following Pully
and poke it into the
MXL
Synchro-Strap,
Figure 1;
3. Install the MXL
Synchro-Strap
carefully;
3.4 3.4.1 1. Open the Upper Figure 1.
Battery Replace the Housing, refer to
Battery
Section 3.1.2.1;

2. Press the button


marked using a
small screwdriver
in Figure 1 to
release the battery

from the holder;

3. The battery should


pop out of the Figure 2.

holder,Figure 2;

4. Replace the battery


and press it tightly
into the holder until
it is retained by the
clip shown.

Page 45 of 67
ACON Laboratories, Inc.

3.5 3.5.1 1. Take out the Figure 1.


Data Reseat the Detection Unit, refer
Transfer FFC to section 3.1.3.5;
Cable Unit connecting 2. Tear the twin
RS232 to adhesive carefully,
Main Figure 1 and Figure
Board 2;
3. Replace a new FFC;

Figure 2.

Figure 3.

Page 46 of 67
ACON Laboratories, Inc.

4. Adhibit one twin Figure 4.


adhesive to the
Slide-rail Holder
carefully, Figure 4;
5. Adhibit the
Slide-rail Holder to
the FFC carefully,
Figure 5;
6. Adhibit another twin
adhesive above the
Slide-rail Holder
carefully, Figure 6;
Figure 5.

Figure 6.

Page 47 of 67
ACON Laboratories, Inc.

7. The FFC is installed Figure 7.


successfully.

3.6 3.5.2 1. Open the Upper Figure 1.


Printer Replace the Housing, refer to
Data Section 3.1.2.1;
Transfer 2. Take off the
Cable Unit Detection Unit, refer
to Section 3.1.3;
3. Take off the main
board, refer to
Section 3.1.4;
4. Remove the 4
screws (specially
used for RS232 Figure 2.
Port) using two Hex
Sockets, Figure1;
5. Replace the Data
Transfer Cable Unit
and tighten all the
screws, Figure2.

Page 48 of 67
ACON Laboratories, Inc.

3.6.1 1. Open the Upper Figure 1.


Reseat the Housing, refer to
connectors Section 3.1.2.1;
of printer 2. Reseat the
connectors marked
in Figure 1.

Page 49 of 67
ACON Laboratories, Inc.

3.6.2 1. Open the Upper Figure 1.


Replace the Housing, refer to
Printer Section 3.1.2.1;
2. Take off the Printer
Board, refer to
Section 3.2.4;
3. Remove all 4
screws marked in
Figure 1;
4. Take off the Printer
Box and remove the
2 screws marked in
Figure 2. Figure 2.
5. Take off the printer
from Printer Box,
Figure 3.
6. Replace a new
printer, plug all the
connectors and
tighten all the
screws.

Figure 3.

Page 50 of 67
ACON Laboratories, Inc.

3.6.3 1. Raise the green Figure 1.


Clean the lever as shown in
Printer Figure 1 and
Roller and remove the printer
Printer paper;
Head 2. Clean the printer
roller and printer
head using cotton
swab dipping in
75% alcohol
solution, Figure 2;
3. After the cleaned
area is dry, position Figure 2.
the green lever
down again, Figure
3.

Figure 3.

Page 51 of 67
ACON Laboratories, Inc.

4. Spare parts
Part No. Name Subassembly Figure

001 LCD LCD

002 Printer Printer

Printer
003 Printer board
board

Page 52 of 67
ACON Laboratories, Inc.

004 Main board Main board

Switching
005 power Switching power Supply
Supply

Detection
006 Detection Unit
Unit

Strip sensor
007 Strip sensor board
board

Page 53 of 67
ACON Laboratories, Inc.

Optical
008 Optical couple
couple

Strip bed
090 and waste Strip bed and waste box
box

Data
010 Transfer Data Transfer Cable Unit
Cable Unit

011 Switch Unit Switch Unit

Page 54 of 67
ACON Laboratories, Inc.

Wire
harness
connecting Wire harness connecting
012 Switching Switching Power Supply
Power to Main Board
Supply to
Main Board

013 Battery Battery

50-pin flat
cable from 50-pin flat cable from
014 Main Board Main Board to Printer
to Printer Board
Board

Cable
connecting Cable connecting RS232 to
015
RS232 to Main Board
Main Board

Page 55 of 67
ACON Laboratories, Inc.

Cable
connecting
Cable connecting LCD to
016 LCD to
Printer Board
Printer
Board

Following
017 Following Pully
Pully

MXL
MXL
018 Synchro-Str
Synchro-Strap
ap

Synchro-Str
019 Synchro-Strap
ap

Page 56 of 67
ACON Laboratories, Inc.

Strip Carry
020 Strip Carry Furcation
Furcation

Strip
021 Scanning Strip Scanning Motor
Motor

Page 57 of 67
ACON Laboratories, Inc.

5. Error messages, possible reasons and trouble shooting methods

Error message Reason Trouble shooting


Fuse is broken Replace the fuse. Refer to section3.1.1.
The input voltage of Switching Power Supply Measure the input voltage of Switching
is abnormal Power Supply.
The output voltage of Switching Power Supply Replace the Switching Power Supply.
No response from
is abnormal. Refer to section 3.1.5.
Analyzer
The wire harness connecting Switching
The input voltage of main board is abnormal Power Supply to main board is loose or
damaged. Refer to section 3.1.6.
Main board is damaged. Replace main board Section3.1.4.
Analyzer keeps
Main board is damaged. Replace main board Section3.1.4.
beeping
1. Reseat the FFC connectors ;
The FFC connecting LCD to Printer board is
2. Replace the FFC. Refer to section
loose or damaged.
3.2.1.
1. Reseat the cable connectors ;
The 50-pin flat cable from main board to Printer
No LCD display 2. Replace the 50-pin flat cable Refer to
board is loose or damaged.
section 3.2.2.
LCD is damaged. Replace the LCD. Refer to section 3.2.3.
Replace the Printer board. Refer to
Printer board is damaged
section 3.2.4.
Main board is damaged. Replace main board. Section3.1.4.
The brightness of the characters on LCD is too Potentiometer adjustment on Main board
dark or contrast is low. is not proper. Refer to section 3.2.5.

1. Reseat the FFC connectors;


2.Replace the FFC Refer to Section3.2.1.
LCD display is 1. Reseat the cable connectors ;
abnormal. 2. Replace the 50-pin flat cable.
Other abnormal appearance
Refer to Section3.2.2.
Replace the LCD. Section3.2.3.
Replace the Printer board Section3.2.4.

Replace the main board. Section3.1.4.

1. Reseat the Strip sensor board


The strip sensor can’t connector. Detection Unit. Refer to
sense the strip properly The sensor can’t respond the strip at all. section 3.3.1
2. Replace the Detection Unit. Refer to
section 3.1.3

Page 58 of 67
ACON Laboratories, Inc.

1. Clean the Strip sensor board. Refer to


section 3.3.2
The sensor can’t sense the strip in partly area.
2. Replace the strip sensor board. Refer
to section 3.3.1

Main board is damaged. Replace main board Section3.1.4;

1. Reseat the motor connector. Refer to


section 3.3.5;
The strip catch motor can’t move at all.
2. Replace the Detection Unit. Refer to
section 3.1.3.

1. Clean the Strip Carry Furcation.


Refer to section 3.3.3
The strip catch motor can’t move smoothly.
2. Replace Detection unit. Refer to
section 3.1.3
The strip catch motor
can’t move properly. 1. Clean the optical coupler Refer to
3.3.8
The strip catch motor can’t drive the belt to 2. Replace Detection unit. Refer to
desired position. section 3.1.3
3. Replace the Synchro-Strap; refer to
section 3.3.12

Main board is damaged. Replace main board Section3.1.4

1. Reseat the motor connector. Refer to


section 3.3.6;
The strip transfer motor can’t move at all..
2. Replace the Detection Unit. Refer to
section 3.1.3.
Replace the Detection Unit. Refer to
The strip transfer motor can’t move smoothly.
The strip transfer section 3.1.3;
motor can’t move 1. Clean the Strip Carry Furcation.
properly Refer to section 3.3.3
The strip transfer motor can’t drive the belt to 2. Replace the Strip Carry Furcation,
desired position. refer to section 3.3.13
3. Replace Detection unit. Refer to
section 3.1.3
Main board is damaged. Replace main board Section3.1.4
1. Reseat the motor connector. Refer to
section 3.3.7
The strip scanning
2. Replace Detection unit, Refer to
motor does not move
The strip scanning motor does not move at all.. section 3.1.3,
properly
3. Replace the strip scanning
motor,refer to section 3.3.14;

Page 59 of 67
ACON Laboratories, Inc.

1. Replace the MXL Synchro-Strap,


refer to section 3.3.11
The strip scanning motor does not move 2. Replace the Following Pully, refer to
smoothly. section 3.3.15
3. Replace Detection unit, Refer to
section 3.1.3;
Replace Detection Unit, Refer to section
The test result behave obviously error .
3.1.3.

Main board is damaged. Replace main board.Section 3.1.4.

Replace the main board. Refer to section


Main control unit failed.
3.1.4..
1. White calibration is not clean, Refer
to Section 3.3.4;
2. The Strip bed and waste box
Optical sensor failed. is positioned incorrectly on Detection
unit.
3. Replace the Detection Unit. Refer to
Self testing failed section 3.1.3
1. Replace the Detection Unit. Refer to
section 3.1.3
Detection Unit failed.
2. Replace main board. Refer to section
3.1.4
1. Replace the Detection Unit. Refer to
section 3.1.3
“Excessive light” failed.
2. Replace the main board. Refer to
section 3.1.4
1. Reseat the FFC connectors;
The FFC connecting LCD to Printer board is
2. Replace the FFC. Refer to section
loose or damaged.
3.2.1;
1. Reseat the cable connectors;
The 50-pin flat cable connecting the main board
2. Replace the cable. Refer to section
to Printer board is loose or damaged.
Touch screen abnormal 3.2.2;

LCD is damaged. Replace the LCD. Refer to section 3.2.3;

Replace the Printer board.


Printer board is damaged.
Refer to section 3.2.4;
Menu display failure Main board is damaged. Replace main board Section3.1.4

Page 60 of 67
ACON Laboratories, Inc.

Battery is out of power. Replace the battery. Refer to section 3.4.1

1. Install the battery again.


Bad contact between battery and battery
2. Replace the main board. Refer to
Wrong configuration contacts.
section 3.1.4
of Date/Time

Main board is damaged. Replace main board Section3.1.4

Wrong external Data Transfer cable. Use correct cable.

Setup the parameters as defined in the


Wrong communication parameters.
user’s manual.
1. Reseat the cable connectors;
FFC connecting RS232 to main board is loose
2. Replace the FFC. Refer to section
Data transfer incorrect or damaged.
3.5.1.
or failed Replace the Data Transfer Cable Unit.
The Data Transfer Cable Unit is damaged.
Refer to section 3.5.2..

Main board is damaged. Replace main board Section3.1.4.

1. The printer is not set as “Internal”;


2. The cable connecting printer to
Printer board is loose or damaged;
Refer to section 3.6.1;
3. The 50-pin flat cable connecting
main board to Printer board is loose
Printer has no response. or damaged; Refer to section 3.2.2;
4. Printer is damaged. Refer to section
3.6.2;
5. Printer board is damaged. Refer to
Internal printer section 3.2.4;
abnormal 6. Main board is damaged. Refer to
section 3.1.4.
1. The cable connecting printer to
Printer board is loose or damaged;
Refer to section 3.6.1;
2. The printer roller and priter head is
Printer result is illegible contaminated; Refer to section 3.6.3;
3. Printer is damaged. Refer to section
3.6.2;
4. The green printer release lever is not
seated.

Page 61 of 67
ACON Laboratories, Inc.

1. The cable connecting printer to


Printer board is loose or damaged;
Refer to section 3.6.1;
2. The 50-pin flat cable connecting main
board to Printer board is loose or
damaged; Refer to section 3.2.2;
Printer does not feed paper.
3. Printer is damaged. Refer to section
3.6.2;
4. Printer board is damaged. Refer to
section 3.2.4;
5. Main board is damaged. Refer to
section 3.1.4.
1. The cable connecting printer to
Printer board is loose or damaged;
Refer to section 3.6.1;
2. The 50-pin flat cable connecting
main board to Printer board is loose
or damaged; Refer to section 3.2.2;
Printer feeds paper, but the printout is blank. 3. Main board is damaged. Refer to
section 3.1.4;
4. The printer paper isn’t placed
properly;
5. The pinter paper isn’t thermal paper;
6. The green printer release lever is not
seated.
1. The cable connecting printer to
Printer board is loose or damaged;
Refer to section 3.6.1;
Printout is incomplete 2. The printer roller and priter head is
contaminated; Refer to section 3.6.3;
3. Printer is damaged. Refer to section
3.6.2.
1. The cable connecting printer to
Printer board is loose or damaged;
Refer to section 3.6.1;
Characters in printedout are narrow. 2. Printer is damaged. Refer to section
3.6.2;
3. Printer board is damaged. Refer to
section 3.2.4.

Page 62 of 67
ACON Laboratories, Inc.

1. The cable connecting printer to


Printer board is loose or damaged;
Refer to section 3.6.1;
2. The 50-pin flat cable connecting
main board to Printer board is loose
or damaged; Refer to section 3.2.2;
Does not feed paper automatically
3. Printer is damaged. Refer to section
3.6.2;
4. Printer board is damaged. Refer to
section 3.2.4;
5. Main board is damaged. Refer to
section 3.1.4.
Printer is not set as “External”. Set printer “External” in menu.
The external printer is not the proper type. Use proper printer.
1. Reseat the cable connectors
The 50-pin flat cable connecting the main board
External printer 2. Replace the cable. Refer to section
to Printer board is loose or damaged.
abnormal. 3.2.2..
Replace the Data Transfer Cable Unit.
RS232 board is damaged.
Refer to section 3.5.2;

Main board is damaged. Replace main board Section3.1.4

The cable connecting the main board to USB 1. Reseat the cable connectors
board is loose or damaged. 2. Replace the cable.
Replace the Data Transfer Cable Unit.
Barcode Reader RS232 board is damaged.
Refer to section 3.5.2;
abnormal
Barcode Reader is damaged. Replace Barcode Reader.

Main board is damaged. Replace main board Section3.1.4

Install the Strip bed and waste box as in


The Strip bed and waste box is not in proper
the user’s manual.
location.
Clean the Strip bed and waste box as in
The Strip bed and waste box is not clean.
the user’s manual.
Clean the white calibration as in the
The white calibration is not clean.
user’s manual.

Test data is incorrect Strip type error Use proper strip.

Replace Detection Unit. Refer to section


High and low frequency is abnornal.
3.1.3.
Replace Detection Unit. Refer to section
The Optical AMP board is damaged.
3.1.3.

Main board is damaged. Replace main board Section3.1.4

Displayed and stored


Main board is damaged. Replace main board Section3.1.4
data is abnormal

Page 63 of 67
ACON Laboratories, Inc.

6. Cleaning and Maintainance


6.1 Daily Cleaning
Cleaning the strip bed and waste box with water after every use.
Daily Cleaning
Procedure Figure
Turn off the analyzer and take out the strip bed
and waste box.

Remove the waste strips in box, and clean the


strip bed and waste box with water

Dry the strip bed and waste box with clean and
dry cotton cloth or paper

Insert the strip bed and waste box to the


analyzer.

6.2 Periodic Cleaning


The dentate rack under strip bed will get dirty after usual testing, which will affect the testing results. So
the dentate rack should be cleaned after one month using or 5000 testing. Refer to section 3.3.3

Page 64 of 67
ACON Laboratories, Inc.

7. Language Chip Replacement Package Insert

Language Chip Replacement Package Insert


Procedure Figure
1. Open the upper Housing, refer to Section
3.1.2.1;
2. Remove the waste tray from the analyzer;

Locate the PLCC44 Chip on the main board.

Remove the original PLCC44 Chip from the holding


socket with the PLCC extractor.

EMBED PBrush
1. Insert the new PLCC44 Chip in the same
direction as the original PLCC44 Chip;
2. Press the PLCC44 Chip into the holding socket
with fingers.

Locate the PLCC32 Chip on the main board.

Remove the original PLCC32 Chip from the holding


socket with the PLCC extractor.

1. Insert the new PLCC32 Chip in the same


direction as the original PLCC32 Chip;
2. Press the PLCC32 Chip into the holding socket
with fingers.

Page 65 of 67
ACON Laboratories, Inc.

Close the Upper Housing,refer to Section 3.1.2.2 .


Insert the waste tray.

Plug the power cord into the analyzer and turn the
Power Switch on.

Press to display the Main Menu.

Press from the Main Menu to view the newly


installed languages on the Language screen. Choose
proper language.

Page 66 of 67
ACON Laboratories, Inc.

8. Detection Optimization
 Detection Optimization will ensure that the analyzer-read results will be consistent with the
visual-read results for one or more of the following analyzes: Leucokyte, Nitrite, Urobilinogen,
Protein, Blood, Ketone, Bilirubin, Glucose and Ascorbic Acid. Specific Gravity and pH detection is
determined by qualitative values. Contact technical support for more details on optimizing the
detection of the U500 Urine Analyzer.

Page 67 of 67