ADE M118600 weight calibration
To enter the calibration menu press the TARE button followed by the ON/ZERO
button and keep the buttons pressed until C 5 will be displayed. A test
sequence will be shown. When stable press the TARE button again. Place the
requested test weights as shown on the display on the scale. 5000g follow by
10000g and 15000g. The scale will show P1 with indicates the end of the
calibration.
To enter the calibration menu press the HOLD button followed by the ON/ZERO
button and keep the buttons pressed until C H will be displayed. A test
sequence will be shown. When stable place the requested test measures
between the foot and head stop.
1
ADE M118600 / M118600-01 drawings
2
Spare parts list ADE M118600 & M118600-01
Overlay: MZ118600-004
3
Cleaning & maintenance
Cleaning
For cleaning just use a damp cloth or common disinfection products. Do not use any abrasive or acid
cleaning fluids. Ensure that no cleaning fluid or water gets into the scale and always follow the
instruction of the manufacturer.
4
Safety instructions
Make sure to use only mains adaptors supplies by ADE. All other equipment might
cause electromagnetic interferences to other equipments.
Waste disposal
Keep all original parts and pieces for an eventual sending back of the scale in order to avoid any
possible damage during transport which automatically leads to no warranty.
Disconnect all cables before transport to avoid unnecessary damages.
5
CE - conformity
Technical data
Capacity: 20 kg
Graduation: 5g
LC Display: size 90 x 25mm
Letter size 25mm
Weighing unit selectable: kg, lb
Battery operation 4pcs 1.5V AA alkaline batteries
Mains adaptor AC/DC adaptor with 6V 1A output