Stephen A. Ifft
McCrometer Inc.
Hemet, California, USA
Abstract
This paper will discuss the approval of flow around a cone positioned in the
the McCrometer V-Cone flowmeter for center of the pipe.
custody transfer measurement in
Canada. In May 1998, Measurement
Canada issued AG-0428, the Notice of
Approval for the V-Cone meter. This
notice states the models and methods
approved for use in Canada. Of
particular note is the need for approved
V-Cone calculation devices.
Introduction
The V-Cone is a differential pressure
flowmeter invented and manufactured by
McCrometer Inc. Patented in 1986 as a Fig. 1 Illustration of a typical V-Cone
new type of differential pressure flow design
meter, the V-Cone is based on the
principles of Bernoulli. Equations for the V-Cone are only
slightly different from standard
The geometry of the V-Cone is a differential pressure equations. The
radically different approach to V-Cone beta ratio follows the same
differential pressure flow metering, see principle as other differential pressure
Figure 1. As with other differential devices. Thus a V-Cone and an orifice
pressure devices, the flow constricts to plate beta ratio are equivalent to each
create a high velocity area, which creates other in terms of open area. The basic
a lower pressure just past the equation of flow for the V-Cone is
constriction. By measuring the pressure similar to standard differential pressure
upstream and downstream of the cone, a equations.
differential pressure is known and can be
related to the flowrate through the
pipeline. The V-Cone’s constriction, Approval Process
however, is not a concentric opening The approval process with the Canadian
through the center of the pipe like government is well defined in
traditional differential pressure documents such as the Specifications for
flowmeters. The V-Cone creates an Approval of Type of Gas Meters and
annular opening, forcing the fluid to Auxiliary Devices (LMB-EG-08). This
The V-Cone is a differential pressure The second reason was the issue of
type device and McCrometer’s intention calibration. Similar to turbine meters,
was to have the V-Cone certified under V-Cone meters need in-line calibrations
the same principles and requirements as for the best accuracy. Each custody
an orifice plate. transfer V-Cone will be flow calibrated
at a laboratory directly traceable to
After receiving the application for type Canadian or American national
approval for the V-Cone, Measurement measurement standards. These
Canada reviewed the product data and laboratories are typically independent
existing performance data regarding the companies whose expertise is in
V-Cone. Even though the V-Cone is compressible gas flow and calibration.
based on the same principles as the
orifice place, the V-Cone could not be
evaluated under the same requirements Performance Test
as the orifice meter’s section. A performance test is a necessary part of
the evaluation process. Measurement
Section 8 of the specifications refers Canada required a witnessed test of the
extensively to the American National performance of the meter and instrument
Standard, ANSI/API 2530, “Orifice system. This test would support the
Metering of Natural Gas and Other documented data already produced
Related Hydrocarbon Fluids.” This during the evaluation process.
standard deals exclusively with the
orifice plate design and could not be A typical turbine meter calibration
correlated to a V-Cone application. would take place in Measurement
Measurement Canada then needed to Canada’s gas flow facility. This system
decide how to deal with the V-Cone and operates at atmospheric pressure and
under what section the evaluation should was not suited for the higher-pressure
proceed. performance test of the V-Cone. With
consent from both Measurement Canada
For two apparent reasons, the V-Cone and McCrometer, the Colorado
application would be evaluated under Engineering Experiment Station, Inc.
Section 7 of the specifications regarding (CEESI) was selected as the test
turbine meters. The first reason was the laboratory for the performance test.
performance specifications of the CEESI is well acquainted with the
V-Cone. The standard accuracy and V-Cone and well equipped for the type
turndown of the V-Cone is stated as of calibration required.
1.50
1.00
0.50
% difference
0.00
-0.50
-1.00
-1.50
-2.00
0 200000 400000 600000 800000 1000000 1200000
Reynolds Number