Professional Documents
Culture Documents
Servo-Air - Service Manual
Servo-Air - Service Manual
Contents
1 Important
2 Introduction
3 Description of functions
4 Disassembling and assembling
5 Service procedures
6 Troubleshooting
7 Preventive maintenance
8 Index
9 Revision history
10 Diagrams
1 Important
Responsibility for the safe functioning of the equipment
1.1 General reverts to the owner or user in all cases in which
service or repair has been done by a non-professional
The information in this Service Manual is based on or by persons who are not employed by or authorized
the SERVO-air Ventilator System of System by Getinge, and when the equipment is used for other
version 4.0. Unless otherwise stated, the Service than its intended purpose.
Manual is valid also for System versions below 4.0.
The system complies with standards and requirements
Maquet Critical Care AB, part of Getinge, is the legal as stated in the User's Manual.
manufacturer of the SERVO ventilator systems.
WARNING! Do not modify this
Service documentation for the SERVO-air Ventilator
equipment without authorization of the
System consists of:
manufacturer.
• User's Manual including Cleaning and Maintenance
bullet
manual. The User's Manual is an indispensable
complement to the Service Manual for proper System version and System software version
servicing.
The System version number can be found in the
• Service Manual
bullet SYSTEM STATUS window on the User interface.
• Installation Instructions
bullet Make sure that the version of the User’s Manual
corresponds to the System version.
• Spare parts information
bullet
• Documentation for optional equipment included in
bullet The System SW version number can be found in the
the system SYSTEM STATUS window on the User interface. For
• Product End-of-Life - Disassembly Instructions
bullet functionality enhancement, the latest released System
(Disassembly and Disposal Instructions). SW version is always recommended.
Maintenance of the system must be performed by 1.8.2 Components with special environmental
Getinge authorized personnel with the following concern
intervals:
Components listed below shall be disposed of in
• Preventive maintenance at least once a year, or
bullet accordance with appropriate industrial and
every 5000 hours of operation, whichever comes environmental standards.
first.
Printed circuit boards
• The Battery modules shall be replaced after
bullet
300 charge cycles or four years (48 months) from
manufacturing date, whichever comes first. • PC 1781 Pressure transducer, 2 pcs
bullet
Only original spare parts from Maquet must be used • PC 2004/PC 2024 Expiratory channel
bullet
in the system. • PC 2020 Main backplane
bullet
• PC 2021 Rear connectors
bullet
1.7 To the responsible service personnel
• PC 2023 Panel
bullet
• Fans, 2 pcs
bullet Others – very small amounts
• Loudspeaker
bullet
• Cable harness (handle as electronic waste)
bullet • Filter paper of fiberglass
bullet
• ABS (Acrylicnitrilebutadienstyrene)
bullet 9. Memory backup batteries
10. Expiratory cassette
• EPDM (Ethylenepropylenedienemonomer)
bullet
11. Expiratory cassette membrane
• FPM (Fluororubber)
bullet
12. O2 cell including filter
• LDPE (Low-density polyethylene)
bullet
Item 1: Consumption approx. 250 pcs/year.
• NBR (Nitrilerubber)
bullet
• PA (Polyamide)
bullet Item 2: Consumption approx. 15 pcs/year.
• PE (Polyethylene)
bullet Item 7: Replaced after 300 charge cycles or four years
• PEEK (Polyether ether ketone)
bullet (48 months) from manufacturing date, whichever
comes first.
• POM (Polyoxymethylene)
bullet
• PP (Polypropylene)
bullet Item 8: Replaced every 10 years.
• PTFE (Polytetrafluoroethylene)
bullet Item 9: Replaced every five years.
• PS (Polystyrene)
bullet
Items 10 – 12: Replaced when needed.
• PUR (Polyurethane)
bullet
• PVC (Polyvinyl chloride)
bullet Articles related to clinical applications, e.g., patient
tubing and nebulizer medication cups are not included
• SI (Silicone)
bullet
in the list above.
SERVO-air Ventilator System, Service Manual 1-5
| 1 | Important |
The power consumption depends on the operating For disassembly and disposal (scrapping) information,
mode and Patient unit settings. refer to the document SERVO-air Ventilator System
Product End-of-Life Disassembly Instructions.
• Typical minimum power consumption (standby):
bullet
26 W
• Loading pallet.
bullet
• Corrugated cardboard.
bullet
• Shock-absorbing material of expanded
bullet
polypropylene, EPP.
2 Introduction
Table of Contents
2.1 Main units | 2-2
2.2 Patient unit | 2-4
2.3 User interface | 2 -10
2.4 Software structure | 2 -12
2.5 Internal communication | 2 -14
f h
a g k l j
S20002
• Alarm output
bullet
• Battery module (extra)
bullet
• Battery charger
bullet
• Mobile cart
bullet
• Mobile cart drawer kit
bullet
• Cabinet (for Mobile cart)
bullet
• Gas cylinder restrainer kit
bullet
• Shelf base
bullet
• Support Arm 179
bullet
• Y piece holder and hook
bullet
• Cable holder
bullet
• Aerogen nebulizers (Aerogen Pro and
bullet
Aerogen Solo)
s
F G
d D K
f S H
h j k l ; A J
S20003
The main section is accessed by removing the top cover and the lid on the PC board rack.
s
g
d
f
h
j
S20004
S20005
The left-hand side of the Patient unit contains the Inspiratory section.
d f
s
g
l
S20006
The right-hand side of the Patient unit contains the AC/DC converter and Mains power inlet, as well as the
Expiratory valve coil.
a s
d
g
S20007
a s
S20008
S20009
The illustration below shows the backside of the User The illustration below shows the inside of the User
interface after it is detached from the User interface interface rear cover.
holder.
a s d S20011
a s
S20012
The System software is divided into the following main • Controls the expiratory cassette handling including
bullet
subsystems: flow measurement performed by the flowmeter in
the cassette.
• Breathing BRE
bullet • Nebulizer control. Controls the Nebulizer function.
bullet
• Monitoring MON
bullet • Controls Pre-use check operations on the Breathing
bullet
• Panel PAN
bullet subsystem
• Power POW
bullet • Maintain a persistent data storage area used to
bullet
save parameter settings.
• Blower BLO
bullet
• System ID
bullet
• Controls some of the system settings, e.g. storing
bullet
default patient category and default mode of
ventilation. The system settings will be stored in
persistent memory.
User interface The Breathing software is stored on PC 1991 Control
PC 2023 and PC 2004/PC 2024 Expiratory Channel. The
software must be re-installed if PC 1991 or
PAN SW
PC 2004/PC 2024 is replaced.
Patient unit
PC 1786
PC 2029
PC 1781
MON SW
BRE SW
Insp.
PC 2031/
PC 1785
PC 1993
PC 1991
PC 1992
PC 2024
BLO SW
PC 2026
PC 2027
PC 2033 (Turbine
module)
PC 2021 PC 2025
S20135
The Monitoring subsystem’s main function is alarm The Power subsystem has the following main
and monitoring, including trends of measured values: responsibilities:
When installing software options, the System ID is • ETHERNET bus using TCP/IP running at
bullet
used to verify that the option is valid for that system. 100 Mbit/s. ETHERNET connects the main
subsystems BRE, MON and PAN using the Ethernet
When replacing PC 2020 Main back-plane, the
switch.
replacement (spare part) must be factory programmed
with the correct System ID to ensure proper ventilation • CAN bus (Controller Area Network) running at
bullet
performance, compatibility with future updates and 500 kbit/s. The CAN is primarily used for the
correct options configuration. Expiratory flowmeter.
3 Description of functions
Table of Contents
3.1 O2 gas module (1) | 3-4
3.2 Turbine module including PC 2027 Motor | 3-6
control (2)
3.3 Inspiratory gas docking (3) | 3-7
3.4 Inspiratory pipe (4) | 3-7
3.5 Inspiratory flowmeter (5) | 3-7
3.6 Inspiratory channel (6) | 3-7
3.7 O2 cell (7) | 3-8
3.8 Safety valve including pull magnet (8) | 3-8
3.9 O2 evacuation fan (9) | 3-9
3.10 PC 2030 Inspiratory channel (10) | 3-9
3.11 PC 1781 Inspiratory pressure transducer (11) | 3 -10
3.12 PC 2029 Inspiratory channel extension (12) | 3 -10
3.13 Expiratory cassette (13) | 3 -10
3.14 Expiratory valve (14) | 3 -11
3.15 Expiratory valve coil (15) | 3 -11
3.16 Expiratory outlet (16) | 3 -11
3.17 PC 1786 Expiratory channel cassette (17) | 3 -11
3.18 PC 2031/PC 1785 Expiratory channel | 3 -11
connector (18)
3.19 PC 2020 Main back-plane (19) | 3 -12
3.20 PC 1781 Expiratory pressure transducer (20) | 3 -12
3.21 PC 2004/PC 2024 Expiratory channel (21) | 3 -12
3.22 PC 1991 Control (22) | 3 -13
3.23 PC 1992 Monitoring (23) | 3 -13
3.24 PC 1993 Ethernet switch (24) | 3 -14
3.25 PC 1994 Nebulizer (25) | 3 -14
3.26 Mains inlet (26) | 3 -14
3.27 AC/DC Converter (27) | 3 -14
3.28 PC 2028 DC/DC & Battery control (28) | 3 -14
3.29 On/Off switch (29) | 3 -15
3.30 PC 2021 Rear connectors (30) | 3 -15
3.31 PC 2026 Battery extension (31) | 3 -16
3.32 PC 2025 Battery back-plane (32) | 3 -16
3.33 Battery module (33) | 3 -16
3.34 Main fan (34) | 3 -17
About this chapter The System electronics contains all PC boards and
electrical components that are not included in the other
The information in this chapter refers to the System main blocks:
overview diagram in chapter Diagrams. Note the
numbering in each heading (in brackets), that 19. PC 2020 Main back-plane
corresponds to the position number in the System 20. PC 1781 Expiratory pressure transducer
overview diagram.
21. PC 2004/PC 2024 Expiratory channel
22. PC 1991 Control
Main blocks
23. PC 1992 Monitoring
The information in this chapter is divided into main 24. PC 1993 Ethernet switch
blocks:
25. PC 1994 Nebulizer
The Inspiratory section conveys the breathing gas 26. Mains inlet
from the gas inlets to the patient breathing system. It 27. AC/DC Converter
contains the following main components: 28. PC 2028 DC/DC & Battery control
1. O2 gas module 29. On/Off switch
2. Turbine module 30. PC 2021 Rear connectors
3. Inspiratory gas docking 31. PC 2026 Battery extension
4. Inspiratory pipe 32. PC 2025 Battery back-plane
5. Inspiratory flowmeter 33. Battery module
6. Inspiratory channel 34. Main fan
7. O2 cell 35. User interface cable
The Expiratory section conveys the breathing gas 38. PC 2023 Panel
from the patient breathing system to the Expiratory 39. Illuminated alarm indicator on top of the User
outlet. It contains the following main components: interface frame.
40. Loudspeaker
13. Expiratory cassette
14. Expiratory valve
15. Expiratory valve coil
16. Expiratory outlet
17. PC 1786 Expiratory channel cassette (integrated
into the Expiratory cassette)
18. PC 2031/PC 1785 Expiratory channel connector
3.2 Turbine module including PC 2027 Motor 7. Nipple for air inlet pressure measurement for
control (2) supervision of pressure drop over the Air inlet
filter.
3.2.1 Turbine
The Turbine module shall be replaced every 10 years.
The Turbine generates the inspiratory air gas flow
(max. 240 l/min). It generates the air flow and pressure Note: The Turbine is factory calibrated. The Turbine
from the surrounding air. This air flow is then mixed must not be disassembled.
together with the O2 flow from the O2 gas module.
The maximum Turbine output pressure is dependent 3.2.2 PC 2027 Motor control
on the ambient pressure and may therefore vary
PC 2027 Motor control’s main purpose is to control
between 65 cmH2O (high altitude) and
and supervise the Turbine:
approx. 100 cmH2O (sea level).
• Powering the Turbine. The power to the motor driver
bullet
s g is supplied through a step-up regulator that elevates
the voltage from the system voltage (approximately
11-17 V) to approximately 28 V. The energy created
when the motor is decelerated is stored in a
a capacitor buffer to enable quick acceleration.
Further excess energy will be delivered to other
parts of the system.
h
X T • Data transferred to BRE via SPI communication for
bullet
further dispatching to MON:
P1 PC 2027 Motor control P - Inspiratory flow and gas temperature from the
Inspiratory flowmeter.
d f h j - Ambient air temperature and humidity from
S20103 sensors on PC 2030.
- Air inlet filter pressure drop measured by a
The Turbine module includes:
differential pressure sensor. Will indicate if the
1. Brushless DC motor powered by a motor driver Air inlet filter is occluded or clogged. Will also
on PC 2027 Motor control. Maximum speed detect if the Air inlet filter is missing during
70 000 r/min. Pre-use check.
2. Turbine wheel. • Passing data from the ambient air inlet filter
bullet
differential pressure sensor.
3. Hall effect sensors to measure motor speed and
phase.
4. Temperature sensor to monitor the Turbine 3.2.3 Air inlet filter including dust filter
temperature. The reading is used to detect
The airflow to the Turbine, and to the patient, is
overheating and to log running conditions.
protected by an Air inlet filter. The Air inlet filter
5. Silencing foam and rubber suspensions to consists of a HEPA (High Efficiency Particulate Air)
minimize vibrations and sound level. filter and a dust filter.
6. Cooling air flow for the motor. Routed from the
Turbine unit out through the Patient unit via a The Air inlet filter must be replaced during the
hose. Preventive maintenance. Check the dust filter and
replace it once a month or more often if needed.
The Inspiratory pipe leads the gas from the Inspiratory The flowmeter is equipped with a flow conditioner on
gas docking to the Inspiratory channel. the inlet to create a laminar flow to enhance the
accuracy.
The Inspiratory pipe contains:
3.6 Inspiratory channel (6)
• Inspiratory flowmeter connector muff.
bullet
The Inspiratory channel connects the Inspiratory pipe
• Inspiratory flowmeter.
bullet
to the patient breathing system. The Inspiratory
• Inspiratory channel connector muff.
bullet channel contains:
The Inspiratory pipe is equipped with a flow conditioner
ahead of the flowmeter, to ensure that the flowmeter • The Inspiratory pressure tube that connects the
bullet
makes reliable measurements. Inspiratory channel with the Inspiratory pressure
transducer. The tube is connected to the bacteria
3.5 Inspiratory flowmeter (5) filter that protects the pressure transducer PC 1781
Inspiratory pressure transducer from contamination.
The Inspiratory flowmeter (inspiratory flow transducer) The bacteria filter must be replaced during
measures the combined air and O2 flow, as well as Preventive maintenance.
temperature, of the inspiratory gas from the O2 gas
module and Turbine module. The flow measurements
• The Safety valve membrane.
bullet
are used as input to control the gas module and the • Housing for the O2 cell.
bullet
Turbine, creating a feedback loop. • The Inspiratory outlet, a 22 mm/15 mm tube
bullet
connector for the Inspiratory tube of the patient
The flowmeter uses a thermal measuring principle, in breathing system.
this case a differential temperature measurement. The
The Inspiratory channel can easily be removed using
basic principle is that two temperature sensors, one
the quick coupling.
mounted before and one after a small heater in the
flow direction:
3.7 O2 cell (7) 3.8 Safety valve including pull magnet (8)
The O2 concentration is measured by an O2 cell. The The movable axis of the Safety valve pull magnet
O2 cell is mounted in a housing on the Inspiratory controls the opening and closing of the Safety valve
channel and is protected by a bacteria filter. membrane in the Inspiratory channel. The pull magnet
is controlled (electrically activated = closed) by
O2 PC 2004/PC 2024 Expiratory Channel.
O2 gas module
O2
Air
O2 cell O2 gas module
Turbine module
Air
Turbine module
S20015
• Turbine module
bullet
• O2 gas module
bullet
• O2 cell
bullet
s • O2 evacuation fan
bullet
• Inspiratory flowmeter
bullet
• Safety Valve
bullet
a s a
S20019
received by the right hand side transducer now acting 3.15 Expiratory valve coil (15)
as a receiver. The time from sending to receiving
ultrasonic sound in downstream expiratory gas flow The movable axis of the Expiratory valve coil controls
is measured. the opening of the Expiratory valve by pushing the
valve membrane into desired position. The power
Then the right hand side transducer (earlier receiving) supply to the coil is regulated so that the remaining
is sending out ultrasonic sound upstream the pressure in the patient system, towards the end of the
expiratory gas flow. The ultrasonic sound is received expiration time, is kept on the PEEP level according
by the left hand side transducer now acting as a to User interface setting.
receiver. The time from sending to receiving ultrasonic
sound in upstream expiratory gas flow is measured. 3.16 Expiratory outlet (16)
The time difference between the downstream and the The gas from the patient system leaves the system
upstream time measurements provides flow via this Expiratory outlet. A thermo sleeve can be
information. attached to the outlet to reduce condensation.
Membrane capacity (counter) can be shown in the For software and hardware requirements and
System status window. Select SYSTEM STATUS > compatibility, see the Compatibility chart in chapter
Expiratory cassette to check Membrane capacity. The Revision history.
operating capacity meter must be reset after
replacement of the membrane. Note that the reset
function is not implemented in System version 1.0.
3.19 PC 2020 Main back-plane (19) 3.21 PC 2004/PC 2024 Expiratory channel
(21)
PC 2020 is the interconnection board for PC boards
and components.
Note: PC 2024 replaces PC 2004. Introduced in
PC 2020 Main back-plane is also connected to production during Q1 2016.
PC 2030 Inspiratory channel via the PC 2029
Inspiratory channel extension for communication with PC 2004/PC 2024 Expiratory channel is part of
components in the Inspiratory section. subsystem Breathing BRE. The main functions are
expiratory flow measurement and expiratory cassette
The System ID (Serial No.), operating time, etc., is handling. Expiratory flow measurement is performed
stored in an EEPROM on PC 2020. Thus, when by the flowmeter in the cassette. PC 2004/PC 2024
replacing PC 2020, a spare part that is factory is connected to the cassette via PC 2031/PC 1785
programmed for the concerned system must be used. Expiratory channel connector.
As the Preventive maintenance time stamp will be lost PC 2004/PC 2024 Expiratory channel contains
when replacing PC 2020, a new time stamp must be electronics including microprocessor for handling of:
set. In order to make this new time stamp correct,
there are two alternatives: • Expiratory flow measurement performed by the
bullet
flowmeter inside the cassette. The output signal
• Perform Preventive maintenance on the system
bullet Exp. Flow is used in sub-systems Control and
and report completed Preventive maintenance in Monitoring.
Service & Settings.
• All electronic connections to and from the cassette.
bullet
or
• Measurement of expiratory pressure via pressure
bullet
the PC 1781 Expiratory pressure transducer.
• If correct value (hours) until next Preventive
bullet
PC 2004/PC 2024 contains holder and electrical
maintenance is known, enter this value in Service
& Settings. connector for the pressure transducer board.
• Nebulizer control.
bullet • Barometric pressure. PC 1992 contains a
bullet
barometric transducer and the measured barometric
A Memory backup battery on PC 1991 power supplies
pressure is supplied to the other sub-units in the
the internal memory on the PC board. If the battery
system.
on PC 1991 is disconnected or if the battery voltage
is too low, startup ventilation configurations made via • Real time clock, RTC.
bullet
the Service & Settings may be erased. The Memory • Temperature monitoring. A thermistor on PC 1992
bullet
backup batteries must be replaced after five years. monitors the temperature inside the Patient unit.
An alarm is activated if the temperature is 77 ±5 °C
Includes an ID PROM. The ID information can be read
(170 ±9 °F) or higher.
by the system.
• System voltage monitoring: The following voltages
bullet
For software and hardware requirements and are supervised:
compatibility, see the Compatibility chart in chapter
Revision history.
• +24 V
bullet
• +12 V
bullet
Note: The System software must be re-installed if • -12 V
bullet
PC 1991 is replaced. • +5 V
bullet
•+3.3 V
bullet
3.23 PC 1992 Monitoring (23) The buzzer on PC 1992 generates the alarm signal
in case of +5 V or +3.3 V power failures. The buzzer
PC 1992 Monitoring is part of subsystem Monitoring
and +5 V/+3.3 V failure logic is powered by backup
MON. The Monitoring subsystem features alarm and
capacitors in case of power failure.
monitoring functions, calculations, trending of
measured values and is also CAN-bus master.
• Maintains a persistent data storage area to save
bullet
logs. The logs are saved once per minute.
PC 1992 Monitoring comprises electronics including
microprocessor for handling of: • RS-232 communication. The system is equipped
bullet
with two RS-232 ports. The RS-232 communication
• Control that all alarms are displayed on the User
bullet is handled by PC 1992. The communication protocol
interface and that the alarm sound is generated. In SCI (SERVO Communication Interface) is described
case of malfunction in the loudspeakers in the User in the 'SERVO Communication Interface –
Reference Manual'.
PC 1994 Nebulizer is the integrated interface for • Switch to battery power supply when Mains
bullet
connection of the Aerogen® Nebulizer Systems, power/External 12 V DC supply is lost, and control
Aerogen Pro and Aerogen Solo. PC 1994 Nebulizer switching between Mains power and external 12 V
contains electronics required to control the nebulizer DC power supply.
units. • Control of Main fan through one thermistor on PC
bullet
2020, and one on PC 2028.
For electrical safety test of the Nebulizer connector,
refer to chapter Preventive maintenance section Safety • Control of O2 evacuation fan through a thermistor
bullet
inspection. on PC 2030.
The power modes in the system are: There are three power supply voltages to the system:
Note: At least one Battery module must always be The following connectors are located on PC 2021 Rear
installed to ensure proper operation. connectors:
1. Mains power
• N29 – RS-232 (1). Isolated RS-232 connector
bullet
(identical to N30). For SCI communication or
2. External +12 V DC supply (if connected) MSync.
3. Battery modules
SERVO-air Ventilator System, Service Manual 3 - 15
| 3 | Description of functions |
Typical recharge time approximately 2 h/battery The Fan will start with half effect at approximately
(90 %), up to 3 h (100 %) if battery is completely 33 °C (91 °F) and with full effect at approximately
discharged. If a battery is fully discharged, e.g. due to 43 °C (109 °F). When the temperature drops below
long storage time, each battery may require up to approximately 37 °C (99 °F), the Fan turns to half
6 hours charging time. effect and when the temperature drops below
approximately 27 °C (81 °F), the Fan stops.
Each battery includes an ID PROM. The ID information
can be read by the system. The air inlet is protected by a filter that must be
cleaned or replaced during the Preventive
CAUTION: Do not expose the batteries maintenance.
to water, fire or excessive heat. Do not
crush, disassemble, puncture or short 3.35 User interface cable (35)
circuit the connector terminals.
This User interface cable connects the Patient unit
and the User interface. The cable is hidden in the
panel arm behind a cover.
3.34 Main fan (34)
The Main fan forces cooling air through the AC/DC Note: The User interface cable must only be
section, and the Main section of the Patient unit. The connected or disconnected following the instructions
cooling air flow inside the Patient unit, is indicated in given in section Hazard notices in chapter
the System overview. Disassembling and assembling.
1. Main fan inlet It is estimated that the LCD Display will maintain an
2. Main fan outlet acceptable brightness level during the unit’s lifetime.
Using Service & Settings, a backlight time meter can
The Main fan inlet (1) is located at the bottom of the be shown.
side cover. The Main fan outlet (2) is located above
the connectors on the backside of the Patient unit. 3.38 PC 2023 Panel (38)
The Main fan is monitored by the temperature sensors The PC 2023 Panel contains drivers for:
on PC 2020, and PC 2028, and controlled by PC 2028.
The value from the temperature sensor, giving the • Touch and display
bullet
highest temperature reading, will be used.
• Loudspeaker
bullet
The Fan is always on when the batteries are charging, • Microphone
bullet
and for 2.5 hours after the charging is complete.
• Microcontroller
bullet
SERVO-air Ventilator System, Service Manual 3 - 17
| 3 | Description of functions |
The multicolor LEDs, visible in the Alarm indicator, Preferably the NC contact should be used, as this will
are located on PC 2023. detect any interruption in the alarm system or
unintentional disconnection of the cable between the
3.39 Alarm indicator (39) system and the ventilator. Furthermore, if both the
NO- and the NC-contacts are utilized, the highest
The Alarm indicator is located on the upper part of the safety level for unforeseen failures can be achieved.
User interface frame. The multicolor LEDs on PC 2023 Note that, in both cases, even an intentional
are visible in the Alarm indicator. The LEDs will flash disconnection of the ventilator will trigger the external
when an alarm is triggered in the system: alarm.
The Alarm output connector, N67 on PC 2021 Rear 3.41.4 Mobile cart
connectors, enables connection of an external alarm
The Mobile cart is designed for carrying the Patient
signal system to the system. The alarm output signal
unit, the User interface and all required optional
is active as long as the alarm audio is active on the
equipment.
system.
The Mobile cart has standard 10x25 mm rails to carry
For pin configuration and signal names in the Alarm
accessories such as Support arm 179 and Humidifier
output connector N67, see Connectors in chapter
holder. The column has vertical slots for installation
Diagrams. For further information, refer to the 'Alarm
of e.g. Gas cylinder restrainers. The optional Drawer
output connector – Reference Manual'.
kit or Cabinet can be mounted on the Mobile cart.
The Alarm output connector has two contact functions:
NO (Normally Open) and NC (Normally Closed). In 3.41.5 Gas cylinder restrainer kit
an alarm situation the open contact will close and the
closed one will open. The contacts are independent The Gas cylinder restrainers are mounted on each
of polarity and can be used both with AC and DC side of the Mobile cart. It is intended for two gas
systems. cylinders, maximum volume 4.5 litres each.
The Shelf base is a mounting device for the Patient MSync is intended for information transfer from the
unit. It allows positioning of the Patient unit on a shelf. Maquet devices to the hospitals information system
by:
3.41.7 Support arm 179
• Importing real-time clinical data from the device
bullet
Support arm 179 is equipped with a clamp intended using its respective communication protocol.
for standard 10x25 mm rails. The support arm has a
central locking mechanism that secures the selected
• Converting device data to HL7 compliant data.
bullet
position. A twin tube holder (15-27 mm diameter) and • Exporting device data in HL7 format to the
bullet
information system.
a filter/tube support (20/25 mm diameter) can be
mounted on the support arm. Maximum load with fully An MSync unit is connected to each Maquet device
extended arm is 1 kg. intended to be integrated with the information system.
Table of Contents
4.1 General | 4-2
4.2 Hazard notices | 4-2
4.3 PC boards and electrical components | 4-2
4.4 Information stored in the system | 4-3
4.5 Assembling guidelines | 4-5
4.6 Front cover | 4-6
4.7 Top cover | 4-6
4.8 Chassis | 4-7
4.9 Main section | 4-8
4.10 Inspiratory section | 4-9
4.11 AC/DC section | 4 -16
4.12 Front section | 4 -21
4.13 Rear section | 4 -22
4.14 PC 2020 Main back-plane | 4 -24
4.15 Expiratory cassette | 4 -25
4.16 User interface | 4 -28
The illustrations in the Spare Parts Lists are useful 4.3 PC boards and electrical components
when disassembling and assembling the system.
4.3.1 Replacing electrical components
After any installation, maintenance or service
intervention in the system, perform a 'Functional check’ As stated in section Hazard notices, mains power and
according to instructions in chapter Important. battery power supply must be disconnected and On/Off
switch set to Off when replacing electrical components.
4.2 Hazard notices If energized electrical components are disconnected
or connected, this may interfere with the operation of
Before disassembling or assembling the system, make the unit.
sure that:
After replacement of PC boards or other electrical
• The system is switched Off.
bullet components, it is recommended to perform an
electrical safety test. For electrical safety tests, refer
• Mains power cable is disconnected.
bullet
to chapter Preventive maintenance, section Safety
• External +12 V DC supply is disconnected.
bullet inspection.
• Backup battery modules are disconnected. Note
bullet
that the backup battery modules will supply power 4.3.2 Handling PC boards
to parts of the system also when the system is
switched Off. Those who come into contact with circuit boards
containing sensitive components must take certain
• Gas supply is disconnected, central gas supply and
bullet
precautions to avoid damaging the components (ESD
backup gas supply.
protection).
• The system, including all gas conveying parts, is
bullet
cleaned. Refer to instructions in the User’s Manual. PC boards (spare parts) must always be kept in
protective packaging for sensitive electronic device.
WARNING! With gas supply connected
to the system, there are pressurized PC boards must not be inserted or removed while the
components inside the unit. All mains power or battery power is applied to the PC
personnel must exercise extreme boards.
caution if fault tracing or adjustments
are performed with gas supply Remove and insert the PC boards very carefully to
connected and with covers removed. avoid damage to the connectors.
• PC 1991 Control
bullet Other technical errors may occur after a software
• PC 1992 Monitoring
bullet installation. These errors will disappear after restarting
the system.
• PC 2004/PC 2024 Expiratory channel
bullet
• PC 2023 Panel
bullet 4.4 Information stored in the system
• PC 2027 Motor control
bullet
All electronic components equipped with an ID PROM
• PC 2028 DC/DC & Battery control
bullet store the information that is specific for the component,
for example article number, version number and serial
Additional actions when replacing PC boards number.
When delivered as spare parts, these components Furthermore, some of the electronic components store
are equipped with a System SW version that may differ information related to the complete system, for
from the version on the system to be repaired. To keep example system software, system info, configurations,
the System SW version used prior to the replacement, logs and so forth. When replacing these components,
the applicable System SW version must be available the stored information may be lost. See table below.
for reinstallation purposes.
Expiratory channel
PC 2004/PC 2024
PC 1991 Control
Stored information
PC 2023 Panel
control
O2 cell
System software X X X X X X
System info: X
§ Serial number
§ Total operating and
ventilation time
§ Next preventive
maintenance
System info: X X
§ Installed software
options
System info: X
§ System version
Configurations X
§ Startup ventilation
configuration
Configurations X
§ Startup alarm
configuration
Configurations X
§ Units, language, date
format, etc.
Date and time setting X
Logs X X2 X
Trends X
Recordings X
Pre-use check results X X X X X
including transducer result calibration calibration calibration calibration
calibrations
Barometer calibration X
Touch screen calibration X
Performed preventive X1
maintenance
Replaced expiratory X1 X
membrane counter
Replaced backlight X1 X3
counter
Backlight operating time X3
Language X
Expiratory cassette: X
§ Date of first use
§ Total operating time
§ Remaining membrane
capacity
O2 cell: X
§ Date of first use
§ Estimated remaining
capacity (O2 cell)
S20104
1
Included in Service report as part of the Service log.
2
Copied to PC 1992 at Power On and continuously when the system is powered.
All parts of the system assembled with screws and Lubricants should normally not be used when servicing
nuts are tightened with a specified torque. Thread the system. If lubricants must be used, use only very
locking adhesives are used if required. small amounts of grease with P/N 95 73 700. This
special grease is O2 compatible.
In order to maintain these specifications over time, it
must be ensured that after any service intervention
WARNING! Due to the increased O2
removed parts are re-assembled and secured
concentration in valves and gas
according to instructions. Make sure to follow the
channels, only special grease intended
guidelines stated below.
for use in high O2 concentrations, are
allowed.
4.5.1 Tightening torque
S20020
1.4 Nm
S20021
4.8 Chassis
4.8.1 Removing the chassis
S20022
s
a
a
1.4 Nm f d
S20023
a a
S20024
0.6 Nm
S20134
1.4 Nm
S20025
When assembling:
• Do not forget to connect the fan cable.
bullet
• Make sure that the Inspiratory gas docking kit
bullet
connects properly to the Inspiratory section lid at
(6). g S20056
• Release the four rubber straps (2), and lift off the
bullet
fan.
S20057
S20058
S20059
f S20109
k
h
S20110
• Pull out the Air inlet filter (4) from the socket.
bullet
• Remove the Gas module lid (3).
bullet f
g
Note: When attaching the Gas module lid to the
Patient unit, make sure the Battery extraction tag (5)
is not stuck under the lid. This will make it hard to 1.4 Nm
S20111
a
Note: Additional actions may be required after
replacement of PC boards (PC 2027 is included in
the Turbine module). Refer to section Replacing PC
boards in this chapter.
S20033
0.5 Nm
S20035
S20036
d
s
a
S20112
a 1.4 Nm
S20027
S20120
a
S20114
S20121
• Release the four rubber straps (2), and lift off the
bullet
fan.
S20037
s d
S20122
d s
a S20046
a
S20116
a
s
S20117
s a
S20031
a
S20124
S20028
S20115
f
S20030
S20123
• Lift the board slightly and then slide it out from the
bullet
front.
a
S20118
S20126
S20041
S20042
S20043
S20044
S20125
S20047
S20048
• Remove the five screws (2) and lift off the User
bullet
interface from the holder.
a s
S20049
S20050
S20051
a
S20052
S20053
f
S20054
5 Service procedures
Table of Contents
5.1 Service key | 5-2
5.2 Service & Settings | 5-3
5.3 Software installation | 5 -58
5.4 Leakage detection | 5 -62
5.5 Battery modules | 5 -68
5.6 Memory backup batteries | 5 -70
The Service key is a USB memory stick provided to If the Touch screen does not function properly, due
personnel trained and authorized by Getinge to to a faulty panel or not being well enough calibrated,
perform installation, service or maintenance of the a mouse can be connected to the system. The mouse
system. can then be used to, e.g., logging in to Service &
Settings to start the Touch screen calibration.
With the Service key it is possible to access different
user levels of the built-in Service & Settings software.
Note: Using a mouse with the system is only
There are two user levels intended for Service that
permitted for service purposes.
are described in this manual:
Notes:
• When saving information to a USB memory stick
bullet
(logs, screenshots, etc), allow this procedure to
complete before removing the memory stick.
Information will be shown on the control panel
when completed.
Notes:
• It is only possible to access Service & Settings
bullet
with the system in Standby mode.
• STATUS
bullet
• LOGS
bullet
• SERVICE REPORT
bullet
• CONFIGURATION
bullet
• COPY CONFIGURATION
bullet
• INSTALLATION
bullet
• CALIBRATION
bullet
• EXIT
bullet
These Service & Settings main functions are described
later in this chapter.
Status
• SYSTEM INFORMATION
bullet
• O2 CELL
bullet
• EXPIRATORY CASSETTE
bullet
• BATTERY STATUS
bullet
Logs
• Alarms
bullet
• Settings
bullet
• Functions
bullet
• Configuration
bullet
Notes:
• Logs can be saved to a USB memory stick.
bullet
• Only a USB memory stick may be connected to
bullet
the USB port.
Service report
Configuration
• SETTINGS
bullet
• STARTUP CONFIGURATION
bullet
• ALARMS
bullet
• NETWORK
bullet
• SET DATE & TIME
bullet
• BIOMED CODE
bullet
Settings
• Height
bullet
• Weight
bullet
Startup configuration
Alarms
Network
SET DATE & TIME is used to set the systems real time
clock.
Biomed code
Copy configuration
Installation
Exit
• STATUS
bullet
• LOGS
bullet
• TESTS
bullet
• SERVICE REPORT
bullet
• CONFIGURATION
bullet
• COPY CONFIGURATION
bullet
• INSTALLATION
bullet
• CALIBRATION
bullet
• EXIT
bullet
These Service & Settings main functions are described
later in this chapter.
Status
• SYSTEM INFORMATION
bullet
• O2 CELL
bullet
• EXPIRATORY CASSETTE
bullet
• BATTERY STATUS
bullet
Logs
• Test results
bullet
• Test details
bullet
• Preventive maintenance
bullet
• Service report
bullet
• Technical alarms
bullet
• Installation
bullet
• Alarms
bullet
• Settings
bullet
• Functions
bullet
• Configuration
bullet
Notes:
• Logs can be saved to a USB memory stick.
bullet
• Only a USB memory stick may be connected to
bullet
the USB port.
Tests
After each test, the user can select Yes if the test
passed, No if the test failed or Cancel to cancel the
test. These results can be saved as a log file on a USB
memory stick.
Service report
Configuration
• SETTINGS
bullet
• STARTUP CONFIGURATION
bullet
• ALARMS
bullet
• NETWORK
bullet
• SET DATE & TIME
bullet
• BIOMED CODE
bullet
Settings
• Date format
bullet
• Decimal separator
bullet
• Language
bullet
• PM reminder (On/Off)
bullet
• Enable biomed service report (On/Off). Will
bullet
enable/disable the SERVICE REPORT function in
Biomed user level. Available in Field service user
level only.
• Airway pressure
bullet
• Flow
bullet
• Compliance
bullet
• Resistance
bullet
• Gas supply pressure
bullet
• Barometric pressure
bullet
• Height
bullet
• Weight
bullet
Startup configuration
Alarms
Network
SET DATE & TIME is used to set the systems real time
clock.
Biomed code
Copy configuration
Installation
Calibration
Barometer
Touch screen
Exit
Software option
installs a new System software version. A System Software option
installation
software installation is not depending on the serial SW Option
Software option
number of the system and does not alter the SW Option
installed Software options. Software option
System software
system. A Software option is individually created for
installation
each system and can only be installed on the
Corrections
intended system. Serial number of the system must and function
be stated when ordering a Software option. improvement
S25085
• Tap INSTALLATION.
bullet
• Tap SOFTWARE INSTALLATION.
bullet
• The software installation activation box will now
bullet
open. Check that correct System SW version to be
installed is stated in this box.
Note 1: Very occasionally, it may seem that the Note 2: After a software installation, technical alarms
installation is interrupted and the Service & Settings may be activated due to reset internal memory on the
login menu will appear. If so, do not cancel the PC boards. Restart the system once more and check
installation. The installation will proceed in the that no technical alarms are activated.
background. Wait 15 minutes before restarting the
system.
• Tap INSTALLATION.
bullet
• Tap PERMANENT OPTIONS.
bullet
When a leakage is detected in the Internal leakage • Gas module silicone connector muff
bullet
test, the Pre-use check will stop and display either • Inspiratory gas docking incl. one-way valve
bullet
Leakage or Excessive leakage. To locate the leakage,
the three leakage tests described below can be used.
• Inspiratory flowmeter connector muff
bullet
• Inspiratory flowmeter
bullet
Test equipment:
• Inspiratory channel connector muff
bullet
• Test tube
bullet • Inspiratory pressure silicone tube
bullet
• Test tube
bullet • Inspiratory pressure transducer
bullet
Insp.
Exp.
S20087
Expiratory channel
Expiratory
valve coil
Part of leakage
detector kit
Expiratory
pressure
transducer PC 1781
P40 N40
P50
MOD1
Manometer
S20088
Expiratory cassette
• Expiratory cassette.
bullet
N17P17
S20090
1. Connect the test equipment to the Inspiratory If no leakage was found in the above mentioned parts,
outlet. remove the Inspiratory channel and restart the Pre-use
2. Pressurize to approx. 50 cmH2O. A pressure drop check. Check that the Safety valve piston is activated
indicates a leakage. If a leakage is indicated, the during the internal leakage test.
parts below should be checked/replaced:
If the piston is activated, continue with Inspiratory gas
• Inspiratory channel.
bullet docking and Inspiratory pipe. If not, check/replace
Note: In order to remove the Inspiratory as follows (one at a time):
channel, the safety valve must be opened (it is
closed during the test). This is done by • Check the safety valve piston including cable and
bullet
cancelling the test and setting the ventilator to connector.
Standby.
• Replace PC 2030 Inspiratory Channel.
bullet
• O2 cell bacteria filter seal.
bullet • Replace PC 2004/PC 2024 Expiratory Channel.
bullet
• Safety valve membrane. Clean if necessary.
bullet
• Inspiratory pressure tube.
bullet
• Inspiratory pressure bacteria filter and holder.
bullet
• Slot number.
bullet
• Battery module S/N.
bullet
• Remaining operating time in minutes for each
bullet
battery.
5.6.2 Preparations
• Cut the cable ties holding the old battery (4) to the
bullet
PC board.
6 Troubleshooting
Table of Contents
6.1 General | 6-2
6.2 Start-up test | 6-2
6.3 Pre-use check | 6-3
6.4 Other tests | 6-6
6.5 Test logs | 6-7
6.6 Technical error codes and messages | 6 -23
6.1 General
The information in this chapter is applicable to System The troubleshooting guides below focus on the
version 1.0 – 4.0. For functionality enhancement, the technical problems. Information about clinically related
latest System software version is recommended. problems can be found in the User's Manual.
• Software check
bullet 1. Restart the system. Do not touch the User interface
during system start-up. Interfering with the touch screen,
• Reading EEPROM
bullet loudspeaker grid and so forth, may affect the internal
• Checksum EEPROM
bullet technical tests.
2. Reinstall the System software.
• Audio test.
bullet
(BRE + MON + PAN)
The system requests the user to start the automatic Checks the following sensors in the inspiratory section:
Pre-use check at every startup of the system. It is also
possible to select the Pre-use check via the Standby • Humidity (0.0–100.0 %). Sensor on PC 2030
bullet
menu. Inspiratory channel.
Checks that the loudspeakers in the User interface Generates failures on different system voltages.
generate an adequate sound level. A sound level is Checks that these failures are detected.
generated and a microphone mounted nearby the
loudspeakers measures the sound. PAN calculates 6.3.2 Barometer test sequence
the measured value to verify that the sound level
meets the requirement. Barometer test (MON)
Monitor audio test (MON) Checks that the barometric pressure measured by the
internal barometer is within 650–1070 hPa (mbar).
Checks that the buzzer in the Patient unit generates
an adequate sound level. A power failure is generated Checks that the measured barometric pressure values
in order to activate the buzzer. A microphone mounted between BRE and MON differ less than 8 hPa (mbar).
nearby the buzzer measures the sound. MON
calculates the measured value to verify that the sound
level meets the requirement.
6.3.3 Internal leakage test sequence • Checks that acceleration (rise time) from approx.
bullet
5000 to 40000 r/min is <0.150 s.
Internal leakage test (BRE)
• Checks that speed is 38000–42000 r/min.
bullet
Checks the internal leakage, with the test tube
connected, using the inspiratory and expiratory
• Checks that deceleration (fall time) from approx.
bullet
40000 to 5000 r/min is <0.150 s.
pressure transducers:
6.3.4 Turbine and Gas supply test sequence • With the inspiratory pressure transducers factory
bullet
default gain used as a reference, a new gain factor
Gas supply test (MON) is set for the expiratory pressure transducer. The
new gain factor may not differ more than ±8% from
Checks gas supply: factory calibration.
6.3.6 Safety valve test sequence • Checks that measured inspiratory flow differs less
bullet
than 5% from the flow delivered from the O2 gas
Monitor safety valve test (MON)
module at a flow of 0.5 l/s (30 l/min).
Checks the hardware signals related to the safety • Checks that the difference between MON and BRE
bullet
valve functions. subsystems is less than 0.005 l/s (<1%).
Breathing safety valve test (BRE) This test requires O2 gas supply.
Checks that the safety valves opening pressure is Breathing flow transducer test (BRE)
approx. 117 cmH2O and calibrates the valve if
required. The test can be repeated up to five times to Checks the offset values for:
achieve a Passed result:
• Inspiratory flow (22.021–24.609 V).
bullet
• Acceptance interval for the first test is
bullet • O2 flow (0.000–0.081 V).
bullet
115.50–118.50 cmH2O.
• Expiratory flow (0.093–0.142 V).
bullet
• Acceptance interval for the following tests is
bullet
114.00–120.00 cmH2O. Flow test (MON)
• The test fails if an opening pressure within
bullet
Monitoring flow test in the S&S Test menu.
114.00–120.00 cmH2O is not reached after five
attempts. Checks the offset values for:
Checks and calibrates the O2 cell: • Checks inspiratory flow (0.9798–1.0198 l/s).
bullet
• Checks O2 flow (0.4842–0.5040 l/s).
bullet
• Checks if the O2 cell is consumed. Expected life
bullet
• Measures expiratory flow and calculates gain
bullet
time 500 000 %hours.
(k value) for the expiratory cassette (0.90–1.15).
• Checks with 100 % O2 (>70.0 %).
bullet
With 100 % O2 concentration and 0.5 l/s gas flow:
• Checks with 21 % O2 (<25.0 %).
bullet
• Checks offset (4–24 mV).
bullet • Checks inspiratory flow BRE and MON
bullet
(0.4750–0.5250 l/s).
• Calculates gain.
bullet
• Checks O2 flow BRE and MON (0.4900–0.5100 l/s).
bullet
This test requires O2 gas supply.
• Checks that the difference between O2 flow BRE
bullet
and MON is less than 0.005 l/s (<1 %).
6.3.8 Flow transducer test sequence
• Checks expiratory flow MON (±5 % of delivered
bullet
Insp flow sensor test (MON) flow).
Sensirion flow sensor test in the S&S Test menu. • Checks that the difference between expiratory flow
bullet
BRE and MON is less than 0.005 l/s (<1 %).
Checks the inspiratory flowmeter:
With 21 % O2 concentration and 0.5 l/s gas flow: 6.3.10 Patient circuit test sequence
Battery switch test (MON) For pin configuration and signal names in the Alarm
output connector N67, see Connectors in chapter
Checks the battery module(s):
Diagrams. For further information, refer to the 'Alarm
output connector – Reference Manual'.
• Checks that a Maquet battery module is connected.
bullet
• Checks the status, age and number of charge
bullet Alarm state test (MON)
cycles.
Checks that no Technical error alarms are active
• Checks that there is a battery module in slot 2 (main
bullet during the Pre-use check.
slot).
This test is not performed if: This test performs the Patient circuit test separately,
as described above.
• Less than 10 minutes backup time remains in the
bullet
connected battery modules.
The Pre-use check, complete test as well as separate Some checks will be repeated if they fail. If they pass
tests, is logged in the Service log. Select Test results when repeated, the check and the complete test
and Test details to show all test result logs. A Service sequence will pass. This will be shown in the log.
key is required to access the Service log. The
description below shows how to interpret the When a check fails due to a user error; a Dialog Choice
Service log. (DIALOG) appears and gives the user the choice to
redo the check by choosing Redo or, to bypass the
If one or more checks within a test fail, the log will show check by choosing Cancel.
which check that failed and also include an error code.
These error codes are called Check Failure Identifier Always perform recommended actions in a
(CFI). All checks that fail during the Pre-use check chronological order. A test that fails can cause
have a unique CFI. Possible causes and recommended subsequent tests to fail.
actions for checks that fail permanently or intermittently
are listed in the following CFI tables.
Comment: 0.0–100.0 %.
2 Turbine temperature is not within accepted limits. 1. Check Air inlet filter.
2. Replace Turbine module.
Comment: 0.0–70.0 °C.
Comment: 0.0–60.0 °C
Comment: 1586–1690.
2 Remote Alarm has not been activated. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.
5 The buzzer has not been turned on. 1. Replace PC 1992 Monitoring.
6 The buzzer has not been turned off. 1. Replace PC 1992 Monitoring.
2 Barometric Pressure (MON) is not within the 1. Calibrate the barometer (S&S).
accepted limits. 2. Replace PC 1992 Monitoring.
Comment: <650 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.
Comment: <0.2000 l
3 The System Volume is too high. 1. Check that the correct test tube is used during the Pre-use
check and check that the test tube is correctly connected.
Comment: >0.4000 l 2. If the ‘Flow transducer test’ also fails, replace Inspiratory
flowmeter.
3. Check that the expiratory cassette is correctly seated in the
cassette compartment.
4. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
5. Check that the pressure transducer tubes/filters are correctly
mounted.
6. Check that the inspiratory gas docking, inspiratory flowmeter
and inspiratory channel are correctly mounted.
7. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
8. Check that the safety valve closes properly when the Pre-use
check starts (distinct clicking sound from the valve). If the
safety valve not closes properly during this check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.
3 The difference between O2 supply pressure 1. Check that the connected O2 supply pressure is stable.
(MAX) and O2 supply pressure (MIN) is too 2. Replace O2 gas module.
large.
2 Expiration Offset could not be stored persistently 1. Replace PC 1781 Expiratory pressure transducer.
- Value out of range. 2. Replace PC 1991 Control.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.
SERVO-air Ventilator System, Service Manual 6 - 15
| 6 | Troubleshooting |
2 Expiration offset >6 cmH2O from factory default. 1. Replace PC 1781 Expiratory pressure transducer.
2. Replace PC 1992 Monitoring.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.
3 Expiration gain >8% from factory default. 1. Cancel the ‘Pressure transducer test’ and continue with the
remaining tests (this will calibrate the safety valve). Restart
the Pre-use check.
2. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Check/replace PC 1781 Pressure transducer (Insp. and Exp.).
To locate the faulty pressure transducer, replace one
transducer at a time.
4 Expiration offset is different at zero pressure and 1. Replace PC 1781 Expiratory pressure transducer.
high pressure.
5 The difference between the values measured in 1. Replace PC 1991 Control.
BRE and MON is too high. 2. Replace PC 1992 Monitoring.
Comment: Differs more than ±0.80 cmH2O.
6 Inspiration pressure transducer could not be 1. Replace PC 1781 Inspiratory pressure transducer.
locked or is faulty.
7 Expiration pressure transducer could not be 1. Replace PC 1781 Expiratory pressure transducer.
locked or is faulty.
8 Inspiratory pressure value out of bounds. 1. Replace Turbine module.
2 Expiratory Pressure too low. 1. Clean or replace Turbine module one-way valve.
2. If possible, replace the expiratory cassette and check if the
Comment: Measures at 40, 20, 10 and 5 cmH2O. new cassette is accepted by the Pre-use check. If the new
Measured values must be within ±15 cmH2O. cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Replace Expiratory valve coil.
3 Voice Coil too strong. 1. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
Comment: Expiratory valve coil not within cassette was accepted by the Pre-use check, the fault was
specified limits; offset 0.06768–0.15684 A, gain located to the cassette membrane.
0.00730–0.01130 A/cmH2O. 2. Replace Expiratory valve coil.
2 The Safety Valve could not be closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.
2 Pressure not rising. Timeout. 1. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
Comment: Not possible to reach opening 2. Check that the inspiratory gas docking, inspiratory flowmeter,
pressure 117 cmH2O. inspiratory channel and turbine module one-way valve are
correctly mounted.
4 The Safety Valve is not closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.
Comment: >25 %.
Error in the log text. Correct text is:
O2 Concentration in AIR Supply too high.
6 - 18 SERVO-air Ventilator System, Service Manual
| Troubleshooting | 6 |
6 O2 supply low. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: <200 kPa 2. Replace O2 gas module.
7 Gas supply not approved. 1. Gas supply test not passed. Redo the Gas supply test.
2 Insp flow sensor test, Bre - Mon comparison 1. Replace PC 1991 Control.
failed. 2. Replace PC 1992 Monitoring.
5 Insp flow sensor test, Bre - Mon Sensirion 1. Replace PC 1991 Control.
comparison failure. 2. Replace PC 1992 Monitoring.
Comment: 22.021-24.609 V
Comment: 0.000-0.081 V
3 Expiration Offset could not be stored persistently 1. Check that the expiratory cassette is correctly seated in the
- Value out of range. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.093-0.142 V
3. Replace PC 2004/PC 2024 Expiratory channel.
Comment: 0.000-0.033 V
3 The Expiration Offset is not within the accepted 1. Check that the expiratory cassette is correctly seated in the
limits. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.024-0.072 V
3. Replace PC 2004/PC 2024 Expiratory channel.
4 Expiration Gain is not within the specs. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: 0.90-1.15 (k value) 2. Check that the cassette is correctly seated in the cassette
compartment. If possible, replace the expiratory cassette and
check if the new cassette is accepted by the Pre-use check.
If the new cassette was accepted by the Pre-use check, the
fault was located to the cassette. The fail with the old cassette
may in this case be due to water collected inside the cassette.
3. Replace Inspiratory flowmeter.
4. Replace O2 gas module.
5. Replace PC 1785/PC 2031 Expiratory channel connector.
6. Replace PC 1991 Control.
7. Replace PC 1992 Monitoring.
8. Replace PC 2004/PC 2024 Expiratory channel.
6 Timeout getting Expiration Gain (k value) 1. Check that the expiratory cassette is correctly seated in the
cassette compartment.
2. Replace the expiratory cassette.
2 The batteries have not been switched on 1. Mains power not disconnected as demanded.
(TIMEOUT).
3 The batteries have been switched back off 1. Mains power not reconnected as demanded.
without mains being back on first. (TIMEOUT).
4 No battery operation detected by BRE. 1. Replace PC 1991 Control.
5 Remaining battery power is insufficient for the 1. Charge or replace Battery modules.
test to be performed.
7 The batteries have been switched on without 1. Replace PC 2028 DC/DC and Battery control.
mains loss first. (TIMEOUT). 2. Replace PC 1992 Monitoring.
8 The batteries have been switched back off 1. Replace PC 2028 DC/DC and Battery control.
without mains being back on first. (TIMEOUT). 2. Replace PC 1992 Monitoring.
10 Power system external power source test failed. 1. Connect mains power.
(POW)
11 Power system battery switch test failed. 1. Connect Battery module in slot 2 (main slot).
(POW) 2. Replace PC 2028 DC/DC and Battery control.
12 Power system no external power source test 1. Replace PC 2028 DC/DC and Battery control.
failed.
(POW)
2 The Compressible Volume is too low. 1. Check or replace the patient circuit.
3 The Compressible Volume is too high. 1. Check or replace the patient circuit.
4 Leakage too high, or the safety valve is either 1. If the internal leakage test has passed, the leakage is to be
not locked or its status could not be read. located to the patient circuit. Check for leakage or replace
the patient circuit.
Comment: >0.0800 l/min
5 System Volume is too low. 1. Check that the internal leakage test passed.
Comment: <0.1500 l
6 System Volume is too high. 1. Check that the internal leakage test passed.
2. Improper patient circuit connected.
Comment: >6.0000 l
2 Remote External Alarm is already active. 1. Check the external alarm system.
2. Replace PC 2021 Rear connectors.
6.6.1 Monitoring
Error
Error message / Possible cause Recommended action
code
1 Power error. 1. Replace PC 2028 DC/DC & Battery control.
Error
Error message / Possible cause Recommended action
code
7 Software error. 1. Replace PC 1991 Control.
2. Replace PC 1992 Monitoring.
Patient category mismatch between Breathing and Monitoring
subsystems. 3. Replace the O2 gas module.
Error
Error message / Possible cause Recommended action
code
25 Communication error. Depending on ID # and Error # stated in the
Technical alarm log.
Communication error between ID PROMs/EEPOTs and
Monitoring subsystem. If repeated, replace the spare part indicated by
the ID # according to the list below:
The ID # is followed by an Error # in the Technical alarm log.
ID #
For Error 6, report to MCC HSC or to your local Getinge
representative for further information. 0: PC 2020 Main back-plane.
For all other Error #, replace the part identified by the ID #. 1: PC 1992 Monitoring.
Error
Error message / Possible cause Recommended action
code
32 Software version conflict. 1. Reinstall the System software.
2. Report to MCC HSC or to your local
Conflict between the software versions installed. Getinge representative for further
information.
Error
Error message / Possible cause Recommended action
code
48 Pre-oxygen time exceeded. 1. Replace PC 1991 Control.
Error
Error message / Possible cause Recommended action
code
60 Charger error. 1. Replace Battery module.
2. Replace PC 2028 DC/DC & battery control.
This error code is an indication of a charger error in the power
supply subsystem.
62 Very high temperature in Turbine module. Ventilation stopped. 1. Replace Turbine module.
Error
Error message / Possible cause Recommended action
code
70 Ambient air temperature sensor range error. 1. Replace PC 2030 Inspiratory channel.
71 Air humidity (RH) sensor range error. 1. Replace PC 2030 Inspiratory channel.
73 Ambient air temperature and humidity sensor error. 1. Replace PC 2030 Inspiratory channel.
Error
Error message / Possible cause Recommended action
code
80 Power error. 1. Replace PC 2028 DC/DC & Battery control.
2. Replace Inspiratory flowmeter.
+5 V to Inspiratory flowmeter too low, i.e. < +4.5 V.
6.6.2 Breathing
Error
Error message / Possible cause Recommended action
code
10001 Memory backup battery depleted. 1. Replace battery on PC 1991 Control.
6.6.3 Panel
Error
Error message / Possible cause Recommended action
code
20002 Backlight error. If the display is lit (and the error code is shown
on the display):
The backlight power consumption is measured. If the power 1. Replace PC 2023 Panel.
consumption is outside specified range, this technical error is If the display is not lit (the display is dark):
activated. 1. Replace Display incl. touchscreen.
This error message may also be logged immediately before 2. Replace PC 2023 Panel.
the No battery capacity alarm when the system is running in
Note: With a failure on the display, or likely
battery mode.
also on PC 2023 Panel, the User interface
display will become dark. This error code will
in such case not be possible to see (but will
be logged).
20003 Activate all pending alarm. Report to MCC HSC or to your local Getinge
representative for further information.
There is a software-related internal communications problem.
20004 Alarm sound level too low. 1. Restart the system. Do not touch the User
interface during system startup. Interfering
The loudspeakers are checked using a microphone during with the Touch screen, loudspeaker grid,
startup. etc., may affect the internal technical tests.
2. Replace loudspeaker.
3. Replace PC 2023 Panel.
Error
Error message / Possible cause Recommended action
code
For systems with PC 2004 Expiratory channel
40001 Expiratory flowmeter error. Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM read/write error:
See information logged together with error code 40001 in the 1. Replace PC 2004 Expiratory channel.
Service log / Technical alarms.
Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM checksum error:
1. Replace Expiratory cassette.
Note that this error may occur if the system is running outside
operating temperature range.
Error
Error message / Possible cause Recommended action
code
40007 Power error. 1. Replace PC 2024 Expiratory channel.
2. Replace PC 2028 DC/DC & Battery control.
+5 V is more than 15 % outside nominal value.
40010 Internal temperature too high, > 77 ºC. 1. Ensure proper cooling from the Main fan:
- Check fan operation
The temperature is too high or temperature sensor error on
PC 2024. - Check the operating temperature
- Check the fan filter.
2. Replace PC 2024 Expiratory channel.
1. Replace O2 cell/sensor.
2. Replace PC 1992 Monitoring.
7 Preventive maintenance
Table of Contents
7.1 General | 7-2
7.2 Preventive maintenance intervals | 7-2
7.3 Required equipment | 7-2
7.4 Preventive maintenance calculation | 7-2
and notification
7.5 Maintenance kit SERVO-air 5000 h | 7-3
7.6 Performing the Preventive maintenance | 7-3
7.5 Maintenance kit SERVO-air 5000 h 7.6 Performing the Preventive maintenance
Only original parts from the manufacturer must be • Disassembling and assembling of the system is
bullet
used. Spare parts and maintenance kits can be required when replacing parts included in the
ordered from your local Getinge representative. Maintenance kit SERVO-air 5000 h. If not stated
otherwise, refer to chapter Disassembling and
When performing this maintenance, the Maintenance
assembling for instructions.
kit SERVO-air 5000 h should be used.
• All steps during Preventive maintenance should be
bullet
The following parts, included in the Maintenance kit documented.
SERVO-air 5000 h, shall be replaced:
• The letters A – D in the text below refers to the
bullet
description of the Maintenance kit above.
7.6.1 Preparations
S20096
S20098
S20099
a s
S20101
7.6.12 Gas supply pressure transducer A. Protective earthing (Protective earth resistance)
B. Touch current (Equipment leakage current –
• Connect the gas supply (O2).
bullet direct method)
• Select SYSTEM STATUS > General.
bullet C. Patient leakage current (Applied part leakage
current – direct method)
• Disconnect the O2 gas supply.
bullet
D. Insulation (Insulation resistance) – Measured
• Check that the O2 pressure value in the SYSTEM
bullet between mains and protective earth.
STATUS -window drops.
E. Insulation (Insulation resistance) – Measured
between isolated connections and protective
7.6.13 Battery modules earth.
The system is classified within the standards listed
• The Battery modules shall be replaced after
bullet
above, and shall be tested according to the
300 charge cycles or four years (48 months) from
manufacturing date, whichever comes first. following:
8 Index
Table of Contents
8.1 Alphabetic index | 8-2
9 Revision history
Table of Contents
9.1 SERVO-air revision history | 9-2
9.2 Compatibility chart | 9-4
9.3 Service Manual revision history | 9-5
Service improvements:
• Now possible to reset the expiratory
bullet
cassette membrane counter in Service
& Settings.
• SERVO Communication Interface (SCI)
bullet
updated to Protocol version 0004.
• Stability enhancements to address the
bullet
occurrence of technical errors
Technical errors 43, 78, 80 and 81.
• Timing improvements in the internal
bullet
communication between Panel and
Monitoring to reduce occurrence of
Technical error 20006.
• Improvements to reduce the
bullet
occurrence of buzzer alarms during
software installation (2.1.4).
• In Service & Settings, STATUS > Parts
bullet
data, the new alternative version of PC
2024 Expiratory channel will be
presented with the correct name
(2.1.4).
PC 2024
PC 2004 Expiratory PC 2028
PC 1991 PC 1992 Expiratory channel PC 2023 DC/DC &
System SW version Control Monitoring channel 68 81 702 or Panel Battery control
SERVO-air requirement for 66 89 785 66 89 790 66 92 477 68 87 462 68 81 737 68 81 577
System SW version options revision # revision # revision # revision # revision # revision #
1.0.1 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.2 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.3 – ≤ 08 ≤ 09 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.2 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.4 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.2.0 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
S20136
10 Diagrams
Table of Contents
10.1 System block diagram |10 - 2
10.2 Connectors |10 - 3
10.3 PC 2020 Main back-plane |10 - 4
10.4 PC 2030 Inspiratory channel |10 - 8
10.5 Service & Settings menus |10 -12
10.6 System overview |10 -13
N34
N33
PC 2026 Baery extension
P34
P33
Baery module
P32
Breathing BRE
Slot 1
Monitoring MON
O₂ evacuaon fan
Safety
valve Panel PAN
P31
Baery module Power POW
T
Slot 2, main slot Blower BLO
Inspiratory
flowmeter
System ID
O₂ cell
P17
P23
P19
P18
P37
N15
Gas module O₂
PC 1781 Inspiratory RH
pressure transducer
Turbine module
PC 2027 Motor control
N16
P1 T
P2
Filter dP
PC 2030 Inspiratory channel
P21
P20
Loud-
speaker
N21
N20
P72
PC 2029 Inspiratory
channel extension
N11
N10
SD
N74
P11
P10
N12
P12
P39
On/Off switch
PC 2023 Panel
N77
N70
N36
HDD
ALM_LED
External
N26
P3
+12 V DC inlet
Alarm output
N67
connector
N78
P76
P92
T
RS-232
N29
T
N2
P2
RS-232
PC 2021 Rear connectors
LCD display
N30
connector
Backlight
N9
P9
N27
P1
PC 2028 DC/DC & Baery control
PC 1991 Control
Ethernet
N28
P4
T
PC 1994 Nebulizer
P40 P41
Nebulizer
Expiratory cassee
P35
P5
P46
P13
cassee channel
P7
connector
AC/DC converter
PC 2033 AC/DC Mains power
supply connector 110-230 VAC
150 W power supply
Main fan
S20013
10.2 Connectors
S20106
1. GND
2. +12V_UNREG_EXT_DC
Recommended cable connector:
LEMO 736-FGG2B302CLAD82Z, Circular Push Pull
Connector.
Recommended cable:
Outer diameter 7.2 mm – 8.2 mm, two wire cable, area min.
2
2.5 mm (AWG13).
S20107
S20108
9 O2 evacuation fan
Inspiratory section
N19
P19
N15 P1
O2 P19 N19 T RH
7
cell
1 O2 Gas O2
ID #6 P18 N18 P23 N23
module inlet
ID #34
3
Inspiratory E2Pot
Inspiratory “0111””1111”
6 Inspiratory channel gas
outlet docking
5 In
Emergency flo spi
wm rat
air intake 8 eteory
4 r 2
Safety T
Safety valve valve Inspiratory Turbine
pull magnet pipe module
ID #32
Air
P17
T
inlet
X
N17
PC 1781 10 PC 2030 Inspiratory
ID #4 channel N16 P1 PC 2027 Motor control P
P20 P21
N20 N21 E2Prom “011”
11
37 PC 2029 Inspiratory P50 Inspiratory
12
channel extension MOD2 pressure
N10 N11 transducer
PC 1994 25
Nebulizer
Nebulizer N1 P32 33 Battery ID #xx
Aerogen 31 PC 2026 PC 2025 module Slot 1
interface Battery extension 32 Battery back-plane
User interface Battery Slot 2
P33 N33 P31 33 ID #xx
P34 module (main slot)
39
N34
P35
P76
28 PC 2028 DC/DC & Battery control
40 N1 RS-232
Interface
Iso. N29
N78 P1 RS-232 (1)
Battery N22 22 PC 1991 Control
P22A/B RS-232
Interface
Iso. N30 RS-232 (2)
CPU
N12
P12
+12V DC
N4 Fuse F1 N26
P4
supply
PC 2004/PC 2024 input
User interface CPU
E2Prom “000”
21 Expiratory channel
P40 N40
Remote alarm
interface
Iso. N67 Alarm
PC 2023 T E2Pot “0010” ID #1024 output
MOD1 P41
N5
P5
PC 2033
PC 2031/ 27
PC 2029
N7 P7 Expiratory
PC 1781
Exp.
channel
PC 2031/ connector
Insp.
PC 1785
P46 Expiratory section
PC 2004/
PC 1993
N46
PC 1991
PC 1992
PC 2024
PC 2026
Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and for life
science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts, healthcare
professionals and medtech specialists, we are improving the everyday life for people - today and tomorrow.
www.getinge.com