Download as pdf or txt
Download as pdf or txt
You are on page 1of 222

Service Manual

Servo-air Ventilator System


| Contents |

Contents

1 Important
2 Introduction
3 Description of functions
4 Disassembling and assembling
5 Service procedures
6 Troubleshooting
7 Preventive maintenance
8 Index
9 Revision history
10 Diagrams

SERVO-air Ventilator System, Service Manual 0-1


| Contents |

0-2 SERVO-air Ventilator System, Service Manual


| Important | 1 |

1 Important
Responsibility for the safe functioning of the equipment
1.1 General reverts to the owner or user in all cases in which
service or repair has been done by a non-professional
The information in this Service Manual is based on or by persons who are not employed by or authorized
the SERVO-air Ventilator System of System by Getinge, and when the equipment is used for other
version 4.0. Unless otherwise stated, the Service than its intended purpose.
Manual is valid also for System versions below 4.0.
The system complies with standards and requirements
Maquet Critical Care AB, part of Getinge, is the legal as stated in the User's Manual.
manufacturer of the SERVO ventilator systems.
WARNING! Do not modify this
Service documentation for the SERVO-air Ventilator
equipment without authorization of the
System consists of:
manufacturer.
• User's Manual including Cleaning and Maintenance
bullet
manual. The User's Manual is an indispensable
complement to the Service Manual for proper System version and System software version
servicing.
The System version number can be found in the
• Service Manual
bullet SYSTEM STATUS window on the User interface.
• Installation Instructions
bullet Make sure that the version of the User’s Manual
corresponds to the System version.
• Spare parts information
bullet
• Documentation for optional equipment included in
bullet The System SW version number can be found in the
the system SYSTEM STATUS window on the User interface. For
• Product End-of-Life - Disassembly Instructions
bullet functionality enhancement, the latest released System
(Disassembly and Disposal Instructions). SW version is always recommended.

The SERVO-air Ventilator System is referred to as Serial number


'the system' in this document.
There are two serial number labels on the unit:
The User’s Manual for the SERVO-air Ventilator
System is referred to as ‘the User's Manual’ in this • One label is attached to the Patient unit above the
bullet
document. mains power supply connector. The serial number
stated on this label is the ID number of the Patient
Maquet Headquarter Support Center is referred to as unit. The serial number is also stored in the software
‘MCC HSC’ in this document. Note that MCC HSC is memory as the 'System ID'.
available only for the Getinge service organization.
For other service organizations, for example hospital
• One label is attached on the rear side of the User
bullet
interface. The serial number stated on this label is
biomedical departments, technical support is provided
the ID number of the User interface.
by the local Getinge representative.

In addition to the Important information given here and


in the related documents, always pay attention to
applicable local and national regulations.

SERVO-air Ventilator System, Service Manual 1-1


| 1 | Important |

1.2 Symbols used in this manual 1.3 Hazard notices


Before disassembling or assembling the system, make
WARNING! Indicates critical
sure that:
information about a potential serious
outcome to the patient or the user.
• The On/Off switch is set to Off.
bullet
• Mains power cable is disconnected.
bullet
CAUTION: Indicates instructions that
must be followed in order to ensure the • External +12 V DC supply is disconnected.
bullet
proper operation of the equipment. • Backup battery modules are disconnected. Note
bullet
that the backup battery modules supply power to
parts of the system also when the system is
Note: ESD sensitive components. switched Off.
When handling ESD-sensitive devices,
established procedures must be • Gas supply is disconnected, central gas supply and
bullet
observed to prevent damage. backup gas supply.

• The system, including all gas conveying parts, is


bullet
cleaned. Refer to instructions in the User’s Manual.
Note: Special waste. This product
contains electronic and electrical WARNING! With gas supply connected
components. Discard disposable, to the system, there are pressurized
replaced and left-over parts in components inside the unit. All
accordance with appropriate industrial personnel must exercise extreme
and environmental standards. caution if fault tracing or adjustments
are performed with gas supply
connected and with covers removed.
Note: Recyclable material. Recycling
must be performed in accordance with
appropriate industrial and WARNING! With power supply
environmental standards. connected to the system, there are
energized electrical components inside
the unit. All personnel must exercise
Note: Trained and authorized
extreme caution if fault tracing or
personnel. Only personnel trained and
adjustments are performed with power
authorized by Getinge shall be
supply connected and with covers
permitted to perform installation,
removed.
service or maintenance of the system.

CAUTION: Do not expose the batteries


Note: Text inside a box is used to highlight
to water, fire or excessive heat. Do not
important information.
crush, disassemble, puncture or short
circuit the connector terminals.

1-2 SERVO-air Ventilator System, Service Manual


| Important | 1 |

1.4 Installation 1.6 Service

Note: Trained and authorized WARNING! The system must not be


personnel. Only personnel trained and serviced or maintained while in use with
authorized by Getinge shall be a patient.
permitted to install the system. The
installation and handing-over
Note: ESD sensitive components.
procedures are described in the
When working with ESD sensitive
Installation instructions.
components, always use a grounded
wrist band and a grounded work
surface. Adequate service tools must
1.5 Functional check always be used.

After any installation, maintenance or service


intervention in the system, perform a Functional check. Note: ESD sensitive components.
The Functional check is divided into two steps: PC boards (spare parts) must always
be kept in a package for sensitive
1. Internal leakage test: The Internal leakage test
electronic devices. Maquet will not
must be performed without O2 connected to the
otherwise assume responsibility for the
system. The test can be performed either as a materials used, the work performed or
separate Internal leakage test in Service & any possible consequences of same.
Settings or by running a Pre-use check. The
Internal leakage test must be successfully passed
before starting step 2 of the Functional check. Note: Special waste. This system
Note: If running a Pre-use check, make sure that contains electronic and electrical
the Internal leakage test was successfully passed. components. Discard disposable,
Other tests can be cancelled as they may fail replaced and left-over parts in
when O2 is not connected. accordance with appropriate industrial
and environmental standards.
2. Pre-use check: Connect O2 and perform a
complete Pre-use check according to instructions
in the User’s Manual. Note: Recyclable material. Recycling
must be performed in accordance with
appropriate industrial and
environmental standards.

Note: Trained and authorized


personnel. The system must be
serviced at regular intervals by
personnel trained and authorized by
Getinge. It is recommended that service
and maintenance is done as a part of
a service contract. Any maintenance
or service must be noted in a log book.

SERVO-air Ventilator System, Service Manual 1-3


| 1 | Important |

Note: All repair or service requests resulting from


1.8 Environmental declaration
a device failure to perform their intended function 1.8.1 Purpose
shall be submitted as complaints.
Please report a potential complaint to your local This environmental declaration applies to a SERVO-air
Getinge representative as soon as possible for basic system including the Mobile cart and one Battery
further processing. module.

Maintenance of the system must be performed by 1.8.2 Components with special environmental
Getinge authorized personnel with the following concern
intervals:
Components listed below shall be disposed of in
• Preventive maintenance at least once a year, or
bullet accordance with appropriate industrial and
every 5000 hours of operation, whichever comes environmental standards.
first.
Printed circuit boards
• The Battery modules shall be replaced after
bullet
300 charge cycles or four years (48 months) from
manufacturing date, whichever comes first. • PC 1781 Pressure transducer, 2 pcs
bullet

• The Memory backup batteries shall be replaced


bullet
• PC 1991 Control, including battery
bullet
every five years. • PC 1992 Monitoring, including battery
bullet
• The Turbine module shall be replaced every 10
bullet • PC 1993 Ethernet switch
bullet
years.
• PC 1994 Nebulizer
bullet

Only original spare parts from Maquet must be used • PC 2004/PC 2024 Expiratory channel
bullet
in the system. • PC 2020 Main backplane
bullet
• PC 2021 Rear connectors
bullet
1.7 To the responsible service personnel
• PC 2023 Panel
bullet

Note: Trained and authorized • PC 2025 Battery back-plane


bullet
personnel. Only personnel trained and • PC 2026 Battery extension
bullet
authorized by Getinge shall be
permitted to perform installation,
• PC 2027 Motor control
bullet

service or maintenance of the system. • PC 2028 DC/DC & Battery control


bullet
• PC 2029 Inspiratory channel extension
bullet
The contents of this document are not binding. If any • PC 2030 Inspiratory channel
bullet
significant difference is found between the system and
this document, please contact Maquet for further • PC 2031/PC 1785 Expiratory channel connector
bullet
information.
Other electronics
We reserve the right to modify the system without
amending this document or advising the user. • LCD display
bullet
• Touch panel (glass)
bullet
• Turbine module
bullet
• O2 cell, containing caustic lime and lead (Pb)
bullet
• O2 gas module, containing multiple PC boards
bullet

1-4 SERVO-air Ventilator System, Service Manual


| Important | 1 |

• Expiratory cassette, containing PC board and other


bullet • TPE (Thermoplastic elastomer)
bullet
electronics
Electronics – total 15%
• Expiratory valve coil
bullet
• Safety valve pull magnet
bullet • Printed circuit boards, cables etc.
bullet
• AC/DC converter, containing PC board
bullet
Battery – total 2%
• Battery module Li-ion
bullet
• Mains power supply connector
bullet • Battery module Li-ion
bullet

• Fans, 2 pcs
bullet Others – very small amounts
• Loudspeaker
bullet
• Cable harness (handle as electronic waste)
bullet • Filter paper of fiberglass
bullet

1.8.3 Construction materials 1.8.4 Articles of consumption

The construction materials used in the system, in % 1. Bacteria filter


of the total weight. 2. Dust filter for the Air inlet filter
3. Filter for the O2 gas module
Metal – total 70%
4. Filter for the Inspiratory pressure transducer
• Aluminium 40%
bullet 5. Nozzle unit for the O2 gas module

• Steel, zink, brass 30%


bullet 6. Air inlet filter for the Turbine module
7. Battery modules
Polymeric material – total 13%
8. Turbine module

• ABS (Acrylicnitrilebutadienstyrene)
bullet 9. Memory backup batteries
10. Expiratory cassette
• EPDM (Ethylenepropylenedienemonomer)
bullet
11. Expiratory cassette membrane
• FPM (Fluororubber)
bullet
12. O2 cell including filter
• LDPE (Low-density polyethylene)
bullet
Item 1: Consumption approx. 250 pcs/year.
• NBR (Nitrilerubber)
bullet
• PA (Polyamide)
bullet Item 2: Consumption approx. 15 pcs/year.

• PBT (Polybutylene terephthalate)


bullet Items 3 – 6: Replaced once a year or every 5000 hours
• PC (Polycarbonate)
bullet of operation, whichever comes first.

• PE (Polyethylene)
bullet Item 7: Replaced after 300 charge cycles or four years
• PEEK (Polyether ether ketone)
bullet (48 months) from manufacturing date, whichever
comes first.
• POM (Polyoxymethylene)
bullet
• PP (Polypropylene)
bullet Item 8: Replaced every 10 years.
• PTFE (Polytetrafluoroethylene)
bullet Item 9: Replaced every five years.
• PS (Polystyrene)
bullet
Items 10 – 12: Replaced when needed.
• PUR (Polyurethane)
bullet
• PVC (Polyvinyl chloride)
bullet Articles related to clinical applications, e.g., patient
tubing and nebulizer medication cups are not included
• SI (Silicone)
bullet
in the list above.
SERVO-air Ventilator System, Service Manual 1-5
| 1 | Important |

1.8.5 Power consumption 1.8.8 Product End-of-Life

The power consumption depends on the operating For disassembly and disposal (scrapping) information,
mode and Patient unit settings. refer to the document SERVO-air Ventilator System
Product End-of-Life Disassembly Instructions.
• Typical minimum power consumption (standby):
bullet
26 W

• Typical power consumption: 36 W


bullet
Pressure control (intended delivered volume 500
ml) with PEEP=5 cmH2O, PAP=10 cmH2O, RR=20
breaths/min, FiO2=30%. Pressure/volume settings,
respiratory rate, air/oxygen mix, leakages and/or
other factors will affect power consumption and
battery backup time. See section Battery modules
for approximate recharge times.

• Typical maximum power consumption: 185 W


bullet

1.8.6 Noise level

A-weighted sound pressure level (LpA): <49 dB,


measured at a distance of 1 m.

A-weighted sound power level (LWA): <57 dB.

1.8.7 Packaging material

Materials for packing:

• Loading pallet.
bullet
• Corrugated cardboard.
bullet
• Shock-absorbing material of expanded
bullet
polypropylene, EPP.

Materials added when packing for tropical climatic


conditions:

• Plywood container with metal reinforcements.


bullet
• Plastic wrapping.
bullet
• Desiccant (humidity absorber).
bullet

The packaging material complies with the demands


stated by the International Plant Protection Convention
(IPPC).

1-6 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2 Introduction

Table of Contents
2.1 Main units | 2-2
2.2 Patient unit | 2-4
2.3 User interface | 2 -10
2.4 Software structure | 2 -12
2.5 Internal communication | 2 -14

SERVO-air Ventilator System, Service Manual 2-1


| 2 | Introduction |

2.1 Main units


The system is configured for different patient The system can be divided into the following main
categories: units:
• Pediatric
bullet • Patient unit. The Patient unit contains pneumatics
bullet
and electronics for gas supply to the patient. Power
• Adult
bullet
supply and battery backup is also contained in the
These patient categories are as standard equipped
Patient unit.
with a number of ventilation modes suitable for each
category. Additional ventilation modes can be installed • User interface. The User interface contains all
bullet
via a Software option installation. controls used to set the ventilation and monitoring
parameters. Ventilation parameters as well as other
important information are shown on the User
interface display.
The User interface cable connects the User interface
and the Patient unit.

2.1.1 System mounted on the optional Mobile cart

f h

a g k l j

S20002

1. Patient unit 7. Battery compartments


2. User interface 8. Patient circuit
3. Expiratory inlet with moisture trap 9. AC mains power
4. Inspiratory outlet 10. O2 supply
5. Emergency air intake 11. Wheel lock
6. Air inlet for turbine

2-2 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.1.2 Optional accessories

There are also several accessories to be used with


the system, e.g.:

• Alarm output
bullet
• Battery module (extra)
bullet
• Battery charger
bullet
• Mobile cart
bullet
• Mobile cart drawer kit
bullet
• Cabinet (for Mobile cart)
bullet
• Gas cylinder restrainer kit
bullet
• Shelf base
bullet
• Support Arm 179
bullet
• Y piece holder and hook
bullet
• Cable holder
bullet
• Aerogen nebulizers (Aerogen Pro and
bullet
Aerogen Solo)

• Servo DuoGuard and Servo Guard filters


bullet
• Expiratory heater, Servo DuoGuard
bullet
• Thermoshell for expiratory cassette
bullet
• Humidifier holder
bullet
• Waterbag/IV pole for humidifier holder
bullet
• MSync (adapter for HL7 communication)
bullet

For further information regarding accessories, contact


your Getinge representative.

SERVO-air Ventilator System, Service Manual 2-3


| 2 | Introduction |

2.2 Patient unit


Items accessible from the outside of the Patient unit are shown in this illustration.

s
F G
d D K

f S H

h j k l ; A J

S20003

1. On/Off switch. The switch must be pulled 11. Battery compartments


downwards before it can be switched. 12. USB ports
2. Expiratory outlet 13. Gas inlet for O2
3. Power indicators 14. Gas inlet for air including air inlet filter
4. RS-232 connectors 15. Inspiratory outlet
5. Equipotentiality terminal 16. Emergency air intake
6. AC mains power supply connector with fuse 17. Nebulizer connector
7. Alarm output connector 18. Expiratory inlet
8. External +12 V DC inlet 19. Cooling fan air inlet with filter (on both sides)
9. Fuse for external DC power 20. Expiratory cassette
10. Ethernet connector 21. Expiratory inlet with moisture trap

2-4 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.1 Main section

The main section is accessed by removing the top cover and the lid on the PC board rack.

s
g
d

f
h

j
S20004

No. Description Remark


1 PC 1993 Ethernet switch Part of subsystem Monitoring MON. Controls the Ethernet communication
in the system.
2 PC 1992 Monitoring Part of subsystem Monitoring MON. Main functions are alarm and
monitoring, calculations, trending of measured values and is also CAN-bus
master. Includes a Memory backup battery.
3 PC 1991 Control Part of subsystem Breathing BRE. Main functions are responsibility for
the patient treatment, i.e. how to regulate the inspiratory and expiratory
gas flow. Includes a Memory backup battery.
4 PC 2004 Expiratory channel/ Part of subsystem Breathing BRE. Main functions are expiratory flow
PC 2024 Expiratory channel measurement and expiratory cassette handling. PC 1781 Expiratory
pressure transducer is connected onto this board. PC 2024 replaces
PC 2004. Introduced in production Q1 2016.
5 Expiratory valve coil connector Cable connector for the Expiratory valve coil.
6 PC 1781 Expiratory pressure Expiratory pressure transducer connected to the Expiratory cassette with
transducer the Expiratory pressure tube.
7 Expiratory channel connector cable Cable connector for PC 2031/PC 1785 Expiratory channel connector.

SERVO-air Ventilator System, Service Manual 2-5


| 2 | Introduction |

2.2.2 PC 2020 Main back-plane

S20005

No. Description Remark


1 PC 2020 Main back-plane The PC boards, as listed above, are directly or indirectly connected to
the PC 2020 Main back-plane.

2-6 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.3 Inspiratory section

The left-hand side of the Patient unit contains the Inspiratory section.

d f

s
g

l
S20006

No. Description Remark


1 Turbine module The Turbine module generates compressed air and delivers air to the
inspiratory gas.
2 O2 gas module The O2 gas module is connected to the O2 supply and delivers O2 to the
inspiratory gas.
3 Inspiratory gas docking Connects the Turbine module and the O2 gas module. Contains the
Turbine module one-way valve.
4 Inspiratory flowmeter The inspiratory flowmeter measures the combined air and O2 flow, as
well as the temperature, of the inspiratory gas from the O2 gas module
and Turbine module.
5 Inspiratory flowmeter cable Cable connector for the Inspiratory flowmeter.
6 PC 1781 Inspiratory pressure Inspiratory pressure transducer connected to the inspiratory pipe with
transducer the inspiratory pressure tube.
7 O2 cell cable Cable connector for the O2 cell.
8 PC 2030 Inspiratory channel Connector board for the inspiratory section. PC 1781 Inspiratory pressure
transducer is connected onto this board.
9 O2 evacuation fan Forces cooling air through the Inspiratory section and evacuates O2 gas
in case of O2 leakage.

SERVO-air Ventilator System, Service Manual 2-7


| 2 | Introduction |

2.2.4 AC/DC section

The right-hand side of the Patient unit contains the AC/DC converter and Mains power inlet, as well as the
Expiratory valve coil.

a s

d
g

S20007

No. Description Remark


1 AC/DC Converter Converts the connected AC Power (mains power) to the internal DC
supply voltage.
2 Expiratory valve coil Controls the opening of the expiratory valve by pushing the valve
membrane in the cassette into desired position.
3 PC 2028 DC/DC & Battery control The main functions are On/Off control, control of the battery modules,
control of fans and connection to the User interface. It also supplies
required internal voltages.
4 Main fan Forces cooling air through the AC/DC section and the main section of
the Patient unit.
5 Mains power supply inlet Mains power connector including fuses.

2-8 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.2.5 Expiratory cassette

a s

S20008

No. Description Remark


1 Expiratory cassette
2 PC 2031/PC 1785 Expiratory channel connector PC 2031 must be used together with PC 2004.
PC 1785 must be used together with PC 2024.

The expiratory cassette is a complete unit. It contains


the following parts:

• Expiratory inlet with moisture trap.


bullet
• PC 1786 Expiratory channel cassette.
bullet
• Ultrasonic flowmeter.
bullet
• Heating foil to avoid condensation in the Ultrasonic
bullet
flowmeter.

• Pressure transducer connection, incl. bacteria filter,


bullet
to connect the expiratory pressure transducer.

• Expiratory valve incl. valve membrane.


bullet

PC 1786 Expiratory channel cassette inside the


Expiratory cassette is electrically connected to
PC 2004/PC 2024 Expiratory channel via
PC 2031/PC 1785 Expiratory channel connector.

The Expiratory valve coil, mounted under the


Expiratory cassette compartment, controls the valve
membrane in the cassette.

SERVO-air Ventilator System, Service Manual 2-9


| 2 | Introduction |

2.3 User interface


The User interface is mounted onto the Patient unit
using a 2-joint swivel arm.

The User interface can be tilted up and down into a


suitable position.

For information regarding operation of the User


interface, refer to the User's Manual.

S20009

The illustration below shows the backside of the User The illustration below shows the inside of the User
interface after it is detached from the User interface interface rear cover.
holder.

a s d S20011

No. Description Remark


1 Loudspeaker grid
2 Loudspeaker connector
3 Loudspeaker The loudspeaker is attached with double-coated adhesive tape.

2 - 10 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

a s

S20012

No. Description Remark


1 Alarm indicator Multicolor LEDs on PC 2023 will flash when an alarm is triggered in the
system
2 PC 2023 Panel Part of subsystem Panel PAN.
3 LCD Display/Touch screen Complete unit including LED backlight.

SERVO-air Ventilator System, Service Manual 2 - 11


| 2 | Introduction |

2.4 Software structure 2.4.2 Breathing BRE

2.4.1 General The Breathing subsystem’s main function is the patient


treatment, i.e. how to regulate and measure the
The System software will contain all available system inspiratory and expiratory gas flow:
functionality. The software is distributed across the
subsystems. The installation procedure ensures that • Controls all ventilation modes, ventilator settings
bullet
all subsystems are properly updated. and delivery of gas mixtures.

The System software is divided into the following main • Controls the expiratory cassette handling including
bullet
subsystems: flow measurement performed by the flowmeter in
the cassette.
• Breathing BRE
bullet • Nebulizer control. Controls the Nebulizer function.
bullet
• Monitoring MON
bullet • Controls Pre-use check operations on the Breathing
bullet
• Panel PAN
bullet subsystem

• Power POW
bullet • Maintain a persistent data storage area used to
bullet
save parameter settings.
• Blower BLO
bullet
• System ID
bullet
• Controls some of the system settings, e.g. storing
bullet
default patient category and default mode of
ventilation. The system settings will be stored in
persistent memory.
User interface The Breathing software is stored on PC 1991 Control
PC 2023 and PC 2004/PC 2024 Expiratory Channel. The
software must be re-installed if PC 1991 or
PAN SW
PC 2004/PC 2024 is replaced.

The Breathing software is executed by the


microprocessors on PC 1991 and PC 2004/PC 2024.
PC 1994
PC 2028 POW SW

Patient unit
PC 1786

PC 2029

PC 2020 SYSTEM ID SW PC 2030


PC 2004/ PC 1781Exp

PC 1781
MON SW
BRE SW

Insp.

PC 2031/
PC 1785
PC 1993
PC 1991

PC 1992
PC 2024

BLO SW
PC 2026

PC 2027
PC 2033 (Turbine
module)

PC 2021 PC 2025

S20135

2 - 12 SERVO-air Ventilator System, Service Manual


| Introduction | 2 |

2.4.3 Monitoring MON 2.4.5 Power POW

The Monitoring subsystem’s main function is alarm The Power subsystem has the following main
and monitoring, including trends of measured values: responsibilities:

• Detects and handles alarms. Contains a backup


bullet • Communication with plug-in battery modules to
bullet
sound buzzer. determine battery status.

• Monitors the system behavior.


bullet • Controlling management of, and voltage to, the
bullet
Inspiratory channel.
• Maintains a persistent data storage area to save
bullet
Alarm settings. • Controlling charging/discharging of plug-in battery
bullet
modules.
• Monitors some of the system settings, e.g. default
bullet
alarm limits. • Selection of power source to be used for the
bullet
system.
• CAN communication with Expiratory channel.
bullet
• Battery module and power source supervision.
bullet • Controlling Main fan and O2 evacuation fan.
bullet
The Power software is stored on PC 2028. The
• Pre-use check: Pre-use check operations on the
bullet
software must be re-installed if PC 2028 is replaced.
Monitoring subsystem

• External communication protocol SCI (SERVO


bullet The Power software is executed by the microprocessor
Communication Interface). on PC 2028.

• Optional functionality such as ventilation mode and


bullet
other features. All options are stored in the 2.4.6 Blower BLO
persistent memory.
The Blower subsystem is responsible for the
The Monitoring software is stored on PC 1992 infrastructure, hardware, and services for control of
Monitoring. The software must be re-installed if the blower. It also supplies the BRE and MON
PC 1992 is replaced. subsystems with sensor data for inspiratory flow and
temperature as well as ambient air humidity and air
The Monitoring software is executed by the
inlet filter pressure drop:
microprocessor on PC 1992.

2.4.4 Panel PAN


• Controlling and monitoring the Turbine module.
bullet
• Managing flow measurement for the Inspiratory flow
bullet
The Panel subsystem controls all user interaction, meter.
e.g.:
• Managing ambient air temperature and air humidity
bullet
measurements.
• Presentation and changes of ventilator settings:
bullet
• Enable Air inlet filter supervision by measuring air
bullet
• Presentation of metrics, e.g. real-time values from
bullet inlet filter pressure drop.
the Monitoring subsystem.
The Blower software is stored on PC 2027 Motor
• Presentation of alarms and their corresponding
bullet controller board. The software must be re-installed if
alarm sounds from all subsystem. Turbine module is replaced.
The Panel software is stored on PC 2023. The
software must be re-installed if PC 2023 is replaced.

The Panel software is executed by the microprocessor


on PC 2023.

SERVO-air Ventilator System, Service Manual 2 - 13


| 2 | Introduction |

2.4.7 System ID 2.5 Internal communication


The System ID is stored on PC 2020 Main back-plane The internal communication is separated into different
and is unique for each system. communication protocols:

When installing software options, the System ID is • ETHERNET bus using TCP/IP running at
bullet
used to verify that the option is valid for that system. 100 Mbit/s. ETHERNET connects the main
subsystems BRE, MON and PAN using the Ethernet
When replacing PC 2020 Main back-plane, the
switch.
replacement (spare part) must be factory programmed
with the correct System ID to ensure proper ventilation • CAN bus (Controller Area Network) running at
bullet
performance, compatibility with future updates and 500 kbit/s. The CAN is primarily used for the
correct options configuration. Expiratory flowmeter.

• SPI bus (Serial Peripheral Interface) primarily used


bullet
for communication between BRE and BLO.

• I2C bus (Inter-Integrated Circuit) primarily used for


bullet
communication with:
- Local E2PROMS.
- The O2 cell.
- The expiratory cassette.
- The inspiratory flowmeter
- The ambient temperature and humidity sensor.

• Digital and analogue point-to-point signals


bullet
connecting subsystems and components.

• The RS-232 is used for the SERVO Communication


bullet
Interface SCI.

• The RS-485 is used for communication between


bullet
the POW and the MON subsystems.

• The SMBus (System Management Bus) is used for


bullet
communication with the POW subsystem and the
battery modules.

2 - 14 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3 Description of functions

Table of Contents
3.1 O2 gas module (1) | 3-4
3.2 Turbine module including PC 2027 Motor | 3-6
control (2)
3.3 Inspiratory gas docking (3) | 3-7
3.4 Inspiratory pipe (4) | 3-7
3.5 Inspiratory flowmeter (5) | 3-7
3.6 Inspiratory channel (6) | 3-7
3.7 O2 cell (7) | 3-8
3.8 Safety valve including pull magnet (8) | 3-8
3.9 O2 evacuation fan (9) | 3-9
3.10 PC 2030 Inspiratory channel (10) | 3-9
3.11 PC 1781 Inspiratory pressure transducer (11) | 3 -10
3.12 PC 2029 Inspiratory channel extension (12) | 3 -10
3.13 Expiratory cassette (13) | 3 -10
3.14 Expiratory valve (14) | 3 -11
3.15 Expiratory valve coil (15) | 3 -11
3.16 Expiratory outlet (16) | 3 -11
3.17 PC 1786 Expiratory channel cassette (17) | 3 -11
3.18 PC 2031/PC 1785 Expiratory channel | 3 -11
connector (18)
3.19 PC 2020 Main back-plane (19) | 3 -12
3.20 PC 1781 Expiratory pressure transducer (20) | 3 -12
3.21 PC 2004/PC 2024 Expiratory channel (21) | 3 -12
3.22 PC 1991 Control (22) | 3 -13
3.23 PC 1992 Monitoring (23) | 3 -13
3.24 PC 1993 Ethernet switch (24) | 3 -14
3.25 PC 1994 Nebulizer (25) | 3 -14
3.26 Mains inlet (26) | 3 -14
3.27 AC/DC Converter (27) | 3 -14
3.28 PC 2028 DC/DC & Battery control (28) | 3 -14
3.29 On/Off switch (29) | 3 -15
3.30 PC 2021 Rear connectors (30) | 3 -15
3.31 PC 2026 Battery extension (31) | 3 -16
3.32 PC 2025 Battery back-plane (32) | 3 -16
3.33 Battery module (33) | 3 -16
3.34 Main fan (34) | 3 -17

SERVO-air Ventilator System, Service Manual 3-1


| 3 | Description of functions |

3.35 User interface cable (35) | 3 -17


3.36 Touch screen (36) | 3 -17
3.37 LCD Display (37) | 3 -17
3.38 PC 2023 Panel (38) | 3 -17
3.39 Alarm indicator (39) | 3 -18
3.40 Loudspeaker (40) | 3 -18
3.41 Optional equipment | 3 -18

3-2 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

About this chapter The System electronics contains all PC boards and
electrical components that are not included in the other
The information in this chapter refers to the System main blocks:
overview diagram in chapter Diagrams. Note the
numbering in each heading (in brackets), that 19. PC 2020 Main back-plane
corresponds to the position number in the System 20. PC 1781 Expiratory pressure transducer
overview diagram.
21. PC 2004/PC 2024 Expiratory channel
22. PC 1991 Control
Main blocks
23. PC 1992 Monitoring
The information in this chapter is divided into main 24. PC 1993 Ethernet switch
blocks:
25. PC 1994 Nebulizer
The Inspiratory section conveys the breathing gas 26. Mains inlet
from the gas inlets to the patient breathing system. It 27. AC/DC Converter
contains the following main components: 28. PC 2028 DC/DC & Battery control
1. O2 gas module 29. On/Off switch
2. Turbine module 30. PC 2021 Rear connectors
3. Inspiratory gas docking 31. PC 2026 Battery extension
4. Inspiratory pipe 32. PC 2025 Battery back-plane
5. Inspiratory flowmeter 33. Battery module
6. Inspiratory channel 34. Main fan
7. O2 cell 35. User interface cable

8. Safety valve including pull magnet


9. O2 evacuation fan
The User interface is the main user interface during
10. PC 2030 Inspiratory channel operation of the system. The User interface contains
11. PC 1781 Inspiratory pressure transducer the following main components:
12. PC 2029 Inspiratory channel extension
36. Resistive Touch screen.
37. LCD Display, 12” TFT display with LED backlight.

The Expiratory section conveys the breathing gas 38. PC 2023 Panel
from the patient breathing system to the Expiratory 39. Illuminated alarm indicator on top of the User
outlet. It contains the following main components: interface frame.
40. Loudspeaker
13. Expiratory cassette
14. Expiratory valve
15. Expiratory valve coil
16. Expiratory outlet
17. PC 1786 Expiratory channel cassette (integrated
into the Expiratory cassette)
18. PC 2031/PC 1785 Expiratory channel connector

SERVO-air Ventilator System, Service Manual 3-3


| 3 | Description of functions |

Memory types used in the system 3.1 O2 gas module (1)


There are four different types of memories used in the The O2 gas module regulates the inspiratory O2 gas
system: flow.

• Flash memory. For System software storage.


bullet
g f d
Present on PC 1991, PC 1992, PC 2004/PC 2024,
PC 2023, PC 2027, PC 2028. The System software
can be re-installed/updated using a System
h
software installation.
s
• RAM. For temporary storage of software and data.
bullet
Present on PC 1991, PC 1992, PC 2004/PC 2024,
PC 2023, PC 2027, PC 2028. a
j
• RAM with battery backup. For settings, trends
bullet
and logs. Present on PC 1991 and PC 1992.

• EEPROM. For PC board information, configuration,


bullet k S20014
calibration data, etc. Present on several PC boards,
and in the O2 cell. 1. Gas inlet
2. Filter
3. Inspiratory valve temperature sensor
4. Supply pressure transducer
5. Flow transducer (Delta pressure transducer and
wire mesh)
6. Nozzle unit with valve diaphragm
7. Inspiratory solenoid
8. Gas module key

Note: The O2 gas module is factory calibrated. The


O2 gas module must not be disassembled further
than described in chapter Preventive maintenance.

3.1.1 Gas inlet

O2 gas supply is connected to the system’s O2 inlet


nipple. The design of the O2 gas inlet nipple varies
according to the standard chosen.

O2 gas is to be connected from hospital central gas


supply or from an O2 gas cylinder.

Refer to Technical data in the User's Manual for O2


gas quality specifications.

3-4 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.1.2 Filter The complete plastic nozzle unit must be replaced


during the Preventive maintenance. After replacement,
The Filter protects the system from particles in the gas allow the diaphragm to adapt to the valve seat by the
delivered to the O2 gas module. The Filter must be spring tension during approximately 10 minutes before
replaced during the Preventive maintenance. gas pressure is connected to the O2 gas module.

A non-return valve for the O2 gas inlet is located in the


filter cover. This valve will suppress short pressure 3.1.7 Inspiratory solenoid
drops in the O2 gas supply.
The gas flow through the O2 gas module is regulated
The non-return valve is also designed to slowly by the Inspiratory solenoid via the Nozzle unit.
evacuate compressed O2 gas from the module, if the
The current supplied to the solenoid is regulated so
O2 gas supply to the O2 gas module is disconnected. that the O2 gas module will deliver a gas flow
according to the settings on the User interface.
3.1.3 Inspiratory valve temperature sensor
3.1.8 Gas module key
The temperature of the supplied O2 gas is measured
by the Inspiratory valve temperature sensor. This The O2 gas module is provided with a mechanical key
sensor is situated in the gas flow. to prevent that gas modules adapted for other gases
are connected to the system.
The output signal from this sensor is used to
compensate for the O2 gas density variations due to The key consists of a plastic guide mounted
temperature. underneath the module and a corresponding guide
mounted in the Patient unit.
3.1.4 Supply pressure transducer
3.1.9 ID PROM
The pressure of the supplied O2 gas is measured by
the Supply pressure transducer. The O2 gas module is provided with an ID-PROM. The
ID information can be read by the system.
The output signal from this transducer is amplified. It
is then used to calculate the absolute pressure of the
O2 gas to compensate for gas density variations due
to pressure.

3.1.5 Flow transducer

The O2 gas flows through a wire mesh (resistance)


which causes a pressure drop. The pressure is
measured on both sides of this net and the differential
pressure value is then amplified.

3.1.6 Nozzle unit

The plastic Nozzle unit contains a valve diaphragm.


The valve diaphragm, controlled by the Inspiratory
solenoid, regulates the gas flow through the O2 gas
module.

SERVO-air Ventilator System, Service Manual 3-5


| 3 | Description of functions |

3.2 Turbine module including PC 2027 Motor 7. Nipple for air inlet pressure measurement for
control (2) supervision of pressure drop over the Air inlet
filter.
3.2.1 Turbine
The Turbine module shall be replaced every 10 years.
The Turbine generates the inspiratory air gas flow
(max. 240 l/min). It generates the air flow and pressure Note: The Turbine is factory calibrated. The Turbine
from the surrounding air. This air flow is then mixed must not be disassembled.
together with the O2 flow from the O2 gas module.

The maximum Turbine output pressure is dependent 3.2.2 PC 2027 Motor control
on the ambient pressure and may therefore vary
PC 2027 Motor control’s main purpose is to control
between 65 cmH2O (high altitude) and
and supervise the Turbine:
approx. 100 cmH2O (sea level).
• Powering the Turbine. The power to the motor driver
bullet
s g is supplied through a step-up regulator that elevates
the voltage from the system voltage (approximately
11-17 V) to approximately 28 V. The energy created
when the motor is decelerated is stored in a
a capacitor buffer to enable quick acceleration.
Further excess energy will be delivered to other
parts of the system.
h
X T • Data transferred to BRE via SPI communication for
bullet
further dispatching to MON:
P1 PC 2027 Motor control P - Inspiratory flow and gas temperature from the
Inspiratory flowmeter.
d f h j - Ambient air temperature and humidity from
S20103 sensors on PC 2030.
- Air inlet filter pressure drop measured by a
The Turbine module includes:
differential pressure sensor. Will indicate if the
1. Brushless DC motor powered by a motor driver Air inlet filter is occluded or clogged. Will also
on PC 2027 Motor control. Maximum speed detect if the Air inlet filter is missing during
70 000 r/min. Pre-use check.

2. Turbine wheel. • Passing data from the ambient air inlet filter
bullet
differential pressure sensor.
3. Hall effect sensors to measure motor speed and
phase.
4. Temperature sensor to monitor the Turbine 3.2.3 Air inlet filter including dust filter
temperature. The reading is used to detect
The airflow to the Turbine, and to the patient, is
overheating and to log running conditions.
protected by an Air inlet filter. The Air inlet filter
5. Silencing foam and rubber suspensions to consists of a HEPA (High Efficiency Particulate Air)
minimize vibrations and sound level. filter and a dust filter.
6. Cooling air flow for the motor. Routed from the
Turbine unit out through the Patient unit via a The Air inlet filter must be replaced during the
hose. Preventive maintenance. Check the dust filter and
replace it once a month or more often if needed.

3-6 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.3 Inspiratory gas docking (3)


The Inspiratory gas docking connects the O2 gas
module, and Turbine module, to the Inspiratory
flowmeter connector muff. The Inspiratory gas docking Gas flow
also includes:
T T
• The Gas module silicone muff connects the O2 gas
bullet
module to the Inspiratory gas docking. a s d s
S20102
• The Turbine module one-way valve that prevents
bullet
gas flow backwards through the Turbine module. 1. Flow conditioner
2. Temperature sensors
3.4 Inspiratory pipe (4) 3. Heater

The Inspiratory pipe leads the gas from the Inspiratory The flowmeter is equipped with a flow conditioner on
gas docking to the Inspiratory channel. the inlet to create a laminar flow to enhance the
accuracy.
The Inspiratory pipe contains:
3.6 Inspiratory channel (6)
• Inspiratory flowmeter connector muff.
bullet
The Inspiratory channel connects the Inspiratory pipe
• Inspiratory flowmeter.
bullet
to the patient breathing system. The Inspiratory
• Inspiratory channel connector muff.
bullet channel contains:
The Inspiratory pipe is equipped with a flow conditioner
ahead of the flowmeter, to ensure that the flowmeter • The Inspiratory pressure tube that connects the
bullet
makes reliable measurements. Inspiratory channel with the Inspiratory pressure
transducer. The tube is connected to the bacteria
3.5 Inspiratory flowmeter (5) filter that protects the pressure transducer PC 1781
Inspiratory pressure transducer from contamination.
The Inspiratory flowmeter (inspiratory flow transducer) The bacteria filter must be replaced during
measures the combined air and O2 flow, as well as Preventive maintenance.
temperature, of the inspiratory gas from the O2 gas
module and Turbine module. The flow measurements
• The Safety valve membrane.
bullet

are used as input to control the gas module and the • Housing for the O2 cell.
bullet
Turbine, creating a feedback loop. • The Inspiratory outlet, a 22 mm/15 mm tube
bullet
connector for the Inspiratory tube of the patient
The flowmeter uses a thermal measuring principle, in breathing system.
this case a differential temperature measurement. The
The Inspiratory channel can easily be removed using
basic principle is that two temperature sensors, one
the quick coupling.
mounted before and one after a small heater in the
flow direction:

• At no flow through the flowmeter, the temperature


bullet
will be equal on both temperature sensors.

• With a flow through the flowmeter, the temperature


bullet
on the two temperature sensors will differ,
consequently the higher flow the bigger temperature
difference.

SERVO-air Ventilator System, Service Manual 3-7


| 3 | Description of functions |

3.7 O2 cell (7) 3.8 Safety valve including pull magnet (8)

The O2 concentration is measured by an O2 cell. The The movable axis of the Safety valve pull magnet
O2 cell is mounted in a housing on the Inspiratory controls the opening and closing of the Safety valve
channel and is protected by a bacteria filter. membrane in the Inspiratory channel. The pull magnet
is controlled (electrically activated = closed) by
O2 PC 2004/PC 2024 Expiratory Channel.
O2 gas module

O2
Air
O2 cell O2 gas module
Turbine module

Air
Turbine module
S20015

Maintenance, including exchange of bacteria filter, is


performed according to the User´s Manual.

The O2 cell gives an output voltage proportional to the


partial pressure of oxygen inside the Inspiratory pipe.
At constant ambient pressure this output is
a s
proportional to the O2 concentration in percent.

The life time of the cell is affected by the O2


concentration. With a concentration (at the cell) in % S20016

and expected cell life time in hours the following


applies at 25 ºC (77 ºF): 1. During startup, the pull magnet is electrically
activated and the pull magnet axis is pushed up,
Expected cell life time = O2 Conc (%) x Time (h) creating a clicking sound. The Safety valve is now
= 500 000 (%hours) closed. This is the normal operating mode, when
the pull magnet is electrically activated and the
The O2 cell is automatically calibrated each time a
safety valve membrane is closed.
Pre-use check is performed (if O2 is connected to the
2. When the Safety valve is not electrically activated,
system).
the weight of the pull magnet axis combined with
If the system has been in use continually for a long the design of the valve membrane causes the pull
time, the measured O2 concentration may drop due magnet axis to be pulled down to the bottom
to normal degradation of the O2 cell. This will activate position. When the pull magnet axis is in the
bottom position, the Safety valve is open and the
a nuisance alarm. For further information, refer to the
inspiratory gas is released from the Inspiratory
User's Manual.
pipe via the Emergency air intake resulting in
An ID PROM is integrated into each O2 cell. Its ID decreased inspiratory pressure. The procedure
information and remaining lifetime can be read by the described is a normal safety (pop-off) function.
system. The Safety valve is opened when:

Notes: • The ventilator is switched Off or to Standby.


bullet
• Perform a Pre-use check to calibrate the O2 cell.
bullet • The pressure inside the inspiratory pipe is 5 cmH2O
bullet
above the preset Upper pressure alarm limit. This
• The O2 Sensor cannot be installed in SERVO-air.
bullet
condition is controlled by the Monitoring subsystem.

3-8 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

• The pressure inside the inspiratory pipe is 7 cmH2O


bullet The O2 evacuation fan air inlet (1) is located at the
above the preset Upper pressure alarm limit. This bottom of the side cover. The air outlets (2) are located
condition is controlled by the Breathing subsystem. above and below the Air inlet filter on the back of the
Patient unit. The air inlet is protected by a filter that
• The pressure inside the inspiratory pipe is above
bullet
must be cleaned or replaced during the Preventive
117 ±7 cmH2O. This is an extra safety function and
maintenance.
this situation does not normally occur.

• The safety valve also opens by other alarms, for


bullet The O2 evacuation fan is monitored by a temperature
example the Airway pressure continuously sensor on PC 2030, and controlled by PC 2028.
high-alarm.
The O2 evacuation fan is always running:
The safety valve opening pressure is calibrated to
117 ±3 cmH2O during each Pre-use check. • When the system is turned on. This is to make sure
bullet
that no O2 gas is present in the Inspiratory section
The Emergency air intake (Safety valve outlet) is while the power is on in case of O2 leakage.
covered by a plastic grid. The Emergency air intake
must be checked for obstructions during the Preventive
• When the batteries are charging, and for 2.5 hours
bullet
after the charging is complete.
maintenance.
Depending on the temperature conditions, the fan can
3.9 O2 evacuation fan (9) also run when the system is turned off but connected
to mains power supply.
The O2 evacuation fan has two purposes:
The fan always runs on half effect, and with full effect
• Force cooling air through the Inspiratory section.
bullet at approx. 43 °C (109 °F). When the temperature drops
• If needed, evacuate O2 gas from the Inspiratory
bullet below approx. 37 °C (99 °F), the fan turns to half
section, to prevent O2 gas from reaching too high effect.
concentrations.
3.10 PC 2030 Inspiratory channel (10)
The cooling air flow inside the Inspiratory section is
Interconnecting board including connectors for:
indicated in the System overview.

• Turbine module
bullet
• O2 gas module
bullet
• O2 cell
bullet

s • O2 evacuation fan
bullet
• Inspiratory flowmeter
bullet
• Safety Valve
bullet

PC 2030 Inspiratory channel is also connected to


a PC 2020 Main back-plane via the PC 2029 Inspiratory
channel extension for communication with components
S20017
in the center part of the Patient unit.
1. O2 evacuation fan inlet
The ambient air temperature and humidity sensor is
2. O2 evacuation fan outlet located on PC 2030. The gas temperature is used as
input to volume calculations and for supervision of the
temperature delivered to the patient.

SERVO-air Ventilator System, Service Manual 3-9


| 3 | Description of functions |

3.11 PC 1781 Inspiratory pressure 1. Ultrasonic transducer


transducer (11) 2. PC 1786 Expiratory channel cassette

PC 1781 Inspiratory pressure transducer is mounted 3. Bacteria filter


onto PC 2030 Inspiratory channel. 4. Expiratory pressure tube connector
5. Heating foil
The pressure, conveyed via the pressure tube
connected to this block, is led to and measured by its
differential pressure transducer. With differential 3.13.1 Expiratory inlet
reference to the ambient pressure, the output signal
A 22 mm/15 mm tube connector for the patient
is proportional to the measured pressure thus giving
breathing system’s Expiratory tube. The inlet is
a linear measurement in the range -40 cmH2O to
designed to make condensed water drip out and allow
+160 cmH2O.
use of a water trap for such water to be collected.
Technical limitation: Pressures exceeding ±400 cmH2O Expiratory inlet bacteria filter can be connected to
protect the cassette from contamination.
must be avoided.

Includes an ID PROM. The ID information can be read 3.13.2 Heating foil


by the system.
An electrical Heating foil is applied on the outside of
the Expiratory pipe where the Ultrasonic flowmeter is
3.12 PC 2029 Inspiratory channel extension
situated. The purpose of the Heating foil is to reduce
(12)
condensation and maintain a stable temperature in
PC 2029 Inspiratory channel extension is a connector the expiratory gas.
board. It connects PC 2030 Inspiratory channel to
PC 2020 Main back-plane. There are no active 3.13.3 Ultrasonic flowmeter
components on PC 2029.
The Ultrasonic flowmeter (expiratory flow transducer)
is a measuring device for the expiratory gas flow, using
3.13 Expiratory cassette (13)
ultrasound technique with two ultrasonic
The expiratory gas conveying parts and PC 1786 transducers/receivers. The measuring process is
Expiratory Channel Cassette are integrated into one controlled from PC 2004/PC 2024 Expiratory channel.
part – the Expiratory Cassette – which can be easily
removed for cleaning or exchange, see the User’s
Manual.

The Expiratory cassette can be interchanged between


different systems. A Pre use check is always required
after exchanging the Expiratory cassette.

a s a

S20019

g d f The left hand side transducer is sending out an


S20018
ultrasonic sound that is reflected against the inner wall
of the Expiratory channel. The ultrasonic sound is
3 - 10 SERVO-air Ventilator System, Service Manual
| Description of functions | 3 |

received by the right hand side transducer now acting 3.15 Expiratory valve coil (15)
as a receiver. The time from sending to receiving
ultrasonic sound in downstream expiratory gas flow The movable axis of the Expiratory valve coil controls
is measured. the opening of the Expiratory valve by pushing the
valve membrane into desired position. The power
Then the right hand side transducer (earlier receiving) supply to the coil is regulated so that the remaining
is sending out ultrasonic sound upstream the pressure in the patient system, towards the end of the
expiratory gas flow. The ultrasonic sound is received expiration time, is kept on the PEEP level according
by the left hand side transducer now acting as a to User interface setting.
receiver. The time from sending to receiving ultrasonic
sound in upstream expiratory gas flow is measured. 3.16 Expiratory outlet (16)
The time difference between the downstream and the The gas from the patient system leaves the system
upstream time measurements provides flow via this Expiratory outlet. A thermo sleeve can be
information. attached to the outlet to reduce condensation.

A temperature sensor inside the cassette measures


the expiratory gas temperature. This temperature
3.17 PC 1786 Expiratory channel cassette
measurement is also used when calculating the
(17)
expiratory flow. The PC 1786 Expiratory channel cassette is a
connection board, integrated into the Expiratory
3.13.4 Bacteria filter and expiratory pressure tube cassette, for the Ultrasonic flowmeter and for the
Heating foil. It connects to PC 2031/PC 1785 mounted
Via a Bacteria filter inside the cassette, the Expiratory
in the Expiratory cassette compartment.
pressure tube connects the cassette to the Expiratory
pressure transducer. The filter and the connector are Includes an ID PROM. The ID information can be read
integrated parts of the cassette. The filter protects the by the system.
transducer on PC 1781 Pressure transducer from
contamination. 3.18 PC 2031/PC 1785 Expiratory channel
connector (18)
3.14 Expiratory valve (14)
Note: PC 1785 replaces PC 2031. Introduced in
The Expiratory valve consists of a membrane in the
production during Q1 2016.
cassette that is operated by the axis of the Expiratory
valve coil. The valve is fully open as long as no power
is supplied to the coil. The PC 2031/PC 1785 Expiratory channel connector
is a connector board, including signal filters (and a
Operating capacity for the membrane is estimated to power regulator on PC 2031). It is mounted in the
10 000 000 breathing cycles. When this limit is passed Expiratory cassette compartment, and connects to
or if the membrane for some reason has become PC 1786 mounted in the Expiratory cassette when the
defective, it must be replaced. Refer to instructions in cassette is docked to the Expiratory cassette
chapter Disassembling and assembling. compartment.

Membrane capacity (counter) can be shown in the For software and hardware requirements and
System status window. Select SYSTEM STATUS > compatibility, see the Compatibility chart in chapter
Expiratory cassette to check Membrane capacity. The Revision history.
operating capacity meter must be reset after
replacement of the membrane. Note that the reset
function is not implemented in System version 1.0.

SERVO-air Ventilator System, Service Manual 3 - 11


| 3 | Description of functions |

3.19 PC 2020 Main back-plane (19) 3.21 PC 2004/PC 2024 Expiratory channel
(21)
PC 2020 is the interconnection board for PC boards
and components.
Note: PC 2024 replaces PC 2004. Introduced in
PC 2020 Main back-plane is also connected to production during Q1 2016.
PC 2030 Inspiratory channel via the PC 2029
Inspiratory channel extension for communication with PC 2004/PC 2024 Expiratory channel is part of
components in the Inspiratory section. subsystem Breathing BRE. The main functions are
expiratory flow measurement and expiratory cassette
The System ID (Serial No.), operating time, etc., is handling. Expiratory flow measurement is performed
stored in an EEPROM on PC 2020. Thus, when by the flowmeter in the cassette. PC 2004/PC 2024
replacing PC 2020, a spare part that is factory is connected to the cassette via PC 2031/PC 1785
programmed for the concerned system must be used. Expiratory channel connector.

As the Preventive maintenance time stamp will be lost PC 2004/PC 2024 Expiratory channel contains
when replacing PC 2020, a new time stamp must be electronics including microprocessor for handling of:
set. In order to make this new time stamp correct,
there are two alternatives: • Expiratory flow measurement performed by the
bullet
flowmeter inside the cassette. The output signal
• Perform Preventive maintenance on the system
bullet Exp. Flow is used in sub-systems Control and
and report completed Preventive maintenance in Monitoring.
Service & Settings.
• All electronic connections to and from the cassette.
bullet
or
• Measurement of expiratory pressure via pressure
bullet
the PC 1781 Expiratory pressure transducer.
• If correct value (hours) until next Preventive
bullet
PC 2004/PC 2024 contains holder and electrical
maintenance is known, enter this value in Service
& Settings. connector for the pressure transducer board.

• Expiratory valve control (with input from Control).


bullet
3.20 PC 1781 Expiratory pressure transducer • Safety valve control (with input from other
bullet
(20) subsystems).
Includes an ID PROM. The ID information can be read
PC 1781 Expiratory pressure transducer is mounted
by the system.
onto PC 2004/PC 2024 Expiratory channel.
For software and hardware requirements and
Function identical to PC 1781 Inspiratory Pressure
compatibility, see the Compatibility chart in chapter
Transducer.
Revision history.

Note: If PC 2004/PC 2024 is replaced, the System


software must be re-installed.

3 - 12 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.22 PC 1991 Control (22) interface, a backup sound generating device


(buzzer) on PC 1992 will be activated. The backup
PC 1991 Control is part of subsystem Breathing BRE. buzzer is monitored by a microphone and tested at
The main responsibility is functions for the patient startup and during Pre-use check.
treatment, i.e. how to regulate the inspiratory and
expiratory gas flow. • Monitoring of alarm limits. Activates pressure
bullet
reducing mechanisms, including activation of the
PC 1991 Control comprises electronics including safety valve, in case of excessive breathing system
microprocessor for handling of: pressure.

• Supervision of delivered pressure and flow.


bullet
• O2 gas module and Turbine module.
bullet
• Calculation of parameters. Providing the system
bullet
• Airway pressure control with input from values
bullet with different measured data.
measured by the Inspiratory and Expiratory
pressure transducer. • Monitors battery module status.
bullet

• Nebulizer control.
bullet • Barometric pressure. PC 1992 contains a
bullet
barometric transducer and the measured barometric
A Memory backup battery on PC 1991 power supplies
pressure is supplied to the other sub-units in the
the internal memory on the PC board. If the battery
system.
on PC 1991 is disconnected or if the battery voltage
is too low, startup ventilation configurations made via • Real time clock, RTC.
bullet
the Service & Settings may be erased. The Memory • Temperature monitoring. A thermistor on PC 1992
bullet
backup batteries must be replaced after five years. monitors the temperature inside the Patient unit.
An alarm is activated if the temperature is 77 ±5 °C
Includes an ID PROM. The ID information can be read
(170 ±9 °F) or higher.
by the system.
• System voltage monitoring: The following voltages
bullet
For software and hardware requirements and are supervised:
compatibility, see the Compatibility chart in chapter
Revision history.
• +24 V
bullet
• +12 V
bullet
Note: The System software must be re-installed if • -12 V
bullet
PC 1991 is replaced. • +5 V
bullet
•+3.3 V
bullet
3.23 PC 1992 Monitoring (23) The buzzer on PC 1992 generates the alarm signal
in case of +5 V or +3.3 V power failures. The buzzer
PC 1992 Monitoring is part of subsystem Monitoring
and +5 V/+3.3 V failure logic is powered by backup
MON. The Monitoring subsystem features alarm and
capacitors in case of power failure.
monitoring functions, calculations, trending of
measured values and is also CAN-bus master.
• Maintains a persistent data storage area to save
bullet
logs. The logs are saved once per minute.
PC 1992 Monitoring comprises electronics including
microprocessor for handling of: • RS-232 communication. The system is equipped
bullet
with two RS-232 ports. The RS-232 communication
• Control that all alarms are displayed on the User
bullet is handled by PC 1992. The communication protocol
interface and that the alarm sound is generated. In SCI (SERVO Communication Interface) is described
case of malfunction in the loudspeakers in the User in the 'SERVO Communication Interface –
Reference Manual'.

SERVO-air Ventilator System, Service Manual 3 - 13


| 3 | Description of functions |

• Remote alarm function. The Remote alarm


bullet The system will automatically adjust to the connected
connector is available on PC 2021 Rear connectors. mains power if the mains power is within specified
The alarm signal used by this option is generated range. No voltage or frequency setting is required.
on PC 1992.
The mains inlet is equipped with two mains power
A Memory backup battery on PC 1992 power supplies
fuses, F11 and F12, rated 4 A.
the internal memory on the PC board. If the battery
on PC 1992 is disconnected or if the battery voltage
3.27 AC/DC Converter (27)
is too low, startup alarm configurations made via the
Service & Settings, all logs, Pre-use check results, The AC/DC Converter, including PC 2033, is a
date and time setting and service reports will be complete unit adapted for the system. The AC/DC
erased. The Memory backup batteries must be Converter converts the connected AC Power (mains
replaced after five years. power) to the internal DC supply voltage
+12V_AC-DC.
Includes an ID PROM. The ID information can be read
by the system. Mains power is supplied to the AC/DC Converter via
the Mains inlet.
For software and hardware requirements and
compatibility, see the Compatibility chart in chapter
Revision history. 3.28 PC 2028 DC/DC & Battery control (28)
This PC includes routing of Nebulizer signals to/from
Note: The System SW must be re-installed if PC PC 1991, and Ethernet/USB signals to the PC 2023
1992 is replaced. Panel board. The PC 2025 Battery back-plane is
connected to PC 2028, via the PC 2026 Battery
3.24 PC 1993 Ethernet switch (24) extension board. A secondary temperature sensor on
the PC 2028 regulates the Main fan and the O2
PC 1993 Ethernet switch controls the Ethernet evacuation fan.
communication in the system. Ethernet is used by the
PAN, BRE and MON subsystems. The main functions for the PC 2028 DC/DC & Battery
control are:
Includes an ID PROM. The ID information can be read
by the system. • On/Off control.
bullet
• Control charging/discharging of the Battery
bullet
3.25 PC 1994 Nebulizer (25) modules.

PC 1994 Nebulizer is the integrated interface for • Switch to battery power supply when Mains
bullet
connection of the Aerogen® Nebulizer Systems, power/External 12 V DC supply is lost, and control
Aerogen Pro and Aerogen Solo. PC 1994 Nebulizer switching between Mains power and external 12 V
contains electronics required to control the nebulizer DC power supply.
units. • Control of Main fan through one thermistor on PC
bullet
2020, and one on PC 2028.
For electrical safety test of the Nebulizer connector,
refer to chapter Preventive maintenance section Safety • Control of O2 evacuation fan through a thermistor
bullet
inspection. on PC 2030.

• Supply required internal voltages.


bullet
3.26 Mains inlet (26) • Connects the User interface control cable.
bullet
Inlet for mains power supply including grounding The ID information is stored in flash memory in the
connection. CPU and can be read by the system.

3 - 14 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.28.1 On/Off control 3.28.3 Internal voltages

The power modes in the system are: There are three power supply voltages to the system:

• At Power up, i. e. when the On/Off switch is turned


bullet • +12V_AC-DC from the AC/DC converter.
bullet
On, all internal voltages will be enabled. When the
On/Off switch is toggled from Off to On, the O2
• +12V_UNREG_EXT_DC from the External +12 V
bullet
DC power supply input.
evacuation fan is activated. The fan will run so that
any O2 build up is evacuated, prior to enabling • +12V_UNREG from the Battery modules.
bullet
voltages in the inspiratory compartment.
PC 2028 converts the power supply voltages into the
• At Power down, the Power supply system will
bullet following internal DC supply voltages:
deactivate the hardware signal Power_Good.H, in
order to let the different subsystems save their • +24 V
bullet
current settings in non-volatile memory. Power
down can be caused by:
• +12 V UNREG_INTERNAL
bullet

- Turning the On/Off switch Off.


• +12 V UNREG_VALVES
bullet

- Mains failure resulting in a switch to battery, • +12 V


bullet
where the backup battery voltage is too low for • -12 V
bullet
proper operation of the system or if there is no • +5 V
bullet
backup battery connected.
• +3.3 V
bullet
- Automatic power down when the battery voltage
becomes too low for operation of the system.
3.29 On/Off switch (29)
In the Off mode, only charging of Battery modules is
enabled (if the system is connected to mains). To allow Switch to Power up or Power down the system. Refer
adequate cooling during battery charging, the Main to section Power supply.
fan and O2 evacuation fan can operate also in Off
mode. All other circuitry is un-powered. 3.30 PC 2021 Rear connectors (30)

Note: At least one Battery module must always be The following connectors are located on PC 2021 Rear
installed to ensure proper operation. connectors:

• N26 – External +12 V DC power supply input. The


bullet
Note: A Power system restart might be necessary connector is equipped with a fuse F1, rated 10 A.
to reset power related alarms. To restart, switch Off There are no alarms indicating power supply failure
the system and disconnect both the mains power related to the External +12 V DC supply.
supply and the External DC power supply.
• N27 – USB (x2). For USB service key, data copy,
bullet
software updating and mouse.
3.28.2 Power supply selection
Note: Using a mouse with the system is only
If the internal DC supply voltage drops, due to power
permitted for service purposes.
supply failure, the power supply source will
automatically switch. The following power supply
source priority is used: • N28 – Ethernet. For Remote Services.
bullet

1. Mains power
• N29 – RS-232 (1). Isolated RS-232 connector
bullet
(identical to N30). For SCI communication or
2. External +12 V DC supply (if connected) MSync.
3. Battery modules
SERVO-air Ventilator System, Service Manual 3 - 15
| 3 | Description of functions |

• N30 – RS-232 (2). Isolated RS-232 connector


bullet To calculate its own status, the battery uses an internal
(identical to N29). For SCI communication or highly accurate voltmeter, amperemeter and time clock
MSync. to measure actual charge in and out of the battery. In
addition, there are algorithms to compensate for the
• N67 – Alarm output. Isolated alarm output
bullet
effects of discharge rate, discharge temperature,
connector. The Alarm output software option is
selfdischarge and charging efficiency, etc.
required to enable the Alarm output function. For a
technical description of the function Alarm output, The system will, among others, monitor:
refer to Alarm output in section Optional equipment.
• Expiration date.
bullet
Pin configuration and signal names can be found in
chapter Diagrams. • Charge cycles.
bullet
• If the operational capacity is too poor for continued
bullet
PC 2021 also hosts the On/Off switch and LEDs for usage.
power-on and mains power connection indication:
In all cases, battery replacement information will be
• A blue mains LED that is lit when mains power is
bullet shown on the User interface.
connected.
Even with this technology, the only time at which the
• A green power on LED that is lit when the On/Off
bullet
battery charge status is absolutely reliable is when it
switch is On.
is either completely full or completely empty. What’s
more, if the battery only sees partial charges and
3.31 PC 2026 Battery extension (31) discharges during its application, then it may not get
the benefit of a 'full' or 'empty' reference point for some
PC 2026 Battery extension is a connector board. It
time, and must rely more and more on its calculated
connects PC 2025 Battery back-plane to PC 2028
figure.
DC/DC & Battery control. There are no active
components on PC 2026. Select SYSTEM STATUS > Batteries on the User
interface to check battery status. For further
3.32 PC 2025 Battery back-plane (32) information, refer to chapter Service procedures,
section Battery modules.
PC 2025 Battery back-plane has connectors for the
battery modules. PC 2025 is connected to PC 2028 The battery life time is 300 charge cycles or four years
DC/DC & Battery control via the PC 2026 Battery (48 months) from manufacturing date, whichever
extension board. comes first. Remaining charge cycles and time
(months) are shown in the STATUS > Battery-window.
3.33 Battery module (33)
The remaining charge cycles are ‘full charge cycles’
The Battery module is a 14.4 V/6.6 Ah Li-ion i.e. from an empty battery (0 %) to a full battery
rechargeable 'smart battery'. Two Battery modules (100 %). If the battery is charged in shorter cycles,
can be connected to the Battery compartment. e.g. from 75 % to 100 %, the cycles are added until
they reach 100 % and then calculated as one full
Note: At least one Battery module must always be charge cycle.
installed for the system to work.
Example: Four charge cycles from 75 % to 100 % is
counted as one full charge cycle.

3 - 16 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

Typical recharge time approximately 2 h/battery The Fan will start with half effect at approximately
(90 %), up to 3 h (100 %) if battery is completely 33 °C (91 °F) and with full effect at approximately
discharged. If a battery is fully discharged, e.g. due to 43 °C (109 °F). When the temperature drops below
long storage time, each battery may require up to approximately 37 °C (99 °F), the Fan turns to half
6 hours charging time. effect and when the temperature drops below
approximately 27 °C (81 °F), the Fan stops.
Each battery includes an ID PROM. The ID information
can be read by the system. The air inlet is protected by a filter that must be
cleaned or replaced during the Preventive
CAUTION: Do not expose the batteries maintenance.
to water, fire or excessive heat. Do not
crush, disassemble, puncture or short 3.35 User interface cable (35)
circuit the connector terminals.
This User interface cable connects the Patient unit
and the User interface. The cable is hidden in the
panel arm behind a cover.
3.34 Main fan (34)
The Main fan forces cooling air through the AC/DC Note: The User interface cable must only be
section, and the Main section of the Patient unit. The connected or disconnected following the instructions
cooling air flow inside the Patient unit, is indicated in given in section Hazard notices in chapter
the System overview. Disassembling and assembling.

3.36 Touch screen (36)


The projected resistive Touch screen implements the
touch function of the User interface screen and is
s interactive with information displayed on the LCD
Display. The Touch screen can be recalibrated in
Service & Settings.

3.37 LCD Display (37)


a
The LCD Display is a 12″ 4:3 TFT display with LED
S20055
backlight for display of picture- and alphanumeric data.

1. Main fan inlet It is estimated that the LCD Display will maintain an
2. Main fan outlet acceptable brightness level during the unit’s lifetime.
Using Service & Settings, a backlight time meter can
The Main fan inlet (1) is located at the bottom of the be shown.
side cover. The Main fan outlet (2) is located above
the connectors on the backside of the Patient unit. 3.38 PC 2023 Panel (38)
The Main fan is monitored by the temperature sensors The PC 2023 Panel contains drivers for:
on PC 2020, and PC 2028, and controlled by PC 2028.
The value from the temperature sensor, giving the • Touch and display
bullet
highest temperature reading, will be used.
• Loudspeaker
bullet
The Fan is always on when the batteries are charging, • Microphone
bullet
and for 2.5 hours after the charging is complete.
• Microcontroller
bullet
SERVO-air Ventilator System, Service Manual 3 - 17
| 3 | Description of functions |

The multicolor LEDs, visible in the Alarm indicator, Preferably the NC contact should be used, as this will
are located on PC 2023. detect any interruption in the alarm system or
unintentional disconnection of the cable between the
3.39 Alarm indicator (39) system and the ventilator. Furthermore, if both the
NO- and the NC-contacts are utilized, the highest
The Alarm indicator is located on the upper part of the safety level for unforeseen failures can be achieved.
User interface frame. The multicolor LEDs on PC 2023 Note that, in both cases, even an intentional
are visible in the Alarm indicator. The LEDs will flash disconnection of the ventilator will trigger the external
when an alarm is triggered in the system: alarm.

• Blue LED – Low priority alarm


bullet The Alarm output software option is required to enable
the Alarm output function. The Alarm output software
• Yellow LED – Medium priority alarm
bullet
is individually created for each specific system and
• Red LED – High priority alarm
bullet can only be installed on this system. The Alarm output
function must be enabled in the configuration software.
3.40 Loudspeaker (40)
3.41.2 Battery module
Loudspeakers for, e.g., alarm. Connected to the
PC 2023 Panel. The Module compartment allows up to two Battery
modules. To guarantee safe battery backup, always
There is one loudspeaker in the User interface. The
use at least one battery.
User interface is mechanically designed to achieve a
good sound output. For further information, refer to section Battery module.

At startup and during Pre-use check the functionality


3.41.3 Battery charger
of the loudspeakers are monitored by a microphone.
During operation the loudspeaker is continuously The Battery charger is an external smart battery
monitored. charger used to charge the SERVO-air Battery
modules. The charger has one slot and can charge
3.41 Optional equipment one battery at a time. For further information, see the
Battery charger manual.
3.41.1 Alarm output

The Alarm output connector, N67 on PC 2021 Rear 3.41.4 Mobile cart
connectors, enables connection of an external alarm
The Mobile cart is designed for carrying the Patient
signal system to the system. The alarm output signal
unit, the User interface and all required optional
is active as long as the alarm audio is active on the
equipment.
system.
The Mobile cart has standard 10x25 mm rails to carry
For pin configuration and signal names in the Alarm
accessories such as Support arm 179 and Humidifier
output connector N67, see Connectors in chapter
holder. The column has vertical slots for installation
Diagrams. For further information, refer to the 'Alarm
of e.g. Gas cylinder restrainers. The optional Drawer
output connector – Reference Manual'.
kit or Cabinet can be mounted on the Mobile cart.
The Alarm output connector has two contact functions:
NO (Normally Open) and NC (Normally Closed). In 3.41.5 Gas cylinder restrainer kit
an alarm situation the open contact will close and the
closed one will open. The contacts are independent The Gas cylinder restrainers are mounted on each
of polarity and can be used both with AC and DC side of the Mobile cart. It is intended for two gas
systems. cylinders, maximum volume 4.5 litres each.

3 - 18 SERVO-air Ventilator System, Service Manual


| Description of functions | 3 |

3.41.6 Shelf base 3.41.11 MSync

The Shelf base is a mounting device for the Patient MSync is intended for information transfer from the
unit. It allows positioning of the Patient unit on a shelf. Maquet devices to the hospitals information system
by:
3.41.7 Support arm 179
• Importing real-time clinical data from the device
bullet
Support arm 179 is equipped with a clamp intended using its respective communication protocol.
for standard 10x25 mm rails. The support arm has a
central locking mechanism that secures the selected
• Converting device data to HL7 compliant data.
bullet

position. A twin tube holder (15-27 mm diameter) and • Exporting device data in HL7 format to the
bullet
information system.
a filter/tube support (20/25 mm diameter) can be
mounted on the support arm. Maximum load with fully An MSync unit is connected to each Maquet device
extended arm is 1 kg. intended to be integrated with the information system.

3.41.8 Aerogen nebulizers

The Aerogen® Nebulizer Systems, Aerogen Pro and


Aerogen Solo, can be connected to the Nebulizer
connector. PC 1994 Nebulizer contains electronics
required to control the nebulizer units.

Aerogen Pro and Aerogen Solo are devices intended


to aerosolize physician-prescribed medications, for
inhalation, that are approved for use with a general
purpose nebulizer.

• Aerogen Pro: Reusable nebulizer unit.


bullet
• Aerogen Solo: Single use nebulizer unit.
bullet
The Aerogen Pro/Aerogen Solo are designed to
operate in-line with the ventilator circuit in acute and
sub acute care environments. It operates without
changing patient ventilator parameters.

3.41.9 Humidifier and Humidifier holder

Recommended active (heated) humidifiers are Fisher


& Paykel MR 810 and MR 850. For further information,
refer to manufacturer's documentation.

The Humidifier holder is equipped with a clamp


intended for standard 10x25 mm rails. Maximum load
is 5 kg.

3.41.10 Waterbag/IV pole

The Waterbag/IV pole is mounted on the optional


Humidifier holder. It includes two hooks. Maximum
load is 1 kg.
SERVO-air Ventilator System, Service Manual 3 - 19
| 3 | Description of functions |

3 - 20 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4 Disassembling and assembling

Table of Contents
4.1 General | 4-2
4.2 Hazard notices | 4-2
4.3 PC boards and electrical components | 4-2
4.4 Information stored in the system | 4-3
4.5 Assembling guidelines | 4-5
4.6 Front cover | 4-6
4.7 Top cover | 4-6
4.8 Chassis | 4-7
4.9 Main section | 4-8
4.10 Inspiratory section | 4-9
4.11 AC/DC section | 4 -16
4.12 Front section | 4 -21
4.13 Rear section | 4 -22
4.14 PC 2020 Main back-plane | 4 -24
4.15 Expiratory cassette | 4 -25
4.16 User interface | 4 -28

SERVO-air Ventilator System, Service Manual 4-1


| 4 | Disassembling and assembling |

4.1 General CAUTION: Do not expose the batteries


Disassembling of the system is described in this to water, fire or excessive heat. Do not
chapter. If not stated otherwise, the assembling crush, disassemble, puncture or short
procedure is the reverse of the described circuit the connector terminals.
disassembling procedure.

The illustrations in the Spare Parts Lists are useful 4.3 PC boards and electrical components
when disassembling and assembling the system.
4.3.1 Replacing electrical components
After any installation, maintenance or service
intervention in the system, perform a 'Functional check’ As stated in section Hazard notices, mains power and
according to instructions in chapter Important. battery power supply must be disconnected and On/Off
switch set to Off when replacing electrical components.
4.2 Hazard notices If energized electrical components are disconnected
or connected, this may interfere with the operation of
Before disassembling or assembling the system, make the unit.
sure that:
After replacement of PC boards or other electrical
• The system is switched Off.
bullet components, it is recommended to perform an
electrical safety test. For electrical safety tests, refer
• Mains power cable is disconnected.
bullet
to chapter Preventive maintenance, section Safety
• External +12 V DC supply is disconnected.
bullet inspection.
• Backup battery modules are disconnected. Note
bullet
that the backup battery modules will supply power 4.3.2 Handling PC boards
to parts of the system also when the system is
switched Off. Those who come into contact with circuit boards
containing sensitive components must take certain
• Gas supply is disconnected, central gas supply and
bullet
precautions to avoid damaging the components (ESD
backup gas supply.
protection).
• The system, including all gas conveying parts, is
bullet
cleaned. Refer to instructions in the User’s Manual. PC boards (spare parts) must always be kept in
protective packaging for sensitive electronic device.
WARNING! With gas supply connected
to the system, there are pressurized PC boards must not be inserted or removed while the
components inside the unit. All mains power or battery power is applied to the PC
personnel must exercise extreme boards.
caution if fault tracing or adjustments
are performed with gas supply Remove and insert the PC boards very carefully to
connected and with covers removed. avoid damage to the connectors.

Note: The PC boards contain


WARNING! With power supply components that are highly sensitive
connected to the system, there are to static electricity. When working with
energized electrical components inside ESD sensitive components, always use
the unit. All personnel must exercise a grounded wrist band and a grounded
extreme caution if fault tracing or work surface. Adequate service tools
adjustments are performed with power must always be used.
supply connected and with covers
removed.

4-2 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.3.3 Replacing PC boards When replacing PC 2004/PC 2024 Expiratory Channel,


two Pre-use checks may be required. The first check
The System SW is distributed to the following PC will calibrate the Safety valve, but fail in other tests.
boards and modules: The second check will pass.

• PC 1991 Control
bullet Other technical errors may occur after a software
• PC 1992 Monitoring
bullet installation. These errors will disappear after restarting
the system.
• PC 2004/PC 2024 Expiratory channel
bullet
• PC 2023 Panel
bullet 4.4 Information stored in the system
• PC 2027 Motor control
bullet
All electronic components equipped with an ID PROM
• PC 2028 DC/DC & Battery control
bullet store the information that is specific for the component,
for example article number, version number and serial
Additional actions when replacing PC boards number.

When delivered as spare parts, these components Furthermore, some of the electronic components store
are equipped with a System SW version that may differ information related to the complete system, for
from the version on the system to be repaired. To keep example system software, system info, configurations,
the System SW version used prior to the replacement, logs and so forth. When replacing these components,
the applicable System SW version must be available the stored information may be lost. See table below.
for reinstallation purposes.

For functionality enhancement, the latest released


System SW version is always recommended.

Before installing a new System SW version, ensure


that the software is fully compatible with all hardware,
software and mechanical components in the system.
Refer to the Compatibility chart in chapter Revision
history.

When replacing PC 1991 Control or PC 1992


Monitoring, turn on the unit and leave the unit on for
two minutes before restart. A certain time is required
for the options and settings to be transferred to the
new PC board.

When replacing PC 1991 Control or PC 1992


Monitoring, the set time and date may need to be
adjusted.

When replacing PC 1992 Monitoring, Technical error


25 may occur. Adjust the Barometric pressure in
Service & Settings to solve this error (Service key
required).

SERVO-air Ventilator System, Service Manual 4-3


| 4 | Disassembling and assembling |

PC 1781 Pressure transducer

PC 1786 Expiratory cassette


PC 2028 DC/DC & Battery
PC 2020 Main back-plane

PC 2027 Motor control


PC 1992 Monitoring

Expiratory channel
PC 2004/PC 2024
PC 1991 Control
Stored information

PC 2023 Panel
control

O2 cell
System software X X X X X X
System info: X
§ Serial number
§ Total operating and
ventilation time
§ Next preventive
maintenance
System info: X X
§ Installed software
options
System info: X
§ System version
Configurations X
§ Startup ventilation
configuration
Configurations X
§ Startup alarm
configuration
Configurations X
§ Units, language, date
format, etc.
Date and time setting X
Logs X X2 X
Trends X
Recordings X
Pre-use check results X X X X X
including transducer result calibration calibration calibration calibration
calibrations
Barometer calibration X
Touch screen calibration X
Performed preventive X1
maintenance
Replaced expiratory X1 X
membrane counter
Replaced backlight X1 X3
counter
Backlight operating time X3
Language X
Expiratory cassette: X
§ Date of first use
§ Total operating time
§ Remaining membrane
capacity
O2 cell: X
§ Date of first use
§ Estimated remaining
capacity (O2 cell)
S20104

1
Included in Service report as part of the Service log.
2
Copied to PC 1992 at Power On and continuously when the system is powered.

4-4 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

3 Excluded in later System versions.

4.5 Assembling guidelines 4.5.3 Special grease

All parts of the system assembled with screws and Lubricants should normally not be used when servicing
nuts are tightened with a specified torque. Thread the system. If lubricants must be used, use only very
locking adhesives are used if required. small amounts of grease with P/N 95 73 700. This
special grease is O2 compatible.
In order to maintain these specifications over time, it
must be ensured that after any service intervention
WARNING! Due to the increased O2
removed parts are re-assembled and secured
concentration in valves and gas
according to instructions. Make sure to follow the
channels, only special grease intended
guidelines stated below.
for use in high O2 concentrations, are
allowed.
4.5.1 Tightening torque

• Thread size M3: 0.95 Nm ±15%


bullet
• Thread size M4: 3.1 Nm ±15%
bullet
• Thread size M4 (in plate thread joints):
bullet
1.4 Nm ±15%. These screw joints are indicated with
a ’1.4 Nm symbol’ in the concerned illustrations.

• Thread size M5: 5.0 Nm ±15%


bullet
• Thread size M6: 7.0 Nm ±15%
bullet
• Gas inlet nipple: 20 Nm ±15%
bullet

Note: Do not fasten the screws too tightly in the


screw joints with plate threads. When tightening the
screws, it is important to turn each alternatively and
gradually. Do not apply specified torque until all the
screws have been positioned correctly into the plate
threads.

4.5.2 Thread locking adhesives

• Loctite 243® on the expiratory valve coil screws.


bullet

SERVO-air Ventilator System, Service Manual 4-5


| 4 | Disassembling and assembling |

4.6 Front cover

• Tilt back the User interface.


bullet
• Release the Front cover hatches, and then release
bullet
and lift off the Front cover.

S20020

4.7 Top cover

• Remove the screw (1).


bullet
Note that the screw remains in the top cover.

• Release and lift off the Top cover.


bullet

1.4 Nm
S20021

4-6 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.8 Chassis
4.8.1 Removing the chassis

• Remove the Front cover.


bullet
• Remove the Top cover.
bullet
• Remove the O2 evacuation fan filter (1) by pressing
bullet
the catch in and down.

• Remove the Main fan filter (2) by pressing the catch


bullet a
in and down.

• Remove the Expiratory cassette.


bullet

S20022

• Remove the five screws (positions 1-4) that secure


bullet
the Chassis.
Note that there are two screws at position (1), one
on each side of the Patient unit.

s
a
a

1.4 Nm f d
S20023

• Pull the Chassis carefully to the sides (1), push


bullet s
backwards (2) and lift it off.

a a

S20024

SERVO-air Ventilator System, Service Manual 4-7


| 4 | Disassembling and assembling |

4.8.2 Replacing the handle

• Remove the Chassis.


bullet
• Remove the screws (1).
bullet
• Pull apart the chassis to release the handle (2).
bullet
s

0.6 Nm

S20134

4.9 Main section


4.9.1 Main section lid

• Remove the Top cover.


bullet a
• Remove the three screws (1-2) that secure the Main
bullet
section lid to the Patient unit.

1.4 Nm

S20025

4-8 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.9.2 PC board rack

• Remove the Main section lid.


bullet a s d f
• Carefully pull out the PC board:
bullet
- PC 1993 Ethernet switch (1)
- PC 1992 Monitoring (2)
- PC 1991 Control (3) g
- PC 2004/PC 2024 Expiratory channel (4)
including PC 1781 Expiratory pressure
transducer (5)
Note: Before removing PC 2004/PC 2024:
- Disconnect Expiratory valve coil cable
connector (6). h j k
S20032

- Disconnect Pressure transducer tube (7).


- Disconnect PC 2031/PC 1785 cable
connector (8).

For software and hardware requirements and


compatibility, see the Compatibility chart in chapter
Revision history.

Note: Additional actions may be required after


replacement of PC boards. Refer to section
Replacing PC boards in this chapter.

4.10 Inspiratory section


4.10.1 Inspiratory section lid

• Remove the Chassis.


bullet 1
• Remove the eight screws (1-3) that secure the
bullet
Inspiratory section lid to the Patient unit. On systems
manufactured from Q3 2017 it is only required to
loosen (not remove) the three screws (2). The holes
in the lid are changed to slots as shown in the 4
illustration. 6

• Loosen the two screws (4).


bullet

Note: Note that the O2 evacuation fan mounted on


the Inspiratory section lid is connected to the Patient 2
unit by the fan cable. Carefully fold out the lid to 3
disconnect the cable.
1.4 Nm
S20026

SERVO-air Ventilator System, Service Manual 4-9


| 4 | Disassembling and assembling |

• Disconnect the O2 evacuation fan cable (5).


bullet

When assembling:
• Do not forget to connect the fan cable.
bullet
• Make sure that the Inspiratory gas docking kit
bullet
connects properly to the Inspiratory section lid at
(6). g S20056

4.10.2 O2 evacuation fan

• Remove the Inspiratory section lid.


bullet
• Disconnect the O2 evacuation fan cable
bullet
connector (1) from the PC 2030 Inspiratory channel.

• Release the four rubber straps (2), and lift off the
bullet
fan.

Note: When mounting the fan, make sure to point


the fan in the right direction. The cooling air flow
must be directed into the Patient unit. a s
S20039

4.10.3 Inspiratory flowmeter

• Remove the Inspiratory section lid.


bullet
• Disconnect the cable (1) from PC 2030.
bullet

Note: Be careful with the connection (2) on the


flowmeter. Make sure not to break it when removing
the flowmeter from the Inspiratory channel muff.

• Remove the flowmeter from the Inspiratory channel


bullet
muff.
s

S20057

4 - 10 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.4 Safety valve

• Remove the Inspiratory section lid.


bullet
• Disconnect the O2 cell cable (1).
bullet
• Loosen and fold out the locking latch (2).
bullet
a

S20058

• Remove the O2 cell (3).


bullet

S20059

• Disconnect the Inspiratory pressure transducer


bullet
tube (4).

• Remove the Inspiratory channel (5).


bullet

f S20109

SERVO-air Ventilator System, Service Manual 4 - 11


| 4 | Disassembling and assembling |

• Disconnect the Safety valve cable (6) from PC 2030.


bullet
• Remove the two screws (7) that secure the Safety
bullet
valve.

• Remove the Safety valve and cable (8).


bullet

k
h

S20110

4.10.5 Gas module lid

• Remove the Top cover.


bullet a
• Remove the screw on the top of the Patient unit (1).
bullet
• Remove the two screws on the side of the
bullet
Inspiratory section (2). On systems manufactured
from Q3 2017 it is only required to loosen (not d
remove) the two screws (2). The holes in the side
s
wall are changed to slots as shown in the illustration.

• Pull out the Air inlet filter (4) from the socket.
bullet
• Remove the Gas module lid (3).
bullet f
g
Note: When attaching the Gas module lid to the
Patient unit, make sure the Battery extraction tag (5)
is not stuck under the lid. This will make it hard to 1.4 Nm

pull out the Battery module. S20029

4 - 12 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.6 O2 gas module

• Remove the Gas module lid.


bullet a 1.4 Nm
• Remove the screw (1).
bullet
• Pull out the O2 gas module (2).
bullet s

S20111

4.10.7 Turbine module

• Remove the Gas module lid.


bullet
• Pull out the Turbine module (1).
bullet

a
Note: Additional actions may be required after
replacement of PC boards (PC 2027 is included in
the Turbine module). Refer to section Replacing PC
boards in this chapter.

S20033

4.10.8 Inspiratory gas docking

• Remove the Inspiratory section lid.


bullet
s
• Remove the Gas module lid.
bullet
• Remove the O2 module.
bullet
• Remove the Turbine module.
bullet
• Disconnect the Inspiratory flowmeter connector
bullet
muff (1).

• Lift off the Inspiratory gas docking (2).


bullet a

Note: Be careful with the connection on the


S20034
flowmeter, make sure not to break it when
disconnecting the Inspiratory flowmeter muff.

SERVO-air Ventilator System, Service Manual 4 - 13


| 4 | Disassembling and assembling |

4.10.9 Turbine module one-way valve

• Remove the Inspiratory gas docking.


bullet g f d s a
• Remove the four screws (1), and lift off the cap (2)
bullet
with O-ring (3).

• Pull out the one-way valve (4) from the Inspiratory


bullet
gas docking housing (5).

0.5 Nm

S20035

To insert the Turbine module one-way valve:

• Pull the strap through the hole in the Inspiratory gas


bullet
docking until it snaps into place.

Note: Make sure all parts are clean before inserting


the Turbine module one-way valve.

S20036

4 - 14 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.10.10 PC 2030 Inspiratory channel

• Remove the Inspiratory gas docking.


bullet
• Disconnect the O2 cell cable (1) from PC 2030.
bullet
• Disconnect the Inspiratory transducer tube (2) from
bullet
PC 1781.

• Remove the four screws (3) that secure PC 2030.


bullet
Note that one screw must be removed through the
Safety valve channel.

d
s
a

S20112

• Lift off PC 2030 (4).


bullet h
Note that PC 2029 (5) Inspiratory channel extension
may come loose when lifting off PC 2030.

4.10.11 PC 1781 Inspiratory pressure transducer


f
g
• Open or cut the cable tie.
bullet
S20113

• Lift off PC 1781 (6).


bullet

SERVO-air Ventilator System, Service Manual 4 - 15


| 4 | Disassembling and assembling |

4.11 AC/DC section


4.11.1 AC/DC section lid

• Remove the Chassis.


bullet

Open the AC/DC section lid:


• Remove the four screws (1) that secure the AC/DC
bullet
section lid to the Patient unit.

a 1.4 Nm
S20027

Note: Note that the AC/DC section lid is still


attached to the hinge at the bottom.

S20120

4 - 16 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

Remove the AC/DC section lid:

• Disconnect the Main fan cable (1).


bullet

a
S20114

• Open the AC/DC section lid to maximum, slide it


bullet
towards the back and then lift it off.

S20121

4.11.2 Main fan

• Open the AC/DC section lid.


bullet a s
• Disconnect the Main fan cable connector (1) from
bullet
the Main back-plane.

• Release the four rubber straps (2), and lift off the
bullet
fan.

Note: When mounting the fan, make sure to point


the fan in the right direction. The cooling air flow
S20038
must be directed into the Patient unit.

SERVO-air Ventilator System, Service Manual 4 - 17


| 4 | Disassembling and assembling |

4.11.3 AC/DC converter

• Open the AC/DC section lid.


bullet
• Disconnect the internal power cable connector (1).
bullet
• Disconnect the signal cable connector (2).
bullet
• Disconnect the mains power cable connector (3).
bullet
• Remove the screws (4) and lift off the AC/DC
bullet
converter. f a s d f
1.4 Nm

S20037

4.11.4 Mains inlet

• Remove the AC/DC converter.


bullet
• Remove the two screws (1) that secure the mains
bullet
power inlet.

• Remove the two nuts (2) that secure the grounding


bullet
cables to the grounding screw (3).

• Remove the mains power inlet.


bullet

s d

S20122

4 - 18 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.11.5 Expiratory cassette holder

• Remove the Chassis.


bullet
• Remove the Main section lid.
bullet
• Remove the four screws (1) that secure the User
bullet
interface cable lid (2) to the Patient unit.

• Lift off the User interface cable lid (2).


bullet
• Disconnect the User interface cable (3) from the
bullet
connector. h
• While holding the User interface, remove the four
bullet
hinge screws (4) that secure the User interface on g
the Chassis. f
• Remove the User interface (6).
bullet
• Remove the screw (5).
bullet

d s

a S20046

• Open the AC/DC section lid (1).


bullet s d f g
• Remove the Expiratory valve coil cable (2) from
bullet
PC 2004/PC 2024.

• Remove the Expiratory flowmeter ribbon cable (3)


bullet
from PC 2004/PC 2024.

• Remove the Expiratory transducer tube (4) from


bullet
PC 2004/PC 2024.

• Remove the plastic cover (5) at the Expiratory


bullet
cassette outlet by bending it towards the right and
up.

a
S20116

SERVO-air Ventilator System, Service Manual 4 - 19


| 4 | Disassembling and assembling |

• Remove the four screws that secure the cassette


bullet
holder.

• Carefully lift at position (1) to release the Expiratory


bullet
cassette holder from the metal bracket.

• Carefully lift off the Expiratory cassette holder and


bullet
at the same time release the two cables and tube
from the silicone guide (2).

a
s
S20117

4.11.6 PC 2031/PC 1785 Expiratory channel


connector

• Remove the Expiratory cassette holder.


bullet
• Remove the two screws that secure the
bullet s
PC 2031/PC 1785 to the Expiratory cassette holder.

• Carefully pull out PC 2031/PC 1785 (1).


bullet a
For software and hardware requirements and
compatibility, see the Compatibility chart in chapter
Revision history.

4.11.7 Expiratory valve coil


S20119
• Remove the Expiratory cassette holder.
bullet
• Remove the four screws that secure the Expiratory
bullet
valve coil to the Expiratory cassette holder.

• Remove the Expiratory valve coil (2).


bullet

4 - 20 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.12 Front section


4.12.1 Nebulizer

• Remove the Chassis.


bullet
• Remove the five screws (1) that secure the
bullet
Nebulizer lid.

• PC 1994 Nebulizer (2) is mounted inside the


bullet
Nebulizer lid and can now be removed from the lid.

s a
S20031

4.12.2 Front section lid including PC 2028 DC/DC


& Battery control

• Remove the Nebulizer.


bullet
• Remove the four screws (1) that secure the User
bullet
s
interface cable lid (2) to the Patient unit. d
• Remove the User interface cable lid.
bullet
• Disconnect the User interface cable (3) from the
bullet
connector.

a
S20124

• Remove the seven screws (1) that secure the Front


bullet
panel lid (incl. PC 2028) to the Patient unit.

Note: Some of the screws may have been removed


earlier.

• Carefully pull out the lower part of the Front section


bullet
lid to release the PC board connectors, then slide
the lid and the board down to remove.

S20028

SERVO-air Ventilator System, Service Manual 4 - 21


| 4 | Disassembling and assembling |

4.12.3 PC 2026 Battery extension board

• Remove the Front section lid including PC 2028


bullet
DC/DC & Battery control

• Carefully pinch the plastic spacer (1) that secures


bullet
PC 2026, i.e. with pliers. a
• Remove PC 2026 (2).
bullet

4.12.4 PC 2025 Battery back-plane

• Remove the Front section lid including PC 2028


bullet d
DC/DC & Battery control
s
• Remove the four screws that secure PC 2025.
bullet
• Lift out PC 2025 (3).
bullet

S20115

4.13 Rear section


4.13.1 Rear connectors lid including PC 2021 Rear
connectors

• Remove the Top cover.


bullet
• Remove the two screws (1).
bullet
a
• Remove the mains power cable restrainer (2).
bullet
Not shown in the illustration.

• Remove the screw (3).


bullet
• Remove the five screws (4) that secure the Rear
bullet
connectors lid (incl. PC 2021) to the Patient unit. d
• Carefully slide out the Rear connectors lid.
bullet
s

f
S20030

4 - 22 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.13.2 On/Off switch

• Remove the Rear connectors lid including PC 2021


bullet
Rear connectors.
a
• Disconnect the On/Off switch cable (1) from
bullet
PC 2021.

• Remove the nut and washer (2) that secure the


bullet
On/Off switch.

• Lift off the On/Off switch.


bullet s

S20123

4.13.3 Battery guides

• Remove the chassis.


bullet
• Remove the gas module lid.
bullet
• Remove the two screws (1).
bullet
• Insert a thin blade between the plastic battery guide
bullet
and the metal wall. This will release the guides. a
• Pull off both plastic guides (2 and 3).
bullet
1
1.4 Nm
2
When assembling, make sure that the leaf springs (4)
are mounted in the left-side guide. f d s S20131

SERVO-air Ventilator System, Service Manual 4 - 23


| 4 | Disassembling and assembling |

4.14 PC 2020 Main back-plane

• Remove the Chassis.


bullet
• Remove the Main section lid and the PC boards.
bullet
• Remove the Inspiratory section lid, and remove:
bullet
- Inspiratory flowmeter
- Safety valve
- Gas module lid
- O2 module
- Turbine module
- Inspiratory gas docking
- PC 2030 Inspiratory channel incl. PC 1781
Inspiratory pressure transducer
- PC 2029 Inspiratory channel extension

• Remove the Nebulizer and the Front section lid.


bullet
• Remove the AC/DC section lid.
bullet
• Remove the Rear connectors lid incl. PC 2021 Rear
bullet
connectors.

• PC 2020 Main back-plane is mounted with seven


bullet
clips (1). To release the board from the clips, pull it
carefully approximately 3 mm towards the front.

• Lift the board slightly and then slide it out from the
bullet
front.

a
S20118

4 - 24 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.15 Expiratory cassette


The Expiratory cassette is a complete unit and must
not be disassembled. There are two spare parts though
that may need to be replaced:

• Moisture trap including O-ring


bullet
• Expiratory cassette valve membrane
bullet

4.15.1 Moisture trap

• Use a screwdriver and carefully bend on both sides


bullet
of the Moisture trap to remove it from the Expiratory
cassette.

S20126

4.15.2 Expiratory cassette valve membrane

Operating capacity for the membrane is estimated to


10.000.000 breathing cycles. When this limit is passed
or if the membrane for some reason has become
defective, it must be replaced.

Membrane capacity (counter) can be shown in the


System status window. Tap SYSTEM STATUS >
Expiratory cassette to check Membrane capacity. The
operating capacity meter must be reset after
replacement of the membrane.

Note: It is very important for the function of the


expiratory valve that the valve membrane is
removed and mounted correctly as described below.

To remove the valve membrane from the cassette

SERVO-air Ventilator System, Service Manual 4 - 25


| 4 | Disassembling and assembling |

• Carefully remove the membrane including retaining


bullet
ring using a suitable retaining ring pliers.

S20041

To mount the valve membrane into the cassette

• Place the retaining ring (1) correctly into the


bullet
membrane. Make sure that the retaining ring (1) and
the washer (2) are correctly mounted.

Note: It is very important that the valve membrane


and the membrane seat in the cassette are clean.
Dirt particles can create leakage in the cassette.

S20042

• Place the membrane onto the cassette as shown in


bullet
the illustration.

S20043

• Squeeze the retaining ring pliers and carefully press


bullet
the membrane in place into the cassette as shown
in the illustration.

S20044

4 - 26 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

• Carefully release and remove the pliers.


bullet
• Check that the membrane is not deformed by the
bullet
retaining ring. If necessary, remove the membrane
and redo the complete mounting procedure.

• Mount the expiratory cassette onto the Patient unit.


bullet
• Reset the operating capacity meter after
bullet
replacement of the membrane. To access the reset
button, tap SERVICE & SETTINGS > SERVICE
REPORT > Report replacement of exp. cassette
membrane. S20045

SERVO-air Ventilator System, Service Manual 4 - 27


| 4 | Disassembling and assembling |

4.16 User interface


Remove the User interface from the Panel arm and
place it on an ESD protected work space, following
the instructions below.

4.16.1 Removing the User interface

• Remove the Front cover.


bullet
• Remove the four screws (1) that secure the User
bullet
interface cable lid (2) to the Patient unit.
s
• Lift off the User interface cable lid.
bullet d
• Disconnect the User interface cable (3) from the
bullet
connector.

S20125

• Tilt back the User interface.


bullet a s
• Remove the four screws (2) that secure the Panel
bullet
arm lid (1), and then lift off the lid.

S20047

• Remove the four hinge screws (1).


bullet

S20048

4 - 28 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

4.16.2 Opening the User interface

• Remove the caps (1) covering the screws on the


bullet
heat sink.

• Remove the five screws (2) and lift off the User
bullet
interface from the holder.

a s
S20049

• Disconnect the User interface cable (1) from the


bullet
User interface.

S20050

SERVO-air Ventilator System, Service Manual 4 - 29


| 4 | Disassembling and assembling |

• Disconnect the loudspeaker cable (1).


bullet a s
• Remove the eight screws (2) and lift off the backside
bullet
of the panel chassis.

S20051

• The loudspeaker (1) is attached to the backside of


bullet
the panel chassis using double-coated adhesive
tape.

a
S20052

• Remove the Alarm indicator (1).


bullet s a
• Remove the eight screws (2) and lift off the metal
bullet
cover.

S20053

4 - 30 SERVO-air Ventilator System, Service Manual


| Disassembling and assembling | 4 |

• Disconnect the cables (2-4), and then remove the


bullet
seven Panel board screws (1). a
• Remove the LCD display (6) and PC 2023 Panel (5),
bullet
and place them on an ESD protected work space.
g s
Note: Additional actions may be required after
replacement of PC boards. Refer to section
'Replacing PC boards' in this chapter. h d

f
S20054

SERVO-air Ventilator System, Service Manual 4 - 31


| 4 | Disassembling and assembling |

4 - 32 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5 Service procedures

Table of Contents
5.1 Service key | 5-2
5.2 Service & Settings | 5-3
5.3 Software installation | 5 -58
5.4 Leakage detection | 5 -62
5.5 Battery modules | 5 -68
5.6 Memory backup batteries | 5 -70

SERVO-air Ventilator System, Service Manual 5-1


| 5 | Service procedures |

5.1 Service key 5.1.1 Connecting a USB mouse to the system

The Service key is a USB memory stick provided to If the Touch screen does not function properly, due
personnel trained and authorized by Getinge to to a faulty panel or not being well enough calibrated,
perform installation, service or maintenance of the a mouse can be connected to the system. The mouse
system. can then be used to, e.g., logging in to Service &
Settings to start the Touch screen calibration.
With the Service key it is possible to access different
user levels of the built-in Service & Settings software.
Note: Using a mouse with the system is only
There are two user levels intended for Service that
permitted for service purposes.
are described in this manual:

1. Connect the mouse to the USB port when the


• Trained biomed. User level intended for hospital
bullet
system is turned Off.
biomed trained and authorized by Getinge.
2. Start the system.
• Field service. User level intended for trained and
bullet
authorized Getinge and Getinge partner’s 3. Perform your tasks, e.g., logging in to Service &
personnel. Settings and starting the Touch screen calibration.
The Service key, and its associated access code, are 4. Disconnect the mouse, and then restart the
personal and must not be handed-over to anyone else. system.
The access code is unique for each Service key.

Notes:
• When saving information to a USB memory stick
bullet
(logs, screenshots, etc), allow this procedure to
complete before removing the memory stick.
Information will be shown on the control panel
when completed.

• Make sure to log out (select EXIT) from Service


bullet
& Settings when completed. If the Service key is
removed while Service & Settings is open,
access remains until manual exit.

5-2 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.2 Service & Settings


Service & Settings (S&S) is a built-in software provided
to facilitate troubleshooting, service and maintenance
of the system.

WARNING! Service & Settings must


not be activated with a patient
connected to the system.

Notes:
• It is only possible to access Service & Settings
bullet
with the system in Standby mode.

• The Service & Settings menus diagram in chapter


bullet
Diagrams gives an overview of the Service &
Settings functions.

5.2.1 To access Service & Settings

• Tap SERVICE & SETTINGS in the extended menu.


bullet
The Service & Settings login menu is displayed.

There are five options in the login menu:

• BIOMED. Tap BIOMED and enter the 4 digit code


bullet
1973 (default) to access this user level. The User’s
Manual describes user level BIOMED.

• SERVICE. Only for personnel trained and authorized


bullet
by Getinge. A Service key connected to the system
and the associated access code is required.

• REMOTE SERVICE. Remote Services start button.


bullet
• LICENSES. Licenses for third party software
bullet
packages. Listed for legal reasons.

• EXIT. To exit Service & Settings.


bullet

SERVO-air Ventilator System, Service Manual 5-3


| 5 | Service procedures |

5.2.2 To access the BIOMED user level:

• Insert the Service key.


bullet
• Tap BIOMED.
bullet
• Enter the access code and verify by tapping the
bullet
green check mark.

• The Service & Settings start screen is displayed.


bullet

The start screen shows the actual user level.

The following choices are available:

• STATUS
bullet
• LOGS
bullet
• SERVICE REPORT
bullet
• CONFIGURATION
bullet
• COPY CONFIGURATION
bullet
• INSTALLATION
bullet
• CALIBRATION
bullet
• EXIT
bullet
These Service & Settings main functions are described
later in this chapter.

5-4 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5-5


| 5 | Service procedures |

Status

STATUS is used for viewing system information,


installed software options and parts data.

The following information is available for System info:

• SYSTEM INFORMATION
bullet
• O2 CELL
bullet
• EXPIRATORY CASSETTE
bullet
• BATTERY STATUS
bullet

The following information is available for Options:


• INSTALLED OPTIONS (software options)
bullet

5-6 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5-7


| 5 | Service procedures |

Logs

LOGS is used for viewing Event logs for a certain


period of time. A date interval can also be set and a
search function is available.

The following search filters are available for Event logs:

• Alarms
bullet
• Settings
bullet
• Functions
bullet
• Configuration
bullet

Notes:
• Logs can be saved to a USB memory stick.
bullet
• Only a USB memory stick may be connected to
bullet
the USB port.

Save logs to USB:

• Insert a USB memory stick.


bullet
• Tap Save to USB. It is not necessary to select logs
bullet
before saving to USB. All required logs will be
copied to the USB memory stick.

• The message SAVE LOGS TO USB will be shown


bullet
during the file transfer. When completed, all log files
are saved in a ‘ventilatorData’ folder on the USB
memory.

• Remove the USB memory stick.


bullet

5-8 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5-9


| 5 | Service procedures |

Service report

SERVICE REPORT is used for reporting service tasks.

The following completed service tasks can be


confirmed:

• Report replacement of exp. cassette membrane


bullet
• Report completed preventive maintenance
bullet
• Report replaced backlight. Not active from System
bullet
version 3.0 and not available from System version
4.0.
Tap Open keyboard to enter comments and a
signature.

With the button Edit, it is possible to set a shorter time


until next Preventive Maintenance. This function can
be used if the time stamp for the previous Preventive
Maintenance has been lost, for example after
replacement of the Main back-plane.

5 - 10 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Configuration

CONFIGURATION is used for viewing and editing the


systems configuration.

The following choices are available and explained


below:

• SETTINGS
bullet
• STARTUP CONFIGURATION
bullet
• ALARMS
bullet
• NETWORK
bullet
• SET DATE & TIME
bullet
• BIOMED CODE
bullet

SERVO-air Ventilator System, Service Manual 5 - 11


| 5 | Service procedures |

5 - 12 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Settings

SETTINGS is divided into two sections; GENERAL


and UNITS.

The following settings are available in GENERAL:

• Remote Services after pre-use check. To set if log


bullet
files should be sent via Remote Services after
pre-use checks.

The unit settings for the following parameters are


available in UNITS:

• Height
bullet
• Weight
bullet

SERVO-air Ventilator System, Service Manual 5 - 13


| 5 | Service procedures |

5 - 14 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Startup configuration

STARTUP CONFIGURATION is used to set the


systems startup configuration.

In this menu, it is also possible to restore factory default


startup configuration.

Refer to the User’s Manual for information on startup


configuration.

SERVO-air Ventilator System, Service Manual 5 - 15


| 5 | Service procedures |

5 - 16 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Alarms

ALARMS is used to set the system alarm limits for:

• Patient category and type of ventilation:


bullet
- Pediatric
- Pediatric NIV
- Adult
- Adult NIV

• Restore default alarm limits.


bullet

• It is also possible to set Alarm sound level or


bullet
Restore default audio level.

SERVO-air Ventilator System, Service Manual 5 - 17


| 5 | Service procedures |

Network

NETWORK is used to display and configure network


parameters when connecting the system to Remote
Services.

In the Network menu, it is possible to select


Auto (DHCP) or Manual configuration.

For installation and setup instructions, refer to the


‘MCare Remote Services, SERVO-air ventilator system
– Installation Instructions’.

5 - 18 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 19


| 5 | Service procedures |

Set date & time

SET DATE & TIME is used to set the systems real time
clock.

5 - 20 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Biomed code

BIOMED CODE is used to set the four-digit code used


to log in as Biomed to Service & Settings.

The default biomed code is 1973.

SERVO-air Ventilator System, Service Manual 5 - 21


| 5 | Service procedures |

5 - 22 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Copy configuration

COPY CONFIGURATION is used to export or import


the systems configuration using a USB memory stick.

This function can for example be used to quickly install


the same configuration to several systems.

SERVO-air Ventilator System, Service Manual 5 - 23


| 5 | Service procedures |

Installation

INSTALLATION is used to install:

• PERMANENT OPTIONS. This function will install


bullet
Software options.

• SOFTWARE INSTALLATION. This function will


bullet
install a new System software version.
Installation is described separately in section ‘Software
installation’.

5 - 24 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 25


| 5 | Service procedures |

Exit

EXIT is used to log out from Service & Settings.

Make sure to log out (select EXIT) from Service &


Settings when completed. If the Service key is removed
with Service & Settings open, access to Service &
Settings remains.

5 - 26 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.2.3 To access the SERVICE user level:

• Insert the Service key.


bullet
• Tap SERVICE.
bullet
• Enter the access code and verify by tapping the
bullet
green check mark.

• The Service & Settings start screen is displayed.


bullet

The start screen shows the actual user level.

The following choices are available:

• STATUS
bullet
• LOGS
bullet
• TESTS
bullet
• SERVICE REPORT
bullet
• CONFIGURATION
bullet
• COPY CONFIGURATION
bullet
• INSTALLATION
bullet
• CALIBRATION
bullet
• EXIT
bullet
These Service & Settings main functions are described
later in this chapter.

SERVO-air Ventilator System, Service Manual 5 - 27


| 5 | Service procedures |

5 - 28 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Status

STATUS is used for viewing system information,


installed software options and parts data.

The following information is available for System info:

• SYSTEM INFORMATION
bullet
• O2 CELL
bullet
• EXPIRATORY CASSETTE
bullet
• BATTERY STATUS
bullet

The following information is available for Options:


• INSTALLED OPTIONS (software options)
bullet

The following information is available for Parts data


(parts with ID-PROM):
• PART (name)
bullet
• ARTICLE NUMBER
bullet
• VERSION NUMBER
bullet
• SERIAL NUMBER
bullet

SERVO-air Ventilator System, Service Manual 5 - 29


| 5 | Service procedures |

5 - 30 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Logs

LOGS is used for viewing Service logs and Event logs


for a certain period of time. A date interval can also be
set and a search function is available.

The following search filters are available for Service


logs:

• Test results
bullet
• Test details
bullet
• Preventive maintenance
bullet
• Service report
bullet
• Technical alarms
bullet
• Installation
bullet

The following search filters are available for Event logs:

• Alarms
bullet
• Settings
bullet
• Functions
bullet
• Configuration
bullet

Notes:
• Logs can be saved to a USB memory stick.
bullet
• Only a USB memory stick may be connected to
bullet
the USB port.

Save logs to USB:

• Insert a USB memory stick.


bullet
• Tap Save to USB. It is not necessary to select logs
bullet
before saving to USB. All required logs will be
copied to the USB memory stick.

• The message SAVE LOGS TO USB will be shown


bullet
during the file transfer. When completed, all log files
are saved in a ‘ventilatorData’ folder on the USB
memory.

• Remove the USB memory stick.


bullet

SERVO-air Ventilator System, Service Manual 5 - 31


| 5 | Service procedures |

5 - 32 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Tests

TESTS is used to perform a Pre-use check, either the


complete check, or any of the separate tests included
in the Pre-use check.

• Tap Select PUC tests to perform a complete Pre-use


bullet
check.

All Pre-use checks are described in chapter


Troubleshooting.

The TEST menu also contains choices to:

• Select breathing tests (selected in the screenshot)


bullet
• Select monitoring tests
bullet
• Select panel tests
bullet
These tests are included in Pre-use check, but when
performed in this menu, they are limited to each
subsystems function.

SERVO-air Ventilator System, Service Manual 5 - 33


| 5 | Service procedures |

5 - 34 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

The tool Manual panel tests is primarily intended for


manufacturing tests, but can also be used during
service or maintenance.

The Manual panel tests requires interaction with the


user and performs the following subtests:

• Device data. Shows device data.


bullet
• Calibration test. Tap on the circles to check the
bullet
touch screen calibration.

• Backlight test. To check the backlight at different


bullet
levels.

• LED test. Tap the swatches and check that the


bullet
corresponding color on the alarm indicator is lit.

• Pixel RGB test. The screen switches between


bullet
different colors to make it possible to check for
defective pixels.

• Sound test. To check different sound levels.


bullet
• Ethernet test. Checks the Ethernet communication
bullet
between the User interface and the Patient unit.
Tap Run to start the test. To continue the test
sequence, swipe diagonally on the screen.

After each test, the user can select Yes if the test
passed, No if the test failed or Cancel to cancel the
test. These results can be saved as a log file on a USB
memory stick.

SERVO-air Ventilator System, Service Manual 5 - 35


| 5 | Service procedures |

5 - 36 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Service report

SERVICE REPORT is used for reporting service tasks.

The following completed service tasks can be


confirmed:

• Report replacement of exp. cassette membrane


bullet
• Report completed preventive maintenance
bullet
• Report replaced backlight. Not active from System
bullet
version 3.0 and not available from System version
4.0.
Tap Open keyboard to enter comments and a
signature.

With the button Edit, it is possible to set a shorter time


until next Preventive Maintenance. This function can
be used if the time stamp for the previous Preventive
Maintenance has been lost, for example after
replacement of the Main back-plane.

SERVO-air Ventilator System, Service Manual 5 - 37


| 5 | Service procedures |

Configuration

CONFIGURATION is used for viewing and editing the


systems configuration.

The following choices are available and explained


below:

• SETTINGS
bullet
• STARTUP CONFIGURATION
bullet
• ALARMS
bullet
• NETWORK
bullet
• SET DATE & TIME
bullet
• BIOMED CODE
bullet

5 - 38 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 39


| 5 | Service procedures |

Settings

SETTINGS is divided into two sections; GENERAL


and UNITS.

The following settings are available in GENERAL:

• Date format
bullet
• Decimal separator
bullet
• Language
bullet
• PM reminder (On/Off)
bullet
• Enable biomed service report (On/Off). Will
bullet
enable/disable the SERVICE REPORT function in
Biomed user level. Available in Field service user
level only.

• Remote Services after pre-use check. To set if log


bullet
files should be sent via Remote Services after
pre-use checks.

• Flashing standby message (On/Off)


bullet

The unit settings for the following parameters are


available in UNITS:

• Airway pressure
bullet
• Flow
bullet
• Compliance
bullet
• Resistance
bullet
• Gas supply pressure
bullet
• Barometric pressure
bullet
• Height
bullet
• Weight
bullet

5 - 40 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 41


| 5 | Service procedures |

Startup configuration

STARTUP CONFIGURATION is used to set the


systems startup configuration.

In this menu, it is also possible to restore factory default


startup configuration.

Refer to the User’s Manual for information on startup


configuration.

5 - 42 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 43


| 5 | Service procedures |

Alarms

ALARMS is used to set the system alarm limits for:

• Patient category and type of ventilation:


bullet
- Pediatric
- Pediatric NIV
- Adult
- Adult NIV

• Restore default alarm limits.


bullet

• It is also possible to set Alarm sound level or


bullet
Restore default audio level.

5 - 44 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Network

NETWORK is used to display and configure network


parameters when connecting the system to Remote
Services.

In the Network menu, it is possible to select


Auto (DHCP) or Manual configuration.

For installation and setup instructions, refer to the


‘MCare Remote Services, SERVO-air ventilator system
– Installation Instructions’.

SERVO-air Ventilator System, Service Manual 5 - 45


| 5 | Service procedures |

5 - 46 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Set date & time

SET DATE & TIME is used to set the systems real time
clock.

SERVO-air Ventilator System, Service Manual 5 - 47


| 5 | Service procedures |

Biomed code

BIOMED CODE is used to set the four-digit code used


to log in as Biomed to Service & Settings.

The default biomed code is 1973.

5 - 48 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

SERVO-air Ventilator System, Service Manual 5 - 49


| 5 | Service procedures |

Copy configuration

COPY CONFIGURATION is used to export or import


the systems configuration using a USB memory stick.

This function can for example be used to quickly install


the same configuration to several systems.

5 - 50 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Installation

INSTALLATION is used to install:

• PERMANENT OPTIONS. This function will install


bullet
Software options.

• SOFTWARE INSTALLATION. This function will


bullet
install a new System software version.
Installation is described separately in section ‘Software
installation’.

SERVO-air Ventilator System, Service Manual 5 - 51


| 5 | Service procedures |

5 - 52 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Calibration

CALIBRATION is used for calibration of:


• Barometer
bullet
• Touch screen
bullet

Barometer

Barometer calibration is used to calibrate the internal


barometer. A barometer or information about the actual
barometric pressure at the installation site is required.

Note: The actual barometric pressure is affected by


altitude and weather. Make sure that the absolute
barometric pressure at your current location and
elevation is used. Do not use barometric pressure
readings from a weather service as these typically
report barometric pressure at sea level.

The barometer must be calibrated:

• If the Barometer test in Pre-use check fails.


bullet
• If a Technical error code indicates Barometer error.
bullet
• If the barometric pressure check performed during
bullet
installation or Preventive maintenance shows that
calibration is required, that is if the Barometric
pressure shown in the SYSTEM STATUS > General
window differs more than ±5% from the actual
barometric pressure.

• After replacement of PC 1992 Monitoring or in case


bullet
of malfunction in the memory backup battery on
PC 1992.

To calibrate the barometer:

• Tap CALIBRATION and then tap BAROMETER.


bullet
• Tap Update.
bullet
• Enter the actual barometric pressure and tap the
bullet
OK button.
After calibration, check that the new Barometric
pressure value corresponds to the set value. Note that
the new Barometric pressure value may differ 1-2 hPa
from the set value.

SERVO-air Ventilator System, Service Manual 5 - 53


| 5 | Service procedures |

5 - 54 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Touch screen

The touch screen must be calibrated:

• If an inaccuracy is seen in the touch screen function.


bullet
• After replacement of PC 2023 Panel.
bullet

To calibrate the touch screen:

• Tap CALIBRATION and then tap TOUCHSCREEN.


bullet
• Tap Start calibration.
bullet

• Use a narrow-tipped tool to tap each of the nine line


bullet
intersections that are presented on the screen.

Note: Do not use sharp tools on the screen.

• Tap each of the nine circles presented on the


bullet
screen.

When the calibration is completed, the Start calibration


menu appears.

Note: If the touch screen is not calibrated well


enough, it can be difficult to enter Service & Settings
to start the calibration procedure. If so, a mouse can
be connected to the system to facilitate the
navigation. For instructions, refer to Connecting a
USB mouse to the Patient unit in chapter Service
procedures, section Service key.

SERVO-air Ventilator System, Service Manual 5 - 55


| 5 | Service procedures |

5 - 56 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Exit

EXIT is used to log out from Service & Settings.

Make sure to log out (select EXIT) from Service &


Settings when completed. If the Service key is removed
with Service & Settings open, access to Service &
Settings remains.

SERVO-air Ventilator System, Service Manual 5 - 57


| 5 | Service procedures |

5.3 Software installation 5.3.3 Check after installation

5.3.1 General • A software installation may change the system


bullet
functionality and therefore require a new version of
• This chapter describes installation of System
bullet the User's Manual.
software and Software options.
• After any software installation in the system, perform
bullet
• Only personnel trained and authorized by Getinge
bullet a Pre-use check according to instructions in the
shall be permitted to install System software and User's Manual.
Software options.
CAUTION: The system must not be
• Software installation is performed in Service &
bullet
switched off during the software
Settings. The Software installation function is
installation process. Such interruption
available in BIOMED and SERVICE user levels.
may cause malfunctions on PC boards.
With BIOMED user level, Service key is not
required.

• For functionality enhancement, the latest released


bullet
5.3.4 Materials/documents required
System software version is always recommended.

• USB memory stick with the System software


bullet
5.3.2 Check before installation version/Software option to be installed. Only Maquet
approved USB memory sticks must be used.
• It is not recommended to install System software
bullet
• User’s Manual.
bullet
with lower version number than already installed in
the system: • Gas supply required for the Pre-use check.
bullet
- Check installed System software version, see
System SW version in SYSTEM STATUS >
General.
- Check the System software version stored on
the USB memory stick.

• Some parts of the configuration may be changed


bullet
during the System software installation. With
System version 4.0 or higher, the Startup
configuration will automatically be copied to and
reinstalled from the USB memory stick as part of
the software installation procedure. For lower
system versions, copy the Startup configuration to
a USB memory stick prior to a software installation
and reinstall the configuration afterwards.

• Make sure that the software is fully compatible with


bullet
all hardware, software and mechanical components
in the system. See the Compatibility chart in chapter
Revision history.

5 - 58 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.3.5 Software information

• System software: A System software installation


bullet

Software option
installs a new System software version. A System Software option

installation
software installation is not depending on the serial SW Option
Software option
number of the system and does not alter the SW Option
installed Software options. Software option

• Software option: A Software option installation


bullet
installs Software options ordered for the concerned
System software

System software
system. A Software option is individually created for

installation
each system and can only be installed on the
Corrections
intended system. Serial number of the system must and function
be stated when ordering a Software option. improvement

S25085

SERVO-air Ventilator System, Service Manual 5 - 59


| 5 | Service procedures |

5.3.6 System software installation

To perform a System software installation:

• Turn the system On.


bullet
• Enter Service & Settings (BIOMED or SERVICE
bullet
user level).

• With System version 4.0 or higher, the Startup


bullet
configuration will automatically be copied to the USB
memory stick as part of the software installation
procedure. For lower system versions, copy the
configuration to the USB memory stick.

• Connect the USB memory stick, with the System


bullet
software version to be installed.

• Tap INSTALLATION.
bullet
• Tap SOFTWARE INSTALLATION.
bullet
• The software installation activation box will now
bullet
open. Check that correct System SW version to be
installed is stated in this box.

• Tap Yes to start the installation. The software


bullet
installer screen will now open and the installation
will start.

• The installer screen now shows the progress of the


bullet
installation.

• When the installation is completed (after 10–11


bullet
minutes), this is notified in the installer screen. See
Note 1 below.

• Restart the system. See Note 2 below.


bullet
• System version 4.0 or higher, the Startup
bullet
configuration will automatically be copied from the
USB memory stick as part of the software installation
procedure. For lower system versions, connect the
USB memory stick containing the configuration and
copy the configuration from the USB memory stick.

• Perform a Pre-use check according to instructions


bullet
in the User's Manual.

Note 1: Very occasionally, it may seem that the Note 2: After a software installation, technical alarms
installation is interrupted and the Service & Settings may be activated due to reset internal memory on the
login menu will appear. If so, do not cancel the PC boards. Restart the system once more and check
installation. The installation will proceed in the that no technical alarms are activated.
background. Wait 15 minutes before restarting the
system.

5 - 60 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.3.7 Software option installation

To perform a Software option installation:

• Turn the system On.


bullet
• Enter Service & Settings (BIOMED or SERVICE
bullet
user level).

• Connect the USB memory stick with the Software


bullet
options to be installed.

• Tap INSTALLATION.
bullet
• Tap PERMANENT OPTIONS.
bullet

• The software installation activation box will now


bullet
open. Furthermore, all software options that will be
available in the system after the installation will be
listed on the screen.

• Tap Yes to start the installation.


bullet
• When the installation is completed (after a few
bullet
seconds), this is notified in the installer screen.

• Wait two minutes to allow the Service log to be


bullet
updated.

• Restart the system.


bullet
• Perform a Pre-use check according to instructions
bullet
in the User's Manual.

SERVO-air Ventilator System, Service Manual 5 - 61


| 5 | Service procedures |

5.4 Leakage detection Test case 3 – Inspiratory section:

5.4.1 General • Nozzle unit in O2 gas module


bullet

When a leakage is detected in the Internal leakage • Gas module silicone connector muff
bullet
test, the Pre-use check will stop and display either • Inspiratory gas docking incl. one-way valve
bullet
Leakage or Excessive leakage. To locate the leakage,
the three leakage tests described below can be used.
• Inspiratory flowmeter connector muff
bullet
• Inspiratory flowmeter
bullet
Test equipment:
• Inspiratory channel connector muff
bullet

• Test tube
bullet • Inspiratory pressure silicone tube
bullet

• Leakage detector tool, P/N 68 81 146


bullet • Safety valve membrane
bullet

• Manometer, local purchase (e.g. a Cuff filling


bullet • Safety valve pull magnet
bullet
measuring device with manometer) • O2 cell bacteria filter seal
bullet
• Inspiratory pressure silicone tube
bullet
5.4.2 Parts tested during the check
• Inspiratory pressure bacteria filter
bullet
Test case 1 – Test tube: • Inspiratory bacteria filter holder incl. tube
bullet

• Test tube
bullet • Inspiratory pressure transducer
bullet

Test case 2 – Expiratory channel:

• Expiratory inlet incl. moisture trap with O-ring


bullet
• Expiratory cassette
bullet
• Expiratory valve membrane
bullet
• Expiratory pressure transducer tube
bullet
• Expiratory pressure transducer
bullet
• Expiratory valve coil
bullet

5 - 62 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.4.3 Starting the leakage detection

• Start the manual leakage detection when the


bullet
Pre-use check has stopped during the 'Internal
leakage test' and displayed either Leakage or
Excessive leakage.

• Do not cancel the Pre-use check, the Redo function


bullet
will be used during the leakage detection.

5.4.4 Test case 1 - Test tube

1. Check the test tube for leakage. Replace if


required.
2. Reconnect the test tube properly.
3. Tap Redo.
If the leakage remains, continue with Test case 2.

Insp.
Exp.

S20087

SERVO-air Ventilator System, Service Manual 5 - 63


| 5 | Service procedures |

5.4.5 Test case 2 – Expiratory channel

Expiratory channel

H PC 1786 Exp. channel cassette


Expiratory inlet Expiratory B A
including moisture cassette PC 2031/
trap with O-ring PC 1785 Expiratory
Expiratory valve
channel membrane
connector

PC 2031/ Expiratory pressure


PC 1785 transducer tube

Expiratory
valve coil
Part of leakage
detector kit

Expiratory
pressure
transducer PC 1781

P40 N40
P50
MOD1

Manometer

PC 2004/PC 2024 Expiratory channel

S20088

1. Connect the test equipment to the expiratory inlet.


2. Tap Redo.
3. When the expiratory valve closes, pressurize to
approx. 50 cmH2O. This must be done quickly
because the expiratory valve coil is closed only
for 2-3 seconds. A pressure drop indicates a
leakage.
If a leakage is detected in Expiratory channel,
continue with Expiratory valve coil. If not, go to Test
case 3.

5 - 64 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Expiratory valve coil

1. Remove the expiratory cassette.


2. Tap Redo.
3. Make sure the expiratory valve coil axis is
activated (pushed up).
If the axis is activated, continue with Expiratory
cassette. If not, check/replace the items listed below,
one at a time:
• Check the expiratory valve coil including cable and
bullet
connector.

• Replace PC 2004/PC 2024 Expiratory Channel.


bullet
• Replace PC 1991 Control.
bullet

Expiratory cassette

1. Connect the test equipment to the expiratory inlet


Press
(while the cassette is removed).
2. Block the expiratory pressure transducer channel
hole on the cassette and close the expiratory valve
by hand, see illustration.
3. Pressurize to approx. 50 cmH2O. If a leakage is
detected check/replace the following parts (one at
a time):

• Expiratory inlet incl. moisture trap with O-ring.


bullet
• Expiratory valve membrane. The membrane
bullet
may not close properly due to cleaning residues
on the membranes sealing surface and/or on
the corresponding sealing surface in the
S20089
cassette.

• Expiratory cassette.
bullet

Expiratory pressure transducer and tube

1. If a leakage is detected in the Expiratory channel


but still remains after performed tests Expiratory
valve coil and Expiratory cassette, check the
Expiratory pressure transducer and Expiratory
pressure transducer tube.

SERVO-air Ventilator System, Service Manual 5 - 65


| 5 | Service procedures |

5.4.6 Test case 3 – Inspiratory section

O2 cell O2 gas module


O2 cell
bacteria Nozzle unit O2
Inspiratory filter seal in gas inlet
outlet Inspiratory module
Inspiratory channel Inspiratory gas
pipe docking
Insp. Ins
flo pi
pressure wm rat
tube eteory
Part of leakage Safety valve r
detector kit membrane Insp. Turbine module
pressure PC 1781
Manometer Inspiratory Air
bacteria inlet
Safety valve filter and holder pressure
pull magnet transducer

N17P17
S20090

Complete Inspiratory section Safety valve piston

1. Connect the test equipment to the Inspiratory If no leakage was found in the above mentioned parts,
outlet. remove the Inspiratory channel and restart the Pre-use
2. Pressurize to approx. 50 cmH2O. A pressure drop check. Check that the Safety valve piston is activated
indicates a leakage. If a leakage is indicated, the during the internal leakage test.
parts below should be checked/replaced:
If the piston is activated, continue with Inspiratory gas
• Inspiratory channel.
bullet docking and Inspiratory pipe. If not, check/replace
Note: In order to remove the Inspiratory as follows (one at a time):
channel, the safety valve must be opened (it is
closed during the test). This is done by • Check the safety valve piston including cable and
bullet
cancelling the test and setting the ventilator to connector.
Standby.
• Replace PC 2030 Inspiratory Channel.
bullet
• O2 cell bacteria filter seal.
bullet • Replace PC 2004/PC 2024 Expiratory Channel.
bullet
• Safety valve membrane. Clean if necessary.
bullet
• Inspiratory pressure tube.
bullet
• Inspiratory pressure bacteria filter and holder.
bullet

5 - 66 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

Inspiratory gas docking and Inspiratory pipe

Gas module silicone O2 gas module


Inspiratory connector muff
channel
connector Nozzle unit O2
muff in O2 gas inlet
Inspiratory module
gas
docking
In
flo spi
Part of leakage wm rat
detector kit eteory
r
Turbine module
Inspiratory pipe
Manometer Air
inlet
Inspiratory
flowmeter
connector
muff
S20091

1. Connect the test equipment to the inspiratory


channel connector muff outlet, i.e. the inspiratory
channel removed, see picture.
2. Pressurize to approx. 50 cmH2O. A pressure drop
indicates a leakage. If a leakage is indicated, the
inspiratory section must be opened for further
investigation. See chapter Disassembling and
Assembling. Check/replace the following parts in
the Inspiratory section:

• Nozzle unit in the O2 gas module.


bullet
• Gas module silicone connector muff.
bullet
• Inspiratory gas docking incl. one-way valve.
bullet
Clean the valve if necessary.

• Inspiratory flowmeter connector muff.


bullet
• Inspiratory flowmeter.
bullet
• Inspiratory channel connector muff.
bullet

Note: The Inspiratory pressure transducer and the


holder for inspiratory pressure bacteria filter are
bypassed in this test. They should be checked
visually while the section is opened.

5.4.7 Completing the Leakage detection

1. Restore the unit including test tube.


2. Select Redo and make sure the test passes.

SERVO-air Ventilator System, Service Manual 5 - 67


| 5 | Service procedures |

5.5 Battery modules


5.5.1 General

• The batteries condition and capacity are verified


bullet
during the Pre-use check and also continuously
during operation of the unit. No further function
checks of the batteries are required.

• Typical recharge time approximately 2 h/battery


bullet
(90 %), up to 3 h (100 %) if battery is completely
discharged. If a battery is fully discharged, e.g. due
to long storage time, each battery may require up
to 6 hours charging time.

• The System software will monitor:


bullet
- Remaining life time. The Battery life time is
300 charge cycles or four years (48 months) from
manufacturing date, whichever comes first.
- If the operational capacity is too poor for
continued usage.
In both cases, battery replacement information will
be shown on the User interface.

CAUTION: Do not expose the batteries


to water, fire or excessive heat. Do not
crush, disassemble, puncture or short
circuit the connector terminals.

5 - 68 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.5.2 Battery status

Select SYSTEM STATUS > Batteries to open the


Battery status window.

The Battery status window shows:

• Battery capacity. Calculated as the sum of the


bullet
estimated backup time for the connected battery
modules reduced with 20 minutes as an extra safety
feature. This estimate may differ from the actual
usable backup time during running. Usable backup
time depends on set mode and selected ventilation
settings.

• Slot number.
bullet
• Battery module S/N.
bullet
• Remaining operating time in minutes for each
bullet
battery.

• Charge indicator for each battery. If the ventilator


bullet
system is running on battery power, the indicator
for the active battery turns yellow.

• Remaining life time (months) and charge cycles for


bullet
each battery.
If Replace battery-message is displayed, the battery
is unreliable, regardless of the operating time displayed
under Batteries. In this situation, replace the battery
even when the STATUS window indicates that
significant operating time remains.

At least one battery module must always be installed.

SERVO-air Ventilator System, Service Manual 5 - 69


| 5 | Service procedures |

5.6 Memory backup batteries


5.6.1 General

• The Memory backup batteries must be replaced


bullet
after five years. A Technical error message will
appear on the screen if the battery voltage level is
too low.

• Always replace both batteries at the same time to


bullet
keep the same replacement date for both batteries.

• After any maintenance or service of the system,


bullet
perform a Pre-use check. Refer to the User's
Manual.

Note: ESD sensitive components.


The Memory backup batteries are
mounted on ESD sensitive PC boards.
Refer to chapter Disassembling and
assembling, section Handling PC
boards for further information regarding
ESD sensitive components.

5.6.2 Preparations

• Turn Off the system.


bullet
• Disconnect the mains power cable.
bullet
• Disconnect the O2 gas supplies (wall and/or
bullet
cylinder).

• Remove patient tubing.


bullet

5 - 70 SERVO-air Ventilator System, Service Manual


| Service procedures | 5 |

5.6.3 Replacing Memory backup battery

• Remove the Patient unit front cover. Refer to chapter


bullet
s d
Disassembling and assembling.

• Carefully pull out PC 1991 and PC 1992.


bullet
Note: Do not remove the cable connector (1).
Certain information stored in the PC backup is
disconnected:
- The battery backup on PC 1992 supplies power
to the real time clock and to the memory
containing logs and settings for the last minutes,
warm and cold start information, all of which is
lost at disconnection.
- The battery backup on PC 1991 supplies power
to the memory containing startup ventilation
configuration, which is lost at disconnection.

• There are two equal battery connectors (2 and 3)


bullet
on the PC boards. Connect the new battery to the a f
S20093
un-used connector. The memory functions are now
secured by the new battery.

• Cut the cable ties holding the old battery (4) to the
bullet
PC board.

• Disconnect and remove the old battery.


bullet
• Mount the new battery onto the PC board using new
bullet
cable ties as shown in the illustration.

• Insert the PC board into the correct PC board slot


bullet
and reassemble the Patient unit.

SERVO-air Ventilator System, Service Manual 5 - 71


| 5 | Service procedures |

5 - 72 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6 Troubleshooting

Table of Contents
6.1 General | 6-2
6.2 Start-up test | 6-2
6.3 Pre-use check | 6-3
6.4 Other tests | 6-6
6.5 Test logs | 6-7
6.6 Technical error codes and messages | 6 -23

SERVO-air Ventilator System, Service Manual 6-1


| 6 | Troubleshooting |

6.1 General
The information in this chapter is applicable to System The troubleshooting guides below focus on the
version 1.0 – 4.0. For functionality enhancement, the technical problems. Information about clinically related
latest System software version is recommended. problems can be found in the User's Manual.

This chapter contains: Possible causes to malfunction, not mentioned in the


• Descriptions of the Start-up test and the Pre-use
bullet following troubleshooting guides, that need to be
check. considered during troubleshooting:

• Information about how to interpret the Test log and


bullet • Incorrectly assembled system after cleaning,
bullet
the Technical error codes as well as recommended maintenance or service.
actions to eliminate malfunctions reported by the • Disconnected or incorrectly connected cable
bullet
system. connectors, PC board connectors or interconnection
Before starting troubleshooting, try to eliminate all boards.
possibilities of operational errors. If the malfunction • Disconnected or defective gas tubes.
bullet
remains, use the troubleshooting guides below as well
as the information in chapter Description of functions In the descriptions and troubleshooting guides below,
to locate the faulty part. Perform actions step by step the tested subsystems are referred to as:
and check that the malfunction is eliminated. • BRE = Breathing
bullet
When the fault is corrected, carry out a complete • MON = Monitoring
bullet
Pre-use check as described in the User's Manual. • PAN = Panel.
bullet

6.2 Start-up test

Internal technical tests Recommended action if the test fails

• Software check
bullet 1. Restart the system. Do not touch the User interface
during system start-up. Interfering with the touch screen,
• Reading EEPROM
bullet loudspeaker grid and so forth, may affect the internal
• Checksum EEPROM
bullet technical tests.
2. Reinstall the System software.
• Audio test.
bullet
(BRE + MON + PAN)

6-2 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.3 Pre-use check Temperature and Moist sensor test (MON)

The system requests the user to start the automatic Checks the following sensors in the inspiratory section:
Pre-use check at every startup of the system. It is also
possible to select the Pre-use check via the Standby • Humidity (0.0–100.0 %). Sensor on PC 2030
bullet
menu. Inspiratory channel.

The User's Manual describes how to perform this


• Turbine temperature (0.0–70.0 °C). Sensor inside
bullet
Turbine module.
Pre-use check. The Pre-use check description on the
following pages gives more detailed information about • Ambient temperature (0.0–55.0 °C). Sensor on
bullet
the Pre-use check. PC 2030 Inspiratory channel.

Some of the recommended actions described below


• Inspiratory gas temperature. (0.0–60.0 °C). Sensor
bullet
inside Inspiratory flowmeter.
refer to Service & Settings (S&S). The Service key is
required to access S&S. Troubleshooting can of Blower inlet test (MON)
course be performed without access to the S&S, but
for some of the recommended actions, S&S will make Checks the turbine module inlet:
troubleshooting faster and easier.
• Inlet resistance. 1.00–3.30 cmH2O/l/s at
bullet
Check if the fault remains after each performed service 2.70–3.30 l/s inspiratory flow.
action. Re-run the complete Pre-use check or run the
concerned test from S&S. • Inlet pressure sensor. Pressure diff. sensor test
bullet
offset steps 1586–1690.
In the Standby menu it is possible to perform a
separate Patient circuit test to evaluate circuit leakage Alarm output connector test (MON)
and measure circuit compliance. This separate test
Remote alarm test in the S&S Test menu.
does not replace the complete Pre-use check.
Checks that the Alarm output connector relay switches
6.3.1 Internal test sequence between open and closed.

Panel audio test (PAN) Power failure test (MON)

Checks that the loudspeakers in the User interface Generates failures on different system voltages.
generate an adequate sound level. A sound level is Checks that these failures are detected.
generated and a microphone mounted nearby the
loudspeakers measures the sound. PAN calculates 6.3.2 Barometer test sequence
the measured value to verify that the sound level
meets the requirement. Barometer test (MON)

Monitor audio test (MON) Checks that the barometric pressure measured by the
internal barometer is within 650–1070 hPa (mbar).
Checks that the buzzer in the Patient unit generates
an adequate sound level. A power failure is generated Checks that the measured barometric pressure values
in order to activate the buzzer. A microphone mounted between BRE and MON differ less than 8 hPa (mbar).
nearby the buzzer measures the sound. MON
calculates the measured value to verify that the sound
level meets the requirement.

SERVO-air Ventilator System, Service Manual 6-3


| 6 | Troubleshooting |

6.3.3 Internal leakage test sequence • Checks that acceleration (rise time) from approx.
bullet
5000 to 40000 r/min is <0.150 s.
Internal leakage test (BRE)
• Checks that speed is 38000–42000 r/min.
bullet
Checks the internal leakage, with the test tube
connected, using the inspiratory and expiratory
• Checks that deceleration (fall time) from approx.
bullet
40000 to 5000 r/min is <0.150 s.
pressure transducers:

• Generates 50 cmH2O pressure.


bullet 6.3.5 Pressure transducer test sequence

• Checks that the measured pressure values differ


bullet Breathing pressure transducer test (BRE)
less than 10 cmH2O between inspiratory and
expiratory pressure transducers. Checks that the offset values for the pressure
transducers do not differ more than ±6 cmH2O (±0.3 V)
• Checks that the system volume required to generate
bullet
from factory calibration.
50 cmH2O is 0.2000–0.4000 l.

• Checks that the leakage at 50 cmH2O is maximum


bullet Monitor pressure transducer test (MON)
0.0200 l/min.
Calibrates and checks the inspiratory and expiratory
To locate a leakage, a leakage detection test can be pressure transducers:
performed, see Leakage detection in chapter Service
procedures. The Leakage detection test is a
• The new offset values for the inspiratory and
bullet
expiratory pressure transducers may not differ more
complement to the recommended actions below.
than ±6 cmH2O (±0.3 V) from factory calibration.

6.3.4 Turbine and Gas supply test sequence • With the inspiratory pressure transducers factory
bullet
default gain used as a reference, a new gain factor
Gas supply test (MON) is set for the expiratory pressure transducer. The
new gain factor may not differ more than ±8% from
Checks gas supply: factory calibration.

• Start with a short gas flush.


bullet • Checks that the difference between BRE and MON
bullet
is less than ±0.8 cmH2O at 50 cmH2O during the
• Checks that the O2 gas supply pressure measured
bullet
test.
by the internal gas supply pressure transducer is
within 200–600 kPa (2000–6000 mbar). • Checks that the pressure generated for the Monitor
bullet
pressure transducer tests is 60 cmH2O (±5 cmH2O).
• Checks that the O2 supply pressure differs less than
bullet
100 hPa (mbar) during the test.
Expiratory valve test (BRE)
• Checks that the measured supply gas pressure
bullet
values between MON and BRE differ less than Checks the expiratory valve at four different pressures:
20 hPa (mbar). 40, 20, 10 and 5 cmH2O. Measured values must be
within ±15 cmH2O.
This test requires O2 gas supply.
Checks offset and gain in the expiratory valve:
Turbine accuracy test (MON)
• Offset (0.06768–0.15684 A).
bullet
Checks the turbine module:
• Gain (0.00730–0.01130 A/cmH2O).
bullet

• Checks that operating hours is <40000 h.


bullet This test requires O2 gas supply.
• Checks that speed is 4750–5250 r/min.
bullet

6-4 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.3.6 Safety valve test sequence • Checks that measured inspiratory flow differs less
bullet
than 5% from the flow delivered from the O2 gas
Monitor safety valve test (MON)
module at a flow of 0.5 l/s (30 l/min).
Checks the hardware signals related to the safety • Checks that the difference between MON and BRE
bullet
valve functions. subsystems is less than 0.005 l/s (<1%).

Breathing safety valve test (BRE) This test requires O2 gas supply.

Checks that the safety valves opening pressure is Breathing flow transducer test (BRE)
approx. 117 cmH2O and calibrates the valve if
required. The test can be repeated up to five times to Checks the offset values for:
achieve a Passed result:
• Inspiratory flow (22.021–24.609 V).
bullet
• Acceptance interval for the first test is
bullet • O2 flow (0.000–0.081 V).
bullet
115.50–118.50 cmH2O.
• Expiratory flow (0.093–0.142 V).
bullet
• Acceptance interval for the following tests is
bullet
114.00–120.00 cmH2O. Flow test (MON)
• The test fails if an opening pressure within
bullet
Monitoring flow test in the S&S Test menu.
114.00–120.00 cmH2O is not reached after five
attempts. Checks the offset values for:

This test requires O2 gas supply. • O2 flow (0.000–0.033 V).


bullet

6.3.7 O2 cell test sequence


• Expiratory flow (0.024–0.072 V).
bullet

With 60 % O2 concentration and 1 l/s gas flow:


O2 cell test (MON)

Checks and calibrates the O2 cell: • Checks inspiratory flow (0.9798–1.0198 l/s).
bullet
• Checks O2 flow (0.4842–0.5040 l/s).
bullet
• Checks if the O2 cell is consumed. Expected life
bullet
• Measures expiratory flow and calculates gain
bullet
time 500 000 %hours.
(k value) for the expiratory cassette (0.90–1.15).
• Checks with 100 % O2 (>70.0 %).
bullet
With 100 % O2 concentration and 0.5 l/s gas flow:
• Checks with 21 % O2 (<25.0 %).
bullet
• Checks offset (4–24 mV).
bullet • Checks inspiratory flow BRE and MON
bullet
(0.4750–0.5250 l/s).
• Calculates gain.
bullet
• Checks O2 flow BRE and MON (0.4900–0.5100 l/s).
bullet
This test requires O2 gas supply.
• Checks that the difference between O2 flow BRE
bullet
and MON is less than 0.005 l/s (<1 %).
6.3.8 Flow transducer test sequence
• Checks expiratory flow MON (±5 % of delivered
bullet
Insp flow sensor test (MON) flow).

Sensirion flow sensor test in the S&S Test menu. • Checks that the difference between expiratory flow
bullet
BRE and MON is less than 0.005 l/s (<1 %).
Checks the inspiratory flowmeter:

• Checks offset (9920–10080).


bullet

SERVO-air Ventilator System, Service Manual 6-5


| 6 | Troubleshooting |

With 21 % O2 concentration and 0.5 l/s gas flow: 6.3.10 Patient circuit test sequence

• Checks O2 flow BRE and MON (-0.0016–0.0016 l/s).


bullet Patient circuit test (BRE)

• Checks inspiratory flow BRE and MON


bullet Checks the system with patient tubing connected:
(0.4900–0.5100 l/s).

• Checks difference between inspiratory flow BRE


bullet • Checks that system volume is 0.1500–6.0000 l at
bullet
and MON is less than 0.005 l/s (<1 %). 50 cmH2O (>40 cmH2O).

• Checks expiratory flow MON (±5 % of delivered


bullet • Checks that compressible volume is
bullet
flow). 0.02–6.00 ml/cmH2O.

• Checks difference between expiratory flow BRE


bullet • Checks that leakage is <0.0800 l/min at 50 cmH2O
bullet
and MON is less than 0.005 l/s (<1 %). (>40 cmH2O).

With 0 l/s gas flow:


• Checks flow resistance at 10 l/min.
bullet
• Checks flow resistance at 60 l/min.
bullet
• Checks inspiratory flow BRE and MON
bullet
The system volume and compressible volume checks
(-0.0016–0.0016 l/s).
can be repeated up to three times to achieve a stable
If O2 is not connected, the check is performed with compressible volume value.
Air. The Flow test passes if the result of this check
corresponds to the old calibration factor from the 6.3.11 Alarm state test sequence
previous Pre-use check. The same expiratory cassette
External alarm system test (MON)
must be used.
If the option Alarm output is enabled, the user can test
6.3.9 Battery switch test sequence the external alarm system.

Battery switch test (MON) For pin configuration and signal names in the Alarm
output connector N67, see Connectors in chapter
Checks the battery module(s):
Diagrams. For further information, refer to the 'Alarm
output connector – Reference Manual'.
• Checks that a Maquet battery module is connected.
bullet
• Checks the status, age and number of charge
bullet Alarm state test (MON)
cycles.
Checks that no Technical error alarms are active
• Checks that there is a battery module in slot 2 (main
bullet during the Pre-use check.
slot).

• Checks that the power supply switches to battery


bullet 6.4 Other tests
when mains power is disconnected.
6.4.1 Patient circuit test sequence
• Checks that the power supply switches back to
bullet
mains power when reconnected. Patient circuit test (BRE)

This test is not performed if: This test performs the Patient circuit test separately,
as described above.
• Less than 10 minutes backup time remains in the
bullet
connected battery modules.

• No battery module is connected.


bullet

6-6 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.5 Test logs


This section describes how to interpret the Test logs
related to the Pre-use check.

6.5.1 Interpreting the Test result logs

The Pre-use check, complete test as well as separate Some checks will be repeated if they fail. If they pass
tests, is logged in the Service log. Select Test results when repeated, the check and the complete test
and Test details to show all test result logs. A Service sequence will pass. This will be shown in the log.
key is required to access the Service log. The
description below shows how to interpret the When a check fails due to a user error; a Dialog Choice
Service log. (DIALOG) appears and gives the user the choice to
redo the check by choosing Redo or, to bypass the
If one or more checks within a test fail, the log will show check by choosing Cancel.
which check that failed and also include an error code.
These error codes are called Check Failure Identifier Always perform recommended actions in a
(CFI). All checks that fail during the Pre-use check chronological order. A test that fails can cause
have a unique CFI. Possible causes and recommended subsequent tests to fail.
actions for checks that fail permanently or intermittently
are listed in the following CFI tables.

Acronyms in test logs


Acronym Test log entry type Acronym Test log entry type
DIALOG Dialog choise SFIN Test sequence result
FIN Test result SSTA Test sequence start
INF Info STA Test start
MEA Measurement value

SERVO-air Ventilator System, Service Manual 6-7


| 6 | Troubleshooting |

Test log example 1

The log example below shows an Insp flow sensor


test. This test is included in the Flow transducer test
sequence:

1. The Insp flow sensor test starts.


2. The measured Insp flow sensor total flow (0.5279
l/s) is outside the tolerance interval.
3. The check of measured Insp flow sensor total flow
failed with CFI 4 in the log post.
4. The test sequence result is failed with CFI 4 in
the log post.
5. The complete Insp flow sensor test result is failed
with CFI 4 in the log post.
2017-10-20 19:37:38 [STA] Insp flow sensor test
2017-10-20 19:37:40 [CHK_SSTA] Insp. flow transducer test sequence
2017-10-20 19:37:40 [CHK_STA] Insp flow sensor test offset check
2017-10-20 19:37:40 [MEA] Insp flow sensor test offset steps = 10001 - Tolerance Interval = [ 9920; 10080 ] ( Unit : - )
2017-10-20 19:37:40 [CHK_FIN] PASSED
2017-10-20 19:37:40 [CHK_SFIN] PASSED
2017-10-20 19:37:43 [CHK_SSTA] Insp flow sensor test O2 measurement check sequence
2017-10-20 19:37:43 [CHK_STA] Insp flow sensor test Insp total flow check
2017-10-20 19:37:43 [MEA] Insp flow sensor test total flow = 0.5279 - Tolerance Interval = [ 0.4736; 0.5235 ] ( Unit : l/s )
2017-10-20 19:37:43 [CHK_FIN] FAILED : Insp flow sensor test, sensirion measurement failure [Error Code = 4]
2017-10-20 19:37:43 [CHK_STA] Insp flow sensor test Bre - Mon Insp total flow check
2017-10-20 19:37:43 [MEA] Insp flow sensor test Bre - Mon Insp total flow: Mon flow = 0.5279 - set flow = 0.5000 - Tolerance interval =
[ 0.5223; 0.5323 ] ( Unit : l/s )
2017-10-20 19:37:43 [CHK_FIN] PASSED
2017-10-20 19:37:43 [CHK_SFIN] FAILED - [Error Code = 4]
2017-10-20 19:37:43 [FIN] Insp flow sensor test FAILED - [Error Code = 4] S20138

6-8 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Test log example 2

The log example below shows the Turbine and Gas


supply test sequence:

1. The Turbine and Gas supply test sequence starts.


2. The Gas supply test starts.
3. The measured O2 supply pressure (2 mbar) is
outside the tolerance interval.
4. The O2 supply pressure is too low. The O2 supply
pressure test failed with CFI 2 in the log post
5. The Gas supply test failed with CFI 2 in the log
post.
6. The different turbine test passed, but the complete
Turbine and Gas supply test sequence failed due
to the low O2 pressure in the Gas supply test.

2018-01-19 10:09:38 [SSTA] Turbine and Gas supply test sequence


2018-01-19 10:09:40 [STA] Gas supply test
2018-01-19 10:09:42 [CHK_STA] Checking O2 supply pressure.
2018-01-19 10:09:42 [MEA] O2 supply pressure = 2 - Tolerance Interval = [ 2000; 6000 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] FAILED : O2 supply pressure too low. [Error Code = 2]
2018-01-19 10:09:42 [CHK_STA] Checking that O2 supply pressure (MAX) - O2 supply pressure (MIN) < Max tolerated difference.
2018-01-19 10:09:42 [MEA] O2 supply pressure (MAX) = 4 - Tolerance Interval ( -,O2 supply pressure (MIN) + 100 ) = [ -; 102 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] PASSED
2018-01-19 10:09:42 [CHK_STA] Checking that |O2 supply pressure (MON) - O2 supply pressure (BRE)| < Tolerance.
2018-01-19 10:09:42 [MEA] O2 supply pressure (BRE) = 5 - Tolerance Interval (O2 supply pressure (MON) +/- 20) = [ -18; 24 ] ( Unit : mbar )
2018-01-19 10:09:42 [CHK_FIN] PASSED
2018-01-19 10:09:56 [ADD_INFO] Gas type: air.
2018-01-19 10:09:56 [FIN] Gas supply test FAILED - [Error Code = 2]
2018-01-19 10:09:56 [STA] Turbine accuracy test
2018-01-19 10:09:56 [CHK_STA] Checking turbine age
2018-01-19 10:09:56 [MEA] Turbine operating hours = 932 - Limit = [ -; 40000 ] ( Unit : hours )
2018-01-19 10:09:56 [CHK_FIN] PASSED
2018-01-19 10:09:58 [CHK_STA] Checking turbine rpm generation
2018-01-19 10:09:58 [MEA] Measured turbine rpm = 4996 - Tolerance Interval = [ 4750.00; 5250.00 ] ( Unit : rpm )
2018-01-19 10:09:58 [CHK_FIN] PASSED
2018-01-19 10:09:58 [CHK_STA] Checking turbine rise fall time
2018-01-19 10:09:58 [MEA] Measured rise time = 0.110 - Limits = [ -; 0.150 ] ( Unit : seconds )
2018-01-19 10:09:58 [CHK_FIN] PASSED
2018-01-19 10:09:59 [CHK_STA] Checking turbine rpm generation
2018-01-19 10:09:59 [MEA] Measured turbine rpm = 40043 - Tolerance Interval = [ 38000.00; 42000.00 ] ( Unit : rpm )
2018-01-19 10:09:59 [CHK_FIN] PASSED
2018-01-19 10:09:59 [CHK_STA] Checking turbine rise fall time
2018-01-19 10:09:59 [MEA] Measured fall time = 0.110 - Limits = [ -; 0.150 ] ( Unit : seconds )
2018-01-19 10:09:59 [CHK_FIN] PASSED
2018-01-19 10:09:59 [FIN] Turbine accuracy test PASSED
2018-01-19 10:10:00 [SFIN] Turbine and Gas supply test sequence FAILED S20137

SERVO-air Ventilator System, Service Manual 6-9


| 6 | Troubleshooting |

6.5.2 Pre-use check CFI

Internal test sequence

CFI Log description Recommended action


Panel audio test (PAN)
- If the test fails, the audio does not work. 1. Replace Loudspeaker.
2. Replace PC 2023 Panel.

Monitor audio test (MON)


1 The buzzer is permanently on. 1. Replace PC 1992 Monitoring.

2 The buzzer could not be turned on. 1. Replace PC 1992 Monitoring.

3 The buzzer could not be turned off. 1. Replace PC 1992 Monitoring.

Temperature and Moist sensor test


1 Humidity is not within accepted limits. 1. Replace PC 2030 Inspiratory channel.

Comment: 0.0–100.0 %.

2 Turbine temperature is not within accepted limits. 1. Check Air inlet filter.
2. Replace Turbine module.
Comment: 0.0–70.0 °C.

3 Ambient temperature is not within accepted 1. Replace PC 2030 Inspiratory channel.


limits.

Comment: 0.0–55.0 °C.

4 Insp gas temperature is not within accepted 1. Replace Inspiratory flowmeter.


limits.

Comment: 0.0–60.0 °C

Blower inlet test


1 Blower inlet resistance is too low. 1. Check that the Air inlet filter is connected.
2. Replace Turbine module.
Comment: <1.00 cmH2O/l/s.

2 Blower inlet resistance is too high. 1. Check Air inlet filter.


2. Replace Dust filter or complete Air inlet filter.
Comment: >3.30 cmH2O/l/s.
3. Replace Turbine module.

3 Pressure diff sensor test: pressure transducer 1. Replace Turbine module.


offset not within accepted limits.

Comment: 1586–1690.

Alarm output connector test (MON)


1 Remote Alarm is already active. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

2 Remote Alarm has not been activated. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

6 - 10 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


3 Remote Alarm has not been deactivated. 1. Replace PC 2021 Rear connectors.
2. Replace PC 1992 Monitoring.

Power failure test (MON)


1 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #1 for 5 V.

2 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #2 for 5 V.

3 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #1 for 3.3 V.

4 PFA alarm is inactive. 1. Replace PC 1992 Monitoring.

Comment: Check #2 for 3.3 V.

5 The buzzer has not been turned on. 1. Replace PC 1992 Monitoring.

6 The buzzer has not been turned off. 1. Replace PC 1992 Monitoring.

7 The remote alarm is active. 1. Replace PC 1992 Monitoring.

Comment: Failed to inactivate the remote alarm.

8 The remote alarm is active. 1. Replace PC 1992 Monitoring.

Comment: Failed to activate the remote alarm.

9 PFA alarm is active. 1. Replace PC 1992 Monitoring.

Barometer test sequence

CFI Log description Recommended action


Barometer test (MON)
1 Barometric Pressure (MON) is not within the 1. Calibrate the barometer (S&S).
accepted limits. 2. Replace PC 1992 Monitoring.
Comment: >1070 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.

2 Barometric Pressure (MON) is not within the 1. Calibrate the barometer (S&S).
accepted limits. 2. Replace PC 1992 Monitoring.
Comment: <650 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.

3 The difference between the values measured in 1. Replace PC 1991 Control.


BRE and MON is too high. 2. Replace PC 1992 Monitoring.
Comment: >8 hPa (mbar). 3. Replace PC 2004/PC 2024 Expiratory channel.
4. Replace O2 gas module.

SERVO-air Ventilator System, Service Manual 6 - 11


| 6 | Troubleshooting |

Internal leakage test sequence

CFI Log description Recommended action


Internal leakage test (BRE)
1 The difference between InspPress and ExpPress 1. Check that the pressure transducer tubes and the inspiratory
is too high. filter are correctly mounted.
2. Check that both PC 1781 Pressure transducer (Insp. and
Comment: >10 cmH2O Exp.) are correctly mounted.
3. If the ‘Pressure transducer test’ also fails, refer to the
recommended actions if ‘Pressure transducer test’ failed.

2 The System Volume is too low. 1. Replace Inspiratory flowmeter.

Comment: <0.2000 l

3 The System Volume is too high. 1. Check that the correct test tube is used during the Pre-use
check and check that the test tube is correctly connected.
Comment: >0.4000 l 2. If the ‘Flow transducer test’ also fails, replace Inspiratory
flowmeter.
3. Check that the expiratory cassette is correctly seated in the
cassette compartment.
4. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
5. Check that the pressure transducer tubes/filters are correctly
mounted.
6. Check that the inspiratory gas docking, inspiratory flowmeter
and inspiratory channel are correctly mounted.
7. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
8. Check that the safety valve closes properly when the Pre-use
check starts (distinct clicking sound from the valve). If the
safety valve not closes properly during this check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

6 - 12 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


4 Inspiratory Pressure is too low. 1. The Leakage detection, described in chapter Service
procedures, can be used to locate the leakage. The Leakage
Comment: Excessive leakage. Required detection is a complement to the recommended actions
pressure 50 cmH2O not obtained, leakage test below.
cannot be performed. 1. Check that the connected gas supply pressure is within
the specified range.
2. Check that the test tube is correctly connected.
3. Check that the expiratory cassette is correctly seated in
the cassette compartment.
4. If possible, replace the expiratory cassette and check if
the new cassette is accepted by the Pre-use check. If
the new cassette was accepted by the Pre-use check,
the fault was located to the cassette.
5. Check that the pressure transducer tubes/filters are
correctly mounted.
6. Check that the inspiratory pipe is correctly mounted in
the inspiratory section.
7. Check that the safety valve membrane is clean and
correctly seated in the inspiratory pipe.
8. Check that the safety valve closes properly when the
Pre-use check starts (distinct clicking sound from the
valve). If the safety valve not closes properly during this
check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

SERVO-air Ventilator System, Service Manual 6 - 13


| 6 | Troubleshooting |

CFI Log description Recommended action


5 The Leakage Value is too high, or the safety 1. Check that the correct test tube is used during the Pre-use
valve is either not locked or its status could not check and check that the test tube is correctly connected.
be read. 2. Check that the expiratory cassette is correctly seated in the
cassette compartment.
Comment: >0.0200 l/min
3. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
4. Check that the pressure transducer tubes/filters are correctly
mounted.
5. Check that the inspiratory gas docking, inspiratory flowmeter
and inspiratory channel are correctly mounted.
6. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
7. Check that the safety valve closes properly when the Pre-use
check starts (distinct clicking sound from the valve). If the
safety valve opens during this test, the opening pressure
may not be correctly calibrated (see 'Safety valve test' below).
Run the 'Safety valve test' and repeat the complete Pre-use
check.
8. The Leakage detection, described in chapter Service
procedures, can be used to locate the leakage. The Leakage
detection is a complement to the recommended actions
below.
1. Check that the test tube is correctly connected.
2. Check that the expiratory cassette is correctly seated in
the cassette compartment.
3. If possible, replace the expiratory cassette and check if
the new cassette is accepted by the Pre-use check. If
the new cassette was accepted by the Pre-use check,
the fault was located to the cassette.
4. Check that the pressure transducer tubes/filters are
correctly mounted.
5. Check that the inspiratory pipe is correctly mounted in
the inspiratory section.
6. Check that the safety valve membrane is clean and
correctly seated in the inspiratory pipe.
7. Check that the safety valve closes properly when the
Pre-use check starts (distinct clicking sound from the
valve). If the safety valve not closes properly during this
check:
- Replace safety valve pull magnet.
- Replace PC 2004/PC 2024 Expiratory channel.

Turbine and Gas supply test sequence

CFI Log description Recommended action


Gas supply test (MON)
1 O2 supply pressure too high. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: >600 kPa (6000 mbar). 2. Replace O2 gas module.

6 - 14 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


2 O2 supply pressure too low. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: <200 kPa (2000 mbar). 2. Replace O2 gas module.

3 The difference between O2 supply pressure 1. Check that the connected O2 supply pressure is stable.
(MAX) and O2 supply pressure (MIN) is too 2. Replace O2 gas module.
large.

Comment: O2 supply pressure variation too


large, >100 hPa (mbar).

4 The difference between O2 supply pressure 1. Replace PC 1991 Control.


(MON) and O2 supply pressure (BRE) is too 2. Replace PC 1992 Monitoring.
large.

Comment: >20 hPa (mbar).

11 Breathing disconnected. 1. Restart the system.


2. Replace PC 1991 Control.

Turbine accuracy test (MON)


1 Turbine age too old. 1. Replace Turbine module.

Comment: Max. 40000 h.

2 Turbine rpm accuracy error. 1. Replace Turbine module.

Comment: 4750–5250 r/min.

3 Turbine acceleration accuracy error. 1. Replace Turbine module.

Comment: Acceleration (rise time) from approx.


5000 to 40000 r/min - max. 0.150 s.

4 Turbine deceleration accuracy error. 1. Replace Turbine module.

Comment: Deceleration (fall time) from approx.


40000 to 5000 r/min - max. 0.150 s.

Pressure transducer test sequence

CFI Log description Recommended action


Breathing pressure transducer test (BRE)
1 Inspiration Offset could not be stored persistently 1. Replace PC 1781 Inspiratory pressure transducer.
- Value out of range. 2. Replace PC 1991 Control.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

2 Expiration Offset could not be stored persistently 1. Replace PC 1781 Expiratory pressure transducer.
- Value out of range. 2. Replace PC 1991 Control.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.
SERVO-air Ventilator System, Service Manual 6 - 15
| 6 | Troubleshooting |

CFI Log description Recommended action


Monitor pressure transducer test (MON)
1 Inspiration offset >6 cmH2O from factory default. 1. Replace PC 1781 Inspiratory pressure transducer.
2. Replace PC 1992 Monitoring.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

2 Expiration offset >6 cmH2O from factory default. 1. Replace PC 1781 Expiratory pressure transducer.
2. Replace PC 1992 Monitoring.
Comment: Differs more than ±6 cmH2O
(±0.300 V) from factory default.

3 Expiration gain >8% from factory default. 1. Cancel the ‘Pressure transducer test’ and continue with the
remaining tests (this will calibrate the safety valve). Restart
the Pre-use check.
2. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Check/replace PC 1781 Pressure transducer (Insp. and Exp.).
To locate the faulty pressure transducer, replace one
transducer at a time.

4 Expiration offset is different at zero pressure and 1. Replace PC 1781 Expiratory pressure transducer.
high pressure.
5 The difference between the values measured in 1. Replace PC 1991 Control.
BRE and MON is too high. 2. Replace PC 1992 Monitoring.
Comment: Differs more than ±0.80 cmH2O.

6 Inspiration pressure transducer could not be 1. Replace PC 1781 Inspiratory pressure transducer.
locked or is faulty.
7 Expiration pressure transducer could not be 1. Replace PC 1781 Expiratory pressure transducer.
locked or is faulty.
8 Inspiratory pressure value out of bounds. 1. Replace Turbine module.

Comment: 45.80–54.20 cmH2O.

Expiratory valve test (BRE)


1 Expiratory Pressure too high. 1. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
Comment: Measures at 40, 20, 10 and 5 cmH2O. cassette was accepted by the Pre-use check, the fault was
Measured values must be within ±15 cmH2O. located to the cassette.
2. Replace Expiratory valve coil.

2 Expiratory Pressure too low. 1. Clean or replace Turbine module one-way valve.
2. If possible, replace the expiratory cassette and check if the
Comment: Measures at 40, 20, 10 and 5 cmH2O. new cassette is accepted by the Pre-use check. If the new
Measured values must be within ±15 cmH2O. cassette was accepted by the Pre-use check, the fault was
located to the cassette.
3. Replace Expiratory valve coil.

3 Voice Coil too strong. 1. If possible, replace the expiratory cassette and check if the
new cassette is accepted by the Pre-use check. If the new
Comment: Expiratory valve coil not within cassette was accepted by the Pre-use check, the fault was
specified limits; offset 0.06768–0.15684 A, gain located to the cassette membrane.
0.00730–0.01130 A/cmH2O. 2. Replace Expiratory valve coil.

6 - 16 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

CFI Log description Recommended action


4 The expiratory valve constants were not stored 1. Replace PC 2004/PC 2024 Expiratory channel.
persistently.

SERVO-air Ventilator System, Service Manual 6 - 17


| 6 | Troubleshooting |

Safety valve test sequence

CFI Log description Recommended action


Monitor safety valve test (MON)
1 The Safety Valve could not be opened. 1. Replace PC 2004/PC 2024 Expiratory channel.
2. Replace PC 1992 Monitoring.

2 The Safety Valve could not be closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

3 The Safety Valve is open. 1. Replace Safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

4 Disable valves activation failed. 1. Replace PC 1992 Monitoring.

5 Disable valves deactivation failed. 1. Replace PC 1992 Monitoring.

Breathing safety valve test (BRE)


1 Pressure superior to allowed max and valve not 1. Replace Safety valve pull magnet.
open. 2. Replace PC 2004/PC 2024 Expiratory channel.
Comment: >120 cmH2O.

2 Pressure not rising. Timeout. 1. Check that the safety valve membrane is clean and correctly
seated in the inspiratory pipe.
Comment: Not possible to reach opening 2. Check that the inspiratory gas docking, inspiratory flowmeter,
pressure 117 cmH2O. inspiratory channel and turbine module one-way valve are
correctly mounted.

3 Calibration Impossible. 1. Replace Safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory channel.

4 The Safety Valve is not closed. 1. Replace Safety valve pull magnet.
2. Replace PC 2004/PC 2024 Expiratory channel.
3. Replace PC 1992 Monitoring.

5 The Safety Valve could not be disabled. 1. Replace PC 1991 Control.

6 The Safety Valve could not be enabled. 1. Replace PC 1991 Control.

O2 cell test sequence

CFI Log description Recommended action


O2 cell test (MON)
1 No O2 Transducer detected. 1. Check that the O2 cell is connected.
2. Replace O2 cell.
3. Replace O2 cell cable.
4. Replace PC 1992 Monitoring.

3 O2 Concentration in AIR Supply too low. 1. Replace O2 cell.

Comment: >25 %.
Error in the log text. Correct text is:
O2 Concentration in AIR Supply too high.
6 - 18 SERVO-air Ventilator System, Service Manual
| Troubleshooting | 6 |

CFI Log description Recommended action


4 O2 Concentration in O2 Supply too low. 1. Check that the connected O2 supply concentration is within
the specified range.
Comment: <70.0 % 2. Replace O2 cell.

6 O2 supply low. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: <200 kPa 2. Replace O2 gas module.

7 Gas supply not approved. 1. Gas supply test not passed. Redo the Gas supply test.

8 Invalid EEPROM data. 1. Replace O2 cell.

9 Lock failed. 1. Replace O2 cell.


2
Comment: Lock failed, I C problem

11 O2 Cell Calibration impossible. 1. Replace O2 cell.

12 The O2 Cell needs to be replaced. 1. Replace O2 cell.

Flow transducer test sequence

CFI Log description Recommended action


Insp flow sensor test (MON) (Sensirion flow sensor test)
1 Insp flow sensor test, measured O2 flow failed. 1. Replace O2 gas module.
2. Replace Inspiratory flowmeter.

2 Insp flow sensor test, Bre - Mon comparison 1. Replace PC 1991 Control.
failed. 2. Replace PC 1992 Monitoring.

3 Sensirion flow transducer offset not within 1. Replace Inspiratory flowmeter.


accepted limits. 2. Replace PC 1992 Monitoring.

4 Insp flow sensor test, sensirion measurement 1. Replace Inspiratory flowmeter.


failure. 2. Replace O2 gas module.

5 Insp flow sensor test, Bre - Mon Sensirion 1. Replace PC 1991 Control.
comparison failure. 2. Replace PC 1992 Monitoring.

6 No gas supply. 1. Check O2 supply.


2. Replace O2 gas module.

SERVO-air Ventilator System, Service Manual 6 - 19


| 6 | Troubleshooting |

CFI Log description Recommended action


Breathing flow transducer test (BRE)
1 Inspiration Air Offset could not be stored 1. Replace Inspiratory flowmeter.
persistently - Value out of range.

Comment: 22.021-24.609 V

2 Inspiration O2 Offset could not be stored 1. Replace O2 gas module.


persistently - Value out of range.

Comment: 0.000-0.081 V

3 Expiration Offset could not be stored persistently 1. Check that the expiratory cassette is correctly seated in the
- Value out of range. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.093-0.142 V
3. Replace PC 2004/PC 2024 Expiratory channel.

Flow test (MON)


2 The Inspiration O2 Offset is not within the 1. Replace O2 gas module.
accepted limits.

Comment: 0.000-0.033 V

3 The Expiration Offset is not within the accepted 1. Check that the expiratory cassette is correctly seated in the
limits. cassette compartment.
2. Replace the expiratory cassette.
Comment: 0.024-0.072 V
3. Replace PC 2004/PC 2024 Expiratory channel.

4 Expiration Gain is not within the specs. 1. Check that the connected O2 supply pressure is within the
specified range.
Comment: 0.90-1.15 (k value) 2. Check that the cassette is correctly seated in the cassette
compartment. If possible, replace the expiratory cassette and
check if the new cassette is accepted by the Pre-use check.
If the new cassette was accepted by the Pre-use check, the
fault was located to the cassette. The fail with the old cassette
may in this case be due to water collected inside the cassette.
3. Replace Inspiratory flowmeter.
4. Replace O2 gas module.
5. Replace PC 1785/PC 2031 Expiratory channel connector.
6. Replace PC 1991 Control.
7. Replace PC 1992 Monitoring.
8. Replace PC 2004/PC 2024 Expiratory channel.

6 Timeout getting Expiration Gain (k value) 1. Check that the expiratory cassette is correctly seated in the
cassette compartment.
2. Replace the expiratory cassette.

6 - 20 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Battery switch test sequence

CFI Log description Recommended action


Battery switch test (MON)
1 Mains is not on. 1. Connect mains power.

2 The batteries have not been switched on 1. Mains power not disconnected as demanded.
(TIMEOUT).
3 The batteries have been switched back off 1. Mains power not reconnected as demanded.
without mains being back on first. (TIMEOUT).
4 No battery operation detected by BRE. 1. Replace PC 1991 Control.

5 Remaining battery power is insufficient for the 1. Charge or replace Battery modules.
test to be performed.

Comment: <10 min.

6 No batteries attached. 1. Connect Battery module in slot 2 (main slot).


2. Replace PC 2028 DC/DC and Battery control.

7 The batteries have been switched on without 1. Replace PC 2028 DC/DC and Battery control.
mains loss first. (TIMEOUT). 2. Replace PC 1992 Monitoring.

8 The batteries have been switched back off 1. Replace PC 2028 DC/DC and Battery control.
without mains being back on first. (TIMEOUT). 2. Replace PC 1992 Monitoring.

9 Unknown battery detected. 1. Connect a Maquet battery module.

10 Power system external power source test failed. 1. Connect mains power.
(POW)

11 Power system battery switch test failed. 1. Connect Battery module in slot 2 (main slot).
(POW) 2. Replace PC 2028 DC/DC and Battery control.

12 Power system no external power source test 1. Replace PC 2028 DC/DC and Battery control.
failed.
(POW)

13 Bad battery detected. 1. Replace Battery module.


(POW)
14 No battery in main slot. 1. Connect Battery module in slot 2 (main slot).
(POW)
15 Battery has expired. 1. Replace Battery module.
(POW)

Comment: >300 charge cycles/>48 months from


manufacturing date.

16 Low capacity (FCC) battery detected. 1. Replace Battery module.


(POW)

SERVO-air Ventilator System, Service Manual 6 - 21


| 6 | Troubleshooting |

Patient circuit test sequence

CFI Log description Recommended action


Patient circuit test (BRE)
1 InspPress is too low. 1. If the internal leakage test has passed, the leakage is to be
located to the patient circuit. Check for leakage or replace
Comment: <40 cmH2O the patient circuit.

2 The Compressible Volume is too low. 1. Check or replace the patient circuit.

Comment: <0.02 ml/cmH2O.

3 The Compressible Volume is too high. 1. Check or replace the patient circuit.

Comment: >6.00 ml/cmH2O.

4 Leakage too high, or the safety valve is either 1. If the internal leakage test has passed, the leakage is to be
not locked or its status could not be read. located to the patient circuit. Check for leakage or replace
the patient circuit.
Comment: >0.0800 l/min

5 System Volume is too low. 1. Check that the internal leakage test passed.

Comment: <0.1500 l

6 System Volume is too high. 1. Check that the internal leakage test passed.
2. Improper patient circuit connected.
Comment: >6.0000 l

Alarm state test sequence

CFI Log description Recommended action


External alarm system test (MON) If this option is installed.
1 Remote External Alarm is not connected. 1. Check the external alarm system.
2. Replace PC 2021 Rear connectors.

2 Remote External Alarm is already active. 1. Check the external alarm system.
2. Replace PC 2021 Rear connectors.

Alarm state test (MON)


- cf alarms error codes Check Technical error alarms logs for further information.

6 - 22 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.6 Technical error codes and messages


The table below shows recommended actions in case
of Technical errors.

6.6.1 Monitoring

Error
Error message / Possible cause Recommended action
code
1 Power error. 1. Replace PC 2028 DC/DC & Battery control.

-12 V too low, i.e. < -13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

2 Power error. 1. Check status of external 12 V supply (if


connected).
-12 V too high, i.e. > -10.8 V. 2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

3 Power error. 1. Check status of external 12 V supply (if


connected).
+12 V too low, i.e. < +10.8 V. 2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

4 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V too high, i.e. > +13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

5 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace the O2 gas module.
+24 V too low, i.e. < 21.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

6 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+24 V too high, i.e. > 26.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

SERVO-air Ventilator System, Service Manual 6 - 23


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
7 Software error. 1. Replace PC 1991 Control.
2. Replace PC 1992 Monitoring.
Patient category mismatch between Breathing and Monitoring
subsystems. 3. Replace the O2 gas module.

8 Insp. pause hold time exceeded. 1. Replace PC 1991 Control.

The max. Insp. pause hold time 30 s exceeded. The system


has not returned to ventilation after Insp. pause hold time.

9 Exp. pause hold time exceeded. 1. Replace PC 1991 Control.

The max. Exp. pause hold time 30 s exceeded. The system


has not returned to ventilation after Exp. pause hold time.

10 Ventilation stopped. 1. Replace PC 2004/PC 2024 Expiratory


channel.
The VALVES_DISABLED signal has stopped power supply to 2. Replace PC 1991 Control.
O2 gas module, Turbine module and Safety valve.
3. Replace PC 2028 DC/DC & Battery control.
4. Replace PC 1992 Monitoring.

11 Safety valve open. 1. Replace safety valve pull magnet.


2. Replace PC 2004/PC 2024 Expiratory
The Safety valve is detected as open without fulfilled opening channel.
conditions.

16 Ventilation error. 1. Replace PC 2004/PC 2024 Expiratory


channel.
PC 2004/PC 2024 Expiratory channel failure.

17 Ventilation stopped. 1. Report to MCC HSC or to your local


Getinge representative for further
Software conflict. Two PC 2004/PC 2024 Expiratory channel information.
detected on the CAN bus.

22 Backup audible alarm error. 1. Replace PC 1992 Monitoring.

Backup alarm buzzer error.

24 Backup audible alarm or Remote alarm error. 1. Replace PC 1992 Monitoring.

Error indicated in the backup alarm or remote alarm capacitor


on PC 1992 Monitoring.

6 - 24 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
25 Communication error. Depending on ID # and Error # stated in the
Technical alarm log.
Communication error between ID PROMs/EEPOTs and
Monitoring subsystem. If repeated, replace the spare part indicated by
the ID # according to the list below:
The ID # is followed by an Error # in the Technical alarm log.
ID #
For Error 6, report to MCC HSC or to your local Getinge
representative for further information. 0: PC 2020 Main back-plane.

For all other Error #, replace the part identified by the ID #. 1: PC 1992 Monitoring.

Error # 2: PC 2028 DC/DC & Battery control.

1. CHK_SUM_ERR 3: PC 2021 Rear connectors.


2. NR_OF_BYTES_IN_CHKSUM_ERR
4: PC 1781 Inspiratory pressure transducer.
3. CHKSUM_FILL_BYTE_ERR
4. WRITABLE_RECORDS_FAILURE 5: PC 1781 Expiratory pressure transducer.
5. WRITABLE_RECORDS_DIFFER
6: O2 cell or O2 cell cable.
6. COMM_ERR
7. VALIDATION_ERR 32: PC 2004/PC 2024 Expiratory channel.
8. WCR_VALIDATION_ERR
9. WRITE_NOT_PERMITTED 33: Turbine module.
10. OTHER_ERR 34: O2 gas module.

35: PC 1992 Monitoring.

256: PC 1991 Control.

512: PC 2023 Panel.

1024: PC 2004/PC 2024 Expiratory channel.

1025: Expiratory cassette.

1026: PC 1993 Ethernet switch.

Note: After replacement of PC 1992, TE 25


may occur. Adjust the barometric pressure in
Service & Settings to solve this error (Service
key required).

27 Backup audible alarm error. 1. Replace PC 1992 Monitoring.

Backup alarm buzzer error.

28 Audible alarm error. 1. Replace loudspeaker.


2. Replace PC 2023 Panel.
Loudspeaker error.

29 Memory backup battery depleted. 1. Replace Memory backup battery on


PC 1992 Monitoring.
The Memory backup battery on PC 1992 Monitoring is depleted.

SERVO-air Ventilator System, Service Manual 6 - 25


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
32 Software version conflict. 1. Reinstall the System software.
2. Report to MCC HSC or to your local
Conflict between the software versions installed. Getinge representative for further
information.

35 Ventilation disabled. 1. Replace PC 2004/PC 2024 Expiratory


channel.
Communication error or PC 2004/PC 2024 Expiratory channel
failure during startup.

37 Software error. 1. Replace PC 2004/PC 2024 Expiratory


channel.
PC 2004/PC 2024 Expiratory channel indicates wrong patient 2. Report to MCC HSC or to your local
category. Getinge representative for further
information.

38 Barometric pressure too high. 1. Calibrate barometer (S&S required).


2. Replace PC 1992 Monitoring.
Barometric pressure above allowed limit or barometer error.

39 Barometric pressure too low. 1. Calibrate barometer (S&S required).


2. Replace PC 1992 Monitoring.
Barometric pressure below allowed limit or barometer error.
Note: This alarm will be activated if the
ambient pressure is below 650 hPa, e.g. on
a high altitude.

40 Measured value invalid. 1. Restart the system.

Invalid measured metric detected. The invalid metric is not


displayed.

41 Real time clock error. 1. Replace PC 1992 Monitoring.

Internal processor real time clock error.

42 Checksum error. 1. Restart the system.


2. Replace Memory backup battery on
Software checksum error during startup. PC 1992 if together with Error code 29.
3. Replace PC 1992 Monitoring.

43 Battery communication error. 1. Update to System SW version 2.1.2 or


higher.
Battery module or PC 2028/PC 1992 communication error. 2. Replace the Battery module(s).
3. Replace PC 1992 Monitoring.
4. Replace PC 2028 DC/DC & Battery control.

46 Alarm output connector error. 1. Replace PC 2021 Rear connectors.


2. Replace PC 1992 Monitoring.
Alarm output connector error or communication error with
PC 1992.

6 - 26 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
48 Pre-oxygen time exceeded. 1. Replace PC 1991 Control.

The max. Pre-oxygenation time during


DISCONNECTION/SUCTION exceeded. The system has not
returned to ventilation using the previous settings after the
Preoxygenation time.

50 Communication error. 1. Replace PC 1992 Monitoring.


2. Replace PC 2020 Main back-plane.
Communication error between ID PROM on PC 2020 Main
back-plane and Monitoring subsystem. System ID,
configuration, operating time, etc., not available.

51 On/Off switch error. 1. Replace On/Off switch including cable.

The On/Off switch is neither in On or Off position during more


than 3 seconds.

55 Communication error. Check for additional information in the Technical


log.
This error code is an indication of one or more of:
With log text
• Internal Ethernet communication failure
bullet "CommunicationBrokenCondition:NodeStatusBreathing":
1. Replace PC 1991 Control.
• Battery status erroneously repeated
bullet
• Subsystem mismatch of patient category, ventilation mode, With log text
bullet
and/or ventilation setting(s) "CommunicationBrokenCondition:NodeStatusPanel":
1. Replace the Control cable.
• Certain problems with the Edi module.
bullet
2. Replace B740 CPU or the complete User
Interface.

For all other errors:

1. Replace Control cable.


2. Update to latest released System SW
version.
3. Replace battery module.
4. Replace Edi module (if present).
5. Replace PC 1992 Monitoring.
6. Replace PC 1991 Control.
7. Replace PC 1993 Ethernet Switch.
8. Report to MCC HSC or to your local
Getinge representative for further
information.

56 Missing SW option data. 1. Report to MCC HSC or to your local


Getinge representative for further
information.

57 Missing SW option data. 1. Reinstall software options.


2. Replace PC 1991 Control.

58 Missing SW option data. 1. Reinstall software options.


2. Replace PC 2023 Panel.

59 SW option installation failure. 1. Reinstall software options.


2. Replace PC 1992 Monitoring

SERVO-air Ventilator System, Service Manual 6 - 27


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
60 Charger error. 1. Replace Battery module.
2. Replace PC 2028 DC/DC & battery control.
This error code is an indication of a charger error in the power
supply subsystem.

The cause is a hardware error for example:

• An SMB-bus MUX error which may cause estimation of


bullet
remaining battery capacity to be inadequate.
• A charger electronics error causing unreasonably high
bullet
current consumption.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

61 Main fan error. 1. Replace Main fan.


2. Replace PC 2028 DC/DC & battery control.
Power supply to Main fan is monitored and this error code
indicates for example:

• Main fan error.


bullet
• Error in the power supply subsystem.
bullet
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

62 Very high temperature in Turbine module. Ventilation stopped. 1. Replace Turbine module.

63 Turbine module communication error. Ventilation stopped. 1. Replace Turbine module.


2. Replace PC 1992 Monitoring.

64 Turbine module stand still detected. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

65 Turbine module failure. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

66 Turbine module under-voltage. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

67 Turbine module over-voltage. 1. Replace Turbine module.

Turbine speed control degraded. Ventilation continues but with


decreased gas delivery accuracy.

68 Turbine module input voltage failure. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

69 Turbine module failure. 1. Replace Turbine module.

Turbine module has stopped, O2 is delivered instead.

6 - 28 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
70 Ambient air temperature sensor range error. 1. Replace PC 2030 Inspiratory channel.

71 Air humidity (RH) sensor range error. 1. Replace PC 2030 Inspiratory channel.

72 Inspiratory flowmeter temperature sensor range error. 1. Replace Inspiratory flowmeter.


2. Replace Turbine.

73 Ambient air temperature and humidity sensor error. 1. Replace PC 2030 Inspiratory channel.

74 Inspiratory flowmeter error. 1. Replace Inspiratory flowmeter.


2. Replace Turbine.

75 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V to Turbine module too high, i.e. > +18 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

76 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Turbine module.
+12 V to Turbine module too low, i.e. < +10 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

77 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+12 V to Inspiratory flowmeter too high, i.e. > +13.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

78 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Inspiratory flowmeter.
+12 V to Inspiratory flowmeter too low, i.e. < +10.8 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

79 Power error. 1. Replace PC 2028 DC/DC & Battery control.

+5 V to Inspiratory flowmeter too high, i.e. > +5.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

SERVO-air Ventilator System, Service Manual 6 - 29


| 6 | Troubleshooting |

Error
Error message / Possible cause Recommended action
code
80 Power error. 1. Replace PC 2028 DC/DC & Battery control.
2. Replace Inspiratory flowmeter.
+5 V to Inspiratory flowmeter too low, i.e. < +4.5 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

81 Power error. 1. Replace PC 2028 DC/DC & Battery control.


2. Replace Inspiratory flowmeter.
-12 V to Inspiratory flowmeter too high, i.e. > -10.8 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

82 Power error. 1. Replace PC 2028 DC/DC & Battery control.

-12 V to Inspiratory flowmeter too low, i.e. < -13.2 V.

A Power system restart might be necessary to reset power


related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

84 O2 evacuation fan error. 1. Replace the O2 evacuation fan.


2. Replace PC 2028 DC/DC & Battery control.
A Power system restart might be necessary to reset power
related alarms. To restart, switch off the system and disconnect
both the mains power supply and the External DC power supply.

6.6.2 Breathing

Error
Error message / Possible cause Recommended action
code
10001 Memory backup battery depleted. 1. Replace battery on PC 1991 Control.

The battery on PC 1991 Control is depleted.

10002 Communication error. 1. Replace PC 1991 Control.

Communication error with ID PROM on PC 1991 Control.

10003 Ventilation stopped. 1. Restart the system.


2. Replace PC 1991 Control.
Error in the internal memory detected.

This technical alarm may occur after a software installation.


Restart the system and check that no technical alarms are
activated.

If TE10003 remains, this indicates a HW error.

10004 Missing SW option data. 1. Reinstall software options.


2. Replace PC 1991 Control.

6 - 30 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

6.6.3 Panel

Error
Error message / Possible cause Recommended action
code
20002 Backlight error. If the display is lit (and the error code is shown
on the display):
The backlight power consumption is measured. If the power 1. Replace PC 2023 Panel.
consumption is outside specified range, this technical error is If the display is not lit (the display is dark):
activated. 1. Replace Display incl. touchscreen.
This error message may also be logged immediately before 2. Replace PC 2023 Panel.
the No battery capacity alarm when the system is running in
Note: With a failure on the display, or likely
battery mode.
also on PC 2023 Panel, the User interface
display will become dark. This error code will
in such case not be possible to see (but will
be logged).

20003 Activate all pending alarm. Report to MCC HSC or to your local Getinge
representative for further information.
There is a software-related internal communications problem.

20004 Alarm sound level too low. 1. Restart the system. Do not touch the User
interface during system startup. Interfering
The loudspeakers are checked using a microphone during with the Touch screen, loudspeaker grid,
startup. etc., may affect the internal technical tests.
2. Replace loudspeaker.
3. Replace PC 2023 Panel.

20005 Communication error. 1. Update to latest released System SW


version.
Ethernet communication failure between PAN and MON. 2. Report to MCC HSC or to your local
Getinge representative for further
May occur with TE 55. information.

20006 Missing SW option data. 1. Restart the system.


2. Update to latest released System SW
If together with TE 56 or TE 10004, see recommended actions version.
for these errors.

SERVO-air Ventilator System, Service Manual 6 - 31


| 6 | Troubleshooting |

6.6.4 Expiratory flowmeter

Error
Error message / Possible cause Recommended action
code
For systems with PC 2004 Expiratory channel

40001 Expiratory flowmeter error. Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM read/write error:
See information logged together with error code 40001 in the 1. Replace PC 2004 Expiratory channel.
Service log / Technical alarms.
Technical alarm Expiration flow meter PC 2004
and Exp. cassette EEPROM checksum error:
1. Replace Expiratory cassette.

Technical alarm Expiration flow meter PC 2004


and PC 2004 thermistor failure:
1. Check Main fan operation for proper
cooling.
2. Replace PC 2004 Expiratory channel.

Technical alarm Expiration flow meter PC 2004


and PC 2004 60 V overrange:
1. Replace PC 2004 Expiratory channel.
2. Replace PC 2028 DC/DC & Battery control.

Technical alarm Expiration flow meter PC 2004


and PC 2004 60 V underrange:
1. Replace PC 1991 Control.
2. Replace PC 2028 DC/DC & Battery control.
3. Replace PC 2004 Expiratory channel.

For systems with PC 2024 Expiratory channel

40001 Internal communication error on PC 2024. 1. Replace PC 2024 Expiratory channel.

40002 Temperature sensor error on PC 2024. 1. Replace PC 2024 Expiratory channel.

Note that this error may occur if the system is running outside
operating temperature range.

40003 Power error. 1. Replace PC 2024 Expiratory channel.

+60 V supply voltage to the ultrasonic transducers in the


Expiratory cassette too high, i.e. > +75 V.

40004 Power error. 1. Replace PC 2024 Expiratory channel.


2. Replace PC 2028 DC/DC & Battery control.
+60 V supply voltage to the ultrasonic transducers in the
Expiratory cassette too low, i.e. < +45 V.

40005 Power error. 1. Replace PC 2024 Expiratory channel.

+1.8 V is more than 15 % outside nominal value.

40006 Power error. 1. Replace PC 2024 Expiratory channel.

+1.2 V is more than 15 % outside nominal value.

6 - 32 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error
Error message / Possible cause Recommended action
code
40007 Power error. 1. Replace PC 2024 Expiratory channel.
2. Replace PC 2028 DC/DC & Battery control.
+5 V is more than 15 % outside nominal value.

40008 Power error. 1. Replace PC 2024 Expiratory channel.

-5 V is more than 15 % outside nominal value.

40009 Power error. 1. Replace PC 2024 Expiratory channel.

+2.5 V is more than 15 % outside nominal value.

40010 Internal temperature too high, > 77 ºC. 1. Ensure proper cooling from the Main fan:
- Check fan operation
The temperature is too high or temperature sensor error on
PC 2024. - Check the operating temperature
- Check the fan filter.
2. Replace PC 2024 Expiratory channel.

40011 Multiple CPU restarts on PC 2024. 1. Restart the system.


2. Replace PC 2024 Expiratory channel.

SERVO-air Ventilator System, Service Manual 6 - 33


| 6 | Troubleshooting |

6.6.5 Other Technical error alarms

Error message / Possible cause Recommended action


Airway pressure continuously high 1. Perform a Pre-use check and follow the recommended
actions if the check fails.
Obstruction leading to constant high airway pressure. 2. If the Pre-use check will pass, the alarm is related to
the patient circuit.

Expiratory cassette error 1. Replace the expiratory cassette.


2. Replace PC 2031/PC 1785 Expiratory channel
Technical error in Expiratory cassette or in the communication connector.
with the cassette. 3. Replace PC 2004/PC 2024 Expiratory channel.
Check the Service log > Technical alarms. If Expiration flow
meter Exp. cassette power failure is logged together with
error code Technical error in Expiratory cassette:

1. Replace PC 1991 Control.


2. Replace PC 2028 DC/DC & Battery control.

Nebulizer hardware error 1. Replace the nebulizer.


2. Replace connection cable.
Technical problem with nebulizer hardware. 3. Check internal PC board connections and cable
connections.
Technical problem with connection cable.
4. Replace PC1994 Nebulizer.
5. Replace PC 1991 Control.

Internal nebulizer error - TE 1. Restart the nebulizer.


2. Replace connection cable.
Technical problem with nebulizer hardware. 3. Check internal PC board connections and cable
connections.
Temperature in the nebulizer patient unit too high.
4. Replace PC1994 Nebulizer.
Technical problem with connection cable. 5. Replace PC 1991 Control.

6 - 34 SERVO-air Ventilator System, Service Manual


| Troubleshooting | 6 |

Error message / Possible cause Recommended action


O2 concentration low 1. Perform a Pre-use check and follow the recommended
actions if the check fails.
O2 concentration high 2. Connect an external gas concentration measuring
instrument to the inspiratory outlet to get a reference
There are two main reasons for these alarms: value:
- If the reference value differs from the set O2
• Gas delivery error - Wrong gas concentration is delivered
bullet
from the system, but the gas concentration is measured
concentration, this indicates a gas delivery error.
correctly. - If the reference value corresponds to the set O2
concentration, this indicates a measurement error.
• Measurement error - Correct gas concentration is
bullet
delivered from the system, but the gas concentration is
measured incorrectly. Gas delivery error – Recommended actions

1. Check gas supply; concentration and pressure. See gas


supply requirements as stated in the User’s Manual.
Also note that:
- Gas supply with incorrect O2 concentration may
cause gas delivery error.
- Insufficient gas supply may cause short intermittent
pressure drops and instable O2 concentrations.
2. Replace nozzle units:
- If O2 concentration high – Start with nozzle unit in O2
gas module.
- If O2 concentration low – Start with nozzle unit in Air
gas module.
3. Replace gas modules:
- If O2 concentration high – Start with O2 gas module.
- If O2 concentration low – Start with Air gas module.
4. Replace PC 1991 Control.

Measurement error – Recommended actions

1. Replace O2 cell/sensor.
2. Replace PC 1992 Monitoring.

6.6.6 User interface errors

Error Recommended action


LCD display not lit 1. Replace Display incl. touchscreen.
2. Replace PC 2023 Panel.

User interface will not start up 1. Check User interface cable.


2. Replace PC 2023 Panel.

Touch screen malfunction 1. Check internal and cable connections.


2. Replace Display incl. Touch screen.
3. Replace PC 2023 Panel.

Alarm indicator malfunction 1. Replace PC 2023 Panel.

SERVO-air Ventilator System, Service Manual 6 - 35


| 6 | Troubleshooting |

6 - 36 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7 Preventive maintenance

Table of Contents
7.1 General | 7-2
7.2 Preventive maintenance intervals | 7-2
7.3 Required equipment | 7-2
7.4 Preventive maintenance calculation | 7-2
and notification
7.5 Maintenance kit SERVO-air 5000 h | 7-3
7.6 Performing the Preventive maintenance | 7-3

SERVO-air Ventilator System, Service Manual 7-1


| 7 | Preventive maintenance |

7.1 General 7.3 Required equipment


It is recommended that a complete cleaning of the • Standard service tools.
bullet
system should be performed before carrying out
Preventive maintenance. The cleaning procedure is
• Barometer (or information about the actual
bullet
barometric pressure).
only permitted by personnel trained and authorized
by Getinge. Refer to instructions in the User's Manual • Access to the Service & Settings. Recommended
bullet
(Cleaning and maintenance). but not required.

In some parts of the Preventive maintenance, as


• Preventive maintenance kit containing all parts
bullet
needed during the maintenance.
described in this chapter, access to Service & Settings
is required. • Isolation test cable kit, P/N 68 83 300.
bullet

Optional equipment not covered by this Service


7.4 Preventive maintenance calculation
Manual, may also demand maintenance actions. Refer
and notification
to the optional equipment’s documentation.
7.4.1 Calculation
WARNING! The system must not be
serviced or maintained while in use with The calculation of time until next Preventive
a patient. maintenance (5000 hours) includes only operating
hours, standby hours are excluded.

Note: Trained and authorized 7.4.2 Notification


personnel. The system must be
serviced at regular intervals by The Preventive maintenance notification (PM
personnel trained and authorized by reminder) is shown on the User interface. The PM
Getinge. It is recommended that service reminder is shown after the Pre-use check dialogue
and maintenance is done as a part of (performed or cancelled).
a service contract. Any maintenance
or service must be noted in a log book. There are two different time limits for the PM reminder:

• The first type of PM reminder will be shown when


bullet
there are 200 hours or less until next Preventive
7.2 Preventive maintenance intervals maintenance.

• Preventive maintenance of the system must be


bullet • The second type of PM reminder will be shown
bullet
performed once a year, or every 5000 hours of when there are 0 hours until next Preventive
operation, whichever comes first. maintenance, i.e. time limit for Preventive
maintenance is exceeded.
• The Battery module life time is four years
bullet
The PM reminder is active by default, but can be
(48 months) from manufacturing date or 300 charge
cycles, whichever comes first. disabled in Service & Settings CONFIGURATION >
SETTINGS > GENERAL > PM reminder.
• The Turbine module shall be replaced every 10
bullet
years.

• The Memory backup batteries shall be replaced


bullet
every five years.

7-2 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.5 Maintenance kit SERVO-air 5000 h 7.6 Performing the Preventive maintenance
Only original parts from the manufacturer must be • Disassembling and assembling of the system is
bullet
used. Spare parts and maintenance kits can be required when replacing parts included in the
ordered from your local Getinge representative. Maintenance kit SERVO-air 5000 h. If not stated
otherwise, refer to chapter Disassembling and
When performing this maintenance, the Maintenance
assembling for instructions.
kit SERVO-air 5000 h should be used.
• All steps during Preventive maintenance should be
bullet
The following parts, included in the Maintenance kit documented.
SERVO-air 5000 h, shall be replaced:
• The letters A – D in the text below refers to the
bullet
description of the Maintenance kit above.

7.6.1 Preparations

• Check and note Serial number, System software


bullet
C version and Operating time.

• Check that a User's Manual corresponding to the


bullet
installed System software version is present. Also
check that operating manuals for optional
equipment connected to the system are present.
A B D • Make a general inspection/visual check of the
bullet
S20094 system for external defects or damages. Replace
defective or damaged parts.
Pos Description Remark
A Gas module filter For the O2 gas module. • Make sure that gas supply hoses and connectors
bullet
B Gas module nozzle unit For the O2 gas module. are correct.
C Bacteria filter For the inspiratory
pressure transducer.
• Check if there are any unexpected Technical alarms
bullet
in the Service log available via Service & Settings.
D Air inlet filter For the Turbine module.
Complete with dust filter. • Perform a Pre-use check to make sure that the
bullet
system work properly before the maintenance.

• Set the On/Off switch to Off.


bullet
• Disconnect the mains power cable.
bullet
• Disconnect the O2 gas supply (wall and/or cylinder).
bullet
• Remove patient tubing.
bullet
• If fitted, remove bacteria filter from the Expiratory
bullet
inlet.

SERVO-air Ventilator System, Service Manual 7-3


| 7 | Preventive maintenance |

7.6.2 Gas module filter (A)

• Replace the filter in the O2 gas module.


bullet
- When replacing the filter, move the rubber seal
from the old to the new filter.
- Make sure that the rubber seal is clean. Dirt
particles on the rubber seal may cause leakage.

Note: Check for moisture in the O2 gas module filter


housing. If moisture is found, this may indicate too
high moisture content in the O2 gas supply. Inform
the customer that the O2 gas supply must fulfill
requirements for medical grade gases according to
applicable standards. Refer to Technical data in the
User's Manual.
S20095

7.6.3 Gas module nozzle unit (B)

• Replace the nozzle unit in the O2 gas module.


bullet
- Make sure that the O-ring on top of the nozzle
unit is correctly seated and clean. Dirt particles
on the nozzle unit may cause leakage.

Note: After replacement of the nozzle unit, wait


10 minutes before connecting pressure to the O2
gas module.

S20096

7.6.4 Inspiratory pressure transducer filter (C)

• Replace the filter for the inspiratory pressure


bullet
transducer.
- Make sure that the filter is correctly seated into
the rubber ring.
- Refer to instructions in the User's Manual
(Cleaning and maintenance).

Note: This filter may already be replaced during


cleaning performed in combination with the
Preventive maintenance. S20097

7-4 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.6.5 Air inlet filter for the turbine (D)

• Replace the air inlet filter for the turbine.


bullet
- Pull out the filter from the socket.

S20098

- Push the new filter into the socket.

S20099

7.6.6 Emergency air intake

• Check that there are no obstructions on the


bullet
Emergency air intake (1). Clean if required.

7.6.7 Fan filters

• Remove all fan filters:


bullet
a
- O2 evacuation fan inlet filter (2).
- Main fan inlet filter (3).

• Check if the filters are damaged. Replace damaged


bullet s d
filters.

• If not damaged, clean the filters. The filters can be


bullet
rinsed in water. Shake out and make sure that the
filters are free from excess water. S20100

• Mount the new/cleaned filters.


bullet

SERVO-air Ventilator System, Service Manual 7-5


| 7 | Preventive maintenance |

7.6.8 Fan tests

• O2 evacuation fan (1) and Main fan (2):


bullet
- Turn the system On.
- Check that the O2 evacuation fan and Main fan
start and run for a few seconds.

a s
S20101

7.6.9 Expiratory cassette

• Operating capacity for the membrane is estimated


bullet
to 10 000 000 breathing cycles. When this limit is
passed or if the membrane for some reason has
become defective, it must be replaced. Refer to
instructions in chapter Disassembling and
assembling.

• Membrane capacity (counter) can be shown in the


bullet
System status window. Select SYSTEM STATUS
> Expiratory cassette to check Membrane capacity.

• The operating capacity meter must be reset after


bullet
replacement of the membrane. Note that the reset
function is not implemented in System version 1.0.

7-6 SERVO-air Ventilator System, Service Manual


| Preventive maintenance | 7 |

7.6.10 User interface • Select SYSTEM STATUS > Batteries to check


bullet
battery status. For further information, refer to
• Check the screen readability:
bullet chapter Service procedures, section Battery
- Transparency modules.
- Surface
- Brightness (backlight). The LED backlight is 7.6.14 Turbine module
designed to maintain an acceptable brightness
level throughout the lifetime of the system. • The Turbine module must be replaced every 10
bullet
years. Replacement is described in chapter
- Open a bright window, for example the Service
Disassembling and assembling.
& Settings login window, and check if pixels on
the screen are defective. A few defective pixels
can be accepted. Check that defective pixels are 7.6.15 Memory backup batteries
not concentrated to a small area which can
reduce the readability in this area. • The Memory backup batteries must be replaced
bullet
every five years. Check manufacturing date on the
• Check the touch functions and make sure that the
bullet Memory backup batteries. Replacement is
software responds to these actions. Tap on different
described in chapter Service procedures.
parts of the touch screen.

7.6.16 Safety inspection


7.6.11 Barometric pressure

• Select SYSTEM STATUS > General. Check that


bullet
• Check all external cables and their connections for
bullet
damage.
the Barometric pressure value corresponds to the
actual barometric pressure value at the local site. • Perform an electrical safety test. Use of an Electrical
bullet
The value shown on the User interface may not Safety Tester and the Isolation test cable kit (P/N
differ more than ±5% from the actual barometric 68 83 300) is recommended.
pressure. The electrical safety test is a standard procedure
according to IEC 60601-1:2005 or IEC 62353:2007
• If the value shown on the User interface differs more
bullet
(or corresponding national standards). Allowable
than ±5% from the actual barometric pressure, the
internal barometer must be calibrated. Barometer values and test methods are defined in the
calibration is described in chapter Service standards.
procedures. The following tests shall be performed (IEC 62353
terminology in brackets):

7.6.12 Gas supply pressure transducer A. Protective earthing (Protective earth resistance)
B. Touch current (Equipment leakage current –
• Connect the gas supply (O2).
bullet direct method)
• Select SYSTEM STATUS > General.
bullet C. Patient leakage current (Applied part leakage
current – direct method)
• Disconnect the O2 gas supply.
bullet
D. Insulation (Insulation resistance) – Measured
• Check that the O2 pressure value in the SYSTEM
bullet between mains and protective earth.
STATUS -window drops.
E. Insulation (Insulation resistance) – Measured
between isolated connections and protective
7.6.13 Battery modules earth.
The system is classified within the standards listed
• The Battery modules shall be replaced after
bullet
above, and shall be tested according to the
300 charge cycles or four years (48 months) from
manufacturing date, whichever comes first. following:

SERVO-air Ventilator System, Service Manual 7-7


| 7 | Preventive maintenance |

• SERVO-air base unit: Class 1, Type B.


bullet
Test A, B, D.

• Nebulizer connector: Type BF. Test C, E.


bullet

7.6.17 Completing the Preventive maintenance

• Perform a Functional check according to


bullet
instructions in chapter Important.

• Reset the Preventive maintenance time counter. In


bullet
Service & Settings, tap SERVICE REPORT >
Report completed preventive maintenance > Save
to log. This information will also be stored in the
log.

• Save system log files. Remote Services is


bullet
recommended.

• Inform the customer that Preventive maintenance


bullet
has been performed.

7-8 SERVO-air Ventilator System, Service Manual


| Index | 8 |

8 Index

Table of Contents
8.1 Alphabetic index | 8-2

SERVO-air Ventilator System, Service Manual 8-1


| 8 | Index |

8.1 Alphabetic index Environmental declaration 1-4


Equipotentiality terminal 2-4
A Ethernet bus 2-14
AC mains power 2-2 Ethernet connector 2-4
AC mains power supply connector with fuse 2-4 Exit 5-26, 5-57
AC/DC converter 2-8, 3-14, 4-18 Expiratory cassette 2-4, 2-9, 3-10, 4-25, 7-6
AC/DC section 2-8, 4-16 Expiratory cassette holder 4-19
AC/DC section lid 4-16 Expiratory cassette membrane 4-25
Aerogen nebulizers 2-3 Expiratory channel connector cable 2-5
Air inlet filter for turbine 3-6, 7-3, 7-5 Expiratory flowmeter 6-32
Air inlet for turbine 2-2 Expiratory heater 2-3
Alarm indicator 2-11, 3-18 Expiratory inlet 2-4
Alarm output 2-3, 3-18 Expiratory inlet with moisture trap 2-2, 2-4, 2-9
Alarm output connector 2-4 Expiratory outlet 2-4, 3-11
Alarms 5-17, 5-44 Expiratory section 3-3
Articles of consumption 1-5 Expiratory valve 2-9, 3-11
Assembling guidelines 4-5 Expiratory valve coil 2-8, 3-11, 4-19
Expiratory valve coil connector 2-5
B
External +12 V DC inlet 2-4
Bacteria filter 2-9, 7-3
Barometer 5-53 F
Barometric pressure 7-7 Fan filters 7-5
Battery charger 2-3, 3-18 Fan tests 7-5
Battery compartments 2-2, 2-4 Fans 4-17
Battery module 3-16, 3-18, 5-68, 7-7 Filter 3-5
Battery module (extra) 2-3 Flash memory 3-4
Battery status 5-69 Flow transducer 3-5
Biomed code 5-21, 5-48 Front cover 4-6
Blower BLO 2-13 Front section lid 4-21
Breathing BRE 2-12, 6-2, 6-30 Functional check 1-3
Fuse for external DC power 2-4
C
Cabinet 2-3 G
Cable holder 2-3 Gas cylinder restrainer kit 2-3, 3-18
Calibration 5-53 Gas inlet for air including air inlet filter 2-4
CAN bus 2-14 Gas inlet for O2 2-4, 3-4
Chassis 4-7 Gas module filter 7-3, 7-4
Completing the Preventive maintenance 7-8 Gas module key 3-5
Components with special environmental concern 1-4 Gas module lid 4-12
Configuration 5-11, 5-38 Gas module nozzle unit 7-3, 7-4
Connectors 10-3 Gas supply pressure transducer 7-7
Construction materials 1-5
Cooling fan air inlet with filter (on both sides) 2-4 H
Copy configuration 5-23, 5-50 Handling PC boards 4-2
Covers 4-6 Hazard notices 1-2
Heating foil 2-9
E Humidifier 3-19
EEPROM 3-4 Humidifier holder 2-3, 3-19
Emergency air intake 2-2, 2-4, 7-5

8-2 SERVO-air Ventilator System, Service Manual


| Index | 8 |

I N28 – Ethernet 10-3


I2C bus 2-14 N29/N30 – RS-232 10-3
ID PROM 3-5 N67 – Alarm output 10-3
Information stored in the system 4-3 Nebulizer 3-14, 3-19, 4-21
Inspiratory channel 3-7 Nebulizer connector 2-4
Inspiratory flowmeter 2-7, 3-7, 4-10 Network 5-18, 5-45
Inspiratory flowmeter cable 2-7 Noise level 1-6
Inspiratory gas docking 2-7, 3-7, 4-13 Nozzle unit 3-5
Inspiratory outlet 2-2, 2-4
Inspiratory pipe 3-7 O
Inspiratory pressure transducer filter 7-4 O2 cell 3-8, 4-11
Inspiratory section 2-7, 3-3, 4-9 O2 cell cable 2-7
Inspiratory section lid 4-9 O2 evacuation fan 2-7, 3-9, 4-10
Inspiratory solenoid 3-4, 3-5 O2 gas module 2-7, 3-4, 4-13
Inspiratory valve temperature sensor 3-5 O2 supply 2-2
Installation 5-24, 5-51 On/Off switch 2-4, 3-15, 4-23
Internal communication 2-14 Optional accessories 2-3
Optional equipment 3-18
L Other Technical error alarms 6-34
LCD Display 3-17
LCD Display/Touch screen 2-11 P
Leakage detection 5-62 Packaging material 1-6
Logs 5-8, 5-31 Panel PAN 2-13, 6-2, 6-31
Loudspeaker 2-10, 3-18 Patient circuit 2-2
Loudspeaker connector 2-10 Patient unit 2-2, 2-4
Loudspeaker grid 2-10 PC 1781 Expiratory pressure transducer 2-5, 3-12
PC 1781 Inspiratory pressure transducer 2-7, 3-10,
M 4-15
Main blocks 3-3 PC 1785 Expiratory channel connector 2-5, 2-9,
Main fan 2-8, 3-17, 4-17 3-11, 4-9
Main section 2-5 PC 1786 Expiratory channel cassette 2-9, 3-11
Main section lid 4-8 PC 1991 Control 2-5, 3-13, 4-9
Main units 2-2 PC 1992 Monitoring 2-5, 3-13, 4-9
Mains inlet 3-14, 4-18 PC 1993 Ethernet switch 2-5, 3-14, 4-9
Mains power supply inlet 2-8 PC 1994 Nebulizer 3-14
Maintenance kit SERVO-air 5000h 7-3 PC 2004 Expiratory channel 2-5, 2-9, 3-12, 4-9
Memory backup batteries 5-70, 7-7 PC 2020 Main back-plane 2-6, 3-12, 4-24, 10-4
Memory types 3-4 PC 2021 Rear connectors 3-15, 4-22
Mobile cart 2-3, 3-18 PC 2023 Panel 2-11, 3-17
Mobile cart drawer kit 2-3 PC 2024 Expiratory channel 2-5, 2-9, 3-12, 4-9
Moisture trap inlet 4-25 PC 2025 Battery back-plane 3-16, 4-21
Monitoring 6-23 PC 2026 Battery extension 3-16, 4-21
Monitoring MON 2-13, 6-2 PC 2027 Motor control 3-6
Mouse 5-2 PC 2028 DC/DC & Battery control 2-8, 3-14, 4-21
MSync (adapter for HL7 communication) 2-3, 3-19 PC 2029 Inspiratory channel extension 3-10, 4-15
PC 2030 Inspiratory channel 2-7, 3-9, 4-15, 10-8
N PC 2031 Expiratory channel connector 2-5, 2-9,
N26 – External +12 V DC supply input 10-3 3-11, 4-9, 4-19
N27 – USB port 10-3
SERVO-air Ventilator System, Service Manual 8-3
| 8 | Index |

PC board rack 4-9 System ID 2-14


PC boards and electrical components 4-2 System overview 10-13
Power consumption 1-6 System software installation 5-60
Power indicators 2-4
Power POW 2-13 T
Pressure transducer connection 2-9 Technical error codes and messages 6-23
Preventive maintenance 7-1 Test case 1 – Test tube (Leakage detection) 5-62,
Preventive maintenance calculation and notification 5-63
7-2 Test case 2 – Expiratory channel (Leakage detection)
Pull magnet 3-8 5-62, 5-64
Test case 3 – Inspiratory section (Leakage detection)
R 5-62, 5-66
RAM 3-4 Tests 5-33
Rear connectors lid 4-22 Thermoshell for expiratory cassette 2-3
Replacing electrical components 4-2 Thread locking adhesives 4-5
Replacing PC boards 4-3 Tightening torque 4-5
RS-232 2-14 To the responsible service personnel 1-4
RS-232 connectors 2-4 Top cover 4-6
RS-485 2-14 Touch screen 3-17, 5-55
Turbine 3-6
S Turbine module 2-7, 3-6, 4-13, 7-7
Safety inspection 7-7 Turbine module incl PC 2027 Motor control 3-6
Safety valve 3-8, 4-11 Turbine module one-way valve 4-14
Service & Settings 5-3
Service & Settings menus 5-3, 10-12 U
Service key 5-2 Ultrasonic flowmeter 2-9
Service Manual revision history 9-5 USB mouse 5-2
Service report 5-10, 5-37 USB ports 2-4
Servo DuoGuard 2-3 User interface 2-2, 2-10, 3-3, 4-28, 7-7
Servo Guard filters 2-3 User interface cable 3-17
SERVO-air revision history 9-2 User interface errors 6-35
Set Date & Time 5-20, 5-47
Settings 5-13, 5-40 V
Shelf base 2-3, 3-19 Valve membrane (Expiratory valve) 2-9
SMBus 2-14
W
Software delivery 5-59
Waterbag/IV pole 2-3, 3-19
Software installation 5-58
Wheel lock 2-2
Software option installation 5-61
Software structure 2-12 Y
Special grease 4-5 Y piece holder and hook 2-3
SPI bus 2-14
Startup configuration 5-15, 5-42
Status 5-6, 5-29
Supply pressure transducer 3-5
Support arm 179 2-3, 3-19
Symbols used in this manual 1-2
System block diagram 10-2
System electronics 3-3

8-4 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9 Revision history

Table of Contents
9.1 SERVO-air revision history | 9-2
9.2 Compatibility chart | 9-4
9.3 Service Manual revision history | 9-5

SERVO-air Ventilator System, Service Manual 9-1


| 9 | Revision history |

9.1 SERVO-air revision history

System version # Status Comments


System version 1.0 Not valid Replaced by System version 2.1 in
mandatory update MCC/17/004/IU.
System version 2.1 Valid Clinical improvements:
• GUI adapted to the SERVO-U 2.1 GUI.
bullet
• Updated PRVC algorithm to include
bullet
improvements for spontaneously
breathing patients.
• Pressure level above PEEP maximum
bullet
value in NIV increased from 32 cmH2O
to 62 cmH2O.
• For pediatric and adult patient category
bullet
outside the 130-200 cm height range,
the entered body weight is used and
presented as VT/BW, instead of
VT/PBW (predicted body weight).
• Swedish, Norwegian and Danish
bullet
languages now included.

Service improvements:
• Now possible to reset the expiratory
bullet
cassette membrane counter in Service
& Settings.
• SERVO Communication Interface (SCI)
bullet
updated to Protocol version 0004.
• Stability enhancements to address the
bullet
occurrence of technical errors
Technical errors 43, 78, 80 and 81.
• Timing improvements in the internal
bullet
communication between Panel and
Monitoring to reduce occurrence of
Technical error 20006.
• Improvements to reduce the
bullet
occurrence of buzzer alarms during
software installation (2.1.4).
• In Service & Settings, STATUS > Parts
bullet
data, the new alternative version of PC
2024 Expiratory channel will be
presented with the correct name
(2.1.4).

System version 2.2 Valid Tryout option periods extended from 2 to


4.

9-2 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

System version # Status Comments


System version 4.0 Valid Software improvements:
• Updates in base software:
bullet
• Driving pressure
bullet
• Decelerating Flow in Volume Control
bullet
and the possibility to lock set values
• Short trends
bullet
• Turbine lifetime validated up to
bullet
50000 hours
• Servo Communication Interface
bullet
(SCI) updated to protocol version
0007.
The configuration is copied
automatically during software
installation.
New options in Servo-air:
• High flow therapy
bullet
• Servo compass
bullet
Servo-air NIV is launched with the
release of system version 4.0.

SERVO-air Ventilator System, Service Manual 9-3


| 9 | Revision history |

9.2 Compatibility chart


The chart contains both software and hardware related
information such as:

• Software version required to support hardware


bullet
changes.

• Minimum software version required for a certain


bullet
option.

To find article numbers and revision numbers of the


PC boards:

• See the labels on the PC boards (part


bullet
number/revision number)

• Go to STATUS > Parts data (ARTICLE


bullet
NUMBER/VERSION NUMBER) in the Service &
Settings menu. Service key required.
These article numbers cannot be used when ordering
spare parts. See the Spare parts list for correct
ordering information.

PC 2024
PC 2004 Expiratory PC 2028
PC 1991 PC 1992 Expiratory channel PC 2023 DC/DC &
System SW version Control Monitoring channel 68 81 702 or Panel Battery control
SERVO-air requirement for 66 89 785 66 89 790 66 92 477 68 87 462 68 81 737 68 81 577
System SW version options revision # revision # revision # revision # revision # revision #

1.0.1 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.2 – ≤ 08 ≤ 09 ≥ 00 – ≥ 01 ≥ 04
1.0.3 – ≤ 08 ≤ 09 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.2 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.1.4 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
2.2.0 – ≥ 06 ≥ 06 ≥ 00 ≥ 01 ≥ 01 ≥ 04
S20136

PC 1785/PC 2031 Expiratory channel connector


limitations:

• PC 1785 Expiratory channel connector must be


bullet
used together with PC 2024.

• PC 2031 Expiratory channel connector must be


bullet
used together with PC 2004.

9-4 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3 Service Manual revision history


9.3.1 Service Manual revision 01

• Intended for internal use only.


bullet
• Not published.
bullet

9.3.2 Service Manual revision 02

• Intended for internal use only.


bullet
• Not published.
bullet

9.3.3 Service Manual revision 03

• First released and published version.


bullet

SERVO-air Ventilator System, Service Manual 9-5


| 9 | Revision history |

9.3.4 Service Manual revision 04

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual New layout.
x All chapters in the Service Manual PC 2024 replaces PC 2004.
x All chapters in the Service Manual PC 1785 replaces PC 2031 (in systems with PC 2024).
x 1-1 - Important Restructured and adapted to new guidelines.
1-6
x 1-5 1.8.4 Articles of consumption Filter for O2 cell no longer included in the Preventive
maintenance kit.
x 2-10 - 2.3 User interface Description of User interface clarified.
2-11
x 3-3 Main blocks Description of Main blocks extended and corrected.
x 3-6 3.2 Turbine module including PC Description of Turbine module and PC 2027 extended.
2027 Motor control
x 3-7 3.5 Inspiratory flowmeter Description of Inspiratory flowmeter extended.
x 3-8 3.7 O2 cell Expected cell life time calculation clarified.
x 3-10 3.12 PC 2029 Inspiratory channel Information regarding PC 2029 introduced.
extension
x 3-16 3.31 PC 2026 Battery extension Information regarding PC 2026 introduced.
x 4-6 4.7.1 Tightening torque Tightening torque and procedure clarified.
x 4-7 4.8 Patient unit Disassembling and assembling instructions restructured and
clarified.
x 6-20 6.3.4 Expiratory flowmeter PC 2004: Recommended actions corrected.
x 6-20 6.3.4 Expiratory flowmeter PC 2024: Error code 40001-40011 introduced.
x 7-2 Required equipment An Isolation test cable kit is now available. To be used during
the electrical safety test.
x 7-7 7.5.16 Safety inspection Information regarding the electrical safety test improved.
x 10-12 10.5 System overview Illustration of Turbine module, Inspiratory gas docking and
Inspiratory pipe extended and clarified.

9-6 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3.5 Service Manual revision 05

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual The Service Manual now states Maquet Critical Care AB as
legal manufacturer but with Getinge as the local service
provider.
x All chapters in the Service Manual The information in the Service Manual is based on System
version 2.2.
x 2-12 2.4 Software structure Software structure illustration introduced.
x 3-4 3.1 O2 gas module (1) Illustration and list of included parts corrected.
CAUTION symbol regarding disassembly corrected to Note.
x 3-6 3.2 Turbine module including PC CAUTION symbol regarding disassembly corrected to Note.
2027 Motor control (2)
x 3-8 3.8 Safety valve including pull Illustration and description of safety valve clarified.
3-9 magnet (8)
x 3-11 3.18 PC 2031/PC 1785 Expiratory Reference to Compatibility chart clarified.
channel connector (18)
x 3-12 3.21 PC 2004/PC 2024 Expiratory Reference to Compatibility chart clarified.
channel (21)
x 3-13 3.22 PC 1991 Control (22) Reference to Compatibility chart clarified.
x 3-14 3.23 PC 1992 Monitoring (23) Reference to Compatibility chart clarified.
x 3-14 3.25 PC 1994 Nebulizer (25) Description reduced to cover only PC 1994. Description of
nebulizer functions moved to 3.41.8 Aerogen nebulizers.
x 3-15 3.30 PC 2021 Rear connectors (30) Information regarding USB connector changed, keyboard not
allowed to connect.
x 3-18 3.41.1 Alarm output Information regarding alarm priority corrected.
x 3-18 3.41.3 Battery charger Battery charger information added.
x 3-19 3.41.8 Aerogen nebulizers Description of nebulizer functions introduced, moved from
3.25 PC 1994 Nebulizer.
x 3-19 3.41.11 MSync MSync information added
x 4-3 4.3.3 Replacing PC boards Reference to Compatibility chart clarified.
x 4-5 4.5.1 Tightening torque Tightening torque for gas inlet nipple introduced.
x 4-8 4.8.2 Replacing the handle New spare part.
x 4-9 4.9.2 PC board rack Reference to Compatibility chart clarified.
x 4-9 4.10.1 Inspiratory section lid Changed disassembly/assembly instructions.
x 4-12 4.10.5 Gas module lid Changed disassembly/assembly instructions.
x 4-20 4.11.6 PC 2031/PC 1785 Expiratory Reference to Compatibility chart clarified.
channel connector

SERVO-air Ventilator System, Service Manual 9-7


| 9 | Revision history |

A B C Page Section Comment


x 4-23 4.13.3 Battery guides New spare part.
x 5-9 5.2.5 Service report Information clarified.
x 5-10 5.2.6 Configuration > Settings Available unit settings changed.
x 5-12 5.2.6 Configuration > Alarms Information clarified.
x 5-18 5.3 Software installation Information clarified.
5-19
x 5-20 5.3.6 System software installation Information clarified.
x 5-21 5.3.7 Software option installation Information clarified.
x 5-31 5.6.3 Replacing Memory backup Information regarding PC 1991 changed.
battery
x 6-3 to 6.3 Pre-use check Detailed description of all sub tests included in Pre-use check
6-6 introduced.
x 6-7 to 6.5 Test logs The previous troubleshooting guides replaced by information
6-20 on how to interpret Test result logs and use Check Failure
Identifier (CFI) to troubleshoot the system.
x 6-21 to 6.6 Technical error codes and Minor corrections of the following error codes:
6-31 messages TE 25
TE 37
TE 43
TE 44 (removed)
TE 55
TE 20005
TE 20006
TE 40001 (for PC 2004)
TE 40010
x 6-32 6.6.5 Other Technical error alarms Information clarified.
x 9-2 9.1 SERVO-air revision history Description of System versions 2.1 and 2.2 added.
x 9-3 9.2 Compatibility chart Compatibility chart introduced.
x 10-2 10.1 System block diagram System block diagram moved from chapter 2.
x 10-3 10.2 Connectors Information regarding N27 - USB port changed.
x 10-12 10.5 Service & Settings menus Information based on System version 2.2.
x 10-13 10.6 System overview MSync introduced.

9-8 SERVO-air Ventilator System, Service Manual


| Revision history | 9 |

9.3.6 Service Manual revision 06

In the table below, the following keywords are used:

A = Added. Information about new functionality added.

B = Changed. Information about changed


functionality.

C = Corrected. Corrections made due to printing


errors or clarifications.

A B C Page Section Comment


x All chapters in the Service Manual The information in the Service Manual is based on System
version 4.0.
x All chapters in the Service Manual Updated the Turbine extended lifetime to 10 years
x 1-4 1.6 Service Added a note with complaint handling information.
x 4-4 4.4 Information stored in the system Added a new note in the table.
x 4-5 4.5.2 Thread locking adhesives Updated with loctite 243.
x 4-25 4.15.2 25 Updated the text for expiratory cassette valve membrane.
x 5-4 5.2.2 4 Added service settings for BIOMED user.
x 5-37 5.2.3 27 Updated the text for service report.
x 5-38 5.3.2 Check before installation Updated text for new system version 4.0.
x 5-20 5.3.6 System software installation Updated text for new system version 4.0.
x 6-3 6.3 Pre-use check Updated Gas supply test and Monitor pressure transducer
test.
x 6-7 to 6.5 Test logs Updated different tables according to system version 4.0.
6-20
x 6-21 to 6.6 Technical error codes and Updated different tables according to system version 4.0.
6-31 messages
x 6-32 6.6.5 Other Technical error alarms Added O2 concentration error in the table.
x 7-7 7.6.10 User interface Updated text for touch function.
x 9-2 9.1 SERVO-air revision history Updated the table with system version 4.0.
x 9-3 9.2 Compatibility chart Updated the table with system version 4.0.
x 10-12 10.5 Service & Settings menus Updated the chart with system version 4.0.

SERVO-air Ventilator System, Service Manual 9-9


| 9 | Revision history |

9 - 10 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

10 Diagrams

Table of Contents
10.1 System block diagram |10 - 2
10.2 Connectors |10 - 3
10.3 PC 2020 Main back-plane |10 - 4
10.4 PC 2030 Inspiratory channel |10 - 8
10.5 Service & Settings menus |10 -12
10.6 System overview |10 -13

SERVO-air Ventilator System, Service Manual 10 - 1


| 10 | Diagrams |

10.1 System block diagram

PC 2025 Baery back-plane


Subsystem allocaon

N34

N33
PC 2026 Baery extension

P34

P33
Baery module

P32
Breathing BRE
Slot 1
Monitoring MON
O₂ evacuaon fan
Safety
valve Panel PAN

P31
Baery module Power POW

T
Slot 2, main slot Blower BLO
Inspiratory
flowmeter
System ID
O₂ cell
P17

P23

P19

P18
P37

N15
Gas module O₂
PC 1781 Inspiratory RH
pressure transducer

Turbine module
PC 2027 Motor control
N16

P1 T

P2
Filter dP
PC 2030 Inspiratory channel
P21

P20

Loud-
speaker
N21

N20
P72

PC 2029 Inspiratory
channel extension
N11

N10
SD
N74

P11

P10

N12
P12

P39
On/Off switch
PC 2023 Panel

N77
N70

N36
HDD

ALM_LED

External
N26

PC 1993 Ethernet switch


N3

P3

+12 V DC inlet

Alarm output
N67

connector
N78
P76

P92

T
RS-232
N29

T
N2

P2

PC 1992 Monitoring connector


Touch screen

RS-232
PC 2021 Rear connectors
LCD display

N30

connector
Backlight

N9

P9

N27

USB port (x2)


N1

P1
PC 2028 DC/DC & Baery control

PC 1991 Control

Ethernet
N28

PC 1781 Expiratory PC 2004/PC 2024


pressure transducer Expiratory channel
N4

P4

T
PC 1994 Nebulizer

P40 P41
Nebulizer

Expiratory cassee
P35

PC 2020 Main P45


N5

P5

back-plane PC 1786 PC 2031/


Expiratory Expiratory Expiratory PC 1785
N46

P46

T valve coil flowmeter channel Expiratory


P14

P13

cassee channel
P7

connector

AC/DC converter
PC 2033 AC/DC Mains power
supply connector 110-230 VAC
150 W power supply
Main fan
S20013

10 - 2 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

10.2 Connectors

N26 – External +12 V DC supply input N27 – USB port (x2)

Non-isolated connector. USB 2.0 Type A.

S20106
1. GND
2. +12V_UNREG_EXT_DC
Recommended cable connector:
LEMO 736-FGG2B302CLAD82Z, Circular Push Pull
Connector.
Recommended cable:
Outer diameter 7.2 mm – 8.2 mm, two wire cable, area min.
2
2.5 mm (AWG13).

N28 – Ethernet N29/N30 – RS-232

Isolated connector. Unshielded RJ45.


Speed: 10 Mbit/s.

S20107

Isolated connector. 9 pin D-Sub.


RS-232C for data communication via the Servo
Communication Interface (SCI).
1. -
2. RXD_ISO
3. TXD_ISO
4. DTR_ISO
5. GND_ISO
6. -
7. RTS_ISO
8. -
9. -

N67 – Alarm output

S20108

Isolated connector. Modular plug 4/4 (4P4C)


Switching capability: Max 40 V DC, Max 500 mA, Max 20 W.
1. NO – Normally Open
2. NC – Normally Closed
3. Common
4. Common

SERVO-air Ventilator System, Service Manual 10 - 3


| 10 | Diagrams |

10.3 PC 2020 Main back-plane

10 - 4 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 5


| 10 | Diagrams |

10 - 6 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 7


| 10 | Diagrams |

10.4 PC 2030 Inspiratory channel

10 - 8 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 9


| 10 | Diagrams |

10 - 10 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

SERVO-air Ventilator System, Service Manual 10 - 11


| 10 | Diagrams |

10.5 Service & Settings menus

10 - 12 SERVO-air Ventilator System, Service Manual


| Diagrams | 10 |

10.6 System overview

9 O2 evacuation fan
Inspiratory section

N19
P19
N15 P1
O2 P19 N19 T RH
7
cell
1 O2 Gas O2
ID #6 P18 N18 P23 N23
module inlet

ID #34
3
Inspiratory E2Pot
Inspiratory “0111””1111”
6 Inspiratory channel gas
outlet docking
5 In
Emergency flo spi
wm rat
air intake 8 eteory
4 r 2
Safety T
Safety valve valve Inspiratory Turbine
pull magnet pipe module
ID #32
Air
P17
T
inlet
X
N17
PC 1781 10 PC 2030 Inspiratory
ID #4 channel N16 P1 PC 2027 Motor control P
P20 P21
N20 N21 E2Prom “011”
11
37 PC 2029 Inspiratory P50 Inspiratory
12
channel extension MOD2 pressure
N10 N11 transducer

PC 1994 25
Nebulizer
Nebulizer N1 P32 33 Battery ID #xx
Aerogen 31 PC 2026 PC 2025 module Slot 1
interface Battery extension 32 Battery back-plane
User interface Battery Slot 2
P33 N33 P31 33 ID #xx
P34 module (main slot)
39
N34
P35

36 Touch screen P10 P11


Alarm 19 PC 2020 Main back-plane
indicator DC/DC
T E2Prom “00X”
37 LCD display +24 V
ID #0
+12 V N3
INTERN. P3
24 PC 1993 Ethernet switch MSync
+12 V (optional)
VALVES
+12 V Ethernet PC 2021 Rear
-12 V E2Prom “000” 30
switch E2Prom “100” ID #1026 connectors
+5 V
P72 N70 38 +3.3 V Mains LED
PC 2023 N2 Power On LED
T
Panel P2 23 PC 1992 Monitoring +12 V LAN
ID #2 Battery N22
X1
P22A/B N27 2xUSB
ID #512 E2Pot “00” E2Prom “101” ID #3
ID #35 CPU
X2 T E2Prom “000” ID #1
PS1 N28 Ethernet
N9
P9

P76
28 PC 2028 DC/DC & Battery control

40 N1 RS-232
Interface
Iso. N29
N78 P1 RS-232 (1)
Battery N22 22 PC 1991 Control
P22A/B RS-232
Interface
Iso. N30 RS-232 (2)
CPU
N12
P12

E2Prom “000” ID #256


35 User interface cable External
N36

+12V DC
N4 Fuse F1 N26
P4
supply
PC 2004/PC 2024 input
User interface CPU
E2Prom “000”
21 Expiratory channel
P40 N40
Remote alarm
interface
Iso. N67 Alarm
PC 2023 T E2Pot “0010” ID #1024 output
MOD1 P41
N5
P5

P50 N41 N39 P39


P13 N13 P14 N14
20 PC 1781
Expiratory ID #5
pressure E2Prom “100” 29 On/Off
transducer switch
Power On/Off
PC 1994 N45
PC 2028 34 P45 26 Mains inlet
Main
Patient unit fan 18 P90 N90
PC 1786

PC 2033
PC 2031/ 27
PC 2029

PC 2020 PC 2030 AC/DC converter Fuse F11, F12


PC 1785
PC 1781

N7 P7 Expiratory
PC 1781
Exp.

channel
PC 2031/ connector
Insp.

PC 1785
P46 Expiratory section
PC 2004/

PC 1993

N46
PC 1991

PC 1992
PC 2024

PC 2026

H 17 PC 1786 Expiratory 13 Expiratory cassette


PC 2033 B channel cassette
A
E2Prom “X00”
ID #1025
PC 2021 PC 2025
14 Expiratory valve
Heating foil

Ultrasonic flowmeter 16 Expiratory


outlet
S20001 - 2017-09-12

Expiratory pressure tube Expiratory


Expiratory 15 valve
inlet coil

SERVO-air Ventilator System, Service Manual 10 - 13


| 10 | Diagrams |

10 - 14 SERVO-air Ventilator System, Service Manual


x
© Maquet Critical Care AB 2019. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice.
Order No. 68 84 130 • Rev 06 • Servo-air • Service manual • 2019-09

Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and for life
science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts, healthcare
professionals and medtech specialists, we are improving the everyday life for people - today and tomorrow.

Manufacturer: Maquet Critical Care AB · Röntgenvägen 2, SE-171 54 Solna, Sweden

www.getinge.com

You might also like