Legal information
Warning notice system
This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent damage
to property. The notices referring to your personal safety are highlighted in the manual by a safety alert symbol, notices
referring only to property damage have no safety alert symbol. These notices shown below are graded according to
the degree of danger.

DANGER
indicates that death or severe personal injury will result if proper precautions are not taken.

WARNING
indicates that death or severe personal injury may result if proper precautions are not taken.

CAUTION
indicates that minor personal injury can result if proper precautions are not taken.

NOTICE
indicates that property damage can result if proper precautions are not taken.
If more than one degree of danger is present, the warning notice representing the highest degree of danger will be
used. A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property
damage.
Qualified Personnel
The product/system described in this documentation may be operated only by personnel qualified for the specific
task in accordance with the relevant documentation, in particular its warning notices and safety instructions.
Qualified personnel are those who, based on their training and experience, are capable of identifying risks and
avoiding potential hazards when working with these products/systems.
Proper use of Siemens products
Note the following:

WARNING
Siemens products may only be used for the applications described in the catalog and in the relevant technical
documentation. If products and components from other manufacturers are used, these must be recommended or
approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance
are required to ensure that the products operate safely and without any problems. The permissible ambient
conditions must be complied with. The information in the relevant documentation must be observed.

Trademarks
All names identified by ® are registered trademarks of Siemens AG. The remaining trademarks in this publication may
be trademarks whose use by third parties for their own purposes could violate the rights of the owner.
Disclaimer of Liability
We have reviewed the contents of this publication to ensure consistency with the hardware and software described.
Since variance cannot be precluded entirely, we cannot guarantee full consistency. However, the information in this
publication is reviewed regularly and any necessary corrections are included in subsequent editions.

Siemens AG Document order number: A5E46619106 Copyright © Siemens AG 2019 - 2021.


Digital Industries Ⓟ 04/2021 Subject to change All rights reserved
Postfach 48 48
90026 NÜRNBERG
GERMANY
Table of contents

1 Getting started ...................................................................................................................................... 5


1.1 Commissioning SIWAREX DB with SIWAREX WP electronic weighing system .......................... 5
1.2 Replacing SIWAREX JB with SIWAREX DB ............................................................................... 5
1.3 Assigning parameters to the electronic weighing system with SIWAREX DB ........................... 6
2 Introduction ........................................................................................................................................... 9
2.1 Purpose of this documentation ............................................................................................ 9
2.2 Document history ................................................................................................................ 9
2.3 Designated use .................................................................................................................... 9
2.4 Industrial use note ............................................................................................................... 9
2.5 Checking the consignment................................................................................................. 10
2.6 Scope of delivery ............................................................................................................... 10
2.7 Security information .......................................................................................................... 10
2.8 Transportation and storage ................................................................................................ 11
2.9 Notes on warranty ............................................................................................................. 11
3 Safety notes ......................................................................................................................................... 13
3.1 Prerequisites for safe use.................................................................................................... 13
3.1.1 Laws and directives ............................................................................................................ 13
3.1.2 Conformity with European directives .................................................................................. 13
3.2 Use in hazardous areas....................................................................................................... 14
3.2.1 Special conditions for use................................................................................................... 14
4 Description........................................................................................................................................... 15
4.1 Applications....................................................................................................................... 15
4.2 Design ............................................................................................................................... 16
4.3 Internal view...................................................................................................................... 18
4.4 LEDs .................................................................................................................................. 19
4.5 Example of nameplates ...................................................................................................... 20
5 Installing/mounting ............................................................................................................................. 21
5.1 Basic safety notes............................................................................................................... 21
5.2 Installing SIWAREX DB ........................................................................................................ 22
6 Connecting .......................................................................................................................................... 23
6.1 Basic safety notes............................................................................................................... 23
6.2 Connect SIWAREX DB to SIWAREX WP electronic weighing system and load cell................... 25

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 3
Table of contents

6.3 Connect the cable .............................................................................................................. 26


7 Service and maintenance .................................................................................................................... 29
7.1 Disposal............................................................................................................................. 30
8 Diagnostics and troubleshooting......................................................................................................... 31
8.1 Meaning of the LED states.................................................................................................. 31
8.2 No connection to the SIWAREX DB...................................................................................... 31
9 Technical data...................................................................................................................................... 33
9.1 Rated conditions ................................................................................................................ 33
9.2 Construction...................................................................................................................... 33
9.2.1 Die-cast aluminum housing................................................................................................ 33
9.2.2 Stainless steel enclosure..................................................................................................... 34
9.2.3 Suitable cables................................................................................................................... 35
9.3 Power supply ..................................................................................................................... 35
9.4 Load cell interface.............................................................................................................. 36
9.5 RS485 interface ................................................................................................................. 36
9.6 Electrical, EMC and climatic requirements........................................................................... 37
9.7 Certificates and approvals................................................................................................... 39
10 Dimension drawings ............................................................................................................................ 41
11 Ordering data ...................................................................................................................................... 43
11.1 Accessories ........................................................................................................................ 43
A Product documentation and support .................................................................................................. 45
A.1 Product documentation ..................................................................................................... 45
A.2 Technical support............................................................................................................... 45
Index .................................................................................................................................................... 47

SIWAREX DB
4 Operating Instructions, 04/2021, A5E46619106-AC
Getting started 1
1.1 Commissioning SIWAREX DB with SIWAREX WP electronic
weighing system

Condition
You have read the following safety notes:
• General safety notes (Page 13)
• Basic safety notes: Installing/mounting (Page 21)
• Basic safety notes: Connecting (Page 23)
Read the entire device manual in order to achieve the optimum performance of the device.

Procedure
1. Open the cover of the SIWAREX DB.
2. Install the SIWAREX DB.
Installing SIWAREX DB (Page 22)
3. Connect the SIWAREX DB to the SIWAREX WP electronic weighing system and the load cells.
Connect SIWAREX DB to SIWAREX WP electronic weighing system and load cell (Page 25)
4. Close the cover of the SIWAREX DB according to the tightening torque (Page 33).
5. Switch on the supply voltage.
6. Assign parameters to the SIWAREX WP electronic weighing system using the SIMATIC HMI,
the SIWATOOL or the SIMATIC S7 control system.
Assigning parameters to the electronic weighing system with SIWAREX DB (Page 6)

1.2 Replacing SIWAREX JB with SIWAREX DB

Condition
You have read the following safety notes:
• General safety notes (Page 13)
• Basic safety notes: Installing/mounting (Page 21)
• Basic safety notes: Connecting (Page 23)
Read the entire device manual in order to achieve the optimum performance of the device.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 5
Getting started
1.3 Assigning parameters to the electronic weighing system with SIWAREX DB

Procedure
1. Switch off the supply voltage of the SIWAREX WP electronic weighing system.
2. Remove all cables from the SIWAREX JB.
3. Remove the SIWAREX JB.
4. Install the SIWAREX DB.
Installing SIWAREX DB (Page 22)
5. Connect the SIWAREX DB to the SIWAREX WP electronic weighing system and the load cells.
Connect SIWAREX DB to SIWAREX WP electronic weighing system and load cell (Page 25)
6. Close the cover of the SIWAREX DB according to the tightening torque (Page 33).
7. Switch on the power supply.
8. Update the software for the SIMATIC HMI, the SIWATOOL and the SIMATIC S7 control system
(function block).
9. Update the firmware of your SIWAREX WP electronic weighing system.
Firmware update SIWAREX WP electronic weighing system (http://
support.automation.siemens.com/WW/view/en/10807015/133100)
10.Assign parameters to the SIWAREX WP electronic weighing system using the SIMATIC HMI,
the SIWATOOL or the SIMATIC S7 control system.
Assigning parameters to the electronic weighing system with SIWAREX DB (Page 6)

1.3 Assigning parameters to the electronic weighing system with


SIWAREX DB

Introduction
You can assign parameters to the SIWAREX WP electronic weighing systems using the
SIMATIC HMI, the SIWATOOL or the SIMATIC S7 control system. The procedure described for you
is oriented to the data records (DS) from SIWATOOL and the SIMATIC S7 control system.

Procedure
1. Set the Load cell type parameter to SIWAREX DB in data record DR03. (Not relevant for
SIWAREX WP321)
2. Set the RS 485 protocol parameter to SIWAREX DB and the RS 485 baud rate to 115 200 bits/
s in data record DR13.
3. Activate the load cell input for every connected load cell in the data record DR70.
4. Specify whether 4- or 6-wire load cells are connected for the Sensor type parameter in data
record DR70.

SIWAREX DB
6 Operating Instructions, 04/2021, A5E46619106-AC
Getting started
1.3 Assigning parameters to the electronic weighing system with SIWAREX DB

5. In order to identify the load cell type and the installation location of your load cells enter the
Load cell manufacturer, Load cell order number and Location Designation of Sensor in the
data record DR70.
6. The details of the data data records and parameters are specified in the parameter
assignment tables.

See also
Electronic weighing system manuals (https://support.industry.siemens.com/cs/ww/en/ps/
17796/man)

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 7
Getting started
1.3 Assigning parameters to the electronic weighing system with SIWAREX DB

SIWAREX DB
8 Operating Instructions, 04/2021, A5E46619106-AC
Introduction 2
2.1 Purpose of this documentation
These instructions contain all information required to commission and use the device. Read the
instructions carefully prior to installation and commissioning. In order to use the device
correctly, first review its principle of operation.
The instructions are aimed at persons mechanically installing the device, connecting it
electronically, configuring the parameters and commissioning it, as well as service and
maintenance engineers.

2.2 Document history


The most important changes in the documentation when compared with the respective
previous edition are given in the following table.

Manual edition Note


04/2021 Revised technical specifications
02/2020 Stainless steel enclosure added

2.3 Designated use


Use the device in accordance with the information on the nameplate and in the Technical data
(Page 33).
If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.

2.4 Industrial use note

NOTICE
Use in a domestic environment
This Class A Group 1 equipment is intended for use in industrial areas.
In a domestic environment this device may cause radio interference.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 9
Introduction
2.7 Security information

2.5 Checking the consignment


1. Check the packaging and the delivered items for visible damages.
2. Report any claims for damages immediately to the shipping company.
3. Retain damaged parts for clarification.
4. Check the scope of delivery by comparing your order to the shipping documents for
correctness and completeness.

WARNING
Using a damaged or incomplete device
Risk of explosion in hazardous areas.
• Do not use damaged or incomplete devices.

2.6 Scope of delivery

  

① Junction box with cover


② 4 x M16 x 1.6 plastic/metal cable glands
③ 3 x M16 x 1.5 plastic/metal blanking plugs
④ 1 x M20 x 1.5 EMC cable gland
Figure 2-1 Scope of delivery

2.7 Security information


Siemens provides products and solutions with industrial security functions that support the
secure operation of plants, systems, machines and networks.

SIWAREX DB
10 Operating Instructions, 04/2021, A5E46619106-AC
Introduction
2.9 Notes on warranty

In order to protect plants, systems, machines and networks against cyber threats, it is necessary
to implement – and continuously maintain – a holistic, state-of-the-art industrial security
concept. Siemens’ products and solutions constitute one element of such a concept.
Customers are responsible for preventing unauthorized access to their plants, systems,
machines and networks. Such systems, machines and components should only be connected to
an enterprise network or the internet if and to the extent such a connection is necessary and only
when appropriate security measures (e.g. firewalls and/or network segmentation) are in place.
For additional information on industrial security measures that may be implemented, please
visit
https://www.siemens.com/industrialsecurity.
Siemens’ products and solutions undergo continuous development to make them more secure.
Siemens strongly recommends that product updates are applied as soon as they are available
and that the latest product versions are used. Use of product versions that are no longer
supported, and failure to apply the latest updates may increase customer’s exposure to cyber
threats.
To stay informed about product updates, subscribe to the Siemens Industrial Security RSS Feed
under
https://www.siemens.com/industrialsecurity.

2.8 Transportation and storage


To guarantee sufficient protection during transport and storage, observe the following:
• Keep the original packaging for subsequent transportation.
• Devices/replacement parts should be returned in their original packaging.
• If the original packaging is no longer available, ensure that all shipments are properly
packaged to provide sufficient protection during transport. Siemens cannot assume liability
for any costs associated with transportation damages.

NOTICE
Insufficient protection during storage
The packaging only provides limited protection against moisture and infiltration.
• Provide additional packaging as necessary.

Special conditions for storage and transportation of the device are listed in Technical data
(Page 33).

2.9 Notes on warranty


The contents of this manual shall not become part of or modify any prior or existing agreement,
commitment or legal relationship. The sales contract contains all obligations on the part of
Siemens as well as the complete and solely applicable warranty conditions. Any statements
regarding device versions described in the manual do not create new warranties or modify the
existing warranty.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 11
Introduction
2.9 Notes on warranty

The content reflects the technical status at the time of publishing. Siemens reserves the right to
make technical changes in the course of further development.

SIWAREX DB
12 Operating Instructions, 04/2021, A5E46619106-AC
Safety notes 3
3.1 Prerequisites for safe use
This device left the factory in good working condition. In order to maintain this status and to
ensure safe operation of the device, observe these instructions and all the specifications relevant
to safety.
Observe the information and symbols on the device. Do not remove any information or symbols
from the device. Always keep the information and symbols in a completely legible state.

Symbol Explanation
Consult operating instructions

3.1.1 Laws and directives


Observe the safety rules, provisions and laws applicable in your country during connection,
assembly and operation. These include, for example:
• National Electrical Code (NEC - NFPA 70) (USA)
• Canadian Electrical Code (CEC) (Canada)
Further provisions for hazardous area applications are for example:
• IEC 60079-14 (international)
• EN 60079-14 (EU)

3.1.2 Conformity with European directives


The CE mark on the device is a sign of conformity with the following European directives:

Electromagnetic Directive of the European Parliament and of the Council on the


compatibility EMC harmonization of the laws of the Member States relating to elec‐
2014/30/EU tromagnetic compatibility
Atmosphère explosible ATEX Directive of the European Parliament and of the Council on the
2014/34/EU harmonization of the laws of the Member States relating to
equipment and protective systems intended for use in potentially
explosive atmospheres
2011/65/EU RoHS Directive of the European Parliament and of the Council on the
restriction of the use of certain hazardous substances in electrical
and electronic equipment
The applicable directives can be found in the EU conformity declaration of the specific device.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 13
Safety notes
3.2 Use in hazardous areas

WARNING
Improper device modifications
Risk to personnel, system and environment can result from modifications to the device,
particularly in hazardous areas.
• Only carry out modifications that are described in the instructions for the device. Failure to
observe this requirement cancels the manufacturer's warranty and the product approvals.

3.2 Use in hazardous areas

Qualified personnel for hazardous area applications


Persons who install, connect, commission, operate, and service the device in a hazardous area
must have the following specific qualifications:
• They are authorized, trained or instructed in operating and maintaining devices and systems
according to the safety regulations for electrical circuits, high pressures, aggressive, and
hazardous media.
• They are authorized, trained, or instructed in carrying out work on electrical circuits for
hazardous systems.
• They are trained or instructed in maintenance and use of appropriate safety equipment
according to the pertinent safety regulations.

WARNING
Use in hazardous area
Risk of explosion.
• Only use equipment that is approved for use in the intended hazardous area and labeled
accordingly.
• Do not use devices that have been operated outside the conditions specified for hazardous
areas. If you have used the device outside the conditions for hazardous areas, make all Ex
markings unrecognizable on the nameplate.

3.2.1 Special conditions for use

WARNING
Transient overvoltages
Take measures to ensure that temporary transient overvoltages of more than 119 V do not
occur.

SIWAREX DB
14 Operating Instructions, 04/2021, A5E46619106-AC
Description 4
4.1 Applications

Overview
SIWAREX DB is a digital junction box for extended diagnostic and monitoring options in
conjunction with the following SIWAREX WP electronic weighing systems:
• SIWAREX WP231
• SIWAREX WP321
Thanks to individual channel monitoring, error states such as wire break and overload can be
identified in a targeted manner. Connecting SIWAREX DB to a SIWAREX WP electronic weighing
system ensures seamless integration in the SIMATIC world.

Application
SIWAREX DB is the optimum solution wherever DMS sensors such as load cells, force sensors and
torque measuring shafts are used for measuring in the SIMATIC environment and optimum
diagnostics options are needed. SIWAREX DB is suitable for all weighing applications,
particularly level measurement, platform weighing and dosing.
SIWAREX DB is used in weighing systems with extra-low voltages ≤ 32 V. SIWAREX DB is used to
connect load cells to the electronic weighing system. With SIWAREX DB, you can connect a scale
with up to 4 load cells to one electronic weighing system.

LC

LC
SIWAREX DB W

LC 

LC
LC Maximum 4 load cells on SIWAREX DB
W Electronic weighing system
① Load cell with 4-wire or 6-wire system
② See cable recommendations Accessories (Page 43)
Figure 4-1 Connection principle of SIWAREX DB and electronic weighing system with up to 4 load cells

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 15
Description
4.2 Design

4.2 Design

  

① Cover ⑤ 2 M16 x 1.5 plastic blanking plugs for the


connection of RS-485 and 24 V DC.
Or use the EMC cable glands, that you can
order here: Accessories (Page 43).
② 4 screws for the cover ⑥ M20 x 1.5 metal EMC cable gland for the
connection of a SIWAREX cable (RS-485 and
24 V DC)
③ Nameplate ⑦ Mounting screw for equipotential bonding
conductor (functional grounding)
④ M16 x 1.5 four plastic cable glands for the
connecting cables of the load cells
Figure 4-2 Design of SIWAREX DB die-cast aluminum housing

 

① Cap nut
② Fitting
Figure 4-3 Cable gland of die-cast aluminum housing

SIWAREX DB
16 Operating Instructions, 04/2021, A5E46619106-AC
Description
4.2 Design

 

  

① Cover ⑤ 2 M16 x 1.5 metal blanking plugs for the


connection of RS-485 and 24 V DC.
Or use the EMC cable glands, that you can
order here: Accessories (Page 43).
② 4 screws for the cover ⑥ M20 x 1.5 metal EMC cable gland for the
connection of a SIWAREX cable (RS-485 and
24 V DC)
③ Nameplate ⑦ Fixing lug including mounting flange for
equipotential bonding conductor (func‐
tional grounding)
④ 4 M16 x 1.5 metal EMC glands for the con‐
necting cables of the load cells
Figure 4-4 Design of SIWAREX DB stainless steel enclosure

   

① Cap nut
② Sealing plug
③ Fitting
④ Lock nut
Figure 4-5 Cable gland of stainless steel enclosure

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 17
Description
4.3 Internal view

4.3 Internal view


 

+SIG- I +SEN- I +EXC-

LC1

+SIG- I +SEN- I +EXC-



OUTPUT
LC2
L+ I M I D+ I D- SHIELD
24VDC I RS485 LC Status +SIG- I +SEN- I +EXC-

 LC1
LC2
LC3
LC3
INPUT +SIG- I +SEN- I +EXC-
LC4

 Err
PWR LC4
SHIELD
Device
Status

① LC1 ... LC4 load cell terminals ⑤ Connection of the supply voltage and
RS-485 terminal
② One shield terminal for each load cell con‐ ⑥ DIP switches 1 and 2:
nection Terminating resistor RS-485
③ DIP switch for RS-485 address of the SIWAR‐
EX DB
④ LEDs (Page 19)
Figure 4-6 Internal view SIWAREX DB

SIWAREX DB
18 Operating Instructions, 04/2021, A5E46619106-AC
Description
4.4 LEDs

4.4 LEDs
+ I M I D+ I D-
VDC I RS485 LC Status
LC1
LC2
LC3
LC4

Err
PWR
Device
Status +SIG- I +SEN- I +EXC-

LC1

+SIG- I +SEN- I +EXC-

OUTPUT
LC2
L+ I M I D+ I D- SHIELD
24VDC I RS485 LC Status +SIG- I +SEN- I +EXC-
LC1
LC2
LC3
LC3
INPUT +SIG- I +SEN- I +EXC-
LC4

Err
PWR LC4
SHIELD
Device
Status

LC1 to LC4 LEDs of load cells


Err Error
PWR Supply voltage
LC Load cell
Figure 4-7 Position and assignment of LEDs

See also
Meaning of the LED states (Page 31)

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 19
Description
4.5 Example of nameplates

4.5 Example of nameplates


The nameplate shows the article number and other important product information.





1PXFS3BUJOH7%$
.BYJNVN1PXFS8
 &ODMPTVSF/&."5:1&9

 ,1'& 217(4
3FGFSUPNBOVBM (




① Manufacturer ⑥ Consult operating instructions


② QR code for mobile website with device-spe‐ ⑦ Protection class
cific information about the product
③ Conformity with country-specific guidelines ⑧ Serial number
④ Manufacturer's address ⑨ Article number
⑤ Place of manufacture ⑩ Product name
Figure 4-8 Example of aluminum nameplate

.)"%

1PXFSSBUJOH7%$
.BYJNVNQPXFS8
ާࣙ5Bࣙާ
IND. CONT. EQ.
&

&ODMPTVSF/&."5ZQF9 *1
%&,3""5&99 
*&$&Y%&,9
**(&9FD**$5
** %&9UD***$5ާ
3FGFSUPNBOVBM

⑪ ATEX symbol and approval numbers


Figure 4-9 Example nameplate, stainless steel with information on the hazardous area

SIWAREX DB
20 Operating Instructions, 04/2021, A5E46619106-AC
Installing/mounting 5
5.1 Basic safety notes

WARNING
Loss of type of protection
Risk of explosion. Damage to device if the enclosure is open or not properly closed. The type of
protection specified on the nameplate or in Technical data (Page 33) is no longer guaranteed.
• Make sure that the device is securely closed.

WARNING
Improper commissioning in hazardous areas
Device failure or risk of explosion in hazardous areas.
• Do not commission the device until it has been mounted completely and connected in
accordance with the information in Technical data (Page 33).
• Before commissioning take the effect on other devices in the system into account.

WARNING
Loss of explosion protection
Risk of explosion in hazardous areas if the device is open or not properly closed.
• Close the device as described in Connecting (Page 23).

WARNING
Opening device in energized state
Risk of explosion in hazardous areas
• Only open the device in a de-energized state.
• Check prior to commissioning that the cover, cover locks, and cable inlets are assembled in
accordance with the directives.
Exception: Devices having the type of protection "Intrinsic safety Ex i" may also be opened in
energized state in hazardous areas.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 21
Installing/mounting
5.2 Installing SIWAREX DB

NOTICE
Incorrect mounting
The device can be damaged, destroyed, or its functionality impaired through improper
mounting.
• Before installing ensure there is no visible damage to the device.
• Make sure that process connectors are clean, and suitable gaskets and glands are used.
• Mount the device using suitable tools. Refer to the information in Technical data (Page 33).

5.2 Installing SIWAREX DB

Introduction
You can mount the junction box in any position. The cable glands should point downward.

Condition
The cover is open.

Procedure
1. Prepare the installation location.
2. Set the holes according to the dimension drawings.
Dimension drawings (Page 41)
3. Screw the junction box securely to the wall with four screws.

SIWAREX DB
22 Operating Instructions, 04/2021, A5E46619106-AC
Connecting 6
6.1 Basic safety notes

WARNING
Unsuitable cables, cable glands and/or plugs
Risk of explosion in hazardous areas.
• Use only cable glands/plugs that comply with the requirements for the relevant type of
protection.
• Tighten the cable glands in accordance with the torques specified in Technical data
(Page 33).
• Close unused cable inlets for the electrical connections.
• When replacing cable glands, only use cable glands of the same type.
• After installation, check that the cables are seated firmly.

WARNING
Improper power supply
Risk of explosion in hazardous areas as result of incorrect power supply.
• Connect the device in accordance with the specified power supply and signal circuits. The
relevant specifications can be found in the certificates, in Technical data (Page 33) or on
the nameplate.

WARNING
Unprotected cable ends
Risk of explosion through unprotected cable ends in hazardous areas.
• Protect unused cable ends in accordance with IEC/EN 60079-14.

WARNING
Improper laying of shielded cables
Risk of explosion through compensating currents between hazardous area and the
non‑hazardous area.
• Shielded cables that cross into hazardous areas should be grounded only at one end.
• If grounding is required at both ends, use an equipotential bonding conductor.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 23
Connecting
6.1 Basic safety notes

WARNING
Connecting or disconnecting device in energized state
Risk of explosion in hazardous areas.
• Connect or disconnect devices in hazardous areas only in a de-energized state.
• Install a suitable switch-off device.
Exceptions:
• Devices having the type of protection "Intrinsic safety Ex i" may also be connected in
energized state in hazardous areas.

NOTICE
Ambient temperature too high
Damage to cable sheath.
• At an ambient temperature ≥ 60 °C (140 °F), use heat-resistant cables suitable for an
ambient temperature at least 20 °C (36 °F) higher.

NOTICE
Condensation in the device
Damage to device through formation of condensation if the temperature difference between
transportation or storage and the mounting location exceeds 20 °C (36 °F).
• Before taking the device into operation, let the device adapt for several hours in the new
environment.

Note
Electromagnetic compatibility (EMC)
You can use this device in industrial environments, households and small businesses.
For metal enclosures there is an increased electromagnetic compatibility compared to high-
frequency radiation. This protection can be increased by grounding the enclosure.

See also
Technical data (Page 33)

SIWAREX DB
24 Operating Instructions, 04/2021, A5E46619106-AC
Connecting
6.2 Connect SIWAREX DB to SIWAREX WP electronic weighing system and load cell

Note
Improvement of interference immunity
• Lay signal cables separate from cables with voltages > 60 V.
• Use cables with twisted wires.
• Keep device and cables at a distance from strong electromagnetic fields.
• Take account of the conditions for communication specified in the Technical data (Page 33).
• Use shielded cables to guarantee the full specification according to HART/PA/FF/Modbus/
EIA-485/Profibus DP.

6.2 Connect SIWAREX DB to SIWAREX WP electronic weighing system


and load cell

Procedure
1. Open the cover of the SIWAREX DB.
2. Screw in an M16 x 1.5 cable gland for each load cell.
3. Screw in an M20 x 1.5 EMC cable gland for the signal cable to the SIWAREX WP electronic
weighing system.
4. Attach the cables from SIWAREX DB to the load cell and the SIWAREX WP electronic weighing
system.
Connect the cable (Page 26)
6,:$5(;'% 6,:$5(;:3HOHFWURQLFZHLJKLQJV\VWHP

/RDGFHOO SHIELD 6LJQDOFDEOH


56 56
:KLWH
SENSE- ' '

EXC- 5HG
' '
SIGNAL-

SENSE+ b9b'& b9b'&


%OXH
EXC+ / /
%ODFN
SIGNAL+ 0 0

(TXLSRWHQWLDOERQGLQJFRQGXFWRU &DEOHJODQG

6VKLHOGFRQQHFWLRQHOHPHQW
Figure 6-1 Connect SIWAREX DB to SIWAREX WP electronic weighing system and load cell

5. Activate the terminating resistor RS-485 ① by setting both DIP switches to ON.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 25
Connecting
6.3 Connect the cable

6. Set the RS485 address by setting the DIP switch 1 ② to ON.


+ -

+SIG- I +SEN- I +EXC-

LC1

+SIG- I +SEN- I +EXC-

OUTPUT + -
LC2
L+ I M I D+ I D- SHIELD
24VDC I RS485 L C Status +SIG- I +SEN- I +EXC-
LC1
+ I M I D+ I D-

LC2

C I RS485 L C Status
LC3
LC3
INPUT +SIG- I +SEN- I +EXC-
LC4

LC1 Err
PWR LC4
Device
+ - SHIELD

LC2 Status

LC3
LC4

Err
PWR

7. Close any unused opening in the enclosure with a blanking plug.


8. Connect the equipotential bonding conductor (functional grounding) to the enclosure
exterior.
Use shielded cable lugs.
9. Close the cover of the SIWAREX DB according to the tightening torque 1.5 to 2 Nm.

See also
Die-cast aluminum housing (Page 33)
Internal view (Page 18)

6.3 Connect the cable

Condition
You require a screwdriver with a maximum blade width of 2.5 mm.
You can use the SIWAREX cable for supply voltage and RS-485 communications. Accessories
(Page 43)

SIWAREX DB
26 Operating Instructions, 04/2021, A5E46619106-AC
Connecting
6.3 Connect the cable

Procedure
1. Place the cable shield in the EMC cable gland over a large area.



Figure 6-2 Place the cable shield in the EMC cable gland.

2. For plastic cable glands, connect the corresponding wire of the SIWAREX load cell to the
shield terminal.
3. Strip the cable ends at least 6 mm.
4. Push the screwdriver into the rectangular opening of the terminals.
5. Insert the stripped cable into the round opening as far as it will go.
6. Remove the screwdriver from the rectangular opening.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 27
Connecting
6.3 Connect the cable

SIWAREX DB
28 Operating Instructions, 04/2021, A5E46619106-AC
Service and maintenance 7
Note
The device is maintenance-free.

WARNING
Dust layers above 5 mm
Risk of explosion in hazardous areas.
Device may overheat due to dust build up.
• Remove dust layers in excess of 5 mm.

NOTICE
Penetration of moisture into the device
Damage to device.
• Make sure when carrying out cleaning and maintenance work that no moisture penetrates
the inside of the device.

WARNING
Impermissible repair of explosion protected devices
Risk of explosion in hazardous areas
• Repair must be carried out by Siemens authorized personnel only.

WARNING
Impermissible accessories and spare parts
Risk of explosion in areas subject to explosion hazard.
• Only use original accessories or original spare parts.
• Observe all relevant installation and safety instructions described in the instructions for the
device or enclosed with the accessory or spare part.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 29
Service and maintenance
7.1 Disposal

WARNING
Humid environment
Risk of electric shock.
• Avoid working on the device when it is energized.
• If working on an energized device is necessary, ensure that the environment is dry.
• Make sure when carrying out cleaning and maintenance work that no moisture penetrates
the inside of the device.

7.1 Disposal

Devices described in this manual should be recycled. They may not be dis‐
posed of in the municipal waste disposal services according to the Directive
2012/19/EC on waste electronic and electrical equipment (WEEE).
Devices can be returned to the supplier within the EC, or to a locally ap‐
proved disposal service for eco-friendly recycling. Observe the specific reg‐
ulations valid in your country.
Further information about devices containing batteries can be found at:
Information on battery/product return (WEEE) (https://
support.industry.siemens.com/cs/document/109479891/)

Note
Special disposal required
The device includes components that require special disposal.
• Dispose of the device properly and environmentally through a local waste disposal
contractor.

SIWAREX DB
30 Operating Instructions, 04/2021, A5E46619106-AC
Diagnostics and troubleshooting 8
8.1 Meaning of the LED states

LED state Error Causes and remedies


LC LED off. Load cell channel not
active.
LC LED lit red. Channel alarm A detailed error message can be obtained from the connected SIWAR‐
EX WP electronic weighing system. Electronic weighing system manuals
(https://support.industry.siemens.com/cs/ww/en/ps/17796/man)
LC LED lit green. Channel active and
no error.
PWR LED lit green. Status supply
24 V DC present /
ready for operation
ERR LED lit red. SIWAREX DB errors A detailed error message can be obtained from the connected SIWAR‐
EX WP electronic weighing system. Electronic weighing system manuals
(https://support.industry.siemens.com/cs/ww/en/ps/17796/man)

See also
LEDs (Page 19)
Load cell interface (Page 36)
Rated conditions (Page 33)
Electrical, EMC and climatic requirements (Page 37)

8.2 No connection to the SIWAREX DB

Introduction
There is no connection between the SIWAREX WP electronic weighing system and the
SIWAREX DB.

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 31
Diagnostics and troubleshooting
8.2 No connection to the SIWAREX DB

Troubleshooting

+ I M I D+ I D-
C I RS485 LC Status
LC1
+SIG- I +SEN- I +EXC-

LC1

+SIG- I +SEN- I +EXC-

OUTPUT
LC2
L+ I M I D+ I D- SHIELD
24VDC I RS485 LC Status +SIG- I +SEN- I +EXC-
LC1
LC2
LC3
LC3
INPUT +SIG- I +SEN- I +EXC-
LC4

Err
PWR LC4
SHIELD
Device
Status

Figure 8-1 DIP switches 1 and 2: Terminating resistor RS-485

• Disable and enable DIP switches 1 and 2.

SIWAREX DB
32 Operating Instructions, 04/2021, A5E46619106-AC
Technical data 9
9.1 Rated conditions

Operating conditions Value


Possible applications Outdoor use (incl. wet locations, not in flood areas) and indoors
Relative humidity 5 ... 95%
Non-condensing, corresponds to relative humidity (RH) degree 2 ac‐
cording to DIN IEC 61131-2
Ambient temperature Die-cast aluminum enclosure: -20 ... +80 °C
Stainless steel enclosure -40 to +85 °C
Storage temperature -50 ... +100 °C
Degree of protection IP66 / NEMA/TYPE 4X for devices without explosion protection
IP65 for devices with explosion protection
Degree of pollution Outside of the enclosure: 3
Inside the enclosure: 2
Height above sea level -1000 ... 2000 m

9.2 Construction

9.2.1 Die-cast aluminum housing

Variable Value
Enclosure material Die-cast aluminum housing
DIN EN 1706 EN AC-AlSi 12 (Fe)
Weight Approx. 1 kg
Color Powder-coated, anthracite

Cable glands Screw connection Terminal area Tightening torque


Fitting and cap nut
(Page 16)
• Load cells M16 x 1.5 mm 4 ... 8 mm 2.5 Nm
• RS-485 M16 x 1.5 mm 4.5 to 10 mm 5.5 Nm
• Signal cable M20 x 1.5 mm 7 ... 12 mm 8 Nm
Spring-loaded terminals
• Vibration test According to EN 60068 2 6: 1996 05
10 ... 150 ... 10 Hz, amplitude 0.35 mm

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 33
Technical data
9.2 Construction

Variable Value
Core cross-section
• Load cell terminal 0.2 ... 1.5 mm² (AWG24 ... AWG16)
• Supply voltage
terminal
• RS-485 terminal
• Shield terminal 0.2 ... 2.5 mm2 (AWG24 ... AWG14)
Tightening torque 1.5 ... 2 Nm
mounting screw of the
cover

9.2.2 Stainless steel enclosure

Variable Value
Enclosure material Stainless steel 1.4301
Weight Approx. 1.5 kg
Color Polished stainless steel
Cable glands Screw connection Terminal area Tightening torque
Fitting and cap nut
(Page 17)
• Load cells M16 x 1.5 mm 5 to 10 mm 5.5 Nm
• Signal cable M20 x 1.5 mm 10 to 14 mm 8 Nm
Spring-loaded terminals
• Vibration test according to DIN EN 60068 2 6: 1996 05
10 ... 150 ... 10 Hz, amplitude 0.35 mm
Core cross-section
• Load cell terminal 0.2 ... 1.5 mm² (AWG24 ... AWG16)
• Supply voltage
terminal
• RS-485 terminal
• Shield terminal 0.2 ... 2.5 mm2 (AWG24 ... AWG14)
Tightening torque 1.5 ... 2 Nm
mounting screw of the
cover

SIWAREX DB
34 Operating Instructions, 04/2021, A5E46619106-AC
Technical data
9.3 Power supply

9.2.3 Suitable cables

Cable Version
Load cell cable SIWAREX load cells are supplied with connecting cables.
Power and signal cable The junction box is connected to the electronic weighing system via
the power and signal cable.

Use Copper Conductors Only


• Shielded cable 4 cores
• Recommended cable SIWAREX cable Li2Y2x0,75St+2x(2x0,34St)-CY
See section Accessories (Page 43)
• Maximum length See Electronic weighing system manual (https://
support.industry.siemens.com/cs/ww/en/ps/17796/man)
• Diameter of individual con‐ ≥ 0.1 mm (hazardous area)
ductors
Equipotential bonding conductor At least 4 mm²
(function earth cable)

Use Copper, Copper-Clad Aluminum, or Aluminum Conductors

9.3 Power supply

Power supply SELV power supply


Rated voltage 24 V DC
Low / high limits Die-cast aluminum housing: 9 ... 32 V DC for device
version without explosion protection
Stainless steel enclosure: 9 ... 30 V DC
Non-periodic overvoltages 40 V DC for 500 ms with a recovery time of 50 s
Overvoltage category according to IEC 60664-1 I
Maximum current consumption 100 mA with 24 V DC
Maximum loss of module 2.4 W

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 35
Technical data
9.5 RS485 interface

9.4 Load cell interface

Load cell interface


Accuracy Corresponds to the technical specifications of the
SIWAREX WP electronic weighing system used.
Electronic weighing system manuals (https://
support.industry.siemens.com/cs/ww/en/ps/
17796/man)
Connection 4-wire or 6-wire
Strain gauge input resistance 330 ... 4 100 Ω

9.5 RS485 interface

RS-485 interface
Standard EIA-485 for connection to SIWAREX WP electronic weighing sys‐
tem
Baud rate 115.2 and 57.6 Kbps
Data bits 8
Parity Even
Stop bits 1
Terminating resistors (can be activa‐ 120 Ω
ted)
Electrical isolation 500 V AC
Cable length ≤ 1000 m at 115.2 Kbps
(SIWAREX cable Accessories (Page 43))

SIWAREX DB
36 Operating Instructions, 04/2021, A5E46619106-AC
Technical data
9.6 Electrical, EMC and climatic requirements

9.6 Electrical, EMC and climatic requirements


Electrical protection and safety requirements

Fulfilled requirement Standards Comments


Safety regulations IEC 61010-1:2010
+C1:2011 + C2:2013
Use in hazardous areas IEC 60079-0:2017 When installing the SIWAREX WP electron‐
IEC 60079-15:2017 ic weighing system in hazardous areas,
observe the special operating conditions
according to SIWAREX Product Informa‐
tion – "Use of SIWAREX modules in a Zone
2 Hazardous Area" A5E02192786A.
Electromagnetic IEC 61326-1: 2012 To comply with electromagnetic compat‐
compatibility ibility requirements, ground all shielded
cables at both ends.
If the shielded cable is routed out of the
hazardous area for explosion-proof equip‐
ment, both ends of the cable shield must
be connected to equipotential bonding.

Electromagnetic compatibility

Table 9-1 Requirements: Interference emission in industrial areas in accordance with IEC 61326-1

Comments Standard Limits


Emission of radio interferences Class A industrial environment: • 30 to 230 MHz, 40 dB (mV/m)
(electromagnetic fields) • CISPR 11: 2009 Q at 10 m
• 230 ... 1000 MHz, 47 dB (mV/
m) Q at 10 m

Table 9-2 Requirements: Interference immunity in industrial areas in accordance with IEC 61000-6-2
and IEC 61326-1

Standard Test standard Connection Severity level


Electrostatic dis‐ IEC 61326-1 Direct ± 4 kV
charge (ESD) Indirect ± 4 kV
IEC 61000-4-2 Electrostatic air discharge ± 8 kV
Radiation IEC 61326-1 Enclosure 80 ... 1000 MHz, 10V/m, AM 1 kHz
IEC 61000-4-3 80 %
1.4 ... 2.0 GHz, 3V/m, AM 1 kHz 80 %
2.0 ... 2.7 GHz, 1 V/m, AM 1 kHz 80 %
IEC 61000-6-2 Enclosure 1.4 ... 6.0 GHz, 3V/m, AM 1 kHz 80 %
Burst pulses IEC 61326-1 Power supply cable ± 2 kV / 5 kHz and 100 kHz 5 ns / 50 ns
IEC 61000-4-4 Load cell cable ± 1 kV / 5 kHz and 100 kHz 5 ns / 50 ns
RS485 cable ± 1 kV / 5 kHz and 100 kHz 5 ns / 50 ns

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 37
Technical data
9.6 Electrical, EMC and climatic requirements

Standard Test standard Connection Severity level


Surge voltage IEC 61326-1 Power supply cable ± 2 kV line to ground ± 1 kV line to
IEC 61000-4-5 ground 1.2 μs / 50 μs
Load cell cable ± 1 kV line to ground 1.2 μs / 50 μs
RS485 cable ± 1 kV line to ground 1.2 μs / 50 μs
Electromagnetic IEC 61326-1 Power supply cable 0.15 ... 80 MHz, 3 V, AM 1 kHz 80 %
RF fields Load cell cable 0.15 ... 80 MHz, 3 V, AM 1 kHz 80 %
IEC 61000-4-6 RS485 cable 0.15 ... 80 MHz, 3 V, AM 1 kHz 80 %

Ambient conditions

Table 9-3 Operating conditions in accordance with IEC 60721

Operation IEC 60721-3-3


Class 4K4H, field device, weatherproof
Storage/transport IEC 60721-3-2
Class 2K4 without precipitation

Table 9-4 Climatic requirements

Comments Ambient conditions Application areas


Atmospheric pres‐ Operation IEC 60068-2-13 1080 to 795 hPa (opera‐
sure tion)
(-1000 to + 2000 m above
sea level)
For transport and IEC 60068-2-13 1 080 to 660 hPa (storage)
storage (‑1 000 to +3 500 m above
sea level)

Reliability
Mean Time Between Failure (MTBF)
The MTBF calculation results in the following values for the modules:

Table 9-5 MTBF

SIWAREX DB MTBF in years


SIWAREX DB 73.01 years @TA = 40 °C

SIWAREX DB
38 Operating Instructions, 04/2021, A5E46619106-AC
Technical data
9.7 Certificates and approvals

9.7 Certificates and approvals

ATEX/IECEx
Certificate number
• DEKRA 21ATEX0025X
• IECEx DEK. 21.0018X
Type of protection "increased safety ec", Zones 2 and 22
Protection by enclosure tb
Identification
• ATEX II 3 G Ex ec IIC T4 Gc
II 3 D Ex tc IIIC T4 Dc
• IECEx Ex ec IIC T4 Gc
Ex tc IIIC T4 c

See also
Certificates (http://www.siemens.com/processinstrumentation/certificates)

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 39
Technical data
9.7 Certificates and approvals

SIWAREX DB
40 Operating Instructions, 04/2021, A5E46619106-AC
Dimension drawings 10

Figure 10-1 Dimension drawing SIWAREX DB aluminum, dimensions in mm [inch]

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 41
Dimension drawings

Figure 10-2 Dimension drawing SIWAREX DB stainless steel, dimensions in mm [inch]

SIWAREX DB
42 Operating Instructions, 04/2021, A5E46619106-AC
Ordering data 11
11.1 Accessories
You can order accessories online: Industry Mall (https://mallstage.industry.siemens.com/
mall/en/b0/Catalog/Products/10038765?tree=CatalogTree)
• EMC cable glands for connecting SIWAREX junction box
• Signal cable
SIWAREX cable standard applications (orange)

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 43
Ordering data
11.1 Accessories

SIWAREX DB
44 Operating Instructions, 04/2021, A5E46619106-AC
Product documentation and support A
A.1 Product documentation
Product documentation for process instrumentation is available in the following formats:
• Certificates (http://www.siemens.com/processinstrumentation/certificates)
• Downloads (firmware, EDDs, software) (http://www.siemens.com/processinstrumentation/
downloads)
• Catalogs and technical data sheets (http://www.siemens.com/processinstrumentation/
catalogs)
• Manuals (http://www.siemens.com/processinstrumentation/documentation)
You have the option to view, open, save or configure the manual.
– "Display": The manual opens in HTML5 format.
– "Configure": You can register and configure the specific documentation for your plant
here.
– "Download": The manual is opened or saved in PDF format.
– "Download as html5, only PC": The manual is opened or saved in HTML5 view on your PC.
You can also find manuals at Industry Online Support (https://support.industry.siemens.com/cs/
ww/de/sc/2067) using the mobile app. Download the app to your mobile device and scan the QR
code.

See also
Generally applicable information/SIMATIC manual overview (https://
support.industry.siemens.com/cs/ww/en/view/109742709)

A.2 Technical support

Technical support
If this documentation does not completely answer your technical questions, you can enter a
Support Request (http://www.siemens.com/automation/support-request).
Additional information on our technical support can be found at Technical Support (http://
www.siemens.com/automation/csi/service).

Service & support on the Internet


In addition to our technical support, Siemens offers comprehensive online services at Service &
Support (http://www.siemens.com/automation/serviceandsupport).

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 45
Product documentation and support
A.2 Technical support

Contact
If you have further questions about the device, contact your local Siemens representative at
Personal Contact (http://www.automation.siemens.com/partner).
To find the contact for your product, go to "all products and branches" and select "Products &
Services > Industrial automation > Process instrumentation".
Contact address for business unit:
Siemens AG
Digital Industries
Process Automation
Östliche Rheinbrückenstr. 50
76187 Karlsruhe, Germany

SIWAREX support
• Email (mailto:hotline.siwarex@siemens.com)
• Phone: +49 (721) 667-1200 CET 8:00 to 17:00

SIWAREX DB
46 Operating Instructions, 04/2021, A5E46619106-AC
Index

C N
Catalog nA, (See ec)
Technical data sheets, 45
Certificates, 13, 45
Climatic requirements, 38 Q
Customer Support, (Refer to Technical support)
Qualified personnel, 14

D R
Disposal, 30
Reliability, 38
Documentation
Edition, 9
Downloads, 45
S
Safety requirements, 37
E Scope of delivery, 10
Service, 45
Electromagnetic compatibility, 37
Service and support
Internet, 45
Support, 45
H Support request, 45
Hazardous area
Laws and directives, 13
Qualified personnel, 14 T
History, 9
Technical support, 45
Hotline, (Refer to Support request)
partner, 46
personal contact, 46
Test certificates, 13
I
Immunity, 37
W
Warranty, 11
L
Laws and directives
Disassembly, 13
Personel, 13

M
Manuals, 45
Modifications
correct usage, 14
improper, 14
MTBF, 38

SIWAREX DB
Operating Instructions, 04/2021, A5E46619106-AC 47
Index