Professional Documents
Culture Documents
V-781381-320 - IFU - VyntusBody - V 01.00 EN
V-781381-320 - IFU - VyntusBody - V 01.00 EN
Vyntus BODY
Vyntus APS
V-781381-320
Version 01.00
for SentrySuite® Software ≥ 3.20
English
Instructions for Use Vyntus BODY/APS
This manual has been composed with the assistance of Adobe® Creative Suite®
programs Adobe InDesign®, Adobe Photoshop®, Adobe Illustrator® and Adobe
Acrobat®. The complete pages have been produced with Adobe InDesign® for PC
and are also available as PDF files. Adobe Acrobat Reader is required in order to
open these PDF files.
Trademarks
Windows is either a registered trademark or a trademark of Microsoft Corporation
in the United States and/or other countries.
Adobe, Creative Suite, InDesign, Photoshop, Illustrator and Acrobat are registered
trademarks of Adobe Systems Incorporated in the USA and/or other countries.
Sidestream is a registered trademark of Respironics UK Ltd.
Contact:
Vyaire Medical GmbH
Leibnizstrasse 7
97204 Hoechberg, Germany
+49 931 4972-0 tel
+49 931 4972-423 fax
E-Mail: MarketingInfo.EU@Vyaire.com
Copyright
All rights, also those of translations, are reserved. Written permission of Vyaire
shall be obtained for the reproduction of the manual or an excerpt thereof by any
means which includes printing, photostatic duplication, microfilms or any other
process.
Note
This document contains copyright information. All rights are reserved. It is not
allowed to copy, duplicate or translate this manual in any other language
without obtaining written approval from Vyaire. Vyaire reserves the right to alter
the information included in this document without notice.
Names of persons mentioned in the context of this manual are fictitious - any
resemblance to living or deceased persons is purely incidental and not intended.
Table of Contents
System Configuration Vyntus BODY...................................................................... 5
System Configuration Option Vyntus APS............................................................. 7
Indications for Use and Usability............................................................................ 8
Practical Hints..................................................................................................... 10
Symbols for Notes on Safety in Instruction Manuals............................................ 11
Declaration of Conformity................................................................................... 11
Switching on/off Vyntus BODY/Vyntus APS......................................................... 12
USS Module - Ultrasonic Sensor.......................................................................... 14
System Start........................................................................................................ 14
The Start Screen/ Home Page.............................................................................. 15
Measurements with the MicroGard Filter............................................................ 19
SeSQM-Quality Management.............................................................................. 19
Ambient Data..................................................................................................... 20
Calibration / Verification Frequency..................................................................... 21
Volume Calibration / Verification......................................................................... 22
Box Calibration................................................................................................... 29
Gas Analyzer Calibration for Vyntus BODY with Diffusion................................... 31
Enter Patient Data............................................................................................... 36
Load Patient Data ............................................................................................. 37
Show Patient Data.............................................................................................. 39
Measurement Programs...................................................................................... 42
Basic Conditions prior to starting a Measurement............................................... 43
Slow Spirometry.................................................................................................. 44
Forced Spirometry............................................................................................... 48
Maximum Voluntary Ventilation (MVV)............................................................... 51
Bodyplethysmography......................................................................................... 53
Diffusion SB Realtime.......................................................................................... 60
Diffusion SB Intrabreath...................................................................................... 67
R Occlusion (Option)........................................................................................... 74
Respiratory Drive P 0.1........................................................................................ 77
MIP / MEP........................................................................................................... 80
SNIP (Option)...................................................................................................... 86
Perform a SNIP Measurement.............................................................................. 86
Compliance (Option)........................................................................................... 89
Preparing for a Measurement.............................................................................. 89
Catheter Positioning........................................................................................... 93
Rhinomanometry (Option)................................................................................. 100
Preparing for the Measurement........................................................................ 100
Perform a measurement.................................................................................... 101
Set Visit Levels.................................................................................................. 106
Quality Assessment/Interpretation/Comments................................................... 107
Bronchoprovocation Test.................................................................................. 110
Optionally, the Vyntus BODY can be extended inside the cabin with
• Diffusion gas analyser for Single Breath Diffusion Real-Time with CO/CH4 incl.
Intrabreath Diffusion
• Vyntus APS aerosol provocation system for bronchial provocation
• ROcc, P0.1, MIP / MEP, SNIP, Rhinomanometry and Compliance measure-
ments.
The applied parts are indicated in the chapter “Disposables, spare parts and acces-
sories”.
Source Of Energy
The Vyntus APS consists of the extended basic module with an integrated com-
pressor placed on the CART of the Vyntus BODY and a nebulizer unit mounted
inside the cabin. The Vyntus APS is connected to the same PC like the Vyntus
BODY via USB interface.
Specific measurement applications of the Vyntus BODY are used to observe the
provocation test.
The applied parts are indicated in the chapter “Disposables, spare parts and
accessories”.
Source Of Energy
Information on Usability:
Ambient requirements:
- Use of the medical product in a medical environment (e.g. doctor‘s office,
clinic, hospital)
- For diffusion single breath (Realtime and Intrabreath) measurements a good
ventilation of the room with fresh air must be ensured.
- Communication with the patient must be ensured (e.g. eye contact, duty of
care)
- Ambient conditions see section “Technical Data“
User requirements:
- Operation only by specially trained physicians or medical technicians
- Reading and comprehension of accompanying documents
- Adequate computer skills incl. operating elements
- Communication with the patient and the medical product must be ensured
(use of e.g. visual and acoustic aids)
Patient requirements:
- Adequate patient cooperation is ensured
- Use with patients who are at least 4 years of age is possible
- No restrictions with regard to height, weight and gender
General requirements:
- The patient must not be left unattended during the measurement.
Practical Hints
Instructing the patient:
Many of the patients will not have performed a pulmonary function test before.
These patients (but also experienced ones) may be agitated and nervous. It will be
in the interest of the operator and patient if the procedure is thoroughly explained
with regard to the time taken and purpose of the test.
Prior to a patient‘s first examination of the day, it is important to give the patient
a waiting period of 10 to 15 minutes to relax and to normalize his/her ventilation
which might be increased due to walking etc.
For more detailed information on the programs refer to the Online Manual
which can be called up from every application program.
Injury to Damage
Degree of Danger Use in case of:
Persons to Property
Declaration of Conformity
“I/O“ switch
Switching on:
Use the “I/O“ switch on the backside of the cabin for switching on.
The Power status is shown at ceiling cover inside the cabin.
I/O Switch
CART 3b:
Switch on the PC, monitor and printer. It is not necessary to switch on the
Vyntus APS or its respective power supply unit separately. As soon as the PC is
turned on, the Vyntus APS device will automatically exit standby mode.
All
All system
system components,
components, which
which connected
connected toto wall
wall sockets
sockets must
must be
be easily
easily
accessible, allows to disconnect easily the system during an incident.
accessible, allows to disconnect easily the system during an incident.
Isolating
Isolating from
from supply
supply mains:
mains: For
For safe
safe disconnection
disconnection from
from the
the mains,
mains, the
the
power supply cord must be pulled.
power supply cord must be pulled.
Switching off:
Use the “I/O“ switch again for switching off the cabin.
CART 3b:
Shut down the PC. Switch off the monitor and the printer. It is not necessary/
possible to switch off the Vyntus APS device or its respective power supply unit
separately. As soon as the PC is switched off, the Vyntus APS device automati-
cally enters standby mode.
Connect the USS MicroGard adapter to the USS Module, then connect the
MicroGard filter to the USS MicroGard adapter (see picture).
Please use a new MicroGard filter and a clean mouthpiece for every patient.
In order to ensure firm hold USS module and the demand valve, press both
components together and slightly turn them while attaching them to the FPV
block.
MicroGard filter
Connection cable
System Start
To start the SentrySuite Software just double-click the <Launch SentrySuite>
symbol on the Windows® desktop.
In the top left-hand corner of the screen you will find the icons for entering and
searching patient data.
Main items containing calibration tools, patient data, measurements, report and
review.
For further information please see Online Manual “Home and Help”
This row shows the device warm up time. For the Vyntus BODY and the Vyntus
APS the warmup takes 5 minutes. For the Vyntus BODY with Diffusion the
warmup takes 45 minutes.
Measurements:
Click the icon to open a menu containing all available measurement programs.
The menu includes depending on the configuration:
- Spirometry
- Bodyplethysmography
- Diffusion SB Realtime
- Diffusion SB Intrabreath
- R Occlusion (optional)
- P 0.1 (optional)
- MIP / MEP (optional)
- SNIP (optional
- Rhinomanometry (optional)
- Compliance (optional)
- Bloodgas Data Transfer (optional)
- Offline data input
- Spirometer Data Transfer
- Bronchial Challenge (optional)
Calibration Tools:
- Volume Calibration
- Gas analyzer Calibration
- Box Calibration
The other groups contain programs that are not routinely used. Clicking the
Button „Configuration and Tools“ delivers following items:
Configuration:
Connectivity... Program for configuring GDT/HIS settings
and for checking the connectivity to other
databases (such as JLAB, Vmax etc.).
Please contact Vyaire tech support for
further advice.
System Configuration Program for hardware/software
configuration
Global Settings... Program for performing settings which
usually are only performed once
Database Handling:
Opens the programs:
- Backup/Restore Program for performing a backup and
restore of the complete SQL database
including the stored patient and
measurement data
- Data Export/Import... Program for exporting or importing data
which has been created in SentrySuite
(see Online Manual “Home and Help”)
- Patient Data Management Program for changing the patient
ID, deleting selected patient and
measurement data, and moving
measurement data (visits, levels,
measurements) (optional)
Caution:
Program only for experienced operators.
User Administration:
Opens the programs:
- User Administration Program for the administration of users
and groups including distribution of
rights (optional)
The programs listed above are not described in this Instructions for Use. For a
detailed description refer to the Online Manual which can be started directly
from the individual programs.
For Vyntus BODY with USS module, the use of the MicroGard filter is
mandatory.
Procedure:
Ticked means: The software corrects the influence of the bacterial filter for the
determination of the volume (BTPS correction).
SeSQM-Quality Management
Ambient Data
The ATP-BTPS correction factor for inspiratory flows and volumes will be
determined from the ambient data. Therefore, ambient data have to be
updated at regular intervals. Incorrect or unprecise ambient data will result in
incorrect measuring results.
If an Ambient Unit is included in the delivery package, the current values for room
temperature, relative humidity and barometric pressure are recorded with the help
USB-Ambient Unit of this module and automatically saved in the system.
In order to assure correct data, the Ambient Unit must neither be exposed to
direct sunlight nor positioned near heating elements.
If the lung function measuring station is not equipped with an Ambient Unit, the
current ambient data are to be entered manually. In this case, room temperature,
relative humidity and absolute barometric pressure need to be determined by
a small home weather station and entered manually in the entry mask of the
Ambient Data screen menu.
If the ambient data change by more than 2 °C / 2 °F for temperature and 10 %
for relative humidity, the current ambient conditions should be entered again in
order to ensure that the parameters, especially inspiratory values, are determined
correctly. The calibration procedure does not need to be repeated.
To enter the current ambient data manually or to check the data transferred into
the system via the Ambient Unit, click on the “Ambient Data“ symbol.
The ambient data program can be selected from each application (e.g.
spirometry, Body...) and each calibration program.
Click on the corresponding blue field of the respective ambient unit and
enter the current ambient data.
Volume Calibration
The measurement recording systems Vyntus APS must be calibrated in order to
determine flow and volume.
As a result of the calibration, the computer will calculate gain factors, which will
be considered in the later measurements in order to assure that the values achieve
the maximum accuracy.
Volume Verification
Verification only serves to check an existing calibration. Calibration data will not
be overwritten and will be preserved.
Only the deviation from the predicted value, i.e. the pump volume of 3 liters, will
be checked, i.e. volume accuracy should be within a range of ± 3.5 %.
Please note:
If the volume calibration is started from the Home Page via “Calibration
Icon” and <Volume Calibration> on a combined measurement system, the
type of system to be calibrated must be selected first.
Depending on the existing system select the corresponding measuring system for
performing a volume calibration/verification.
Before starting the actual calibration/verification, the USS Module including USS
MicroGard adapter and MicroGard filter must be connected to the calibration
syringe (see picture below). If the measurement is performed without using a filter,
like Vyntus APS, calibration must also be performed without a filter.
During volume verification using the Vyntus BODY with Diffusion option, the drying
tube needs to be connected to the adapter nipple.
USS
MicroGard Adapter
Vyntus APS
APS unit
Make sure to pump slowly and regularly and do not exceed the pump flow of
2 L/s.
If auto start has been set, calibration/verification can be started directly. Otherwise,
start the calibration/verification with <F1> “Start”. If Vyntus BODY is selected as
measurement device, then automatically volume verification (instead of calibration)
is started.
On the screen you are asked to move the pump piston smoothly back and
forth. It is important to pump without interruption from impact to impact.
The first pump stroke is not relevant for the calibration/verification result
and will be discarded.
1 pump stroke = pump twice, i.e. from impact to impact
The upper part of the screen shows the results of the last saved calibration “Prev“
and of the current calibration “Act“ as well as the difference “%Diff“ between
“Prev“ and “Act“. The current volume values must be within an accuracy of
±6 %.
The mean values are calculated from the individual pump strokes (column “1-4“)
and displayed in the “Act“ column.
The lines “Gain ex“ and “Gain in“ show the gain factors calculated for expiration
and inspiration.
The middle part of the screen shows the individual pump strokes as a Flow-Volume
curve on the left side and as a bar diagram in liters on the right side.
The accuracy range of ± 6 % is indicated by the vertical lines in the graph to the
left and the right.
If all quality criteria are met, the following message will be displayed:
“Gain and quality factors within their limits“
The “Calibration result“ is “Ok“.
Calibration result:
If calibration meets all relevant criteria, this is indicated by the bar indicator in the
bottom right part of the screen: the blue rectangle is at the right end of the bar
indicator (“OK“).
If calibration has not been successful, the user is prompted to repeat the trial.
If the quality criteria are not met, a corresponding prompt will appear and the
calibration result is assessed to be “Not Ok“.
If, for example, “Quality factor out of range“ is displayed, the calibration must
be repeated with <F1> “Start”.
Calibration means:
The existing calibration data will be completely replaced by the results of the new
calibration.
Verification means:
Verification only serves to check an existing calibration. Calibration data will not
be overwritten and will be preserved.
Only the deviation from the predicted value, i.e. the pump volume of 3 liters, will
be checked, i.e. volume accuracy should be within a range of ± 3.5 %. It is recom-
mended to use a 3-L pump.
The volume accuracy can be set within the settings of the program volume calibra-
tion. Volume accuracy can be 3.5%, 3.0% or 2.5%.
Note: During the verification process a maximum of two strokes beyond the
verification limit are ignored. The operator has two more strokes to achieve the
verification limit.
Only the deviation from the predicted value, i.e. the set, certified pump volume,
will be checked. That means the volume accuracy should be within a range of
± 3.5 %.
The ± 3.5 % tolerance is made up of 0.5 % maximum variation of the pump
volume and 3 % acceptable spread of the flow-volume sensor.
As a rule, the pump volume used for verification corresponds to the volume used
for calibration.
In general please check that your syringe is not leaking and re-calibration cycle
of the syringe is ensured.
If all above does not help please contact Global Customer Support.
You can find the contact data at the end of your Instruction for Use.
Box Calibration
The program “Box Calibration“ allows the calibration of the sensor system (shift
volume) and the verification of the leak tightness of the box (time constant).
The Box Calibration should be done at least once a day, preferably 10 minutes
after having switched on the unit. As a result, the computer will calculate
important calibration factors.
The box calibration program can either be started by calling up the “Calibration
Icon“ on the Home Page and selecting <Box Calibration> or directly from the
measurement program.
We recommend to not skip the two minutes waiting time. This ensures that
there are stable pressure and temperature conditions in the box.
The right part of the screen shows the quality of the single parameters by means
of two bar indicators. “Gain” and “time constant” should be in the green range.
Furthermore, the blue rectangle in the window “Calibration result” at the
bottom of the screen should be positioned on “Ok”.
It is possible to repeat the complete Box Calibration or only the Tau verification or
the shift volume calibration if required.
The gas calibration should be done once a day, preferably after having switched on
the unit (note the warmup time of 45 minutes of the gas analyzers). The computer
calculates correction factors as a result of each calibration.
These correction factors will be considered in later measurements in order to
ensure that the recorded values are correct.
For the calibration, the drying tube has to be removed from the USS MicroGard
Adapter for gas sampling and connected to the cal gas port on the analyzer unit.
Important note for the Vyntus BODY with Diffusion in the corner:
Analyzer unit Analyzer unit
Important note for the Vyntus BODY with Diffusion in the ceiling:
Housing for gas ports Housing for gas ports
As already mentioned above, the gas calibration can either be started from the
Home Page via “Calibration Icon“ and <Gas Analyzer Calibration> or directly
from the measurement programs
• Diffusion SB Intrabreath
• Diffusion SB Realtime
Click <F1> “Start” to start the calibration. The calibration of CO/CH4 starts
automatically. The following screen display appears:
If the gas calibration on the Vyntus BODY was more than 36 hours ago,
the gas path is automatically flushed for 3 minutes with the calibration gas
before the actual gas calibration begins. This ensures that only pure, undilut-
ed calibration gas is used for gas calibration. A time bar is displayed for this.
The flushing process can be skipped with <Cancel>.
If the calibration has been successful, a “green tick“ appears behind the
respective gas.
The bottom part shows on the left the quality check in text form.
If calibration meets all quality criteria, this is indicated by the bar in the bottom
right part of the screen: the blue frame is at the right end of the bar (“Ok“).
If criteria have not been met, the blue frame is at the left end of the bar (“Not
Ok“).
The user is requested to repeat the calibration.
By clicking <Return to ...> the calibration will be concluded and the last opened
measurement program will appear.
Note:
The C2H2 curves displayed in the screenshots are only visible with the Vyntus ONE device and are not valid for the Vyntus BODY with Diffusion
option.
Before measuring a patient for the first time, the patient‘s personal data must
be stored in the database. Predicted values are calculated from biometric data,
so please verify that the entered data is correct.
Incorrect patient data produce incorrect predicted values!
For the calculation of the predicted values, the patients biological gender is
relevant. For more detailed information regarding the gender settings, please
see the Online Manual “Patient Data“, chapter “Gender Settings“.
This information is relevant for SentrySuite® Software ≥ 3.10
The Patient Chart is displayed. The cursor flashes in the first entry field
“Identification”. The program is ready for entering data.
The entry lines from “Identification“ to “Weight“ marked with an “*“ are
mandatory fields. These data are the patient’s master data including biometric
details used to calculate predicted values. All other entries are optional.
Enter data in a field and confirm each entry with the <TAB key>. The cursor will
then jump to the next box.
When a patient whose data is already stored in the database comes for
another visit, you can reload his/her data from the database. You do not have
to enter the data again.
A list of all patients whose data have been accessed recently or on the same day is
shown. The desired patient can be selected and loaded immediately.
If an existing last name and/or first name and/or ID number is entered, the
respective patient data will be displayed in the “Search result“ window as soon
as <Search> is clicked.
Another method to search for a patient is to enter the first character(s) in the
boxes Identification, Last Name, First name or Physician. Please note the following:
Enter the first character(s) in a box (for example “Sm“ for “Smith“ in the
“Last Name“ box).
Characters can be replaced by an “*“ (for example “*ith“ for Smith; “*1205*“
for JohnSmith120570).
It is also possible to search for patients who were measured within a specific
period of time, for example, all patients measured today, yesterday, within the
past 7 days, the past 30 days. A user-specific period of time (from - to) may also be
entered.
Click <Search> without entering any search criteria to open a list of all patients.
It is advised to always use at least one alpha-numeric character to expedite your
search, especially with large databases.
Click on the desired patient. By clicking <Select> the appropriate patient data and
existing visits will be displayed.
If the desired patient cannot be found, click <New Patient> in order to enter the
patient data of the new patient.
Click <Add new visit> and the patient data such as height and weight which
were entered prior to the last examination will be accepted.
Therefore, enter the current patient‘s height and weight in the appropriate
fields prior to selecting <Add new visit>.
Height and weight are required for the calculation of the patient‘s reference
values.
The patient details can also be edited any time. The blue fields can be modified
directly. The gray fields are write-protected. Click <Enable fields> to remove the
write protection.
What is a Visit?
Visit 3
Visit 2
Visit 1
1. General data
- Date of visit
- Time when the visit was created
- Biometric data of the patient stored to this visit
2. Levels
The type of visit is indicated by levels.
- One or several performed measurement(s) is/are assigned to each level.
- A visit can consist of several levels.
- Each level is assigned the time the level was created.
The following levels are available:
- Test = Measurements without any further definition
- Pre = Measurements before bronchodilation
- Post = Measurements after bronchodilation
- Base = Base provocation measurement (option)
- Prov = Provocation measurement (option)
- Dilat. = Bronchodilation measurement during a provocation test
(option)
3. Measurements
The performed measurements are assigned to each level.
Spiro = Measurements in the Spirometry program
- SVC = Slow Spirometry
- FVC = Forced Spirometry or Flow-Volume curve
- MVV = Maximum Voluntary Ventilation MVV
Rocc = Measurements in the R Occlusion program
- Rocc = Airway resistance measured with the shutter pressure
method R Occ
4. Visit status
The following icons indicate the visit status:
red means: Patient data and a new visit have been created,
but no measurement has been performed so far.
yellow means: At least one measurement has been performed.
gray means: The visit is complete and at least one preliminary report
has been saved.
wine-red means: The visit is complete and at least one preliminary report
has been saved. In the measurement program in the tab
Comments/Technician “Test Review” has been activated.
blue means: The visit is complete and at least one preliminary report plus
a doctor‘s comment or interpretation has been saved.
green means: The visit is complete and at least one final report has been
saved.
green/red means: The visit is complete and at least one corrected report has
been saved. (Only final reports can be corrected).
Measurement Programs
SentrySuite is capable of performing different types of measurements.
Depending on the system configuration, different programs are displayed on the
screen.
- Slow spirometry
- Forced spirometry (flow-volume measurement)
- MVV measurement
- R Occlusion
- Bodyplethysmography
- Bronchial Challenge
- ...
For Vyntus BODY with USS module, the use of the MicroGard filter is
mandatory.
Slow Spirometry
This measurement records a volume versus time curve. All diagnostically important
parameters of Slow Spirometry can be determined.
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
As soon as the system detects stable tidal breathing after a minimum of five
breaths, two vertical lines will automatically be set in the lower spirogram window
indicating the space to be filled by the breathing maneuver. In addition, the end
expiratory volume will be marked by a horizontal dashed line.
Recordings to the left of the first vertical line are used to determine the slow
spirometry parameters VT and BF.
Breathing for the slow vital capacity maneuver is limited to the space between
the vertical lines (30 s).
If the vertical lines are not displayed as the patient‘s tidal breathing is not stable
enough, it is also possible to mark the range for the maximum maneuver manually
with <F2> “Switch”.
Depending on the system setting, the slow vital capacity can be performed and
recorded as an ERV maneuver (see following screenshot) or as an IC maneuver.
ERV maneuver
Instruct the patient to slowly exhale to a maximum (ERV maneuver) until a
constant level is reached. Then the patient should slowly inhale (VC IN) and slowly
exhale to a maximum (VC EX). According to the ATS Guidelines the end-expiratory
level to be expected will be reached after 6 seconds.
After few normal breaths the breathing maneuver is complete.
If the two bars are within the green range, the expiratory maneuver
has been performed correctly, which is additionally indicated by the
green checkmark as an “End of test“ criteria for the forced exhalation.
In the right middle window, the patient is given textual instructions for performing
the test. These instructions change depending on the type of maneuver.
Below these diagrams the measurement is displayed online as a spirogram.
Forced Spirometry
Forced Spirometry is one of the basic lung function tests. With various breathing
maneuvers during which the patient exhales as much and as fast as possible, the
ventilated lung volumes and the breathing flows are measured.
Results are represented in a flow-volume and a volume-time curve.
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
Please note: During the whole examination the patient must not leave the
mouthpiece.
If the two bars are within the green range, the expiratory maneuver
has been performed correctly, which is additionally indicated by the
green checkmark as an “End of test“ criteria for the forced exhalation.
In the right middle window, the patient is given textual instructions for performing
the test. These instructions change depending on the type of maneuver.
Below these diagrams the measurement is displayed online as a spirogram
(volume-time curve).
* Literature:
- MR Miller et al. Series “ATS/ERS Task Force: Standardisation of Lung Function
Testing“, Standardisation of Spirometry, Eur Respir J 2005; 319-338
Copyright © ERS Journals Ltd. 2005
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
The left part of the chart section shows the recording of the flow-volume curve.
The right part shows a lung animation of inhalation and exhalation and two
bar indicators showing the variation of the tidal volume (VT) and the breathing
frequency (BF).
In the bottom part of the chart section the patient is given textual instructions
for performing the test. These instructions change depending on the type of
maneuver.
Below these diagrams the measurement is displayed online as spirogram. The two
vertical lines mark the MVV measuring range (12 s).
If the patient hyperventilates for at least 6 seconds, the parameter MVV6 will be
available.
Screen display after the measurement:
Bodyplethysmography
The measurement program “Bodyplethysmography“ consists of three
measurement phases which allow a comprehensive analysis of the respiratory
mechanics of the lung.
In the first phase, important specific resistances (sReff, sRtot, ...) are determined
from the registered “breathing loops“. In the following second phase, the lung
volume (FRCpleth) is determined during an occlusion maneuver. The airway
resistances (Reff, Rtot, ...) will then automatically be calculated from those
parameters. If the FRCpleth occlusion measurement is “linked“ to a maximal
maneuver, important absolute lung volumes (RV, TLC) can be calculated as well.
The bodyplethysmography is usually be concluded with the third phase, in which
additional forced spirometry parameters (FEV1, FVC, FEFxx, ...) are determined.
Perform a Measurement
Measurement Sequence: Body Standard
sReff FRCpleth FRCpleth FRCpleth forced Spirometry forced Spirometry forced Spirometry
Start 1. Shutter maneuver 2. Shutter maneuver „Linked“ Maneuver 2. Forced Spirometry n. Forced Spirometry
Breathing loops „non-linked“ „non-linked“ FRCpleth + slow/forced Spirometry with door open
Note:
• The Breathing loops (sReff) should be recorded until they are reproducible. In this case, the spirogram
displays steady breathing.
• We recommend performing at least two “non-linked” FRCpleth measurements. The shutter is
activated with <F1> “Continue”.
• Now, any number of “linked maneuver” can be performed. We recommend to perform a FRCpleth
measurement combined with a forced spirometry maneuver. After the last volume determination
additional forced spirometry measurements can be performed.
• Switch to the measurement phase flow/volume curve with <F2> “Switch” and open the door.
• We recommend performing at least three forced spirometry measurements.
Ask the patient to take a seat in the box (but not to approach the mouthpiece,
yet). The intercom system is activated automatically.
Door opener Inform the patient about the button to open the door from inside in case of
emergency.
Ask the patient to adapt to the mouthpiece and close the nose with the nose
clip. Instruct the patient to breathe normally.
Once stable tidal breathing has been obtained, the earliest after four breaths, a
red dashed line representing FRC stability is drawn indicating the consistency of
the end-expiratory level.
Expiration Inspiration
The last five (adjustable via the settings menu) loops of the specific resistance
(sReff) are shown on the left side of the screen. The spirogram appears in the
lower third of the screen. The right part of the screen shows three bars (indicators)
for evaluating the trial as well as a lung model. From top to bottom, the bars
indicate the stability (CV%) of the respiratory loops (sRaw), the stability of the tidal
volume (dVT) and the breathing frequency (BF).
Best results are obtained when the three bars are all in the green zone as
indicated above!
As soon as reproducible loops are shown and the spirogram displays regular
breathing (ideally with FRC stability), click <F1> “Continue“ to continue with
the volume measurement in which FRCpleth will be determined by an occlusion
maneuver.
After activation of the volume measurement, the shutter is closed at the end
of expiration.
Ask the patient to continue breathing normally even after the shutter is
closed.
Pulling or pressing against the shutter must not be forceful!
Click <F1> „Continue“ to continue with the next closure (non-linked maneuver).
The two moving Stabilitrac bars should be in the green zone. A green
tick mark indicating „End of Test“ shows that the breathing maneuver
was performed correctly.
A red cross means criteria have not been met
If the previously performed FRCpleth measurements are not comparable yet, click
<F1> „Continue“ to start the next „linked“ maneuver.
Click <F7> to deactivate the door magnets. The door will open automatically.
first closure as
„non-linked“ maneuver
second forced third forced
second closure as spirometry spirometry
„non-linked“ maneuver
Result display:
The best trial of all trials is automatically calculated and displayed on the
screen. The best curve is always represented in black.
Diffusion SB Realtime
The measurement program “Diffusion Realtime“ allows the determination of the
diffusion capacity (DLCO-SB) and the alveolar volume (VA) using the single-breath
method (SB). It also allows the determination of additional parameters such as
the Krogh factor (KCO) and the absolute volumes (TLC-SB, FRC-SB, and RV-SB).
Realtime means that the concentration of the test gases is already recorded in the
mouth during the measurement because of quick gas analyzers.
Analyzer unit
FPV (Flow Path Valve) USS MicroGard
block Adapter for gas
sampling
MicroGard
filter
Demand valve
Housing
Clamp
Drying tube for gas
analysis
USS MicroGard
Adapter for gas
sampling
MicroGard
filter
Demand valve
Housing
Clamp
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
The patient needs to breathe normally until an even tidal breathing has been
obtained. Once FRC stability has been reached - no earlier than after 4 breaths
- a red dashed line will be drawn on the screen indicating consistency of the
endexpiratory level.
<F1> “Maneuver” should only be pressed once the patient is actually in the
deep expiration phase; however, <F1> “Maneuver” should not be pressed
too late in the maneuver as the system requires a reaction time to release the
measurement gas.
As soon as the residual volume (RV) has been reached after the maximal
exhalation, the patient is asked to inhale quickly to the maximum (VIN) until he or
she reaches his or her total lung capacity (TLC).
At that moment the flap valve is closed automatically.
The current ATS/ERS standards for the determination of the diffusion capac-
ity*1 require a rapid inhalation to TLC, where 85% of VIN must be inspired in
< 4.0 sec. and VIN is at least 90% of VCmax.
The horizontal lines plotted in the “Spirogram and Mouth pressure“ window (see
image below) have the following meaning:
continuous line indicates the inspiratory volume.
If a spirometry measurement has been performed
prior to the examination, the window shows the VIN
best value which was measured. If there is no VIN
value available during the examination, the predicted
value is displayed.
dashed line indicates the inspiratory volume at 90 %
(adjustable) of the patient related best value
According to the ERS/ATS 2017 standards*1 the patient should inhale at least
90 % of the VIN. Alternatively, a VIN ≥ 85% of VCMAX is acceptable if the
VA of the trial is within 200 ml or 5% (whichever is greater) of the largest VA
from another acceptable trial.
The right chart displays the remaining breath hold time. It gives the operator an
indication of how much longer the shutter occlusion will last.
Furthermore, the mouth pressure during the breath hold time is recorded. The
objective is to keep the mouth pressure in the green range, i.e. the patient should
be relaxed during the breath hold maneuver and should not perform inspiratory or
expiratory efforts against the shutter.
During the breath hold time the inhaled single-breath gas mixture is evenly
distributed throughout the lungs and the CO gas diffuses into the blood via
the alveolocapillary membrane.
Existing distribution disorders of the lungs influence the CO gas distribution in
the alveolar gas space.
After the preset breath holding time has expired, the shutter opens automatically.
The patient is required to exhale quickly and completely without hesitation.
In the case of very small vital capacities, the sample volume (SV) is rescaled
automatically.
Below the spirogram the displayed pressure channel (red) shows possible breathing
efforts of the patient against the shutter. If there is any higher expiratory or
inspiratory pressure, the diffusion measurement can be influenced. The patient
should be made aware to keep pressure as neutral as possible during the breath
hold. Afterwards the measurement should be repeated.
Discard volume (DV) and sample volume (SV) are adjusted according to the
setting. The analysis of the measured gas concentration takes place in the area
of the sample volume. The parameters are calculated fully automatically.
According to the ERS/ATS 2017 standards*1 ,VA can be calculated using the
mass balance equation on the complete exhalation of the tracer gas CH4. If
the setting “mass balance” is enabled, the alveolar volume (VA_EX) will be
calculated based on the exhaled CH4 (FACH4end).
If the setting for quality criterion is set to „ERS/ATS 2017“ the four minutes wait-
ing time can be skipped and the user can decide by themselves how long the
patient should wait between the trials. According the „ERS/ATS 2017“ standard
the SentrySuite software has an algorithm implemented to recognize if the patient
is enough washed out or not. The following warning message will appear earliest
with the second trial during the tidal breathing phase if the patients wasn‘t suf-
ficiently washed out.
Please observe the safety notes on CO toxicity in the enclosed Instructions for
Use.
Click <F4> “Edit” to edit or deactivate trials or to enter the value for hemoglobin.
For more detailed information please refer to the Online Manual.
Diffusion SB Intrabreath
The measurement program “Diffusion SB Intrabreath“ allows the determination
of the diffusion capacity (DLCO IB) and the alveolar volume (VA) using the
intrabreath method (IB). It also allows the determination of additional parameters
as the Krogh factor (KCO IB) and the absolute volumes (TLC-IB, FRC-IB, and RV).
Intrabreath means that the maximal inspiration is followed by a slow and even
expiration to the maximum. Thus, a breath-holding time is not necessary.
USS MicroGard
Adapter for gas
sampling
MicroGard
filter
Demand valve
USS module
FPV silicone valve
USS MicroGard
Adapter for gas
sampling
MicroGard
filter
Demand valve
USS module
FPV silicone valve
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
The patient has to breathe normally until he or she has reached an even tidal
breathing. As soon as there is a FRC stability, yet at the earliest after 4 breaths, a
red dashed line indicating the end expiratory volume appears.
<F1> “Maneuver” should only be pressed once the patient is actually in the
deep expiration phase; however, <F1> “Maneuver” should not be pressed
too late in the maneuver as the system requires a reaction time to release the
measurement gas.
As soon as the residual volume (RV) is reached after the maximal expiration,
the patient is required to inhale quickly and maximally (VIN) until he or she has
reached his or her total lung capacity (TLC).
The current ATS/ERS standards for the determination of the diffusion capac-
ity*1 require a quick inspiration which is completed in less than 4 seconds.
The horizontal (gray) lines plotted in the “Spirogram & Flow“ window have the
following meaning:
continuous line indicates the patient-related inspiratory vital capacity
If a spirometry measurement has been performed
prior to the examination, the window shows the VIN
value of the patient which was actually measured.
If there is no VIN value available during the
examination, the predicted value is displayed.
dashed line indicates the patient-related inspiratory vital capacity
at 85 % (adjustable).
After the maximal inspiration, ask the patient to expire slowly and evenly
(important!) to maximum expiration. Make sure that the flow during expira-
tion is between 0.3 L/s and 0.7 L/s.
The gray bar indicates the range in which the expiration flow should be
(0.3 - 0.7 L/s). It is the aim to reach an even expiration with a constant flow, i.e.
the red flow curve should run horizontally within the grey range.
The inhaled gas mixture is evenly distributed throughout the lungs and the
CO diffuses into the blood via the capillary membrane. Existing distribution
disorders of the lungs influence the CO gas distribution in the alveolar gas
space and therefore influence the result of the diffusion measurement.
The measured gas concentration of the sample volume is analyzed and the
results will be calculated automatically.
According to ATS 1995 and ATS/ERS 2005 regulations at least two trials should be
performed in order to ensure reproducibility.
Click <F4> “Edit” to edit or deactivate trials or to enter the value for hemoglobin.
For more detailed information please refer to the Online Manual.
Note: The shown green C2H2 curves to calculate cardiac output in the screen shots
are possible with the device Vyntus ONE only but are not part of the device Vyntus
BODY with diffusion.
R Occlusion (Option)
The measurement program “R Occlusion“ allows to perform an airway
resistance measurement according to the shutter or shutter pressure method. The
measurement only requires tidal breathing, therefore “R Occ“ can be determined
almost independently of patient cooperation.
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
In the bottom part of the chart section the patient is given textual instructions
for performing the test.
Below these diagrams the measurement is displayed online as a spirogram.
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
At the earliest after five breaths a red dashed line appears indicating an even tidal
breathing. As soon as the system has recognized an even tidal breathing, click on
<F1> “Start” to start the maneuver.
Please explain the maneuver to the patient:
Instruct the patient to inhale and exhale calmly and evenly.
At the start of an inspiration the shutter will automatically close and after 0.1
seconds the inspiratory mouth pressure “P0.1“ is measured. The shutter only
closes for a fraction of a second and does not hamper the patient‘s breathing.
In subsequent measurements the shutter is not set to close with each breath, but
irregularly (random number generator) in order to avoid reactions of the patient.
In order to obtain precise measurement results several occlusions should be
performed during one trial. At least 5 occlusions per trial are recommended and
preset.
Shutter
The left part of the operator area shows the recording of the pressure-time curve
and underneath the current trend graph. The right part shows a lung model giving
a graphic display of inspiration and expiration as well as two bars indicating the
regularity of breathing and the breathing frequency (BF).
Ideally the two bars are within the green range as shown above.
Expiration Inspiration
In the bottom part of this area the breathing maneuver is displayed as a volume-
time curve.
The parameters shown in the Best column represent the mean value
(median) of all performed trials.
The curve which comes closest to the mean value (median) is displayed in black.
MIP / MEP
The “MIP / MEP“ program allows the measurement of the maximal inspiratory
and expiratory mouth pressure. The measured pressure values are taken for
assessing respiratory muscle strength.
MIP Measurement
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
At the earliest after five breaths a red dashed line appears indicating an even tidal
breathing. As soon as the system has recognized an even tidal breathing, the
patient is instructed to exhale deeply.
Do not click <F1> “Continue” until it is visible that the patient is actually
exhaling deeply.
After a deep expiration, the patient is asked to breathe in as fast and strongly
as possible against the closed shutter and to maintain the pressure for at least
2 seconds.
For another trial ask the patient to continue to breathe normally until the red
dashed line appears indicating an even tidal breathing.
Now the maneuver can be repeated by clicking <F1> “Continue” while the
patient is breathing out deeply again.
The central screen displays the pressure-time curve on the left hand side and below
that window the trend graph of each trial of the MIP measurement.
The lower section displays the quality assessment.
The parameters shown in the Best column represent the highest measured values
from all trials.
All in all, 10 trials can be performed and saved. If an 11th trial is performed,
the trial with the lowermost value is overwritten.
MEP Measurement
The measurement can either be called up directly after the MIP measurement or by
selecting <MEP>.
Just before starting the measurement, the patient should close his/her nose
with the nose-clip, take the mouthpiece between his/her teeth and keep his/
her lips tightly around the mouthpiece. Check the correct position of the
mouthpiece!
At the earliest after five breaths a red dashed line appears indicating an even tidal
breathing. As soon as the system has recognized an even tidal breathing, the
patient is instructed to inhale deeply.
Do not click <F1> “Continue” until it is visible that the patient is actually
inhaling deeply.
After a deep inspiration, the patient is asked to breathe out as fast and
strongly as possible against the closed shutter and to maintain the pressure for
at least 2 seconds.
For another trial ask the patient to continue to breathe normally until the red
dashed line appears indicating an even tidal breathing. Now the maneuver can be
repeated by clicking <F1> “Continue” while the patient is breathing in deeply
again.
The central screen displays the pressure-time curve on the right hand side and
below that window the trend graph of each trial of the MEP measurement.
The lower section displays the quality assessment.
The parameters shown in the Best column represent the highest measured values
from all trials.
All in all, 10 trials can be performed and saved. If an 11th trial is performed,
the trial with the lowermost value is overwritten.
SNIP (Option)
The “SNIP“ program (SNIP = Sniff Nasal Inspiratory Pressure) allows the non-
invasive measurement of the inspiratory muscle strength.
4. Make sure that the nose plug occludes the respective nostril pressure-tight. If
necessary, replace the nasal plug with another size to customize the fit.
Nasal plug
(single use item)
FPV Control Unit
Connection nipple
Pressure tube
Just before starting the measurement, the patient should breathe normally
with his/her mouth closed. Check the correct position of the nasal plug!
As soon as the first snip maneuver is recognized, the countdown timer starts to
count down 30 seconds. It is recommended* to observe a resting period of 15-30
seconds until the next trial is started via <F1> “Start“. Usually, one maneuver per
trial is sufficient.
Further maneuvers can be performed within a trial at any time and as often as
required. With each recording of a snip maneuver the countdown timer is reset to
its initial value of 30 s.
Ideally, the grey bar is within the green range as shown below.
Remaining
waiting time
between each
sniff
Countdown
Timer
currently
measured
value in %
of predicted
value
previous best
value of the
current trial in
% of predict-
ed value
100 % Pred
Pressure/Time
curve of all
maneuvers of
the current
trial
In order to ensure high measuring accuracy, at least 5-7 trials should be recorded.
Optimal collaboration is indicated by a plateau of the recent measurements’ SNIP
peak values.
The best maneuver of the first trial is automatically displayed as a colored bullet in
the middle window on the left.
All in all, 10 trials can be saved. If further trials are performed, the trial with the
so far lowermost value will be replaced.
Compliance (Option)
This program measures dynamic and/or static compliance from the esophagus
pressure-volume curve. The compliance is measured in the esophagus because this
pressure is approximately the same as the thoracic (pleural) pressure. Compliance
is the reciprocal value for the elasticity of the lung tissue. The measurement of the
dynamic compliance takes place during tidal breathing (not dependent on patient
cooperation). The measurement of the static compliance can be carried out over
the whole range of vital capacity.
The catheter can be used for children older than 10 years and for adults.
2. Connect the blue adapter to the adapter of the catheter with a rotary
movement.
“open” position
4. In order to remove the air, put two thirds of the catheter into a jar filled with
water. As soon as the air is removed, set the 3-Way-Stopcock in the „closed“
position.
“closed” position
Pneumotach
Esophagus catheter
Lung
Heart
Esophagus balloon
Diaphragm
Of course, other methods can be applied as well. The method described above
is only one of many examples.
“open” position
Pressure transducer
„Connection nipple“
Esophagus catheter
FPV Control Unit
Catheter Positioning
Just before starting the measurement, close off the patient‘s nose with the
nose-clip. The patient should take the mouthpiece between his/her teeth and
keep his/her lips tightly around the mouthpiece. Check the correct position of
the mouthpiece! In order to avoid shifting of the catheter towards the stomach,
secure the catheter to prevent it from moving.
Screen display:
The window „Spirogramm“ records the curves for volume and transpleural pres-
sure (Ptp) against the time. Both signals should be synchronous. If a signal shift is
noticed, the esophagus balloon is located in the stomach and needs to be pulled
back slightly. Short and regular pressure deflections of the Ptp signal indicate the
influence of the heart beat; usually, they can be reduced to a minimum by moving
the catheter a little deeper.
The window „Guidance & Stabilitrac“ shows the transpleural pressure. Ideally, the
bar should be in the green range.
If the transpleural pressure is below -0,5 kPa, a blue arrow down appears indicat-
ing that the catheter should be slightly moved downwards.
The upper left window shows the curves of the dynamic compliance in a volume/
pressure (-Ptp) display. The curves should be egg-shaped, slightly tilted and run
counterclockwise. The more they tile the smaller the dynamic compliance and the
less flexible are the lungs.
The upper left window shows the curves of volume and transpleural pressure (Ptp)
related to the time. Both signals should be synchronous. If this is not the case,
refer to chapter „Catheter Positioning“ for more information.
After a few breaths, the trial is ended automatically and is indicated in the
„Trend“ window as a colored bullet. At the same time, the next trial is started
automatically. The patient is asked to continue breathing slowly and evenly.
The window „Guidance & Stabilitrac“ shows the stability of the dynamic
compliance (d Cdyn) as well as the breathing frequency (BF). Ideally, the respective
bars should be in the green range.
In order to increase measuring accuracy, several trials should be successively
performed.
If the „Trend“ window shows a plateau of the measurement values, you can
switch to the measurement of the static compliance.
Click <F2> to switch to the static compliance measurement. At the same time,
the measurement is started automatically.
The upper left window shows the curve of the static compliance online in a vol-
ume/pressure display. The curve should be „S“-shaped and slightly tilted. The more
they tile, the smaller the static compliance and the less flexible are the lungs.
The lower window shows the curves for volume and transpleural pressure (Ptp)
against the time. Both signals should be synchronous. If this is not the case, refer
to chapter „Catheter Positioning“ for more information.
Click <F1> to end the first trial. The next trial is started automatically.
Result screen:
Best = Best value, determined from the median of all valid trials
The left graphic window displays the dynamic The right graphic window displays the static
compliance curves of all trials overlayed. The compliance curves of all trials overlayed. The
best trial is displayed as a black curve. best trial is displayed as a black curve.
The trend is displayed below. The trend is displayed below.
Rhinomanometry (Option)
The measurement program „Rhinomanometry“ allows the measurement and
evaluation of the ventilatory function of the nose.
USS Module
MicroGard II C Filter
Cone adapter
FPV Controll Unit
Connection nipple
Face mask
tube adapter
Nasal plug
Cone adapter
Tube adapter
Hose to connect on the
FPV Control Unit
Perform a measurement
As depicted, the patient’s right nostril is occluded using the nasal plug. The left
nostril will be measured.
Make sure that the nasal plug occludes the respective nostril pressure-tight but
does not deform it too much. If necessary, replace the nasal plug with another
size to customize the fit.
Ask the patient to place his/her hands around the face mask and to press it against
his/her face.
Make sure that the face mask fits tightly but does not deform the nose
and that the pressure hose located inside the mask is not bent. Ask the
patient to regularly breathe in and out through the open nostril 3 to 5
times per trial while keeping the mouth closed tightly.
The left upper window displaying the ”Flow/Pressure“ graph successively shows
the flow-pressure curve of those breaths performed within the trial as a grey curve.
The bottom window ”Spirogram“ displays the individual breaths against the time.
The “Guidance & Stabilitrac“ window on the right incorporates two bars. The
upper bar indicates the stability of the tidal volume VT. The lower bar indicates the
breathing frequency BF recorded during normal breathing. The animation shows
the correct position of the nasal plug as well. The side of the nostril which is not
occluded is measured.
Before starting another trial, the patient should briefly leave the mask
to avoid dyspnea. Breathing freely through the mouth creates the pre-
requisite for an optimal next trial.
Record further trials as described above. Each new trial is started via <F1> “Start”
and ended via <F1> “Present”.
The trend graph shows the trend information for flow or resistance depending on
the settings. The presentation of the inspiratory flow is recommended.
The measurement of the first side of the nose can be terminated if the test
results show a plateau.
The patient’s left nostril is occluded using the nasal plug. The right nostril will
be measured.
Click <F1> „Start“ to start the measurement and click <F1> “Present” to end
the trial. Perform several trials as described above.
Result screen:
The parameter table presents the predicted value calculated from the patient data
as well as the best values which are the final results of the test and the recorded
values of all valid trials.
Best = Best value, determined from the median of all valid trials
The left graphic window displays the The right window shows the flow
flow-pressure graphs of all trials (left and and resistance trend of all valid
right side of the nose) superimposed. The trials (left and right side of the
respective best trial for the left and for the nose).
right nostril is displayed as a black curve.
left nostril
right nostril
It is recommended to use the preset level “Pre“ for the first measurement of a
visit.
Quality Assessment/Interpretation/Comments
In all measurement programs there are different tabs in the lower part of the
screen:
- Guidance
- Quality
- Auto Interpretation
- Interpretation/Comments
Quality Assessment
When clicking the “Quality“ tab, the quality assessment result according to
the recommendations of “ATS and ERS 2005“ is indicated for the performed
measurement.
Auto Interpretation
When clicking the “Auto Interpretation“ tab, the automatic interpretation is
displayed as a text on the left hand side of the screen.
Clicking the down arrow opens a list of interpretation modules. The measurement
program saves the selected module as the standard module for the next
examination.
P 0.1, MIP / MEP and SNIP classification is based on P 0.1**, MIP / MEP** and SNIP** as follows:
severe = P 0.1 or MIP / MEP or SNIP ≤ -4 Z-Scores or ≥ 4 Z-Scores
moderate severe = P 0.1 or MIP / MEP or SNIP -4 to -3 Z-Scores or 3 to 4 Z-Scores
moderate = P 0.1 or MIP / MEP or SNIP -3 to -2 Z-Scores or 2 to 3 Z-Scores
mild = P 0.1 or MIP / MEP or SNIP -2 to -1.64 Z-Scores or 1.64 to 2 Z-Scores
normal = P 0.1 or MIP / MEP or SNIP - 1.64 to 1.64 Z-Scores
Rhinomanometry classification is based on the logarithmized effective resistance Log10Reff (LReff L/R) of
the 4-phase rhinomanometry according to K. Vogt**** :
Class Color Nasal Resistance log. resistance area
1 green Normal < 0,7
2 green/yellow/ Slightly increased ≥ 0,7 and < 0,9
3 yellow Moderately increased ≥ 0,9 and < 1,1
4 red Severe restriction ≥ 1,1 and < 1,4
5 dark red Very severe restriction ≥ 1,4
Obstruction classification is based on the Z-Score Raw (Best Airway Resistance - depending on
the setting Reff, Rtot…) for patients ≤ 18 years and based on the absolute limit values of Raw
for patients > 18 years:
For patients ≤ 18 years**
Normal = Z-Score Raw < 1.64
Mild = 2 > Z-Score Raw ≥ 1.64
Moderate = 3 > Z-Score Raw ≥ 2
Severe = 4 > Z-Score Raw ≥ 3
Very severe = Z-Score Raw ≥ 4
Interpretation/Comments
The “Interpretations/Comments“ tab enables the user to edit comments or
findings. Prior to entering a comment or findings, select (Technician) or (Physician)
by clicking the corresponding tab. This selection regulates where the comment is
registered. Now text can be entered manually or by using the templates which can
be configured by the user.
If the visit is complete and at least one preliminary report has been saved, “Test
Review” is displayed.
If “Test Review” is activated, the visit has the status “Test Review”.
Reference:
* Pellegrino R. et. al. Interpretative strategies for lung function tests; Series “ATS/ERS Task Force: Standardisation of lung
function testing“ Eur Respir J 2005; 26:948-968
** M Horstmann et. al. Transfer factor for carbon monoxide; European Respiratory Monograph 31, April 2005; 10: 127-145
*** Criée CP et al. Empfehlungen zur Ganzkörperplethysmographie. Empfehlungen der Atemwegsliga und der DGP. Dustri-
Verlag Dr. Karl Feistle 2009; ISBN 3-87185-394-1.
**** K. Vogt et al. Four-phase rhinomanometry: a multicentric retrospective analysis of 36,563 clinical measurements. Rhinology
2015: Published online.
Bronchoprovocation Test
Assessment of bronchial hyperreactivity in terms of increased airway
responsiveness to various stimuli plays an important role in the diagnostics of
dyspnea at rest and during physical exercise, chronic cough, asthma and for
medical reports.
Possible stimuli are specific (allergens) and non-specific pharmacological stimuli
(methacholine, histamine, carbachol, hypertonic salt solutions), but also indirect
stimuli such as isocapnic hyperventilation, physical exercise or cold air.
The degree of bronchial reactivity or hyperreactivity is quantified by means of
protocols applying the mentioned stimuli, combined with one or various lung
function tests, which provide a so-called dose-effect curve (response graph).
Use the Sidestream nebulizer.
FEV1
[%]
PD/PC-20 FEV1
FEV1 “PD20”
100
An inhalational bronchoprovocation test
is regarded to be positive if, for example,
90
FEV1 is reduced by 20 % as compared to
the initial value.
80 PD/PC-20
PD20
PD= Provocation Dose
70 PC= Provocation Concentration
60
Bronchial response tests are reserved for more in-depth diagnostic issues
regarding the presence of or the therapy effect on bronchial reactivity or
hyperreactivity.
Preventive Measures
Generally, standardized provocation tests with pharmacological substances
and tests in lung function labs require the same preventive measures. The staff
should be familiar with the treatment of airway obstructions and asthmatic
attacks. That means, oxygen and inhalational ß2-sympathicomimetica can be
administered without any delay, if required. A doctor which is experienced in
emergency treatment should also be available at any time.
At no time during the examination must the patient be unattended. The
patient should be informed to interrupt the provocation test immediately
in case of dyspnea . After the test the patient should only leave the lab if
the obstruction - documented by a lung function test - was reversible either
spontaneously or after bronchospasmolysis (FEV1 > 90 % of the reference
value before provocation is desirable). It is recommended to continuously
supervise children‘s gas exchange by transcutaneous pO2 and pCO2
measurement and/or measurement of the O2 saturation (pulse oximetry).
- ATS Guidelines for Methacholine and Exercise Challenge Testing - 1999. Am J Respir Crit Care
Med, 2000; Vol 161, 309 - 329
- ERS Airway Responsiveness
P.J. Sterk et al. Standardized challenge testing with pharmacological, physical and sensitizing
stimuli in adults.
Eur Respir J. 1993, 6, Suppl. 16, 53-63
APS Unit
Nebulization occurs solely during inspiration and the biggest portion of the
substance is taken up by the patient.
The in-line aerosol filter with a high absorption capacity prevents further
contamination of the environment.
If the APS unit is used regularly, a compressed air check must be performed
weekly by the operator (see chapter “Compressed air check“).
It must further be ensured that the applied jet nebulizers are new or not used
for more than half a year.
Preparation
P1 P2 P3 P4
3. Attach the nebulizer vessel containing the medication for the first provocation
step.
4. Connect compressed air tube to nebulizer vessel.
5. Connect new, disinfected mouthpiece and new aerosol filter.
Aerosol filter
Jet nebulizer
Special caution is advised during the filling and the handling of the nebulizer
vessel. Avoid the spillage of liquids!
Performing Bronchoprovocation
In the following, a bronchoprovocation test will be described.
Settings*:
Protocol: _1-conc_4-step_M-Aid
Observation module: _Spir_FEV1
Observation measurement: Spirometry
Before starting provocation make sure that the correct patient has been loaded.
If the wrong patient is displayed, click <Search> to load a patient already stored
in the SentrySuite database. If the patient has not been saved yet, click <New> to
store a new patient.
Corresponding to the set protocol, the first protocol level B1 is a reference test, i.e.
the measurement is performed without nebulization.
If, for example, the actual values for FEV1 are less than 70 % of the predicted
value, this is a contraindication. A provocation test should not be performed!
The predicted value check relates to the parameter limits set in the selected
observation module.
The threshold values are set in the generation program.
The observation parameters of the measurement programs Spirometry,
R Occlusion and Bodyplethysmography may be considered for assessment of a
contraindication.
Contraindication
The threshold values generated in an observation module aid you in deciding
whether or not a provocation test should be performed in the case of limited
reference values. Example: FEV1 < 70 % of the predicted value
If the reference test yields an FEV1 value of less than 70 % of the predicted
value, a corresponding warning will be displayed for this contraindication after
clicking <Return to Bronchial Challenge>.
Now instruct the patient to approach the mouthpiece and to breathe via the
APS unit. Click <OK> to start nebulization.
Dose administration is marked by a red bar. Number of nebulizations performed and yet to be performed
The patient should slowly breathe in and out as evenly as possible. His
breathing curve should be within the shown auxiliary lines (usually less
than 0.5 L/s). Therefore, it is easier for the patient when he can observe his
breathing pattern on screen.
Please note:
If a nebulizing breath is not performed correctly, i.e. the patient stops
inspiration before the nebulization per breath is terminated, the flow
observation system of the APS recognizes the problem, stops nebulization and
waits for your decision whether nebulization is to be continued or repeated.
Choose “Retry the breath“ to ignore the already administered dose and to
repeat the last breath.
If the programmed dose was administered completely, the APS compressor will
switch off automatically.
Nebulization can be aborted at any time during the nebulization phase by clicking
<F3>.
As soon as the complete dose has been administered, the APS compressor
switches off automatically.
After the exposure time has expired, click <F5> to switch to the observation
measurement in the set measurement program, in our example “Spirometry“.
At this point we relinquish the description of the provocation step P2 with the
corresponding control measurement and we proceed with the provocation
step P3.
After the exposure time has expired, click <F5> to switch to the observation
measurement in the set measurement program, in our example “Spirometry“.
Message:
The results of the provocation measurement after the third provocation step P3
have been entered automatically.
Result:
An inhalational bronchial provocation test is to be regarded as positive if FEV1 is
reduced by 20 % as compared to the initial value.
If the calculation of a threshold dose (PD/PC) for each of the selected parameters
(in our example for FEV1) is generated in the observation module, these values will
be displayed in the response table.
PD = Provocation dose
PC = Provocation concentration
Please note:
The limit values as well as the threshold values can be generated freely and are
part of the observation module.
Dilatation
Click <F1> to activate the next level, in our example level “D“ = Dilatation.
On the screen you are asked to administer the dose defined in the protocol:
For our example, 2 puffs of Sultanol which are not administered via the APS but
through a spray.
Administer the substance and click <OK> to confirm.
After the exposure time has expired, click <F5> to switch to the observation
measurement in the set measurement program, in our example “Spirometry“.
Only those measurement applications are switched active and are selectable to
which observation parameters have been assigned.
In our example <F5> spirometry and <F8> impulse oscillometry are active. As it is
recommended to perform examinations at rest prior to forced maneuvers, impulse
oscillometry is examined first which is started with <F8>.
After returning from impulse oscillometry into the bronchial challenge program,
the desired oscillometric observation parameters are already registered in the
parameter table, however, the values of the measurement application forced
spirometry are still missing. Then start and perform the missing measurement
application, in our example spirometry by clicking <F5>.
If the setting “Allow to skip observation measurements” is activated then
within the provocation steps the provocation program will query if the second
observation measurement should be proceeded or skipped.
If the observation parameters have been determined completely, the subsequent
provocation step suggested by the system is permitted to remain active.
Nebulization is then triggered via <F1>.
rubber plug
3. Connect (empty) nebulizer vessel and close the opening at the patient‘s
side with a rubber plug.
4. Start the check by clicking <OK>.
Report
After a measurement, the results can be reported by being displayed on the
screen, outputted to a printer, stored at any location as PDF, TIF, or JPG, or
allocated to the corresponding visit in an individually definable report*1.
Click on one of the tabs available in every measurement program.
The tabs indicate the output type and the report names assigned to the individual
tabs.
The following output types are available:
A Preview PDF of the report is displayed on screen and can then be printed out
and/or saved in the database to the corresponding visit with the status Preliminary,
Final or Corrected.
A Full View of the Report is displayed as PDF on screen and further offers the
following options:
- select a different visit
- save report with the status Preliminary, Final or Corrected
- print out report
- save report as a PDF file at any location
- save report as a TIF, JPEG or RTF file at any location (optional)
- enter a comment on the measurement
- send report to the Hospital Information System
The report is saved to the database to the corresponding visit with the status
Preliminary, Final or Corrected.
The following output types are preset:
- Quick View
- Save
Click <...> to select different reports. It is further possible to freely define the
3 tabs.
Calibration is not repeatable! - AirLife filter from Vyntus APS is perhaps dirty
- pump volume does not correspond to the set volume
- enter XX mL for the pump volume
Volume drift during occlusion too large. - let the patient leave the mouthpiece - initiate zero adjust-
ment - repeat measurement
- test the system for leaks
Zero adjustment failed. The patient must not yet approach the mouthpiece.
Repeat zero adjustment.
No measurement possible. Enter patient data.
No Cal. gas connected or pressure too low. Open the gas cylinder or replace the empty gas cylinder.
Calibration gas not stable.
Incorrect calibration value(s), calibration stop- Please check the settings and the measurement device and
ped. repeat the calibration.
Compare the specifications on the cylinder (certificate of
analysis) with the specifications in the setting of the calibration
program and make the respective corrections.
Clean or replace the sample line.
Replace the sample line with a new or a dry sample line.
As the setup for the data exchange with Vyaire JLab and Vmax systems via the
connectivity software typically strongly depends on the local laboratory and IT
environment please contact Vyaire tech support for further advice.
Online Manual
Before dispatching your Vyntus system, we have already installed an exact Online
Manual on the hard disk of the delivered computer. This manual can be loaded
from the start screen or any SentrySuite application program.
To launch the Online Manual, click “?“ in the menu bar, then User manual and
then Table of contents.
The “Acrobat Reader“ will be loaded automatically, displaying the “Online
Manual“ on the screen.
This Instructions for Use is part of the instrument, and should always be kept
on hand.
The instruction manual describes the present state of the device/system including
software and accessories with regard to the fundamental requirements of the
MDD 93/42/EC.
Exact adherence to the instructions issued is a prerequisite for proper and intended
functioning of Vyaire instruments.
Electrical Safety
As far as Protection Category I equipment/systems are concerned, their
touchable and maybe voltage-carrying metal parts are electrically conductively
connected to each other and connected to the grounding cable of the mains
power supply. A circuit breaker will be tripped should the insulation fail.
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
the floor.
- Equipment (for example vacuum cleaner, radio...) and devices which are not
part of the medical device/system must not be connected to the multi-plug
socket.
- Additional multi-plug sockets or extension cables must not be connected to the
ME system.
EMC Notes
This Vyaire device meets the regulations according to IEC 60601-1-2 regarding
the interference radiated and received. The device should not be installed in the
vicinity of high-frequency devices, X-ray equipment, motors or transformers with
high installed power rating since electric or magnetic interference fields may falsify
the result of measurements or make taking measurements impossible. Due to this,
the vicinity of power lines is to be avoided as well.
Changes or modifications to the Vyntus devices may result in increased emissions
or decreased immunity of the Vyntus devices in relation to EMC performance.
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Vyntus devices, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
WARNING: Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
WARNING: This equipment has been tested for radiated RF immunity only at
selected frequencies, and use nearby of emitters at other frequencies could result
in improper operation.
Keep a distance of about 2 meters from possible error sources when using the
device.
WARNING: Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
The Vyntus Devices have essential performance (see “Technical Data”). This can
be verified during the verification and calibration process. In the unlikely event of a
Fast Transient / ESD event occurring, the device should be reset and located away
from the source of interference.
Ambient Conditions
The medical device must not be operated in rooms or in the presence of
flammable anesthetic mixture with air or flammable anesthetic mixture with
oxygen or nitrous oxide. The device is not to be used in Oxygen enriched
environments (>25 Vol% O2).If gases and pharmacological substances (aerosol)
are used, a good ventilation is necessary.
The medical device must be operated in rooms where only non-conductive
pollution occurs; however, occasional temporary conductivity due to condensation
is to be expected. The medical device is designed for operation in medically used
rooms.
The medical device has to be effectively protected against moisture. Ventilation
slots must be kept free of obstructions in order to enable air circulation.
Medical Supervision
A qualified physician has to reassess all Vyntus /SentrySuite measurements. An
interpretation by Vyntus/SentrySuite is only significant by confirmation of other
clinical findings.
The operator is responsible for the change of predicted modules, authors or the
entry of new predicted value formulas.
Contraindications
The patient‘s case history has to be matched with the contraindications possibly
existing for the individual measurement methods in order to minimize the risk to
the patient.
In reference to humidity and water that may get inside the units, the Vyaire
devices correspond to the safety degree IP 20. This means, the device can be
cleaned with a damp (but not soaking) cloth which does not produce lint. More
detailed information can be found under “Hygiene“.
Chemicals required for operation or care of the unit must always be stored, pre-
pared, and made available in specially marked vessels to prevent any accidental
use.
Maintenance
Information regarding regularly exchanged parts is found in the Instructions for
Cleaning and Disinfection. Do not service or maintain any part of the medical
device while it is in use with a patient. Use only Vyaire approved accessories and
spare parts for this medical device. Service and Maintenance of the medical device
may not be performed during intended use. If applied parts (e.g. USS module)
have been exposed to extreme mechanical stress, a function test (e.g. volume
verification) has to be performed. If function is lost, the defective part is to be
replaced. Damaged parts, e.g. frayed plugs, receptacles, pressure regulators, a
damaged USS module, and defective cables should be replaced immediately by an
authorized specialist or engineers of the Vyaire Global Customer Support or,
if necessary, be put out of service after notifying the Global Customer Support. If
it is opened without authorization the guarantee entitlement expires. The Vyaire
Global Customer Support is always at your disposal with help and assistance
in case of problems. The system needs to be switched off and disconnected from
the power supply network. Make sure that the system can not be switched on
accidentally.
All connections and protective measures that have been disconnected or changed
during the maintenance phase must be rebuild to the condition that the safety
of the medical device / system is guaranteed on reactivation. (see installation
instructions)
For all medical devices by Vyaire the primary reference values will be detected,
documented and attached prior to dispatch according to IEC 60601-1. Only units
which have been maintained orderly and within regular periods of time are safe.
Maintenance and inspection can be carried out by the Vyaire Global Customer
Support, also as part of a maintenance contract.
Technical documentation, schematic circuit diagrams, stock lists, and descriptions
are available on request at the Vyaire Global Customer Support. An authorized
organization must receive training and a certification by the Vyaire Global
Customer Support.
All serious incidents which are associated with the product should be reported to
the manufacturer and the responsible authority where the user and/or the patient
is located.
Recurrent Test
For the secure operation of medical electrical equipment IEC 62353 (Medical
electrical equipment - Recurrent test and test after repair of medical electrical
equipment) is valid in principle. IEC 62353 defines that a recurrent test has to
be performed at regular intervals and after a repair of the medical electrical
equipment. Put the device out of service and notify the Global Customer Support
if aberrations beyond the limit values occur.
Vyaire specifies an interval of 3 years.
The calibration syringe itself has to be re-calibrated at regular intervals. For further
information see enclosed certificate.
In case of aberrations contact Vyaire Global Customer Support for assistance.
Perform system verification according section “Calibration / Verification
Frequency”.
Recycling
Observe the WEEE directive and follow all Federal, State and Local regulations or
contact Vyaire Global Customer Support for assistance when disposing the medical
device, its accessories or batteries.
Patient environment:
A patient environment of 1.5 meters to the Vyntus CART 3.b has to be kept.
That means the patient has to keep a distance of at least 1.5 meters from all
open interfaces, connectors and the Vyntus CART 3.b to avoid any contact with
electrical voltage.
Brakes:
- Set the four brakes during operation.
- Release the four brakes for the transport.
Patient environment:
If a Vyntus CART 3.0N / 3.1N is used in combination with a Vyntus BODY (without
a Vyntus APS) an additional potential equalization conductor has to be installed
to the potential equalization between the single ME devices or to the potential
equalization of the building wiring installation in order to prevent differences in
the electrical potential.
In this combination a patient environment of 1.5 meters to the Vyntus CART 3.0N
/ 3.1N is not necessary, nevertheless, the Vyntus BODY is powered by a separate
wall socket.
Brakes:
- Set the four brakes during operation.
- Release the four brakes for the transport.
Keyboard
Clamping elements
Clamping device
Use the height adjustment activator of the swivel chair to move into the correct
seating position of the patient.
The patient must be supervised at all times during the measurement procedure
and must never be unattended.
Door opener
The bench in the box can be loaded up to a maximum of 250 kg (551 LBS).
APS pressure
connection APS Pressure: 1,4 bar
SB (RT) IN
SB (RT) IN supply CO: 3000 ppm
USB PC
connection
Potential equalization
connection
The door and all the panes on the Vyntus BODY are made of tempered glass.
Thermal treatment makes tempered glass especially stable. If it breaks, it crumbles
into small granular chunks to effectively minimize injuries from cuts and impact.
Please note:
- Keep the pivot range of the door completely free of objects it can poten-
tially hit
- Keep the door from striking any hard objects, e.g.: chair legs and bases,
office chair rollers, stones, etc.
- Use only soft, lint-free cloths for cleaning – do not use anything that
scratches, e.g., scrapers, scrubbing sponges, and the like.
Front side: LED status Handling of the “Extended Basic Module“ for Vyntus APS
- status LEDs (description see “LED status table“)
- Disconnect the system while manipulating the interfaces (disconnection or
connection).
- Please note the color coding and coding during the handling of the interfaces.
connection for
future options
Rear side:
connection
pressure hose connection APS head
APS head
USB-PC
connection
= LED is illuminated
= flashes slowly
= flashes quickly
Esophagus Catheter
Please observe the manufacturer‘s instructions for the esophagus catheter (item
no. 705003).
Graphical Symbols
Alternating current
Earthed conductor
Potential equalization
Attention!
Year of Production
Manufacturer
Use before
CAUTION:
Rx ONLY FEDERAL U.S. LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A
PHYSICIAN.
Serial number
Reference number
Batch code
Direct current
Medical device
Please observe all federal, state and local regulations for safe handlin of gases.
it from tipping).
- Place the full cylinder in the cylinder rack.
- Remove the cylinder cap and the lock nut/sealing cap, if applicable.
- Follow the instruction leaflet for exchanging the sealing ring of the pressure
reducer. (Instruction leaflets and sealing rings are delivered with every gas cylin-
der delivered by Vyaire).
- Check the connection for tightness as follows:
- Briefly open the cylinder valve and close it again.
- After 5 minutes open the cylinder valve again while watching the high-pressure
manometer display. If the manometer clearly indicates a pressure reduction,
there must be a leakage.
- The high-pressure manometer display should indicate a constant pressure.
- However, a pressure reduction of approx. 30 % within 12 hours is considered
as normal.
- After tightness of all gas connectors has been checked, open the cylinder valve.
Please note: Slowly open the cylinder valve as far as possible. (The system is now
under pressure again). Turn back the cylinder valve by a 1/4 revolution of the
hand-wheel (thus preventing the valve from getting stuck).
Notes on Using CO
Trade name: Carbon Monoxide
Chemical characteristics:
CO, carbon monoxide, very potent haemotoxin, combustible, highly
flammable, gaseous, colorless, odorless. It is fatal to inhale large
quantities of carbon monoxide. Carbon monoxide may possibly
damage the unborn fetus.
Toxicology:
Very potent haemotoxin. During inspiration the gas combines with hemoglobin so
that oxygen cannot be absorbed (suffocation).
Toxin class 3 DL50 = 50 to 500 mg/kg
Literature:
- Notes on using CO
Hommel: Handbuch der gefährlichen Güter, Sheet 116
Kühn-Birett: Merkblätter gefährlicher Arbeitsstoffe, Sheet K19
Before any maintenance activity at the system the gas bottle valve
must be closed. Then disconnect the pressure reducer from the cylinder
valve.
Access to the gas cylinders must be ensured at all times and the
fittings of the pressure reducers must be visible.
Literature
IEC 60601-1: 2012 Medical Electrical Equipment,
Part 1: General requirements for basic safety and essential performance
IEC 62353:2014 Medical electrical equipment - Recurrent test and test after repair
of medical electrical equipment
Reference source:
VDE-Verlag GmbH, Bismarckstr. 33, 10625 Berlin or Merianstrasse 29, 63069 Of-
fenbach, Germany, e-mail: kundenservice@vde-verlag.de
BEUTH Verlag GmbH, Burggrafenstrasse 6, 10787 Berlin, e-mail: info@beuth.de
Technical Data
Product Lifetime:
If the complete product is used and maintained in the condition it was delivered, the expected
product lifetime is at least 7 years.
Flow:
Vyntus BODY
Type Ultrasound
Methodological Simultaneous measurement of ultrasound transit time in and against flow direction
Peculiarities
Sample rate True 1000 Hz for flow, achieved by 2000 ultrasound transit time measurements (double
shot technology)
Range 0 to 18 L/s bidirectional
Accuracy Exhalation 0 to 14 L/s:
1.5 % or 0.05 L/s (whichever is greater)
Inhalation 0 to 14 L/s:
2.0 % or 0.05 L/s (whichever is greater)
Precision 1 % or 0.1 L/s (whichever is greater)
Resolution 1 mL/s
Total resistance <0.150 kPa*s/L at 14 L/s
(MicroGard II filter +
USS Module + FPV
block)
Dead space USS 66 mL
Module
Volume:
Vyntus BODY
Type Software volume integration of flow signal (ultrasound)
Range 30 L (software limited)
Accuracy Exhalation and inhalation 0.5 to 14 L/s: 2.5% or 0.075 L (75 ml)
Precision 1 % or 0.05 L (whichever is greater)
Resolution 1 mL
Mouth Pressure:
Vyntus BODY
Type piezo resistive
Range ± 20 kPa
Accuracy ± 1 % or 0.01 kPa (0.075mmHg) (whichever is greater)
Resolution 0.003 kPa (0.225 mmHg)
Pressure reducer (Diffusion pressure reducer for demand valve - Vyntus BODY
with Diffusion):
Inlet pressure 200 bar
Outlet pressure 7 bar (fixed)
Flow rate at least 500 L/min
at most 700 L/min
2 outlets G 1/4 internal screw thread
Power Supply:
Vyntus BODY (cabin)
Mains voltage 100 to 240 V, AC 50 to 60 Hz
Power input max. 80 VA
Electrical safety Protection class I
Mains plug Used for isolating all poles simultaneously from supply mains
Power Supply:
Vyntus APS,
Vyntus Power Module
Type Magic Power MPM-X125
Mains input voltage 100-240 V, AC 47-63 Hz
Power consumption 1.5 - 1 A
Output voltage 24 V DC
Output 120 VA / 5 A
Electrical safety Protection class I
Mains plug Used for isolating all poles simultaneously from supply mains
Moisture protection:
Vyntus APS IP 20
Vyntus CART 3.b no IP protection
Vyntus CART 3.0N, 3.1N no IP protection
Vyntus BODY IP 20
Ambient conditions:
Altitude: ≤ 3000 m
Temperature: +10 °C to +34 °C (+50 °F to 93.2 °F)
Rel. humidity: 20 to 80 % RH, non-condensing
Ambient pressure: 700 to 1060 hPa (525 to 795 mmHg)
Operating mode:
MicroGard II Series:
v yaire.co m
© 2 0 2 1 V y a i re M e d i c a l , I n c. o r o n e o f i ts a f f i l i a tes . A l l r i g h ts res e r v e d . V y a i re, t h e Vy a i re M e d i c a l l o g o a n d a l l o t h e r
t ra d e m a r ks a re t ra d e m a r ks o r re g i s te re d t ra d e m a r ks of Vy a i re M e d i c a l , I n c. o r o n e of i ts a f f i l i a tes . M e d i c a l d ev i c e s
c l a ss I I a a c c o rd i n g to M e d i c a l D e v i c e s D i re c t i v e 93 / 4 2 / E E C. P l e a s e re a d t h e c o m p l e te I n s t r u c t i o n s F o r U s e t h a t
c o m e w i t h t h e d e v i c e s o r fo l l o w t h e i n s t r u c t i o n s o n t h e p ro d u c t l a b e l i n g .