lutech Engineering Pvt.Ltd.

CC-2891

Certificate of Calibration Page 1 Off1


Certificate No: FEPL/C/ 2022 /0848 Date Calibrated: 08-09-2022
Lab Work ID: FEPL/220483 *Recommended Cal Due Date: 08-09-2023

ULRNO:CC289122000010702F Date ofIssue: 08-09-2022


Field of Parameter: Dimensions

Name &Address of Customer Customer Reference:Through RK Technologies


Harpitha Engineering8 Reference DC NO : 0134
Sy.No129,Shed No.34/A Near Bhatru Hotel. Reference No Date : 08-09-2022
Date of Receipt :08-09-2022
BMTC 9th Depot,4th phase,
: Satisfactory
Peenya Industrial Area,Bangalore-560058. Status ofitem on Receipt
Device Under Calibration Details
Name of The DUC :Digital Inclinometer |Least count of the Device :0.1°
Make/Model Freemans Operating Range :0-9 0
ID No./Serial No. :2022-15-FM1 Units of Measurement : Degrees

Range 0-90 Place of Calibration : In-House (Dimension Lab)


& IS :6231-1971 (By comparing readings of
Calibration Procedure:_ Done as per FEPL Ref: LCP No: LCP-M-D-023 in accordance with IS:4239:1970
indicated by Reference Standard with the Device under calibration
Reference Standard Used Standard1
Angle Gauge Block Set
Nomenclature
: KCP
Manufacturer
Model/serial No : 10202
Traceability/Certificate No : KCP/38/20-21/2580

Calibration Valid Upto : 14-09-2022


20.7 °C HumidityY 50.2
Environmentalcondition: Temp
Calibration Results
Bevel Protractor Reading in degrees Average Deviation
Angle Gauge
Reading in degrees Trial 1 Trial 2 Trial 3 Trial 4 in degrees
0.0
in degrees
0.0
0.0 0.0 0.0
.0 1.0 1.0 1.0 1.0 0.0
3.0 0.0
3.0 .0 3.0
.0 5.0 5.0 5.0 5.0 0.0
0.0
15.0 15.0 15.0 15.0 15.0
30.0 30.0 30.0
30.0 30.0 0.0
0.0
45 45.0 45.0 45.0 45.0 45.0
60.0 60.0 60.0 60.0 0.0
60 60.0 89.9 0.1
90 89.9 89.9 89.9 89.9
95%. Appoximately with coverage
The Estmated Uncertainty of measurements associated with
the results is 1.38 min calculated at a confidence level

factorK=2
Calibrated By
Authorised ignatory

KN Kamath/P Mobandas/$saná
Prabath Shetty (CEO/QM/ TM).
(Calibration Engineer)
Remarks: Scan the QR code for NABL Accreditation Certificate
and scope of our Laboratory.

*Therecalibration interval is determined by User based on the usage and conditions of operation.
National/International standards as requlred in ISO/IEC 17025(E)
Reference Standards used are traceable to
All in
calibrated & shall not be reproduced except without the written
full,
*This Certificate refers only to the item
-560043.
approval ofthe FEPL Bangalore item submitted for calibration and valid at
the time of and under the stated
Results reported refers only to this
conditions of measurement.
and not for legal purpose or for advertisement.
The certificate is intended only for guidance
Calibration Certificate.
Corrections/Erasing will invalidate the

Certificate is issued only for scientific &Industrial purpose only


ANEEA
8 o f Certifiea

Bangalore

Flutech Eneering Pvt.Ltd.


Kalyan Nagar, Bangalore- 560 043. Ph.: +91 96329 34165, 97402 76354, 73381 83749
# 431/11, Glass Factory Road, Chelikere,
Regd. Office:# 57/A,5th Main. Kodihalli,
1st Floor, Bangalore -560 008 E-mail: info@flutech.in web: www.flutech.in
U29253KA201OPTCO54098
GST:29AABCF5836B1Zzs
CIN No.:

You might also like