Assessment of suport surfaces (static & mixed) for neonates

Conference Paper · May 2014


DOI: 10.6084/m9.figshare.977866

CITATION READS

1 75

7 authors, including:

Montse Sanchez Pablo García-Molina


CONSELLERIA DE SANITAT University of Valencia
14 PUBLICATIONS   17 CITATIONS    51 PUBLICATIONS   225 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Proyecto de investigación sobre redistribución de presión en SEMP de sedestación y SEMP neonatales View project

All content following this page was uploaded by Carlos Vergara-Hernández on 05 February 2015.

The user has requested enhancement of the downloaded file.


ASSESSMENT OF SUPPORT
SURFACES (STATIC & MIXED)

FP-50002 EWMA/GNEAUPP 2014


FOR NEONATES
M. M. Sánchez-Lorente1; P. García-Molina2-3; C. Vergara-Hernández2;
E. Sanchis-Sánchez2; J. Martos-Torres2; A. Álvarez-Ordiales4;
& E. Balaguer-López2-3

1. Conselleria de Sanitat, Valencia (SPA).


2. University of Valencia, Valencia (SPA).
3. Hospital Clínico Universitario, Valencia (SPA).
4. Hospital Universitari i Politècnic “La Fe”, Valencia (SPA).

Presenting author: Mr. Pablo García-Molina


Background

FP-50002 EWMA/GNEAUPP 2014


Existence of several neonatal support surfaces

however…

Lack of evaluation
regarding…

 Interface pressure & pressure redistribution


 Efficacy in the prevention or treatment of PU
Aim 5
2

FP-50002 EWMA/GNEAUPP 2014


1

• To assess the interface pressure and pressure redistribution


of five devices of constant low pressure (static and mixed
support surfaces).
3
4
Methods

FP-50002 EWMA/GNEAUPP 2014


Study design
• Laboratory study with repeated measures design.

Sample
• Eight neonatal simulation dummies with different characteristics
regarding head circumference, weight and height.
Methods

FP-50002 EWMA/GNEAUPP 2014


Support devices
Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D
Viscoelastic Viscoelastic Viscoelastic Viscoelastic Polyester
Structure
“T” units Block Block Block filaments

Layers 1 2 2 2 1

Density (Kg/m3) 50 50/50 50/50 40 NA

Hardness (Kpa) 1.6 1.01/5 1.01/3.6 2.5 NA

Compression factor 2 2/2 2/2 - NA


Methods

FP-50002 EWMA/GNEAUPP 2014


Variables
• Mean pressure (mmHg).
• Maximum pressure (mmHg).
• Mean pressure in head region (mmHg).
• Area of contact (cm2).

Equipment
• XSENSOR X3 PX100:40.40.002 device.

Procedure
• See: E-poster no. 50006.
Methods

FP-50002 EWMA/GNEAUPP 2014


Statistical analysis
• Descriptive statistics.

• Due to a small sample size, a non-parametric test (Friedman test) was carried
out to assess differences among groups. If a significant difference was found,
post hoc comparisons among groups were performed using the Wilcoxon-
Nemenyi-McDonald-Thompson test.

• The analysis was carried out by means of the R environment (The R


Foundation for Statistical Computing).

• Data set and code employed for the analysis can be found at
www.figshare.com (*).
*http://dx.doi.org/10.6084/m9.figshare.977866
Results
Lowest values!

FP-50002 EWMA/GNEAUPP 2014


Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Mean pres., x̅ (SD) 16.26 (2.77) 17.51 (3.32) 18.12 (3.32) 17.11 (3.07) 17.82 (2.58)

Max. pres., x̅ (SD) 24.58 (7.39) 26.79 (8.88) 28.15 (7.91) 25.49 (7.55) 27.53 (6.12)

Head pres., x̅ (SD) 17.25 (3.37) 18.58 (4.17) 19.53 (4.45) 17.95 (3.71) 18.7 (2.92)

Area, x̅ (SD) 19.29 (11.29) 22.11 (12.13) 22 (9.75) 23.52 (13.21) 29.23 (12.46)

Highest values!
Results

FP-50002 EWMA/GNEAUPP 2014


Friedman test χ2 Df P-value

Mean pres., x̅ (SD) 14,3 4 0,006

Max. pres., x̅ (SD) 11,2 4 0,02

Head pres., x̅ (SD) 8,7 4 0,07

Area, x̅ (SD) 18,51 4 < 0,001


Results

FP-50002 EWMA/GNEAUPP 2014


Post hoc analysis: mean pressure (p-value)

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,24 - - - -

Visc-Gray 0,005 0,61 - - -

Visc-Blue 0,71 0,93 0,17 - -

Pol-3D 0,056 0,97 0,93 0,61 -


Results
Results

FP-50002 EWMA/GNEAUPP 2014


Post hoc analysis: maximum pressure (p-value)

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,32 - - - -

Visc-Gray 0,014 0,71 - - -

Visc-Blue 0,88 0,88 0,17 - -

Pol-3D 0,32 1 0,71 0,88 -


Results
Results

FP-50002 EWMA/GNEAUPP 2014


Post hoc analysis: pressure area (p-value)

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,71 - - - -

Visc-Gray 0,95 0,98 - - -

Visc-Blue 0,27 0,95 0,71 - -

Pol-3D < 0,001 0,053 0,01 0,27 -


Results
Conclusion

The “T” device presents the best values of pressure


relief.
The polyester fiber device presents the best values of
pressure redistribution.

Adding an upper layer of polyester fiber to the “T” device


could result in a better support surface.
Next step: studying the properties of the new support
surface in healthy subjects under clinical conditions.