4 CVP

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

CENTRAL VENOUS PRESSURE

NCM 118: CRITICAL CARE NURSING (SKILLS)


BS Nursing | SEM 1 2022

CENTRAL VENOUS PRESSURE PROCEDURES


(Action & Rationale)
- Measure of the pressure within
the right atrium of the heart. It 1. Wash hands.
measures the ability of the right
side of the heart to deal with the Reduces transmission of microorganism
systemic fluid load
- Low CVP is an indicator of 2. Wears gloves and wear mask
hypovolemia
- High CVP can be caused by Practice aseptic technique
hyperkalemia or poor cardiac
function 3. Gather equipment needed

● Sterile gloves
CVP CAN BE MEASURED ● Manometer set/transducer
1. A transducer ● Marking pen indelible ink
2. Use of manometer attached to the ● Mask
intravenous fluid ● Normal saline
● Tape
● IV tubing
● Stopcock
PHLEBOSTATIC AXIS
- Located at the right atrium at the Maximizes efficiency and minimizing the
mid axillary line about one-third of chance of breaking sterility once started
the distance from the anterior to
the posterior chest wall in the 4. Position the client a supine or flat
position if this is not tolerated and
fourth intercostal space
the client is placed in semi-fowler's
positions.
Take all measurements at the same
SPECIAL CONSIDERATION angle. Mark the right atrium with an
“X” using indelible ink pen
❖ CVP reading: The patient must (phlebostatic axis) to identify the
always be positioned in the same level of the atrium
place, relative to the right atrium
when measurement is performed The manometer should always be zeroed
❖ Patient lie in supine position without at the “X”: to minimize variance in
a pillow measurements.
❖ Position should be documented to
ensure that all nurses caring for the 5. Connect the IV fluid to a three-way
patient perform measurement in the stopcock and flush the other two
same manner parts. (Normal Saline)

Forces air out of the stopcock. Fluids with


glucose are sticker and should be avoided
COMPLICATION
- Infection 6. Connect the CVP manometer to the
- Air embolism upper port of the stopcock

Inserts the CVP manometer into the central

1
CENTRAL VENOUS PRESSURE
NCM 118: CRITICAL CARE NURSING (SKILLS)
BS Nursing | SEM 1 2022

line entering the fluid column or the


client and to prevent contamination
7. Connect the CVP tubing from the of the manometer
client to the second side port of the
stopcock The top of the manometer is open to the
air. If the manometer is not properly stored,
Establish IV line from normal saline to CVP contaminations or air can enter the
catheter manometer and be flushed into the client.

8. Allow normal saline to drip rapidly 14. Wash hands


into client for few seconds, with
stopcock close to manometer Prevent the spread of microorganism

Establishes that the CVP line is patent. 15. Document reading


Fluids must flow freely for reading to be
accurate Provides continuity of care

9. Turn stopcock off to client and fill


manometer with normal saline to
the 20-cm mark above the
anticipated reading

The normal CVP reading varies from 5 to


12 cm of water

10. Hold manometer at the phlebostatic


axis and turn the stopcock off to the
normal saline

System is open from the manometer to the


patient

11. Watch as the fluid falls in the


manometer. Take the CVP reading
when the fluid stabilizes

The fluid stabilizes at a level equal to the


pressure in the central veins or right atrium.
If the fluid level fluctuates with the client’s
respirations, take the reading at the end of
client’s respirations

12. Turn the stopcock off to the


manometer

Re-establishes fluid flow from the IV to the


client

13. Store the manometer in an upright


position(usually hanging from the IV
pole) to prevent air bubbles from

You might also like