Professional Documents
Culture Documents
2022-06-14 Analytical Sensitivity Study Omicron Variant BA.4 - BA.5
2022-06-14 Analytical Sensitivity Study Omicron Variant BA.4 - BA.5
2022-06-14 Analytical Sensitivity Study Omicron Variant BA.4 - BA.5
The purpose of this study was to validate whether the assay performance will be affected by SARS-
CoV-2 variant of concern, lineage BA.4 and BA.5.
2. Reagent
Toda Coronadiag Ag®
Lot 1: I2111176; Lot 2: I2111178; Lot 3: I2111179.
3. Specimen
a. Positive controls used in the study
Recombinant SARS-CoV-2 nucleocapsid protein (NCP)
4. Study Design
a. LOD screening
1) 10-fold dilutions of each positive control were made in assay buffer.
2) Apply 100uL of each dilution onto the sample well of test device.
5) The concentration demonstrating 9 of 9 positives was selected for LOD ranging finding.
2) Apply 100uL of each dilution onto the sample well of test device.
c. LOD confirmation
1) A total of 20 replicates of the selected concentration based on LOD range finding were tested with
three kit lots as described above, to obtain a total of 60 testing results.
2) The LOD was confirmed when 95% of testing results were all positive.
5. Results
Page 2/4
Version 1.0: Analytical Sensitivity study for SARS-CoV-2 Omicron Variant
2.9×101 + + + + + + + + + 9/9
2.9 - - - - - - - - - 0/9
The SARS-CoV-2 variant BA.5 isolate
Page 3/4
Version 1.0: Analytical Sensitivity study for SARS-CoV-2 Omicron Variant
6. Conclusion
The study demonstrated that the assay performance would not be affected by testing SARS-CoV-2
variant BA.4 and BA.5.
Based on the above testing results, the limit of detection of the Toda Coronadiag Ag® for SARS-CoV-2
variant BA.4 was 29pg/ml and for the SARS-CoV-2 variant BA.5 was 3.68pg/ml. And the testing results
were met the criteria. Note that the results are also applicable for the self-test TODA® Autotest Nasal
COVID-19.
Approval:
Page 4/4