Organo Phosphonate Test Method
Organo Phosphonate Test Method
Phosphonate PP 316
0.2 - 125 mg/l PO4
Persulfate UV Oxidation Method
Material
Required material (partly optional):
Application List
• Cooling Water
Preperation
1. All glassware must first be rinsed with diluted Hydrochloric acid (1:1) and then
rinsed with deionised water. Do not use detergents with phosphates.
Notes
1. During UV digestion Phosphonates are converted to ortho-Phosphates. This step
is normally completed in 10 minutes. Organic highly-loaded samples or a weak UV
lamp can cause incomplete phosphate conversion to take place.
2. UV lamp available on request.
NLEN
Handboek
Handbook
vanofMethoden
Methods 01/19
Phosphonate PP / 316
3. For handling of the UV lamp see manufacturer‘s manual. Do not touch the surface
of the UV lamp. Fingerprints will erode the glass. Wipe the UV lamp with a soft and
clean cloth between measurements.
4. The reagent Vario Phosphate Rgt. F10 is not completely dissolved.
5. The given reaction time of 2 minutes refers to a sample temperature of more than
15 °C. At a sample temperature lower than 15 °C, a reaction time of 4 minutes is
required.
NL Handboek
EN Handbook van
of Methods
Methoden01/19
01/19
Phosphonate PP / 316
Digestion
Select the appropriate volume of sample according to the following table:
0 - 2.5 50 0.1
0 - 5.0 25 0.2
0 - 12.5 10 0.5
0 - 25 5 1.0
0 - 125 1 5.0
25 ml
With the selected sample Fill one of the digestion vials Add Vario Potassium Per-
volume fill a 50 ml measu- with 25 ml of prepared sulfate F10 powder pack.
ring cylinder. If necessary, sample.
fill up with demineralised
water to 50 ml and mix.
NLEN
Handboek
Handbook
vanofMethoden
Methods 01/19
Phosphonate PP / 316
Close digestion vial Swirl around to dissolve the Keep the UV lamp in the
powder. sample. Note: wear UV
safety goggles!
10 min
12
11 1
10 2
9 3
8 4
7 5
6
Turn on the UV lamp. Wait for 10 minute(s) reac- The UV lamp is switched
tion time. off when the countdown is
finished.
NL Handboek
EN Handbook van
of Methods
Methoden01/19
01/19
Phosphonate PP / 316
10 ml 10 ml
Blank Sample
Blank
Prepare two clean 24 mm Fill blank with 10 ml prepa- Fill sample vial with 10
vials. Mark one as a blank. red, not digested sample. ml prepared, digested
sample.
Blank Sample
2 min
Zero
12
11 1
10 2
9 3
8 4
7 5
6
Blank
Place blank in the sample Press the ZERO button. Wait for 2 minute(s) reac-
chamber. • Pay attention to tion time.
the positioning.
Once the reaction period is finished, the measurement takes place automatically.
NLEN
Handboek
Handbook
vanofMethoden
Methods 01/19
Phosphonate PP / 316
Test
Sample
Remove the vial from the Place sample vial in the Press the TEST (XD:
sample chamber. sample chamber. • Pay START) button.
attention to the positioning.
The result in mg/l PO43- appears on the display.
NL Handboek
EN Handbook van
of Methods
Methoden01/19
01/19
Phosphonate PP / 316
Analyses
The following table identifies the output values can be converted into other citation
forms.
Chemical Method
Persulfate UV Oxidation Method
Appendix
Interferences
Interference from / [mg/l] Influence
Aluminium (from 100 mg/l) 1000
Arsenic in all
concentrations
Benzotriazoles 10
HCO3- 1000
Br-
100
Ca 5000
CDTA 100
Cl- 5000
CrO4 2-
100
Cu 100
CN- 100
Diethanoldithiocarbamate 50
EDTA 100
Fe 200
NO3- 200
NTA 250
PO43- 15
NLEN
Handboek
Handbook
vanofMethoden
Methods 01/19
Phosphonate PP / 316
a)
determination of free, combined and total | b) Reactor is necessary for COD (150 °C), TOC (120 °C) and total
-chromium, - phosphate, -nitrogen, (100 °C) | c) MultiDirect: Adapter is necessary for Vacu-vials® (Order code 19 20
75) | d) Spectroquant® is a Merck KGaA Trademark | e) alternative reagent, used instead of DPD No.1/No.3 in case of
turbidity in the water sample caused by high concentration of calcium and/or high conductivity | f) additionally required
for determination of bromine, chlorine dioxide and ozone in the presence of chlorine | g) Reagent recovers most inso-
luble iron oxides without digestion | h) additionally required for samples with hardness values above 300 mg/l CaCO3 |
i)
high range by dilution | # including stirring rod, 10 cm
NL Handboek
EN Handbook van
of Methods
Methoden01/19
01/19