Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Instrumentation 

& Measurement    _____________________________________________________   _            AAiT 

Chapter 3
Transducers and Their Response
3.1 Transducer

Instrument Society of America defines a sensor or transducer as a device which provides a usable
output in response to a specified measured. Here the measured is a physical quantity and the output
may be an electrical quantity, mechanical and- optical.

Sensor
An element that senses a variation in input energy to produce a variation in another or same
form of energy is called a sensor.

Transducer
Transducer converts a specified measured into usable output using transduction principle.
For example, a properly cut piezoelectric crystal can be called a sensor where as it becomes
a transducer with appropriate electrodes and input/output mechanisms attached to it. So,
the sensor is the primary element of a transducer.

Transducers is a devices used to transform one kind of energy to another. When a transducer
converts a measurable quantity (temperature, pressure, level, optical intensity, magnetic field, etc)
to an electrical voltage or an electrical current we call it a sensor. Energy information conversion
is the objective of a sensor. The information available in one energy form must be converted
into the same or another energy form, with exactly the same information content as the
originating energy form.

The sensor or the sensing element is the first element in a measuring system and takes information
about the variable being measured and transforms it into a more suitable form to be measured.

Sensor is sometimes called a primary measuring element, it can be found simply as a mercury
thermometer to measure the temperature. It may be embedded in the transducer to perform its
function. That means the transducer consists of a primary element (sensor) plus a secondary

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 1 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

element (signal conditioning circuit) that transforms the passive change or small voltage signal
into active signal range that can be easily used in other chains of the control loop.

A transducer can be defined as a device capable of converting energy from one form into another.
Transducers can be found both at the input as well as at the output stage of a measuring system.

The input transducer is called the sensor, because it senses the desired physical quantity and
converts it into another energy form.

The output transducer is called the actuator, because it converts the energy into a form to which
another independent system can react, whether it is a biological system or a technical system. So,
for a biological system the actuator can be a numerical display or a loudspeaker to which the visual
or aural senses react respectively. For a technical system the actuator could be a recorder or a laser,
producing holes in a ceramic material. The results can be interpreted by humans.

Actuators are important in instrumentation. They allow the use of feedback at the source of the
measurement. However we will pay little attention to them in this course. The study of using
actuators and feedback belongs to a course in Control theory.

3.2 Types of Energy Form

We can distinguish six different energy domains: (1) radiant, (2) mechanical, (3) thermal, (4)
electrical, (5) magnetic and (6) chemical.
If certain information is already available in the electrical domain it can be claimed that it
requires no energy conversion, but in general there is 'shape' conversion left and this is just the
domain which belongs to the field of electronics and electrical science and engineering.

A good example of such a sensor only sensitive to electrical energy is the probe of an
oscilloscope, with which a good adaptation to the signal source is realized. In the modifier
stage we meet other examples of shape converters, for instance the A/D and D/A converters.

Table 1.6 Energy types and corresponding measured


_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 2 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Energy Measurands
Length, area, volume, force, pressure, acceleration,
Mechanical torque, mass flow, acoustic intensity and so on.
Thermal Temperature, heat flow, entropy, state of matter
Electrical Charge, current, voltage, resistance, inductance,
capacitance, dielectric constant, polarization,
frequency, electric field, dipole moment, and so on.
Magnetic Field intensity, flux density, permeability, magnetic
moment, and so on.
Radiant Intensity, phase, refractive index, reflectance,
transmittance, absorbance, wavelength, polarization,
and so on.
Chemical Concentration, composition, oxidation/reduction
potential, Reaction rate, pH and the like.

3.3 Classification of transducers

The transducers may be classified based on


i. The physical effect employed
ii. The physical quantity measured
iii. The source of energy

i. Classification based on physical effect


The physical quantity applied as measurand (quantity to be measured) to the transducer causes
some physical changes in its element. By this physical effect the transducer converts the physical
quantity in to electrical quantity. For example, a change in temperature to be measured causes
variation of resistance (physical change) in a copper wire (element) and this effect could, be used
to convert temperature in to an electrical output,
The physical effects commonly employed are
a. Variation of resistance
b. Variation. of inductance
c. Variation of capacitance
d. Piezo electric effect
e. Magnetostrictive effect
_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 3 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

f. Elastic effect
g. Hall effect

a. Variation of resistance

The resistance of a length of metallic wire is given by

l
R
a
Where,
R = Resistance in ohm.
 =Resistivity (or specific resistance) of the material in ohm-m.
l = length of wire in m.
a = Area of cross-section in m2.

As resistance is a function of  , l , a (i.e) R  f (  , l , a ) , with any change in anyone of the


physical quantities  , a or l due to variation in resistance, a variable resistance transducer
can be designed to convert physical quantity.

Some of the transducers based on this principle are potentiometer, strain gauge, resistance
thermometer, carbon microphone, and photoconductive cell.
 The resistance thermometer is based upon thermo resistive effect which is the
change in electrical resistivity of a metal or semiconductor due to change in
temperature co-efficient of resistivity.
 Carbon microphone works on the principle of change in contact resistance due to
applied pressure.
 Photoconductive cell is based on photoconductive effect which is the change in
electrical conductivity due, to incident light.
 Potentiometer works on the principle of change in resistance due to linear or
rotational motion.
 Strain gauge works on the principle of change in resistance due to applied pressure.

b. Variation of inductance

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 4 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

The inductance of a coil is given by

d
LN
dt

N 2 o r A
L
l
where,
L = inductance in henry
N = No., of turns
o = absolute permeability
r = relative permeability
A = area of cross section of the core
l = length of magnetic path
d
= rate of change of magnetic flux.
dt

As L is a function, of N, r , A, l ,

(i.e) L = f (N, r ,A, l ), when anyone of these quantities changes, the inductance changes.

This leads to the design of a variable inductance transducer.

Some of the transducers based on variation of inductance are induction potentiometer,


linear variable differential transformer (LVDT) and synchros.

c. Variation of capacitance

Capacitive sensors consist of two parallel metal plates in which the dielectric between the
plates is either air or some other medium. The capacitance between two conductor plates
is given by

 o r A
C
d

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 5 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Where

C = capacitance in farad

 o = absolute permittivity

 r = relative permittivity of the separating medium

A = area of cross-section of the plates

As C is a function of  r , A, d i.e C = (  r , A, d ) when anyone of these quantities changes, the

capacitance varies. This leads to the design of a variable capacitance transducer.

Capacitive devices in which the distance between the plates is variable are primarily used as
displacement sensors. Motion of the moveable capacitive plate relative to a fixed one changes the
capacitance. Such devices can be used directly as a displacement sensor by applying the motion to
be measured to the moveable capacitor plate. Capacitive displacement sensors commonly form
part of instruments measuring pressure, sound, or acceleration.

In the alternative form of capacitor, the distance between the plates is fixed. Variation in
capacitance is achieved by changing the dielectric constant of the material between the plates in
some way. One application is where the dielectric medium is air and the device is used as a
humidity sensor by measuring the moisture content of the air. Another common application is as
a liquid level sensor, where the dielectric is part air and part liquid according to the level of the
liquid that the device is inserted in. This principle is used in devices to measure moisture content,
humidity values, and liquid level, as discussed in later chapters.

d. Piezoelectric effect

When a piezoelectric crystal like quartz or Rochelle salt is subjected to mechanical stress, an
electric charge is generated. This is known as piezoelectric effect. The transducer based on this
effect is piezoelectric transducer.

They can also operate in the reverse mode where an applied voltage produces an output force.
They are frequently used as ultrasonic transmitters and receivers. They are also used as
displacement transducers, particularly as part of devices measuring acceleration, force, and
pressure. In ultrasonic receivers, the sinusoidal amplitude variations in the ultrasound wave

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 6 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

received are translated into sinusoidal changes in the amplitude of the force applied to the
piezoelectric transducer. In a similar way, the translational movement in a displacement transducer
is caused by mechanical means to apply a force to the piezoelectric transducer.

Piezoelectric transducers are made from piezoelectric materials. These have an asymmetrical
lattice of molecules that distorts when a mechanical force is applied to it. This distortion causes a
reorientation of electric charges within the material, resulting in a relative displacement of positive
and negative charges. The charge displacement induces surface charges on the material of opposite
polarity between the two sides. By implanting electrodes into the surface of the material, these
surface charges can be measured as an output voltage. For a rectangular block of material, the
induced voltage is given by:

kFd
V
A
where F is the applied force in g, A is the area of the material in mm, d is the thickness of the
material, and k is the piezoelectric constant. The polarity of the induced voltage depends on
whether the material is compressed or stretched.

e. Magnetostrictive effect

When a magnetic material is subjected to mechanical stress, its permeability changes. This effect
is magnetostrictive effect and the transducer based on this effect is magnetostrictive transducer.

Magnetic sensors utilize the magnetic phenomena of inductance, reluctance, and eddy currents to
indicate the value of the measured quantity, which is usually some form of displacement.

f. Elastic effect

When an elastic member is subjected to mechanical stress it is deformed. The transducer based on
this effect is called elastic transducer.

g. Hall effect

Basically, a Hall effect sensor is a device that is used to measure the magnitude of a magnetic field.
It consists of a conductor carrying a current that is aligned orthogonally with the magnetic field,
as shown in Figure 3.1

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 7 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Figure 3.1 Principles of Hall effect sensor

This produces a transverse voltage difference across the device that is directly proportional to the
magnetic field strength. For an excitation current I and magnetic field strength B, the output voltage
is given by V = KIB, where K is known as the Hall constant.

The conductor in Hall effect sensors is usually made from a semiconductor material as opposed to
a metal, because a larger voltage output is produced for a magnetic field of a given size. In one
common use of the device as a proximity sensor, the magnetic field is provided by a permanent
magnet that is built into the device. The magnitude of this field changes when the device becomes
close to any ferrous metal object or boundary. The Hall effect is also commonly used in computer
keyboard push buttons. When a button is depressed, a magnet attached underneath the button
moves past a Hall effect sensor. This generates an induced voltage in the sensor which is converted
by a trigger circuit into a digital output. Such push-button switches can operate at high frequencies
without contact bounce.

ii. Classification based on physical quantity measured


The transducers may be classified based on the physical quantity they measure as follows:

 Temperature transducers  Transducers used to measure temperature


 Pressure transducers  To measure pressure
 Flow transducers  To measure flow
 Liquid level transducers  To measure liquid level
 Force/Torque transducers  To measure force & Torque
 Velocity/Speed transducers  To measure velocity & speed
_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 8 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

 Humidity transducers  To measure humidity


 Acceleration/vibration transducers  To measure acceleration & vibration
 Displacement transducers  To measure displacement

iii. Classification based on source of energy


Transducers may be, classified based on source of energy into two types.
• Active transducer
• Passive transducer

Passive transducer
A component whose output energy is supplied entirely or almost entirely by its input signal is
called a passive transducer. A passive transducer is the one which absorbs energy from the input
medium and converts it directly into the output signal.
Example
A Thermocouple extracts heat energy from the input medium and converts it into electrical
energy (voltage).
Active Transducer
An active transducer has an auxiliary source of power which supplies a major part of the output
power while the input signal supplies only an insignificant portion (i.e) this transducer uses the
energy it absorbs from the input medium as a control signal to transfer energy from the power
supply to produce a proportional output.
Example
Strain gauge
The energy extracted from the strained member is very small. The energy for the output
signal is supplied by an external power source.

Figure 1.1 Active and passive transducers


_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 9 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Selection of Transducers
Transducers are used for the measurement of physical quantities. The selection of transducers for
particular measurand is very important. The selection of transducers may be based on the following
factors for effective measurement.
1. The physical quantity to be measured (measurand),
2. The range of input quantity,

Based on physical quantity to be measured

The correct type of transducer should be selected for measuring the physical quantity. The
following table shows the physical quantity and the corresponding transducer types available.

Table 1.2 classification based on physical quantity


No. Physical quantity Transducers available
1 Temperature Bimetallic element
Fluid expansion systems
i. Liquid-in-steel bulb thermometers
ii. Liquid-in-glass thermometers
iii. Vapour pre-ssure thermometers
Thermoresistive elements
i. Resistance Temperature detector (RTD)
ii. Thermistor
Thermocouple
Linear-Quartz thermometer
Pyrometry
2 Pressure U-tube and ball type manometers
Ring balance manometer
Metallic diaphragms
Capsules and bellows
Bourdon tubes
Membranes
3 Force (weight) Spring balance
Cantilever
Diaphragms
Pneumatic and hydraulic load cells
Column and proving ring load cells
4 Torque Torsion bar
Flat spiral springs
Dynamometer
Gyroscope

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 10 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

5 Density of liquids Hydrometer


Air bubbler system
U-tube weighing system
6 Liquid level Float elements.
Manometer system
Diaphragms
Container weight
7 Viscosity Capillary tube
Concentric cylinder system
8 Flow rate of fluids Pitot static tube
Flow-obstruction elements
Rotating vane system
Rotameter float system
9 Displacement Flapper nozzle system
10 Absolute displacement, Seismic system
Velocity and acceleration
Vehicle attitude Gyroscope

3.4 Characteristics of Transducers

The selection of most suitable transducer from commercially available instruments is very
important in designing an Instrumentation system. For the proper selection of transducer,
knowledge of the performance characteristics of them are essential. The performance
characteristics can be classified into two namely
i. Static characteristics
ii. Dynamic characteristics
Static characteristics are a set of performance criteria that give a meaningful description of the
quality of measurement without becoming concerned with dynamic descriptions involving
differential equations.
Dynamic characteristics describe the quality of measurement when the measured quantities vary
rapidly with time. Here the dynamic relations between the instrument input and output must be
examined, generally by the use of differential equations.

For further reading on this subtopic please read the additional material

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 11 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

3.5 Calibration

Calibration consists of comparing the output of the instrument or sensor under test against the
output of an instrument of known accuracy when the same input (the measured quantity) is applied
to both instruments. This procedure is carried out for a range of inputs covering the whole
measurement range of the instrument or sensor.

Calibration ensures that the measuring accuracy of all instruments and sensors used in a
measurement system is known over the whole measurement range, provided that the calibrated
instruments and sensors are used in environmental conditions that are the same as those under
which they were calibrated. For use of instruments and sensors under different environmental
conditions, appropriate correction has to be made for the ensuing modifying inputs.

Instruments used as a standard in calibration procedures are usually chosen to be of greater inherent
accuracy than the process instruments that they are used to calibrate. Because such instruments are
only used for calibration purposes, greater accuracy can often be achieved by specifying a type of
instrument that would be unsuitable for normal process measurements. For instance, ruggedness
is not a requirement, and freedom from this constraint opens up a much wider range of possible
instruments. In practice, high accuracy, null-type instruments are very commonly used for
calibration duties, because the need for a human operator is not a problem in these circumstances.

Instrument calibration has to be repeated at prescribed intervals because the characteristics of any
instrument change over a period. Changes in instrument characteristics are brought about by such
factors as mechanical wear, and the effects of dirt, dust, fumes, chemicals, and temperature
changes in the operating environment. To a great extent, the magnitude of the drift in
characteristics depends on the amount of use an instrument receives and hence on the amount of
wear and the length of time that it is subjected to the operating environment. However, some drift
also occurs even in storage, as a result of aging effects in components within the instrument.

A proper course of action must be defined that describes the procedures to be followed when an
instrument is found to be out of calibration, that is when its output is different to that of the
calibration instrument when the same input is applied. The required action depends very much
upon the nature of the discrepancy and the type of instrument involved. In many cases, deviations

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 12 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

in the form of a simple output bias can be corrected by a small adjustment to the instrument
(following which the adjustment screws must be sealed to prevent tampering). In other cases, the
output scale of the instrument may have to be redrawn, or scaling factors altered where the
instrument output is part of some automatic control or inspection system. In extreme cases, where
the calibration procedure shows up signs of instrument damage, it may be necessary to send the
instrument for repair of even scrap it.

Whatever system and frequency of calibration is established; it is important to review this from
time to time to ensure that the system remains effective and efficient. It may happen that a cheaper
(but equally effective) method of calibration becomes available with the passage of time, and such
an alternative system must clearly be adopted in the interests of cost efficiency. However, the main
item under scrutiny in this review is normally whether the calibration interval is still appropriate.
Records of the calibration history of the instrument will be the primary basis on which this review
is made. It may happen that an instrument starts to go out of calibration more quickly after a period
of time, either because of aging factors within the instrument or because of changes in the operating
environment. The conditions or mode of usage of the instrument may also be subject to change.
As the environmental and usage conditions of an instrument may change beneficially as well as
adversely, there is the possibility that the recommended calibration interval may decrease as well
as increase.

3.5.1 Control of Calibration Environment

Any instrument that is used as a standard in calibration procedures must be kept solely for
calibration duties and must never be used for other purposes. Most particularly, it must not be
regarded as a spare instrument that can be used for process measurements if the instrument
normally used for that purpose breaks down. Proper provision for process instrument failures must
be made by keeping a spare set of process instruments. Standard calibration instruments must be
totally separate.

To ensure that these conditions are met, the calibration function must be managed and executed in
a professional manner. This will normally mean setting aside a particular place within the
Instrumentation Department of a Company where all calibration operations take place and where
all instruments used for calibration are kept. As far as possible this should take the form of a
_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 13 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

separate room, rather than a sectioned-off area in a room used for other purposes as well. This will
enable better environmental control to be applied in the calibration area and will also offer better
protection against unauthorized handling or use of the calibration instruments. The level of
environmental control required during calibration should be considered carefully with due regard
to what level of accuracy is required in the calibration procedure, but should not be over specified
as this will lead to unnecessary expense. Full air conditioning is not normally required for
calibration at this level, as it is very expensive, but sensible precautions should be taken to guard
the area from extremes of heat or cold, and also good standards of cleanliness should be
maintained.

While it is desirable that all calibration functions are performed in this carefully controlled
environment, it is not always practical to achieve this. Sometimes, it is not convenient or possible
to remove instruments from process plant, and in these cases, it is a standard practice to calibrate
them in situ. In these circumstances, appropriate corrections must be made for the deviation in the
calibration environmental conditions away from those specified. This practice does not obviate the
need to protect calibration instruments and maintain them in constant conditions in a calibration
laboratory at all times other than when they are involved in such calibration duties on plant.

As far as management of calibration procedures is concerned, it is important that the performance


of all calibration operations is assigned as the clear responsibility of just one person. That person
should have total control over the calibration function, and be able to limit access to the calibration
laboratory to designated, approved personnel only. Only by giving this appointed person total
control over the calibration function, can the function be expected to operate efficiently and
effectively. Lack of such definite management can only lead to unintentional neglect of the
calibration system, resulting in the use of equipment in an out-of-date state of calibration and
subsequent loss of traceability to reference standards. Professional management is essential so that
the customer can be assured that an efficient calibration system is in operation and that the accuracy
of measurements is guaranteed.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 14 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Calibration is an essential process to be undertaken for each instrument and measuring system
frequently. A reference standard at least ten times more accurate than the instrument under test is
normally used.
Calibration is the process where the test instrument (the instrument to be calibrated) is compared
with the standard instrument. It consists of reading the standard and test instruments
simultaneously when the input quantity is held constant at several values over the range of the test
instrument.
Generally, certification of an instrument manufactured by an industry is undertaken by the
National Physical Laboratory and other authorized laboratories where the secondary standards
and the working standards are kept.

3.5.2 Procedure for calibration


a) Examine the construction of the instrument, and identify and list all the possible
inputs.
b) Decide, which of the inputs will be significant in the application for which
the instrument is to be calibrated.
c) Select the apparatus that will allow you to vary all the significant inputs over
the ranges considered necessary. Select standards to measure each input.
d) By holding some inputs constant, varying others and recording the outputs develop
the desired static input-output relations.

3.6 Mathematical Models of System Response


3.6.1 Basic System Models
 Modelling is the process of representing the behaviour of a system by a collection of
mathematical equations & logics.
o Modelling is comprehensively utilized to study the response of any system.
o Response of a system is a measure of its fidelity to its purpose.
 Simulation is the process of solving the model and it performed on a computer.
 Equations are used to describe the relationship between the input and output of a system.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 15 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

 Analogy approach is also widely used to study system response

3.6.2 Mechanical System Elements


A) Spring

F=kx
 F = Force (tension or compression),
 x = Displacement (extension or compression),
 k = Spring constant. The bigger the value of k the greater the forces required to
stretch or compress the spring & so the greater the stiffness.

B) Dashpot

dx
F  cv  c
dt
 F = Force opposing the motion at velocity v,
 c = Damping coefficient. Larger the value of c the greater the damping force at a
particular velocity.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 16 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

C) Mass

dv d 2x
F  ma  m  m 2
dt dt
 F = Force required to cause acceleration, a,
 m = Mass of the element that is distributed throughout some volume. However, in
many cases, it is assumed to be concentrated at a point.

In general, the response of measurement instruments under dynamic conditions can be complex.
The fundamental concepts of dynamic response, however, can be understood by studying
relatively simple mathematical models.
We will consider three mathematical models for dynamic system response: zeroth, first and second
order systems.

3.6.3 Zeroth Order Systems

Imagine a thermometer that measures the temperature in a room of an office building.


For practical purposes, the thermometer will indicate the current temperature at the location where
it has been installed.
The fact that the output of the instrument follows the input “exactly” is the defining characteristic
of zeroth order systems.

Mathematically, if we let f(t) be the input to the system as a function of time and y(t) be the output,
then the relationship between them is

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 17 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

In the example above f(t) would be the actual temperature of the room and y(t) would be the
indicated temperature. K is a constant that multiplies the input to generate the output. If y(t) is the
temperature as displayed in a readout device and the thermometer is calibrated correctly, then K
would ideally be equal to one.

On the other hand, if the output of the thermometer is, for example, an electrical signal, then K
would be a constant with units of Volts per degree Fahrenheit.
The electrical signal could be used to actuate a valve that directs either cold or hot air from the air-
conditioning system to the room. K is usually called the static sensitivity.

Note that the output of zeroth order systems is not affected by the speed at which f(t) changes. The
above Equation is always valid, so the results of static calibration are sufficient to characterize the
response of the system.

3.6.4 First Order Systems

Let’s consider the example of an oral thermometer used at a clinic to measure body temperature
in patients. Prior to use, the thermometer is at room temperature. When the thermometer is put in
the patient’s mouth it experiences a sudden increase in temperature. Generally, we have to wait
for a while before reading the temperature. Unlike the case of a thermometer monitoring the
examination room’s temperature, the situation at hand cannot be represented by a zeroth order
model. Why?

Let’s consider a common glass bulb thermometer to explain. The thermometer in originally at
room temperature, which will be denoted by To.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 18 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Figure 1.2 Thermometer measuring body temperatures

It is then put in the mouth of a patient represented by the area inside the dashed line which is at
temperature T1. In order for the thermometer to work, the mercury in the bulb must be heated to
T1. The thermal expansion of the mercury will cause the column of mercury in the stem of the
thermometer to increase in length. Measuring this length with the scale marked in the glass gives
a temperature reading T(t).

It takes a while, however, for the temperature of the mercury to reach the value T1, so we must
wait that long before the thermometer indicates the correct temperature. In order to write an
equation to model the response of the thermometer we need a little background in thermodynamics
and heat transfer.

 Heat is a form of energy. We will represent it by Q. It flows from a hot place to a cold one.
The energy in the mercury in the bulb of the thermometer, which we will call E, increases
as heat travels from the mouth of the patient to the bulb.
 Due to conservation of energy, the rate of change of energy in the bulb with respect to time
is equal to how fast heat is flowing in. In mathematical terms:

 As E increases, the temperature of the mercury, T, rises in proportion. How fast the
temperature increases depends on how much mercury the bulb holds (the mass, m, of the
mercury) and a property of mercury called the ‘specific heat,’ cv.
_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 19 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

 The increase in energy is related to the increase in temperature by

Heat must travel through the glass walls of the bulb on its way from the patient’s mouth to the
mercury.

How fast heat can flow through the walls of the bulb depends on a property of glass called the
convection heat transfer coefficient, h, the surface area of the bulb, A, and the current temperature
difference between the mercury and the mouth of the patient. In equation form,

Rearranging the above two equation

This is a differential equation that governs what the temperature of the mercury is at any time.
Since the length of the column of mercury is proportional the temperature, the equation also
governs what the indicated temperature is.

In general, the equation of a first order system is given

For the example above, we have

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 20 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

As you can see, the equation governing the behavior of a first order system is a first order
differential equation, so called because the highest derivative of the output variable in the equation
is the first with respect to time.

3.6.5 Second Order Systems

When you go to the grocery store, note that weight scales are available at the produce department
so you can get an idea how many pounds of potatoes you will have to pay for. The scales
generally have a pointer, which indicates weight on a big dial.

Figure 1.3 Spring Balance

If you drop a bag with potatoes on the scale, chances are that the pointer will oscillate a bit before
settling and indicating the correct weight. The reason for the oscillation of the pointer is that the
scale has mass, hence inertia, and its behavior is dictated by Newton’s laws of motion.

You may now see where the second order label comes from if you recall that acceleration is the
second derivative of displacement ... Let’s analyze the balance and see what we get.

The balance can be represented by the mass-spring-dashpot system shown in Figure

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 21 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Figure 1.4 Free body diagram of spring balance

The balance can be represented by the mass-spring-dashpot system shown in Figure. To simplify
things a bit, the dial has been replaced by the straight scale at the left. We can write the equation
of motion for the balance using basic dynamics as follows:

 The mass m represents the object weighted so that the weight W is equal to mg
where g is the acceleration of gravity.
 The spring, attached near the left edge of the mass, is a mechanical element which
develops force in proportion to how much it has been stretched. Since the top of the
spring is fixed to the ceiling, the force is proportional to the displacement y of the
mass.
 If the zero value of y corresponds to the position of the spring when it is unloaded,
then the force Fs required to stretch the spring a distance y is given by

 where k is called the spring constant.

The balance needs a way to dampen the oscillations of the pointer after a weight is dropped. The
damping is provided by the dashpot, which is attached to the mass near the right. Dashpots are like
shock absorbers in cars.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 22 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

You can imagine one as a piston inside a closed cylinder. The cylinder is filled with a viscous
fluid. As the piston moves, a small gap between the piston and the cylinder allows the fluid to flow
from one side of the piston to the other.

A force is needed to move the piston due to the viscosity of the fluid. The net result is that dashpots
produce a force, Fd which is proportional to the speed of the piston relative to the cylinder.

This force-speed relation can be written as

 where c is called the damping coefficient.

The free body diagram of the mass is shown in Figure. Applying Newton’s second law we obtain
the equation

 which can be re-written as

This is the differential equation that governs the motion of the scale. Since the weight is indicated
by the displacement of the scale, the equation also governs the indicated weight.

In general, the equation for a second order system is given by


_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 23 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

Where

3.6.6 Solution to the Differential Equations

In order to see how these systems behave for a given input we need to solve them. The solutions
to the differential equations depend, of course, on how the input varies with time.

 Does it increase, decrease or cycle? Smoothly or abruptly?


 If smoothly, are the changes slow or fast?

Although an infinite number of possible inputs exist, we will consider only two, which are most
telling about the characteristics of the system response. These two inputs are: the step input and
the harmonic input.

We will not go over the method of solution of the differential equations, instead we will just quote
the results. If you know how to solve differential equations, it is reasonably straight forward to
obtain the solutions. If not, there is no need to learn now, just make sure you can use the solutions
presented below.

3.6.7 First Order Systems


3.6.7.1 Step input response

A step input is used to represent situations when the input changes from one, constant,
value to another constant value in a very short period of time. Using Laplace transform
for first order equation

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 24 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

 Note that at t = 0, y(t) = yo,


and that as

 The transition between the initial and the final values of y(t) is
exponential as shown in Figure.

Figure 1.5 Instrument responses for first order step input

 The parameter  is called the time constant of the system.


 It indicates how fast the system responds to a change in the input.
 Note that after a time of one time constant, y(t) is 63.2% of the way to its
final value.
 Generally, y(t) is considered to have achieved its final value after five time
constants.
 As an example, suppose that an oral thermometer takes at least 3 minutes to
indicate a patient’s temperature.
 Then we can guess that the time constant of the thermometer is in the order
of 0.6 minutes.

3.6.7.2 Harmonic input response

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 25 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

It means that the input is a sine wave with constant frequency and amplitude. In other
words the input is of the form(s)

All three definitions are equivalent. In the first, T is the period in seconds. In the
second, f is the frequency in cycles per second or Hertz. In the third, is the circular
frequency in radians per second.

Why should we be concerned with the response of measurement systems to harmonic


inputs?

The answer is that many natural and man-made processes are periodic in nature. For
example, think about day and night, the seasons repeating every year, the waves at the
beach, etc. In the engineering world think about the vibration of a motor, the
movement of pistons in an internal combustion engine, the turning of wheels, the air
conditioning system at your house turning on and off, and so on.

The steady state solution to this equation

with the input

Where is called the magnitude ratio and

is called the phase angle.

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 26 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

The magnitude ratio tells us how the amplitude of the output of the system depends
on the frequency of the input. The phase angle tells us how far the system output is
falling behind the input.

Figure 1.6 Instrument responses for first order harmonic input

3.6.7.3 Second order systems

Now, let us consider the response of a second order system to a step input. In other
words we need the solution to

In fact, we need to consider three different cases depending on the value of the
parameter  called the damping ratio.

 Case 1. Underdamped system   1

 Case 2. Critically damped case   1

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 27 of 28 
 
Instrumentation & Measurement    _____________________________________________________   _            AAiT 

 Case 3. Over damped case   1

Figure 1.7 Instrument responses for second order

_____________________________________________________________________________________
Compiled by Yidnekachew M.                                                                                                                               Page 28 of 28 
 

You might also like