Site Readiness for the Kodak DryView

5800 Laser Imager


Site Checklist
Use the following checklist to ensure that you have accomplished all of
the tasks required prior to installation of the 5800 Laser Imager.

✔ Item Responsibility Description Page


Installation Customer, The Imager must be installed by a
Carestream Carestream Health, Inc. Service Provider or
Health, Inc., or the customer’s qualified service personnel.
Service Provider
Imager Customer Provide adequate space for normal operator See “Location” on
Location functions and sufficient isolation from page 2.
patients, flammable material, liquids, and
other equipment.
Environmental Customer Ensure that the proper operating See “Environmental
Requirements environment is provided (temperature, Requirements” on
humidity, altitude, floor vibration, and floor page 4.
levelness).
Power Customer Provide a single-phase, 15 A power source, See “Network
Requirements with grounding, within 2.5 m (8 ft) of the Requirements” on
rear left corner of the Imager. page 6.
Ethernet Customer Install, test, and certify an Ethernet See “Network
Network Network prior to installation. Connection” on
page 6.
Network Customer Provide network (host) name, DICOM port See “Network
Configuration number, IP address, subnet mask, and Information” on
default gateway. page 6.
Modality Customer Provide the required information for each See “Modality
Configuration modality. Configuration” on
page 7.
Remote Customer Provide the required information for sites See “Remote
Management that will install the Remote Management Management Services”
Services Services. on page 9.

July 20, 2007 2G0735 1


Site Readiness

✔ Item Responsibility Description Page


Film Customer Order film from Carestream Health, Inc. See “Film” on page 10.
prior to installation. Observe storage
requirements and precautions listed on the
film package.
Seismic Kit Customer or A Seismic Kit is available for installations See “Seismic Kit” on
Carestream requiring earthquake protection. page 10.
Health, Inc.

Location
Equipment
Specifications Unpacked Packed
Height 62 cm (24.25 in.)
Width 62 cm (24.25 in.)
Depth 66 cm (26.0 in.) without film trays
?? cm (?? in.) with film trays
Weight 70 kg (155 lbs)

Operating Space The following is the minimum space required around the Imager to perform
Requirements normal operator functions. Service functions will require more space.

7.6 cm 60.96 cm 7.6 cm


(3.0 in.) (24.0 in.) (3.0 in.)

12.7 cm 5800 Imager


(5.0 in.)
60.96 cm
(24.0 in.)

134.62 cm
(63.0 in.)

60.96 cm
(24.0 in.)

76.2 cm
(30.0 in.)

2 2G0735 July 20, 2007


Site Readiness

Other Location Additional factors that influence where the Imager is located include
Considerations proximity to patients, flammable materials, liquids, and other equipment.

CAUTION:
This equipment is not contained in a sealed cabinet. Do not use this
equipment in locations where it can come in contact with liquids,
including body fluids.

IMPORTANT:
• This equipment is intended to connect to other medical devices. Only
an authorized Service Provider of Carestream Health, Inc. Products or
Customer Qualified Service Personnel may install this equipment.
• Do not locate the Imager within 2 m (6.56 ft) of where a cell phone
will be used, even if the cell phone is separated from the equipment
by a wall.
• Do not locate the Imager within 4 m (13.12 ft) of a microwave oven.
Electromagnetic radiation from a microwave oven is only an issue if,
after the oven door is closed and latched, the seal does not maintain
an electromagnetic tight fit between the oven door and the oven main
housing. Determining if the seal has an electromagnetic tight fit
requires special detection equipment.
• Do not locate the Imager in the presence of flammable anesthetics,
oxygen, or nitrous oxide. The Imager does not have a gas-sealed
electronics enclosure and could ignite any flammable or explosive
gases present in the Imager’s environment.
• Do not locate the Imager in close contact with MRI devices due to
possible high magnetic fields near an MRI unit. The magnetic field in
the area where the Imager is installed must be less than 50 Gauss.

1.83 m (6 ft.)

• Do not locate the Imager within the patient environment. Therefore,


do not locate the Imager closer than 1.83 m (6 ft) from a patient bed
or chair.
• Do not substitute or modify any part of the Imager without prior
written approval of Carestream Health, Inc.

July 20, 2007 2G0735 3


Site Readiness

Environmental Requirements
Temperature • Operating: 15° to 33° C (59° to 91° F)
• Storage: -40° to 60° C (-40° to 140° F)

Relative Humidity • Operating: 20% to 80% RH, noncondensing


• Storage: 10% to 90% RH, noncondensing

Altitude • 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level

Surface Condition A level (must be level within 1°) table top or counter capable of supporting
the weight of the Imager. For the Imager to operate with no image
degradation, surface vibration levels cannot exceed the following.

G2/Hz Freq. Range Direction


1 x 10-7 1-50 Hz Vertical or horizontal
6 x 10-5 50-200 Hz Vertical or horizontal
5 x 10-6 200-650 Hz Vertical or horizontal
1 x 10-6 650-1000 Hz Vertical or horizontal

Environmental Effects
• Heat Dissipation: total heat dissipation for the Imager is 850
BTU/hour
• Acoustical Noise:
– Less than or equal to 75 dB-A at 1 m during a sound burst of 1
sec. or less
– Less than or equal to 65 dB-A at 1 m during normal operation
– Less than or equal to 50 dB-A at 1 m during idle mode

4 2G0735 July 20, 2007


Site Readiness

Laser Specifications

Item Specification
Wavelength 810 nm
Power 50 mW
Radiation class 3B
Type Diode

Power Requirements
The Imager power supply has an auto-sensing and auto-switching feature for
both voltage and frequency. The current draw is 10 A at 120 V AC, and the
power consumption is a maximum of 1275 W.
One of the following single-phase, 15 A power sources, with grounding,
must be provided within 2.5 m (8 ft) of the Imager.
Area Power
North America 120 V AC +6% to -10%, 60 Hz ±3Hz
Europe 240 V AC ±10%, 50 Hz ±3Hz
Japan 100 V AC ±10%, 50/60 Hz ±3Hz
The wire must be insulation-rated for 600 V. A dedicated line is
recommended.

July 20, 2007 2G0735 5


Site Readiness

Network Requirements
Network Connection The 5800 Laser Imager receives digital images from medical imaging
devices (modalities) over a 10/100Base-T Ethernet Network. A single
10/100Base-T Ethernet Network connection via a CAT 5 UTP cable
terminating in an RJ-45 plug must be located less than 9.5 m (31 ft.) from
the bottom rear of the Imager.

Network Information During the installation, you will enter the network information for the
Imager to let it communicate across the network.
Gather and record the information in the following table prior to the
installation of the Imager. Have this page available when you start the
installation. See your site’s network administrator for help to gather the
information.
Required Information for the 5800 Laser ImagerFill in the table:
Host (Computer) Name
IP Address - Use the format: xxx.xxx.xxx.xxx
Subnet Mask - Use the format:
xxx.xxx.xxx.xxx
Default Gateway - Use the format:
xxx.xxx.xxx.xxx
DICOM Port Number - The default port
number, 5040, is correct for most installations.
If necessary you can change it for your site.
You will enter the port number in 2 places: at
the modality and at the Imager.

6 2G0735 July 20, 2007


Site Readiness

Modality Prior to installation, you must set up the image source (a modality such as
Configuration CT, MRI, workstation, or other imaging device) to send print jobs to the
5800 Laser Imager. At the modality, the Imager must be added as a print
destination. If you need help adding the Imager, contact the modality’s
vendor.
A list of the modalities that will connect to the Imager and the configuration
for each modality must be available at installation. Use the following to
record the information that will be required.

1 2
Manufacturer: ________________________ ________________________
Model: ________________________ ________________________
Modality Type (CT, DR, MR, etc.): ________________________ ________________________
GSDF Compliant (Yes or No): ________________________ ________________________
If GSDF Compliant, does the modality
send a PLUT with all images (Yes or No): ________________________ ________________________
Film Type: ________________________ ________________________
Film Size: ________________________ ________________________
Modality ID (descriptive name assigned
by the user for use on the Local Panel): ________________________ ________________________

3 4
Manufacturer: ________________________ ________________________
Model: ________________________ ________________________
Modality Type (CT, DR, MR, etc.): ________________________ ________________________
GSDF Compliant (Yes or No): ________________________ ________________________
If GSDF Compliant, does the modality
send a PLUT with all images (Yes or No): ________________________ ________________________
Film Type: ________________________ ________________________
Film Size: ________________________ ________________________
Modality ID (descriptive name assigned
by the user for use on the Local Panel): ________________________ ________________________

July 20, 2007 2G0735 7


Site Readiness

5 6
Manufacturer: ________________________ ________________________
Model: ________________________ ________________________
Modality Type (CT, DR, MR, etc.): ________________________ ________________________
GSDF Compliant (Yes or No): ________________________ ________________________
If GSDF Compliant, does the modality
send a PLUT with all images (Yes or No): ________________________ ________________________
Film Type: ________________________ ________________________
Film Size: ________________________ ________________________
Modality ID (descriptive name assigned
by the user for use on the Local Panel): ________________________ ________________________

7 8
Manufacturer: ________________________ ________________________
Model: ________________________ ________________________
Modality Type (CT, DR, MR, etc.): ________________________ ________________________
GSDF Compliant (Yes or No): ________________________ ________________________
If GSDF Compliant, does the modality
send a PLUT with all images (Yes or No): ________________________ ________________________
Film Type: ________________________ ________________________
Film Size: ________________________ ________________________
Modality ID (descriptive name assigned
by the user for use on the Local Panel): ________________________ ________________________

8 2G0735 July 20, 2007


Site Readiness

Remote Management Services


The Remote Management Services enables the Service Provider to perform
many of the tasks that the on-site Field Engineer performs. These tasks
include remotely installing software, configuring the Imager, performing
diagnostics, and checking image quality. The Service Provider can isolate a
problem without customer interaction and facilitate problem resolution or
the dispatch of a Field Engineer to provide on-site service.
To utilize the Remote Management Services, the Imager must be connected
to the internet. It is recommended that a high-speed connection be used.
Outbound connections from the Remote Management Services Agent are
initiated securely behind the customer’s firewall over port 443. The Agent
does not require any special port or a fixed IP address. Contact your Sales
Representative or Service Provider for more information about the Remote
Management Services.
Use the following to record information that will be required during the
installation of the Remote Management Services.
Customer’s Site Name: ________________________
Customer’s Country: ________________________
Customer’s Location: ________________________
Location at the customer site where the
Imager will be installed: ________________________
Is a proxy server required? Yes ____ No ____
Proxy Type: Normal _____ NTLM _____
Proxy Location: ________________________
Proxy Port Number: ________________________
Proxy Username: ________________________
Proxy Password: ________________________
Proxy Domain: ________________________
Proxy Server Name: ________________________
DNS Configuration: Yes _____ No _____
DNS Domain Name: ________________________
Domain Server IP Addresses: ________________________

July 20, 2007 2G0735 9


Site Readiness

Film
Prior to installation you must order and have on hand Kodak DryView Laser
Imaging Film. Contact your sales representative if you have questions.

Film Types The 5800 Laser Imager accommodates the following film types:
• Kodak DryView DVB Laser Imaging Film
• Kodak DryView DVC Laser Imaging Film
• Kodak DryView DVB+ Laser Imaging Film
• Kodak DryView DVB+ Premium Laser Imaging Film
NOTE: Not all film types are available in every country.

Film Sizes The 5800 Laser Imager accommodates the following film sizes:
• 20 x 25 cm (8 x 10 in.)
• 25 x 30 cm (10 x 12 in.)
• 28 x 35 cm (11 x 14 in.)
• 35 x 35 cm (14 x 14 in.)
• 35 x 43 cm (14 x 17 in.)

Film Storage and Film must be stored at 35° C (95° F) and 85% RH or less. The Imager prints
Throughput up to 80 films per hour.

Seismic Kit
Some installation sites require that the Imager be fastened to the floor to
prevent movement due to earthquake or MR attraction. The brackets and
hardware required to secure the Imager are supplied in a Seismic Kit,
Catalog Number xxxxxx. The Seismic Kit satisfies the requirements of
California Building Code Title 24, Volume 2, Chapter 16A - CCR Uniform
Building Code 2001.

10 2G0735 July 20, 2007