Measurement Liquid Flow Closed Conduits Transit-Time Ultrasonic Flowmeters
Measurement Liquid Flow Closed Conduits Transit-Time Ultrasonic Flowmeters
Measurement Liquid Flow Closed Conduits Transit-Time Ultrasonic Flowmeters
ANSVASME MFC-5M-1985
REAFFIRMED 1994
REAFFIRMED 2001
FOR CURRENT COMMITTEE PERSONNEL
PLEASE E-MAIL CS@asme.org
SPONSOREDANDPUBLISHEDBY
T H EA M E R I C A NS O C I E T Y OF M E C H A N I C A LE N G I N E E R S
United
Engineering
Center 3 4 5 East 47th
Street
New York, N. Y. 1001 7
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
Date of Issuance: July 15, 1985
This Standard will be revised when the Society approves the issuance of a new edition.
There will be no addenda or written interpretations of the requirements of this Standard
issued to this Edition.
This code or standard was developed under procedures accredited as meeting the criteria for
American National Standards. The Consensus Committee that approved the codeor standard was
balanced to assure that individuals from competent and concerned interests have had an oppor-
tunity to participate. The proposed code or standard was made available for public review and
comment which provides an opportunity for additional public input from industry, academia, reg-
ulatory agencies, and the public-at-large.
ASME does not "approve," "rate," or "endorse" any item, construction, proprietary device, or
activity.
ASME does not take any position with respect to the validity of any patent rights asserted in
connection with any items mentioned in this document, and doesnot undertake to insure anyone
utilizing a standard against liability for infringement of any applicable Letters Patent, nor assume
any such liability. Users of a code or standard are expressly advised that determination of the
validity of any such patent rights, and the risk of infringement of such rights, is entirely their own
responsibility.
Participation by federal agency representativek) or person(s) affiliated with industry is not to
be interpreted as government or industry endorsement of this code or standard.
ASME accepts responsibility for only those interpretations issued in accordance with governing
ASME procedures and policies which preclude the issuance of interpretations by individual vol-
unteers.
Copyright 0 1985 by
THE AMERICANSOCIETY OF MECHANICAL ENGINEERS
All Rights Reserved
Printed in U.S.A.
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
FOREWORD
The 'need for a document describing measurement of liquid flows by means of transit-time
ultrasonic flowmeters has been recognized for many years. This document represents the first
formal attempt to establish common ground between the users and manufacturers of these
meters. Doppler flowmeters are not covered by this document.
This Standard, which was approved by the ASME Standards Committee on Measurement of
Fluid Flow in Closed Conduits, was approved and designated as an American National Standard
by the American National Standards Institute on April 12, 1985.
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ASME STANDARDSCOMMITTEE
Measurement of Fluid Flow in Closed Conduits
(The following is the roster of the Committee atthe time of approval of this Standard.)
OFFICERS
R. W. Miller, Chairman
W. F. 2. Lee, Vice Chairman
K. Wessely. Secretary
COMMITTEEPERSONNEL
R. B. Abernethy, Pratt & Whitney Aircraft, West Palm Beach, Florida
J. W. Adam, Houston, Texas
N. A. Alston, Diederich Standard Corp., Boulder, Colorado
H. P. Bean, El Paso, Texas
S. R. Beitler, The Ohio State University, Columbus, Ohio
M. Bradner, The FoxboroCo., Foxboro, Massachusetts
E. E. Buxton, St. Albans, West Virginia
J. S. Castorina. Naval Ship System Engineering Station, Philadelphia, Pennsylvania
G . P. Corpron, Rosemount Inc., Eden Prairie, Minnesota
C. F. Cusick, Philadelphia, Pennsylvania
L. A. Dodge, Richmond Heights, Ohio
R. B. Dowdell, University of Rhode Island, Kingston, Rhode Island
R. L. Galley, Antioch, California
D. Halmi, D. Halmi & Associates Inc., Pawtucket, Rhode Island
R. M. Hickox, S.O.R., Inc., Olathe, Kansas
H. S. Hillbrath, Boeing Aerospace Co., Santa Clara, California
E. H. Jones, Jr., Gulf Research and Development Co., Houston, Texas
L. J. Kemp. Palos Verdes Estates, California
D. R. Keyser, Aero-Mechanical Branch, Warminster, Pennsylvania
C. P. Kittredge, Princeton, New Jersey
W. F. 2. Lee, Rockwell International, Pittsburgh, Pennsylvania
E. D. Mannherz, Fischer & Porter Co., Warminster, Pennsylvania
G. E. Mattingly, National Bureau of Standards, Washington, D.C.
R. W. Miller, The Foxboro Co., Foxboro, Massachusetts
R. V. Moone, Union Carbide Corp., Tonawanda, New York
L. C. Neale, Jefferson, Massachusetts
M. H. November, Hacienda Heights, California
W. M. Reese, Jr., Fluidic Techniques Inc., Mansfield,Texas
P. G. Scott, Observer, The Foxboro Co., Foxboro, Massachusetts
H. E. Snider, AWWA Standards Committee, Kansas City, Missouri
D. A. Sullivan, Fern Engineering, Bourne, Massachusetts
R. G. Teyssandier, Daniel Industries Inc., Houston, Texas
J. S. Yard, Fischer & Porter Co., Warminster, Pennsylvania
V
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
SUBCOMMITTEE 9 - ULTRASONIC FLOWMETERS
R. G. Teyssandier, Chairman, Daniel Industries, Inc., Houston, Texas
R. W. Burdett, Kennecott Minerals Co., Salt Lake City, Utah
C. G. Clyde, Utah State University, Logan, Utah
G. P. Corpron, Rosemount Inc., Eden Prairie, Minnesota
L. D. DiNapoli, Leeds & Northrup Co., North Wales, Pennsylvania
D. J. Grant, NASA, Greenbelt, Maryland
D. Halrni, D. Halmi & Associates Inc., Pawtucket, Rhode Island
R. E. Johnson, BC Hydro & Power Authorities, Surrey, British Columbia, Canada
L. J. Kernp, Palos Verdes Estates, California
R. B. Kirnball, Technical Sales Development Co., San Francisco, California
F. C. Lowell, Ferranti ORE Inc., Falmouth, Massachusetts
L. C. Neale, Jefferson, Massachusetts
P. H. Nelson, Bureau of Reclamation, Denver, Colorado
W. E. VanOver. Transducers Inc., Cerritos, California
J. H. Vignos, The FoxboroCo., Foxboro, Massachusetts
E. M. Zacharias, Jr., MAPCO Controls Co., Tulsa, Oklahoma
vi
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
CONTENTS
...
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Committee
Roster ............................................... v
General
1 ................................................... 1
2 Flowmeter Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3 Error
Sourcesand Their Reduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4 Application
Guidelines ......................................... 8
5 Meter FactorDetermination
and Verification . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Figures
Wetted
1 Transducer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 ProtectedConfiguration With Cavities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 ProtectedConfiguration With Protrusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4 ProtectedConfiguration With Smooth Bore . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
5 Acoustic Path Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
6 A Typical Cross Path Ultrasonic Flowmeter Configuration (Paths
May Be Multipath Diametric or Chordal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Appendices
AAcousticTransit Timein a Nonuniform FluidVelocity Field . . . . . . . . . . . . . . . . 11
B Selection Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
vii
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled wh
ANSllASMEMFC-5M-1985
ANAMERICANNATIONALSTANDARD
++
liquid and the speed of sound in that same liquid at rest
is directly proportional to the liquid’s instantaneous tup = (3)
velocity. As a consequence, a measure of the average
c, - v, cos eo
velocity of the liquid along a path can be obtained by
transmitting an acoustic pulse along the path and sub-
sequently measuring its transit time. and
The volumetric flow rate of a liquid flowing in a com-
pletely filled closed conduit is defined as the average
(4)
velocity of the liquid over a cross section multiplied by
the area of the cross section. Thus, by measuring the sin’ e, + V , cos e,
average velocity of a liquid along one or more acoustic
paths and combining the measurements with knowledge
where 1, is the straight line distance between the centers
of the cross-sectional area and the velocity profile over
of the faces of the acoustic transmitterand receiver.
the cross section, it is possible to obtain an estimate of
Taking the difference between the reciprocals of these
the volumetric flow rate of the liquid in the conduit.
transit times leads to
The theoretical considerations involved in implement-
ing these concepts is the subject of this section.
1 1
- 2va, cos e,
2.1.2 Fluid Velocity Measurement. Several techniques tup tdown IO
can be used to obtain a measure of the average effective
speed of propagation of an acoustic pulse in a moving and, onrearranging, to
liquid in-order to determine the average axial flow ve-
locity (V,) along an acoustic path. Two approaches,
transit-time difference and frequency difference, will be
discussed here.
2.1.2.1 Transit-TimeDifference. Thebasis of this
technique is the direct measurement of the transit time
of acoustic signals as they propagate between a trans- -
mitter and a receiver, both of which are assumed (see where A t = tup - tdown. Since v,,
is constant, = v,, v,
para. 2.1.3) to be in direct contact with the liquid. Dif- (the average flow velocity).
ferent transmitter/receiver arrangements are described in This analysis becomes more complicated in theab-
para. 2.1.3 below. For an acoustic signal traveling up- sence of plug flow. Nevertheless, to the degree that
stream, the apparent sound speed at any point along the (Vax/Co)&ax.<< 1, the result for the averageof the
line of transmission, assuming only axial flow, is axial liquid velocity along the acoustic path is (v,)
identical (see Appendix A), i.e.,
2
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled whe
MEASUREMENT OF LIQUIDFLOW IN CLOSEDCONDUITS ANSI/ASME MFC-5M-1985
USINGTRANSIT-TIMEULTRASONICFLOWMETERS ANAMERICANNATIONALSTANDARD
transittime beingreplacedby the reciprocal of a con- 2.1.4 Estimating Volumetric Flow. Oncethe average
stant (Kf)times
the frequencies, i.e., axial flow
velocity
along acoustic
an path
has been
found, the volumetric flow rate can be calculated from
- Kf 10 the following equation:
v, = (fdown - fup) (6)
2 cos eo n
3
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
ANSllASMEMFC-5M-1985 MEASUREMENT OF LIQUIDFLOW IN CLOSEDCONDUITS
ANAMERICANNATIONALSTANDARD USINGTRANSIT-TIMEULTRASONICFLOWMETERS
Upstream
transducer
element
Intervening
material
\ \
\
Downstream
transducer
element
I I
I Measurement I
section
FIG. 3 PROTECTED CONFIGURATIONWITH
PROTRUSIONS
FIG. 1 WETTED TRANSDUCER
CONFIGURATION
Intervening Intervening
material material
'\
\ L?\
FIG. 2 PROTECTED CONFIGURATION
WITH
FIG. 4 PROTECTED CONFIGURATION
WITH
CAVITIES SMOOTH BORE
4
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
MEASUREMENTOFLIQUIDFLOW IN CLOSEDCONDUITS ANSI/ASME MFC-5M-1985
USINGTRANSIl-TIMEULTRASONICFLOWMETERS ANAMERICANNATIONALSTANDARD
Single Path
Multipath
Chordal Diametric
types areaxial, diametric,andchordal, as shown in characteristic point of the received signal. Exact details
Fig. 5. vary from one manufacturer to another.
2.2.2 SecondaryDevice. The secondary device com- 2.2.2.3 Processing of Data. The processing section,
prises the electronic equipment required to operate the in addition to estimating the flow rate from measured
transducers, make the measurements, process the mea- transittimes,should be capable of rejecting invalid
sured data, and display or record the results. measurements, noise, etc. The indicated flow rate may
be the result of oneormore individual flowvelocity
2.2.2.1 Operation of Transducers. Transducers may determinations.
be excited simultaneously oralternatelywithoneor
more transmissions in each direction. The acoustic fre- 2.2.2.4DisplaysandOutputs. Most meters have
quency and pulse repetition rate may vary. The intercon- several outputs available, either as standard features or
necting cable lengthsbetweenthe secondary and pri- as optional additions to the equipment. Displays may
mary devices are an important consideration, and their show flow rate, integrated flow volume and/or direction,
maximum length is usually defined by the manufacturer. and may be analog or digital. Signal outputs usually in-
clude one or more of thefollowing: current, voltage, dig-
2.2.2.2 Measurement Method. The transit time of ital, and a pulse rate proportional to flow. These outputs
an acoustic pulse is usually taken to be the time interval may or may not be electrically isolated. Flowmeters may
between initial excitation of the transmitter and some also include alarms and diagnostic aids.
5
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
ANSIIASMEMFC-5M-1985 MEASUREMENTOFLIQUIDFLOWIN CLOSEDCONDUITS
ANAMERICANNATIONALSTANDARD USINGTRANSIT-TIMEULTRASONICFLOWMETERS
3 ERROR SOURCES AND THEIR REDUCTION using the same detection electronics and transducer pairs
for both directions of propagation.
The purpose of this Section is to describe possible er-
ror sources for the ultrasonic flowmeters covered by this 3.1.2.2 Signal Detection. Acoustic transit-time mea-
Standard. Although these errors may not be significant surements may be affected by inconsistencies in recogni-
in some cases, they should all be addressed in detail tion of the received acoustic signal caused by variations
when analyzing the error for a particular flowmeter. in received signal levelof waveform and noise.
(a) Variations inreceivedsignallevel or waveform
3.1 Axial Velocity Estimate can occur as the acoustic properties ofthe liquid in
the measurement section changebecause of excessive
Axial velocity errors are those errors that are made in amounts of entrained air, suspended solids, temperature,
thedetermination of Fax across an acoustic path [see pressure, etc., or as transducer fouling occurs. These vari-
Eqs. (5)-(8)]. . ations may result in uncertainty in determining the tran-
3.1.1 AcousticPathLengthandAngle. The determi- sit time, thus causing uncertainty in. ,V
nation of axial flow velocity (Fa) is based on the acous- ( b ) Noise can affect the accuracy of the transit-time
tic path length (1,) and angle (6). The error in v, is in measurement. Noise sources may be either electrical or
acoustic, and either external or self-generated. Generally,
direct proportion to the uncertainty in the acoustic path
length and angle. Acoustic path length and angle errors externally generated electronic or acoustic noise is ran-
may be static due to inaccuracies in the initial measure- dom with respect to the received signal. Self-generated
ments, or they may be dynamic due to changesin the acoustic noise, however, is usually synchronized with the
measurement section or, in the case of refractive sys- received signal and is therefore much harder to compen-
tems, changes in the index of refraction of the materials sate for in the secondary device.
in the acoustic path. Signal detection errors are reduced by operating with
Errors in the acoustic path .length or angle for nonre- high signal-to-noise ratios. Methods can be provided that
fractive systems canbe reduced by accurate geometric will reject those signals which are excessively attenuated
and acoustic measurements. For flowmeters in which or which are distorted by noise.
sound energy undergoes refraction,errors in acoustic If the background noise is caused by external sources,
path length or anglecanbe reduced by design and/or the signal-to-noise ratio canbe improved by increasing
compensation based upon knowledge of the speed of the transmitted signal level. However, in many cases, the
sound in the process liquid and intervening materials be- most troublesome noise is self-generated acoustic noise.
tween the transducer element and the process liquid. This noise generally increases as the level of the trans-
In certain applications, compensation for changes in mitted signalis increased. The signal-to-noise ratio, in
acoustic path length and angle which result from temper- these cases, may beimproved by acoustically isolating
ature or pressure induced pipe deformation may be pro- the transducers from the measurement section by ap-
vided for in either refractive or nonrefractive systems. plication of damping materials and/orother suitable
Systems where the transducers are field mounted can techniques.
achieve accuracies comparable to systems where the
3.1.3 Sound Speed Dependency. The speed of sound
transducers are factory mounted if precise postinstalla-
in the liquid and in any intervening materials along the
tion measurements are made and the scaling of the elec-
acoustic path varies with composition, temperature, and
tronics is adjusted to reflect these as-built dimensions.
pressure. Depending upon a particular ultrasonic flow-
3.1.2 Transit Time meter’sdesign, I,, 8 , and to (f,) [Eqs. (7) and (811
3.1.2.1 Timing. Uncertainty in the transit-time mea- may be affected.
surement due to limits in internal timing accuracy - and In nearly all cases, the errors caused by sound speed
resolution lead to a corresponding uncertainty in V,. variations in the liquid are negligible for a properly im-
Errors in the measurement of transit time may be re- plemented, nonrefractive, wetted transducer system.
duced by the use of stable and accurate high frequency Changes in the speed of sound in the intervening ma-
oscillators and averaging many individual transit-time terial may, however, require compensation for non-
measurements. Transit-time measurement errors are the wetted transducer systems.
most amenable of all potential errors to analysis. In refractive systems, changes in the speed of sound in
Transit-time measurement errors due to differences intervening materials and the process liquid affect the
between upstream to downstream and downstream to acoustic path length and angle. It is possible to compen-
upstream electronic signal paths may be reduced by sate for these effects. In rare cases, changes in the speed
6
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
MEASUREMENTOFLIQUIDFLOW IN CLOSEDCONDUITS ANSllASME MFC-5M-1985
USINGTRANSIT-TIMEULTRASONICFLOWMETERS ANAMERICANNATIONALSTANDARD
/
/ \
/ \
of sound in the liquid may refract the beam so much structural loading. It may also be due to the formation
that it misses the opposing transducer. Accurate knowl- of deposits or growths, such as algae,'in the measurement
edgeof the speed of sound in a particular liquid can section. Usually it is due to combinations of the items
reduce this possibility. mentioned above.
Cross section dimensional errors can be reduced by
v,
3.1.4 Secondary Flow. Secondary flow can produce
manufacturing or choosing a measurement section which
an error in the determination of since it is normally
has constant dimensions along its length, can be accu-
assumed in the calculations that all flow is in the axial
rately measured, and has a stable surface so that cross
direction. Secondary flow is a result of flow perturba-
section changes with time due to corrosion, material
tions occurring upstream or downstream of the measure-
buildup, or loss of protective coatings willbe small.
ment section because of devices such as elbows, valves,
Additionally, if temperatures or pressures are expected
and pumps. Secondary flow may result inan error in
to be substantially different from reference conditions,
determining
- the transit time which ultimately affects
it may be necessary to adjust the measured dimensions
vu. to compensate for dimensional changes that may occur
The most effective way to reduce secondary flow er-
under operating conditions.
rors is to avoid installations where severe secondary flow
In circular pipes, cross section dimensional errors can
exists. Reduction of secondary flow may require long,
be reduced by minimizing the effects of out-of-roundness
straight runs of pipe and/or the use of flow straighteners,
through averaging of diameter measurements made at
depending upon the nature of the secondary flow source
the upstream, middle, and downstream ends of the mea-
and the accuracy required. Secondary flow errors can
surement section.
alsobe reduced by the use of an appropriate acoustic
The measurement section should be inspected period-
path orientation or by computing line velocities on ap-
propriate multiple acoustic paths, i.e., crossed paths as ically to determine if the dimensional factor should be
illustrated in Fig. 6 , and averaging. adjusted to compensate for observed changes.
3.22 AcousticPathLocation. T h e acoustic path lo-
cation, particularly in multipath flowmeters, is an impor-
3.2Integration tant contributor to overall flowmeter accuracy. The un-
Integration error is the error in the flow rate estimate certainty in the position of the acoustic path can cause
which occurs in the computation of the flow rate from errors through improper assignment of a weighting fac-
Fa,, A , S,and W ias shown in Eq. (10). tor (Wi)and by causing unnecessary sensitivity of i, to
the velocity profile through nonoptimum placement of
3.2.1 CrossSectionDimensionalErrors. Error in the transducers.
assumed cross-sectional area of the measurement section Errors due to acoustic path location can be reduced
causes an error in the flow rate estimate. This error may by suitable manufacturing techniques in the case of a
be due to initial measurement section shape irregularities, prefabricated measurement section or by accurately
such as out-of-roundness; or it may be due to changes in determining the acoustic pathfor systems where the
the initial shape caused by temperature, pressure, or transducers are field mounted.Pathlocations can be
7
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
ANSllASMEMFC-5M-1985 MEASUREMENT OF LIQUIDFLOW IN CLOSEDCONDUITS
ANAMERICANNATIONALSTANDARD USINGTRANSIT-TIMEULTRASONICFLOWMETERS
determined in a variety of ways, but probably the most 3.5 Equipment Degradation
accurate involves optical determination of both the
Performance errors may arise due to fouling or phys-
acoustic path location and angle within the measurement
ical degradation oftheequipment. Equipment design
section.
should include reasonable tolerance to changes in com-
3.2.3 Velocity Profile. Ultrasonic flowmeters are af- ponent values and process conditions. The equipment
fected by variations in flow profile because uncertainty should also indicate when degradation of flowmeter
in the velocity profile causesan error in that may performance occurs. The probability of error canbe
not be compensated for by the velocity profile correc- reduced considerably by including suitable self-test or
tion factor S. This error may affect both the linearity diagnostic circuits in the equipment.
and the value of the flow rate estimate. Velocity profile
variations can be caused by changes in flow rate (both
transient and steady state), wall roughness, temperature, 4 APPLICATION GUIDELINES
viscosity, upstream or downstream hydraulic conditions, 4.1 PerformanceParameters
transducer projections, and transducer cavities. In the
absence of upstream and downstream hydraulic effects, By evaluating the performance parameters listed
the Reynolds number and friction factor of the measure- below, a user should be better able to predict the per-
ment section would be sufficient to determine the veloc- formance of a given ultrasonic flowmeter in a specific
ity profile correction factor S. application, as well as compare its performance to other
There usuallyis a difference between the actual ve- ultrasonic flowmeters. These parameters are limited
locity profile and that assumed in the flowmeter’s com- to those which canbe substantiated by test or well-
putations. Since most flowmeter computations assume a established computational methods. (It should beem-
fully developed velocity profile, errors can be reduced phasized that the following performance parameters are
byplacing the measurement section as far as possible normally determined by the manufacturer under specific
from bends, valves, tees, transitions, etc. These errors reference conditions that will, in general, differ from the
can also be reduced by using a more accurate model of user’s actual conditions.) Subjective or nonperformance
the actual velocity profile, or in general, by increasing parameters, such as convenience of installation, reliabil-
the number of acoustic paths. ity, and cost, are left to the individual user for evalua-
tion. In each case below, the range and conditions ap-
plicable to each parameter should be specified.
indication (seealsoANSI/ASME MFC-lM, Glossary of 4.2.5 Secondary Flow. Secondary flow directly af-
Terms Used in the Measurement of Fluid Flow in Pipes). fects a meter’s performance and should be considered in
the design of the installation and in the selection and
4.1.4 Stability. Stability is a measure of change in ac-
orientation of a meter. In general, the meter should be
curacy with time. It should be reported as the maximum
placed as far as possible from upstream elbows, transi-
deviation in accuracy, as a percentage of actual flow,
tions, valves, etc.
which canbe expected to occur over a specified time
period (subject to constant hydraulic conditions). 4.2.6 Integration. Flow profile changes and dimen-
sionalchangesin the measurement section, including
4.1.5 Resolution. Resolution is the minimum change
those caused by corrosion, erosion, or material buildup,
in actual flow required to produce an observable change
in the output of the equipment. directly affect meter performance and should be con-
sidered in the selection, location, and orientation of a
4.1.6 Rangeability. Rangeability is the maximum and meter.
minimum flow rates over which the performance is spec- The measurement section should be inspected period-
ified (see also ANSI/ASME MFC-lM, Glossary of Terms ically to determine if the cross section area factor K or
Used in the Measurement of Fluid Flow in Pipes). profile correction factor S should be adjusted to com-
4.1.7. Response Time. Response time is the time it
pensate for observed changes.
takes, following a step change in the flow rate, for the 4.2.7 Calibration. Installation considerations and the
flowmeter’s output to indicate a changeinflow rate required installed accuracy usually determine the method
equal to 63%of the step change. of calibration.
4.1.8 PowerRequirements. The power requirements
of the flowmeter, including voltage and frequency toler-
ances necessary for proper performance, as well as the
flowmeter’s power consumption, should be clearly 5 METERFACTOR DETERMINATION
specified. ANDVERIFICATION
There are three principal methods of meter factor
4.2 Installation Considerations determination:
(a) laboratory calibration
4.2.1 Introduction. Many of the error sources listed in ( b ) field calibration
Section 3 can be reduced or eliminated by proper instal- (c) analytical procedures
lation. Errors the user should address during the design The first two can be used to verify meter performance.
phase of a project are listed below.
4.2.2 AcousticPathLengthandAngle. Changes in 5.1 Laboratory Calibration
acoustic path length and angle can be caused by signifi-
cant temperature or pressure changes as well as external Laboratory calibrations should be conducted at fa-
loading of the meter section. The installation location cilities where the procedures are in accordance with
should be chosen to minimize these effects. national or international standards.
The calibration tests should generallybe run using
4.2.3 Signal Detection. Suspended solids, fouling, en-
water that is free from acoustically interfering entrained
trained air, or cavitation (caused by upstream equipment
air or solid particles. Standard calibrations should be
or even the meter itself) may degrade accuracy or pre-
conducted using flows that are as free as possible from
vent operation by attenuating the acoustic signal. Elec-
swirl, nonaxisymmetric flow, and pulsation. Generally,
trical interference and acoustic noise caused by mechan-
these conditions have been achieved by using sufficient
ical vibration or cavitation can also interfere with the
lengths of straight pipe upstream and downstream of the
meter’s operation.
measurement section and, if necessary, by installing up-
4.2.4 Multiple Fluids. Metering fluids with widely dif- stream flow conditioners.
fering acoustic properties may require multiple primary The extent to which the above conditions have been
devices (spool pieces) due to excessive acoustic beam an- achieved can be determined by noting the sensitivity of
gular variations in refractive systems and/or excessive the meter factor to rotation and translation of the pri-
signalloss due to acoustic mismatch or attenuation. mary device.
Under these conditionsthe manufacturer should be A statistically significant number of runs should be
consulted. made over a range of flow rates. Flowmeter accuracy,
9
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled
ANSllASMEMFC-5M-1985 MEASUREMENTOFLIQUIDFLOWINCLOSEDCONDUITS
ANAMERICANNATIONALSTANDARD USINGTRANSIT-TIMEULTRASONICFLOWMETERS
within the uncertaintyofthelaboratory standards, degree. of uncertainty in the accuracy of the standards
should be determined by the combined random and sys- that are employed. In some cases, these secondary stan-
tematicerrors in the measurement ofthe volumetric dards may be considerably less accurate than the ultra-
flow rate. sonic flowmeter being calibrated.
Special calibration tests may also be performed for
those cases where the piping in the final installation may
5.3 Analytical Procedures
produce nonaxisymmetric flow or where other flow ir-
regularities are suspected. This will require appropriate Analytical procedures are oftenthe only available
modeling of upstream and downstream piping. technique for meter factor determination. This is partic-
ularly true for very high flow rates and large line sizes.
These procedures require physical measurements as well
5.2 Field Calibration
as instructions and data supplied by the manufacturer.
Field calibration, as opposed to laboratory calibration, The uncertainty in the meter performance should reflect
has 'the advantage thattrue operating conditions are uncertainties associated with these procedures as well as
encountered. The major disadvantage may be a greater those error sources outlined in Section 3.
10
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
APPENDIX A
ACOUSTIC TRANSIT TIME IN A NONUNIFORM FLUID VELOCITY FIELD
In the case of a nonuniform axial flow field, it can be shown that the length 1 of the acoustic
path, which in this case is not a straight line, is given to second order by:
where 1, is the straight line distance connecting the centers of the faces of the acoustic trans-
mitter and receiver (see Fig. 1) and
where 8, is the angle of 1, with respect to the pipe axis, and vaxand Vax2
-
are averages of the
liquid axial velocity profile. To the degree that a << 1, the effective path of the acoustic ray
can be taken to be a straight line equal to I,. This being the case, the apparent speed of sound
at any given point, for an acoustic pulse traveling upstream along this straight line path, is given
bY
The transit times for upstream and downstream acoustic pulses (tup, tdown)are given
respectively by:
11
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
and
and
where
2V, cos eo
-
[1- cosz eo ($J]
N
1,
12
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
Assuming the worst case, i.e., that velocity profile which results in the most stringent condi-
13
tions, these inequalities reduceto:
Copyrighted material licensed to Stanford University by Thomson Scientific (www.techstreet.com), downloaded on Oct-05-2010 by Stanford University User. No further reproduction or distribution is permitted. Uncontrolled w
APPENDIX B
SELECTIONGUIDELINES
(This Appendix is not an integral part of ANSllASME MFC-5M-1985 and is included for information purposes only.)
This Appendix is intended to assist in selecting the install than insert systems. Since the pipe inner wallis
most appropriate ultrasonic flowmeter fora particular undisturbed, there is no flow disturbance in the vicinity
application. Since there are many variations and differ- of the transducers. They can also be easily removed with-
ences even among the same types of flowmeters, this out requiring shutdown of the process. Since these sys-
Appendix addresses itself only to the major differences tems utilize a beam that refracts into the liquid, they
between the types. It issuggested that the application may be affected by variations in the sonic properties of
conditions be discussed with the manufacturers prior to the liquid andthe pipe. The manufacturer should be
a decision on a particular type of flowmeter. contacted to ensure that the expected application con-
ditions can be handled with satisfactory performance.
Insert type systems can be of two types - those using
B1 SINGLE PATH VERSUS MULTIPATH wetted transducers and those where the transducers are
INSTRUMENTS installed in a protective well. Where the transducer faces
and protective surfaces are orthogonal to the acoustic
Single path instruments are usually lower in cost than path, these systems can offer greater immunity to
multipath instruments. They are also less complex, and changes in the sonic velocity of the liquid since refrac-
therefore usually somewhat simpler to install. Multipath tion of the beam does not take place. Since these sys-
flowmeters can exhibit better performance than single tems usually either have the transducer recessedin a
path meters under the variable and/or nonideal velocity cavity or protrudingintothe liquid beyond the pipe
profile distribution conditions caused by changing Reyn- wall, a local flow disturbance results which may affect
olds number, changes in friction factor, and the effects the meters performance. Although reportedly rare, the
caused by upstream and downstream elbows, valves, or transducer cavity could collect debris.
other sources of flow disturbance. The choice of path Those insert systems where this well is filed with a
orientations varies among manufacturers. It can be either protective window result in refraction ofthe beam
crossed or parallel, either chordal or diametric. Choice of through the liquid and require either relatively constant
the most appropriate path distribution should be made sonic velocity or compensation in the secondary device.
after a full discussion of the application conditions with
the various manufacturers. NOTE: If verylargechanges in sonic velocity occur, refaaction
may cause the sonic beam to miss the opposing transduces.
14