True Sense DO-66 Pen Type Dissolved Oxygen Meter

The DO-66 type pen type intelligent Dissolved Oxygen Analyzer is designed to facilitate the
user to carry the operation on the spot. The instrument can be divided into two parts: the
sensor and electronic unit. The sensor adopts thin film dissolved oxygen electrode. The
electronic unit is an integrated operational amplifier + microcomputer chip with automatic
temperature compensation function.

Product Code : DO-66

Package Includes:

• 1 x Dissolved Oxygen Electrode


• 1 x Dissolved Oxygen Membrane
• 1 x Dissolved Oxygen Filling Solution (30ML)
• 1 x Anaerobic Water (50ML)

Website : www.truesense.net.in ; www.truesense.in


Email : sales@truesense.in ; truesense.sales@gmail.com
Technical Features :

1.Measurement Range:
DO: 0~199.9%; 0~16.50mg/L
Temperature :0~50℃; 32~122℉

2.Resolution:
DO: 0.1%; 0.01mg/L
Temperature :0.1℃; 32.18℉

3.Accuracy:
DO: ±1.5% F.S
Temperature :±1℃;±33.8℉

4.Residual Current: No more than 0.15mg/L


5.Response Time: No more than 15 seconds
6.Temperature Compensation: 0℃~50℃
7.Power Supply: 2 *AAA batteries
8.Working Temperature: 0℃~50℃
9.Dimensions: 207*32*36mm
10.Weight: 85 Grams

How To Use :

1. Installing batteries.
2. Connect the DO electrode to the corresponding input.
3. Press ON/OFF key to turn on the machine, take off the electrode protector and insert the
electrode into the solution to be tested.
4. Press [% and mg / L] key for % and mg/L display conversion.
5. The dissolved oxygen concentration of the solution (mg/L or %) can only be read after the
display of the instrument is stable.
6. Press the ‘°C/°F’ key to switch between Celsius and Fahrenheit.
7. Press the ON/OFF key to turn off the power.

Supplementary Note For Calibration:

1. Preparation of oxygen free water


Configure 250ml of 5% sodium sulfite solution, you can add the right amount of cobalt
chloride as a catalyst.

2. Preparation of saturated dissolved oxygen water


Take 300 to 500 ml of distilled water, at a relatively stable temperature, use an air
pump to aerate the distilled water for at least 15 minutes.

Website : www.truesense.net.in ; www.truesense.in


Email : sales@truesense.in ; truesense.sales@gmail.com
Calibration :

1. Rinse the electrode well with DI water or rinse solution.


2. Press ON/OFF to turn on the machine. Dip the electrode into the zero oxygen solution for
at least 3 minutes, and keep shaking gently until the display value stabilizes.
3. Press and hold the ON/OFF key until the second line of display shows "CAL" character,
continue to shake the electrode gently, when the display value is stable, press [% / L/mg]
key, it shows DO value 0%, complete the zero calibration.
4. Clean the electrode, dip the electrode into the saturated dissolved oxygen solution and
keep shaking the electrode gently until the displayed value is stable.
5. Shake gently and press ON/OFF key until the second line of the display shows "CAL"
character, then press ℃/°F key, the DO value of 100.0% is shown at this time, the slope
correction is finished.

Website : www.truesense.net.in ; www.truesense.in