Professional Documents
Culture Documents
Quality Control Procedure (Sectra-Branded) - Philips MicroDose L30 SW 8.4 8.4P1
Quality Control Procedure (Sectra-Branded) - Philips MicroDose L30 SW 8.4 8.4P1
00
Quality Control Procedures
As part of our commitment to providing the highest quality products and services, we would
like to encourage feedback on the quality of this documentation.
If you think we can improve this document in any way, please send your suggestions to
info.imtec@sectra.se
Date 2010-10-22
Contents
1. Introduction..................................................................................................................... 5
1.1 Purpose ...................................................................................................................... 5
1.2 Scope ......................................................................................................................... 5
1.3 Acceptance testing .................................................................................................... 5
1.4 General information about quality control procedures ............................................. 5
1.5 Intended readers ........................................................................................................ 6
1.6 Related documents .................................................................................................... 6
1.7 References ................................................................................................................. 6
1.8 Symbols used in this document ................................................................................. 7
1.9 Terms, definitions and typographic conventions ...................................................... 8
2. Equipment and preparatory information..................................................................... 9
2.1 Equipment required for procedures performed at least monthly............................... 9
2.2 Equipment for procedures performed less frequently than once a month ................. 9
2.3 Sending to PACS of For Processing (unprocessed) images .................................... 11
2.4 Copy of For Processing (unprocessed) images to external memory ....................... 11
3. Daily (User).................................................................................................................... 12
3.1 Daily quality control ............................................................................................... 12
3.2 Guidance system daily procedure ........................................................................... 14
4. Weekly (QC Technologist) ........................................................................................... 15
4.1 Full calibration ........................................................................................................ 15
5. Monthly (QC Technologist) ......................................................................................... 17
5.1 Spatial resolution..................................................................................................... 17
5.2 Contrast-to-noise ratio............................................................................................. 19
5.3 Visual checklist ....................................................................................................... 22
6. Semi-annual (Medical physicist).................................................................................. 26
6.1 Thickness indication ............................................................................................... 26
6.2 Compression force .................................................................................................. 28
6.3 X-ray tube output (Air kerma) ................................................................................ 30
6.4 Air kerma reproducibility........................................................................................ 32
6.5 AEC system: Breast thickness and exposure .......................................................... 33
7. Annual (Medical physicist) .......................................................................................... 37
7.1 AEC system: Density compensation ....................................................................... 37
7.2 Tube Voltage ........................................................................................................... 39
7.3 Half value layer ....................................................................................................... 40
7.4 Contrast detail resolution ........................................................................................ 42
7.5 Detector linearity..................................................................................................... 44
7.6 Image field and x-ray field agreement .................................................................... 46
7.7 Missed tissue at chest wall ...................................................................................... 50
7.8 Guidance system precision control ......................................................................... 51
8. Unscheduled calibrations (User).................................................................................. 53
8.1 Quick calibration ..................................................................................................... 53
9. Summary of Quality control procedures and Tests ................................................... 55
Appendix A – Weekly record ............................................................................................... 58
Appendix B – Monthly record .............................................................................................. 59
1. Introduction
1.1 Purpose
This document, delivered together with Sectra MicroDose Mammography L30, describes
Quality Control Procedures that have to be performed on regular basis.
1.2 Scope
The information in this document applies to the Sectra MicroDose Mammography L30.
This document describes all necessary quality control procedures that shall be performed in
order to secure a stable operation and verify that the product comply with regulatory
requirements. Country specific regulatory requirements may exist.
This document contains all the quality control procedures for the L30, which shall be
performed to verify that the system fulfils the specification for image and radiation quality.
The procedures require special equipment for dosimetry and radiation quality control. This
equipment is not delivered with the system.
Chapter 3-5 in this document contains quality control procedures that shall be performed at
least once a month. Since these procedures can be performed by a regular user and a QC
Technologist, the content of chapter 3-5 is also included in the User Manual [1].
Appendix A, B and C contain pre-printed records for the weekly, monthly, semi-annual and
annual measurements.
The term “quality control” has been applied liberally in this document and contains some
elements that may be considered calibration procedures. These calibrations are an important
part of the quality assurance.
The procedure for handling any deviations is described in each test case. If any doubt, contact
the supplier.
When the user is instructed to create a new examination while performing a test it is
recommended that a “patient” named “default”, “test” or similar is used in order to
avoid confusion.
For extended sequences of measurement exposures it is recommended to keep the
anode rotating while inside an examination. In the Tools menu in the application, select
Service Tool - Anode brake settings. Uncheck Brake anode and click OK. Be sure to
restore the original setting after the measurement session is completed.
1.7 References
The quality control procedures follow certain international guidelines and standards [2]-[6].
When possible, the same or similar tests are performed as for film-based mammography. In
some cases, these must be modified for a digital system. They also include tests that are only
relevant to digital mammography systems. In these cases, [2] has served as a guide.
[2] European Guidelines for Quality Assurance in Breast Screening and Diagnosis, 4th
Edition, 2006.
[3] International standard, IEC 61223-2-10, Evaluation and routine testing in medical
imaging departments – Part 2-10: Consistency tests – X-ray equipment for
mammography, 1999.
[4] MQSA – Mammography Quality Standards Act, U.S. Public Law 102-539, U.S.
Public Law 105-248.
[5] Mammography Quality Control Manual, American College of Radiology, 1999.
[6] European Protocol on Dosimetry in Mammography, EUR 16263 EN, 1996.
Caution! Caution is used when danger for patient, personnel or system exists.
Exposure
Table 2 Equipment needed for quality checks performed less frequent than monthly.
Aluminium filter Purity: > 99.9%. Thickness accuracy Approx. 4 Not specific.
<3%.
A number of 0.5, 0.2, 0.1 and 0.05
mm filters with a minimum area of 4
cm2. It should be possible to vary the
thickness between 0.3 and 0.7 mm in
0.05 mm steps.
Contrast detail phantom PMMA block that contains discs of 1 CDMAM 3.4
varying thickness and diameter
Compression force Accuracy: < 5%. Range up to at 1 Not specific
measurement tool least 300 N.
3. Daily (User)
The calibrations and quality controls in this chapter are to be performed by the user on a
regular basis. This chapter is also included in the User Manual [1] in each local language.
The second part of Daily quality control (i.e. Image Quality) should also be performed when a
Quick calibration (Section 8.1) or a Full calibration (Section 4.1) has been performed.
Performed by
User.
Frequency
Before the first examination of the day.
If the patient support or the collimator collides with something and the
user suspects that the part moved as a result of this collision, a new Daily
Notice quality control shall be performed.
If the result of the control is not satisfactory, stop performing
examinations. The user should contact technical support immediately.
Procedure
Compression Paddle
Visually check all compression paddles for scratches or sign of wear. The compression
paddles must not have scratches, cracks or other defects that may distort the image or harm
the patient.
Image Quality
1. The standard or high edge compression paddle must be used.
2. Select the Daily quality control menu option in the application (or press the Quality
Control shortcut key on the keypad). Daily quality control examination automatically
opens.
3. Place the Daily quality phantom on the patient support. The phantom has a metal
guiding edge. Slide the phantom on the patient support so that the guiding edge of the
phantom reaches the front end of the patient support. The contrast markers should be
on the chest wall side; see Figure 1.
Figure 1 How to place the Daily quality phantom on the patient support.
4. Follow the displayed instructions and lower the compression paddle till it stops
against the phantom.
5. Press Compression done.
6. Perform an exposure with the default settings.
7. Study the acquired phantom image. Search for vertical or horizontal lines or other
irregularities in the image. Pay no attention to the added contrast markers. Check that
the image is homogeneous and without artifacts. Document any findings in the
examination notes.
8. If no disturbing irregularities are found, the system is ready for patient examinations.
The user must approve the image actively by pressing the Approve button. Then an
additional dialog box will appear. Here, the user should approve or disapprove the
image.
(Pressing the Approve button in the dialog box will approve the image and send it to
PACS, pressing the Disapprove button will disapprove the image and send it to
PACS and the Cancel button will cancel the operation.)
Expected results
The compression paddles should not have scratches or other defects that can affect image
quality or harm the patient.
The acquired phantom image should not have any disturbing irregularities.
Handling irregularities
If scratches or other defects are discovered on a compression paddle, it may no longer be used
for examinations and must be replaced.
If the image is unacceptable, perform a Quick calibration (Section 8.1), and then perform a
new Daily quality control.
Repeat this cycle (Quick calibration + Daily quality control) up to two times if the system still
does not pass the Daily quality control.
If this fails, a Full calibration (Section 4.1) must be performed followed by a Daily quality
control. If the image still is unacceptable upon completion of the calibrations, do not perform
any examinations. Report the irregularity to technical support.
Performed by
User.
Frequency
If the guidance system is used, perform this procedure before the first Wire Localization
examination of the day.
If the user suspects that guidance system has been exposed to undesired
forces, e.g. guidance system has been dropped to the ground, a Guidance
system precision check shall be performed according to section 7.8 in this
Notice document.
If the result of the control is not satisfactory, stop performing
examinations. The user should contact technical support immediately.
Procedure
Guidance arm
Visually check the Guidance arm. The guidance arm, including the rounded mounting
surface, must not be physically damaged or deformed.
Laser unit
Visually check the Laser unit. The laser unit must not be physically damaged or deformed.
Check that the laser unit is properly attached to the laser unit magnet foot.
Power on the laser unit and check that the laser beam is lit and properly reflected by the
mirror.
Expected results
The guidance arm or laser unit must not be physically damaged or deformed. The laser beam
must be properly projected.
Handling irregularities
If guidance arm or laser unit is physically damaged or deformed, it may no longer be used for
examinations and must be replaced. Report the irregularity to technical support.
If laser unit is not lit, change the battery according to instructions in the User manual.
Performed by
QC Technologist.
Frequency
A Full calibration is to be performed weekly, preferably Monday morning, before the first
examination. (The exact day for the weekly Full calibration can be changed depending on
local conditions, but performing this calibration once each Monday morning is the
recommended schedule.)
This calibration can also be necessary to perform in case of special circumstances (See
“Handling irregularities” in section 3.1) or upon recommendation of the supplier’s technical
support.
If a Full calibration is not performed within two weeks, the system will not
Notice
allow image acquisition without performing a Full calibration.
Figure 2 Calibration phantom placed on the patient support (seen from front)
4. Follow the displayed instructions and lower the compression paddle till it stops
against the phantom.
5. Press the Compression done button.
6. Wait until the system is ready for exposure (the exposure button is lit).
7. Press and hold the exposure button until the X-ray warning symbol (shown in the
lower part of the dialog) has appeared and disappeared.
8. The system will now process the data of the phantom image for a few moments.
9. Repeat step 6 to 8 above by following the on screen instructions. Please note that
pressing the Compression done button is not required between exposures.
Expected results
The expected result of a normal Full calibration is that the calibration procedure is performed
without any errors or warnings.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
QC Technologist.
Frequency
Performed once a month.
Procedure
1. Place the line pair phantom on top of its 4.5 cm base.
2. The line pair phantom should be laterally centered and the centre of the phantom’s
pattern approximately 6 cm from the chest wall. The bars should be slightly slanted
(2-5 degrees) with respect to the chest wall edge. See Figure 3 below.
9. Use zoom and adjust window setting to optimize contrast and brightness.
10. For each direction there are several groups of lines. The density of the lines increases
gradually.
For each direction start with the group with the widest lines and go to narrower (and
denser) lines until the space between the lines in a group is no longer discernable. The
groups with the narrowest visible patterns correspond to the systems linear resolution
(1) along and (2) perpendicular to chest wall.
Note the greatest visible line density in each direction and enter it in the record. See
example image in Figure 4 below.
Expected result
In both directions, the visible line density should be at least 6 lp/mm.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
QC Technologist.
Frequency
Performed once a month.
Procedure
1. Place the Daily quality phantom on the patient support. The phantom has a metal
guiding edge. Slide the phantom on the patient support so that the guiding edge of the
phantom reaches the front end of the patient support. The contrast markers should be
on the chest wall side; see Figure 5 below.
Contrast markers
(3 light and
3 dark)
Figure 5 How to place the Daily quality phantom on the patient support.
2. Choose “Daily quality control” from the Edit menu in the application or press the
shortcut key on the keypad.
3. Follow the displayed instructions and lower the compression paddle till it stops
against the phantom.
4. Press Compression done.
5. Acquire an image.
6. Right click on the image and select “Relabel\Raw”
Tip: If necessary, it is possible to adjust window level by pressing and holding the
scroll wheel while moving the mouse.
Expected result
Depending on the tube voltage used, the acquired CNR value must be multiplied with a factor
in order to compare acquired CNR and reference CNR. Note that the reference CNR should
also be multiplied with the factor.
These are the factors to use:
26 kV: 1.13
29 kV: 1.06
32 kV: 1
35 kV: 0.95
38 kV: 0.90
The sum of the CNR-values should be within ±10% of the reference CNR. The reference
CNR is to be measured at installation in conjunction with the contrast-detail resolution test
(section 7.4). The reference CNR is measured in the same way as described above, but should
be averaged over at least three images.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
QC Technologist.
Frequency
Once a month.
Procedure
Review each item on the visual checklist and indicate the status.
Date and initial the checklist when indicated.
Expected result
All items are expected to pass the test and receive a check mark.
Handling irregularities
If any item does not pass the test, the source of the problem must be identified and corrective
action must be taken before the next examination is performed. Any item missing from the
room must be replaced immediately. Malfunctioning equipment shall immediately be reported
to technical support and either be replaced or repaired before the next examination.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once every six months.
Procedure
1. Create a new examination in the application.
2. Position the compressible material centred laterally and with one edge along the chest
wall edge (so that it simulates a breast positioned in the CC view).
3. Compress and then measure the distance, 'x' in Figure 9, between the patient support
and the compression paddle using the vernier calliper.
20–30 mm
40–60 mm
70–90 mm
Table 3 Average error in thickness indication
Expected result
For each measurement, the difference between the indicated and the measured breast
thickness should be less than or equal to 3 mm. If a measurement indicates a bigger difference
than 3 mm, try again with the same combination of thickness and force in order to make sure
it was not due to an error.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
Medical physicist or Sectra SE. (This test can also be performed by QC Technologist.)
Frequency
Performed twice a year.
Procedure
Testing the systems Compression force indication
1. Place the compression force measurement tool on the patient support. Protect the
patient support from scratches.
2. Place the compressible material between the compression force measurement tool and
the compression paddle.
4. Look at the dial of the compression force measurement tool and note at which force
the compression was disengaged.
Expected result
The largest difference between indicated and measured force must be less than 20 N.
During compression maintained for a minute, the measured compression force may not
fluctuate more than 5 N.
It should not be possible to compress more than 20 daN.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once every six months.
Procedure
See [6] for general information on dosimetry. Air kerma 4.5 cm above the patient support is
to be measured for all five clinically available tube voltages and for three mAs values. (I.e. a
total of 15 measurements.)
Before exposure
1. Place the dose meter on the patient support, laterally centered and 6 cm in
from the edge of the chest wall. Make sure it is secured. Note the height above the
patient support.
2. Create a new examination in the application.
3. Use manual exposure parameters.
4. Select Service Tool – Anode brake settings in the Tools menu. Uncheck Brake anode
and click OK.
Exposure
5. Do the following for each value of tube voltage:
a. For three mAs values (minimum, maximum and one in between):
Place the compression paddle at about 9 cm height from the patient support.
Note the compression paddle must not be removed.
Reset the dose meter, perform an exposure and read the measured air kerma.
b. Calculate the average of the three air-kerma measurements in units of mGy/mAs.
6. Use the inverse square law to rescale the values to a point 4.5 cm above the patient
support. I.e. if the reference plane of the dose meter is D mm above the patient
support, the measured value should be scaled by a factor (source to patient support
distance is 640 mm.1)
1
The position of the focal spot is marked on the tube housing. In order to view this, the right C-arm
cover must be removed.
Expected result
These values shall not differ by more than ±15% from the values estimated by the system,
which can be found in the Auto exposure control status window. See example of values in
Figure 10.
The tube output at 32 kV shall be at least 0.11 mGy/mAs.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result
is not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once every six months.
Procedure
1. Place the dose meter on the patient support, laterally centered and 6 cm in
from the edge of the chest wall. Make sure it is secured.
2. Create a new examination in the application.
3. Use manual exposure parameters. Choose 32 kV.
Select an intermediate mAs-value.
4. Place the compression paddle at about 9 cm height from the patient support.
Note: The compression paddle must not be removed.
5. Reset the dose meter, perform an expose and note the measured air kerma.
6. Repeat 5 times for the same mAs and tube voltage.
7. Calculate the mean and standard deviation of the six measured air kerma values.
Enter the values in the record.
8. Calculate the relative standard deviation (= standard deviation / mean).
Expected result
The relative standard deviation of the series of air kerma values should be less than 5%.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once every six months.
Procedure
Please note that the PMMA plates used for this procedure are not
equivalent to a typical breast. In addition, a flat non-compressible
large object will introduce a bias in the thickness measurement.
We therefore need to override the measured compression height
with the equivalent breast thickness that the PMMA simulates.
By choosing PMMA 20, PMMA 30 etcetera in the drop-down
menu (Figure 11) in the application the equivalent breast thickness
tabulated in [2] are used. Figure 11 Drop-down menu
Assessment of exposure parameters
1. Create a new examination in the application with Examination Code “QACODE1”
2. Place 20 mm PMMA laterally centered on the patient support.
3. Select the image type Phantom and select the PMMA 20 option in the drop-down
menu.
4. Choose the SmartAEC option.
5. Lower the compression paddle and press Compression done.
6. Acquire an image.
7. Note the used exposure parameters displayed in the image (tube voltage, mAs).
8. Add 10 mm of PMMA and repeat steps 3 to 8 until 70 mm of PMMA has been used
(choose PMMA30 for 30 mm of PMMA etc)
Tip: If necessary, it is possible to adjust window level by pressing and holding the scroll
wheel while moving the mouse.
10. A result box is shown in connection with the ROI pair. Note down the CNR value in
the record.
11. Repeat step 4-10 for 30 to 70 mm PMMA, by adding PMMA above the filter in 10
mm steps.
12. Normalize the values by dividing with the result for 50 mm PMMA.
2
For the C120 dose configuration, the value is 85%.
3
For the C120 dose configuration, the value is 70%.
AGD calculation
The AGD - that a corresponding breast of normal composition should have received - can be
calculated. Use the HVL measured in the section Half value layer (Section 7.3) or take the
values from the last record. Take Air kerma values from section X-ray Tube Output (Section
6.2).
Details on dose calculation and tables for converting input dose to AGD are found in [2] and
[6]. Also remember to use breast thickness and not PMMA thickness.
Enter the calculated AGD value for each thickness in the record.
Also enter the systems AGD value for each thickness displayed in the acquired images.
Expected result
The measured relative CNR values must be above the CNR threshold in Table 4.
The limiting values for AGD must be below the achievable AGD as specified in [2] and
found in Table 4.
The calculated AGD value for each thickness must not deviate more than +/- 10% from the
systems AGD value displayed in the corresponding image.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected
result is not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
1. Create a new examination in the application with Examination Code “QACODE1”
2. Place 30 mm of PMMA on the patient support and put the compression paddle at a
height of approximately 9 cm.
3. Choose the SmartAEC option. Select the image type Phantom and select
the PMMA 40 option in the drop-down menu.
Use PMMA 40 throughout the whole test. The system should always
Notice assume a mean composition and compensate for the variation in
glandularity (modeled by a change of thickness).
4. Acquire an image.
5. Right click on the image and select “Relabel\Raw”
7. A result box is shown in connection with the ROI pair. Note down the ROI 1 (Signal)
SNR value in the record.
8. Add 5 mm PMMA. Repeat step 3–7 until 50 mm of PMMA has been exposed.
9. Enter all the SNR values, representing 30 to 50 mm of PMMA, in the record.
Expected results
All measured SNR values must be within ±15% from the average value.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected
result is not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
Tube voltage is to be measured at all tube voltages used clinically. Place the kVp meter on the
patient support laterally cantered about 6 cm from the front edge of the patient support.
Consult the manual of the kVp meter for correct handling.
Accuracy
1. Create a new examination.
2. Choose manual exposure parameters. Use an arbitrary mAs-value between
min and max.
3. Measure the voltage for each tube voltage used clinically.
4. Calculate the accuracy for each voltage.
Accuracy = | kVmeasured – kVset |.
5. Average the accuracy of all selectable tube voltages. Enter the value in the record.
Reproducibility
1. Create a new examination.
2. Choose manual exposure parameters. Use an arbitrary mAs-value between
min and max.
3. Choose one of the available tube voltages.
4. Measure the tube voltage at least six times.
5. Calculate the reproducibility, defined as the standard deviation of the measurement
series. Enter the value in the record.
Pay attention to that the scanning can influence the measurement. Meters
use several detectors with different filtration, which may not be exposed
Notice
simultaneously depending on how the detector is placed. Be sure that the
kVp meter can handle this effect.
Expected result
Reproducibility better than 0.5 kV.
Accuracy better than 1 kV.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected
result is not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
HVL is to be measured for all clinically available tube voltages.
Before measurement
The compression paddle should be included.
The dose meter should be positioned similar as when measuring X-ray Tube Output
(Section 6.2).
“Good geometry” must be used in order to avoid scattered radiation reaching the
detector. I.e. the radiation must be collimated at or before the compression paddle. To
achieve this, put the compression paddle at a height of about 9 cm above the patient
support.
Put the lead sheet on top of the compression paddle.
Acquire an image to verify that the collimator is lined up with the dose meter. See
Figure 15 for an example of good line-up.
c. Place the compression paddle at about 9 cm height from the patient support.
Expose.
d. Repeat step “b” and “c” but choose filters that reduce the Air kerma to just
below 50%.
Calculate the half value layer with the following formula
X 1 ln 2 K 2 / K 0 X 2 ln 2 K 1 / K 0
HVL
ln K 2 / K 1
K0 is the Air kerma without filter. K1 and K2 respectively represents Air
kerma with filter. X1 and X2 are the filter thickness above and below 50%
respectively.
4. Note the HVL values in the record.
Figure 15 Example of good line-up of dose meter (ion chamber) and lead collimator.
Expected result
The measured HVL-value should not deviate more than ±0.03 mm from the reference value.4
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected results are
not met.
4
± 0.03 mm is a more strict limit than ref [5] where the requirement is defined as
kV/100+0.03 < HVL (mm) < kV/100 + C. C=0.3given for W/Rh anode filter combination. There is no
value for W/Al, therefore the value for W/Rh is used.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
The phantom, including its 4 cm PMMA base, corresponds to 5 cm PMMA and 6 cm
breast [2].
1. Place the CDMAM phantom with its base of 4 cm PMMA so that the long side meets
the edge of the image field at the chest wall, see Figure 16. The phantom should be 2
cm above the patient support, i.e. placed on 2 cm PMMA with 2 cm on top. The
largest discs should be to the left seen from the patient’s position.
2. Choose PMMA 50 from the drop-down menu in order to override the system’s
measured breast thickness with the equivalent breast thickness of 50 mm PMMA as
specified in [2].
3. Study the image on the diagnostic workstation using clinical reading conditions. Use
zoom and adjust window levels (contrast and brightness) in order to maximize
visibility of the details in the displayed images. A minimum of two observers should
score three images each.
4. For each image:
a. For each disc diameter (1.60, 1.00, 0.63, 0.40, 0.25 and 0.1 mm):
Start with the thickest disc. There is a disc in the centre and one in one of the four
corners. Try to determine in which corner the disc is located.
When you no longer judge it possible to determine in which corner the disc is located,
proceed with at least one more disc.
Compare to the proof of the phantom.
Note the thickness of the thickest disc that could not be identified correctly.
5. For each diameter:
Average the threshold thickness from the different images and observations.
Note the value in the record.
Expected result
The threshold thickness shall be lower than that required in table for 5 of 6 of the diameters
for the C100 and C120 configuration.
Diameter Required thickness
(mm) threshold (µm) for
C100 and C120 dose
configurations
1.6 0.074
1.0 0.091
0.63 0.123
0.4 0.189
0.25 0.352
0.1 1.68
Table 5 Required threshold thickness for the C100 and C120 configuration for different disc
diameters. The C100 and C120 values correspond to the acceptable limiting values as
specified in [2].
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
This test verifies that the squared signal-to-noise ratio is linear with the input dose (mAs).
Since the tube current is constant, this is achieved by varying the scan time.
1. Create a new examination in the application with Examination Code “QACODE1”
2. Place the Daily quality control phantom laterally centered on the patient support,
see Figure 5 on page 19.
3. Follow the displayed instructions and lower the compression paddle till it stops
against the phantom.
4. Press Compression done.
5. Determine the min and max mAs-values for the system by choosing manual exposure
and 32 kV in the application. The minimum and maximum values can be found in the
drop-down menu. Note the values.
6. Acquire an image using “phantom,” 32 kV and minimum mAs.
7. Right click on the image and select “Relabel\Raw”.
8. Right click on the image and select "Measurements\CNR measurement”, see Figure
17 for a reference.
a. Place ROI 1 (Signal) right between the top two discs (do not resize the ROI)
b. Place ROI 2 (Background) anywhere on the image (it will not be used)
Figure 17 Placing the ROIs, not that the background ROI placement is not
important
9. A result box is shown in connection with the ROI pair. Use the ROI 1 (Signal) SNR
value to calculate SNR2 and note down this value and the mAs used (will be used for
later calculations).
10. Return to step 5 (above) and repeat this procedure for maximum mAs and at least five
evenly distributed mAs-values between min and max. (I.e. total of 7 measurements.)
11. The squared correlation coefficient is defined by (using y = SNR2 and x = mAs)
2
n xi y i xi y i
R2 i i i
2
2
n x i xi n y i y i
2 2
i i i i
where n is the number of measurements.
Functions for calculating the correlation coefficient are included in most spreadsheet
programs.
Note R2 in the record.
Expected result
R2 > 0.99
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
Medical physicist (or Sectra SE).
Frequency
Annually, when changes in image and x-ray field congruence may be suspected, or after
preventative maintenance or service which may impact image and x-ray field congruence.
Procedure
Congruence of indicated and actual image field
1. Center coins on each of the four edges of the indicated image field on the breast
support. (The indicated image field is outlined by the painted area on the side and rear
edge of the field.)
2. Place the coins inside the indicated field and tangential to the edge (Figure 18).
Double-sided tape may be used to hold the coins in place if they tend to slide toward
the center.
Place at least 0.5 cm PMMA on top of the breast support (this allows the actual image
border to be seen on the mammographic image).
Figure 18 Photograph of top surface of patient breast support and showing placement
of coins and painted outline of the indicated image field.
Figure 19 Radiograph of coins showing appropriate alignment with the edge of the
actual image field.
8. Remove the acrylic sheet.
9. Measure the difference between the outline of the indicated and actual image fields
(indicated by the edges of the coins) on the surface of the patient breast support using
a ruler. (Figure 20)
Figure 22 X-ray and image field congruence test for chest wall edge.
The radiation field should not extend beyond the actual image field by more than 5 mm on the
right and left sides, and along the rear of the image receptor, i.e., the intensifying screens
should not produce light during the scan.5
No radiation must be visible beyond the chest wall edge of the patient support.
If any of the above criteria are not met, contact the service engineer for Corrective Action.
5
MQSA regulations (MQSA, 1999) require 2% of the SID which is equivalent to 13 mm.
Performed by
Medical physicist (or Sectra SE).
Frequency
Performed once a year.
Procedure
1. Use the image from the “Image field and x-ray field agreement” procedure
(Section 7.6) or acquire an image following the instructions found there.
2. On the chest wall side, estimate the difference between the indicated image
field and the actual image field (should be possible within 0.25 mm).
3. Use the vernier calliper to measure the distance from the edge of the patient
support to the edge of the indicated (coloured) image field (should be possible
within 0.5 mm).
4. Together, the two values above indicate the distance from the edge of the
patient support to the edge of the actual image field, i.e. how much tissue is
missed during imaging. Record this value in the record.
Alternative procedure
Use a phantom including objects specifically designed to measure chest wall dead area (e.g.
small steel balls). Follow the instructions given by the phantom manufacturer.
Expected result
The level of missed tissue at the chest wall should be less than 5 mm.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected
result is not met.
Performed by
QC Technologist.
Frequency
If the guidance system is used, perform this procedure annually or when the user suspects that
guidance system has been exposed to undesired forces, e.g. if the guidance system has been
dropped to the ground.
Procedure
1. Create and open a new examination in application.
2. Mount the matrix compression paddle and select matrix compression paddle in
application.
3. Adjust the compression height to 75 mm.
4. Mount the Guidance system.
5. Position the guidance laser unit to point at the tip of the “A” marker line in the
compression paddle coordinate system.
6. Remove the matrix compression paddle.
7. Position the radio opaque marker on the patient support so that the guidance laser unit
is pointing at the tip of this position marker too. Secure the marker position with tape.
8. Re-mount the matrix compression paddle and verify that the laser beam still point at
the tip of the “A” marker line in the compression paddle coordinate system. Seen
from above the “A” marker line and the radio opaque marker shall point towards each
other.
9. Repeat step 5 to 8 for the “F” marker line on the opposite side of the compression
paddle coordinate system.
10. Remove guidance system.
11. Press Compression done.
12. Perform an exposure.
13. Magnify the acquired x-ray image and measure the distance between the tip of the
marker lines and the tip of the radio opaque markers with the measurement tool.
Figure 1: How to align the marker lines and the radio opaque markers
Expected result
The distance between the tip of the marker lines and the tip of the radio opaque markers shall
be less than 1.5 mm.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
Performed by
User.
Frequency
This calibration routine should only be started if the Daily quality
control is unsatisfactory or if a disturbing image artefact is detected
during normal operation.
Procedure
1. The standard or high edge compression paddle must be used.
2. Select the menu option Quick calibration in the application (or press the Quick
calibration shortcut key on the keypad).
3. Instructions appear on the screen describing how to place the Calibration phantom on
the patient support, see Figure 23. The phantom has a metal guiding edge. Slide the
phantom on the patient support so that the guiding edge of the phantom touches the
front end of the patient support.
Figure 23 Calibration phantom placed on the patient support (seen from front).
4. Follow the displayed instructions and lower the compression paddle till it stops
against the phantom.
Expected results
The expected result of Quick calibration is that the calibration procedure is performed without
any errors or warnings.
Handling irregularities
Report all irregularities to technical support. Discontinue operation if the expected result is
not met.
5.1 Spatial QC Tech Study an image Minimum: pattern with Contact support
resolution of the line pair density 6 lp/mm must be Discontinue
phantom visible. operation.
10 min
each
5.2 Contrast- QC Tech Measure the CNR within ±10% of Contact support
to-Noise Ratio CNR using ref. CNR measured at Discontinue
contrast discs installation operation
in the Daily
quality
phantom
5.3 Visual QC Tech Visual check of 20 min Error-free execution Contact support
checklist the system Discontinue
operation
Semi-annual procedures
Measure distance between
6.1 Thickness Medical compression paddle and patient
Difference < 3 mm
indication physicist support using vernier calliper and
compare to indicated value
Less than 20 N deviation from
indicated force.
6.2 Compression Medical Compression force maintained
Measure force with scale
force6 physicist for one minute.
Compression force may not
exceed 200 N.
6.3 X-ray Tube
Medical Measurement with dose meter, e.g. Tube output at 32 kV shall be
Output
physicist ionization chamber > 0.11 mGy/mAs
(Air kerma)
6.4 Air kerma Medical Measure air kerma for six exposures
<5 % relative standard deviation
reproducibility physicist using the same exposure parameters.
6.5 AEC system: Medical CNR measurement using of Al filter
Breast thickness on PMMA for different PMMA See Table 4.
physicist
and exposure thicknesses.
Annual procedures
7.1 AEC system:
Medical SNR measurement for different ±15% from the average
Density
Pysicist PMMA thicknesses. SNR value.
compensation
Inaccuracy < 1.0 kV
Medical
Measurement using kVp meter at: Reproducibility < 0.5 kV
7.2 Tube Voltage lowest, typical and highest clinical (Complies to FDA regulation
physicist
tube voltages 900.12 (e)(5)(ii). Tolerance +/-
5%.)
HVL > kV/100+0.03
7.3 Half value Medical
Standard procedure HVL within 0.03 mm from the
layer physicist
reference value.
7.4 Contrast detail Medical Evaluate image quality using
See Table 5.
resolution physicist CDMAM 3.4 phantom
7.5 Detector Medical Measure SNR in ROI for different Correlation mAs-SNR2
linearity physicist mAs values R2 > 0.99
6
Can also be performed by a QC Technologist.
Unscheduled procedures
Note: In addition to the above-described tests for medical physicists, inspection of the
equipment is performed annually by Sectra SE for maintenance purposes.
Comments:
Perform Daily
Tuesday Morning
quality control
Perform Daily
Wednesday Morning quality control
Perform Daily
Thursday Morning
quality control
Perform Daily
Friday Morning
quality control
Micro-califications
Masses
Total score
7
The serial number of the stand can be found (1) in documentation delivered with the system, (2) on a
label on the door of the side cabinet or (3) in the “About box” found in the application.
Comments:
Contrast-to-Noise Ratio
Measured Measured Passed
Parameter Sum
in ROI 1A in ROI 1B (OK / not OK)
Mean (m)
NA No requirement
Standard
NA No requirement
deviation (s)
CNR
Reference CNR NA NA
Conversion
NA NA
factor used
Spatial resolution
Greatest visible line density Passed
Line direction
(per mm) (OK / not OK)
Along chest wall
Perpendicular to chest
wall
8
The serial number of the stand can be found (1) in documentation delivered with the system, (2) on a
label on the door of the side cabinet or (3) in the “About box” found in the application.
Visual checklist
Installation: Facility:
Comments:
Month:
Day:
Time:
Initials:
C-arm
(7) C-arm
Stand
Stand
9
The serial number of the stand can be found (1) in documentation delivered with the system, (2) on a
label on the door of the side cabinet or (3) in the “About box” found in the application.
AW table
Comments:
Equipment used
Type Specification Model/serial number
Semi-annual tests
Passed
Name Result Comments
(OK / not OK)
Thickness Average error: mm
indication
4.5 cm above patient support
26 kV: _______ mGy/mAs
29 kV: _______ mGy/mAs
X-ray tube output
(Air kerma) 32 kV: _______ mGy/mAs
35 kV: _______ mGy/mAs
38 kV: _______ mGy/mAs
Used kV:
Used (max) mAs:
Air kerma Reproducibility: %(max mAs)
reproducibility
Used (min) mAs:
Reproducibility: %(min mAs)
10
The serial number of the stand can be found (1) in documentation delivered with the system, (2) on a
label on the door of the side cabinet or (3) in the “About box” found in the application.
Annual tests
Passed Commen
Name Result
(OK / not OK) ts
Height SNR
(mm PMMA) Comments:
30
AEC system: 35
Density 40
compensation
45
540
Passed: (OK / not OK)
Reproducibility: kV
Tube voltage
Accuracy: kV
Including compression paddle
26 kV: _______ mm Al
29 kV: _______ mm Al
HVL
32 kV: _______ mm Al
35 kV: _______ mm Al
38 kV: _______ mm Al
kV: mAs:
D Mean Required Pass
(mm) threshold threshold (y/n)
thickness thickness
(µm) (µm) for
C100/
C120
Contrast detail 1.6 0.074
resolution 1.0 0.091
0.63 0.123
0.4 0.189
0.25 0.352
0.1 1.68
Detector linearity R2 =
Guidance system
mm
precision control
www.sectra.com/medical
© Sectra 2010