HI TECH CALIBRATION SERVICES, NO.

209, VGP NAGAR, MUGAPPAIR WEST,


Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 1 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Permanent Facility
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
1 Sourcing 1 A to 10 A 0.07 % to 0.1 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
2 Sourcing 10 µA to 100 µA 0.26 % to 0.064 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
3 Sourcing 100 µA to 100 mA 0.064%
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 2 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
4 Sourcing 100 mA to 1 A 0.064 % to 0.07 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
5 Sourcing 10 A to 20 A 0.1 % to 0.21 %
Current (< 1 5 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
6 Sourcing 1 A to 10 A 0.041 % to 0.1 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 3 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
7 Sourcing 10 µA to 100 µA 0.26 % to 0.035 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
8 Sourcing 10 A to 30 A 0.1 % to 0.14 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
9 Sourcing 100 µA to 100 mA 0.035%
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 4 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
10 Sourcing 100 mA to 1 A 0.035 % to 0.041 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 kHz
11 Sourcing 100 µA to 100 mA 0.091 % to 0.09 %
Current (< 1 to 30 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
12 Multimeter by Direct 10 µA to 100 µA 0.67 % to 0.25 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 5 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
13 Multimeter by Direct 100 µA to 3 A 0.25 % to 0.31 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
14 Multimeter with 3 A to 20 A 0.31 % to 0.35 %
Current (< 1 1 kHz)
Shunt by V/I Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating
15 AC Current (50 Hz) Multimeter with 20 A to 30 A 0.35 % to 0.48 %
Current (< 1
Shunt by V/I Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating
16 AC Current (50 Hz) Multimeter with 30 A to 100 A 0.48 %
Current (< 1
Shunt by V/I Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 6 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
17 Probe with DMM by 1 kV to 6 kV 1.9 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
18 Probe with DMM by 15 kV to 28 kV 5.8 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
19 Probe with DMM by 6 kV to 15 kV 1.9 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
20 Sourcing 1 mV to 10 mV 0.16 % to 0.049 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 7 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
21 Sourcing 10 mV to 100 mV 0.049 % to 0.012 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
22 Sourcing 100 mV to 1000 V 0.012%
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
23 Sourcing 1 mV to 10 mV 0.15 % to 0.038 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 8 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
24 Sourcing 10 mV to 100 mV 0.038 % to 0.008 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
25 Sourcing 100 mV to 100 V 0.008%
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
26 Sourcing 1 mV to 10 mV 0.48 % to 0.36 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 9 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
27 Sourcing 10 mV to 100 mV 0.36 % to 0.064 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
28 Sourcing 100 mV to 100 V 0.064 % to 0.064 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
29 Sourcing 1 mV to 10 mV 2.77 % to 2.36 %
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 10 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
30 Sourcing 10 mV to 100 mV 2.36 % to 1.27 %
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
31 Sourcing 100 mV to 10 V 1.27%
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (40 Hz to
32 Sourcing 100 V to 1000 V 0.009 % to 0.018 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 11 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
33 Multimeter by Direct 1 mV to 10 mV 2.46 % to 0.35 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
34 Multimeter by Direct 10 mV to 100 mV 0.35 % to 0.1 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
35 Multimeter by Direct 100 mV to 750 V 0.1 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating
36 Capacitance Sourcing 1 mF to 100 mF 0.082 % to 0.094 %
Current (< 1
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 12 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
37 Meter by Direct / 1 nF to 100 µF 0.06 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
38 Meter by Direct / 10 pF 0.23 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
39 Meter by Direct / 100 pF 0.06 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Inductance (@ 1
40 Meter by Direct & 100 µH to 10 H 0.08 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 13 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating 1 ohm to 10 kohm @
41 Resistance Meter by Direct / 0.062 % to 0.028 %
Current (< 1 1 kHz
Comparison Method
GHz)
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
42 Alternating 2.99 A to 20 A 0.08 % to 0.2 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
43 Alternating 3.29 mA to 329 mA 0.12 % to 0.06 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
44 Alternating 30 µA to 329 µA 0.53 % to 0.26 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
45 Alternating 329 µA to 3.29 mA 0.26 % to 0.12 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 14 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
46 Alternating 329 mA to 2.99 A 0.06 % to 0.14 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
47 Alternating AC Current (50 Hz) 5522A with Current 100 A to 1000 A 0.36 %
Current (< 1 Coil 5500A by Direct
GHz) (Source) Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
48 Alternating AC Current (50 Hz) 5522A with Current 20 A to 100 A 0.87 % to 0.36 %
Current (< 1 Coil 5500A by Direct
GHz) (Source) Method
ELECTRO-
Using Multiproduct
TECHNICAL- AC Power 1Ø (50 Hz)
Calibrator Fluke
49 Alternating - Lead / Lag ± 0.2 to 0.12 W to 4800 W 1 % to 0.07 %
5522A by Direct
Current (< 1 UPF
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Power Factor 1Ø Calibrator Fluke 0.0021 PF to 0.0013
50 Alternating ± 0.2 to UPF
(50 Hz) - Lead / Lag 5522A by Direct PF
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 15 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
51 Alternating 1 mV to 3 mV 0.71 % to 0.3 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
52 Alternating 3 mV to 30 mV 0.3 % to 0.044 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
53 Alternating 30 mV to 300 mV 0.044 % to 0.02 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
54 Alternating 300 mV to 300 V 0.02 % to 0.024 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
55 Alternating 300 V to 1000 V 0.024 % to 0.036 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 16 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
56 Alternating 0.39 nF to 0.6 nF 3.57 % to 2.5 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
57 Alternating 0.6 nF to 1 nF 3.57 % to 1.7 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
58 Alternating Capacitance Box by 1 nF to 100 µF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
59 Alternating 1 nF to 7 nF 1.7 % to 0.5 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
60 Alternating Capacitance Box by 10 pF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 17 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
61 Alternating Capacitance Box by 100 pF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
62 Alternating 7 nF to 330 nF 0.5 % to 0.4 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 100 Calibrator Fluke
63 Alternating 0.33 µF to 30 µF 0.47 % to 0.67 %
Hz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 100 Calibrator Fluke
64 Alternating 30 µF to 300 µF 0.67 % to 0.72 %
Hz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Inductance (@ 1
65 Alternating Inductance Box by 100 µH to 10 H 1.19 %
kHz)
Current (< 1 Direct Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 18 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
66 DIRECT DC Current Sourcing 0.01 µA to 0.1 µA 4.62 % to 0.47 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
67 DIRECT DC Current Sourcing 0.1 µA to 1 µA 0.47 % to 0.050 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
68 DIRECT DC Current Multimeter by Direct 0.1 µA to 10 µA 0.7 % to 0.16 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 19 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
69 DIRECT DC Current Sourcing 1 µA to 10 µA 0.050 % to 0.0071 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
70 DIRECT DC Current Sourcing 1 A to 10 A 0.023 % to 0.025%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0071 % to 0.0014
71 DIRECT DC Current Sourcing 10 µA to 100 µA
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 20 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
72 DIRECT DC Current Multimeter by Direct 10 µA to 100 µA 0.16 % to 0.09 %
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
73 DIRECT DC Current Sourcing 10 A to 30 A 0.025 % to 0.082 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0014 % to 0.0050
74 DIRECT DC Current Sourcing 10 mA to 100 mA
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 21 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
75 DIRECT DC Current Sourcing 100 µA to 10 mA 0.0014%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
76 DIRECT DC Current Multimeter by Direct 100 µA to 100 mA 0.09 % to 0.065 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
77 DIRECT DC Current Multimeter with 100 A to 500 A 0.06 % to 0.5 %
CURRENT Shunt by V/I Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
78 DIRECT DC Current Sourcing 100 mA to 1 A 0.0050 % to 0.023 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 22 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
79 DIRECT DC Current Multimeter by Direct 100 mA to 3 A 0.065 % to 0.28 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
80 DIRECT DC Current Multimeter with 3 A to 30 A 0.2 % to 0.06 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
81 DIRECT DC Current Multimeter with 30 A to 100 A 0.14 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
82 DIRECT DC Current Multimeter with 500 A to 1000 A 0.5 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using High Voltage
83 DIRECT DC High Voltage Probe with DMM by 1 kV to 6 kV 1.3 % to 1.6 %
CURRENT Direct method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 23 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using High Voltage
84 DIRECT DC High Voltage Probe with DMM by 15 kV to 40 kV 1.6 % to 2.6 %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using High Voltage
85 DIRECT DC High Voltage Probe with DMM by 6 kV to 15 kV 1.6 %
CURRENT Direct method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
86 DIRECT DC Resistance Multimeter by Direct 0.1 ohm to 1 ohm 4.6 % to 0.47 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
87 DIRECT DC Resistance Multimeter by Direct 1 Mohm to 10 Mohm 0.02 % to 0.05 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
1 ohm
88 DIRECT DC Resistance Multimeter by Direct 0.47 % to 0.06 %
to 10 ohm
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 24 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
10 Mohm to 100
89 DIRECT DC Resistance Multimeter by Direct 0.05 % to 0.93 %
Mohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
10 ohm
90 DIRECT DC Resistance Multimeter by Direct 0.06 % to 0.02 %
to 100 ohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 Mohm to 1
91 DIRECT DC Resistance Multimeter by Direct 0.93 % to 3.5 %
Gohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 ohm
92 DIRECT DC Resistance Multimeter by Direct 0.02 %
to 1 Mohm
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
93 DIRECT DC Voltage Sourcing 0.01 mV to 0.1 mV 2.35 % to 0.24 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 25 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
94 DIRECT DC Voltage Sourcing 0.1 mV to 1 mV 0.24 % to 0.024 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
95 DIRECT DC Voltage Multimeter by Direct 0.1 mV to 1 mV 4.1 % to 0.41 %
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
96 DIRECT DC Voltage Sourcing 1 mV to 10 mV 0.024 % to 0.0028 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
97 DIRECT DC Voltage Multimeter by Direct 1 mV to 10 mV 0.41 % to 0.05 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 26 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00036 % to
98 DIRECT DC Voltage Sourcing 1 V to 10 V
0.00032 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0028 % to 0.00063
99 DIRECT DC Voltage Sourcing 10 mV to 100 mV
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 DIRECT DC Voltage Multimeter by Direct 10 mV to 1000 V 0.05 % to 0.0054 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 27 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00032 % to
101 DIRECT DC Voltage Sourcing 10 V to 100 V
0.00051 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00063 % to
102 DIRECT DC Voltage Sourcing 100 mV to 1 V
0.00036 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00051 % to
103 DIRECT DC Voltage Sourcing 100 V to 1000 V
0.00054 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 28 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.1 mohm to 1
104 DIRECT Resistance Sourcing 4.85 % to 0.49 %
mohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
105 DIRECT Resistance Sourcing 1 Gohm to 10 Gohm 0.017 % to 0.18 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0010 % to
106 DIRECT Resistance Sourcing 1 Mohm to 10 Mohm
0.0012%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 29 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
107 DIRECT Resistance Sourcing 1 mohm to 10 mohm 0.49 % to 0.048 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0016 % to 0.0010
108 DIRECT Resistance Sourcing 1 ohm to 10 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
10 Mohm to 100 0.0012 % to 0.0031
109 DIRECT Resistance Sourcing
Mohm %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 30 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
10 mohm to 100
110 DIRECT Resistance Sourcing 0.048 % to 0.0063 %
mohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0010 % to 0.0009
111 DIRECT Resistance Sourcing 10 ohm to 100 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 kohm to 1 0.0009 % to 0.0010
112 DIRECT Resistance Sourcing
Mohm %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 31 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 Mohm to 1
113 DIRECT Resistance Sourcing 0.0031 % to 0.017 %
Gohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0063 % to 0.0016
114 DIRECT Resistance Sourcing 100 mohm to 1 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 ohm to 100
115 DIRECT Resistance Sourcing 0.0009%
kohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 32 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
116 DIRECT DC Current 10 µA to 100 µA 0.25 % to 0.04 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
117 DIRECT DC Current 100 µA to 329 µA 0.04 % to 0.024 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
118 DIRECT DC Current 5522A with Current 100 A to 1000 A 0.32 % to 0.3 %
CURRENT Coil 5500A by Direct
(Source) Method
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
119 DIRECT DC Current 2.99 A to 20 A 0.046 % to 0.12 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
120 DIRECT DC Current 5522A with Current 20 A to 100 A 0.53 % to 0.32 %
CURRENT Coil 5500A by Direct
(Source) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 33 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
121 DIRECT DC Current 3.29 mA to 329 mA 0.014 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
122 DIRECT DC Current 329 µA to 3.29 mA 0.024 % to 0.014 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
123 DIRECT DC Current 329 mA to 2.99 A 0.014 % to 0.046 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Fluke 5522A
TECHNICAL-
Multiproduct
124 DIRECT DC Power 1 W to 19950 W 0.03 % to 0.08 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
125 DIRECT DC Resistance Resistance Box by 0.01 ohm to 1 ohm 2.4 % to 0.58 %
CURRENT Direct Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 34 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Decade
126 DIRECT DC Resistance Resistance Box by 1 ohm to 10 ohm 0.58 % to 0.05 %
CURRENT Direct Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
127 DIRECT DC Resistance 5522A & Decade 10 ohm to 30 ohm 0.05 % to 0.011 %
CURRENT Resistance box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
290 Mohm to 1
128 DIRECT DC Resistance 5522A & Decade 0.4 % to 1.8 %
Gohm
CURRENT Meg Ohm Box by
(Source) Direct Method
Using Multiproduct
ELECTRO- Calibrator Fluke
TECHNICAL- 5522A, Decade
129 DIRECT DC Resistance Resistance Box & 3 kohm to 300 kohm 0.011 % to 0.008 %
CURRENT Decade Meg Ohm
(Source) Box by Direct
Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
130 DIRECT DC Resistance 5522A & Decade 3 Mohm to 30 Mohm 0.013 % to 0.04 %
CURRENT Meg Ohm Box by
(Source) Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 35 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Using Multiproduct


TECHNICAL- Calibrator Fluke
30 Mohm to 290
131 DIRECT DC Resistance 5522A & Decade 0.04 % to 0.4 %
Mohm
CURRENT Meg Ohm Box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
30 ohm
132 DIRECT DC Resistance 5522A & Decade 0.011 %
to 3 kohm
CURRENT Resistance Box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
300 kohm to 3
133 DIRECT DC Resistance 5522A & Decade 0.008 % to 0.013 %
Mohm
CURRENT Meg Ohm Box by
(Source) Direct Method
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
134 DIRECT DC Voltage 0.1 mV to 1 mV 1.2 % to 0.12 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
135 DIRECT DC Voltage 1 mV to 329 mV 0.12 % to 0.0023 %
5522A by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 36 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.0023 % to 0.0016
136 DIRECT DC Voltage 329 mV to 1000 V
5522A by Direct %
CURRENT
Method
(Source)
ELECTRO-
Using Decade
TECHNICAL-
Megohm Box 1 Gohm to 100
137 DIRECT Resistance 2.44 % to 5.85 %
8400HV by Direct Gohm
CURRENT
Method
(Source)
Oscilloscope -
ELECTRO-
Amplitude (Vertical Multiproduct
TECHNICAL-
Axis Deflection Calibrator Fluke
138 ELECTRICAL 1 mV to 130 V 4.74 % to 0.13 %
Factor) AC: 1 kHz 5522A, SC600 &
EQUIPMENT
Square Wave @ 1 Direct Method
(Source)
Mohm
ELECTRO-
Using Multiproduct
TECHNICAL-
Oscilloscope - Calibrator Fluke 5 mV to 5.5 V @ (50
139 ELECTRICAL 3.63 % to 5.11 %
Bandwidth 5522A & SC600 by kHz to 500 MHz)
EQUIPMENT
Direct Method
(Source)
ELECTRO-
Oscilloscope - Time Using Multiproduct
TECHNICAL-
Base (Horizontal Calibrator Fluke
140 ELECTRICAL 2 ns to 5 s 0.5 % to 0.12 %
Axis Deflection 5522A & SC600 by
EQUIPMENT
Factor) Direct Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 37 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Oscilloscope
Using Multiproduct
TECHNICAL- Amplitude (Vertical
Calibrator Fluke
141 ELECTRICAL Axis Deflection 1 mV to 130 V 4.68 % to 0.058 %
5522A & SC600 by
EQUIPMENT Factor) DC @ 1
Direct Method
(Source) Mohm
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
B-Type
142 TEMPERATURE Sourcing 600 °C to 1820 °C 0.044°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
B-Type Calibrator Fluke
143 TEMPERATURE 600 °C to 1820 °C 0.6 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
E-Type Calibrator Fluke (-) 200 °C to 1000
144 TEMPERATURE 0.12 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 38 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
E-Type
145 TEMPERATURE Sourcing -200 °C to 1000 °C 0.011°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
J-Type Calibrator Fluke (-) 200 °C to 1200
146 TEMPERATURE 0.14 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
J-Type
147 TEMPERATURE Sourcing -200 °C to 1200 °C 0.014°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
K-Type Calibrator Fluke (-) 200 °C to 1370
148 TEMPERATURE 0.21 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 39 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
K-Type
149 TEMPERATURE Sourcing -200 °C to 1370 °C 0.035°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
N-Type Calibrator Fluke (-) 200 °C to 1300
150 TEMPERATURE 0.17 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
N-Type
151 TEMPERATURE Sourcing -250 °C to 1300 °C 0.06°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
R-Type Calibrator Fluke
152 TEMPERATURE 50 °C to 1760 °C 0.59 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 40 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
R-Type
153 TEMPERATURE Sourcing 50 °C to 1765 °C 0.05°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
154 TEMPERATURE RTD PT 100 Multimeter by Direct (-) 200 °C to 600 °C 0.18 °C
SIMULATION Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
155 TEMPERATURE RTD Pt 100 Sourcing -200 °C to 800 °C 0.004°C
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
S-Type Calibrator Fluke
156 TEMPERATURE 50 °C to 1760 °C 0.58 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 41 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
S-Type
157 TEMPERATURE Sourcing 50 °C to 1765 °C 0.045°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
T-Type Calibrator Fluke
158 TEMPERATURE (-) 100 °C to 400 °C 0.11 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
T-Type
159 TEMPERATURE Sourcing -100 °C to 400 °C 0.086°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
B-Type Calibrator Fluke
160 TEMPERATURE 600 °C to 1820 °C 0.6 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 42 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
E-Type Calibrator Fluke
161 TEMPERATURE (-) 200 °C to 950 °C 0.12 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
J-Type Calibrator Fluke (-) 200 °C to 1150
162 TEMPERATURE 0.14 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
K-Type Calibrator Fluke (-) 200 °C to 1370
163 TEMPERATURE 0.2 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
N-Type Calibrator Fluke (-) 100 °C to 1250
164 TEMPERATURE 0.17 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
R-Type Calibrator Fluke
165 TEMPERATURE 5 °C to 1750 °C 0.53 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 43 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
166 TEMPERATURE RTD PT 100 (-) 200 °C to 800 °C 0.11 °C
5522A by Direct
SIMULATION
method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
S-Type Calibrator Fluke
167 TEMPERATURE 5 °C to 1750 °C 0.58 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
T-Type Calibrator Fluke
168 TEMPERATURE (-) 100 °C to 350 °C 0.11 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0013 % to 0.00035
169 TIME & Frequency Sourcing 1 Hz to 10 Hz
%
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 44 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00012 % to
170 TIME & Frequency Sourcing 1 kHz to 100 MHz
0.00012%
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
171 TIME & Frequency Multimeter by Direct 10 Hz to 100 Hz 0.05 % to 0.01 %
FREQUENCY Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00035 % to
172 TIME & Frequency Sourcing 10 Hz to 100 Hz
0.00012 %
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
173 TIME & Frequency Multimeter by Direct 100 Hz to 300 kHz 0.01 %
FREQUENCY Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 45 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00012 % to
174 TIME & Frequency Sourcing 100 Hz to 1 kHz
0.00012 %
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
175 TIME & Frequency Multimeter by Direct 3 Hz to 10 Hz 0.081 % to 0.05 %
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Frequency
176 TIME & Frequency, Period Counter by Direct 1 Hz to 300 MHz 2 ppm
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Time Totaliser
177 TIME & Timer, Stopwatch by Comparison 1 s to 1000 s 0.0022 s to 0.063 s
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Time Totaliser
178 TIME & Timer, Stopwatch by Comparison 1000 s to 5000 s 0.063 s to 0.25 s
FREQUENCY Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 46 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Time Totaliser
179 TIME & Timer, Stopwatch by Comparison 5000 s to 86400 s 0.25 s to 1 s
FREQUENCY Method
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.0011 % to 0.00051
180 TIME & Frequency 1 Hz to 1000 Hz
5522A by Direct %
FREQUENCY
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.00065 % to 0.029
181 TIME & Frequency 100 kHz to 2 MHz
5522A by Direct %
FREQUENCY
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.00051 % to
182 TIME & Frequency 1000 Hz to 100 kHz
5522A by Direct 0.00065 %
FREQUENCY
Method
(Source)
Centrifuge, RPM, Using Digital
MECHANICAL-
Speed (Indicator, Tachometer as per 1000 RPM to 10000
183 ACCELERATION 0.8 RPM
Sensor, Meter, SANAS TR 45-02 by RPM
AND SPEED
Source) Comparison Method
Centrifuge, RPM, Using Digital
MECHANICAL-
Speed (Indicator, Tachometer as per 10000 RPM to 30000
184 ACCELERATION 1.5 RPM
Sensor, Meter, SANAS TR 45-02 by RPM
AND SPEED
Source) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 47 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Centrifuge, RPM, Using Digital


MECHANICAL-
Speed (Indicator, Tachometer as per
185 ACCELERATION 6 RPM to 1000 RPM 0.16 RPM
Sensor, Meter, SANAS TR 45-02 by
AND SPEED
Source) Comparison Method
Using Digital
MECHANICAL-
Tachometer as per 1000 RPM to 10000
186 ACCELERATION Contact Tachometer 0.31 RPM
SANAS TR 45-02 by RPM
AND SPEED
Comparison Method
Using Digital
MECHANICAL-
Tachometer as per 10000 RPM to 15000
187 ACCELERATION Contact Tachometer 1.3 RPM
SANAS TR 45-02 by RPM
AND SPEED
Comparison Method
Using Digital
MECHANICAL-
Tachometer as per
188 ACCELERATION Contact Tachometer 6 RPM to 1000 RPM 0.16 RPM
SANAS TR 45-02 by
AND SPEED
Comparison Method
Non-Contact
Tachometer, Using Digital
MECHANICAL-
Stroboscope, RPM, Tachometer as per 1000 RPM to 10000
189 ACCELERATION 0.31 RPM
Speed (Indicator, SANAS TR 45-02 by RPM
AND SPEED
Sensor, Meter, Comparison Method
Source)
Non-Contact
Tachometer, Using Digital
MECHANICAL-
Stroboscope, RPM, Tachometer as per 10000 RPM to 50000
190 ACCELERATION 1.4 RPM
Speed (Indicator, SANAS TR 45-02 by RPM
AND SPEED
Sensor, Meter, Comparison Method
Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 48 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Non-Contact
Tachometer, Using Digital
MECHANICAL-
Stroboscope, RPM, Tachometer as per 50000 RPM to 99950
191 ACCELERATION 2.5 RPM
Speed (Indicator, SANAS TR 45-02 by RPM
AND SPEED
Sensor, Meter, Comparison Method
Source)
Non-Contact
Tachometer, Using Digital
MECHANICAL-
Stroboscope, RPM, Tachometer as per
192 ACCELERATION 6 RPM to 1000 RPM 0.16 RPM
Speed (Indicator, SANAS TR 45-02 by
AND SPEED
Sensor, Meter, Comparison Method
Source)
Using Sound Level
MECHANICAL- 94 dB & 114 dB @ 1
193 Sound Level Meter Calibrator as per IS 0.21 dB
ACOUSTICS kHz
15575 (Part-2)
BS / IP / RF U Tube Using Standard
MECHANICAL-
Reverse Flow Capillary Viscometer
194 DENSITY AND 1 cSt/s 0.3 %
Viscometer @ 135 by Comparison
VISCOSITY
°C Method
MECHANICAL- Cannon Manning Using Viscosity
200 cP/s to 6000
195 DENSITY AND Vacuum Viscometer Standards by Direct 1.65 %
cP/s
VISCOSITY @ 60 °C Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 49 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Density Hydrometer,
Brix Hydrometer,
Baume Hydrometer,
MECHANICAL- Using Cuckow's
Twaddle
196 DENSITY AND Method and ASTM E 0.6 g/ml to 2 g/ml 0.000093 g/ml
Hydrometer, Specific
VISCOSITY 126
Gravity Hydrometer,
Lactometer,
Alcoholmeter
Using Weighing
MECHANICAL- Balance
600 kg/m^3 to 1800
197 DENSITY AND Density of Liquid (Readability: 0.1 mg) 0.065 %
kg/m^3
VISCOSITY as per OIML G 14 by
Gravimetric Method
Using Weighing
Balance
MECHANICAL- (Readability: 0.1 mg)
500 kg/m^3 to
198 DENSITY AND Density of Solid as per OIML G 14 / 0.065 %
15000 kg/m^3
VISCOSITY OIML R 111-1 by
Hydrostatic
Weighing Method
MECHANICAL- Glass Capillary Using Viscosity
0.002 cSt/s to 0.008
199 DENSITY AND Viscometer (Direct Standards by Direct 0.2 %
cSt/s
VISCOSITY Flow) Method
MECHANICAL- Glass Capillary Using Viscosity
0.008 cSt/s to 100
200 DENSITY AND Viscometer (Direct Standards by Direct 0.18 % to 0.44 %
cSt/s
VISCOSITY Flow) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 50 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Glass Capillary Using Viscosity


0.004 cSt/s to 0.035
201 DENSITY AND Viscometer (Direct Standards by Direct 0.5 %
cSt/s
VISCOSITY Flow) @ -20 °C Method
MECHANICAL- Glass Capillary Using Viscosity
0.008 cSt/s to 0.05
202 DENSITY AND Viscometer (Direct Standards by Direct 0.5 %
cSt/s
VISCOSITY Flow) @ -40 °C Method
MECHANICAL- Glass Capillary Using Viscosity
0.002 cSt/s to 300
203 DENSITY AND Viscometer (Reverse Standards by Direct 0.09 % to 0.69 %
cSt/s
VISCOSITY Flow) Method
MECHANICAL-
Saybolt Furol Using Viscosity
204 DENSITY AND Up to 2500 SFS 1%
Viscometer Standard
VISCOSITY
MECHANICAL-
Saybolt Universal
205 DENSITY AND Viscosity Standard Up to 1000 SUS 1%
Viscometer
VISCOSITY
MECHANICAL-
Viscometers
206 DENSITY AND Viscosity Standard Up to 60 x 10^6 cP 0.4 %
(Dynamic Viscosity)
VISCOSITY
MECHANICAL- Viscosity Cup
Using Viscosity
207 DENSITY AND (Orifice Diameter: 1 Up to 1800 cSt 0.2 %
Standard
VISCOSITY mm to 6 mm)
MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS
208 Bevel Protractor 0 ° to 360 ° 1.25 minute of arc
MEASURING 4239 by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 51 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Bore Dial Gauge - for Using Dial
(BASIC Transmission Calibration Tester as
209 0 to 2 mm 1.6 µm
MEASURING Accuracy only (L.C.: per JIS B 7515 by
INSTRUMENT, 0.0001 mm) Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Coating Thickness Using Standard Foils
(BASIC
210 Gauge (L.C. : 0.1 by Comparison 0 to 0.1 mm 0.8 µm
MEASURING
µm) Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Coating Thickness Using Standard Foils
(BASIC
211 Gauge (L.C.: 0.01 by Comparison 2 mm to 6 mm 0.03 mm
MEASURING
mm) Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Coating Thickness Using Standard Foils
(BASIC
212 Gauge (L.C.: 0.01 by Comparison 6 mm to 10 mm 0.033 mm
MEASURING
mm) Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 52 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Standard Foils
(BASIC Coating Thickness
213 by Comparison 0.1 mm to 2 mm 3.5 µm
MEASURING Gauge (L.C.: 0.1 µm)
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC
214 Combination Set Projector by 0 ° to 180 ° 1.25 minute of arc
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Optical Flat &
(BASIC Lever Dial Gauge as
215 Comparator Stand 250 mm x 250 mm 4 µm
MEASURING per IS 7599 by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital Level
(BASIC Cube Mould
216 as per IS 10086 by 0 ° to 90 ° 0.18 °
MEASURING (Angular)
Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 53 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Digital Caliper
(BASIC
217 Cube Mould (Linear) as per IS 10086 by Up to 150 mm 0.02 mm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL- Using Slip Gauge Set
DIMENSION Cylindrical & Digital
(BASIC Measuring Pin, Pin Comparator as per
218 0.17 mm to 25 mm 0.7 µm
MEASURING Gauge, Thread IS 11103 & IS 6311
INSTRUMENT, Measuring Wire by Comparison
GAUGE ETC.) Method
MECHANICAL-
Using Digital
DIMENSION
Cylindrical Setting Comparator &
(BASIC
219 Master FCDMM as per IS 3 mm to 200 mm 4.3 µm
MEASURING
(Concentricity) 4349 by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL- Using Digital
DIMENSION Comparator, Slip
(BASIC Cylindrical Setting Gauge Set & Long
220 3 mm to 200 mm 1 µm
MEASURING Master (Diameter) Gauge Block Set as
INSTRUMENT, per IS 4349 by
GAUGE ETC.) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 54 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Depth Indent
Using Digital
(BASIC Measurement
221 Comparator by 0 to 25 mm 3.02 µm
MEASURING Device for Hardness
Comparison Method
INSTRUMENT, Tester (L.C.: 0.2 µm)
GAUGE ETC.)
MECHANICAL-
Using Mic Check Set,
DIMENSION
Slip Gauge Set &
(BASIC Depth Micrometer
222 Length Bar Set as 0 to 150 mm 3.7 µm
MEASURING (L.C.: 0.001 mm)
per BS 6468 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL- Using Mic Check Set,
DIMENSION Slip Gauge Set, Long
(BASIC Depth Micrometer Gauge Block Set &
223 150 mm to 300 mm 6.8 µm
MEASURING (L.C.: 0.01 mm) Length Bar Set as
INSTRUMENT, per BS 6468 by
GAUGE ETC.) Comparison Method
MECHANICAL-
DIMENSION Using Slip Gauge
Dial / Digital
(BASIC Set, Standard Foils
224 Thickness Gauge 0 to 25 mm 0.4 µm
MEASURING by Comparison
(L.C.: 0.0001 mm)
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 55 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Dial / Digital Using Slip Gauge Set
(BASIC
225 Thickness Gauge & Standard Foils by 0 to 25 mm 0.7 µm
MEASURING
(L.C.: 0.001 mm) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Thickness Using Slip Gauge Set
(BASIC
226 Gauge (L.C.: 0.1 by Comparison 0 to 75 mm 0.04 mm
MEASURING
mm) Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital Caliper
(BASIC
227 Elongation Gauge as per IS 2386 by 14.7 mm to 81 mm 16.3 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Mic Check Set,
DIMENSION
Slip Gauge Set &
(BASIC External Micrometer
228 Optical Flat as per IS 0 to 25 mm 0.14 µm
MEASURING (L.C.: 0.0001 mm)
2967 by Comparison
INSTRUMENT,
Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 56 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Mic Check Set,


DIMENSION Slip Gauge Set,
(BASIC External Micrometer Optical Flat &
229 0 to 150 mm 1 µm
MEASURING (L.C.: 0.001 mm) Length Bar Set as
INSTRUMENT, per IS 2967 by
GAUGE ETC.) Comparison Method
Using Mic Check Set,
MECHANICAL-
Slip Gauge Set,
DIMENSION
Optical Flat, Long
(BASIC External Micrometer
230 Gauge Block Set & 100 mm to 200 mm 1.9 µm
MEASURING (L.C.: 0.001 mm)
Length Bar Set as
INSTRUMENT,
per IS 2967 by
GAUGE ETC.)
Comparison Method
Using Mic Check Set,
MECHANICAL-
Slip Gauge Set,
DIMENSION
Optical Flat, Long
(BASIC External Micrometer
231 Gauge Block Set & 200 mm to 500 mm 5.9 µm
MEASURING (L.C.: 0.01 mm)
Length Bar Set as
INSTRUMENT,
per IS 2967 by
GAUGE ETC.)
Comparison Method
Using Mic Check Set,
MECHANICAL-
Slip Gauge Set,
DIMENSION
Optical Flat, Long
(BASIC External Micrometer 500 mm to 1000
232 Gauge Block Set & 10.8 µm
MEASURING (L.C.: 0.01 mm) mm
Length Bar Set as
INSTRUMENT,
per IS 2967 by
GAUGE ETC.)
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 57 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Digital
DIMENSION
Comparator &
(BASIC
233 Feeler Gauge Digital Micrometer 0.03 mm to 2 mm 0.8 µm
MEASURING
as per IS 3179 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS 4.89 mm to 33.9
234 Flakiness Gauge 6.98 µm
MEASURING 2386 by Comparison mm
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Slip Gauge Set
(BASIC & Digital
235 Flush Pin Gauge Up to 200 mm 3 µm
MEASURING Comparator by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Slip Gauge
DIMENSION
Set, Optical Flat &
(BASIC Gauge Block
236 Digital Comparator Up to 500 mm 1.4 µm
MEASURING Accessories Set
as per IS 4440 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 58 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Profile
(BASIC Glass Scale (L.C.:
237 Projector by Up to 300 mm 3.7 µm
MEASURING 0.05 mm)
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Slip Gauge Set
Groove Dial Gauge &
(BASIC & Slip Gauge
238 Dial Caliper Gauge 5 mm to 150 mm 5 µm
MEASURING Accessories Set by
(Internal / External)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital
(BASIC 0
239 Hegmann Gauge Comparator by 1.58 µm
MEASURING to 100 µm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker & Slip
(BASIC Height Gauge (L.C.:
240 Gauge Set as per IS 0 to 300 mm 8 µm
MEASURING 0.01 mm)
2921 by Comparison
INSTRUMENT,
Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 59 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Caliper
DIMENSION
Checker & Slip
(BASIC Height Gauge (L.C.:
241 Gauge Set as per IS 300 mm to 600 mm 9.27 µm
MEASURING 0.01 mm)
2921 by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker, Slip Gauge
(BASIC Height Gauge (L.C.: 600 mm to 1000
242 Set & Length Bar Set 14.5 µm
MEASURING 0.01 mm) mm
as per IS 2921 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Slip Gauge
DIMENSION
Height Master, Set, Length Bar Set
(BASIC
243 Height Micrometer & Lever Dial Gauge Up to 310 mm 4.5 µm
MEASURING
(L.C.: 0.001 mm) as per IS 13907 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Industrial Gauge Using Profile
(BASIC
244 Plain Work Piece Projector by 0 ° to 360 ° 1.25 minute of arc
MEASURING
(Angle) Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 60 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Industrial Gauges - Using Profile
(BASIC
245 Plain Work Piece Projector by 0.1 mm to 150 mm 3.2 µm
MEASURING
(Diameter) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Industrial Gauges - Using Profile
(BASIC
246 Plain Work Piece Projector by 45 µm to 200 mm 3.2 µm
MEASURING
(Length) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL- Using Slip Gauge
DIMENSION Accessories Set, Slip
Internal Micrometer 100 mm to 1200
(BASIC Gauge Set & Long
247 - Stick Type (L.C.: mm (In steps of 300 1.6 µm
MEASURING Gauge Block Set as
0.001 mm) mm)
INSTRUMENT, per IS 2966 by
GAUGE ETC.) Comparison Method
MECHANICAL- Using Slip Gauge
DIMENSION Accessories Set, Slip
Internal Micrometer
(BASIC Gauge Set & Long
248 - Stick Type (L.C.: 13 mm to 100 mm 0.9 µm
MEASURING Gauge Block Set as
0.001 mm)
INSTRUMENT, per IS 2966 by
GAUGE ETC.) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 61 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Dial
(BASIC Lever Dial Gauge Calibration Tester as
249 0 to 2 mm 1.5 µm
MEASURING (L.C.: 0.001 mm) per IS 11498 by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
250 by Comparison 0 to 25 mm 0.4 µm
MEASURING with Indicator (L.C.:
Method
INSTRUMENT, 0.0001 mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
251 by Comparison 0 to 50 mm 1.5 µm
MEASURING with Indicator (L.C.:
Method
INSTRUMENT, 0.001 mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
252 & Length Bar Set by 50 mm to 600 mm 6.7 µm
MEASURING with Indicator (L.C.:
Comparison Method
INSTRUMENT, 0.001 mm)
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 62 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Scale & Tape
(BASIC Calibrator as per IS 14 x sqrt (L/1000)
253 Measuring Scale 0 to 1000 mm
MEASURING 1481 by Comparison µm (where L in mm)
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS
254 Measuring Scale 0 to 250 mm 3.6 µm
MEASURING 1481 by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Scale and
(BASIC Tape Calibrator as 1000 mm to 3000 33 x sqrt (L/1000)
255 Measuring Scale
MEASURING per IS 1481 by mm µm (where L in mm)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Scale & Tape
(BASIC Measuring Tape, PI Calibrator as per IS 16.6 x sqrt (L/1000)
256 0 to 5 m
MEASURING Tape 1269 by Comparison µm (where L in mm)
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 63 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Scale & Tape
184.9 x sqrt
(BASIC Measuring Tape, PI Calibrator as per IS
257 5 m to 100 m (L/1000) µm (where
MEASURING Tape 1269 by Comparison
L in mm)
INSTRUMENT, Method
GAUGE ETC.)
Using Slip Gauge
MECHANICAL-
Set, Digital
DIMENSION
Comparator, Long
(BASIC Micrometer Setting
258 Gauge Block Set & 100 mm to 275 mm 2.6 µm
MEASURING Rod, Length Bar
Length Bar Set as
INSTRUMENT,
per IS 7014 by
GAUGE ETC.)
Comparison Method
MECHANICAL- Using Slip Gauge
DIMENSION Set, Digital
(BASIC Micrometer Setting Comparator &
259 25 mm to 100 mm 0.9 µm
MEASURING Rod, Length Bar Length Bar Set as
INSTRUMENT, per IS 7014 by
GAUGE ETC.) Comparison Method
Using Slip Gauge
MECHANICAL-
Set, Digital
DIMENSION
Comparator, Long
(BASIC Micrometer Setting
260 Gauge Block Set & 275 mm to 975 mm 8.8 µm
MEASURING Rod, Length Bar
Length Bar Set as
INSTRUMENT,
per IS 7014 by
GAUGE ETC.)
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 64 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Slip Gauge Set
(BASIC Penetrometer (L.C.:
261 by Comparison 0 to 40 mm 0.07 mm
MEASURING 0.1 mm)
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Slip Gauge Set
(BASIC Pistol Caliper (L.C.:
262 by Comparison 0 to 100 mm 0.045 mm
MEASURING 0.1 mm)
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS
263 Pitch Gauge 0 ° to 60 ° 1.14 minute of arc
MEASURING 4211 by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS
264 Pitch Gauge 0.4 mm to 7 mm 3.6 µm
MEASURING 4211 by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 65 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Profile
(BASIC Pitch Micrometer
265 Projector by 50 ° to 60 ° 2.8 minute of arc
MEASURING (Angular)
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Mic Check Set
(BASIC Pitch Micrometer
266 & Slip Gauge Set by 0 to 50 mm 0.95 µm
MEASURING (Linear)
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Digital
DIMENSION
Comparator & Slip
(BASIC
267 Plain Plug Gauge Gauge Set as per IS 1 mm to 110 mm 1.75 µm
MEASURING
3455 by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL- Using Digital
DIMENSION Comparator, Slip
(BASIC Gauge Set & Long
268 Plain Plug Gauge 110 mm to 200 mm 3.4 µm
MEASURING Gauge Block Set as
INSTRUMENT, per IS 3455 by
GAUGE ETC.) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 66 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Dial
DIMENSION
Plunger / Digital Dial Calibration Tester &
(BASIC
269 Gauge (L.C.: 0.0001 Slip Gauge Set as 0 to 25 mm 0.4 µm
MEASURING
mm) per IS 2092 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Dial
DIMENSION
Plunger / Digital Dial Calibration Tester &
(BASIC
270 Gauge (L.C.: 0.0005 Slip Gauge Set as 0 to 50 mm 0.7 µm
MEASURING
mm) per IS 2092 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Dial
DIMENSION
Plunger / Digital Dial Calibration Tester &
(BASIC
271 Gauge (L.C.: 0.01 Slip Gauge Set as 50 mm to 100 mm 5.1 µm
MEASURING
mm) per IS 2092 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Length Bar Set
(BASIC Proximity Sensor
272 & Slip Gauge Set by 0 to 1000 mm 11 µm
MEASURING (L.C.: 0.01 mm)
Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 67 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Profile
(BASIC Projector as per IS
273 Radius Gauge 0.6 mm to 25 mm 3.6 µm
MEASURING 5273 by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Slip Gauge Set
(BASIC
274 Snap Gauge as per IS 3455 by 2 mm to 200 mm 2.1 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Profile
DIMENSION
Projector, Slip Gauge
(BASIC
275 Spline Ring Gauge Set & Measuring Pin Up to 120 mm 4.4 µm
MEASURING
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital
(BASIC
276 Standard Foils Comparator by 0.004 mm to 10 mm 0.6 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 68 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Electronic
(BASIC Level as per IS 2220 2.3 x sqrt (L/150)µm
277 Straight Edge Up to 5000 mm
MEASURING & IS 12937 by (where L in mm)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC Taper Scale (L.C.:
278 Projector by 1 mm to 150 mm 8.81 µm
MEASURING 0.1 mm)
Comparison Method
INSTRUMENT,
GAUGE ETC.)
Using FCDMM,
MECHANICAL-
Thread Measuring
DIMENSION
Wires & Cylindrical
(BASIC Taper Thread Plug
279 Setting Master as 2 mm to 100 mm 6 µm
MEASURING Gauge
per IS 554, IS 8999
INSTRUMENT,
& ASME B 1.20.1 by
GAUGE ETC.)
Comparison Method
MECHANICAL-
Using Slip Gauge Set
DIMENSION
& Digital
(BASIC Test Mandrel
280 Comparator as per Up to 50 mm 1.4 µm
MEASURING (Diameter)
IS 4349 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 69 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Slip Gauge Set
DIMENSION
& Digital
(BASIC Test Mandrel
281 Comparator as per 20 mm to 150 mm 1.4 µm
MEASURING (Length)
IS 2063 by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using FCDMM &
(BASIC Test Mandrel (Run Lever Dial Gauge as
282 Up to 50 mm 5.5 µm
MEASURING out Dia) per IS 2063 by
INSTRUMENT, Comparison Method
GAUGE ETC.)
Using Profile
MECHANICAL- Projector & Digital
DIMENSION Caliper as per IS 460
(BASIC (Part 1, 2, 3), ASTM 0.025 mm to 100
283 Test Sieves 4.2 µm
MEASURING E 11, ASTM E 323 & mm
INSTRUMENT, BS/ISO 3310 (Part 1,
GAUGE ETC.) 2) by Comparison
Method
Using Profile
MECHANICAL- Projector & Digital
DIMENSION Caliper as per IS 460
(BASIC (Part 1, 2, 3), ASTM
284 Test Sieves 100 mm to 200 mm 28 µm
MEASURING E 11, ASTM E 323 &
INSTRUMENT, BS/ISO 3310 (Part 1,
GAUGE ETC.) 2) by Comparison
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 70 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using FCDMM,
MECHANICAL- Cylindrical Setting
DIMENSION Master & Thread
(BASIC Measuring Wires as
285 Thread Plug Gauge 1 mm to 100 mm 4.12 µm
MEASURING per IS 4218 (Part-3),
INSTRUMENT, IS 14962 (Part-1, 3)
GAUGE ETC.) & IS 2334 by
Comparison Method
MECHANICAL-
DIMENSION Using Slip Gauge Set
(BASIC Ultrasonic Thickness & Long Gauge Block
286 0 to 200 mm 0.07 mm
MEASURING Gauge Set by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL- Using Caliper
DIMENSION Checker & Slip
(BASIC Vernier Caliper (L.C.: Gauge Set as per IS
287 0 to 300 mm 7.8 µm
MEASURING 0.01 mm) 3651 (Part 1, 2, 3)
INSTRUMENT, by Comparison
GAUGE ETC.) Method
MECHANICAL- Using Caliper
DIMENSION Checker & Slip
(BASIC Vernier Caliper (L.C.: Gauge Set as per IS
288 300 mm to 600 mm 9.6 µm
MEASURING 0.01 mm) 3651 (Part 1, 2, 3)
INSTRUMENT, by Comparison
GAUGE ETC.) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 71 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Caliper


DIMENSION Checker, Slip Gauge
(BASIC Vernier Caliper (L.C.: Set & Length Bar Set 600 mm to 1000
289 15 µm
MEASURING 0.01 mm) as per IS 3651 Part mm
INSTRUMENT, 1, 2, 3) by
GAUGE ETC.) Comparison Method
MECHANICAL- Using Caliper
DIMENSION Checker & Length
(BASIC Vernier Caliper (L.C.: Bar Set as per IS
290 0 to 2000 mm 24.4 µm
MEASURING 0.02 mm) 3651 (Part 1, 2, 3)
INSTRUMENT, by Comparison
GAUGE ETC.) Method
MECHANICAL- Using Slip Gauge
DIMENSION Vernier Depth Set, Long Gauge
(BASIC Gauge / Depth Block Set & Length
291 0 to 600 mm 9.8 µm
MEASURING Vernier (L.C.: 0.01 Bar Set as per IS
INSTRUMENT, mm) 4213 by Comparison
GAUGE ETC.) Method
MECHANICAL-
DIMENSION
Using Profile
(BASIC
292 Weld Fillet Gauge Projector by 0 to 25 mm 10.07 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 72 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Profile
(BASIC
293 Weld Gauge Projector by 0 to 60 mm 8.74 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC
294 Weld Gauge (Angle) Projector by 0 ° to 90 ° 3 minute of arc
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC Wet / Dry Film 0
295 Projector by 3.6 µm
MEASURING Applicator to 3000 µm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Digital
(BASIC Micrometer as per IS
296 Width Gauge 0.5 mm to 10 mm 3 µm
MEASURING 3455 by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 73 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Profile
(BASIC
297 Wire Gauge Projector by 0 to 10 mm 9.02 µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
Using Slip Gauge
MECHANICAL-
Set, Length Bar Set
DIMENSION
298 Caliper Checker & Lever Dial Gauge 0 to 300 mm 3 µm
(PRECISION
by Comparison
INSTRUMENTS)
Method
Using Slip Gauge
MECHANICAL-
Set, Length Bar Set
DIMENSION
299 Caliper Checker & Lever Dial Gauge 300 mm to 600 mm 5 µm
(PRECISION
by Comparison
INSTRUMENTS)
Method
MECHANICAL- Using Digital
Dial Calibration
DIMENSION Comparator as per
300 Tester, Micrometer 0 to 25 mm 1.1 µm
(PRECISION IS 9483 by
Head (L.C.: 0.2 µm)
INSTRUMENTS) Comparison Method
Using Electronic
MECHANICAL- Level & Robust
Electronic Level
DIMENSION Tilting Table as per
301 (Sensitivity: 0.001 ± 2 mm/m 5.1 µm/m
(PRECISION IS 5706 & JIS B 7510
mm/m)
INSTRUMENTS) by Comparison
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 74 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
MECHANICAL- Electronic Level, Level & Robust
DIMENSION Spirit Level Tilting Table as per
302 ± 20 mm/m 18.93 µm/m
(PRECISION (Sensitivity: 0.01 IS 5706 & JIS B 7510
INSTRUMENTS) mm/m) by Comparison
Method
MECHANICAL- Profile Projector, Using Angular
DIMENSION Microscope, Video Graticules as per JIS
303 0 ° to 360 ° 0.4 minute of arc
(PRECISION Measuring Machine B 7184 by
INSTRUMENTS) (Angular) Comparison Method
MECHANICAL- Profile Projector,
Using Glass Scale as
DIMENSION Microscope, Video
304 per JIS B 7184 by 300 mm x 200 mm 2.8 µm
(PRECISION Measuring Machine
Comparison Method
INSTRUMENTS) (Linear)
MECHANICAL- Profile Projector,
Using Glass Scale as
DIMENSION Microscope, Video
305 per JIS B 7184 by Up to 100 X 0.3 %
(PRECISION Measuring Machine
Comparison Method
INSTRUMENTS) (Magnification)
IRHD Rubber
Hardness Tester
MECHANICAL- As per ASTM D 1415
306 (Indent Length 0 to 100 IRHD 3.09 µm
DUROMETER & ISO 48-9
Measurement
Method)
IRHD Rubber
Hardness Tester
MECHANICAL- As per ASTM D 1415 Minor Load 0.3 N to
307 (Spring Force 2.16 mN to 3.14 mN
DUROMETER & ISO 48-9 Total Load 5.7 N
Measuement
Method)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 75 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Shore Hardness
Tester, Durometer
(Type A, B, C, D, E,
As per ASTM D 2240,
MECHANICAL- M, O, OO, OOO,
308 ISO 48-9 & JIS S 0 to 100 Shore 0.27 Shore
DUROMETER OOO-S, AM, AO, DO,
6050
C1L): Indent Length
Measurement
Method
Shore Hardness
Tester, Durometer
(Type A, B, C, D, E,
As per ASTM D 2240,
MECHANICAL- M, O, OO, OOO,
309 ISO 48-9 & JIS S 0 to 100 Shore 0.38 Shore
DUROMETER OOO-S, AM, AO, DO,
6050
C1L): Spring Force
Measurement
Method
Using Newton
MECHANICAL-
Push Pull Gauge, Weights as per
MOBILE FORCE
310 Force Gauge, Spring VDI/VDE 2624 0.1 N to 50 N 0.006 N
MEASURING
Balance (Part-2.1) & JIS B
SYSTEM
7602
Using Newton
MECHANICAL-
Push Pull Gauge, Weights as per
MOBILE FORCE
311 Force Gauge, Spring VDI/VDE 2624 50 N to 500 N 0.07 N
MEASURING
Balance (Part-2.1) & JIS B
SYSTEM
7602

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 76 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Newton
MECHANICAL-
Push Pull Gauge, Weights as per
MOBILE FORCE
312 Force Gauge, Spring VDI/VDE 2624 500 N to 1000 N 0.6 N
MEASURING
Balance (Part-2.1) & JIS B
SYSTEM
7602
Characterization / Using Fluke Piston
MECHANICAL-
Calibration of Gauge 2468 as per
PRESSURE
Absolute Dead Euramet cg -3 by 0.014 bar (abs) to
313 BALANCE OR 0.006%
Weight Tester Cross Float Method 3.5 bar (abs)
DEAD WEIGHT
(Pneumatic - (Effective Area &
TESTER
Absolute Pressure) Pressure Method)
Characterization / Using Fluke Piston
MECHANICAL-
Calibration of Gauge 2468 as per
PRESSURE
Absolute Dead Euramet cg -3 by 0.14 bar (abs) to 70
314 BALANCE OR 0.0060%
Weight Tester Cross Float Method bar (abs)
DEAD WEIGHT
(Pneumatic - (Effective Area &
TESTER
Absolute Pressure) Pressure Method)
Using DH-Budenberg
MECHANICAL- Dead Weight Tester
Characterization /
PRESSURE as per Euramet cg-3
Calibration of Dead
315 BALANCE OR & OIML R 110 by 6 bar to 1200 bar 0.0087 % rdg
Weight Tester
DEAD WEIGHT Cross Float Method
(Hydraulic Pressure)
TESTER (Effective Area &
Pressure Method)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 77 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using DH-Budenberg
MECHANICAL- Characterization / Dead Weight Tester
PRESSURE Calibration of Dead as per Euramet cg-3
(-) 35 mbar to (-)
316 BALANCE OR Weight Tester & OIML R 110 by 0.0126 % rdg
1000 mbar
DEAD WEIGHT (Pneumatic Cross Float Method
TESTER Pressure) (Effective Area &
Pressure Method)
Using DH-Budenberg
MECHANICAL- Characterization / Dead Weight Tester
PRESSURE Calibration of Dead as per Euramet cg-3
317 BALANCE OR Weight Tester & OIML R110 by 0.2 bar to 25 bar 0.0062 % rdg
DEAD WEIGHT (Pneumatic Cross Float Method
TESTER Pressure) (Effective Area &
Pressure Method)
Using DH-Budenberg
MECHANICAL- Characterization / Dead Weight Tester
PRESSURE Calibration of Dead as per Euramet cg-3
35 mbar to 1000
318 BALANCE OR Weight Tester & OIML R 110 by 0.0077 % rdg
mbar
DEAD WEIGHT (Pneumatic Cross Float Method
TESTER Pressure) (Effective Area &
Pressure Method)
Using Fluke Piston
Characterization /
MECHANICAL- Gauge 2468 as per
Calibration of
PRESSURE Euramet cg -3 &
Pneumatic Dead
319 BALANCE OR OIML R 110 by Cross 0.014 bar to 3.5 bar 0.006%
Weight Tester
DEAD WEIGHT Float Method
(Pneumatic -
TESTER (Effective Area &
Pressure)
Pressure Method)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 78 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Fluke Piston


Characterization /
MECHANICAL- Gauge 2468 as per
Calibration of
PRESSURE Euramet cg -3 &
Pneumatic Dead
320 BALANCE OR OIML R 110 by Cross 0.14 bar to 70 bar 0.006%
Weight Tester
DEAD WEIGHT Float Method
(Pneumatic -
TESTER (Effective Area &
Pressure)
Pressure Method)
Using Fluke Piston
Characterization /
MECHANICAL- Gauge 2468 & as
Calibration of
PRESSURE per Euramet cg -3 &
Vacuum Dead
321 BALANCE OR OIML R 110 by Cross -0.014 bar to -1 bar 0.006%
Weight Tester
DEAD WEIGHT Float Method
(Pneumatic -
TESTER (Effective Area &
Vacuum)
Pressure Method)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
Pressure
MECHANICAL- Using Fluke Piston
Transmitter,
PRESSURE Gauge 2468 as per
322 Pressure 0.014 bar to 3.5 bar 0.006%
INDICATING DKD R 6-1 by Direct
Transducer,
DEVICES Method
Pressure Recorder,
Manometer,
Magnehelic Gauge
(Pneumatic-
Pressure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 79 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
MECHANICAL- Pressure Using Fluke Piston
PRESSURE Transmitter, Gauge 2468 as per 0.014 bar (abs) to
323 0.006%
INDICATING Pressure DKD R 6-1 & OIML R 3.5 bar (abs)
DEVICES Transducer, 97 by Direct Method
Pressure Recorder,
Barometer
(Pneumatic–Absolute
Pressure)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
MECHANICAL- Pressure Using Fluke Piston
PRESSURE Transmitter, Gauge 2468 as per 0.14 bar (abs) to 70
324 0.006%
INDICATING Pressure DKD R 6-1 & OIML R bar (abs)
DEVICES Transducer, 97 by Direct Method
Pressure Recorder,
Barometer
(Pneumatic–Absolute
Pressure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 80 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
Pressure
MECHANICAL- Using Fluke Piston
Transmitter,
PRESSURE Gauge 2468 as per
325 Pressure 0.14 bar to 70 bar 0.006%
INDICATING DKD R 6-1 by Direct
Transducer,
DEVICES Method
Pressure Recorder,
Manometer,
Magnehelic Gauge
(Pneumatic -
Pressure)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
MECHANICAL- Using DH-Budenberg
Pressure
PRESSURE Dead Weight Tester
326 Transmitter, 6 bar to 60 bar 0.0087 % rdg
INDICATING as per DKD R 6-1 by
Pressure
DEVICES Direct Method
Transducer,
Pressure Switch,
Pressure Recorder
(Hydraulic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 81 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
MECHANICAL- Using DH-Budenberg
Pressure
PRESSURE Dead Weight Tester
327 Transmitter, 60 bar to 1200 bar 0.0091 % rdg
INDICATING as per DKD R 6-1 by
Pressure
DEVICES Direct Method
Transducer,
Pressure Switch,
Pressure Recorder
(Hydraulic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
MECHANICAL- Pressure Using DH-Budenberg
PRESSURE Transmitter, Dead Weight Tester
328 0.2 bar to 25 bar 0.0082 % rdg
INDICATING Pressure as per DKD R 6-1 by
DEVICES Transducer, Direct Method
Pressure Switch,
Pressure Recorder,
Manometer
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 82 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
Pressure
Using Absolute
MECHANICAL- Transmitter,
Pressure Calibrator
PRESSURE Pressure 0 bar (abs) to 4 bar
329 as per DKD R 6-1 & 0.35 mbar (abs)
INDICATING Transducer, (abs)
OIML R-97 by
DEVICES Pressure Switch,
Comparison Method
Pressure Recorder,
Manometer,
Barometer (Absolute
Pressure -
Pneumatic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
Pressure
MECHANICAL- Using DH-Budenberg
Transmitter,
PRESSURE Dead Weight Tester 35 mbar to 1000
330 Pressure 0.0075 % rdg
INDICATING as per DKD R 6-1 by mbar
Transducer,
DEVICES Direct Method
Pressure Switch,
Pressure Recorder,
Manometer,
Magnehelic Gauge
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 83 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
MECHANICAL- Pressure Using Pressure
PRESSURE Transmitter, Calibrator as per
331 0 bar to 1000 bar 0.013 % rdg
INDICATING Pressure DKD R 6-1 by
DEVICES Transducer, Comparison Method
Pressure Switch,
Pressure Recorder
(Hydraulic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter, Using Pressure
MECHANICAL-
Pressure Calibrator as per
PRESSURE (-) 35 mbar to 35
332 Transducer, DKD R 6-1 & DKD R 0.021 mbar
INDICATING mbar
Pressure Switch, 6-2 by Comparison
DEVICES
Pressure Recorder, Method
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 84 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter,
MECHANICAL- Using Pressure
Pressure
PRESSURE Calibrator as per
333 Transducer, 0 bar to 0.6 bar 0.0125 % rdg
INDICATING DKD R 6-1 by
Pressure Switch,
DEVICES Comparison Method
Pressure Recorder,
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter,
MECHANICAL- Using Pressure
Pressure
PRESSURE Calibrator as per
334 Transducer, 0 bar to 25 bar 0.0082 % rdg
INDICATING DKD R 6-1 by
Pressure Switch,
DEVICES Comparison Method
Pressure Recorder,
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 85 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure
Gauge,Pressure
Indicator,Pressure
Calibrator,Pressure
MECHANICAL- Using Pressure
Transmitter,Pressure
PRESSURE Calibrator as per
335 Transducer,Pressure -25 mbar to 25 mbar 0.005mbar
INDICATING DKD R 6-1 by
Switch,Pressure
DEVICES Comparison Method
Recorder,Magnehelic
Gauge,Level
Gauge,Manometer(P
ressure-Pneumatic)
Vacuum Gauge,
Vacuum Indicator,
Vacuum Calibrator,
MECHANICAL- Vacuum Controller, Using Fluke Piston
PRESSURE Vacuum Transmitter, Gauge 2468 as per
336 -0.014 bar to -1 bar 0.006%
INDICATING Vacuum Transducer, DKD R 6-1 by Direct
DEVICES Vacuum Recorder, Method
Manometer
(Pneumatic -
Vacuum)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 86 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Vacuum Gauge,
Vacuum Indicator,
Vacuum Calibrator, Using Pressure
MECHANICAL-
Vacuum Controller, Calibrator as per ISO
PRESSURE
337 Vacuum Transmitter, 3567, ISO 27893 & (-)1 bar to 0 bar 0.0133 % rdg
INDICATING
Vacuum Transducer, DKD R 6-2 by
DEVICES
Vacuum Switch, Comparison Method
Vacuum Recorder
(Pneumatic)
Vacuum Gauge,
Vacuum Indicator,
Vacuum Calibrator, Using DH -
MECHANICAL- Vacuum Controller, Budenberg Dead
PRESSURE Vacuum Transmitter, Weight Tester as per (-) 35 mbar to (-)
338 0.012 % rdg
INDICATING Vacuum Transducer, ISO 3567, ISO 27893 1000 mbar
DEVICES Vacuum Switch, & DKD R 6-2 by
Vacuum Recorder, Direct Method
Manometer
(Pneumatic)
Torque Wrenches /
Screw Drivers, Using Torque Tool
MECHANICAL-
Pneumatic / Calibration machine
TORQUE
339 Electrical Screw and torque Sensors 0.02 Nm to 10 Nm 0.5%
GENERATING
Drivers/ Torque as per IS
DEVICES
Gauges (Type I & 16906-2018
Type II)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 87 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Torque Wrenches /
Screw Drivers, Using Torque Tool
MECHANICAL-
Pneumatic / Calibration machine
TORQUE
340 Electrical Screw and torque Sensors 0.02 Nm to 10 Nm 0.53 %
GENERATING
Drivers/ Torque as per ISO 6789
DEVICES
Gauges (Type I & -2017
Type II)
Torque Wrenches /
Screw Drivers,
Using Torque Tool
MECHANICAL- Torque Multipliers,
Calibration machine
TORQUE Pneumatic /
341 and torque Sensors 10 Nm to 3000 Nm 0.35%
GENERATING Electrical Screw
as per IS
DEVICES Drivers/ Torque
16906-2018
Gauges (Type I &
Type II)
Torque Wrenches /
Screw Drivers,
Using Torque Tool
MECHANICAL- Torque Multipliers,
Calibration machine
TORQUE Pneumatic /
342 and torque Sensors 10 Nm to 3000 Nm 0.63%
GENERATING Electrical Screw
as per ISO 6789
DEVICES Drivers/ Torque
-2017
Gauges (Type I &
Type II)
Using Electronic
Balance with
MECHANICAL- Bulk Density Distilled Water as
343 3000 ml to 30000 ml 4 ml
VOLUME Cylinder per IS 2386 (Part-3)
by Gravimetric
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 88 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
Balance with
MECHANICAL- Burette, Separating Distilled Water as
344 1 ml to 10 ml 0.84 µl
VOLUME Funnel per ISO 4787 &
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- Burette, Separating Distilled Water as
345 10 ml to 50 ml 4 µl
VOLUME Funnel per ISO 4787 &
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- Burette, Separating Distilled Water as
346 50 ml to 100 ml 6 µl
VOLUME Funnel per ISO 4787 &
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- Measuring Cylinder,
347 Distilled Water as 5000 ml to 10000 ml 1.7 ml
VOLUME Measuring Jar
per ISO 4787 by
Gravimetric Method
Using Electronic
Measuring Cylinder, Balance with
MECHANICAL-
348 Measuring Jar, Distilled Water as 2000 ml to 5000 ml 0.51 ml
VOLUME
Pycnometer per ISO 4787 by
Gravimetric Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 89 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Measuring Cylinder, Using Electronic


Measuring Jar, Balance with
MECHANICAL- Pycnometer, Specific Distilled Water as
349 1 ml to 10 ml 0.7 µl
VOLUME Gravity Bottle/Cup, per ISO 4787 &
Le-Chatlier Flask, ASTM E 542 by
Weight per Litre cup Gravimetric Method
Measuring Cylinder, Using Electronic
Measuring Jar, Balance with
MECHANICAL- Pycnometer, Specific Distilled Water as
350 10 ml to 100 ml 0.04 ml
VOLUME Gravity Bottle/Cup, per ISO 4787 &
Le-Chatlier Flask, ASTM E 542 by
Weight per Litre cup Gravimetric Method
Measuring Cylinder, Using Electronic
Measuring Jar, Balance with
MECHANICAL- Pycnometer, Specific Distilled Water as
351 100 ml to 2000 ml 0.13 ml
VOLUME Gravity Bottle/Cup, per ISO 4787 &
Le-Chatlier Flask, ASTM E 542 by
Weight per Litre cup Gravimetric Method
Using Electronic
Balance with
MECHANICAL-
352 Measuring Jar Distilled Water as 10000 ml 1.7 ml
VOLUME
per ISO 4787 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL-
353 Micro Pipette Distilled Water as 1 ml to 10 ml 0.64 µl
VOLUME
per ISO 8655-6 by
Gravimetric Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 90 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
Balance with
MECHANICAL-
354 Micro Pipette Distilled Water as 10 µl to 50 µl 0.08 µl
VOLUME
per ISO 8655-6 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL-
355 Micro Pipette Distilled Water as 10 ml to 25 ml 3.5 µl
VOLUME
per ISO 8655-6 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL-
356 Micro Pipette Distilled Water as 25 ml to 50 ml 4.6 µl
VOLUME
per ISO 8655-6 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- 50 µl
357 Micro Pipette Distilled Water as 0.4 µl
VOLUME to 1000 µl
per ISO 8655-6 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- Distilled Water as
358 Pipette 0.1 ml to 1 ml 0.67 µl
VOLUME per ISO 4787 &
ASTM E 542 by
Gravimetric Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 91 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
Balance with
MECHANICAL- Distilled Water as
359 Pipette 1 ml to 50 ml 3.2 µl
VOLUME per ISO 4787 &
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
MECHANICAL- Distilled Water as
360 Pipette 50 ml to 100 ml 6.5 µl
VOLUME per ISO 4787 &
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
Standard Flask,
MECHANICAL- Distilled Water as
361 Conical Flask, 1 ml to 10 ml 2.2 µl
VOLUME per ISO 4787 &
Beaker
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
Standard Flask,
MECHANICAL- Distilled Water as
362 Conical Flask, 10 ml to 50 ml 5.7 µl
VOLUME per ISO 4787 &
Beaker
ASTM E 542 by
Gravimetric Method
Using Electronic
Standard Flask, Balance with
MECHANICAL-
363 Conical Flask, Distilled Water as 2000 ml to 5000 ml 0.3 ml
VOLUME
Beaker per ISO 4787 by
Gravimetric Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 92 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
Balance with
Standard Flask,
MECHANICAL- Distilled Water as
364 Conical Flask, 50 ml to 500 ml 0.02 ml
VOLUME per ISO 4787 &
Beaker
ASTM E 542 by
Gravimetric Method
Using Electronic
Balance with
Standard Flask,
MECHANICAL- Distilled Water as
365 Conical Flask, 500 ml to 2000 ml 0.09 ml
VOLUME per ISO 4787 &
Beaker
ASTM E 542 by
Gravimetric Method
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
366 E2 Class & Coarser (Readability 0.01 1g 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
367 E2 Class & Coarser (Readability 0.01 10 g 0.02 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 93 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
368 E2 Class & Coarser (Readability 0.1 mg) 100 g 0.1 mg
WEIGHTS
as per OIML R 111,
ASTM E 617 & NIST
TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
369 E2 Class & Coarser (Readability 0.01 2g 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
370 E2 Class & Coarser (Readability 0.01 20 g 0.02 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
371 E2 Class & Coarser (Readability 0.1 mg) 200 g 0.1 mg
WEIGHTS
as per OIML R 111,
ASTM E 617 & NIST
TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 94 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
372 E2 Class & Coarser (Readability 0.01 5g 0.02 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
373 E2 Class & Coarser (Readability 0.01 50 g 0.02 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
374 F1 Class & Coarser (Readability 1 mg) 1 kg 1 mg
WEIGHTS
as per OIML R 111,
ASTM E 617 & NIST
TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
375 F1 Class & Coarser (Readability 0.01 1 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 95 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
376 F1 Class & Coarser (Readability 0.01 g) 10 kg 10 mg
WEIGHTS
as per OIML R-111,
ASTM E 617 & NIST
TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
377 F1 Class & Coarser (Readability 0.01 10 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
378 F1 Class & Coarser (Readability 0.01 100 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
379 F1 Class & Coarser (Readability 0.01 2 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 96 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
380 F1 Class & Coarser (Readability 0.01 20 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
381 F1 Class & Coarser (Readability 0.01 200 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
382 F1 Class & Coarser (Readability 0.01 5 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
383 F1 Class & Coarser (Readability 0.01 50 mg 0.01 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 97 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
384 F1 Class & Coarser (Readability 1 mg) 500 g 1 mg
WEIGHTS
as per OIML R 111,
ASTM E 617 & NIST
TG 577
Using E1 Class
Standard Weight &
Electronic Balance
MECHANICAL-
385 F1 Class & Coarser (Readability 0.01 500 mg 0.008 mg
WEIGHTS
mg) as per OIML R
111, ASTM E 617 &
NIST TG 577
Using E2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
386 F2 Class & Coarser (Readability 0.01 g) 2 kg 10 mg
WEIGHTS
as per OIML R-111,
ASTM E 617 & NIST
TG 577
Using E2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
387 F2 Class & Coarser (Readability 0.1 g) 20 kg 100mg
WEIGHTS
as per OIML R-111,
ASTM E 617 & NIST
TG 577

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 98 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
388 F2 Class & Coarser (Readability 0.01 g) 5 kg 10 mg
WEIGHTS
as per OIML R-111,
ASTM E 617 & NIST
TG 577
Using F2 Class
Standard Weight &
Electronic Balance
MECHANICAL-
389 M1 Class & Coarser (Readability 1 g) as 50 kg 1g
WEIGHTS
per OIML R-111,
ASTM E 617 & NIST
TG 577
Humidity Indicator,
Humidity
THERMAL- Transmitter, Thermo
Using Humidity
390 SPECIFIC HEAT Hygrometer, 0.5 % RH @ 23 °C 0.35 % RH
Standard
& HUMIDITY Humidity Indicator
with Sensor
(Relative Humidity)
Humidity Indicator,
Humidity
THERMAL- Transmitter, Thermo
Using Humidity
391 SPECIFIC HEAT Hygrometer, 10 % RH @ 23 °C 0.35 % RH
Standard
& HUMIDITY Humidity Indicator
with Sensor
(Relative Humidity)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 99 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Humidity Indicator, Using Temperature


Humidity & Humidity
THERMAL- Transmitter, Thermo Generator,
15 % RH to 95 % RH
392 SPECIFIC HEAT Hygrometer, Temperature & 0.92 % RH
@ 25 °C
& HUMIDITY Humidity Indicator Humidity Indicator
with Sensor with Sensor by
(Relative Humidity) Comparison Method
Humidity Indicator, Using Temperature
Humidity & Humidity
THERMAL- Transmitter, Thermo Generator,
20 % RH to 95 % RH
393 SPECIFIC HEAT Hygrometer, Temperature & 1.44 % RH
(@ 20 °C to 50 °C)
& HUMIDITY Humidity Indicator Humidity Indicator
with Sensor with Sensor by
(Relative Humidity) Comparison Method
Humidity Indicator,
Humidity
THERMAL- Transmitter, Thermo
Using Humidity
394 SPECIFIC HEAT Hygrometer, 5 % RH @ 23 °C 0.35 % RH
Standard
& HUMIDITY Humidity Indicator
with Sensor
(Relative Humidity)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 100 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Humidity Indicator,
Humidity
Transmitter, Thermo
Using Temperature
Hygrometer,
THERMAL- & Humidity
Humidity Indicator 5 °C to 50 °C @ 50
395 SPECIFIC HEAT Generator, SSPRT 0.18 °C
with Sensor, % RH
& HUMIDITY with 6½ DMM by
Temperature
Comparison Method
Indicator with
Internal Sensor
(Temperature)
Using IR
THERMAL-
396 Black Body Source Thermometer by 50 °C to 500 °C 2 °C
TEMPERATURE
Comparison Method
Using SSPRT with
THERMAL- 6½ DMM & Liquid (-) 80 °C
397 Glass Thermometer 0.08 °C
TEMPERATURE Bath by Comparison to 30 °C
Method
Using SSPRT with
THERMAL- 6½ DMM & Liquid
398 Glass Thermometer (-) 80 °C to 50 °C 0.28 °C
TEMPERATURE Bath by Comparison
Method
Using SSPRT with
THERMAL- 6½ DMM & Liquid 150 °C
399 Glass Thermometer 0.11 °C
TEMPERATURE Bath by Comparison to 250 °C
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 101 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using SSPRT with


THERMAL- 6½ DMM & Liquid 30 °C
400 Glass Thermometer 0.03°C
TEMPERATURE Bath by Comparison to 150 °C
Method
Using SSPRT with
THERMAL- 6½ DMM & Liquid
401 Glass Thermometer 50 °C to 250 °C 0.21 °C
TEMPERATURE Bath by Comparison
Method
IR Themometer, Using Black Body
THERMAL- Non-Contact Source & IR 50 °C
402 2 °C
TEMPERATURE Thermometer, Thermometer by to 500 °C
Pyrometer Comparison Method
Using R-Type
THERMAL- Temperature Bath, Thermocouple with
403 1200 °C to 1500 °C 3.75 °C
TEMPERATURE Dry Block Calibrator Indicator by
Comparison Method
Using R-Type
THERMAL- Temperature Bath, Thermocouple with
404 660 °C to 1200 °C 1.63 °C
TEMPERATURE Dry Block Calibrator Indicator by
Comparison Method
Temperature Bath, Using SSPRT with
THERMAL- (-) 60 °C
405 Liquid Bath, Dry 6½ DMM by 0.02 °C
TEMPERATURE to 150 °C
Block Calibrator Comparison Method
Temperature Bath, Using SSPRT with
THERMAL-
406 Liquid Bath, Dry 6½ DMM by (-) 80 °C to (-) 60 °C 0.12 °C
TEMPERATURE
Block Calibrator Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 102 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Temperature Bath, Using SSPRT with


THERMAL- 150 °C
407 Liquid Bath, Dry 6½ DMM by 0.3 °C
TEMPERATURE to 250 °C
Block Calibrator Comparison Method
Temperature Bath, Uing SSPRT with 6½
THERMAL-
408 Liquid Bath, Dry DMM by Comparison 250 °C to 660 °C 0.34 °C
TEMPERATURE
Block Calibrator Method
Thermometer, RTD
Sensor,
Thermocouple, Using Temperature
THERMAL- Temperature Source (LN2), SSPRT
409 (-) 196 °C 0.04 °C
TEMPERATURE Indicator / Controller with 6½ DMM by
/ Transmitter / Comparison Method
Recorder with
Sensor
Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM & Liquid (-) 30 °C
410 Indicator / Controller 0.11 °C
TEMPERATURE Bath by Comparison to 250 °C
/ Transmitter /
Method
Recorder with
Sensor,
Temperature Gauge

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 103 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM & Liquid (-) 60 °C
411 Indicator / Controller 0.05 °C
TEMPERATURE Bath by Comparison to 150 °C
/ Transmitter /
Method
Recorder with
Sensor,
Temperature Gauge
Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM, Liquid (-) 80 °C
412 Indicator / Controller 0.058 °C
TEMPERATURE Bath by Comparison to (-) 60 °C
/ Transmitter /
Method
Recorder with
Sensor,
Temperature Gauge
Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM & Dry 250 °C
413 Indicator / Controller 0.34 °C
TEMPERATURE Block Calibrator by to 660 °C
/ Transmitter /
Comparison Method
Recorder with
Sensor,
Temperature Gauge

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 104 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Thermometer, RTD
Sensor,
Thermocouple, Using R-Type
Temperature Thermocouple with
THERMAL- 660 °C
414 Indicator / Controller Indicator & Dry 1.63 °C
TEMPERATURE to 1200 °C
/ Transmitter / Block Calibrator by
Recorder with Comparison Method
Sensor,
Temperature Gauge

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 105 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Site Facility
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
1 Sourcing 1 A to 10 A 0.07 % to 0.1 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
2 Sourcing 10 µA to 100 µA 0.26 % to 0.064 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
3 Sourcing 100 µA to 100 mA 0.064%
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 106 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
4 Sourcing 100 mA to 1 A 0.064 % to 0.07 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (1 kHz to
5 Sourcing 10 A to 20 A 0.1 % to 0.21 %
Current (< 1 5 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
6 Sourcing 1 A to 10 A 0.041 % to 0.1 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 107 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
7 Sourcing 10 µA to 100 µA 0.26 % to 0.035 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
8 Sourcing 10 A to 30 A 0.1 % to 0.14 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
9 Sourcing 100 µA to 100 mA 0.035%
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 108 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 Hz to
10 Sourcing 100 mA to 1 A 0.035 % to 0.041 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Current (10 kHz
11 Sourcing 100 µA to 100 mA 0.091 % to 0.09 %
Current (< 1 to 30 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
12 Multimeter by Direct 10 µA to 100 µA 0.67 % to 0.25 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 109 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
13 Multimeter by Direct 100 µA to 3 A 0.25 % to 0.31 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Current (50 Hz to
14 Multimeter with 3 A to 20 A 0.31 % to 0.35 %
Current (< 1 1 kHz)
Shunt by V/I Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating
15 AC Current (50 Hz) Multimeter with 20 A to 30 A 0.35 % to 0.48 %
Current (< 1
Shunt by V/I Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating
16 AC Current (50 Hz) Multimeter with 30 A to 100 A 0.48 %
Current (< 1
Shunt by V/I Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 110 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
17 Probe with DMM by 1 kV to 6 kV 1.9 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
18 Probe with DMM by 15 kV to 28 kV 5.8 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using High Voltage
Alternating AC High Voltage (50
19 Probe with DMM by 6 kV to 15 kV 1.9 %
Current (< 1 Hz)
Direct Method
GHz)
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
20 Sourcing 1 mV to 10 mV 0.16 % to 0.049 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 111 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
21 Sourcing 10 mV to 100 mV 0.049 % to 0.012 %
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (1 kHz to
22 Sourcing 100 mV to 1000 V 0.012%
Current (< 1 10 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
23 Sourcing 1 mV to 10 mV 0.15 % to 0.038 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 112 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
24 Sourcing 10 mV to 100 mV 0.038 % to 0.008 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 Hz to
25 Sourcing 100 mV to 100 V 0.008%
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
26 Sourcing 1 mV to 10 mV 0.48 % to 0.36 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 113 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
27 Sourcing 10 mV to 100 mV 0.36 % to 0.064 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (10 kHz
28 Sourcing 100 mV to 100 V 0.064 % to 0.064 %
Current (< 1 to 100 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
29 Sourcing 1 mV to 10 mV 2.77 % to 2.36 %
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 114 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
30 Sourcing 10 mV to 100 mV 2.36 % to 1.27 %
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (100 kHz
31 Sourcing 100 mV to 10 V 1.27%
Current (< 1 to 1 MHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating AC Voltage (40 Hz to
32 Sourcing 100 V to 1000 V 0.009 % to 0.018 %
Current (< 1 1 kHz)
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 115 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
33 Multimeter by Direct 1 mV to 10 mV 2.46 % to 0.35 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
34 Multimeter by Direct 10 mV to 100 mV 0.35 % to 0.1 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digit
Alternating AC Voltage (50 Hz to
35 Multimeter by Direct 100 mV to 750 V 0.1 %
Current (< 1 1 kHz)
Method
GHz)
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO-
8588A by Direct
TECHNICAL-
Method (For
Alternating
36 Capacitance Sourcing 1 mF to 100 mF 0.082 % to 0.094 %
Current (< 1
Instruments) /
GHz)
Comparison Method
(Measure)
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 116 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
37 Meter by Direct / 1 nF to 100 µF 0.06 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
38 Meter by Direct / 10 pF 0.23 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Capacitance (@ 1
39 Meter by Direct / 100 pF 0.06 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating Inductance (@ 1
40 Meter by Direct & 100 µH to 10 H 0.08 %
Current (< 1 kHz)
Comparison Method
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 117 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using Precision LCR
Alternating 1 ohm to 10 kohm @
41 Resistance Meter by Direct / 0.062 % to 0.028 %
Current (< 1 1 kHz
Comparison Method
GHz)
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
42 Alternating 2.99 A to 20 A 0.08 % to 0.2 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
43 Alternating 3.29 mA to 329 mA 0.12 % to 0.06 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
44 Alternating 30 µA to 329 µA 0.53 % to 0.26 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
45 Alternating 329 µA to 3.29 mA 0.26 % to 0.12 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 118 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
AC Current (45 Hz to Calibrator Fluke
46 Alternating 329 mA to 2.99 A 0.06 % to 0.14 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
47 Alternating AC Current (50 Hz) 5522A with Current 100 A to 1000 A 0.36 %
Current (< 1 Coil 5500A by Direct
GHz) (Source) Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
48 Alternating AC Current (50 Hz) 5522A with Current 20 A to 100 A 0.87 % to 0.36 %
Current (< 1 Coil 5500A by Direct
GHz) (Source) Method
ELECTRO-
Using Multiproduct
TECHNICAL- AC Power 1Ø (50 Hz)
Calibrator Fluke
49 Alternating - Lead / Lag ± 0.2 to 0.12 W to 4800 W 1 % to 0.07 %
5522A by Direct
Current (< 1 UPF
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Power Factor 1Ø Calibrator Fluke 0.0021 PF to 0.0013
50 Alternating ± 0.2 to UPF
(50 Hz) - Lead / Lag 5522A by Direct PF
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 119 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
51 Alternating 1 mV to 3 mV 0.71 % to 0.3 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
52 Alternating 3 mV to 30 mV 0.3 % to 0.044 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
53 Alternating 30 mV to 300 mV 0.044 % to 0.02 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
54 Alternating 300 mV to 300 V 0.02 % to 0.024 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
AC Voltage (45 Hz to Calibrator Fluke
55 Alternating 300 V to 1000 V 0.024 % to 0.036 %
1 kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 120 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
56 Alternating 0.39 nF to 0.6 nF 3.57 % to 2.5 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
57 Alternating 0.6 nF to 1 nF 3.57 % to 1.7 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
58 Alternating Capacitance Box by 1 nF to 100 µF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
59 Alternating 1 nF to 7 nF 1.7 % to 0.5 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
60 Alternating Capacitance Box by 10 pF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 121 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Decade
Capacitance (@ 1
61 Alternating Capacitance Box by 100 pF 1.13 %
kHz)
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 1 Calibrator Fluke
62 Alternating 7 nF to 330 nF 0.5 % to 0.4 %
kHz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 100 Calibrator Fluke
63 Alternating 0.33 µF to 30 µF 0.47 % to 0.67 %
Hz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Capacitance (@ 100 Calibrator Fluke
64 Alternating 30 µF to 300 µF 0.67 % to 0.72 %
Hz) 5522A by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Inductance (@ 1
65 Alternating Inductance Box by 100 µH to 10 H 1.19 %
kHz)
Current (< 1 Direct Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 122 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
66 DIRECT DC Current Sourcing 0.01 µA to 0.1 µA 4.62 % to 0.47 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
67 DIRECT DC Current Sourcing 0.1 µA to 1 µA 0.47 % to 0.050 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
68 DIRECT DC Current Multimeter by Direct 0.1 µA to 10 µA 0.7 % to 0.16 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 123 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
69 DIRECT DC Current Sourcing 1 µA to 10 µA 0.050 % to 0.0071 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
70 DIRECT DC Current Sourcing 1 A to 10 A 0.023 % to 0.025%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0071 % to 0.0014
71 DIRECT DC Current Sourcing 10 µA to 100 µA
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 124 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
72 DIRECT DC Current Multimeter by Direct 10 µA to 100 µA 0.16 % to 0.09 %
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
73 DIRECT DC Current Sourcing 10 A to 30 A 0.025 % to 0.082 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0014 % to 0.0050
74 DIRECT DC Current Sourcing 10 mA to 100 mA
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 125 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
75 DIRECT DC Current Sourcing 100 µA to 10 mA 0.0014%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
76 DIRECT DC Current Multimeter by Direct 100 µA to 100 mA 0.09 % to 0.065 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
77 DIRECT DC Current Multimeter with 100 A to 500 A 0.06 % to 0.5 %
CURRENT Shunt by V/I Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
78 DIRECT DC Current Sourcing 100 mA to 1 A 0.0050 % to 0.023 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 126 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
79 DIRECT DC Current Multimeter by Direct 100 mA to 3 A 0.065 % to 0.28 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
80 DIRECT DC Current Multimeter with 3 A to 30 A 0.2 % to 0.06 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
81 DIRECT DC Current Multimeter with 30 A to 100 A 0.14 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
82 DIRECT DC Current Multimeter with 500 A to 1000 A 0.5 %
CURRENT Shunt by V/I Method
(Measure)
ELECTRO-
TECHNICAL- Using High Voltage
83 DIRECT DC High Voltage Probe with DMM by 1 kV to 6 kV 1.3 % to 1.6 %
CURRENT Direct method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 127 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using High Voltage
84 DIRECT DC High Voltage Probe with DMM by 15 kV to 40 kV 1.6 % to 2.6 %
CURRENT Direct Method
(Measure)
ELECTRO-
TECHNICAL- Using High Voltage
85 DIRECT DC High Voltage Probe with DMM by 6 kV to 15 kV 1.6 %
CURRENT Direct method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
86 DIRECT DC Resistance Multimeter by Direct 0.1 ohm to 1 ohm 4.6 % to 0.47 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
87 DIRECT DC Resistance Multimeter by Direct 1 Mohm to 10 Mohm 0.02 % to 0.05 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
1 ohm
88 DIRECT DC Resistance Multimeter by Direct 0.47 % to 0.06 %
to 10 ohm
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 128 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digit
10 Mohm to 100
89 DIRECT DC Resistance Multimeter by Direct 0.05 % to 0.93 %
Mohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
10 ohm
90 DIRECT DC Resistance Multimeter by Direct 0.06 % to 0.02 %
to 100 ohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 Mohm to 1
91 DIRECT DC Resistance Multimeter by Direct 0.93 % to 3.5 %
Gohm
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 ohm
92 DIRECT DC Resistance Multimeter by Direct 0.02 %
to 1 Mohm
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
93 DIRECT DC Voltage Sourcing 0.01 mV to 0.1 mV 2.35 % to 0.24 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 129 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
94 DIRECT DC Voltage Sourcing 0.1 mV to 1 mV 0.24 % to 0.024 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
95 DIRECT DC Voltage Multimeter by Direct 0.1 mV to 1 mV 4.1 % to 0.41 %
CURRENT Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
96 DIRECT DC Voltage Sourcing 1 mV to 10 mV 0.024 % to 0.0028 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
97 DIRECT DC Voltage Multimeter by Direct 1 mV to 10 mV 0.41 % to 0.05 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 130 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00036 % to
98 DIRECT DC Voltage Sourcing 1 V to 10 V
0.00032 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0028 % to 0.00063
99 DIRECT DC Voltage Sourcing 10 mV to 100 mV
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
100 DIRECT DC Voltage Multimeter by Direct 10 mV to 1000 V 0.05 % to 0.0054 %
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 131 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00032 % to
101 DIRECT DC Voltage Sourcing 10 V to 100 V
0.00051 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00063 % to
102 DIRECT DC Voltage Sourcing 100 mV to 1 V
0.00036 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00051 % to
103 DIRECT DC Voltage Sourcing 100 V to 1000 V
0.00054 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 132 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.1 mohm to 1
104 DIRECT Resistance Sourcing 4.85 % to 0.49 %
mohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
105 DIRECT Resistance Sourcing 1 Gohm to 10 Gohm 0.017 % to 0.18 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0010 % to
106 DIRECT Resistance Sourcing 1 Mohm to 10 Mohm
0.0012%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 133 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
107 DIRECT Resistance Sourcing 1 mohm to 10 mohm 0.49 % to 0.048 %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0016 % to 0.0010
108 DIRECT Resistance Sourcing 1 ohm to 10 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
10 Mohm to 100 0.0012 % to 0.0031
109 DIRECT Resistance Sourcing
Mohm %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 134 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
10 mohm to 100
110 DIRECT Resistance Sourcing 0.048 % to 0.0063 %
mohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0010 % to 0.0009
111 DIRECT Resistance Sourcing 10 ohm to 100 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 kohm to 1 0.0009 % to 0.0010
112 DIRECT Resistance Sourcing
Mohm %
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 135 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 Mohm to 1
113 DIRECT Resistance Sourcing 0.0031 % to 0.017 %
Gohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0063 % to 0.0016
114 DIRECT Resistance Sourcing 100 mohm to 1 ohm
%
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
100 ohm to 100
115 DIRECT Resistance Sourcing 0.0009%
kohm
CURRENT Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 136 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
116 DIRECT DC Current 10 µA to 100 µA 0.25 % to 0.04 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
117 DIRECT DC Current 100 µA to 329 µA 0.04 % to 0.024 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
118 DIRECT DC Current 5522A with Current 100 A to 1000 A 0.32 % to 0.3 %
CURRENT Coil 5500A by Direct
(Source) Method
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
119 DIRECT DC Current 2.99 A to 20 A 0.046 % to 0.12 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
120 DIRECT DC Current 5522A with Current 20 A to 100 A 0.53 % to 0.32 %
CURRENT Coil 5500A by Direct
(Source) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 137 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
121 DIRECT DC Current 3.29 mA to 329 mA 0.014 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
122 DIRECT DC Current 329 µA to 3.29 mA 0.024 % to 0.014 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
123 DIRECT DC Current 329 mA to 2.99 A 0.014 % to 0.046 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Fluke 5522A
TECHNICAL-
Multiproduct
124 DIRECT DC Power 1 W to 19950 W 0.03 % to 0.08 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
125 DIRECT DC Resistance Resistance Box by 0.01 ohm to 1 ohm 2.4 % to 0.58 %
CURRENT Direct Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 138 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Decade
126 DIRECT DC Resistance Resistance Box by 1 ohm to 10 ohm 0.58 % to 0.05 %
CURRENT Direct Method
(Source)
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
127 DIRECT DC Resistance 5522A & Decade 10 ohm to 30 ohm 0.05 % to 0.011 %
CURRENT Resistance box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
290 Mohm to 1
128 DIRECT DC Resistance 5522A & Decade 0.4 % to 1.8 %
Gohm
CURRENT Meg Ohm Box by
(Source) Direct Method
Using Multiproduct
ELECTRO- Calibrator Fluke
TECHNICAL- 5522A, Decade
129 DIRECT DC Resistance Resistance Box & 3 kohm to 300 kohm 0.011 % to 0.008 %
CURRENT Decade Meg Ohm
(Source) Box by Direct
Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
130 DIRECT DC Resistance 5522A & Decade 3 Mohm to 30 Mohm 0.013 % to 0.04 %
CURRENT Meg Ohm Box by
(Source) Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 139 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Using Multiproduct


TECHNICAL- Calibrator Fluke
30 Mohm to 290
131 DIRECT DC Resistance 5522A & Decade 0.04 % to 0.4 %
Mohm
CURRENT Meg Ohm Box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
30 ohm
132 DIRECT DC Resistance 5522A & Decade 0.011 %
to 3 kohm
CURRENT Resistance Box by
(Source) Direct Method
ELECTRO- Using Multiproduct
TECHNICAL- Calibrator Fluke
300 kohm to 3
133 DIRECT DC Resistance 5522A & Decade 0.008 % to 0.013 %
Mohm
CURRENT Meg Ohm Box by
(Source) Direct Method
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
134 DIRECT DC Voltage 0.1 mV to 1 mV 1.2 % to 0.12 %
5522A by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
135 DIRECT DC Voltage 1 mV to 329 mV 0.12 % to 0.0023 %
5522A by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 140 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.0023 % to 0.0016
136 DIRECT DC Voltage 329 mV to 1000 V
5522A by Direct %
CURRENT
Method
(Source)
ELECTRO-
Using Decade
TECHNICAL-
Megohm Box 1 Gohm to 100
137 DIRECT Resistance 2.44 % to 5.85 %
8400HV by Direct Gohm
CURRENT
Method
(Source)
Oscilloscope -
ELECTRO-
Amplitude (Vertical Multiproduct
TECHNICAL-
Axis Deflection Calibrator Fluke
138 ELECTRICAL 1 mV to 130 V 4.74 % to 0.13 %
Factor) AC: 1 kHz 5522A, SC600 &
EQUIPMENT
Square Wave @ 1 Direct Method
(Source)
Mohm
ELECTRO-
Using Multiproduct
TECHNICAL-
Oscilloscope - Calibrator Fluke 5 mV to 5.5 V @ (50
139 ELECTRICAL 3.63 % to 5.11 %
Bandwidth 5522A & SC600 by kHz to 500 MHz)
EQUIPMENT
Direct Method
(Source)
ELECTRO-
Oscilloscope - Time Using Multiproduct
TECHNICAL-
Base (Horizontal Calibrator Fluke
140 ELECTRICAL 2 ns to 5 s 0.5 % to 0.12 %
Axis Deflection 5522A & SC600 by
EQUIPMENT
Factor) Direct Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 141 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Oscilloscope
Using Multiproduct
TECHNICAL- Amplitude (Vertical
Calibrator Fluke
141 ELECTRICAL Axis Deflection 1 mV to 130 V 4.68 % to 0.058 %
5522A & SC600 by
EQUIPMENT Factor) DC @ 1
Direct Method
(Source) Mohm
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
B-Type
142 TEMPERATURE Sourcing 600 °C to 1820 °C 0.044°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
B-Type Calibrator Fluke
143 TEMPERATURE 600 °C to 1820 °C 0.6 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
E-Type Calibrator Fluke (-) 200 °C to 1000
144 TEMPERATURE 0.12 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 142 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
E-Type
145 TEMPERATURE Sourcing -200 °C to 1000 °C 0.011°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
J-Type Calibrator Fluke (-) 200 °C to 1200
146 TEMPERATURE 0.14 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
J-Type
147 TEMPERATURE Sourcing -200 °C to 1200 °C 0.014°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
K-Type Calibrator Fluke (-) 200 °C to 1370
148 TEMPERATURE 0.21 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 143 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
K-Type
149 TEMPERATURE Sourcing -200 °C to 1370 °C 0.035°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
N-Type Calibrator Fluke (-) 200 °C to 1300
150 TEMPERATURE 0.17 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
N-Type
151 TEMPERATURE Sourcing -250 °C to 1300 °C 0.06°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
R-Type Calibrator Fluke
152 TEMPERATURE 50 °C to 1760 °C 0.59 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 144 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
R-Type
153 TEMPERATURE Sourcing 50 °C to 1765 °C 0.05°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
154 TEMPERATURE RTD PT 100 Multimeter by Direct (-) 200 °C to 600 °C 0.18 °C
SIMULATION Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
155 TEMPERATURE RTD Pt 100 Sourcing -200 °C to 800 °C 0.004°C
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
S-Type Calibrator Fluke
156 TEMPERATURE 50 °C to 1760 °C 0.58 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 145 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
S-Type
157 TEMPERATURE Sourcing 50 °C to 1765 °C 0.045°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
T-Type Calibrator Fluke
158 TEMPERATURE (-) 100 °C to 400 °C 0.11 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
T-Type
159 TEMPERATURE Sourcing -100 °C to 400 °C 0.086°C
Thermocouple
SIMULATION Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
Using Multiproduct
TECHNICAL-
B-Type Calibrator Fluke
160 TEMPERATURE 600 °C to 1820 °C 0.6 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 146 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
E-Type Calibrator Fluke
161 TEMPERATURE (-) 200 °C to 950 °C 0.12 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
J-Type Calibrator Fluke (-) 200 °C to 1150
162 TEMPERATURE 0.14 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
K-Type Calibrator Fluke (-) 200 °C to 1370
163 TEMPERATURE 0.2 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
N-Type Calibrator Fluke (-) 100 °C to 1250
164 TEMPERATURE 0.17 °C
Thermocouple 5522A by Direct °C
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
R-Type Calibrator Fluke
165 TEMPERATURE 5 °C to 1750 °C 0.53 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 147 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke
166 TEMPERATURE RTD PT 100 (-) 200 °C to 800 °C 0.11 °C
5522A by Direct
SIMULATION
method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
S-Type Calibrator Fluke
167 TEMPERATURE 5 °C to 1750 °C 0.58 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
T-Type Calibrator Fluke
168 TEMPERATURE (-) 100 °C to 350 °C 0.11 °C
Thermocouple 5522A by Direct
SIMULATION
Method
(Source)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.0013 % to 0.00035
169 TIME & Frequency Sourcing 1 Hz to 10 Hz
%
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 148 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00012 % to
170 TIME & Frequency Sourcing 1 kHz to 100 MHz
0.00012%
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
171 TIME & Frequency Multimeter by Direct 10 Hz to 100 Hz 0.05 % to 0.01 %
FREQUENCY Method
(Measure)
Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00035 % to
172 TIME & Frequency Sourcing 10 Hz to 100 Hz
0.00012 %
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
173 TIME & Frequency Multimeter by Direct 100 Hz to 300 kHz 0.01 %
FREQUENCY Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 149 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using 8½ Digit
Multimeter Fluke
ELECTRO- 8588A by Direct
TECHNICAL- Method (For
0.00012 % to
174 TIME & Frequency Sourcing 100 Hz to 1 kHz
0.00012 %
FREQUENCY Instruments) /
(Measure) Comparison Method
(For Measuring
Instruments)
ELECTRO-
TECHNICAL- Using 6½ Digit
175 TIME & Frequency Multimeter by Direct 3 Hz to 10 Hz 0.081 % to 0.05 %
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Frequency
176 TIME & Frequency, Period Counter by Direct 1 Hz to 300 MHz 2 ppm
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Time Totaliser
177 TIME & Timer, Stopwatch by Comparison 1 s to 1000 s 0.0022 s to 0.063 s
FREQUENCY Method
(Measure)
ELECTRO-
TECHNICAL- Using Time Totaliser
178 TIME & Timer, Stopwatch by Comparison 1000 s to 5000 s 0.063 s to 0.25 s
FREQUENCY Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 150 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Time Totaliser
179 TIME & Timer, Stopwatch by Comparison 5000 s to 86400 s 0.25 s to 1 s
FREQUENCY Method
(Measure)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.0011 % to 0.00051
180 TIME & Frequency 1 Hz to 1000 Hz
5522A by Direct %
FREQUENCY
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.00065 % to 0.029
181 TIME & Frequency 100 kHz to 2 MHz
5522A by Direct %
FREQUENCY
Method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator Fluke 0.00051 % to
182 TIME & Frequency 1000 Hz to 100 kHz
5522A by Direct 0.00065 %
FREQUENCY
Method
(Source)
Centrifuge, RPM, Using Digital
MECHANICAL-
Speed (Indicator, Tachometer as per 1000 RPM to 10000
183 ACCELERATION 0.8 RPM
Sensor, Meter, SANAS TR 45-02 by RPM
AND SPEED
Source) Comparison Method
Centrifuge, RPM, Using Digital
MECHANICAL-
Speed (Indicator, Tachometer as per 10000 RPM to 30000
184 ACCELERATION 1.5 RPM
Sensor, Meter, SANAS TR 45-02 by RPM
AND SPEED
Source) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 151 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Centrifuge, RPM, Using Digital


MECHANICAL-
Speed (Indicator, Tachometer as per
185 ACCELERATION 6 RPM to 1000 RPM 0.16 RPM
Sensor, Meter, SANAS TR 45-02 by
AND SPEED
Source) Comparison Method
MECHANICAL-
DIMENSION
2D Electronic Height Using Length Bar Set
(BASIC
186 Gauge (L.C.: 0.001 & Slip Gauge Set by 0 to 600 mm 6.16 µm
MEASURING
mm) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
187 by Comparison 0 to 25 mm 0.4 µm
MEASURING with Indicator (L.C.:
Method
INSTRUMENT, 0.0001 mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
188 by Comparison 0 to 50 mm 1.5 µm
MEASURING with Indicator (L.C.:
Method
INSTRUMENT, 0.001 mm)
GAUGE ETC.)
MECHANICAL-
DIMENSION LVDT Probe,
Using Slip Gauge Set
(BASIC Electronic Probe
189 & Length Bar Set by 50 mm to 600 mm 6.7 µm
MEASURING with Indicator (L.C.:
Comparison Method
INSTRUMENT, 0.001 mm)
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 152 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Electronic
(BASIC Level as per IS 2220 2.3 x sqrt (L/150)µm
190 Straight Edge Up to 5000 mm
MEASURING & IS 12937 by (where L in mm)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Electronic
1.85 x sqrt {(L+W) /
(BASIC Level as per IS 7327 3000 mm x 5000
191 Surface Plate / Table 150} µm (L & W are
MEASURING & IS 12937 by mm
in mm)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL- Using Caliper
DIMENSION Checker & Length
(BASIC Vernier Caliper (L.C.: Bar Set as per IS
192 0 to 2000 mm 24.4 µm
MEASURING 0.02 mm) 3651 (Part 1, 2, 3)
INSTRUMENT, by Comparison
GAUGE ETC.) Method
MECHANICAL- Profile Projector, Using Angular
DIMENSION Microscope, Video Graticules as per JIS
193 0 ° to 360 ° 0.4 minute of arc
(PRECISION Measuring Machine B 7184 by
INSTRUMENTS) (Angular) Comparison Method
MECHANICAL- Profile Projector,
Using Glass Scale as
DIMENSION Microscope, Video
194 per JIS B 7184 by 300 mm x 200 mm 2.8 µm
(PRECISION Measuring Machine
Comparison Method
INSTRUMENTS) (Linear)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 153 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Profile Projector,


Using Glass Scale as
DIMENSION Microscope, Video
195 per JIS B 7184 by Up to 100 X 0.3 %
(PRECISION Measuring Machine
Comparison Method
INSTRUMENTS) (Magnification)
Using Slip Gauge
MECHANICAL-
Set, Length Bar Set
DIMENSION Scale and Tape
196 & Lever Dial Gauge 0 to 1000 mm 10.2 µm
(PRECISION Calibrator
by Comparison
INSTRUMENTS)
Method
IRHD Rubber
Hardness Tester
MECHANICAL- As per ASTM D 1415
197 (Indent Length 0 to 100 IRHD 3.09 µm
DUROMETER & ISO 48-9
Measurement
Method)
IRHD Rubber
Hardness Tester
MECHANICAL- As per ASTM D 1415 Minor Load 0.3 N to
198 (Spring Force 2.16 mN to 3.14 mN
DUROMETER & ISO 48-9 Total Load 5.7 N
Measuement
Method)
Using Precision
Direct Verification of Length Measuring
MECHANICAL-
Brinell Hardness Instrument as per
HARDNESS
199 Tester (Calibration ISO 6506-2, IS 0 to 7 mm 0.004 mm
TESTING
of Diameter 1500-2, ASTM E 10
MACHINES
Measuring System) & JIS B 7724 by
Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 154 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Load Cell with


Direct Verification of
MECHANICAL- Indicator as per ISO
Brinell Hardness
HARDNESS 6506-2, IS 1500-2,
200 Tester (Calibration 980.7 N to 29.42 kN 0.2 %
TESTING ASTM E 10 & JIS B
of Test Force - Test
MACHINES 7724 by Direct
Load)
Method
Using Precision
Direct Verification of Length Measuring
Micro Vickers, Instrument as per
MECHANICAL-
Vickers, Knoop ISO 6507-2, IS
HARDNESS
201 Hardness Tester 1501-2, IS 1754, ISO 0 to 2 mm 0.004 mm
TESTING
(Calibration of 4545-2, IS 6885-2,
MACHINES
Diameter Measuring ASTM E 384 & JIS B
System) 7725 by Direct
Method
Using Load Cell with
Direct Verification of
Indicator & ISO
MECHANICAL- Micro Vickers,
6507-2, IS 1501-2, IS
HARDNESS Vickers, Knoop
202 1754, ISO 4545-2, IS 0.1 N to 49.03 N 0.3 %
TESTING Hardness Tester
6885-2, ASTM E 384
MACHINES (Calibration of Test
& JIS B 7725 by
Force - Test Load)
Direct Method
Using Load Cell with
Direct Verification of
Indicator as per ISO
MECHANICAL- Micro Vickers,
6507-2, IS 1501-2, IS
HARDNESS Vickers, Knoop
203 1754, ISO 4545-2, IS 49.03 N to 490.3 N 0.2 %
TESTING Hardness Tester
6885-2, ASTM E 384
MACHINES (Calibration of Test
& JIS B 7725 by
Force - Test Load)
Direct Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 155 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Direct Verification of Using Precision


Rockwell Hardness Length Measuring
MECHANICAL-
Tester, Rockwell Instruments as per
HARDNESS
204 Superficial Hardness ISO 6508-2, IS 0 to 0.3 mm 0.001 mm
TESTING
Tester (Calibration 1586-2, ASTM E 18
MACHINES
of Depth Measuring & JIS B 7726 by
System) Direct Method
Direct Verification of Using Precision
Rockwell Hardness Length Measuring
MECHANICAL-
Tester, Rockwell Instruments as per
HARDNESS
205 Superficial Hardness ISO 6508-2, IS 0.3 mm to 0.8 mm 0.002 mm
TESTING
Tester (Calibration 1586-2, ASTM E 18
MACHINES
of Depth Measuring & JIS B 7726 by
System) Direct Method
Direct Verification of
Using Load Cell with
Rockwell Hardness
MECHANICAL- Indicator as per ISO
Tester, Rockwell
HARDNESS 6508-2, IS 1586-2,
206 Superficial Hardness 29.42 N & 98.07 N 0.2 %
TESTING ASTM E 18 & JIS B
Tester (Calibration
MACHINES 7726 by Direct
of Test Force -
Method
Preload)
Direct Verification of
Using Load Cell with
Rockwell Hardness
MECHANICAL- Indicator as per ISO
Tester, Rockwell
HARDNESS 6508-2, IS 1586-2,
207 Superficial Hardness 147.1 N to 1471 N 0.2 %
TESTING ASTM E 18 & JIS B
Tester (Calibration
MACHINES 7726 by Direct
of Test Force - Test
Method
Load)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 156 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Load Cell with


Indicator, Precision
Length & Angle
Measuring
MECHANICAL- Verification of Instruments,
IMPACT Impact Testing Stopwatch as per IS
208 Up to 450 J 0.77 %
TESTING Machine (Izod & 3766 , ISO 148, ISO
MACHINE Charpy) 180, ISO 13802, BS
131-4, ASTM E 23,
ASTM D 6110 &
ASTM D 256 by
Direct Method
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure Controller,
Pressure
Using Absolute
MECHANICAL- Transmitter,
Pressure Calibrator
PRESSURE Pressure 0 bar (abs) to 4 bar
209 as per DKD R 6-1 & 0.35 mbar (abs)
INDICATING Transducer, (abs)
OIML R-97 by
DEVICES Pressure Switch,
Comparison Method
Pressure Recorder,
Manometer,
Barometer (Absolute
Pressure -
Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 157 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
MECHANICAL- Pressure Using Pressure
PRESSURE Transmitter, Calibrator as per
210 0 bar to 1000 bar 0.013 % rdg
INDICATING Pressure DKD R 6-1 by
DEVICES Transducer, Comparison Method
Pressure Switch,
Pressure Recorder
(Hydraulic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter, Using Pressure
MECHANICAL-
Pressure Calibrator as per
PRESSURE (-) 35 mbar to 35
211 Transducer, DKD R 6-1 & DKD R 0.021 mbar
INDICATING mbar
Pressure Switch, 6-2 by Comparison
DEVICES
Pressure Recorder, Method
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 158 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter,
MECHANICAL- Using Pressure
Pressure
PRESSURE Calibrator as per
212 Transducer, 0 bar to 0.6 bar 0.0125 % rdg
INDICATING DKD R 6-1 by
Pressure Switch,
DEVICES Comparison Method
Pressure Recorder,
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)
Pressure Gauge,
Pressure Indicator,
Pressure Calibrator,
Pressure
Transmitter,
MECHANICAL- Using Pressure
Pressure
PRESSURE Calibrator as per
213 Transducer, 0 bar to 25 bar 0.0082 % rdg
INDICATING DKD R 6-1 by
Pressure Switch,
DEVICES Comparison Method
Pressure Recorder,
Magnehelic Gauge,
Level Gauge,
Manometer
(Pneumatic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 159 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure
Gauge,Pressure
Indicator,Pressure
Calibrator,Pressure
MECHANICAL- Using Pressure
Transmitter,Pressure
PRESSURE Calibrator as per
214 Transducer,Pressure -25 mbar to 25 mbar 0.005mbar
INDICATING DKD R 6-1 by
Switch,Pressure
DEVICES Comparison Method
Recorder,Magnehelic
Gauge,Level
Gauge,Manometer(P
ressure-Pneumatic)
Vacuum Gauge,
Vacuum Indicator,
Vacuum Calibrator, Using Pressure
MECHANICAL-
Vacuum Controller, Calibrator as per ISO
PRESSURE
215 Vacuum Transmitter, 3567, ISO 27893 & (-)1 bar to 0 bar 0.0133 % rdg
INDICATING
Vacuum Transducer, DKD R 6-2 by
DEVICES
Vacuum Switch, Comparison Method
Vacuum Recorder
(Pneumatic)
Compression /
MECHANICAL- Using Newton
Universal Testing
UTM, TENSION Weights as per IS
Machine, Spring
CREEP AND 1828 (Part-1), ISO
216 Testing Machine, 0.1 N to 2 N 0.58 %
TORSION 7500-1, ASTM E 4 &
Flexural Testing
TESTING JIS B 7721 by
Machine
MACHINE Comparison Method
(Compression Mode)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 160 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Compression / Using Newton


MECHANICAL-
Universal Testing Weights & Load Cell
UTM, TENSION
Machine, Spring with Indicator as per
CREEP AND
217 Testing Machine, IS 1828 (Part-1), ISO 10 kN to 100 kN 0.23 %
TORSION
Flexural Testing 7500-1, ASTM E 4 &
TESTING
Machine JIS B 7721 by
MACHINE
(Compression Mode) Comparison Method
Compression / Using Newton
MECHANICAL-
Universal Testing Weights & Load Cell
UTM, TENSION
Machine, Spring with Indicator as per
CREEP AND
218 Testing Machine, IS 1828 (Part-1), ISO 100 kN to 2000 kN 0.16 %
TORSION
Flexural Testing 7500-1 & JIS B 7721
TESTING
Machine by Comparison
MACHINE
(Compression Mode) Method
Compression /
MECHANICAL-
Universal Testing Using Newton
UTM, TENSION
Machine, Spring Weights & Load Cell
CREEP AND
219 Testing Machine, with Indicator as per 100 kN to 500 kN 0.16 %
TORSION
Flexural Testing ASTM E 4 by
TESTING
Machine Comparison Method
MACHINE
(Compression Mode)
Compression / Using Newton
MECHANICAL-
Universal Testing Weights & Load Cell
UTM, TENSION
Machine, Spring with Indicator as per
CREEP AND
220 Testing Machine, IS 1828 (Part-1), ISO 2 N to 10 kN 0.1 %
TORSION
Flexural Testing 7500-1, ASTM E 4 &
TESTING
Machine JIS B 7721 by
MACHINE
(Compression Mode) Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 161 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Tensile / Universal
Using Newton
MECHANICAL- Testing Machine,
Weights & Load Cell
UTM, TENSION Spring Testing
with Indicator as per
CREEP AND Machine,
221 IS 1828 (Part-1), ISO 0.1 N to 2 N 0.58 %
TORSION Tensometer,
7500-1, ASTM E 4 &
TESTING Flexural Testing
JIS B 7721 by
MACHINE Machine (Tension
Comparison Method
Mode)
Tensile / Universal
Using Newton
MECHANICAL- Testing Machine,
Weights & Load Cell
UTM, TENSION Spring Testing
with Indicator as per
CREEP AND Machine,
222 IS 1828 (Part-1), ISO 2 N to 250 kN 0.14 %
TORSION Tensometer,
7500-1, ASTM E 4 &
TESTING Flexural Testing
JIS B 7721 by
MACHINE Machine (Tension
Comparison Method
Mode)
Tensile / Universal
MECHANICAL- Testing Machine, Using Load Cell with
UTM, TENSION Spring Testing Indicator as per IS
CREEP AND Machine, 1828 (Part-1), ISO
223 250 kN to 500 kN 0.4 %
TORSION Tensometer, 7500-1, ASTM E 4 &
TESTING Flexural Testing JIS B 7721 by
MACHINE Machine (Tension Comparison Method
Mode)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 162 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Verification of
Using Precision
UTM, TENSION Displacement
Length Measuring
CREEP AND Measuring System &
224 Instruments as per 0 to 50 mm 0.0013 mm
TORSION Devices used in
ASTM E 2309 by
TESTING Material Testing
Comparison Method
MACHINE Machine
MECHANICAL- Verification of
Using Precision
UTM, TENSION Displacement
Length Measuring
CREEP AND Measuring System &
225 Instruments as per 200 mm to 600 mm 0.0051 mm
TORSION Devices used in
ASTM E 2309 by
TESTING Material Testing
Comparison Method
MACHINE Machine
MECHANICAL- Verification of
Using Precision
UTM, TENSION Displacement
Length Measuring
CREEP AND Measuring System &
226 Instruments as per 50 mm to 200 mm 0.003 mm
TORSION Devices used in
ASTM E 2309 by
TESTING Material Testing
Comparison Method
MACHINE Machine
MECHANICAL- Verification of
Using Precision
UTM, TENSION Displacement
Length Measuring
CREEP AND Measuring System & 600 mm to 1500
227 Instruments as per 0.012 mm
TORSION Devices used in mm
ASTM E 2309 by
TESTING Material Testing
Comparison Method
MACHINE Machine

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 163 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Precision


UTM, TENSION Verification of Length Measuring
CREEP AND Extensometer used Instruments as per
228 0 to 50 mm 0.0011 mm
TORSION in Uniaxial Testing ASTM E 83, IS
TESTING Machine 12872, ISO 9513 by
MACHINE Comparison Method
MECHANICAL- Using Precision
UTM, TENSION Verification of Length Measuring
CREEP AND Extensometer used Instruments as per 200 mm to 1000
229 0.15 mm
TORSION in Uniaxial Testing ASTM E 83, IS mm
TESTING Machine 12872, ISO 9513 by
MACHINE Comparison Method
MECHANICAL- Using Precision
UTM, TENSION Verification of Length Measuring
CREEP AND Extensometer used Instruments as per
230 50 mm to 200 mm 0.003 mm
TORSION in Uniaxial Testing ASTM E 83, ISO
TESTING Machine 12872, ISO 9513 by
MACHINE Comparison Method
MECHANICAL- Using Precision
UTM, TENSION Length Measuring
Verification of Speed
CREEP AND Instruments &
231 for Material Testing 0 to 50 mm/minute 0.0025 mm/minute
TORSION Stopwatch as per
Machine
TESTING ASTM E 2658 by
MACHINE Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 164 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Precision


UTM, TENSION Length Measuring
Verification of Speed
CREEP AND Instruments & 1000 mm/minute to
232 for Material Testing 0.022 mm/minute
TORSION Stopwatch as per 1500 mm/minute
Machine
TESTING ASTM E 2658 by
MACHINE Comparison Method
MECHANICAL- Using Precision
UTM, TENSION Length Measuring
Verification of Speed
CREEP AND Instruments & 200 mm/minute to
233 for Material Testing 0.016 mm/minute
TORSION Stopwatch as per 1000 mm/minute
Machine
TESTING ASTM E 2658 by
MACHINE Comparison Method
MECHANICAL- Using Precision
UTM, TENSION Length Measuring
Verification of Speed
CREEP AND Instruments & 50 mm/minute to
234 for Material Testing 0.005 mm/minute
TORSION Stopwatch as per 200 mm/minute
Machine
TESTING ASTM E 2658 by
MACHINE Comparison Method
MECHANICAL- Weighing Scale & Using E1 Class
WEIGHING Balance Standard Weights as
235 Up to 21 g 0.015 mg
SCALE AND (Readability: 0.001 per OIML R 76, ASTM
BALANCE mg) E 898
MECHANICAL- Weighing Scale & Using E1 Class
WEIGHING Balance Standard Weights as
236 Up to 220 g 0.08 mg
SCALE AND (Readability: 0.01 per OIML R 76 &
BALANCE mg) ASTM E 898

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 165 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using E2 Class


Weighing Scale &
WEIGHING Standard Weights as
237 Balance Up to 22 kg 0.1 g
SCALE AND per OIML R 76 &
(Readability: 0.1 g)
BALANCE ASTM E 898
MECHANICAL- Using E2 & F1 Class
Weighing Scale &
WEIGHING Standard Weights as
238 Balance Up to 60 kg 1g
SCALE AND per OIML R 76 &
(Readability: 1 g)
BALANCE ASTM E 898
MECHANICAL- Using E1 Class
Weighing Scale &
WEIGHING Standard Weights as
239 Balance Up to 1 kg 1 mg
SCALE AND per OIML R 76 &
(Readability: 1 mg)
BALANCE ASTM E 898
MECHANICAL- Using E2, F1 & F2
Weighing Scale &
WEIGHING Class Standard
240 Balance Up to 300 kg 10 g
SCALE AND Weights as per OIML
(Readability: 10 g)
BALANCE R 76 & ASTM E 898
MECHANICAL- Using E1 & E2 Class
Weighing Scale &
WEIGHING Standard Weights as
241 Balance Up to 10 kg 10 mg
SCALE AND per OIML R 76 &
(Readability: 10 mg)
BALANCE ASTM E 898
MECHANICAL- Using E1 & E2 Class
Weighing Scale &
WEIGHING Standard Weights as
242 Balance Up to 6 kg 10 mg
SCALE AND per OIML R 76 &
(Readability: 10 mg)
BALANCE ASTM E 898
MECHANICAL- Using F2 & M1 Class
Weighing Scale &
WEIGHING Standard Weights as
243 Balance Up to 1000 kg 100 g
SCALE AND per OIML R 76 &
(Readability: 100 g)
BALANCE ASTM E 898

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 166 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using F2 & M1 Class


Weighing Scale &
WEIGHING Standard Weights as
244 Balance Up to 2000 kg 200 g
SCALE AND per OIML R 76 &
(Readability: 200 g)
BALANCE ASTM E 898
MECHANICAL- Using E2, F1 & F2
Weighing Scale &
WEIGHING Class Standard
245 Balance Up to 100 kg 5g
SCALE AND Weights as per OIML
(Readability: 5 g)
BALANCE R 76 & ASTM E 898
MECHANICAL- Using F1, F2 & M1
Weighing Scale &
WEIGHING Class Standard
246 Balance Up to 600 kg 50 g
SCALE AND Weights as per OIML
(Readability: 50 g)
BALANCE R 76 & ASTM E 898
Humidity Chamber, Using Temperature
THERMAL-
Environmental & Humidity Data 20 % RH to 95 % RH
247 SPECIFIC HEAT 2.4 % RH
Chamber, Climatic Logger (Multi (@ 20°C to 50 °C)
& HUMIDITY
Chamber Position Calibration)
Humidity Indicator /
Recorder / Controller
Using Temperature
THERMAL- with Sensor of
& Humidity Indicator 5 % RH to 15 % RH
248 SPECIFIC HEAT Humidity Chamber, 0.92 % RH
with Sensor (Single @ 25 °C
& HUMIDITY Environmental
Position Calibration)
Chamber, Climatic
Chamber

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 167 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Humidity Indicator,
Recorder, Controller
Using Temperature
THERMAL- with Sensor of
& Humidity Indicator 15 % RH to 95 % RH
249 SPECIFIC HEAT Humidity Chamber, 0.92 % RH
with Sensor (Single @ 25 °C
& HUMIDITY Environmental
Position Calibration)
Chamber, Climatic
Chamber
Humidity Indicator,
Recorder, Controller
Using Temperature
THERMAL- with Sensor of
& Humidity Indicator 20 % RH to 95 % RH
250 SPECIFIC HEAT Humidity Chamber, 1.44 % RH
with Sensor (Single (@ 20°C to 50 °C)
& HUMIDITY Environmental
Position Calibration)
Chamber, Climatic
Chamber
Using N-Type
Autoclave, Chamber,
THERMAL- Thermocouples with 100 °C
251 Water Bath, Hot Air 2.48 °C
TEMPERATURE Data Logger (Multi to 500 °C
Oven, Furnace
Position Calibration)
Using IR
THERMAL-
252 Black Body Source Thermometer by 50 °C to 500 °C 2 °C
TEMPERATURE
Comparison Method
Using N-Type
THERMAL- Chamber, Hot Air Thermocouples with 500 °C
253 4.07 °C
TEMPERATURE Oven, Furnace Data Logger (Multi to 1200 °C
Position Calibration)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 168 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Freezer, Deep
Using RTD Sensor
Freezer, Incubator,
THERMAL- with Datalogger (-) 80 °C
254 Autoclave, Chamber, 1.53 °C
TEMPERATURE (Multi position to 100 °C
Water Bath, Hot Air
Calibration)
Oven
Using SSPRT with
THERMAL- 6½ DMM & Liquid
255 Glass Thermometer 50 °C to 250 °C 0.21 °C
TEMPERATURE Bath by Comparison
Method
IR Themometer, Using Black Body
THERMAL- Non-Contact Source & IR 50 °C
256 2 °C
TEMPERATURE Thermometer, Thermometer by to 500 °C
Pyrometer Comparison Method
Using R-Type
THERMAL- Temperature Bath, Thermocouple with
257 1200 °C to 1500 °C 3.75 °C
TEMPERATURE Dry Block Calibrator Indicator by
Comparison Method
Using R-Type
THERMAL- Temperature Bath, Thermocouple with
258 660 °C to 1200 °C 1.63 °C
TEMPERATURE Dry Block Calibrator Indicator by
Comparison Method
Temperature Bath, Using SSPRT with
THERMAL- (-) 60 °C
259 Liquid Bath, Dry 6½ DMM by 0.02 °C
TEMPERATURE to 150 °C
Block Calibrator Comparison Method
Temperature Bath, Using SSPRT with
THERMAL-
260 Liquid Bath, Dry 6½ DMM by (-) 80 °C to (-) 60 °C 0.12 °C
TEMPERATURE
Block Calibrator Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 169 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Temperature Bath, Using SSPRT with


THERMAL- 150 °C
261 Liquid Bath, Dry 6½ DMM by 0.3 °C
TEMPERATURE to 250 °C
Block Calibrator Comparison Method
Temperature Bath, Uing SSPRT with 6½
THERMAL-
262 Liquid Bath, Dry DMM by Comparison 250 °C to 660 °C 0.34 °C
TEMPERATURE
Block Calibrator Method
Temperature
Using R-Type
Indicator / Recorder
THERMAL- Thermocouple with 1200 °C
263 / Controller with 3.71 °C
TEMPERATURE Indicator (Single to 1500 °C
Sensor of Chamber,
Position Calibration)
Furnace
Temperature
Indicator / Recorder Using R-Type
THERMAL- / Controller with Thermocouple with
264 400 °C to 1200 °C 1.5 °C
TEMPERATURE Sensor of Chamber, Indicator (Single
Hot Air Oven, Position Calibration)
Furnace
Temperature
Indicator / Recorder
/ Controller with
Using RTD Sensor
Sensor of Freezer,
THERMAL- with Indicator
265 Deep Freezer, (-) 90 °C to 400 °C 0.18 °C
TEMPERATURE (Single Position
Incubator,
Calibration)
Autoclave, Chamber,
Water Bath, Hot Air
Oven, Furnace

This is annexure to 'Certificate of Accreditation' and does not require any signature.
HI TECH CALIBRATION SERVICES, NO. 209, VGP NAGAR, MUGAPPAIR WEST,
Laboratory Name :
CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2151 Page No 170 of 170
Validity 11/09/2019 to 10/09/2021* Last Amended on 08/12/2020

*The validity is extended for one year up to 10.09.2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM & Liquid (-) 30 °C
266 Indicator / Controller 0.11 °C
TEMPERATURE Bath by Comparison to 250 °C
/ Transmitter /
Method
Recorder with
Sensor,
Temperature Gauge
Thermometer, RTD
Sensor,
Thermocouple,
Using SSPRT with
Temperature
THERMAL- 6½ DMM & Dry 250 °C
267 Indicator / Controller 0.34 °C
TEMPERATURE Block Calibrator by to 660 °C
/ Transmitter /
Comparison Method
Recorder with
Sensor,
Temperature Gauge
Thermometer, RTD
Sensor,
Thermocouple, Using R-Type
Temperature Thermocouple with
THERMAL- 660 °C
268 Indicator / Controller Indicator & Dry 1.63 °C
TEMPERATURE to 1200 °C
/ Transmitter / Block Calibrator by
Recorder with Comparison Method
Sensor,
Temperature Gauge
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.