Professional Documents
Culture Documents
X and Gamma Reference Radiation For Calibrating Dosemeters and Doserate Meters and For Determining Their Response As A Function of Photon Energy
X and Gamma Reference Radiation For Calibrating Dosemeters and Doserate Meters and For Determining Their Response As A Function of Photon Energy
4037-1:1996
ICS 17.240
BS ISO 4037-1:1996
National foreword
Summary of pages
This document comprises a front cover, an inside front cover, pages i and ii,
the ISO title page, pages ii to iv, pages 1 to 36 and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on
the inside front cover.
© BSI 10-1999
Contents
Page
National foreword Inside front cover
Foreword iv
Text of ISO 4037-1 1
© BSI 10-1999 i
ii blank
BS ISO 4037-1:1996
Contents
Page
Foreword iii
1 Scope 1
2 Normative references 1
3 Definitions 1
4 Continuous reference filtered X radiation 3
5 Fluorescence X radiation 9
6 Gamma radiation emitted by radionuclides 12
7 Photon radiation with energy between
eV and 9 MeV15
Annex A (informative) Bibliography 35
Figure 1 — Low air-kerma rate series spectra 19
Figure 2 — Narrow-spectrum series 21
Figure 3 — Wide-spectrum series 23
Figure 4 — High air-kerma rate series spectra 24
Figure 5 — Uranium spectrum 26
Figure 6 — Schematic diagram of a K-fluorescence X-ray
installation 27
Figure 7 — Example of a collimated installation 28
Figure 8 — Energy levels and emission probabilities of
photon radiation from the decay of 16N (left) and from the
de-excitation of 16O for an incident proton energy
of 340,5 keV on 19F (right)[22] 29
Figure 9 — Thin-target photon yield as a function of proton
energy for the 19F(p, !*)16O reaction 30
Figure 10 — Example of the photon fluence spectrum
of 6 MeV to 7 MeV reference radiation, proton energy 2,7 MeV 31
Figure 11 — Example of the photon fluence spectrum of
the 4,4 MeV reference radiation, proton energy 5,5 MeV 32
Figure 12 — Photon spectrum of “titanium beam” 33
Figure 13 — Photon spectrum of a “nickel beam” 34
Figure 14 — Example of an irradiation facility for production
of capture gamma radiation 34
Table 1 — List of X and gamma reference radiation and
their mean energies 2
Table 2 — Specifications of filtered X radiation 5
Table 3 — Characteristics of low air-kerma rate series 5
Table 4 — Characteristics of narrow-spectrum series 6
Table 5 — Characteristics of wide-spectrum series 6
Table 6 — Characteristics of high air-kerma rate series 7
Table 7 — Approximate characteristics of high air-kerma
rate series 7
Table 8 — Inherent filtration 8
Table 9 — Metal properties 8
Table 10 — Radiators and filters used for K-fluorescence
reference radiation 10
Table 11 — Examples of air-kerma rates and extraneous
radiation measured 30 cm from the radiator centre with
a tube current of 10 mA 12
Table 12 — Radionuclide properties 13
Table 13 — Specific activity and recommended chemical
form of radioactive nuclides 13
Table 14 — Example of composition of aperture alloy used
in the collimator of Figure 7 14
ii © BSI 10-1999
BS ISO 4037-1:1996
Page
Table 15 — Typical photon yields and air-kerma rates for
specified proton energies and 14A proton current 16
Table 16 — Principal photon yields of titanium and nickel
per 100 neutron captures 18
Table 17 — Capture gamma radiation — Examples of targets,
air-kerma rates and reference energies obtained[26] 18
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national standards bodies (ISO member bodies). The work of
preparing International Standards is normally carried out through ISO technical
committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee.
International organizations, governmental and non-governmental, in liaison with
ISO, also take part in the work. ISO collaborates closely with the International
Electrotechnical Commission (IEC) on all matters of electrotechnical
standardization.
Draft International Standards adopted by the technical committees are circulated
to the member bodies for voting. Publication as an International Standard requires
approval by at least 75 % of the member bodies casting a vote.
International Standard ISO 4037-1 was prepared by Technical Committee
ISO/TC 85, Nuclear energy, Subcommittee SC 2, Radiation protection.
This first edition of ISO 4037-1, along with ISO-4037-2, cancels and replaces the
first edition of ISO 4037:1979, which has been technically revised.
ISO 4037 consists of the following parts, under the general title X and gamma
reference radiation for calibrating dosemeters and doserate meters and for
determining their response as a function of photon energy.
— Part 1: Radiation characteristics and production methods;
— Part 2: Dosimetry of X and gamma reference radiation for radiation protection
over the energy ranges 8 keV to 1,3 MeV and 4 MeV to 9 MeV.
Annex A of this part of ISO 4037 is for information only.
iv © BSI 10-1999
BS ISO 4037-1:1996
1 Scope 3 Definitions
This part of ISO 4037 specifies the characteristics For the purposes of this part of ISO 4037, the
and production methods of X and gamma reference following definitions apply:
radiation for calibrating protection-level dosemeters 3.1
and rate dosemeters at air kerma rates mean photon energy, E
from 10 4Gy·h–1 to 10 Gy·h–1 and for determining
their response as a function of photon energy. The ratio defined by the formula:
methods for producing a group of reference
radiations for a particular photon-energy range are
described in four sections which define the
characteristics of these radiations. The four groups
of reference radiation are:
a) in the energy range from about 7 keV
to 250 keV, continuous filtered X radiation and where ÍE is the derivative of the fluence ÍE of the
the gamma radiation of americium-241; primary photons of energy E with respect to
energies between E and E + dE[1], defined as
b) in the energy range 8 keV to 100 keV,
fluorescence X radiation;
c) in the energy range 600 keV to 1,3 MeV,
gamma radiation emitted by radionuclides;
d) in the energy range 4 MeV to 9 MeV, gamma In this part of ISO 4037, this definition is
radiation produced by reactors and accelerators. abbreviated to “mean energy”
These reference radiations should be selected from 3.2
Table 1. spectral resolution, RE (full width at
half maximum)
2 Normative references
ratio, expressed as a percentage, defined by the
The following standards contain provisions which, formula:
through reference in this text, constitute provisions
of this part of ISO 4037. At the time of the
publication, the editions indicated were valid. All
standards are subject to revision, and parties to
agreements based on this part of ISO 4037 are where increment %E is the spectrum width
encouraged to investigate the possibility of applying corresponding to half the maximum ordinate of the
the most recent editions of the standards indicated spectrum
below. Members of IEC and ISO maintain registers
NOTE In the case where fluorescence radiation is present in
of currently valid International Standards. the spectrum, the spectrum width measured is based upon the
ISO 197-1:1983, Copper and copper alloys — Terms continuum only.
and definitions — Part 1: Materials. In this part of ISO 4037 this definition is
ISO 1677:1977, Sealed radioactive sources — abbreviated to resolution
General. 3.3
ISO 3534-1:1993, Statistics — Vocabulary and half-value layer (air kerma), HVL or HVLx[2]
symbols — Part 1: Probability and general thickness of the specified material which attenuates
statistical terms. the beam of radiation to an extent such that the air
ISO 8963:1988, Dosimetry of X and gamma kerma rate is reduced to half of its original value
reference radiations for radiation protection over the in this definition, the contribution of all scattered
energy range from 8 keV to 1,3 MeV. radiation, other than any which might be present
ICRU Report 10b, Physical Aspects of Irradiation, initially in the beam concerned, is deemed to be
National Bureau of Standards Handbook 85(1964). excluded
© BSI 10-1999 1
BS ISO 4037-1:1996
Table 1 — List of X and gamma reference radiation and their mean energies
Values in kiloelectronvolts
Fluorescence X Mean energy, filtered X radiation Gamma radiation,
radiation, mean mean energy
energy Low air-kerma Narrow-spectrum Wide-spectrum High air-kerma
rate series series series rate series
8,6 8,5 8 7,5
9,9
12 13
15,8 16
17,5 17
20 20
23,2 24
25,3 26
30
31
33
37,4 37
40,1 45
48 48
49,1
59,3 60 57 57 59,5 (241 Am)
65
68,8
75,0 79
98,4 87 83
100 104 102
109
118 122
149 137 146
147
164 173
185 208 208
211 250
662(137Cs)
1 173 and
1 333 (60Co)
4 440 (12C)
6 000 (Ti)
6 130a
16 *
O and 16N
8 500 (Ni)
a
When produced by protons of energy near the reaction theshold, see 7.1).
2 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 3
BS ISO 4037-1:1996
Details of the operating conditions for each of the 4.2.2 Tube potential
four series are given in Table 3, Table 4, Table 5 and The reference laboratory shall calibrate, at several
Table 6. Table 7 shows an example of the additional points and under operating conditions, the
filtration required to produce the radiation qualities equipment used to indicate the tube potential. The
of the high air-kerma rate series for particular best methods employ an appropriately calibrated
values of the fixed filtration. resistor chain or involve the measurement of the
For the lower air-kerma rate, the narrow-spectrum maximum photon energy by high resolution
and the wide-spectrum series, a “reference spectrometry. If the calibration is determined by
laboratory” shall verify, by a spectrometric study, spectrometry, the tube potential shall be found from
that the value of the mean energy produced is the intersection of the extrapolated linear
within ± 3 %, and the resolution, RE, of the spectra high-energy part of the spectrum with the energy
is within ± 10 % of the values listed in Table 3, axis. The conventionally true value of the tube
Table 4 and Table 5. potential shall be known to within ± 2 %.
For reference radiation for these three series having For laboratories without these facilities, it is
mean energies lower than 30 keV, the mean possible to set the tube potential to produce any of
energies shall be within ± 5 % and the resolutions the radiation described in Table 3, Table 4 and
within ± 15 % of the values in Table 3, Table 4 and Table 5.
Table 5. For reference radiation using additional This may be accomplished in one of the
filtration of 1 mm AI or less, the target angle, target following ways.
condition and air path strongly influence the values
of the mean energies, resolutions and HVLs. a) For radiation generated at potentials
below 116 kV (i.e. below the K-absorption edge of
If a laboratory does not have a spectrometry system, uranium at 115,6 keV), the voltage-measuring
the high voltages and filtration characteristics equipment or meter can be calibrated using
listed in Table 3, Table 4 and Table 5 shall be used techniques based on the excitation of the
and the reference radiation produced shall be characteristic radiation from a selected element.
checked by the simple method described in 4.3.
b) Alternatively, and for tube potentials
For the high air-kerma rate series, the quality of the above 116 kV, using the method described in 4.3.
reference radiation is specified in terms of the X-ray The inherent filtration shall be determined as
tube potential, and the first HVL. The method for described in 4.2.3 and the fixed filtration shall be
producing the high air-kerma rate series is adjusted to the required value with an additional
described in 4.4. aluminium filter (the total being regarded as
4.2 Conditions and methods for producing constituting the new fixed filtration). The tube
reference radiation potential calibration shall be determined by
4.2.1 Characteristics of the X-ray units achieving the reference HVL by the method
specified in 4.3.
X radiation shall be produced by an X-ray unit
whose tube potential shall have a ripple of less 4.2.3 Filtration
than 10 %. It is preferable to use an X-ray unit NOTE The total filtration is made up of the fixed filtration and
having a ripple as low as possible. X-ray units are the additional filtration. For radiation having the three lowest
mean energies of 8,5 keV, 17 keV and 26 keV of the low
commercially available which have a ripple of < 1 %. air-kerma rate series and for radiation having the five lowest
It should be possible to display the value of this tube mean energies of 8 keV, 12 keV, 16 keV, 20 keV and 24 keV of
potential to whithin ± 1 %. the narrow-spectrum series, the fixed filtration comprises the
recommended inherent filtration of the tube of 1 mm Be. Other
The target of the X-ray tube shall be made of values of the tube filtration may be used [see footnote 3) of
tungsten, shall be of the “reflection” type and shall Table 3 and Table 4].
be orientated at an angle of not less than 20° to the 4.2.3.1 For all other reference X radiation, the fixed
direction of the bombarding electrons. filtration comprises:
During irradiation, the mean value of the tube a) the inherent filtration of the tube, plus that
potential shall be stable to within ± 1 %. due to the monitor ionization chamber, if
NOTE The X-ray tube should be operated in such a way that applicable, plus the aluminium filters which are
ageing effects are minimized, since these effects increase the added to obtain a total fixed filtration equivalent
inherent filtration (see 4.2.3). to that of 4 mm of aluminium at 60 kV. These
aluminium filters shall be placed after the
additional filtration (i.e. furthest from the X-ray
focal spot) in order to reduce fluorescence
radiation from the additional filtration;
4 © BSI 10-1999
BS ISO 4037-1:1996
E RE mm mm
keV % kV Pb Sn Cu Al
b) the inherent filtration of the tube is due to the 4.2.3.2 Determination of the inherent filtration
various constituent elements (glass of the bulb, shall be made by measuring, with aluminium
oil, window, etc.) and is expressed, for a given absorbers of 99,9 % purity, the first HVL of the
voltage, as the thickness of an aluminium filter beam produced by the tube without additional
which, in the absence of the constituent elements filtration, at 60 kV, in the following way.
of the tube, would supply a radiation having the a) The method of measurement of the HVL should
same first HVL. A tube whose inherent filtration be in accordance with ICRU Report 10b and
exceeds 3,5 mm of aluminium should be not used; reference [9].
c) the inherent filtration shall be checked b) If a monitor ionization chamber is used during
periodically in order to ensure that this limit is the measurement of inherent filtration, it should
not reached (because of tube ageing) and to be placed between the two sets of beam
proceed to the adjustment of the fixed filtration. collimators and be followed by the aluminium
absorbers in such a manner that it does not
correspond to radiation backscattered from the
absorbers.
© BSI 10-1999 5
BS ISO 4037-1:1996
keV % kV Pb Sn Cu Al mm mm
keV % kV mm mm mm
Sn Cu
6 © BSI 10-1999
BS ISO 4037-1:1996
mm
kV Al Cu
10 0,04
20 0,11
30 0,35
60 2,4 0,077
100 0,29
200 1,7
250 2,5
280 b 3,4
300 3,4
a
The constant potential is measured under load.
b
This reference radiation has been introduced as an alternative to that generated at 300 kV, for use when 300 kV cannot be
attained under conditions of maximum load.
c The HVLs are measured at 1 m from the focal spot.
kV Al Cu Air Al Cu Al Cu keV
c) The first HVL shall be determined using an d) The inherent filtration measurements shall be
ionization chamber with a known response per made in a manner such that negligible scattered
unit air-kerma rate over the energy range of radiation from the aluminium absorbers reaches
interest. Corrections shall be applied for any the detector, since such radiation would increase
variation in detector response with changes in the measured HVL. For radiation produced at
the photon spectrum as the thickness of the potentials above 100 kV, extrapolation to
aluminium absorber is increased. infinitely small field size should be made.
© BSI 10-1999 7
BS ISO 4037-1:1996
8 © BSI 10-1999
BS ISO 4037-1:1996
A monitor chamber shall be used in order to permit 4.5 Field uniformity and scattered radiation
application of corrections for fluctuations in the air 4.5.1 Field diameter
kerma rate.
The diameter of the field shall be sufficient to
4.3.3 Measurement procedure completely and uniformly irradiate the detector at
For selected reference radiation corresponding to the point of test closest to the focus, usually not
the conditions specified in Table 3, Table 4 and closer than 50 cm. The field may remain unchanged
Table 5, the following procedure shall be carried out. for all other experimental points of test or may be
Plot the attenuation curve loge (Id) = f(d) where Id is reduced to be just sufficient to irradiate the detector
the value of the air-kerma rate which is transmitted uniformly.
through a filter having a thickness d. 4.5.2 Field uniformity
From the attenuation curve, determine the first and The air-kerma rate at each point of test shall not
second HVLs. vary by more than 5 % over the entire sensitive
If values of these layers agree within ± 5 % with volume of the detector under test.
those listed in Table 3, Table 4 and Table 5, it shall 4.5.3 Scattered radiation
be assumed that the quality of the reference
radiation complies with this part of ISO 4037. Note Both the following tests shall be carried out to check
for the low air-kerma rate series only the first HVLs that, at the experimental distances the contribution
need comply with the values given in Table 3. due to scattered radiation is less than 5 % of the
total air-kerma rate. These tests shall be carried out
If this is nost the case, the voltage used shall be with the aid of a secondary standard ionisation
adjusted and the measurements repeated until chamber of adequate sensitive whose variations in
the ± 5 % criterion is met. response per unit air kerma as a function of photon
4.4 Production of high air-kerma rate series energy and direction are small within the spectrum
4.4.1 Fixed filtration range considered.
For tube potentials up to and including 60 kV, the 4.5.3.1 Test 1
total filtrations (inherent + additional) are less than Measure the air-kerma rates on the central axis of
the equivalent of 4 mm of aluminium, so an X-ray the beam at the various points of test. The
tube with low inherent filtration is required to air-kerma rates, after corrections for air attenuation
generate the lower energy radiation. At potentials and for chamber size if applicable, shall be
of 60 kV and above, the fixed filtration shall be proportional within 5 % to the inverse square of the
adjusted to the equivalent of 4 mm of aluminium. focus to detector distance.
The aluminium filter used to supplement the 4.5.3.2 Test 2
inherent filtration of the tube shall be placed after
the copper filter in order to reduce any fluorescent At each distance employed in test 1, measure the air
radiation arising from the copper. The thickness of kerma rate after displacing the chamber, in a plane
aluminium employed shall not be less than 0,5 mm. perpendicular to the axis of the beam, by a distance
which is equal to twice the radius of the beam plus
4.4.2 Additional filtration its penumbra. The air-kerma rates of the scattered
At a given potential, the thickness of the additional radiation outside the direct beam shall be less than
filtration shall be adjusted so that the measured or equal to 5 % of the corresponding air-kerma rates
first HVL lies within ± 10 % of that specified, for on the central axis.
radiation generated up to and including 30 kV and
within ± 5 % for the higher energy radiation. The 5 Fluorescence X radiation
minimum purity of the additional filters and the
5.1 Principle
absorbers used to determine the HVL shall
be 99,9 %, except in the case of aluminium used at The calibration of dosemeters and doserate meters
and below 20 kV when the minimum shall by means of fluorescence radiation makes use of the
be 99,99 %. Examples of additional filtration for the K fluorescence lines of certain materials having
high air-kerma rate series are given in Table 7. energies between 8,6 keV and 100 keV and which
are given, as a first approximation, by that of their
K!1 line (see Figure 5). The contribution of the K"
lines is made negligible with the aid of secondary
filters whose K-absorption edges lie between the K!
and K" lines (see Table 10).
© BSI 10-1999 9
Table 10 — Radiators and filters used for K-fluorescence reference radiation
10
Theoretical Total primary
Radiator Secondary filtration
energy, filtration
© BSI 10-1999
BS ISO 4037-1:1996
1) Here the environment is taken to consist of the walls, the supports and other accessories of the installation.
© BSI 10-1999 11
BS ISO 4037-1:1996
The air-kerma rates may be reduced by varying the The values of the air-kerma rate at the point of test
following parameters: shall be measured. Then the ionization chamber
a) tube current: reduction to approximately 1 mA shall be displaced in a plane perpendicular to the
(subject to preserving the reference radiation axis of the beam by a distance equal to twice the
characteristics); radius of the beam plus its penumbra. Two
diametrically opposite measurements shall be
b) area of the fluorescence source: use of a smaller
carried out in this plane. The results of either of
diameter primary diaphragm, which must not be
these measurements shall not exceed 5 % of the
les than the focus diameter;
air-kerma rate measured at the point of test. If this
c) radiator-detector distance: use of distances up limit is exceeded, the effectiveness of the X-ray
to approximately 1 m. If distances > 1 m are used shielding shall be checked. For this purpose, the
it will be necessary to recheck the spectral purity residual air-kerma rate at the point of test shall be
of the reference radiation. measured with the secondary beam completely
NOTE Extraneous radiation includes characteristic emissions absorbed. The measured air-kerma rate should be
other than the K! radiation of the radiator and the scattered less than 0,5 % of that due to the fluorescence
radiation originating from the radiator itself and its support,
from the diaphragms and from the filters. It does not include the radiation.
radiation scattered from the environment, mentioned later. 5.5 Guidance on the use of the reference
Table 11 — Examples of air-kerma rates and radiation
extraneous radiation measured 30 cm from
It should be noted that the extraneous radiation
the radiator centre with a tube current
referred to in Table 11 is specified in terms of the
of 10 mA
quantity air kerma. For quantities that refer to the
Air-kerma Air kerma due to measurement at a depth in a material, where the
Energy obtained rate 30 cm from extraneous attenuation of the fluorescent X-ray lines may be
centre of radiator radiationa
significant the spectral distribution of Compton
keV mGy·h–1 % scattered impurities may have a dominant influence
From 10 to 25 60 to 130 u 10 upon the device being calibrated. As a consequence
of such spectral impurities, it is extremely difficult
From 25 to 26 to 60 u 10 to quantify the dose and effective mean energy of
98,4 penetrating quantities and necessitates precise
a
Defined in note to 5.3.2 and given by the formula: spectral information of the radiation beams,
Air kerma of extraneous radiation × 100 weighted in terms of the appropriate dosimetric
--------------------------------------------------------------------------------------------------------------------------------------------------------------------
K radiation air kerma + air kerma of extraneous radiation quantity. Also, for comparative purposes,
measurements performed using different tube
5.4 Measurement of scattered radiation potentials would produce ambiguous results.
Therefore the radiation qualities numbered 1
An appropriately calibrated ionization chamber,
and 11 in Table 10 shall not be used for determining
selected for the energy and air-kerma rate ranges in
an instrument’s response with respect to a
question, shall be used to determine the
dosimetric quantity at a depth of 1 cm. For the same
contribution due to scattered radiation at the points
reason, caution should be exercised in using other
of test.
low-energy fluorescence radiation.
The contribution due to scattered radiation shall be
less than 5 % of the air-kerma rate due to the 6 Gamma radiation emitted by
fluorescence radiation. radionuclides
The ionization chamber’s variations in response per
6.1 Radionuclides used for the production of
unit air kerma as a function of the spectral quality
gamma radiation
and direction of the radiation shall be small and
known over the energy range in question. Calibrations of dosemeters and rate dosemeters by
means of gamma radiation emitted by radionuclides
shall be carried out with radiation from the
radionuclides listed in Table 12.
12 © BSI 10-1999
BS ISO 4037-1:1996
2) This kind of installation produces at most 5 % scattered photons for 137Cs and less for 60Co.
© BSI 10-1999 13
BS ISO 4037-1:1996
The safety enclosure shall be made of lead of 6.3.3 Variation of air-kerma rate by means of
sufficient thickness to reduce the fluence of the lead attenuators
extraneous beam passing through the enclosure to Instead of using sources with different activities,
one-thousandth of the useful beam. For 60Co, this the air-kerma rate may also be varied by means of
thickness is 12,5 cm and for 137Cs, it is 6,5 cm. lead attenuators for collimated beams of 137Cs
These values may have to be increased in order to and 60Co. The attenuators shall be placed in close
limit radiation exposure of users to acceptable vicinity to the diaphragm. A sequence of lead
levels. attenuators with thicknesses of
A collimator shall be employed to define the shape about 20 mm, 40 mm, 60 mm, etc.
and size for the photon beam. The collimated and 38 mm, 76 mm, 114 mm etc. leads to a
installation shown in Figure 7 has a collimator that reduction in air-kerma rate by successive orders of
is conical in shape with the source at the apex. It is magnitude for Cs and Co, respectively. The above
made up of a succession of at least six apertures, figures serve merely as a guideline. The exact extent
having a total thickness of about 90 mm and of the attenuation depends on geometrical
separated from each other by 20-mm interstices parameters such as field size. Therefore the value of
which serve as traps for the photons scattered by the the air-kerma rate at the point of test shall be
edges of the preceding aperture. The final aperture determined by dosimetric measurements.
has a thickness of 3 mm and a diameter which is The range of attenuation may cover six orders of
slightly greater than the cross-section of the beam a magnitude or more. Despite an increased fraction of
that point. These apertures are made of tungsten photons which have undergone a scattering event
alloy. An example for the composition of such an with increasing attenuator thickness, the spectral
alloy is given in Table 14. purity of the radiation is maintained as the fluence
NOTE In an improved method of operation, the collimator is spectra of all photons become progressively
extended by an output tube. The tube-collimator assembly, which
is closed at its ends by thin windows made of ethylene glycol
narrower, i.e. the mean energy approaches more
polyterephthalate, forms an enclosure in which a vacuum can be and more that of the emission line(s)[15, 16].
created, thereby reducing air scatter whithin this assembly. In order to eliminate the influence of a possibly
Table 14 — Example of composition of distorted electronic equilibrium at the point of test
aperture alloy used in the collimator the distance between the lead attenuator and point
of Figure 7 of test should be at least 100 cm. Alternatively, the
Element Content lead attenuator may be covered by a layer of
graphite of at least 0,5 cm thickness.
%
6.4 Checking installation conformity
Tungsten 89
The following test shall be carried out in order to
Nickel 7 check that, at the various experimental distances,
Copper 4 the contribution due to scattered radiation
extraneous to that from the source capsule does not
The beam cross-section shall be larger than that of exceed 5 % of the total air-kerma rate.
the detectors to be irradiated. The distance dl The air-kerma rates shall be measured on the axis
(see Figure 7) shall be greater than or equal of the beam at the various points of test. After
to 30 cm. The distance d2 shall be sufficiently great correcting for air attenuation, the air-kerma rates
for the contribution to the total air-kerma rate of shall be proportional within 5 % to the inverse
photons backscattered by the walls of the room to be square of the distance from the source centre to the
compatible with the requirements given in 6.4. detector centre.
14 © BSI 10-1999
BS ISO 4037-1:1996
7 Photon radiation with energy Relative photon emission rate (yield) as a function of
between 4 MeV and 9 MeV proton energy is illustrated in Figure 9. As target
thickness (and thus proton energy loss in the target)
7.1 General is increased, the yield increases and the photon
Reference radiation in the energy range spectrum changes, as the protons undergoing
between 4 MeV and 9 MeV is provided because of interactions with the fluorine have decreasing
the 6 MeV photon fields produced by many nuclear energies with increasing depth. The energy of the
power stations and other nuclear reactor systems, photons emitted is high enough for their
as well as by other high-energy photon sources. attenuation in the target to be considered negligible.
Further energies are not specified, since the Depending on the required yield, the proton energy
variation in response of most dosemeters and rate chosen for the production of the reference radiation
dosemeters with photon energy shows no shall be either one of the resonance
discontinuity over this energy range. energies (340,5 keV or 872,1 keV) or a convenient
7.2 Production of reference radiation energy between 2 MeV and 3 MeV. If a high yield is
required and a contamination contribution to the air
Photon reference radiation shall be produced by one
kerma of approximately 4 % can be tolerated,
of the following reactions:
protons, of energy close to 2,7 MeV, incident on a
a) de-excitation of 16O in the 19F target of approximately 6 mg·cm–2 in thickness,
(p, !*) 16O reaction, (see 7.2.1)[17, 18, 19, 20]; should be used (see also 7.4.3). For the purest
b) de-excitation of 12C (see 7.2.2)[20]; possible reference radiation, 340,5 keV protons
c) thermal neutron-capture gamma radiation should be used, provided that the lower air-kerma
(see 7.2.3)[21]; rates are acceptable. For the 340,5 keV proton
resonance, calibration shall be carried out both
d) decay of 16N (see 7.2.4)[24, 25].
on-resonance and off-resonance by – 10 keV in order
Examples of photon fluence spectra for these to allow for the effect of any low-energy and
reference radiations are shown in Figure 10, non-resonant radiation originating from the
Figure 11, Figure 12 and Figure 13. accelerator. The difference between the
7.2.1 Photon reference radiation on-resonance and the off-resonance calibrations
from de-excitation of 16O in the 19F(p, !*)
!* shall be taken as due only to the 6,13 MeV photon
16O reaction radiation and to associated knock-on electrons.
This reference radiation shall be produced using a Care should be taken to prevent fluorine other than
particle accelerator to bombard a fluorine target that of the target from being introduced into the
(usually CaF2) with protons using accelerator.
the 19F(p,!*)16O reaction. Typical yields and air-kerma rates are given in
The energy levels and the relative emission Table 15 for four different incident proton energies,
probabilities resulting from this reaction for a proton current of 1 4A and a target thickness
for 340,5 keV protons incident on a thin target are of approximately 6 mg·cm–2.
shown in Figure 8. At this proton energy, the NOTE The proton energy loss in such a target is
probability for the decay of the excited 16O state by approximately 600 keV for a 2,7 MeV incident proton.
emission of 6,13 MeV photons is 97 %, 2,5 % for A typical photon fluence spectra produced
the 7,717 MeV level and 0,5 % for the 6,917 MeV with 2,7 MeV protons and a target thickness of
level, while the 6,05 MeV photon emission is approximately 6 mg·cm–2 is shown in Figure 10[8l.
negligible. The deviation from isotropic emission of
these photons is less than 3,5 %. At the higher
proton energies the relative contribution of
the 6,13 MeV photons decreases in favour of the
higher energy photons, and there is an increase in
the contribution by contaminant reactions, for
example (p, p½*) and pair production.
© BSI 10-1999 15
BS ISO 4037-1:1996
Table 15 — Typical photon yields and air-kerma rates for specified proton
energies and 1 4A proton current
Typical air-kerma rate at 1 m from
Proton energy Photon yield
target
16 © BSI 10-1999
BS ISO 4037-1:1996
The subsequent beta-decay of 16N (see also 7.2) with 7.4.2 Contamination of reference radiation
a half-life of 7,1 s leads to the excited states of 16O, common to all methods of production of
in this case yielding 6,13 MeV and 7,12 MeV reference radiation
photons, with relative emission probabilities of 68 % 7.4.2.1 Photons with energies of 0,511 MeV are
and 5 % respectively, and 10,4 MeV beta-radiation. produced by positron annihilation after pair
Photon energies and relative emission probabilities production events in the target chamber and in the
are shown in Figure 8 walls of the calibration room and in filter materials,
In a practical set-up, water is pumped continuously if used.
through the reactor core in a closed loop at a 7.4.2.2 Beta particles created in the target as a
flowrate of 30 l·s–1. This loop is brought out through consequence of a nuclear reaction, or secondary
the biological shield of the reactor and acts as the electrons created by photons at or near the target
radiation source[24]. The photon emission rate per and in the intervening air space, cause considerable
megawatt of thermal power and per kilogram of contamination of the reference radiation. Further
water is of the order of 1 × 108 s–1 and the associated contamination can arise from associated
air-kerma rate is approximately 50 mGy·h–1 at 1 m. bremsstrahlung.
7.3 Beam diameter and uniformity of 7.4.2.3 Scattering of photons of the reference
radiation field radiation in the target and the material in its
The requirements shall be identical to those vicinity produces lower-energy photons contributing
specified in 4.5, except that the term “tube focus” at least 1 % to the air-kerma rate.
shall be replaced by “target”. If the area of the field 7.4.3 Additional contamination of accelerator
is not sufficient to irradiate the dosemeter or produced reference radiation from
phantom completely and uniformly, they should be de-excitation of 16O (see 7.2.1)
scanned across the beam. This technique is not
always applicable to air-kerma rate instruments. In addition to the forms of contamination listed
in 7.4.2, which can be reduced by decreasing the
7.4 Contamination of photon reference target chamber mass, discrete gamma radiations
radiation from nuclear reactions are induced in the target by
7.4.1 General the proton beam. At proton energies between 2 MeV
Contamination of the reference radiation by and 3 MeV, photons with energies between
neutrons, electrons and by photons of energy other about 0,1 MeV and 1,5 MeV are produced by the
[19F(p, p½*)19F] reaction, with yields increasing with
than the reference energy shall be assessed, and its
proton energy. At a proton energy of 2,7 MeV, this
influence on the readings of the dosemeters and rate
reaction contributes about 4 % of the air-kerma rate
dosemeters being calibrated shall be determined.
from the 6 MeV to 7 MeV reference radiation. The
Both the photon reference radiation and the use of a filter constructed of a high atomic-number
associated photon contamination may be assessed material, e.g. lead, around the target to eliminate
from measurements of the pulse-height the low-energy photon contamination increases
distributions (see Figure 10, Figure 11, Figure 12 contamination by secondary electrons and
and Figure 13). Since the variation in response of annihilation radiation.
most dosemeters and rate dosemeters with photon
energy is small and shows no discontinuities over 7.4.4 Additional contamination of reference
the energy range from 4 MeV to 9 MeV, radiation produced by reactor neutrons
contamination by photon energies differing by up to 7.4.4.1 Contamination of reference radiation
around 1 MeV from the reference beam energy may produced by thermal neutron-capture gamma
be tolerated. In instruments containing beryllium, reactions in titanium or nickel (see 7.2.3)
tin or lead, the effects caused by photonuclear In addition to the forms of contamination listed
reactions in these materials will be negligible. in 7.4.2, the reference beams contain discrete
Methods of reducing the contamination of the radiation of low-energy photons (see Table 16) and
reference beam are described in photons produced by Compton scattering, especially
references [17], [18], [21], [24] and [25]. The most in the filters added to reduce the low-energy photon
prevalent forms of contamination of the reference contamination. The discrete radiation of photons of
radiation produced by the methods specified are energies below 5 MeV in the reference radiation
given in 7.4.2 to 7.4.4. produced by titanium and below 6,8 MeV in the
reference radiation produced by nickel shall be
reduced by means of added filtration, such that the
kerma rate from photons below these energies shall
not exceed 10 % of the total air-kerma rate.
© BSI 10-1999 17
BS ISO 4037-1:1996
Table 16 — Principal photon yields of titanium and nickel per 100 neutron captures
Titanium Nickel
a
Photon energy Number of photons Photon energy Number of photonsa
keV keV
mm kg %
In the examples given in reference [23], using 7.4.4.2 Contamination of reference radiation
additional filtration of 30 cm aluminium reduces produced by decay of 16N (see 7.2.3)
the air-kerma rate due to such photon In addition to the forms of contamination listed
contamination, so that the effective mean energy in 7.4.2, this reference radiation contains
increases from 4,5 MeV to 6,4 MeV for titanium and low-energy photons emitted by the various
from 7,0 MeV to 8,1 MeV for the nickel target activation products from the contaminants in the
reference beam. cooling medium, the main contribution arising
There also is contamination of the reference from 24Na with energies at 2,754 MeV
radiation by neutrons produced by and 1,369 MeV.
photodisintegration; this shall be determined from
measurements made with a neutron detector
(e.g. a neutron dose equivalent ratemeter).
18 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 19
BS ISO 4037-1:1996
20 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 21
BS ISO 4037-1:1996
22 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 23
BS ISO 4037-1:1996
24 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 25
BS ISO 4037-1:1996
26 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 27
BS ISO 4037-1:1996
28 © BSI 10-1999
BS ISO 4037-1:1996
Figure 8 — Energy levels and emission probabilities of photon radiation from the
decay of 16N (left) and from the de-excitation of 16O for an incident proton energy
of 340,5 keV on 19F (right)[22]
© BSI 10-1999 29
BS ISO 4037-1:1996
30 © BSI 10-1999
BS ISO 4037-1:1996
NOTE Two groups of reactions can be seen: in addition to the 6 MeV to 7 MeV radiation, there is an anihilation reactior
with an energy of 511 keV.
Figure 10 — Example of the photon fluence spectrum of 6 MeV to 7 MeV reference
radiation, proton energy 2,7 MeV
© BSI 10-1999 31
BS ISO 4037-1:1996
32 © BSI 10-1999
BS ISO 4037-1:1996
© BSI 10-1999 33
BS ISO 4037-1:1996
34 © BSI 10-1999
BS ISO 4037-1:1996
Annex A (informative)
Bibliography
[1] International Commission on Radiation Units and Measurements. Radiation Quantities and Units.
ICRU Report 19, 1971.
[2] International Commission on Radiation Units and Measurements. Radiation Dosimetry. X-Rays
Generated at Potentials of 5 to 150 kV. ICRU Report 17, 1970.
[3] ILES, W.J. Conversion Coefficients from Air Kerma to Ambient Dose Equivalent for the International
Standard Organization — Wide Narrow and Low Series of Reference Filtered X Radiation. NRPB Report
R 206, 1987.
[4] PEAPLE, L.H.J., BIRCH, R. and MARSHALL, M. Measurements of the ISO Series of Filtered Radiations.
United Kingdom Atomic Energy Authority Report R 13424, 1989.
[5] SEELENTAG, W.W., PANZER, W., DREXLER, G., PLATZ, L. and SANTNER, F. Catalogue of Spectra for the
Calibration of Dosemeters. GSF Bericht 560, Munich: Gesellschaft far Strahlen und Umweltforschung
mbH, 1979.
[6] LAITANO, R.F., PANI, R., PELLEGRINI, R. and TONI, M.P. Energy Distributions and Air-Kerma Rates of
ISO and BIPM Reference Filtered X-Radiations, ENEA Publication RT/AMB/90, 1990.
[7] READ, L.R. and HAGREEN, M. The ISO Narrow Spectrum Series of Filtered X-Radiations at NPL. NPL
Report RS (EXT) 92, 1987.
[8] BUEERMANN, L., GULDBAKKE, S. and KRAMER, H.M. To be published.
[9] TROUT, E.D., KELLY, J.P. and LUCAS, A.C. Determination of Half-Value Layer. Am. J.
Roentgenology, 85, 1960, p. 933.
[10] TAYLOR, L.S. Physical Foundations of Radiology, 2nd Edition, 1959, pp. 227–257.
[11] ISO 3534-1:1993, Statistics — Vocabulary and symbols — Part 1: Probability and general statistical
terms.
[12] ISO 8963:1988, Dosimetry of X and gamma reference radiations for radiation protection over the energy
range from 8 keV to 1,3 MeV.
[13] WOODS, M.J. The Half-life of 137Cs, A critical review. Nucl. Instr. Methods, A, 286, 1990,
pp. 576–583.
[14] International Commission on Radiation Units and Measurements. Radiation Quantities and Units.
ICRU Report 33, 1980.
[15] ROOS, M. and GROSSWENDT, B. A Simplified Method for Large Range Variations of Dose Rate
for 137Cs Gamma Radiation. Radiat. Prot. Dosim., 18, 1987, pp. 147–151.
[16] ROOS, M. and GROSSWENDT, B. Variation der Dosisleistung von 60Co-Gammastrahlung bei geringer
Beeinflussung des Spektrums. Medizinische Physik, ed. D. Harder, 1990, pp. 290–291.
[17] HALL, R.S. and POLLE, D.H. A Radiation Source using a Positive Ion Accelerator. Central Electricity
Generating Board, Berkeley Nuclear Laboratories, Report RD/B/N265, 1967.
[18] ROGERS, D.A. Nearly Mono-energetic 6 to 7 MeV Photon Calibration Source. Health Phys., 45(1), 1983
pp. 127–137.
[19] DUVALL, K.C., SOARES, C.G., HEATON II, H.T. and SELTZER, S.M. The Development of a 6 and 7 MeV
Photon Field for Instrument Calibration. Nucl. Instrum. Methods, B10/11, 1985, pp. 942–945.
[20] GULDBAKKE, S. and SCHÄFFLER, D. Properties of High-energy Photon Fields to be Applied for
Calibration Purposes. Nucl. Instrum. Methods, A299, 1990, pp. 367–371.
[21] BERMANN, F. et al. Capture Gamma Ray Beam for the Calibration of Radioprotection Dosemeters
between 5 and 9 MeV. Radiat. Prot. Dosim., 30(4), 1990, pp. 237–243.
[22] LONE, M.A., LEAVITT, R.A. and HARRISON, D.A. Prompt Gamma-rays from Thermal Neutron Capture.
Atomic Data and Nuclear Data Tables, 26, 1981, p. 511.
[23] VORBRUGG, W. and ZILL, H.W. Erzeugung hochenergetischer Photonenbündel durch Einfang
thermischer Neutronen. 6th Int. Congr. IRPA ‹‹Radiation — Risk Protection» Berlin (West), 7-12 Mai, 1984,
pp. 1158–1160.
© BSI 10-1999 35
BS ISO 4037-1:1996
[24] BECK, J. et al. Test Verschiedener Gamma-Detektoren zum Nachweis der N-16 Strahlung bei
Leckagen in Wärmeübertragungssystem von Kernkraftwerken. Atomkernenerg./Kerntech., 34, 1979,
pp. 57–60.
[25] NEAULT, P.J. The Dosimetry of Nitrogen-16. M. Sc. Thesis, Lowell Technological Institute, 1980.
[26] BERMANN, F. and TROESCH, G. Étalonnage de détecteurs de radioprotection avec des * de haute energie:
Utilisation d’un faisceau de * de capture. VIIIéme Congres International de la Société française de
radioprotection, Saclay, March 23–26 1976, pp. 538–565.
36 © BSI 10-1999
blank
BS ISO
4037-1:1996
BSI — British Standards Institution
BSI is the independent national body responsible for preparing
British Standards. It presents the UK view on standards in Europe and at the
international level. It is incorporated by Royal Charter.
Revisions
It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.
BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Buying standards
Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.
Information on standards
Copyright
Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.
This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
written permission of BSI must be obtained.