Service manual
Warranty
Tisenc guarantees a warranty for the ACCRE-6 analyzer from the date of
delivery to the user. For this period it is guaranteed that the Analyzer works
within quality requirements valid on the date of manufacture release.
Tisenc is prepared to compensate for faulty material or malfunctions within
the limitations of the warranty. Tampering with the internal components,
damage due to operating errors, misuse and overlooking essential
information with respect to warnings and precautions described in this User
Manual will invalidate the warranty.
Products will be repaired with charge during the warranty period in
following cases;
1> Improper use or misuse
2> Consumer’s intentional abuse or neglect of the products
3> Unauthorized repair or parts replacement
4> Lots product warranty or changes in contents
5> Damages or defects due to fire, pollution, earthquake or other natural
disasters.
The acknowledgment of claims shall immediately be reported to your local
authorized representatives.
1. User’s Maintenance Procedure
Please check if the printer frequently, and if the printer paper is out,
change the printer paper according to the following procedure:
Open the thermo printer cover on the right side of the analyzer, Fig 2-1
Fig 2-1 Open the thermo printer cover
Remove the used printer paper reel, and install the new printer paper
reel into the printer, please note the direction of paper reel, as Fig 8-2, then
place the setting plate of the printer into its original position. Close the cover
of printer and cut the extra paper roll outside.
3.Specifications of fuse
The fuse is shown in the red box, as fig 3,
Specifications:fusing,250V.6A,5*20mm.
Check the fuse if it’s working normally.
Fig 4 Fuse of the power seat
5. Trouble shooting
Listed below are all of the error messages programmed into the ACCRE
analyzer.
• If error messages other than the ones below appear, call Tisenc Technical
Assistance or your local representative.
• In some cases, there are suggestions for corrective action. Where no
corrective action impossible, or if the corrective actions fail, the final action
is to call Tisenc Technical Assistance or your local representative.
Se the table below :
Error Error Analyzer’s
SN Error Description Resolution for Users
Level category Respond
Fail to access the assay
1
parameters
Fail to access the Internet
2
connection parameters
3 Fail to access the QC parameters
Documents Fail to access the Calibration
4
Error parameters
5 Fail to access the QC data Error Message
6 Fail to access the TEST data will display in
7 Fail to access the SYSTEM data software
Fail to access the Users Account screen, Restart the analyzer by
8
data analyzer will turn-off the power switch
9 Motion Error Error code 0-1000 pause on going and turn-on the power
10 Reading Initialization Error test if error switch again.If the error
11 Level 1 Error Reading Error happens, Use occurs again please
12 Syringe initialization Error should read the contact authorized
13 Pipette initialization Error error message representative or
PMT vertical position and Click”OK” technical service of
14
initialization Error to continue, Tisenc
PMT horizon position analyzer will
15
initialization Error restart after
6. Software update
In order to make sure the Software function work perfectly , it’s
recommended to update the software when there is new software.
Procedures:
1) prepare the USB disk(USB2.0,FAT32 form), and put the all upgrading
files to the root dictionary(no folders).
on the machine , after hearing one beep sound, release the button,
then you will hear one more short beep sound , and one long beep sound,
after long beep sound , the machine will reboot, the updating is finished.
5) Check the software version by Search>> version, if not succeed, please
check the USB disk or the USB port, and repeat the steps again.