Test parameters
D_Bil Version number 3 . 1
Flow
Blank type Yes Primary dilution 1 + 0 Dispensed volume 150
CE-marked application for Indiko™ by Thermo Fisher Scientific Oy, applicable for
use outside the US. Prior to change.
Page 2 / 3
Test parameters
D_Bil Version number 3 . 1
352 2.0 30
Sample Volume (µl) Dispense with Extra volume (µl) Extra wash
5 Water 20 No
300 297
0 *
Reagent Reagent Volume (µl) Dispense with Extra volume (µl) Syringe speed Replacing reagent
353 2.0 30
300 297
End-point Main wavelength (nm) Side wavelength (nm) Residual net abs. (A)
measurement
540 None 0
Dilution Limits
Measuring range (µmol/l) Sec. dilution 1 +
Dilution with Water
Min Max Min Max
Primary dilution 1 + 0
Dilution limit * 170.00 * 6.0
CE-marked application for Indiko™ by Thermo Fisher Scientific Oy, applicable for
use outside the US. Prior to change.
Page 3 / 3
Test parameters
D_Bil Version number 3 . 1
Calibration
QC
Procedure 1 QC profile
Wash definition
Preceding reagent Following reagent Wash type Wash reagent Tag
CE-marked application for Indiko™ by Thermo Fisher Scientific Oy, applicable for
use outside the US. Prior to change.