Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

1

INTRODUCTION note any colors and odors. If foulants


The physical dissection of a reverse are present on the membrane surface, a
osmosis (RO) membrane and the sample is removed for further
laboratory analysis of it's foulants can evaluation.
be one of the most definitive methods
of troubleshooting an RO system. CHEMICAL ID OF FOULANTS
An FTIR spectrometer is used to
Identifying the type of membrane determine the nature of any organic
foulant is the first step in optimizing foulants. This instrument scans the
system performance. With the foulant surface with infrared radiation
Proc edure s in an recommendations in the Avista report, and plots the absorbance as a function
Avista Me mbrane steps can be taken to determine how to of wavelength. The spectrum acts as a
begin correcting frequent cleanings, fingerprint of the organic material and
Elem ent Aut opsy: loss of productivity, reductions in is compared with over 10,000
permeate quality, system downtime, or compounds to determine it's closest
frequent membrane replacement. match.
Ø External Visual Examination
An EDX is used to identify inorganic
THE AUTOPSY PROCEDURE foulants. A SEM is used to take
Ø Wet Test of the Membrane Element An Avista autopsy includes a photographs of the foulant layer and
comprehensive examination of both the
these pictures are included in the
Ø Internal Visual Examination exterior and interior of the membrane completed report.
as well as an analysis of the foulant
Ø Chemical Identification of layer removed from the membrane LOSS ON IGNITION (LOI)
Surface Foulants Using: surface. The following examinations An LOI test is used to obtain a
and tests are performed: relationship between the amount of
• FTIR Spectrometer organic foulant vs. inorganic foulant.
EXTERNAL VISUAL EXAM
• Energy Dispersive CELL AND DYE TEST
A thorough examination is made of the
X-Ray Spectrometer (EDX) exterior of the membrane element. A cut sample of the membrane is tested
• Scanning Electron Microscope The inspection looks for damage or for flow and rejection. This data is
defects in the o-rings and brine seal. compared with the test results obtained
(SEM)
An examination of the feed and during the full element wet test. A dye
concentrate ends and the outer test is also performed to determine if
Ø Loss-on-Ignition (LOI) fiberglass wrapping is also completed. the membrane has been exposed to
oxidation or is physically damaged.
Ø Cell and Dye Test WET TEST
The membrane is placed in a single FUJIWARA TEST
Ø Fujiwara Test element pressure vessel and operated If the membrane is a polyamide, then a
under laboratory conditions. Both feed Fujiwara test is conducted. This is
Ø Summary and Recommendations and pressure drops as well as salt currently the most valid indicator of
rejection are measured. Fluxes are halogenated organics such as chlorine.

................. measured and normalized. The data


from this test is compared to the cell SUMMARY &
test data to differentiate between the RECOMMENDATIONS
Avista Technologies, Inc. condition of the full membrane The data from each test is combined to
133 North Pacific Street element vs the cut membrane sample. provide an overall summary of the
San Marcos, CA 92069 U.S.A. laboratory results. The final segment
Telephone: (760) 744-0536 INTERNAL VISUAL EXAM of the report is from our technical
Facsimile: (760) 744-0619 The outer fiberglass wrapping is personnel whose recommendations are
www.AvistaTech.com removed and the membrane is geared toward reducing the membrane
dissected to examine glue lines and to fouling and its associated costs.

1 . . . . . . . . . . . . . . . . . . . . . . . .

You might also like