CARESCAPE B450 and B650

Software Settings Transfer Guide


Software version 3 host build v3.1

CARESCAPE B450 and B650


English
2nd edition
2109358-005
© 2018 General Electric Company.
All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies,
Inc. and GE Healthcare Finland Oy.

2 CARESCAPE B450 and B650 2109358-005


2018-08-28
Contents
Intended use of this guide ............................................................................ 5
About this guide .......................................................................................... 5
New settings in software version 3.1 ........................................................... 7
Extended settings: New default value for SV trend scale......................... 7
Extended settings: Temp Alarm Priority..................................................... 7
Extended setting: V Tach alarm priority .................................................... 7
Change in Beat source................................................................................ 8
Changed setting: CO2 average................................................................... 9
Change in ECG 12SL options ...................................................................... 9
Changes in HR alarm priorities................................................................... 9
Changed settings: Invasive Pressures ....................................................... 10
Changed setting: Minimum HR/min........................................................... 10
Changed setting: NIBP cuff size ................................................................. 11
Change in Pacemaker detection setting ................................................... 11
Change in Printing settings......................................................................... 12
Change in Snapshot sweep speed............................................................. 12
Changed setting: SpO2 Ch1 Sensitivity/SpO2 Ch2 Sensitivity ................... 12
New default value for Cardiac Artifact alarm............................................ 13
New CARESCAPE RAD options .................................................................... 13
New critical alarms ..................................................................................... 13
New ECG options......................................................................................... 14
New ECG 12SL options................................................................................ 14
New Gases options ..................................................................................... 15
New QT/QTc alarm priority ......................................................................... 15
New Telemetry option................................................................................. 15
New Impedance respiration options.......................................................... 15
New Invasive pressures options................................................................. 17
New Spirometry options ............................................................................. 18
New SpO2 options ....................................................................................... 18
New settings: V Tach Criteria ..................................................................... 19
New Temperature option............................................................................ 19

2109358-005 CARESCAPE B450 and B650 3


New Trends options .................................................................................... 20
New UI options ............................................................................................ 21
New Ventilator disconnected alarm priority.............................................. 21
New settings in software version 2.3 ........................................................... 23
Change in non-invasive and invasive blood pressure alarm
priorities ....................................................................................................... 23
New option: Snapshot sweep speed ......................................................... 26
New option: 12 lead print speed ................................................................ 27
Extended settings: Allowed alarm priorities .............................................. 27
New settings in software version 2.0.8 ........................................................ 29
New option: ECG lead off alarm priority .................................................... 29
Extended settings: PVC, SVC, Impedance Resp, ST and Temp alarm
priority.......................................................................................................... 29
Change in Art and ICP Invasive pressure labels ........................................ 30
Change in other Invasive pressure labels .................................................. 31
Extended settings: ECG Printout format .................................................... 31
New settings in software version 2.0.7 ........................................................ 33
Extended settings: Pacemaker Detection with PDM................................. 33
Extended settings: 12SL Display Format ................................................... 33
Extended settings: New default value for Spirometry sensor
type .............................................................................................................. 33
Change in Menu structure (Diagnostic Tool).............................................. 34
New global options ..................................................................................... 34
New options in NICU software package with CARESCAPE Respiratory
module ......................................................................................................... 34

4 CARESCAPE B450 and B650 2109358-005


Intended use of this guide
This guide provides information about software settings that require manual
configuration when transferring clinical settings between CARESCAPE B450 and B650
monitors with different CARESCAPE software versions. The new or changed settings
will remain in factory defaults in the destination monitor when you transfer the
settings from a source monitor with an older software version. You must configure the
settings listed in this document manually for all the software packages and profiles in
each destination monitor.
You cannot transfer platform settings between monitors with software version 2 and
software version 3. For information on configuring the platform settings manually,
refer to the monitor’s service manual.

About this guide


In this document, the following conventions are used to refer to different software
versions:
● Software version 2.0.6 refers to software version 2 with host build 2.0.6.
● Software version 2.0.7 refers to software version 2 with host build 2.0.7.
● Software version 2.0.8 refers to software version 2 with host build 2.0.8.
● Software version 2.3 refers to software version 2 with host build 2.3.
● Software version 3.1 refers to software version 3 with host build 3.1.
The following table lists the chapters in this document where you can find the new
and changed software settings, depending on the software version of your previous
monitor (source monitor).
Source monitor SW Information about new and changed settings
version
Software version 2.3 See the following chapters:
● New settings in software version 3.1
Software version See the following chapters:
2.0.8
● New settings in software versions 3.1
● New settings in software version 2.3
Software version See the following chapters:
2.0.7
● New settings in software versions 3.1
● New settings in software version 2.3
● New settings in software version 2.0.8
Software version See the following chapters:
2.0.6
● New settings in software versions 3.1
● New settings in software version 2.3
● New settings in software version 2.0.8
● New settings in software version 2.0.7

For more detailed information regarding all settings, refer to the supplemental
information manual.

2109358-005 CARESCAPE B450 and B650 5


Use this guide together with the monitor’s service manual. The service manual has
useful service information and important safety precautions.

6 CARESCAPE B450 and B650 2109358-005


New settings in software version 3.1
Extended settings: New default value for SV trend scale
Setting Selection in Selection in Default value Menu Notes
version 2.3 version 3.1
SVI 50 - 150 20 - 120 20 - 120 Trends > View > Profile setting.
ml/m2 0 - 100 0 - 100 Graphic > Trend Separate for each
Scales > Cardiac user profile.
0 - 250 50 - 150 Output The settings have
0 - 250 been extended
with the new 20 -
120 option, which
is also the new
default. When
transferring
settings to
software version
3.1, the previous
user selection is
preserved.

Extended settings: Temp Alarm Priority


Settings Selection Selection Default Menu Notes
in version in version value
2.3 3.1
Temp Escalating Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
high/low Medium Priorities > Other for each user profile.
Informa- Parameters
tional Low Selections have been
extended with Low
Informa- and Medium. When
tional transferring settings to
software version 3.1, the
previous user selection is
preserved.
The Informational
selection is only available
if Temperature Alarms
Informational Allowed is
Enabled.

Extended setting: V Tach alarm priority


Settings Selection Selection Default Menu Notes
in v2.3 v3.1 value in
v3.1
V Tach High High High Alarm Setup > Arrhytmia Profile settings. Separate
alarm Medium Medium for each user profile.
priority Selections have
Low Low
been extended with
Informa- Informational and Off
tional (NICU only).

2109358-005 CARESCAPE B450 and B650 7


Settings Selection Selection Default Menu Notes
in v2.3 v3.1 value in
v3.1
Off (NICU Selection needs to be
only) activated in Default
Setup > Care Unit
Settings > Parameters
> Parameter Alarms >
Allowed Priorities > ECG.
NICU only: To enable
the Off selection, the
selection needs to be
activated in Default
Setup > Care Unit
Settings > Parameters >
ECG.

Change in Beat source


Setting Selection Selection in Default value Menu Notes
in version 3.1
version
2.3
Beat Primary Primary HR Primary HR Parameter Setup > Profile setting.
Source HR ECG > ECG > Setup Separate for each
ECG user profile.
ECG
Ch x Site Name When transferring
Art x (for IP aterial settings to software
Fem channels x when version 3.1, the
UAC Ch x Site Name is following user
configured to Art, selections are
Pleth Fem or UAC) preserved:
Pleth(SpO2)
● Primary HR
Pleth(2)(SpO2(2))
● ECG
● Pleth

When transferring
settings to software
version 3.1, the
following user
selections revert to
Primary HR (default):
● Art
● ABP
● Fem
● UAC

8 CARESCAPE B450 and B650 2109358-005


Changed setting: CO2 average
Setting Selection in Selection in Default Menu Notes
version 2.3 version 3.1 value
CO2 average Single 10 s Parameter Profile setting. Separate for
Breath 20 s 10 s Setup > each user profile.
10 s Gases > CO2 Options have been reduced:
> Setup Single Breath has been
20 s removed. When transferring
settings to software version
3.1, the Single Breath
selection changes to 10 s.
Other selections remain the
same.

Change in ECG 12SL options


Setting Selection Selection Default Menu Notes
in version in version value
2.3 3.1
Display 4x2.5 - 1 If 12 Lead If 12 Lead Parameter Setup > ECG Profile setting. Separate
Format Rhythm Print Speed Print > ECG > Setup > 12 Lead for each user profile.
4x2.5 - 3 is 25mm/s: Speed is Analysis > Settings Display format settings
Rhythm 4x2.5 - 1 25mm/s: for 25mm/s and for
12 Rhythm 4x2.5 - 3 50mm/s in version 3.1
Rhythms 4x2.5 - 3 Rhythm are two different settings.
Rhythm If 12 Lead If you change the Display
2x6 Format for 25mm/s, it
Rhythms 12 Print
Speed is does not have any effect
Cabrera Rhythms on the 50mm/s display
50mm/s:
2x6 range and vice versa.
Rhythms 12
Rhythms When transferring
Cabrera settings to software
If 12 Lead version 3.1, the previous
Print Speed selection is changed
is 50mm/s: to the Display Format
selected for 25mm/s.
12 Display Format for
Rhythms 50mm/s is set to the
2x6 default value.
Rhythms
Cabrera

Changes in HR alarm priorities


Setting Available values Default Menu Notes
value
HR high Escalating NICU: High Alarms Setup > Alarm Profile setting. Separate
alarm High ED, ICU, OR, Priorities > ECG > for each user profile.
priority and PACU: Tachy/PR High Previously, the HR
Medium
Escalating alarm priority setting
Low determined both the
high limit and the low
limit alarm priorities. In

2109358-005 CARESCAPE B450 and B650 9


Setting Available values Default Menu Notes
value
software version 3.1,
there are two separate
settings for the HR alarm
priorities.
HR low Escalating NICU: High Alarms Setup > Alarm Profile setting. Separate
alarm High ED, ICU, OR, Priorities > ECG > for each user profile.
priority and PACU: Brady/PR Low Previously, the HR
Medium
Escalating alarm priority setting
Low determined both the
high limit and the low
limit alarm priorities. In
software version 3.1,
there are two separate
settings for the HR alarm
priorities.

Changed settings: Invasive Pressures


Label specific settings have been changed to channel specific settings. There are
also label specific defaults for certain settings, but these are only applied if the label
for a channel is changed. When transferring settings to software version 3.1, the
user selections for each pressure channel are preserved. The SPV Source selection is
reverted to the default value (Off).

Changed setting: Minimum HR/min


Setting Selection in Selection in v3.1 Default Menu Notes
v2.3 value
Minimum ICU, OR, With PDM: 90-300 100 Alarms Setup > Profile setting.
HR/min PACU, and (adjustable) NICU: 160 Arrhytmia > Lethal Separate for each
ED: 100 (not With combination Alarms > V Tach user profile.
adjustable) monitoring: Criteria This setting has been
NICU: 160 (not ● For patients who locked. To change
adjustable) are 14 years or the value, the setting
older: >100-300 must be unlocked
(adjustable in in Default Setup >
steps of 10) Locking settings >
Parameters.
● For patients
younger than
14 years: Not
adjustable

10 CARESCAPE B450 and B650 2109358-005


Changed setting: NIBP cuff size
Setting Selection Selection Default value Menu Notes
in version in version
2.3 3.1
Cuff Size AUTO (Not (Not Selected) Parameter Setup > Profile setting.
Adult Selected) NIBP > Setup Separate for each
Adult user profile.
Child
Child Selections have been
Infant updated:
Infant
● PSM specific AUTO
has been removed.
● New default is now
(Not Selected).
When transferring
settings to software
version 3.1, AUTO
changes to (Not
Selected). Otherwise,
the previous user
selection is preserved.

Change in Pacemaker detection setting


Setting Selection Selection Default Menu Notes
in version in version value
2.3 3.1
Pacemaker Off For PDM Sensitive Parameter Setup > ECG Profile setting. Separate
Detection Pace 1 v2.4 (or (Pace 2) > ECG > Advanced for each user profile.
later): PSM and Tram specific
Show Off selections have been
(Pace 1) removed, and previous
Normal
Normal (Pace 1) values are mapped to
(Pace 1) new selections.
Sensitive
Pace 2 (Pace 2) When transferring
settings to software
On (Pace 2) For PDM version 3.1, the
v2.3: PDM/Telemetry values
Sensitive
Off remain as such.
(Pace 2)
On (Pace 2) When using PSM:
Hide
● v2 PSM values are
mapped as follows:
■ Hide changes to
Sensitive
■ Show (Pace 1)
changes to Normal
■ Sensitive (Pace
2) changes to
Sensitive
● v2 Tram values are
mapped as follows:

2109358-005 CARESCAPE B450 and B650 11


Setting Selection Selection Default Menu Notes
in version in version value
2.3 3.1
■ Off remains Off
■ Pace 1 changes to
Normal
■ Pace 2 changes to
Sensitive

Change in Printing settings


When transferring settings to software version 3.1, the network printing devices
and previous user selections for printing devices (for waveform, alarm, telemetry
waveform, numeric trend, reports printouts) are reverted to the default values and
have to be reconfigured by the user. The default values have not changed.

Change in Snapshot sweep speed


Setting Selection Selection Default Menu Notes
in version in version value
2.3 3.1
Sweep 12.5 mm/s 12.5 mm/s 12.5 mm/s Trends > View > Profile settings. Separate
Speed 25 mm/s 25 mm/s Snapshot > Snapshot for each user profile.
Setup When transferring
50 mm/s
settings to software
version 3.1, 50 mm/s is
mapped to 25 mm/s.

Changed setting: SpO2 Ch1 Sensitivity/SpO2 Ch2 Sensitivity


Setting Selection in Selection in Default value Menu Notes
version 2.3 version 3.1
Sensi- Normal Normal Normal Primary channel: Profile setting.
tivity Maximum Parameter Setup Separate for each
Maximum > SpO2 > Setup user profile.
Secondary SpO2 sensitivity
channel: was changed
Parameter Setup from a patient
> SpO2(2) > Setup case specific
setting (which
reverts to the
default after each
case) to a profile
setting. Thus, the
selected value
can now be saved
and persisted
after discharge.
This selection
is Masimo
technology
specific.

12 CARESCAPE B450 and B650 2109358-005


New default value for Cardiac Artifact alarm
Setting Selection in Selection in Default value Menu Notes
version 2.3 version 3.1
Cardiac Alarm On Alarm On Alarm On Parameter Profile setting. Separate for
Artifact Alarm Off Alarm Off Setup > each user profile.
Impedance When transferring settings
Respiration to software version 3.1, the
> Alarms previous user selection is
reverted to the new default
(Alarm On).

New CARESCAPE RAD options


Setting Available Default Menu Notes
values value
Remote High High and Default Setup > Care Unit Care unit setting.
Alarm Device High and Medium Settings > Alarms > Remote
Priority Medium Alarm Device

Operation Intermittent Intermittent Default Setup > Care Unit Care unit setting.
Mode Settings > Alarms > Remote
Continuous Alarm Device

New critical alarms


Setting Available values Default value Menu Notes
HR critical 20 - 300 ED and Monitor Setup > Parameter Profile setting. Separate for
alarm limits ICU: 180/40 Setup > ECG > HR/PR each user profile.
(Neuro profile: Alarms
150/40;
Pediatric
profile:
200/55;
Infant profile:
210/65)
NICU: 230/75
(Infant profile:
210/65)
OR: 190/40
(Neuro profile:
150/40;
Adolescent
profile:
180/40;
Pediatric
profile:
200/55;
Infant profile:
210/65)
PACU: 180/40
(Neuro profile:
150/55;
Pediatric
profile:

2109358-005 CARESCAPE B450 and B650 13


Setting Available values Default value Menu Notes
200/55;
Infant profile:
210/65)
HR critical Alarm On Alarm Off Monitor Setup > Parameter Profile setting. Separate for
alarm on Alarm Off Setup > ECG > HR/PR each user profile.
Alarms
SpO2 critical 30 -100 OFF/88 Monitor Setup > Parameter Profile setting. Separate for
alarm limits Setup > SpO2 > Alarms each user profile.
SpO2 critical Alarm On Alarm Off Monitor Setup > Parameter Profile setting. Separate for
alarm on Alarm Off Setup > SpO2 > Alarms each user profile.

SpO2 critical 30 -100 OFF/88 Monitor Setup > Parameter Profile setting. Separate for
alarm limits Setup > SpO2(2) > Alarms each user profile.
for the
secondary
(SpO2(2))
channel
SpO2 critical Alarm On Alarm Off Monitor Setup > Parameter Profile setting. Separate for
alarm on Alarm Off Setup > SpO2(2) > Alarms each user profile.
primary and
secondary
(SpO2(2))
channel

New ECG options


Setting Available Default Menu Notes
values value
Beat Tone Soft ED, ICU, NICU: Care Unit Settings > Care unit setting.
Sound Sharp Soft Parameters
OR, PACU:
Sharp
PR (SpO2/IP) 0s 5s Default Setup > Profile Profile setting. Separate for
high/low 5s Settings > Alarm Delays each user profile.

10 s
15 s
20 s

New ECG 12SL options


Setting Available Default Menu Notes
values value
Allow Disabled Disabled Default Setup > Care Unit Care unit setting.
Baseline Enabled Settings > Parameters > 12
Stabilization Lead ECG
Stabilize Disabled Disabled Parameter Setup > ECG > Care unit setting.
Baseline Enabled ECG tab > Setup > 12 Lead The Stabilize Baseline
Analysis selection is only available if

14 CARESCAPE B450 and B650 2109358-005


Setting Available Default Menu Notes
values value
Allow Baseline Stabilization
is Enabled.

New Gases options


Setting Available Default Menu Notes
values value
Measure- Automatic Automatic Parameter Setup > Gases > Profile setting. Separate for
ment Source (A) (A) CO2 > Setup each user profile.
Gas Module
(Mod)
CO2 device
(Dev)

New QT/QTc alarm priority


Setting Available Default Menu Notes
values value
QT/QTc alarm Medium Low Alarms Setup > Alarm Profile setting. Separate for
priority Low Priorities > ECG each user profile.
Informa- Informational selection
tional needs to be activated in
Default Setup > Care Unit
Settings > Parameter Alarms
> Allowed Priorities > ECG.
QT/QTc Enabled Disabled Default Setup > Care Unit Care unit setting.
Alarms In- Disabled Settings > Parameter Alarms
formational > Allowed Priorities > ECG
Allowed

New Telemetry option


Setting Available values Default value Menu Notes
Enhanced Enabled Enabled Care Unit Settings > Care unit setting.
ASY Disabled Telemetry
Detection

New Impedance respiration options


Setting Available Default Menu Notes
values value
Resp (Imped.) Escalat- Escalating, Default Setup > Care Unit Care unit setting.
Meas. ing, High, High Settings > Parameter Alarms
Paused Medium, > Allowed Priorities > Other
Low, Infor- Parameters
mational,
Off
Escalat-
ing, High,

2109358-005 CARESCAPE B450 and B650 15


Setting Available Default Menu Notes
values value
Medium,
Low, Infor-
mational
Escalating,
High,
Medium,
Low
Escalating,
High,
Medium
Escalating,
High
Resp (Imped.) Escalating Escalating Alarms setup > Alarm Profile setting. Separate for
measure- High Priorities > Other each user profile.
ment paused Parameters The options shown for the
Medium
setting depend on the Resp
Low (Imped.) measurement
Informa- paused setting.
tional If the Resp (Imped.)
measurement paused
Off setting is changed and
the new allowed priority
setting does not include the
currently selected value, the
new value for Resp (Imped.)
measurement paused is
Escalating.
Insp./Exp. On On Parameter Setup > Profile setting. Separate for
Markers Off Impedance Respiration each user profile.
> Setup
RR (Imped.) 0s 15 s Default Setup > Profile Profile setting. Separate for
high 5s (NICU: 10 s) Settings > Alarm Delays each user profile.
10 s
15 s
20 s
25 s
30 s
RR (Imped.) 0s 15 s Default Setup > Profile Profile setting. Separate for
low 5s (NICU: 10 s) Settings > Alarm Delays each user profile.
10 s
15 s
20 s
25 s
30 s
Imped. Resp On Off Parameter Setup > Profile setting. Separate for
Smoothing Off (NICU: On) Impedance Respiration each user profile.
> Setup

16 CARESCAPE B450 and B650 2109358-005


New Invasive pressures options
Setting Available Default Menu Notes
values value
Primary AUTO AUTO Parameter Setup > Invasive Profile setting. Separate for
Arterial Ch x Site Pressures > Advanced each user profile.
Source Name The rows 2-9 for Primary
Arterial Source contain the
options for any Ch x Site
Name configured to Art, Fem,
or UAC. The options are listed
according to the channel
number, starting from lowest.
Art On On Care Unit Settings > Care unit setting.
Disconnect Off Parameters Setting has changed from
profile setting to care unit
setting. When transferring
settings to software version
3.1, the previous selections
are lost.
Fem On On Care Unit Settings > Care unit setting.
Disconnect Off Parameters Setting has changed from
profile setting to care unit
setting. When transferring
settings to software version
3.1, the previous selections
are lost.
UAC On On Care Unit Settings > Care unit setting.
Disconnect Off Parameters Setting has changed from
profile setting to care unit
setting. When transferring
settings to software version
3.1, the previous selections
are lost.
PCWP kPa 0-4 0-4 Trends > Trend Scales > Profile setting. Separate for
0-8 IP/NIBP each user profile.
Active if Blood Pressure
unit is kPa. The options
have been separated for
different measurement units.
When transferring settings
to software version 3.1, the
previous selections are lost.
PCWP mmHg 0 - 30 0 - 30 Trends > Trend Scales > Profile setting. Separate for
0 - 60 IP/NIBP each user profile.
Active if Blood Pressure
unit is mmHg. The options
have been separated for
different measurement units.
When transferring settings
to software version 3.1, the
previous selections are lost

2109358-005 CARESCAPE B450 and B650 17


Setting Available Default Menu Notes
values value
CPP kPa 0 - 13 0 - 26 Trends > Trend Scales > Profile setting. Separate for
0 - 20 (NICU: 0 - 13) IP/NIBP each user profile.
0 - 26 Active if Blood Pressure
unit is kPa. The options
have been separated for
different measurement units.
When transferring settings
to software version 3.1, the
previous selections are lost.
CPP mmHg 0 - 100 0 - 200 Trends > Trend Scales > Profile setting. Separate for
0 - 150 (NICU: 0 - 100) IP/NIBP each user profile.
0 - 200 Active if Blood Pressure
unit is mmHg. The options
have been separated for
different measurement units.
When transferring settings
to software version 3.1, the
previous selections are lost.

New Spirometry options


Setting Available Default Menu Notes
values value
Humidifica- HME HME Parameter Setup > Profile setting. Separate for
tion Type Spirometry/ Gas Exchange > each user profile.
Humidifier Setup
None
Mvexp OFF OFF Default Setup > Profile Profile setting. Separate for
Settings > Alarm Absolute each user profile.
0.6 to 30 in Limits
0.2 steps Mvexp alarm configuration
now depends on the selected
sensor type Adult/Pediatric.

New SpO2 options


Setting Available Default Menu Notes
values value
Averaging 5s 12 s Parameter Setup > SpO2 > Profile setting. Separate for
12 s SpO2 > Setup each user profile.
Averaging (SpO2 Ch1 GE)
is used for channel 1 when
CARESCAPE SpO2 is the
measurement source.
SpO2(2) 15 colors Cyan Monitor Setup > Colors > Profile setting. Separate for
Common each user profile.

18 CARESCAPE B450 and B650 2109358-005


Setting Available Default Menu Notes
values value
Variable Beat Off Off Parameter Setup > SpO2 > Profile setting. Separate for
Tone Source SpO2 Setup each user profile.
SpO2(2) Variable Beat Tone Source
selection is available for all
supported SpO2 channels.
SpO2 high 0s 5s Default Setup > Profile Profile setting. Separate for
5s Settings > Alarm Delays each user profile.
10 s
15 s
SpO2 low 0s 5s Default Setup > Profile Profile setting. Separate for
5s Settings > Alarm Delays each user profile.
10 s
15 s

New settings: V Tach Criteria


Setting Available Default value Menu Notes
values
Event Duration Off Off Alarms Setup > Profile setting. Separate
5s Arrhytmia > Lethal for each user profile.
Alarms > V Tach The Event Duration
10 s Criteria selection is only
15 s available if V Tach
20 s Event Duration Criteria
25 s is Allowed.
30 s
V Tach Criteria AND AND Alarms Setup > Profile setting. Separate
OR Arrhytmia > Lethal for each user profile.
Alarms > V Tach The V Tach Criteria
Criteria selection is only
available if V Tach
Event Duration Criteria
is Allowed.
V Tach Event Allowed Not allowed Monitor Setup > Profile setting. Separate
Duration Criteria Not Default Setup > Profile for each user profile.
allowed Settings > Alarm
Delays

New Temperature option


Setting Available Default Menu Notes
values value
Show Tblood On On Parameter Setup > Profile setting. Separate for
- T3 Temperatures > Tblood each user profile.
Off > Setup

2109358-005 CARESCAPE B450 and B650 19


New Trends options
Setting Available Default Menu Notes
values value
SPO2 Range: 0 - 96 70 Trends > View > Histogram > Profile setting. Separate for
Lower Value Histogram Setup > SpO2 each user profile.
SPO2 Range: 4 -100 100 Trends > View > Histogram > Profile setting. Separate for
Upper Value Histogram Setup > SpO2 each user profile.
SPO2 Target 0 - 98 87 Trends > View > Histogram > Profile setting. Separate for
zone Limits: Histogram Setup > SpO2 each user profile.
Lower Limit
SPO2 Target 2 - 100 94 Trends > View > Histogram > Profile setting. Separate for
zone Limits: Histogram Setup > SpO2 each user profile.
Upper Limit
HR Range: 0 - 280 30 Trends > View > Histogram > Profile setting. Separate for
Lower Value Histogram Setup > HR each user profile.
HR Range: 20 - 300 160 Trends > View > Histogram > Profile setting. Separate for
Upper Value Histogram Setup > HR each user profile.
HR Target 0 - 290 50 Trends > View > Histogram > Profile setting. Separate for
zone Limits: Histogram Setup > HR each user profile.
Lower Limit
HR Target 10 - 300 150 Trends > View > Histogram > Profile setting. Separate for
zone Limits: Histogram Setup > HR each user profile.
Upper Limit
Time Scale 20min 2h Trends > View > Histogram Profile setting. Separate for
1h ED: 1 h each user profile.
2h
4h
6h
8h
10h
12h
24h
Custom
With 72 Hour
trending
option:
36h
48h
72h
With High
Resolution
Trends
option:
2min
4min

20 CARESCAPE B450 and B650 2109358-005


New UI options
Setting Available Default Menu Notes
values value
SPO2(2) & On On Default Setup > Care Unit Care unit setting.
SVO2/ScvO2 Settings > Show in Screen
Off Setup

New Ventilator disconnected alarm priority


Setting Available Default value Menu Notes
values
Ventilator Medium OR and PACU: Alarms Setup > Alarm Profile setting. Separate for
disconnected High Medium Priorities > Other Parameters each user profile.
alarm priority ICU, NICU, and
ED: High

2109358-005 CARESCAPE B450 and B650 21


22 CARESCAPE B450 and B650 2109358-005
New settings in software version 2.3
Change in non-invasive and invasive blood pressure alarm
priorities
Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
NIBP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Art alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Fem alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
FemV alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
PA alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

2109358-005 CARESCAPE B450 and B650 23


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
CVP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
LAP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
RAP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
ICP alarm Priorities > Other each user profile.
priority Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
CPP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
RVP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

24 CARESCAPE B450 and B650 2109358-005


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
UAC alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
UVC alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P1 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P2 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P3 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P4 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

2109358-005 CARESCAPE B450 and B650 25


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
P5 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P6 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P7 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P8 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

New option: Snapshot sweep speed


Settings Available Selection Default Menu Notes
setting in v2.3 value in
v2.3
Snapshot Sweep 12.5 mm/s 12.5 mm/s Trends > View > Profile settings. Separate
Setup Speed 25 mm/s Snapshot > Snapshot for each user profile.
Setup
50 mm/s

26 CARESCAPE B450 and B650 2109358-005


New option: 12 lead print speed
Setting Selection Selection Default value in v2.3 Menu Notes
in v2.0.8 in v2.3
12 Lead 50 mm/s 25 mm/s 25 mm/s Default Setup > Care unit setting.
Print Not 50 mm/s Care Unit Settings Separate for each
Speed possible > Parameters > 12 user profile.
to lead ECG
change.

Extended settings: Allowed alarm priorities


Setting Selection Selection Default Menu Notes
in version in version value in
2.08 2.3 version 2.3
NIBP alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Priorities > Other Separate for each user
High High Medium Parameters profile.
Medium Medium Selection needs to be
activated in Default
Low Low
Setup > Care Unit
or Settings > Parameters
> Parameter Alarms
High > Allowed Priorities >
Medium parameter group.

IP alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority High Medium Priorities > Invasive Separate for each user
High Pressures profile.
Medium
Medium Selection needs to be
Low activated in Default
Low Setup > Care Unit
or
Settings > Parameters
High > Parameter Alarms
> Allowed Priorities >
Medium parameter group.
CO2 alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Medium Priorities > Other Separate for each user
High High Parameters profile.
Medium Medium Selection needs to be
activated in Default
or Setup > Care Unit
Settings > Parameters
High > Parameter Alarms
Medium > Allowed Priorities >
parameter group.
FiAA alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Medium Priorities > Other Separate for each user
High High Parameters profile.
Medium Medium Selection needs to be
activated in Default
or Setup > Care Unit
Settings > Parameters
High

2109358-005 CARESCAPE B450 and B650 27


Setting Selection Selection Default Menu Notes
in version in version value in
2.08 2.3 version 2.3
> Parameter Alarms
Medium > Allowed Priorities >
parameter group.

28 CARESCAPE B450 and B650 2109358-005


New settings in software version 2.0.8
New option: ECG lead off alarm priority
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ECG lead Selection Low Low Alarm Setup > Alarm Profile setting. Separate
off alarm not Informa- Priorities > ECG for each user profile.
priority available tional The new selection needs
to be activated in Default
Setup > Care Unit
Settings > Parameters
> Parameter Alarms >
Allowed Priorities > ECG

Extended settings: PVC, SVC, Impedance Resp, ST and Temp


alarm priority
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
PVC alarm Escalating Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
priority High High Priorities > ECG for each user profile.
Medium Medium
Low Low
Informa-
tional
SVC alarm Escalating Escalating Low Alarm Setup > Alarm Profile setting. Separate
priority High High Priorities > ECG for each user profile.
Medium Medium
Low Low
Informa-
tional
RR (Imped.) Escalating Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
high/low High High Priorities > Other for each user profile.
alarm Parameters The new selection needs
priority Medium Medium
to be activated in Default
Low Low Setup > Care Unit
Informa- Settings > Parameters
tional > Parameter Alarms
> Allowed Priorities >
Other Parameters

2109358-005 CARESCAPE B450 and B650 29


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ST alarm High High Medium Alarm Setup > Alarm Profile setting. Separate
priority Medium Medium Priorities > ECG for each user profile.
Low Low The new selection needs
to be activated in Default
Informa- Setup > Care Unit
tional Settings > Parameters
> Parameter Alarms >
Allowed Priorities > ECG
Parameters
Temp Selection Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
alarm not Informa- Priorities > Other for each user profile.
priority available tional Parameters The new selection needs
to be activated in Default
Setup > Care Unit
Settings > Parameters
> Parameter Alarms
> Allowed Priorities >
Other Parameters

Change in Art and ICP Invasive pressure labels


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
Art Art can be Art can be Multiple Parameter Setup > Profile setting. Separate
assigned to assigned to Invasive Pressure > Px > for each user profile.
one InvBP multiple In- Setup If you assign additional
channel at vBP chan- (Px = P1, P2, P3...) Art channels, the settings
a time nel simul- are copied from the
taneously original Art channel.
ICP ICP can be ICP can be Multiple Parameter Setup > Profile setting. Separate
assigned to assigned to Invasive Pressure > Px > for each user profile.
one InvBP multiple In- Setup If you assign additional
channel at vBP chan- (Px = P1, P2, P3...) ICP channels, the settings
a time nel simul- are copied from the
taneously original ICP channel.

30 CARESCAPE B450 and B650 2109358-005


Change in other Invasive pressure labels
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ABP available not N/A N/A You need to check the
available settings manually after
upgrading. *)
UAC available available N/A Parameter Setup > Profile setting. Separate
only in in OR, Invasive Pressure > Px > for each user profile.
NICU SW ICU,NICU, Setup Settings transferred in
package ED and (Px = P1, P2, P3...) NICU SW package only.
PACU SW
packages
UVC available available N/A Parameter Setup > Profile setting. Separate
only in in OR, Invasive Pressure > Px > for each user profile.
NICU SW ICU,NICU, Setup Settings transferred in
package ED and (Px = P1, P2, P3...) NICU SW package only.
PACU SW
packages

*) NOTE: Some of the ABP settings are transferred from the original Art and ABP
settings. To ensure proper settings configuration, GE recommends that you check
the settings manually after upgrading.

Extended settings: ECG Printout format


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ECG 4x2.5 – 1 4x2.5 – 1 4x2.5 – 1 Default Setup > Care Unit Profile setting. Separate
Printout Rhythm Rhythm Rhythm Settings > Parameters > for each user profile.
format 4x2.5 – 3 4x2.5 – 3 ECG
Rhythm Rhythm
12 12
Rhythms Rhythms
25mm/s 25mm/s
12 12
Rhythms Rhythms
50mm/s 50mm/s
Cabrera 2x6
Rhythms
50mm/s
Cabrera
50mm/s

2109358-005 CARESCAPE B450 and B650 31


32 CARESCAPE B450 and B650 2109358-005
New settings in software version 2.0.7
Extended settings: Pacemaker Detection with PDM
Settings Selection Selection Default value Menu Notes
in version in version
2.0.6 with 2.0.7 with
PDM PDM 2.4
Pace- On Sensitive Sensitive (with PDM Parameter Setup > Profile setting.
maker Off Normal 2.4) ECG > Advanced Separate for each
detection (New) On (with PDM 2.3 or user profile.
Off older) New selection
Sensitive (with available only with
E-PSM(P)) PDM 2.4 and software
version 2.0.7.
Pace2 (with TRAM)

Extended settings: 12SL Display Format


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 2.0.7
12SL 4x2.5 - 1 4x2.5 - 1 4x2.5 - 3 Parameter Setup > ECG Profile setting. Separate
Display Rhythm Rhythm Rhythm > 12 Lead Analysis > for each user profile.
Format 4x2.5 - 3 4x2.5 - 3 Settings
Rhythm Rhythm
12 12
Rhythms Rhythms
Cabrera 2x6
Rhythms
Cabrera

Extended settings: New default value for Spirometry sensor


type
Settings Selection New default value in Menu Notes
in version version 2.0.7
2.0.6
Spirometry Adult All user profiles in NICU Parameter Setup Profile setting. Separate
sensor type Pediatric software package: > Spirometry/Gas for each user profile.
Pediatric Exchange
User profiles Infant and
Pediatric in ED, ICU,
OR, and PACU software
package: Pediatric
User profiles other than
Infant and Pediatric in
ED, ICU, OR, and PACU
software package: Adult

2109358-005 CARESCAPE B450 and B650 33


Change in Menu structure (Diagnostic Tool)
Settings Selection Combined Default Menu Notes
in version menu in value in
2.0.6 version version
2.0.7 2.0.7
ACITIPI On/Off ACITIPI ACITIPI Parameter Setup > ECG Moved to other menu
ACS > 12 Lead Analysis > destination.
ACS On/Off Settings
Off Separate for each user
profile.

New global options


Settings New Default Menu Notes
available value
setting in
version 2.0.7
Citrix Server Support for Default Webmin > Configuration > Platform setting. For all user
version Citrix server server Citrix profiles.
versions version:
4.0/4.5 and 4.0/4.5
5.0.
Citrix Server Support Default IP Webmin > Configuration > Platform setting. For all user
address 2–4 for 4 Citrix address: Citrix profiles.
servers with Empty
4 different IP
addresses.
Alarm New Immediate Default Setup > Care Unit Care Unit setting. For all user
Activation for available Settings > Parameters > profiles.
PR selection: Allowed priorities > Other
Immediate Parameters
or Delayed
Print Inter- New Enabled Parameter Setup > ECG > 12 Profile setting. Separate for
pretations available Lead Analysis > Settings each user profile.
selection:
Enabled or
Disabled
Low-Pass New 150Hz Parameter Setup > ECG > 12 Same default for all user
Filter available Lead Analysis > Filter profiles.
selections: Not possible to change and
20Hz or 40Hz save as profile of unit settings.
or 100Hz or
150Hz

New options in NICU software package with CARESCAPE


Respiratory module
NOTE: Not available in the US and its territories.
NOTE: Not available in China.

34 CARESCAPE B450 and B650 2109358-005


Settings Available New default Menu Notes
setting setting in
version 2.0.7
Alarm Limit New factory 2 Parameter Setup > Available only with
default for Spirometry > MV/Vent “E-sC(Ai)O(V) module NICU
Mvexp High Alarms SW” license.
Alarm Limit Profile setting. Separate for
each user profile.
New factory 30 Parameter Setup > Available only with
default for Spirometry > PAW Alarms “E-sC(Ai)O(V) module NICU
Ppeak High SW” license.
Alarm Limit Profile setting. Separate for
each user profile.
New factory 8 Parameter Setup > Available only with
default for Spirometry > PAW Alarms “E-sC(Ai)O(V) module NICU
PEEPe High SW” license.
Alarm Limit Profile setting. Separate for
each user profile.

2109358-005 CARESCAPE B450 and B650 35


36 CARESCAPE B450 and B650 2109358-005
content
GE Healthcare Finland Oy
Kuortaneenkatu 2
FI-00510 Helsinki
Finland
Tel: + 358 10 39411

www.gehealthcare.com