Professional Documents
Culture Documents
Aestiva MRI
Aestiva MRI
Aestiva MRI
I. GENERAL INFORMATION
The following planned maintenance procedures are to be performed every 6 months, except where
specifically stated differently.
WARNING: This procedure contains limited warnings and cautions. Refer to the product
Service Manuals for a full listing of precautions.
4. Remove the screws securing the electronic compartment cover and remove the cover.
5. Disconnect the battery cable connector at the power supply PCB.
6. Remove the battery cable securing clamp at the upper perforated bracket.
7. Slide the battery out of its holding bracket.
8. Record the replacement date on the new battery and install the new battery.
9. Reassemble by reversing the above procedure. Ensure the battery cable is secured to
the upper perforated bracket to keep it clear of the AC Inlet connector. Also ensure the
AC Inlet module securing screws cannot be loosened without the use of a tool.
C. Switch Checks
1. From the Service Mode main menu select Display I/O Signals.
2. Test each switch listed by the following methods. Ensure the status of each switch
changes appropriately.
a. Control Panel Cover: Open and close the control panel or move the switch lever with
your finger.
b. CO2 Bypass: Push on the switch button located in the cavity near the front and top of
the absorber canister.
c. ACGO Switch, if equipped: Cycle the ACGO lever.
d. Bag/Vent: Cycle the bag/vent switch.
e. O2 Sensor: Disconnect and connect the O2 sensor.
f. Canister: Open and close the absorber canister handle.
g. Breathing Circuit Switches 1-4: Use a small piece of paper or similar small object to
block the IR light beam of each switch. The switch numbers are (from left to right) 1,
3, 2. Switch 4 is not used. All switches are normally ON.
h. Verify DPL Switch is OFF.
i. O2 Flush: Push the O2 flush button. Must have O2 pressure.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 3 of 17 09/19/02 Number: 2.01
F. Diagnostic Tests
1. Select and run the following tests from the Service Mode Main Menu:
a. Test Panel Switches
b. Test Serial Ports (internal test only)
c. Test Flow Valve
d. Test Gas Inlet Valve
e. Test Pressure Limit Switch
G. Calibration
1. Select and run the following calibration routines from the Service Mode Main
a. Cal O2 Sensor
b. Cal Flow Sensor
c. Cal Pressure Sensitivity
d. Cal Flow Valve
e. Cal Bleed Resistor
2. Exit the Service Mode.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 4 of 17 09/19/02 Number: 2.01
NOTE: This test is only applicable to those gases with cylinder supplies. If there are
no cylinder supply yokes on the machine, skip this section and proceed to
step VII.
Note: There are two different methods for performing the Pressure Sensor/Low O2 Alarm
test. Determine the output pressure of the gas supply module regulator and
perform the test that is applicable.
1. 97 PSI REGULATORS
a. Verify all pipeline supplies are connected and open all cylinder valves.
b. Ensure all cylinder gauges indicate pressure above the red zone of the gauge
c. Close cylinder valves only for gases which also have pipeline supplies.
d. Set all flowmeters for 3 L/min and observe the cylinder gauges.
e. Verify that the cylinder gauges Drop to 0 psi for gases with pipeline supplies.
f. Remove the oxygen pipeline supply and verify the following events occur in sequence:
1. Ventilator audio alarm activates.
2. Flowmeter floats for all gases, except Air, fall to zero. Air should continue to flow
without disruption.
g. Ensure that the message “No O2 Pressure” is displayed on the ventilator.
h. Close all flow control valves.
i. Re-establish the O2 pipeline supply.
j. Verify that the audio alarm silences and the message clears.
k. Proceed to Step F, O2 Flush Valve Test.
2. 45 PSI REGULATORS
a. Verify all pipeline supplies are connected and open all cylinder valves.
b. Close cylinder valves.
c. Ensure all cylinder gauges indicate pressure above the red zone of the gauge.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 6 of 17 09/19/02 Number: 2.01
•Insert the gauge at a corner as shown and test all along the surface to the other
corner.
•Minor interference can be removed with extra fine sandpaper.
•If the clearance cannot be achieved with light sanding, the toggle should be
replaced.
B. Flow Sensor Module
1. Inspect the housing, tubing and connectors of both flow sensors. Ensure the tubing is not
kinked.
2. Inspect the flow sensor connector mounted in the absorber. Ensure the electrical and
pneumatic connection points are clean with no residue.
3. Replace any components that look damaged or worn.
4. Set aside for later installation.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 7 of 17 09/19/02 Number: 2.01
5. For circle circuit modules, inspect the drain valve o-ring and seal. Replace as necessary.
Check for proper operation. Clean any contamination that may be present.
6. Reassemble the breathing circuit module.
7. Reinstall the O2 sensor and set the module aside.
8. Repeat steps 1-7 for every breathing circuit module available.
D. Bag Arm
1. (For long bag arm only) Check the bag arm for free rotation at both ends. Check the
rotation end stops for proper function.
2. Remove the bag arm from the absorber and set aside.
G. Bellows Inspection
1. Remove the bellows assembly and completely disassemble.
2. Inspect all removed components. Check that the bellows housing is not cracked, chipped
or damaged. Also check the condition of the housing lettering. Replace any components
that appear damaged, distorted or worn. Clean any that are contaminated.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 8 of 17 09/19/02 Number: 2.01
1. Remove the exhalation valve assembly from the absorber and completely disassemble.
2. Inspect all exhalation valve components. Replace any that look damaged, distorted or
worn. Clean any that are contaminated. Sparingly lubricate the standpipe U-cup seal with
Kryrox.
3. Inspect the AGSS “Y” seal. Replace it if it looks damaged, worn, or distorted. Ensure it is
fully seated on the absorber sub-floor.
4. Reassemble the exhalation valve and set aside for a later test.
M. Absorbent Canister
1. Release the canister lock ensuring that the handle operates properly and smoothly.
Lubricate the absorber canister locking mechanism.
2. Push down the bottom dish and remove the absorbent canisters.
3. Remove the upper dish/lower dish assembly with corrugated tube attached and inspect.
Replace any parts that look damaged, distorted or worn. Clean any that are contaminated.
Ensure corrugated tube is tightly secured at both ends.
4. Inspect the transfer tube and seal in the rear of the canister area. Ensure the seal is
properly attached to the scavenging pipe. Replace any parts that look damaged, distorted
or worn. Clean any that are contaminated.
5. Reassemble the canister assembly and canisters in the absorber in the reverse order.
Lock the canister in place. Ensure the canister handle operates smoothly and locks into
position.
NOTE: The old style relief valve may allow a negative pressure indication up to -2
cmH2O. If you find an old style relief valve, replace it with the new style valve
assembly.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 10 of 17 09/19/02 Number: 2.01
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 11 of 17 09/19/02 Number: 2.01
NOTE: If the MOPV Valve doesn’t pass the above test, adjust the height of the white
standpipe in the exhalation valve port. Adjusting the height of the standpipe will
eliminate the drive pressure limit switch from the circuit and allow the test to be
performed.
R. Reassembly of Absorber
1. Install the exhalation valve.
2. Install the breathing system manifold.
3. Install the bellows assembly.
4. Close the breathing system control panel.
5. Install the bag arm.
6. Install the breathing circuit module and connect the O2 sensor cable.
7. Install the flow sensor module.
8. Install the gas scavenging (AGSS) module.
NOTE: For the 12 and 24 month PM, the AGSS module must be left out for accessibility
to the AGSS area later in this procedure. It will be reinstalled in the Ventilator
section.
4. When the APL valve is set to minimum (fully counter-clockwise), the pressure gauge
should read close to zero (approximately 0.5 – 1.1 cmH20).
5. Push O2 flush and ensure the pressure maintains near zero.
6. Remove the tube between the inhalation and bag arm ports.
(Every 12 Months)
1. Check and zero the pressure gauge as required (Refer to section 6.6).
2. Ensure that the AGSS is on and operational.
3. Plug the patient connection by connecting it to the test port on the
handle.
4. Plug the bag arm with a test plug.
5. Fully open the APL to MIN.
6. Set the O2 flow to 3 L/min. The pressure indicated on the pressure
gauge should not exceed 3.0 cm H2O.
7. Adjust the total fresh gas flow to 30 L/min*. The pressure indicated on
the pressure gauge should not exceed 4.0 cm H2O.
8. Check the valve at mid-range settings of about 20, 30, and 50 while
flowing 30 L/min. Note that scale markings are only approximate. The
pressure should stabilize at each setting; however, fluctuations within
an 8 cm H2O range (min to max) are acceptable**. Ensure that the APL
valve exhibits a tactile detent feel when turning in the 30–70 cm H2O
range and that the knob does not rotate on its own.
9. Set the APL to 70 while flowing 30 L/min. The pressure indicated on the
pressure gauge should be between 60–85 cm H2O, including any
fluctuations**.
10. Set the APL to MIN.
11. Remove the plug from the bag arm and replace it with the bag. The bag should fully
inflate.
12. Reduce the O2 flow to minimum and all other flow controls off.
13. The bag should remain inflated.
* To achieve 30 L/min total fresh gas flow, set the O2 flow to 15 L/min and the
Air flow to 15 L/min. Alternatively, set the O2 flow to 15 L/min and the N2O flow
to the maximum stop less approximately 1/2 turn.
WARNING Nitrous oxide flows through the system during this test. Use a safe
and approved procedure to collect and remove it.
** Pressure fluctuations can occur with this test due to the steady high flow rate
used. However, there should be no fluctuations or spikes outside of an 8 cm
H2O range (min to max) at any setting. This will not occur during normal
bagging operation with lower fresh gas flow rates.
WARNING Make sure that there are no test plugs or other objects caught in the
breathing system.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 13 of 17 09/19/02 Number: 2.01
WARNING: Use the hospital evacuation system with the machine AGSS. If hospital
evacuation system is not available, perform service in a well-ventilated area.
5. Remove the flowmeter verification test device from the common gas outlet.
6. Install the breathing circuit module.
7. Install the flow sensor module.
C. PROPORTIONING SYSTEM
1. Connect a hose between the inhalation and exhalation ports of the flow sensor module.
2. Occlude the bag arm with a rubber stopper.
3. Set the Bag/Vent switch to Bag.
4. Adjust the APL Valve to minimum (fully counter clockwise).
5. Ensure all flow control knobs are set to minimum flow (fully clockwise). The O2 flow should
be between 25 and 75 mL/min.
6. Expose the O2 sensor to room air and perform the O2 calibration from the calibration menu
on the ventilator.
7. Reinstall the O2 sensor into the breathing system.
8. Use the N2O Flow Control Valve and perform the tests in Table A progressively from low to
high flows. Do not overshoot any setting. Allow the System O2 monitor to stabilize between
readings.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 14 of 17 09/19/02 Number: 2.01
TABLE A
Set the N2O The O2 monitor must read between
Flow to (L/min) the following percentages
System O2 Monitor
0.15 21% Minimum
0.5 21% Minimum
0.8 21 - 30%
1.0 21 - 30%
3.0 21 - 30%
6.0 21 - 30%
9.0 21 - 30%
TABLE B
Set the O2 The O2 monitor must read between
Flow to (L/min) the following percentages
System O2 Monitor
3.0 21 - 30%
2.0 21 - 30%
1.0 21 - 30%
0.3 21 - 30%
12. Decrease N2O flow to minimum. The O2 float should not decrease at this time.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 15 of 17 09/19/02 Number: 2.01
13. Check for additional gas proportioning by using the appropriate table(s) below.
TABLE C
Oxygen Air Oxygen
Monitor Range
4 L/min 3 L/min 61-71%
3.5 L/min 6 L/min 45-55%
1.5 L/min 8 L/min 28-38%
TABLE D
Oxygen Heliox (25% Oxygen, 75% Oxygen
Helium Monitor Range
4.0 L/min 3.0 L/min 63-73%
3.5 L/min 6.0 L/min 48-58%
1.5 L/min 8.0 L/min 32-42%
TABLE E
Oxygen Carbon Dioxide Oxygen
Monitor Range
2 L/min 0.2 L/min 86-96%
0.5 L/min 0.5 L/min 45-55%
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 16 of 17 09/19/02 Number: 2.01
TABLE F
Ventilator Settings Volume Readings (mL)
VT (mL) Rate I:E Minimum Nominal Maximum
500 10 1:3 450 500 550
1000 10 1:3 900 1000 1100
1000 10 1:5 900 1000 1100
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 17 of 17 09/19/02 Number: 2.01
NOTE: Perform electrical Safety Tests every time the cover to a machineis removed.
NOTE: If tests are conducted with power from an isolated electrical power system, conduct the
tests as detailed below. Do not use an electrical jumper plug to defeat the isolation of
the power system.
A. Preparation
1. Turn the System Master Switch to the STANDBY position.
2. Disconnect all pipeline hoses from the medical gas outlets.
3. Disconnect all electrical monitors and accessories from the unit under test.
4. Plug the unit under test into the safety analyzer. Plug the analyzer into an AC wall outlet.
5. Set the safety analyzer to measure line voltage. Verify normal polarity is indicated.
6. Connect the grounding clip assembly to the safety analyzer.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
04/2007
AA Aestiva/5 MRI 18 of 17 09/19/02 Number: 2.01
1. Disconnect all test devices and reconnect pipeline hoses, monitors and accessories.
2. Verify correct operation of the light bar and/or gooseneck light (if equipped).
3. Ensure the following:
a. All vaporizers are OFF.
b. All flow control valves are at the minimum stop.
c. The APL valve is open (minimum setting, fully counterclockwise).
d. All panels and tray top are attached.
e. The pipeline hoses are connected.
f. All cylinder gauges are at zero.
g. All units are plugged into an appropriate power source.
4. Place the appropriate labels on the machine advising that routine service has taken place and
that the controls have been adjusted.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
App A 04/2007
AA Aestiva/5 MRI 1 of 2 09/19/02 Number: 2.01
Appendix A
Required Tools
Except where indicated by an asterisk (*), locally purchased tools of similar type and specifications may
be used.
Ensure test equipment is properly and currently calibrated. Test equipment accuracy must be taken into
account for all tests and checks.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
App A 04/2007
AA Aestiva/5 MRI 2 of 2 09/19/02 Number: 2.01
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
App B 04/2007
AA Aestiva/5 MRI 1 of 1 09/19/02 Number: 2.01
Appendix B
Required Parts
The following is a list of parts requiring periodic/scheduled replacement on the Aestiva MRI Anesthesia
Machine and are referred to in this procedure.
This list does not contain any non-scheduled parts replaced on an as-required basis. For these
parts, refer to Aestiva Service Manuals 1006-0452-000 and 1006-0453-000.
• Every 6 Months
1. Backup Battery 1503-3045-000
• Every 12 Months
1. Vaporizer Port O-Ring Kit 1102-3016-000
• Every 24 Months
1. Free Breathing Valve 0211-1454-100
2. O-Ring for Free Breathing Valve 1503-3208-000
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
App C 04/2007
AA Aestiva/5 MRI 1 of 2 09/19/02 Number: 2.01
Appendix C
Tools/Adapters to Assist in Servicing the Aestiva 3000
The following list is provided for convenience only should troubleshooting or repair be required during the
PM procedure.
Except where indicated by an asterisk (*), locally purchased tools of similar type and specifications may
be used.
Ensure test equipment is properly and currently calibrated. Test equipment accuracy must be taken into
account for all tests and checks.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Mfr. Description Page Date Issued: Revision Date:
App D 04/2007
AA Aestiva/5 MRI 1 of 1 09/19/02 Number: 2.01
Appendix D *
Maintenance Procedure Revision Log
* Note: This document has been under revision control, but a revision history has not
been required to be included with this procedure. From this revision going forward, this
log will be updated to inform the user of changes made to this document.
MP 1032 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.