Professional Documents
Culture Documents
IDT Foc Error
IDT Foc Error
May 2002
ALL PAGES OF THIS DOCUMENT CONTAIN PROPRIETARY AND CONFIDENTIAL INFORMATION OF PHILIPS MEDICAL SYSTEMS
(CLEVELAND), INC. THE DOCUMENTS ARE INTENDED FOR CURRENT PHILIPS MEDICAL SYSTEMS’ PERSONNEL OR ARE LICENSED TO
PHILIPS MEDICAL SYSTEMS’ CUSTOMER FOR USE BY THE CUSTOMER'S IN-HOUSE SERVICE EMPLOYEE ON EQUIPMENT LOCATED AT THE
CUSTOMER'S DESIGNATED SITE. COPYING, DISCLOSURE TO OTHERS OR OTHER USE IS PROHIBITED WITHOUT THE EXPRESS WRITTEN
AUTHORIZATION OF PHILIPS MEDICAL SYSTEMS’ LAW DEPARTMENT. REPORT VIOLATIONS OF THESE REQUIREMENTS TO PHILIPS
MEDICAL SYSTEMS’ LAW DEPARTMENT, HIGHLAND HEIGHTS, OHIO
Copyright © Siemens AG. All rights reserved. For internal use only.
5=INIT_TIMEOUT
6=PREPARE_TIMEOUT
Copyright © Siemens AG. All rights reserved. For internal use only.
given with the mode tables
7: FOC –15.00V
8: FOC + 5.00V
Alle Rechte vorbehalten. Nur für internen Gebrauch.
9: FOC +15.00V
P2: 0
P3: not used
E 0xA0 F 0x32 COMP both sensors P1: 0
P2: 0
Both COMP sensors detect both
P3: 0
end positions at the same time.
E 0xA0 W 0x33 COMP timeout P1:
1=during moving OUT
The compensator does not reach
2=during moving IN
the position in the estimated time
P2: 0
P3: 0
E 0xA0 W 0x34 XRAY count=0 P1: 0
P2: 0
FOC has no XRAY detected
P3: 0
E 0xA0 F 0x35 COMP must exist (TAA) P1: 0
P2: 0
FOC can not detect a functional
P3: 0
working compensator after
receiving table TAA
E 0xA0 F 0x36 wrong block count (TAB) P1:
1=less than 2 blocks
At least block 1 and block 2 must
2= gap in used blocks detected
be in use in Table TAB. A gap of
unused blocks is not allowed. P2: 0
P3: 0
E 0xA0 F 0x37 UHR both sensors P1: 0
P2: 0
Both UHR sensors detect both end
P3: 0
positions at the same time
Copyright © Siemens AG. All rights reserved. For internal use only.
jaws if bit TAAzcodebug is set.
P3: 0
P3: 0
Alle Rechte vorbehalten. Nur für internen Gebrauch.
Copyright © Siemens AG. All rights reserved. For internal use only.
P2: 0
During the ZCO test no stepper
P3: 0