TENDER ENQUIRY DOCUMENT

FOR PURCHASE OF
MEDICAL EQUIPMENT

FOR AND ON BEHALF OF


EMPLOYEE STATE INSURANCE CORPORATION
UNDER THE ADMINISTRATIVE CONTROL OF
MINISTRY OF LABOUR AND EMPLOYMENT
GOVT. OF INDIA

HLL/PCD/ESIC-20/10-11

BY

HLL Lifecare Limited


(A GOVT. OF INDIA ENTERPRISE & Formerly HINDUSTAN LATEX LTD.)
Procurement & Consultancy Services Division
B-14A, Sector-62, Noida-201 307, Uttar Pradesh, India
Ph: 0120-4071500; Fax: 0120-4071513
URL: www.lifecarehll.com
Email: pcd@lifecarehll.com
HLL Lifecare Limited

INDEX
Section Topic Page No.

Section I – Notice inviting Tender (NIT) ---------------------------------------------------------- 03


Section II – General Instructions to Tenderers (GIT) --------------------------------------------- 07
Section III – Special Instructions to Tenderers (SIT) ---------------------------------------------- 24
Section IV – General Conditions of Contract (GCC) ---------------------------------------------- 25
Section V – Special Conditions of Contract (SCC) ----------------------------------------------- 40
Section VI – List of Requirements ------------------------------------------------------------------- 41
Section VII – Technical Specifications --------------------------------------------------------------- 45
Section VIII – Quality Control Requirements ------------------------------------------------------- 141
Section IX – Qualification Criteria ------------------------------------------------------------------- 142
Section X – Tender Form ----------------------------------------------------------------------------- 144
Section XI – Price Schedules -------------------------------------------------------------------------- 145
Section XII – Questionnaire ---------------------------------------------------------------------------- 149
Section XIII – Deleted
Section XIV – Manufacturer’s Authorisation Form -------------------------------------------------- 150
Section XV – Bank Guarantee Form for Performance Security /CMC Security ---------------- 151
Section XVI – Contract Form (A & B) ------------------------------------------------------------------152
Section XVII – Proforma of Consignee Receipt Certificate ----------------------------------------- 156
Section XVIII– Proforma of Final Acceptance Certificate by the Consignee --------------------- 157
Section XIX – Details of Shipping arrangement for Liner Cargoes in respect of C&F/CIF/
Turnkey/F.O.R. Contracts for Import-------------------------------------------------- 159
Section XX – Check List for the Tenderers ----------------------------------------------------------- 163
Section XXI – Consignee address ---------------------------------------------------------------------- 166

HLL/PCD/ESIC-20/10-11 Page 2 of 166 01.09.2010


HLL Lifecare Limited NIT

SECTION I

NOTICE INVITING TENDERS (NIT)


HLL LIFECARE LIMITED
(A GOVERNMENT OF INDIA ENTERPRISE)
Procurement & Consultancy Services Division
B-14A, Sector-62, Noida-201 307, Uttar Pradesh, India
Ph: 0120-4071500; Fax: 0120-4071513
Email: pcd@lifecarehll.com

Tender Enquiry No. HLL/PCD/ESIC-20/10-11 Date: 01.09.2010


1. Procurement & Consultancy Services Division of HLL Lifecare Limited (Formerly Hindustan Latex
Limited), for and on behalf of Director General of Employee State Insurance Corporation (ESIC),
invites sealed tenders, from eligible and qualified tenderers for supply of following Medical
Equipment for ESI Hospitals at different locations within India:

Total
Sl. EMD Amount
Item Description Quantity
No. Rs.
(Nos.)
1 Biphasic Defibrilator 8 56,000.00
2 Fiber Optic Bronchoscope with Camera 6 144,000.00
3 Pulse Oximeter 3 3,000.00
4 Anaesthesia Workstation 2 100,000.00
5 ECG Machine 12 channel 2 6,000.00
6 Pharmaceutical Refrigerator for O.T. 1 3,000.00
7 O T Tables for Orthopaedics 5 290,000.00
8 O T Tables for General Surgery & Gynaecology 13 520,000.00
9 Electro Surgical Unit 5 45,000.00
10 O T Table for Minor O T 1 7,000.00
11 Boyle's Appratus 1 20,000.00
12 Multipara Monitor (6Nos) with Central Station (1No) 2 144,000.00
13 Portable Ventilator 1 10,000.00
14 BIPAP 2 12,000.00
15 Binocular Microscope 6 27,000.00
16 Urine Analyzer 1 8,500.00
17 ELISA Reader with Washer 5 50,000.00
Cell Counter(Fully automated 5 part differential
18 2 28,000.00
analyser)
19 Biochemistry Semi- Auto Analyser 2 6,000.00
20 Coagulation Analyzer 2 20,000.00
21 Electrolyte Analyzer 1 8,000.00
22 500 M.A. X-ray machine 1 18,000.00
23 Portable X-ray machine (Mobile) 3 18,000.00
24 Colour Doppler Digital 1 40,000.00
25 Infant Resuscitator 1 10,000.00
26 Neonatal Monitor 6 48,000.00
27 Ventilator Critical Care 1 26,000.00
28 Open Care System 6 84,000.00

HLL/PCD/ESIC-20/10-11 Page 3 of 166 01.09.2010


HLL Lifecare Limited NIT

Total
Sl. EMD Amount
Item Description Quantity
No. Rs.
(Nos.)
29 Transport Incubators 1 14,000.00
30 Plaster cutter with Aspirator 1 10,500.00
31 Tourniquet Electronic 1 3,800.00
32 Hot Pack Unit 1 6,200.00
33 Cold Pack Unit 1 6,200.00
34 Traction Unit 1 18,000.00
35 Operating Laparoscope Set for Gynaec 1 70,000.00
36 Hysteroscope System 1 20,000.00
37 Cryo Surgical Unit 4 52,000.00
38 Vacuum Extractor 4 36,000.00
39 Fetal Monitor 7 56,000.00
40 ENT Examination Unit 4 200,000.00
41 Operating Laparoscope Set for General Surgery 2 140,000.00
42 Multipara Bed side Monitor 1 14,000.00

43 Emergency Resuscitation Kit-Adult(imported) 6 66,000.00

44 Portable Spirometer 1 12,000.00


45 Infusion Pump Syringe 5 6,000.00
46 PFT Machine(Spirometer & Computerised) 2 48,000.00
47 Fully Automated Biochemistry Analyzer 1 14,000.00
48 Centrifuge Machine 24-36 tubes 3 8,400.00
49 Computed Radiography System 1 100,000.00
50 Advancved Colour Doppler Ultrasound Scanner 1 120,000.00
51 Donor Chair 2 24,000.00
52 Blood Collection Monitor 3 12,000.00
53 Blood Bank Refrigerator 3 15,000.00
54 Tube Sealer 2 4,800.00
55 Cell Counter - 3 Parameter 1 8,000.00
56 Refrigerated Centrifuge 1 4,000.00
57 Plasma Expressor(E) 2 8,000.00
58 Dielectric Tube Sealer Table Top 1 3,000.00
59 PLATELET INCUBATOR & AGITATOR 1 4,000.00
AUTOMATED PLASMA THAWING
60 1 6,000.00
EQUIPMENT
61 LAMINAR FLOW- VERTICAL 1 8,000.00
62 Deep Freezer -40 Deg. 2 8,000.00
63 Deep Freezer -80 Deg. 1 6,000.00
64 Cell Separator 1 8,000.00
65 ICU Ventilator (High End) 1 24,000.00
66 ABG Analyzer 2 28,000.00
67 Arthroscopy System 1 90,000.00
68 Color Doppler Ultrasound Unit 1 36,000.00
69 Video Bronchoscopy System 1 80,000.00
70 High Frequency Mobile X- Ray Unit 1 20,000.00

HLL/PCD/ESIC-20/10-11 Page 4 of 166 01.09.2010


HLL Lifecare Limited NIT

Total
Sl. EMD Amount
Item Description Quantity
No. Rs.
(Nos.)
71 Digital Mamography System 1 180,000.00
72 Holter Monitoring System 2 32,000.00
73 Stress Test System(TMT) 1 2,400.00
74 Volumetric Infusion Pump 10 12,000.00
75 Biosafety Cabinet 1 16,000.00
76 Laprocator 2 20,000.00
77 Ultrasound Machine 3D with Color Doppler 2 108,000.00
78 Phototherapy Machine 1 15,000.00
79 Microscope 6 3,600.00
80 Yag Laser 1 50,000.00
81 Cell Separator 1 40,000.00
82 Cell Washer 1 14,000.00
83 Donor Couch 1 3,000.00
84 Platelet Incubator & Agitator 1 10,000.00
85 Sterile Tube Connecting Device 1 19,000.00
86 General Instruments Set 24 696,000.00
87 Surgery Instrument Set for Minor O.T. 4 60,000.00
88 Orthopaedic Surgery Instrument Set 1 70,000.00
89 Vaginal Hysterectomy 4 20,000.00
90 Abdominal Hysterectomy set 1 3,200.00
91 Delivery and Ceasarian set 4 24,000.00
92 General Surgical Instruments for O T 1 40,000.00
93 Instruments for Dental Dept. 1 14,000.00
94 Instruments for Eye Implant 1 20,000.00

2. Tender No.: HLL/PCD/ESIC-20/10-11

Sl
Description Schedule
No.
Dates of sale of tender enquiry 03.09.2010 to 04.10.2010, in all working
i.
documents days, during 1000 Hrs. to 1600 Hrs. (IST)
HLL Lifecare Limited
Place of sale of Tender Enquiry
ii. Procurement & Consultancy Services Divn.
Documents
B-14A, Sector-62, Noida -201 307
iii. Cost of the Tender Enquiry Document Rs. 3,000.00/ USD 75.00
iv. Pre Tender Meeting Date & Time 10.09.2010, 1100 Hrs. (IST)
v. Pre Tender Meeting Venue Same as given in 2 (ii)
Closing date & time for receipt of
vi. 05.10.2010, 1430 Hrs. (IST)
Tender
Time and date of opening of Techno-
vii. 05.10.2010, 1500 Hrs. (IST)
Commercial tenders
Venue of Opening of Techno-
viii. Same as given in 2 (ii)
Commercial Tender

HLL/PCD/ESIC-20/10-11 Page 5 of 166 01.09.2010


HLL Lifecare Limited NIT

3. Interested tenderers may obtain further information about this tender from the office of Head
(P&CD), HLL Lifecare Ltd., Noida. Tender Enquiry Documents may be purchased on payment of
non-refundable fee of Rs. 3,000.00/ USD 75.00 per set in the form of account payee Demand
Draft/Pay Order/Cashier’s Cheque/Banker’s Cheque, drawn on a scheduled bank in India, in favour
of “HLL Lifecare Limited” payable at New Delhi.

4. If requested, the Tender Enquiry Documents will be mailed by Registered Post/Speed Post to the
domestic tenderers and by international airmail to the foreign tenderers, for which extra expenditure
per set will be Rs 100.00 for domestic post and USD 50.00 for international airmail. The tenderer is
to add the applicable postage cost in the non-refundable fee mentioned in Para 3 above. However,
HLL Lifecare Ltd. shall not be responsible for any postal loss/delay.

5. Tenderer may also download the tender enquiry documents from the web site www.esic.nic.in or
www.lifecarehll.com and submit its tender by utilizing the downloaded document, along with the
required non-refundable fee as mentioned in Para 3 above.

6. All prospective tenderers may attend the Pre Tender meeting. The venue, date and time indicated in
the Para 2 above

7. Tenderers shall ensure that their tenders, complete in all respects, are dropped in the Tender Box
located at HLL Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector-62,
Noida -201307, Uttar Pradesh on or before the closing date and time indicated in the Para 2 above,
failing which the tenders will be treated as late tender and rejected. The tenders sent by post/ courier
must reach the above said address on or before the closing date & time indicated in Para 2 above,
failing which the tenders will be treated as late tender and rejected.
8. In the event of any of the above mentioned dates being declared as a holiday / closed day for the
purchase organisation, the tenders will be sold/received/opened on the next working day at the
appointed time.

9. The Tender Enquiry Documents are not transferable.

10. All Tenders must be accompanied by EMD as mentioned against each item. Tenders without EMD
shall be rejected.

For and on behalf of Employee State Insurance Corporation


Head (P & CD)
HLL Lifecare Limited,
Procurement and Consultancy Division
B-14A, Sector -62, Noida -201307,
Uttar Pradesh.

HLL/PCD/ESIC-20/10-11 Page 6 of 166 01.09.2010


HLL Lifecare Limited GIT

SECTION - II

GENERAL INSTRUCTIONS TO TENDERERS (GIT)

CONTENTS
Sl. Page
Topic
No. No.
A PREAMBLE
1 Definitions and Abbreviations 9
2 Introduction 10
3 Deleted --
4 Language of Tender 11
5 Eligible Tenderers 11
6 Eligible Goods and Services 11
7 Tendering Expense 11
B TENDER ENQUIRY DOCUMENTS
8 Contents of Tender Enquiry Documents 11
9 Deleted --
10 Clarification of Tender Enquiry Documents 12
C PREPARATION OF TENDERS
11 Documents Comprising the Tender 12
12 Tender Currencies 13
13 Tender Prices 13
14 Indian Agent 15
15 Firm Price / Variable Price 16
16 Deleted --
Documents Establishing Tenderer’s Eligibility and
17 16
Qualifications
Documents Establishing Good’s Conformity to Tender 16
18
Enquiry Document
19 Earnest Money Deposit (EMD) 17
20 Tender Validity 17
21 Signing and Sealing of Tender 17
D SUBMISSION OF TENDERS
22 Submission of Tenders 18
23 Late Tender 18
24 Alteration and Withdrawal of Tender 18

HLL/PCD/ESIC-20/10-11 Page 7 of 166 01.09.2010


HLL Lifecare Limited GIT

E TENDER OPENING
25 Opening of Tenders 19
F SCRUTINY AND EVALUATION OF TENDERS
26 Basic Principle 19
27 Preliminary Scrutiny of Tenders 19
28 Deleted --
29 Discrepancies in Prices 20
30 Discrepancy between original and copies of Tender 20
31 Qualification Criteria 20
32 Conversion of Tender Currencies to Indian Rupees 20
33 Deleted --
34 Comparison of Tenders 21
Additional Factors and Parameters for Evaluation and
35 21
Ranking of Responsive Tenders
36 Tenderer’s capability to perform the contract 21
37 Contacting the Purchaser 21
G AWARD OF CONTRACT
Purchaser’s Right to Accept any Tender and to Reject 21
38
any or All Tenders
39 Award Criteria 22
40 Variation of Quantities at the Time of Award 22
41 Notification of Award 22
42 Issue of Contract 22
Non-receipt of Performance Security and Contract by 22
43
the Purchaser/Consignee
44 Return of EMD 22
45 Publication of Tender Result 23
46 Corrupt or Fraudulent Practices 23

HLL/PCD/ESIC-20/10-11 Page 8 of 166 01.09.2010


HLL Lifecare Limited GIT

A. PREAMBLE
1. Definitions and Abbreviations

1.1 The following definitions and abbreviations, which have been used in these documents shall have
the meanings as indicated below:

1.2. Definitions:

(i) “Purchaser” means the organization purchasing goods and services as incorporated in
the Tender Enquiry document.
(ii) “Tender” means Bids / Quotation / Tender received from a Firm / Tenderer / Bidder.
(iii) “Tenderer” means Bidder/ the Individual or Firm submitting Bids / Quotation / Tender
(iii) “Supplier” means the individual or the firm supplying the goods and services as
incorporated in the contract.
(iv) “Goods” means the articles, material, commodities, livestock, furniture, fixtures, raw
material, spares, instruments, machinery, equipment, medical equipment, industrial plant
etc. which the supplier is required to supply to the purchaser under the contract.
(v) “Services” means services allied and incidental to the supply of goods, such as
transportation, installation, commissioning, provision of technical assistance, training,
after sales service, maintenance service and other such obligations of the supplier
covered under the contract.
(vi) “Earnest Money Deposit” (EMD) means Bid Security/ monetary or financial guarantee
to be furnished by a tenderer along with its tender.
(vii) “Contract” means the written agreement entered into between the purchaser and/or
consignee and the supplier, together with all the documents mentioned therein and
including all attachments, annexure etc. therein.
(viii) “Performance Security” means monetary or financial guarantee to be furnished by the
successful tenderer for due performance of the contract placed on it. Performance
Security is also known as Security Deposit.
(ix) “Consignee” means the Hospital/Dispensaries/Institute/Medical College/ person to
whom the goods are required to be delivered as specified in the Contract. If the goods
are required to be delivered to a person as an interim consignee for the purpose of
despatch to another person as provided in the Contract then that “another” person is the
consignee, also known as ultimate consignee.
(x) “Specification” means the document/standard that prescribes the requirement with
which goods or service has to conform.
(xi) “Inspection” means activities such as measuring, examining, testing, gauging one or
more characteristics of the product or service and comparing the same with the specified
requirement to determine conformity.
(xii) “Day” means calendar day.

1.3 Abbreviations:

(i) “T E Document” means Tender Enquiry Document


(ii) “NIT” means Notice Inviting Tenders.
(iii) “GIT” means General Instructions to Tenderers
(iv) “SIT” means Special Instructions to Tenderers
(v) “GCC” means General Conditions of Contract
(vi) “SCC” means Special Conditions of Contract

HLL/PCD/ESIC-20/10-11 Page 9 of 166 01.09.2010


HLL Lifecare Limited GIT

(vii) “DGS&D” means Directorate General of Supplies and Disposals


(viii) “NSIC” means National Small Industries Corporation
(ix) “PSU” means Public Sector Undertaking
(x) “CPSU” means Central Public Sector Undertaking
(xi) “LSI” means Large Scale Industry
(xii) “SSI” means Small Scale Industry
(xiii) “LC” means Letter of Credit
(xiv) “DP” means Delivery Period
(xv) “BG” means Bank Guarantee
(xvi) “ED” means Excise Duty
(xvii) “CD” means Custom Duty
(xviii) “VAT” means Value Added Tax
(xix) “CENVAT” means Central Value Added Tax
(xx) “CST” means Central Sales Tax
(xxi) “RR” means Railway Receipt
(xxii) “BL” means Bill of Lading
(xxiii) “FOB” means Free on Board
(xxiv) “FCA” means Free Carrier
(xxv) “FOR” means Free On Rail
(xxvi) “CIF” means Cost, Insurance and Freight
(xxvii) “CIP (Destinations)” means Carriage and Insurance Paid up to named port of
destination. Additionally the Insurance (local transportation and storage) would be
extended and borne by the Supplier from ware house to the consignee site for a period
including 3 months beyond date of delivery.
(xxviii) “DDP” means Delivery Duty Paid named place of destination (consignee site)
(xxix) “INCOTERMS” means International Commercial Terms as on the date of Tender
Opening
(xxx) ”ESIC” means Employee State Insurance Corporation.
(xxxi) “DG” means Director General of ESIC under the administrative control of Ministry of
Labour and Employment
(xxxii) “CMC” means Comprehensive Maintenance Contract (labour, spare and preventive
maintenance)
(xxxiii) “RT” means Re-Tender.

2. Introduction

2.1 The Purchaser has issued these TE documents for purchase of goods and related services as
mentioned in Section – VI – “List of Requirements”, which also indicates, interalia, the required
quantity, delivery schedule, terms and place of delivery.
2.2 This section (Section II - “General Instructions to Tenderers”) provides the relevant information
as well as instructions to assist the prospective tenderers in preparation and submission of
tenders. It also includes the mode and procedure to be adopted by the purchaser for receipt and
opening as well as scrutiny and evaluation of tenders and subsequent placement of contract.
2.3 The tenderers shall also read the Special Instructions to Tenderers (SIT) related to this purchase,
as contained in Section III of these documents and follow the same accordingly. Whenever there
is a conflict between the GIT and the SIT, the provisions contained in the SIT shall prevail over
those in the GIT.
2.4 Before formulating the tender and submitting the same to the purchaser, the tenderer should read
and examine all the terms, conditions, instructions, checklist etc. contained in the TE documents.
Failure to provide and/or comply with the required information, instructions etc. incorporated in
these TE documents may result in rejection of its tender.

HLL/PCD/ESIC-20/10-11 Page 10 of 166 01.09.2010


HLL Lifecare Limited GIT

3. Deleted
4. Language of Tender
4.1 The tender submitted by the tenderer and all subsequent correspondence and documents relating
to the tender exchanged between the tenderer and the purchaser, shall be written in the English
language, unless otherwise specified in the Tender Enquiry. However, the language of any
printed literature furnished by the tenderer in connection with its tender may be written in any
other language provided the same is accompanied by a notarised English translation and, for
purposes of interpretation of the tender, the English translation shall prevail.
4.2 The tender submitted by the tenderer and all subsequent correspondence and documents relating
to the tender exchanged between the tenderer and the purchaser, may also be written in the Hindi
language, provided that the same are accompanied by notarised English translation, in which
case, for purpose of interpretation of the tender etc, the English translations shall prevail.
5. Eligible Tenderers
This invitation for tenders is open to all suppliers who fulfil the eligibility criteria specified in
these documents.
6. Eligible Goods and Services
All goods and related services to be supplied under the contract shall have their origin in India or
any other country with which India has not banned trade relations. The term “origin” used in this
clause means the place where the goods are mined, grown, produced, or manufactured or from
where the related services are arranged and supplied.
7. Tendering Expense
The tenderer shall bear all costs and expenditure incurred and/or to be incurred by it in
connection with its tender including preparation, mailing and submission of its tender and for
subsequent processing the same. The purchaser will, in no case be responsible or liable for any
such cost, expenditure etc regardless of the conduct or outcome of the tendering process.

B. TENDER ENQUIRY DOCUMENTS


8. Content of Tender Enquiry Documents
8.1 In addition to Section I – “Notice inviting Tender” (NIT), the TE documents include:
Section II – General Instructions to Tenderers (GIT)
Section III – Special Instructions to Tenderers (SIT)
Section IV – General Conditions of Contract (GCC)
Section V – Special Conditions of Contract (SCC)
Section VI – List of Requirements
Section VII – Technical Specifications
Section VIII – Quality Control Requirements
Section IX – Qualification Criteria
Section X – Tender Form
Section XI – Price Schedules
Section XII – Questionnaire
Section XIII – Deleted
Section XIV – Manufacturer’s Authorisation Form
Section XV – Bank Guarantee Form for Performance Security/CMC Security
Section XVI – Contract Forms A & B
Section XVII – Proforma of Consignee Receipt Certificate

HLL/PCD/ESIC-20/10-11 Page 11 of 166 01.09.2010


HLL Lifecare Limited GIT

Section XVIII – Proforma of Final Acceptance Certificate by the consignee


Section XIX – Details of Shipping arrangement for Liner Cargoes in respect of
C&F/CIF/Turnkey/F.O.R. Contracts for Import
Section XX – Check List for the Tenderers
Section XXI – Consignee List
8.2 The relevant details of the required goods and services, the terms, conditions and procedure for
tendering, tender evaluation, placement of contract, the applicable contract terms and, also, the
standard formats to be used for this purpose are incorporated in the above-mentioned documents.
The interested tenderers are expected to examine all such details before submission of the tender.
9. Deleted
10. Clarification of TE documents
A tenderer requiring any clarification or elucidation on any issue of the TE documents may take
up the same with the purchaser in writing. The purchaser will respond to such request provided
the purchaser receives the same minimum 1 (one) hour prior to the scheduled time of pre-bid
meeting. Clarification for the same will be published on the websites as mentioned in para 5 of
NIT (Section I) within 4 days after the prebid meeting.

C. PREPARATION OF TENDERS
11. Documents Comprising the Tender
11.1 The Two Tender System, i.e. “Techno – Commercial Tender” and “Price Tender” prepared by
the tenderer shall comprise the following:
A) Techno – Commercial Tender (Un priced Tender)
i) Earnest money furnished in accordance with GIT clause 19.
ii) Tender Form as per Section X.
iii) Documentary evidence, as necessary in terms of clauses 5 and 17 establishing that the
tenderer is eligible to submit the tender and, also, qualified to perform the contract if its
tender is accepted.
iv) Tenderer/Agent who quotes for goods manufactured by other manufacturer shall furnish
Manufacturer’s Authorisation in the prescribed format as per Section - XIV.
v) Power of Attorney/Authorisation in favour of signatory of TE documents.
vi) Documents and relevant details to establish in accordance with GIT clause 18 that the
goods and the allied services to be supplied by the tenderer conform to the requirement of
the TE documents.
vii) Performance Statement as per section IX along with relevant copies of purchase orders
and end users’ satisfaction certificate.
viii) Price Schedule(s) as per Section XII filled up with all the details including Qty., Make,
Model, Contry of origin, etc. of the goods offered with prices blank (without indicating
any prices).
ix) Certificate of Incorporation of the bidder.
x) Checklist as per Section XX.
xi) Statement of deviations parameter wise from tendered technical specifications, if any.

B) Price Tender:
The information given at clause no. 11.1 A)- viii) above should be reproduced with the prices
indicated.

HLL/PCD/ESIC-20/10-11 Page 12 of 166 01.09.2010


HLL Lifecare Limited GIT

N.B.
1. All pages of the Tender should be page numbered and indexed.
2. It is the responsibility of tenderer to go through the TE document to ensure furnishing all
required documents in addition to above, if any.
11.2 The authorized signatory of the tenderer must sign the tender duly stamped at appropriate places
and initial all the remaining pages of the tender.
11.3 A tender, which does not fulfil any of the above requirements and/or gives evasive
information/reply against any such requirement, shall be liable to be ignored and rejected.
11.4 Tender sent by fax/telex/cable/electronically shall be ignored.
12. Tender currencies
12.1 The tenderer supplying indigenous goods or already imported goods shall quote only in Indian
Rupees.
12.2 For imported goods if supplied directly from abroad, prices shall be quoted in any freely
convertible currencies say USD, Euro, GBP or Yen. As regards price(s) for allied services, if any
required with the goods, the same shall be quoted in Indian Rupees only if such services are to be
performed /undertaken in India. Commission for Indian Agent, if any and if payable shall be
indicated in the space provided for in the price schedule and will be payable in Indian Rupees
only.
12.3 Tenders, where prices are quoted in any other way shall be treated as non-responsive and
rejected.
13 Tender Prices
13.1 The Tenderer shall indicate on the Price Schedule provided under Section XI all the specified
components of prices shown therein including the unit prices and total tender prices of the goods
and services it proposes to supply against the requirement. All the columns shown in the price
schedule should be filled up as required. If any column does not apply to a tenderer, it should be
clarified as “NA” by the tenderer.
13.2 The tenderer has the option to submit its quotation for any one or more item (s) in the List of
Requirements. However, separate sealed cover to be used for each item for price bid.

13.3 The quoted prices for goods offered from within India and that for goods offered from abroad are
to be indicated separately in the applicable Price Schedules attached under Section XI.

The price quoted by the tenderer for indigenous goods shall not be higher than the lowest price
charged for the goods of the same nature, class or description to an individual/ firm/ organisation
or department of Govt. of India.

For imported goods, the price quoted shall not be higher than the lowest price charged by the
tenderer for the goods of the same nature, class or description to a purchaser, domestic or foreign
or to any organisation or department of Govt. of India.

If it is found at any stage that the goods as stated have been supplied at a lower price, then that
price, with due allowance for elapsed time will be applicable to the present case and the
difference in cost would be refunded by the supplier to the purchaser, if the contract has already
been concluded.

13.4 While filling up the columns of the Price Schedule, the following aspects should be noted for
compliance:

HLL/PCD/ESIC-20/10-11 Page 13 of 166 01.09.2010


HLL Lifecare Limited GIT

13.4.1 For domestic goods or goods of foreign origin located within India, the prices in the
corresponding price schedule shall be entered separately in the following manner:
a) the price of the goods, quoted ex-factory/ ex-showroom/ ex-warehouse/ off-the-shelf, as
applicable, including all taxes and duties like sales tax, CST VAT, CENVAT, Custom Duty,
Excise Duty etc. already paid or payable on the components and raw material used in the
manufacture or assembly of the goods quoted ex-factory etc. or on the previously imported
goods of foreign origin quoted ex-showroom etc;
b) any sales or other taxes and any duties including excise duty, which will be payable on the
finished goods in India if the contract is awarded;
c) charges towards Packing & Forwarding, Inland Transportation, Insurance,
Loading/Unloading and other local costs incidental to delivery of the goods to their
final destination as specified in the List of Requirements and Price Schedule;
d) the price of Incidental Services, as mentioned in List of Requirements and Price Schedule;
e) the prices of Turnkey ( if any), as mentioned in List of Requirements, Technical Specification
and Price Schedule; and
f) the price of CMC, as mentioned in List of Requirements, Technical Specification and Price
Schedule.
13.4.2 For goods offered from abroad, the prices in the corresponding price schedule shall be
entered separately in the following manner:

a) the price of goods quoted FOB port of shipment, as indicated in the List of Requirements and
Price Schedule;
b) the amount of freight and insurance and price of goods quoted CIP port of entry in India as
indicated in the List of Requirements and Price Schedule;
c) the price of goods quoted should be on DDP basis at consignee site in India as indicated
in the List of Requirements, Price Schedule and Consignee List;
d) wherever applicable, the amount of custom duty with CDEC applicable on CIP value on the
goods to be imported; ESIC will issue Customs Duty Exemption Certificate (CDEC) where
applicable.
e) the charges for Loading/Unloading, Inland transportation, Insurance and other local costs,
Incidental cost to delivery of the goods from the port of entry in India to Consignee Site, as
specified in the List of Requirements and Price Schedule;
f) the charges for Incidental Services, as in the List of Requirements and Price Schedule;
g) the prices of Turnkey ( if any), as mentioned in List of Requirements, Technical Specification
and Price Schedule; and
h) the price of CMC, as mentioned in List of Requirements, Technical Specification and Price
Schedule.
13.5 Additional information and instruction on Duties and Taxes:

13.5.1 If the Tenderer desires to ask for excise duty, sales tax/ VAT, Service Tax, Works Contract Tax
etc. to be paid extra, the same must be specifically stated. In the absence of any such stipulation
the price will be taken inclusive of such duties and taxes and no claim for the same will be
entertained later.
13.5.2 Excise Duty:
a) If reimbursement of excise duty is intended as extra over the quoted prices, the supplier must
specifically say so also indicating the rate, quantum and nature of the duty applicable. In the
absence of any such stipulation it will be presumed that the prices quoted are firm and final
and no claim on account of excise duty will be entertained after the opening of tenders.

HLL/PCD/ESIC-20/10-11 Page 14 of 166 01.09.2010


HLL Lifecare Limited GIT

b) If a Tenderer chooses to quote a price inclusive of excise duty and also desires to be
reimbursed for variation, if any, in the excise duty during the time of supply, the tenderer
must clearly mention the same and also indicate the rate and quantum of excise duty included
in its price. Failure to indicate all such details in clear terms may result in rejection of that
tender.
c) Subject to sub clauses 13.5.2 (a) & (b) above, any change in excise duty upward/downward
as a result of any statutory variation in excise duty taking place within contract terms shall be
allowed to the extent of actual quantum of excise duty paid by the supplier. In case of
downward revision in excise duty, the actual quantum of reduction of excise duty shall be
reimbursed to the purchaser by the supplier. All such adjustments shall include all reliefs,
exemptions, rebates, concession etc. if any obtained by the supplier.
13.5.3 Sales Tax:
If a tenderer asks for sales tax/ VAT, Service Tax and Works Contract Tax to be paid extra, the
rate and nature of sales tax applicable should be shown separately. The sales tax / VAT, Service
Tax and Works Contract Tax will be paid as per the rate at which it is liable to be assessed or has
actually been assessed provided the transaction of sale is legally liable to sales tax / VAT,
Service Tax and Works Contract Tax and is payable as per the terms of the contract. If any
refund of Tax is received at a later date, the Supplier must return the amount forth-with to the
purchaser.

13.5.4 Octroi Duty and Local Duties & Taxes:


Octroi duty, terminal tax and other levies of local bodies (like town body, municipal body etc.) if
not exempted shall be reimbursed to the supplier on production of such proof of payment along
with the final bill.
13.5.5 Customs Duty:
The tenderer shall quote the Customs duty wherever applicable, considering availability of
CDEC. The applicable rates and amount of the Custom Duty and the corresponding Indian
Customs Tariff number should be shown separately in the price schedule. In case of non-
availibility of CDEC, the supplier has to custom cleared the goods paying the applicable Custom
Duty and any difference of duty from the applicable CDEC rate shall be reimbursed separately
on production of such proof of payment alongwith the final bill.

13.6 For transportation of imported goods offered from abroad, relevant instructions as incorporated
under GCC Clause 10 shall be followed.
13.7 For insurance of goods to be supplied, relevant instructions as provided under GCC Clause 11
shall be followed.
13.8 Unless otherwise specifically indicated in this TE document, the terms FCA, FOB, FAS, CIF,
CIP, DDP etc. for imported goods offered from abroad, shall be governed by the rules &
regulations prescribed in the current edition of INCOTERMS, published by the International
Chamber of Commerce, Paris
13.9 The need for indication of all such price components by the tenderers, as required in this clause
(viz., GIT clause 13) is for the purpose of comparison of the tenders by the purchaser and will no
way restrict the purchaser’s right to award the contract on the selected tenderer on any of the
terms offered.
14. Indian Agent
14.1 If a foreign tenderer has engaged an agent in India in connection with its tender, the foreign
tenderer, in addition to indicating Indian agent’s commission, if any, in a manner described under
GIT sub clause 12.2 above, shall also furnish the following information:

HLL/PCD/ESIC-20/10-11 Page 15 of 166 01.09.2010


HLL Lifecare Limited GIT

a) The complete name and address of the Indian Agent and its permanent income tax
account number as allotted by the Indian Income Tax authority.
b) The details of the services to be rendered by the agent for the subject requirement.
c) Details of Service outlets in India, nearest to the consignee(s), to render services during
Warranty and CMC period.

15. Firm Price


15.1 Unless otherwise specified in the SIT, prices quoted by the tenderer shall remain firm and fixed
during the currency of the contract and not subject to variation on any account.
15.2 However, as regards taxes and duties, if any, chargeable on the goods and payable, the conditions
stipulated in GIT clause 13 will apply.
16. Deleted

17 Documents Establishing Tenderer’s Eligibility and Qualifications


17.1 Pursuant to GIT clause 11, the tenderer shall furnish, as part of its tender, relevant details and
documents establishing its eligibility to quote and its qualifications to perform the contract if its
tender is accepted.
17.2 The documentary evidence needed to establish the tenderer’s qualifications shall fulfil the
following requirements:
a) in case the tenderer offers to supply goods, which are manufactured by some other firm, the
tenderer has been duly authorised by the goods manufacturer to quote for and supply the
goods to the purchaser. The tenderer shall submit the manufacturer’s authorization letter to
this effect as per the standard form provided under Section XIV in this document.
b) the tenderer has the required financial, technical and production capability necessary to
perform the contract and, further, it meets the qualification criteria incorporated in the
Section IX in these documents.
c) in case the tenderer is not doing business in India, it is duly represented by an agent stationed
in India fully equipped and able to carry out the required contractual functions and duties of
the supplier including after sale service, maintenance & repair etc. of the goods in question,
stocking of spare parts and fast moving components and other obligations, if any, specified in
the conditions of contract and/or technical specifications.
d) in case the tenderer is an Indian agent/authorized representative quoting on behalf of a
foreign manufacturer for the restricted item, the Indian agent/authorized representative is
already enlisted under the Compulsory Enlistment Scheme of Ministry of Finance, Govt. of
India, operated through Directorate General of Supplies & Disposals (DGS&D), New Delhi.

18. Documents establishing Good’s Conformity to TE document.


18.1 The tenderer shall provide in its tender the required as well as the relevant documents like
technical data, literature, drawings etc. to establish that the goods and services offered in the
tender fully conform to the goods and services specified by the purchaser in the TE documents.
For this purpose the tenderer shall also provide a clause-by-clause commentary on the technical
specifications and other technical details incorporated by the purchaser in the TE documents to
establish technical responsiveness of the goods and services offered in its tender.
18.2 In case there is any variation and/or deviation between the goods & services prescribed by the
purchaser and that offered by the tenderer, the tenderer shall list out the same in a chart form
without ambiguity and provide the same along with its tender.

HLL/PCD/ESIC-20/10-11 Page 16 of 166 01.09.2010


HLL Lifecare Limited GIT

18.3 If a tenderer furnishes wrong and/or misguiding data, statement(s) etc. about technical
acceptability of the goods and services offered by it, its tender will be liable to be ignored and
rejected in addition to other remedies available to the purchaser in this regard.

19. Earnest Money Deposit (EMD)

19.1 Pursuant to GIT clauses 8.1 and 11.1(A) the tenderer shall furnish along with its tender, earnest
money for amount as shown in the List of Requirements. The earnest money is required to
protect the purchaser against the risk of the tenderer’s unwarranted conduct as amplified under
sub-clause 19.7 below.
19.2 Deleted
19.3 The earnest money shall be denominated in Indian Rupees or equivalent currencies as per GIT
clause 12.2. The earnest money shall not be accepted in any other form except the following:
i. Account Payee Demand Draft or
ii. Banker’s cheque
19.4 The demand draft or banker’s cheque shall be drawn on any commercial bank in India or country
of the tenderer, in favour of the “HLL Lifecare Limited” payable at New Delhi.
19.5 Deleted.

19.6 Unsuccessful tenderers’ earnest money will be returned to them without any interest, after expiry
of the tender validity period, but not later than thirty days after conclusion of the resultant
contract. Successful tenderer’s earnest money will be returned without any interest, after receipt
of performance security from that tenderer.

19.7 Earnest Money is required to protect the purchaser against the risk of the Tenderer’s conduct,
which would warrant the forfeiture of the EMD. Earnest money of a tenderer will be forfeited, if
the tenderer withdraws or amends its tender or impairs or derogates from the tender in any
respect within the period of validity of its tender or if it comes to notice that the
information/documents furnished in its tender is incorrect, false, misleading or forged without
prejudice to other rights of the purchaser. The successful tenderer’s earnest money will be
forfeited without prejudice to other rights of Purchaser if it fails to furnish the required
performance security within the specified period.
20. Tender Validity
20.1 If not mentioned otherwise in the SIT, the tenders shall remain valid for acceptance for a period
of 120 days (One hundred and twenty days) after the date of opening of techno-commercial
tenders prescribed in the TE document. Any tender valid for a shorter period shall be treated as
unresponsive and rejected.
20.2 In exceptional cases, the tenderers may be requested by the purchaser to extend the validity of
their tenders up to a specified period. Such request(s) and responses thereto shall be conveyed by
surface mail or by fax/ email followed by surface mail. The tenderers, who agree to extend the
tender validity, are to extend the same without any change or modification of their original tender
and they are also to extend the validity period of the EMD accordingly. A tenderer, however,
may not agree to extend its tender validity without forfeiting its EMD.
20.3 In case the day up to which the tenders are to remain valid falls on/ subsequently declared a
holiday or closed day for the purchaser, the tender validity shall automatically be extended up to
the next working day.
21. Signing and Sealing of Tender
21.1 The tenderers shall submit their tenders as per the instructions contained in GIT Clause 11.

HLL/PCD/ESIC-20/10-11 Page 17 of 166 01.09.2010


HLL Lifecare Limited GIT

21.2 Unless otherwise mentioned in the SIT, a tenderer shall submit two copies of its tender marking
them as “Original” and “Duplicate”. Duplicate tenders may contain all pages including Technical
Literature/Catalogues as in Original tenders.
21.3 The original and duplicate copies of the tender shall either be typed or written in indelible ink
and the same shall be signed by the tenderer or by a person(s) who has been duly authorized to
bind the tenderer to the contract. The letter of authorization shall be by a written power of
attorney, which shall also be furnished along with the tender.
21.4 All the copies of the tender shall be duly signed at the appropriate places as indicated in the TE
documents and all other pages of the tender including printed literature, if any shall be initialled
by the same person(s) signing the tender. The tender shall not contain any erasure or overwriting,
except as necessary to correct any error made by the tenderer and, if there is any such correction;
the same shall be initialled by the person(s) signing the tender.
21.5 The tenderer is to seal the original and copy of the tender in separate envelopes, duly marking the
same as “Original”, “Duplicate” and so on and writing the address of the purchaser and the
tender reference number on the envelopes. The sentence “NOT TO BE OPENED” before
__________ (The tenderer is to put the date & time of tender opening) are to be written on these
envelopes. The inner envelopes are then to be put in a bigger outer envelope along with envelope
containing EMD, which will also be duly sealed, marked etc. as above. If the outer envelope is
not sealed and marked properly as above, the purchaser will not assume any responsibility for its
misplacement, premature opening, late opening etc.
21.6 TE document seeks quotation following two Tender System, in two parts. First part will be
known as ‘Techno - Commercial Tender’, and the second part ‘Price Tender’ as specified in
clause 11 of GIT. Tenderer shall seal ‘Techno - Commercial Tender (along with envelope
containing EMD)’ and ‘Price Tender’ separately and covers will be suitably super scribed.
Both these sealed covers shall be put in a bigger cover and sealed and procedure prescribed in
Paras 21.1 to 21.5 followed.

D. SUBMISSION OF TENDERS
22. Submission of Tenders
22.1 Unless otherwise specified, the tenderers are to deposit the tenders in the tender box kept for this
purpose at HLL Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector -
62, Noida -201307, Uttar Pradesh. In case of bulky tender, which can not be put into tender
box, the same shall be submitted by the tenderer by hand to Head (P&CD) or his nominee, HLL
Lifecare Limited, Procurement and Consultancy Division, B-14A, Sector -62, Noida -
201307, Uttar Pradesh. The officer receiving the tender will give the tenderer an official receipt
duly signed with date and time.
22.2 The tenderers must ensure that they deposit their tenders not later than the closing time and date
specified for submission of tenders. It is the responsibility of the tenderer to ensure that their
Tenders whether sent by post or by courier or by person, are dropped in the Tender Box by the
specified clearing date and time. In the event of the specified date for submission of tender falls
on / is subsequently declared a holiday or closed day for the purchaser, the tenders will be
received up to the appointed time on the next working day.
23. Late Tender
23.1 A tender, which is received after the specified date and time for receipt of tenders will be treated
as “late” tender and will be ignored.

24. Alteration and Withdrawal of Tender


24.1 The tenderer, after submitting its tender, is permitted to alter / modify its tender so long as such
alterations / modifications are received duly signed, sealed and marked like the original tender,

HLL/PCD/ESIC-20/10-11 Page 18 of 166 01.09.2010


HLL Lifecare Limited GIT

within the deadline for submission of tenders. Alterations / modifications to tenders received
after the prescribed deadline will not be considered.

24.2 No tender should be withdrawn after the deadline for submission of tender and before expiry of
the tender validity period. If a tenderer withdraws the tender during this period, it will result in
forfeiture of the earnest money furnished by the tenderer in its tender.

E. TENDER OPENING
25. Opening of Tenders
25.1 The purchaser will open the tenders at the specified date and time and at the specified place as
indicated in the NIT.

In case the specified date of tender opening falls on / is subsequently declared a holiday or closed
day for the purchaser, the tenders will be opened at the appointed time and place on the next
working day.

25.2 Authorized representatives of the tenderers, who have submitted tenders on time may attend
the tender opening provided they bring with them letters of authority from the corresponding
tenderers.
The tender opening official(s) will prepare a list of the representatives attending the tender
opening. The list will contain the representatives’ names & signatures and corresponding
tenderers’ names and addresses.

25.3 Two-Tender system as mentioned in para 21.6 above will be as follows. The Techno-
Commercial Tenders are to be opened in the first instance, at the prescribed time and date as
indicated in NIT. These Tenders shall be scrutinized and evaluated by the competent committee/
authority with reference to parameters prescribed in the TE document. During the Techno -
Commercial Tender opening, the tender opening official(s) will read the salient features of the
tenders like brief description of the goods offered, delivery period, Earnest Money Deposit and
any other special features of the tenders, as deemed fit by the tender opening official(s).
Thereafter, in the second stage, the Price Tenders of only the Techno-Commercially acceptable
offers shall be opened at a latter date which will be notified to such tenderers. The prices, special
discount if any of the goods offered etc., as deemed fit by tender opening official(s) will be read
out.

F. SCRUTINY AND EVALUATION OF TENDERS


26. Basic Principle
26.1 Tenders will be evaluated on the basis of the terms & conditions already incorporated in the TE
document, based on which tenders have been received and the terms, conditions etc. mentioned
by the tenderers in their tenders. No new condition will be brought in while scrutinizing and
evaluating the tenders.
27. Preliminary Scrutiny of Tenders
27.1 The Purchaser will examine the Tenders to determine whether they are complete, whether any
computational errors have been made, whether required sureties have been furnished, whether
the documents have been properly signed stamped and whether the Tenders are generally in
order.
27.2 Deleted.

HLL/PCD/ESIC-20/10-11 Page 19 of 166 01.09.2010


HLL Lifecare Limited GIT

27.3 Deleted
27.4 The tenders will be scrutinized to determine whether they are complete and meet the essential
and important requirements, conditions etc. as prescribed in the TE document. The tenders,
which do not meet the basic requirements, are liable to be treated as non – responsive and will be
summarily ignored.
27.5 The following are some of the important aspects, for which a tender shall be declared non-
responsive and will be summarily ignored;
(i) Tender form as per Section X (signed and stamped) not enclosed
(ii) Tender is unsigned.
(iii) Tender validity is shorter than the required period.
(iv) Required EMD have not been provided.
(v) Tenderer has quoted for goods manufactured by other manufacturer(s) without the required
Manufacturer’s Authorisation as per Format given in Section XIV.
(vi) Tenderer has not agreed to give the required performance security.
(vii) Goods offered are not meeting the tender enquiry specification.
(viii) Tenderer has not agreed to other essential condition(s) specially incorporated in the tender
enquiry like terms of payment, liquidated damages clause, warranty clause, DDP clause,
Delivery period clause, dispute resolution mechanism applicable law.
(ix) Poor/ unsatisfactory past performance.
(x) Tenderers who stand deregistered/banned/blacklisted by any Govt. Authorities.
(xi) Tenderer is not eligible as per GIT Clauses 5.1 & 17.1.
(xii) Tenderer has not quoted for the entire quantity as specified in the List of Requirements for
the quoted item (s).
28. Deleted
29 Discrepancies in Prices
29.1 If, in the price structure quoted by a tenderer, there is discrepancy between the unit price and the
total price (which is obtained by multiplying the unit price by the quantity), the unit price shall
prevail and the total price corrected accordingly, unless the purchaser feels that the tenderer has
made a mistake in placing the decimal point in the unit price, in which case the total price as
quoted shall prevail over the unit price and the unit price corrected accordingly.
29.2 If there is an error in a total price, which has been worked out through addition and/or subtraction
of subtotals, the subtotals shall prevail and the total corrected; and
29.3 If there is a discrepancy between the amount expressed in words and figures, the amount in
words shall prevail, subject to sub clause 29.1 and 29.2 above.
30. Discrepancy between original and copies of Tender
In case any discrepancy is observed between the text etc. of the original copy and that in the
other copies of the same tender set, the text etc. of the original copy shall prevail.

31. Qualification Criteria


Tenders of the tenderers, who do not meet the required Qualification Criteria prescribed in
Section IX, will be treated as non-responsive and will not be considered further.
32. Conversion of tender currencies to Indian Rupees
In case the TE document permits the tenderers to quote their prices in different currencies, all
such quoted prices of the responsive tenderers will be converted to a single currency viz., Indian
Rupees for the purpose of equitable comparison and evaluation, as per the exchange rates
established by the Reserve Bank of India for similar transactions, as on the date of ‘Price
Tender’ opening.

HLL/PCD/ESIC-20/10-11 Page 20 of 166 01.09.2010


HLL Lifecare Limited GIT

33. Deleted
34. Comparison of Tenders
34.1 Unless mentioned otherwise in Section – III – Special Instructions to Tenderers and Section – VI
– List of Requirements, the comparison of the responsive tenders shall be carried out on Delivery
Duty Paid (DDP) consignee site basis. The quoted turnkey prices and CMC prices will also be
added for comparison/ranking purpose for evaluation.
35. Additional Factors and Parameters for Evaluation and Ranking of Responsive Tenders
35.1 Further to GIT Clause 34 above, the purchaser’s evaluation of a tender will include and take into
account the following:
i) In the case of goods manufactured in India or goods of foreign origin already located in India,
sales tax & other similar taxes and excise duty & other similar duties, Customs Duties, Service
Tax, Works Contract Tax etc which will be contractually payable (to the tenderer), on the
goods if a contract is awarded on the tenderer; and
ii) in the case of goods of foreign origin offered from abroad, customs duty and other similar
import duties/taxes, which will be contractually payable (to the tenderer) on the goods if the
contract is awarded on the tenderer.
35.2 The purchaser’s evaluation of tender will also take into account the additional factors, if any,
incorporated in SIT in the manner and to the extent indicated therein.
35.3 Deleted
36. Tenderer’s capability to perform the contract
36.1 The purchaser, through the above process of tender scrutiny and tender evaluation will determine
to its satisfaction whether the tenderer, whose tender has been determined as the lowest evaluated
responsive tender is eligible, qualified and capable in all respects to perform the contract
satisfactorily. If, there is more than one item in the List of Requirements, then, such
determination will be made separately for each item.
36.2 The above-mentioned determination will, interalia, take into account the tenderer’s financial,
technical and production capabilities for satisfying all the requirements of the purchaser as
incorporated in the TE document. Such determination will be based upon scrutiny and
examination of all relevant data and details submitted by the tenderer in its tender as well as such
other allied information as deemed appropriate by the purchaser.
37. Contacting the Purchaser
37.1 From the time of submission of tender to the time of awarding the contract, if a tenderer needs to
contact the purchaser for any reason relating to this tender enquiry and / or its tender, it should do
so only in writing.

37.2 In case a tenderer attempts to influence the purchaser in the purchaser’s decision on scrutiny,
comparison & evaluation of tenders and awarding the contract, the tender of the tenderer shall be
liable for rejection in addition to appropriate administrative actions being taken against that
tenderer, as deemed fit by the purchaser.

G. AWARD OF CONTRACT
38. Purchaser’s Right to accept any tender and to reject any or all tenders
The purchaser reserves the right to accept in part or in full any tender or reject any or more
tender(s) without assigning any reason or to cancel the tendering process and reject all tenders at

HLL/PCD/ESIC-20/10-11 Page 21 of 166 01.09.2010


HLL Lifecare Limited GIT

any time prior to award of contract, without incurring any liability, whatsoever to the affected
tenderer or tenderers.

39. Award Criteria


Subject to GIT clause 38 above, the contract will be awarded to the lowest evaluated responsive
tenderer decided by the purchaser in terms of GIT Clause 36.
40. Variation of Quantities at the Time of Award/ Currency of Contract
40.1 At the time of awarding the contract, the purchaser reserves the right to increase or decrease by
up to twenty five (25) per cent, the quantity of goods and services mentioned in the schedule (s)
in the “List of Requirements” without any change in the unit price and other terms & conditions
quoted by the tenderer.
40.2 If the quantity has not been increased at the time of the awarding the contract, the purchaser
reserves the right to increase by up to twenty five (25) per cent, the quantity of goods and
services mentioned in the contract without any change in the unit price and other terms &
conditions mentioned in the contract, during the currency of the contract or within one-year from
the date of Notification of Award..

41. Notification of Award

41.1 Before expiry of the tender validity period, the purchaser will notify the successful tenderer(s) in
writing, by registered / speed post or by fax/ telex/cable (to be confirmed by registered / speed
post) that its tender for goods & services, which have been selected by the purchaser, has been
accepted, also briefly indicating therein the essential details like description, specification and
quantity of the goods & services and corresponding prices accepted. The successful tenderer
must furnish to the purchaser the required performance security within twenty-one (21) days
from the date of dispatch of this notification, failing which the EMD will be forfeited and the
award will be cancelled. Relevant details about the performance security have been provided
under GCC Clause 5 under Section IV.

41.2 The Notification of Award shall constitute the conclusion of the Contract.

42. Issue of Contract


42.1 Promptly after notification of award, the Purchaser/Consignee will mail the contract form (as per
Section XVI) duly completed and signed, in duplicate, to the successful tenderer by registered /
speed post.
42.2 Within twenty one days from the date of the contract, the successful tenderer shall return the
original copy of the contract, duly signed and dated, to the Purchaser by registered / speed post.
42.3 The Purchaser reserves the right to issue the Notification of Award consignee wise.
43. Non-receipt of Performance Security and Contract by the Purchaser/Consignee
Failure of the successful tenderer in providing performance security and / or returning contract
copy duly signed in terms of GIT clauses 41 and 42 above shall make the tenderer liable for
forfeiture of its EMD and, also, for further actions by the Purchaser/Consignee against it as per
the clause 24 of GCC – Termination of default.

44. Return of E M D
The earnest money of the successful tenderer and the unsuccessful tenderers will be returned to
them without any interest, whatsoever, in terms of GIT Clause 19.6.

HLL/PCD/ESIC-20/10-11 Page 22 of 166 01.09.2010


HLL Lifecare Limited GIT

45. Publication of Tender Result


The name and address of the successful tenderer(s) receiving the contract(s) will be mentioned in
the notice board/bulletin/web site of the purchaser.

46. Corrupt or Fraudulent Practices


It is required by all concerned to observe the highest standard of ethics during the procurement
and execution of such contracts. In pursuance of this policy, the Purchaser: -
(a) defines, for the purposes of this provision, the terms set forth below
as follows:
(i) “corrupt practice” means the offering, giving, receiving or soliciting of any thing of
value to influence the action of a public official in the procurement process or in
contract execution; and
(ii) “fraudulent practice” means a misrepresentation of facts in order to influence a
procurement process or the execution of a contract to the detriment of the Purchaser, and
includes collusive practice among Tenderers (prior to or after Tender submission)
designed to establish Tender prices at artificial non-competitive levels and to deprive the
Purchaser of the benefits of free and open competition;
(b) will reject a proposal for award if it determines that the Tenderer recommended for award has
engaged in corrupt or fraudulent practices in competing for the contract in question;
(c) will declare a firm ineligible, either indefinitely or for a stated period of time, to be awarded a
contract by the purchaser if it at any time determines that the firm has engaged in corrupt or
fraudulent practices in competing for, or in executing the contract.

HLL/PCD/ESIC-20/10-11 Page 23 of 166 01.09.2010


HLL Lifecare Limited SIT

SECTION - III
SPECIAL INSTRUCTIONS TO TENDERERS
(SIT)

The following Special Instructions to Tenderers will apply for this purchase. These special instructions
will modify/substitute/supplement the corresponding General Instructions to Tenderers (GIT)
incorporated in Section II. The corresponding GIT clause numbers have also been indicated in the text
below:
In case of any conflict between the provision in the GIT and that in the SIT, the provision contained in the
SIT shall prevail.

HLL/PCD/ESIC-20/10-11 Page 24 of 166 01.09.2010


HLL Lifecare Limited GCC

SECTION - IV

GENERAL CONDITIONS OF CONTRACT (GCC)

TABLE OF CLAUSES
Sl No. Topic Page
1 Application 26
2 Use of contract documents and information 26
3 Patent Rights 26
4 Country of Origin 26
5 Performance Security 26
6 Technical Specifications and Standards 27
7 Packing and Marking 27
8 Inspection, Testing and Quality Control 28
9 Terms of Delivery 28
10 Transportation of Goods 28
11 Insurance 29
12 Spare parts 29
13 Incidental services 30
14 Distribution of Dispatch Documents for Clearance/Receipt of Goods 30
15 Warranty 31
16 Assignment 32
17 Sub Contracts 32
18 Modification of contract 32
19 Prices 33
20 Taxes and Duties 33
21 Terms and mode of Payment 33
22 Delay in the supplier’s performance 35
23 Liquidated Damages 36
24 Termination for default 36
25 Termination for insolvency 37
26 Force Majeure 37
27 Termination for convenience 37
28 Governing language 38
29 Notices 38
30 Resolution of disputes 38
31 Applicable Law 38
32 General/Miscellaneous Clauses 38

HLL/PCD/ESIC-20/10-11 Page 25 of 166 01.09.2010


HLL Lifecare Limited GCC

GENERAL CONDITIONS OF CONTRACT (GCC)


1. Application
The General Conditions of Contract incorporated in this section shall be applicable for this
purchase to the extent the same are not superseded by the Special Conditions of Contract
prescribed under Section V, List of requirements under Section VI and Technical Specification
under Section VII of this document.
2. Use of contract documents and information
2.1 The supplier shall not, without the purchaser’s prior written consent, disclose the contract or any
provision thereof including any specification, drawing, sample or any information furnished by or
on behalf of the purchaser in connection therewith, to any person other than the person(s)
employed by the supplier in the performance of the contract emanating from this TE document.
Further, any such disclosure to any such employed person shall be made in confidence and only so
far as necessary for the purposes of such performance for this contract.
2.2 Further, the supplier shall not, without the purchaser’s prior written consent, make use of any
document or information mentioned in GCC sub-clause 2.1 above except for the sole purpose of
performing this contract.
2.3 Except the contract issued to the supplier, each and every other document mentioned in GCC sub-
clause 2.1 above shall remain the property of the purchaser and, if advised by the purchaser, all
copies of all such documents shall be returned to the purchaser on completion of the supplier’s
performance and obligations under this contract.
3. Patent Rights
The supplier shall, at all times, indemnify and keep indemnified the purchaser, free of cost,
against all claims which may arise in respect of goods & services to be provided by the supplier
under the contract for infringement of any intellectual property rights or any other right protected
by patent, registration of designs or trademarks. In the event of any such claim in respect of
alleged breach of patent, registered designs, trade marks etc. being made against the purchaser, the
purchaser shall notify the supplier of the same and the supplier shall, at his own expenses take
care of the same for settlement without any liability to the purchaser.
4. Country of Origin
4.1 All goods and services to be supplied and provided for the contract shall have the origin in India
or in the countries with which the Government of India has trade relations.
4.2 The word “origin” incorporated in this clause means the place from where the goods are mined,
cultivated, grown, manufactured, produced or processed or from where the services are arranged.
4.3 The country of origin may be specified in the Price Schedule
5. Performance Security
5.1 Within twenty-one (21) days from date of the issue of notification of award by the
Purchaser/Consignee, the supplier, shall furnish performance security to the Purchaser/Consignee
for an amount equal to ten percent (10%) of the total value of the contract, valid up to sixty (60)
days after the date of completion of all contractual obligations by the supplier, including the
warranty obligations.
5.2 The Performance security shall be denominated in Indian Rupees or in the currency of the contract
as detailed below:
a) It shall be in any one of the forms namely Account Payee Demand Draft drawn from any
Scheduled bank in India or Bank Guarantee issued by a Scheduled bank in India, in the

HLL/PCD/ESIC-20/10-11 Page 26 of 166 01.09.2010


HLL Lifecare Limited GCC

prescribed form as provided in section XV of this document in favour of the


Purchaser/Consignee.
b) In the case of Bank Guarantee furnished from banks outside India (i.e. foreign Banks), it
should be authenticated and countersigned by any nationalised bank in India by way of
back-to-back counter guarantee.
5.3 In the event of any failure /default of the supplier with or without any quantifiable loss to the
consignee/purchaser including furnishing of consignee wise Bank Guarantee for CMC security as
per Proforma in Section XV, the amount of the performance security is liable to be forfeited. The
Administration Department may do the needful to cover any failure/default of the supplier with or
without any quantifiable loss to the Government.
5.4 In the event of any amendment issued to the contract, the supplier shall, within twenty-one (21)
days of issue of the amendment, furnish the corresponding amendment to the Performance
Security (as necessary), rendering the same valid in all respects in terms of the contract, as
amended.
5.5 The supplier shall enter into Comprehensive Maintenance Contract as per the ‘Contract Form - B’
in Section XVI with respective consignees, 3 (three) months prior to the completion of Warranty
Period. The CMC will commence from the date of expiry of the Warranty Period.
5.6 Subject to GCC sub – clause 5.3 above, the Purchaser/Consignee will release the Performance
Security without any interest to the supplier on completion of the supplier’s all contractual
obligations including the warranty obligations & after receipt of Consignee wise bank guarantee
for CMC security in favour of Head of the Hospital/ Institute/ Medical College of the consignee as
per the format in Section XV.
6. Technical Specifications and Standards
6.1 The Goods & Services to be provided by the supplier under this contract shall conform to the
technical specifications and quality control parameters mentioned in ‘Technical Specification’ and
‘Quality Control Requirements’ under Sections VII and VIII of this document.
7. Packing and Marking
7.1 The packing for the goods to be provided by the supplier should be strong and durable enough to
withstand, without limitation, the entire journey during transit including transhipment (if any),
rough handling, open storage etc. without any damage, deterioration etc. As and if necessary, the
size, weights and volumes of the packing cases shall also take into consideration, the remoteness
of the final destination of the goods and availability or otherwise of transport and handling
facilities at all points during transit up to final destination as per the contract.
7.2 The quality of packing, the manner of marking within & outside the packages and provision of
accompanying documentation shall strictly comply with the requirements as provided in
Technical Specifications and Quality Control Requirements under Sections VII and VIII and in
SCC under Section V. In case the packing requirements are amended due to issue of any
amendment to the contract, the same shall also be taken care of by the supplier accordingly.
7.3 Packing instructions:
Unless otherwise mentioned in the Technical Specification and Quality Control Requirements
under Sections VII and VIII and in SCC under Section V, the supplier shall make separate
packages for each consignee (in case there is more than one consignee mentioned in the contract)
and mark each package on three sides with the following with indelible paint of proper quality:
a. contract number and date
b. brief description of goods including quantity
c. packing list reference number
d. country of origin of goods

HLL/PCD/ESIC-20/10-11 Page 27 of 166 01.09.2010


HLL Lifecare Limited GCC

e. consignee’s name and full address and


f. supplier’s name and address
8. Inspection, Testing and Quality Control
8.1 The purchaser and/or its nominated representative(s) will, without any extra cost to the purchaser,
inspect and/or test the ordered goods and the related services to confirm their conformity to the
contract specifications and other quality control details incorporated in the contract. The purchaser
shall inform the supplier in advance, in writing, the purchaser’s programme for such inspection
and, also the identity of the officials to be deputed for this purpose. The cost towards the
transportation, boarding & lodging will be borne by the purchaser and/or its nominated
representative(s).
8.2 The Technical Specification and Quality Control Requirements incorporated in the contract shall
specify what inspections and tests are to be carried out and, also, where and how they are to be
conducted. If such inspections and tests are conducted in the premises of the supplier or its
subcontractor(s), all reasonable facilities and assistance, including access to relevant drawings,
design details and production data, shall be furnished by the supplier to the purchaser’s inspector
at no charge to the purchaser.
8.3 If during such inspections and tests the contracted goods fail to conform to the required
specifications and standards, the purchaser’s inspector may reject them and the supplier shall
either replace the rejected goods or make all alterations necessary to meet the specifications and
standards, as required, free of cost to the purchaser and resubmit the same to the purchaser’s
inspector for conducting the inspections and tests again.
8.4 In case the contract stipulates pre-despatch inspection of the ordered goods at supplier’s premises,
the supplier shall put up the goods for such inspection to the purchaser’s inspector well ahead of
the contractual delivery period, so that the purchaser’s inspector is able to complete the inspection
within the contractual delivery period.
8.5 If the supplier tenders the goods to the purchaser’s inspector for inspection at the last moment
without providing reasonable time to the inspector for completing the inspection within the
contractual delivery period, the inspector may carry out the inspection and complete the formality
beyond the contractual delivery period at the risk and expense of the supplier. The fact that the
goods have been inspected after the contractual delivery period will not have the effect of keeping
the contract alive and this will be without any prejudice to the legal rights and remedies available
to the purchaser under the terms & conditions of the contract.
8.6 The purchaser’s/consignee’s contractual right to inspect, test and, if necessary, reject the goods
after the goods’ arrival at the final destination shall have no bearing of the fact that the goods have
previously been inspected and cleared by purchaser’s inspector during pre-despatch inspection
mentioned above.
8.7 Goods accepted by the purchaser/consignee and/or its inspector at initial inspection and in final
inspection in terms of the contract shall in no way dilute purchaser’s/consignee’s right to reject the
same later, if found deficient in terms of the warranty clause of the contract, as incorporated under
GCC Clause 15.
8.8 Principal/ Foreign supplier shall also have the equipment inspected by recognised/ reputed agency
like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch at the supplier’s cost
and furnish necessary certificate from the said agency in support of their claim.
9. Terms of Delivery
Goods shall be delivered by the supplier in accordance with the terms of delivery specified in the
contract.
10. Transportation of Goods

HLL/PCD/ESIC-20/10-11 Page 28 of 166 01.09.2010


HLL Lifecare Limited GCC

10.1 Instructions for transportation of imported goods offered from abroad:


The supplier shall not arrange part-shipments and/or transhipment without the express/prior
written consent of the purchaser. The supplier is required under the contract to deliver the goods
under DDP at consignee site; the shipment shall be made by Indian flag vessel or by vessels
belonging to the conference lines in which India is a member country through India’s forwarding
agents/coordinators. In case the forwarding agent/coordinators are unable to provide timely
adequate space in Indian flag vessel or by vessels belonging to the conference lines, the supplier
shall arrange shipment through any available vessel to adhere to the delivery schedule given in the
contract.

In case of airlifting of imported goods offered from abroad, the same will be done only through
the National Carrier i.e. Air India wherever applicable. In case the National Carrier is not
available, any other airlines available for early delivery may be arranged.

Goods will be custom cleared by the supplier/ Indian agent and transported to the consignee’s site
as per the contract terms.
Instructions for transportation of domestic goods including goods already imported by the supplier
under its own arrangement:

In case no instruction is provided in this regard in the SCC, the supplier will arrange
transportation of the ordered goods as per its own procedure.

The goods shall be custom cleared by the Supplier/ Indian Agent and transported to the
consignee’s site as per contract terms. The supplier shall be responsible for safe and timely
delivery of ordered goods under his own arrangement.
11. Insurance:
Unless otherwise instructed in the SCC, the supplier shall make arrangements for insuring the
goods against loss or damage incidental to manufacture or acquisition, transportation, storage and
delivery in the following manner:
i) in case of supply of domestic goods on Consignee site basis, the supplier shall be responsible
till the entire stores contracted for arrival in good condition at destination. The transit risk in
this respect shall be covered by the Supplier by getting the stores duly insured. The insurance
cover shall be obtained by the Supplier and should be valid till 3 months after the receipt of
goods by the Consignee.
ii) in case of supply of the imported goods on DDP Basis, the supplier shall arrange and pay for
marine/ air insurance making the consignee as beneficiary. The additional extended Insurance
(local transportation and storage) would also be borne by the Supplier from the port of entry to
the consignee site for a period including 3 months beyond date of delivery.

12. Spare parts


12.1 If specified in the List of Requirements and in the resultant contract, the supplier shall
supply/provide any or all of the following materials, information etc. pertaining to spare parts
manufactured and/or supplied by the supplier:
a) The spare parts as selected by the Purchaser/Consignee to be purchased from the supplier,
subject to the condition that such purchase of the spare parts shall not relieve the supplier
of any contractual obligation including warranty obligations; and

HLL/PCD/ESIC-20/10-11 Page 29 of 166 01.09.2010


HLL Lifecare Limited GCC

b) In case the production of the spare parts is discontinued:


i) Sufficient advance notice to the Purchaser/Consignee before such discontinuation
to provide adequate time to the purchaser to purchase the required spare parts etc.,
and

ii) Immediately following such discontinuation, providing the Purchaser/Consignee,


free of cost, the designs, drawings, layouts and specifications of the spare parts, as
and if requested by the Purchaser/Consignee.
12.2 Supplier shall carry sufficient inventories to assure ex-stock supply of consumable spares for the
goods so that the same are supplied to the Purchaser/Consignee promptly on receipt of order from
the Purchaser/Consignee.
13. Incidental services
Subject to the stipulation, if any, in the SCC (Section – V), List of Requirements (Section – VI)
and the Technical Specification (Section – VII), the supplier shall be required to perform the
following services.
i) Installation & commissioning, Supervision and Demonstration of the goods
ii) Providing required jigs and tools for assembly, minor civil works required for the
completion of the installation.
iii) Training of Consignee’s Doctors, Staff, operators etc. for operating and maintaining
the goods
iv) Supplying required number of operation & maintenance manual for the goods

14. Distribution of Dispatch Documents for Clearance/Receipt of Goods


The supplier shall send all the relevant despatch documents well in time to the
Purchaser/Consignee

Unless otherwise specified in the SCC, the usual documents involved and the drill to be followed
in general for this purpose are as follows.

A) For Domestic Goods, including goods already imported by the supplier under its own
arrangement

Within 24 hours of despatch, the supplier shall notify the purchaser, consignee, and others
concerned if mentioned in the contract, the complete details of despatch and also supply the
following documents to them by registered post / speed post (or as instructed in the contract):
(i) Four copies of supplier’s invoice showing contract number, goods description,
quantity, unit price and total amount;
(ii) Two copies of packing list identifying contents of each package;
(iii) Inspection certificate issued by the nominated Inspection agency, if any.
(iv) Certificate of origin;
(v) Insurance Certificate as per GCC Clause 11.
(vi) Manufacturers/Supplier’s warranty certificate & In-house inspection certificate.

B) For goods imported from abroad


Within 24 hours of despatch, the supplier shall notify the purchaser, consignee, and others
concerned if mentioned in the contract, the complete details of despatch and also supply the

HLL/PCD/ESIC-20/10-11 Page 30 of 166 01.09.2010


HLL Lifecare Limited GCC

following documents to them by airmail/ registered post / speed post (or as instructed in the
contract).
(i) Four copies of supplier’s invoice showing contract number, goods description,
quantity, unit price and total amount;
(ii) Original and four copies of the negotiable clean, on-board Bill of Lading/Airway
bill, marked freight pre paid and four copies of non-negotiable Bill of
Lading/Airway bill;
(iii) Four Copies of packing list identifying contents of each package;
(iv) Insurance Certificate as per GCC Clause 11.
(v) Manufacturer’s/Supplier’s warranty certificate;
(vi) Inspection Certificate for the despatched equipments issued by recognized/ reputed
agency like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch
(vii) Manufacturer’s own factory inspection report;
(viii) Certificate of origin
(ix) Port of Loading;
(x) Port of Discharge and
(xi) Expected date of arrival.
15. Warranty
15.1 The supplier warrants comprehensively that the goods supplied under the contract is new, unused
and incorporate all recent improvements in design and materials unless prescribed otherwise by
the purchaser in the contract. The supplier further warrants that the goods supplied under the
contract shall have no defect arising from design, materials or workmanship or from any act or
omission of the supplier that may develop under normal use of the supplied goods under the
conditions prevailing in India.
15.2 This warranty shall remain valid for 2 (Two) years in general, after the goods or any portion
thereof as the case may be, have been delivered to the final destination and installed and
commissioned at the final destination and accepted by the Purchaser/Consignee in terms of the
contract, unless specified otherwise in the SCC.
a. No conditional warranty like mishandling, manufacturing defects etc. will be acceptable.
b. Warranty as well as Comprehensive Maintenance contract will be inclusive of all
accessories and Turnkey work and it will also cover the following items:-
i. X-ray and CT tubes and high-tension cables.
ii. Helium replacement
iii. Any kind of motor
iv. Plastic & Glass parts
v. All kinds of sensors including oxygen sensors
vi. All kinds of coils, probes and transducers including ECG cable, BP transducers,
SpO2 Probes, Ultrasound and Color Doppler Transducers/probes, BP Cuffs,
Defribrillator internal paddles, chart recorders, ventilator reusable patient circuits,
servo humidifier with chamber, electrodes and probes for blood gas analyser, MRI
coils.
vii. All kinds of flat panel sensors and cassettes for Digital Radiography & Computer
Radiography systems and patients handling trolleys, etc.
viii. Printers and imagers including laser and thermal printers with all parts.
ix. UPS including the replacement of Batteries.
x. Air-conditioners
c. Replacement and repair will be under taken for the defective goods.
d. Proper marking has to be made for all spares for identification like printing of installation
and repair dates.

HLL/PCD/ESIC-20/10-11 Page 31 of 166 01.09.2010


HLL Lifecare Limited GCC

15.3 In case of any claim arising out of this warranty, the Purchaser/Consignee shall promptly notify
the same in writing to the supplier. The period of the warranty will be as per G.C.C clause number
15.2 above irrespective of any other period mentioned elsewhere in the bidding documents.
15.4 Upon receipt of such notice, the supplier shall, within 8 hours on a 24(hrs) X 7 (days) X 365
(days) basis respond to take action to repair or replace the defective goods or parts thereof, free of
cost, at the ultimate destination. The supplier shall take over the replaced parts/goods after
providing their replacements and no claim, whatsoever shall lie on the purchaser for such replaced
parts/goods thereafter. The penalty clause for non-rectification will be applicable as per tender
conditions
15.5 In the event of any rectification of a defect or replacement of any defective goods during the
warranty period, the warranty for the rectified/replaced goods shall be extended to a further
period as mentioned under clause 15.2 from the date such rectified / replaced goods starts
functioning to the satisfaction of the purchaser.
15.6 If the supplier, having been notified, fails to respond to take action to repair or replace the
defect(s) within 8 hours on a 24(hrs) X 7 (days) X 365 (days) basis, the purchaser may proceed to
take such remedial action(s) as deemed fit by the purchaser, at the risk and expense of the supplier
and without prejudice to other contractual rights and remedies, which the purchaser may have
against the supplier.
15.7 During Warranty period, the supplier is required to visit at each consignee’s site at least once in 6
months commencing from the date of the installation for preventive maintenance of the goods
15.8 The Purchaser/Consignee reserve the rights to enter into Annual Comprehensive Maintenance
Contract between Consignee and the Supplier for the period as mentioned in General Points for
Technical Specifications, Section VII (para-4), after the completion of warranty period.
15.9 The supplier along with its Indian Agent and the CMC provider shall ensure continued supply of
the spare parts for the machines and equipments supplied by them to the purchaser for 10 years
from the date of installation and handing over.
15.10 The Supplier along with its Indian Agent and the CMC Provider shall always accord most
favoured client status to the Purchaser vis-à-vis its other Clients/Purchasers of its
equipments/machines/goods etc. and shall always give the most competitive price for its
machines/equipments supplied to the Purchaser/Consignee.
16. Assignment
16.1 The Supplier shall not assign, either in whole or in part, its contractual duties, responsibilities and
obligations to perform the contract, except with the Purchaser’s prior written permission.
17. Sub Contracts
17.1 The Supplier shall notify the Purchaser in writing of all sub contracts awarded under the contract
if not already specified in its tender. Such notification, in its original tender or later, shall not
relieve the Supplier from any of its liability or obligation under the terms and conditions of the
contract.
17.2 Sub contract shall be only for bought out items and sub-assemblies.
17.3 Sub contracts shall also comply with the provisions of GCC Clause 4 (“Country of Origin”).
18. Modification of contract
18.1 If necessary, the purchaser may, by a written order given to the supplier at any time during the
currency of the contract, amend the contract by making alterations and modifications within the
general scope of contract in any one or more of the following:
a) Specifications, drawings, designs etc. where goods to be supplied under the contract are
to be specially manufactured for the purchaser,
b) Mode of packing,
c) Incidental services to be provided by the supplier

HLL/PCD/ESIC-20/10-11 Page 32 of 166 01.09.2010


HLL Lifecare Limited GCC

d) Mode of despatch,
e) Place of delivery, and
f) Any other area(s) of the contract, as felt necessary by the purchaser depending on the
merits of the case.
18.2 In the event of any such modification/alteration causing increase or decrease in the cost of goods
and services to be supplied and provided, or in the time required by the supplier to perform any
obligation under the contract, an equitable adjustment shall be made in the contract price and/or
contract delivery schedule, as the case may be, and the contract amended accordingly. If the
supplier doesn’t agree to the adjustment made by the Purchaser/Consignee, the supplier shall
convey its views to the Purchaser/Consignee within twenty-one days from the date of the
supplier’s receipt of the Purchaser’s/Consignee’s amendment / modification of the contract.
19. Prices
19.1 Prices to be charged by the supplier for supply of goods and provision of services in terms of the
contract shall not vary from the corresponding prices quoted by the supplier in its tender and
incorporated in the contract except for any price adjustment authorised in the SCC.
20. Taxes and Duties
20.1 Supplier shall be entirely responsible for all taxes, duties, fees, levies etc. incurred until delivery
of the contracted goods to the purchaser.
20.2 Further instruction, if any, shall be as provided in the SCC.

21. Terms and Mode of Payment


21.1 Payment Terms
Payment shall be made subject to recoveries, if any, by way of liquidated damages or any
other charges as per terms & conditions of contract in the following manner.
A) Payment for Domestic Goods Or Foreign Origin Located Within India.
Payment shall be made in Indian Rupees as specified in the contract in the following manner:
a) On delivery:
90 % payment of the contract price shall be paid on receipt of goods in good condition and
upon the submission of the following documents:
(i) Four copies of supplier’s invoice showing contract number, goods description, quantity,
unit price and total amount;
(ii) Consignee Receipt Certificate as per Section XVII in original issued by the authorized
representative of the consignee;
(iii) Two copies of packing list identifying contents of each package;
(iv) Inspection certificate issued by the nominated Inspection agency, if any;
(v) Insurance Certificate as per GCC Clause 11;
(vi) Certificate of origin.
b) On Acceptance:
Balance 10 % payment would be made against ‘Final Acceptance Certificate’ as per
Section XVIII of goods to be issued by the consignees subject to recoveries, if any, either
on account of non-rectification of defects/deficiencies not attended by the Supplier or
otherwise.
B) Payment for Imported Goods:

HLL/PCD/ESIC-20/10-11 Page 33 of 166 01.09.2010


HLL Lifecare Limited GCC

Payment for foreign currency portion shall be made in the currency as specified in the contract in
the following manner:
a) On delivery:
Ninety (90) % of the net CIP price (CIP price less Indian Agency commission) of the goods
shipped shall be paid through irrevocable, non-transferable Letter of Credit (LC) opened in
favour of the supplier in a bank in his country and upon submission of documents specified
hereunder:
(i) Four copies of supplier’s invoice showing contract number, goods description, quantity,
unit price and total amount;
(ii) Original and four copies of the negotiable clean, on-board Bill of Lading/ Airway bill ,
marked freight pre paid and four copies of non-negotiable Bill of Lading/Airway bill;
(iii) Four Copies of packing list identifying contents of each package;
(iv) Insurance Certificate as per GCC Clause 11 and documents also to be submitted for
payment of LC confirming that dispatch documents has already been sent within 24
hours to all concerned as per the contract;
(v) Manufacturer’s/Supplier’s warranty certificate;
(vi) Manufacturer’s own factory inspection report and
(vii) Certificate of origin by the chamber of commerce of the concerned country;
(viii) Inspection Certificate for the despatched equipments issued by recognized/ reputed
agency like SGS, Lloyd or equivalent (acceptable to the purchaser) prior to despatch.
(ix) Consignee Receipt Certificate as per Section XVII in original issued by the authorized
representative of the consignee
b) On Acceptance:
Balance payment of 10 % of net CIP price of goods would be made against ‘Final Acceptance
Certificate’ as per Section XVIII to be issued by the consignees through irrevocable, non-
transferable Letter of Credit (LC) opened in favour of the Foreign Principal in a bank in his
country, subject to recoveries, if any.
c) Payment of custom duty amount with Custom Duty Exemption Certificate (CDEC), if
applicable, customs clearance and handling charges, loading/ unloading, inland
transportation, incidental costs till consignee site & incidental services (including
installation & commissioning, supervision, demonstration and training) will be paid in
Indian Rupees to the Indian agent at actual not exceeding the quoted rates after 100 %
payment to the foreign principal.

d) Payment of Indian Agency Commission:


Indian Agency commission will be paid to the manufacturer’s agent in the local currency for
an amount in Indian rupees indicated in the relevant Price Schedule (as per prevailing rate of
exchange ruling on the date of Contract) and shall not be subject to further escalation /
exchange variation. Payment shall be made after 100% payment to the Foreign Principal.

C) Payment of Turnkey, if any:

Turnkey payment will be made to the manufacturer’s agent in Indian rupees indicated in the
relevant Price Schedule (as per prevailing rate of exchange ruling on the date of Contract) and
shall not be subject to further escalation / exchange variation. Payment shall be made to the Indian
Agent after 100 % payment to the Foreign Principal.

HLL/PCD/ESIC-20/10-11 Page 34 of 166 01.09.2010


HLL Lifecare Limited GCC

D) Payment for Annual Comprehensive Maintenance Contract (CMC) Charges:

The consignee will enter into CMC with the supplier at the rates as stipulated in the contract. The
payment of CMC will be made on six monthly basis after satisfactory completion of said period,
duly certified by the consignee on receipt of bank guarantee for an amount equivalent to 2.5 % of
the cost of the equipment as per contract in the prescribed format given in Section XV valid till 2
months after expiry of entire CMC period.

21.2 The supplier shall not claim any interest on payments under the contract.
21.3 Where there is a statutory requirement for tax deduction at source, such deduction towards income
tax and other tax as applicable will be made from the bills payable to the Supplier at rates as
notified from time to time.
21.4 Irrevocable & non-transferable LC shall be opened by ESIC/ Purchaser. However, if the supplier
requests specifically to open confirmed LC, the extra charges would be borne by the supplier. If
LC is required to be extended and/or amended for reasons not attributable to the purchaser/
consignee, the charges thereof shall be borne by the supplier.
21.5 The payment shall be made in the currency / currencies authorised in the contract.
21.6 The supplier shall send its claim for payment in writing, when contractually due, along with
relevant documents etc., duly signed with date, to the purchaser.
21.7 While claiming payment, the supplier is also to certify in the bill that the payment being claimed
is strictly in terms of the contract and all the obligations on the part of the supplier for claiming
that payment has been fulfilled as required under the contract.
21.8 While claiming reimbursement of duties, taxes etc. (like sales tax, excise duty, custom duty) from
the Purchaser/Consignee, as and if permitted under the contract, the supplier shall also certify that,
in case it gets any refund out of such taxes and duties from the concerned authorities at a later
date, it (the supplier) shall refund to the Purchaser/Consignee forthwith.
21.9 In case where the supplier is not in a position to submit its bill for the balance payment for want of
receipted copies of Inspection Note from the consignee and the consignee has not complained
about the non-receipt, shortage, or defects in the supplies made, balance amount will be paid by
the paying authority without consignee’s receipt certificate after three months from the date of the
preceding part payment for the goods in question, subject to the following conditions:
(a) The supplier will make good any defect or deficiency that the consignee (s) may report within
six months from the date of despatch of goods.
(b) Delay in supplies, if any, has been regularized.
(c) The contract price where it is subject to variation has been finalized.
(d) The supplier furnishes the following undertakings:
“I/We, __________ certify that I/We have not received back the Inspection Note duly receipted by the
consignee or any communication from the purchaser or the consignee about non-receipt, shortage or
defects in the goods supplied. I/We ______ agree to make good any defect or deficiency that the
consignee may report within three months from the date of receipt of this balance payment.
22. Delay in the supplier’s performance

22.1 The supplier shall deliver the goods and perform the services under the contract within the time
schedule specified by the Purchaser/Consignee in the List of Requirements and as incorporated in
the contract.
22.2 Subject to the provision under GCC clause 26, any unexcused delay by the supplier in maintaining
its contractual obligations towards delivery of goods and performance of services shall render the
supplier liable to any or all of the following sanctions:
(i) imposition of liquidated damages,

HLL/PCD/ESIC-20/10-11 Page 35 of 166 01.09.2010


HLL Lifecare Limited GCC

(ii) forfeiture of its performance security and


(iii) termination of the contract for default.
22.3 If at any time during the currency of the contract, the supplier encounters conditions hindering
timely delivery of the goods and performance of services, the supplier shall promptly inform the
Purchaser/Consignee in writing about the same and its likely duration and make a request to the
Purchaser/Consignee for extension of the delivery schedule accordingly. On receiving the
supplier’s communication, the Purchaser/Consignee shall examine the situation as soon as
possible and, at its discretion, may agree to extend the delivery schedule, with or without
liquidated damages for completion of supplier’s contractual obligations by issuing an amendment
to the contract.
22.4 When the period of delivery is extended due to unexcused delay by the supplier, the amendment
letter extending the delivery period shall, interalia contain the following conditions:
(a) The Purchaser/Consignee shall recover from the supplier, under the provisions of the clause 23
of the General Conditions of Contract, liquidated damages on the goods and services, which
the Supplier has failed to deliver within the delivery period stipulated in the contract.
(b) That no increase in price on account of any ground, whatsoever, including any stipulation in
the contract for increase in price on any other ground and, also including statutory increase in
or fresh imposition of customs duty, excise duty, sales tax/ VAT, Service Tax and Works
Contract Tax or on account of any other tax or duty which may be levied in respect of the
goods and services specified in the contract, which takes place after the date of delivery
stipulated in the contract shall be admissible on such of the said goods and services as are
delivered and performed after the date of the delivery stipulated in the contract.
(c) But nevertheless, the Purchaser/Consignee shall be entitled to the benefit of any decrease in
price on account of reduction in or remission of customs duty, excise duty, sales tax/ VAT,
Service Tax and Works Contract Tax or any other duty or tax or levy or on account of any
other grounds, which takes place after the expiry of the date of delivery stipulated in the
contract.
22.5 The supplier shall not dispatch the goods after expiry of the delivery period. The supplier is
required to apply to the Purchaser/Consignee for extension of delivery period and obtain the same
before despatch. In case the supplier dispatches the goods without obtaining an extension, it would
be doing so at its own risk and no claim for payment for such supply and / or any other expense
related to such supply shall lie against the purchaser.
23. Liquidated damages
23.1 Subject to GCC clause 26, if the supplier fails to deliver any or all of the goods or fails to perform
the services within the time frame(s) incorporated in the contract, the Purchaser/Consignee shall,
without prejudice to other rights and remedies available to the Purchaser/Consignee under the
contract, deduct from the contract price, as liquidated damages, a sum equivalent to 0.5% per
week of delay or part thereof on delayed supply of goods and/or services until actual delivery or
performance subject to a maximum of 10% of the contract price. Once the maximum is reached
Purchaser/ Consignee may consider termination of the contract as per GCC 24.
During the above-mentioned delayed period of supply and / or performance, the conditions
incorporated under GCC sub-clause 22.4 above shall also apply.
24. Termination for default
24.1 The Purchaser/Consignee, without prejudice to any other contractual rights and remedies available
to it (the Purchaser/Consignee), may, by written notice of default sent to the supplier, terminate
the contract in whole or in part, if the supplier fails to deliver any or all of the goods or fails to

HLL/PCD/ESIC-20/10-11 Page 36 of 166 01.09.2010


HLL Lifecare Limited GCC

perform any other contractual obligation(s) within the time period specified in the contract, or
within any extension thereof granted by the Purchaser/Consignee pursuant to GCC sub-clauses
22.3 and 22.4.
24.2 In the event of the Purchaser/Consignee terminates the contract in whole or in part, pursuant to
GCC sub-clause 24.1 above, the Purchaser/Consignee may procure goods and/or services similar
to those cancelled, with such terms and conditions and in such manner as it deems fit and the
supplier shall be liable to the Purchaser/Consignee for the extra expenditure, if any, incurred by
the Purchaser/Consignee for arranging such procurement.
24.3 Unless otherwise instructed by the Purchaser/Consignee, the supplier shall continue to perform the
contract to the extent not terminated.
25. Termination for insolvency
25.1 If the supplier becomes bankrupt or otherwise insolvent, the purchaser reserves the right to
terminate the contract at any time, by serving written notice to the supplier without any
compensation, whatsoever, to the supplier, subject to further condition that such termination will
not prejudice or affect the rights and remedies which have accrued and / or will accrue thereafter
to the Purchaser/Consignee.
26. Force Majeure
26.1 Notwithstanding the provisions contained in GCC clauses 22, 23 and 24, the supplier shall not be
liable for imposition of any such sanction so long the delay and/or failure of the supplier in
fulfilling its obligations under the contract is the result of an event of Force Majeure.
26.2 For purposes of this clause, Force Majeure means an event beyond the control of the supplier and
not involving the supplier’s fault or negligence and which is not foreseeable and not brought about
at the instance of, the party claiming to be affected by such event and which has caused the non –
performance or delay in performance. Such events may include, but are not restricted to, acts of
the Purchaser/Consignee either in its sovereign or contractual capacity, wars or revolutions,
hostility, acts of public enemy, civil commotion, sabotage, fires, floods, explosions, epidemics,
quarantine restrictions, strikes excluding by its employees, lockouts excluding by its management,
and freight embargoes.
26.3 If a Force Majeure situation arises, the supplier shall promptly notify the Purchaser/Consignee in
writing of such conditions and the cause thereof within twenty-one days of occurrence of such
event. Unless otherwise directed by the Purchaser/Consignee in writing, the supplier shall
continue to perform its obligations under the contract as far as reasonably practical, and shall seek
all reasonable alternative means for performance not prevented by the Force Majeure event.
26.4 If the performance in whole or in part or any obligation under this contract is prevented or delayed
by any reason of Force Majeure for a period exceeding sixty days, either party may at its option
terminate the contract without any financial repercussion on either side.
26.5 In case due to a Force Majeure event the Purchaser/Consignee is unable to fulfil its contractual
commitment and responsibility, the Purchaser/Consignee will notify the supplier accordingly and
subsequent actions taken on similar lines described in above sub-paragraphs.
27. Termination for convenience
27.1 The Purchaser/Consignee reserves the right to terminate the contract, in whole or in part for its
(Purchaser’s/Consignee ’s) convenience, by serving written notice on the supplier at any time
during the currency of the contract. The notice shall specify that the termination is for the
convenience of the Purchaser/Consignee. The notice shall also indicate interalia, the extent to
which the supplier’s performance under the contract is terminated, and the date with effect from
which such termination will become effective.
27.2 The goods and services that are complete and ready in terms of the contract for delivery and
performance within thirty days after the supplier’s receipt of the notice of termination shall be

HLL/PCD/ESIC-20/10-11 Page 37 of 166 01.09.2010


HLL Lifecare Limited GCC

accepted by the Purchaser/Consignee following the contract terms, conditions and prices. For the
remaining goods and services, the Purchaser/Consignee may decide:
a) To get any portion of the balance completed and delivered at the contract terms, conditions
and prices; and / or
b) To cancel the remaining portion of the goods and services and compensate the supplier by
paying an agreed amount for the cost incurred by the supplier towards the remaining portion
of the goods and services.
28. Governing language
28.1 The contract shall be written in English language following the provision as contained in GIT
clause 4. All correspondence and other documents pertaining to the contract, which the parties
exchange, shall also be written accordingly in that language.
29. Notices
29.1 Notice, if any, relating to the contract given by one party to the other, shall be sent in writing or by
cable or telex or facsimile and confirmed in writing. The procedure will also provide the sender of
the notice, the proof of receipt of the notice by the receiver. The addresses of the parties for
exchanging such notices will be the addresses as incorporated in the contract.
29.2 The effective date of a notice shall be either the date when delivered to the recipient or the
effective date specifically mentioned in the notice, whichever is later.
30. Resolution of disputes
30.1 If dispute or difference of any kind shall arise between the Purchaser/Consignee and the supplier
in connection with or relating to the contract, the parties shall make every effort to resolve the
same amicably by mutual consultations. The jurisdiction for the settlement of disputes will be at
New Delhi, India.
30.2 If the parties fail to resolve their dispute or difference by such mutual consultation within twenty-
one days of its occurrence, then, unless otherwise provided in the SCC, either the
Purchaser/Consignee or the supplier may give notice to the other party of its intention to
commence arbitration, as hereinafter provided the applicable arbitration procedure will be as per
the Arbitration and Conciliation Act, 1996 of India. In the case of a dispute or difference arising
between the Purchaser/Consignee and a domestic Supplier relating to any matter arising out of or
connected with the contract, such dispute or difference shall be referred to the sole arbitration of
an officer, appointed to be the arbitrator by the Director General of ESIC. The award of the
arbitrator shall be final and binding on the parties to the contract subject to the provision that the
Arbitrator shall give reasoned award in case the value of claim in reference exceeds Rupees One
Lac (Rs. 1,00,000/-)
30.3 Venue of Arbitration: The venue of arbitration shall be New Delhi, India.
31. Applicable Law
The contract shall be governed by and interpreted in accordance with the laws of India for the
time being in force.
32. General/ Miscellaneous Clauses
32.1 Nothing contained in this Contract shall be constructed as establishing or creating between the
parties, i.e. the Supplier/its Indian Agent/CMC Provider on the one side and the Purchaser on the
other side, a relationship of master and servant or principal and agent.
32.2 Any failure on the part of any Party to exercise right or power under this Contract shall not
operate as waiver thereof.
32.3 The Supplier shall notify the Purchaser/Consignee of any material change would impact on
performance of its obligations under this Contract.

HLL/PCD/ESIC-20/10-11 Page 38 of 166 01.09.2010


HLL Lifecare Limited GCC

32.4 Each member/constituent of the Supplier/its Indian Agent/CMC Provider, in case of consortium
shall be jointly and severally liable to and responsible for all obligations towards the
Purchaser/Consignee for performance of contract/services including that of its Associates/Sub
Contractors under the Contract.
32.5 The Supplier/its Indian Agent/CMC Provider shall at all times, indemnify and keep indemnified
the Purchaser against all claims/damages etc. for any infringement of any Intellectual Property
Rights (IPR) while providing its services under CMC or the Contract.
32.6 The Supplier/its Agent/CMC Provider shall, at all times, indemnify and keep indemnified the
Purchaser/Consignee against any claims in respect of any damages or compensation payable in
consequences of any accident or injury sustained or suffered by its employees or agents or by any
other third party resulting from or by any action, omission or operation conducted by or on behalf
of the supplier/its associate/affiliate etc.
32.7 All claims regarding indemnity shall survive the termination or expiry of the contract

HLL/PCD/ESIC-20/10-11 Page 39 of 166 01.09.2010


HLL Lifecare Limited SCC

SECTION – V

SPECIAL CONDITIONS OF CONTRACT (SCC)

The following Special Conditions of Contract (SCC) will apply for this purchase. The corresponding
clauses of General Conditions of Contract (GCC) relating to the SCC stipulations have also been
incorporated below.

These Special Conditions will modify/substitute/supplement the corresponding (GCC) clauses.


Whenever there is any conflict between the provision in the GCC and that in the SCC, the provision
contained in the SCC shall prevail.

HLL/PCD/ESIC-20/10-11 Page 40 of 166 01.09.2010


HLL Lifecare Limited List of Requirement

SECTION – VI

LIST OF REQUIREMENTS

Part I:
Consignee wise Qty.
Sl. Total
Item Description Basai Qty
No. Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana
Darapur
1 Biphasic Defibrilator 2 1 5 8
2 Fiber Optic Bronchoscope with Camera 1 1 1 1 1 1 6
3 Pulse Oximeter 2 1 3
4 Anaesthesia Workstation 2 2
5 ECG Machine 12 channel 1 1 2
6 Pharmaceutical Refrigerator for O.T. 1 1
7 O T Tables for Orthopaedics 1 1 1 1 1 5
O T Tables for General Surgery &
13
8 Gynaecology 2 3 3 1 4
9 Electro Surgical Unit 2 3 5
10 O T Table for Minor O T 1 1
11 Boyle's Appratus 1 1
Multipara Monitor (6Nos) with Central
2
12 Station (1N0) 1 1
13 Portable Ventilator 1 1
14 BIPAP 2 2
15 Binocular Microscope 1 3 2 6
16 Urine Analyzer 1 1
17 ELISA Reader with Washer 1 1 1 1 1 5
Cell Counter(Fully automated 5 part
2
18 differential analyser) 1 1
19 Biochemistry Semi- Auto Analyser 1 1 2
20 Coagulation Analyzer 1 1 2
21 Electrolyte Analyzer 1 1
22 500 M.A. X-ray machine 1 1
23 Portable X-ray machine (Mobile) 1 2 3
24 Colour Doppler Digital 1 1
25 Infant Resuscitator 1 1
26 Neonatal Monitor 1 2 2 1 6
27 Ventilator Critical Care 1 1
28 Open Care System 3 1 1 1 6
29 Transport Incubators 1 1
30 Plaster cutter with Aspirator 1 1
31 Tourniquet Electronic 1 1
32 Hot Pack Unit 1 1
33 Cold Pack Unit 1 1
34 Traction Unit 1 1
35 Operating Laparoscope Set for Gynaec 1 1
36 Hysteroscope System 1 1
37 Cryo Surgical Unit 1 1 1 1 4
38 Vacuum Extractor 1 1 1 1 4
39 Fetal Monitor 1 2 2 2 7
40 ENT Examination Unit 1 1 2 4
Operating Laparoscope Set for General
2
41 Surgery 1 1
42 Multipara Bed side Monitor 1 1

HLL/PCD/ESIC-20/10-11 Page 41 of 166 01.09.2010


HLL Lifecare Limited List of Requirement

Consignee wise Qty.


Sl. Total
Item Description Basai Qty
No. Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana
Darapur
Emergency Resuscitation Kit-
6
43 Adult(imported) 1 2 2 1
44 Portable Spirometer 1 1
45 Infusion Pump Syringe 1 4 5
PFT Machine(Spirometer &
2
46 Computerised) 1 1
47 Fully Automated Biochemistry Analyzer 1 1
48 Centrifuge Machine 24-36 tubes 3 3
49 Computed Radiography System 1 1
Advancved Colour Doppler Ultrasound
1
50 Scanner 1
51 Donor Chair 2 2
52 Blood Collection Monitor 2 1 3
53 Blood Bank Refrigerator 1 2 3
54 Tube Sealer 1 1 2
55 Cell Counter - 3 Parameter 1 1
56 Refrigerated Centrifuge 1 1
57 Plasma Expressor(E) 2 2
58 Dielectric Tube Sealer Table Top 1 1
59 PLATELET INCUBATOR & AGITATOR 1 1
AUTOMATED PLASMA THAWING
1
60 EQUIPMENT 1
61 LAMINAR FLOW- VERTICAL 1 1
62 Deep Freezer -40 Deg. 1 1 2
63 Deep Freezer -80 Deg. 1 1
64 Cell Separator 1 1
65 ICU Ventilator (High End) 1 1
66 ABG Analyzer 1 1 2
67 Arthroscopy System 1 1
68 Color Doppler Ultrasound Unit 1 1
69 Video Bronchoscopy System 1 1
70 High Frequency Mobile X- Ray Unit 1 1
71 Digital Mamography System 1 1
72 Holter Monitoring System 2 2
73 Stress Test System(TMT) 1 1
74 Volumetric Infusion Pump 10 10
75 Biosafety Cabinet 1 1
76 Laprocator 2 2
Ultrasound Machine 3D with Color
2
77 Doppler 2
78 Phototherapy Machine 1 1
79 Microscope 6 6
80 Yag Laser 1 1
81 Cell Separator 1 1
82 Cell Washer 1 1
83 Donor Couch 1 1
84 Platelet Incubator & Agitator 1 1
85 Sterile Tube Connecting Device 1 1
86 General Instruments Set 4 8 8 4 24
87 Surgery Instrument Set for Minor O.T. 1 1 1 1 4
88 Orthopaedic Surgery Instrument Set 1 1
89 Vaginal Hysterectomy 1 1 1 1 4
90 Abdominal Hysterectomy set 1 1
91 Delivery and Ceasarian set 1 1 1 1 4

HLL/PCD/ESIC-20/10-11 Page 42 of 166 01.09.2010


HLL Lifecare Limited List of Requirement

Consignee wise Qty.


Sl. Total
Item Description Basai Qty
No. Adityapur Indore Joka Baddi Manesar Bhiwadi Ludhiana
Darapur
92 General Surgical Instruments for O T 1 1
93 Instruments for Dental Dept. 1 1
94 Instruments for Eye Implant 1 1

Part II: Required Delivery Schedule:

a) For Indigenous goods or for imported goods if supplied from India:


Within 60 days from date of Notification of Award to delivery at consignee site. The date of
delivery will be the date of delivery at consignee site (Tenderers may quote earliest delivery
period).

b) For Imported goods directly from abroad:


Within 90 days from date of opening of L/C. The date of delivery will be the date of delivery at
consignee site (Tenderers may quote earliest delivery period).

Part III: Scope of Incidental Services:


Installation & Commissioning, Supervision, Demonstration, Trial run and Training etc. as specified
in GCC Clause 13. Installation & Commissioning shall be completed within 15 days of handing
over the site of installation, complete in all respect by the consignee. The date of handing over of
the site has to be intimated by the supplier to the purchaser. The delay on the part of the supplier to
install and commission the equipment will attract the provisions as contained in the liquidated
damage clause.

Part IV:
Turnkey (if any) as per details in Technical Specification. The tenderer shall also specify the time
schedule for completion of Turnkey work.

Part V:
Annual Comprehensive Maintenance Contract (CMC) as per details in General Technical
Specifications para 4. The cost of CMC need not be quoted for items in sl. no. 86 to 94 and these
items will be evaluated for ranking purpose without the CMC price.

Part VI:
Required Terms of Delivery and Destination.
a) For Indigenous goods or for imported goods if supplied from India:
Delivery required at Consignee Site.
Insurance (local transportation and storage) would be borne by the Supplier from warehouse
to the consignee site for a period including 3 months beyond date of delivery

b) For Imported goods directly from abroad:


The foreign tenderers are required to quote their rates on DDP at consignee’s site basis
giving break up of the price as per the Proforma prescribed in the Price Schedule.
Custom clearance, handling, unloading & loading and transportation to the consignee’s site
shall be the responsibility of the supplier/ Indian agent.
Insurance (local transportation and storage) would be extended and borne by the Supplier
from ware house to the consignee site for a period including 3 months beyond date of
delivery.
HLL/PCD/ESIC-20/10-11 Page 43 of 166 01.09.2010
HLL Lifecare Limited List of Requirement

c) Destination/Consignee details are given in Section XXI

Part VII:
Inspection:
a) For Indigenous goods or for imported goods if supplied from India:
At consignee site by the respective In-charge of the Hospital or his authorised representative.

b) For Imported goods directly from abroad:


Pre-despatch inspection at manufacturer’s premises as per GCC clause 8.8 and inspection
after arrival in India at consignee site by the respective In-charge of the Hospital or his
authorised representative.

HLL/PCD/ESIC-20/10-11 Page 44 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Section – VII
Technical Specifications

Item No. 1

Biphasic Defibrilator

• Biphasic
• Waveform display (min 3)
• External energy selection from 2 J to 200 J, biphasic
• Charging time less than 8 seconds @ 200 J (with a charged battery)
• Synchronizer and cardio version
• Unique disarm button (in addition to automatic time delay)
• Should come with high resolution monitor.
• Should come with 3 lead ECG that can be measured from cables, adult external
paddles, paediatric paddles.
• Heart rate: 20 to 300 bpm BPM with user selectable alarms.
• Should come with external pacing, demand and asynchronous modes
• Should display CPR in real time
• Should have large internal memory that stores and prints 25 ECG events.
• Battery to last not less than 2 hours of continuous ECG
monitoring or 30 full energy discharges.
• Battery indicator on display and self test on battery
Accessories needed
• Paddles with remote energy selection, charge and discharge buttons on paddles.
• Pediatric paddles adapters (set of 2)
• A/c 240 v V 50 hz charger and mains power source. Adult paddles and test paddles.
• 3 lead patient cable
• One spare battery
• Roll of 50mm recording paper x 10 rolls
• 5 oz tube of defibrillation gel x 5 tubes
• ECG cable with leads- 2 sets.
• External disposable pacing pads-2 sets
• Operation manual, service manual complete.

Item No. 2

Fiber Optic Bronchoscope with Camera

Bronchoscope should have :


Field of view : 120deg.
Depth of field : 3-50mm
Tip deflection up/down : 180/130deg.
Distal diameter : 4.9 mm
Insertion tube diameter : 4.8mm
Diameter of working channel : 2.2mm
Insertion tube working length : 600mm
Total length : 900mm
Sharp, smooth image
Ergonomic control section
Easy insertion and excellent maneuverability

HLL/PCD/ESIC-20/10-11 Page 45 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Compatible to electrosurgical treatment


Light Source should have :
Compatible light source (Halogen) with fibre-optic cable & back-up lamp
Standard Accessories like :
Foreign Body Retriever Forceps
Tempo-nation Balloons
Magnet Forceps
Biopsy Forceps
Single Chip camera of high resolution > 500 lines
Incorporate 1/3" high sensitivity CCD and should have anti moire filter
C-mount camerahead to attach bronchoscope and 3 windows size.
14" colour Medical grade LCD monitor for images

Item No. 3
Pulse Oximeter

· It monitors from -neonate to adult.


. To track oxygen saturation even during low perfusion condition.
· Set seconds alarm management technology avoids unnecessary alarm & distraction
during monitoring
of severe motion patients.
. Display for better monitoring.
· Dual colour LED for differentiating normal mode & alarm mode.
· Internal battery provides minimum of 8 hours backup.
· 10 segment LED bar graph for pulse strength monitoring.
· Facility for computer interface & printer connectivity
· Individual alarm settings for Sao2 & pulse rate
· Compatible with the complete range of disposable & reusable sensors.
· Built-in battery charger with charger indication.

Item No. 4

Anaesthesia Workstation

Technical specifications:
Anesthesia system should be high end three gas system with three gas Oxygen,
Nitrous Oxide and Medical Air
double scale flowmeter with high and low flow and minimal flow provisions.
Should have an independent Oxygen flow meter for Oxygen delivery and an integrated
variable flow suction unit.
System should have at least three drawers and an additional writing surface that can
be easily accessed.
Drawers shall have the ability to lock , and shall be easily removed for the purposes of
cleaning and
sterilisation.
Pipeline, cylinder and Airway pressures should all be displayed on colour coded
gauges and be visible at all times during operation.
Should have provision to attach 2 cylinders 1 each for O2 and N2O.
Should have facility of delivering basal flow of oxygen on switiching on the machine.
System should have a second user accessible port for extraction of Anesthetic gas
when using a nonrebreathing
patient circuit. System should also provide the option of returning sample gas to the
scavenging
system with a dedicated port.
A single pneumatic/electric on/off switch should activate the gas flow and vaporization.

HLL/PCD/ESIC-20/10-11 Page 46 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

The unit should have a battery back up facility for the ventilator in the event of power
failure and should operate for a minimum of one hour.
In the event of complete power loss and battery failure it shall still be possible to
manually ventilate and deliver anaesthetic agent.
System ahould have easily accessible common gas outlet in the event of an
emergency and for use of alternate breathing circuits.
Should have unlockable Oxygen flush to deliver oxygen flow of approximately 40l/min.
Should have built in safety features like O2 failure alarm, N2O cutoff, Low O2 pressure
etc.,
Should have motion sensitive back lighting for vaporizer dial adjustment. Should also
have mandatory
illumination of the writing table.
The frame should have integrated power outlets to supply a minimum of four external
devices.
Should have locking of the front castors by a single central brake mechanism.
Gas Flow
The unit shall have a mechanical hypoxic guard system to control the ratio of Oxygen
and Nitrous oxide to
ensure a minimum of 25% of oxygen delivery at all times to avoid delivery of hypoxic
mixture.
It shall be possible to deliver Air with only basal flow oxygen independent of the above
mentioned hypoxic
control.
Gas flow shall be controlled mechanically to avoid errors during power failure and
electronic malfunction.
Visual display of the gas flow shall be by physical means independent of electrical
power.
Cascade or dual flow tubes should be available for all gases to allow suitable resolution
and accurate control at
low total fresh gas flows.
Flow meters should have backlight and antiglare illumination.
The unit should have an independent measurement and display of fresh gas flow
offering safety for low and
minimal flow anaesthesia.
A bag arm with height and positional adjustment shall be available as an option.
Vaporizers
The unit should accommodate two vaporizers for anesthetic agent delivery to allow
easy selection of agent to
be used. A third vaporiser storage area shall be available as an option.
Vaporiser should be selectatec type, tool free installation and vaporiser of user choice
can be mounted at will
with interlocking facility to allow operation of only one vaporiser at one time.
Vaporizers supplied with the unit shall be routine maintenance free for the life of the
product.
Should provide Isoflurane and Sevoflurane key filled vaporisers.
Breathing System
All parts of the breathing system that are in contact with patient gas should be latex
free and autoclavable.
Should not require tools when dismantled for cleaning and sterilization.
Should accept large and small volume absorber canisters.
The ventilator bellows shall be clearly visible and should ascend on expiration to
provide a quick visual indicator
for system leaks.
Breathing system should have the option of C02 Absorber bypass control that will allow
the absorber canisters
to be removed without introducing system leaks.

HLL/PCD/ESIC-20/10-11 Page 47 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Should have bag / vent selecting valve integrated onto the absorber and should
automatically turn on the
ventilator when positioned to vent mode.
Ventilatior
Ventilator should be pneumatically driven, electronically controlled and should be
ascending bellows type.
Ventilator should automatically change drive gas should there be a gas depletion.
Ventilator shall have a color display with touch screen user interface.
Ventilator should have the following ventilation abilities, volume control, decelerating
flow pressure control,
SIMV with pressure support and pressure support.
Ventilator should be capable of ventilating diverse range of patient groups from
neonates to adult patients with
restictive airways with tidal volume range between 20 ml to 1500 ml with single bellows
system.
Assisted modes of breathing should be flow triggered.
Ventilator should have an active proportional exhalation valve to prevent the potential
of over delivery during
pressure modes of ventilation.
Ventilator should have a leak and compliance test that can be done independently of
the full system check.
On switching on, the ventilator system should be able to and shall give the user a
choice of doing a unit test or bypassing in the case of an emergency.
Ventilator shall compensate for fresh gas flow and compliance of the entire circuit
dynamically.
Measurement at the patient end of the circuit ( sensor at the patient end) should be
provided to compensate for
small leakages and compressible volume variability that occur during ventilation.
User should also have the option of setting a pre set compliance correction where
similar circuits are used constantly.
Should provide constant fresh gas flow into the breathing circuit during the inspiratory
phase as mandatory.
Ventilator should have the ability to set and store a hospital default as well as individual
user preferences for
easy selection of ventilation parameters and include screen layout, alarm preferences
and ventilation
settings.
User should be able to set their own password.
Apnea alarms must be user adjustable to allow for all operating conditions and phases
during Anesthesia.
Ventilator should have the ability to display and store Patient Spirometery loops
including Flow-Volume and
Pressure Volume curves.
Ventilator should also display waveforms for flow and airway pressure.
Ventilator should display measured fresh gas independent of the flow meters.
Ventilator should display a dynamic compliance measurement.
Integrated Monitoring system:
Anesthesia Monitoring system should be of modular type and capable of monitoring
adult, pediatric and
neonatal patients. Should be from the same manufacturer as of the anesthesia system.
Monitor should have minimum 19” independent flat panel display with multi color touch
screen user interface
to ensure all parameters are visible simultaneously.
Module rack / housing should be independent and should be able to be placed near to
the patient.
Should be capable of 8 traces display. Should have facility to monitor: ECG, NIBP,
SpO2, Respiration,

HLL/PCD/ESIC-20/10-11 Page 48 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Invasive pressures (3), temperatures (2), Capnography and Bispectral index. Should
have Cardiac output
port enabled.
Should have automatic identification and measurement of anesthetic agents, EtCo2,
O2 and N2O and MAC value.
Should have depth of anesthesia monitoring using Bispectral index.
Cardiac output monitoring facility with all accessories.
ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in
arrhythmia
monitoring on all 12 leads
Inbuilt ST segment analysis and arrhythmia detection for all the leads should be
available.
Should have haemodynamic, oxygenation and drug dose calculations.
EtCO2 should have both mainstream and side stream in one module.
Respiration should be available with Cardio Vascular Artifact filter.
OCRG(oxy cardio respiro gram) should be available for monitoring neonates.
Alarm parameter should flash red in the presence of high priority alarms (e.g.
ventricular fibrillation and
asystole) and flash yellow in the presence of medium or low priority alarms (e.g. noisy
signal, etc.)
24 hours trend data should be displayed.
All monitors including central station should have similar user interface for usage
among all clinicians.
Modules should be compatible with transport monitors.
Monitor shall provide capability to remote view of real time waveforms via the internet.
Should be able to
upgrade to softwares for electronic flow sheet and full dislcosure of all waveforms.
On-screen keyboard for entering this data should have USB ports to connect mouse,
key
board, bar code scanner.
Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and
actual parameter alarm
settings must be displayed on-screen when alarms are on.
Position of the displayed waveforms and color of the waveform must be user
configurable.
Monitor shall permit the optional ability to receive and display information from other
patient devices such as
ventilators, infusion pumps and other standalone devices.
All modules should be compatible with all monitors quoted.
Should be supplied with necessary accessories for adult , pediatric and neonatal
accessories.
Should be US FDA Approved
Should be compatible with HIS and Should be HL7 compliant
Monitor should have capability to accommodate remote viewing of real time waveforms
through internet.
Accessories and spares
ECG / respiration: 5 lead ECG cable and lead wire set and 10 lead ECG cable and lead
wire set per monitor
NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor
SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal sensor
with cable per monitor
IBP: Include 10 nos of disposable pressure transducer with bracket and interface cable
per monitor
Temperature: Skin and nasopharyngeal probes per monitor
BIS: 25 nos of disposable sensors per monitor
Environmental factors:
Safe disposal system : AGSS – Anesthetic Gas Scavenging System, should be in place

HLL/PCD/ESIC-20/10-11 Page 49 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 5

ECG Machine 12 channel

· Simultaneous acquisition of up to 12 leads.


· Real time continuous recording of 3, 6 and 12 channel.
· Recording speeds of 10, 25 or 50 mm/sec
· Extensive ECG quality control by AC Noise Filtering and Baseline.
· A4 size reports for convenient reading and filing.
· Colour coded keys for case of operation.
· Convenient battery operation for greater mobility.
· Versatile report formats and speed options to provide auto reports or rhythm reports.
· User configurable filters.
· Preview signal quality prior to printing. Saving time and paper.
· Keyboard entry for patient ID information.
· Capability to generate any number of ECG copies possible for filing and distribution.
· Adult and paediatric analysis programs std.
· Automatic interpretations of ECG data.
· Availability of adult, paediatric and neonatal accessories.
· Accessories e.g., stand, cables, electrodes etc. Should be quoted separately.
· ECG Paper roll for 1000 patients
· Service and operation manual complete.

Item No. 6

Pharmaceutical Refrigerator for O.T.

Good quality double steel walled cabinet


Capacity 500 litre
Forced air circulation to maintain chamber temp. uniformity.
Temp. in range of 2 to 8 degree centigrade
Low noise level
Automatic defrosting & digital temp.display
Audiovisual alarm for low & high temp.
Input voltage 220 to 240 volts
Inbuilt electronic voltage stabilizer

Item No. 7

O T Tables for Orthopaedics

Electro-hydraulic battery operated mobile operating and extension table for orthopaedic
interventions including open and intra-medullary nailing and shall be fully compatible
with C-Arm Image Intensifier.
Should include back plate adjustment, with SFC padding, including. basic equipment -
2 traction bars short and long, 1 rotation-tilt clamp, 2 elongation bars, 1 rotation and
traction stirrup clamp, 1 screw tension device, 1 counter traction post for neck of femur,
1 foot plate support, 1 side rail elongation.
Should have the following features:
- Battery-powered or mains operated electrically controlled hydraulic drive.

HLL/PCD/ESIC-20/10-11 Page 50 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

- Base covers and column casing, table top frame, traction bars and basic
accessories made of CrNi steel.
Operating table with easy-rolling castors for longitudinal and lateral movement.
Operating table top subdivided into 9 sections: head rest , 3-section back plate with 2
detachable shoulder segments, seat plate with 2 detachable buttock supports, pair of
leg plates, detachable.
- 2 traction bars beneath the seat plate with 2 separately adjustable pivot joints each
for extension of the extremities with unimpeded use of the image intensifier.
- Guide rails beneath the seat plate allowing X-ray cassette insertion from either the
right- or the left-hand side for the hip area.
- Detachable pads made of special foam core, 80 mm thick, radio translucent and
electrically conductive on plastic supporting plates.

Should meet the following Electrical standards:


Special-design batteries with a capacity for approx. 2 weeks ( 50 - 80 operations).
- Battery charge level: electronic monitoring with optical indicator and acoustic
signal
- Recharging the batteries via mains supply 100 - 240 V AC (adjustable),
- 50-60Hz, via mains cable
- Class II, Type B; the enclosure leakage current meets the requirements of the
patient leakage current for CF conditions according to EN 60601-1.
Should meet following Technical data:

- Length of table top incl. head rest : 1965 mm


(With standard back plate)

- Width of seat / back plate without side rails : 530 mm

Permissible load approx : 180kg

Should have following Electro hydraulic adjustments:


Control of the electro hydraulic motions from outside the sterile area via cordless IR
remote control optional hand control.

- Height without padding : 800 – 1175 mm

- Trendelenburg / reverse Trendelenburg : + 25° / - 30°

- Lateral tilt : 20°

- Back plate: + 60 °

- 0-position (horizontal alignment of the table top)

Should supply the following with Basic table:

- IR remote control comprising: cordless, battery-powered hand control with film


keypad and mobile charging station (to recharge the battery of the hand control) with
mains cable; wall mounting should be possible.
- Cable-connected hand control, detachable, with film keypad; for connection to the
operating table via flexible coiled cable
- Head rest
- Foot plates, width adjustable, with Velcro strap and heel pad, pair
- Arm board
- Anaesthesia screen
- All necessary clamps or adaptors

HLL/PCD/ESIC-20/10-11 Page 51 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

- Accessory stand
Should supply the following with Extension accessories:

FEMUR NAILING IN SUPINE POSITION:


a) Counter traction post for femur: 1 No.

FOR TIBIA & FIBULA NAILING


a) Counter traction post for tibia : 1 No.
b) Guide bar : 1 No.
c) Condoyle fixation : 1 No.
FOR TOTAL HIP END PROSTHESIS
a) Fixtures for body support : 2 Nos.
b) Back- buttock support with pad : 1 No
c) Lateral support with pad : 1 No.

KNEE REPLACEMENTS
a) Motorised Knee positioning device
including radial clamp : 1 No
SPINE IN KNEE ELBOW POSITION
a) Positioning device for spinal surgery, height adjustable and horizontal adjustment : 1
No.
b) Trolley : 1 No.
c) Sitting bracket with lateral supports : 1 No
d) Radial setting clamps : 2 Nos.

CERVICAL SPINE SURGERIES


a) Connection Fixture : 1 No
b) Horse shoe shaped head rest ( two pc.) : 1 No.
c) Guiding roller for head side traction : 1 No.
INTRAMEDULLARY NAILING OF HUMERUS
a) Humerus positioning device : 1 No.
b) Countertraction post for humerous : 1 No
c) Weinberger hand traction device : 1 No.

OPERATIONS ON THE HAND.


a) Hand operating table : 1 No.
b) Radial setting clamp. : 1 No.

Item No. 8

O T Tables for General Surgery & Gynaecology

- Should be Mobile, modular operating table for any surgery, electro-hydraulic


drive, with corded hand control, with upper back plate, with leg plates, with head rest,
with SFC padding of 80 mm.

Should have following features:


- Battery (1 week/ 60-70 operations on Single Charge) and mains operation.
- Stable base construction with 4 double swivel castors for easy motion and
manoeuvring (locking of the double swivel castors via a foot pedal)
- Base cover and cover for the override control panel made of CrNi steel, resistant
to impact, breakage and disinfectants
- SFC padding (Special Foam Core) with multi-layer construction for exceptional
comfort during patient positioning and utmost patient safety

HLL/PCD/ESIC-20/10-11 Page 52 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

- Safe patient positioning due to optimum covering of table top joints


- Column casing made of CrNi steel
- Supporting bars for the seat and back section, leg plate sockets, joint covers and
side rails made of CrNi steel
- Operating table top subdivided into 5 sections: head rest, upper back plate, lower
back plate, seat plate & leg plates
- If there are any malfunctions or if the corded hand control is defective, the
operating table can be controlled using the override panel.Entire table top without
crossbars, allowing intra-operative fluoroscopy
Operating elements: corded hand control, override panel, foot switch

Should meet following Electrical Standards:


- Battery charge level: electronic monitoring with optical indicator
Recharging of the batteries and supply of the operating table by means of a mains
- cord
- Nominal mains voltage /220/230-240 V frequency 50/60 Hz
- IEC 60601-1
Should meet following Technical data:
-Length of table top : 2075 mm
-Width of seat/lower back plate without side rails : 510 mm
-Width of seat/lower back plate with side rails: 560 mm
Max. Patient Weight -180 Kg

Should have following Electro hydraulic adjustments:


Via corded hand control or override control panel, with status indicators for battery
charge level and base brake feature, located at the column head.

Adjustment options via corded hand control or IR remote control:


-Height with padding: 600 – 950 mm
-Trendelenburg/reverse Trendelenburg: + 25° / - 25°
-Lateral Tilt :15°
-Lower back plate: 60° / 50°
-Leg plates: 0° / 95 °
-Longitudinal Shift: 290 mm
-Body Elevator: 100 mm
-0-position (horizontal alignment of the table top)

Should have following Manual adjustments:


-Standard back plate : + 90° / - 45°
-Standard head rest : + 45° / - 45°
-Base braking (lock / unlocking)

Should be supplied with Basic equipment:


- Corded hand control
- Head rest, inclinable, SFC padding
- Retrofit kit, side rails, for head rest for subsequent lateral mounting
- Arm Boards pair
- Anesthesia Screen
- Clamps or Adaptors
- Leg plates with SFC padding
- Gopel knee crutch with clamp pair
- Side supports pair

HLL/PCD/ESIC-20/10-11 Page 53 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 9
Electro Surgical Unit

Should be suitable for all types of surgeries.


Digital system with automatic patient plate monitoring.
Display: Digital
Monopolar cut: 300 to 400 W
Not less than two blend modes
Provision for Spray, Dessication
Bipolar Coagulation
Facility for underwater cutting
Facility for simultaneous coagulation
Audio visual alarm for breakage of contact between patient and plate
Accessories:
Double pedal foot switch
Single Pedal Foot switch
Patient plate with cable x1
Autoclavable handles: 3 sets
Electrodes: 3 sets
Bipolar forceps with cord x 1
All accessories should be from same manufacturer to ensure compatibility.
All instruments should be autoclavable or Single Use. Single Use Disposables if offered
should be
sufficient for 20 surgeries.
The equipment should be US FDA Approved
Complete instruction and service manual should be supplied.

Item No. 10

O T Table for Minor O T

O T Table with following specs


Electro Mechanical Adjustments
Height 760 mm to 960 mm,
Trendlenberg/Rev Trendelberg,25/25 deg
Lateral Tilt.-15 deg
Manual Adjustments
Backrest with gas spring up/down-65/25 deg
Headrest Detachable
Legrest Detachable
Standard. Four Section Mattress – 1 Set
Accessories
1-Arm Board 02
2-Anaesthesia Screen 01
3-Body Strap 01
4- Radial setting clamp 03
5- Goepel Knee crutches 02
6-Perineal cutout filter 01

HLL/PCD/ESIC-20/10-11 Page 54 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 11

Boyle's Appartus

Anaesthesia system should be high end three gas system with three gas Oxygen,
Nitrous Oxide and Medical Air with double scale flowmeter for O2 & N2O.
System should have at least two drawer.
Pipeline, cylinder and Airway pressures should all be displayed on colour coded
gauges and be visible at all times during operation.
Should have provision to attach 2 cylinders 1 each for O2 and N2O.
The unit should have a battery back up facility for the ventilator in the event of power
loss and should operate for a minimum of one hour.
In the event of complete power loss and battery failure it shall still be possible to
manually ventilate and deliver anaesthetic agent.
Should have unlockable Oxygen flush to deliver oxygen flow of approximately 40l/min.
Should have built in safety features like O2 failure alarm, N2O cutoff, Low O2 pressure
etc.,
The unit shall have a hypoxic guard system to control the ratio of Oxygen and Nitrous
oxide to ensure a minimum of 25% of oxygen delivery at all times to avoid delivery of
hypoxic mixture.
The unit should have an independent measurement and display of fresh gas flow
offering safety for low and minimal flow anaesthesia.
The unit should accommodate two vaporizers for anesthetic agent delivery to allow
easy selection of agent to be used.
Vaporiser should be selectatec type, tool free installation and vaporiser of our choice
can be mounted at will with interlocking facility to allow operation of only one vaporiser
at one time.
Vaporizers supplied with the unit shall be routine maintenance free for the life of the
product.
Should provide Isoflurane key filled vaporisers.
All parts of the breathing system that are in contact with patient gas should be latex
free and canister and bellow should be autoclavable.
The ventilator bellows shall be clearly visible and should ascend on expiration to
provide a quick visual indicator for system leaks.
Breathing system should have C02 Absorber.
Should have bag / vent selecting valve integrated onto the absorber and should
automatically turn on the ventilator when positioned to vent mode.
Ventilator should be pneumatically driven, electronically controlled and should be
ascending bellows type.
Ventilator should automatically change drive gas should there be a gas depletion.
Ventilator shall have a large display with touch screen user interface.
Ventilator should have the following ventilation abilities, volume control, SIMV with
pressure support .
Ventilator should be capable of ventilating diverse range of patient groups from
neonates to patients with restictive airways with tidal volume range between 20 ml to
1500 ml with single bellows system.
Assisted modes of breathing should be flow triggered.
Ventilator shall have an active proportional exhalation valve to prevent the potential of
over delivery during pressure modes of ventilation.
Ventilator shall compensate for fresh gas flow and compliance of the entire circuit
dynamically.
Measurement at the patient end of the circuit ( sensor at the patient end) should be
provided to compensate for small leakages and compressible volume variability that
occur during ventilation.
User should also have the option of setting a pre set compliance correction where
similar circuits are used constantly.

HLL/PCD/ESIC-20/10-11 Page 55 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Should provide constant fresh gas flow into the breathing circuit during the inspiratory
phase as mandatory.
Apnea alarms must be user adjustable to allow for all operating conditions and phases
during Anesthesia.
Ventilator should also display waveforms for flow and airway pressure.
Ventilator shall display measured fresh gas independent of the flow meters.
Ventilator shall display a dynamic compliance measurement.

Item No. 12

Multipara Monitor with Central Monitoring Station

Patient monitor system should be of modular type and capable of monitoring adult,
pediatric & neonatal patients.
Monitor should have 17” independent flat panel display.
Touch screen user interface .
Module rack / housing should be independent and shall be able to be placed near to
the patient.
Should be capable of 8 traces display.
Monitor must be capable of simultaneously monitoring the following parameters which
should be present as standard: ECG, NIBP, SpO2, invasive pressures (2),
temperatures (2)
Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to
be offered as optional
ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in
arrhythmia monitoring on all leads.
Inbuilt ST segment analysis and arrhythmia detection for all the leads should be
possible.
Haemodynamic and drug dose calculations should be available.
Arrhythmia should be grouped based on classifications – and should show no of
arrhythmias occurred.
Respiration should be available with Cardio Vascular Artifact filter.
ICP monitoring should be possible.
Alarm parameter should flash red in the presence of high priority alarms (e.g.
ventricular fibrillation and asystole) and flash yellow in the presence of medium or low
priority alarms (e.g. noisy signal, etc.)
24 hours trend data should be displayed.
All monitors including central station should have similar user interface for easy usage
among all clinicians.
Monitor shall provide the capability to interact with alarms at remote bedsides.
Monitor shall provide the capability to receive and display real-time waveforms, trended
data and alarm status from other bedside or telemetry units on the patient monitoring
network.
Monitor shall provide the capability enter patient information at the bedside or central
monitor.
On-screen keyboard for entering this data is preferable. Should have USB ports to
connect mouse, key board, bar code scanner.
Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and
actual parameter alarm settings must be displayed on-screen when alarms are on.
Position of the displayed waveforms must be user configurable.
Waveform color changing should be user configurable.
Monitor shall permit the optional ability to receive and display information from other
patient devices such as ventilators, infusion pumps and other standalone devices.
All modules should be compatible with all monitors quoted.
Bed to bed communication between the monitors should be possible with out a central

HLL/PCD/ESIC-20/10-11 Page 56 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

station.
Networking to central station should be possible and price of central station should be
offered as optional
Patient monitoring network shall use standard TCP/IP protocol and be capable of
residing on hospital’s network infra-structure.
Should be compatible with HIS and should be HL7 compliant.
Monitor should provide remote viewing of real time waveforms through internet.
Patient monitoring network shall be able to support up to 1,000 monitoring nodes.
Should be supplied with necessary accessories for adult , pediatric and neonatal
accessories.
Accessories and spares
1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor
2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor
3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal
sensor with cable per monitor
4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface
cable per monitor
5. Temperature: Skin and nasopharyngeal probes per monitor.
The equipment should be CE & US FDA Approved.

Central Monitoring Station for multipara monitor


System should have minimum 16 beds capability.
Central station should have 17” color display.
Should have drug dose and hemodynamic calculations.
It should have possible to view information such as vital signs, alarm status, arrhythmia
analysis, trended
parameters, patient data etc for any selected bed from the central station.
Should have separate computer keyboard and 4 channel thermal array recorder.
should have default alarm limits and customizable parameter settings.
Central station should have full bed review capability.
Central station should be able to be configured as a bedside monitor if required.
Should have 24 hours trends.
Should have capability for HL7 interface. Should be capable of monitoring telemetry
modules.
All system should have CE & US FDA certifications.
Should be supplied with a On-line suitable UPS
Note: Price of MULTI PARAMETER MONITOR and CENTRAL MONITORING
STATION
should be quoted separately.

Item No. 13

Portable Ventilator

The Ventilator should be portable.

The ventilator should be capable of ventilating from pediatric patients to adults.

The modes should include Assist Control, SIMV in Volume control and Pressure control
modes.

It should also have CPAP and PSV available.

The tidal volume should range from 50ml to 2000ml.

HLL/PCD/ESIC-20/10-11 Page 57 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Peak Inspiratory Pressure - 5 to 55 mbar

Pressure Support – 5 to 55 mbar

Insp. Time – 0.3 to 2.4 secs

Breath Rate – 0-60 BPM

Insp. Sensitivity – 1 to 5

Exp. Sensitivity – 5 – 95%

Flow Pattern should be Square, descending and Sinusoidal

PEEP should be 0.5mbar to 20mbar

Rise Time should be from 1 to 4 with increments of 1.

I:E ratio should be 1:4 to 1:1

I / T should be 20% to 50%

The Apnea time should be 1- 60 secs and backup rate of 5 to 40 BPM.

Vt SIGH should be in single to double multiplier of Vt.

The ventilator should have bright display with backlight

The ventilator should have Waveform display of Pressure Vs. Time and Flow Vs. Time.

Ventilator should be able to be used invasively through an artificial airway, or


noninvasively through a mask or other noninvasive interface.

The ventilator should have choice to select the type of breathing circuit (Paed. OR
Adult) for Circuit volume compensation.

The ventilator should have internal battery back up for at least 11 hours with Real-time
battery life indicator

The ventilator noise level should be < 30 dB at 1 m.

The ventilator should have Automatic atmospheric adjustment

The ventilator should have recording facility of patient data trends through USB for 3
months to 1 year time with following parameters;

Monitoring: pressure, inspired flow, exhaled flow and leak measurement


Trends: leaks, VTI, VTE, Rate, I/T, M. Vol, P MAX and PEEP measurements
The Machine should be operable with oxygen flow from B type oxygen cylinders as well
as central
oxygen supply.

HLL/PCD/ESIC-20/10-11 Page 58 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 14

BIPAP

Non invasive ventilation/ bipap with LCD display of parameters with following features
•IPAP range 6-30 cms
• EPAP range 4-20 cms
• Respiratory rate can be set to 4-40 bpm
• Spontaneous / CPAP/ spontaneous with time/ timed
• Connectors for et ET tube available for direct connection
• Should have CPAP, PSV ST, PCV, PACV modes
• Should have facility to set target tidal volume
• Inspiratory time should be possible to set allowing critical patients to breathe out.
The rise time from EPAP to IPAP can be set and varied
The system should be supplied with operator’s manual, ultra mirage mask (full mask) ,
hose pipe and
power cable

Item No. 15

Binocular Microscope

1. Antimould/Antifungal type microscope


2. Colour corrected infinity optical system
3. Nose piece Quintuple reversed inword facing.
4. Objectives Planachromatic and springloaded 4X (1pece), 10X (1perce), 20X(1piece),
40X(1piece), oil immersion 100X (2 piese).
5. Eye piece wide field, 10X, one pair each with preferably with pointer.
6. Field of view>20mm.
7. Trinocular eye piece tube to facilitate camera attachment. The eye piece tube should
be
siedentopf type, 30 degree inclined and rotatable by 360 degree.
8. Dioptre adjustment of both eye pieces.
9. Inbuilt arrangement of illumination with halogen lamp (6V/20W) fitted directly under
file
lenses (Kohler’s system) with intensity control.
10. Condenser- Bright field Abbe’s NA 1.25 with iris diaphragm and filter holder.
11. Coaxial fine and coarse adjustment with adjustable tension.
12. Double stage- Double slide holder low position and coaxial, movement
13. Power supply 220/240 volts.
14. Spare halogen lamps-6 no’s to be supplied with each microscope
15. Power cord
16. Dust cover and preferable with box

Item No. 16

Urine Analyzer

Two Different strips can run on the system


Combur 10M : Glucose,protein,pH, Leucocytes,Nitrite,Ketones
Bilirubin,Urobilinogen,Blood,SG (10 parameter)
Combur 9 :Glucose,protein,pH, Leucocytes,Nitrite,Ketones,Blood,Bilirubin,Urobilinogen

HLL/PCD/ESIC-20/10-11 Page 59 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Features

· Sample : Uncentrifused Fresh Urine


· Analysis time – less than1 min
· Sample throughput – approx. 60test strips/hours (normal Mode)
· Calibration through dry strips for better results .
· Totally Maintenance Free System
· Light Source - LED
· Wavelenght measurements - 470nm,555nm, 620 nm
· Data display on Touch screen monitor
· Data printout on fast low noise thermal printer
· Memory for 1000 patient results with time and date,3×100 Controll
· Simple Touch Screen operation
· Flagging of abnormal results
· Interfaces - 1 x RS 232
· 110V to 240V AC operation
· CE / CB / UL/CUL approved

Item No. 17
ELISA Reader with Washer

1. Description of Function
1.1 ELISA Reader is required to Read the Colour Density known as OD(Optical
Density) in ELISA (Enzyme Linked Immuno-Sorbent Assay.)Plates.
2 Operational Requirements
2.1 Only ELISA Reader is required.
3 Technical Specifications
3.1 OPTICAL SYSTEM
Digital light control
8 measurement channels including 1 reference.
Single and dual wavelength measurement with facility for kinetic measurement
8 s maximum measurement time for single and dual wavelength and 5
s(+/_1Sec.) for kinetic
Measurement Range 400-700nm
Indication Range 0-2.999 abs
Accuracy Plus/Minus 2% or Plus/Minus 0.005 abs
Resolution 0.001 abs
Inbuilt Filters: Narrow band interference
Should have the following filters – 405, 450, 492(+/_2nm), 540, 620 (+/-10nm)
and 690 nm
Should measure end point, curves and kinetic.
3.2 SOFTWARE:
Storage of immediately preceding measurement At least 15 user programmable
tests permanently stored
Time programmable between each measurement. Agitation programmable
before each reading
Bidirectional printer interface.
Data memory through computer
Built in Windows based software programming software.
3.3 MEASUREMENT MODES
Plate shaking mode for sample mixing selectable speed and time)
Flexible blank mode setting
Matrix Modes: Matrix -/x/t, Matrix-/0-0 (Range ),
Matrix-/f/(Floating cut off)

HLL/PCD/ESIC-20/10-11 Page 60 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Difference Mode: Absorbance of each well in even numbered subtracted from


those of odd numbered columns
Curve fit Modes: LIN/LIN.LIN/LOG.LOG/LOG or auto curve transformation
with ability to add the standard curve; 8 to 12 way string orientation or kinetic
modes
Table of optical densities, Delta DD, Graphic, Reaction rate/V-Max
3.4 Adjustable for different micro plate geometrics
3.5 Halogen Lamp 20 - 40 W.
3.6 16 digit alphanumeric fluorescent display
3.7 Membrane keyboard.
3 .7 Technical Specifications for washer
3.1 Auto strip washer for 96 well plates / strips
3.2 1 x 8 strips/ 1x12 strips.
3.3 Dispensable wash volume 50 - 300 µl.
3.3a Residual wash Volume -<0.5µl
3.4 Aerosol Shield for user safety.
3.5 In built shaking facility
4 System Configuration Accessories, spares and consumables
8-12 channel manifold, all tubing sets, wash, rinse and waste bottles
Maintenance kit to be provided.
4 System Configuration Accessories, spares and consumables
4.1 System as specified-
4.2 Halogen Lamps : 2
4.3 Printer inbuilt or external to be supplied along with 10 Rolls/Z Fold
4.4 Dust cover.
5 Environmental factors
5.1 The unit shall be capable of being stored continuously in ambient temperature of 0
-50deg C and relative humidity of 15-90%
5.2 The unit shall be capable of operating continuously in ambient temperature of 10 -
40deg C and relative humidity of 15-90%
6 Power Supply
6.1 Power input to be 220-240VAC, 50Hz fitted with Indian plug
6.2 Resettable over current breaker shall be fitted for protection
6.3 Suitable voltage corrector/stabilizer
6.4 Suitable UPS with maintenance free batteries for minimum one-hour back-up
should be supplied with the system.
7 Standards and Safety
7.1 Comprehensive training for lab staff and support services till familiarity with the
system.
7.2 Should be compliant to ISO 13485: Quality systems - Medical devices - Particular
requirements for the application of ISO 9001 applicable to manufacturers and
service providers that perform their own design activities.
7.3 Should be FDA or CE or ISI approved product
8 Documentation
8.1 User/Technical/Maintenance manuals to be supplied
8.2 Certificate of calibration and inspection from factory.
8.3 List of Equipments available for providing calibration and routine maintenance
support as per manufacturer documentation in service / technical manual.
8.4 List of important spare parts and accessories with their part number and
costing.
8.5 Log book with instruction for daily, weekly, monthly and quarterly
maintenance checklist.
The job description of the hospital technician and company service engineer
should be clearly spelt out

HLL/PCD/ESIC-20/10-11 Page 61 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 18

Cell Counter(Fully automated 5 part differential haematology


analyser)

· Automated hematology analyzer should include 24 parameters including histogram for


RBC, WBC and platelet.
· Should have impedance principle for counting and photometer for hemoglobin.
· It should read at least 60 samples per hour or more.
· Should have dual channel measurement.
· Double dilution chamber
· Sample volume less than 200 micro litres in whole blood and pre –dilute mode.
· It should have various types of discrete mode and real time random access analysis
to save
reagent consumption and analysis time.
· Sampling needle should have automatic wash from inside and outside.
· LCD / VGA Monitor with graphical user interface (GUI) for easy operation.
· Large illuminated colored VGA or LCD should display the result of all parameters and
histogram together.
· Should have sample manual and capillary mode.
· Should have capacity to store at least 20000 numeric patient results and 5000
graphics.
· Should have inbuilt / External graphic printer.
· Should have RS232 serial and parallel port can be connected with LAN and laser
printer.
· Should have a membrane keyboard for routine operations and maintenance with
option to
attach external key board for patient demographic entry at instrument operation.
· Should have three dimensional technology or Flow cytometry for differential analysis
to
maximize resolution, specificity and efficiency.
· Should have extended analysis time for cytopenic sample. .
· Should be able to integrate with optional automated slide maker and stainer.
· Should have zero routine maintenance with automatic electronic aperture cleaning
and
back flush after each sample.
· Instrument should accept all types of vacutainer tubes.
· The instrument should have option for auto sampler, bar code reader.
· Reagent cost per cycle including start up and shutdown if 200 & 500 samples are
processed at a time should be submitted separately in the financial bid.
· There should be automatic storage of calibration data and extensive quality control
programme with LJ plot for at least 8 control lots and at least 25 runs per lot.
Basic common necessities:
· Input Voltage 230 volts 50 Hz as per Indian standard.
· Service manual and technical data with all necessary passwords without any
obligation.
· Instruction and operational manuals without any obligation.
· UPS preferably sine wave based with maintenance free batteries with duration two
hours.

Item No. 19

Bio-Chemistry Semi-Auto Analyzer

HLL/PCD/ESIC-20/10-11 Page 62 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Automatic flow-through Cell Analyzer


Easy to use high quality precision Analyzer
Monochromatic and Bichromatic measurements
(340 - 630) º
Interference filters 340-405-500-546-578-630mm
Bandwidth 8 nm.
Optical measuring pass of 10 nm.
Measures in absorbances, Concentration kinetics,
Fixed Time, Rate and Differential MULTI-STANDARDS
Zero set: fully automatic
Fully programmable directly from the keyboard
Storage capacity - 100 complete tests.
Reading volume need: 500 ul
Disposable cuvette programmable with flow through
Cell: 500- 1000ul
Built in printer
Thermostated block by peltier effect at 25-30-37 °C
for 10 cuvettes.
External interface to be connected to computer.
WITH STANDARD ACCESSORIES

Item No. 20

COAGULATION ANALYZER (COAGULOMETER)

Fully automated, open random access, 4 detector channel Blood Coagulation Analyzer
having clotting,
chromogenic and immunological assay channels.
Clotting detection methods by recording change of viscosity and/or change of light
transmitability and
permitting both chromogenic and immunogenic assays
Parameters:
Provision for simultaneous random analysis for at least 15 parameters such as PT,
APTT, Fibrinogen,
Protein S , Protein C, APC resistance, ProC, Antithrombin III, Heparin, Plasminogen
Activator inhibitor,
D-Dimer Plus, von Willebrand Antigen, Lupus Anticoagulant, and coagulation factors
VIII & IX .
Should have throughput of at least 100 tests per hour for PT.
Should be provided with a compatible printer.
Should have a intra assay reproducibility of:
Normal:
PT<2.0%
APTT<2.0%
Fibrinogen<4%
Patient:
PT<2.0%
APTT<2.0%
Fibrinogen<5%
Additional Accessories required :1) All standard accessories 1 set extra.
2) Compatible Online UPS with1 hour back up.

HLL/PCD/ESIC-20/10-11 Page 63 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 21

Electrolyte Analyzer

Fully automated Ph/blood gas/electrolytes analyzer, measures the following


parameters PH, PCO2,
PCO2 and PO2 and barometric pressure, NA, K, Ca, Cl, Ct, THB, HCL
• Analyses per hour should not be less than 36
• Sample volume should not exceed 110 µl for all parameter
• All calibration and cleaning cycle should be fully automated with user selectable
calibration
times
• The electrodes provide should be zero maintenance including the reference
electrode.
• The system should have onboard data management to store all patient result and
calibrations.
• A power fault protection for 20 minute to keep all calibration and programmed data
• The system should have a screen to access all system software and to display the
patient
results.
• A built –in printer should be provided to print out patient result
• The system should have a patient data, new reagents and new electrodes. Startup kit
1000
samples, essential spares.
• The system should work on discrete testing, i.e. selectable parameter testing

Item No. 22

500 M.A. X-ray machine


A. Generator:
1. Generator should be high frequency/inverter type for constant output.
2. Output 50 kw or more.
3. Kv range 40 kv – 120 kv.
4. Output at 100 kv should be 500 ma or more.
5. It should have automatic exposure control device.
6. It should have digital display of kv & mas.
7. Anatomical programming radiography should be possible.
8. It should have over loading protection.
B. X – ray tube and collimator:
1. The x-ray tube should be rotating anode high speed, compatible with the generator
and must have dual
focus. Focal spots of following sizes:
Large focus: 1.2/2.0 mm or better.
Small focus: 0.6/1.0 mm or better.
Tube with anode heat storage capacity 150 khu or more.
2. Motorized collimator having additional filters (for dose reduction) and auto shut
provision for the light.
C. X – ray table:
1. Horizontal table with floating table top.
2. It should have transverse + 10 cm or more and longitudinal movements + 35 cm or
more with
electromagnetic brakes.
3 It should be thin table top.
4. It should be provided with bucky which can hold all standard sizes of cassettes upto
14”x17”.

HLL/PCD/ESIC-20/10-11 Page 64 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

5. Bucky should have a grid ratio 10:1 or more with 65 lines per cm.
D. Vertical bucky stand:
1. It should have provision to do chest radiography with grid.
E. Essential accessories: the following essential accessories to be provided with the
unit.
1. Lateral cassette holder – one.
Should be supplied with following:
· Whole body (upper and lower) lead apron (standard make) – 8 nos.
· Lead screen (with lead glass window)– 2 nos.
· Lead lining for walls and door
· Apron & glove hanger
Heavy duty multiple glove and apron hanger wall or floor mounted. Stainless steel
construction. Holds two pairs of
gloves and two lead aprons. Dimensions: 76 x 46 x 25 cm approximately.
All necessary aerb / barc approvals.

Item No. 23

Portable Xray unit

2.5 kw HF General purpose,single tank diagnostic mobile


X-ray Equipment,suitable for radiography of standing,
sitting or recumbend patients as well as for patients in bed or
operation table & as stand by unit.
Comparising of High Frequency X-ray generator
2.5 with the following out put-
Output: 2.5 kw as per Is 7620
Kv Range : 40 kv-100kv in 24 steps
mA Range :16mA-60mA, mAs Range : 0.32mA's-200mAs
Exposure Time : Min 20 m sec
The unit will be calibrated within the permissible limit of x-ray tube
X -RAY TUBE
stationary anode x-ray tube having 1.4*1.4mm focal spot
Collimato:-
Brightfield light beam collimator with 100w halogen
lamp and auto shut-off facility
Power Supply:-
Single phase 195v-265v, 50Hz power supply with line
resitance<0.8 ohms
Mobile Stand
Light weight,low height mobile stand with easy manoeuvarbility

Item No. 24

Colour Doppler Digital


1. Beam former
Universal Digital broadband former accepting routine Phased array sector, Convex,
Linear and Endovaginal probes.
2. Monitor

HLL/PCD/ESIC-20/10-11 Page 65 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Should have high-resolution 15” medical grade Flat Panel monitor with articulating arm
3. Digital Processing Channels
Atleast 1024 channels

4. Gray Scale
System should have a minimum of 256 gray levels with system dynamic range to be
more than 200dB
5. Display modes
With B, 2B, M, Color M mode, PW, HPRF and Color Doppler with Power Doppler.
Phase Inversion Tissue Harmonics should be available on Convex and Phased Array
probes. steering on Color / PW modes on Linear probe should be available.
Trapezoidal Image. Triplex mode .
6. Cine Review
Standard cine memory providing 500 frames on 2D mode and up to 30 seconds
Doppler Cine. To be confirmed.
7. Transducer technology
System should be capable of handling 1 – 15 MHz
Multifrequency imaging with independent selection of 2D / Color / Spectral doppler
frequency should be offered.
8. Exam Presets
Should be able to store atleast 30 user specific Image presets.
9. Image Optimisation on B and M modes
System should have the following

· Selectable frequency on probes


· Up / Down & Right / Left Image rotation
· Muiltiple steps of Edge Enhancement settings
· Upto 25cm dept.
· levels of persistence
10. Color and Spectral Doppler Optimisation
System should have the following

· Selectable frequency on all probes


· Gate adjustments on Spectral modes.
· Filter adjustments should be specified.
11. Zoom function
When a more detailed view is needed, the system should reoptimises and rescans the
selected region to actually increase the acoustic information content of the image upto
10 times.
12. Deep Doppler and color performance
HPRF performance at depth
13. Measurements and Calculations.
System should have atleast 6calipers with depth information and extensive,
customizable measurement and report packages including complete Obstetrics,
Gynecology studies.

Distance, Area, % stenosis on B mode


Distance, Time, Heart Rate, Slope on M mode
Velocity, Acceleration time, Slope, PI, RI, S/D Ratio with Auto Doppler on Doppler
mode
14. Image Management
System should have facilities to store atleast 100,000 images, should have the facility
of performing calculations and annotations on stored images. System should have
inbuilt CD / DVD writer for image transfer.
15. User Interface
PC – based operating principles.
Task light with back-lit illumination of control panel.

HLL/PCD/ESIC-20/10-11 Page 66 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

System should have on screen menu for easy and immediate access to imaging
controls. Live image should be supported with side by side reference / stored images
on the Flat panel monitor.

16. Power requirements


System should operative between 200 – 240 VAC, 50 Hz. Maximum power
consumption should not exceed 700 VA
17. Required Accessories

· Convex Probe 2 – 5 MHz (Harmonic Imaging should be available)


· Linear Probe 3.0 – 9 MHz (Harmonic Imaging should be available)
· Endovaginal probe 4-8 Mhz

18. Accessories
· Thermal printer
· Inkjet Printer

Item No. 25

Infant Resuscitator

Manometer Range: 20 to 80 cm H2O (-20 to 7.8 kPa)


Maximum Pressure relief should be 5 to 70 cm h2O at 8 LPM
Peak inspiratory Pressure (PIP) should be 5 to 70 cm H2O at 8 LPM
Positive-end expiratory pressure PEEP) should be 1 to 9 cm H2O at 8 LPM.
Gas inlet flow range:-
5 LPM (min) to 15 LPM (max) If the gas inlet flow rate increases from 5 to 15 LPM, the
peak inspiratory
pressure typically increases approximately 8cm H2O (0.8 kPa).
Delivered oxygen Concentration: - Up to 100% depending on gas supply
Operating time (400 L cylinder):- @ 8 LPM 50 minutes
Recommended body weight: - Up to 10 kg (22 lb)
Should be supplied with all related tubing’s and consumable (Preferable reusable) for
at least 50
cases (application)

Item No. 26

NEONATAL MONITOR

Patient monitor system should be of modular type and capable of monitoring adult,
pediatric & neonatal patients.
Monitor should have 17” independent flat panel display.
Touch screen user interface .
Module rack / housing should be independent and shall be able to be placed near to
the patient.
Should be capable of 8 traces display.
Monitor must be capable of simultaneously monitoring the following parameters which
should be present as standard: ECG, NIBP, SpO2, invasive pressures (2),
temperatures (2)
Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to
be offered as optional

HLL/PCD/ESIC-20/10-11 Page 67 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in
arrhythmia monitoring on all leads.
Inbuilt ST segment analysis and arrhythmia detection for all the leads should be
possible.
Haemodynamic and drug dose calculations should be available.
Arrhythmia should be grouped based on classifications – and should show no of
arrhythmias occurred.
Respiration should be available with Cardio Vascular Artifact filter.
ICP monitoring should be possible.
Alarm parameter should flash red in the presence of high priority alarms (e.g.
ventricular fibrillation and asystole) and flash yellow in the presence of medium or low
priority alarms (e.g. noisy signal, etc.)
24 hours trend data should be displayed.
All monitors including central station should have similar user interface for usage
among all clinicians.
Monitor shall provide the capability to interact with alarms at remote bedsides.
Monitor shall provide the capability to receive and display real-time waveforms, trended
data and alarm status from other bedside or telemetry units on the patient monitoring
network.
Monitor shall provide the capability enter patient information at the bedside or central
monitor.
On-screen keyboard for entering this data is preferable. Should have USB ports to
connect mouse, key board, bar code scanner.
Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and
actual parameter alarm settings must be displayed on-screen when alarms are on.
Position of the displayed waveforms must be user configurable.
Waveform color changing should be user configurable.
Monitor shall permit the optional ability to receive and display information from other
patient devices such as ventilators, infusion pumps and other standalone devices.
All modules should be compatible with all monitors quoted.
Bed to bed communication between the monitors should be possible with out a central
station.
Networking to central station should be possible and price of central station should be
offered as optional
Patient monitoring network shall use standard TCP/IP protocol and be capable of
residing on hospital’s network infra-structure.
Should be compatible with HIS and should be HL7 compliant.
Monitor should provide remote viewing of real time waveforms through internet.
Patient monitoring network shall be able to support up to 1,000 monitoring nodes.
Should be supplied with necessary accessories for adult , pediatric and neonatal
accessories.
Accessories and spares
1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor
2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor
3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal
sensor with cable per monitor
4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface
cable per monitor
5. Temperature: Skin and nasopharyngeal probes per monitor.
The equipment should be CE & US FDA Approved.

HLL/PCD/ESIC-20/10-11 Page 68 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 27
Ventilator Critical Care

Microprocessor controlled Ventilator for Neonates, Paediatrics and Adults patient with
invasive & non-
Invasive ventilation in both pressure and volume based modes.
Should be expandable and up gradable.
Should have the both pressure & flow trigger sensitivity.
Minimum of following Modes of ventilation should be present: -
• CAMV – controlled Assisted mechanical ventilation
• SIMV – with pressure and volume support mode (VS)
• Pressure controlled ventilation
• Tube compensation
• PRVC
• BIPAP/Bi-level or equivalent with pressure support
• CPAP
• PAV+ (Proportional Assist Ventilation) or equivalent mode
Should have following Parameters: -
Tidal volume: 5 to 2000 ml
• Frequency: 2 to 150 b/m
• Pressure support: 0 to 70 cmH2O
• Inspiratory time 0.2 – 8 sec.
• Inspiratory flow: 3-150 L/min.
• Inspiratory pressure : 5-90 cmH2O
• Exhalation Sensitivity: 1-80% of Spont. Peak Flow
• Oxygen cone. :- 21 to 100%
• PEEP / CPAP : 0-45cmH2O
Following parameters should be monitored:-
a) Volume: Exp. Tidal volume & M.V
b) Tinsp.
c) Frequency
d) FiO2
e) Pressure: peak, plateau, peep, mean
f) Resistance and compliance
g) Ti/ Total & RSBI, P0.1 and Vital capacity
Should have user programmable Apnea back-up & should have detection of severe
patient occlusion to
protect patient against excessive airway pressure, terminate normal ventilation and
allow patient to
exhale through inspiratory limb by opening safety valve.
Should have at least 10” size integrated colored touch screen.
Should display: -
a) Wave forms: P x t, f x t, v x t
b) Loops: p x v, f x v
c) Should have different color for different breath
Alarms: - Audio visual Alarms for low air pressure, low oxygen pressure, low and high
inspiratory
pressure, low and high rate, leak rate, disconnection, apnea alarm time 15-60 sec, low
battery etc.
Should have reusable auto cleavable heated bacterial filter exhalation isolation system
Inbuilt Battery back up for ventilator for at least 30 min.
Essential Accessories:-
A) Reusable heated bacterial filter/cassette exhalation isolation system- 10no.
B) Humidifier- Heated temperature controlled, preferably temperature monitored with
alarms- 1 no.
C) Reusable humidifier chamber- 01no

HLL/PCD/ESIC-20/10-11 Page 69 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

D) Heater wires – 01no.


E) Heater wire adapter- 01no.
F) Flow sensor if applicable with flow sensor cables- 10no.
G) Reusable Breathing Circuits adult = 05no.
H) Reusable breathing Circuits pediatrics = 05no.
I) Reusable breathing circuits neonatal = 01no.
J) Nebulizer-
Ultrasonic nebulizer with pore size up to 2 micro meters to deliver medicament. It
should not
affect Ventilator parameter delivered to pt. When in use & it has both timed and
Continuous
nebulisation mode.
K) Compressor
a. Should be of same make as of ventilator.
b. Should be oil free, medical grade and silent (less then 60db at 1/meter) and flow
upto 160LPM.
c. Should have high temperature and low-pressure alarms.
d. Compatible to be connected to compressed air from hospital central gas supply.
L) Trolley
The equipment should be US FDA Approved.

Item No. 28

Open Care System


It should micro processed control and Temperature range should be 25’C to 38’C
Temperature indication range 20’C to 45’C
Power Adjustment (Manual Mode) 0-100%
Display resolution Skin Temp. Should be 0.1’C
Control Precision (skin Temp.) +/- 0.2’C
Alarms: - Power failure
High temp. - +1’C +/- 0.2’C
Low temp. - -1’C +/- 0.2’C
Safety High temp. 15 Min. on Maximum power
Probe failure, dislodged skin probe
Manual Mode – Every 10 Min.
It should have indication for Heating Power bar graph, Sound inhibid, manual mode,
servo mode.
It should have metabolic scale with capacity 7Kg display resolution 2gm.
Heater power 560 W
It should have rotary reflector 180◦
Voltage 220 V 50/60 Hz
Height till mattress area should be 1000mm (+/-100mm)*
Mattress dimensions Approx. 565mm X 810mm
Bed Inclination grade up to 12◦
It should have height adjustment mortised
It should have X ray cassette tray.
It should have 4 drawers with tray.
It should have observation lamp.
It should have blender set for O2 support with two flow meters with one humidification
flask and O2 mask with 1.5 meter hose.

HLL/PCD/ESIC-20/10-11 Page 70 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

It should have aspirator with vacuum meter mounted on the front panel.
It should have gas panel with two O2 and two compressed air and one vacuum outlet,
ABNT slandered.
It should have I.V. Pole.
It should have upgar timer 0 to 10 Min.
It should have tray for monitor

Item No. 29

TRANSPORT INCUBATOR
· Built-in membrane pump & receiver bottle complete within the incubator chassis,
with ECG, SPO2, NIBP monitor.
· Oxygen flow meter & 1 lit O2 cylinder with clear front panel display of pressure
completely integrated within the incubator.
· Manual temp. setting with over temp (40º C) audible & visual alarm.
· Internal lighting
· Removable infant tray.
· Special straps to fix the transport incubator securely to all stretchers in both
ambulance & helicopters.
· Harness to secure the infant on the tray.
· Power supply: Main supply and dry battery (12V) or mains power supply of
ambulance/automobile/helicopter.
· Should have a 25x10 mm rail at the back of transport incubator.
· Relative humidity up to 60% C distilled water.
· Weight with 1 lit O2 tank not to exceed 25 kgs.
· Unit should be completely detachable from trolley for putting in plane.
· Unit should be complete with all accessories and startup kit, extra dry battery for
12 hours work, probes electrodes etc.

Item No. 30

Plaster cutter with Aspirator

1.The plaster cutter hand piece should be ergonomic design.


2. The cutter should have two speed control and with less noise
3.the cutter should have impact resistant housing and it should be versatile i.e
can be used with vacuum or directly plugged into the wall/point.
4.the vacuum should have mobile on 5 castors for stability and transportation.
5.the vacuum should have facility to connect the hose with the cast cutter.
6.locking mechanism should be there in the system to lock the hose with the cutter.
7.the length of the hose should be 8 feet.
8.one complete system should be supplied with following items/accessories-
I. Cast cutter hand piece-1 no.
II. Cast cutter blades-20 nos(different sizes)
III. Cast vacuum with hose and mobile stand-1 no.
IV. Blade mounting toll-2 nos
V. Small cast spreader-1 no
VI. Large cast spreader-1 no.

HLL/PCD/ESIC-20/10-11 Page 71 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 31

Tourniquet Electronic

1. Electrically operated system with two hoses and battery back up of upto 45
min.
2. Should have option for bier"s block and bilateral procedures.
3. Seven sizes 8'' to 42" of cylindrical cuffs should meet individual requirements
of thin and fat patient for arm and thigh. Cuffs should be autoclavable.
4. Small and light weight unit that can be mounted on i.v pole or placed on a
table.
5. Audible and visual alarms when pressure variation is detected
6. Unit should perform self-calibration check every time it is turn on.
7. 50 to 475 mm hg pressure setting
8. 6 to 240 minutes time setting.
9. Positive locking connectors, leak free inflation cuff sleeves, which help
reduce wrinkles pinching and shearing of soft tissues.
10. Cuff with lockout features, which should ensure that the cuffs stays inflated

Item No. 32

HOT PACK UNIT

· Must be full stainless steel unit with wheels with option to easily change water
and heating capacity of upto 12 large packs at one time and temperature preset at 167
degree F.
· Should have a PVC coated and low-water cutoff feature with special insulate to
preserve heat and help conserve power consumption
· Must come with built in adjustable thermostats and provide lighted on/off
switches
· Should have concelead elements, coved bottoms, coated racks and hospital-
grade power cord
· Unit must have inside mesh and hangers to hold the pads for proper stable
heating to transfer
· Unit should be provided with different size white clay pads and able to hold
temperature upto 30minutes of deep moist heat.
· Hot Packs able to be reused for hundreds of treatments and provide 12 months
of warranty
· Should come with durable custom size and shapes Terry covers for hot packs
with different sizes for cervical, lower back, extremities and more
· UL-listed, ETL/CE and CSA –approved with 220 volts

Item No. 33

COLD PACK UNIT

· Must be Five Cubic Feet of Storage and able to hold 12 Gel packs
· Should have adjustable thermostatic control and drain for defrosting
· Dimension approx 27” deep, 34 “ high has to be a cooler and not a freezer

HLL/PCD/ESIC-20/10-11 Page 72 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

· Have to provide compressed cold therapy pack for extremities able to 360
degree around the injured area made out of durable Nylon outer chamber.
· Must provide Body ice packs with non-freezing gel
· Must be made out of PVC Vinyl exterior and available in different sizes for
different body parts cervical, lumbar, and extremities
· Should able to hold temperature up to 30 minutes
· UL-listed, ETL/CE and CSA –approved with 220 volts option available

Item No. 34

Traction Unit

· Must be FDA approved decompression-traction unit


· Should able to deliver decompression therapy along with 3-channel of light
therapy to muscle stimulation and blood circulation
· Should able to deliver light therapy treatment and decompression
simultaneously.
· Should have a touch screen interactive display for treatment set-ups and easy
angle selection and must come with treatment protocol manual
· Must provide along with the package Angle reference chart
· Should able to automatically calculate and digitally display the rope pull angle for
decompression and traction as per the treatment protocol
· Must provide a 8”x 10’ Infrared light Pad unit for muscles relaxing and muscle
spasms
· Must have built in training information for traction and light therapy protocols
· Should provide protocol manual for light therapy and lumbar and cervical
protocol manual
· Must be a build in computerized software package and protection against
accidental setting of force-must have a safety switch for emergency shut of
· Should come with Flexion stool, Knee bolsters, Cervical pillow, Ankle bolsters,
and decompression belts thoracic and pelvic.
· Should provide with a 4 section motorized table hi/lo with clamps, frame
attachments for connecting the traction unit
· Should be a certified class device with all CE mark and US FDA approved Unit,
and must provide proprietary certificate and IEC 60601-1(CE) and CSA/NRTL

Item No. 35

Operating Laparoscope Set for Gynaec

Laparoscopy TROLLEY
Automatic Light source
220 V,300 W. Xenon Bulb
Bulb Working life more than 500 Hrs.
Bulb life counter on light source
Stand By Mode via push button on light source console
Universal Jaw Assembly to adapt any make of fiber optic cable without use of any adapters.
Fiber optic Cable
High Definition Monitor

HLL/PCD/ESIC-20/10-11 Page 73 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

High definition monitor, resolution 1280 x 1024 DVI input.


Option for wall mounting and desktop in same unit.
CO2 Insufflator
Minimum 40 Liter of high flow
Microprocessor controlled unit
Soft Approach Pressure control for safe recovery of abdominal pressure
Gas heating
HD Camera System
Camera console with universal coupler & Autoclavable camera head
Pure Digital signal with high definition video(1280*1024 native resolution)
Progressive scan technology both on camera head & console
Digital zoom & white balance on camera head
Video Output
1. 2 DVI output
2. 2 SVHS & 1 RGB out put
3. One Composite out put
Ideal Eye/HD Laparoscopes, Fully Autoclavable with working length 300mm
Wide angled distortion free view
Universal adaptor for other light cable
High quality sapphires and the latest rod lens technology for excellent transmission and detail
recognition
30 degree, 10mm x 1
0 degree, 10mm x 1
0 degree, 5mm x 1
Instrument set:
Laparoscopic hand instruments (reusable) and rotatable
Veress needle 12 cm length 1
Veress needle 15 cm length 1
Trocars sleeves 11 mm 3
Reducer 11/5 mm 2
Trocars sleeves 5.5 mm 3
Trocars (pyramidal tip) 10 mm 2
Trocars (pyramidal tip) 5 mm 2
Maryland dissector 5mm with unipolar diathermy 1
Atraumatic graspers, 5mm 2
Tooth Grasper 5mm 2
Metzenbaum scissors (5cm) with unipolar diathermy 2
Laproscopic cautery lead 2
Laproscopic bowel grasper 5mm, length 33-36 cm 1
Laproscopic suction cannuala, 10 mm 1
Laparoscopic suction cannula 5 mm 1
Bipolar Forceps 5mm Cilpenger type 1
Bipolar cable 1
Kelly Dissecting Forceps Long Jaws 5mm 1
Fenestrated Long Jaw Forceps 5mm 1
Dissecting & Grasping Forceps for ovaries 5mm 1
PCOD Dissecting Needle 5mm 1
All the Units should be US FDA Approved
Cidex Tray (Indian)-2
Formalin Chamber (Indian)-2

HLL/PCD/ESIC-20/10-11 Page 74 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 36
HYSTEROSCOPY SYSTEM

Hysteroscopy set – telescope 30 deg.


Diameter 4mm length 30 cm, Autoclavable x 1 No.
2. Examination sheath Diameter approx 5.4 mm x 1 No.
3. Continuous flow operating sheath approx 6mm x 1 No.
4. Operating sheath approx 5.4 mm x 1 No.
5. Deleted
6. Biopsy forceps, trough cutting single action
jaws, approx 5FR x 2 No.
7. Scissor, signal action jaws, approx 5FR x 2 No.
8. Biopsy and grasping forceps approx 5FR x 2 No.
9. Hysteroscope pump for Diagnostic & operative Hysteroscopy, Roller Pump working
220 – 240
VAC, pressure pre – Selection : - 150 - 200mm Hg, Flow regulation: 0 – 500ml/mm.
Equipments
should be supplied along with set for at least 300 procedures.
x 1 No.
10. Suction Unit with following Essential accessories:
i. One pedal foot switch one stage x 1 No.
ii. Suction bottle 51 x 1 No.
iii. Bottle cap x 1 No.
iv. Bottle stand x 1 No.
v. Bottle stand holder x 1 No.
RESECTOSCOPE
To be compatible with Hysteroscope mentioned above.
1. Working element set consisting of x 1 set
i. One working element set.
ii. Two cutting loops.
iii. Two coagulating electrode.
iv. Two high frequency cord.
v. One protection tube.
2. Resectoscope sheath including connecting tubes
For in and out flow, approx 26 FR oblique beak,
Inner tube with ceramic insulation, rotatable x 1 No.
3. 3. Standard obturator for use with approx 26FR sheath x 1 No.

Item No. 37

Cryo Surgical Unit

1. Facility to cutoff gas line without closing cylinder while inter changing probes.
2. Twin Trigger system to frost and defrost.
3. Cleaning the inner hypodermic needle should be much easier
4. System should be operated by CO2 or NO2 gas.
5. Console to house cylinder with temperature indicator
6. Probes Tips: Exo Cervical, Endo Exo Endo cervical and long Endo Exo Cervial,
HPV, Flat and Round in different shapes.
7. System high quality and reliability.

HLL/PCD/ESIC-20/10-11 Page 75 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 38

Vacuum Extractor

1. Noiseless suction unit have fast vacuum built up


2. Vacuum should have maximum -90 k pa/-675 mm Hg, Suction capacity
50ltr/min/1kg/cm2 & bottle
capacity 3ltrs
3. Suction system should have piston / cylinder ( self lubricating )
4. Should have mechanical overflow protection system .
5. Set of silicon cups 50mm & 60mm-2each
6. Set of bird cups, stainless steel 40mm, 50mm & 60mm-2each
7. Machine can be operated on 220-240 V AC single phase
8. Should be provided with a gauge to display of vacuum generated
9. Should have vacuum release valve.

Item No. 39

Fetal Monitor

Cardiotocograph machine with twin monitoring capability should meet the following
specification and capabilities:
· FHR twin monitoring using external ultrasound
· Direct ECG and maternal ECG measurements.
· Uterine activity using an external toco transducer or IUP catheter.
· Fetal movement profile parameter to record accurately the fetal movements using the
ultrasound
channel without additional procedures or transducers and statistics for advance
information on
fetal well-being.
· Low ultrasound energy to the fetus.
· Audible alert indication of fetal bradycardia and tachycardia
· Audible indications of paper out and NST time complete.
· Should have a feature to provide more accurate and continuous fetal heart rate (FHR)
thereby
reducing the need for repositioning the ultrasound transducer.
· Should have the facility of cross channel verification when two channels are picking
up the same
signal.
· Should have signal quality indicators guiding to obtain the strongest and most
continuous
ultrasound HR signal.
· Built-in multi channel high resolution thermal array recorder with visual and audible
paper end
detection and should annotate time of day, date and paper.
Should be supplied with the following accessories:
· Mobile cart with two drawers and integrated mounting rail.
· 2 x ultrasound transducers
· 1 x external toco transducer
· 1 x ECG module with degl & MECG adapter cable
· 20 x 250 g bottle of gel.

HLL/PCD/ESIC-20/10-11 Page 76 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

· 100 numbers of disposable signal spiral fetal scalp electrodes, quick connect type
· 80 packs of recording paper. (to be supplied as per the usage; i.e. in a manner that
they should
not get faded without being used.)

Item No. 40

ENT Examination Unit

· Hose connection for irrigation of maxillary duct.


· Splash protector
· Water warmer, 37 deg with electric temperature control.
· Compressed air device, complete with 2 medicament sprays, one powder blower,
adapter and spray
holder
· Jet connetion for water irrigation handle, short, medium, and angles.
· Connection box for automatic switching on/off water supply by master switch.
· Mains supply unit, with connection box, and installation kit.
· Head light with exit source light.
· Cotton swabs dispenser.
· Tongue patches dispenser.
· Holder for ear funnels
· Water filter, complete for warm water equipment.
· Cold light source, with additional rear light outlet, for operating microscope.
· Operating microscope mounted on unit, with double pantograph arm, straight
binocular head.
· Endoscope holder and warmer.
· Air flow for cold light, to prevent misting
· Mirror pre warmer.
Instrument cabinet all steel casing comprising of:
· Suturing complete set
· 4 large instrument trays
· Aluminium/SS trays
· Shallow drawer
· Deep drawer
· Used instrument deposit
Swivel chair adjustable high with back rest
Should be supplied complete with treatment couch and compressed air & vacuum.

Item No. 41

Operating Laparoscope Set for Gen Surgery

Laparoscopy TROLLEY
Automatic Light source
220 V,300 W. Xenon Bulb
Bulb Working life more than 400 Hrs.
Bulb life counter on light source
Stand By Mode via push button on light source console
Universal Jaw Assembly to adapt any make of fiber optic cable without use of any adapters.
Fiber optic Cable
High Definition Monitor

HLL/PCD/ESIC-20/10-11 Page 77 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

High definition monitor, resolution 1280 x 1024 dvi input.


Option for wall mounting and desktop in same unit.
CO2 Insufflator
Minimum 40 Liter of high flow
Microprocessor controlled unit
Soft Approach Pressure control for safe recovery of abdominal pressure
Gas heating
HD Camera System
Camera console with universal coupler & Autoclavable camera head
Pure Digital signal with high definition video(1280*1024 native resolution)
Progressive scan technology both on camera head & console
Digital zoom & white balance on camera head
Video Output
1. 2 DVI output
2. 2 SVHS & 1 RGB out put
3. One Composite out put
High Definition/ Ideal Eye Laparoscopes, Fully Autoclavable with working length 300mm
Wide angled distortion free view
Universal adaptor for other light sources
High quality sapphires and the latest rod lens technology for excellent transmission and detail
recognition
0 degree, 10mm x 1
30 degree, 10mm x 1
0 degree, 5mm x 1

Instrument set:
Laparoscopic hand instruments (reusable) and rotatable
Veress needle 12 cm length 1
Veress needle 15 cm length 1
Trocars sleeves 11 mm 3
Reducer 11/5 mm 2
Trocars sleeves 5.5 mm 3
Trocars (pyramidal tip) 10 mm 2
Trocars (pyramidal tip) 5 mm 2
Maryland dissector 5mm with unipolar diathermy 2
Atraumatic graspers, 5mm 2
Tooth Grasper 5mm 2
Metzenbaum scissors (5cm) with unipolar diathermy 2
Laproscopic cautery lead 2
L shaped hook electrode 5mm 2
Laproscopic bowel grasper 5mm, length 33-36 cm 1
Laproscopic suction cannuala, 10 mm 1
Laparoscopic suction cannula 5 mm 1
Clip applicator 10 mm for Large, Medium Large, Small Clips
Claw Forcep 10mm 1
Babcock forcep 1
All the units should be US FDA Approved
Cidex Tray (Indian)-2
Formalin Chamber (Indian)-2

Item No. 42

Multipara Bediside Monitor

HLL/PCD/ESIC-20/10-11 Page 78 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Patient monitor system should be of modular type and capable of monitoring adult,
pediatric & neonatal patients.
Monitor should have 17” independent flat panel display.
Touch screen user interface .
Module rack / housing should be independent and shall be able to be placed near to
the patient.
Should be capable of 8 traces display.
Monitor must be capable of simultaneously monitoring the following parameters which
should be present as standard: ECG, NIBP, SpO2, invasive pressures (2),
temperatures (2)
Should be compatible with Capnography, Cardiac output, EEG, and BIS and prices to
be offered as optional
ECG should have capability for 3, 5 and / or 10 lead monitoring and should have built in
arrhythmia monitoring on all leads.
Inbuilt ST segment analysis and arrhythmia detection for all the leads should be
possible.
Haemodynamic and drug dose calculations should be available.
Arrhythmia should be grouped based on classifications – and should show no of
arrhythmias occurred.
Respiration should be available with Cardio Vascular Artifact filter.
ICP monitoring should be possible.
Alarm parameter should flash red in the presence of high priority alarms (e.g.
ventricular fibrillation and asystole) and flash yellow in the presence of medium or low
priority alarms (e.g. noisy signal, etc.)
24 hours trend data should be displayed.
All monitors including central station should have similar user interface for usage
among all clinicians.
Monitor shall provide the capability to interact with alarms at remote bedsides.
Monitor shall provide the capability to receive and display real-time waveforms, trended
data and alarm status from other bedside or telemetry units on the patient monitoring
network.
Monitor shall provide the capability enter patient information at the bedside or central
monitor.
On-screen keyboard for entering this data is preferable. Should have USB ports to
connect mouse, key board, bar code scanner.
Alarm limit status (ON/OFF) must be indicated on-screen for each parameter and
actual parameter alarm settings must be displayed on-screen when alarms are on.
Position of the displayed waveforms must be user configurable.
Waveform color changing should be user configurable.
Monitor shall permit the optional ability to receive and display information from other
patient devices such as ventilators, infusion pumps and other standalone devices.
All modules should be compatible with all monitors quoted.
Bed to bed communication between the monitors should be possible with out a central
station.
Networking to central station should be possible and price of central station should be
offered as optional
Patient monitoring network shall use standard TCP/IP protocol and be capable of
residing on hospital’s network infra-structure.
Should be compatible with HIS and should be HL7 compliant.
Monitor should provide remote viewing of real time waveforms through internet.
Patient monitoring network shall be able to support up to 1,000 monitoring nodes.
Should be supplied with necessary accessories for adult , pediatric and neonatal
accessories.
Accessories and spares
1. ECG / respiration: 5 lead ECG cable and lead wire set per monitor
2. NIBP: Adult: 2 sizes and Pediatric 2 sizes and neonatal, 1 size per monitor

HLL/PCD/ESIC-20/10-11 Page 79 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

3. SPo2 Sensor: Adult sensor with cable, pediatric sensor with cable and neonatal
sensor with cable per monitor
4. IBP: Include 10 nos of disposable pressure transducer with bracket and interface
cable per monitor
5. Temperature: Skin and nasopharyngeal probes per monitor.
The equipment should be CE & US FDA Approved.

Item No. 43

Emergency Resuscitation Kit (imported)

1. To have Retromolar Intubation fiberscope for unexpected difficult airways.


a. Tip Distal Bending 40°.
b. To be movable eyepiece
c. To have a light source connection
d. With length 40-42cms and dia 5-6 cms.
e. ET tube holder should be provided
f. Should take min. 5.5 size of ET tube
2. Portable LED light source should be provided
i. with illumination not less than 50000 Lux
ii. should run on two 3v photo batteries
iii. burning life should be more than 100
minutes
iv. ergonomically designed and can be
connected to both the fibrescopes
v. life of LED should be close to 50000 hrs
3. One Laryngoscope with rechargeable battery pack and blade with fibreoptic
mechanism should be provided to be used on both adult and pediatric patients with
charger.
4. Other accessories like, magill forceps should be provided.
5. Should have Emergency Cricothyroidotomy for pediatric and adult
i. disposable blades
ii. dialator
6. Should have Combitube size 37Fr.
i. with complete kit
7. Should have Intubating Laryngeal Mask Airways with Following Components:
a. ILMA Sizes 3 & 4.
b. ILMA Tubes ID 7mm & 7.5mm.
c. Tube Stabilizing rod
d. Cuff deflator
8. Should have Laryngeal Mask Airways
i. sizes 1,2 and 4
9. Handy and strong brief case/bag should be provided to keep all the instruments
safe.
10. Set of disposable percutaneous tracheotomy kit for adult and pediatric.
11. Should have standard AMBU bag for pediatric and adult.
12. Mechanical suction pump with suction catheter and stomach tubes.
13. Should have Aluminum Oxygen reservoir 2 Liter with oxygen tube and cather.
14. Oxygen pressure reducer, regulable 0-15 liter with coupler for respirator.
15. Ventilating bag
16. Lubricant
17. Blood pressure meter, boso K-II

HLL/PCD/ESIC-20/10-11 Page 80 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

18. Stethoscope
19. Rescue blanket gold/silver
20. Infusion system.

Item No 44

Portable Spirometer
• The unit should be small, portable, with a built in caring handle convenient for
Ward rounds and on site patient screening.
• The unit should Menu driven and operate on 220V AC
• The unit should incorporate a Fleisch type pneumotach autoclavable sensor with no moving
Parts.
• Should have built in storage up to many thousands of patients with their life time data for
individual or batch printing.
• The unit should incorporate with large touch screen displays with easy to follow instructions,
flow/volume & volume/time graph, post test, incentive display and an inbuilt printer.
• The unit should have facility to direct connectivity to PC and able to transfer the data from unit
to PC and printout facility through external printer.
• The unit should perform single breath test, VC, and closed circuit spirometry.
• The unit should measure the following parameter : VC, FVC, FEV.5, FEV.5%, FEV1,
FEV1/VC%, FEV1/FVC%, FEV/PEF, FEV3, PEF, FEF25-75, FEF25-75%, FEF25%,
FEF50%, FEF75%, FMFT, FET, MVVind, FIVC, FIVC/FVC%, PIF, FIF 25%, FIF50%, TV,
RV, IRV, ERV, IC, etc.
• The print out should be configurable with choice of curves, parameters printed, interpretations
and test quality control message .
• The equipment warranty should be of one year.
• The unit should be compliant to all standard like ATS 94, ERS 94, CE, FDA, and EN60601.

Item No 45

Infusion Pump Syringe

It should be Light weight , Compact & should be able to adopt syringes from 5 to 60ml
horoizontal
Programming : Should be ml/hr, ml/hr x volume limit & time x volume limit
Flow rate should be 450 ml/hr with 0.1ml/hr increment
Volume Limit should be 999.9 ml
Must have Occlusion Pressure at least 3 levels
KVO should be 1ml/hr or less
Bolus rate should be 450 ml/hr
Time limit from 1min to 99hrs & 59 minutes.
Special Functions Like : Clear Volume
Should have pre alarms such as End of Battery, End of Infusion, & visual & sound
alarms like Syringe
positioning error, occlusion, infusion completed, syringe error, programming error,
Block error, clamp error , low battery etc
Flow rate deviation should not be more than +/- 1% of the programmed flow rate.
Battery Rechargeable with battery life 3hrs , Should be able to work on 230 Volts
The product should have CE & ISO 13485 certifications

Item No 46

PFT Machine(Spirometer & Computerised)

HLL/PCD/ESIC-20/10-11 Page 81 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

The unit should be light weight, exclusively designed to carry out examination of
pulmonary system and to measure & display lung function parameters, operable on
USB port power , not requiring large installation space ( compact).

It should come with an incentive display for cooperation free, lung testing.

It should be designed for the following measurement protocols

Spirometery with pre/post comparison, Flow/Vol and Vol/Time loops/curves and Trend
Loops.
Individual interpretation assistance.
MVV
Patient Co-operation Display

It should measure following options without using additional gas analyzer

Capno-Volumetery measurement
Emphysema diagnostic
Determination of anatomical and functional dead spaces

The system should incorporate a light weight, Ultrasonic sensor (non-heated) free from
any kind of frictional inefficiencies which is absolutely insensitive to moisture. It should
not require any warm-up time and calibration with following measurement ranges :-

Flow Measurement
Range : 0 to +/- 16 Ltrs/sec with accuracy better than +/- 3% or 100 ml/sec .
Volume measurement
Range : 0 to 10 ltrs with resolution of 10 ml
Corrections
Flow Volume : EGKS or ATS
Inspiratory Gas qty : BTPS (ambient module)
The machine should measure the following Parameters:
SVC , V C ex, V C max, IRV, ERV, VT, IC, Frequency, MVV, , ti, te, ti/te, FVC, FEV1,
FEV 0.5, FEV 1/ VC max, FEV 1/ FVC ex, MEF25, MEF 25-75, PEF, PIF, MVV and
pre-post comparisons.
Should measure CO2 max %, Vm25-50, Vm50-75 etc.

The unit should be supplied with 50 no of disposable breathing tubes.

The unit should be supplied complete with hardware and software including a PC as
per below given specifications from a respected brand like IBM, Compaq or HP.

a) CPU – Pentium IV Core 2 DUO Processor.


b) RAM – 1 GB, At least 160 GB HDD
c) At least 14” TFT Color Monitor
d) DVD R/W Drive, Keyboard, Mouse
e) Serial / Parallel interface PCI SLOT
f) Windows XP
g) Inkjet Printer.
h) Suitable UPS for the computer system

The manufacturer should be an EN ISO accredited company and the product should be
duly CE marked to MDD 0124 for medical devices.

HLL/PCD/ESIC-20/10-11 Page 82 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No 47

Fully Automated Biochemistry Analyzer

Fully automated, latest and bench top analyzer to perform the analysis of substrates, enzymes
and special parameters from whole blood, serum, plasma and urine samples

System should be Discrete, fully selective random access with a provision to test STAT samples

System should have four different on-board technologies ( Photometry, Potentiometry,


Fluorescence Polarization and Turbidimetry ) to measure substrates, enzymes, Homogeneous
immunoassays, TDM’s and Drugs.

System should have facility for programming 125 - 150 different test parameters and the
reagents should be available from the same manufacturer.

System should have a routine throughput of 400 tests / hr

Onboard sample capacity should be at least 90 or more

Flexibility to use different sample containers like primary tubes with different sizes, sample cups,
micro cups and cup on tube for easy processing.

Facility to keep reagent bottles / cassettes for at least 30 common tests with on board
refrigeration is must.

Sample volumes should be less than 2 - 10 ul per test.

System should have high sensitive pressure sensors to detect any incorrect pipetting even at 2 ul
sample volume

Onboard sample and calibrator dilution should be available ( 1 – 100 times )

System must use disposable cuvettes to prevent any carryover without using any onboard
washing

System should be used for testing special parameters like HbA1c, Lactate, hsCRP, D-Dimer,
Ferritin, IgA, IgM, IgG, ASO, Cyclosporine, MPA and electrolytes ( Na, K and Cl ), TDM, DAT
tests besides the routine clinical parameters

On-board reagent stability should be for at least 3 months and calibration of the parameter
should be typically with lot. No daily calibration should be required by the system to save the
reagents.

System should have 12 wavelength photometer with mono and bi-chromatic measurements

Light source should be 20 W halogen lamp with lamp save feature

System should have external windows NT based data control work station with flat screen
monitor for programming the tests and entering the patient data.

System should external printer to take printout of patient results

Patient samples and Reagents can be scanned with barcode scanner for easy operation.

System should have 1 x RS 232 bidirectional interface and in-built modem for remote diagnostics
access.

HLL/PCD/ESIC-20/10-11 Page 83 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Power supply – 220 V / 50 Hz. Suitable online UPS should be supplied along with the system

Item No 48

Centrifuge Machine 24-36 tubes

1 Description of Function
1.1 Centrifuges are required in the Laboratory to separate various components of Blood
for analysis.
2 Operational Requirements
2.1 Aerodynamic compact construction for vibration free performance
2.2 Table top version
3 Technical Specifications
3.1 Angle Head rotor with Tube Capacity Size 5 – 15 ml
3.2 To have capacity to hold 24-36 tubes at a time.
3.3 Should have a digital timer
3.4 Body should be made of strong fabricated & corrosion resistant steel
3.5 Control panel – for start/stop switch, dynamic brakes, step less speed regulator with
zero
start switch & speed indicator with timer and protective fuses.
3.6 Door interlock
3.7 Maintenance-free brushless drive motor with exact speed pre selection and
alphanumerical
interactive LCD digital display in control panel of RPM & RCF. Speed range 100 to
6000
rpm and above.
3.8 Choice of acceleration and braking profiles.
3.9 Imbalance detection and auto shut down.
4 System Configuration Accessories, spares and consumables
4.1 Tube Holders as appropriate
5 Environmental factors
5.1 Shall meet IEC-60601-1-2:2001(Or Equivalent BIS) or General Requirements of
Safety for
Electromagnetic Compatibility.
5.2 The unit shall be capable of operating continuously in ambient temperature of 10 -
40deg C
and relative humidity of 15-90%
5.3 The unit shall be capable of being stored continuously in ambient temperature of 0 -
50deg C
and relative humidity of 15-90%
6 Power Supply
6.1 Power input to be 220-240VAC, 50Hz as appropriate fitted with Indian plug
7 Standards and Safety
7.1 The supplier should be preferably ISO certified for quality standards.
7.2 Product should preferably be FDA/CE or ISI approved
7.3 Should preferably comply with IEC/TR 61010-3-020 :Safety requirements for
electrical
equipment for measurement, control, and laboratory use - Part 3-020: Conformity
verification
report for IEC 61010-2-020:1992 Particular requirements for laboratory centrifuges"
7.4 Should preferably comply with IEC 61010-1 Safety requirements for electrical
equipment for
measurement, control, and laboratory use - Part 1: General requirements
8 Documentation
8.1 User manual in English

HLL/PCD/ESIC-20/10-11 Page 84 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

8.2 Service manual in English


8.3 Certificate of calibration and inspection from factory
8.4 List of Equipments available for providing calibration and routine maintenance
support as per
manufacturer documentation in service / technical manual
8.5 List of important spare parts and accessories with their part number and costing
Log book with instructions for daily, weekly, monthly and quarterly maintenance
checklist. The job
description of the hospital technician and company service engineer should be clearly
spelt out.

Item No 49

Computed Radiography System

Description of function
Radiography system to replace conventional Film/screen based X-Ray processing
techniques with Photostimulable Phosphor Plate technology based digital technology.

Operational requirements
The system shall be able to record X-Ray images on Imaging Plates(IP)
The IP Shall be housed in CR Cassettes that have a Radio frequency Tag to store
demographic data.
Operationally and functionally equivalent to and better than the present film based
system
Convert these images from the IP into digital values and transfer these values to an
image evaluation computer with predefined Image Processing Parameters.
Should record Paient Identification data on the on the image.
Maintain and manage data bank of all patient and image data.
Retrieve and reproduce accurate, high quality high resolution images from stored data
without loss of image quality.
Present CR images on a workstation as well as on hard copies.
show full image in the X-Ray room for preview purposes.
Read and Write in CD.DVD for data Storage and review.
Appropriate technology to provide uniform and thick slice thickness.

Technical Specifications
Image Reader will have the following :
Cassete Machanism to Load and Unload IP.
Scanning mechanism to read, erase and process the images from the imaging plate.
(IP)
Including autorouting newly acquired images to desired preview monitor.
-IP processing rate> 50 plates/hr.
-Panel for indicating online status of the CR Reader in case of machine malfunction.
-Emergency Mode for accepting exposed cassettes
without patient demographics for casualty trauma workflow requirements.
-Capability of retrieving at least 10 scanned images and quick check of the exam data
and the image of at least the last four cassettes scanned at the X-Ray room.
- Verification of the connectivity status of configured image desination.
-Spatial resolution of digital image 6-10 pixels/mm..
-Scanning resolution for all the IP Plates should be specification in the quote.
-Should enable 12 bit images.
-X-Ray Generator compatibility with reputed manufacturers..
-Image matrix at standard resolution (14 ´ 17) - 2000 ´ 2500 Row ´ Column
-Image matrix at standard resolution (14 ´ 17) - 3000 ´ 4000 Row ´ Column

HLL/PCD/ESIC-20/10-11 Page 85 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

CR Workstation will have the following.:


--Capable of Archieving and printing selected images to a standard DICOM destination
in DICOM 3.0 format.
--Sroring images in the local disk for predefined period.
-Mechanism for accepting new images when the local disc is full..
-Sorting of patient image based on name, date, exam etc.
-Advance Processing Software.
Using predefined parameters or user defined and stored image parameters.
-Correcting typographical in patient demographic module, in case RIS connection was
down and manual data entry was done.
-Capability of changing R/L, Flipping, Rotating, Zooming, Collimating, annotating the
incoming image..
-Multi image and slide formats.
-Capability of storing in CD/DVD.
-Software for Advance Image processing, applications, dispay and quality monitoring.
-Connectivity and compatibility to communicate to RIS/HIS and DICOM Compatible
devices such as MR/CT/DSA Work station,
-Shouild provide for HL-7 compatible interface.
-Output grayscale resolution- 16 bits/pixel.
Remote ID and Preview station. Should have the following:
-Auto detection of cassette.
-Mechanism of writing /reading data on/from RF Tag..
-HIS/RIS/DICOM Compatibility.
-Preview scanned images on predefined preview terminal.
-Retrieving capality of last 10 patient ID on the terminal.
-Identification of overxposure on preview module.
-Mechanism for user release in case of autorouting images. To predefined DICOM
destinations
-System should be able to support minimum 5 review terminals
-Preview display time <45 sec.

Color dry view imaging printer


(film based) with the following:
-Print Image from CR workstation. In DICOM 3 format.
-Mechanism to print images to 14x17 and 8x10 film sizes simultaneously.
-Docked in processor.
-Resolution> 600 DPI.
-Multiple Image and slide printing capability.
-Shall be able to swtich between Receiver Mode and Processor mode.

System Configuration Accessories, spares and consumables


IP/Cassettes size- qty
35 cmx35 cm 03
24 cmx30 cm 03
15 cmx30 cm 03
20 cmx25 cm 04
25 cmx30 cm 08
Image Reader system 01
CR Workstation 01
RIS Interface 01
Remote ID and Preview station. 01
Archieving System 01
(film based)

HLL/PCD/ESIC-20/10-11 Page 86 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Black and white thermal video printer


Black and white laserjet printer for reporting

Environmental factors
0
The unit shall be capable of operating continuously in ambient temperature of 30 C
and relative humidity of 80%

Power supply
Power input to be 220-240VAC, 50Hz, fitted with Indian plug
Resettable overcurrent breaker shall be fitted for protection
Spike protector of appropriate rating should be provided
UPS of suitable rating conforming to IS-302 shall be supplied

Standards and safety


Should be FDA or CE approaved product
Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450
Comprehensive guarantee for 5 years of complete system and all other parts for which
order will be placed with uptime warranty of 98%. Comprehensive Guarantee period
will be extended by double the downtime if it exceeds more than 2% in a year.
Rates of comprehensive AMC for complete system and all parts for which order will be
placed with an undertaking of 98% uptime and extension of AMC period by double the
downtime if it exceeds more than 2%.

Documentation
User manual in English
Service manual in English
List of important spare parts and accessories with their part number and costing.
Certificate of calibration and inspection from factory.
Log book with instruction for daily, weekly, monthly and quarterly maintenance checklist
.
The job description of the hospital technician and company service engineer should be
clearly spelt out

Item No 50

Advancved Colour Doppler Ultrasound Scanner

Description of function
Fir whole body Ultrasound Anatomical studies, Blood flow Studies and 3D studies.

Operational requirements
Digital Ultrasound system platform for excellent 2D, Colour & Power Doppler and 3D
Imaging capability.

Technical Specification
Must have minimum 1000 digital channels*
Should have 15" or more high resolution monitor with tilt and swivel.
Line density 512 lines

HLL/PCD/ESIC-20/10-11 Page 87 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Dynamic range>160dB
Penetration upto 30 cms
Upto 4 selectable frequencies in each probe
Tissue harmonic imaging with phase inversion, pulse inversion, or wide-bandwidth
imaging technology.
Inbuilt 3D imaging with hand acquisition and auto sweep.
Must have minimum of 4 rendering modes with measurements.
Panaromic imaging upto 240 cms or 180 deg with colour and cine facility.
Should automatically equalize gain and brightness with touch of one button.
Ability to enhance 2D and tissue harmonic penetration and colour sensitivity
momentarily to improve visualization in difficult patients.
Imaging with multiple line of sight combined to a single line of sight to imrove
resolution.**
Should provide for vascular imaging enhancing by using power Doppler to enhance B
Mode image.
Appropriate technology to provide uniform and thick slice thickness.
Software for various applications including Vascular, Abdomen, Foetal echo,
Transcranial and cardiac studies should be available.

Accessories, spares and consumables


1-4 MHz Convex Array probe.
5-13 MHZ Liner Array Probe
4-10 MHz Endocavity probe
Volume probe
PC Based Image management system
Color dry view imaging printer
(film based)
Black and white laserjet printer for reporting

Environmental factors
The unit shall be capable of operating continuously in ambient temperature of 300 C
and relative humidity of 80%

Power supply
Power input to be 220-240VAC, 50Hz, fitted with Indian plug.
Resettable overcurrent breaker shall be fitted for protection
Spike protector of appropriate rating should be provided
UPS of suitable rating conforming to IS-302 shall be supplied

Standards and safety


Should be FDA or CE approad product
Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450
Comprehensive guarantee for 5 years of complete system including all transducers and
all other parts for which order will be placed with uptime warranty of 98%.
Comprehensive Guarantee period will be extended by double the downtime if it
exceeds more than 2% in a year.
Rates of comprehensive AMC for complete system and all prats for which order will be
places with an undertaking of 98% uptime and extension of AMC period by double the
downtime if it exceeds more than 2%.

Documentation
User manual in English

HLL/PCD/ESIC-20/10-11 Page 88 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Service manual in English


List of important spare parts and accessories with their part number and costing.
Certificate of calibration and inspection from factory.
Log book with instruction for dail, weekly, monthly and quarterly mantenance checklist.
The job description of the holpital technician and company service engineer should be
clearly spelt out

Item No 51

Donor Chair

1. Description of Function
Blood Donor Couch is a completely automatic enveloping, variable tilt chair and
specially designed to make blood withdrawals easier, safe and functional, and also for
other diagnostic and therapeutic areas
2. Operational Requirements
1) Provides a comfortable position for the donor.
2) Variable positioning for either arm with Comfortably wide armrests.
3) Armrests have swinging out as well as up and down moving facility.
4) Reclining and upright body positions with a smooth shifting to any position.
5) Both sides have supporting brackets.
6) Drawers provided for the upkeep of equipment & consumables.
7) If a vasovagal attack occurs the Donor's head needs to be lowered immediately and
his legs lifted above his heart level so that blood can flow back to the brain and other
vital organs. This facility should be available
3. Technical Specifications
3.1 Comfortable chair type with soft padding for cushioning and rexin cover.
3.2 Seat, back rest and leg rest size designed for donor comfort. It should have step
less electric remote controlled height adjustment.
3.3 Adjustable arm rest for donor’s comfort and phlebotomist friendly
3.4 Easily tilted to head low position, electrically operated
3.5 Comfortable working level for the operator. Lifting capacity - Approx 200 kg.
3.6 4 Lockable castors for easy mobility
3.7 Storage Drawers for storing consumables & Blood Collection Monitors
3.8 UP/DOWN control
3.9 OPTIONS:
(i). A paper roll holder can be fixed on the' upper part of the chair.
(ii). Melodious musical Headphone can be integrated for patient relaxation while blood
donation is in progress.
(iii). Preferable to have inbuilt trays & stands for keeping all blood collection
accessories.
3.10 Should have interface for blood collection monitor (optional)
4. System Configuration Accessories, spares and consumables
4.1 Donor Couch -01
4.2 Dust Cover -01
4.3 Power cable -01
4.4 Arm Rests (pair) -01 pair
4.5 Remote control -01
5. Environmental factors
5.1 The unit shall be capable of operating continuously in ambient temperature of 10 –
400 C and relative humidity of 15-90%
5.2 The unit shall be capable of being stored continuously in ambient temperature of 0 -
40 C and relative humidity of 15-90%
5.3 Shall meet IEC-60601-1-2: 2001 (Or Equivalent BIS) General Requirements of
Safety for Electromagnetic Compatibility.

HLL/PCD/ESIC-20/10-11 Page 89 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

6. Power Supply
6.1 Power input: 220-240V/ 50 Hz AC Single phase or 380-400V AC 50 Hz Three
phase fitted with appropriate Indian plugs and sockets.
6.2 Resettable over current breaker shall be fitted for protection
6.3 Suitable Servo controlled Stabilizer/CVT
Standards and Safety
7.1 Should be FDA or CE approved product
7.2 Electrical safety conforms to standards for electrical safety IEC-60601 / IS-13450
7.3 Manufacturer should have ISO certification for quality standards.
7.4 All electrical actuators and mechanisms should be housed inside the structure
making the product safer
7.5 Comprehensive warranty for 2 years and 5 years AMC after warranty
8. Documentation
8.1 User manual in English
8.2 Service manual in English
8.3 List of important spare parts and accessories with their part number and costing.
8.4 Certificate of Calibration and inspection from the factory
8.5 Log book with instructions for daily, weekly, monthly and quarterly maintenance
checklist. The job description of the hospital technician and company service engineer
should be clearly spelt out.
8.6 List of Equipments available for providing calibration and routine maintenance
support as per manufacturer documentation in service / technical manual.
8.7 Original Information Brochure should be provided.

Item No 52

Blood Collection Monitor

• Weighing range 100—999ml


• Automatic tare to zero for the bag weight.
• Adjustable low and high flow alarms.
• Adjustable donation time out up to 20 minutes.
• Adjustable default volume.
• Automatic clamp of tubing at the end of the
donation.
• Weighing accuracy +/- 2%.
• Power supply 115/230 VAC 50/60Hz
• Power consumption Max 10VA
• Dimensions - 290(L) X 253(W)X 150(H)mm
• Weigh: about approx 5kg incl battery
· Should have a data memory of approx 30000 characters
· Provision to attach bar code reader(optional) for capturing external data related to
donor
· Internal fuses to be PTC-self recovery to ensure continuous operations
· Automatic Calibration Feature

Item No 53

Blood Bank Refrigerator

General
Should operate at 4 ° C with +/- 1 °C temperature uniformity
Must be designed for blood bank use. Commercial or modified commercial refrigerators
are not
acceptable

HLL/PCD/ESIC-20/10-11 Page 90 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Should be able to pass through standard door heights of 201 cm (79”) with casters
without requiring
the unit to be tipped or laid on its side.
Should be CE marked.
Construction
Should have an interior and exterior that is constructed of minimum 20 gauge,
galvanized steel.
Should incorporate)bacteria resistant, powder coated interior, exterior and door handle.
Should have a minimum non-CFC urethane insulation
Should utilize a self-closing door with full-length handle, key lock and non-CFC
urethane insulation.
Should include swivel-locking casters as a standard feature.
Should incorporate a recessed interior floor to contain spills.
Should have a chamber access port in the top of the unit.
Should have an interior fluorescent light with control panel mounted switch as a
standard feature.
Should have a light bulb that can be changed without removing the drawers.
Should have dual-pane, glass door and key lock with a right hand hinge.
Should have self-closing door system
Independent Temperature Controller
Should utilize an independent, microprocessor temperature controller that is
programmable from +2o
C to +8o C.
Should have a Microprocessor controlled temperature readout, readable in 0.1o C
increments.
Should have a stainless steel, RTD temperature probe that is located in the chamber.
Should have all functions accessible through a touch pad on the control panel.
Should have refrigeration system “On” indicator provided as a standard feature.
Independent Alarm / Monitor System
Should be able to program the high and low temperature alarms.
Should have audible and visual high and low temperature alarms as a standard
feature.
Should have a stainless steel RTD temperature probe located in the top portion of the
chamber in a
product simulation bottle.
Should have audible and visual door ajar alarm as a standard feature.Must have all
functions
accessible through a touch pad on the control panel. Should have an alarm silence
button.
Should have alarm disable switch. Should have remote alarm contacts as a standard
feature.
Should have battery backup with a minimum of 2 hours life.
Should have a power fail alarm as a standard feature.
1.5 Temperature Recorder
Must have four inch, 7-day, ink-less, pressure-sensitive circular chart recorder.
Must have chart recorder temperature range of -5oC to +20oC.
Must incorporate a separate battery backup to ensure continuous operation of the chart
recorder
during power failure.
Must have temperature recorder probe that is independent from other probes.
Must have power status indicator.
Must have an optional deduction of the chart recorder.
. Should have LCD Temprature Graphs that should display 24 continuous hours of data
and
event logging of door openings and alarm conditions.
. Should have adjustable alarm volumes and password protected configurations.
Refrigeration System
Must incorporate a heavy-duty, air-cooled refrigeration system designed to operate on
230 volt

HLL/PCD/ESIC-20/10-11 Page 91 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

50/60 Hz.
Must utilize non-CFC, commercially available refrigerant.
Must have an automatic condense evaporator as a standard feature.
Must have an internal evaporator fan that shuts off when the door is opened.
Must have a compressor that can maintain required chamber temperatures when
operating between
200-240 volts and 50 Hz.
Must incorporate a defrost system that requires no defrost timer, electric heaters or
defrost down
time.
Must keep the refrigerator free of frost without elevating the chamber temperature.
Drawers
Must have solid bottom and liquid tight stainless steel drawers for containment of spills
Must incorporate Scratch-Guard drawer edge protectors that keep the glass from being
scratched.
Must have fully extendable drawer slides.
Must have shelf standards with a clear powder-coated finish to guard against rust and
corrosion.
Must have drawers that are adjustable
Must have optional drawer dividers available as an accessory.
Must have a cabinet designed to accommodate available optional half-size wire
shelves, full-size
wire shelves and rollout wire baskets with no cabinet modifications.
Electrical
External transformers are not acceptable

Item No 54

Tube Sealer

1. Should be a hand held sealer for apheresis, Stem cell, leucoreduction processeas
and should
have anywhere mobility for multiple application.
2. Should be supplied with one power source, hand held sealing head and one NICD
rechargeable battery pack.
3. Should do 1000 seals per fully charged battery pack and battery should fully be
charged
within two hours.
4. Should be a smart sealer to adjust for different sized tubing.
5. Tear seal feature to make segments that can be separated by hand.
6. Should be certified for patient connected use.
7. Lightweight and compact for ease of mobility. Total weight approx.3kgs.
8. Should have PTC overload protection internal fuses to ensure continuous operations

Item No 55

Cell Counter - 3 Parameter

Fully automated Hematology Cell Counter 18 Parameters with Histogram for


3parameters, should have highest linearity of PLT more than 1.8million/micro litter to
measure PLT Values in Platelet Reach Plasma (PRP) :-

INSTRUMENT SHOULD HAVE THE FOLLOWING ESSENTIAL FEATURES

HLL/PCD/ESIC-20/10-11 Page 92 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

TECHNOLOGY
: Electronic Impedance Principle for counting and HGB by photometer.

PARAMETERS
: WBC with LYM%, LYM#, MON%, MON#, GRA%, GRA#, RBC, HGB, HCT,
MCV, MCH, MCHC, RDW, PLT, MPV, PDW & PCT.

HISTOGRAM
: RBC, WBC, PLT DISPLAY AND PRINT-OUT.

CALIBRATION /QC.
: L-J graph for quality control mode with possibility to perform
automatic & manual calibration .

MESURING SYSTEM
: Dual channel measurement.

VALVES
: Minimum number of Maintenance Free Valve.

FLUIDICS
: Shortest fluidics to minimize maintenance and contamination.

APPERTURE CLEANING
: Automatic Back Flush in case of clog.

HGB MEASUREMENT
: By mean of LED with automatic blanking.

SAMPLE VOLUME/MODE
: Less than 20MCL in whole blood mode.

SAMPLING NEEDLE
: With automatic wash from inside and outside.

THRU-PUT
: 60Sample/Hour.

DISPLAY/OPERATION
: LCD Graphic Display with Touch Screen or Keypad operation.

MEMORY
: AT LEAST LAST 1000 Result memory with Histogram.

COMMUNICATION PORT
: RS-232C Serial Port or USB Port

PRINTER
: Built-in or External Printer.

REAGENT CONSUMPTION
: Technical Bid must show each Reagents consumed in ML as:-
Start-Up, Shut Down: Diluent -? Lyse -? In ML, If any other reagent needed to be
mentioned.
Per Cycle: Reagent Consumed in ML such as Diluent-?
,Lyse-? Or any other reagent has to be mentioned.

HLL/PCD/ESIC-20/10-11 Page 93 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Instrument must be supplied with suitable UPS & Start up reagents for at least
1000Samples.

Item No 56

Refrigerated Centrifuge

1 For separation of blood components like packed cells, platelet rich plasma, platelet
concentrate,
Cryprecipitate & Buffy Coat
2 Programmable memories with tamper proof facility
3 Rotor head with swing out buckets, windshielded.
4 Oval Shaped metal buckets with removable plastic oval cups to hold double/triple/
quadruple
blood bags
5 Refrigerated centrifuge with suitable rotor head & buckets should
accommodate/process
minimum 8 blood bags of 450 ml capacity simultaneously
6 Manufacturer to indicate volume & capacity of plastic bucket
7 Centrifugal Force -- Max ceiling 5000g
8 Microprocessor Controlled rotor speed to with in 10 rpm of set value
9 Different Acceleration & Deceleration profiles should be available
10 Temperature range 0 deg C to + 40 deg C, Microprocessor controlled rotor
temperature with
in +/- 1deg C regardless of centrifuge speed
11 Programmable time should be from 0-99 minutes with minimum revolution of 1
minute.
12 Digital display of temperature, speed & time & other relevant parameters
13 Automatic shut down of centrifuge , if rotor load is out of balance with indicator
14 Safety Key lock to prevent unauthorized use
15 The equipment should have lockable castors
16 The Refrigerated centrifuge should have built in memory to store atleast 50 data of
centrifugation during the processing of Blood Components. The data from the
centrifuge can
be transferred to a computer using a RS 232 data interface
17 Power Supply Requirement :- 220-240 Volts 50 Hz Single Phase
18 Servo Controlled heavy duty line voltage corrector should form part of the standard
configuration
19 A line voltage corrector of appropriate rating giving all the specifications should be
supplied
along with the unit
20 Should be a CE marked product & from a ISO , WHO-GMP compliant manufacturer.

Item No 57

Plasma Expressor(E)

Product certification & approval


CE mark and S mark as per EN 61010-1 standard
Ensures safety against electrical shock hazards, fire hazards, mechanical hazards,
electromagnetic interference etc
Manufacturing standard
Manufactured at ISO 9001:2000 Certified facility
Power source
AC - DC power adapter supplied along with the equipment or power from another

HLL/PCD/ESIC-20/10-11 Page 94 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Electronic Plasma Expressor of same make


Input for AC-DC power adapter
120-240VAC, 50/60 Hz, 0.4A
Plasma Expressor input
12VDC, 1A
Power Consumption
12W (Max)
Dimensions
Approx.(WxDxH) in mm
165 x 230 x 280
Sensor type
Infrared beam interrupt opto-sensor
Bag pressing mechanism
Stainless steel spring loaded transparent acrylic sheet
Clamping power source
Spring loaded clamp with geared DC motor
Clamp closing time
Less than 1sec
Flow rate
180 - 210ml/minute
Controls
Power Switch, Clamp Up, Clamp Down, Start
Indication lamps
Power, Start , Clamp Down
Modes of operation
Manual & electronic modes
No. of electronic plasma connected using a single power supply
4 Nos
Interface
Provision for interfacing with CopmoScale for separating a definite quantity of plasma
to
the satellite bag

Item No 58

Dielectric Tube Sealer Table Top

Bench top model.


1. Compatible with all blood bags.
2. No warm up time, automatic sealing when tube is in position.
3. Seal minimum 2 mm wide.
4. Easy to clean electrodes, easily accessible & protected by cover.
5. 220-240 volts.
6. Indication for sealing progress.

Item No 59

PLATELET INCUBATOR & AGITATOR

PLATELET INCUBATOR
PLATELET INCUBATOR SHOULD HAVE THE PROVISION TO STORE THE
AGITATOR FOR 48
PLATELET BAGS AGITATOR.
SHOULD HAVE CLEAR VIEW SINGLE PANE TEMPERED GLASS
AGITATOR SHOULD STOP AUTOMATICALLY ONCE THE DOOR IS OPENED.

HLL/PCD/ESIC-20/10-11 Page 95 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

SHOULD HAVE MICROPROCESSED CONTROLLED LED


DISPLAY,TEMPERATURE GRAPH
DISPLAY,
SHOULD HAVE STAINLESSSTEEL RTD SENSOR PROBES
SHOULD HAVE PROVISION FOR 4”7DAY INKLESS CHART RECORDER WITH
BATTERY
BACKUP FOR CONTINEOUS OPERATION DURING POWER FAILURE.
SHOULD HAVE ALL CONTROLS IN ONE CONVENIENT LOCATION INCLUDING
CHART
RECORDER AND ALARM KEY
SHOULD BE ABLE TO MAINTAIN A TEMPERATURE OF 22 DEGREES WITH +_.
1DEGREES
VARIATION.
PLATELET AGITATOR
SHOULD BE ABLE TO STORE MINIMUM 48 RANDOM PLATELET BAGS OR
APHERESIS BAGS
OR BAGS OF DIFFERENT SIZES. With GENTLE SIDE TO SIDE MOTION (1 ½”
38MM)
SHOULD HAVE SINGLE FAN FOR FORCED AIR CIRCULATION.
SHOULD BE STURDY ONE PIECE DRAWERS WITH HOLES FOR COMPELETE AIR
CIRCULATION ACROSS BOTH SURFACES OF PALATELET BAGS
Should be CE marked

Item No 60

AUTOMATED PLASMA THAWING EQUIPMENT

1. SHOULD BE ABLE TO THAW 8 – 12 PLASMA BAGS


(FFP/APHERESISOR PLASMA BAGS OF ANY SIZE or ANY MAKE)
2. SHOULD HAVE WATER BATH BASED SYSTEM WHICH SHOULD BE
OPERATIONAL AT
4 DEGREE TEMPERATURE TO 37 DEGREE CELCUIS PRECISELY.
3. SHOULD BE COMPACT IN SIZE.
4. SHOULD HAVE RACK HOLDERS WITH BUILT-IN FINGERS FOR SECURELY
HOLDING
THE PLASMA BAGS OF ALL SIZES.
5. SHOULD HAVE AN ALARM WHEN THE PLASMA BAGS ARE THAWED
6. SHOULD HAVE THE PROVISION FOR SELECTING PROGRAMMED TIME
SETTING FOR
THE LENGTH OF THAWING CYCLE. .
7. SHOULD HAVER DIGITAL TIMER CLEARLY DISPLAYINGTHE PROGRAMMED
SET TIME
OR REMAINING CYCLE IN MINUTES
8. SHOULD HAVE ALARM SYSTEM FOR ADJUSTABLE OVERTEMPERATURE
ALARM
SETTING, AUDIBLE AND VISUAL ALARM WARNINGS,
9. SHOULD HAVE A DEEP THAWING CHAMBER FOR INCREASED HEAT
TRANSFER
EFFICIENCIES, WHICH RESULTS IN FASTER FFP THAWING TIMES. THE CLEAN
STREAMLINED DESIGN OF THE HEAVY GAUGE STAINLESS STEEL CHAMBER
SIMPLIFIES
ROUTINE CLEANING.
10. SHOULD HAVE A CHAMBER DRAIN SYSTEM WITH A HIGH FLOW RATE TO
DRAIN THE
CHAMBER WITHIN 2-3MINUTS.
11. SHOULD HAVE A TERMPRATURE CONTROLLER
12. SHOULD HAVE A SERVO CONTROLLED VOLTAGE STABILIZER OF ATLEAST

HLL/PCD/ESIC-20/10-11 Page 96 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

3KVA
13. Should be CE approved

Item No 61

LAMINAR FLOW- VERTICAL

Hepa Filter : 99.999 %efficiency for particles >0.3 µm


Pre-Filter : 85 %efficiency for particles >0.5 µm
Particle Count : Better than US Fed Std 209B Class10 and VDI 2083 Class 3
Cabinet : Laminated High Quality Wooden Board
Work Table : AISI 304 Stainless Steel
Airflow Speed Control : Speed Controller (Three Step Speed Controller)
Blower : High efficient centrifugal type with lifetime lubricated bearings
Light : High intensity,low wattage >800 lux
Noise Level : <55 dBA
Standard Accessories : Air/gas cock and .mains power socket (16A)
Power Supply : 220-230 V,50 Hz.
Power Consumption : 400 w
Internal Work Space: 600mmx600mmx600mm
900mmx600mmx600mm

Item No 62

Deep Freezer -40 Deg.

VOLTAGE: 220V/50Hz Capacity 310 Litres


- TEMP: -20 TO -40degC, MICROPROCESSOR CONTROL
- ACCESS PORT
- HEAVY-DUTY SWIVEL CASTERS
- HEAVY-GAUGE STEEL CABINET WITH LONG-LASTING POWDER PAINT
FINISH
- SINGLE DOOR +KEYLOCK, DIGITAL DISPLAY
- REFRIGERATING FLUID & INSULATION: CFC FREE.
- ROUNDED INTERIOR CORNERS
- EYE-LEVEL CONTROLS
- TEMP: ALARM
- MICROPROCESSOR ALARM
- BATTERY BACK-UP
- AUDIBLE/VISUAL OVERTEMPERATURE ALARM
- LOW VOLTAGE BOOSTER
- AIR INSULATED INNER DOORS
- 3 ADJUSTABLE Shelves
- BASE MOUNTED CONTROLS
- MULTI-POINT GASKET SEALS
- AUTOMATIC VOLTAGE BOOSTER
- SET POINT SECURITY SYSTEM
- INDEPENDENT OPERATING TEMPERATURE AND HIGH/LOW LIMIT
ALARM
- AUTOMATIC VOLTAGE BOOST
- ON-BOARD MONITORING
- AIR-COOLED CASCADE REFRIGERATION SYSTEM
- EFFICIENT DOWNFEED EVAPORATOR
- HIGH CAPACITY AIR-COOLED CONDENSER

HLL/PCD/ESIC-20/10-11 Page 97 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

- HEATED DOOR SEALS (HOT GAS) MINIMIZES FROST BUILD-UP


AROUND DOOR GASKET
- DOOR LATCH CAM ACTION WITH KEY LOCK, ONE HAND OPERATION
- DOOR HANDLE FULL LENGTH FOR EASY ACCESS
- CHART RECORDER 7 DAY 6 ”
- SERVO CONTROLLED VOLTAGE STABILIZER OF 3 KVA SHOULD BE
INCLUDED

Item No 63

Deep Freezer -80 Deg.

The Deep Freezer should achieve low temperature of -80degC


Should be heavy duty refrigeration system, maintenance free, with hermetically sealed
refrigeration
compressors reliable refrigeration with minimum noise & vibration.
Integrated digital temperature cum controller with inbuilt 7 days circular chart recorder
with battery
back up of 3-4 hours to ensure display & recording of temperature even during power
failures.
Construction of double wall with efficient insulation to minimize temerature loss, inner
chamber should
be made of AISI 304 grade non corrosive stainless steel & outer made of high quality
C/R sheet
Servo controlled line voltage corrector of 5 KVAshould form part of the standard
configuration.
Should be a CE marked product & from a ISO, WHO-GMP compliant manufacturer.
The Deep Freezer should be vertical upright with a capacity to hold minimum 185-200
plasma bags

Item No 64

Cell Separator

Should be Fully automatic, Microprocessor controlled with easy access operator control
panel and
has a large touch screen.
Should Perform both Single and Double needle Apheresis.
During single needle procedure the equipment shpuld continue to process the whole
blood during
the return cycle to reduce the procedure time and increase the efficiency.
Should have auto elutriation separation technique to be able to collect platelets faster
with high
collection efficiency to help process less blood volume.
The disposable kit for the machine should be True Closed system disposable with pre-
attached
factory fitted ACD, Normal Saline and Needles as required.
Equipment should ensure all donor safety parameters before starting the procedure
and all time
during operation.
Should be Capable of collecting various single donor blood components including
Peripheral Blood
Stem Cells (MNC). Should be able to collect both single and double needle Platelet
apheresis along
with concurrent Plasma and/or RBC
Should be capable of fully automatic PBSC (MNC) collection .. Should be Capable of

HLL/PCD/ESIC-20/10-11 Page 98 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

doing Prime
only with Normal Saline and / or mixture of Normal Saline and ACD.
Should have a Inbuilt Cuff pressure and prompt grip for donor comfort and adequate
blood flow.
Should have a Facility to use platelet additive solution and / or normal saline for re-
suspension and
storage fluid in place of plasma.
Machine should have a Advance help menu available at any time during alarm
conditions.
Extra corporeal volume should not be more than, 21 Oml & 205ml in case of both
single and double
needle apheresis respectively ..
Should be equipped with a Yield estimator to help decide yield, volume to be
processed and
suggested storage fluid and should have a optical sensor at PRP line for online
monitoring of
component collection against the desired yield.
Should be Capable of downloading or printing full procedure report any time after
procedure
Should be Capable to connect bar code reader if desired.
Should have rechargeable battery to store data and restart in case of power failure.
Should have Continuous monitoring of collection to avoid any contaminations through
Interface
detector.
Inlet and return flow rates should be up to 100 ml./min.
The separation of blood in the machine should be able to automatically maintain a
constant
hematocrit to improve collection efficiency and reduce contamination

Item No 65

ICU Ventilator (High End)

1. Latest state of art microprocessor technology based on the Biphasic Ventilation


concept.
2. Can provide respiratory support to Adults, Pediatric & Infant patients with following
modes,
CMV (A/C), SIMV, PCV, PCMV, PSIMV, SPONT, NIV, ASV, APRV, BIPAP, APVcmv,
APVsimv etc.
3. Have fully Automatic Mode ASV (Adaptive support Ventilation) effective for Passive
as well for active patients
4. Have Pulmonary Function Assessments (P-V Tools) provision to have idea for
optimum PEEP.
5. Have Special functions like TRC (Tube Resistance compensation), Manual Breath,
Stand by & Sigh etc.
6. Can be upgraded for ETCO2 Monitoring.
7. Can be upgraded for Heliox Therapy Option.
8. Have in- built color TFT Touch screen size 15 inches for the display of the various
lung parameters, curves (Volume vs time, flow vs time, pressure vs time) & Loops (P-V,
F-V, P-F).
DISPLAY: display Real time visualizations of Lungs with representations of Tidal
Volume, Lung Compliance, resistance and patient activity along with Ventilator
dependency grouped into oxygenation, CO2 elimination, Simultaneous display of at
least 4 waveforms or 2 loops based on volume, flow, airway pressure, auxiliary
pressure or CO2.
9. Continuously monitors 37 parameters including Rapid shallow index (RSBI),
Occlusion pressure (P01), Static compliance, leakage, RC exp etc.

HLL/PCD/ESIC-20/10-11 Page 99 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

10. Have the following user selectable parameters:


Tidal Volume: 10-2000 ml
I:E Ratio : 1:9 to 4:1,
Automatic Base flow: 4-30 L / min.
P ramp: 25-200 ms
Flow: 0-180 L/min.
Apnea time: 15-60 s
Expiratory Trigger Sensitivity: 5-70% of insp. Peak flow
Rate: 1-120 b/min
Flow Patterns: Sine, Square, 100% decelerating, 50 decelerating.
11. Have all the alarm setting for the various parameters with Altitude compensation up
to 3000 m. (9843 ft).
12. Supplied with the reusable autoclavable patient circuits for Adult, pediatric & Infant
patients separately.
13. Supplied with inbuilt Nebulizer & Blender and have internal battery backup of 1
hours.
14. Have the 1, 12 & 24 hours trends for the all the monitored parameters.

15. Have Storage of 1000 events with the time stamp.


16. Have upgradation for Interface connecting facility like USB, Compact Flash for
screen shots & DVI with VGA Output.

Optional Accessories:
External Compressor

A) Non Invasive Pressure Support Ventilation:

1. Spontaneous / CP AP mode.
2. IPAP Range: 2-25 cm H2o with increment of 0.2 cm.
3. EPAP Range: 2-25 cm H2o with increment of 0.2 cm.
4. Reusable Full face masks.
5. Variable rise time available for weaning away patient from invasive Ventilator.
6. Respiratory Back-up upto 10 BPM.
7. Adjustable Trigger and Cycle sensitivity - High, medium and low sensitivity setting.
8. Vsync automatically minimizes any loss of pressure delivery due to leak.

B) Diagnostic Capabilities:

1. Display Respiratory Rate on breath by breath basis.


2. Display Tidal Volumes.
3. Displays Masks or Circuit Leaks.
4. Displays Delivered Pressure.
5. Alarms for Mask off.

C) Patient Safety Mechanics:

1. Inspiratary time can be setting allowing critical patient to breathe out.


2. IPAP Max Range (0.1-4 sec)
3. IPAP min available to force patient to stay at IPAP pressure for user defined time
interval. IPAP Min Range (0.1-IPAP Max)
4. The Rise Time from EPAP to IPAP can be set and varied (Min, 150-900ms).
5. The world’s most comfortable Full Face masks available.
6. Ramp Time available for 45 minutes.
7. 365 nights patient compliance data can be recorded.

HLL/PCD/ESIC-20/10-11 Page 100 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

8. Hour meter to display total number of hours used.


9. Humidification option available, Humidifier optional-Passover or Heated.
10. Automatic smart start ON/OFF; Patient Triggered or manual.

D) Machine Fault Protection and other Mechanics:

1. SMPS for voltage regulation 95-295V built-in.


2. Power Supply Input Range 100-240V, 50-60Hz, <100 VA.
3. Carrying case included.
4. Alarms available

Item No 66

ABG Analyzer

Fully Automatic Microprocessor Controlled Latest Blood Gas Analyser with all the
Necessary accessories having following features
Measured Parameters
pH, pCO2, pO2, Barometric Pressure
+ +
Na , K , CI-, Ca2+
Hematocrit (Hct)
tHb and SaO2

Calculated Parameters

Up to 30 calculated parameters can be programmed


pH at patient temp
pCO2 at patient temp
pO2 at patient temp
Base Excess (BEecf)
BE at actual SaO2 (BEact)
Buffer base (BB)
Actual bicarbonate (HCO3)
Standard Bicarbonate (ST HCO3)
Total CO2 (ctCO2-B & P)
Standard pH (ST pH)
Oxygen Content (cO2)
Hydrogen lon Concentration (cH+ )
AaDO2
Respiratory Indes (RI)
P50
Anion Gap (AG)
FIO2
Qs / Qt
Sample throughput - 30 / hour
Fully automatic Liquid calibration of all parameters at fixes/user defined intervals. No
calibration gases required
Miniature type maintenance free electrodes
Individual switch off / on facility of whole module or parameter
Data display on built in flat color 6” LCD display screen (touch screen)
Data printout on fast low noise thermal printer with graphics capability

HLL/PCD/ESIC-20/10-11 Page 101 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Storage of data on built in Microprocessor


Optional PCMCIA Card for data Storage
Built in Barcode-by facility for economical operation
Low consumption of reagents during the operation
Interfaces 2´Rs 232, 1´Ethemet Connection, 1´Barcode Scanner
Built in voltage stabilizer for the voltage range from 100-240 VAC/50Hz Only 3 types of
reagents Required for calibration & Measurement. Reagent Onboard Shelf life should
be minimum 45 days
Should be FDA Approved
Activating & deactivating parameters by only key-stroke

Item No 67

Arthroscopy System

Camera system
1. Camera console 220 v three chip with universal coupler & autoclavable camera head
2. Pure digital signal with high definition video(1280x1024 native resolution)
3. Resolution-2000 horizontal lines
4. 8 specialty settings
5. Integrated flexible scope filter
6. Signal to noise ratio-70 db
7. Progressive scan technology both on camera head & console
8. Brightness control on console & camera head
9. Aperture control on console
10. Inbuilt 16 step digital image enhancer on console
11. Digital zoom & white balance on camera head
12. Integrated gain/shutter/enhancement with brightness control
13. Two peripheral control on camera head
Xenon light source 300 watts
Light source xenon, 300 watts lamp,color temperature 6000 k, universal
jaw for accepting any make
fiber optic cable adjustable light intensity from 0 to 100 percent, Standby mode;
Fiber optic cable
High definition monitor 21"
High definition monitor, screen 21”, resolution 1280 x 1024 dvi input.
Option for wall mounting and desktop in same unit.
Arthroscope set
Telescope 4mm, 30 deg
quick latch / speed lock dual connect arthroscope sheath 5.9mm.
Obturator for sheath
hook probe
straight punch, 3.4mm
15 deg upbiter 3.4mm
3.4mm 30 deg right cutting punch
3.4mm 90 deg left cutting punch
3.4mm 90 deg right cutting punch
3.4mm foreign body grasper with lock
Shaver system
HLL/PCD/ESIC-20/10-11 Page 102 of 166 01.09.2010
HLL Lifecare Limited Technical Specifications

Electronic control unit


Foot switch
Connecting cable for large bone handpiece
Shaver Handpiece autoclavable rpm 12000
Microdriver with drilling, wiring, micro saw and reciprocating saw
attachment
Heavy Duty (Drill & Reamer) and Sagittal Saw handpieces for large bone
(Arthroplasty)
Jacob's chuck,reamer (for IM & Acetabular reaming) and pin driver
attachment
Aggressive cutter (6 nos.)
Full radious resector ( 6 nos.)
Oval bur ( 6 nos.)
All the items should be from one company
The equipment should be U.S. FDA approved
Imaging movable trolley – Indian

Item No 68

Color Doppler Ultrasound Unit

The system should be state of the art with full digital beam former technology and
should be for whole body application which would include Cardiac, Obs/Gyn,
abdominal, peripheral vascular. Small parts Imaging such as Ophthalmic Imaging,
Thyroid, Intracavity application, etc.
Essential features :
· The system should have 256 Greyscale or more.
· The system should have a dynamic range of 150dB or more.
· System should have advanced image processing algorithms to reduce the
speckle and artifacts for improved image quality.
· Should have Integrated Flat Panel Display with min. 17” screen size with tilt and
swivel facility
General requirements :
1. The system should incorporate facility for high facility for high resolution 2D,M
mode, PW, color flow imaging, power Doppler angio imaging modes.
2. All transducers should be multi frequency broad band technology for High
resolution 2D and Colour Doppler imaging.
3. Zoom should be both read and write type up to 30 times
4. Imaging Depth should be atleast 32 cms or more.
5. Presets : Minimum 50
6. The system should have harmonic imaging for tissues for hard to image
patients, All transducer should have Tissue Harmonic Imaging as standard
7. The system should have a full alphanumeric keyboard with illuminated keys and
status display.
8. System should have cine loop review facility in individual and mixed modes with
memory upto minimum 300 images and 10 seconds of Doppler / M Mode Cine
9. Should have facility of Thumbnail Menu, Support thumbnail images and digital
storage and retrieval of B/W & color image data (both frozen and cine loop) on built-in
and removable media (Min. 40GH HDD, DVD, CD-R/W
10. The system should have automatic real time quantification of Doppler
parameters like velocity, frequency, time heart rate slope, flow volume, pulsatility index,
jpeak velocity, average value, point value, area and diameter flow volume etc.

HLL/PCD/ESIC-20/10-11 Page 103 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

11. Directional Power Doppler angio imaging for perfusion studies should be
available for visualization of flow in small vessels.
12. The system should have extensive calculation software package for generic
measurement, cardiac, Ob/Gyn, vascular etc.
13. Should have direct connectivity to inkjet colour printer for printing images and
report.
14. Equipment with above features to be offered with the following broad bandwidth
probes
a. Broad band convex array transducer with frequency range 2-5MHz
b. Broad band linear array probe with frequency range between 2-4 MHz
c. Broad band phased array probe with frequency range between 2-4 MHz
d. Broad band TV/TR Probe with frequency Range between 5-8 MHz, (Quote
optionally)
15. The system should have the following documentation devices and accessories:
a. B&W Thermal Printing, Color Inkjet Printer and suitable UPS to be provided

Item No 69

Video Bronchoscopy System

(A) THERAPEUTIC VIDEO BRONCHOSCOPY SYSTEM SHOULD HAVE


FOLLOWING ITEMS:
1. Video Bronchoscope
2. Light Source (Xenon)
3. Video Processor
4. Monitor
5. Bronchical Accessories
Scope Features:
1. Remote Switche on the scope
2. Biopsy port below the control section
3. Light weight

Video Broncohscope Should:

Have latest CCD chip Technology


Be of high resolution with maximum Pixels
Be compatible with electrosurgical instruments & Laser with ceramic tip
Have grip fit design for single hand operation
Have autoclavable suction valves for reduction of crass infection possibility
Be fully immersible

Video Broncohscope specification:

CCD Type: Colour chip


Field of view: 115 - 125 Deg
Depth of view: 3-100 mm or better
Distal end: Outer diameter: 5.8 mm - 5.9 mm
Insertion tube: Outer diameter: 5.8 mm -6 mm
Bending section: Angulation range: Up 180 Deg, Down 130 Deg
Working length: 590 mm - 600 mm
Total length: 860 mm - 900 mm
Instrument channel: Inner diameter: 2.8 mm or more

(B) LIGHT SOURCE


1. Should be independent / integrated with processor

HLL/PCD/ESIC-20/10-11 Page 104 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

2. Illumination: Main Lamp-xenon, Emergency lamp as a stand by lamp with automatic


switch over Lamp Life indicator

(C) VIDEO PROCESSOR:

Should be a latest clour chip system with following features:


Should be independent / Integrated from light source
Should be light weight
Should be capable of storing at least 40 patient’s data
Compatible with different scopes including paediatric for future
upgradtion
Video output: RGB, B NC and Y/C
(D) MONITOR: Colour Monitor grade (14”) compatible with this system
(E) Computer / software: Compatible computer system with recording
and printing facility for providing reports patient / storing for future
intervention / reference
(F) Endoscopic trolley with good caster wheels with breaks, hanger to
hang scope
(G) Bronchical Accessories: Compatible with video Bronchoscope
Biopsy Forcep - 2 No.
Biopsy forcep with needle - 2 No.
TBNA Needle - 5 No.
Cytoloty - 5 No.
Foreign Body Rat tooth - 2 No.
Foreign Body Basket - 2 No.
Tamponade Balloon - 1 No.
Injector - 2 No.

Item No 70

High Frequency Mobile X- Ray Unit

X-RAY GENERATOR:
The generator should be High Frequency with output of more than 3 kw.
It should work on constant 40 khz frequency output.
The x-ray generator should have LCD display of KV, mA and mAs.
The radiograph KV should be from 40 ti 100 KV.
The radiograph mA should be 70.
The esxposure time should be in mAs from 0.2 to 320mAs.
The generator should have be stationary anode with focal spot of 1.5mm.
It should operate on single phase 230Volts.
It should be user friendly with facility to select factors based on body anatomical chart.
It should be compatible with Digital X-ray system of CR system.
It should have facility to store the X- ray factors as per the use into the memory.

STAND:
It should be ergonomically designed spring balanced stand.
It should have hand brake for locking the over all machine.
It should have easy movement or ramp, lift and form corridor.
It should have parking position of tube arm in extreme lower position.

HLL/PCD/ESIC-20/10-11 Page 105 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No 71

Digital Mamography System

Description of function
1.1 Mammography system to replace conventional Film/Screen based Mammography
Studies with digital imag
2.1 Full Digital Mammography System consisting of exposure stand with attached swivel
system, separate console with radiation shield, automatic exposure control and
mammography X-Ray Tube.
2.2 An integrated direct-to-digital Flat Detector based on amaorphous silicon technology.
2.3 A separate workstation for image positioning and patient demographic data is
required.
2.4 The workstation should be able to send, receive and print according to DICOM
standards.
2.5 The workstation should also be able to obtain DICOM modality, work list from
connected information system and send information about performed procedure to the
connected information system
2.6 Read and Write in CD/DVD for data Storage and review.
Technical Specifications
3.1 Mammography System 01
The system should consists of a tube head and detector assembly that has isocentric
rotation for every positioning.
The iso-centric movements should be motorized. The patient Compresion device
should have automatic multi-speed variable compression system which senses the
breast density and adjust the compression force.
The maximum compression thickness should be 18 cm or more.
The patient table should have motorised grid movement.
Magnification devices of ratio 1.5 and 1.8 should offered.
Digital display of compression force and compression thickness should be available.
3.2 X-Ray Generator and Tube
The X-ray generator should be high frequency with the following parameters:
kV range: at least 25-35 kV in steps of 1 kV
mAS range: 0-750 mAS or more
Exposure time: 0-700ms
Maximum mA: 180mA or more
X-Ray tube unit:
Dual focus rotating anode tube with the following parameters:
focal spot size: 0.1mm and 0.3mm
Anode heat storage: 150 kHU or more
Tube heat Storage: 1.3MHU or more
Anode material: Molybdenum and Tungsten
Please mention the filter material used in the tube
3.3 Flat Panel Detector:
Type of detector: Amorphour selenium preferred
Detector size: 24cmx29cm or more with two image formats
Pixel size: 70i or less
Image matrix in pixels: large size-3Kx 4K or more Small size: 2Kx 3K or more
3.4 Workstation for image Acquisition:
The workstation should enable immediate image display for general survey
for patient positioning. It should be able to store around 10000 images. The
networking should on TCPIP protocol.
The following image processing should be possible on the workstation:
Image display:

HLL/PCD/ESIC-20/10-11 Page 106 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Freely selectable screen layout


Windows settings (contrast and brightness setting)
Magnification, stepped and dynamic zoom
Image inversion (black/white)
Annotation:
Left/right marking
Text additions
Lines
Rectangles and circles
Measurements:
Distance
Angle
Density
Image evaluation:
contrast enhancement(with table)
Display of histogram
Length measurements
Before /after comparison
Filter
Administration:
The demographic patient data should be retrieved directly from a
HIS/RIS system
The demographic patient data can be entered manually
Retrieval of images from CD, DVD or PACS
- Printing of images on DICOM – compatible printers
The workstation should be fully DICOM compatible
High Contrast 1Kx 1K TFT monitor should be provided with workstation.

Item No 72

Holter Recorders and Analyzers

1 It should have FDA clearance and CE certification


2 It should meet and exceed the requirements of ANSI / AAMI EC 38
Holter Recorders
1 It should be small and lightweight recorder
2 It should be capable of simultaneous real time acquisition of 3
channel and 12 channel recording.
3 Recorder should have LCD display to preview ECG waveforms during
the patient hook up and have lead quality check function.
4 it should have capability of continuous recording for 24 hours and
48 hours / 7 days.
5 It should have at least 24 hours with single battery.
6 The recorder should be water resistant.
7 Recorder should be battery operated and have single AA/AAA
battery or rechargeable in built sealed battery with recharging unit
supplied.
8 It should have internal memory for 99 full disclosure readings.
9 it should have advanced signal processing algorithms to provide
superior accuracy in beat detection, labeling and noise rejection.
10 It should be a 12 channel analyzing system.
11 it should have capacity to download 24 hours data through USB port
/ standard connection port in less than 3 minutes.
12 Device should be defibrillator protected.

HLL/PCD/ESIC-20/10-11 Page 107 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

13 Recorder should have 1000s/sec/channel digital sampling rate for


standard recording and internal storage.
14 Recorder should have compact flash card memory card.
15 it should be capable of pacemaker spike detection.
16 it is preferable if it supports voice recording capability.
17 It should include 2 sets of electrodes and patient cable to enable 3
and 12 channel recording.
18 It should come in 2 sets of patient hook up pouch and hook up
accessories.
19 Should include compact flash card as required for each recorder
with adequate capacity to store entire recording cycle.
20 It should include cable for connecting and downloading data to PC.
21 It should include compatible flash card reader.
22 It should include users guide and technical manual as well as quick
guide and patient hook up instruction posters for easy reference
and use.
23 It should include 100 sets of patient event recording diary and pens.
24 It should have two years warranty on recorder / memory card and 6
months for cable.
Analysis software + Hardware.
1 It should have multiple scanning options like retrospective,
prospective and superimposition scanning modes, event and
template review and be customizable.
2 It should have facility to convert final report into PDF and XML
format that enables connectivity to data management systems.
3 It should have trend graphs for HR, RR interval, RR variance, 12
lead ST, SVPB, VPB etc and provide graphic display options.
4 It should have different beat classifications.
5 It should have various rhythm analysis and ST measurements.
6 It should have 3 and 12 lead ST segment measurement analysis.
7 It should be capable of Pacemaker analysis including atrial,
ventricular and dual chamber pacing and under / over sensing and
capture failure.
8 It should support customizable report format including patient
data, 24 hour profile, selective printing of rhythm strips and trends,
summary statistics in tabular and narrative format and support user
defined acronyms for comments.
9 The PC + Printer provided should be Intel Pentium IV 2.5 GHz and
above and include minimum hard disk of 250 GB / DVD+RW drive or
Blue Ray disc writer / USB port/4GB RAM/10"TFT color monitor,
Laser printer. Should have one year on site warranty.
10 It should have capacity to archive into hard disk and DVD+RW /
High density DVD / Blue Ray disk.
11 It should include licensed operating software for ready to use status
along with reliable anti virus software. These should have back up
provided on CD / DVD.
12 It should include external data back up / archiving facility and
hardware.
13 All trainer manuals should be supplied and on site installation to
complete satisfactions of end user.
14 Training of 2 technicians to operate and use the equipment and
software to be provided at the company expense.
15 Assurance of 95% uptime and prompt provision of service back up
for hardware and software with replacement to maintain
operational equipment as required.

HLL/PCD/ESIC-20/10-11 Page 108 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No 73

Stress Test System(TMT)

Stress testing system should be complete with PC, Software, Tread Mill, Patient
acqusition module and necessary cables.
System should be based on windows platform with 17" color monitor having minimum
resolution of 1280 x 1024, 80 GB HDD, CD-RW, Mouse, UPS for analyser.
System should acquire and analyze 12 leads and should store the full 12 lead resting
/ stress ECG
Should show a pictogram and indicate which electrode has a bad contact before
resting / Stress recording.
Should be able to review retrospective average beat and ST values.
Should have user settable protocols.
Patient module should have USB connectivity.
System should provide standard Full interpretation of Supine ECG with reasoning.
Should provide display of real time 12 lead diagnostic quality ECG waveform,
average complexes beat of all 12 leads with superimposed comparison along with ST
level & slope and ST trend graph.
Automatic detecton, display, storage and review of arrhythmia, heart rate. Should
have running trends of ST available on screen during the test process.
Should have ability of comprehensive auto-measurement package including RR
intervals, P-wave duration, PQ interval and QRS width.
System should have high quality filters for muscle and baseline noise without
influence on the ST results.
Should have alarm levels for ST, HR and Blood Pressure.
Should have keyboard short cuts.
Capability to pause ECG screen view to find a past ECG event without loosing track
of current real time ECG.
Capability to insert event marks during stress.
Capability to change ST points during stress.
Should have manual measurement cursors for P-wave, QRS and T-wave. Should allow
to measure width (P, QRS, T-wave) and intervals (PR, QT etc.,)
Should have alarm levels for ST, HR and Blood Pressure.
Should be able to export QRS intervals from 12 lead Rhythm ECG, Stress / Resting
ECG report in word / RTF format and shluld be able to export raw ECG data.
Printing:
Should be able to print strips during stress or after.
Strips to be printed in 3 channel,s, 2x6 channels, 12 channels.
Print reports should be with summary table, event markers and stage information.
Should have average beat report with ST measurements.
Should have Trend graphs
System should provide multiple and customizable printing formats as per users
choice on A4 size normal plain printing paper.
Heavy duty treadmill - imported. Noise free with speed ranging from 0.8 to 19 kph
and have an elevation range of 0 to 25%.
Treadmill should provide smooth and safe operation.
Should have automatic belt alignment.
Treadmill should have 175 kgs weight(approx.) capacity with all metal chasis.
Should have emergency stop button.
Other Technical specs:
ECG sampling rate: 1000 Hz
ECG input voltage: 16mm Vpp
Noise voltage: 20uV

HLL/PCD/ESIC-20/10-11 Page 109 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

10 lead patient cable


Treadmill interface
Sugnal frequency range: 0.05Hz to 250 Hz.
Windows operating system.
PC Pentium latest workstation.
All consummables required for installation and standardisation of system to be part
of the system.
Should have US FDA and CE certifications
Display of following parameters in all stages of exercise & recovery:-
1. METs
2. Heart Rate
3. Target Heart Rate with % of completion Heart Rate
4. ST measurement mode (Manual/Auto)
5. Current ST level & ST slope
6. Exercise protocol stage name
7. "Switch-off" facility for patient
8. Automatic BP monitoring system

Item No 74

Volumetric Infusion Pump

Flow rate should be from 0.1 to 990 ml/hr with 0.1ml/hr increment
It should have choice of Micro & Macro drops
Volume Limit should be more than 9900ml
It should use conventional tubing (4 mm), PVC any Indian Make
Immediate display of choice of programming after purge
Must have function for immediate forward & reverse of screen/programming
parameters
Programming :
ml/hr,
time x volume limit.
Weight x concentration x dose
Must have Occlusion Pressure from 0 to 120 kPa
KVO should be 0.1 ml to 3ml/hr
Bolus rate should be more than 900 ml/hr
Must have Special functions like Titration , Fluid Balance
Time limit from 1min to 99hrs & 59 minutes.
Should have pre alarms – Ultrasonic sensor type
Visual / sound alarms like End of Battery Power , Incorrect flow rate, Door open, No
drops, Occlusion , programming error, Running in KVO, Fluid balance etc.
Flow rate deviation with standard tubing should not be more than +/- 6% of the
programmed flow rate.
Battery Rechargeable more than 5 hrs of operation
Should be able to work on 220 Volts
Unit should be CE Certified, ISO company & must have ISO 13485 certification

Item No 75

Bio-safety Cabinet

HLL/PCD/ESIC-20/10-11 Page 110 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

simple operation for ultimate safety with 60% less energy consumption and heat output
that complies with the EN 12469
Exterior dimensions with stand Approx. (w x h x d) 1300 x 2200 x 795 mm (51.2 x 86.6
x 31.3 in)
Interior dimensions Approx. (w x h x d) 1200 x 780 x 495 mm (47.2 x 30.7 x 19.5 in)
Work surface with adjustable stand 750 to 960 mm (30 to 38 in)
Interior work surface area Approx.0.56 m2 (930 sq. in)
Working height of front window Approx.200 mm (8 in)
Maximum lifting height of front window Approx.535 mm (21 in)
dimensions Approx.(w x h x d) 1410 x 1700 x 925 mm
Ventilation System Exhaust/inflow air volume 400 m3/h (230 CFM)
Heat emission at 25°C ambient ~0.15 kW
Filter Specification Supply/exhaust air filter HEPA H 14 EN 1822,
Additional exhaust filter option (AEF) HEPA H 14 EN 1822,
Performance Certification EN 12469; GS Nord Cert-TÜV
Sound pressure level <55 dB (A)
Lighting power >1200 lx
Electrical Data Voltage 1/N/PE 230 V
Frequency 50 Hz
Power consumption 0.4 kW
Current consumption 1.7 A
Protection class I / IP 20
Protective measure Conductor connection Conductor connection
Individual precautions on customer side Lead fuse (slow blow) T 16 A or circuit breaker
B 16. The local electrical regulations in the country of
use as well as the relevant connection conditions must be observed. The national
regulations for
electrical engineering as well as the relevant technical connection conditions must be
taken into account.
Supply Management Supply requirement 230 V, 50/60 Hz standard supply. Total
requirement including interior sockets 13-16 Amps.
Receptacles The receptacles have a load capacity of up to 5 A and are protected with
T 5 A fuses. When all
receptacles are in use simultaneously, they must not exceed the maximum total load
capacity of 5 A.
Radio interference Circuit is interference free in accordance with EN 55 014
Service valves Up to 4 (installed through access ports)
Receptacles One double, right side

Item No 76

Laprocator

Laparoscope with operating channel. Dia 11 mm. Parallel eyepiece. Should be CE


marked - 1 no.
Trocar 12 mm with obturator – 1 no.
Spare washers for trocar – 10 nos.
Fiber optic light cable – 6 mm fiber bundle size, 300 mm long for high light output.
Light source, Halogen 250 watt, twin bulb, constant color temperature. Light should not
change whiteness on reducing intensity. Condenser lenses to focus light on fiber optic
cable.
Single puncture Ring applicator -2 nos
Veress Needle 120 mm long – 2 nos
CO2 insufflator flow rate – 30 LPM.

HLL/PCD/ESIC-20/10-11 Page 111 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No 77

Ultrasound Machine 3D with Color Doppler

It should be robust state of art, fully digital system capable of performing imaging
application in abdominal, obs/gynae , musculo sketeal, cardiovascular, small part etc.
It should be eaisly traspoartable and capable of performing bed side examination
in operation theater, endoscopic and intervention suits.
1. System should have board band all digital beam former capable of processing
singles from 2-18 MHz with tissue specific preset and enhanced spatial focus and
resolution and complex pulse design and wave form excitation.
2. System should have advance image processing softwere algorithm to analyse target
and artifacts so as to sharpen target anatomy and reduce/eliminate artifact and improve
resoluction for example separate two objest 0.2mm apart laterally (laterl resoluction ) at
11 MHz approx.
3. System should in corporate facility for high resoluction 2D , M-mode , PW,CW,color
flow imaiging modes, colour sdge processing.
4. System should have real time triplex mode facility in 2D, colour and Doppler modes
with at least 20 frames/sec. or higher .
5. System shoyld have color campare mode color /color power mode and normal gray
scale mode side by side or equivalent.
6.System should have quad &thumbanail display etc.
7. System should have dynamic range of 180db or higher. Higher will be preferred.
8. MI & TI should be displayed on screen.
9. System should have more then 500 frames per second and high PRF (5k) mention
rate highest will be preferred.
10. System should have scan dept of 30cm or more.
11.System should have 256 shades of gray display and a frame rate of 100/Sec or
higher.
12. System should have facility for real time or frozen, pan or point zoom upto 8times.
13. System should have cineloop review possible on all modes for individual and mixed
loops for more then 5000 frames and more 30 seconds of spectral doppler and m-
mode strip.
14. System should have 120 or more transmit channels and 20,000 or more receiving
channels with on site upgradability to higher number. Full spectrum imaging where
whole of band width is utillized.
15.Sytem should have automatic as well as manually adjusable touch screen guided
control of :-
i) Focus Transmit and dynamic recive at multiple dept individually and simultaneously.
ii) Frame rate for B mode, color, compounded resolution ect.
iii) Dynamic range compression.
iv) Pre-processing such as edge enhancing, beam line density.
v) Post processing such as image persistance, speckle reduction.
vi) Real time compound resolution image no of viewing angles and lines per frame (<9)
vii) Audio adjustments.
viii)Adjusttable "packet" size in phase shift analyses for example 10 or move transmit
and receive sequences.
16. System should have panoramic extended field of view and trapezoidal view on
linear propes.
17. System should independent steering of B mode and color on linear probe.
18. System should have 3D acquisition and analysis package for e.g. free hand 3D
,static 3D, multislice view capabilities.
a. Data Acquisition : free hand, mechanecal & electronic and gated techiniques.
b. Transducer : matrix acquisition for faster scan rate with better damping of
traansducer element for sharper focus.

HLL/PCD/ESIC-20/10-11 Page 112 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

19. It should have advanced issue harmonic imaging on all probes such as
fundamental,phase inversion , coded Octave and Intravenous US contrast tuned
imaging and quantification.
20. It should have cardiac package and probe for adult and fetal echo cardiography
e.g. myocardial velocity, stress & strain quantification, quality intima media thickness,
tissue velocity maping and peripheral vascular study optimization programmes.
21. It should have extensive software and automatic and user programmable
calculation packeg for gray scale color Doppler, 3D applications.
22. It should have a 17"hight resolution medical grade LCD screen monitor with
articulated arm.
23. It should be provided with following transducers, board band full spectrum
functioning, with colour, covering 2 to 18 MHz.
• Convex Abdominal 3-6 MHz Approximately.
• Endocavity (TVS+TRUS)5-9 MHZ approx. with 200*R.
• Linear high frequency 6-18 MHz approx.
• Micro convex phased array 2-5/3-6 with cardiac capabilities.
The configuration may vary according to manufacturer but all function should be
served.
24. It should be capable of supporting at least three or more transducers ports with
switching from consol.
25. System should have built in image Management Software, for off line application
when patient has gone after examination, such as image Manipulation, Multi Planner
reformating, etc. its should have hard disk memory of 360GB or more with built in
CD/DVD read write and limitless connectivity e.g. USB/HDD/CD/DVD.
26. For paralle processing of imaging Data system should be provided with a separate
latest Pentium i3 (2.9GHz) configuration with 1Tera Byte Hard based station with USB
and serial port with 19" LCD medical grade monitor with very high quality image
management Software (e.g.radion) with same capabilities as main machine such as
retriving data along Demography, Zoom, Pan, retrieving information from CD/DVD
wretriving and Teleradiology Software exporting JPG & AVI file format to link other
state in the hospital information service (HIS), RIS & PACS server. iIt should have
minimum memory storage for about 100 patients per day and cine loops for 6 month's
work.
27. System should be provided with black & white tharmal printer & colour laser printer
capacity of printing 3000 sheets per cartride and 3000sheet for starting work.
28. On line UPS for 2KVA or more as appropriate for supporting main equipment &all
line accessosies for 30 minutes should be provided with system machine should have
high inherent voltage fluctuation protection.
29. Standard accessories such as various cuffs, elbow rest, standardization phantoms,
bio guides & needles etc. should also be provided.
30. Syatem should be upgradable to :
a. Elastography and sound speed measurement free of cost as and when it
become available with the Company.
b. Comparison/super imposition/fusion with CT/MRI images obtained by teleradio.

Item No 78

PHOTOTHERAPHY MACHINE

1. Dimensions of the chamber should be at least 6 feetx 3 feet x 3feet.


2. Phototheraphy chamber of 18 UVA+18 NB UVB tubes designed for providing even
irradiation of the body in the treatment area.
3. UV chockes must be provided to provide long life to the tube light and cooling fans
for effective cooling of the unit.
4. integrated dosimeter system for easy calculation of irradiation levels.
5. The equipment have CE or FDA or ISI certification.

HLL/PCD/ESIC-20/10-11 Page 113 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

6. Advanced micro computerized electronic LCD/TFT Controller which allows setting of


joules/time for UVA and milli Joules/time for NB UVB tubes.
7. Automatic computation of irradiation time from joules/time for NB UVB tubes.
8. Dose limit can be preset and cumulative dose is displayed instntaneously with
provision of storage of data. Provision of 'software backup' is preferable.
9. Variation in irradiation is taken care by built in UVA/NB UVB sensors which should
be able to detect all irradiation completely and uniformly.
10. Switches the system 'off' automatically with warming alarm at the end of set
irradiation time.
11. Built in memory system that helps to avoide error in treatement.
12. Body to be of a metal which is rust free so as to ensure long rust free life of the unit.
13. Automated and/or mechaical safety mechanism to prevent excess irradiation to the
pantients so as to avoid/prevent burns etc.
14. Electrical leakage circuit tripper/breaker in each panel to ensure maximum safety of
the patient.
15. Open top unit to ensure maximum ventilation and prevent claustrophobia.
16. Mechanism to provide information to the patient regarding duration of treatment and
time left for exposure during their treatment.
17. Computer for patient data management with software and interface for the
phototheraphy chamber which is Rs.232 compatible.
18. To be supplied with suitable stabilizer.
19. Black UV Goggles and Eye pads cover (3 pairs each for adult and 3 pair for
children ) as protective+

Item No 79

Binocular microscope

a) Optical system infinity color corrected optics, antifungus treated.


Eye pieces: 10X wide fied ( FV 22 or more ) with inter pupillary distance 48-75mm with
dioptic adjustment both side, eye guards, eye level riser.
a) Objectives: Bright field infinity color corrected optics, antifungus treated
4X,10X,40X,100X oil immersion. In changing from one objective to anther or
reintroducing the same objective by rotation of the nosepiece, the center of the field
should not apper displaced by more then 0.02mm in object plane.
b) Nosepice : Revolving, reversed (inward) tilt
c) Tubes : Siedento f tiltable Binocular tubes with minimum inclination 25-30 degrees .
d) Stage : uniformly horizontal, scratch resistant, rackless, rotatable stage with right
hand operation & single slide holder with a stage upper limit stopper.
e) Condenser : issuing out universal with numerical aperture of 0.9/1.25 with position
for bright field should have a removable filter holder, swing in, blue filter for bright field.
f) lllumination system : The system should have a bulit in, variable, low voltage light
source, the circuit for the light source should include a constant voltage supply. The
system should be provided with a step down transformer and on/off switch and intensity
control. the lamp should be provided with a lamp socket, which has the facility for easy
replacement of the bulb. The housing of the microscope. Halogen bulb -12v/20-30w.
The illuminator should have a built in field diaphragm for kohler illumination.
Power supply : Voltage 220V, 50hZAC should have one on- off power switch, power
cord with a 3pin male plug. The system should have an inbuilt protective/ safety
devioce to withastand fluctuations of voltage from 140v to 280v. The fuse the halogen
lamp should be easily accessible.
g) Arm rest, Left and right.

Item No 80

HLL/PCD/ESIC-20/10-11 Page 114 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Yag Laser

A. PHOTOCOAGULATOR
MODE OF OPERATION : Continuous wave
TYPE : Solid state laser b
WAVELEGTH : 532nm
COOLING : Air cooled
AIMING BEAM : Continuous wave, diode laser, 635nm

B. SLIT LAMP
SPOT SIZE : 50-1000µm
PROTECTION FILTER FOR SAFETY
MAGNIFICTION : 5step
ILLUMINATION UNIT : Slit width and length 0-14mm
Filters : blue, red-free, heat absorbing
Halogen lamp
POWER SUPPLY : 240V, 50-60Hz
BASE UNIT to alloew longitudinal, lateral and vertical movements
C. SLIT-LAMP ADAPTER
D. Motorrized table
E. Mainster and PRP lenses
F. UPS 1000 VA

Item No 81

CELL SEPARATOR

1. The apheresis machine should have facility for all Blood components collection with
single /double access including 1. Platelepheresis, 2. Peripheral blood stem cells, 3.
Therapeutic Plasma Exchange. And 4.Red Cell Apheresis.
Protocols for the above said proceduresis.
2. The cell separator should be fully automatic, microprocessor controlled with access
operator control panel such as touch screen.
3. It should be based on intermittent or continuous flow technology during duble/ single
needle access.
4. The Cell separator should ensure all donor safety parameters before starting the
procedure and at all time during operation.
5. The cell separator should be able to perform priming with normal salineor with
mixture of anticoagulant and self test as a part of donor's safety.
6. The cell separator should have inbuilt cuff pressure and promt grip for donor comfort
and adequate blood flow.
7. It should have auto cuff mechanism for automatic inflation and deflation.
8. Advance help menu should be available at any time during alarm conditions.
9.Extra corporeal volume not exceed 250 ml.
10. The yield estimator to help decide yield, volume to be processed and suggested
storege fiuid.
11.Interphace detector should continuously monitor the collection line to avoid any
contamination during esch protocol.
12. Capable of downloading or printing full procedure report any time time after the
procedure.
13. The cell separator should have rechargeble battery to store data and re-start in
case of power filure.
14. Inlet and return flow rates upto 100 ml +/-15 ml per minute.
15. In case inlet line occlusion, machine should be able to re-start automatically.

HLL/PCD/ESIC-20/10-11 Page 115 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

16. It should have fluid leak detector for donor's safety.


17. It should have provision for saline re-infusion to donor.
18. It should be able to regulate ACD delivery, should not have bolus return.
19. It should have automatic door lock for centrifuge during the procedure.
20. Lickable castors/ trolley for easy mobillity.
21. Should have Automatic kit loading facillity.
22. The disposable kit should be self sufficeient in terms of collection bag, transfer beg,
needles, anticoagulant and normal saline.
23. Should be a totally closed system.
24. Should be able to provlbe Leucoreduced platelet products
25.Power supply -220-240 volts/50Hz.
26. Suitable UPS with maintenance free batteries for minimum one hour back up.
27. FDA/CE approved
28. Five year Warranty. Five year CMC
29. Disposable kits should be available at fixed price for three years from date of
installation

Item No 82

CELL WASHER

1. Microprocessor controlled fully automatic cell washer


2. Touch pad with digital LCD display
3. Programmes for wash cycles, saline volume, spin time and RPM
4. Save and store multiple programmes
5. To accommodate 12 tube of standard 12x75 mm size
6. Brushless motor for quiet operation
7. Saline detec system with audible low saline warining
8. Agitate cycle to ensure complete re-suspension of cells
9. Audible and visible alert at the end of process
10. Power 220 v/50 Hz
11. CE/FDA/BIS approved

Item No 83

DONOR COUCH

1. Provide comfortable position to donors


2. Chair type with soft padding for cushioning and rexin cover
3. Ability to adjust donor's position automatically by operating a simple switch
4. Easily tilted to head low position
5.Easily adjustable arm rest with swinging out as well as moving up and down facility
6. Arm rest wide, sturdy to hold donor's arm in a comfortable position throught the
procedure of blood donation, provided on both sides
7. Lifting capacity : minimum 130 Kilograms
8. Four lockable castors for easy mobility
9. Trays and stands for keeping all blood collection accessories
10. Input power -220V/50Hz
11. CE/FDA/BIS approved

Item No 84

HLL/PCD/ESIC-20/10-11 Page 116 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

PLATELET INCUBATOR AND AGITATOR

1. Coountertop model for safe storage of platelets


2. Temperature set to 22+/- 2 degree C
3. Capacity : minimum 96 random bags or 32 apheresis bags
4. Power coated MS exterior, stainless stell interior
5. Transparent glass door
6. Flate bed agitator
7. Oscillations: 70 to 80 cycles per minute
8. Agitation stops when the incubator is opened and restarts automatically
9. Pull out easily removable, stainless steel, perforated doawers
10. Microprocessor controlled temperature indicator with digital display of temperature
11. Alarm for temperature deviation, door open, power failure, sensor failure
12. Built in vasible and autdible motion alarm
13. 24x7 days inkless temperature chart recorder
14.Power 220v/50 Hz
15. Voltage stabilizer of appropriate required specifications conmpatible with the unit to
be provided
16.All disposables to be provided at fixed rate for three years from of installation
17. CE/FDA/BIS approved

Item No 85

STERILE TUBE CONNECTING DEVICE

1. Compatible with all standard of different types of blood bag


2. Able to weld wet to wet, dry and wet to dry tubing combinations
3. In built sensor to monitor the temperature during entering welding process
4. LCD display for continuous monitoring
5. Ensure complete sterility of transferred blood
6. No Particle or chemical residues should be formed during the welding procedure
7. Welds should have optimal tube alignment and no leakage of blood of during the
process
8. Automatic replacement of wafers after each use
9. Built in wafer disposal box
10. Input power -220V/50Hz
11. CE/FDA/BIS approved

Item No. 86

General Instruments Set

CATHETERISATION TRAY X 10
Sponge holder 8" 2
Artery forceps straight 9" 1
S.S. Bowl 10 cm 2
S.S.Kidney tray 10" 1
S.S. Adoptor 1
SPINAL TRAY X 10
Sponge holder 8" 2
Artery forceps curved 7" 1

HLL/PCD/ESIC-20/10-11 Page 117 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

S.S. Bowl 10 cm 2
S.S.Kidney tray 10" 1
BASIC SET x 10
B.P. Handle No.3 2
B.P. Handle No.4 2
Dissecting Forceps Plain 7” 2
Dissecting Forceps Plain 8” 1
Dissecting forceps Toothed 7" 1
Dissecting Forceps Toothed 8” 1
Adson Dissecting Forceps Plain 6" 1
Adson Dissecting Forceps Toothed 6" 1
Towel clip 5" 6
Cd. Scissor Suture Cutting 1
Mayo Scissor Cd. 9" TC 1
METZScissor Cd. 8" TC 1
Mayo scissor 8" st. 1
Mayo Scissor 8” 1
Needle Holder 7” Mayo hegar 1
Needle Holder 8” fine 1
Needle Holder 7” Mayo hegar 1
Artery Forceps Cd. 8” 2
Mosq. Artery Forceps Cd. 4
Mosq. Artery Forceps st. 2
Artery Forceps Cd. 7” 6
Artery Forceps St. 7" 2
Allis Forceps 7” 4
Babcock Tissue Holding Forceps 67” 2
Probe and director 1
Suction Tip No. 1, 2, 3, 4 1 Each
Yaunker's Suction with detatchable tip 1
Mixture Clamp 7” 1
Langenback Ret MEDIUM 2
Langenback Ret SMALL 2
“C” Shaped Retractor ( pair ) small & med 2 Each
Sponge Holder 8" 4
Skin hook sharp 2
Vein loops 2
S.S.Bowls 10 cm 4
S.S. Kidney tray 12" 2
PILES EXTRAS x 3
Sims speculum med 1
Sims speculum small 1
Sims speculum large 1
Speculam with one handle 1
Proctoscope small 2
Proctoscope med 2
Proctoscope big. 2
Artery forcep cd 8" 2
Venesection cannula 16,18,20,22 1 Each
Proctoscope small 2
Proctoscope med 2
Proctoscope big. 2
Piles needle 1

HLL/PCD/ESIC-20/10-11 Page 118 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No. 87

Minor Surgery Instrument Set

Minor Surgery Set

B.p.handle no.3 4
B.p.handle no.4 2
Dissecting forceps plain 7" 2
Dissecting forceps toothed 7" 2
Adson Dissecting Forceps Plain 2
Adson Dissecting Forceps Toothed 2
Towel Clip 4” 8
Iris Scissor st. 2
Iris Scissor cd. 2
Mosq. Artery forcep cd 6" 12
Artery Forceps Mosq. St 6" 8
Artery Forceps st. 7” 8
Artery forceps 7” Cd. 8
Allis forceps 7” 8
Babcock tissue forceps 4
Needle holder 8" mayo hegar 2
Needle holder 8" fine 2
Mayo scissor 8" st. 2
Metz. Scissors cd 8" 2
Mayo Scissor cd. 2
Venesection scissors 4" curved 2
Sponge holder 10" 8
Suction tip no1,2,3,4 2
Scoop Double Ended 2
Probe and director 2
Kilner retractor 4
Skin Hook 4
Lagenback retractor medium 4
Sinus forcep 2
Venesection cannula 16,18,20,22 2
S.S.bowls 10 cm 8
S.S.kidney tray 10" 4

CATHETERISATION TRAY

Sponge holder 8" 10


Artery forceps straight 7" 5
S.S. Bowl 10 cm 10
S.S.Kidney tray 10" 5
S.S. Adoptor 5

SPINAL TRAY

Sponge holder 8" 10


Artery forceps curved 6" 5

HLL/PCD/ESIC-20/10-11 Page 119 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

S.S. Bowl 10 cm 10
S.S.Kidney tray 10" 5

BASIC SET

B.P. Handle No.3 10


B.P. Handle No.4 10
Dissecting Forceps Plain 7” 10
Dissecting Forceps Plain 8” 5
Dissecting forceps Toothed 7" 5
Dissecting Forceps Toothed 8” 5
Adson Dissecting Forceps Plain 5
Adson Dissecting Forceps Toothed 5
Towel clip 5" 30
Cd. Scissor Suture Cutting 5
Mayo Scissor Cd. 8" TC 5
METZScissor Cd. 8" TC 5
Mayo scissor 8" st. 5
Mayo Scissor 8” 5
Needle Holder 7” Mayo hegar 5
Needle Holder 8” fine 5
Needle Holder 8” Mayo hegar 5
Artery Forceps Cd. 8” 10
Mosq. Artery Forceps Cd. 20
Mosq. Artery Forceps st. 10
Artery Forceps Cd. 7” 30
Artery Forceps St. 7" 10
Allis Forceps 7” 20
Babcock Tissue Holding Forceps 7” 10
Probe and director 5
Suction Tip No. 1, 2, 3, 4 5 Nos Each
Yaunker's Suction with detatchable tip 5
Mixture Clamp 7” 5
Langenback Ret MEDIUM 10
Langenback Ret SMALL 10
“C” Shaped Retractor ( pair ) small & med 10 Each
Sponge Holder 8" 20
Skin hook sharp 10
Vein loops 10
S.S.Bowls 10 cm 20
S.S. Kidney tray 12" 10

Item No. 88

INSTRUMENTS SETS FOR ORTHOPAEDIC SURGERY

Wire Instrument Set 1 Set


Sterilising Tray with lid 1
Spare Tray 1

HLL/PCD/ESIC-20/10-11 Page 120 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Lower Tray 1
Drill Bit, 2.0mm 1
Triple Drill Guide 2.0 with 3 holes, opposite side 1 hole 1
Wire Passer, 45mm bending diameter 1
Wire Passer, 70mm bending diameter 1
Wire Tightener with handle and two pegs, 1
Holding Forceps for Cerclage Wires, 1
Wire Bending Pliers, 1
Parallel Pliers, flat nosed 1
Wire Cutter, large, 1
Wire Cutter, short, 1
Bending Iron, for Kirschner Wires 1
Medullary Reamer Set 1 Set
Flexible medullary reamer 6.0mm 1
Flexible medullary reamer 6.5mm 1
Flexible medullary reamer 7mm 1
Flexible medullary reamer 7.5mm 1
Flexible medullary reamer 8.0mm 1
Flexible medullary reamer 8.5mm 1
Flexible medullary reamer 9mm 1
Flexible medullary reamer 9.5mm 1
Flexible medullary reamer 10mm 1
Flexible medullary reamer 10.5mm 1
Flexible medullary reamer 11mm 1
Flexible medullary reamer 11.5mm 1
Flexible medullary reamer 12mm 1
Flexible medullary reamer 12.5mm 1
Flexible medullary reamer 13mm 1
Flexible medullary reamer 13.5mm 1
Flexible medullary reamer 14mm 1
Flexible medullary reamer 14.5mm 1
Flexible medullary reamer 15mm 1
Flexible medullary reamer 15.5mm 1
Flexible medullary reamer 16mm 1
Bone Forcep Set For Reduction 1 Set
Verbrugge forceps 240mm 2
Verbrugge forceps 260mm 2
Verbrugge forceps 280mm 2
Reduction forceps broad,130mm 2
Reduction forceps W point,205mm 2
Reduction forceps,serrated,170mm 2
Reduction forceps,serrated,240mm 2
Aluminium Case,Red 1
Tray (Lower/upper) 1
Tray (Lower/upper) 1

DYNAMIC SCREW/DCS HIP SCREW INSTRUMENT Set 1 Set


Sterilising Tray with lid 1
Upper Tray 1
Lower Tray 1
DHS/DCS Threaded Guide Wire, 2.5mm dia., L 230/5mm 10
DHS Angled Guide 135º 1
DHS Angled Guide 150º 1

HLL/PCD/ESIC-20/10-11 Page 121 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

DHS/DCS Direct Measuring Device 1


DHS/DCS Wrench for one-step insertion L 230mm 1
DHS/DCS T-Handle with quick coupling, L 80mm 1
DHS Triple Reamer 1
DHS/DCS Impactor for one-step insertion, L 260mm 1
DHS/DCS Tap, L 220mm 1
DHS/DCS Centering Sleeve, locking 1
Coupling Screw, cannulated 1
DCS Angled Guide 1
DCS Triple Reamer 1
Small Fragement Instrument Set 3.5mm 1 Set
Sterilising Tray with lid 1
Lower Tray for Instruments 1
Tray for Plates 1
Upper Tray 1
Drill Bit, 2.5mm dia., L 110/85mm for quick coupling 2
Drill Bit, 3.5mm dia., L 110/85mm, for quick coupling 2
Countersink Shaft 3.5, L 72mm 1
Tap for 3.5mm Cortex Screws L 50/110mm 1
Tap for 4.0mm Cancellous Bone Screws L 110mm 2
T-Handle with quick coupling, L 80mm 1
Double Drill Sleeve 3.5/2.5 1
Insert Drill Sleeve 3.5/2.5, L 42mm Drill Bit 2.5mm dia. 1
Screwdriver, hexagonal, small, with Holding Sleeve 1
Screwdriver Shaft, hexagonal, small, L 100mm, for quick coupling 1
Screwdriver, hexagonal, small, with groove, L 200mm 1
Holding Sleeve, 1
Depth Gauge for 2.7mm to 4.0mm Screws 1
Sharp Hook, L 155mm 1
Holding Clip 4.5 - 7.0mm 2
Screw Forceps, self-retaining, L 85mm 1
DCP® Drill Sleeve 3.5 for neutral and load position 1
Bending Iron, slit widths 4.5/2.5mm, L 150mm, for Plates 2.7 and 3.5 1
Bending Iron, slit widths 2.5/4.5mm, L 150mm, for Plates 2.7 and 3.5 1
Bending Pliers for Plates 2.4 to 4.0, L 230mm 1
Bending Template for DCP® 3.5 and LC-DCP 3.5, L 87mm 1
Bending Template for DCP® 3.5 and LC-DCP 3.5, L 114mm 1
Wire Bending Pliers, L 155mm 1
Bending Iron, for Kirschner Wires 1.25 to 2.5mm dia., L 120mm 1
Reduction Forceps with points, L 130mm 1
Reduction Forceps with points, wide, ratchet lock, L 132mm 1
Reduction Forceps, toothed, ratchet lock, L 140mm 1
Bone Holding Forceps, self- centering, speed lock, L 190mm 1
Retractor, small, 8mm wide, short narrow tip, L 160mm 2
Periosteal Elevator, round edge, 6mm wide, L 200mm 1
Retractor, 15mm wide, L 160mm 2
Basic Instrument Set Large Fragment 4.5mm 1 Set
Sterilising Tray with lid 1
Upper Tray for Instruments 1
Spare Tray, without contents, 1
Lower Tray 1
Drill Bit, 3.2mm dia., L 145/120mm for quick coupling 3
Drill Bit, 4.5mm dia., L 147/120mm for quick coupling 2

HLL/PCD/ESIC-20/10-11 Page 122 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

T-Handle with quick coupling, L 80mm 1


Tap for 4.5mm Cortex Screws, L 70/125mm 2
Tap for 6.5mm Cancellous Bone Screws L 195mm 1
Double Drill Sleeve 4.5/3.2 1
Insert Drill Sleeve 4.5/3.2, L 80mm 1
Double Drill Sleeve 6.5/3.2 1
Screwdriver Shaft, hexagonal, 1
Screwdriver, hexagonal, large, 1
Holding Sleeve, large, L 120mm 1
Depth Gauge for 4.5 to 6.5mm Screws 1
Sharp Hook, L 155mm 1
Tension Device, articulated 1
Combination Wrench, 1
DCP® Drill Sleeve 4.5 1
LC-DCP Drill Sleeve 4.5 1
Universal Drill Sleeve 4.5 1
Bending Template for DCP® 4.5 and LC-DCP 4.5, L 210mm 1
Bending Template for DCP® 4.5 and LC-DCP 4.5, L 120mm 1
Bending Template for DCP® 4.5 and LC-DCP 4.5, L 155mm 1
Basic Intramedullary Nailing Insts set for Femur & Tibia 1 Set

Universal Socket Wrench 1


Hex Driver for 4.5/5.0 mm screw,6,4mm,5.0mm can screw,measuring guage for nails 1
Medullary Exchange tube 1
internal fracture Alignement Device 1
Depth Guage 1
Cannulated Drill Bit,150mm,3.5mm,4.0mm,4.8mm 1
Drill Bit,150mm,2.7,3.5mm,4.0mm,4.8mm 1
Twist Bit,305mm,3.5mm,4.0mm,4.8mm 1
Trocar 4.0X240mm 1
Krischner-wire with thread 3.2X 305mm 1
Guide Wire for cannulated screw 1.8X350mm 1
cannulated wire for ILN 2.4x900mm,3.0x900,Guide Wirw for reamer 2.0x900,Guide 1
wire with curved Tip 3.0X900mm,2.4X900mm
Open End Wrench SW 17/ 1
Measuring Gauge 1
Measuring Gauge for nails 1
Supine Driver 1
Handle with jacobs chuck 1
Tibial Proximal Device 1
Skin Protector 1
Curved Awl 1
Trocar 3.2mm,4.0mm,4.8mm,8.0mm 1
Tapered Reamer,Cannulated 13.5,15.5mm 1
Cannulated Reamer 9.0mm 1
Drill Sleeve 2.1mm,3.5mm,4.0mm,4.8mm,8.0mm 1
Slide Hammer 1
Tibial Extractor Bolt 1
Guide Rod For Cannulated Screws 1.8X350 1
Universal Proximal Device for Femural Nails 1
Nail Adapter Bolt 1
Length Guage 1
Twist Drill Cannulated 305mm,4.0mm,4.8mm 1
Reamer 1

HLL/PCD/ESIC-20/10-11 Page 123 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

General Instrument Set 1 Set


Sterilising Tray with lid 1
Spare Tray, 1
Lower Tray 1
Bone Hook, sharp, small, L 230mm 1
Bone Hook, sharp, medium, L 230mm 1
Retractor, 8mm wide, short narrow tip, L 220mm 2
Retractor, 18mm wide, short narrow tip, L 235mm 2
Retractor, 24mm wide, long and wide tip, L 270mm 1
Periosteal Elevator, curved shaft, 14mm wide, L 200mm 1
Periosteal Elevator, round edge, 6mm wide, L 200mm 1
Periosteal Elevator, straight shaft, 14mm wide, L 200mm 1
Hammer 500g, L 230mm 1
Chisel Handle, L 185mm 1
Chisel Blade, 10mm wide, thickness 0.9mm, L 81mm 1
Chisel Blade, 16mm wide, thickness 0.9mm, L 81mm 1
Chisel Blade, 25mm wide, thickness 0.9mm, L 81mm 1
Gouge, curved, for cancellous bone graft harvest., 10mm wide, L 250mm 1
Damaged screw,Broken Nail Removal Instrument set 1 Set
Aluminium Case, white, small, perforated, without contents 1
Tray, subdivided 1
Tray for Instruments, without contents 1
Hollow Reamer for 3.5/4.0mm Screws 1
Spare Reamer Tube 1
Extraction Bolt for 3.5/4.0mm Screws 1
Extraction Screw, conical, for 2.7mm, 3.5mm and 4.0mm Screws 1
Hollow Reamer for 4.5mm Screws 1
Spare Reamer Tube 1
Extraction Bolt for 4.5mm Screws 1
Extraction Screw, conical, for 4.5/6.5mm Screws 1
Hollow Reamer for 5.0/6.0/6.5/7.0mm Screws 1
Spare Reamer Tube 1
Extraction Bolt, for 5.0/6.0/6.5/7.0mm Screws 1
Aluminium Plate, anodized 1
Sharp Hook, L 155mm 1
Forceps for Screw Removal, L 205mm 1
Gouge, 10mm wide, L 205mm 1
Extract-Hook f/Nails ø9-11 1
Extract-Hook f/Nails ø11+larger 1
Extract-Hook long f/Nails ø11-14 1
Extract-Hook L480 f/Nails ø9 1
Connector f/Extract-Hook 1
T handle 1
T-Handle with quick coupling, L 80mm 1
Locking Compression Instrument set Large& Small 1 Set
Torque Screw Driver 3.5mm 1
Torque Screw Driver 5.0mm 1
Drill with Quick Coupling 2.5x150/30mm 2
Drill with Quick Coupling 2.7x200/40mm 2
Drill with Quick Coupling 3.2x150/30mm 2
Drill with Quick Coupling 4.1x200/30mm 2
Conical Screw Extractor 2.5mm 1

HLL/PCD/ESIC-20/10-11 Page 124 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Conical Screw Extractor 3.5mm 1


Screw Driver Shaft 2.5mm Hex 2
Screw Driver Shaft 3.5mm Hex 2
Holding Sleever 3.5mm/ 4.0mm 1
Drill Sleever 2.8/3.5mm 2
Quick Coupling Handle 1
Universal Drill Guide 3.5mm 1
Drill Guide Handle 1
Drill Bit 4.5mm / 145mm Quick Coupling 2
Drill With Quick Coupling 4.3x 195mm 2
Screw Driver Shaft 2.5mm Hex 100mm 1
Screw Driver Shaft 3.5mm Hex 100mm 1
Drill Sleever 2.5mm - load 2
Drill Sleever 4.1mm - load 2
Drill Sleever 2.5mm with Compression 2
Drill Sleever 4.1mm with Compression 2
Universal Drill Guide 4.5mm/3.2mm 1
Locking Screw Drill Sleever 3.5mm 1
Locking Screw Drill Sleever 5.0mm 1
Tap with Quick Coupling 3.5 2
Tap with Quick Coupling 4.5 2
Locking Screw Tap with Quick Coupling 3.5mm 2
Locking Screw Tap with Quick Coupling 5.0mm 2
Hexagonal Screw Driver 2.5mm 1
Hexagonal Screw Driver 3.5mm 1
Countersink with Quick Coupling 3.5mm 1
Countersink with Quick Coupling 4.5mm 1
Depth Gauge 3.5mm 1
Depth Gauge 4.5mm/6.5mm 1
Bending Iron 2
Quick Coupling for Jacob Chuck 1
Tap / Drill Sleeve 2.5mm 1
Tap / Drill Sleeve 3.5mm 1
Tap / Drill Sleeve 4.5mm 1
Tap / Drill Sleeve 6.5mm 1
Fusion Lock Set - Case Large Fragment Instruments 1
Cannulated Screws Instruments set 4.5mm 1 Set
Sterilising Tray with lid 1
Threaded Guide Wire, 1.6mm 10
Drill Bit, 1.5mm dia. 1
Drill Bit, 3.2 2
Countersink Shaft, cannulated, 1
T-Handle with quick coupling, 1
Tap, cannulated, for 4.5mm Cannulated Screws 1
Double Drill Sleeve 4.5/3.2 1
Parallel Wire Guide, 1
Trocar 1.6mm dia. 1
Screwdriver, hexagonal, large, 1
Drill Sleeve 1
Protection Sleeve 1
Screwdriver Shaft, hexagonal, large 1
Screwdriver, hexagonal, cannulated 1
Holding Sleeve 1
Measuring Device, 1

HLL/PCD/ESIC-20/10-11 Page 125 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Stylet, 1.6mm 1
Holding Clip 4.5 - 7.0mm for screw 1
Cannulated Screws Instrument Set 3.5 mm 1 Set
Vario Case for instrumrnts 1
Lid Stainless Steel size½ f/VC 1
Hold-Sleeve f/ 314.070 314.290 314.550+3 1
DoubleDrillGuide 3.5/2.5 1
ScrDriverShaft-hex-small ø2.5 1
Tap-cannul f/Cannul-Scr ø3.5+4 1
Drill Bit ø3.5/1.35 cannul L160/130 4flu 1
Drill Bit ø2.7/1.35 cannul L160/130 4flu 1
Handle w/Quick-Coupl L110 1
DoubleDrillGuide 2.7/1.25 1
Countersink cannulated f/Cannul-Scr ø3.5 1
ScrForceps self-hold L85 1
Sharp Hook L155 1
Direct Measur-Device f/Cannul-Scr ø3.5+4 1
Clean-Brush ø1.35 f/Cann-Instr 1
Hold-Clip f/Washers 1
GuideWire ø1.25 w/thread-tip w/trocar L1 1
ScrDriver-hex-cann f/Cannul-Scr ø3.5+4 1

Item No. 89

Vaginal Hysterectomy
B.P.Handle No.3 1
B.P.Handle No.4 1
Diss. Forceps Plain 16cm 1
Diss. Forceps Plain 20cm 1
Diss. Forceps Tooth 20cm fine 1
Diss. Forceps Tooth 20cm 1
Diss. Forceps Tooth 16cm fine 1
Diss. Forceps Attri. 20Cm 1
Diss. Forceps Fine Plain 20cm 1
Artery Forceps St.16cm 2
Artery Forceps Curved.16cm 6
Artery Forceps Mosquito curved 5
Artery Forceps Mosquito curved 5
Artery Forceps Mosquito St. 2
Heaany's Clamp 6
Valsalum Forceps Curved 2
Allis Forceps 20cm 6
Allis Forceps 16cm 6
Babcock Forceps 20cm 2
Babcock Forceps 16cm 2
Tanaculum Forceps 25cm 3
Artery Forceps Long Curved 23cm 2

HLL/PCD/ESIC-20/10-11 Page 126 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Kochers Clamp 20cm St. 2


Kochers Clamp 20cm Cu. 4
Hysterectomy Clamp 23cm 6
Needle Holder 20cm 1
Needle Holder 20cm 2
Needle Holder 15cm 1
Needle Holder 16cm 1
Scissors Mayo St. 15cm 1
Scissors Mayo St. 18cm 1
Scissor Mayo Curved 17cm 1
Scissor Mayo Curved 18cm 1
Scissor Mayo Curved 20cm 1
Sponge Holder 10" St. 2
Sponge Holder 10" Cd. 2
Avard Vaginal Spaculam Handle 1
Avard Vaginal Spaculam Weight 1
Avard Vaginal Spaculam Blade 1
Avard Vaginal Spaculam Blade 1
Sim's Spaculam 1
Sim's Spaculam Blade Guide 1
Sim Speculam Medium 1
Sim Speculam Large 1
Uterus Holding Forceps 1
Anterior Vaginal Wall Retractor 1
Jackson Retractor 2
Suction Cannula Trizer No.4 1
Towel Clip 6
S S Bowls 10 cm 2
S S Kidney tray 12" 1

Item No. 90

Abdominal Hysterectomy set

1 Sponge Holder x4
2 Needle holder x3
3 Bab cock x2
4 Artery forceps 6.1/2” x10
5 Artery forceps7.1/2” x4
6 Allis Forceps x6
7 Straight Hyst. Clamp x5
8 Curved clamp x5
9 Mosquitoes forceps 4.1/2” x6
10 Doyn’s Retractor Big x1
11 Self Retaining Forceps (Balfour) x1
12 Deverse Small x1
13 Deverse Big x1

HLL/PCD/ESIC-20/10-11 Page 127 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

14 Towel clip x5
15 Steel Basin x2
16 S.S Bowl x3
17 Steel Tray Big x2
18 Myo curved scissor 8” x2
19 Straight scissor 6.5” x2
20 Tooth dissecting forceps 6.5” x1
21 Non Tooth dissecting forceps 6.5” x1
22 Metzenbaum curved scissor 8” x2

Item No. 91

Delivery and Ceasarian set

B.P.Handle No.3 1
B.P.Handle No.4 1
Dissecting Forceps Tooth 15cm 2
Dissecting Forceps Plain 15cm 1
Scissor Mayo St. 18cm 1
Scissor Mayo St. & Curved 16cm 1 Each
Scissor Episectomy 1
Needle Holder 18cm 2
Placenta Puncture Forceps 1
Sponge Holder 20cm 2
Sponge Holder 26cm 1
Anterior Vaginal Retractor 1
Artery Forceps Curved 16cm 2
Cord Clips 1
Sim's Speculum Medium 1
Sim's Speculum Large 1
Towel Clip 4
S. S. Bowl 6 cm 2
Big Kidney Tray 1
Allis FORCEPS 8 “ 2
Mosquito Artery cd 2
Ceasarian set
B.p handle no-3 1
B.p handle no-4 1
Dissecting forceps fine toothed 14cm 1
Dissecting forceps adson plain 12.5cm 1
Dissecting forceps adson toothed 12.5cm 1
Dissecting forceps Atragrip 20cm 1
Dissecting forceps Atragrip15cm 1
Dissecting forceps plain 15cm 1
Dissecting forceps toothed 20cm 2
Cord clip 1

HLL/PCD/ESIC-20/10-11 Page 128 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Towel clip backhus 5


Allis forceps 16cm 6
Allis forceps 20cm 2
Babcork forceps 16cm 2
Babcork forceps 20cm 2
Artery forcpes st 16cm 8
Artery forcpes cd 20cm 4
Artery forcpes cd 16cm 12
Needle holder 18cm 1
Needle holder 20cm 1
Needle Holder15cm 1
scissors mayo cd 16cm 1
Scissors mayo st 20cm 1
Episectomy scissors 1
McIndo scissors cd 23cm 1
Suture cutting wave scissors 18cm 1
Uterine scissors (sims) 20cm 1
kocher's clamp cd 2
Kocher's clamp st 4
Green Armitage forceps 6
Sponge holder 26cm 4
Yankurs suctions tip 2
Frizer suction tip cannula NO-3 1
C' shaped retractor 3
Doyans retractor 2
Morries retractor 1
Doyen's skin clip 1

Item No 92

General Instruments for O T

B.P.HANDLE NO.3 22
B.P.HANDLE NO.4 20
B.P.HANDLE NO.7 10
BLADDER SOUND 5
CANNULA OVONOTOMY TROCAR FOR (HYEDRO) 5
CANNULA SUCTION SET 15 SETS
CERVICAL NEEDLE 4
CHIMKO 2
DILATORS MALE 2 SETS
DILATORS HEGARS(UTERINE) 14 SETS
DOYEN MYOMA SCREW 4
FORCEPS PLAIN ADSON 5
FORCEPS TOOTH ADSON 5
FORCEPS ALLES 5" STRAIGHT 40

HLL/PCD/ESIC-20/10-11 Page 129 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

FORCEPS ALLES 5" CURVED 4


FORCEPS ALLES 6" 106
FORCEPS ALLES 7" 44
FORCEPS ALLES 8" 142
FORCEPS ALLES 10" 42
FORCEPS ARTERY STRAIGHT 5" 64
FORCEPS ARTERY CURVED 5" 44
FORCEPS ARTERY CURVED 10" 75
FORCEPS ARTERY CURVED 7" 140
FORCEPS ARTERY CURVED 9" 10
FORCEPS ARTERY CURVED 10" 100
FORCEPS ARTERY CURVED 8" 120
FORCEPS ARTERY CURVED 6" 70
FORCEPS ARTERY STRAIGHT 10" 10
FORCEPS ARTERY STRAIGHT 8" 28
FORCEPS ARTERY STRAIGHT 6" 70
FORCEPS ARTERY STRAIGHT 7" 44
FORCEPS BABCOCK 10" 10
FORCEPS BABCOCK 7" 10
FORCEPS BABCOCK 8" 30
FORCEPS BABCOCK 6" 20
FORCEPS BONNEY VAGINAL 5
FORCEPS BERKELEY BONNEY HYSTERECTIOMY 8" CURVED ON FLAT 5
FORCEPS COLLIN UTERUS HOLDING 10" 5
FORCEPS CHITTLE 12" 54
FORCEPS CHITTLE 10" 10
FORCEPS DISSECTING (TOOTH) 50
FORCEPS DISSECTING TOOTH 20CM. 50
FORCEPS DISSECTING 15CM.(6") 100
FORCEPS FISTULA DISSECTING 10
FORCEPS GLAND HOLDING 4
FORCEPS GREEN ARMYTAGE 4
FORCEPS GWILLINGAM HYSTERECTOMY CORVED 8" 5
FORCEPS KOCHER ARTERY CURVD10" 50
FORCEPS KOCHER ARTERY CURVED 8" 90
FORCEPS KOCHER ARTERY CURVED 6" 50
FORCEPS KOCHER ARTERY CURVED 7" 30
FORCEPS KOCHER ARTERY CURVED 12" 15
FORCEPS KOCHER ARTERY STRAIGHT 12" 25
FORCEPS KOCHER ARTERY STRAIGHT 8" 10
FORCEPS KOCHER ARTERY STRAIGHT 6" 15
FORCEPS KOCHER ARTERY STRAIGHT 7" 10
FORCEPS LUCS BIOPSY SMALL 2
FORCEPS LUCS BIOPSY MEDIUM 2
FORCEP MAGILIS 10
FORCEPS MOSQUITO CURVED 65
FORCEPS MOSQUITO CURVED 4" 20
FORCEPS MOSQUITO CURVED 5" 30
FORCEPS MOSQUITO STRAIGHT 25
FORCEPS MOSQUITO STRAIGHT4" 10
FORCEPS MOSQUITO STRAIGHT5" 10
FORCEPS OVUM 10" 15
FORCEPS PUNCH BIOPSY 10
FORCEPS SPONGE HOLDING6" 20

HLL/PCD/ESIC-20/10-11 Page 130 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

FORCEPS SPONGE HOLDING 8" 10


FORCEPS SPONGE HOLDING 10" 10
FORCEPS SPONGE HOLDING 12" 20
FORCEPS SWAB HOLDING6" 25
FORCEPS SWAB HOLDING12" 50
FORCEPS TOOTH4" 10
FORCEPS TOOTH5" 55
FORCEPS TOOTH6" 35
FORCEPS TOOTH7" 60
FORCEPS TOOTH8" 10
FORCEPS PLAIN5" 15
FORCEPS PLAIN6" 50
FORCEPS PLAIN7" 55
FORCEPS PLAIN8" 50
FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH CURVED 5
FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH STRAIGHT 5
FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH CURVED 5
FORCEPS HEANEYS HYSTERECTOMY CLAMP 8" DOUBLE TOOTH STRAIGHT 5
FORCEPS LANCE TISSUE 5" 20
FORCEPS LANCE TISSUE 6" 20
FORCEPS LANCE TISSUE 7" 10
FORCEPS RIGHT ANGLED LONG 1
FORCEPS RIGTH ANGLED MEDIUM 1
FORCEPS SINUS 6" 30
FORCEPS UTERUS HOLDING SOMER 4
FORCEPS UTERINE PACKING 10" 4
FORCEPS VOLSELLUM 20
FORCEPS VOLSELLUM 10" 88
FORCEPS VOLSELLUM 12" 2
FORCEPS RAMPLEY SPONGE HOLDNG
FORCEPS BOTH END CURRETOR(VOLKMANSCOOP) 5PAIRS
FISTULA DIRECTOR5" 5
FISTULA DIRECTOR6" 5
FG- 655-01-V BONNEY UTERINE POLYPUS FORCEPS 10.1/2 (267MM) 4
FG- 660-01-E VANT2 UTERINE POLYPUS FORCEPS 10.1/2 (267MM) 4
FG- 765-01-H NOVAK SUCTION CURETTE LUER 254MM(10") 2SET(1TO12
NO.)
FG- 768-01-V RANDALL SUCTION CURETTE 4MM DIA. LEUR 10" (254MM) 2SETS
(1TO12NO.)
FG- 795-01-Y LEECH WIKINSON CERVICAL BIOPSY PUNCH 8" (203MM) SHAFT 2
FG- 815-01-M VAN DOREN CERVICAL BIOPSY PUNCH WITH CURVED BASKET 4
JAW ANGLED SHANKS 267MM(10.1/2")
FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED 2
SMALL13/16" X3"(21MMX76MM)7"(17MM)
FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED 2
MEDIUM1.1/8" X3.3/8"(29MMX86MM)7"(178MM)
FG-185-01-E-SIMS DUCKBELL SPECULAM SINGLE ENDED CHROMIUM PLATED 2
LARGE1.5/8" X3.3.1/2"(419MMX89MM)7"(178MM)
LEECH WILKINSON INTRAUTERINE CANNULA MEDIUM BASE OF CONE 1/2" 2
(13MM)1.3/16"(30MM),10.1/2"(267MM)
I.U.D.REMOVING HOOK 12.1/2" 6
SCHUMAKER CERVICAL BIOPSY PUNCH 9.1/2 SHAFT 5
HSG CANNULA SMALL 4
HERNIA RING HOLDER 2

HLL/PCD/ESIC-20/10-11 Page 131 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

INSTRUMENT PIN 6
M.T.P.CANNULA SET OF FOUR 5 SET
MAYO OLIVE STRIPPER COMPLETE SET 2
NEEDLE HOLDER 5" 15
NEEDLE HOLDER 6" 40
NEEDLE HOLDER 7" 55
NEEDLE HOLDER 8" 55
NEEDLE HOLDER FINE9" 5
PROCTOSCOPE LARGE 10
PROCTOSCOPE MEDIUM 10
PROCTOSCOPE SMALL 10
RESPIRATORY RIGHT 1
RESPIRATORY LEFT 1
RETRACTOR ABDOMINAL WALL 4
RETRACTOR CATSPAW TWO PRONG & THREE PRONG LARGE 1
RETRACTOR CATSPAW TWO PRONG & THREE PRONG SMALL 2
RETRACTOR CATSPAW TWO PRONG & THREE PRONG MEDIUM 12
RETRACTOR RIB 1
RETRACTOR CZERNEY DYONS 15
RETRACTOR CZERNEY 8" 20
RETRACTOR DEVARS 10
RETRACTOR JERNEYS 10
RETRACTOR JOLLS THYROID 2
RETRACTOR LONDON VAGINAL BLADE 3.5'(89MMX25MM) 2
RETRACTOR MASTOID SELF RETAINING 1
RETRACTOR ANTERIOR VAGINAL WALL 50
RETRACTOR CGARNIS 6" 15
RETRACTOR LANGENBACK 10
RETARCTOR SKIN D'S(SKIN HOOK) 2
SCHUMAKER CERICAL BIOPSY PUNCH 9.1/2"SHAFT 5
SCISORS POINTED 20
SCISSORS ROND MAYO 15
SCISSORS CURVED SMALL 20
SCISSORS CURVED MEDIUM 20
SCISSORS SHARP STRAIGHT 6
SCISSORS DRESSING 6" 5
SCISSOR DRESSING SHARP & BLUNT 5" 12
SCISSOR DRESSING SHARP & BLUNT 6" 10
SCISSOR DRESSING SHARP & BLUNT 7" 10
SCISSOR DRESSING SHARP & BLUNT 8" 10
SCISSOR DRESSING SHARP 5" 10
SCISSOR DRESSING SHARP 6" 15
SCISSOR DRESSING SHARP 7" 10
SCISSORS EPISIOTOMY(INDIAN) 15
SCISSORS FINE CURVED 3" 10
SCISSORS FINE CURVED 4" 10
SCISSORS FINE CURVED 6" 10
SCISSORS FINE CURVED 8" 10
SCISSORS GAUZE CUTING 10
SCISSORS MAYO'S CURVED 5" 10
SCISSORS MAYO'S CURVED 6" 30
SCISSORS MAYO'S CURVED 7" 20
SCISSORS MAYO'S CURVED 8" 5
SCISSORS MAYO'S STRAIGHT 5" 10

HLL/PCD/ESIC-20/10-11 Page 132 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

SCISSORS MAYO'S STRAIGHT 6" 25


SCISSORS MAYO'S STRAIGHT 7" 15
SCISSORS MAYO'S STRAIGHT 8" 15
SCISSORS METAZOLIC SMALL CURVED 10
SCISSORS METZENBAUM CURVED BLUNT TIP 6" 20
SCISSORS METZENBAUM CURVED BLUNT TIP 7" 20
SCISSORS METZENBAUM CURVED BLUNT TIP 8" 20
SCISSORS METZENBAUM CURVED BLUNT TIP 9" 10
SCISSORS METZENBAUM STRAIGHT POINTED 4" 20
SCISSORS METZENBAUM STRAIGHT POINTED 6" 30
SCISSORS METZENBAUM STRAIGHT POINTED 7" 20
SCISSORS METZENBAUM STRAIGHT POINTED 9" 10
SCISSORS METZENBAUM STRAIGHT POINTED 8" 20
SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 6" 20
SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 7" 20
SCISSORS METZENBAUM STRAIGHT CURVED BLUNT TIP 8" 20
SCISSORS CURVED SHARP 10
SCISSORS STITCH CUTTING 30
SCISSORS STRAIGHT ROUND POINT 30
SCISSORS STRAIGHT SHARP 10
SCISSORS SYMPATHECTOMY 4
SCISSORS TAILOR 2
SCISSORS TAILOR 10" 4
SCISSORS TAILOR 12" 20
SCISSORS TAILOR 6" 2
SHIRODKAR CERVICAL CANNULA MEDIUM 2
SHIRODKAR CERVICAL CANNULA SMALL 2
SKIN GRAFTING KNIFE BLADE HANDLE 5
SPECULUM RECTAL MEDIUM 5
SPECULUM CUSKOS 30
SPECULUM RECTAL SMALL 5
SPECULUM SIM'S LARGE 15
SPECULUM SIM'S MEDIUM 30
SPECULUM SIM'S SMALL 20
SKIN HOOKS 10
TOWEL CLIP 155
UTERINE SOUND 15
UTERINE ELEVATOR 20
SPINAL TRAY 3X4 20
CLAMPS BULLDOG CURVED 8" 2
CLAMPS BULLDOG STRAIGHT 8" 2
CLAMPS INTESTINAL CRUSHNG WITH RUBBER END10" 10
CLAMPS SATENSKYS 2
CURETTE SHARP & BLUNT 25
CLAMPS COOLIES 2
NOVAK SUCTION CURETTE LEUR254MM(10") 2
RANDALL SUCTION CURETTE 4MM DIA. LEUR 10" (254MM) 2
FG- 550-01-S SIMS UTERINE CURETTE PLIABLS SHAFT SHARP 10" (254MM) 4SETS
SET OF 3 SIZES
FG-551-01-W SMALL 1/2"(6MM)
FG-551-02-Y MEDIUM
3/2"(10MM)
FG-551-03-B LARGE 1/2"(13MM)
BRISTOL-DIRECTOR PROBEAMD RASPATORY 7.25"(184MM) 2

HLL/PCD/ESIC-20/10-11 Page 133 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

MACDONALD DISSECTOR,DOUBLE-ENDED 7.5"(191MM) 2


FISTULA DIRECTOR WITH GUARD AND FRAENUMSLIT SIN (127MM) 2
BRODIC DIRECTO PROBE-ENDED MALLEABLE 6.5"(165MM) 2
WASTON CHYNE DISSECTOR FINE WITH PROBE7.5"(191MM) 2
TRUNES WARWICK RETRACTOR BLACK FINISH 2" DIAM*1.5" WIDE 2
(51MM*38MM)
SWIFT JOLY RETRACTOR BLADE 1" *4"(25MM*102MM) 5
SWIFT JOLY RETRACTOR BLADE 1" *2.5"(25MM*64MM) 5
DEAVER ABDOMINAL RETRACTOS SET OF 5 (SCALE .25) 2
DURHAM RETRACTOR 3/8 *3/4"(10MM*19MM) BLADE 4
LONDON HOSPITAL RETRACTOR BLADE 3/4*11/4(19MM*32MM) 9"(229MM) 4
VOLKMANMM RETRACTOR 2 PRONG,BLUNT 4
3/8"*1/2"(10MM*13MM)8.5"(216MM)
VOLKMANMM RETRACTOR 4 PRONG SHARP 4
VOLKMANMM RETRACTOR 4 PRONG,BLUNT 6
BACKHAUS TOWEL CLIP 3.5"(89MM) 20PAIRS
MAYO SAFETY PIN FORCEPS HOLDER 4.5"(114MM) 25
ANEURYSM NEEDLE 8"(203MM) 3
JOLL ANEURYSM NEEDLE SHALLOW CURVE 8"(203MM) 2
NESTED TROCARS 2,5 AND 8 AND GANGE, MEKET SILVER CANNULAC, SET 2SETS
OF 3" METEL CASE 6.5"(165MM)
ADSON DISSECTING FORCEPS 6" 10
POOLE SUCTION TUBE SET WITH LAWRIE EXTENSION PIECE 16.5" 2
TINKLER (SIMPSON SMITH) SUCTION TUBE WITH ON/OFF VOLVE REMOVABLE 2
PERFORATED OUTER SHEATH AND HANDLE 5…..M*10"(150MM*255MM)
TINKLER (SIMPSON SMITH) SUCTION TUBE WITH ON/OFF VOLVE REMOVABLE 2
PERFORATED OUTER SHEATH AND HANDLE 5…..M*10"(150MM*255MM)
TINKER SUCTION TUBE 10" (254MM) CB 250 -08-Y (TINKER SUCTION TUBE 2
NEEDLE)
SAVAGE INTESTTINAL DECOMPRESSOR, TROCAR AND CIIROMIUM PLATED 2
YANKUAR SUCTION TUBE CHROMIUM PLATED 12" (305MM) 12
LAHEY CHOLECYSTOMY FORCEPS CURVED LONGITUDIND SERRATIONS 4
7,,,,,"(197MM)
MOYNIIIAN CHOLECYSTOMY FORCEPS 11.5"(292MM) 2
LOWERS CHOLECYSTOMY FORCEPS 8" (293MM) 2
DESJARDINES GALL STONE FORCEPS, SMALL,SCREW JOINT 8….(220MM) 2
DESJARDINES GALL STONE FORCEPS LARGE SCREW-JOINT 9…..'(238MM) 2
OSCHER GALL BLADDER TOCAR AND CANNULA WITHH SIDE SUCTION
CONECTOR 6MM DIAM 9…"(234MM)
OSCHER GALL BLADDER TOCAR AND CANNULA WITH SIDE SUCTION 2
CONECTOR 4.5MM DLAM 9…."(238MM)
GRAEME ANDERSON RECTO PROTOSCOPY SPECULUM DA-001-01Y 2
GABRIEL RECTAL SPECULAMDA-010-01-A 2
NAUMTON MORGEN RECTAL SPECULUM
DC -065-17-Q NAUNTON MORGEN RECTAL SPECULUM WITH ABTURALOR 2
CHILD CHROMIUM 17MM "WITH FIBES LIGHT COMM.
DC-015-20-R
WALTON RECTAL BIOPSY FORCPS WITH LATERALLY ANGLED SHANKS TO 2
IMPROVE VISION 8"(203MM)
DC-020-25L 2
YEOMAN BIOPSY FORCEPS (BASKET JAWED) TO USE THROUGH
SIGMOIDASCOPE 14"(355MM)
BARRON HAEMORRHOID LIGATOR SET COMPRISING LIGATOR 8.5" (216MM) 1
MC-GIVNEY HAEMMORRHOID SET 10.5" (272MM) STEM AND NANDLE 1
ORLAY HAEMORRHOID INSTRUMENT -DUAL PURPOSE FOR 1

HLL/PCD/ESIC-20/10-11 Page 134 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

LACKHART MUMMERY FISLULA DIRECTORS 6.5" (165MM) NO. 1. 1


STRAIGHT
NO. 2.45*CURVE
NO. 3.90*CURVE
NO. 4. 4 RETROGRADE
DD-130-14-A 2
ST. MARK'SBLACK,NYLON,HALLOW
(9/16"*3/4") (14MM*19MM) 3.5"
(89MM) DD-130-16-E 5/8"*1"(16MM*25MM)
DD-150-38-B 2
LORD ANALDILATOR 11/2"
(38MM) DIAM, 4.5"(105MM)
HED DRICK KIDNEY PEDIC CLAMP 1*2 SERRATION DOUBLE ANGLE 10.5" 2
(267MM)
MAYO CLINIC PEDICLE CLAMP SCREW JOINT 9,3/4"(248MM) 2
COOLEY RENAL ARTERY FORCEPS 7,3/4"(197MM) 2
RANDALL RENAL CALCULUS 2
FORCEPS SCREW JOINT TENESTRATED JAWS STIGHT CURVE NO.1,9.5"
(230MM)
RANDALL RENAL CALCULUS FORCEPS SCREW JOINT FENES TRATED JAWS 2
1/4
DEEP CURVE NO.2, 8 (208MM)
RANDALL RENAL CALCULUS FORCEPS SCREW JOINT 2
RANDALL RENAL CALCULUS FORCEPS TAKE APART JOINT 2
TENESTRATED JAWS DEEP RETRO CURVE NO. 4 7 1/8
TURNER WARWICK STONE FORCEP, SHALLOW CURVE FENESTRATED JAWS 2
7.5"(184MM)
EA-033-02-K 2
TURNER WARWICK RENAL STONE FORCEPS DEEP CURVE FENESTRATED
JUAWS 7 1/4 "(184MM)
EA-042-01-J TINKER URETEROTITHO TOMY FORCEPS 45* BASKET JAW BIN 2
(203MM)
EE-085-01-Y 2
INTRODUCER FOR FOLEY BALLOON CATHETERS 18" (457MM)
EF-015-01-T
TOOMY BLADDER SYRING GLASS WITH METAL NOZZLE 50
EF-100-01-X 2
SUPRAPUBLE TROCAR STRAIGHT WITH NICKEL SILVER CANNULA AND METAL
HANDLE 25FG 4.5"(114MM)
EF-150-01-A 2
HYDRO CELL TOCAR (SILVER CANNULA IN METAL NANDLE NO.3 NG GAUGO
(2.5MM DIAM) 6" (152 MM)
EH-044-04-C
TURNER WAR-WICK NEEDLE HOLDER CURVED LEFT HAND MODEL FUNSTEN 2
CURBIDE INSERT JAWS, BLACK FINISH 8.5"(216MM)
LONDON VAGINAL RETRACTOR BLADE 3.5"(89MM*25MM 2
FG-185-01-E 2
SIMS DUCKBILL SPECULAM SINGLE ENDED CHROMIUM PLATED, SMALL,
13/16"*3"(21MM*76MM) 7" (17MM)
FG-185-02-G 2
SIMS DUCKBILL SPECULAM SINGLE ENDED, CHROMIUM PLATED MEDIUM 1
1/8"*3 3/8" (29MM*86MM) 7" (178MM)
FG-185-03-J 2
SIMS DUCKBILL SPECULUM PLATED, LARGE 15/8"*3.5" (41MM*89MM) 7"
(178MM)
FG-655-01-V 4
BONNEY UTERINE POLYPUS FORCEPS 10.5"(267MM)

HLL/PCD/ESIC-20/10-11 Page 135 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

FG-660-01-E 4
VANT UTERINE POLYPUS FORCEPS 10.5"(267MM)
FG-765-01-V 1 TO 12 NO.
NAVAK SUCTION CURETTE FUER 254MM (10") 2 SET
FG-768-01-V 1 TO 12 NO.
RANDALL SUCTION CURETTE 4MM DIAM, LUER, 10"(254MM) 2 SET
FG-795-01-Y
LEECH WIKINSON CERVICAL BIOPAY PUNCH 8"(203MM) SOFT 2

FG-815-01-M 2
VAN DOREN CERVICAL BIOPAY PUNCH WITH CURVED BASKET JAW ANGLED
SHANKS 267MM(10.5")
DOYAN RETACTOR FIBRE OPTIC 1SET
2"
2.1/2"
3"
3.1/2
MORRIS RETRACTOR(FIBRE OPTIC) 4
CUSCOS SPECULUM (FIBRE OPTIC) SMALL 2SETS EACH

MEDIUM
LARGE
B.P.INSTRUMENT (PEDESTRAL) 20
DRESSING DRUM LARGE 20
DRESSING DRUM MEDIUM 25
DRESSING DRUM SMALL 15
S.STRAY BIG WITH COVER 35
S.S TRAY MEDIUM WITH COVER 45
S.S TRAY SMALL WITH COVER 30
STEEL TRAY BIG FOR RECEIVING BABY 4
STEEL BOWL LARGE 35
STEEL BOWL MEDIUM 70
STEEL BOWL SMALL 55
KIDNEY TRAY MEDIUM 60
KIDNEY TRAY SMALL 50
INHALATION KETTLE STEEL 10
ENEMA CAN STEEL 5
STEEL TRAY WITH COVER WITH STAND 10
STEEL BASIN 20
STEEL BASIN WITH STAND 30
STEEL BOWL WITH COVER SMALL 30
STEEL BOWL WITH COVER MEDIUM 30
STEEL BOWL WITH COVER BIG 35
ELECTRIC NEEDLE SYRINGE DESTROYER 8
WEIGHING MACHINE 8
SUCTION MACHINE 1/2HP, DOUBLE JAR 8
FORMALINE CHAMBER, LARGE 5
FORMALINE CHAMBER MEDIUM 2

HLL/PCD/ESIC-20/10-11 Page 136 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

Item No 93

Instruments for Dental Dept.

ADSON TISSUE FORCEPS


TOOTHED 1
PLAIN 1
AMALGAM CARRIER 2
ARTERY FORCEPS-STRAIGHT 1
B.P.HANDLE NO.3 2
BALL BURNISHER 2
BONE CHISEL 1
BONECURETTE 1
BONE FILE 1
BONE NIBBLER (RONGEUR) 1
CEMENT SPATULA 2
CHIP SYRINGE 1
CHITLE FORCEPS 1
COMPOSITE FILLING INSTRUMENT (TEFLON) 2
COTTON HOLDER 2
COTTON WASTE RECEIVER 2
DIAMOND CARVER 1
DRESSING SCISSORS-STRAIGHT 1
ELLIS TISSUE FORCEPS 1
EXCAVATOR DISCOID NO. 131 & 132 2EACH
EXCAVATOR SPOON NO. 220 & 221 2EACH
EXTRACTION FORCEPS SET 1SET
ADULT
PAEDIATRIC
GRACY CURETTE-ANTERIOR 1
POSTERIOR 1
GUTTA PERCHA PLUGGER 2
HAND SCALERS SET 1
HOLLEN BACK'S CARVER 1
MALLET (SURGICAL HAMMER) 1
MATRIX BANDS NO. 1 & 8 1PKT EACH
MATRIX RETAINER NO. 1 & 8 1EACH
MOSQUITO FORCEPS-STRAIGHT 2
MOUTH MIRROR HANDLES 12
MOUTH PROP 1
NEEDLE HOLDER 1
PARALLELOGRAM CONDENSER (PF151) 2
PERISOTEAL ELEVATORL 1DOZEN (12)
PLASTIC INSTRUMENT (PF21) 2
PROBE-(EXPLORER)NO.5 12
PROBE-STRAIGHT NO.9 6
ROOT ELEVATOR SET(WINTER CRIER TYPE) 3
ROOT ELEVATOR(STRAIGHT) (HOSPITAL PATTERN) 6
ROUND CONDENSER (PF153) 6
SPIRIT LAMP 1
SUTURE CUTTING SCISSORS 2
TOOTH EXTRACTION FORCEPS SET
A.ADULT SIZE 2SETS
B.PAEDIATRIC SIZE 1SET

HLL/PCD/ESIC-20/10-11 Page 137 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

TWIZZERS 12
UNIVERSAL SCALERS 6
WARD'S CARVER 2

Item No 94

Instruments for Eye Implant

UNIVERSAL MICRO SPRING CORNEAL SCISSORS-TITANIUM 10NO.


MICRO NEEDLE HOLDER-TITANIUM 6 NO.
LIM'S CORNEAL FORCEPS-TITANIUM 10 NO.
MACPHERSON'S FORCEPS-TITANIUM 10 NO.
LENS DIALER-TITANIUM 6 NO.
SUTURE TYING FOR CAPS WITH PLATFORM (STRAIGHT) -TITANIUM 10 NO.
VANNAS MICRO SCISSORS 4 NO.
ONE TOOTH FORCEPS 10 NO.
SEMCO CANNULA FOR SUCTION & IRRIGATION 23 GAUGE 10 NO.
UTRATA FORCEPS FOR CAPSALORHEXIS 6 NO.
AIR INJECTION CANNULAS 24 NOS
REVERSE CUTTING NEEDLES 12MM 24 NOS
SPATULATED NEEDLES 12MM 24 NOS
CONUNCTIVAL SCISSORS (SHARP TIP)(WESTCOTT TENOTOMY) STRAIGHT & 10 NOS
CURVED EACH
SMALL EYE SCISSORS4.5" (MICRO) STRAIGHT & CURVED 8 NOS EACH

EPILATION FORCEPS (CILIA) 04NO.


FOREIGN BODY SPUD 04NO.
IRRIGATION CANNULA (BISHOP HARMAN) 23,24,26(G) 12NO.
LACRIMAL SYRINGING CANNULA (BLUNT TIP) 23G 24NO.
PUNCTUM DIALATOR (WILDER) SIZE 1, 2, 3 06NO.
BIPOLAR CAUTERY FORCEPS (OPH USG) 06NO.
"J" SHAPED I/A CANNULA 0.3MM ASPIRATION PORT 02NO.
TOOTH FORCEP (CASTROVELE A SMALL ONE TOOTH WITH TYING 06NO.
PLATFORM 0.12MM)
KALT'S NEEDLE HOLDER WITH LOCK STRAIGHT 10NO.
KIRBY,S LENS EXPRESSOR HOOK 06NO.
BINOCULAR LOUPE WITH HEAD BAND BINOMAG TYPE 04NO
CHALAZION CLAMP 02NO.
NEEDLE HOLDER FOR EYE SURGERY 02NO.
LACRIMAL PROBE 02NO.
FOREIGN BODY NEEDLE 02NO.
RETROVERTED S.S NEEDLE NO 23 10NO.
EYE SHIELD PLASTIC MODE 25NO.
SHIOTZ TONOMETER 02NO.
OPHTHALMOSCOPE BULB HEINE MIROFLEX-2 BULB NO 030 03NO.

HLL/PCD/ESIC-20/10-11 Page 138 of 166 01.09.2010


HLL Lifecare Limited Technical Specifications

GENERAL TECHNICAL SPECIFICATIONS

GENERAL POINTS:
1. Warranty:
a) Comprehensive Warranty as stated in GCC clause No. 15 (in Section – IV) for complete
equipment (including X ray tubes, HT Cable, Probes, Electrodes, Detectors, Batteries for
UPS, other vacuumatic parts wherever applicable) and Turnkey Work from the date of
satisfactory installation, commissioning, trial run & handing over of equipment to Hospital/
Dispensaries/ Institute/ Medical College.
b) 98% up time Warranty of complete equipment with extension of Warranty period by double
the downtime period on 24 (hrs) X 7 (days) X 365 (days) basis.
c) All software updates should be provided free of cost during Warranty period.
2. After Sales Service:
a) After sales service centre should be available at the city of Hospital/Institution/Medical
College on 24 (hrs) X 7 (days) X 365 (days) basis. Complaints should be attended properly,
maximum within 8 hrs. The service should be provided directly by Tenderer/Indian Agent.
Undertaking by the Principals that the spares for the equipment shall be available for at least
10 years from the date of supply.
b) The service provider should have the necessary equipments recommended by the
manufacturer to carry out preventive maintenance test as per guidelines provided in the
service/maintenance manual.
3. Training:
On Site training to Doctors/ Technicians/ staff is to be provided by Principal/ Indian Agents (if
they have the requisite know-how) for operation and maintenance of the equipment to the
satisfaction of the consignee.
4. Annual Comprehensive Maintenance Contract (CMC) of subject equipment with Turnkey:
a) The cost of Annual Comprehensive Maintenance Contract (CMC) which includes
preventive maintenance including testing & calibration as per technical/ service /operational
manual of the manufacturer, labour and spares, after satisfactory completion of Warranty
period to be quoted for next 5 years on yearly basis for complete equipment (including X
ray tubes, HT Cable, Probes, Electrodes, Detectors, Batteries for UPS, other vacuumatic parts
wherever applicable) and Turnkey (if any). The supplier shall visit each consignee site as
recommended in the manufacturer’s technical/ service /operational manual, but at least once
in six months during the CMC period
b) The cost of CMC to be quoted along with taxes applicable on the date of Tender Opening.
The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the
price will be taken inclusive of such taxes and no claim for the same will be entertained later.
c) Cost of CMC will be added for Ranking/Evaluation purpose.
d) The payment of CMC will be made on six monthly basis after satisfactory completion of said
period, duly certified by end user on receipt of bank guarantee for 2.5 % of the cost of the
equipment as per Section XV valid till 2 months after expiry of entire CMC period.
e) There will be 98% uptime warranty during CMC period on 24 (hrs) X 7 (days) X 365 (days)
basis, with penalty, to extend CMC period by double the downtime period.
f) During CMC period, the supplier is required to visit at each consignee’s site at least once in 6
months commencing from the date of the successful completion of warranty period for
preventive maintenance of the goods.
g) All software updates should be provided free of cost during CMC.
HLL/PCD/ESIC-20/10-11 Page 139 of 166 01.09.2010
HLL Lifecare Limited Technical Specifications

h) Failure of the above [4. e) to 4. g)] by the supplier, may lead to the forfeiture of the Bank
Guarantee for Annual CMC.
i) The payment of CMC will be made as stipulated in GCC Clause 21.

5. Turnkey:
Turnkey is indicated in the technical specification of the respective items, wherever required.
The Tenderer shall examine the existing site where the equipment is to be installed, in
consultation with HOD of Hospital/Institution/Medical College concerned. Turnkey details of
each Hospital/Institution/Medical College are given at the end of Technical Specification. The
Tenderers to quote prices indicating break-up of prices of the Machine and Turnkey Job of each
Hospital/Institution/Medical College. The Turnkey costs (wherever applicable) may be
quoted in Indian Rupee will be added for Ranking Purpose.

The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the
price will be taken inclusive of such duties and taxes and no claim for the same will be
entertained later.

The Turnkey Work should completely comply with AERB requirement, if any.

HLL/PCD/ESIC-20/10-11 Page 140 of 166 01.09.2010


HLL Lifecare Limited QC Requirement

Section – VIII
Quality Control Requirements
(Proforma for equipment and quality control employed by the manufacturer(s)

Tender Reference No.


Date of opening
Time
Name and address of the Tenderer:

Note: All the following details shall relate to the manufacturer(s) for the goods quoted for.

01 Name of the manufacturer


a. full postal address
b. full address of the premises
c. Email ID
d. telephone number
e. fax number

02 Plant and machinery details

03 Manufacturing process details

04 Monthly (single shift) production capacity of goods quoted for


a. normal
b. maximum

05 Total annual turn-over (value in Rupees)

06 Quality control arrangement details


a. for incoming materials and bought-out components
b. for process control
c. for final product evaluation

07 Test certificate held


a. type test
b. BIS/ISO certification
c. any other

08 Details of staff
a. technical
b. skilled
c. unskilled

Signature and seal of the Tenderer

HLL/PCD/ESIC-20/10-11 Page 141 of 166 01.09.2010


HLL Lifecare Limited Qualification Criteria

Section – IX
Qualification Criteria

01. The Tenderer must be a Manufacturer or its authorized Agent.


02. (a) The Manufacturer should have supplied and installed in last Five years from the date of
Tender Opening, atleast 100% of the quoted quantity of the similar equipment meeting major
specification parameters which is functoning satisfactorily. The foreign Manufacturer
satisfying the above criteria should also have supplied and installed in last Five years from
the date of Tender Opening, at least 50% (or one No. where the schedule of requirement is
one no.) of quoted quantity of similar model which is functoning satisfactorily any where
outside the country of manufacture.
02. (b) The Tenderers quoting as authorized representative of the manufacturer meeting the above
criteria 02 (a) should have supplied and installed in last Five years from the date of Tender
Opening, atleast 50% (or one No. where the schedule of requirement is one no. ) of the
quoted quantity of similar equipment which is functoning satisfactorily, any where in India
of the same manufacturer.

Note:

1. In support of 2 (a) & 2 (b), the Tenderer shall furnish Performance statement in the enclosed
Proforma ‘A’.

The manufacturer as well as the Tenderer/ Indian Agent shall furnish Satisfactory Performance
Certificate in respect of above, duly translated in English and duly notarized, alongwith the
tender. The performance certificate should be in accordance with the requirement of clause 2 (a) /
2 (b) as stated above.

2. The Tenderer shall furnish a brief write-up, packed with adequate data explaining and establishing
his available capacity/capability (both technical and financial) to perform the Contract (if
awarded) within the stipulated time period, after meeting all its current/present commitments. The
Tenderer shall also furnish details of Equipment and Quality Control in the enclosed Section VIII.

3. Notwithstanding anything stated above, the Purchaser reserves the right to assess the Tenderer’s
capability and capacity to perform the contract satisfactorily before deciding on award of
Contract, should circumstances warrant such an assessment in the overall interest of the
Purchaser.

4. The Purchaser reserves the right to ask for a free demonstration of the quoted equipment at a pre
determined place acceptable to the purchaser for technical acceptability as per the tender
specifications, before the opening of the Price Tender.

HLL/PCD/ESIC-20/10-11 Page 142 of 166 01.09.2010


HLL Lifecare Limited Qualification Criteria

PROFORMA ‘A’
PROFORMA FOR PERFORMANCE STATEMENT
(For the period of last five years)

Tender Reference No.: _________________________________

Date & Time of opening: _________________________________

Name and address of the Tenderer: _________________________________

Name and address of the manufacturer: _________________________________

Date of Remarks Have the


Purchaser/Consig
Order placed by

completion of indicating goods been


(full address of

Value of order
Description of
ordered goods

ordered goods
Order number

Contract reasons for functioning


and services

and services
Quantity of

delay if Satisfactorily
and date

contract
any (attach

Actual
As per
(Rs.)

documentary
nee)

proof)**
1 2 3 4 5 6 7 8 9

Signature and seal of the Tenderer

** The documentary proof will be a certificate from the consignee/end user with cross-
reference of order no. and date in the certificate duly notarised certification authenticating
the correctness of the information furnished. If at any time, information furnished is proved
to be false or incorrect, the earnest money and or performance security furnished will be
forfeited. Such certificates from a third party or middleman other than actual end user will
not be accepted. The satisfactory performance implies working satisfactorily without any
complaint since the date of installation, commissioning & handing over to the end user as
per the standard format enclosed.

HLL/PCD/ESIC-20/10-11 Page 143 of 166 01.09.2010


HLL Lifecare Limited Tender Form

Section – X
TENDER FORM

Date__________
To,
Head (P & CD)
HLL Lifecare Limited
Procurement and Consultancy Division
B-14A, Sector -62, Noida -201307, Uttar Pradesh

Ref. Your TE document No. _________dated ___________

We, the undersigned have examined the above mentioned TE document, including
amendment/corrigendum No. __________, dated ________ (if any), the receipt of which is hereby
confirmed. We now offer to supply and deliver___________ (Description of goods and services) in
conformity with your above referred document.

If our tender is accepted, we undertake to supply the goods and perform the services as mentioned
above, in accordance with the delivery schedule specified in the List of Requirements.

We further confirm that, if our tender is accepted, we shall provide you with a performance security of
required amount in an acceptable form in terms of GCC clause 5, read with modification, if any, in
Section - V – “Special Conditions of Contract”, for due performance of the contract.

We agree to keep our tender valid for acceptance as required in the GIT clause 20, read with
modification, if any in Section - III – “Special Instructions to Tenderers” or for subsequently extended
period, if any, agreed to by us. We also accordingly confirm to abide by this tender up to the aforesaid
period and this tender may be accepted any time before the expiry of the aforesaid period. We further
confirm that, until a formal contract is executed, this tender read with your written acceptance thereof
within the aforesaid period shall constitute a binding contract between us.

We further understand that you are not bound to accept the lowest or any tender you may receive
against your above-referred tender enquiry.

We confirm that we do not stand deregistered/banned/blacklisted by any Govt. Authorities.

We confirm that we fully agree to the terms and conditions specified in above mentioned TE
document, including amendment/ corrigendum if any

(Signature with date)

(Name and designation) Duly authorised to sign tender for and on behalf of

HLL/PCD/ESIC-20/10-11 Page 144 of 166 01.09.2010


HLL Lifecare Limited Price Schedule

SECTION – XI PRICE SCHEDULE


A) PRICE SCHEDULE FOR DOMESTIC GOODS OR GOODS OF FOREIGN ORIGIN LOCATED WITHIN INDIA

1 2 3 4 5 6
Price per unit (Rs.)

Incidental Services
Ex - factory/ Ex - Excise Sales Tax/ Transportation, Insurance charges for
Item Brief Description Country of Quantity (including Installation & Unit Price (at Total Price (at
warehouse /Ex- Duty (if VAT (if loading/ a period including 3
Sl. of Goods (with Origin (Nos.) Commissioning, Consignee Consignee
showroom /Off - any) any) unloading and months beyond the
No. make & model) Supervision, Site) basis Site) basis
the shelf [%age & [%age & Incidental costs date of delivery
Demonstration and (Rs.)
value] value] till consignee’s
Training) at the
site
Consignee’s site (g)
(a) (b) (c) (d) (e)
(f) =a+b+c+d+e+f 4 x 5(g)

Total Tender price in Rupees: ___________________________________________________________________________________________________


In words: ____________________________________________________________________________________________________________________________
Note: -
1. If there is a discrepancy between the unit price and total price THE UNIT PRICE shall prevail.
2. The charges for Annual CMC after warranty shall be quoted separately as per Section-XI – Price Schedule C
Name________________________
Business Address________________________
Place: ___________________________ Signature of Tenderer________________________
Date: _________________________ Seal of the Tenderer________________________

HLL/PCD/ESIC-20/10-11 Page 145 of 166 01.09.2010


HLL Lifecare Limited Price Schedule

B) PRICE SCHEDULE FOR GOODS TO BE IMPORTED FROM ABROAD


1 2 3 4 5
Price per unit (Currency)
Loading/ Unit price on DDP
Custom
Unloading, Installation basis at consignee’s
Net Duty
Brief Gross FOB price inland commissioning site
Amount and CIP by amount as Custom
Item Description of at sea/air port of Net FOB transportation, , supervision.
percentage of Air/ % of Net Clearance
Goods (with Country Quantity Lading (inclusive (excluding insurance as Demonstration
Sl. of Origin (Nos.) Agency Insurance Sea at CIP &
of Agency Agency per Clause 11 & training at
No. make & Commission & Freight the (amount Handling In foreign
Commission) Commission) of GCC & the In Indian
model) **
(a-b)
port of with Charges
incidental cost consignee’s currency
entry CDEC as ** Rupees
till consignee’s site
(c+d) applicable)
site **
**
**
=(b+f+g+
(a) (b) (c) (d) (e) (f) (g) (h) (i) =(e)
h+i)
** to be quoted in Indian Currency

Total price at Consignee’s site


(A) In foreign currency : column (4 x e)____________________________________________________________________________ (In figures and words) plus
** (B) In Indian Rupees : column 4 x (b+f+g+h+i) Rs__________________________________________________________________(In figures and words)
________________________ _______________________________________________________________________________
Note: -
1. The Tenderer will be fully responsible for the safe arrival of the goods at the consignee site in good condition as per terms of contract.
2. The bidders break up of prices under various columns is for comparison of prices up to delivery of goods at consignee’s site for tender evaluation.
3. The quoted price should be supported with original proforma invoice from the foreign manufacturers. The proforma invoice should indicate the percentage of agency commission included in the FOB
prices. Indian Agent to be paid in Indian Currency.
4. All the components of the DDP price will be paid by the tenderer. The purchaser will make the payment of DDP price after receipt of goods at consignee’s site in good condition as per payment terms
in the contract.
5. The prices quoted in foreign currency in column (e) shall be converted in Rupees at the selling rate of exchange applicable on the date of tender opening .The customs duty amount so worked out as
percentage of net CIP value in rupees will be taken for evaluation and comparison of tenders
6. The charges for Annual CMC after warranty shall be quoted separately as per Section-XI – Price Schedule C

Name_____________________________
Business address____________________
Place:________________________________________ Signature of Tenderer___________________

Date:_____________________________________ Seal of Tenderer________________________

HLL/PCD/ESIC-20/10-11 Page 146 of 166 01.09.2010


HLL Lifecare Limited Price Schedule

C) PRICE SCHEDULE FOR COMPREHENSIVE MAINTENANCE CONTRACT AFTER WARRANTY PERIOD


1 2 3 4 5
Comprehensive Maintenance Contract
Item Cost for Each Unit year wise*. Total Comprehensive Maintenance
Brief Description of the Quantity
Sl. Contract Cost for 5 (or as specified) Years
Goods (Nos.) 1st 2nd 3rd 4th 5th
No. [3 x (4a+4b+4c+4d+4e)]
a b c d e

* After completion of Warranty period


NOTE:-
1. In case of discrepancy between unit price and total prices, THE UNIT PRICE shall prevail.
2. The cost of Comprehensive Maintenance Contract (CMC) which includes preventive maintenance including testing & calibration as per technical/
service /operational manual and labour, after satisfactory completion of Warranty period may be quoted for next 5 (or as specified) years on yearly
basis for complete equipment and Turnkey (if any).
3. The cost of CMC may be quoted along with taxes applicable on the date of Tender Opening. The taxes to be paid extra, to be specifically stated. In
the absence of any such stipulation the price will be taken inclusive of such taxes and no claim for the same will be entertained later.
4. Cost of CMC will be added for Ranking/Evaluation purpose.
5. The payment of CMC will be made as per clause GCC clause 21.1 (D).
6. The uptime warranty will be 98 % on 24 (hrs) x 7 (days) x 365 (days) basis or as stated in Technical Specification of the TE document.
7. All software updates should be provided free of cost during CMC period.
8. The stipulations in Technical Specification will supersede above provisions
9. The supplier shall keep sufficient stock of spares required during Comprehensive Maintenance Contract period. In case the spares are required to
be imported, it would be the responsibility of the supplier to import and get them custom cleared and pay all necessary duties.
Name________________________
Business Address________________________
Place: ___________________________ Signature of Tenderer________________________
Date: _________________________ Seal of the Tenderer________________________

HLL/PCD/ESIC-20/10-11 Page 147 of 166 01.09.2010


HLL Lifecare Limited Price Schedule

D) PRICE SCHEDULE FOR TURNKEY

Item BRIEF TURNKEY DESCRIPTION CONSIGNEE Turnkey price


Sl. No. OF GOODS CODE

Note: -

1. The cost of Turnkey as per Technical Specification (Section VII) may be quoted on lump sum along with taxes applicable on the date of Tender
Opening. The taxes to be paid extra, to be specifically stated. In the absence of any such stipulation the price will be taken inclusive of such taxes and no
claim for the same will be entertained later.
2. Cost of Turnkey will be added for Ranking/Evaluation purpose.
3. The payment of Turnkey will be made as per clause GCC clause 21.1 (c).
4. The stipulations in Technical Specification will supersede above provisions

Name________________________
Business Address________________________
Place: ___________________________ Signature of Tenderer________________________
Date: _________________________ Seal of the Tenderer________________________

HLL/PCD/ESIC-20/10-11 Page 148 of 166 01.09.2010


HLL Lifecare Limited Questionnaire

SECTION – XII
QUESTIONNAIRE

Fill up the Section XX – Check List for Tenderers and enclose with the Tender
1. The tenderer should furnish specific answers (alongwith mention of relevant page nos. of
tender) to all the questions/issues mentioned in the Checklist. In case a question/issue does not
apply to a tenderer, the same should be answered with the remark “not applicable”

2. Wherever necessary and applicable, the tenderer shall enclose certified copy as documentary
proof/ evidence to substantiate the corresponding statement.

3. In case a tenderer furnishes a wrong or evasive answer against any of the question/issues
mentioned in the Checklist, its tender will be liable to be ignored.

HLL/PCD/ESIC-20/10-11 Page 149 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

SECTION – XIV
MANUFACTURER’S AUTHORISATION FORM

To,
Head (P & CD)
HLL Lifecare Limited
Procurement and Consultancy Division
B-14A, Sector -62, Noida -201307, Uttar Pradesh

Dear Sir,

Ref. Your TE document No ____________, dated _____________

We, ___________________________________ who are proven and reputable manufacturers


of___________________________(name and description of the goods offered in the tender) having
factories at___________________________________________________________, hereby authorise
Messrs______________________________(name and address of the agent) to submit a tender,
process the same further and enter into a contract with you against your requirement as contained in the
above referred TE documents for the above goods manufactured by us.

We further confirm that no supplier or firm or individual other than Messrs.


________________________ (name and address of the above agent) is authorised to submit a tender,
process the same further and enter into a contract with you against your requirement as contained in the
above referred TE documents for the above goods manufactured by us.

We also hereby extend our full warranty, CMC as applicable as per clause 15 of the General
Conditions of Contract, read with modification, if any, in the Special Conditions of Contract for the
goods and services offered for supply by the above firm against this TE document.
Yours faithfully,

[Signature with date, name and designation]


for and on behalf of Messrs___________________________

[Name & address of the manufacturers]

Note: 1. This letter of authorisation should be on the letterhead of the manufacturing firm and should
be signed by a person competent and having the power of attorney to legally bind the
manufacturer.
2. Original letter may be sent.

HLL/PCD/ESIC-20/10-11 Page 150 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

SECTION – XV

BANK GUARANTEE FORM FOR PERFORMANCE SECURITY/ CMC


SECURITY
To
Head of Hospital/Institute/Medical College of ESIC

WHEREAS _____________________________ (Name and address of the supplier) (Hereinafter


called “the supplier”) has undertaken, in pursuance of contract no________________________ dated
_____________ to supply (description of goods and services) (herein after called “the contract”).

AND WHEREAS it has been stipulated by you in the said contract that the supplier shall furnish you
with a bank guarantee by a scheduled commercial bank recognised by you for the sum specified
therein as security for compliance with its obligations in accordance with the contract;
AND WHEREAS we have agreed to give the supplier such a bank guarantee;

NOW THEREFORE we hereby affirm that we are guarantors and responsible to you, on behalf of the
supplier, up to a total of. ________________________ (Amount of the guarantee in words and
figures), and we undertake to pay you, upon your first written demand declaring the supplier to be in
default under the contract and without cavil or argument, any sum or sums within the limits of
(amount of guarantee) as aforesaid, without your needing to prove or to show grounds or reasons for
your demand or the sum specified therein.

We hereby waive the necessity of your demanding the said debt from the supplier before presenting
us with the demand.

We further agree that no change or addition to or other modification of the terms of the contract to be
performed there under or of any of the contract documents which may be made between you and the
supplier shall in any way release us from any liability under this guarantee and we hereby waive
notice of any such change, addition or modification.

This guarantee shall remain valid till 2 (two) months beyond the warranty period, i.e up to -----------
(indicate date)

…………………………….
(Signature with date of the authorised officer of the Bank)
………………………………………………………….
Name and designation of the officer
………………………………………………………….
………………………………………………………….
Seal, name & address of the Bank and address of the Branch

HLL/PCD/ESIC-20/10-11 Page 151 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

SECTION – XVI
CONTRACT FORM - A
CONTRACT FORM FOR SUPPLY, INSTALLATION, COMMISSIONING, HANDING
OVER, TRIAL RUN, TRAINING OF OPERATORS & WARRANTY OF GOODS

(Address of the Purchaser’s/Consignee’s


office issuing the contract)

Contract No___________ dated______________


This is in continuation to this office’s Notification of Award No_______ dated ______
1. Name & address of the Supplier: ______________________________
2. Purchaser’s TE document No________ dated____________ and subsequent Amendment
No____________, dated_________ (if any), issued by the purchaser
3. Supplier’s Tender No_________ dated__________ and subsequent communication(s)
No____________ dated _________ (if any), exchanged between the supplier and the purchaser in
connection with this tender.
4. In addition to this Contract Form, the following documents etc, which are included in the documents
mentioned under paragraphs 2 and 3 above, shall also be deemed to form and be read and construed
as integral part of this contract:

(i) General Conditions of Contract;


(ii) Special Conditions of Contract;
(iii) List of Requirements;
(iv) Technical Specifications;
(v) Quality Control Requirements;
(vi) Tender Form furnished by the supplier;
(vii) Price Schedule(s) furnished by the supplier in its tender;
(viii) Manufacturers’ Authorisation Form (if applicable for this tender);
(ix) Purchaser’s Notification of Award

Note : The words and expressions used in this contract shall have the same meanings as are
respectively assigned to them in the conditions of contract referred to above. Further, the definitions
and abbreviations incorporated under clause 1 of Section II – ‘General Instructions to Tenderers’ of
the Purchaser’s TE document shall also apply to this contract.
5. Some terms, conditions, stipulations etc. out of the above-referred documents are reproduced below
for ready reference:
(i) Brief particulars of the goods and services which shall be supplied/ provided by the supplier
are as under:
Item
Brief description of Accounting Quantity to Unit Total Terms of
Sl.
goods/services unit be supplied Price price delivery
No.

HLL/PCD/ESIC-20/10-11 Page 152 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

Any other additional services (if applicable) and cost thereof: _____________________
Total value (in figure) ____________ (In words) ___________________________
(ii) Delivery schedule
(iii) Details of Performance Security
(iv) Quality Control
(a) Mode(s), stage(s) and place(s) of conducting inspections and tests.
(b) Designation and address of purchaser’s inspecting officer
(v) Destination and despatch instructions
(vi) Consignee, including port consignee, if any
(vii) Warranty clause
(viii) Payment terms
(ix) Paying authority
____________________________
(Signature, name and address
of the Purchaser’s/Consignee’s authorised official)
For and on behalf of__________________
Received and accepted this contract
(Signature, name and address of the supplier’s executive
duly authorised to sign on behalf of the supplier)
For and on behalf of _________________________
(Name and address of the supplier)
(Seal of the supplier)

Date: _________________________

Place: _________________________

HLL/PCD/ESIC-20/10-11 Page 153 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

CONTRACT FORM – B
CONTRACT FORM FOR COMPREHENSIVE MAINTENANCE CONTRACT
Comprehensive Maintenance Contract No._______________________dated_________________
Between

(Address of Head of Hospital/Institute/Medical College)


And

(Name & Address of the Supplier)


Ref: Contract No___________ dated______________ (Contract No. & date of Contract for
supply, installation, commissioning, handing over, Trial run, Training of operators &
warranty of goods)
In continuation to the above referred contract

a) The Contract of Comprehensive Maintenance is hereby concluded as under: -


1 2 3 4 5
Comprehensive
Maintenance Contract Total Comprehensive
BRIEF Cost for Each Unit year Maintenance Contract
Item Quantity
DESCRIPTION OF wise*. Cost for 5 (or as
Sl. No. (Nos.)
GOODS specified) Years
1st 2nd 3rd 4th 5th
[3 x (4a+4b+4c+4d+4e)]
a b c d e

Total value (in figure) ____________ (In words) ___________________________


b) The CMC commence from the date of expiry of all obligations under Warranty i.e.
from______________ (date of expiry of Warranty) and will expire on ______________ (date
of expiry of CMC)
c) The cost of Comprehensive Maintenance Contract (CMC) which includes preventive
maintenance and labour, after satisfactory completion of Warranty period may be quoted for
next 5 (or as specified) years as contained in the above referred contract on yearly basis for
complete equipment (including X ray tubes, Batteries for UPS, other vacummatic parts, _____
& _____) and Turnkey (if any).
d) There will be 98% uptime warranty during CMC period on 24 (hrs) X 7 (days) X 365 (days)
basis, with penalty, to extend CMC period by double the downtime period.
e) During CMC period, the supplier shall visit at each consignee’s site for preventive maintenance
including testing and calibration as per the manufacturer’s service/ technical/ operational
manual. The supplier shall visit each consignee site as recommended in the manufacturer’s
manual, but at least once in 6 months commencing from the date of the successful completion
of warranty period for preventive maintenance of the goods.
f) All software updates should be provided free of cost during CMC.
g) The bank guarantee valid till ______________ [(fill the date) 2 months after expiry of entire
CMC period] for an amount of Rs. _______________ [(fill amount) equivalent to 2.5 % of the
cost of the equipment as per contract] shall be furnished in the prescribed format given in

HLL/PCD/ESIC-20/10-11 Page 154 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

Section XV of the TE document, along with the signed copy of CMC within a period of 21
(twenty one) days of issue of CMC failing which the proceeds of Performance Security shall be
payable to the Purchaser/Consignee.
h) If there is any lapse in the performance of the CMC as per contract, the proceeds CMC bank
guarantee for an amount of Rs. __________ (equivalent to 2.5 % of the cost of the
equipment as per contract) shall be payable to the Consignee.
i) Payment terms: The payment of CMC will be made against the bills raised to the
consignee by the supplier on six monthly basis after satisfactory completion of said period,
duly certified by the HOD concerned. The payment will be made in Indian Rupees.
j) Paying authority: ______________________ (name of the consignee i.e. Hospital/
Institute /Medical College’s authorised
official)
____________________________
(Signature, name and address of
Hospital/Institute/Medical College’s authorised official)
For and on behalf of__________________
Received and accepted this contract
(Signature, name and address of the supplier’s executive
duly authorised to sign on behalf of the supplier)
For and on behalf of _________________________
(Name and address of the supplier)
(Seal of the supplier)

Date: _________________________

Place: _________________________

HLL/PCD/ESIC-20/10-11 Page 155 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

SECTION – XVII

CONSIGNEE RECEIPT CERTIFICATE


(To be given by consignee’s authorized representative)

The following store (s) has/have been received in good condition:


1) Contract No. & date :______________________________
2) Supplier’s Name :______________________________

3) Consignee’s Name & Address with


telephone No. & Fax No. : ______________________________

4) Name of the item supplied :______________________________

5) Quantity Supplied :______________________________


6) Date of Receipt by the Consignee :______________________________
7) Name and designation of Authorized
Representative of Consignee :______________________________
8) Signature of Authorized
Representative of Consignee with
date :______________________________

9) Seal of the Consignee :______________________________

HLL/PCD/ESIC-20/10-11 Page 156 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

SECTION – XVIII
Proforma of Final Acceptance Certificate by the Consignee
No _______________ Date_______________

To
M/s____________________
_______________________
_______________________

Subject: Certificate of commissioning of equipment/plant.

This is to certify that the equipment(s)/plant(s) as detailed below has/have been received in
good conditions along with all the standard and special accessories and a set of spares
(subject to remarks in Para no.02) in accordance with the contract/technical specifications.
The same has been installed and commissioned.
(a) Contract No______________________________________ dated__________________
(b) Description of the equipment(s)/plants: _______________________________________
(c) Equipment(s)/ plant(s) nos.: ________________________________________________
(d) Quantity:_______________________________________________________________
(e) Bill of Loading/Air Way Bill/Railway
Receipt/ Goods Consignment Note no___________________ dated _________________
(f) Name of the vessel/ Transporter: ___________________________________________
(g) Name of the Consignee: ___________________________________________________
(h) Date of commissioning and proving test: _____________________________________

Details of accessories/spares not yet supplied and recoveries to be made on that account.
Sl. Description of Item Quantity Amount to be recovered
No.

The proving test has been done to our entire satisfaction and operators have been trained to operate
the equipment(s)/plant(s).
The supplier has fulfilled its contractual obligations satisfactorily ## or
The supplier has failed to fulfil its contractual obligations with regard to the following:
He has not adhered to the time schedule specified in the contract in dispatching the
documents/drawings pursuant to ‘Technical Specifications’.
He has not supervised the commissioning of the equipment(s)/plant(s)in time, i.e. within the period
specified in the contract from date of intimation by the Purchaser/Consignee in respect of the
installation of the equipment(s)/plant(s).
The supplier as specified in the contract has not done training of personnel.

The extent of delay for each of the activities to be performed by the supplier in terms of the contract

HLL/PCD/ESIC-20/10-11 Page 157 of 166 01.09.2010


HLL Lifecare Ltd. Various Forms

is______________________________________________________________________________.
The amount of recovery on account of non-supply of accessories and spares is given under Para
no.02.
The amount of recovery on account of failure of the supplier to meet his contractual obligations
is__________________ (here indicate the amount).

Signature
Name:
Designation with stamp

## Explanatory notes for filling up the certificate:

o He has adhered to the time schedule specified in the contract in dispatching the
documents/drawings pursuant to ‘Technical Specification’.
o He has supervised the commissioning of the equipment(s)/plant(s) in time, i.e. within the
time specified in the contract from date of intimation by the Purchaser/Consignee in respect
of the installation of the equipment(s)/plant(s).
o Training of personnel has been done by the supplier as specified in the contract
o In the event of documents/drawings having not been supplied or installation and
commissioning of the equipment(s)/plant(s) having been delayed on account of the supplier,
the extent of delay should always be mentioned in clear terms.

HLL/PCD/ESIC-20/10-11 Page 158 of 166 01.09.2010


HLL Lifecare Ltd. Shipping Arrangement

SECTION – XIX

DETAILS OF SHIPPING ARRANGEMENT FOR LINER CARGOES IN RESPECT OF


C&F/CIF/TURNKEY/F.O.R CONTRACTS FOR IMPORTS

1. SHIPMENT FROM PORTS OF U.K INCLUDING NORTHERN IRELAND (ALSO EIRE), FROM
THE NORTH CONTINENT OF EUROPE (GERMANY, HOLLAND, BELGIUM, FRANCE,
NORWAY, SWEDEN, DENMARK, FINLAND AND PORTS ON THE CONTINENTAL SEABOARD
OF MEDITERRANIAN (I.E. FRENCH WESTERN ITALIAN PORTS), TO PORTS IN INDIA.

The Seller should arrange shipment of the goods by vessels belonging to the member lines of the India-
Pakistan-Bangladesh Conference. If the Seller finds that the space on the ‘Conference Lines’ vessels is
not available for any specific shipment, he should take up with India-Pakistan-Bangladesh Conference.
Conferity House, East Grinstead, Sussex (UK), for providing shipping space and also inform the Shipping
Co-ordination Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW
DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)
The Seller should arrange shipment through the Government of India’s Forwarding Agents, M/s Schenker
& Co., 2000-Hamburg (Cable: SCHENKER CO., HAMBURG) OR obtain a certificate from them to the
effect that shipment has been arranged in accordance with instructions of the Ministry of Surface
Transport, (TRANSCHART), New Delhi.
2. SHIPMENT FORM PORTS OF U.K. INCLUDING NORTHERN

Goods under this contract would be shipped by the national shipping companies of the Contracting Parties
operating bilateral shipping service and vessels under the flag of third countries in accordance with the
Agreement between the Government of German Democratic Republic and the Government of the
Republic of India in the Field of Merchant Shipping signed on 9.1.1979, as amended up-to-date.
3. ISHIPMENT FROM ADRIATIC PORTS OF EASTERN ITALY AND YUGOSLAVIA

The seller should arrange shipment of the goods by vessels belonging to the following Indian member
lines;
1. The Shipping Purchaser of India Ltd.
2. The Scindia Steam Navigation Co., Ltd
3. India Steamship Co., Ltd

For the purpose of ascertaining the availability of suitable Indian vessels and granting dispensation in the
event of their non-availability, the Seller should give adequate notice about the readiness of each
consignment from time to time at least six weeks in advance of the required position to M/s Schenker &
Co. 2000 HAMBURG (Cable: SCHENKER CO., HAMBURG) and also endorse a copy thereof to the
Shipping Co-ordination Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART,
NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)
The seller should arrange shipment through the Government of India’s Forwarding Agents M/s Schenker
& Co. 2000 HAMBURG (Cable: SCHENKER CO., HAMBURG) or obtain certificate from them to the
effect that shipment has been arranged in accordance with the instructions of the Ministry of Surface
Transport, (TRANSCHART), New Delhi.

4. SHIPMENT FROM POLAND & CZECHOSLOVAKIA

(i) IMPORTS FROM POLAND

HLL/PCD/ESIC-20/10-11 Page 159 of 166 01.09.2010


HLL Lifecare Ltd. Shipping Arrangement

Shipment under this contract would be made by the National flag lines of the two parties and vessels of
the third flag conference lines, in accordance with the agreement between the Govt. of the Republic of
India and the Govt. of the Polish People’s Republic regarding Shipping Co-operation dated 27.6.1960 as
amended up-to-date.

(ii) IMPORTS FROM CZECHOSLOVAKIA

Goods under this contract would be signed by the National flag lines of the two parties and vessels of the
third flag conference lines, in accordance with the Agreement Co-operation in shipping between India and
Czechoslovakia signed on 3.11.1978 and ratified on 19.12.1979, as amended up-to-date.
Shipping arrangement should be made by the Sellers in consultation with Resident Representative of the
Indian Shipping Lines in Gdynia, Co., Morska Agencja W. Gdyniul, Pulaskiego 8, P.O. Box 246, Gdynia
(Poland) – Telex : MG PL. 054301, Tel.: 207621, to whom details regarding contract number, nature of
cargo , quantity, port of lading, discharging, name of Government consignee, expected date of readiness
of each consignment etc. should be furnish at least six weeks in advance of the required position, with a
copy thereof endorsed to the Shipping Co-ordination Officer, Ministry of Surface Transport, (Chartering
Wing), New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158,
61159)

5. SHIPMENT FROM U.S.S.R

Shipment under this contract should be made in accordance with the agreement between the Government
of the Republic of India and the Government of U.S.S.R on Merchant Shipping 1976, as amended up-to-
date, by vessels of Indo-Soviet shipping Service.

6. SHIPMENT FROM JAPAN

The shipment of goods should be made of India vessels to the maximum extent possible subject to the
minimum of 50%.
The Seller should arrange shipment of the goods in consultation with the Embassy of India in Japan,
Tokyo to whom details regarding contract number, nature of cargo, quantity, port of loading/discharge,
name of Govt. consignee, expected date of readiness of each consignment etc. should be furnished at least
six weeks in advance of the required position.
Note: The copies of such contracts are to be endorsed both to the Attached (commercial) embassy of
India in Japan, Tokyo, and the shipping Co-ordination Officer, Ministry of Surface Transport, New Delhi.

7. SHIPMENT FROM AUSTRALIA, ALGERIA, BULGARIA, ROMANIA, EGYPY

The Seller shall arrange shipment of the goods by Indian flag vessels to the maximum extent possible
subject to a minimum of 50 %. For the purpose of ascertaining the availability of suitable Indian vessels,
the seller shall give adequate notice of not less than six weeks about the readiness of each consignment to
the Shipping Purchaser of India Ltd., SHIPPING HOUSE, 245, Madame Cama Road, Bombay – 400 021
(CABLE: SHIPINDIA BOMBAY) and also endorse a copy thereof to the Shipping Co-ordination
Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex:
VAHAN IN – 031 – 61157, 61158, 61159)
8. SHIPMENT FROM PAKISTAN

HLL/PCD/ESIC-20/10-11 Page 160 of 166 01.09.2010


HLL Lifecare Ltd. Shipping Arrangement

The shipment of cargoes should be made by Indian vessels to the maximum extent possible subject to a
minimum of 50 %.
Shipment arrangement should be made by the sellers in consultation with M/s Mogul Line Ltd., 16-Bank
Street, Fort, Bombay – 400023 (Cable: MOGUL BOMBAY: Telex: 011 – 4049 MOGUL), to whom,
details regarding contract number, nature of cargo, quantity, port of lading discharging, name of
government consignee, expected date of readiness of each consignment etc. should be furnish at least six
weeks in advance of the required position, with a copy thereof endorsed to the Shipping Co-ordination
Officer, Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex:
VAHAN IN – 031 – 61157, 61158, 61159)
9. SHIPMENT FROM U.S ATLANTIC & GULF PORTS

The Seller should arrange shipment of the goods by vessels belonging to the member lines of the India –
Pakistan – Bangladesh – Ceylon and Burma Outward Freight Conference. If the Seller finds that the space
of the ‘Conference Lines’ vessels is not available for any specific shipment he should take up with India –
Pakistan- Bangladesh – Ceylon and Burma Outward Freight Conference, 19, Rector Street, New York,
N.Y. 10006 USA, for providing shipping space and also inform the Shipping Co-ordination Officer,
Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN –
031 – 61157, 61158, 61159)

10. SHIPMENT FROM ST. LAWRENCE AN EASTERN CANADIAN PORTS

The Seller should arrange shipment of the goods by vessels belonging to the following shipping lines;
1. The shipping Purchaser of India Ltd.
2. The Scindia Steam Navigation Co., Ltd

If the Seller finds that the space in the vessels of these Lines is not available for any particular
consignments, he should inform the Shipping Co-ordination Officer, Ministry of Surface Transport, New
Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN – 031 – 61157, 61158, 61159)
immediately so that dispensation from the shipping lines concerned to use alternative lifting may be
sought.

11. SHIPMENT FROM WEST COAST PORTS OF U.S.S CANADA AND OTHER AREAS NOT
SPECIFICALLY MENTIONED ABOVE

The Seller should arrange shipment of the goods by Indian vessels to the maximum extent possible
subject to a minimum of 50 %. For the purpose of ascertaining the availability of suitable Indian vessels
and granting dispensation in the event of their non-availability, the Seller should furnish the details
regarding contract number, nature of cargo, quantity, port of lading, discharging, name of government
consignee, expected date of readiness of each consignment etc. to the Shipping Co-ordination Officer,
Ministry of Surface Transport, New Delhi, (Cable: TRANSCHART, NEW DELHI, Telex: VAHAN IN –
031 – 61157, 61158, 61159) at least six weeks in advance of the required position.

B) BILLS OF LADING:

(i) C.I.F./C&F/TURNKEY SHIPMENTS

The Bills of lading should be drawn to indicate Shipper and ‘Consignee’ as under:
SHIPPER: The C.I.F (C&F)/TURNKEY SUPPLIERS concerned.

HLL/PCD/ESIC-20/10-11 Page 161 of 166 01.09.2010


HLL Lifecare Ltd. Shipping Arrangement

CONSIGNEE: As per consignee’s particulars in the contract (The name an address of the
‘Port Consignee’ and ‘Ultimate’ both should be indicated).

(ii) F.O.R SHIPMENTS

The Bills of lading should be drawn indicating shipper Consignee as under:

SHIPPER: The F.O.R suppliers Concerned

CONSIGNEE: Supplier’s Indian Agent on order

Note:
1. Moreover the name of the ‘Purchaser’ and ‘Ultimate’ Consignee should appear in the body of
the Bills of Lading as the ‘Notify’ or as a remark.
2. Two non-negotiable copies of the Bills of Lading indicating the freight amount and discount, if
any allowed, should be forwarded to The Shipping Co-ordination Officer, Ministry of surface
Transport (Chartering Wing), New Delhi after the shipment of each consignment is effected.
3. The seller should avoid the use of over-aged vessels for the shipment of the goods under the
contract and if so used the cost of additional. Insurance, if any, shall be borne by the seller.

HLL/PCD/ESIC-20/10-11 Page 162 of 166 01.09.2010


HLL Lifecare Ltd. Check List

SECTION – XX
CHECKLIST
Name of Tenderer:
Name of Manufacturer:
Page No. in
Sl. Yes/ No/
Activity the Tender Remarks
No. NA
document
1. Have you enclosed EMD of required amount
for the quoted schedules?
2(a). Have you enclosed clause-by-clause technical
compliance statement for the quoted goods vis-
à-vis the Technical specifications?
2(b). In case of Technical deviations in the
compliance statement, have you identified and
marked the deviations?
3. Have you kept validity of 120 days from the
Techno Commercial Tender Opening date as
per the TE document?
4(a). Have you enclosed duly filled Tender Form as
per format in Section X?
4(b). Have you enclosed Power of Attorney/
Authorisation in favour of the signatory?
5. Have you submitted manufacturer’s
authorization as per Section XIV?
6. Have you submitted the certificate of
incorporation?
7(a). In case of Indian Tenderer, have you furnished
Income Tax Account No. as allotted by the
Income Tax Department of Government of
India?
7(b). In case of Foreign Tenderer, have you
furnished Income Tax Account No. of your
Indian Agent as allotted by the Income Tax
Department of Government of India?

HLL/PCD/ESIC-20/10-11 Page 163 of 166 01.09.2010


HLL Lifecare Ltd. Check List

Page No. in
Sl. Yes/ No/
Activity the Tender Remarks
No. NA
document
8. Have you intimated the name and full address
of your Banker (s) along with your Account
Number
9. Have you furnished Annual Report (Balance
Sheet and Profit & Loss Account) for last three
years prior to the date of Tender opening?
10. Have you submitted the Quality Control
Requirements as per Proforma given in Section
VIII of TE document?
11. Have you accepted delivery period as per TE
document?
12. Have you accepted the terms of delivery as per
‘DDP at consignee site basis’?
13. Have you accepted the warranty/CMC as per
TE document?
14. Have you accepted all terms and conditions of
TE document?
15. Have you fully accepted payment terms as per
TE document?
16(a) Have you submitted prices of goods, turnkey
(if any), CMC etc. in the Price Schedule as per
Section XI?
16(b) Have you submitted satisfactory end user
performance certificate as per the Proforma for
performance statement in Sec. IX of TE
document in respect of all orders?
17. Have you submitted copy of the order(s)
against the above end user certificate (s)?
18. Have you furnished documents establishing
your eligibility & qualification criteria as per
TE documents?

HLL/PCD/ESIC-20/10-11 Page 164 of 166 01.09.2010


HLL Lifecare Ltd. Check List

N.B.
1. All pages of the Tender should be page numbered and indexed.
2. The Tenderer may go through the checklist and ensure that all the documents/confirmations
listed above are enclosed in the tender and no column is left blank. If any column is not
applicable, it may be filled up as NA.
3. It is the responsibility of tendered to go through the TE document to ensure furnishing all
required documents in addition to above, if any.

(Signature with date)

(Full name, designation & address of the person duly authorised sign on behalf of the
Tenderer)
For and on behalf of

(Name, address and stamp of the tendering firm)

HLL/PCD/ESIC-20/10-11 Page 165 of 166 01.09.2010


HLL Lifecare Ltd. Consignee addresses

Section – XXI

Consignee addresses

Consignee Code Consignee Address Telephone No.

Medical Superintendent
ESIC Hospital & Occupational Disease
Joka Centre (E.Z.) 033-2467 1764
Diamond Harbour Road
Joka, Kolkata - 700 104
Medical Superintendent
ESI Hospital
Basai 011-25100664
Ring Road, Basaidarapur
New Delhi - 110 015
Medical Superintendent
Baddi ESIC Hospital --
Baddi, Himachal Pradesh
Medical Superintendent
Bhiwadi ESIC Hospital --
Bhiwadi, Rajasthan
Medical Superintendent
Manesar ESIC Model Hospital --
Manesar, Haryana
Medical Superintendent
0161-2772435/36 (T)
Ludhiana ESIC Model Hospital
0161-2774357 (F)
Bharat Nagar, Ludhiana-141 001
Medical Superintendent 0731-2559091 (T)
Indore ESIC Model Hospital and ODC 0731-2559080 (F)
Nanda Nagar, Indore – 452011 (M.P.)
Medical Superintendent
ESIC Hospital 0657-2383866 (T)
Adityapur
Adityapur, Jamshedpur, 0657-2383866 (F)
Jharkhand-831013

NB: The purchaser/consignee will ensure timely issue of CDEC, Octroi Exemption
Certificates, Road Permits & Entry Tax Exemption Certificates, wherever
applicable, to the suppliers.

HLL/PCD/ESIC-20/10-11 Page 166 of 166 01.09.2010