Professional Documents
Culture Documents
Operating Instructions Photometer 5010: Robert Riele GMBH & Co KG
Operating Instructions Photometer 5010: Robert Riele GMBH & Co KG
Operating Instructions Photometer 5010: Robert Riele GMBH & Co KG
SYMBOLS
The packaging material, the type plate on the instrument and the manual may contain the following symbols or
abbreviations:
Manufactured by:
Symbol for the marking of important information for appropriate handling of the device
Biohazard
Samples containing material of human origin must be treated as potentially infectious.
The relevant laboratory guidelines on safe use must be observed.
Symbol for the marking of electrical and electronics devices according to § 7 ElektroG
INSTRUMENT APPROVALS
The Photometer 5010 meets the requirements of Directive 98/79/EC on in vitro diagnostic medical devices (IVDD).
Furthermore, the Photometer 5010 is manufactured according to the special safety requirements for IVD medical devices
stated in EN 61010.
SAFETY INFORMATION
Operator qualification
Only appropriately trained operators are qualified to operate the device.
Environmental conditions
The Photometer 5010 is approved for indoor use only.
For further environmental conditions see chapter 10.1.
Patient ambience
The Photometer 5010 may not be used in the patient ambience.
Electrical Safety
This device was examined and left the factory in perfect technical condition. To preserve this and protect safe and
faultless operation, the operator must follow the orders and remarks of this operating manual.
Connect the device to grounded power outlets only. All peripheral devices that are connected to the Photometer
5010 must comply with safety standard EN 60950. Before connecting read the documentation of the peripheral
devices.
Opening covers or removing parts of the instrument, except where this can be achieved manually without the use
of any tool, may expose voltage-carrying components. Connectors can be live, too. Never try to maintain or repair
an open instrument, which is carrying voltage.
Only authorized specialist staff may carry out repairs at the device including replacement of the Lithium battery.
Through improper repairs, the warranty extinguishes, and the operator can be heavily jeopardized.
If suspected the device can no longer be operated safely, turn it off and take steps to ensure that no one will
subsequently attempt to use it.
Electromagnetic waves
Devices that emit electromagnetic waves may affect measured data, or cause the Photometer 5010 to malfunc-
tion. Do not operate the following devices in the same room where the Photometer 5010 is installed: mobile phone,
transceiver, cordless phone, and other electrical devices that generate electromagnetic waves.
Reagents
Regarding reagents follow the safety as well as the operating instructions of the manufacturers.
Pay attention to the currently valid German “Gefahrstoffverordnung” (GefStoffV)!
Biological safety
Liquid waste is potentially biologically hazardous. Always wear gloves if handling those materials. Do not touch
parts of the device other than those specified. Consult the laboratory protocol for handling biohazardous materials.
Pay attention to the currently valid German “Biostoffverordnung” (BioStoffV)!
Waste
Handle liquid waste properly, according to legislation on water pollution, and on the treatment of drainage and
waste matter.
MANUFACTURER’S WARRANTY
ROBERT RIELE GmbH & Co KG warrants Photometer 5010 against defects in material and workmanship.
For further information contact the local distributor.
At the end of the life or utilization time the device and the accessories can be given back to the manufacturer with
costs for an environmental waste disposal. The previous professional decontamination has to be proved with a
certificate.
ROBERT RIELE GmbH & Co KG maintains a quality management system according to ISO 13485, certified by
mdc medical device certification GmbH.
CONTENTS
4 PROGRAM SELECTION............................................................................. 14
4.1 Measurement with programmed methods ....................................................................................... 14
4.2 Measurement with basic methods ................................................................................................... 15
4.3 Method editor .................................................................................................................................. 15
4.4 Utility programs ............................................................................................................................... 16
4.5 Line feed [LF] .................................................................................................................................. 16
8 MAINTENANCE ........................................................................................... 66
8.1 CLEANING INSTRUCTION ........................................................................................................... 66
8.2 CALIBRATING MEASURING SYSTEM ........................................................................................ 66
8.3 ADJUSTMENT OF BUBBLE DETECTOR .................................................................................... 66
8.4 CALIBRATION OF PERISTALTIC PUMP .................................................................................... 66
8.5 REPLACEMENT OF PAPER ROLL ............................................................................................. 66
8.6 REPLACEMENT OF ASPIRATION TUBE ................................................................................... 67
8.7 REPLACEMENT OF LINE FUSES............................................................................................... 67
This device is a semi-automated, programmable photometer for In Vitro Diagnostic (IVD) by qualified laboratory
staff.
It is operated via touchscreen. Remote control is possible by a serial data interface (chapter 7.2.5 - Menu serial
com – REMOTE CONTROL).
For measuring methods several programmed methods with open parameters are available (chapter 5 -
CALCULATION PROCEDURES and chapter 12 - METHOD LIST).
Besides, up to 231 methods - built up on the basic methods - can be established and stored by the operator with
a method editor. A list of methods can be printed out (chapter 6 - METHOD EDITOR).
Up to 50 nonlinear calibration curves with maximum 20 sets of points can be stored (chapter 7.2.2 - Multi-standard
functions).
Photometer 5010 offers a flexible cuvette concept: A special flow-through cuvette and a peristaltic pump provide
for speedy operation with an optimal tempering of the measuring solution. Alternatively, the measuring solution
can be measured in one-way or glass cells in the provided standard cuvette adaptor.
By means of a combination of transistor and Peltier element the solution reaches fast and accurately one of the
three selectable temperatures, 25 °C, 30 °C or 37 °C.
The Photometer 5010 is standard equipped with six optical filters of the wavelengths 340, 405, 492, 546, 578 and
623 nm. If required, they can be exchanged against any wavelength within the range of 340-730 nm. Three
additional filters, e.g. 670 nm, can be installed.
The device is equipped with a thermal printer.
The measuring data can be stored and managed in the Photometer 5010 (chapter 7.2.8 - Stored results).
According to a GLP conformal documentation the names of lab and operator can be printed out as well as
transferred to EDP (chapter 7.2.5 - Menu serial com – EDP ON/OFF).
In Photometer 5010 up to 50 methods can be supervised with a quality control (chapter 7.2.6 - Quality control).
Numerous utility programs permit the individual configuration of the device. Function tests support the analysis of
sources of error.
Photometer 5010 is future-proof by FLASH MEMORY technology: The operating system can be updated with
program novelties and/or improvements comfortably, without having to open the equipment (please ask distributor
for further information).
2 INSTALLATION
2.1 DELIVERY
Check the device and contents of the enclosed box as follows on visible transport damages and completeness:
1 Operator’s Manual
1 Aspiration tube
1 Dust cover
2 Fuses for line power
1 Mains cable
2 Thermal printer paper
1 Pump tube with joints
1 Standard cuvette adaptor
1 Top cover small for printer
1 Waste tube
Inform the sales office immediately about transport damages. Keep the original packaging for a possible
return.
2.2 PREPARATION FOR INSTALLATION
Place the device on a stable, level surface. Do not obstruct the input air at the bottom and the waste air at the
back plate to guarantee the ventilation of the device.
If the device was exposed to extraordinary fluctuation in temperature and/or humidity, it must acclimatize suffi-
ciently before operation.
Before connecting the waste tube to the pump tube remove pump tube at both ends from the metal clamps.
The waste tube of the flow-through system must be led through the tunnel to the backside of the device (chapter
3.2 - BACK) and then into any drain tank.
2.3 INSTALLATION
Photometer 5010 operates at any line voltage between 100 VAC and 240 VAC at 50/60 Hz. The device plug of the
mains cable must be put into the socket at the back of the device and the mains plug into a grounded mains
socket.
While connecting or disconnecting an external device (PC, printer) to Photometer 5010 both devices must
be switched off.
Switch on Photometer 5010 by the mains switch at the back.
Greeting screen:
( c ) R O B E R T R I E L E
After switching on copyright, website, type of
G m b H & C o K G device and version designation are
displayed and - in the case of activated
P H O T O M E T E R 5 0 1 0 printer - printed out.
V 7 . X a d d / m m / y y D
w w w . r i e l e . d e
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
If errors appeared during operation, first of all they have to be confirmed with [E] before remedy (chapter 9 -
ERROR MESSAGE / CORRECTION).
2.4 LOADING PRINTER PAPER
With initial operation or if the colored end of the paper roll appears, printer paper must be inserted:
• Open the printer cover.
• Put the green head-up lever in the up position.
• Remove the rest of paper.
• Put the green head-up lever in the down position.
• Put printer paper axis into the new printer paper reel.
• Insert the paper inside the printer. The roller will automatically feed the paper for about 4 cm.
• Press [LF] several times for line-feed until the paper has a length of about 5 cm. In case of no reaction
the printer may be deactivated.
• Insert printer paper reel into the axis guide.
• Push the printer paper through the slot in the printer cover and close the printer with the cover.
3 OPERATING ELEMENTS
3.1 FRONT
printer cover
lid
working area
touchscreen
aspiration tube sipping lever [P]
3.2 BACK
3.3 TOUCHSCREEN
The touchscreen shows applications and
information. It is contact-sensitive and reacts
to the pressure exerted on it. In order to
execute a function, the desired range on the
screen must be touched
waste tubing
REF 1704834001 pump tube with joints
REF 5010-050
aspiration tubing
REF 5010-065
If a flow system is installed, disconnect the bubble detector, remove the aspiration tube from the metal tube and
disconnect the outlet tube cuvette from the joints of the pump tube.
Another adaptor can easily be inserted. Before check the cleanliness of the lense and the underside of the
adaptor. When inserting the adaptor electrical contacts in the device are closed. Turn milled screw only by hand!
After replacing of a cuvette adaptor by another one an optic adjustment should be done after reaching operating
temperature (see chapter 7.2.1 - Optic adjustment).
Before connecting the tubes remove pump tube at both ends from the metal clamps. The pump tube must
not be tensed during longer time out (fig. below).
The lid of the cuvette compartment may stay open. The optic is not sensitive to stray light.
Tension release:
The upper connection of
the pump tube can be
loosened easily from the
metal clamps.
Plug the bubble detector to the socket in the working area. It must sit as close as possible at the connection of
the flow-through cuvette. For working without bubble detector switch off this function (chapter 7.2.4.3 - Bubble
detector ON / OFF).
While working with the flow-through system the tubes must not be sharply bended. No residues should be
inside. Check from time to time that the tubing and the connections are leakproof. After each replacement of
tubing execute a Pump calibration (chapter 7.2.4.2).
Before as well as after all measuring it is absolutely necessary to wash the tube system repeatedly with
distilled water or another appropriate rinse solution by pressing [WASH] or sipping lever [P]. This is also necessary
after a method change (chapter 8.1 - CLEANING INSTRUCTION). For the procedure within a measuring series
see the application regulation.
In order to aspirate solution into the measuring system put the aspiration tube deeply enough into the respective
vessel.
Before setting to zero press [ZERO]. Trigger setting to zero by sipping lever [P].
Trigger a normal measuring by sipping lever [P]. Repeat a measuring of a solution which is already aspirated by
[RESULT].
To work in optimized volume mode use the function Volume optimized ON / OFF (chapter 7.2.4.3.2). This
function makes it possible, for example, to pump two consecutive times 500 µl from 1000 µl sample volume.
3.5.3 Working with standard cuvettes
The optical path is directed from left to right of the device. Insert single cuvette according to the drawing
OPTIC CONSTRUCTION in TECHNICAL DATA.
The lid of the cuvette compartment may stay open. The optic is not sensitive to stray light.
Trigger setting to zero by [ZERO].
Pay attention to the correct connection of aspiration tube and outlet tube cuvette.
Switch on or off bubble detector by the function Bubble detector ON / OFF (chapter 7.2.4.3).
4 PROGRAM SELECTION
After completion of a method or execution of a utility program the program always returns to the main menu.
M A I N M E N U Main menu:
Down in the status line from left to right
following is shown:
M E A S U R E W I T H P R O G R . M E T H O D S
• Current temperature of the cuvette
M E A S U R E W I T H B A S I C M E T H O D S adaptor in °C.
In the case of switched off tempering the
M E T H O D S N E W / C H A N G E / C O P Y display changes between --.--C und
xx.xxC.
In the case of switched on tempering and
U T I L I T I E S L F
instable temperature the display
changes between --.--C and e.g. 37.3°C.
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9 In the case of stable temperature the
current temperature of e.g. 37.0°C is
shown. Small fluctuations of the value
are normal.
• Date in the format day/month/year
• Time
M E T H O D : G l u c o s e
The valid range for a method number lies
U N I T : m g / d l between 20 and 250.
1 2 3 4 5 6 7 8 Scroll all existing methods by [+] or [-]. If no
method is programmed, a plain text error
E S C 9 0 . + - E message (chapter 9.3 - PLAINTEXT
ERROR MESSAGES) is shown.
Further information:
Application sheets of reagent manufacturers
• Absorbance measurement
C O N C . W . F A C T O R P A G E
• Concentration measurement / end point
measurement
C O N C . W . F A C T O R R B
• Fixed time kinetic / two point kinetic
C O N C . W . F A C T O R S B • Kinetic
C O N C . W . F A C T O R R B S B E X I T • Transmission
Scrolling through all methods is possible by
[PAGE]. The current page is shown at the
right upper screen corner. By [END] the
program returns to the main menu.
• CONC. = concentration
• KIN = kinetic
• FTK = fixed time kinetic
• F = factor
• STD = standard
• RB = reagent blank
• SB = sample blank
Further information:
Chapter: 5 - CALCULATION
PROCEDURES
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9 Further information:
Chapter: 6 - METHOD EDITOR
P R I N T E R
P U M P M E N U E X I T
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
M E A S U R E W I T H B A S I C M E T H O D S
M E T H O D S N E W / C H A N G E / C O P Y
U T I L I T I E S L F
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
5 CALCULATION PROCEDURES
• Deviations from normal operation, caused by the device or by the operator, are notified by
“ERROR”. They always have to be confirmed by [E] (chapter 9 - ERROR MESSAGE / CORRECTION).
• After switching on the tempering it lasts up to 10 minutes until a constant temperature of 25 °C, 30 °C or
37 °C is reached.
• The current temperature of the flow-through cuvette or of the cuvette adaptor is shown at the Iower edge
of the touchscreen. For meaning of the display see chapter 4 - PROGRAM SELECTION /MAIN MENU.
A temperature instable or out of tolerance during measuring is marked by an asterisk (*) at the utmost
right position in the corresponding print line.
To avoid deviations due to temperature influence a delay between triggering and actual measuring can
be programmed in each method.
• For a quick mode of operation all temperature-sensitive samples, reagents and washing solutions should
be externally tempered by Incubator T12/T16 (REF 500-002 / 500-001) or a water bath.
0.615 S T [ A ]
standard is shown.
By [RESULT] a measuring is
triggered.
• The determined resulting factor of a standard measurement is stored together with the corresponding
method number. After renewed selection of this method the last resulting factor is offered as "OLD STD".
• The principle of the multiple measurement can also be expanded to all measurements. The correspond-
ing entry can be set invoking a basic method. The parameter is definable in preprogrammed methods
(chapter 6 - METHOD EDITOR).
In each time interval (Delta or Δt) the difference between the relating absorbance values as well as the gradient
of the curve are calculated.
To obtain the alteration per minute AS,Minute the gradients must be averaged. This is done by a simple linear
regression calculation also giving an indicator for the linearity of the test. This indicator is called the coefficient of
correlation R. For practical reasons, the square of the coefficient of correlation R^2 or coefficient of determination
is taken in a Kinetic calculation. The value of R^2 can vary between 0 and 1. An R^2 value of 1 indicates perfect
linearity and a value of 0 indicates absolute non-linearity. Already values < 0.9 indicate a bad linearity and
therefore an incorrect test. In order to improve the linearity of the kinetic only the best three deltas are considered
in the calculation procedure of the regression calculation. Therefore, at least five deltas are required when
programming a new method. If this procedure does not lead to an improvement all deltas are reconsidered in the
calculation procedure.
In practice, linear tests show values of R^2 near to 1. In the example for Calculation procedure 11 (KIN/F/Rb)
values of R^2 ≥ 0.998 are permitted. Results with smaller R^2 values could be caused by temperature instability,
pollution, expired reagents, unfavorable delay time, etc.
For a better monitoring the number of deltas (deltas or Δt) should be bigger than specified for the manual
procedure. The classic three-minutes-test with three deltas of 60 s can be replaced by 15 deltas of 12s.
When programming a new method, which is based on CP 11 or CP 12, it is possible to set lower and upper limits
for the measurement result within the method editor (see chapter 6 - METHOD EDITOR, Fig. 6.5). This can be
achieved setting the parameters MIN. VALUE and MAX. VALUE. If the measured value exceeds the MAX.
VALUE, a message RANGE MAX. is shown, and if the measured value falls below, MIN. VALUE message
RANGE MIN. is shown. Also a lower limit for R^2 can be entered by setting MIN. R^2, if the obtained R^2 value
falls below the entered value a message NON-LINEAR is shown.
In order to get positive results at tests with decreasing absorbance (see Fig. 5.1.7.1), a negative factor has to be
entered. Only if MAX. VALUE is set and the sign of the measured value is not equal to the sign of the entered
MAX. VALUE a message RANGE +/- is shown.
The parameters MIN. VALUE, MAX. VALUE and MIN. R^2 are deactivated entering a zero value.
NO. TIME [s] ABS. In the next step a chart including all data
1 0 0.718
2 30 0.734 points is printed.
3 60 0.750
4 90 0.767
5 120 0.785
6 150 0.805
7 180 0.825
ABS.1: 0.642
ABS.2: 0.750 Printout after confirming [MODE], [MODE]
1: 0.0312 and [DETAIL].
2: 0.0320
3: x 0.0345 ABS/min
4: x 0.0364
5: x 0.0387 As of five programmed deltas in case of an
6: 0.0417
DELTA REG: 0.0365 improvement of the linearity the three best
deltas are considered and marked with a
“x”. If this procedure does not lead to an
improvement all deltas are reconsidered
in the calculation procedure and none of
the deltas are marked.
E D I T M E A S .
N E W S A M P L E N O . 1 - 9 9
1 2 3 4 5 6 7 8
E S C 9 0 . + - N / I E
5.2 ABBREVIATIONS
Explanations:
CP-No. ........................Number of the calculation procedure (chapter 6 - METHOD EDITOR)
Characteristic ..............Name of the calculation procedure (chapter 12.1 - BASIC METHOD)
Calculation formula .....Calculation basis of basic method
M E A S U R E S A M P L E G l u c o s e
reading window
numerator
16.5 Additional method
specific information
3 m m o l / l
ID number I D - N O . 1 3 4 6 7
dimension
action keys
E X I T M O D E W A S H Z E R O R E S U L T
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
info line
[WASH] With activated pump the wash volume defined to the method is pumped
Mode functions:
[NUM.] Edit sample numerator or ID-NO. (chapter 5.1.9)
[LF] Line Feed
[QC] Quality Control functions
[PRN] Switch printer ON/OFF / printout of current time by [TIME]
[DETAIL] Display/printout of detailed results
[M-STD] Multi standard functions
[E1/E2] Change to measuring E2 (chapter 5.4.14)
[RETRY] Repeat a measurement
[RETURN] Return to normal functions
ST[A] 1: 1.485
ST[A] 2: 1.521 →Insert / measure standard blank
ST[A] 3: 1.495 →Insert / measure standard 1
→Insert / measure standard 2 (optional)
ST[A]: 1.501 →Insert / measure standard 3 (optional)
STb[A]: 0.479
DELTA ST: 1.022 (Averaged standard)
FACTOR: 8.74 (standard blank)
(Averaged standard minus standard blank)
NO. ABS. RESULT (Resulting factor)
1 1.495 7.89
Sb[A]: 0.489
2 1.542 7.89 →Insert / measure sample blank
Sb[A]: 0.535 →Insert / measure sample
3 1.394 8.39
Sb[A]: 0.329 →Insert / measure sample blank
→Insert / measure sample
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [NO].
After each measurement the next sample can be measured with [NEXT].
With [RESULT] it is possible to measure the same sample again.
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [ENTER].
After each measurement the next sample can be measured with [NEXT].
With [RESULT] it is possible to measure the same sample again.
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [ENTER].
During the procedure the dialog asks for the use of a reagent blank. The default value is OFF. To continue with-
out reagent blank press [ENTER].
This calculation procedure has special parameters that allow a time controlled measuring process. These
parameters are: time/delta T1, measure time T2, delay T3, reagent time #2 and reagent time #3. By setting a
time/delta (value between 10s and 255s) the other parameters for the time controlled measuring will be used. In
the time controlled mode the quantity of samples is determined by the measure time and the time/delta, e.g. with
a measure time of 60s and a time/delta of 10s it is possible to measure 6 samples (without sample blank). The
measure time should be chosen greater than or equal to the time/delta.
Before starting a new E1/E2 course a new zero measurement has to be done.
The reagent time #3 is usable only if the reagent time #2 is set. In that case the reagent mode will be entered, i.e.
the photometer also takes control of the timing for dispensing the reagent before measuring the samples. The
quantity of samples will be determined by the reagent time #2 and the time/delta. The reagent time #2 should be
chosen less than or equal to the reagent time #3 and greater than or equal to the time/delta.
Figure 5.4.14.1 shows the time sequence of a time controlled measuring process with N samples, a delay time
T3 and without reagent time.
... T3 T1 T1
... T1 T1
...
T2
t: Time
T1: Time/Delta : automatic process
T2: Measure time
T3: Delay, optional : manual process
A: Start of time controlled measuring
E1: Start of first course (E1)
E2: Start of second course (E2)
AS = E2 – Fdil * E1
whereby Fdil = (a + b) / (a + b + c)
D E L T A R 1 R 2
reading window Number of
0 S A M P L E S : 3 samples
F d i l : 1 . 0 0 0 0
Z [ A ]
T I M E 4 s Current time
dimension N E X T A C T I O N A T T I M E 5 s
Instructions
D I S P E N S E R E A : I N S A M P L E # 1 for next
action
action keys
E X I T M O D E - > R 2 Z E R O R E S U L T
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
info line
NO. RESULT
→The result will be displayed
1 -0.116
2 -0.116
3 -0.116
6 METHOD EDITOR
By the method editor the daily laboratory work can be substantially facilitated. Based on the 15 calculation
procedures up to 231 user-defined methods with their setting parameters can be saved. With the functions of the
editor a method can be established, changed or deleted.
Fig. 6.1
M E T H O D N E W / C H A N G E / C O P Y In the main window of the method editor
following options are available:
Fig. 6.2
[#→#] All methods from no 20 can be
copied to a new method no. First the method
to be copied is queried in Fig. 6.3. Its
M E T H O D C O P Y parameters can be changed starting with
Fig. 6.4.
Fig. 6.3
Inquiry window of the desired method.
The method used last is suggested. With [+]
M E T H O D N O . or [-] the methods can be scrolled. A numeric
input of the method number is possible at
M E T H O D : ( N A M E ) any time. A known method is indicated with
U N I T : ( D I M . )
name and dimension.
Fig. 6.7
Query of the desired method number, under
which the new method is to be stored. The
next free method number is indicated. How-
M E T H O D N O . ever each free method number can be
N E X T F R E E N O . x x selected within the range of 20 to 250.
7 UTILITY PROGRAMS
M A I N M E N U
Main menu:
M E A S U R E N E W / C H A N G E / C O P Y
U T I L I T I E S L F
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
Utilities Description in
chapter
Calibration procedure:
Wait for about 40s until adjustment is
D A R K A D J U S T M E N T finished.
8 3 9 2 1 9 6 The function cannot be interrupted. After
completion the program returns to the utility
P L E A S E W A I T program level.
Fig. 7.2.2.1
Main window of multi-standard functions
U T I L I T I E S S U B M E N U E
[MEASURE CURVE] After query of a curve
M E A S U R E C U R V E L I S T number and the first standard the program
changes automatically to the method
P R I N T C U R V E selection window. There the given
parameters of the respective method can
be again controlled and/or changed. All
E D I T C U R V E
further standards are queried during the
following procedure.
S H O W C U R V E E X I T
The program for measuring the multi-
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9 standards branches out automatically to the
calculation procedures as follows:
CP1 → CP5
CP2 → CP6
CP3 → CP7
CP4 → CP8
CP9 → CP10
CP11 → CP12
CP14 → CP14
Print-out of curve:
For measuring in another standard
method than suggested, start measuring
multi-standards in each other selected
standard method (before measuring the first
standard) by [MODE] [MODE] [M-STD].
Curve number and standards are queried
accordingly.
E X I T
1 2 3 4 5 6 7 8
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
E X I T W A S H
B A U D R A T E E X I T
3 7 . 0 ° C 0 1 / 1 3 / 1 5 1 1 : 3 9
E X I T S 1 S 2
DATE: 09/03/17
TIME: 08:44:12
25 GLUCOSE 13.03
21 HDL-C 367
27 UREA COL 197.2
29 CK-MB 1128.1
31 GOT 189.9
M O N T H L Y M E M O R Y S 2
In the overview window of the selected
M E T H O D 2 5 G L U C O S E method all data of quality control are visible.
S E R U M 1 L T - S Y S a b n o r m a l In the line above the keys following
L O T G 3 1 2 information the current reading is indicated:
C O M P A N Y L A B O R + T E C H
D A T E M A Y 0 6
R E Q U I R E D 1 4 . 4 m m o l / l (# 1) → Numerator of the monthly memory.
R A N G E L : 1 2 . 1 O : 1 6 . 7 The oldest reading corresponds to the 1.
C A L C U L . s : 1 . 0 1 2 C V : 7 . 2 9 8
m : 1 3 . 8 6 0 n : 2 0
(01.27.06) → Date of reading
1 0 7 . 0 9 . 0 7 1 3 . 3 0 - 1 s 1
E X I T M O R E D E L E T E - + (13.30) → Reading
Output dialog
If at least 20 readings are stored in the
P R N - N monthly memory, these are indicated in the
Levey Jennings plot. In this representation
the deviations can be controlled visually and
P R N - E
thus tendencies or systematic errors be
better recognized.
7.2.6.4 QC EDP
Not implemented
E X I T S T A R T
******ACTUAL*SETS*******
DATE: 09/03/17
TIME: 08:44:12
PHOTOMETER 5010 #20001
V7.1a dd/mm/yy D
/T1.7
CORE V1.0
PCB LAYOUT c
ADC COUNTS (DARK ADJ.
S: 8394148 A: 8392514
FILTER
1: 340 2: 405 3: 492 The percentage is proportional to the
21/21% 18/23% 12/18%
4: 546 5: 578 6: 623 brightness level.
13/20% 17/26% 16/24% The percentage value left represents the S
7: 999 8: 999 9: 999 adapter and the right one the A adapter.
0/ 0% 0/ 0% 0/ 0%
TEMPERATURE S
25C 8939 3000
30C 10165 3000
37C 11874 3000
TEMPERATURE A
25C 8687 3000
30C 9892 3000
37C 11601 3000
PUMP
MS AIR 178
MS/50µl 56
PUMP SPEED (1-30) 25
B-DET. 6
VOLUME OPTIMIZED OFF
BATTERY: OK
ADC CORRECTION 250
EDP ON (CR-LF)
TOUCH
Mx148 My192 Fx120 Fy100
LANGUAGE
1: ENGLISH
2: GERMAN
KEY SIGNAL ON
T.COUNTER : 334:45
S.COUNTER : 402
PROGR. METHODS 0
STORED RESULTS 123
S A V E T S A D J U S T M E N T
E X I T O K
Hint: If the device is maladjusted, this function can directly be called during the switching on routine:
Switch on the device. After greeting screen (chapter 2.3 - INSTALLATION) is displayed, keep the touchscreen or
the zipping lever [P] pressed. After some seconds a deep signal tone sounds and the text message
“TOUCHSCREEN ADJUSTMENT” will be shown at the first line of the greeting screen. Release the touchscreen
within one second. Execute the adjustment of the touchscreen as described above. Select the type of photometer
after touchscreen adjustment is completed.
E X I T O K S E L E C T
ESC 9 0 . + - E
7.2.17 Language
The status of the language is indicated in the
first line.
• LANGUAGE : ENGLISH
E X I T O K S E L E C T • LANGUAGE : GERMAN
Save setting permanently by [OK].
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
8 MAINTENANCE
This chapter provides necessary information concerning general maintenance by the user.
If any faults should occur which cannot be remedied, then service should be contacted. Repairs at the
device may be carried out only by authorized specialist staff. Through improper repairs the warranty extinguishes,
and the user can be heavily jeopardized.
Liquid waste is potentially biologically hazardous. Always wear gloves if handling those materials. Do not
touch parts of the device other than those specified. Consult the laboratory protocol for handling biohazardous
materials.
Take care that no liquid enters the device! There is no protection against penetrating of liquids
(Code IP X0). Check from time to time that the tubing and the connections are leakproof.
The flow-through system of the Photometer 5010 has to be washed with dist. water regularly before and after
measuring 2 – 3 times. Depending on sample material and reagent, and always at the end of a working day, the
Photometer 5010 has to be cleaned additionally with a phosphate-free detergent, e.g. with approx. 5ml of Biogent-
A (REF 5010-024), and afterwards rinsed with 5ml distilled water.
Persistent residues are to be removed with a combined alkaline-acid treatment. The following procedure is
advised step by step:
1. 5 x wash with NaOH 1 N
2. 5 x rinse with dist. water
3. 5 x wash with HCl 0.5 N
4. 5 x rinse with dist. water
If the system is contaminated severely, it may alternatively be cleaned with hypochlorite. The following procedure
is advised step by step:
1. 5 x rinse with dist. water
2. 2 – 3 x wash with hypochlorite (1:20 diluted solution) or ISE Cleaning Solution,
undiluted, let it possibly affect up to 20 minutes
3. 5-10 x rinse with dist. water
For device cleaning and surface decontamination purposes use commercial decontaminating solution which are
usually available in clinical chemistry laboratories like Mikrozid ® AF Liquid, Bacillol® plus, 3 % Kohrsolin® or similar
solutions. Switch off the device and disconnect it from the mains voltage. Then clean the device with soft cloth
and decontaminating solution.
Empty the drain tank at the end of daily measurement, or whenever filled.
mains switch
fuse holder
In the display errors are shown as plaintext (chapter 9.3 - PLAINTEXT ERROR MESSAGES)
RANGE MIN. The programmed low limit was under-run by the measurement.
RANGE MAX. The programmed upper limit was exceeded by the measurement.
NON-LINEAR The square of the correlation coefficient R lies at the kinetic measuring below the
programmed low limit.
RANGE +/- At the kinetic measuring the procedure of the kinetic is wrong (increasing / falling).
NO METHOD Dialed method is not programmed. Select other method according to method list.
9 check sum error in the data record of the device basic automatic error remedy
setting (status, ADC correction)
10 division by a too small value (< 0.001 A) check filter; check standard;
check measuring solution
11 invalid calibration curve Select valid number
12 setting to zero not possible (zero value is < 32768 cycle) check LED; check filter;
check zero solution
13 setting to zero not possible (zero value is > 983039 cycle) repeat optic adjustment (chapter 7.2.1);
check LED; check filter;
check zero solution
14 invalid standard measure valid standard solution
15 no parameter memory vacant (too little memory for non- delete a no longer actual nonlinear
linear methods) method
16 method no. is occupied select other method no.; delete a no
longer actual nonlinear method
17 check sum error in the parameter memory (nonlinear program method newly
method)
18 at calculation overflows in nonlinear method check factor; check parameter
19 clock malfunction
20 overflow at Kinetic check measuring solution
21 overflow at Kinetic check measuring solution
22 overflow at Kinetic check measuring solution
23 overflow at Kinetic check measuring solution
24 overflow at Kinetic check measuring solution
25 overflow at Kinetic check measuring solution
26 overflow at Kinetic check measuring solution
27 overflow at Kinetic check measuring solution
28 overflow at Kinetic check measuring solution
29 wrong input of deltas or time per delta restart method
30 battery empty contact service partner
31 communication: wrong data format contact service partner
32 communication: sent data not plausible / not interpretable contact service partner
33 communication: mentioned module does not answer in a check interconnecting cable;
certain time check mentioned module
34 communication: overflow send/receive buffer reduce amount of data at the
communication partner
35 remote control: wrong method number external software problem
36 remote control: unknown command external software problem
37 remote control: wrong data format external software problem
38 check sum of operating system bank 0 damaged contact service partner
39 check sum of operating system bank 1 damaged contact service partner
40 timeout at reception from module 2 switch off/on device
41 check sum error at reception from module 2 switch off/on device
42 NAK at reception from module 2 switch off/on device
10 TECHNICAL DATA
A Identification
DECLARATION OF CONFORMITY:
The above mentioned absorption photometer is in conformity with the following metrological description.
W. Riele
B Metrological description
OPTIC CONSTRUCTION
The optical path is directed from left to right of the device. Insert standard cuvette accordingly.
Diaphragm
Lens 1 Lens 2
Cuvette Photodiode
LED
Cuvette adaptor Filter
12 METHOD LIST