203-JN9000 02 005
203-JN9000 02 005
EUT DESCRIPTION Portable imaging diagnostics instrument for veterinary labs with optional wireless
networking. Both Bluetooth & Wifi
TÜV SÜD America Inc, as an independent testing laboratory, declares that the equipment tested as specified above
conforms to ETSI EN 300 328 v1.8.1 “Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband
transmission systems; Data transmission equipment operating in the 2,4 GHz ISM band and using wide band
modulation techniques; Harmonized EN covering essential requirements under article 3.2 of the R&TTE Directive”
TÜV SÜD America, Inc and its professional staff hold government and professional organization certifications and are members of AAMI, ACIL,
AEA, ANSI, IEEE, NARTE, and VCCI.
Not Transferable
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 1 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
1 TEST RESULT SUMMARY
Tables 1 and 2 summarize the device’s test results with respect to the essential requirements defined in EN 300 328
v1.8.1. clause 4.3, Technical requirements. All sub clauses may not be applicable or not tested because the
manufacturer has previous test data that demonstrates compliance.
1.1. Transmitter
Table 1: Transmitter results summary
Not Not
EN Clause
Transmitter parameters Pass Fail applicable Tested
Bluetooth
4.3.1.1 Maximum transmit power
4.3.1.2 Duty Cycle, Tx-sequence, Tx-gap
4.3.1.3 Dwell time, minimum frequency occupation & hopping
sequence
4.3.1.4 Hopping frequency separation
4.3.1.5 Medium Utilization factor
4.3.1.6 Adaptivity
4.3.1.7 Occupied channel bandwidth
4.3.1.8 Transmitter unwanted emissions in the out-of-band
domain
4.3.1.9 Transmitter unwanted emissions in the spurious domain
Wifi
4.3.2.1 RF output power
4.3.2.2 Power Spectral Density
4.3.2.3 Duty Cycle, Tx-sequence, Tx-gap
4.3.2.4 Medium Utilisation (MU) factor
4.3.2.5 Adaptivity
4.3.2.6 Occupied Channel Bandwidth
4.3.2.7 Transmitter unwanted emissions in the out-of-band
domain
4.3.2.8 Transmitter unwanted emissions in the spurious domain
1.2 Receiver
Table 2: Receiver results summary
EN Not Not
Clause Transmitter parameters Pass Fail applicable Tested
Bluetooth
4.3.1.10 Receiver spurious emissions
4.3.1.11 Receiver blocking
Wifi
4.3.2.9 Receiver spurious emissions
4.3.2.10 Receiver Blocking
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 2 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
REVISION RECORD
TOTAL
REVISION NUMBER DATE DESCRIPTION
OF PAGES
98 27 November 2013 Initial Release
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 3 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
2. Introduction
This testing is intended to verify that the model 89-91921-02, IDEXX Portable imaging diagnostics instrument is
performing in accordance with the requirements of EN 300 328 V1.8.1.
All pages have been numbered consecutively and bear TÜV SÜD America, the test report number, the date, the test
specification in its current version as well as the type designation of the EUT. The total number of pages in this report
is 98.
The tests were carried out on a representative assembly and in accordance with the test methods stated in:
EN 300 328 V1.8.1
3. Testing laboratory
The tests were carried out at TÜV SÜD America:
Accredited by:
NVLAP
Accreditation number 200696-0
Oakwood Lab
nd
29959 592 Street
Millville MN 55957-0255
- Medium Test Site Open Area Test Site
Accredited by:
NVLAP
Accreditation number 200695-0
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 4 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
4. Applicant
TÜV SÜD America Inc reports apply only to the specific samples tested under stated test conditions. TÜV SÜD
America Inc shall have no liability for any deductions, inferences or generalizations drawn by the client or others from
TÜV SÜD America Inc issued reports.
TÜV SÜD America assumes the sample to comply with the requirements of EN 300 328 for the respective test sector,
if the test results are positive.
Comments: ---
Tested By Reviewed By
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 5 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
7.0 ENVIRONMENTAL CONDITIONS IN THE LAB
Actual
Temperature: : 19-21°C
Atmospheric pressure : 98 kPa
Relative Humidity : 57-58%
MEASUREMENT UNCERTAINTY
Table 3: Measurement uncertainties
Parameter Maximum Measurement Uncertainty Actual
Radio Frequency ±1 × 10-7 ±3 × 10-9
Radiated emission of transmitter valid up to 4 GHz ±6 dB ±4.8 dB
Radiated emission of receiver valid up to 4 GHz ±6 dB ±4.8 dB
Temperature ±1 °C ±0.8 °C
Humidity ±5 % ±4.6 %
The equipment comprising the test systems is calibrated on an annual basis.
SIGN EXPLANATIONS
- not applicable
- applicable
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 6 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8. Technical requirements for Frequency Hopping equipment (Bluetooth)
Test location
New Brighton environmental lab
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
NBLE10440 E4446A HP Spectrum Analyzer 44GHz US44300488 25 Sep 14
NBLE02238 SH27 ENVIRONTRONICS 27 Cu Ft T/H Chamber 09963482-5 Code B 29-Oct-13
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
The equivalent isotropic radiated power (e.i.r.p.) shall be equal to or less than 100 mW or 20 dBm. This limit shall apply
for any combination of power level and intended antenna assembly.
Test Data
conducted measurements
DFSK, DH1, and High channel determined worst case from FCC data
15°C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 7 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
23°C
30°C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 8 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.2 Duty Cycle, Tx-sequence, Tx-gap
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
These requirements do not apply for equipment with a maximum declared RF Output power of less than 10 dBm
e.i.r.p.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 9 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.3 Dwell time, minimum frequency occupation and hopping sequence
Test summary
The requirements are: - MET - NOT MET
The dwell time is 2.9 msec.
The minimum frequency occupation time is 2.9 msec in 4(.0029)(79), or 916 msec.
There are 79 channels.
Test limit
The accumulated Dwell Time on any hopping frequency shall not be greater than 15 ms within any period of 15 ms
multiplied by the minimum number of hopping frequencies (N) that have to be used.
The Minimum Frequency Occupation Time shall be equal to one dwell time within a period not exceeding four times
the product of the dwell time per hop and the number of hopping frequencies in use.
The hopping sequence(s) shall contain at least N hopping frequencies where N is 15 or 15 divided by the minimum
Hopping Frequency Separation in MHz, whichever is the greater.
Test Data
Dwell time limit = 15 msec in 1185 msec (15 ms x 79 channels), or 1.185 seconds.
Maximum number of pulses 21.75 in 31.6 seconds, or 1 pulse every .815 seconds, or 1.45 pulses in 1.185 seconds
1.45 x 2.9 msec = 4.205 msec in 1.185 seconds - < 15
Minimum Frequency Occupation Time limit = 1 dwell time / 4 x 2.9 msec x 79, or 916.4 msec.
Maximum number of pulses 21.75 in 31.6 seconds, or 1 dwell time every 815 msec - < 916.4
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 10 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.4 Hopping frequency separation
Test summary
The requirements are: - MET - NOT MET
The hopping frequency separation is 1 MHz – minimum occupied bandwidth is 847 kHz.
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
Test limit
The minimum Hopping Frequency Separation shall be equal to Occupied Channel Bandwidth (see clause 4.3.1.7) of a
single hop, with a minimum separation of 100 kHz.
Test Data
conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 11 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.5 Medium Utilization factor
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
These requirements do not apply for equipment with a maximum declared RF Output power of less than 10 dBm
e.i.r.p.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 12 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.6 Adaptivity
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
These requirements do not apply for equipment with a maximum declared RF Output power of less than 10 dBm
e.i.r.p.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 13 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.7 Occupied channel bandwidth
Test summary
The requirements are: - MET - NOT MET
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
Limit
The Occupied Channel Bandwidth for each hopping frequency shall fall completely within the band 2.4 – 2.4835 GHz
Test Data
conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 14 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
3DH5, Pi_8-DPSK, Mid channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 15 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Plot is from FCC test report
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 16 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.8 Transmitter unwanted emissions in the out-of-band domain
Test summary
The requirements are: - MET - NOT MET
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
NBLE10440 E4446A HP Spectrum Analyzer 44GHz US44300488 25 Sep 14
NBLE02238 SH27 ENVIRONTRONICS 27 Cu Ft T/H Chamber 09963482-5 Code B 29-Oct-13
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
The transmitter unwanted emissions in the out-of-band domain but outside the allocated band, shall not exceed the
values provided by the mask
Test Data
See following pages
Hopping mode, conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 17 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
2dh5 pi_4-dkpsk 1.231MHz obw
15°C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 18 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 19 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 20 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
dh5 gfsk 847kHz obw
15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 21 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 22 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 23 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.9 Transmitter unwanted emissions in the spurious domain
Test summary
The requirements are: - MET - NOT MET
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
Frequency range Maximum power, Bandwidth
e.r.p. (≤ 1 GHz)
e.i.r.p. (> 1 GHz)
30 MHz to 47 MHz 36 dBm 100 kHz 30 MHz to 47 MHz
47 MHz to 74 MHz 54 dBm 100 kHz 47 MHz to 74 MHz
74 MHz to 87,5 MHz 36 dBm 100 kHz 74 MHz to 87,5 MHz
87,5 MHz to 118 MHz 54 dBm 100 kHz 87,5 MHz to 118 MHz
118 MHz to 174 MHz 36 dBm 100 kHz 118 MHz to 174 MHz
174 MHz to 230 MHz 54 dBm 100 kHz 174 MHz to 230 MHz
230 MHz to 470 MHz 36 dBm 100 kHz 230 MHz to 470 MHz
470 MHz to 862 MHz 54 dBm 100 kHz 470 MHz to 862 MHz
862 MHz to 1 GHz 36 dBm 100 kHz 862 MHz to 1 GHz
1 GHz to 12,75 GHz 30 dBm 1 MHz 1 GHz to 12,75 GHz
Test Data
See following pages, conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 24 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
3dh5 8-dpsk
Low channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 25 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 26 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.10 Receiver spurious emissions
Test summary
The requirements are: - MET - NOT MET
Minimum margin of compliance is 3.16 dB at 567.781 MHz
Test location
Wild River Lab Large Test Site (Open Area Test Site)
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03204 EM-6917B Electro-Metrics Biconicalog Periodic 102 30-May-14
WRLE02670 8447D Hewlett-Packard Preamplifier 2443A03954 Code B 11-Jan-14
WRLE02689 8566B Hewlett-Packard Spectrum Analyzer 2416A00321 22-Apr-14
WRLE03295 85662A Hewlett-Packard Analyzer Display 2349A06144 22-Apr-14
NBLE02683 85650A Hewlett-Packard Quasi-peak Adapter 2430A00495 30-May-14
NBLE02652 SMY02 Rhode & Schwarz Signal Gen 9 KHz-2.08 GHz DE11277 20 May 14
WRLE03236 UHAP-10dB Schwarzbeck Dipole Antenna 300-1000 164 Code Y
WRLE02075 3115 EMCO Ridge Guide Ant. 1-18 GHz 9001-3275 12-Feb-14
WRLE10527 SL18B4020 Phase 1 Microwave Preamplifier 1 – 18 GHz 0001 Code B 08-Jan-14
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
Frequency range Maximum power, Bandwidth
e.r.p. (≤ 1 GHz)
e.i.r.p. (> 1 GHz)
30 MHz – 1 GHz -57 dBm 100 kHz
1 GHz – 12.75 GHz -47 dBm 1 MHz
Test Data
Radiated measurements, 30 – 1000 MHz
Measurement summary for limit1: EN 300 328 <1GHz Rx 3m (Qp)
FREQ LEVEL CABLE / ANT / PREAMP / FINAL POL / HGT / AZ FINAL
(dBuV) ATTEN (dBuV / m) (m)(DEG) (dBm)
(dB)
567.781 MHz 41.18 Qp 2.34 / 18.63 / 27.11 / 0.0 35.04 V / 1.51 / 139 -60.5
543.097 MHz 39.1 Qp 2.3 / 18.08 / 27.18 / 0.0 32.3 V / 1.50 / 180 -63.2
617.155 MHz 36.17 Qp 2.43 / 19.54 / 26.96 / 0.0 31.17 V / 1.39 / 105 -64.3
419.671 MHz 37.3 Qp 1.95 / 16.43 / 27.23 / 0.0 28.45 V / 1.00 / 0 -67
641.827 MHz 31.85 Qp 2.47 / 19.38 / 26.89 / 0.0 26.81 V / 1.00 / 0 -68.6
444.355 MHz 35.2 Qp 2.02 / 16.48 / 27.29 / 0.0 26.42 H / 1.50 / 90 -68.9
358.357 MHz 33.75 Qp 1.77 / 15.0 / 27.25 / 0.0 23.27 V / 1.00 / 0 -72
172.807 MHz 36.7 Qp 1.18 / 9.19 / 27.42 / 0.0 19.64 V / 1.00 / 90 -75.6
176.97 MHz 36.2 Qp 1.2 / 9.37 / 27.41 / 0.0 19.35 V / 1.00 / 0 -75.8
123.438 MHz 34.79 Qp 0.91 / 8.9 / 27.53 / 0.0 17.06 V / 1.00 / 0 -78
Sample substitution performed at 567.781 MHz matching 41.2 dBuV at analyzer input
Signal generator = -48 dBm
Coax loss = 6.3 dB
1/2 wave dipole gain = -6.2 dBi
-48 dBm - 6.3 dB + -6.2 dBi = -60.5 dBm eirp
Limit = -57 dBm
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 27 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
8.11 Receiver Blocking
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
These requirements do not apply for equipment with a maximum declared RF Output power of less than 10 dBm
e.i.r.p.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 28 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9 Technical requirements for other types of Wide Band modulation (Wifi)
Test location
Wild River Lab Large Test Site (Open Area Test Site)
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
NBLE10986 436A Hewlett-Packard Power Meter 2347A17580 01 Nov 13
NBLE10447 8482A HP Power Sensor 3318A26753 11-Mar-14
WRLE03058 2 Inmet 20 dB Attenuator 18N20W-20dB Code B 04-Feb-14
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
The maximum RF output power shall be 20 dBm
Test Data
conducted measurements
Low channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 29 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Mid channel
High channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 30 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.2 Power Spectral Density
Test summary
The requirements are: - MET - NOT MET
The maximum power spectral density is 7.64 dBm / MHz or -52.36 dBm / Hz
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
WRLE03058 2 Inmet 20 dB Attenuator 18N20W-20dB Code B 04-Feb-14
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
The maximum Power Spectral Density is limited to 10 dBm / MHz or -50 dBm / Hz
Test Data
PSD levels displayed in dBm / Hz. Add 60 for dBm / MHz
conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 31 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
1 Mbps, Mid channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 32 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
6 Mbps, Low channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 33 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
6 Mbps, High channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 34 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
11 Mbps, Mid channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 35 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
36 Mbps, Low channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 36 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
36 Mbps, High channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 37 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
54 Mbps, Mid channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 38 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
MCS0, Low channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 39 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
MCS0, High channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 40 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
MCS7, Mid channel
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 41 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.3 Duty Cycle, Tx-sequence, Tx-gap
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
Duty Cycle is defined as the ratio of the total transmitter 'on'-time to a 1 second observation period.
Tx-sequence is defined as a period in time during which a single or multiple transmissions may occur and which shall
be followed by a Tx-gap.
Tx-gap is defined as a period in time during which no transmissions occur.
NOTE: The maximum Duty Cycle at which the equipment can operate, is declared by the supplier.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 42 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.4 Medium Utilization factor
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
MU = (P/100 mW) × DC
where: MU is Medium Utilisation.
P is the RF output power as defined in clause 4.3.2.1.1 expressed in mW.
DC is the Duty Cycle as defined in clause 4.3.2.3.1 expressed in %.
Limit
This requirement does not apply to adaptive equipment unless operating in a non-adaptive mode.
For non-adaptive equipment using wide band modulations other than FHSS, the maximum Medium Utilisation factor
shall be 10 %.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 43 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.5 Adaptivity
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
This requirement does not apply to adaptive equipment unless operating in a non-adaptive mode..
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 44 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.6 Occupied channel bandwidth
Test summary
The requirements are: - MET - NOT MET
Occupied channel bandwidth compliance is revealed in the conducted measurement data for transmitter unwanted
emissions in the out-of-band domain. Emission levels at the 2.4 GHz & 2.4835 GHz band edges are at least 20 dB
below the maximum fundamental levels.
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
Limit
The Occupied Channel Bandwidth shall fall completely within the band 2.4 – 2.4835 GHz
Test Data
See conducted measurement plots for transmitter unwanted emissions in the out-of-band domain, section 9.7
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 45 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.7 Transmitter unwanted emissions in the out-of-band domain
Test summary
The requirements are: - MET - NOT MET
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
NBLE10440 E4446A HP Spectrum Analyzer 44GHz US44300488 25 Sep 14
NBLE02238 SH27 ENVIRONTRONICS 27 Cu Ft T/H Chamber 09963482-5 Code B 29-Oct-13
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
The transmitter unwanted emissions in the out-of-band domain but outside the allocated band, shall not exceed the
values provided by the mask
Test Data
See following pages
Conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 46 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 1 Mbps, 16.7 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 47 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 1 Mbps, 16.7 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 48 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 6 Mbps, 19.7 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 49 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 11 Mbps, 16.7 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 50 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 11 Mbps, 16.7 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 51 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 36 Mbps, 19.7 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 52 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 54 Mbps, 19.7 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 53 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 54 Mbps, 19.7 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 54 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, MCS0, 21 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 55 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, MCS7, 21 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 56 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, MCS7, 21 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 57 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 1 Mbps, 16.7 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 58 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 6 Mbps, 19.7 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 59 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 6 Mbps, 19.7 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 60 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 11 Mbps, 16.7 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 61 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 36 Mbps, 19.7 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 62 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 36 Mbps, 19.7 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 63 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 54 Mbps, 19.7 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 64 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, MCS0, 21 MHz obw, 15° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 65 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, MCS0, 21 MHz obw, 30° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 66 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, MCS7, 21 MHz obw, 23° C
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 67 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.8 Transmitter unwanted emissions in the spurious domain
Test summary
The requirements are: - MET - NOT MET
Test location
Wild River Lab, Tech area
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03371 E4440A Agilent Spectrum Analyzer MY43362222 06-Nov-13
WRLE03058 2 Inmet 20 dB Attenuator 18N20W-20dB Code B 04-Feb-14
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
Frequency range Maximum power, Bandwidth
e.r.p. (≤ 1 GHz)
e.i.r.p. (> 1 GHz)
30 MHz – 47 MHz -36 dBm 100 kHz
47 MHz – 74 MHz -54 dBm 100 kHz
74 MHz – 87.5 MHz -36 dBm 100 kHz
87.5 MHz – 118 MHz -54 dBm 100 kHz
118 MHz – 174 MHz -36 dBm 100 kHz
174 MHz – 230 MHz -54 dBm 100 kHz
230 MHz – 470 MHz -36 dBm 100 kHz
470 MHz – 862 MHz -54 dBm 100 kHz
862 MHz – 1 GHz -36 dBm 100 kHz
1 GHz – 12.75 GHz -30 dBm 1 MHz
Test Data
See following pages, conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 68 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 1 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 69 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 1 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 70 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 6 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 71 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 6 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 72 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 11 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 73 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 11 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 74 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 36 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 75 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 36 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 76 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, 54 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 77 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, 54 Mbps
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 78 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, MCS0
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 79 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, MCS0
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 80 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Low channel, MCS7
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 81 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
High channel, MCS7
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 82 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.9 Receiver spurious emissions
Test summary
The requirements are: - MET - NOT MET
Minimum margin of compliance is 3.16 dB at 567.781 MHz
Test location
Wild River Lab Large Test Site (Open Area Test Site)
Test equipment
TUV ID Model Manufacturer Description Serial Cal Due
WRLE03204 EM-6917B Electro-Metrics Biconicalog Periodic 102 30-May-14
WRLE02670 8447D Hewlett-Packard Preamplifier 2443A03954 Code B 11-Jan-14
WRLE02689 8566B Hewlett-Packard Spectrum Analyzer 2416A00321 22-Apr-14
WRLE03295 85662A Hewlett-Packard Analyzer Display 2349A06144 22-Apr-14
NBLE02683 85650A Hewlett-Packard Quasi-peak Adapter 2430A00495 30-May-14
NBLE02652 SMY02 Rhode & Schwarz Signal Gen 9 KHz-2.08 GHz DE11277 20 May 14
WRLE03236 UHAP-10dB Schwarzbeck Dipole Antenna 300-1000 164 Code Y
WRLE02075 3115 EMCO Ridge Guide Ant. 1-18 GHz 9001-3275 12-Feb-14
WRLE10527 SL18B4020 Phase 1 Microwave Preamplifier 1 – 18 GHz 0001 Code B 08-Jan-14
Code B = Calibration verification performed internally. Code Y = Calibration not required when used with other calibrated equipment
Test limit
Frequency range Maximum power, Bandwidth
e.r.p. (≤ 1 GHz)
e.i.r.p. (> 1 GHz)
30 MHz – 1 GHz -57 dBm 100 kHz
1 GHz – 12.75 GHz -47 dBm 1 MHz
Test Data
Radiated measurements, 30 – 1000 MHz
Measurement summary for limit1: EN 300 328 <1GHz Rx 3m (Qp)
FREQ LEVEL CABLE / ANT / PREAMP / FINAL POL / HGT / AZ FINAL
(dBuV) ATTEN (dBuV / m) (m)(DEG) (dBm)
(dB)
567.781 MHz 41.18 Qp 2.34 / 18.63 / 27.11 / 0.0 35.04 V / 1.51 / 139 -60.5
543.097 MHz 39.1 Qp 2.3 / 18.08 / 27.18 / 0.0 32.3 V / 1.50 / 180 -63.2
617.155 MHz 36.17 Qp 2.43 / 19.54 / 26.96 / 0.0 31.17 V / 1.39 / 105 -64.3
419.671 MHz 37.3 Qp 1.95 / 16.43 / 27.23 / 0.0 28.45 V / 1.00 / 0 -67
641.827 MHz 31.85 Qp 2.47 / 19.38 / 26.89 / 0.0 26.81 V / 1.00 / 0 -68.6
444.355 MHz 35.2 Qp 2.02 / 16.48 / 27.29 / 0.0 26.42 H / 1.50 / 90 -68.9
358.357 MHz 33.75 Qp 1.77 / 15.0 / 27.25 / 0.0 23.27 V / 1.00 / 0 -72
172.807 MHz 36.7 Qp 1.18 / 9.19 / 27.42 / 0.0 19.64 V / 1.00 / 90 -75.6
176.97 MHz 36.2 Qp 1.2 / 9.37 / 27.41 / 0.0 19.35 V / 1.00 / 0 -75.8
123.438 MHz 34.79 Qp 0.91 / 8.9 / 27.53 / 0.0 17.06 V / 1.00 / 0 -78
Sample substitution performed at 567.781 MHz matching 41.2 dBuV at analyzer input
Signal generator = -48 dBm
Coax loss = 6.3 dB
1/2 wave dipole gain = -6.2 dBi
-48 dBm - 6.3 dB + -6.2 dBi = -60.5 dBm eirp
Limit = -57 dBm
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 83 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
9.10 Receiver Blocking
Test summary
The requirements are: - MET - NOT MET - NOT APPLICABLE
This requirement does not apply to non-adaptive equipment or adaptive equipment operating in a non-adaptive mode.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 84 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
10. Test-setup photo(s):
Radiated emissions
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 85 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Radiated emissions
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 86 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Extreme measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 87 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Extreme measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 88 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Conducted measurements
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 89 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
11. GENERAL INFORMATION
Test Methodology
Emission testing is performed according to the procedures in EN 300 328.
Justification
The Equipment Under Test (EUT) is configured in a typical user arrangement in accordance with the
manufacturer's instructions. A cable is connected to each available port and either terminated with a peripheral
into its characteristic impedance or left unterminated. When appropriate, the cables are manually manipulated with
respect to each other to obtain maximum emissions from the unit.
It is the manufacturer's responsibility to assure that additional production units of this model are manufactured with
identical electrical and mechanical characteristics. Any modifications necessary for compliance made during testing on
the above mentioned date(s) must be implemented in all production units for compliance to be maintained.
Radiated Emissions
The spectrum analyzer uses a quasi-peak detector for frequencies up to and including 1 GHz. For measurements
above 1 GHz, peak and average detectors are used. The bandwidth used are equal to or greater than 100 Hz from 9
kHz to 150 kHz, 9 kHz from 150 kHz to 30 MHz, 100 kHz from 30 MHz to 1000 MHz, and 1 MHz from 1 GHz to 40
GHz. Video bandwidth are at least three times greater than the IF bandwidth
The final level, in dBV/m, equals the reading from the spectrum analyzer (Level dBV), adding the antenna
correction factor and cable loss factor (Factor dB) to it, and subtracting the preamp gain (and duty cycle correction
factor, if applicable). This result then has the limit subtracted from it to provide the Delta, which gives the tabular
data as shown in the data. Intentional radiators are rotated through 3 orthogonal axes to determine the test position
yielding the maximum emission levels.
Example:
FREQ LEVEL CABLE/ANT/PREAMP FINAL POL/HGT/AZ DELTA1
(MHz) (dBuV) (dB) (dB/m) (dB) (dBuV/m) (m) (deg)
60.80 42.5Qp + 1.2 + 10.9 - 25.5 = 29.1 V 1.0 0.0 -10.9
Test Equipment
All measurement instrumentation is traceable to the National Institute of Standards and Technology and is calibrated
according to internal procedure.
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 90 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
12. CONSTRUCTIONAL DATA FORM
Test Report NC1306936.2 EN 300 328 v1.8.1 for IDEXX Laboratories model 89-91921-02 Page 91 of 98
TÜV SÜD AMERICA INC 19333 Wild Mountain Road Taylors Falls MN USA 55084-1786 Tel: 651 638 0297 Fax: 651 638 0298
Form
General Equipment Description -- NOTE: This information will be input into your test report as shown below.
EUT Description Portable, battery powered, imaging diagnostics instrument used in veterinary labs.
Instrument activates tests on animal fluid samples in disposable snap cartridges
and captures images of the chemical reactions. Instrument can optionally be
networked wirelessly (WiFi or Bluetooth) or wired (Ethernet or USB).
EUT Name IDEXX SNAP Pro
Model No.: 89-91921-02 Serial No.:
Product Options: N/A
Configurations to be tested: 1 - typical
Equipment Modification (If applicable, indicate modifications since EUT was last tested. If modifications are made
during this testing, submit revised TP/CDF after testing is complete.)
Test Objective(s): Please indicate the tests to be performed, entering the applicable standard(s) where noted.
EMC Directive 2004/108/EC (EMC) FCC: Class A B Part
Std: EN 55024, EN 55022:2010 and FCC VCCI: Class A B
testing up to 2.4GHz intentional
Machinery Directive 89/392/EEC (EMC) BSMI: Class A B (Separate Report)
Std: Canada: Class A B
Medical Device Directive 93/42/EEC (EMC) Australia: Class A B
Std: Other: CE Mark and Japan
Vehicle Directive - 2004/104/EC (EMC) Ag Directive *2009/64/EC (EMC)
Other Vehicle Std:
FDA Reviewers Guidance for Premarket
Notification Submissions (EMC)
Attendance
Test will be: Attended by the customer Unattended by the customer
Failure - Complete this section if testing will not be attended by the customer.
If a failure occurs, TÜV SÜD America should:
Call contact listed above, if not available then stop testing. (After hrs phone):
Continue testing to complete test series.
Continue testing to define corrective action.
Stop testing.
Power Requirements
Regulations require testing to be performed at typical power ratings in the countries of intended use. (i.e.,
European power is typically 230 VAC 50 Hz or 400 VAC 50 Hz, single and three phase, respectively)
Voltage: 3.7V (If battery powered, make sure battery life is sufficient to complete testing.)
# of Phases: N/A
Current Current
(Amps/phase(max)): 1.46A (Amps/phase(nominal)): 0.6A
Other
Connector Port
Type Type Termination Type Termination
Equipment Under Test (EUT) Operating Modes to be Tested -- list the operating modes to be used during test.
It is recommended the equipment be tested while operating in a typical operation mode. FCC testing of personal computers and/or
peripherals requires that a simple program generate a complete line of upper case H’s. Provide a general description of all software,
firmware, and PLD algorithms used in the equipment. List all code modules as described above, with the revision level used during
testing. Consult with your TÜV Product Service Representative if additional assistance is required.
1. Linux Kernel running on SOM with microSD Card, applications, and files.
2.
3.
Equipment Under Test (EUT) System Components -- List and describe all components which are part of the EUT.
For FCC & Taiwan testing a minimum configuration is required. (ie. Mouse, Printer, Monitor, External Disk Drive, Motherboard, etc)
Description Model # Serial # FCC ID #
ASSEMBLY, DM3730 Torpedo SOMDM3730- YKP1024119
SOM with 802.11 A/B/G/N and IDX01-0301R-A 10029A-1024119
Bluetooth capability
ASSY, IDEXX, TOP BOARD 2053-79-91946-02
ASSY, IDEXX, VERTICAL BRD 2053-79-91945-02
ASSY, IDEXX, INTERFACE BRD 2053-79-91944-01
ASSY, IDEXX, BATTERY DOOR 2053-79-91948-01
BRD
ASSY, IDEXX, BATTERY FRONT 2053-79-91941-00
BRD
ASSY, IDEXX, LED FLEX BOARD 2053-79-91951-01
Truly 3.5" TFT Display w/Touch TFT2N2126
Oscillator Frequencies
Derived
Manufacturer Frequency Frequency Component # / Location Description of Use
Power Supply
Manufacturer Model # Serial # Type
N/A
N/A
EMC Critical Detail -- Describe other EMC Design details used to reduce high frequency noise.
N/A
System Configuration Block Diagram -- Provide a line drawing identifying the EUT, simulators, support equipment, I/O
cables, power cables, and any other pertinent components to be used during testing. Use a dashed line to separate the equipment
in the testing field versus equipment outside testing field.
Authorization Signatures