User manual
UV- BOX BK 850
Inhalt:
V 2.01e Page 2
12. Voltage Converter _________________________________________________________ 32
13. Circuit diagram ____________________________________________________________ 34
14. Appendix _________________________________________________________________ 35
V 2.01e Page 3
1.1 Signs within the manual
Attention!
Warning UV radiation
V 2.01e Page 4
2. Description of the device
With the UV curing chamber you have purchased a modern device for irradiating UV-
reactive materials. Please read the instructions carefully and follow the advice. Make sure
to consult the instructions for the proper use and safety instructions. The applications of the
UV chamber are very diverse. In addition to the usual tasks, such as curing and bonding,
the chamber used for irradiation for chemical and biological applications.
The UV chamber consists of the device housing (1) with an electric or manual closure
device (device version information) and a built-in appliances shell lamp unit (2). A robust
closure and appropriate safety interlocks ensure job security.
V 2.01e Page 5
Abb. 2: UV irradiation chamber with electric shutter and timer
Note:
The UV chamber is equipped with a safety interlock which prevents opening of the UV
chamber during the UV irradiation or makes starting of the UV lamp with the door open
impossible.
V 2.01e Page 6
(1) (2)
Abb. 3: (1) = Irration chamber, (2) = power cable
V 2.01e Page 7
3. Application area
The UV irradiation chamber BK 850 is exclusively for curing UV - sensitive materials such
as paints, varnishes, adhesives and potting compounds. You can similar both in production
and in the laboratory to produce or repair be used.
One application for other drying processes, such as the drying of solvent-based materials is
not permitted!
With an appropriate concentration of the solvent can cause explosions! The UV irradiation
chamber is not suitable for use in hazardous areas. Any use other than described in this
manual is prohibited.
V 2.01e Page 8
4. Safety information
The UV-Bridge corresponds to the actual technique standard and to the accepted safety
rules. Nevertheless dangers can appear.
The voltage for ignition of the UV-Lamp is ca. 4-5 kV. Therefor during all
tasks on the device, turn off the main switch and separate from power
supply - disconnect the plug! Danger for life! After Turn off, wait for
approx. 5 minutes for discharge of the condensers! Let the filter pane at
the front of the device cool down!
Operate the UV-Hand lamp only in a proper condition and take notice of
the user manual.
4.1 UV-Radiation
UV-Radiation produces the generation of ozone. Therefor take care about a sufficient
ventilation for fresh air.
Avoid a direct or indirect radiation to the skin by using suitable clothes. Additional to the
dosage of the radiation, the spectrum is from great meaning.
V 2.01e Page 9
4.1.3 Effect of UV-Radiation to the eyes
Avoid direct and indirect radiation of the eyes by using suitable safety precautions (i.e.
safety goggles).
Basically, safety precautions against UV-Radiation are not difficult, since it is „non transient
noxe“, which follows simple physically rules of extension. As well, the so-called indirect
possible imperilments of a short wave UV-Radiation, due to the production of ozone, can be
simply dominated by suction near to the source. It may be irradiated are no living beings of
any kind into this unit!
• Mark the areas, where especially high performance UV-C-lamps are installed.
• Carefully check for metallic items with smooth surface within the radiation area, which
can generate extensive reflections. (cover or paint)
• The use of light density protection covers, which only can be opened by tooling and
which are perhaps equipped with interlock switches.
• Flaps for material or similar easy to open orifices should be equipped with interlock
switches (safety switches) in any case.
Staff should be informed about the available UV-Radiation sources and the safety
precautions.
Therefor we recommend the letters of the Bundesanstalt für Arbeitsschutz "Schutz vor
ultravioletter Strahlung" ( ISBN 3-88314-352-9 ).
For generation of UV-Light, the lamp contains a small amount of quicksilver. If the lamp will
be destroyed or breaks, steam of quicksilver drops out.
Safety precautions: Leave the room and ventilate properly. After that, take the rest of
quicksilver using suitable materials and waste. We recommend
the corresponding UV-Safety kit of the manufacturer.
V 2.01e Page 10
5. Warranty, Liability and Safety
Warranty and Liability:
Basically, the general conditions of the manufacturer apply. Warranty and liability claims for
personal injury or material damage. Especially when these are due to the following reasons:
Safety
V 2.01e Page 11
6. Installation of the system
On the back of the UV irradiation chamber is a connector for the power supply (230V /
50Hz). Note. Before connecting the technical data in Section 9
When installing the lamp, make sure that the fan is not delivered. Make sure that the air
inlets and outlets are free.
Connect the foot switch, if you want to start with this lamp.
Abb. 6
V 2.01e Page 12
7. Startup
Check before you connect, whether the information on the nameplate with the supply
voltage.
3
4
Title:
1) irradiation chamber
2) front panel
3) Operating hours counter
4) Timer (only for version "electric lock)
5) Foot switch for starting the irradiation cycle (only for version "electric lock)
V 2.01e Page 13
7.1 Turning on the lamp
Switching on / off
The device is switched on or off at the switch (1). The UV lamp is turned on by
the switch (3) + (2) button off.
Note:
After switching off the radiator by button (2), the lamp should be about 3-5
minutes. Post-cooled before being completely off using the power button (1).
• The UV lamp on button 3 switch. The radiator is ignited. A short delay the ignition is
normal. The ignition can only take place with the shutter closed.
• The lamp now requires about 3 minutes to reach full power. After this time, the lamp is
ready.
V 2.01e Page 14
7.2 Turn off the system
• Switch the UV lamp on button 2. The spotlight goes out. The fan will continue to run until
the device with the device switch (1) is switched off completely.
• The lamp unit now takes about 3-5 minutes to cool sufficiently again. The UV chamber
complete with switch 1 off after this time, otherwise there is a risk of heat stress.
To set the mode (starting with foot switch or start button on the switch), please proceed as
follows:
The setting of the irradiation time, you can also change as follows:
V 2.01e Page 15
You can switch between hours and minutes display as follows:
Press and hold for 3 seconds to get into the programming mode. Press 1 x and 1 x
to move to the mode setting. he appears the display . While you can hold
down the buttons or select the desired setting. min = display in minutes - Hour =
display in hours. Complete adjustment by 1 x and 1 x .
Briefly press the button - the display will begin to flash. Now you can select the or
buttons to select the setting. By continuously holding the keys you can accelerate the speed
of change. By default, the steps to "1". If you want to change this, see the instructions,
below 7.6. Programming mode.
V 2.01e Page 16
7.6 Digital Timer
V 2.01e Page 17
V 2.01e Page 18
V 2.01e Page 19
7.3 UV Dosis
The UV irradiation chamber is operated with 50 or 100% power. Dosing is by changing the
exposure time of the product.
Intensität in
mW/cm²
900
800
700
600
500
400
300
200
100
0
0 5 10 15 20 25 30 35 40 45 50 55 60
Abstand in cm
V 2.01e Page 20
7.4 UV Measuring
The UV irradiation chamber is designed so that when the door is either a) closed a probe or
b) may express a probe and run the appropriate cable to the outside.
.
Opening for measuring sensor or sensor cable to the front of the UV irradiation
chamber
V 2.01e Page 21
8. Maintenance
For all maintenance and repair work, the UV irradiation chamber is disconnected from the
mains and secured against unintentional switching on again. The discharge of the
capacitors - about 3 minutes - is to be seen.
To avoid damage to the UV irradiation chamber and to ensure safe operation, use only
original Beltron UV light can be fitted. When using other manufacturers' approval and void
your warranty.
V 2.01e Page 22
The change of the UV lamp is carried out in the following steps:
When lamp replacement, always wear cotton gloves. Never touch the lamp with bare
hands. If the change of the UV lamp in the UV irradiation chamber are necessary, proceed
as follows:
Change UV Lamps:
Turn off power and disconnect the power cable from the power supply, hot parts such as.
Let cool radiator and note the capacitors to discharge.
Lamp unit holder solve (not Pull out the screws - just loosen)
Remove the lamp housing screw and rotate 180 °; so that the lamp opening points
upwards
V 2.01e Page 23
Framework and remove filter disk,
Strahler kann getauscht werden durch leichten seitlichen Druck nach rechts oder links und
danach Strahler nach oben herausnehmen, der Saugnippel zeigt in Richtung Reflektor. 4.
emitters can be replaced by a slight lateral pressure to the right or left and then remove the
radiator up the teat pointing towards the reflector.
V 2.01e Page 24
Please note while lamp replacement:
In the UV lamps used by Beltron GmbH is the equipment covered by electric Act provided
for disposal. The end user is entitled to dispose of them for free at their local yard / local
collection point, since Beltron GmbH and our suppliers are registered with the Electrical and
Electronic Equipment tab.
V 2.01e Page 25
8.3 Inspection and cleaning
Depending on the location and use a cleaning of the UV chamber at regular intervals is
necessary. Free the fans and the rest of the housing area of dust. In particular, make sure
that the vents be inspected and cleaned if necessary. The cleaning of the filter disc, the
reflection chamber and the lamp should be done with a lint-free cloth. For fixed
contamination clean them with alcohol.
The door lock (locking the door open, closure) must also be checked regularly. In this case,
make sure that no UV light can escape from the irradiation chamber.
V 2.01e Page 26
9. Technical data
9.1 Dimensions
Dimensions of chamber
V 2.01e Page 27
9.2 Weight
V 2.01e Page 28
10. Spare parts
V 2.01e Page 29
11. Debugging
Fault:
After turning on the UV irradiation chamber at the Switch (N) no reaction
Possible causes:
• Check the power cord
• Built-in backup device plug check
Fault:
After pressing the button 3 (lamp ON), no ignition of the UV lamp
Note: A short flash of the fault lamp and ignition delay is normal
Possible causes:
• Switch (N) is not turned on
• Spotlight is not yet sufficiently cooled
• UV lamp defective
• Lamp closure not in "Close"
Fault:
After prolonged use of the UV lamp switches off automatically
Possible causes:
• Overheating of the lamp - Check whether the vents are clean and free of
• Ambient temperature is too high
Fault:
There is no sufficient UV drying instead
Possible causes:
• UV lamps used
• exposure time too short
• Lamp may need (power 100% UV) are switched to full load
• dirty filter disc / reflectors
V 2.01e Page 30
Problem:
Shutter will not open
Solution:
The shutter can only be opened when the door is closed (mechanical interlock)
Problem:
Door will not open
Solution:
The door can only be opened when the UV lamp when the shutter is fully closed (electric
lock) - lock to position "CLOSE“
V 2.01e Page 31
12. Voltage Converter
The UV dryer is supplied with an additional voltage converter to operate at 115V / 60 Hz.
Item 1
Back view
The voltage converter is connected to the mains plug to 115V / 60 Hz outlet. The UV -
Dryer is connected with the supplied plug and socket (Item 1) with the voltage converter.
Item 2
Front view
With the switch (item 2), the voltage converter can be switched “ON” and "OFF", as
required. The converter is protected by a fuse 10A. To change this, the housing must be
opened. (Before opening the housing the plug must be unplugged!).
Note: The operating voltage of the converters is only 115V / 60 Hz. If the UV dryer is used
with other voltages / frequencies it may be damaged!
V 2.01e Page 32
Integrated Fuse
When a short circuit or overcurrent the fuse (F1) triggers in the ballast. This is a circuit
breaker which can be switched on after the removal of the fault. For this purpose, please
unscrew the housing.
Note: Before opening the housing unit from the power supply disconnect. Pull the power
plug!
Specifications
Type autotransformer
800 VA
V 2.01e Page 33