Thermal Desorption Guide - Environmental Air Monitoring
Thermal Desorption Guide - Environmental Air Monitoring
www.markes.com
Introduction to Markes What is TD?
International Ltd.
Since the early 1980s, thermal desorption has
Formed in 1997, Markes International Ltd. is one of provided the ultimate versatile sample
the world’s leading suppliers of thermal desorption introduction technology for GC / GC-MS. It
(TD) equipment for monitoring trace toxic and combines selective concentration enhancement
odorous chemicals in air, gas and materials. Serving with direct extraction into the carrier gas and
fast growing markets from environmental health and efficient transfer / injection all in one fully
safety to materials testing and from food / flavour / automated and labour-saving package.
fragrance to defence / forensic, Markes’ global
customer base includes major industry, government
agencies, academia and the service laboratory sector.
Trichloroethylene
Methylene Chloride
effectively monitored using pumped sampling onto
Toluene
HC
Oxygenated HC
multi-sorbent tubes / traps with automated
Benzene
TD-GC-MS(SCAN) analysis.
Std. method: US EPA Method TO-17
1,1,1-trichloroethane
Dimethyl Sulphide
Typical TD-GC analytical conditions:
CCL2F2
Freon 113
CCl3F
Sampling: Pumped sorbent tube
C2Cl4
SO2
CCl4
TD system: ULTRA-UNITY
CH3Br
CH3Cl
COS
HC
CHClF2
CS2
HC
10
12
16
(~500 ml/min) pumped sampling of the vapour
11
fraction onto tubes packed with a combination of
13
15
8
9
quartz wool and carbon sorbents followed by TD-
7
GC-MS analysis allows measurement of PAH at low
14
ppt levels.
1
6
5
4
MEK 580 583 580 Desorption: 5 mins at 330ºC or 280ºC (if Tenax)
Benzene 0.14 0.18 0.18 Trap: Tenax / carbon or 2 carbons 30 to 300ºC
Toluene 94 91 93 Split: 3,000:1 double split with SecureTD-Q
Ethyl benzene 30 30 29 Analysis: GC-MS(SCAN) or GC-FID
PGMEA 43 43 43
Xylene 274 275 271 Reference: TDTS77 on stack emissions
DMS 28 28 28 monitoring
Trimethylbenzene 43 44 42
Vapour-phase semi-volatiles by Background:
Thermal Desorption Thermal desorption is usually associated with
analysis of volatile organic chemicals. However,
the short, inert, heated flow path of Markes TD
systems also ensures quantitative recovery of
semi-volatiles up to n-C40.
C18
TD system: ULTRA-UNITY
C36
Desorption: 15 mins at 350ºC
C40
Trap: High boilers trap: -10 to 350ºC
Split: Typically 50:2.5 during trap desorption only
Complete recovery of n-C40 through Markes UNITY TD Analysis: GC-MS(SCAN)
CS2
H2S
Dimethyl disulphide
compounds are usually sampled on-line or in
canisters / bags and analysed using TD-GC-FPD.
Ethyl mercaptan
Std. method: Korean Government Guidance
Method - Standard Method for Off-Odour Analysis
Dimethyl sulphide (2005)
Typical TD-GC analytical conditions:
Methyl mercaptan
Typical analytes:
Freons such as CF4, C2F6 and C3F8
Toluene
e.g. 3 to 14 days. When combined with
subsequent automated TD-GC(-MS) analysis, this
Hexane
Pentane
Xylene
Typical TD-GC analytical conditions:
Sampling: Diffusive (passive) tubes
Sorbent: Carbograph 1TD, Carbopack X or other
TD system: ULTRA-UNITY
2-week diffusive sampling around a refinery perimeter. Desorption: 5 mins at 320ºC
VOCs detected include benzene, toluene & xylene
Trap: Dual carbon black -10 to 320ºC
Split: Low split during trap desorption only
Analysis: GC-MS(SCAN) or GC-FID
Typical analytes:
• Benzene Reference: TDTS49 on fence-line monitoring
• 1,3-butadiene
• Mixed solvents and hydrocarbons
Concentrations: ppb to ppm
In-situ monitoring of Background:
underground fuel leaks Underground fuel or chemical leaks present a
grave environmental risk. Soil probes containing
standard diffusive tube samplers allow cost-
effective, in-situ screening of large areas of land
including active production sites. They can also be
placed along the length of fuel pipelines to provide
early warning of a leak. Diffusive tubes inside the
soil probes monitor the soil gases for ~24 hours.
Automated TD-GC analysis allows rapid
identification of the nature, source and spread of
ground contamination.
Typical TD-GC analytical conditions:
Sampling: Diffusive tubes inside soil probes
Sorbent: Tenax TA
TD system: ULTRA-UNITY
Desorption: 5 mins at 280ºC
Trap: Tenax TA or dual bed -10 to 300ºC
VOC-Mole™ Soil Probes arranged in a grid pattern
around an industrial site allow low-cost mapping of Split: Double split
contaminated ground Analysis: GC-MS(SCAN) or GC-FID
α-pinene
Butanoic acid ethyl ester
Toluene
sorbent tube using a gas syringe. Subsequent
Redesorption blank gentle desorption on an inert ULTRA-UNITY TD set
Limonene
at low (<100ºC) flow path temperatures allows
Decane
measurement of trace target analytes including
Butan-2-ol
Chloroethane
thiols.
1,1- & 1,2-dichloroethene
Butan-1-ol
Nonane
Official guidance: UK Env. Agency publication
Dimethyl sulphide
Xylene
Dimethyl disulphide
1,1,1-trichloroethane
Vinyl chloride
Trichloroethene
‘Monitoring trace components in landfill gas.’
1,1-dichloroethane
1-pentene
Carbon disulphide
Benzene
Trimethylbenzene
Toluene
Hexane
Trichlorofluoromethane
with subsequent TD-GC-MS analysis for profiling of
Ethyl benzene
ppt-ppb level VOCs.
1,1-difluoro-1-chloroethane
Benzyl alcohol
Freon 113
Dichloromethane
Tetrachloroethane
1,1,1,2-tetrafluoroethane
o/p xylene
Std. methods: US EPA Method TO-17, EN ISO
Nonanal
Dichlorodifluoromethane
Ethylcyclohexane
Undecane
16017-1, ASTM D 6196
α-pinene
Ethyl acetate
Dodecane
Tetrachloromethane
Tridecane
1,1-difluoroethane
Limonene
Styrene
Tetradecane
Propanol
Decane
Isopentane
Acetone
Hexanal
Sorbent: Typically multi-sorbent
TD system: ULTRA-UNITY
Desorption: 5 mins at 280ºC (depends on sorbent)
Clean indoor air pumped onto multi-sorbent tube and Trap: Tenax TA or dual bed -10 to 300ºC
analysed by TD-GC-MS
Split: During trap desorption only ~15:1
Analysis: GC-MS(SCAN)
Dimethylbenylamine
the quality of cabin air and reducing emissions
from vehicle trim components. This will reduce
Styrene m-/p-xylene
2-(2-butoxy-ethoxy)ethanol
Toluene
Trimethyl benzene
C11
Cubebene / Copaene
n-Decane
sampling onto sorbent tubes with TD-GC-MS
C12
Methylcyclohexane
11/12
C13
analysis is the method of choice for profiling car
Ethyl benzene
C
Ethyl toluene
Silyl ester
n-Hexane
cabin air.
Cyclohexane
n-Nonane
o-Xylene
n-Octane
Isooctane
Benzene
NN DMF
Heptane
n-C13
Std. methods: EN ISO 16017-1, ISO 16000-6,
MEK
ASTM D6196
TD-GC analytical conditions:
n-decane
Dimethylbenzylamine
Saloon car cabin air 40°C C12 Sampling: Pumped sampling of 2 L volume
2-(2-butoxy-ethoxy)ethanol
TVOC 9.7 ppm
C11/12
Trimethylbenzene
C13
TD system: ULTRA-UNITY
m-/p-xylene
C11
n-C13
Desorption: 6 mins at 280ºC
n-C14
Methylcyclohexane
Silyl ester
n-nonane o-xylene
Dodecane
Copaene
Styrene
Split: 75:1
NN DMF
Analysis: GC-MS(SCAN)
PMCP
First
reduces monitoring costs. After sample collection
PDCB
Floor
tubes are analysed via TD-GC-MS/ECD. The rise
and subsequent decay in tracer gas concentrations
PMCH allows the building ventilation (rate of air exchange)
to be monitored.
PMCH
Typical TD-GC analytical conditions:
Ground Sampling: Diffusive or pumped
Floor Sorbent: Carbograph 1TD (40-60 mesh)
TD system: ULTRA-UNITY
PMCP
Desorption: 5 mins at 320ºC
PDCB Trap: Carbograph 1TD -10 to 300ºC
Split: ~10:1
Analysis: GC-MS or ECD
FAN
PNCB
1PPD
TD-GC(-MS) analysis provides a solvent-free, safe
4A
analytical option with ~1000 x more sensitivity
than conventional charcoal tube / CS2 extraction
Chloroform
Sulphur dioxide
4N
methods. TD tubes are also reusable indefinitely.
1-chloro-1,3-butadiene
Chlorpyrifos
Methylchlorpyrifos
Methacrifos
Methylpirimifos
Dichlorvos
Diazinon
Etrimfos
Malathion
(glass or Silcosteel) tubes together with TD-GC-MS
analysis provides a reliable and highly sensitive
Phosphamidon
Fenitrothion
monitoring method.
Std. methods: UK MDHS series, EN ISO 16017-1,
ASTM D 6196, NIOSH 2549
Typical TD-GC analytical conditions:
Sampling: Pumped
Sorbent: Tenax in glass or Silcosteel tubes
TD system: ULTRA-UNITY
Desorption: 10 mins at 280ºC
Trap: Tenax -10 to 300ºC
Primary and repeat analysis of pesticides for personal
exposure monitoring Split: ~10:1 during trap desorption only
Analytes: Analysis: GC-MS
2-butanone
and inhalation. Alveolar breath sampling using the
disposable Bio-VOC allows large-scale, non-invasive
Acetone
Toluene
guidance notes available from UK HSL.
Xylene
Typical TD-GC analytical conditions:
Sorbent: Tenax TA (35-60 mesh)
TD system: ULTRA-UNITY
Desorption: 5 mins at 250ºC
Skin-absorbed solvents in the breath of shoe workers Trap: Tenax TA -10 to 280ºC
collected using the Bio-VOC™
Split: ~10:1
Analysis: GC-MS(SCAN)
Typical analytes include:
Reference: TDTS13 Evaluation of the
• Halogenated solvents Bio-VOC, TDTS48 Overview of breath
• Styrene sampling
• Ketones
• Aromatics
The Markes International advantage
Trademarks
UNITY™, ULTRA™, Air Server™, µ-CTE™, SafeLok™, DiffLok™, VOC-Mole™,
Bio-VOC™, TT24-7™, TC-20™, UniCarb™ and SecureTD-Q™ are trademarks of
Markes International Ltd., UK The Markes International team
Tenax TA™ is a trademark of Buchem B.V., Netherlands
Carbograph 1TD™ is a trademark of LARA s.r.l., Italy
Carbopack X™ is a trademark of Supelco Inc., USA
Silcosteel™ is a trademark of Restek Inc., USA
Chromosorb 106™ is a trademark of Manville Corp., USA