TECHNICAL DATA
Measurement specifications
Temperature range2) 50 ... 400 °C (3ML)
(scalable via programming 100 ... 600 °C (3MH)
keys or software) 150 ... 1000 °C (3MH1)3)
General specifications 200 ... 1500 °C (3MH2)3)
250 ... 1800 °C (3MH3)3)
Environmental rating IP 65 (NEMA-4)
Spectral range 2.3 μm
Ambient temperature1) –20 ... 85 °C (sensing head, 50 °C with laser ON)
–20 ... 85 °C (electronics) Optical resolution 60:1 (3ML)
(90 % energy) 100:1 (3MH)
Storage temperature – 40 ... 125 °C (sensing head) 300:1 (3MH1 – 3MH3)
– 40 ... 85 °C (electronics)
System accuracy4) ±(0.3 % of reading +2 °C)
Relative humidity 10 – 95 %, non-condensing (at ambient temp. 23 ±5 °C)
Vibration (sensor) IEC 60068-2-6 (sinus shaped) Repeatability ±(0.1 % of reading +1 °C)
IEC 60068-2-64 (broadband noise) (at ambient temp. 23 ±5 °C)
Shock (sensor) IEC 60068-2-27 (25G and 50G) Temperature resolution 0.1 K
(digital)
Weight 600 g (sensing head) / 420 g (electronics)
Exposure time5) (90 % signal) 1 ms
Electrical specifications
Emissivity/ Gain 0.100 – 1.100
Outputs / analog 0 / 4 – 20 mA, 0 – 5 / 10 V, thermocouple J, K (adjustable via programming
keys or software)
Outputs / alarm 24 V / 50 mA (open collector)
Transmissivity/ Gain 0.100 – 1.100
Optional Relay: 2 x 60 V DC / 42 V ACeff; 0.4 A; (adjustable via programming
optically isolated keys or software)
Outputs / digital USB, RS232, RS485, CAN, Profibus DP, Ethernet Signal processing (parameter Peak hold, valley hold, average; extended hold
(optional) adjustable via programming function with threshold and hysteresis
Output impedances mA max. 500 Ω (with 8 – 36 V DC) keys or software, respectively)
mV min. 100 kΩ load impedance
Software optris Compact Connect
thermocouple 20 Ω
1) The
functioning of the LCD display may be limited in ambient temperatures
Inputs Programmable functional inputs for external
below 0 °C
emissivity adjustment, ambient temperature
2) T
compensation, trigger (reset of hold functions) object >Tsensing head +25 °C
3) Specification valid at T
object ≥ start of measurement range +50 °C
Cable length 3 m (standard), 8 m, 15 m 4) ε = 1, response time 1 s
Optical specifications
S 20 11.0 0.85 9.5 16.8 29.1 41.4 53.6 78.2 103 128 152 177 (mm)
S 20 18.3 16.6 14.9 13.2 11.4 9.7 8 6.3 4.5 7.3 13 19 24 (mm)
Dimensions