Test Report Page 1 of 1

27/11/2019 10:57
Royal Liverpool Hospital Version: 4.1.1 DB: 3.8.2
Medical Engineering
Contact: SUE
Phone:
E-mail:

Control#: 123 Work order:


Manufacturer: GE Location: Liverpool
Device type: 3500 Procedure ID:
Serial#: 123 Technician: SUE
Date of purchase: Tester ID: ESA615, 4737041
Standard: IEC62353_Dir Class/Type: I B
AP designation: Module1: ECG Leads, CF - 1, ID:
MAINS VOLTAGE: Current Consumption: 0A
L1-Earth 242.6 V
L2-Earth 0.6 V Protective Earth Resistance [Ohm]: >2 [0.3] *F*
L1-L2 241.9 V Test Current: 0.2 A
INSULATION RESISTANCE [MOhm]:
Iso Mains >100 [-]
DIRECT EQUIPMENT LEAKAGE CURRENT [uA]:
NORMAL
No Earth 1 [500]
REVERSED
No Earth 1 [500]
DIRECT PATIENT LEAKAGE CURRENT [uA]:
1-ALL
NORMAL 0 [50]
REVERSED 1 [50]
Comments: We are doing training with Sue in office at liverpool.

UNIT FAILED