Leak Tes te r

MFY-01

INSTRUCTION MANUAL
for
Leak Tester
MFY-01

Labthink

1
Leak Tes te r

MFY-01

Notice

The information contained in this manual is derived from proprietary data from Labthink. This
manual has been prepared expressly for the purpose of assisting operation and maintenance
personnel in their use and general maintenance of MFY-01 Leak Tester.

Publication of this information does not imply any rights to reproduce or use it for purposes
other than installing, operating, or maintaining the instrument described herein.

These provisions are intended to state all of the rights and responsibilities between
Labthink and the customer. They take place of and supersede all warranties, expressed
or implied, and whether of merchantability, fitness or otherwise. The remedies contained
in this instruction manual are exclusive.

2
Leak Tes te r

MFY-01

Contents
Preface ................................................................................................................... 4
1. Introduction ........................................................................................................ 5
Standards ................................................................................................................ 5
Specifications .......................................................................................................... 5
Principles ................................................................................................................ 5
Unpacking ............................................................................................................... 6
2. Instrument .......................................................................................................... 7
Instrument Appearance ........................................................................................... 7
Control Panel .......................................................................................................... 8
Installation ............................................................................................................... 9
Instrument System .................................................................................................. 9
3. Operation .......................................................................................................... 10
Procedures............................................................................................................ 10
An Example........................................................................................................... 10
4. Maintenance ..................................................................................................... 13
General Maintenance ............................................................................................ 13
Technical Support ................................................................................................. 13
Appendix: Structure of the Vacuum Chamber .................................................. 14
Customer Service Regulation ............................................................................ 15

3
Leak Tes te r

MFY-01

Preface
About This Guide Welcome to the Labthink PARAM® MFY-01 Leak Tester system. This
manual contains a description of the operation and maintenance of
your Leak Tester.

Safety Precautions Observe the following safety precautions when you operate or perform
service on the MFY-01 Leak Tester:

Please see the instrument nameplate for power supply and it should be
well grounded. This instrument SHOULD NOT be operated with other
high-power electrical appliances.
This instrument SHOULD NOT be operated in intensely corrosive,
combustible, explosive, or similar environment, or under intensive
vibration or strong electromagnetic interference.

Disassembly of the instrument or modification of the system without


authorization is strictly PROHIBITED.

Please perform operation only according to this instruction manual.


Users who are unable to comply with this manual take full responsibility
for any accidents and damages.

Operating personnel should be present during the test process.

Check the filter at the gas inlet side. Remove the filtered water if any.

Clean the vacuum chamber and the filter to avoid plugging

Replace water in the vacuum chamber to keep it clear.

The compressed gas should be clean and without any visible liquid
drop or solid substances which may cause damage to the instrument.

DO NOT wipe up the vacuum chamber with organic solvent.

4
Leak Tes te r

MFY-01

1. Introduction
MFY-01 Leak Tester is professionally designed for the leakage tests of packages for food, drugs,
medical instruments, daily chemical products, cars, electronic components, stationeries and
other industrial products. The instrument also can be used to test seal performance of
specimens after falling and compression tests.

Standards
This test instrument conforms to the standards: GB/T 15171, ASTM D3078

Specifications
Items Specifications
Vacuum Degree 0 ~ -90 KPa
Accuracy 1% FS
Φ270 mm x 210 mm (H) (standard)
Vacuum Chamber Effective Φ360 mm x 585 mm (H) (optional)
Sizes Φ460 mm x 330 mm (H) (optional)
Note: customization is available for other sizes
Air Source Pressure 0.7 MPa (outside of supply scope)
Port Size Ф6 mm PU Tubing
Instrument Dimension 300 mm (L) x 380 mm (W) x 450 mm (H)
Power Supply See instrument nameplate
Net Weight 12 kg

Principles
Firstly submerge the specimen in the water within the vacuum chamber, and then evacuate the
vacuum chamber to form differential pressure between the inside and outside of specimen. The
leakage property could be obtained by observing the steady progression of bubbles from the
specimen and how the specimen expands and restores to its original shape after vacuum
release.

5
Leak Tes te r

MFY-01

Unpacking

Remove the instrument from the shipping carton and inspect for any signs of damage. Contact
the delivery agent immediately if any damage observed. Leave the instrument in the shipping
carton for the delivery agent to examine the damage.

If there is no visible damage, check the contents of your shipment against the packing slip.
Notify Labthink immediately if there are any discrepancies.

The instrument is shipped in a special designed box to prevent any damage. If


re-shipment is necessary, the instrument should be repacked in the original box. Please
save the box properly.

6
Leak Tes te r

MFY-01

2. Instrument

Instrument Appearance

Figure 1 Instrument overview

7
Leak Tes te r

MFY-01

Control Panel

Figure 2 Control Panel

Function keys

“SETUP” At standby mode, press this key to maintain the current vacuum degree
“STOP” 1. Confirm and save the maintaining time of a specific vacuum degree at
preset mode; 2. Lock the current test status; 3. Press this key to enter
standby mode when “REVERSE” key is pressed or “reverse” status is
activated after the test.
“REVERSE” Press this key to manually terminate the test when “STOP” is pressed
“TEST” Start a test
“+/-” Increase/decrease the vacuum maintaining time (the figure shown on the
left side of the LCD in min)

Note:
1. “SETUP” is only valid at standby mode;
2. “REVERSE” key is only valid when “STOP” key is pressed;
3. “TEST” key is only valid when “STOP” key is pressed to exit parameter setup or to cease
“reverse” mode.

8
Leak Tes te r

MFY-01

Installation
The system should be placed with enough space to allow installation and operation.

The instrument should not be operated under strong electromagnetic interferences.

The instrument should not be operated with other high-power electrical appliances

Please see the instrument nameplate for power supply and it should be well grounded.

Operating personnel should perform the test according to this instruction manual.

Please the instrument on a steady table with at least 0.5 m surrounding to allow installation.

Remove the 2 pcs ofФ6 PU tubing (100mm) installed at the connector of pressure regulator.

Connect the power cord to the instrument and plug in the power supply.

Instrument System
Note: all function keys should not be pressed for more than 1 second; otherwise, system may
fail. If it happens, please contact Labthink for technical support.

1. At evacuating status, LED displays a constant number, generally, 00.

2. Press “ ” key and LED displays “P_1” while the number “XXXX” blinks. Employ

“ / “ key to set up the lower limit of vacuum degree, which should be 1 kPa lower than
desired value.

3. Press “ ” key and LED displays “P_2” while the number “XXXX” blinks. Employ

““ / “ key to set up the desired value of vacuum degree.

4. Press “ ” key and LED displays the pressure inside vacuum chamber and the vacuum
switch is open again.

9
Leak Tes te r

MFY-01

3. Operation

Procedures
1. Hook up the gas supply.
2. Turn on the instrument.
3. Set up test parameters, including vacuum degree and maintaining time at a specific vacuum
degree. Press “STOP” key to confirm.
4. Open the hood of vacuum chamber and load test specimen.
5. Place the hood back and seal the chamber.
6. Press “TEST” key to initiate test. The system automatically evacuates the chamber until it
reaches the preset vacuum degree, and this status is maintained for the preset time period.
7. When test is done, system automatically “reverses” the chamber and injects air to the
atmospheric pressure. If the test is terminated by “REVERSE” key, press “STOP” key to
manually “reverse” the chamber.
8. Open the hood and unload the test specimen.
9. Repeat the above procedures to test the next specimen if necessary.
10. When all finished, turn off the instrument and power supply.

An Example
Test requirement: vacuum degree— -60 kPa; maintaining time—10 min.

 Load test package

Add a specific amount of water to the vacuum chamber where the water level is about 10
mm above the porous plate (see Appendix). Place the test specimen into the chamber and
seal the chamber with the hood.

 Set up parameters

Vacuum degree:

Press “ ” key P_1 XXX.XX

10
Leak Tes te r

MFY-01

Employ “ / “ key to set P_1 to -59

Press “ ” key P_2 XXX.XX

Employ “ / “ key to set P_2 to -60

Press “ ” key to display “00”, which does not blink.

Maintaining time:
Press “SETUP” key at standby mode

The number on the left of “-- --“ blinks in LED

Employ ““+/-“ key to set to 10.0

Press “STOP” and LED displays 10.0——0600

 Start the test

Press “TEST” key

The system is being vacuumed

11
Leak Tes te r

MFY-01

Vacuum degree changes from 00 to -60

The vacuum valve is closed and vacuum degree is maintained

The number “600” (s) on the right side of LED decreases

When “0” (s) is reached, system enters “reverse” mode

Vacuum chamber returns to the atmospheric pressure

Press “STOP” key to cease “reversing” and enters standby mode

Remove the test specimen and load a new one

Repeat the above procedures to perform a new test.

Note: if test specimen fails before the desired vacuum degree or the preset maintaining
time is reached, users need to press “STOP” key to terminate test and “REVERSE” key
to re-inject the vacuum chamber. Once “00” is reached, press “STOP” key again to enter
standby mode

12
Leak Tes te r

MFY-01

4. Maintenance

General Maintenance
1. Read the instruction manual thoroughly before any operation.
2. Please see the instrument nameplate for power supply and it should be well grounded. .
3. Do not disassemble the instrument without authorization.
4. Power off the instrument and cover with a sheet after test.
5. Perform calibration every once a year.

Note:
1. Clean the vacuum chamber and the filter to avoid plugging
2. Replace water in the vacuum chamber to keep it clear.
3. DO NOT wipe up the vacuum chamber with organic solvent.

Technical Support

Web support
Please visit our website for more details of general problems and technical supports at
http://www.labthinkinternational.com.cn

Contact us
You can also contact us by telephone or e-mail:
Tel: +86 531 85063398
E-mail: afterservice@labthink.com

NOTICE:

Before you call us for help, please make sure:


Read the Instruction Manual thoroughly and check all possible problems;
Place instrument and Instruction Manual at hand;
Prepare the following information: model, purchase date and manufacturing code.

13
Leak Tes te r

MFY-01

Appendix: Structure of the Vacuum Chamber

14
Leak Tes te r

MFY-01

Customer Service Regulation

 General Provisions
※Purpose: This regulation is instituted to satisfy the need of customers with professional and
timely service.
※Content: This regulation includes items of general provisions, service instructions and
response time.
※ Service Aim: Provide customers with professional, timely and continuous service. Try our
best to satisfy the requirements of customers.
※Service Principle: Integrity and sense of responsibility, sophisticated skill and customer
oriented.
※Performance Pledge:
1. Notice of lack of configuration should be provided in written form within thirty days from
AWB date. Otherwise, it is considered as a conform information. Responses will be made within
one workday after receiving written notification.
2. Products are warranted with free service and replace components of the same brand, type
and specifications for a period of one year from the AWB data against defects of quality.
3. Replaced parts are warranted with free service of replacement or maintenance for a period
of 90 days from the date of purchase against defects of quality.
4. Technical whole-life service will be provided outside warranty period. While cost of
maintenance is free, cost of material as well as freight out and home is charged.
5. Repaired instruments are warranted with free service of replacement or maintenance for a
period of 180 days from the return date of the repaired instrument against defects of quality.
6. Responses will be made within one hour after receiving service application. Disposing
suggestions and solution program will be provided within one working day.
※Contact information (8:00-17:00 Monday to Friday Peking time):
 Tel: +86-531-85068566
 Service Tel: +86-531-58702798
 Fax: +86-531-85062108
 E-mail: trade.en@labthink.com
 Http://www.labthinkinternational.com.cn
 Zip Code: 250031
 Address:No.144 Wuyingshan Road, Jinan, China

15
Leak Tes te r

MFY-01
※ Exclusion:
For products not sold by LABTHINK directly, the sales regulation abides by the service
agreement and quality guarantee regulations between distributors and customers.
Division of authority and responsibility as well as the quality guarantee regulations
between corresponding distributor and LABTHINK follow the content of contract (oral
agreement) between these two sides.

 Service instructions
※Instructions for Initial Use
1. Open-package Inspection: Check the required configuration of product in contract with that
of packing list. Make sure that all the parts mentioned in the packing list are included. (When
taking out the products from its package, pay attention to the caution sign such as “Do not hold
here”.) Be sure all seals on the instrument are intact. Otherwise please contact LABTHINK as
soon as possible.
2. Product Commissioning: Please review the instruction manual and other materials
supplied with the instruments carefully. Test the instruments under specified environment,
parameters and operating mode. If there is any problem about our products, please don't
hesitate to clarify it with LABTHINK before commissioning. (Software of instrument computer
has been installed before delivery. Users can use the computer directly. Discs supplied with
instrument should be kept carefully for later use)
※Confirmation of Warranty Period
Warranty period is supported from AWB date (Buyer and seller as well as the product mode are
given clear indication). Without such supporting document, the warranty start date will be
Ex-factory Date of the purchased products recorded by LABTHINK. In any of the following
cases, there will be no maintenance free of charge.
1. Products are not purchased from LABTHINK or LABTHINK’s agents and distributors are not
covered in the service regulation.
2. Defects of instrument or other components caused by the use of accessories or peripheral
equipments not produced or authorized by LABTHINK.
3. Defects or damages of instruments or other components resulted from abnormal reasons
(including unsatisfied power supply, working environment, not the first delivery, long term
storage, outside force and foreign body entering instrument).
4. Defects or damages of instruments or other components caused by natural influences (such
as earthquake or fire) or accidents (such as stealing and lost)

16
Leak Tes te r

MFY-01
5. Defects or damages of instruments or other components caused by dismantling, repair, and
refit without LABTHINK permission.
6. Defects or damages of instruments or other components caused by plugging and
connecting power cord , typing cord, communication cord and other connecting wire without
cutting the power supply.
7. According to the international convention, correlation insurance company is directly
responsible for charges of maintenance or complete damage resulting from serious damage
during international transportation. Customers must inform LABTHINK within one working day
after receiving the product. (For uninsured cargo, it is recommended that customers take out
insurance)
※Expiration of Warranty Period
Warranty period is expired in one year from the AWB date (indicating the transaction sides and
instrument type). Without above supporting document, the warranty start date will be the
Ex-factory Date of the purchased instrument recorded by LABTHINK.
※ Charge of Service
1. Training
a. LABTHINK provides customers with demonstration video of the sold products free of
charge (attached with instrument or transmitted through Internet)
b. On the premise that customers need and are willing to bear correlative expenses,
LABTHINK engineer may arrive at the nominated places for product installation, debugging,
training and commission. While maintenance service is free of charge, Users must undertake
relating travel expense of the engineer (including round-trip air ticket, ticket, accommodation
and other travel expense). With the signed (or reached) service contract (oral agreement)
between two sides and prepaid relating expenditures, LABTHINK will send its engineer for
contract service.
2. Maintenance
a. For problem caused by product quality within warranty period, LABTHINK will provide
product maintenance, technical support, freight out and home as well as cost of replaced
components free of charge.
b. Problem not caused by product quality within warranty period or maintenance outside
warranty period is only charged for replaced components. However, freights home and out is at
the customer’ side (only the replaced component is charged if transportation of instrument is not
needed).
c. Door-to-door service outside warranty period requires signed (or reached) service contract
(oral agreement) between two sides. Users must undertake relating travel expense of the

17
Leak Tes te r

MFY-01
engineer (including round-trip air ticket, ticket, accommodation and other travel expense). With
the signed (or reached) service contract (oral agreement) between two sides and prepaid
relating expenditures, LABTHINK will send its engineer for contract service.

 Service Content
1. Please contact LABTHINK as soon as possible if there is problem during the process of
operation. To help service man providing accurate and timely solution for your problem or
trouble, be sure to make the following preparation before contact:
(1)Check product type on product certification (or in instruction manual and manufacturer
certificate);
(2)Having reviewed instruction manual carefully and having it at hand at the time of contact
(3)Make sure the protective tube and all the connecting wire are in good condition when there
is malfunction.
(4)Collect information about symptoms of problem (including displaying content and operation
condition), problem time and other detailed information (customers who conceal actual
information should be responsible for the lost)
2. LABTHINK will make telephone call or door-to-door return visit about using condition of
product periodically.

 Response Time
1. LABTHINK’s personnel will response to help users solving the problem within one hour after
receiving telephone call or written notice. It will give customers technical support through
remote technologies such as telephone call, E-mail, and demonstration video. If component
replacement is needed, the required component will be sent out within one working day (if there
is charged fees, the required component will be sent out after receiving fax of transferred
account).
2. If instruments need to be sent back for maintenance, LABTHINK will notify the user within
one working day after receiving the product about the inspection information. Instruments will
be repaired within two working days and send out within one working day after repairing (if there
is charged fees, the product will be sent out after receiving fax of transferred account).
3. If technical support cannot be realized through remote technology while it is inconvenient to
send instruments back, LABTHINK can offer field service. Charges and regulations are as
above stated. Unless there are special circumstances, the response time is:

18
Leak Tes te r

MFY-01
 Notify customers the personal information of LABTHINK engineer within one workday after
receiving customer notice and reaching agreement.
 On the premise that customers provide corresponding invitation and other necessary help,
LABTHINK will get the engineer’s visa as soon as possible.
 LABTHINK engineer reach the conferred place within one week of the visa date.

 Others
 This regulation is the only open service regulations for customers. Herein other ones with
special purpose or beyond the above items are refused.
 The right to explain this regulation is reserved by LABTHINK INSTRUMENTS CO., LTD.
 The right to revise this regulation is reserved by LABTHINK INSTRUMENTS CO., LTD.
 Distributors (agents) should refer to cooperation agreement additionally.

19
Leak Tes te r

MFY-01

LABTHINK INSTRUMENTS CO., LTD.


144 Wuyingshan Road, Jinan 250031 China
Website: http://www.labthikinternational.com.cn
Tel: +86-531-85068566
Service Tel: +86-531-58702798
Fax: +86-531-85062108

20