Professional Documents
Culture Documents
142 Blood Leak Alarm
142 Blood Leak Alarm
Blood Leak
Blood Leak Detection
A blood leak during a treatment on the AK200 is indicated by blood leak constant light
and constant alarm
Note: Flashing blood leak button indicates blood is no longer detected: press blood leak
button to confirm
Recommended Procedure
o Macroscopic : Visible
o Microscopic : Positive haematest from red dialysate line sample port.
o (Negative haematest retest; 2nd negative - see below)
Do not carry out a heat disinfection cycle after blood leak (it will caramelise the blood)
Possible causes of a negative haematest result in the presence of a blood leak alarm
include
o Expired haematest agent may show negative in the event of a true microscopic
blood leak. Small particles in blood leaking housing (avoided with regular
maintenance procedures. Should be referred to a technician)
o Failure of blood leak detector system during treatment. The blood leak detector
system is routinely and automatically tested by AK 200 as part of the machine
function check. In addition, the system is checked every 5 minutes to ensure that
it is able to detect non-transparency. If not, the system will act as a if a blood leak
alarm has occurred and enter a patient safe condition. (Should be referred to a
technician)