Professional Documents
Culture Documents
01 HMASeries
01 HMASeries
HMA Series
● Targets
- Monitoring wastewater from factories
- Monitoring environmental water quality
● Major heavy metals
- High priority : Pb, Hg, Cd, Cr/Cr6+
- Other parameters : Ni, Cu, Zn, Mn
4
What type of analyzers are required?
● Surface water
Multi Single · Measure river and lake water
measurement measurement
· Simultaneous and multiple parameters
· Low concentration measurement
Factory Factory · High accuracy is required
· High price range
Centra HM HM
WWTP ● Industrial wastewater
lized
Factory in
WWT
factory
· Measure industrial wastewater
P
· Parameters depending on type of facility
Factory
· Single parameter
· Wide measurement range
· Low price range
Targeted Factory and
Measurement Parameter
Cr6+
Cu Ni Mn
TCr
Electroplating ○ ○ ○
Mining and smelting ○ ○ ○
Iron and steel ○
Semiconductor ○
Battery ○
Tan and fur machining ○
What is Measurement principle?
Colorimetric
-HMA Series-
M
+
= Total
coloring reagent
Features
● High Reliability
● Stable Measurement
Available for Selling
in Your Country
I. Main Specifications
II. Configuration
III. Is Measurement “Total”?
Appendix 1
Main Specifications
-HMA Series-
Industrial wastewater
Measurement (Final effluent water)
600 105
object
Cr6+, Total Cr, Cu, Ni, and Mn
Measurement (single-parameter )
parameter
Measurement Colorimetric method
principle Cr6+ 0.01 - 2 mg/L, 0.04-5mg/L
900
Configurations
-HMA Series-
Controller
(Touch panel, DC4-20mA, RS485, etc.)
Digestion unit
Colorimeter
Sample
water
Reaction
Sampling cell Drain tank
pump
Reagent tank
Appendix 3
Is Measurement “Total”?
-HMA Series-
M O M A-
+
H Competitors products measures
hydroxides salts
only dissolved metal ions
M
+
ion
Measured data without digestion
M O M
+ may have some difference to the
oxides complexes lab data
M
+
≠ Total