Professional Documents
Culture Documents
GammaMedplus Ix User Guide
GammaMedplus Ix User Guide
GammaMedplus Ix User Guide
GammaMedplus iX ™
User Guide
™
including the GammaMedplus 3/24 iX
™
and MammoSource iX
P/N GM11020140
Rev 1.1 – March 2009
GM10 UG Book.book Page ii Tuesday, March 24, 2009 11:22 AM
Abstract The GammaMedplus iX User Guide (P/N GM11020140) provides GammaMedplus iX system users with
basic instructions for using the afterloader control software with the GammaMedplus iX and
GammaMedplus 3/24 iX high dose rate remote afterloaders.
Help Desk Support If you cannot find information in this user guide, you may contact us at one of the toll free numbers
available in your area:
Australia Belgium Canada Denmark Finland France
1 800 144 130 0800 74 248 1 888 226 8633 80 88 07 45 0800 11 68 92 0800 905 397
Indonesia
Germany Holland Hong Kong Hungary Ireland
001 803 0441
0800 182 6937 0800 022 5072 0800 96 21 42 06 80013 318 1 800 551 716
1109
Taiwan UK USA
0080 04 40 99 0800 068 06 88 1800 360 7909
For countries not listed above, Technical Support may be obtained via the following numbers:
UK: USA:
Phone: +44 1293-601-327 Phone: +1-434-979-1540
Fax: +44-1293-601-397 Fax: +1-434-979-1481
eFax: +1-866-385-1322
Technical Support E-mail: brachyhelp@varian.com
Notice Information in this user guide is subject to change without notice and does not represent a commitment on
the part of Varian Medical Systems, Inc. Varian Medical Systems is not liable for errors contained in this
user guide or for incidental or consequential damages in connection with furnishing or use of this material.
This document contains proprietary information that is protected by copyright and is the sole property of
Varian Medical Systems. No part of this document may be reproduced, translated, or transmitted without
the express written permission of Varian Medical Systems, Inc.
FDA 21 CFR 820 Varian Medical Systems products are designed and manufactured in accordance with the requirements
specified within these federal regulations.
Quality System
Regulations (CGMPs)
ISO 13485:2003 Varian Medical Systems products are designed and manufactured in accordance with the requirements
specified within the ISO 13485:2003 quality standard.
CE 0086 Varian Medical Systems products meet the requirements of Council Directive MDD 93/42/EEC. European
representative:
Varian Medical Systems UK Ltd. +44-1293-601-200 Phone
Gatwick Road, Crawley +44-1293-891-386 Fax
West Sussex, RH10 9RG
United Kingdom
Legal Varian Medical Systems, Inc. +1-888-666-7847 Phone
3100 Hansen Way +1-434-979-1481 Fax
Manufacturer Palo Alto, CA 94304 USA
ii
GM10 UG Book.book Page iii Tuesday, March 24, 2009 11:22 AM
CHAPTER SUMMARY
Introduction 1
System Configuration 4
Emergency Procedures 5
iii
GM10 UG Book.book Page iv Tuesday, March 24, 2009 11:22 AM
Unauthorized Software
The GammaMedplus iX system and its associated hardware and software
constitute medical devices. The software installed on the control PC by Varian
Medical Systems has been through rigorous testing in a controlled system
configuration. The use and/or addition of other software to this computer
violates the approved system configuration and risks potential system
malfunction. Although there are safeguards in the system to minimize the risk
of hazardous operation, the system may still shut down should it detect
corrupted data files or improper computer operation, preventing the completion
of patient treatment. The control PC must operate in “real time” to properly
administer a patient treatment. Any third party software or computer viruses
introduced by non-approved software may cause system instability and a
subsequent system shut down requiring a billable service call to restart.
Unauthorized Applicators
All Varian approved applicators have been tested against and passed strict safety
criteria. Under no circumstances should any applicator or catheter that has not
been approved by Varian be used in conjunction with the GammaMedplus iX
system. Varian cannot be responsible for the operation of the system if this
warning is ignored. Any service calls required as a result of such misuse will be
considered additional to Varian’s standard warranty or service contracts.
Unauthorized Sources
All Varian approved source wires have been tested against and passed strict
safety criteria. Under no circumstances should any source wire that has not been
approved by Varian be used in conjunction with the GammaMedplus iX system.
Varian cannot be responsible for the operation of the system if this warning is
ignored. Any service calls required as a result of such misuse will be considered
additional to Varian’s standard warranty or service contracts.
Extreme care should therefore be taken to ensure the wire path is not constrained
by less than 1.3 cm.
iv
GM10 UG Book.book Page v Tuesday, March 24, 2009 11:22 AM
v
GM10 UG Book.book Page vi Tuesday, March 24, 2009 11:22 AM
Manufacturer location
Date of manufacture
Serial number
SN
vi
GM10 UG Book.book Page vii Tuesday, March 24, 2009 11:22 AM
Alarm condition
Note in text
vii
GM10 UG Book.book Page viii Tuesday, March 24, 2009 11:22 AM
viii
GM10 UG Book.book Page ix Tuesday, March 24, 2009 11:22 AM
Contents
ix
GM10 UG Book.book Page x Tuesday, March 24, 2009 11:22 AM
xi
GM10 UG Book.book Page xii Tuesday, March 24, 2009 11:22 AM
xiii
GM10 UG Book.book Page xiv Tuesday, March 24, 2009 11:22 AM
Chapter 1 Introduction
WARNING 1-1: The afterloader control software enables you to import, edit, and
deliver a predetermined treatment prescription. The software
does NOT perform any dosimetric calculations related to the
determination of the proper treatment plan except for computing
source decay. The actual treatment plan is usually created using
a brachytherapy treatment planning system, which, with the
exception of its file interface to the control software, is beyond
the scope of this document. Be sure to check the integrity and
clinical accuracy of any data that you import or enter before
delivering a treatment. Failure to do so may result in incorrect
treatment and may cause serious injury to the patient.
In This Chapter
Topic Page
Glossary of Regulatory Terms 1-2
Audience 1-3
Important Safety Information 1-3
GammaMedplus iX System Components 1-11
Starting the GammaMedplus iX System 1-30
Overview of the Afterloader Control Software 1-31
1-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Audience
The GammaMedplus iX User Guide is intended for radiation oncologists,
radiation therapists, and medical physicists who use the GammaMedplus iX 1
system to deliver patient treatments at a site or clinic.
Operation
The GammaMedplus iX should be operated only by those persons who have
training in HDR radiation safety and who have been instructed in the use of this
device.
Introduction 1-3
GM10 UG Book.book Page 4 Tuesday, March 24, 2009 11:22 AM
Safety Notices
The GammaMedplus iX system is a licensed medical device used for
delivering ionizing radiation to a patient. In order to prevent unintended
exposure to radiation, the safety notices and directives listed below shall be
followed:
• The emergency container shall be located near the device in the treatment
room along with the wire cutter and pair of long handled forceps.
• The GammaMedplus iX is certified as a Type A container and may be
transported with an active source installed. In such case (even when 1
moving the device between two treatment rooms), the source shall be
secured in the device by inserting the source locking wire into
Channel 1 and switching the battery supply off by turning the battery key
switch to position 0 (OFF). Refer to Appendix E “Transporting the
GammaMedplus iX” for more information.
Introduction 1-5
GM10 UG Book.book Page 6 Tuesday, March 24, 2009 11:22 AM
Potential Hazards
Electrical Shock
Personal Injury
Radiation Concerns
The HDR sources are highly radioactive. As such, radiation safety is a high
priority when working with and around the GammaMedplus iX device. The
device is subject to USNRC, Agreement State, and other Competent Authority
regulations for the handling of radioactive material. Personnel shall follow all
safety instructions exactly for safety and regulatory reasons.
Introduction 1-7
GM10 UG Book.book Page 8 Tuesday, March 24, 2009 11:22 AM
• Source Exchange
Only service personnel authorized by Varian Medical Systems shall
perform the source exchange. The sources will be delivered to the clinic in
approved shipping containers.
• Source Strength
The maximum source strength that can be installed or used is defined in
your site specific license.
• Source Extension
GammaMedplus iX users shall only extend the active wire during patient
treatments and during safety and quality checks of the system.
• Transportation
When shipping the sources in the approved container, the appropriate
national or international regulations for the safe transport of radioactive
material shall be followed. The device and, if applicable, its packaging
shall be labeled according to these regulations.
• Emergency Situations
Personnel shall be familiar with, and trained in, all HDR safety procedures.
Personnel shall plan for and practice the emergency response procedures
developed at your site to assure an immediate and quick response in case
of an emergency. Refer to Chapter 5 “Emergency Procedures” for more
information.
• Safety Interlocks
The proper functioning of all GammaMedplus iX safety interlocks and
interlock checks shall be verified before patient preparation and patient
treatment begin.
WARNING 1-4: Any attempt to disable the safety interlock system could result
in injury. Modification of this safety system is prohibited by
regulations.
Radioactive Contamination
The Ir-192 source assembly is designed and built as special form radioactive
material. As such, it should never leak radioactive material. However, should
there be any suspicion that the source is damaged or leaking radioactive 1
material, the Radiation Safety Officer (RSO) at your site and a Varian service
representative must be informed immediately and contamination checks shall
be performed.
Fire
If a fire occurs during patient treatment, retract the source into the shielded
position immediately. Once the radiological and physical safety of the patient
has been established, proceed with normal fire fighting procedures.
Explosion
WARNING 1-5: Do not use the device in the presence of anesthetics or other
flammable gases, or in an oxygen-enhanced atmosphere, as
there is a risk of explosion.
Introduction 1-9
GM10 UG Book.book Page 10 Tuesday, March 24, 2009 11:22 AM
WARNING 1-6: Under rare circumstances, such as a defect in the circuit that
charges the battery, the battery may produce explosive gas.
For this reason, keep the device away from sparks and flames.
Adequate ventilation must be provided.
Operating Environment
The GammaMedplus iX is designed to be operated and stored in an
environmentally controlled area such as a hospital or laboratory. The
GammaMedplus iX should not be exposed to extreme moisture, high or low
temperatures, or to abnormal amounts of dust and debris. The recommended
environmental conditions are as follows:
Introduction 1-11
GM10 UG Book.book Page 12 Tuesday, March 24, 2009 11:22 AM
GammaMedplus iX Afterloader
The GammaMedplus iX afterloader is equipped with an Iridium-192 (Ir-192)
source. The source can be moved into 24 channels in succession via the
indexer head. In each of these channels, the source may be programmed to stop
in up to 60 dwell positions. The applicator in the patient is connected to the
afterloader via one or more source guide tubes.
The wires used for the dummy and active source are moved by electrical
motors coupled to drive wheels that are in contact with the wire. This drive
system allows for precise wire movement with a minimum of components. All
normal wire movements, both in and out, are performed by the motors. The
emergency retract system is only used upon failure of the main drive system.
Indexer head
Chassis
Trailing wheels
Swivel wheels
The GammaMedplus iX rests on four wheels that you can lock in place by
means of the central foot brake lever. When the foot brake is released, you can
easily move the device with the help of the large lateral handles.
Afterloader keypad
Handcrank
Foot brake
Introduction 1-13
GM10 UG Book.book Page 14 Tuesday, March 24, 2009 11:22 AM
The battery key switch is located at the bottom of the afterloader near the foot
brake. Usually the key switch must be in the I (ON) position. If you are
transporting the device, you must switch the key to the 0 (OFF) position. Refer
to Appendix E “Transporting the GammaMedplus iX” for more information.
Afterloader Keypad
The afterloader keypad is located on the back of the GammaMedplus iX. The
keypad consists of the following areas:
1
• SOURCE section and key switch
• DUMMY section
• HEAD section
• RADIATION section
• EMERGENCY button
DUMMY section HEAD section
DUMMY wire control Height adjustment of device housing
OUT = extension UP = raise
IN = retraction DOWN = lower
SOURCE section
RELEASED
Source ready
RADIATION section
for extension
RADIATION light
Indicates presence
LOCKED
of radiation
Source locked
Key switch used
to release or
lock source
EMERGENCY button
POWER Provides rapid
AC power retraction of source
supply okay in case of
emergency
CHARGE
Storage battery is
being charged
The SOURCE section contains the key switch used to lock the Ir-192 source
or to release it for extension during treatment and testing. The LOCKED
indicator light flashes when the key is in the vertical position. The
RELEASED indicator light turns on when the key is in the horizontal position.
Introduction 1-15
GM10 UG Book.book Page 16 Tuesday, March 24, 2009 11:22 AM
Note: For safety reasons, the key switch should remain in the
LOCKED position with the key removed and in a secure
location unless you are performing a treatment or test.
The POWER light indicates that the afterloader is connected to AC power, and
the CHARGE light indicates that the storage battery is being charged.
Note: You should not normally disconnect the power to the system
unless there is a specific need to do so. If the device has been
unplugged for a long time, the storage battery may require
recharging before use.
DUMMY Section
In conjunction with the Channel Editor in the control software, you can use the
DUMMY section of the keypad to change the distal position of the source.
Refer to the “Change the Distal Position of the Source” section in Chapter 2
“Patient and Treatment Data” for more information.
HEAD Section
The HEAD section contains the UP and DOWN buttons that control the
motorized height adjustment of the afterloader. The adjustment range is
between 90 and 130 cm (35 and 51 inches) from the floor to the center of the
indexer head.
CAUTION: When transporting the device, lower the source head fully in
order to provide maximum stability.
RADIATION Section
EMERGENCY Button
The access for the dummy wire is located on the right side below the
afterloader keypad. The dummy wire can only be replaced by Varian
authorized service personnel.
Introduction 1-17
GM10 UG Book.book Page 18 Tuesday, March 24, 2009 11:22 AM
Service opening
for dummy wire
Control Console
The Control Console is the main control panel for the GammaMedplus iX
afterloader. This device should be located next to the control PC. You use the
Control Console to enable power to the other GammaMedplus iX devices, to
activate special functions mode, and to monitor the status of the source,
treatment, and alarm circuits. A red light on the Control Console may indicate
an alarm situation requiring immediate intervention, yellow indicates the
presence of radiation, and green indicates correct functioning.
The POWER, MODE, and SOURCE sections on the left side of the Control
Console are labeled in the graphic below and then described.
NORMAL
All treatment functions able to
MODE Key Switch operate
SPECIAL
Service mode with special functions
MAINS
POWER Key Switch Main AC power supply okay
BATTERY
Function of charged battery okay
Introduction 1-19
GM10 UG Book.book Page 20 Tuesday, March 24, 2009 11:22 AM
When the POWER key is switched on, an automatic test of the Control
Console indicators is performed. Refer to the “Starting the GammaMedplus iX
System” section below for more information.
The MODE section contains a second, unique key switch that is used by
Varian authorized service personnel for test and service purposes and to
activate special functions, such as changing the distal offset of the source via
the IN and OUT buttons on the DUMMY section of the afterloader keypad.
With the key in the horizontal position, the SPECIAL indicator light turns on.
For all other functions of the GammaMedplus iX system, the key switch must
be in the vertical position, with the NORMAL indicator lit.
WARNING 1-10: For safety reasons, insert the MODE key only when
necessary and guard it against unauthorized use.
SOURCE Section
The SOURCE section indicates the status of the source and serves as a safety
indicator. The yellow RADIATION indicator turns on if radiation is detected
in the treatment room. The yellow OUT indicator light flashes as the active
source is being extended and illuminates permanently when the source has
reached a dwell position. When the source moves again, the OUT indicator
flashes until it reaches another dwell position or returns to the shielded
position. The SAFE indicator light turns on when the source is retracted fully
into its shielded position. The LOCKED indicator light flashes when the
source is locked at the key switch on the afterloader keypad.
WARNING 1-11: If one of the two yellow indicators does not light up during a
treatment, there is a possibility of a system malfunction, and
an error message is displayed on the control PC. The
treatment must be interrupted immediately. Check for the
cause of the error and correct the problem. Failure to do so 1
may result in injury to the patient.
The TREATMENT and ALARM sections on the right side of the Control
Console are labeled in the graphic below and then described.
INDICATOR LIGHTS
STARTED
Flashing light: before pressing the
START button START button
Starts treatment
Steady light: treatment in progress
COMPLETED
INTERRUPT button Treatment completed
Interrupts treatment INTERRUPTED
Treatment interrupted
ALARM signal
Flashing alarm indicates
malfunction and possible DOOR
radiation hazard Steady light if door of the treatment
room is opened during treatment
ALARM RESET button DOSE
Resets the audible alarm signal Steady light if the external dose
measurement unit has exceeded a
BUZZER OFF button warning threshold
Turns off audible warning signal BUZZER OFF
Indicates audible warning signal
turned off
TREATMENT Section
Introduction 1-21
GM10 UG Book.book Page 22 Tuesday, March 24, 2009 11:22 AM
By pressing the START button, the audible signal ceases and the STARTED
indicator changes to a steady light, showing that treatment is in progress. At the
end of the treatment, the COMPLETED indicator light turns on and an
intermittent signal is audible. You can turn off the audible signal by pressing
the BUZZER OFF button.
You can interrupt the treatment by pressing the INTERRUPT button, which
causes the source to be retracted at normal speed. The INTERRUPTED
indicator light flashes and an audible signal can be heard.
After other interruptions (e.g. when the door of the treatment room has been
opened) the INTERRUPTED indicator light flashes, accompanied by the
audible signal (and, in this case, the DOOR light).
You can usually continue an interrupted treatment by pressing the START
button on the Control Console. Refer to Chapter 5, “Emergency Procedures”
for detailed information on interrupting a treatment during an emergency.
ALARM Section
Urgent alarms are indicated in the ALARM section. If an error in the treatment
sequence occurs or if there is a malfunction of the GammaMedplus iX (e.g. if
the source is not automatically retracted), the red alarm symbol flashes and a
simultaneous audible alarm signal sounds.
You can turn off the audible signal by pressing the BUZZER OFF button. In
this case, the BUZZER OFF indicator light turns on. If the error condition
remains uncleared, the audible signal will sound again in 5 minutes.
After clearing the cause of the alarm, you can reset the visible and audible
alarm signals by pressing the ALARM RESET button. If no alarm reset is
executed, but the cause of the alarm is cleared, the alarm will be reset when you
press the START button.
Note: If the cause of the alarm is not cleared, the alarm signal is
repeated after pressing the START button.
The red DOSE indicator light turns on if an external dose measurement unit
has exceeded a set warning threshold. The GammaMedplus iX retracts the 1
source immediately into the shielded position. Before continuing the treatment,
the external device must be reset to start up with initial values and to set the
dose input of the GammaMedplus iX to a ready state.
The red DOOR indicator light turns on if the door of the treatment room is
opened at any time. If the door is opened during a treatment, the treatment is
interrupted automatically and the source returns to its shielded position. Verify
that no persons other than the patient are in the room. After closing the door,
the treatment can be continued by pressing the START button.
The extended sound functionality provides audible support for determining the
status of a treatment. The sound system uses three distinct tones for identifying
the different states of the afterloader:
• Standard tone — a flat, steady tone can be heard at each movement of the
source drive and when each user input is required
• Attention tone — a rising tone can be heard if any normal interruption in
treatment occurs and when a treatment is completed normally
Introduction 1-23
GM10 UG Book.book Page 24 Tuesday, March 24, 2009 11:22 AM
• Error tone — an alternating low-high tone can be heard if any error occurs
during treatment
The Last Man Out system acts as an additional safety feature for supervision
of the door of the treatment room. This system is active as long as the power to
the afterloader is switched on. The Last Man Out button is located inside the
treatment room, next to the door. In order for a treatment to commence, this
button must be pressed prior to closure of the treatment room door.
After preparing the patient for treatment, you need to verify that no one except
the patient is in the treatment room. You then press the Last Man Out button
before exiting the treatment room. The green indicator light inside the Last
Man Out button begins to flash. Once the Last Man Out button is pressed,
you have a predefined period of time to leave the treatment room and close the
door. This time period is set by service personnel.
If you correctly follow the Last Man Out sequence within the given time
period, treatment may begin. The flashing green indicator light on the Last
Man Out button turns to solid green. The DOOR indicator light on the Control
Console is switched off.
After pressing the Last Man Out button, if you do not close the door or do not
close the door within the given time period, the room is not accepted as
certified for treatment. The flashing green light on the Last Man Out button
turns off, and treatment may not begin. The DOOR indicator light on the
Control Console remains illuminated.
The printer is connected to the control PC via a standard USB interface. The
printer helps you document the steps taken during treatment and also produces
reports on any malfunction of the system. You can print data disclosure
reports, treatment history, error reports, diagnostic reports, and other reports
related to system usage. The printer should be switched on for normal
operation and should contain an adequate supply of paper. Refer to the user
manual supplied with the printer for details on usage.
Safety Interlocks
The GammaMedplus iX system contains a series of safety interlocks and
interlock checks that must be satisfied before you can begin treatment. Some
of these interlocks are represented by icons in the Treatment Delivery Strip on
the Treatment Delivery Page. The icons for satisfied interlocks display in green
and for unsatisfied interlocks display in red.
Introduction 1-25
GM10 UG Book.book Page 26 Tuesday, March 24, 2009 11:22 AM
• Door interlock
This interlock interrupts the treatment in progress and retracts the source
into the shielded position if the treatment room door is opened while a
treatment is in progress. After the treatment room is cleared of personnel
and the door is closed, the treatment can be resumed by pressing the
START button on the Control Console. This interlock also prevents
initiation of a treatment if the door is already open.
• Emergency return switch
This interlock detects the status of the EMERGENCY RETURN
switches (at least one is located in the control room with a second one in
the treatment room) and disables or interrupts treatment if this switch has
been pushed and remains pressed down. Refer to the “Automatic and
Emergency Retraction of the Source” section of Chapter 5 “Emergency
Procedures” for more information.
Note: If you hover the mouse pointer over an individual icon, the
name and status (Passed/Failed) of the interlock displays to the
left of the icons in the Treatment Delivery Strip.
The following conditions must also be satisfied before treatment can begin:
Introduction 1-27
GM10 UG Book.book Page 28 Tuesday, March 24, 2009 11:22 AM
Viewing and intercom systems are required for observation of the patient
during treatment. If you are using a video camera, you need to adjust the
camera in such a way that the patient can be observed. An intercom is also
required for two-way verbal communication with the patient at all times.
Note: Viewing and intercom systems are not part of the supply by
Varian unless specifically ordered. See the manufacturer’s
instructions for these devices.
Treatment room
Emergency Container
The container is made of steel with a lead lining and a sealing lead lid. Cutters
attached to the container allow for cutting off the source from the extension
wire. A pair of long handled forceps can be used for picking up the source. The
forceps allow you to hold the source at a greater distance from the body, thus
reducing exposure to radiation.
Introduction 1-29
GM10 UG Book.book Page 30 Tuesday, March 24, 2009 11:22 AM
To start the GammaMedplus iX system, insert the security key in the key
switch of the POWER section on the Control Console and turn it to the
horizontal position. The GammaMedplus iX automatically performs an
indicator test in which all indicator lights on the Control Console (and any
other external consoles) flash briefly, and at the same time an audible signal is
generated. When the indicator test is complete, several indicator lights remain
lit on the Control Console — the MAINS and BATTERY indicators in the
POWER section, the NORMAL indicator in the MODE section, and the
SAFE indicator in the SOURCE section. In addition, the LOCKED indicator
in the SOURCE section of the Control Console and afterloader keypad
flashes.
When you power up the control PC, the afterloader control software
automatically starts up on the PC. During the startup sequence, the control
software performs a series of self-tests. If any of the self-tests fail, the control
software displays detailed information describing the problem, probable
causes, and suggested actions to take.
The administrator for the afterloader control software specifies which users
have access to the system. Each user has a login account that belongs to a user
group. Members of a group can perform certain functions based on the access
privileges associated with the group. The user interface allows for flexibility in
assignment of user privileges, including the privileges of the administrator
group.
Parts of the control software are designated as controlled areas that can only be
accessed by authorized personnel, such as Varian authorized service personnel
and physicists. Utilities are available for performing safety and maintenance
related functions, and for testing the system and measuring crucial parameters.
Refer to the “Define User Access and Location Information” section in
Chapter 4 “Introduction” for more information.
Introduction 1-31
GM10 UG Book.book Page 32 Tuesday, March 24, 2009 11:22 AM
Note: For safety reasons, only Varian authorized service personnel are
permitted to exit to the operating system from the control
software.
Start Page
The control software provides two major workflows and a comprehensive set
of system configuration options that you initiate from a single point — the Start
Page. This page is displayed automatically when you turn on the control PC.
The workflows and configuration options are represented by the Open Patient
File, Create New Patient File, and System Configuration buttons on the
Start Page. The Start Page also has a Shut Down button that enables you to
gracefully shut down the control software and power down the control PC.
Elements of a Page
Each page of the user interface is divided into a set of horizontal strips. Each
strip is devoted to either displaying or entering related sets of data, such as
patient demographic information, channel dwell information, and system 1
status. Some strips, such as the System Status Strip, are displayed on all pages
of the interface, providing constant information on the general status of the
system.
Dates and times on any given page are displayed in standard ISO 8601 format
— yyyy-mm-dd for dates and hh:mm:ss for times. Numeric formats reflect
the settings chosen in the Windows Regional Settings dialog during system
configuration.
Introduction 1-33
GM10 UG Book.book Page 34 Tuesday, March 24, 2009 11:22 AM
Major Workflows
The workflows available from the Start Page are:
The User Login Page, which is displayed each time you press the Open
Patient File, Create New Patient File, or System Configuration button, is
described below.
The User Login Page is the first page shown regardless of the selected
workflow. This page enables you to enter your username and password for
logging into the control software, specify the physical location of the 1
afterloader, change your password, and log out of the control software. The
current date and time are also displayed. The Workflow field indicates the
current workflow.
Introduction 1-35
GM10 UG Book.book Page 36 Tuesday, March 24, 2009 11:22 AM
Edit a Password
To change your password, press the Edit Password button and enter the
appropriate data in the Operator, Current password, New password, and
Confirm password fields of the Edit Password Strip. The password must be
between 4 and 15 characters in length with no spaces. Then press the Save
button. User names and passwords are set in the system configuration options
by your administrator.
Note: The first time you log into the control software, you must
change the password given to you by your administrator. You
will be prompted with a message. When you confirm the
message, the Edit Password Strip is displayed.
The Cancel button closes the Edit Password Strip without saving your
changes.
Choose a Location
For transportable devices, the Location drop-down list enables you to select
your current location from a list of locations that your administrator specifies
in the system configuration options.
Note: When you log on, the control software checks to see if there are
any reasons why you cannot deliver a patient treatment and
displays any relevant messages at the bottom of the screen.
Introduction 1-37
GM10 UG Book.book Page 38 Tuesday, March 24, 2009 11:22 AM
Important: Do not use the value of the Days since last source exchange to
calculate source decay. This value may not provide an accurate
calculation, since it decrements beginning with the date of
source wire installation. Refer to the ”Enter the Calibrated
Source Strength” section in Chapter 4 ”System Configuration”
and the ”Source Decay” and ”Transit Dose” sections in
Appendix B ”GammaMedplus iX Calculations” for more
information.
Network Connectivity
The afterloader control software supports an optional connection to a local area
network. This connection enables you to access shared remote directories for
treatment plans, treatment tests, and treatment history. It also allows you to
back up the patient database to a backup server and print to a network printer.
This chapter describes all of the processes involved in working with patient
and treatment data. The chapter is divided into three major sections:
In This Chapter
Topic Page
Open Patient File Workflow 2-2
Create New Patient File Workflow 2-24
View and Edit Channel Data 2-30
2-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Location in the
Tasks
Control Software
This “Open Patient File Workflow” section describes the tasks associated with
the Patient Selection and Fraction View Pages. The “View and Edit Channel
Data” section of this chapter describes how to view and edit channel data using
the Treatment Delivery Page. Chapter 3 “Delivery of Patient Treatment”
provides detailed information on the remainder of the tasks associated with the
Treatment Delivery Page. Refer to the section called “Overview of the
Afterloader Control Software” in Chapter 1 for information on the User Login
Page.
2
The User Info Strip provides read-only data about who is currently logged into
the control software, including the Operator, Location, and Current date/
time. It also provides buttons for securing or logging out of the system.
Note: The Secure System and Log Out buttons are disabled if a
treatment is currently in progress.
The Patient Selection Page enables you to perform the following tasks:
You can click on a treatment record to select and highlight it in the table. Click
on the header of a field to sort the records in ascending or descending order
using the values in the selected field. By default, the records are sorted on
Console entry date and time in descending order. The Selected patient data
above the Active treatment records table displays additional demographic
data about the selected patient.
Note: You are only permitted to browse patient records for those
locations for which you have access rights. Your administrator
specifies your location or locations in the system configuration
options.
You can edit Selected patient data by selecting a treatment record in the
Active treatment records table and pressing the Edit Record button. All
static fields (except the treatment file name) for the patient switch to editable
text boxes. Click on the field or fields you want to edit and make any necessary
changes. Press the Save Changes button to save your changes to the database
or the Cancel Changes button to discard the edits. You cannot perform any
other actions while an edit is in progress, except canceling or saving the edits.
Note: In order to edit patient records, you must have the Edit
treatment plan privilege, as assigned by your administrator.
Note: The Edit Record button is not available if the Plan Import
Wizard or Treatment Record Maintenance Strip is open.
Note: The control software displays yellow text boxes for all required
fields throughout the user interface.
WARNING 2-1:In order to ensure accurate dwell time scaling for treatments
delivered by the GammaMedplus iX system, BrachyVision™
must be set so that plans are based on nominal source activity
when storing treatment files in DICOM format. This
requirement also applies to plan data created in BrachyVision™
and stored in an ARIA database.
When you press the Import Plan button at the bottom of the Patient Selection
Page, the Plan Import Wizard Strip opens. This strip steps you through the
process of importing a treatment plan file, including:
Your administrator specifies one or more file import locations in the system
configuration options. A location can be a network drive, local hard drive,
USB, or some other removable media. If ARIA integration is enabled, the
ARIA option also appears as a file import location.
When you press the File Location button, you see the current list of locations.
Choose the appropriate location for your treatment files.
When you choose ARIA from the File Location list, the Plan Import Wizard
displays one of two pages, depending on the ARIA Communications Settings
in the system configuration options.
• If Enable ARIA integration and Enable appointment integration are
activated, the Plan Import Wizard displays an Appointments table listing
the appointments that match the most recent selection criteria. By default,
the table displays the appointments for the current date. The Appointments
table shows the following information: patient ID, patient name, machine
name, procedure, start time, end time, and status.
• If only Enable ARIA integration is activated, the Plan Import Wizard
displays the Change Patient Filter Page.
On the Appointments page, select the appropriate appointment and press the
Continue button. The control software displays a table of Plans associated
with the selected appointment. Choose the appropriate plan and press the
Continue button. Continue with the Plan Import Wizard until the plan is
imported.
Important: You cannot import a plan if a patient has multiple plans with the
same Plan ID. In this case, you must assign the plan of interest
a unique Plan ID in BrachyVision™ before the plan import
from ARIA is allowed.
When only the Enable ARIA Integration check box is activated in the system
configuration options, the Plan Import Wizard displays the Change Patient
Filter Page where you can use a Patient ID1 or Last Name to query the patient
plans directly. See the description of the No Appointment button above for
more information.
2
Note: If the current search does not display the record of interest, you
can press the Change Filter button again and enter new search
criteria.
For all treatment files in locations other than the ARIA database, the Plan
Import Wizard displays a table containing information about each treatment
file in the currently selected location. The table shows the following
information: patient ID, patient name, treatment site, treatment file name, and
file date.
Note: The filenames for files that do not pass initial validity checks
upon import are pre-pended with an exclamation point (!).
Select the file in the plan import table and press the Continue
button to view an error message that describes the problem.
Choose the patient file of interest from the table and press the Continue button.
The Continue button performs a check on the data to verify that the treatment
parameters are within the operational limits of the afterloader. The verification
includes the number of channels, the number of dwell positions within each
channel, the spacing between dwell positions, and the channel length. If the
treatment file does not meet the verification requirements, a validation error
message is displayed.
If the treatment file is missing the patient ID, an error message is displayed and
the file cannot be imported. If a treatment file you want to import shares the
same name as a treatment file in the Active treatment records database, a
warning message is displayed before you can proceed. To avoid importing the
same file twice, the control software deletes the treatment file from its original
location if possible when you press the Continue button.
There is no limit on the number of patient treatment files that can be imported
and displayed in the Plan Import Wizard or the number of patient records that
can reside in the Active treatment records database. The maximum number
of patient files is limited only by the available hard disk space on the local
drive.
The Plan Import Wizard also displays buttons for deleting a treatment plan file,
canceling the import, and continuing to the next page of the wizard.
When you select a treatment file, the Continue and Delete File buttons are
enabled, and if a corresponding record for the selected plan already exists in
the database, the record is highlighted in the active treatment records table. The
Delete File button displays a confirmation message before the deletion takes
place.
Important: A deleted file is removed from the selected file location and
cannot be recovered unless you have stored a copy of it
elsewhere.
The Cancel Import button closes the Plan Import Wizard Strip without
importing any treatment files.
The Data Verification Page of the Plan Import Wizard appears after you
select a treatment file and press the Continue button. This page contains the
read-only patient ID and editable text boxes for last and first names.
The Last name is a required field, and the First name is optional. If possible,
these fields are populated with data from the treatment file currently selected
for import. If data is missing from the selected file, the fields are populated
with data from the patient database if a record exists for the patient ID.
The buttons displayed on the Data Verification Page depend on whether or not
you have previously imported a treatment file for a particular patient ID. If this
is the first plan import for an ID, the Cancel Import, Previous, and Done
buttons appear. Otherwise, a Continue button appears in place of the Done
button.
The Cancel Import button closes the Plan Import Wizard Strip without
importing any treatment files. The Previous button returns to the previous
page of the Plan Import Wizard. The Done button completes the plan import
and updates the Active treatment records database.
If this is the first plan import for a particular patient, click on the field or fields
you want to edit, make any necessary changes, and press the Done button to
close the Plan Import Wizard. The new treatment record appears and is
highlighted in the Active treatment records database.
If data already exists in the Active treatment records database for the patient
ID whose treatment file you are importing, make any necessary edits in the text
boxes and press the Continue button. If you make changes to the Last name
or First name, a message appears asking you to confirm if you would like to
use the new values. The next section, “Reconciliation with Existing Plans”,
describes your options for reconciling the record you are importing with the
record that already exists for the patient ID.
When you press the Continue button on the Data Verification Page of the Plan
Import Wizard, the Data Reconciliation Page appears with two option buttons:
The first option adds the imported plan to the most recent course of treatment
for the patient. When the import is completed, the Delivery status column in
the Active treatment records table is updated to reflect the added fraction.
The second option makes the imported plan the first fraction of a new course
of treatment for this patient. When the import is completed, a new record
appears and is highlighted in the Active treatment records database. The
Delivery status column in the plan database shows 0 of x fractions for this
plan, where x is the total number of fractions specified in the treatment file.
Note: If the imported plan has been exported from the control
software (.xml format) and contains one or more fractions that
have already been delivered, only the second option above is
available.
Choose an option and then press the Done button to complete the import and
close the Plan Import Wizard.
There are three types of record maintenance tasks you can perform:
2
Note: In order to perform record maintenance, you must have the Edit
treatment plan privilege, as assigned by your administrator. If
you do not have this privilege, the Record Maintenance button
is disabled.
For each type of task, you can select one or more records. Select a single record
by clicking on it in the database table. You can select a group of adjacent
records by highlighting the first record, pressing and holding down the Shift
key, and then highlighting the last record. To select a group of non-adjacent
records, highlight one, then press the Ctrl key and hold it down while you
individually highlight all the other records you want for the group.
The control software enables you to designate the status of a record as Active
or Inactive. The active status is used for courses of treatment that are currently
in progress. Active records are displayed in the Active treatment records
table. The inactive status is used for courses of treatment that are either
complete or dormant. Inactive records are displayed in the Inactive treatment
records table.
You can change the status of a record and toggle between the Active and
Inactive treatment records tables using the Activation status group box of
the Treatment Record Maintenance Strip. Press the Record Type button and
select the Active or Inactive option to toggle between the two tables.
The export function enables you to copy one or more selected treatment files
to a specified directory. Plans exported in this way can be used as back up
copies or reimported into another control PC using the Plan Import Wizard.
You can export files from the Active treatment records or the Inactive
treatment records table. Your administrator designates one or more Export
Locations in the control software system configuration options. You use the
Export group box in the Treatment Record Maintenance Strip to select an
Export Location and initiate the export.
1. Press the Export Location button and select the appropriate export
directory from the drop-down list. By default, the Export Location list
displays the most recently used export directory.
2. Highlight the record or group of records you want to export in the Active
or Inactive treatment records table and press the Export button. A 2
confirmation message is displayed when the export is complete.
The export function creates one XML file for each treatment file you export.
The name of the XML file is the treatment file name displayed on the Patient
Selection Page. The Modified date for the exported file reflects the date the file
was created in the control software. You can import the XML file back into the
control software by selecting the export file location in the Plan Import Wizard,
providing your administrator has configured your import file location this way.
Note: XML files are text files that are readable in any text editor. Use
due diligence to ensure that patient identification information
and other sensitive data are protected.
Delete a Record
The Report Page and report format are similar for all reports in the control
software. Under the title of each report is an information line that includes the
name of the user logged on at the time the report is printed, clinic (facility
name, if available), machine serial number, and the date and time the report is
printed. Each report (except Data Disclosure) displays the patient name and
fraction number (if applicable) in the page header and the type of afterloader,
version and build number of the control software, and firmware version of the
afterloader in the page footer.
The Report Page for all reports in the control software includes a number of
buttons that enable you to navigate a report, including First, Previous, Next,
and Last buttons. Options for printing the report are also included — Entire
Document, Current Page, and Page Range.
The Save to File Option enables you to save a report in PDF format. Press the
File Location button and select a folder from the list. Folder locations are
designated in the Folder Setup tab of the User & Location strip in System
Configuration. Press the Save button to save the file.
Press the Print button to print a hard copy of a report and the Close button to
close the report and return to the current page.
You can view and print patient Data Disclosure and Treatment History Reports
for a selected record using the buttons in the Reports box on the Patient
Selection Page.
When you press the Data Disclosure button, the Data Disclosure Report for
the selected record appears. This report tracks all changes made to the patient
demographic data for a particular treatment file since it was imported into the
control software. The report is divided into two main sections — Patient/
Treatment Data and Access History. The Patient Data includes Patient Last
Name, Patient First Name, Patient Identification, Sex, and Patient Birthdate.
The Plan Data includes Total Planned Fractions, Physician, Treatment Site,
Applicator, and Plan File. The Access History describes each time the record
was changed. The Access History table includes:
When you press the Treatment History report button, the Treatment History
Report appears. This report tracks the details of the treatments delivered to the
patient. Each report contains a summary page consisting of Patient/ Treatment
Data and a Treatment Summary. The Patient/ Treatment Data summary section
includes Patient Data (Patient Last Name, Patient First Name, Patient
Identification, Sex, and Patient Birthdate) and Plan Data (Total Planned
Fractions, Physician, Treatment Site, Applicator, and Plan File).
The Treatment Details include the Channel Length, Dwell Positions (total),
Planned Time, Scaled Time, Delivered Time, Remaining Time, and a table
showing the dwell Positions, Planned dwell times, Scaled dwell times, and
Delivered dwell times for each channel.
The Source Data section includes the Remaining Active Wire Cycles,
Remaining Source Days, Calibration Date, and Wire Serial number.
If the treatment was interrupted for any reason, an additional Interruption Data
section is included for each fraction. This section includes Time of
Interruption, error Code, error Class, error Description, Channel number,
Dwell Location, Time Spent at Dwell, and Continuation Option.
When you are finished with the Patient Selection Page, press the Proceed
button to display the Fraction View Page, where you can view and modify
fractions.
Note: You cannot proceed to the next page if the Plan Import
Wizard or Treatment Record Maintenance Strip is open.
The User Info Strip provides read-only data about who is currently logged into
the control software, including the Operator and Location. It also provides
buttons for securing or logging out of the system. The Secure System and Log
Out buttons are disabled when a treatment is in progress. Refer to the “Select
or Import a Treatment File” section in this chapter for information on the
Secure System button.
The Patient Info Strip provides read-only data on the currently selected patient,
including Patient, Patient ID, Birth date, Treatment site, and Physician. Refer
to the “System Status Strip” section in Chapter 1 for more information on the
System Status Strip.
The top of the Fraction View Strip provides data on the currently selected
fraction, including the following fields:
• Fraction — the number of the currently selected fraction and the total
number of fractions for this plan (read-only field)
• Planned strength — the source strength assumed by the planning system
when determining dwell times. Scaled dwell times are adjusted at delivery
time according to the current source strength. (read-only field)
• Operator — for fractions whose Delivery status is Complete, the name
of the person logged on when the treatment is delivered; for a fraction that
has not been delivered, the field is blank
• Planned by — the ID of the person who created the plan (editable field)
• Applicator — the applicator specified in the treatment plan (if any)
(read-only field)
The middle of the Fraction View Strip displays the Fraction View Table, which
includes the following information for each fraction:
The Fraction View Page enables you to perform the following tasks:
You can copy an existing fraction and add the copy to the end of the list of
fractions in the Fraction View Table. Select the fraction you want to copy in
the table and press the Copy and Add Fraction button. The new fraction
appears in the table, and the data in the fields at the top of the strip is updated
accordingly. The Matches field indicates which fraction was used for the copy.
The new fractions are not added to the database until you press the Save button.
When you press the Print Fraction Report button, the Selected Fraction
Report appears. This report provides details for the selected fraction. The
report is divided into the following main sections — Patient/ Treatment Data,
Planned/ Actual Comparison, Treatment Details, and Source Data.
In the Patient/ Treatment Data section, the Patient Data includes Patient Last
Name, Patient First Name, Patient Identification, Sex, and Patient Birthdate.
The Plan Data includes Fraction, Physician, Treatment Site, Applicator, Plan
File, Plan Author, and Console Entry Date. The Delivery Data (if available)
includes Delivery Status, Decay Scale Factor, Operator, Delivery Date, Start
Time, and End Time.
The Treatment Details include the Channel Length, Dwell Positions (total),
Planned Time, Scaled Time, Delivered Time, Remaining Time, and a table
showing the dwell Positions, Planned dwell times, Scaled dwell times, and
Delivered dwell times for each channel.
The Source Data section includes the Remaining Active Wire Cycles,
Remaining Source Days, Calibration Date, and Wire Serial number.
If the treatment was interrupted for any reason, an additional Interruption Data
section is included at the end of the report. This section includes Time of
Interruption, error Code, error Class, error Description, Channel number,
Dwell Location, Time Spent at Dwell, and Continuation Option.
Press the Print button to print a hard copy of the report or the Close button to
close the report and return to the Fraction View Page.
When you press the Cancel button, all additions and deletions to the Fraction
View Table performed since the last save are undone.
When you press the Save button, all information displayed in the Fraction
View Table is written to the treatment history file for the selected patient,
including current modifications.
The Previous and Proceed buttons enable you to return to the Patient
Selection Page or advance to the Treatment Delivery Page. If you use the Copy
and Add Fraction button or Delete Fraction button, the Previous and
Proceed buttons are disabled until you Cancel or Save your changes.
When you press the Proceed button, the control software performs an integrity 2
check on the selected fraction to ensure it will run properly on the attached
afterloader and displays a validation error if necessary.
Refer to the “View and Edit Channel Data” section at the end of this chapter
for information on how to view and edit channel data using the Treatment
Delivery Page, or to Chapter 3 “Delivery of Patient Treatment” for detailed
information on delivering a treatment for the selected fraction.
The Create New Patient File workflow also enables you to use a standard plan
as a model or starting point for a new plan. Standard plans are created using the
system configuration options and can be selected on the Fraction Creation Page
in the Create New Patient File workflow.
The pages that comprise this workflow, along with the set of tasks associated
with each page, are shown in the table below.
Location in the
Tasks
Control Software
This “Create New Patient File Workflow” section describes the tasks
associated with the Patient Creation and Fraction Creation Pages. The “View
and Edit Channel Data” section of this chapter describes how to view and edit
channel data using the Treatment Delivery Page. Chapter 3 “Delivery of
Patient Treatment” provides detailed information on the remainder of the tasks
associated with the Treatment Delivery Page. Refer to the section called
“Overview of the Afterloader Control Software” in Chapter 1 for information
on the User Login Page.
2
WARNING 2-2: While you can create new patient treatment plans using the
Create New Patient File workflow, the afterloader control
software does NOT perform any dosimetric calculations related
to the determination of the proper treatment plan except for
computing source decay. Be sure to check the integrity and
clinical accuracy of any data that you enter before delivering a
treatment. Failure to do so may result in incorrect treatment and
may cause serious injury to the patient.
The User Info Strip provides read-only data about who is currently logged into
the control software, including the Operator and Location. It also provides
buttons for securing or logging out of the system. The Secure System and Log
Out buttons are disabled when a treatment is in progress. Refer to the “Select
or Import a Treatment File” section in this chapter for information on the
Secure System button.
The Patient Creation Strip enables you to enter patient demographic data. This
strip is described in detail in the following sections. Refer to the “System
Status Strip” section in Chapter 1 for more information on the System Status
Strip.
You can specify the following patient data using the editable text boxes and
drop-down lists in the Patient Creation Strip:
The patient First name, Last name, and Patient ID1 fields are required. When
you have entered data in these fields, the Proceed button is enabled. When you
have finished entering data in the Patient Creation Page, press the Proceed
button to continue to the Fraction Creation Page.
When you press the Proceed button, the database is queried to see if a patient
with the specified ID already exists. If so, a warning is issued indicating that
the plan about to be created will constitute a new course of treatment for this
patient.
The User Info Strip provides read-only data about who is currently logged into
the control software, including the Operator and Location. It also provides
buttons for securing or logging out of the system. The Secure System and Log
Out buttons are disabled when a treatment is in progress. Refer to the “Select
or Import a Treatment File” section in this chapter for information on the
Secure System button. The Patient Info Strip provides read-only data on the
currently selected patient, including patient name, patient ID, birth date,
treatment, and physician.
You can specify the following fraction data using the editable text boxes and
drop-down lists in the Fraction Creation Strip:
The Number of fractions and Planned strength fields are required. Once you
have entered data in these fields and pressed the Save button, the Proceed
button is enabled. Once you press the Save button, the Previous button is
disabled.
When you have finished entering data in the Fraction Creation Page, press the
Proceed button to continue to the Treatment Delivery Page. The control
software validates all editable fields to ensure they fit within allowed
constraints.
Refer to the “View and Edit Channel Data” section below for information on
how to view and edit a channel using the Treatment Delivery Page, or to
Chapter 3 “Delivery of Patient Treatment” for detailed information on
delivering a treatment for the selected fraction.
This section describes how to use the Treatment Delivery Page to view and edit
channel data. Refer to Chapter 3 “Delivery of Patient Treatment” for detailed
information on each of the other tasks in the list above.
By default, the Treatment Delivery Page consists of the User Info, Patient Info,
Fraction Info, Channel View, Treatment Initiation, Channel Edit, and System
Status Strips. The Treatment Delivery Strip is available once you initiate a
treatment. The Partial Treatment Strip becomes available if a treatment is
cancelled or interrupted for any reason.
This section focuses on the Channel View and Channel Edit Strips.
You can click on a channel to select and highlight it in the Channel View Strip.
You can use the Up and Down arrow keys and the Page Up and Page Down
keys on the keyboard to move up and down through the channels and select
them. A vertical scroll bar appears if necessary and enables you to scroll
through all of the channels.
When you open the Treatment Delivery Page, by default, the control software
ensures that the most distal dwell position of the first occupied channel is
visible in the Channel View Strip. For other channels, if the proximal and distal
dwell positions are within the viewable range, the display marks these two
positions, as seen in Channel 03 below. If either position lies outside the
current viewable range, small black numbers indicate the left and/or right-most
boundary of this range, as seen in Channel 04 below.
Dwell points are shown as histogram bars within the graphical representation
of each channel, where the height of the bar indicates the total planned time at
that position. Refer to the “Deliver and Monitor the Treatment” section of
Chapter 3 “Delivery of Patient Treatment” for detailed information on the
appearance of the channel representation and the values for the read-only fields
during treatment delivery.
Upon initial entry, the Channel View Strip shows the dwell times as specified
in the treatment plan. Once you press the Initiate Treatment button, the
planned dwell times are updated to scaled dwell times based on current source
strength. Refer to the “Deliver and Monitor the Treatment” section in
Chapter 3 “Delivery of Patient Treatment” for more information on initiating
a patient treatment.
2
The Channel Edit Strip
The Channel Edit Strip opens when you press the Edit Channel button on the
Treatment Delivery Page. This strip enables you to add and delete channels
from a selected fraction and add, delete, and modify dwell positions and times
within a selected channel. This section describes how to edit a channel using
the Channel Edit Strip. Refer to the “Deliver and Monitor the Treatment”
section in Chapter 3 for detailed information on the other tasks associated with
the Treatment Delivery Page.
Note: For the Create New Patient File workflow, if there are no
dwell positions currently defined, the Channel Edit Strip opens
automatically when you access the Treatment Delivery Page.
The Channel Edit Strip consists of text boxes, static text, buttons, and a
graphical representation of the dwell positions and times in the current
channel.
Text Boxes
The following text boxes are included in the Channel Edit Strip:
• Name — enter the name of the channel. When you press the Save button,
the Channel name is displayed in the first column of the Channel View
Strip.
• Length (cm) — a read-only field specifying the length of the channel
(which is fixed at 130 cm for all channels of the GammaMedplus iX)
• Step size (cm) — enter the step size in cm between the range of 0.1 and
1.0 cm. The control software accepts only 1 decimal place.
Important: There is no default step size setting for unplanned channels, i.e.,
channels that you add using the Channel Edit Strip. You must
enter a step size before you can add any dwell positions when
creating new plans or adding new channels to an existing plan.
Note: When you change the step size for a selected channel, if one or
more adjacent distances is not a multiple of the new step size,
all of the dwell positions within the channel will be erased, and
you will be required to re-enter them by hand. The control
software displays a confirmation message before the dwell
positions are erased.
• Origin (cm) — You can change the distal position of the source separately
in each channel to adjust for patient anatomy. Select the channel you want
to modify, and enter the offset from the distal position in the Origin field.
A value of 0.0 cm represents the full 130.0 cm channel length; a value of
20.0 cm represents a 20.0 cm offset from the 130.0 channel length, or a
110.0 cm distal position. As an alternative, you can press the Origin button
at the bottom of the Channel Editor, and follow the prompts at the bottom
of the screen. Refer to the “Change the Distal Position of the Source”
section below for more information.
Note: The offset value must be in the range of 0.0 to 40.0 cm with a
resolution of 0.1 cm.
Static Text
• Dwell time scale (s) — the maximum dwell time currently set for any
single position in the current channel
• Total dwell time (s) — the sum of the dwell times for all of the dwell
positions set for the current channel
2
• Total dwell positions — the total number of dwell positions set for the
current channel, up to a maximum of 60
Channel Representation
The current channel is displayed as a series of vertical bars, with each bar
representing a dwell position. The Channel number appears on the left side
of the channel representation. At both ends of the channel representation are
Directional scroll buttons.
The Dwell Selector enables you to view and enter the dwell times for the dwell
positions in the channel. The Dwell Selector Range contains a Boundary
Lock button that enables you to secure the treatment range and designate the
allowable left and right boundaries for the dwell selector. These tools are
discussed in detail in the next section.
The up and down arrows enable you to scroll up and down through the set of
available channels in the current fraction. When you click on an arrow, the
Channel number updates to indicate the current channel. The selected
channel is highlighted in the Channel View Strip. You can also select and
highlight a channel by clicking on it in the Channel View Strip.
The left and right arrows enable you to scroll left and right through the dwell
positions. Each time you click on an arrow, the Dwell Selector moves one
position in the direction you choose according to the current step size. You can
also use the left and right arrow keys on your keyboard to scroll left and right
or the Enter key. Refer to the “Move the Dwell Selector” section below for
more information.
The page left and page right arrows enable you to page a full view at one time.
If the boundary lock is open, you can page to the end of the channel. If the lock
is closed, you can page as far as the first or last dwell position.
The Channel Edit Strip enables you to perform the following tasks:
• Edit a channel
• Copy and paste channel data
• Change the distal position of the source
• Delete a channel
• Save or cancel changes
Edit a Channel
In addition to using the text boxes to edit a channel, you can use the Dwell
Selector to create or modify dwell positions and the buttons at the bottom of
the Channel Edit Strip to Copy, Paste, and Delete channel data.
The Dwell Selector enables you to view dwell time and dwell position settings
for existing dwell points and to create new dwell positions within the current
channel.
Dwell Time
Dwell Point
2
Dwell Position
Important: If the most distal dwell position is set at the end of the channel
(130 cm), the Channel View and Channel Edit strips will
display this position at 130 cm, but the actual treatment position
will be 129.9 cm, due to a 1 mm offset from the end of the
channel.
There are several ways to move the Dwell Selector along the length of the
channel representation. You can click on the scroll left and right buttons in the
Channel Edit Strip. You can also click once on the channel representation and
then use the left and right arrow keys on the keyboard to scroll left and right.
As you move the Dwell Selector to each adjacent Dwell Point, the Dwell
Position at the bottom of the selector updates according to the designated step
size. The Dwell Point number also updates with each position. The
GammaMedplus iX system supports up to 60 dwell positions per channel.
Note: The dwell positions are numbered from 1 to 60 starting from the
most distal position.
You can click on the Scroll Handle in the top right corner of the Dwell
Selector and hold down the mouse button as you drag the selector to the
position you want. When you click on a dwell position, the Dwell Selector
snaps to that position.
When entering dwell times, you can also use the Enter key to move the Dwell
Selector one position at a time to the left or right. The direction (left or right)
depends on the direction last used with the scroll left and right buttons or the
scroll handle.
To enter or modify a Dwell Time, click on the Dwell Time Lock to change the
status to unlocked. Enter the dwell time in seconds in the Dwell Time field,
using one decimal place if necessary. The dwell time must be between 0.10
seconds and 9999.9 seconds.
Dwell times are not saved in the control software database until you press the
Save button, at which time the Channel Edit Strip closes and the changes are
displayed in the Channel View Strip.
Note: If you enter a Dwell Time in the Dwell Selector that exceeds
the current Dwell Time Scale, the control software readjusts the
scale of all of the dwell positions in the channel representation
when you press the Enter key or the Dwell Time Lock. The
new dwell time is displayed in the Dwell Time Scale field, and
the height of the displayed dwell times is normalized to this new
maximum value.
If the currently selected channel has at least one designated dwell position, the
Dwell Selector Range appears at the bottom of the channel representation.
When two or more dwell points are designated, the Dwell Selector Range
displays the right and left most dwell positions in the channel in cm.
When the Boundary Lock button is unlocked, you can scroll the dwell selector
outside the boundaries of the dwell selector range. When the Boundary Lock
button is locked, you can only scroll the dwell selector within the dwell selector
range. Click on the lock to toggle the status.
You can copy and paste all of the dwell positions from one channel to another.
First, select the channel whose data you want to copy by highlighting the
channel in the Channel View Strip or using the up and down arrows in the
Channel Edit Strip. Then press the Copy button. The control software displays
a confirmation message. After you confirm the message, select the channel to
which you want to copy the data and press the Paste button. The control
software displays a confirmation message again. Confirm the message. The
copied data appears in the channel representation of the Channel Edit Strip.
When you press the Save button, the Channel Edit Strip closes and the updated
channel data appears in the Channel View Strip.
Note: If the copied channel has a Channel Name, this name is not
pasted into the new channel; instead, the Name field is left
blank.
You can change the distal position (origin) of the source separately in each
channel to adjust for patient anatomy once the applicator is inserted. Select the
channel you want to modify, press the Origin button at the bottom of the
Channel Editor, and follow the prompts at the bottom of the screen.
This procedure requires that you turn the MODE key switch to the SPECIAL
position and press the START button on the Control Console.
The dummy wire extends fully in the selected channel and the control software
displays a message prompting you to use the IN and OUT buttons in the
DUMMY section of the afterloader keypad to adjust the origin offset.
When you press the IN or OUT button, the dummy wire moves in 0.5 cm
increments. As you adjust the offset, the position is updated in the message
box. The maximum offset for the origin is 40.0 cm.
WARNING 2-3: The control software does not check to see if the adjusted
position of the origin is correct. You must be aware of the
potential risk of a mistake when adjusting the origin and must
carefully check the adjusted position in each channel that you
change. Failure to do so may result in serious injury to the
patient.
When you are finished, press the Save button. The dummy wire retracts fully
and the channel editor indicates the new setting for the distal position.
You can cancel the procedure at any time by pressing the Cancel button.
Delete a Channel
You can delete a selected channel by pressing the Delete button and
confirming the message. Then press the Save button to save the change, close
the Channel Edit Strip, and display the updated channel information in the
Channel View Strip.
Press the Save button to save all of the channel modifications and close the
Channel Edit Strip. The control software displays the updated channel
information in the Channel View Strip.
Note: If you make any changes to dwell positions or dwell times and 2
press the Save button, the control software displays a message
indicating that the dose assigned to the plan may no longer be
correct and that the changes apply only to the current fraction.
The message asks you to confirm that you want to save the
changes.
Press the Cancel button to discard the channel modifications you made since
the last time you pressed the Save button and close the Channel Edit Strip. In
this case, the Channel View Strip is not updated.
In This Chapter
Topic Page
Typical HDR Treatment Sequence 3-2
Perform Treatment Day Tests and Checks 3-3
Insert the Applicator 3-4
Determine the Treatment Plan 3-5
Prepare the Patient, Afterloader, and Control Software 3-6
Deliver and Monitor the Treatment 3-13
Complete the Treatment Delivery 3-29
Interrupt a Treatment 3-33
Partial Treatment Strip 3-34
3-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Important: Patient preparation must not begin until all safety checks and
tests are completed successfully.
2. Insert the Applicator
In accordance with good clinical practice and the relevant applicator
instructions for use, insert the applicator into the patient. Verify the correct
positioning of the applicator.
3. Determine the Treatment Plan
Under the guidance of a physician, with reference to a treatment planning
system, determine the treatment positions and dwell times for each
channel.
4. Prepare the Patient, Afterloader, and Control Software
According to the treatment plan, place the patient in the proper treatment
position with respect to the afterloader, connect the source guide tubes to
the correct channels of the indexer head, and prepare the control software
for treatment delivery. If necessary, secure the applicator and instruct the
patient in treatment procedures.
Refer to the “Treatment Day Tests and Checks” section in Appendix D “Safety
Information and Tests” for more information on each of the safety checks
listed above. The treatment day tests and checks are a part of the overall quality
assurance and safety procedures implemented at your site. Refer to the
“Perform Physics and Service Functions” section in Chapter 4 “System
Configuration” for more information on creating and running physics test
treatments.
Important: All source guide tubes and catheters with clamping adapters or
combinations of source guide tubes and applicators must be
tested for the correct length of 130 cm with the help of a
calibrated measurement instrument. Before every use, these
combinations must be tested unconditionally for being intact. In
addition, pay attention that there are no liquids in the source
guide tubes or applicators.
WARNING 3-1: The treatment plan must be determined under the guidance
of an experienced physician who has been trained in the
proper use of the GammaMedplus iX system. Failure to do
so may result in serious injury to the patient.
WARNING 3-2: For safety reasons during preparation, the key switch in the
SOURCE section of the afterloader keypad must be in the
LOCKED (vertical) position.
WARNING 3-3: For maximum stability during movement, the height of the
GammaMedplus iX must be in the lowest position when you
roll it into place.
When moving the device, pay attention to any connecting
cables. The cables must be placed appropriately to prevent
tripping. Do not run over the cables. Coil the cables loosely
near the GammaMedplus iX, but never around the handles.
Failure to follow these steps may result in injury to you or
the patient.
WARNING 3-4: Due to risk of injury to the patient, perform the height
adjustment carefully if the indexer head is situated above
the patient.
Note: Before you connect the source guide tubes to the indexer head,
use the length gauge to check the correct source guide
tube/applicator probe combination length. Refer to the
instructions for use supplied with the source guide tubes and
applicator for detailed information.
Connect each source guide tube or applicator to the correct channel of the
indexer head according to the channel number determined in the treatment
planning software. Using slight pressure, insert the coupling of each source
guide tube into the channel opening of the indexer head until the coupling is
engaged. Ensure that all source guide tubes are fully seated in the indexer head.
WARNING 3-5: Be careful to connect the correct source guide tube to the
corresponding channel of the indexer according to the
treatment plan. For more information, refer to the source
guide tube instructions. Interchanging source guide tube
connections will lead to incorrect treatment of the patient.
1
2
23 3
22 4
21 5
6
20
7
19 8
18 9
17 10
16 11
15 14 13 12
The source guide tubes must be clean and free of foreign bodies and moisture.
Water vapor, especially on the inside of the tubes, can lead to damage of the
afterloader.
Free passage of the active wire through the source guide tube and applicator
may be impeded if a tube that is kinked, crushed, stretched, or damaged in any
other way is used. Source guide tubes should be inspected prior to use, and if
they show any signs of defects, they should be discarded.
drive mechanism and the curve and the type of tissue surrounding the source
guide tube. If a curve constriction is higher than the drive can overcome, an
automatic retraction will occur.
WARNING 3-6: If necessary, adjust the height of the indexer head so that the
source guide tubes between the patient and the afterloader
are not kinked, twisted, or stressed in any way. Failure to do
so may cause the treatment to terminate prematurely.
WARNING 3-7: The channel number for each applicator or source guide tube
as determined in the treatment planning software must be
followed unconditionally. Do not mix up the connections.
Interchanging the source guide tube connections will lead
to incorrect treatment of the patient.
3
1
2
23 3
22 4
21 5
6
20
7
19 8
18 9
17 10
16 11
15 14 13 12
Screw the source guide tube support into the upper opening of the indexer head
and tighten it by using a 22 mm open-end wrench. Position the hinge of the
source guide tube support suitably and guide the source guide tubes through
the support arm.
1
2
23 3
22 4
21 5
6
20
7
19 8
18 9
17 10
16 11
15 14 13 12
WARNING 3-10: For safe operation, the tubes must be positioned without
loops, double windings, or kinks. Failure to do so may cause
the treatment to terminate prematurely.
WARNING 3-11: Verify that the applicator components are secured to the
patient or to the treatment table to prevent them from
moving or pulling out.
The “Deliver and Monitor the Treatment” section below begins with the
Initiate Treatment button on the Treatment Delivery Page.
Each of these tasks is described below. Refer to the “View and Edit Channel
Data” section in Chapter 2 “Patient and Treatment Data” for an introduction to
the Treatment Delivery Page and detailed information on how to view and edit
channel data using the Channel View and Channel Edit Strips.
Initiate a Treatment
The Treatment Initiation Strip on the Treatment Delivery Page steps you
through the process of initiating a patient treatment. This strip is closed by
default.
When you press the Initiate Treatment button, the control software
automatically renames the Planned dwell time column in the Channel View
Strip to Scaled dwell time and displays the scaled dwell times based on the
decay scale factor. The control software also displays a message to this effect
at the bottom of the screen. The histogram bars in the channel representation
for each channel are updated, if necessary, to reflect these changes. Refer to the
“Scaled Dwell Time” section in Appendix B, “GammaMedplus iX
Calculations” for more information.
The control software performs checks on the data to verify that the treatment
parameters are within the operational limits of the afterloader and displays the
Planned Treatment Report. The control software ensures that all treatment
preconditions, including safety interlock checks supported by the attached
afterloader, are satisfied prior to allowing treatment delivery to begin. Refer to
the “Safety Interlocks” section in Chapter 1, “Introduction” for more
information. The control software also verifies that the afterloader and control
PC clocks are synchronized to within 15 minutes of each other. If any of these
checks should fail, a message is displayed.
3
Note: There are limitations on the minimum usable dwell times,
specifically when using very small dwell times (<0.4 sec). The
minimum dwell time that can be delivered for any dwell
position will vary depending on the distance between adjacent
positions. If one or more dwell times are less than the minimum
recommended dwell time, the control software displays a
message to this effect after you press the Initiate Treatment
button. The treatment is disallowed until you comply with
minimal dwell times that will allow for correct operation as
outlined in the “Minimum Dwell Time Restrictions” section in
Appendix B “GammaMedplus iX Calculations”.
Note: There are limitations on the maximum usable dwell times — the
maximum time per dwell position is 9999.9, the maximum time
per channel is 9999.9, and the maximum time per fraction is
99,999.9. If the treatment plan exceeds any of these limits, the
control software displays a message to this effect after you press
the Initiate Treatment button. The treatment is disallowed
until all time limits are satisfied.
The report is divided into the following main sections — Patient/ Treatment
Data, Planned/ Actual Comparison, Treatment Details, and Source Data.
In the Patient/Treatment Data section, the Patient Data includes Patient Last
Name, Patient First Name, Patient Identification, Sex, and Patient Birthdate.
The Plan Data includes Fraction, Physician, Treatment Site, Applicator, Plan
File, Plan Author, and Console Entry Date. The Delivery Data includes
Delivery Status and Decay Scale Factor.
The Treatment Details include the Channel Length, Dwell Positions (total),
Planned Time, Scaled Time, Delivered Time, Remaining Time, and a table
showing the dwell Positions, Planned dwell times, and Scaled dwell times for
each channel.
The Source Data section includes the Remaining Active Wire Cycles,
Remaining Source Days, Calibration Date, and Wire Serial number.
This report also includes the randomly generated Treatment Code which is
necessary to deliver the treatment. This code is discussed in the next section.
There is space for approval signatures and dates at the bottom of the report.
Press the Print button to print a hard copy of the report and then the Close
button to close the report and return to the Treatment Delivery Page. In the
absence of a printer, you can use the soft copy of the report to review the data
and obtain the Treatment Code.
A suggested partial checklist for comparing the treatment plan report with the
Planned Treatment Report is included below. Follow the verification
procedure for your facility.
Before you deliver the treatment, perform any remaining pre-treatment checks
required at your facility. A sample checklist follows.
• The patient data has been verified and approved for treatment delivery.
• All persons except the patient have left the treatment room.
• The door to the treatment room is closed.
• The viewing and intercom systems are operational.
If you decide to cancel the treatment delivery for any reason, press the Cancel
button. The control software displays a confirmation message. Once you
confirm the message, the Treatment Delivery Strip closes and the Treatment
Initiation Strip appears. All scaled dwell times revert to planned dwell times.
WARNING 3-13: All persons except the patient must leave the treatment
room. Failure to do so may cause serious injury.
1. Insert the SOURCE key into the key switch of the afterloader keypad (if
necessary) and turn the key from the LOCKED to the RELEASED
(horizontal) position. The RELEASED indicator light turns on.
LOCKED
2. Verify that there is no one in the treatment room except the patient. Press
the Last Man Out button (if installed), exit the treatment room, and close
the treatment room door.
1. Verify that the POWER key is inserted in the Control Console and turned
to the MAINS (horizontal) position with the MAINS indicator light on.
(This step must be completed earlier, before treatment initiation can
begin.)
2. Verify that the MODE key is inserted in the Control Console and turned
to the NORMAL (vertical) position with the NORMAL indicator light
on.
3. Enter the 5-digit Treatment Code into the Treatment Code field of the
3
Treatment Delivery Strip. This code is located at the bottom of the Planned
Treatment Report.
4. Press the Deliver Treatment button. The STARTED indicator light on
the Control Console flashes and an audible signal is generated.
Note: Once you press the Deliver Treatment button, the control
software is locked against further input, and the Log Out and
Secure System buttons are disabled.
Note: All data for the channels in the current fraction are sent to the
GammaMedplus iX afterloader when you press the Deliver
Treatment button. There is a steady exchange of data during
the treatment delivery process.
The treatment status and alarm indicators on the Control Console, afterloader
keypad, and other remote consoles provide constant feedback on the status of
the treatment. In the event of an afterloader error condition, the control
software displays a dialog box containing the error code and class, error
description, suggested actions, and optional site-specific notes at the bottom of
the screen. Refer to the “Status and Error Messages” section below for more
information on errors and error conditions. Refer to the “Interrupt a Treatment”
section for information on how to interrupt a treatment.
The movement of the dummy and active wires is displayed in two locations on
the Treatment Delivery Page — in the graphical representation of each channel
in the Channel View Strip and in the graphical representation of the Wire
position in the Treatment Delivery Strip. The Wire position representation is
displayed after you press the Treatment Delivery button. In both locations,
the movement of the dummy wire is displayed in green, and the active wire is
displayed in yellow.
When the dummy or active wire has reached the most proximal dwell position
in the Wire position graphical display, the movement of the wire is also
displayed in the Channel View Strip. If all of the dwell positions for a
particular channel are not visible in the Channel View Strip at once, the display
of the channel shifts to the left or right in the Channel View as necessary to
indicate the current wire position.
When the active source is dwelling at a given position, the dwell point
histogram indicator (bar) in the channel representation of the Channel View
Strip incrementally fills with yellow to graphically indicate the remaining
dwell time. The dwell positions for channels that have been treated are black
in color. Untreated positions are blue in color. Refer to “The Channel View
Strip” section in Chapter 2, “Patient and Treatment Data” for more information
on this strip.
When you begin the treatment delivery, an Initial dummy wire test may be
performed in all programmed channels, depending on the setting for the
Dummy wire sequence in the system configuration options.
There are two options for the Dummy wire sequence:
• Dummy all/ Dummy each/ Active each —sends the dummy wire out on
all programmed channels, and then alternates between the dummy wire
and active wire for each programmed channel
• Dummy each/ Active each — alternates sending the dummy wire and
then the active wire out on each programmed channel
Note: If you have selected the first option, and an error is encountered
in a particular channel during the dummy wire test on all
channels, you must fix the problem and press the START
button to resume the dummy test. The test resumes with the first
used channel in the treatment plan and tests each channel used
in sequence. If multiple errors are encountered, the dummy test
always resumes with the first used channel and tests each used
channel in sequence.
If the dummy wire test is successful, the source wire extends in the channel to 3
deliver the treatment. If the dummy wire test fails, you must clear the cause of
the error and resume the treatment using the Partial Treatment Strip, described
in the “Partial Treatment Strip” section included later in this chapter.
After a successful dummy wire test on a given channel, the active source
extends in that channel and at the same time a short warning tone can be heard.
In addition, the radiation monitor warning light in the treatment room
illuminates. In the control software, the radiation symbol flashes in the
Treatment Delivery Strip. The Wire field in the Treatment Delivery Strip
indicates Active and the Status field indicates Extending, Dwelling, or
Retracting. Refer to the “Dwell Time, Dwell Position, and Wire Status
Information” section below for more information on the indicators at the
bottom of the Channel View Strip.
On the Control Console, the SAFE indicator light in the SOURCE section
extinguishes, and the yellow OUT indicator light illuminates.
Note: The OUT light indicates the extension of the active wire from
its shielded position. This light is activated via mechanical
contacts. The yellow radioactive symbol also indicates an
unshielded source, but for safety reasons, it is activated when
the Geiger-Müller system detects radiation.
WARNING 3-14: If either of the yellow indicators — the OUT light or the
RADIATION symbol — in the SOURCE section of the
Control Console does not light during source extension, a
hazardous failure condition may exist. Interrupt the
treatment immediately by pressing the INTERRUPT
button or the EMERGENCY RETURN switch. Determine
the cause of the failure and eliminate the problem before
resuming treatment. Failure to do so may cause injury to the
patient.
Note: If a deviation of more than 1.4 seconds from the scaled dwell 3
time occurs, the treatment is interrupted and the Partial
Treatment Options are displayed.
• For each channel, a dummy wire test is automatically executed and the
operator is signaled if any obstructions or other errors are encountered.
• If the wire test is successful, the afterloader automatically treats each
channel with the active source wire.
• This process is repeated until:
The Channel View table displays the following information for each channel
in the current fraction:
The bottom of the Channel View Strip includes the following read-only fields:
Note: The Radiation symbol flashes in the Status Strip when the
active wire is extending, dwelling, and retracting.
The control software displays status and error messages at the bottom of the
screen, above the System Status Strip. The following message types are
supported:
• Directives — Displayed when the user needs to perform a task, like “Press
the START button”.
• Status (or information) messages — Displayed for information only, like
“Downloading plan to the afterloader...please wait.”. No interaction is
required by the user.
• Interrupt messages — Displayed when a treatment is interrupted by the
afterloader or the user. Usually, the treatment can be continued by pressing
the START button on the Control Console.
• Emergency messages — Displayed when a treatment cannot be
continued. The treatment can be continued using the Partial Treatment
Strip (refer to the “Partial Treatment Strip” section below).
Directives are usually displayed with a green background, and status messages
are displayed with a beige background. Interrupt and emergency messages are
displayed with a red background. In the event of simultaneous error conditions,
the most severe error is displayed.
Interrupt and emergency messages are initially displayed with the error text
and one or two buttons — View Detail and Clear Error.
Press the View Detail button to see an additional dialog box containing the
error description, error code and class, any possible actions you might take, and
any notes that the administrator has entered using the annotate message feature
of the system configuration options. Refer to the “Annotate Error Messages”
section in Chapter 4 “System Configuration” for more information on how to
customize your error messages.
After you view the message details, press the Done button to return to the
original error dialog box.
All afterloader errors are written to an error log. Other loggable events, such as
user log on, log off, treatment initiation, and wire exchanges are written to the
event log. You can use the Error Log tab in the Console & Afterloader Strip
in the system configuration options to view and print these logs. Refer to the
“Review Error and Event Logs” section in Chapter 4 “System Configuration”
for more information.
TED
Important: Before entering the treatment room, verify that the room
radiation warning light is off, the RADIATION symbols on the
Control Console and afterloader keypad are off, and the
SOURCE SAFE indicator light is on.
In the Patient/Treatment Data section, the Patient Data includes Patient Last
Name, Patient First Name, Patient Identification, Sex, and Patient Birthdate.
The Plan Data includes Fraction, Physician, Treatment Site, Applicator, Plan
File, Plan Author, and Console Entry Date. The Delivery Data includes
Delivery Status, Decay Scale Factor, Operator, Delivery Date, Start Time, and
End Time.
The Planned/ Actual Comparison section includes a table comparing the
Planned and Actual values (if available) for Source Strength, Total Time, and
Total Curie seconds (Ci-s)/GBq-s/Total air kerma strength, indicating the
Difference and Percent Difference for each value. The system units for source
strength are designated in the system configuration options.
The Treatment Details include the Channel Length, Dwell Positions (total),
Planned Time, Scaled Time, Delivered Time, Remaining Time, and a table
showing the dwell Positions, Planned dwell times, Scaled dwell times, and
Delivered dwell times for each channel.
The Source Data section includes the Remaining Active Wire Cycles,
Remaining Source Days, Calibration Date, and Wire Serial number.
If the treatment was interrupted for any reason, an additional Interruption Data
section is included at the end of the report. This section includes Time of
Interruption, Error Class, Error Code, Error Description, Channel, Dwell
Location, Time Spent at Dwell, and Continuation Option.
There is a space for an approval signature and date at the bottom of the report.
Press the Print button to print a hard copy of the report or the Close button to
close the report and return to the Patient Selection Page.
For each subsequent fraction, upon completion of the fraction, the control
software displays a message that asks you if you would like to search for the
appointment that corresponds to the current fraction. If you respond
affirmatively, the control software displays a list of the appointments 3
scheduled for the patient on the day the fraction was completed in an
Appointments table. Select the relevant record in the table and press the OK
button to update the treatment status in the ARIA database.
1
2
23 3
22 4
21 5
6
20
7
19 8
18 9
17 10
16 11
15 14 13 12
WARNING 3-15: In order to assure safety, place the indexer head in the lowest
position before moving the GammaMedplus iX.
Collect all source guide tubes and applicators and place them in a suitable
container for cleaning and sterilization.
Switch the POWER key to the OFF (vertical) position on the Control Console
and remove the key.
Switch the SOURCE key on the afterloader keypad to the LOCKED (vertical)
position and remove the key.
3
For normal storage (after treatment), the battery key stays in ON position to
secure the source and charge the battery. If you need to move the afterloader to
another room, keep the battery key switch in the ON position. For longer
distances (e.g., using an elevator), local regulations must be taken in
consideration. For information on transporting the GammaMedplus iX, refer
to Appendix E, “Transporting the GammaMedplus iX”.
Interrupt a Treatment
You may need to interrupt a treatment in an emergency or in the absence of an
emergency situation, for example, in order to change the position of the patient.
Press the INTERRUPT button on the Control Console.
• Abort this fraction, but distribute the undelivered portion over all
remaining fractions — determines how much of the current fraction
could not be delivered and distributes this amount equally over the
remaining fractions. This option is only available if each of the remaining
fractions is identical to the current fraction. This option is disabled if there
are no remaining fractions when the treatment is interrupted, or if at least
one subsequent fraction does not match the current one.
Important: The option above may have dose and radiobiological effects on 3
the remaining fractions and should only be chosen after careful
consideration of these changes.
Once you choose an option and press the Proceed button, the control software
generates a Treatment Delivery Report that includes an Interruption Data
section. Refer to the “Review Treatment Results” section above for more
information on the Treatment Delivery Report.
If you choose to continue the treatment at a later session, the control software
displays the Patient Selection Page after you close the Treatment Delivery
Report.
Note: The Log Out button is disabled when the Partial Treatment
Strip is open.
The stored data is displayed in an Interruption Data section at the end of the
Treatment Delivery Report. This section includes Time of Interruption, Error
Class, Error Code, Error Description, Channel number, Dwell Location, Time
Spent at Dwell, and Continuation Option. Refer to the “Review Treatment
Results” section above for more information on the Treatment Delivery
Report.
When you press the Clear Error button in the message box, the following
startup message is displayed:
The last patient treatment was interrupted. Log on to the “Open Patient File”
workflow to select a treatment continuation option.
After you log onto the Open Patient File workflow, the Treatment Recovery
Report is displayed, followed by the Partial Treatment Strip. After you choose
a partial treatment option, the Treatment Delivery Report will be generated
documenting the current status of the treatment. Refer to “Review Treatment
Results” section in this chapter for more information on the Treatment
Delivery Report.
The Treatment Recovery Report is divided into the following main sections —
Patient/ Treatment Data and Recovery Data. The Patient Data section includes
Patient Last Name, Patient First Name, Patient Identification, Sex, and Patient
Birthdate. The Plan Data includes Fraction, Physician, Treatment Site,
3
Applicator, Plan File, Plan Author, and Console Entry Date.
The Recovery Data portion of the report includes Start time, Channel number,
Active Wire Position, Time Spent at Dwell, Time Remaining at Dwell, and
Time Spent on Patient.
Note: You cannot deliver a treatment of any kind (patient or test) until
you have selected a partial treatment option.
The control software provides a variety of options that help you configure the
GammaMedplus iX system to meet the needs of your facility. These system
configuration options enable a user with appropriate access privileges to set up
new user accounts, set system preferences, and perform maintenance and other
routine functions. The options also support Varian service personnel in such
tasks as problem diagnosis and wire replacement.
• User & Location — enables you to configure user accounts, group access
privileges, afterloader location data, file location data, and owner
information
• Console & Afterloader — enables you to view the system status, view
error and event logs, annotate error messages, define standard plans,
specify applicator definitions, specify locations for and perform backup
and restore operations, and configure system settings
• Source & Wire — enables the appropriate user to set up and administer
physics test plans, view and edit information regarding the source strength,
and view source and dummy wire information. Service engineers may
specify data about source and dummy wire installations and specify
service engineer parameters.
In This Chapter
Topic Page
Access the System Configuration Page 4-2
Define User Access and Location Information 4-3
Set Up the Console and Afterloader 4-16
Perform Physics and Service Functions 4-38
4-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
The User Info Strip provides read-only data about who is currently logged into
the control software, including the Operator and Location. It also provides
buttons for securing or logging out of the system. Refer to the “Open Patient
File Workflow” section in Chapter 2 for more information on the Secure
System button and to the “System Status Strip” section in Chapter 1 for
information on the System Status Strip.
The User & Location, Console & Afterloader, and Source & Wire Strips are
described in detail below.
The Save and Cancel buttons work the same way throughout the System
Configuration Page.
The Cancel button discards all entries made since opening the current tab. A
confirmation message is displayed before the entries are discarded.
Note: Once you make a change in any tab, you must press the Save or
Cancel button before you can select another tab in the current
strip. 4
If another strip is currently open, press the User & Location button to open the
User & Location Strip.
Define Users
The control software supports controlled access to all product features through
user and group assignments and mandatory passwords for all users. The Users
tab enables you to create new user accounts or modify or delete existing ones.
This tab consists of a sortable table, with each row corresponding to a user
login account, and a set of buttons for maintaining user accounts.
Note: The user, group, location, and password fields are required for 4
each record.
You can click on a record to select and highlight it in the table. Click on the
header of a field to sort the records in ascending or descending order using the
values in the selected field.
The Add button at the bottom of the Users tab enables you to add a new user
to the user table. When you press this button, an empty record appears at the
bottom of the table. Enter the new user login ID in the User field. Each login
ID must be unique for a specific location. Login IDs are not case sensitive.
Press the Enter key or click on the drop-down arrow in the Group field and
select the group to which you want to assign the new user.
By default, there are four types of users defined in the control software based
on the functions they perform and an additional option for specifying Inactive
users:
Note: In addition to the four user groups specified above, there is also
a Service engineer account for providing routine maintenance
and service functions on the GammaMedplus iX system. For
safety reasons, access to and set up of this account is restricted
to Varian service engineers.
Refer to the next section, “Specify Access Privileges for Groups” for
information on how to set group privileges and create a new group.
Click on the drop-down arrow in the Location field and select the treatment
location. Enter a password in the Password field. Passwords must be at least 4
characters and no more than 15 characters in length, with no spaces. Passwords
are not case sensitive.
Note: When you create a new user account, the initial value you enter
in the Password field is visible until you press the Save button.
Following that, the Password field displays only asterisks.
Users with Administrator or Service engineer group
privileges can reinitialize a user's password as needed using this
field.
Press Enter or click on the Password expires field and enter the time limit in
days for the user password. After the time limit expires, the user is required to
enter a new password. Set the expiration to 0 for a password that never expires.
The maximum time limit is 365 days.
4
Delete a User
The Delete button at the bottom of the Users tab enables you to delete the
selected record from the user table. When you press the Save button, the user
is removed from the control software database.
Select a Group
Click on the Group button at the bottom of the tab to display the current list of
groups. Click on a group name to select it.
The selected group appears in the Group name field and the check boxes in
the Allowed field for each record are updated to reflect the privilege
assignments for that group.
You can click on an option to select and highlight it in the table. Click on the
Allowed check box to enable or disable the corresponding privilege. Click on
the header of the Privilege field to sort the privileges in ascending or
descending order.
You can create a new group and assign privileges to it. Press the Add button
and enter a unique name in the Group name field.
Use the Allowed check box for each privilege to enable or disable the
privileges for the new group. Press the Save button to save your changes. The
name of the group appears in the Group drop-down list at the bottom of the
tab. There is no limit on the number of access groups.
Note: You cannot edit privileges or add new privileges to the list.
Delete a Group
The Delete button enables you to delete the selected group. The control
software displays a confirmation message before the deletion occurs.
4
Note: The control software ensures that there is always at least one
group assigned the right to define users and groups. You cannot
delete any of the default system groups — Administrator,
Oncologist, Physicist, or Radiation therapist.
Copy a Group
The Copy button enables you to copy the selected group. Enter a unique name
for the new group in the Group name field. Make any necessary changes to
privileges and press the Save button. The new name appears in the Group
drop-down list.
The Add button at the bottom of the Locations tab enables you to add a new
location to the table. When you press this button, an empty record appears at
the bottom of the table. Enter the name for the location in the Location field.
Each location name must be unique. To add information to the other fields in
the table or to the Location address or Location notes fields, click on the field
and enter the relevant information.
You can click on a record to select and highlight it in the table. The Location
address and Location notes fields will automatically display the data for the
selected location. Click on the header of a field to sort the records in ascending
or descending order using the values in the selected field.
4
Edit Location Information
You can modify an existing location by clicking on the field you want to edit
in the location table and entering the updated information. You can update the
Location address or Location notes by clicking on these fields and entering
new information.
Note: If you change the location name, all users associated with that
location are updated to the new location name.
Delete a Location
The Delete button enables you to delete the selected location. The control
software displays a confirmation message before the deletion occurs.
• File type — the type of file that will be located in the folder
• Alias — the name you want to use to represent the folder location in the
user interface
• Folder — the full path and directory listing for the folder location, such as
G:\Import files (must be a valid location)
• Patient treatment plans — the location where you store the treatment
files you want to import into the control software. The alias for this
location appears when you press the File Location button in the Plan
Import Wizard Strip on the Patient Selection Page. Refer to the “Select or
Import a Treatment File” section of Chapter 2 for more information.
• Treatment plan export — the location where you store exported
treatment files. The alias for this location appears when you press the
Export Location button in the Treatment Record Maintenance Strip on
the Patient Selection Page. Refer to the “Select or Import a Treatment File”
section of Chapter 2 for more information.
4
• Error log export — the location for exported error logs. The alias for this
location appears when you press the Select Log button on the Error Log
tab of the Console & Afterloader Strip on the System Configuration Page.
Refer to the “Review Error and Event Logs” section below for more
information.
• System backup — the location where you store backup files or files you
might want to restore. The alias for this location appears when you press
any of the File Location buttons on the Backup & Restore tab of the
Console & Afterloader Strip on the System Configuration Page. Refer to
the “Back Up and Restore Files” section below for more information.
• Report file export — the location for reports saved in PDF format. The
alias for this location appears when you press the File Location button in
the Save to File area on any Report Page. Refer to the “View and Print
Reports” section in Chapter 2 “Patient and Treatment Data” for more
information.
• Diagnostic file export — the location for exported diagnostic files. The
alias for this location appears when you press the File location button in
the Export diagnostic files area on the Backup & Restore tab of the
Console & Afterloader Strip on the System Configuration Page. Refer to
the “Back Up and Restore Files” section below for more information.
You can click on a record to select and highlight it in the table. Click on the
header of a field to sort the records in ascending or descending order using the
values in the selected field.
The Add button at the bottom of the Folder Setup tab enables you to add a
new folder location to the table. When you press this button, an empty record
appears at the bottom of the table. Select the type of file that will be located in
the folder from the drop-down list in the File type column:
Click on the Alias field to select it and enter the name you want the control
software to use as an alias for this location. For example, you can enter “Import
treatment files” as the alias for the location where you store the patient
treatment plans you want to import into the control software.
Press Enter or click on the Folder field and enter the location for the folder
including the full path and all directories. The path must begin with a drive
letter and colon or two backslashes. Use a valid folder on a network drive or
removable media.
Note: If you export or save a file to the local hard drive (e.g., C:\), only
a service engineer can retrieve it via the Windows operating
system. It is strongly recommended that you use a valid network
drive or removable media for folder locations.
Note: The File type, Alias, and Folder fields are required for each
record.
Click on the field you want to modify and enter or choose the new information.
The Delete button at the bottom of the Folder Setup tab enables you to delete
the selected record from the table.
Click on the field you want to modify and enter the new information.
Note: Only users with service engineers privileges may change the
Machine serial number.
• System Status — enables you to view and print read-only data about the
control PC and afterloader
• Error Log — enables you to view, filter, and print the various errors and
events logged on the GammaMedplus iX system
• Annotate Messages — enables you to add site-specific annotations to
standard afterloader error messages
• Standard Plans — enables you to define, edit, and delete standard plans.
Standard plans are collections of channel and dwell information that can
be applied to patient treatment plans.
• Applicator Definitions — enables you to define, edit, and delete
applicator definitions from the control software database
• Backup & Restore — enables you to define the frequency with which
system data files are backed up and the destination backup drive. You can
also specify parameters for restoring files and restore the backed up files.
• Status item — a list of each status item, grouped by topics: Login status,
General console data, General afterloader data, Current error data, Source
data, Active wire data, and Dummy wire data
• Current value — the value for the status item. The values are derived
from different sources, presenting an up-to-the-minute snapshot of the
current system status.
The Print button at the bottom of the System Status tab enables you to view
and print the System Status Report.
Press the Print button to print a hard copy of the System Status Report and the
Close button to close the report and return to the System Status tab.
This tab consists of a sortable table, with each row corresponding to a logged
error or event, a set of controls for filtering the records in the table, and a set of
buttons for exporting and printing the logs.
There are three different ways to view the logged errors and events:
Refer to Appendix A “System Errors and Messages” for a complete list of error
messages.
4
Click on the Select Log button at the bottom of the tab to display the options
for the types of logs. Click on a log name to select it.
The selected log name appears above the log table and the records are updated
to reflect the selected log.
You can click on a record to select and highlight it in the table. Click on the
header of a field to sort the records in ascending or descending order using the
values in the selected field. By default, the table is sorted in reverse
chronological order on Date occurred.
You can specify a condition for filtering the records in any of the logs to view
a subset of the data in the log. You set up the condition using a field name, a
function (or operator), and a value that you want to match or compare in the
selected field.
Click on the Select Field button to display the list of fields. Click on a field in
the list to select it.
Click on the Select Function button to display the list of functions. Click on a
function to select it.
Click on the Enter value field and enter the value that you want to match or
compare.
The control software automatically applies the filter and displays the filtered
records in the table. To remove the filter, click on the Filter on check box to
deactivate the filter (remove the check mark) and display the full list of records.
Export a Log
4
You can export the currently displayed log table to a comma separated value
(.CSV) file. Press the Export Location button and select the appropriate
export directory from the drop-down list. By default, the Export Location list
displays the most recently used export directory.
Click on the Export button to export the table to a text file. A confirmation
message indicating the name of the export file is displayed when the export is
complete.
Print a Log
The Print button at the bottom of the Error Log tab enables you to view and
print a report of the currently selected log table, including any filter
information.
Press the Print button to print a hard copy of the report and the Close button
to close the report and return to the Error Log tab.
You can click on a record to select and highlight it in the table. Click on the
header of a field to sort the records in ascending or descending order using the
values in the selected field. By default, the table is sorted in ascending order on
Error code.
Click on the error message you want to annotate or edit to select it. Click on
the Notes field and enter the textual description (up to 3 lines). Press the Enter
key on your keyboard to enter the text and then press the Save button.
Refer to the “View and Define Fraction Data” under “Create New Patient File
Workflow” in Chapter 2 for information on how to use a standard plan as the
starting point for a patient treatment file.
The Standard plans table contains the following fields:
• Plan name — the name that was entered when the standard plan was
created
• Treatment site — a drop-down list of treatment sites
Only physicist group users having the Run physics functions privilege can
add, delete, modify, and sort standard plans. Other users may only view this
page.
The New button at the bottom of the Standard Plans tab enables you to add a
new standard plan to the Standard plans table. When you press this button, an
empty record appears at the top of the table. Enter the plan name in the Plan
name field.
Press the Enter key or click on the drop-down arrow in the Treatment site list
and select the type of treatment for the plan:
Click on the drop-down arrow in the Applicator list and select the type of
applicator for the plan:
Click on the Source strength field and enter the planned source strength for 4
the standard plan.
Note: The Treatment site and Applicator are optional for standard
plans.
Press the Save button to save your changes and continue working in the
Standard Plans tab.
Click on the patient treatment plan you want to use to select it in the Patient
plans table. Enter the name you want to use to identify the new standard plan
in the Standard plan name field. Enter the number of the fraction you want to
use in the Fraction field. By default, the value in the Fraction field is the
number of the next fraction to be delivered (or “1” if no fractions have been
delivered) for the selected patient, as displayed in the Delivery status field.
Note: Use the Print button to view the Standard Plan Report and
confirm that the correct fraction has been chosen.
Note: If you are basing the new standard plan on an existing patient
plan, no patient demographic data is included in the new
standard plan.
Press the Save button to save your changes and continue working in the
Standard Plans tab.
Click on a record in the Standard plans table to select and highlight the plan
in the table. Press the Set Up>> button to display the Treatment Delivery Page.
You can view existing channel information in the Channel View Strip or edit
channel information in the Channel Edit Strip. Refer to the “The Channel Edit
Strip” section in Chapter 2 for detailed information on how to use the Channel
Editor.
Press the Save button in the Channel Edit Strip when you are finished adding
or editing the channel(s) of interest. Press the Previous button in the Treatment
Delivery Page to return to the Standard Plans tab. Press the Cancel button in
the Channel Edit Strip to discard your changes and the Previous button to
return to the Standard Plans tab.
Click on the field you want to modify and select or enter the new information
in the Standard plans table.
Delete a Record
The Delete button at the bottom of the Standard Plans tab enables you to
delete the selected record from the standard plan table. The control software
displays a confirmation message before the deletion occurs.
The Print button at the bottom of the Standard Plans tab enables you to view
the Standard Plan Report.
Press the Print button to print a hard copy of the report and the Close button
to close the report and return to the Standard Plans tab. 4
Click on the Applicators button at the bottom of the tab to display a list of the
currently defined applicators. A small set of applicators are included in the
database by default. Click on an applicator name to select it.
The selected applicator name appears in the Applicator name field and the
records in the table are updated to reflect the selected applicator.
The applicator table contains the following fields:
You can click on a record to select and highlight it in the table. Click on the
header of the Channel number and Channel name fields to sort the records
in ascending or descending order using the values in the selected field. By
default, the table is sorted in numerical order on Channel number.
Add an Applicator
The Add button at the bottom of the Applicator Definitions tab enables you
to add a new applicator to the Applicator table. When you press this button,
the control software clears the editable text fields and table so that you can add
the new information.
Enter the name of the applicator you want to define in the Applicator name
field. Enter the name of the manufacturer in the Applicator manufacturer
field.
Press the Add Channel button to add a new channel to the applicator table.
Enter the new channel number in the Channel number field. Press the Enter
key or click on the Channel name field and enter the name for the channel.
The Channel length is fixed at 130 cm for the GammaMedplus iX.
Repeat this process until you have added all of the channels for the applicator.
Note: You cannot define an applicator that uses more than the 24
channels available on the GammaMedplus iX system. 4
If you want to delete a channel, click on the channel in the applicator table to
select it and press the Delete Channel button. The control software displays a
confirmation message before the deletion occurs.
Click on the record you want to edit to select it. Click on the field you want to
modify and enter the new information.
Delete an Applicator
The Delete button at the bottom of the Applicator Definitions tab enables you
to delete the selected applicator from the table. The control software displays
a confirmation message before the deletion occurs.
The Copy button at the bottom of the Applicator Definitions tab enables you
to copy the selected applicator definition. Enter a unique name for the new
applicator definition in the Applicator name field. The new name appears in
the Applicators drop-down list.
The backup procedure includes treatment history files, physics test files,
registry entries, error logs, and other essential data (such as system
configuration options).
Select the destination root folder for the files that are automatically backed up
by pressing the File Location button in the Automatically backup console 4
system files group box and choosing the alias name that represents the folder
from the list:
The folder names that appear in the File Location list are configured in the
Folder Setup tab of the User & Location Strip. By default, the most recently
selected location is displayed.
Click on the Backup frequency (days) text field and enter a number. The
number 1 represents an automatic backup every day, 2 is for every other day,
3 for every third day, etc.
Note: Each time the Start Page displays, the control software checks
to see if an automatic backup is required, and, if so, performs the
backup.
Select the destination root folder for the files that are manually backed up by
pressing the File Location button in the Manually backup console system
files group box and choosing the alias name that represents the folder from the
list.
The root folder contains the backup folder, which in turn contains a series of
folders and associated files. Enter the name of the backup folder in the Backup
file field.
Restore Files
Select the root source folder from which the files will be restored by pressing
the File Location button in the Restore console system files group box and
choosing the alias name that represents the folder from the list.
Select the backup folder by pressing the Backup File button and choosing the
name of the folder from the list.
Press the Restore button to restore the files. The control software displays a
confirmation message before the restore occurs. Select No to cancel the
restore. If necessary, you can take this opportunity to back up the data first.
Select Yes to begin the restore. When the restore is complete, a confirmation
message is displayed. Press OK to restart the control software.
CAUTION: When you press the Restore button, you will see a warning
indicating that all current data and settings are about to be
replaced by whatever is specified in the recovery file. It is
strongly recommended that you perform a manual backup
before you initiate the file recovery. A restore operation cannot
be undone.
You can export control software diagnostic data to a location you specify. The
export of diagnostic files only needs to be performed under the direction of a
Varian Medical Systems service engineer in order to facilitate problem
diagnosis. The exported data can be sent to the service engineer.
Select the destination folder for the files to be exported by pressing the File
Location button in the Export diagnostic files group box and choosing the
alias name for the folder from the list. This location is specified in the Folder
Setup tab of the User & Location Strip.
Press the Export button to export the files. A confirmation message appears
when the export is complete.
The Date and time setting enables you to change the date and time in order to
make allowances for Daylight Savings Time or to synchronize the control PC
clock and GammaMedplus iX afterloader clock if necessary. Double-click on
the Current value field for Date and time and make any necessary changes,
using yyyy-mm-dd format for the date and a 24 hour clock. The updated time
values will be applied to the control PC and the GammaMedplus iX
afterloader.
Note: If the updated time value differs more than 60 minutes from the
original time value, the control software displays a warning that
the new value could affect the accuracy of the current source
calibration. The change will be accepted and logged in the event
log.
Note: The new time value cannot differ more than +/- 75 minutes from
the original time value.
Select a Printer
Double-click on the Current value field for the Selected Printer to display a
drop-down list of the printers that are currently available. Select the printer you
want to use for hard copy reports.
You can select which units to display on screen and in reports for the following
parameters:
WARNING 4-1: The Conversion factor, when specified in the System Units
section, must match the conversion factor used in the
treatment planning system. The control software uses the
conversion factor when scaling the dwell positions. Failure to
match these values may cause the delivered treatment to differ
from the planned treatment.
• Dummy each/ Active each — alternates sending the dummy wire and
then the active wire out on each channel
• Dummy all/ Dummy each/ Active each —sends the dummy wire out on
all channels first, and then alternates between the dummy wire and active
wire for each channel
Double-click on the Current value field for the Dummy sequence and choose
the appropriate dummy wire sequence.
Note: The option you select applies only to patient treatment plans. It
does not apply to physics test treatments. For test treatments, the
setting always defaults to Dummy each/ Active each.
• Test Plans — set up and execute physics test treatments (physicists only)
• Source Strength — enter the calibrated source strength (physicists only)
• Source Exchange — specify data about source and dummy wire
installations (Varian service engineers only)
• Service Parameters — specify data relating to tasks performed by service
engineers (Varian service engineers only)
• System Commands — exit to the operating system and perform other
tests (Varian service engineers only)
WARNING 4-2: Physics test plans must not be used for the treatment of
patients. Doing so may cause serious injury to the patient. It is
essential that you develop and use a fail proof method for
distinguishing test treatment and patient treatment files.
Varian Medical Systems accepts no responsibility if a
physics test plan is delivered to a patient as treatment.
• Plan name — the name of the test plan that was assigned when the plan
was created
• Source strength — the planned source strength
• Absolute value dwell times — when this check box is activated, the
control software does not scale the dwell times to account for the current
source strength; when the check box is deactivated, the control software
adjusts the dwell times to account for the current source strength
• Console entry date — the date the test plan was created
If you have physicist group privileges, you can click on the header of a field to
sort the records in ascending or descending order using the values in the
selected field.
The New button at the bottom of the Test Plans tab enables you to add a new
test plan to the Test plans table. When you press this button, the Physics Test
Plan Setup Strip appears. You can use the options in this strip to create a
fraction with a set of channels having identical dwell characteristics.
• Plan name — the name you want to use to identify this specific test plan
• Planned source strength — source strength to use for the test plan (value
must be between 0.000 and 15.000 Ci). Your test plans should reflect the
source strength that is allowed by your license.
Note: This field is editable only if the check box for Absolute value
dwell times is not activated (not checked).
• Absolute value dwell times — if checked, specified dwell times will not
be scaled to account for the current source strength
• Number of channels — the number of channels for the test plan (value
must be between 1 and 24)
• Channel length (cm) — length is fixed at 130 cm for the
GammaMedplus iX
• Starting dwell position (cm) — the location for the first dwell position
(range of values depends on the Distal Offset and Step Size)
• Number of dwells — the number of dwell positions in each channel (value
must be between 1 and 60)
• Time per dwell (s) — the dwell time for each position (value must be
between 0.1 and 9999.9 sec)
• Step size (cm) — the distance between each dwell position in cm (value
must be between 0.1 and 1.0 cm)
• Distal offset — the offset from the distal position of the source. A value of
0.0 cm represents the full 130.0 cm channel length; a value of 20.0 cm
represents a 20.0 cm offset from the 130.0 cm channel length, or a
110.0 cm distal position (value must be between 0.0 and 40.0 cm)
Press the Save button to save your Test Plan setup and continue working in the
Test Plans tab.
4
When you press the Print button at the bottom of the Test Plans tab, the
Physics Test Report appears.
Press the Print button to print a hard copy of the report or the Close button to
close the report and return to the Test Plans tab.
You can use an existing patient treatment plan as the basis for a new physics
test plan. Click on the Copy Patient Plan button. The control software
displays a table containing the patient treatment records that are currently
active in the patient database.
Click on the patient treatment plan you want to use to select it in the Patient
plans table. Enter the name you want to use to identify the new test plan in the
Test plan name field. Enter the number of the fraction you want to test in the
Fraction field. By default, the value in the Fraction field is the first
undelivered fraction, or the last fraction (if all fractions have been delivered),
as displayed in the Delivery status field.
Note: If you are basing the new test plan on an existing patient plan,
no patient demographic data is included in the new test plan.
Press the Print button to print a hard copy of the report or the Close button to
close the report and return to Patient Plans area of the Test Plans tab.
Press the Save button to save your changes and continue working in the Test
Plans tab. Press the Cancel button to return to the Test Plans tab without
saving your changes.
In order to execute a test treatment, press the Run Plan button at the bottom of
the Test Plans tab to display the Treatment Delivery Page. The control
software prominently displays the following warning message near the top of
this page:
The workflow on the Test Plan Initiation and Test Plan Delivery Strips on the
Test Plan Delivery Page for physics test plans is similar to the workflow for
patient treatment plans, with the following exceptions:
1. When you press the Start Test button, a Planned Treatment Report is not
generated. The dwell times are adjusted using the current source strength
only if you have deactivated the Absolute value dwell times check box in
the Test plans table on the Test Plans tab.
2. In the Test Plan Delivery Strip, enter your logon Password instead of a
Treatment Code.
WARNING 4-4: All persons must leave the treatment room before the
delivery of a physics test plan begins.
In the Test plans table, click on the test plan you want to edit to select it. Click
on the field you want to modify and enter the new information. Use the
Channel Editor to edit channel and dwell data for the test plan. You can access
the Channel Editor by pressing the Run Plan button on the Test Plans tab. For
more information on editing a channel, refer to the “The Channel Edit Strip”
section in Chapter 2 “Patient and Treatment Data”.
The Delete button at the bottom of the Test Plans tab enables you to delete the
selected test plan from the table. The control software displays a confirmation
message before the deletion occurs.
Note: You must enter a Site calibrated source strength value after a
source exchange before delivery of any patient treatment in
order to meet calibration requirements set forth by the USNRC
and other local Competent Authorities. After a source
exchange, the control software displays a message to this effect
when you log on until you enter a site calibrated source strength.
Important: Use the Site calibrated source strength as the base line source
strength to calculate the current source strength with a source 4
decay table. Refer to the “Source Decay” section in Appendix B
“GammaMedplus iX Calculations” for more information. The
Site calibrated source strength is the value that must be
entered in the BrachyVision treatment planning software.
To start this procedure, turn the MODE key switch on the Control Console to
the horizontal position. The SPECIAL indicator light turns on.
The control software decays the source by one day at 12:00 midnight according
to the formula in the “Source Decay” section of Appendix B “GammaMedplus
iX Calculations”. An Ir-192 Source Decay Table is also included in
Appendix B.
You can use the calibrated source strength and calibration date on the
manufacturer’s certificate or enter the source strength determined during your
own source strength calibration. Enter the value of the source strength in the
Site calibrated source strength field in units of either Ci, U, or GBq,
depending on the System units setting for the Source strength in the System
Settings tab of the Console & Afterloader Strip. Enter the Calibration date
that corresponds to the source strength in yyyy-mm-dd format. Enter your
logon Password and press the Save button to save the calibrated source
strength.
The control software checks to see if the new source strength is within an
acceptable range, as follows:
Once accepted, the new source strength is used as the basis for the decay
calculations. This change is logged in the event log with the current user’s
logon ID. The Current source strength in the System Status Strip is
recalculated to reflect the new calibration. Relevant printouts and system
information will indicate the current source strength.
When finished, return the MODE key switch to the vertical (NORMAL)
position, remove the key, and place it in a secure location.
This tab consists of two tables, one for active wires and one for the dummy
wires. There is also a set of buttons for maintaining the tables.
Note: The record for the most recently installed active and dummy
wire appears at the top of the respective table. The record of the
wire currently in use is highlighted.
When you press the Print button at the bottom of the Source Exchange tab,
the Wire Exchange Report appears. The top section of the report includes the
user logged on at the time the report is printed, clinic (institution name),
afterloader serial number, and the date and time the report is printed.
The report is divided into the following main sections — Maximum Strength
Values, Active Wire Exchange History, and Dummy Wire Exchange History,
with each section displaying the information contained in the active and
dummy wire tables and group box.
Press the Print button to print a hard copy of the report or the Close button to
close the report and return to the Source Exchange tab. 4
Click on the Wire Type button at the bottom of the tab and select the type of
wire whose data you wish to view (active or dummy).
The Physical field in the Maximum source strength values group box
displays the value for the maximum source strength for which the afterloader
is designed. The At exchange edit field displays the maximum source strength
that can be installed in the afterloader as allowed by the USNRC, Agreement
State, or Competent Authority. The At treatment field displays the maximum
source strength that can be used at treatment as allowed by the USNRC,
Agreement State, or Competent Authority. The service engineer specifies the
value for the At exchange and At treatment fields.
When you press the Print button at the bottom of the Service Parameters tab,
the Service Parameters Report appears. The top section of the report includes
the user logged on at the time the report is printed, clinic (institution name),
serial number, and the date and time the report is printed.
The report is divided into two main sections — Wire Drive – Specific
Parameters and Wire Cycles. Each section displays the cycles for the active
and dummy drives and wires, respectively.
WARNING 5-1: After an emergency, the physical condition of the source wire cannot
be known unless it is inspected by a Varian Medical Systems service
engineer. If there is any possibility that a GammaMedplus iX source
wire has been damaged, you must contact Varian Medical Systems
before treatment may resume. A damaged wire may fail to properly
track through the GammaMedplus iX unit.
In This Chapter
Topic Page
5-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
For all sites outside of the United States and Canada, please contact your
regional service office, as applicable.
Definition of Emergency
In the context of this document, an emergency is defined as a radiation hazard
to the operator or patient that occurs as the result of a system malfunction or an
error in a treatment sequence. In such an emergency, you must follow the
sequence below without skipping a step:
Precautions
Limitations
The procedures recommended in this chapter may or may not encompass every
emergency you may encounter. For example, you may experience an
emergency situation resulting from a fire or earthquake, neither of which is
addressed in this chapter. This information is provided as guidance only.
Formal emergency procedures must be adopted by your facility in compliance
with your license and local regulations.
Prerequisites
All of the safety features in the “Important Safety Information” section of
Chapter 1, “Introduction” must be satisfied in addition to the emergency
equipment and radiation safety officer prerequisites stated below. Only
properly trained and authorized persons should operate the HDR unit. All
personnel working with or around the HDR unit must be trained in radiation
safety.
Emergency Equipment
The radiation safety officer is responsible for ensuring that all radiation safety
procedures and policies are appropriate. The RSO should have proper training
in the operation of the HDR unit and should actively participate in the initial
and annual practice of HDR emergency procedures.
Alarm Indicators
When an error condition occurs, the red ALARM indicator in the ALARM
section of the Control Console flashes, and an audible warning signal can be
heard.
ALA
ALARM
ARM
On the control PC, a red message box at the bottom of the screen indicates an
error condition and the possible cause of the alarm.
ALA
ALARM
LARM
ARM
DOS
ZER
FF
Check the error message on the control PC for an indication of the possible
cause of the alarm, and resolve the problem. Press the View Detail button in
the error message box to see additional information. Refer to the “Status and
Error Messages” section in Chapter 3 “Delivery of Patient Treatment” for
more information.
WARNING 5-3: If the SAFE indicator light is not lit, and the RADIATION
indicator is on, follow the emergency procedures at your
site. Refer to the “Automatic and Emergency Retraction of
the Source” section below for more information.
Note: If you do not correct the cause of the alarm, the alarm signal is
repeated when you press the START button.
Radiation Indicators
Indication of radiation from an unshielded source is detected and displayed by
multiple methods on the GammaMedplus iX system.
On the Control Console, the yellow SOURCE OUT indicator and the yellow
RADIATION symbol illuminate when the source is outside the shielded safe.
5
Independent Radiation Monitoring System
There are situations in which you might need to manually interrupt a treatment,
such as an emergency that requires immediate entry into the treatment room,
or in the unlikely event that automatic source retraction does not occur when
necessary.
You can initiate an emergency retraction of the source wire using the following
methods:
If emergency retraction of the source is not possible using the actions stated
above, manual recovery of the source must be performed. Manual recovery
means removing the source from the patient without the use of the
INTERRUPT button, EMERGENCY RETURN switch, EMERGENCY
button, or handcrank. Emergency retraction and manual recovery procedures
are outlined in Figure 5-1 and described in detail in the sections below.
retracted RETURN
switch. SWITCH
If source still
not retracted
EMERGENCY
button
Enter treatment room
with survey meter. DUMM
DUMM
Y-CHAN
Y-WEC
GE
HSEL
If source still
not retracted
EMERGENCY RETURN
Switch
ALA
ALARM
ARM
INTERRUPT Button
WARNING 5-5: Verify that the SAFE indicator light in the SOURCE
section of the Control Console is on, and the RADIATION
symbol is off. If not, implement the next step of the
emergency response procedures as described in Figure 5-1
above.
When you press the EMERGENCY button on the afterloader keypad, the
GammaMedplus iX does the following:
WARNING 5-7: When you press the EMERGENCY button on the afterloader
keypad, the external power supply of the GammaMedplus iX
is switched off, and the source is retracted at maximum speed.
After that, all remote monitors, including the Control Console
and on-board radiation monitoring, are also turned off. The
only indication whether the source is in the shielded position
is the independent radiation monitoring system installed in the
treatment room and your portable survey meter. Confirm that
radiation levels are at background level. If not, initiate the next
step of the emergency response procedures as described in
Figure 5-1.
Failure to follow these procedures may result in the following:
• Source may still be in the patient and you will not be aware
of this.
5
• Source may be outside the shielded position and will
expose the operator and patient to high levels of external
radiation.
In the unlikely event that emergency retraction of the source does not occur
after pressing the INTERRUPT button, the EMERGENCY RETURN
switch, or the EMERGENCY button, the source can be retracted manually
into the shielded position using the handcrank.
Follow these steps to manually retract the source using the handcrank:
1. Enter the treatment room with a portable survey meter and go straight to
the GammaMedplus iX.
2. Pull the handle out of the handcrank and turn the crank in the direction of
the arrow until a distinct resistance is noticeable and the room radiation
level is safe.
5
Note: If the source wire has been retracted and the radiation level is
still high or if manual retraction of the source is not possible
using the handcrank, follow the procedures in the “Perform
Manual Recovery of the Source” section below. These
procedures are also outlined in Figure 5-1.
1. Enter the treatment room with a portable survey meter and personal
dosimeter.
2. Calm the patient.
3. Open the Emergency Container.
4. Remove the applicator carefully from the patient using the long handled
forceps whenever possible. Maintain the maximum possible distance from
the patient and the afterloader.
Note: In all cases above, remove the patient from the immediate area,
survey the patient, and, if safe to do so, evacuate the patient
from the room. Post a warning and notify all emergency
contacts immediately. Since, in most jurisdictions, you will be
expected to report on estimated exposure to staff and patient as
a result of the incidents, you would be advised to estimate times
and dose rates involved to produce a reasonable estimate of
exposure to all concerned.
Follow these steps to restart the system after interrupting the treatment:
The GammaMedplus iX resumes the treatment from the last dwell position,
and only the remaining dwell positions in the current fraction will be delivered.
Important: In the case where you have pressed the INTERRUPT Button
and need to stop and resume the treatment at a later time, you
can press the EMERGENCY RETURN switch, the
EMERGENCY button, or turn the POWER key switch on the
Control Console off and then back on (after waiting sufficient
time for the system to shut down). Each of these methods causes
the control software to display the Partial Treatment Options
where you can choose to abort the treatment.
Follow these steps to restart the system after pressing the EMERGENCY
button:
• System installation
• Annual maintenance and service visits 6
• Source and dummy wire exchange
• System check after disruption of service
Important: You must perform a backup of all control software data files
before your service engineer arrives on site. Refer to the “Back
Up and Restore Files” section in Chapter 4, “System
Configuration” for more information.
In This Chapter
Topic Page
System Installation 6-2
Annual Maintenance and Service Visits 6-2
Source and Dummy Wire Exchange 6-3
System Check After Disruption of Service 6-4
6-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
System Installation
The installation and initiation of operation of the GammaMedplus iX system
may only be performed by a qualified Varian service engineer according to the
guidelines set forth in the GammaMedplus iX Service and Maintenance
Manual. After the installation, the service engineer checks the entire
GammaMedplus iX system for functionality and safety using a test protocol.
The protocol for the annual maintenance includes such tasks as removing the
source wire, removing the dummy wire, removing and inspecting the source
and dummy drives (replacing them if necessary), checking guide tubes,
checking mechanical limit switches, checking the handcrank, checking the
indexer, checking the chassis, performing a functional test of the control PC,
running the ServPlus test program, testing safety features, performing a
radiation survey, performing a test treatment, checking the independent
radiation monitoring device, measuring the operating voltages, and making a
list of damaged and replaced parts.
Note: There are no user serviceable parts. Please contact your local
Varian service representative in the event of a system failure.
• Verify the current system status by reviewing the error logs, system
parameters and system settings
• Perform system tests to check the system performance prior to
maintenance activities
• Perform higher level maintenance as applicable
• Install a new source wire
• Complete the appropriate radiation surveys
• Calibrate the active source wire position
• Prepare the old source for return shipment
• System handover: source documentation and field service paperwork
6
Source and Dummy Wire Exchange
A new sealed source is recommended every three to four months due to the
half-life of Iridium-192 (about 74 days). The source exchange must be
performed by Varian Medical Systems service engineers licensed by the
USNRC or by those persons trained by Varian and having a radioactive
materials license to do so. Only source wires delivered from Varian Medical
Systems may be used. A new source wire is delivered with a source certificate
that states the source strength (Air Kerma Rate at a reference date) and attests
that the source is leak free.
Note: When performing a leak test, wipe only the surfaces that have
made contact with the source. Such surfaces are the interior
areas of applicators and source guide tubes.
The used source wire is returned to the manufacturer along with the certificate
of return. For safety reasons, the maximum duration of use for the source wire
is limited to 5000 cycles.
The control software enables the service engineers to specify or review the
following data about a new source wire: serial number, source strength,
exchange date, maximum number of wire cycles, used number of wire cycles,
and comments.
For safety reasons, the maximum duration of use for the dummy wire is limited
to 7000 cycles or a maximum of 2 years regardless of cycles. In the case of
visible wear, external damage, or kinks, the dummy wire must also be
replaced. Only dummy wires delivered from Varian Medical Systems may be
used. After a dummy wire exchange, the distal position of the dummy wire
must be inspected.
The control software enables the service engineers to specify or review the
following data about a new dummy wire: serial number, exchange date,
maximum number of wire cycles, used number of wire cycles, and comments.
This chapter describes features and procedures that are applicable to the
GammaMedplus 3/24 iX system only. Except in the areas described below, the
GammaMedplus 3/24 iX does not differ in its basic operation and function
from the 24-channel system. These operations and functions are described in
detail in the other chapters in this user guide.
In This Chapter 7
Topic Page
System Description 7-2
The Channel View Strip 7-3
Insertion of the Applicator into the Patient 7-4
Source Guide Tube Connection 7-4
Transporting the 3/24 7-6
7-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
System Description
The GammaMedplus 3/24 iX remote-controlled afterloader is equipped with
an Iridium-192 (Ir-192) source that is fixed to a stainless steel cable and moved
into an applicator or catheter inserted into the patient. The source can be moved
into 3 of 5 available channels in succession via the indexer head. In each of
these channels, the source may be programmed to stop in up to 60 dwell
positions for individual dwell times of 0.1 to 9999.9 seconds. The applicator in
the patient is connected to the afterloader via one or more source guide tubes.
Indexer head
Chassis
Trailing wheels
Swivel wheels
You can click on a channel to select and highlight it in the Channel View Strip.
After a channel is selected, you can use the Up and Down arrow keys on the
keyboard to move up and down through the channels and select them. The left
and right scroll buttons enable you to scroll horizontally through a channel
when necessary.
Refer to the “View and Edit Channel Data” section in Chapter 2 “Patient and
Treatment Data” for detailed information on the Channel View and Channel
Edit Strips.
Note: Before you connect the source guide tubes to the indexer head,
use the length gauge to check the correct source guide
tube/applicator probe combination length. Refer to the
instructions for use supplied with the source guide tubes and
applicator for detailed information.
Connect each source guide tube or applicator to the correct channel of the
indexer head according to the channel number determined in the treatment
planning software. Using slight pressure, insert the coupling of each source
guide tube into the channel opening of the indexer head until the coupling is
engaged.
The source guide tubes must be clean and free of foreign bodies and moisture.
Water vapor, especially on the inside of the tubes, can lead to damage of the
afterloader.
Free passage of the active wire through the source guide tube and applicator 7
may be impeded if a tube that is kinked, crushed, stretched, or damaged in any
other way is used. Source guide tubes should be inspected prior to use, and if
they show any signs of defects, they should be discarded.
Refer to the “Prepare the Patient, Afterloader, and Control Software” section
in Chapter 3 “Delivery of Patient Treatment” for more information on setting
up the GammaMedplus iX 3/24 iX system for patient delivery.
This chapter provides a list of critical afterloader and control PC errors and
messages for the GammaMedplus iX. The control software displays error
messages at the bottom of the screen, above the System Status Strip. The
messages are divided into the sections listed in the table below. Refer to the
“Status and Error Messages” section in Chapter 3 “Delivery of Patient
Treatment” for more information on system errors.
WARNING A-1: Always survey the radiation level surrounding the patient
and in the treatment room after a system failure to ensure
that there is no radiation hazard. Failure to do so may result
in serious injury to the operator or patient. If radiation is
detected, proceed with the Emergency Response procedure
as outlined in Chapter 5 “Emergency Procedures”.
In This Chapter
Topic Page
Afterloader Errors and Messages A-2
A
Control Software Errors and Messages A-6
A-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
02 Interrupt DOOR Door of the treatment room Close door or activate Last Man
opened during treatment Out system
03 Interrupt DOSE External dose measurement unit Correct the error and reset the
has exceeded a set warning dose measurement unit
threshold
06 Battery voltage is less Battery voltage too low in order Charge the battery via AC power,
than 20 V to start the treatment check the charge indication; verify
the battery key switch position is
ON
07 Mains voltage is missing Interrupted AC power supply of Check the mains cable
GammaMedplus
11 Limit switch ESOQ defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
12 Limit switch ESOD defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
13 Limit switch ESUQ defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
14 Limit switch ESRQ defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
15 Limit switch ESRD defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
16 Emergency drive defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
18 Limit switch ESUD defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
19 Dummy drive overcurrent Electrical or mechanical fault Call Varian BrachyTherapy service
in channel [x]
20 ESOD not reached Electrical or mechanical fault Call Varian BrachyTherapy service
21 ESRQ not reached or Electrical or mechanical fault Call Varian BrachyTherapy service
spring defective
22 ESRD not reached Electrical or mechanical fault Call Varian BrachyTherapy service
23 Limit switch ESVQ defec- Electrical or mechanical fault Call Varian BrachyTherapy service
tive
24 Broken dummy Electrical fault or broken tip of Call Varian BrachyTherapy service
the dummy wire
25 Source pushed in Source guide tube kinked, length Place source guide tube without
channel [x] at [y] cm of source guide tube too small kinks, pay attention to correct
length of source guide tubes
50 Incorrect program number Error while transferring the treat- After restarting the system, restart
ment plan the treatment
51 Expected character incor- Error while transferring the treat- After restarting the system, restart
rect ment plan the treatment
A
52 Wrong page number Error while transferring the treat- After restarting the system, restart
ment plan the treatment
53 Transmission test failed Error while transferring the treat- After restarting the system, restart
ment plan the treatment
54 Incorrect instruction code Error while transferring the treat- After restarting the system, restart
ment plan the treatment
56 Incorrect checksum in Error while transferring the treat- After restarting the system, restart
channel [x] ment plan the treatment
57 Incorrect cycle in page 00 Error while transferring the treat- After restarting the system, restart
ment plan the treatment
58 Incorrect factor in page 00 Error while transferring the treat- After restarting the system, restart
ment plan the treatment
0b Dummy pushed in Source guide tube kinked, length Place source guide tube without
channel [x] at [y] cm of source guide tube too short kinks, pay attention to correct
length of source guide tube
0c Dummy drive defective Electrical or mechanical fault Call Varian BrachyTherapy service
0d Source drive overcurrent Electrical or mechanical fault Call Varian BrachyTherapy service
in channel [x]
0e Source drive defective Electrical or mechanical fault Call Varian BrachyTherapy service
1a/1b Wrong distance measure- Electrical or mechanical fault Call Varian BrachyTherapy service
ment source/dummy
1c Source travel time error Source needs more than 9 sec Call Varian BrachyTherapy service
to reach distal position
1d Dummy travel time error Dummy needs more than 9 sec Call Varian BrachyTherapy service
to reach distal position
1e Key switch head LOCKED SOURCE key on afterloader Turn key switch to RELEASED
keypad in LOCKED position position
1f ESOQ not reached Electrical or mechanical fault Call Varian BrachyTherapy service
5a Fault in real-time clock or Incorrect check sum in the Please recreate the treatment plan
its RAM memory or hardware defect
5b Fault in data transmission Error while transferring the treat- After restarting the system, restart
ment plan the treatment
5c No echo received Interrupted data transfer from After restarting the system, restart
PC to GammaMedplus the treatment
5d Waited too long for recep- Interrupted data transfer from After restarting the system, restart
tion PC to GammaMedplus the treatment
5e Incorrect operation mode Irradiation was executed in the Use new treatment plan
wrong mode (HDR <> PDR)
a1 RADIATION After the end of a treatment, Check to see if the source is in the
radiation is detected shielded position and follow the
EMERGENCY RESPONSE
instructions
a2 EMERGENCY MANUAL EMERGENCY RETURN switch After restarting the system, restart
pressed the treatment
a3 EMERGENCY MANUAL Emergency button on after- After restarting the system, restart
FROM TROLLY loader keypad pressed the treatment
a4 TIMING ERROR; no GammaMedplus did not receive Check cable connection between
response from PC. Check a signal as expected from the GammaMedplus and control PC;
all data and power control PC synchronize clocks on Gamma-
connections. Medplus and control PC; verify
that power to the GammaMedplus
is on
a6 CONSOLE DEFECTIVE One or more remote consoles is Call Varian BrachyTherapy service
defective
a7 POWER OFF (Console) Power outage on control PC Switch on control PC and refer to
DURING IRRADIATION during treatment delivery the “Continue Treatment after
Control Software Failure” in
A
Chapter 3
a8 Treatment is aborted by Under certain circumstances, You will be prompted to turn the
control PC command ‘@’ the control software will issue an afterloader off and then on at the
abort treatment command to the key switch. The control software
afterloader will display the treatment recovery
options. Check the Treatment
Delivery Report for the cause of
the error and, if necessary, call
Varian BrachyTherapy service.
c8 Afterloader power outage The afterloader has been turned If necessary, turn the afterloader
detected - check key off at the key switch on.
switch
e8 Afterloader has attempted Afterloader firmware or hard- Compare the scaled dwell time
to exceed scaled dwell ware error. against the delivered dwell time for
time by more than one all positions in the Treatment
second Delivery Report and make any
necessary adjustments in the
treatment plan. Call Varian
BrachyTherapy service.
e9 Afterloader has reported Afterloader firmware or hard- Compare the scaled dwell time
dwell complete while one ware error. against the delivered dwell time for
second or more is still all positions in the Treatment
remaining Delivery Report and make any
necessary adjustments in the
treatment plan. Call Varian
BrachyTherapy service.
A patient with this ID already This message appears in the Create Refer to the “Create New Patient File
exists in the database. A new New Patient File workflow if you try Workflow” section in Chapter 2
course of treatment will be to create a new patient with a “Patient and Treatment Data” for more
created for this patient. patient ID that matches a patient_ID information.
Continue? already in the database.
Are you sure you want to The Conversion factor, when spec- Select YES to confirm the value you
change the conversion factor? ified in the System units group box of entered. Select NO to verify the value
Remember that this value the System Settings tab, must match and re-enter it.
should equal the one used in the conversion factor used in the
your planning system. treatment planning system.
At least one channel cannot be This message appears when initi- You can delete the channel using the
delivered, because all dwell ating treatment delivery if the sum of Channel Editor and try again, or
times scale to a value less all the dwell times in a given channel generate a new plan via the treatment
than the minimum supported is zero (after scaling). planning system.
by the afterloader.
At least one dwell time was This plan specifies small dwell times Select YES to continue with the current
scaled to a value less than the that are converted to zero when dwell times or NO to cancel the opera-
minimum supported by the scaled to match the current source tion.
afterloader. These values will strength.
be converted to zero.
Continue?
Battery RAM access failed - This message may appear if a Check all connections with the after-
Could not retrieve power fail communication error occurs while loader, including the serial cable on
data. Console reboot recom- the control software is reading power back of the control PC. If necessary,
mended. failure data from the afterloader. reboot the control software.
Cannot load the failing fraction While recovering from a power It is recommended that you answer
[fraction number]. Clear power failure, the control software tries to NO, and call Varian BrachyTherapy A
fail data from the afterloader? determine the fraction number of the service because the message indi-
plan being delivered. If it cannot find cates a possible corruption in the treat-
the corresponding fraction number in ment history file.
the treatment history file for the plan,
this message will result.
Changing the origin will shift all This message appears if you change Select YES if you want to continue or
dwell positions in channel the origin using the Origin field or NO to cancel the operation. Refer to
[channel number]. Do you the Origin button (and IN/OUT the “The Channel Edit Strip” section in
want to change the origin? buttons on the afterloader keypad) in Chapter 2 “Patient and Treatment
the Channel Editor. Data” for more information.
Changing the step size will This message appears if you enter a Refer to the “The Channel Edit Strip”
clear all dwell positions in new Step size in the Channel Edit section in Chapter 2 “Patient and
channel [channel number]. Do Strip. Treatment Data” for more information.
you want to change the step
size?
Could not reset the error. This message may appear after an After you have been logged off, restart
Treatment will be suspended, unsuccessful restart of the after- the afterloader. Log back into the Open
and you will be logged off. loader following an interruption in Patient File workflow, and you will be
treatment. guided into the Partial Treatment
options.
In certain localities, including If you change the Location on the Recalibrate the source strength. The
the US, regulations require User Login Page, the control soft- control software does not check to
that the afterloader be recali- ware displays this message indi- ensure the recalibration has been
brated when it is moved to a cating that the source strength may completed. For countries other than
new location. Please check need to be recalibrated. If the after- the United States, check the regula-
with your Radiation Safety loader is moved to any location tions set forth by the local Competent
Officer if you are uncertain of outside the facility, the USNRC Authority.
the requirements for your requires that prior to patient treat-
area. ment a full calibration be conducted.
Each time the unit is relocated, it
must undergo full calibration.
Patient [first/last] name does This message appears in the Plan Refer to the “Import a Patient Treat-
not match the corresponding Import Wizard if there is a mismatch ment File” section in Chapter 2 “Patient
value in the database for this in patient first/last name. and Treatment Data” for more informa-
Patient ID. Use this new value tion.
from now on?
The afterloader family-model When the control software boots up, Call Varian BrachyTherapy Technical
is invalid, or the description file it checks the registry. If indetermi- Support
for it could not be found. nate data is returned, this message
Please contact Varian Tech- appears in a Windows dialog box.
nical Support for assistance.
When you click on 'Ok', the
console application will shut
down.
The afterloader is disabled. This message appears if the EMER- Pull the EMERGENCY RETURN
Check the emergency stop GENCY RETURN SWITCH is SWITCH out and press the START
button. pressed down when you try to deliver button.
a treatment.
The afterloader's active wire There is a mismatch in the wire serial Call Varian BrachyTherapy service
does not match the console number, possibly due to swapping of
computer's. the old control PC with a new one.
The console application The control software is attempting to Follow the response as designated in
cannot start. There is a serious start up and there is an error or the error message.
problem with the console data- possible corruption in the database.
base (error code nn). Please
call Varian Technical Support.
Press No to attempt a reboot
with the current database.
Press Yes to replace the
current database with an
empty database. If you have a
recent system backup, you
should attempt to restore it
using the Backup & Restore
tab.
The console can not communi- The communications link between Check the cable connections and/or
cate with the afterloader. the control PC and and the after- turn on the afterloader.
loader is defective or detached; or
the afterloader is turned OFF.
The current source strength This message appears if the control Call Varian BrachyTherapy service
exceeds the value allowed for software determines that the source
patient treatment (call Varian). strength exceeds the allowed value.
The entered time is more than This message may appear when you Refer to the “Set the System Date and A
[n] minutes different from the are entering a new system time in Time” section in Chapter 4 “System
current time. This may affect the System Settings tab. Configuration” for more information.
the accuracy of the source
calibration.
The following components are This message indicates that the Call Varian BrachyTherapy service
missing from the current instal- control software cannot find one or
lation: [list of components]. more files necessary for safe opera-
Please contact Varian Tech- tion. The message should never
nical Support for further assis- appear.
tance. When you click on 'Ok',
the console application will
shut down.
The new source strength This message appears if the entered Select YES to continue with the source
differs by more than [n] source strength is 5-10% greater strength as entered. Select NO to enter
percent from the manufac- than the Manufacturer measured a new value.
turer's value. Continue? source strength.
The plan contains dwell This message appears if you make Select YES if you want to save the
changes. Please note that changes to dwell positions or dwell changes to the dwell positions or NO to
these changes only apply to times in the Channel Editor. cancel the operation. Refer to the “The
the current fraction. Do you Channel Edit Strip” section in
want to save these changes? Chapter 2 “Patient and Treatment
Data” for more information.
The source either needs to be When a new source wire is installed, Calibrate the source and enter the new
site calibrated, or the manu- it must be calibrated before you can value in the New site calibrated
facturer measured value use it for a patient treatment. source strength field of the Source
should be confirmed. Strength tab.
The total scaled dwell time for The total dwell time has exceeded Modify the dwell times to within range
the plan is [n] seconds, which 99999.9 seconds. using the Channel Editor or treatment
exceeds the maximum planning system.
allowed [nn].
The treatment history file for This message appears if you select a Call Varian BrachyTherapy technical
this patient could not be found. patient from the Patient Selection support for a recovery procedure.
Do you want to remove this list, and the underlying treatment
record from the database? history file can not be found. It
means that the treatment history file
was deleted or moved without using
the control software. It should not
happen under normal usage circum-
stances.
The wire serial number in the This message could appear if a As the message indicates, cancel the
afterloader has changed since different afterloader is connected to treatment delivery and re-initiate the
the plan was initiated. Please the control PC after a treatment is treatment.
cancel delivery and re-initiate. initiated but before the treatment
delivery begins.
There are not enough This message appears if there are Call Varian BrachyTherapy service to
remaining dummy or active insufficient wire cycles when you install a new dummy or active wire; or
cycles to perform this treat- attempt to initiate a treatment. possibly a new motor drive.
ment.
This patient has [n] plans with You have selected a patient in the Use BrachyVision™ to assign a unique
the same Plan ID. To import Plan Import Wizard that has multiple Plan ID to the plan. In the control soft-
this plan, please use your plans with the same Plan ID number. ware, press the Cancel Import or
treatment planning system to You cannot import the selected plan Previous button on the Plans Strip of
give this plan a unique Plan ID. without first assigning a unique plan the Plan Import Wizard and proceed
identifier to the plan. with the plan import process.
Treatment file for the inter- This message can appear if the If you answer YES, the last-treated
rupted plan could not be attached afterloader’s on-board record will be deleted from the after-
found. Reset power failure memory indicates that the last treat- loader. If you answer NO, the record
data? ment was interrupted, but the corre- will be kept on the afterloader – but no
sponding treatment history file does patient treatments will be allowed on
not exist in the control software. the afterloader until the record is
cleared.
Unable to set the source This message appears if you attempt Enter a value in range.
strength. The new site cali- to enter a New site calibrated
brated source strength range source strength that is out of range
is between [n] and [nn]. of the values in parentheses to the
right of the field.
Unable to retrieve patient The communication link between Check the cable connections and the
appointments. ARIA and the control PC is defective communications settings.
or detached. There might be an error
in the communications settings for A
ARIA or the control PC, specifically
in the worklist service settings.
Unable to retrieve plan. The communication link between Check the cable connections and the
ARIA and the control PC is defective communications settings.
or detached. There might be an error
in the communications settings for
ARIA or the control PC, specifically
in the database service settings.
Unable to retrieve patients. The communication link between Check the cable connections and the
ARIA and the control PC is defective communications settings.
or detached. There might be an error
in the communications settings for
ARIA or the control PC, specifically
in the database service settings.
Validation Error: File The treatment plan file is corrupted. Recreate the plan in the treatment
checksum is invalid. planning system.
Validation Error: Missing The treatment plan file is corrupted. Recreate the plan in the treatment
end-of-channels indicator. planning system.
Validation Error: Missing The treatment plan file is corrupted. Recreate the plan in the treatment
end-of-file indicator. planning system.
Validation Error: Parse error The treatment plan file is corrupted. Recreate the plan in the treatment
encountered while reading the planning system.
file.
Validation Error: The sum of all This message may appear when Recreate the plan in the treatment
dwell times in channel importing a treatment plan. It indi- planning system.
[number] is declared to be [n], cates a file corruption.
but is actually [nn].
Appendix B GammaMedplus iX
Calculations
This appendix contains information about the calculations in the table below.
In This Chapter
Topic Page
Scaled Dwell Time B-2
Source Decay B-3
Transit Dose B-5
B-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Ts = ( Sp ⁄ Sc ) × Tp
where:
Ts = scaled time
Sp = planned source strength
Sc = current source strength
Tp = planned time
Source Decay
The control software calculates source decay once each day at 12:00 midnight
according to the formula below:
( – λt )
S = So e
where:
S = current source strength
So = known source strength (from
manufacturer)
e = exponential (~2.7183)
λ = (ln 2)/T 1 ⁄ 2 (where T 1 ⁄ 2 = 73.83 days)
t = days since So is known
The following formula is used to back calculate the Site calibrated source
strength to the date of the Manufacturer measured source strength at 12:00
noon in the local time zone:
( – λt )
So = S ⁄ e
where: B
S = current source strength
So = known source strength (from
manufacturer)
e = exponential (~2.7183)
λ = (ln 2)/T 1 ⁄ 2 (where T 1 ⁄ 2 = 73.83 days)
t = days since So is known
Transit Dose
When planning treatments, you may wish to determine the expected transit
dose, and, if necessary, allow for transit dose in dose calculations. This section
describes the transit dose algorithm for the GammaMedplus iX system and
provides specific air kerma measurements for transit absorbed dose.
The irradiation time of a given position starts with the movement from the
previous position to the given position and ends with the start of movement to
the next position. The irradiation during transit to the most distal dwell position
is not considered.
When using the GammaMedplus iX system, the minimum dwell time that can
be delivered for any given dwell position will vary depending on the distance
between adjacent positions. The minimum dwell time assigned to any given
dwell position must be equal to or greater than the recommended minimum
dwell time listed in Table B-2 below. This table is based on the limited wire
speed between adjacent active dwell positions.
The dwell time for a position begins to elapse when the dwell time for the
previous dwell position reaches zero seconds. With this design, the
GammaMedplus iX system ensures that no additional transit dose is delivered.
Important: The control software checks all dwell times and distances
between dwell positions when you press the Initiate
Treatment button. Treatment is not allowed if scaled dwell
times at the increments specified in Table B-2 are included in
the treatment plan. Proceed as recommended below.
It is recommended to set dwell times to zero for those positions that do not
fulfill the minimum required dwell time and to add the corresponding nominal
time to the following or previous dwell position to keep the total prescribed
dose constant. You should also change the nominal dwell times in your
treatment planning system accordingly and verify again the dose
distribution before you deliver the treatment. Nominal dwell times are
related to a 10 Ci source strength.
1-3 0.1
4-13 0.2
14-32 0.3
33-58 0.4
59-92 0.5
93-133 0.6
134-182 0.7
183-238 0.8
239-302 0.9
303-373 1.0
374-452 1.1
453-538 1.2
539-600 1.3
• The approximate air kerma at a position 10 mm from the axial center of the
source applicator (assuming an applicator of 100 mm length, a single dwell
position at the distal tip, and a source strength of 10 Ci) is 6.01 mGy.
• The approximate air kerma at a position 1 m from the axial center of the
channel (assuming a channel length of 1.2 m and a source strength of
10 Ci) is 20.52 µGy.
Note: The values above are for informational purposes only and
should not be used for treatment planning.
In This Chapter
Topic Page
Afterloader Description C-2
System Component Connections C-4
C-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Afterloader Description
Pertinent afterloader specifications are referenced in the following chart.
WARNING C-1: Do not connect any items that are not specified as part of the
system. Doing so may put the system in an unsafe condition.
Power supply circuits and wiring pathways must be provided by the customer
in accordance with the GammaMedplus iX specifications and local electrical
requirements. The necessary component connections of the
GammaMedplus iX system are made by Varian Medical Systems service
engineers at installation. In case of failure, or later structural alterations, the
connections must meet the requirements of the circuit diagram.
EMERGENCY RETURN
Door
Last Man Out Switch Switch
Interlock
Junction Control Control PC Printer
GammaMedplus iX box Console and monitor
STRAHLENQUELLE BESTRAHLUNG
STRAHL-
STELLUNG START
GESTARTET
RUHE- BEENDET
STELLUNG
UNTER-
BRE-
VER- CHUNG UNTERBROCHEN
RIEGELT
BETRIEBSART ALARM
NORMAL
SPEZIAL
ALARM T R
EIN / AUS
DOSIS
NETZ HUPE
AUS
BATTERIE HUPE AUS
STEUERPULT
Optional
EMERGENCY RETURN B4
Switch
B0
B3 B2
B1
C0
C1
D0
D1
In This Chapter
Topic Page
Additional Safety Information D-2
Treatment Day Tests and Checks D-4
Cleaning the Device D-11
Emergency Response Procedures (for posting) D-12
D-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
IEC Requirements
Note: Although this system has been tested to the EMC requirements
of EN60601-1-2:2001 and complies with this standard, it may
still be susceptible to electromagnetic interference or generate
electromagnetic interference. Medical electrical equipment
needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information
provided in the installation documents.
• Radioactive sources and source applicators for this equipment must only
be obtained from Varian Medical Systems. It is not permissible to use
radioactive sources from any other supplier. Active wire removal and
loading must only be performed by Varian Medical Systems service
engineers or by those persons authorized by Varian and having a
radioactive materials license to do so.
• The nominal transit speed of the active wire is up to 630 mm/sec,
depending on the travel distance. However, the speed is allowed to fall
while overcoming a path constriction. Inability to overcome the
constriction will cause an automatic retraction of the source.
• Users must be aware that the requirements of the safety standard IEC
60601-2-17 cannot guarantee that the presence of a radioactive source
outside the storage container of the GammaMedplus iX will be indicated
under all fault conditions. The use of an independent radiation monitor in
the same room as the patient is recommended. This independent radiation
monitor should be capable of giving both a visual and audible indication
of radiation, and should be able to continue to function and give an
indication of radiation for several hours after a power failure, for example
by battery back-up. Consult the regulations set forth by the Competent
Authority in your area.
• The positional accuracy of the active wire relative to the distal position is
calibrated to +/- 1 mm. Provided that the catheter connections are secure,
the length of the catheter is accurately measured and the catheter is not
subsequently stretched, the system will achieve a positional accuracy of
+/- 1.5 mm in any direction relative to the target within the applicator.
• Only cables and components supplied by Varian should be used in order to
comply with the requirements of IEC 60601-1-2:2001 36.201
(Electromagnetic Compatibility Emissions) and IEC 60601-1-2:2001
36.202 (Electromagnetic Compatibility Immunity Test). Detailed cable
information is provided in the service manual.
• The use of cables and equipment other than those specified, with the
exception of cables supplied by Varian Medical Systems as replacement
parts for internal components, may result in increased emissions or D
decreased immunity of the equipment.
• The maximum lengths of cables are:
RS232 PC to junction box, maximum connection length is 15 m
RS422 junction box to afterloader, maximum connection length is 50 m
RS485 control console to junction box, maximum connection length is 5 m
WARNING D-1: All tests involving the extension of the active source wire
must only be performed while no person is in the treatment
room.
Interruption of Treatment
Press the INTERRUPT button on the Control Console during the test
treatment.
EMERGEN
CY
RE
TURN
The treatment is interrupted immediately, and the source retracts into its
shielded position.
Door Interlock
With the GammaMedplus iX system idle (i.e., not carrying out a treatment),
open the treatment room door. If the afterloader is switched on, the Last Man
Out system is enabled. The door indicator light is supported as long as the
afterloader is powered on.
Open the door to the treatment room during the test treatment. The treatment
should be interrupted automatically and the source should retract into the
shielded position.
ALA
ARM
ALARM
ALARM DOOR
RESET
DOSE D
BUZZER
OFF BUZZER OFF
Key Tests
When you switch the system on by turning the POWER key to the horizontal
position, the GammaMedplus iX automatically performs an indicator test in
which all indicator lights on the Control Console (and any other external
consoles) flash briefly, and at the same time an audible signal is generated.
When the indicator test is complete, several indicator lights remain lit on the
Control Console — the MAINS and BATTERY indicators in the POWER
section, the NORMAL indicator in the MODE section, and the SAFE
indicator in the SOURCE section. In addition, the LOCKED indicator in the
SOURCE section of the Control Console and afterloader keypad flashes if the
SOURCE key in the afterloader keypad is in the LOCKED position.
POWE
D
SOURCE Key Switch
With the SOURCE key switch in the LOCKED position, attempt to initiate a
treatment. Verify that you cannot start the test treatment with the key in this
position.
Note: The room radiation monitor is not part of the supply by Varian
unless specifically ordered. See the manufacturer’s instructions
for these devices.
Note: The viewing equipment and intercom are not part of the supply
by Varian unless specifically ordered. See the manufacturer’s
instructions for these devices.
When you press the EMERGENCY button on the afterloader keypad, the
GammaMedplus iX does the following:
WARNING D-4: When you press the EMERGENCY button on the afterloader
keypad, the external power supply of the GammaMedplus iX
is switched off, and the source is retracted at maximum speed.
After that, all remote monitors, including the Control Console
and onboard radiation monitoring, are also turned off. The
only indication whether the source is in the shielded position
is the independent radiation monitoring system installed in the
treatment room and your portable survey meter. Confirm that
radiation levels are at background level. If not, initiate the next
step of the emergency response procedures as described in
Figure D-1.
Failure to follow these procedures may result in the following:
• Source may still be in the patient and you will not be aware
of this.
In the unlikely event that emergency retraction of the source does not occur
after pressing the INTERRUPT button, the EMERGENCY RETURN
switch, or the EMERGENCY button, the source can be retracted manually
into the shielded position using the handcrank.
D
Follow these steps to manually retract the source using the handcrank:
1. Enter the treatment room with a portable survey meter and go straight to
the GammaMedplus iX.
2. Pull the handle out of the handcrank and turn the crank in the direction of
the arrow until a distinct resistance is noticeable and the room radiation
level is safe.
Note: If the source wire has been retracted and the radiation level is
still high or if manual retraction of the source is not possible
using the handcrank, follow the procedures in the “Perform
Manual Recovery of the Source” section below. These
procedures are also outlined in Figure D-1.
1. Enter the treatment room with a portable survey meter and personal
dosimeter.
2. Calm the patient.
3. Open the Emergency Container.
4. Remove the applicator carefully from the patient using the long handled
forceps whenever possible. Maintain the maximum possible distance from
the patient and the afterloader.
retracted RETURN
switch. SWITCH
If source still
not retracted
EMERGENCY
button
Enter treatment room DUMMY-
CHANGE
If source still
not retracted
If source still
not retracted
This chapter describes the steps required to prepare the GammaMedplus iX for
transport. Note that the steps are the same for the GammaMedplus iX and the
GammaMedplus 3/24 iX systems.
In This Chapter
Topic Page
Preparing the GammaMedplus iX for Transport E-2
Inserting the Source Locking Wire E-3
Moving the GammaMedplus iX E-5
E-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
With the power off, uncouple the data cable and the mains cable from the
GammaMedplus iX and coil the cables around the rear cable take-up stand of
the device.
Turn the battery main switch to the OFF position to interrupt the battery
supply. Otherwise you will hear a warning signal. Remove the key to the
battery switch. During transport, the battery supply of the GammaMedplus iX
must be switched off to prevent any extension of the source in case of an
accident. The correct position of the source in the shielded position usually
ensured by the battery is now effected mechanically by inserting the source
locking wire.
2. Then insert the source locking wire into the channel far enough that about
5-10 cm remains outside.
Connector
Wire
2 23 24 2 23 24
21 2 21 2
20 20
19 19
2 2
18 18
3 17 3
17
16 4 16 4
5 5
6 6
13 13
7 12 7
12 8
11 1 8 11 1 9
0 9 0
1. 2.
23 24 2 23 24
2
21 2 21 2
20 20
19 19
2 2
18 18
3 17 3
17
16 4 16 4
5 5
6 6
13 13
7 12 7
12 11 1 8
11 1 8 0 9
0 9
3. 4.
This chapter describes features and procedures that are applicable to the
MammoSource iX model. Except in the areas described below, the
MammoSource iX does not differ in its basic operation and function from the
24-channel GammaMedplus iX system. These operations and functions are
described in detail in the other chapters in this user guide.
In This Chapter
Topic Page
System Description F-2
The Channel View Strip F-2
Insertion of the Applicator into the Patient F-3
F-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
System Description
The MammoSource iX is configured so that the source can be moved into 11
of 13 available channels in succession via the indexer head. In each of these
channels, the source may be programmed to stop in up to 60 dwell positions
for individual dwell times of 0.1 to 9999.9 seconds. The applicator in the
patient is connected to the afterloader via one or more source guide tubes.
Note: The MammoSource iX has channel Nos. 1-11 and 23-24 available,
which are a subset of the 24 channels in the full featured
GammaMedplus iX afterloader.
Refer to the “View and Edit Channel Data” section in Chapter 2 “Patient and
Treatment Data” for detailed information on the Channel View and Channel
Edit Strips.
Glossary
Access Privileges Permissions to use particular features of the control
software. Privileges are granted by an administrator at
the group level, such that all members of a group have
the same access privileges.
Active Wire The wire that contains the high activity Iridium-192
source.
Applicator Any closed ended delivery system through which the
source wire is intended to pass.
Catheter A flexible form of applicator.
Customer Purchaser of the Varian afterloader system. In the
case of the transportable system, the customer will
have multiple treatment locations to which the system
is transported and used.
Dummy Wire In addition to the active wire, the GammaMedplus iX
is equipped with a second, inactive wire. This dummy
wire serves to check the free passage and overall
length of the source guide tube/applicator assembly of
130 cm.
Dwell Location A position along an applicator channel where the
active wire will be placed and remain for a certain
amount of dwell time.
Dwell Time Amount of time the active wire will remain at one
dwell location before being moved to the next dwell
location or retracted into the safe.
Fraction A set of one or more channel plans to be executed in
a single treatment session.
Group Named sets of users with the same access privileges
for each member. The system defaults to having 4
groups (administrators, oncologists, physicists, and
radiation therapists) plus a special service group for
use only by Varian field service personnel.
Inactive Wire See dummy wire.
F
Glossary-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Glossary-3
GM10 UG Book.book Page 4 Tuesday, March 24, 2009 11:22 AM
Index
Index-1
GM10 UG Book.book Page 2 Tuesday, March 24, 2009 11:22 AM
Channel View Strip 2-31, 3-26, 7-3, F-2 Database port 4-37
Channel View table 3-26 Database service IP address 4-37
CHARGE light 1-16 Date and time setting 4-34
Chassis 1-12 Date cleared field 4-19
Class field 4-19 Date delivered field 2-20
Clean the GammaMedplus D-11 Date occurred field 4-19
Clear Error button 3-28, 3-36 Dates and times 1-33
Code field 4-19 Daylight Savings Time 4-34
Comma separated value (.CSV) file 4-21 Days since last source exchange icon 1-38
Competent Authority 1-2, 1-3, 1-7, 6-2 Decay Frequency button 4-46
COMPLETED indicator 1-22, 3-29 Default button 4-9
Connecting cables 3-7 Delete Channel button 4-29
Console & Afterloader Strip 4-1, 4-16 Delete File button 2-10
Console entry date field 2-20, 4-40 Delete Fraction button 2-21
Contact name field 4-11 Delete Record button 2-15
Contact phone field 4-11 Deliver Treatment button 3-19
Contamination checks 1-9 Delivery status field 2-12, 2-20, 4-25, 4-41
Continue button 2-10, 2-11 Description field 4-19, 4-20
Control Console 1-18, 2-40, 3-19, 3-24, Device housing 1-4, 1-12, 1-13
3-29, 3-32, 3-33, 4-46, 5-8, 5-11, 5-14, 5-17, Diagnostic file export 4-13
D-12, D-13, E-2 Directional scroll buttons 2-35
Control PC description 1-25 Directives 3-27
Control software description 1-31 Distal offset field 4-41
Control Software Errors and Messages A-6 Distal offset of the source 1-20
Conversion factor option buttons 4-34 Distal position of the source 1-16, 2-34,
Copy and Add Fraction button 2-21 2-39, 4-41
Copy button 2-39, 4-9 DOOR indicator 1-24
Copy Patient Plan button 4-25, 4-41 Door interlock 1-26
Create New Patient File button 1-32, 2-26 Door interlock test D-7
Create New Patient File workflow 1-34, 2-1, DOOR light 1-22, 1-23
2-24, 2-25, 2-30, 2-33, 4-23 DUMMY section 1-16, 2-40
Current dwell time field 3-26 Dummy sequence 4-36
Current source strength 1-38 Dummy wire 1-12, 1-17, 2-40, 3-20, 3-22,
Current source strength icon 1-38 3-23, 6-4, 7-4, F-3, F-1
Current value field 4-17, 4-33 Dummy wire sequence 3-22
Cycles icon 1-37 Dummy wires table 4-50
D Dwell points 2-32, 2-37, 3-27
Data Disclosure button 2-16 Dwell position 1-1, 1-12, 2-37, 3-5
Data Disclosure Report 2-16 Dwell positions field 3-26
Data Reconciliation Page 2-12 Dwell Selector 2-36
Data Verification Page 2-11 Dwell Selector Range 2-35, 2-39
Database Service Settings Dwell time 1-1, 2-33, 2-38, 3-5
Database AE Title 4-37 Dwell Time Lock 2-38
Index-3
GM10 UG Book.book Page 4 Tuesday, March 24, 2009 11:22 AM
Index-5
GM10 UG Book.book Page 6 Tuesday, March 24, 2009 11:22 AM
Index-7
GM10 UG Book.book Page 8 Tuesday, March 24, 2009 11:22 AM
X
XML file 2-6, 2-15