Doc No : JEL – QMS – P24

JURONG ENGINEERING LIMITED Rev : A


Quality Management System Page : 1 of 3

Effective Date :

Title : Calibration of Measuring Tapes Procedure

Rev. Description of Change Clause

A Initial Release for implementation. N.A.

Prepared By

QMS Engineer
_________________________ _______________________________ ______________________________
(Name / Signature) (Designation) (Date)

Reviewed By

Head, QMS
_________________________ _______________________________ ______________________________
(Name / Signature) (Designation) (Date)

Approved By

Management Representative
_________________________ ______________________________ ______________________________
(Name / Signature) (Designation) (Date)
Doc No : JEL – QMS – P24
JURONG ENGINEERING LIMITED Rev : A
Quality Management System Page : 2 of 3

Title : Calibration of Measuring Tapes Procedure

1. SCOPE

To establish a procedure for calibration of measuring tapes used for various


measurements during the process of fabrication or during all stages of inspection.

2. RESPONSIBILITY

The QC Section shall be responsible for implementation and maintenance of this


procedure.

3. PROCEDURE

a. Calibration

(i) The master tape shall have:


- the manufacturer's certificate of calibration, or
- a calibration certificate from an authorised agency.
(ii) The master tape shall be under the custody of QC Section and shall not
be used for routine measurements.

(iii) The following steps shall be followed for calibration:

Step 1 : The master tape shall be laid out on the floor.


Step 2 : The tape to be calibrated shall be laid parallel to the master tape.
Step 3 : Both the tapes shall be pulled by hand such that they are
Straight. The difference in reading between master tape (tm)
and new tape (tn) shall be recorded at intervals specified
below: interval for calibration of tapes with length less than
or equal to 10 metres shall be 5 metres. Interval for calibration
of tapes with length longer than 10 metres shall be 10 metres
.
Step 4: The actual ambient temperature and result of calibration (tm-tn)
shall be recorded in form no. F04I28.
Step 5: All tapes calibrated as specified above shall be labelled with a
label stating the control number of the tape, due date of re-
calibration and signature of the QC Supervisor.

Doc No : JEL – QMS – P24


JURONG ENGINEERING LIMITED Rev : A
Quality Management System Page : 3 of 3
Title : Calibration of Measuring Tapes Procedure

b. Acceptance criteria

Tapes with a difference higher than the following shall be rejected and shall not
be used for any measurement work.

Tape length (tn) in mm. Tolerance (tm-tn) in mm.

tn  5,000 ± 1.0

5,000 < tn  10,000 ± 1.5

10,000 < tn  20,000 ± 2.0

20,000 < tn  30,000 ± 2.5

30,000 < tn  50,000 ± 3.0

c. Frequency of calibration

All tapes shall be calibrated once in twelve months.

4. DOCUMENTATION

List of measuring tapes