Professional Documents
Culture Documents
Rxfa14g Manual
Rxfa14g Manual
Magnetic
Flow Converter
Contents
1. INTRODUCTION ...................................................................................1-1
1.1 Using the Magnetic Flowmeter Safely.............................................................1-2
1.2 Warranty..............................................................................................................1-3
2. HANDLING PRECAUTIONS.................................................................2-1
2.1 Checking Model and Specifications...............................................................2-1
2.2 Accessories.......................................................................................................2-1
3. INSTALLATION......................................................................................3-1
3.1 Installation Location.........................................................................................3-1
3.2 Mounting............................................................................................................3-1
4. WIRING..................................................................................................4-1
4.1 Wiring Precautions............................................................................................4-1
4.2 Cables................................................................................................................4-1
All Rights Reserved. Copyright © 2006, Rota Yokogawa i IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
CONTENTS
5.3.1 Setting Example for Selection-Type Data: Flow rate units................................ 5-5
6. Parameter Description.........................................................................6-1
6.1 Parameters.........................................................................................................6-1
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 ii All Rights Reserved. Copyright © 2006, Rota Yokogawa
CONTENTS
8.4 Parameters.........................................................................................................8-3
9. Actual Operation...................................................................................9-1
9.1 Pre-operation Zero Adjustment.......................................................................9-1
All Rights Reserved. Copyright © 2006, Rota Yokogawa iii IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
CONTENTS
10. MAINTENANCE.................................................................................10-1
10.1 Maintenance...................................................................................................10-1
10.3 Troubleshooting.............................................................................................10-4
11. Outline..................................................................................................2-1
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 iv All Rights Reserved. Copyright © 2006, Rota Yokogawa
1. INTRODUCTION
1. INTRODUCTION
This instrument has been adjusted at the factory • If the customer or any third party is harmed by
before shipment. the use of this product, Yokogawa assumes no
responsibility for any such harm owing to any
To ensure correct use of the instrument, please
defects in the product which were not predict-
read this manual thoroughly and fully understand
able, or for any indirect damages.
how to operate the instrument before operating it.
NOTE NOTE
This manual describes the hardware and soft- Please refer to manual IM 01R21D01-01E-E for
ware configurations of the RXFA14 converter and information of the RXF Remote Flowtube.
the software configuration of the RXF integral
flowmeter.
Safety and Modification Precautions
The contents of “basic operating procedures”, • The following general safety precautions must
“parameter description”, “operation via BRAIN be observed during all phases of operation,
terminal (BT200)”, “operation via HART communi- service, and repair of this instrument. Failure to
cator”, and “actual operation” for the RXF integral comply with these precautions or with specific
flowmeter are the same as the RXFA14, see WARNINGS given elsewhere in this manual vio-
Chapter 5 to 9 in this manual. lates safety standards of design, manufacture,
For the hardware configuration of the RXF integral and intended use of the instrument. Yokogawa
flowmeter, see the user’s manual of the RXF assumes no liability for the customer's failure to
Integral Flowmeter/Remote Flowtube [Hardware comply with these requirements. If this instru-
Edition] (IM 01R21D01-01E-E). ment is used in a manner not specified in this
For FOUNDATION Fieldbus protocol (Output manual, the protection provided by this instru-
Signal and Communication suffix code "F") ment may be impaired.
please refer to IM 01R21F02-01E. • Yokogawa will not be liable for malfunctions or
damage resulting from any modification made
to this instrument by the customer.
Regarding This User's Manual • The following safety symbol marks are used in
• This manual should be provided to the end user. this user's manual and instrument.
• Before use, read this manual thoroughly to
comprehend its contents. WARNING
• The contents of this manual may be changed
without prior notice. A WARNING sign denotes a hazard. It calls at-
• All rights are reserved. No part of this man- tention to procedure, practice, condition or the
ual may be reproduced in any form without like, which, if not correctly performed or adhered
Yokogawa's written permission. to, could result in injury or death of personnel.
• Yokogawa makes no warranty of any kind with
regard to this material, including, but not limited
to, implied warranties of merchantability and CAUTION
suitability for a particular purpose.
A CAUTION sign denotes a hazard. It calls atten-
• All reasonable effort has been made to ensure
tion to procedure, practice, condition or the like,
the accuracy of the contents of this manual.
which, if not correctly performed or adhered to,
However, if any errors or omissions are found,
could result in damage to or destruction of part
please inform Yokogawa.
or all of the product.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• Please note that this user's manual may not be
revised for any specification changes, construction
changes or operating part changes that are not
considered to affect function or performance.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 1-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
1. INTRODUCTION
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 1-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
1. INTRODUCTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 1-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
1. INTRODUCTION
Blank Page
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 1-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
2. HANDLING PRECAUTIONS
2. HANDLING PRECAUTIONS
This instrument has been inspected carefully at
the factory before shipment. When the instrument
2.2 Accessories
is delivered, visually check that no damage has Check that the parts shown below are included in
occurred during transportation. the package:
Read this section carefully as it contains important • Hexagonal wrench: 1 piece (nominal size 3 mm)
information on handling this instrument. Refer to • Mounting hardware: 1 set
the relevant sections for information not contained
in this section. If you have any problems or ques-
tions, please contact Yokogawa sales office.
2.3 Storage Precautions
If the instrument is to be stored for a long period
2.1 Checking Model and of time after delivery, observe the following points.
F0201.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 2-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
2. HANDLING PRECAUTIONS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 2-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
3. INSTALLATION
3. INSTALLATION
1. Fix the instrument on the mounting fixture using
WARNING
four screws.
Installation of the magnetic flowmeter must be 2. Fix the mounting fixture with the instrument
performed by expert engineers or skilled person- installed on a 2-inch pipe using a U-bolt.
nel. No operator shall be permitted to perform
procedures relating to installation.
• Horizontal Mounting
Mounting fixture
Nut
All Rights Reserved. Copyright © 2006, Rota Yokogawa 3-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
3. INSTALLATION
Blank Page
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 3-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
4. WIRING
This chapter describes the wiring on the converter • Cover each shield of the signal cable with
side only. For information relating to wiring on the vinyl tube or vinyl tape to avoid contact be-
flowtube side, refer to the user’s manual of the tween two shields or between a shield and a
RXF Remote Flowtube (IM 01R21D01-01E-E). case.
• In case of 24V power supply version, it comes
WARNING with a plug. Use this plug to cover the unused
wiring port when wiring the instrument with only
The wiring of the magnetic flowmeter must be
one, four-core cable.
performed by expert engineers or skilled person-
• Be sure to turn power off before opening the
nel. No operator shall be permitted to perform
cover.
procedures relating to wiring.
• Before turning the power on, tighten the cover
securely.
• The terminal box cover is locked by the
CAUTION
special screw. In case of opening the terminal
Once all wiring is complete, check the connec- box cover, use the hexagonal wrench attached.
tions before applying power to the instrument. For handling the locking screw, refer to Figure
Improper arrangements or wiring may cause a 4.4.1.
unit malfunction or damage. • Be sure to lock the cover by the special screw
using the hexagonal wrench attached after
installing the cover. For handling the locking
4.1 Wiring Precautions screw, refer to Figure 4.4.12.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
4. WIRING
B
A
50 (1.97)
SA
SB
90 (3.54)
90 (3.54)
70 (2.76)
150+5
(5.9)
(5.9)
On the On the
converter flowtube 0.5 to 1.5 mm2
side side
In case of power cable, Green/Yellow covered
L (Specified Dimensions)
f10.5 (0.4)
EX2
EX1
F0403.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 4-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
Conduit Wiring
When wiring the conduits, pass the conduit F0408.EPS
through the wiring connection port and utilize the Figure 4.4.1 Removing the Terminal Box Cover
waterproof gland to prevent water from flowing in.
Place the conduit pipe on an angle as shown in
4.4.2 Terminal Configuration
Figure 4.3.4. Install a drain valve at the low end
of the vertical pipe, and open the valve regularly. When the cover is removed, the connection
terminals will be visible. The terminal configura-
tion labels are attached in the locations shown in
Figure 4.4.2.
Drain valve
F0407.EPS
F0409.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
4. WIRING
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 4-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
796 (2610)
600 (1970)
446 (1460)
200 ( 660)
0
20.4 22 24 26 28.8
Usable range E (V)
Cable cross section area: 1.25 mm2
Cable cross section area: 2 mm2
F0411.EPS
Protective grounding terminals
F0412.EPS
IMPORTANT
Set the local commercal power frequency in or- 4.4.6 Wiring the Remote Flowtube with
der to eliminate the effect of induction noise from RXFA14 Converter
the power supply.
Refer to “Chapter 6: Parameter Description” in
WARNING
this manual.
Parameter No.: J30 and J31 Before wiring, be sure that the power supply for
RXFA14 converter has been turned off to prevent
an electrical shock.
4.4.5 Grounding
Connect wiring as shown in the figure below.
CAUTION RXFA14 Converter EX1
Be sure to connect the protective grounding of the EX2
Excitation cable
RXFA14 with a cable of 2mm² or larger cross C
section in oder to avoid electrical shock to the RXFC dedicated
signal cable SA
operators and maintenance engineers and to
prevent the influence of external noise. A
Connect the grounding wire to the mark. The B
grounding should satisfy Class D requirements SB
(ground resistance, 100W or less).
Remote
Converter
flowtube
SA Taping*
A A
A EX2 B B
SB Taping*
IMPORTANT B C C
EX1 EX1
When a lightning protector (option code: /A) is EX1 EX2 EX2
C
selected, use a grounding with a resistance of 10 Ω Remote flowtube
* Individually tape and insulate
the shields corresponding to
or less. SA and SB on the remote
flowtube side. F0413.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
4. WIRING
4.4.7 Connecting to External Instru- • When the active pulse output (optional code /
ments EM) is selected, the range of pulse rate must
be set to 2 pps maximum.
• To avoid communication (BRAIN/ HART) failure,
WARNING
it is recommended to use the shielded cable.
Before wiring with external instruments, be
sure to turn off the power supply for RXFA14
converter and any external instruments. NOTE
RXFA14 RXFA14
l+ DO+
Receiver
l- Instrument PULSE OUT Load Electronic Counter
DO-
Resistive Load Max. 1 kOhm
30V DC, 0.2A. max
Communication Resistive load
BRAIN 250 to 450 Ohm
HART 250 to 600 Ohm F0416.EPS
Mechanical Counter
PULSE OUT DO-
IMPORTANT
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 4-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
IMPORTANT
Load
DO- (or DIO-)
F0419.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-7 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
4. WIRING
Blank Page
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 4-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
WARNING
Be sure to set parameters as “Protect” on the write protect function after finish of
parameter setting work.
Under extremely rare case, the infra-red switches may respond unexpectedly in
such conditions as sticking ball of water or extraneous substances on the surface
of display panel glass according to the principle of infra-red switch operation.
Its probability rises in such cases as sticking rain water by storm or other similar
situation and washing up work near flowmeter installation place.
Either to illuminate or stop illuminating the infra-red switches by the flashlight may
cause the mis-reaction.
Refer to “Menu P: Parameter Protection Items” and section “10.2.2” how to use the
write protect function in detail.
IMPORTANT
Operate the display unit under the condition where direct sunlight, etc.... do not
shine to the setting switches directly when the parameter setting operation is
carried out.
NOTE
(1) Always use the setting switches with the cover of the RXFA14 or the RXF
integral flowmeter closed.
(2) Use these switches with them covered by the glass window.
(3) If dirt, dust or other substances surfaces on the display panel glass, wipe
them clean with a soft dry cloth.
(4) The operation with dirty gloves may cause a switch response error.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
Actual instantaneous flow rate FR Displays the actual reading for instantaneous flow rate.
Instantaneous flow rate: mA FR Displays the instantaneous flow rate for the span as a current output value.
Bar graph indicating None Displays the instantaneous flow rate for the span using a percentage
instantaneous flow rate bar graph.
Totalized forward-direction flow rate FTL Displays the totalized value for flow rate in the forward direction.
Totalized reverse-direction flow rate RTL Displays the totalized value for flow rate in the reverse direction.
Totalized differential flow rate DTL Displays the totalized value for flow rate between forward
totalization and reverse totalization.
Diagnosis of electrode ADH Displays the adhesion condition in the form of a bar graph.
adhesion (See the description for parameters K10 through K15 from
Chapter 6: Parameter Description for more details.)
Communication COM Displays the communication type.
T0501.EPS
(4)
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 5-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
· Before changing any settings, be sure to check the corresponding setting details.
in Chapter 6: Parameter Description.
5.2.1 Display Mode → Setting Mode
Display Mode will be adopted when the power is turned on, and the Setting Mode
can be activated using the following procedure.
+ +
Select “No”
Setting modes
* The term "mode" is used to refer to a situation
where display and setting are possible.
+
Press the SET switch while holding down the SHIFT switch. F0502.EPS
Sample Display: Procedure for moving from Display Mode to Setting Mode
NOTE
(*1)
Entry Mode When other operations are carried out, the
system will automatically return to the Display
Touch SHIFT + key Mode. When no operations in this display are
carried out for 20 seconds, the system will
automatically return to the Display Mode.
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
NOTE
Setting mode
Yes NOTE
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 5-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
If no operations are carried out for a period of 10 min in this mode, the system will
automatically return to the Display Mode.
(i) Selection-type B21:Base Flow Unit The desired data item is selected from a predefined list.
m3
m3
kl(Kiloliter)
(ii) Numeric-type B23:Flow Span Data is set using the values in each digit and using
100 l/min the decimal point.
000100 l/min
Rng:0.00001 32000
Regarding the alphanumeric-type format (iii), the following alphanumerics are displayed in the following sequence:
Setting Mode Setting mode Major Item Parameter Search Mode has been
Start:
Major Item Major item
P:Protect accessed in this screen.
B:Easy Setup
Parameter parameter C:Basic Setup
Search Mode Touch the switch to access B: Easy Setup.
F0504-1.EPS
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
Sub item B21:Base Flow Unit In this screen, Parameter Replacement Mode
Parameter Current unit setting m3
Replacement Mode
Available selections
m3 has been called up using the switch.
kl(Kiloliter)
Sub item B21:Base Flow Unit Touch the switch to move the cursor to the
Current unit setting m3
Available selections
l(Liter) unit item for selection. In this example, the
cm3
switch is touched twice to select l (Liter)
X2
NOTE
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 5-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
NOTE
F0505-1.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-7 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
NOTE
Sub-item Major item B:Easy Setup When no operations are carried out for 20
Parameter 22:Base Time Unit
Sub-item 23:Flow Span seconds in the flashing state, the system will
Search Mode Selection (B) 24:Flow Decimal Pnt automatically return to the Sub-item Parameter
Search Mode.
B23:Flow Span When the operations except
100 l/min
Incorrect setting: 1200000 l/min are carried out, the parameter cannot be
Rng:0.00001 32000 set.
Out of range.
Touch any key. The system returns automatically to Sub-Item
Selection Screen (B).
B23:Flow Span
100 l/min NOTE
Incorrect setting: 0001.0. l/min
Rng:0.00001 32000
If the input value is outside the valid selection
range, the message “Out of range. Touch any
Invalid value.
Touch any key. key.” will be displayed. In such a case, touch
any switch to return to Parameter Replacement
F0505-2.EPS
Mode and redo the setting.
NOTE
If more than one decimal point has been input,
the message “Invalid value. Touch any key.” will
be displayed. In such a case, touch any switch
to return to Parameter Replacement Mode and
redo the setting.
NOTE
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 5-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-9 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
Blank Page
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 5-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
6. PARAMETER DESCRIPTION
6.1 Parameters NOTE
With the exception of parameters that were In order to ensure that correct flow-rate data can
specified by the customer upon ordering, all of the be acquired, it is crucial that the nominal size,
internal parameters will initially be set to default flow-rate span, and meter factor of the combined
values. Actions such as the modification of display remote flowtube are set. In cases where a re-
details can then be carried out whenever necessary. mote flowtube is ordered at the same time as the
For FOUNDATION Fieldbus protocol, please refer RXFA14, or where the RXF integral flowmeter is
to IM 01E20F02-01E. ordered, the nominal size and meter factor will
be set upon shipment from the manufacturing
For PROFIBUS PA protocol, please refer to
plant, and these will not require additional setting.
IM 01E20F12-01E.
If the RXFA14 is ordered separately, the default
value will be set for the meter factor; accordingly,
IMPORTANT it will be necessary to change this setting to the
meter factor indicated on your remote flowtube
Make sure to keep the instrument’s power on at data plate.
least for 30 s after you have set the parameters. If a flow-rate span was specified upon ordering,
If you turn the power off immediately after the this will be set before shipment. If this is not the
parameters are set, the settings will be canceled. case, however, it will be necessary for the ap-
propriate value to be set by the user.
R: Read only
W: Writable
Parameter number.
T0601.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
T0605.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
T0607-1.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-7 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
T0609.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-9 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
K00 Diagnosis
(DIAGNOSIS)
K10 Adhesion Check W No No Selects whether or not to perform diagnosis
(ADHESION CHK) Yes of adhesion to the electrode.
K11 Adhesion Level1 W 0.00 to 100.00 M ohm 2 0.10 Sets the resistance value for adhesion
(ADH LEVEL1) Level 1 to the electorode.
K12 Adhesion Level2 W 0.00 to 100.00 M ohm 2 0.50 Sets the Level 2 for the diagnosis of
(ADH LEVEL2) adhesion to the electrode.
K13 Adhesion Level3 W 0.00 to 100.00 M ohm 2 1.00 Sets the Level 3 for the diagnosis of
(ADH LEVEL3) adhesion to the electrode.
K14 Adhesion Level4 W 0.00 to 100.00 M ohm 2 3.00 Sets the Level 4 for the diagnosis of
(ADH LEVEL4) adhesion to the electrode.
K15 Adh Measure Value R — M ohm 2 Displays the diagnosis value for
(ADH MEAS VAL) adhesion to the electrode.
K60 — R Good
(SELF CHECK) Error See “6.5 Alarm Functions”.
T0611.EPS
M00 Adjustment
(ADJUSTMENT)
M10 Auto Zero Exe W No Execution No Execution Selects whether or not automatic
(AUTOZERO EXE) Execution zero adjustment is carried out.
Linked with B50.
M11 Magflow Zero R/W -99.999 to 99.999 3 0.000 Displays the result of the automatic
(MAGFLOW ZERO) zero adjustment, or sets the zero point.
M60 — R Good
(SELF CHECK) Error See “6.5 Alarm Functions”.
T0612.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
P00 Protect
(PROTECT)
P10 Key Code W 0 to 9999 0 Parameter of the display
(KEY CODE) restriction
P20 Write Protect R No No Displays whether or not overwriting
(W PROTECT) Yes of parameter data is prohibited.
P21 Enable Wrt Passwd W ASCII 8 characters Sets the correct password to release
(ENABLE WRITE) write protection function.
P22 New Password W ASCII 8 characters Sets the password for write protection
(NEW PASSWORD) function.
P23 Software Seal R Break Keep Displays whether or not a Joker
(SOFT SEAL) Keep password was used (Break).
P60 — R Good See “6.5 Alarm Functions”.
(SELF CHECK) Error
T0613-1.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-11 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
6.4 Parameter Description [B22: Base Time Unit] Selection of time units for
the flow rate span
(1) Menu B: Easy Setup items → This setting is linked with that of parameter C41.
This parameter selects the time units of the flow
Those parameters with a high frequency of use
rate span; however, if “meter” has been selected in
have been grouped together in Easy Setup. All
B21 (or C40) for the flow rate units, “/s” is auto-
basic functions can be controlled using only the
matically set for this parameter.
parameters from this block. Parameters from Menu
B share identical names with those from other
[B23: Flow Span] Setting of the flow rate span
menus; however, modification of one such pa-
→ This setting is linked with that of parameter C42.
rameter will result in the other being automatically
The span can be set for the forward flow rate in the
modified.
range 0 to 32,000. The units set using B21/C40:
[B10: Language] Selection of language used for Base Flow Unit and B22/C41: Base Time Unit
the display unit will be displayed at this time.
→ This setting is linked with that of parameter H30.
One of the following languages can be selected
for use by the display unit. NOTE
Data Range If the flow rate units, time units, and flow rate
Setting item Description span are specified upon ordering, these param-
English All parameters, alarm messages etc. eters will be setup before shipment; however, if
dispayed in English this is not the case, it will be necessary for the
Japanese All parameters, alarm messages etc. appropriate values to be set up by the user.
dispayed in Japanese katakana
French All parameters, alarm messages etc.
NOTE
dispayed in French
German All parameters, alarm messages etc. Flow rate span is the value for instantaneous
dispayed in German
flow rate that corresponds to a current output
Italian All parameters, alarm messages etc.
dispayed in Italian of 20 mA. The following factors should be taken
Spanish All parameters, alarm messages etc. into consideration when deciding on the flow-rate
dispayed in Spanish span.
• In the case of applications with large variations
[B20: Flow Damping] Setting of the damping time in flow rate, the maximum flow rate should be
constant set. If a flow rate in excess of the flow rate span
→ This setting is linked with that of parameter C11. has to occur, output would be possible up to
The damping time constant should be modified to an upper limit of 108 %, and beyond this, error
suppress an output fluctuation or to change the would occur. Note that the same applies to
response time. This time constant has an effect pulse output and totalization.
on analog output and on the flow rate display (i.e., • In the case of applications that have a relatively
actual instantaneous flow rate, %, current value), stable flow rate, a flow rate span of 1.5 to 2.0
and in addition, it also affects pulse output and
times larger than the normal flow rate may be
totalization. However, when “No Damping” has
considered suitable.
been set for J24: T/P Damp Select, there will be
• The flow rate span to be adopted should - upon
no effect on pulse output or totalization.
conversion to flow velocity - correspond to a
→ Time constant: The time required for the output
to reach 63.2 % from 0 %. value within the range of 0.1 to 10 m/s. The
flow velocity can be confirmed using sizing data
[B21: Base Flow Unit] Selection of flow units for or with parameter C44: Velocity Check, and in
the flow rate span the latter case, the value obtained when span
→ This setting is linked with that of parameter C40. is converted to flow velocity will be displayed. •
This parameter selects the flow rate span units. (In Regardless of the position of the decimal point,
case of mass flow, the setting of density is also the largest value that can be set on the display
required. Refer to C46: Mass Flow Density for unit is 32,000. Furthermore, it is not possible
more details.) to set a number of 4 or greater for the highest
order digit. Similarly, if 3 is set for this highest
order digit, it will not be possible to set a num-
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-12 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
ber of 2 or greater for the next digit to the right, Item Description
regardless of the position of the decimal point. n Unit/P 10-9 x FU
Example: A value of 333.33 is represented by U Unit/P 10-6 x FU
the character string 33333, and since this ex-
m Unit/P 10-6 x FU
ceeds 32000, it cannot be set. In such a case,
Unit/P FU
the value 333.3 should be set instead.
k Unit/P 103 x FU
[B24: Flow Decimal Pnt] Setting of the decimal M Unit/P 106 x FU
point position for the instantaneous flow rate Pulse/s Number of pulses to be output for one
→ This setting is linked with that of parameter C43. second at 100% output
This parameter sets the position of the decimal FU: Flow rate unit selected in B21/C40: Base Flow Unit
point for instantaneous flow rate values in terms
of the number of digits. When set using “Auto”, the [B31: Total Scale] Setting of the totalization scale
decimal point position will be automatically de- → This setting is linked with that of parameter D11.
termined in accordance with the setting value for The flow rate is totalized in individual counts in
B23/C42: Flow Span as shown below. accordance with this parameter’s setting. If 0 is
FLOW SPAN ≤ 9 Decimal point position: 3 digits selected, it indicates that the totalization function
9 < FLOW SPAN ≤ 90 Decimal point position: 2 digits is not to be used.
90 < FLOW SPAN ≤ 900 Decimal point position: 1 digit
NOTE
900 < FLOW SPAN Decimal point position: no digits
(i.e., no decimal point) If a totalization scale is specified upon ordering,
this parameter is set up before shipment; howev-
When an item other than “Auto” is set, the er, if this is not the case, it will be necessary for
selected number of digits for the decimal point the appropriate value to be set up by the user.
position is used.
With the decimal point removed, 6 digits are avail- NOTE
able for the instantaneous flow-rate value, and
display is possible up to 999,999. If an overflow • By setting the scale of the totalization pulse,
occurs as a result of the setting adopted for the flow rate totalized value is displayed on the
decimal point position, the warning 84: Disp Over display unit. The totalization scale is determined
Wng will be displayed to provide notification of in accordance with the settings of B30/D10:
this condition. Total Unit and B31/D11: Total Scale.
• The maximum value that can be displayed is
Example: When 1000 m3/h is set for B23/C42: 99999999, and if this is exceeded, the value 0
Flow Span is counted once again. However, counting
Item Display content for instantaneous flow rate stops at 99999999 when the totalization switch
value function is used.
Auto 1000 m³/h • If multiple ranges are being used, the flow rate
0 1000 m³/h span for the smallest range becomes the stan-
1 1000.0 m³/h dard for the D13: Total Low Cut setting value.
2 1000.00 m³/h • Totalization for the reverse direction and for the
3 With the decimal point removed, 7 digits are not
differential flow rate is carried out only when
available for the instantaneous flow rate value;
therefore, a warning is displayed. “Fwd/Rev Ranges” is selected for F20: DO
Function or F21: DIO Function.
[B30: Total Unit] Setting of units for totalization • The totalized units are indicated on the display
scale unit when B31/D11 is 0.001, 0.01, 0.1, 1, 10,
→ This setting is linked with that of parameter D10. 100, or 1000. In the case of other setting val-
This parameter selects the flow-rate units for use ues, the totalized units are not indicated.
in totalization. Example 1: To count in 1 Ml (mega-liter) steps
with flow rate span = 100 m3/h
Since 1 Ml = 103 x m3, k Unit/P is
set for B30/D10, and 1 is set for
B31/D11. “x103 m3” is indicated for
the totalized units in the Display Mode.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-13 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[B33: Pulse Scale] Setting of pulse scale [B40: Display Select1] Setting of the first line for
→ This setting is linked with that of parameter E11. display unit
Pulse output is performed in individual counts in → This setting is linked with that of parameter H10.
accordance with this parameter’s setting. If 0 is This parameter selects the display content of the
selected, it indicates that the pulse output function first line for display unit. The size of the characters
is not to be used. which are displayed will depend on the selections
made for B41/H11: Display Select2 and B42/
NOTE H12: Display Select3 as described below. (For
more details, refer to Chapter 5: Basic Operating
If a pulse scale is specified upon ordering, this
Procedures.)
parameter is setup before shipment; however, if
this is not the case, it will be necessary for the
CAUTION
appropriate value to be setup by the user.
It is not possible to set Display Select1 to “Off”.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-14 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[B41: Display Select2] Setting of the second line If a flow rate span was specified upon ordering,
for display unit this will be set before shipment. If this is not the
→ This setting is linked with that of parameter H11. case, however, it will be necessary for the appro-
This parameter selects the display content for the priate value to be set by the user.
display unit’s second line. When “Off” is selected,
one-line display will be adopted regardless of the
[C10: Tag No] Setting of the tag number
setting made for B42/H12: Display Select 3.
→ The setting for this parameter corresponds to
one of the ordered items.
[B42: Display Select3] Setting of the third line for Up to a maximum of 16 characters can be entered
display unit. for the display unit. For more details regarding the
→ This setting is linked with that of parameter H12. actual characters that can be used, refer to Section
This parameter selects the display content of the 5.2.2: Setting Mode.
third line for display unit. When “Off” is selected for
this parameter, two-line display is adopted. NOTE
[B50: Auto Zero Exe] Execution of the automatic If a tag number is specified upon ordering, this
zero adjustment function parameter is set up before shipment; however, if
→ This setting is linked with that of parameter M10. this is not the case, it will be necessary for the
This parameter executes the automatic zero appropriate value to be set up by the user.
adjustment function: If “Execution” is selected, this
function will be started. “Now Auto Zero Executing...”
is indicated while the Auto Zero function is being [C11: Flow Damping] → Setting of the damping
carried out. The result of the automatic zero time → Refer to the description of parameter B20.
adjustment is confirmed using M11: Magflow Zero,
and if the result exceeds the rated value, the [C20: Measure Mode] Selection of dual frequency
warning 82: Auto Zero Wng will be displayed. excitation mode
(For more details, refer to Chapter 9: Operation.) Item Description
Setting Function Standard DF Standard dual frequency exitation
No Execution No Execution Enhanced DF Enhanced dual frequency exitation
Execution Automatic zero adjustment is started
Enhanced DF (i.e., enhanced dual frequency exci-
tation) is set to stabilize measurement for difficult
(2) Menu C: Basic Setting items applications, such as for high concentration slurry
Menu C principally contains the basic setting items or low conductivity fluid. Note that this parameter
for the flowtube. is only valid when this product is used in combina-
tion with an RXF flowtube, as only those flowtubes
will support the enhanced dual frequency excitation.
NOTE
(Optional code:/HF1 or /HF2)
In order to ensure that correct flow rate data can
be acquired, it is crucial that the nominal size,
NOTE
flow rate span, and meter factor of the combined
remote flowtube are set. In cases where a re- • When this product is used in combination with
mote flowtube is ordered at the same time as the any flowtube that does not support enhanced
RXFA14, or where the RXF integral flowmeter is dual frequency excitation and "Enhanced DF" is
ordered, the nominal size and meter factor will selected, a setting alarm will be displayed. The
be set upon shipment from the manufacturing setting should be returned to "Standard DF" in
plant, and these will not require additional setting. such a case.
If the RXFA14 is ordered separately, the default • When dual frequency excitation mode is
value will be set for the meter factor; accordingly, changed, perform zero adjustment. For details
it will be necessary to change this setting to the on zero adjustment refer to chapter 9.
meter factor indicated on your remote flowtube
data plate.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-15 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[C22: High MF] Setting of the high-frequency [C24: High MF (EDF)] Setting of the high-frequen-
meter factor cy meter factor for EDF
This parameter sets the high-frequency meter fac- This parameter sets the high-frequency meter
tor for standard dual frequency exciation. factor as required when Enhanced DF (i.e., en-
hanced dual frequency excitation) is selected.
NOTE
[C31: Nominal Size Unit] Setting of the nominal
Meter Factor Settings
size units
Flowtube data plate
Magnetic
Flow Tube
(PS=Pmax at 20°C) This parameter selects the units used for setting
Meter factors of the nominal size.
of flowtube
(1) Confirm that the serial number indicated by [C41: Base Time Unit] Selection of time units for
COMB. NO. on the RXFA14 converter’s data the flow rate span
plate corresponds with the RXF remote flow- → Refer to the description of parameter B22.
tube’s serial number.
(2) Set the values that are marked in the ME- [C42: Flow Span] Setting of the flow rate span
TER FACTOR fields on the data plate for the → Refer to the description of parameter B23.
Remote Flowtube.
(3) The meter factors are crucial in ensuring that [C43: Flow Decimal Pnt] Setting of the decimal
the electromotive force is correctly in propor- point position for the instantaneous flow rate
tion to the flow velocity and are determined → Refer to the description of parameter B24.
at the manufacturing plant by actual-flow
calibration. [C44: Velocity Check] Display of the span flow
velocity
[C23: Low MF (EDF)] Setting of the low-frequency This parameter displays the span for the maximum
meter factor for EDF range in m/s units.
This parameter sets the low-frequency meter
factor as required when Enhanced DF (i.e., en-
[C45: Density Unit] Setting of the density units
hanced dual frequency excitation) is selected. If
for mass flow rate. This parameter selects the
“Standard DF” has been selected for C20: Mea-
units for density as required when making settings
sure Mode, neither C23: Low MF (EDF) nor C24:
using C46: Mass Density.
High MF (EDF) is displayed, and if “Enhanced
DF” is selected, the four parameters from C21 to
C24 will be displayed.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-16 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[C46: Mass Density] Setting of the density for (3) Menu D: Total Setting items
mass flow rate Menu D contains parameters that are relevant to
This parameter is necessary in situations where totalization function settings.
t, kg, g, klb or lb has been selected as the mass
unit in B21/C40: Base Flow Unit. If a mass unit is [D10: Total Unit] Setting of units for totalization
selected in B21/C40: Base Flow Unit and a value scale → Refer to the description of parameter B30.
of 0 is set for this parameter, the setting alarm
“57: Density Set Err.” will be displayed. In such a [D11: Total Scale] Setting of the totalization scale
case, ensure that the density is set correctly. → Refer to the description of parameter B31.
[C47: User Span Select] Selection of the special [D12: Total Decimal Pnt] Setting of the decimal
setting for units point position for totalization display.
This parameter selects whether or not special This parameter sets the position of the decimal
units are used for instantaneous flow rate. Actual point for totalization display in terms of the number
setting of these units is carried out using C48: of digits. Except in cases where 0 is selected, the
Flow User Unit and C49: Flow User Span. totalized units are not displayed.
[C48: Flow User Unit] Setting of the special flow Example: When totalized value is 12345678 m3
rate units. Item Totalization display
This parameter is used to select the special units 0 12345678 m³
up to maximum 8 characters in length. These units 1 1234567.8
are displayed when instantaneous flow rate is se- 2 123456.78
lected in the Display Mode, and they are displayed 3 12345.678
for A20: FLOW RATE when BRAIN communica- 4 1234.5678
tion is being carried out. 5 123.45678
6 12.345678
7 1.2345678
[C49: Flow User Span] Setting of the special flow
T0619.EPS
rate span.
This parameter sets the special span to be dis-
[D13: Total Low Cut] Setting of the totalization
played for 100% output in the maximum range.
stop range
This parameter allows the settings to be made
NOTE that prevent totalization when the flow rate is
Example : To set the special flow rate span to at or below the low-cut setting value. In cases
100 dl/s where there are multiple ranges or forward/reverse
Since 100 dl (deci-liter)=10 l (liter), ranges, low cut is carried out at the setting value
“l (Liter)” is set for B21/C40: Base for the smallest span (i.e. an integer between 0
Flow Unit, and 100 %).
“/s” is set for B22/C41: Base Time Example: When the first range = 10 m3/h, the
Unit, second range = 100 m3/h, and the Total
“10” is set for B23/C42: Flow Span, Low Cut = 3 %, no totalization is carried
“Yes” is set for C47: User Span out at flow rates of 0.3 m3/h or lower.
Select,
“dl/s” is set for C48: Flow User [D20: Total Execution] Operation setting for the
Unit, totalization function
“100” is set for C49: Flow User This parameter sets ""Start" and "Stop" of the
Span. totalization function, in addition to performing the
“100 dl/s” is indicated for 100% out preset function for the forward totalized value and
put in the Display Mode. the reverse totalized value.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-17 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
*: The preset function starts the count for totaliza- status output, refer to the descriptions of param-
tion from the set value. eters F20 and F21.)
Item Description If this function is set up, the totalization count will
Start (initial Starts totalization stop at 99999999. D23 sets the lower 6 digits of
value) the 8-digit totalization switch value.
Stop Stops totalization
Preset Total Sets the preset value for totalization display that [D24: Ttl Switch Upper] Setting of the totalization
has been specified as the forward totalized value.
Preset value are determined using D21: Ttl Set switch value (upper 2 digits).
Val Lower and D22: Ttl Set Val Upper. In case
that “Start” is sellected, the count for totalization
This parameter sets the upper 2 digits of the
starts from the preset value. Setting of zero as the 8-digit totalization switch value.
preset value allows the zero-reset function to be
implemented.
Preset Rev Total Sets the preset value for totalization display that [D30: Ttl User Select] Selection of the use of
has been specified as the reverse totalized value.
Preset value are determined using D21: Ttl Set special totalization unit.
Val Lower and D22: Ttl Set Val Upper. In case This parameter specifies whether or not special
that “Start” is sellected, the count for totalization
starts from the preset value. Setting of zero as the units are used for totalization units. Actual setting
preset value allows the zero-reset function to be of these units is carried out using D31: Ttl User
implemented.
Unit.
T0620.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-18 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
(4) Menu E: Pulse Setting items [E13: Pulse Low Cut] Setting of the pulse-output
Menu E contains items relevant to pulse output. stop range
This parameter allows the settings to be made
which prevent pulse output when the flow rate
is at or below the low-cut setting value. In cases
NOTE
where there are multiple ranges or forward/reverse
ranges, low cut is carried out at the setting value
For pulse output from the DO terminal, set F20: for the smallest span (i.e., an integer between 0
DO Function to “Pulse Output.” and 100 %).
Example: When the first range = 10 m3/h, the
second range = 100 m3/h, and the Pulse
[E10: Pulse Unit] Setting of the pulse units Low Cut = 3 %, no pulse output is
→ Refer to the description of parameter carried out at flow rates of 0.3 m3/h or
B32: Pulse Unit lower.
NOTE
Open (OFF)
Pulse width F0603.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-19 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-20 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Note 1: When these functions are selected, two terminals become necessary for status output, the DO and
DIO terminals which cannot be used for other functions.
T0623.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-21 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[F22: DO Active Mode] Setting of the active Example 1: When the “Pulse Output” function is
mode for DO terminal selected for the DO terminal and the
E12: Pulse Width is 1 ms, the following
Operations are performed in accordance with the
signals are output from the terminal.
following table when the active mode has been set
to “Closed (On) Act” using this parameter. Operat-
Closed (On)
ing patterns are reversed when the active mode
has been set to “Open (Off) Act.” Open (Off)
1 ms 1 ms
Selected Condition of DO terminal
Closed (On) Active
Open (Off) Active
function Open (Off) Closed (On) F0604-1.EPS
T0624.EPS
[F23: DIO Active Mode ] Setting of the active
mode for DIO terminal Example 3: When the “Total Switch (O)” function is
selected for the DO or DIO (O) termi-
Operations are performed in accordance with the nal, the following signals are output
following table when the active mode has been set from the terminal.
to “Closed/Short Act” using this parameter. Oper-
Closed (On)
ating patterns are reversed when the active mode
Open (Off)
has been set to “Open/Open Act.”
Equal or above setting Equal
value or above setting value
Selected Condition of DIO (O) terminal
Closed (On) Active Open (Off) Active
function Open (Off) Closed (On) F0604-3.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-22 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Multiple ranges setting · When the instantaneous flow rate input exceeds
100% of the range, transition to the next range
Parameters from F30 to F41 are used with (up to four ranges) is carried out automatically.
the automatic multiple ranges and the multiple Furthermore, when the flow is in reverse, the
ranges switching via an external status input. reverse range is automatically selected.
The followings will describe the setting method · Range switching can be confirmed according to
for each range. the status of the DO and DIO status output ter-
minals. Refer to Table 6.4.1: Status Output for
Automatic Range Switching for details of status
The multiple ranges use the following parameters: output conditions for each range.
[B23: Flow Span] Setting of the flow rate span
[Setting of the forward No. 1 range] Status Output for Automatic Multiple Ranges
[F30: Forward Span 2] Setting of the forward Switching
No2. range Operations are performed in accordance with the
[F31: Forward Span 3] Setting of the forward following table when the active mode has been
No.3 range set to “Closed (On) Act” using F22: DO Active
[F32: Forward Span 4] Setting of the forward Mode or “Closed/Short Act” using F23: DIO Ac-
No.4 range tive Mode. Operating patterns are reversed when
[F33: Reverse Span 1] Setting of the reverse the active mode has been set to “Open (Off) Act”
No.1 range using F22: DO Active Mode or “Open/Open Act”
using F23: DIO Active Mode.
[F34: Reverse Span 2] Setting of the reverse
No.2 range Table 6.4.1 Status Output for Automatic Multiple Ranges
[F40: Auto Range Hys] Setting of the automatic Switching
multiple ranges hysteresis width. Status output
[F41: Bi Direction Hys] Setting of the forward/ Parameter setting
Function DO
DIO
reverse flow measurement hysteresis width. F20 F21 DIO (O) DIO (I)
No No Forward single
– – –
Function Function range
Fwd/Rev No Forward Open – –
Rngs (O) Function Auto forward/ Reverse Closed – –
No Fwd/Rev reverse 1 range Forward – Open –
Function Rngs (O) Reverse – Closed –
Auto 2 No Forward 1 range Open – –
Rngs (O) Function Auto forward Forward 2 range Closed – –
No Auto 2 2 ranges Forward 1 range – Open –
Function Rngs (O) Forward 2 range – Closed –
Forward 1 range Open Open –
Auto 3 Auto 3 Auto forward
Forward 2 range Closed Open –
Rngs (O) Rngs (O) 3 ranges
Forward 3 range Open Closed –
Forward 1 range Open Open –
Auto 4 Auto 4 Auto forward Forward 2 range Closed Open –
Rngs (O) Rngs (O) 4 ranges Forward 3 range Open Closed –
Forward 4 range Closed Closed –
Forward 1 range Open Open –
Fwd/Rev Auto 2 Forward 2 range Open Closed –
Rngs (O) Rngs (O) Reverse 1 range Closed Open –
Auto forward/ Reverse 2 range Closed Closed –
reverse 2 rangesForward 1 range Open Open –
Auto 2 Fwd/Rev Forward 2 range Closed Open –
Rngs (O) Rngs (O) Reverse 1 range Open Closed –
Reverse 2 range Closed Closed –
*: “No Function” is the default value. Only DO or DIO is used for
single or dual ranges; accordingly, the unused terminal remains
at the default value (i.e., No Function) and can therefore be used
for other functions.
*: “DIO (O)” indicates DIO function is used for output. T0627.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-23 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
B23: Flow F30: Forward F31:Forward Minimum span % within multiple and
Span Span 2 Span 3 forward/reverse (set using F41) F0606.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-24 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Table 6.4.2 Multiple Ranges Switching via External Status Set the spans for the ranges to be used.
Input No.1 range≤No.2 range
Status output
Parameter setting
Function DIO If “instantaneous flow rate % (FR)” is selected for Display
DO
F20 F21 DIO (O) DIO (I) Mode and A10: FLOW RATE (%) is selected for BRAIN
External forward communication, the instantaneous flow rate % and following
No Ext 2 2 ranges, Forward 1 range – – Open symbols will be displayed only for multiple ranges and
Function Ranges (I) Forward 2 range – – Short forward/reverse flow measurement.
Forward 1 range Open – Open Forward No.1 range: [F1] Reverse No.1 range: [R1]
Fwd/Rev Ext 2 External forward/ Forward 2 range Open – Short Forward No.2 range: [F2] Reverse No.2 range: [R2]
Rngs (O) Ranges (I) reverse 2 ranges Reverse 1 range Closed – Open F0607.EPS
Reverse 2 range Closed – Short
Ext 2 Ext 2 External forward
Forward 1 range Open – Open
Answer (O) Ranges (I) 2 ranges, with Forward 2 range Closed – Short [F40: Auto Range Hys] Setting of automatic
answerback
*: “No Function” is the default value. Only DIO is used for dual range-switching hysteresis width
ranges; accordingly, the unused DO terminal remains at the Automatic switching takes place for multiple range
default value (i.e., No Function) and can therefore be used for
other functions. switching when 100 % of the range is exceeded,
*: “DIO (I)” indicates DIO function is used for input. T0628.EPS and this parameter allows a hysteresis width to be
set for this switching.
Refer to Table 6.4.1: Multiple Ranges and Hyster-
esis Widths.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-25 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
NOTE
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-26 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[%] [%]
100 100 High High Alarm
90
80 80
High Alarm High Alarm
20
Low Low Alarm
0 Time (t) 0 Time (t)
H H
DO DO
L Alarm set to L Alarm set to
-110% indicates -110% indicates
that the alarm is that the alarm is
LL disabled. HH disabled.
DIO DIO
HH Alarm set to 110% indicates
LL Alarm set to -110% indicates
that the alarm is disabled.
that the alarm is disabled.
NOTE
• Although the same items can be selected us-
ing the DO terminal (selected with F20) and
the DIO terminal (selected with F21), output is
identical for both.
• Setting values of -110 % and 110 % are used to
disable corresponding functions; and according-
ly, status output can be customized for specific
purposes.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-27 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[G14: H/L Alarm Hys] Setting of upper/lower [G22: 4-20 mA Burn Out] Display of the current
alarm value hysteresis width output during a CPU failure.
This parameter sets the hysteresis width for upper This parameter sets the current output direction
and lower alarm value, using a % value of the for a CPU error (i.e., burnout). Note that communi-
maximum span. cation will not be possible if such an error occurs.
With the standard specification, this is set to High
and 25 mA is output when an error occurs. Low is
Output Example setup for optional code C1, and in such a case, 0
The hysteresis width is set to 5 %. mA is output when a failure occurs.
Settings are:
NOTE
G10: Low Alarm = 30 %
G11: High Alarm = 80 % The current output direction for a CPU failure
G12: Low Low Alarm = 20 % (i.e., burnout) can be changed. Refer to selection
10.2.1: Setting of Burnout Switch.
G13: High High Alarm = 90 %
G14: H/L Alarm Hys = 5 %
[G30: Alm-Setting] Alarm recognition of "Setting
[%] Alarm"
100 High High Alarm
This parameter specifies whether the setting alarm
90 will be recognized as an alarm.
85%
80
75%
High Alarm
Setting Function
Instantaneous flow rate No Not recognized as an alarm
50 Yes Recognized as an alarm
Low Alarm T0630.EPS
35%
30
25%
20 [G31: Alm-Sig Over] Alarm recognition of "Signal
Low Low Alarm Overflow Alarm"
0 Time (t) This parameter specifies whether the signal over-
75% 35%
H L flow process alarm will be recognized as an alarm.
DO
A signal overflow occurs when there is an error in
85% 25%
the input signal.
HH LL
DIO
Setting Function
Select “H/L Alarm(O)” for F20: DO Function
Select “HH/LL Alarm(O)” for F21: DIO Function No Not recognized as an alarm
Select “Closed (On) Act” for F22: DO Active Mode Yes Recognized as an alarm
Select “Closed/Short Act” for F23: DIO Active Mode
F0611.EPS T0631.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-28 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-29 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[G42: Alm Record Time 1] Display the operation (7) Menu H: Display Setting items
time of alarm record 1 Menu H contains setting items relevant to the
This parameter is used to display the operation display unit.
time at which the alarm indicated by G41: Alm
Record1 was occurred. For example, “1D23:45”
indicates that an alarm was triggered at the op- [H10: Display Select1] Setting of the first line for
eration time of 1 day, 23 hours, and 45 minutes. display unit
→ Refer to the description for parameter B40
[G43: Alm Record 2] Alarm record 2 This parameter selects the display content of the
This parameter is used to display the second first line for display unit.
most-recent alarm, and the alarms that can be
displayed are the same as those for G41: Alm [H11: Display Select2] Setting of the second line
Record1. of display unit
→ Refer to the description for parameter B41
[G44: Alm Record Time 2] Display the operation This parameter selects the display content of the
time of alarm record 2 second line for display unit.
This parameter is used to display the operation [H12: Display Select3] Setting of the third line for
time at which the alarm indicated by G43: Alm display unit
Record2 was occurred. For example, “1D23:45” → Refer to the description for parameter B42
indicates that an alarm was triggered at the op- This parameter selects the display content of the
eration time of 1 day, 23 hours, and 45 minutes. third line for display unit.
[G45: Alm Record 3] Alarm record 3 [H20: Display Cycle] Setting of the display cycle
This parameter is used to display the third This parameter sets the cycle for the display unit’s
most-recent alarm, and the alarms that can be display-response speed of display unit. Settings
displayed are the same as those for G41: Alm should be made in accordance with the measure-
Record1. ment environment by, for example, setting a longer
display cycle when using the equipment in low
[G46: Alm Record Time 3] Display the operation temperatures.
time of alarm record 3
This parameter is used to display the operation [H30: Language] Selection of display language
time at which the alarm indicated by G45: Alm → Refer to the description for parameter B10
Record3 was occurred. For example, “1D23:45” This parameter can be used to select the lan-
indicates that an alarm occurred at the operation guage for use on the display unit.
time of 1 day, 23 hours, and 45 minutes.
(8) Menu J: Auxiliary Function Setting
[G47: Alm Record 4] Alarm record 4 items
This parameter is used to display the fourth
Menu J contains setting items such as the flow
most-recent alarm, and the alarms that can be
direction, rate limits, and current output limits.
displayed are the same as those for G41: Alm
Record1.
[J10: 4-20 mA Low Cut] Setting of the low-cut
[G48: Alm Record Time 4] Display the operation range for current output
time of alarm record 4 This parameter is used to force current output to
This parameter is used to display the operation 0 % (i.e., 4mA) in the vicinity of 0 % output, and
time at which the alarm indicated by G47: Alm setting for the current (4-20 mA) output low cut is
Record4 was occurred. For example, “1D23:45” made using a percentage of the smallest flow rate
indicates that an alarm occurred at the operation span. However, the low cut function will be termi-
time of 1 day, 23 hours, and 45 minutes. nated if this parameter is set to 0 %.
The indications of the instantaneous flow rates (%,
Actual instantaneous flow rate, mA, Bar graph) on
the display unit are the same action.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-30 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Example: Situation where low cut is set to 10 % The indications of the instantaneous flow rates (%,
Output Setting value Actual instantaneous flow rate, mA, Bar graph) on
the display unit are the same action.
5.6mA(10%)
Example: Situation where low limit is set to 10 % [J20: Flow Direction] Setting of the flow direction
Output Upon shipment from the manufacturing plant, the
system is setup such that flow in the same direc-
tion, as shown by the direction of the arrow mark
5.6mA(10%) on the flowtube, will be measured as forward flow;
4mA however, this parameter can be used to set “Re-
Input
verse” so that flow in the opposite direction to the
F0655.EPS
arrow mark will be treated as forward.
Note: This function does not apply to measure-
NOTE ment in both the forward and reverse direc-
tions. If you like to activate dual direction
If “2.4 mA or less” has been set for G21:4-20mA measurement capability please use “Fwd/
Alarm Out, 2.4 mA or less will be output upon Rev Rngs (O)” from either F20: DO Func-
an alarm occurrence, regardless of the low limit tion or F21: DIO Function.
setting.
Setting Function
NOTE Forward Forward direction corresponds with arrow mark
Reverse Forward directions is opposite to arrow mark
• If the setting value for the low limit is not
less than the high limit value (as set using T0638.EPS
J12: 4-20mA High Lmt), the setting alarm [J21: Rate Limit] Setting of the rate limit value
“54: 4-20 Lmt Err” will be displayed. • This parameter is used in situations where sud-
• This parameter has no effect on pulse output or den noise cannot be eliminated by increasing
the totalization function. the damping time constant.
• In situations where step signals or sudden noise
[J12: 4-20 mA High Lmt] Setting of the high limit signals caused by slurry or the like are entered,
for current output. This parameter is used to this parameter is used to set the standard for
restrict high current portions of current (4-20 determining whether an input corresponds to
mA) output, and it is initially set to 120 %. Set- a flow measurement or noise. Specifically, this
ting should be performed when a lower value is determination is made using upper and lower
required for the higher limit. rate limits and using the dead time.
• Rate limit values are set using a percentage of
the smallest range. The range of deviation per
one calculation cycle should be input.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-31 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[J22: Dead Time] Setting of dead time ing, and therefore the output turns out to be as
This parameter sets the time for application of the indicated by the solid line.
rate limit, and if a value of 0 is set, the rate limit (2) Subsequent flow values within the dead time
function will be terminated. zone correspond to signals of post-damping
flow value + rate limit value (1%).
NOTE
(3) Since input signals do not return to within
Determining rate limit value and dead time the rate limit value during the dead time, it is
determined at (c) that this signal is a flow-rate
signal.
Rate limit value: Dead time (T0):
Determines the level for output This is to be determined using the (4) The output signal becomes a damped curve
fluctuation cutoff. For example, if output fluctuation width. If noise
and compliance with the step signal begins.
this is set to 2 %, noise above exceeds the dead time as shown
and below 2 % will be in the diagram below, the dead Three seconds after determination of a flow-
eliminated as shown in the time should be made longer.
rate signal in the above figure, a level of 63.2
T0
% is reached.
2%
2% Example 2: Slurry noise
T0 Input: 0 to 10%
F0613.EPS Slurry noise Damping time constant: 1 s
Dead time: 1 s
Flow-rate value
after rate limit Rate limit value: 1%
• Signal processing method: +1% processing
A fixed upper and lower limit value is setup with -1%
In the figure on the left, it
is determined that the
respect to the primary delay response value for slurry noise signal is not a
Flow-rate value after damping
the flow-rate value obtained during the previous flow-rate signal.
Time
sampling, and if the currently sampled flow rate Dead time: 1 s F0615.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-32 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[J30: Power Synch] Setting of power synchroni- [J40: Memo 1] Setting of memo 1
zation [J41: Memo 2] Setting of memo 2
This parameter selects whether or not the internal
[J42: Memo 3] Setting of memo 3
frequency is to be synchronized with that of the
power supply. These parameters are used with the memo func-
tion, and up to 16 characters can be set for each.
T0642-2.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-33 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[K11: Adhesion Level 1] Setting the resistance B50/M10: Auto Zero Exe.
value for adhesion diagnostic level 1 Specifically, the correction values displayed,
This parameter sets the resistance value (in MΩ) and it is also possible to directly enter correction
for judgment of Level 1. values.
(11) Menu N: Loop Test Setting items
[K12: Adhesion Level 2] Setting the resistance
Menu N contains items that are relevant to loop
value for adhesion diagnostic level 2
testing.
This parameter sets the resistance value (in MΩ)
for judgement of Level 2. [N10: Test Mode] Setting for loop test execution
Setting Function
[K13: Adhesion Level 3] Setting the resistance Normal No execution of loop testing
value for adhesion diagnostic level 3 Test Loop testing is started
This parameter sets the resistance value (in MΩ) T0641.EPS
for judgment of Level 3.
*: The warning 80: Adhesion Wng is displayed
IMPORTANT
when the adhesion level reaches Level 3.
*: If “Warning Output” has been selected for F20: (1) Test output has priority over flow rate mea-
DO Function or F21: DIO Function, then sta- surement signals. When carrying out flow
tus output will be performed when the adhesion rate measurements, be sure to always return
level reaches Level 3. to “Normal”.
(2) Upon entry to the Test Mode, all output termi-
[K14: Adhesion Level 4] Setting 4 for adhesion nals will simultaneously adopt test condition.
diagnostic level (3) “Normal” will be restored when the power
This parameter sets the resistance value (in MΩ) is turned off or when 30 min have elapsed
for judgment of Level 4. since entry to Test Mode.
*: The process alarm 33: Adhesion Alm is dis- (4) In Test Mode, the warning 83: Fix Cur Wng
played when the adhesion level reaches Level will be displayed as a warning message. (For
4. more details, refer to Section 6.5 Alarm Func-
*: Alarm output will be preformed if “Alarm Output” tions.)
has been selected for F20: DO Function or
F21: DIO Function, and “Yes” for G34: Alm-
Adhesion. [N11: Test Output Value] Setting for test output
values
[K15: Adh Measure Value] Displays resistance During loop testing, current output (4-20 mA),
value for the adhesion diagnose. totalization, and pulse will be output in accordance
This parameter displays the value measured using with this parameter’s setting, and values can be
the adhesion diagnostic function (in MΩ). set when “Test” has been selected for N10: Test
When "No" is selected for K10: Adhesion Check, Mode. With multiple ranges or when performing
his parameter displays the indetermination value. forward/reverse flow measurements, setting should
be done using a percentage of the maximum
range.
(10) Menu M: Automatic Zero Adjust-
ment Function Setting items
[N30: Test DO] Setting for DO terminal condition
Menu M contains items that are relevant to auto- during testing
matic adjustment of the zero point. This parameter sets the condition of the DO termi-
nal during loop testing. Setting is possible when
[M10: Auto Zero Exe] Execution of automatic zero “Test” has been selected for N10: Test Mode.
adjustment function
Setting Function
→ Refer to the description of parameter B50.
Open (Off) DO terminal in Open (Off) condition
Closed (On) DO terminal in Closed (On) condition
[M11: Magflow Zero] Results of automatic zero Pulse Outputs pulses as were specified with
adjustment Menu E. *
This parameter is used to display the results T0642.EPS
obtained from
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-34 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
*: If the pulse scale is 0 pps or there is a pulse This parameter is used to indicate whether or not
setting error, the DO terminal is “Open (Off).” write protection is currently on.
[N31: Test DIO (O)] Setting for the DIO terminal
Default setting (Enable)
condition during testing
P20:Write Protect
This parameter sets the condition of the DIO ter- No
minal during loop testing. Setting is possible when
“Test” has been selected for N10: Test Mode. F0617.EPS
Input mode Must be set when N32: Test DIO(I) is used P20:Write Protect
to display the condition of the DIO terminal. Yes
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-35 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
To change a password, first of all use the pass- Upon occurrence of an alarm
word originally set with P21: Enable Wrt Passwd (example: a System alarm)
to release the write protect function, and then set Alarm name
the new password. Alternatively, if it is desired to Description
NOTE
6.5.2 Alarm Selection
If you should forget your password, the joker
password can be used to temporarily release writ The display and output differs depending on the
protection function. To obtain the joker password, alarm levels. Certain types of alarm may or may
please contact your nearest YOKOGAWA sales not be recognized as alarms, according to the set-
office. tings of certain parameters. The parameters that
are relevant to this function as follows.
6.5.1 Alarm Levels [F21: DIO Function] Setting of the function for the
DIO status input/output terminal
Alarms are classified into the following four differ-
ent types based on level. [F22: DO Active Mode] Setting of the active
mode for DO terminal
Alarm Level Description
[F23: DIO Active Mode] Setting of the active
System Major Device error or inability to obtain correct
alarm breakdown measurements. Replacement will be required. mode for DIO terminal
Intermediate Device is normal but process-related errors
Process level break
make correct measurement impossible. [G21: 4-20mA Alarm Out] Setting of the current
alarm down Maintenance or the like will be required. output during an alarm occurring.
Device is normal but errors have been made
Setting Minor
in the setting of parameters. Functions not [G30: Alm-Setting] Alarm recognition of “Setting
related to the incorrect settings are operating
alarm breakdown
normally. Alarm”
The incorrect settings must be corrected.
Device and measurements are normal but a [G31: Alm-Sig Over] Alarm recognition of “Signal
Warning Warning warning is occurred. Overflow Alarm”
T0645.EPS
When an alarm has been occurred, the corre- [G32: Alm-Emp Pipe] Alarm recognition of
sponding name, description, and suitable counter- “Empty Pipe alarm”
measure will be displayed on the display unit. The [G33: Alm-HH/LL] Alarm recognition of “HH/LL
normal Display Mode and Alarm Mode may be Alarm” (Refer to the descriptions of G12 and G13
displayed alternatively. When a warning has been for more details regarding HH and LL alarms.)
issued, the corresponding content will be shown in
[G34: Alm-Adhesion] Alarm recognition of “Adhe-
the third line in the Display Mode.
sion Alarm”
[G41: Alm Record 1] Alarm record 1
[G43: Alm Record 2] Alarm record 2
[G45: Alm Record 3] Alarm record 3
[G47: Alm Record 4] Alarm record 4
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-36 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Selection
Alarm Alarm 4-20 mA Totali- Pulse Display Alarm
description(parameter output output zation output unit record
number)
YES Open Alarm Mode
(Off ) Fixed Stopped Stopped (Message) Recorded
(G31)
30 Sig Overflow Input signal error
NO Closed Continu- Continu- Continu-
(G31) (On) ous (*) ous (*) ous (*) Display Mode No
YES Open Alarm Mode
(G32) (Off ) Fixed Stopped Stopped (Message) Recorded
Flowtube is not
31 Empty Pipe filled with fluid NO Closed Continu- Continu- Continu-
(G32) (On) ous (*) ous (*) ous (*) Display Mode No
YES Open Alarm Mode
(G33) (Off ) Normal Normal Normal (Message)
32 HH/LL Alm HH/LL Alarm No
NO Closed operationoperationoperation
(On) Display Mode
(G33)
YES Open Alarm Mode
Electrode (G34) (Off ) Fixed Stopped Stopped (Message) Recorded
33 Adhesion Alm adhesion alarm
NO Closed Continu- Continu- Continu-
(G34) (On) ous (*) ous (*) ous (*) Display Mode No
Note: • Alarm is output only when F20: DO Function or F21: DIO Function is set to “Alarm Output.”
• The operation when “Open (Off ) Act” is set for F22: DO Active Mode or “Open/Open Act“ is set for
F23: DIO Active Mode is shown in above table.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA
Alarm Out.
*: Although outputs are continuous, output values are not guaranteed. T0647.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-37 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Note: • Alarm is output only when F20: DO Function or F21: DIO Function is set to “Alarm Output.”
• The operation when “Open (Off ) Act” is set for F22: DO Active Mode or “Open/Open Act“ is set for
F23: DIO Active Mode is shown in above table.
• 4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA
Alarm Out. T0648.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-38 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Process Alarms (Device is normal but process-related errors make correct measurement impossible.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Carry out an investigation as follows:
· Check the signal cable for breakage.
Check signal cable · Check for contact between signal cable,
30:Sig Overflow Input signal error power cable, and excitation cable.
and grounding
· Check for stray currents in the fluid.
· Check the grounding.
Fill flow tube with Flowtube is not filled with
31:Empty Pipe Fill the flowtube with fluid.
fluid fluid
Flow rate alarm for greater
32:HH/LL Alm Check the flow rate and than High-High limit value Check the flow rate and setting value of
setting value or less than Low-Low limit High High limit and Low Low limit.
value.
33:Adhesion Alm Clean electrodes Electrode adhesion alarm Clean the electrodes.
T0650.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-39 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Setting Alarms (Device is normal but errors have been made in the setting of parameters.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Check whether parameters
Check parameter C40, C41, Span flow velocity setting is 11 m/s or C40, C41, and C42 are
50:Span > 10m/s correct.
and C42 more
In case that multiple range or
forward and reverse flow
Check parameter C40, C41, Span flow velocity setting is 0.05 m/s or measurement functions is
51:Span < 0.1m/s less used, check whether
and C42
parameters F30 through F34
Check parameter D10 and are correct.
52:TTL>10000p/s Totalization rate is 11000 pps or more
D11 Check whether parameters
Check parameter D10 and D10 and D11 are correct.
53:TTL<0.0001p/s Totalization rate is 0.00005 pps or less
D11
Check parameter J11and J12 The condition [4-20 low limit (J11) < Check whether parameters J11
54:4-20 Lmt Err
4-20 high limit (J12)] is not satisfied. and J12 are correct.
The condition [No. 1 range < No. 2
Check whether parameters
55:Multi Rng Err Check parameter F30 to F34 range < No. 3 range < No. 4 range] is
F30 through F34 are correct.
not satisfied for multiple ranges.
The condition [High Alarm (G11) –
Low Alarm (G10) > H/L Alarm Hys
Check whether parameters
56:H/L HH/LL Set Check parameter G10 to G14 (G14)] or the condition [High High
G10 through G14 are correct.
Alarm (G13) – Low Low Alarm (G12)
> H/L Alarm Hys (G14)] is not satisfied.
Mass units have been selected for Base
Check parameter C40, C45, Flow Unit (C40) but density is set to 0. Check whether parameters
57:Dens Set Err C40, C45, and C46 are correct.
and C46
Pulse rate is 11000 pps or more with
Check parameter E10, E11, 50% duty selection.
60:PLS > 10000p/s Pulse rate is 10000 pps or more with
and E12
0.05 ms selection.
Check parameter E10, E11, Pulse rate is 5000 pps or more with
61:PLS > 5000p/s 0.1 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 1000 pps or more with
62:PLS > 1000p/s 0.5 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 500 pps or more with
63:PLS > 500p/s 1 ms selection.
and E12 Check whether parameters
Check parameter E10, E11, Pulse rate is 25 pps or more with E10, E11, and E12 are correct.
64:PLS > 25p/s 20 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 15 pps or more with
65:PLS > 15p/s 33 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 10 pps or more with
66:PLS > 10p/s 50 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 5 pps or more with
67:PLS > 5p/s 100 ms selection.
and E12
Check parameter E10, E11,
70:PLS<0.0001p/s Pulse rate is 0.00005 pps or less.
and E12
Measure Mode (C20) is set to Enhanced
Check whether parameter C20
71:Meas Mod Set Check parameter C20 DF without selecting an optional code
is correct.
HF1 or HF2.
A value of 3000.1 mm or more is set for Check whether parameter C32
72: Size Set Err Check parameter C32 Nominal Size (C32). is correct.
The condition in Adhesion detection Check whether parameters K11,
73: Adh Set Err Check parameter K11 to K14 level, Level:1<Level:2<Level:3<Level:4 K12, K13 and K14 are correct.
is not satisfied.
T0651.EPS
Setting Alarms (Device and measurements are normal but a warning is issued.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Slight adhesion to Clean and check the electrodes.
80:Adhesion Wng – electrodes. Refer to parameter K13.
Carry out adjustment as follows:
· Check if the flowtube is filled with
Results of automatic zero fluid.
82:Auto Zero Wng – adjustment are higher than · Check if the flow velocity is
the rated values. completely zero.
· Check the condition of grounding.
Confirm whether the flow rate is in
excess of the upper limit (108 %) or
83:Fix Cur Wng – The current value is fixed. below the lower limit (-8 %), or
whether upon entry to the Test Mode
or not.
84:Disp Over Wng Overflow in the display digits during Check whether parameter C43 is
–
(only for display unit) instantaneous flow rate display. correct.
Investigate whether the display unit
90:Disp SW Wng Display unit switches are not
cover is fitted or whether the cover’s
(only for display unit) – operating. glass surface is dirty.
T0652.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-40 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
6.6 Precautions for the (1) Confirm that the meter factor indicated on
your RXF integral flowmeter’s data plate has
RXF Integral Flowme- been set to the parameters.
ter (2) The meter factors are crucial in ensuring that
the electromotive force is correctly in propor-
Although the RXF integral flowmeter’s software tion to the flow velocity and are determined
configurations are the same as the RXFA14, care at the manufacturing plant by actual-flow
must be taken in setting the following parameters: calibration.
NOTE
Setting the meter factor
Meter factor for the RXF integral flowmeter
(PS=Pmax at 20°C)
-10 to 90˚C
mA (0 - 750 Ω )
VDC 0.2A MAX
F0623.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-41 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Blank Page
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 6-42 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
Power screen.
ON/OFF key (See Table 7.1.)
ENT key
• Enters selected b) Entering letters (i.e., A through Z)
items Press a shift key followed by an alphanumeric
• Sets up input data
on the connected key to enter the letter shown on the same side
Alphanumeric keys
instrument. as the shift key. The shift key must be pressed
• Enter digits
• Runs BT200
• Enter letters in for each letter being input.
functions.
combination with
the shift keys. Shift keys
F0701.EPS Left-side letter on an Right-side letter on
alphanumeric key an alphanumeric key
Figure 7.1 Key Layout
F0704.EPS
If BATTERY is displayed on
the screen, it indicates that the Entry Key-in sequence
battery voltage has dropped.
W
Menu Panel Messages
MENU
A:DISPLAY
BATTERY IC
B:EASY SETUP Menu choices
C:BASIC SETUP
D:TOTAL SET
(up to six are J. B
Panel titles E:PULSE SET displayed) T0702.EPS
F:STATUS FUNC
HOME SET ADJ ESC Function commands
Parameter Panel
PARAMETER
A10:FLOW RATE(%)
80.0% Parameters
A11:FLOW RATE (up to three are
100m3/h
A12:FLOW RATE(mA) displayed)
1417
DATA DIAG PRNT ESC Function commands
F0702.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
Use the function key [F2] CAPS to select between Function Command List
uppercase and lowercase (for letters only). The Command Function
case toggles between uppercase and lowercase
ADJ Displays the ADJ menu
each time you press [F2] CAPS.
CAPS/caps Selects uppercase or lowercase
Entering uppercase Entering lowercase
CODE Selects symbols
CODE CAPS CLR ESC CODE caps CLR ESC
CLR Erases inut data or deletes all data
DATA Updates parameter data
Entry Key-in Sequence DEL Deletes one character
to lower case DIAG Calls the self-check panel
Boy ESC Returns to the most recent display
( B ) ( o ) ( y )
HOME Displays the menu panel
F0807.EPS
NO Quits setup and returns to the pre-
vious display
Use the function key [F1] CODE to enter symbols.
OK Proceeds to the next panel
The following symbols will appear in sequence,
one at a time, at the cursor each time you press PARM Enters the parameter number setup
mode
[F1] CODE:
SET Displays the SET menu
/ . – , + * ) ( ’ & % $ # ” !
SLOT Returns to the slot selcection panel
To enter characters next to these symbols, press [
> ] to move the cursor. UTIL Calls the utility panel
*COPY Prints out parameters on display
Entry Key-in Sequence
*FEED Paper feed
symbol command
l/m *LIST Lists all parameters in the menu
(I) (/) (m) *PON/POFF Automatic printout mode on or off
T0803.EPS
*PRNT Changes to the print mode
(2) Function Keys *GO Starts printing
The functions of the function keys depend on the *STOP Cancels printing
function commands on display. * Available on BT200-P00 (with printer)
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC Function comman
Function keys
F0808.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 7-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
G
IMPORTANT
1 N/– POWER
EX SUPPLY
EXCITATION 2 L/+
CURRENT
C + OUTPUT
SIGNAL
SA
A
–
+
I
After approximately 5 min of inactivity, the
Auto Power-Off function will operate to turn your
DO
B –
SB + DIO
–
BT200 off.
4-20 mA
Instrument room
DC signal
Relay transmission Terminal Receiving
NOTE
I terminals line board instrument
I
In case of BT200, the parameters are displayed
Receiving in English only.
resistance:
250 to 450 Ω Even if the language with the exception of English
is selected at B10/H30: Language, the param-
BT200 BT200 BT200 BT200
F0705.EPS
eters are displayed in English upon BT200.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
(2) Upload/download function of BT200 7.2.3 BT200 Screens & Flow Data Dis-
Upload function is used when the parameters of play
one RXFA14 are copied to the BT200. And down- Use the following procedure to display flow data
load function is used when the parameters copied on the BT200.
to the BT200 are set to another RXFA14.
• The display of flow data is updated every 5 s.
For details, refer to BT200 User’s Manual (IM
1C0A11-01E).
When the BT200 is turn
The targeted parameters for upload and download –––WELCOME––– on, the message“Please
are following. BRAIN Terminal
ID:BT200
wait ...” is displayed for
several seconds, and the
Check cable connection the screen on the left is
and press the Ent key. shown.
UNTIL FEED Press the key
Item Name (BRAIN) Item Name (BRAIN)
C11/B20 FLOW DAMPING D10/B30 TOTAL UNIT
C31 SIZE UNIT D11/B31 TOTAL SCALE
C32 NOMINAL SIZE E10/B32 PULSE UNIT
PARAMETER Press the key
C40/B21 FLOW UNIT E11/B33 PULSE SCALE 01:MODEL to display initial data as
AXFA11-A00 shown on the left.
C41/B22 TIME UNIT H10/B40 DISP SELECT1 02:TAG NO
FI-1101 The tag number
C42/B23 FLOW SPAN H11/B41 DISP SELECT2 03:SELF CHECK
GOOD specified upon
C43/B24 FLOW DECIMAL H12/B42 DISP SELECT3 OK ordering is found
T0705.EPS here.
Press the (
OK )o
key.
or
( ESC
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 7-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
IMPORTANT
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
MENU When the cursor has been SETTING When selected, “l (Liter)”
A:DISPLAY moved to B: EASY SETUP, B21:FLOW UNIT will begin to flash on and
B:EASY SETUP m3
C:BASIC SETUP press the key to l(Liter) off: Press the
D:TOTAL SET access Parameter Screen key once again to confirm
E:PULSE SET PRINTER OFF
(Sub-Item Parameter F2:PRINTER ON your selection.
F:STATUS FUNC
Search Mode). FEED POFF NO In case of cancel, press the
HOME SET ADJ ESC
( NO ) key.
X2
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 7-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
MENU When the cursor has been SETTING The data will begin to flash
A:DISPLAY moved to B: EASY SETUP. B23:FLOW SPAN on and off: Press the
B:EASY SETUP 100.000l/min
C:BASIC SETUP Press the key to + 120 key once again
D:TOTAL SET to confirm your entry.
E:PULSE SET access Parameter Screen PRINTER OFF
(Sub-Item Parameter F2:PRINTER ON In case of cancel, press the
F:STATUS FUNC
HOME SET ADJ ESC Search Mode). FEED POFF NO ( NO ) key.
In the parameter list, the The value for flow span has
PARAMETER SETTING
B10:LANGUAGE parameter for setting of flow B23:FLOW SPAN been changed to
English rate span is B23: FLOW 120.000l/min “120.000 l/min.”
B20:FLOW DAMPING
3.0s SPAN. Press the Press ( OK ) key to
B21:FLOW UNIT exit Setting Screen
l(Liter) key four times to move the
DATA DIAG PRNT ESC
cousor to B23. FEED NO OK
(Parameter Replacement
(Note that the key Mode), and then the display
returns to Parameter
can be used to move the
Screen (Sub-Item
cursor to the top of the next
Parameter Search Mode) of
screen.
Menu B items.
In case of redo of setting,
press the ( NO ) key
or
to return the Setting Screen
(Parameter Replacement
Mode).
PARAMETER After moving the cursor to
B22:TIME UNIT
/min B23: FLOW SPAN, press the ( OK )
B23:FLOW SPAN key to access
100.000l/min
B24:FLOW DECIMAL Setting Screen (Parameter
l Replacement Mode). PARAMETER The changed setting for
DATA DIAG PRNT ESC B22:TIME UNIT flow rate span can now be
/min confirmed.
B23:FLOW SPAN
120.000l/min
B24:FLOW DECIMAL Press the ( ESC ) key
l
to return to Menu Screen
DATA DIAG PRNT ESC
(Major Item Parameter
SETTING In this mode, new values can Search Mode).
B23:FLOW SPAN
100.000l/min be entered into the
+ 0 highlighted field using the F0708.EPS
alphanumeric keys.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-7 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
X4
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 7-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
NOTE
This chapter describes the RXFA14 as an example. The same attention must be paid to the RXF inte-
gral flowmeter. Please note that commands and comments displayed on HART interface products will be
refered to as AXF... . The device description (DD) of AXF is identical to RXF. Please select from HART
interface model AXF for usage with model RXF.
The DD revisions for RXFA14 and Configuration Tool’s can confirm in accordance with the following steps.
If the correct DD is not installed in the HART Configuration Tool, download them from the official HART pro-
gramming sites, otherwise, contact the respective vendors of the Configuration Tool for its upgrade informa-
tion.
1. Confirmation of the device revision for RXFA14
(1) Connect the Configuration Tool to the RXFA14. The device revision can be checked as follows.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
G Control room
1 N/– POWER
EX
EXCITATION 2 L/+ SUPPLY 4 to 20 mA DC signal line
C +
CURRENT
OUTPUT
I
Terminal board
SA –
SIGNAL A +
DO
+
B –
SB + DIO
– – Relaying
terminals
USB
FieldMate
Receiving instrument
I+ I- modem
load resistance:
250 Ohm to 600 Ohm
USB (including cable
RXFA14 resistance)
PC/FieldMate
F0802.EPS
WARNING IMPORTANT
Be sure to set parameters as “Protect” on the If the power of flowmeter is turned off within 30
write protect function after finish of parameter seconds after parameters have been set, these
setting work. settings will be canceled. Accordingly, please
Under extremely rare case, the infra-red switches keep the power on for at least 30 seconds after
may respond unexpectedly in such conditions as setting parameters.
sticking ball of water or extraneous substances
on the surface of display panel glass according
to the principle of infra-red switch operation. NOTE
Its probability rises in such cases as sticking
rain water by storm or other similar situation Before updating any setting, remember to always
and washing up work near flowmeter installation check the data content you want to change as
place. described in Chapter 6: Parameter Description.
Either to illuminate or stop illuminating the infra-
red switches by the flashlight may cause the
mis-reaction. NOTE
Refer to Chapter 6 “Menu P: Parameter Protec-
In case of HART Configuration, the parameters
tion Items” and section “10.2.2” how to use the
are displayed in English only.
write protect function in detail. This menu in the
Even if the language with the exception of Eng-
275 is located at [Hot key] → [2 Wrt Protect
lish is selected at “Language” setting display, the
Menu].
parameters are displayed in English upon HART
Configuration Tool.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Device Status
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
• Burst Mode
IMPORTANT
RXFA14 continuously sends the data via HART
The D/A trim should be executed only at single communication when the burst mode is set on
range mode. If the D/A trim is executed at Bi (any one of PV, % range/current, or process vars/
direction mode, it is possible that the current crnt). The data is sent intermittently as a digital
output becomes 108%. signal at 3 times a second.
The burst mode is set as follows.
(1) Setting the data to be sent
IMPORTANT
• Procedure to call up the display
When “D/A trim” or “Scaled D/A trim” is carried
DD Device Setup → Detailed setup
out, the warning message “83: Fix Cur Wng” is → HART output → Burst option
displayed on the display unit. DTM Configuration → HART → Burst option
T0810.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
NOTE
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
Parameter
Read/Write of BRAIN Data(*1)
protocol
F0803.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-7 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Parameter
Read/Write of BRAIN Data(*1)
protocol
Enhanced DF W C20 –
Low MF (EDF) W C23 –
High MF (EDF) W C24 –
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
Parameter
Read/Write of BRAIN Data(*1)
protocol
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-9 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Parameter
Read/Write of BRAIN Data(*1)
protocol
5 Review 1 Review1 R – –
2 Review2 R – –
3 Review3 R – –
R – –
4 Review4
*1 : Data Renewing
P : Periodic Date Renewing
u/d : Discretionary Renewing
– : Other (Method etc...) F0806.EPS
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
F0807.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-11 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Root Menu
Process Variables A
Device Status B
Diag and Service C
Easy Setup D
Configuration E
Calibration F
Write Protect G
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
A
Process Variables PV R A20 P
PV % rnge R A10 P
PV AO R A21 P
PV Span R B23 P
PV Damping R C11/B20 P
Totl R A30 P
Reverse Totl R A31 P
Dif Totl R A32 P
B
Device Status Process Variable PV
PV % rnge
Totl
Reverse Totl
Dif Totl
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-12 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
C
Diag and Service Output Test Loop test – – –
Test Mode W N10 –
Test DO W N30 –
Test DIO (I) R N32 P
Test DIO (O) W N31 –
D
Easy Setup Tag W C10 u/d
Language W B10/H30 u/d
PV Damping W B20/C11 u/d
Base Flow Unit W B21/C40 u/d
Base Time Unit W B22/C41 u/d
PV Span W B23/C42 u/d
Flow Decimal Pnt W B24/C43 u/d
Total Unit W B30/D10 u/d
Total Scale W B31/D11 u/d
Pulse Unit W B32/E10 u/d
Pulse Scale W B33/E11 u/d
Display Select1 W B40/H10 u/d
Display Select2 W B41/H11 u/d
Display Select3 W B42/H12 u/d
Auto Zero Exe – – –
F0809.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-13 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
E
Configuration Characterize Meter Nominal Size Unit W C31 u/d
Nominal Size W C32 u/d
Measure Mode R C20 –
Low MF R C21 –
High MF R C22 –
Low MF (EDF) R C23 –
High MF (EDF) R C24 –
MF Set – – –
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-14 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. Operation via Hart Communicator
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-15 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
8. Operation via Hart Communicator
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
F
Calibration Auto Zero Exe – M10/B50 –
Magflow Zero R/W M11 –
D/A trim – – –
Scaled D/A trim – – –
G
Write Protect Write protect R P20 P
Enable Wrt Password W P21 –
New Password W P22 –
Software Seal R P23 –
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 8-16 All Rights Reserved. Copyright © 2006, Rota Yokogawa
9. ACTUAL OPERATION
9. ACTUAL OPERATION
After you have installed the flowtube into the
process piping, wired the input/output terminals,
set up the required parameters, and performed
a pre-operation zero adjustment, the magnetic
flowmeter should output an accurate flow signal
from its terminals as soon as flow of the fluid to
be measured begins. This section describes zero
adjustment and the corresponding procedures.
For FOUNDATION Fieldbus protocol, please refer
to IM 01R21F02-01E.
IMPORTANT
All Rights Reserved. Copyright © 2006, Rota Yokogawa 9-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
9. ACTUAL OPERATION
Parameter M10:Auto Zero Exe Touch the switch to move the cursor to
No Execution
Replacement No Execution
“Execution”.
(execute) Mode Execution
NOTE
When no operations are carried out for 20 seconds in the flashing state, the system
will automatically return to the Sub-item Parameter Search Mode. When the opera-
tions except are carried out, the parameter cannot be set.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 9-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
9. ACTUAL OPERATION
Sub-item Major item M:Adjustment When zero adjustment function has been com-
Parameter Sub-item
selection (D)
10:Auto Zero Exe pleted, the system returns automatically to the
Search Mode 11:Magflow Zero
sub-item selection screen (D).
F0901.EPS
NOTE
CAUTION
In certain cases where the multiple range function is being used with other status
inputs, it may not be possible to perform settings for automatic zero adjustment.
For more details, refer to the description of multiple ranges from Chapter 6: Pa-
rameter Description.
In order to carry out zero adjustment via external status input, it will be necessary
to set “Ext Auto Zero (I)” using F21: DIO Function. The following describes the
setting procedure.
Start:
Major Item
Parameter Once in Setting Mode, use the switch to
Search Mode
move the cursor to F: Status Function.
Setting Mode Setting Mode
E:Pulse Set
Major item
parameter
F:Status Function
G:Alarm Touch the switch to access Sub-item
Parameter Search Mode.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 9-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
9. ACTUAL OPERATION
X12
F21:DIO Function
No Function Touch the switch to select “Ext Auto Zero
Ext Auto Zero (I)
Ext Ttl Preset (I) (Zero adjustment via external status input)”.
NOTE
Zero adjustment will be started if the DIO terminals are shorted (when the active
mode has been set to “Closed/Short Act” using F23: DIO Active Mode).
This process will end after approximately 30 s.
Now Auto Zero Auto zero adjustment function is being executed.
Executing...
(about 30 s.)
F0902-3.EPS
NOTE
When the DIO terminals continue to be shorted, this zero adjustment is automati-
cally repeated.
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 9-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
10. MAINTENANCE
10. MAINTENANCE
10.1 Maintenance 10.1.2 Changing the Direction of the-
Display Unit
WARNING 10.1.2.1 Removing the Cover
• Maintenance work must be carried out by the (1) Turn off the power.
trained personnel having knowledge of safety (2) Loosen cover locking screw 1 clockwise us-
standard and not by operators. ing a hexagonal wrench (nominal size 3) to
• When opening the cover, wait for more than unlock the cover. (Upon shipment from the
10 minutes after turning off the power. Further- manufacturing plant, the cover is locked.) Hold
more, opening of the cover must also be carried the flowmeter with your hand and remove the
out by the trained personnel having knowledge cover by turning it in the direction of the arrow
of safety standard. as shown below.
CAUTION
• The display cover is locked by the special
screw. In case of opening the display cover use
the hexagonal wrench attached.
• Be sure to lock the cover by the special srew
using the hexagon wrench attached after install-
ing the cover. Cover locking
screws
IMPORTANT F1002.EPS
• As a rule, maintenance of this flowmeter should Figure 10.1.1 Removing the Display Cover
be implemented in a maintenance service shop
where the necessary tools are provided.
• The amplifier assembly contains sensitive parts
that may be damaged by static electricity.
Excercise care so as not to directly touch the
electronic parts or circuit patterns on the board,
for example, by preventing static electrification
by using grounded wrist straps when handing
the assembly. Also take precautions such as
placing a removed amplifier assembly into a
bag with an antistatic coating.
CAUTION
Please contact Yokogawa's service office for fuse
replacement. Also be sure to use the spare fuse
that was supplied with the product, or ones sup-
plied by Yokogawa's sales or service offices.
All Rights Reserved. Copyright © 2003, Rota Yokogawa 10-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
10.1.2.2 Changing the Display Unit Direction 10.1.2.3 Installing the Cover
90 Degrees
(1) Taking care not to entangle the cables, install
(1) Hold the display unit with your hand and re- the cover to the flowmeter by turning it in the
move th two mounting screws. direction of the arrow as shown below.
(2) Turn de display unit 90 degrees clockwise and (2) Tighten cover locking screw 1 counterclock-
confirm the assembling position, taking care of wise using a hexagona wrench (nominal
the connector and wire of the display unit. At size 3) to lock the cover.
this time, do not remove the connector.
(3) Secure the display unit using its two mounting
screws.
*1
Cover locking
screws
F1005.EPS
Clockwise 90
10.2 Setting of Switches
F1004.EP
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 10-2 All Rights Reserved. Copyright © 2003, Rota Yokogawa
10. MAINTENANCE
Table 10.2.1 Output Setting Pins for Burnout 10.2.2 Setting of Write Protect Switch
Position of Burnout Burnout
Remarks
Pin Direction Output By setting the write protect function to “Protect” it
Low High Set to High before is possible to prevent the overwriting of param-
High 25 mA eters. Write protection can be carried out using
shipment
Low High either the hardware switch on the CPU board (i.e.,
Set to Low for optional
Low 0 mA code /C1 Switch 2) or software parameter settings. If either
T1001.EPS
of these items is set to “Protect,” the overwriting of
parameters will be prohibited.
NOTE
NOTE
On the amplifier’s CPU board, the burnout set-
ting switch (i.e., Switch 1) and the write protect • If the hardware switch is set to “Protect”
switch (i.e., Switch 2) are located adjacent to it will not be possible to overwrite param-
each other. Accordingly, special care should be eters; furthermore, this condition will be
taken when making switch settings. In maintained until the switch is set to “Enable”.
case of FOUNDATION Fieldbus protocol, burnout • In case of FOUNDATION Fieldbus protocol,
setting switch is not supplied. write protect setting switch is always set as "En-
Low High able".
Switch 1 2 Burnout setting switch For more details regarding usage of the write pro-
Switch 2 1 Write protect setting switch
Enable Protect tect function and the software’s parameter switch-
es, refer to “Chapter 6: Parameter Description” in
this user’s manual.
F1007.EPS
All Rights Reserved. Copyright © 2003, Rota Yokogawa 10-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
10.3 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when
the instrument is not operated correctly. This section describes troubleshooting pro-
cedures where the cause of the breakdown is identified through receiver indication.
10.3.1 No Indication
START
NO
Is anything else NO
being displayed?
YES
Does the
NO
arrow on the tube match the
flow direction?
Is the flowtube NO
full of fluid?
YES
Change the piping or the
Check the resistance between mounting condition so that
the flowtube’s A, B, and C the tube fills with the fluid.
terminals using a multimeter.
Is the NO
resistance 100 MV or more
in each case?
YES
Is foreign NO
matter adhering to
the walls?
YES
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 10-4 All Rights Reserved. Copyright © 2003, Rota Yokogawa
10. MAINTENANCE
START
YES
Are valves leaking? Replace the valve(s).
NO
All Rights Reserved. Copyright © 2003, Rota Yokogawa 10-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
START
Are parameters NO
Set the parameters correctly.
set correctly?
YES
Is the NO
Completely ground the converter and
flowmeter completely flowtube (Refer to the section on
grounded? grounding in the relevant user’s manual).
YES
Does the NO
conductivity stay within the Use within the specified conductivity range.
specified range?
Signal terminals A-C and B-C: NO Dry the tube fully to restore insulation
100 MV or more? resistance.
Excitation terminals EX1-C:
1 MV or more?
YES
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 10-6 All Rights Reserved. Copyright © 2003, Rota Yokogawa
11. Outline
11. Outline
STANDARD SPECIFICATIONS Lightning Protector:“
The lightning protector is built into the excitation current
Converter
output, the current output, the signal common, pulse/
The contents of (*1) and (*2) described in the converter alarm/status input and output terminals as standard.
specifications are follows.
Protection
(*1): Select two points from: one pulse output, one alarm General-purpose Use / Sanitary type:
output, one status input, or two status outputs. IP66, IP67
(*2:) For models without an indicator, the hand-held Coating/Paint:
terminal is necessary to set parameters. Case and Cover: Corrosion-resistant coating
Coating Color: Mint green (Munsell 5.6 BG 3.3/2.9 or its
Excitation Method (Combined with RXF Remote equivalent)
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-1 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
11. Outline
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 11-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
11. Outline
High 450
grade
- 1000
Accuracy
1 Safety Requirement Standards:
0.6
EN 61010-1
0.5 EN 61010-2-030
• Altitude at installation site: Max. 2000 m above sea level
0
0 2.5 5 7.5 10
• Installation category based on IEC 1010:
V[m/s] F03.EPS
Overvoltage category II (“II” applies to electrical equip-
ment which is supplied from a fixed installation-like
distribution board.)
Current Output Accuracy: • Pollution degree based on IEC 1010
Pulse output accuracy plus ±0.05 % of Span Pollution degree 2 (“Pollution degree” describes the
Repeatability: degree to which a solid, liquid or gas which deteriorates
±0.1 % of Rate (V ≥ 1 m/s (3.3 ft/s)) dielectric strength or surface resistivity is adhering. “2”
±0.05 % of Rate ± 0.5 mm/s (V < 1 m/s (3.3 ft/s)) applies to a normal indoor atmosphere.)
Maximum Power Consumption: EMC Conformity Standards:
Combined with RXF Remote Flowtube: 12 W EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
Insulation Resistance:
EN61000-3-2 Class A
Between power supply terminals and ground
EN61000-3-3
terminal: 100 MΩ at 500 V DC
Between power supply terminals and input/output/
excitation current terminals: 100 MΩ at 500 V DC
Between ground terminal and input/output/excitation
current terminals: 20 MΩ at 100 V DC
Between input/output/excitation current terminals:
20 MΩ at 100 V DC
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-3 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
11. Outline
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 11-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
11. Outline
Stainless Steel Stainless steel tag plate (wired onto converter). Choose this option when an SS tag plate is required
Tag Plate in addition to the standard nameplate with the tag number inscribed on it. /SCT
Dimensions: (Height x Width) : Appr.30 (76.2) x 70 (177.8) mm (in)
Active Pulse Active pulses are output in order to drive an external electromagnetic or electronic counter
Output directly using the converter’s internal power supply. (Nullifies the standard transistor contact
pulse output.)
Output voltage: 24 V DC ±20 % /EM
Pulse specifications:
• At the drive current of 30 to 150 mA
Pulse rate: 0.0001 to 2 pps (pulse/s); pulse width: 20, 33, 50, or 100 ms
Epoxy Resin Coating Epoxy resin coating which has alkali-resistance instead of standard polyurethane resin coating. /X1
The color is same as standard type.
T05.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-5 IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
11. Outline
EXTERNAL DIMENSIONS
Remote Converter AXFA14G/C
Ground Terminal (M4) Unit: mm (approx. inch)
197(7.76)*1
28 103(4.06) 66(2.6)
(1.1) 31(1.22) 180(7.09)
ш128(5.04)
57(2.24) 49(1.92)
243.5(9.59)
139.5(5.49)
ш87(3.43)
55(2.17)
A
B
E
A
B
SA
SB
C
F
90(3.54)
90(3.54)
C
(5.91)
150 +0.20
150 +0.20
(5.91)
CONVERTER
FLOW TUBE
ш10.5(0.41)
L(SPECIFIED LENGTH)
20(0.79)
Model D E F G
RXFC AXFC-4 70(2.76) 25(0.98) 50(1.97) 60(1.36)
F07.EPS
AXFC-0-L
ш10.5(0.41)
CAP
RXFC
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 11-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
All Rights Reserved. Copyright © 2006, Rota Yokogawa IM 01R21C02-01E-E 2nd edition October 01, 2013 -00
YOKOGAWA ELECTRIC CORPORATION YOKOGAWA ELECTRIC ASIA Pte. LTD. Yokogawa has an extensive sales and
World Headquarters 5 Bedok South Road
9-32, Nakacho 2-chome, Musashino-shi Singapore 469270 distribution network.
Tokyo 180-8750 Singapore
Japan www.yokogawa.com/sg Please refer to the European website
www.yokogawa.com
(www.yokogawa.com/eu) to contact your
YOKOGAWA CORPORATION OF AMERICA YOKOGAWA CHINA CO. LTD. nearest representative.
2 Dart Road 3F Tower D Cartelo Crocodile Building
Newnan GA 30265 No.568 West Tianshan Road Changing District
USA Shanghai, China
www.yokogawa.com/us www.yokogawa.com/cn
IM 01R21C02-01E-E 2nd edition October 01, 2013 -00 All Rights Reserved. Copyright © 2006, Rota Yokogawa
Subject to change without notice