Model Centiva/5 Centiva/5 Plus Engström Evita-XL Evita-4 Evita-2 Dura Savina Servo-I
User
Display size 6.5" 12" 12" 15" 12" 6" 6" monochrome 12"
Physical Quick keys, Touch Screen Quick Keys Touch Screen Touch Screen LED, Keys and LED, Keys and Touch Screen
Characteristic CommWheel CommWheel CommWheel Adjustment Wheel, quick Wheel Wheel Direct adjustment
s Tilts Tilts Tilts and rotates Wheel keys Wheel and quick
Quick keys Tilts keys
Detachable Yes Yes No Yes Yes No No Yes
Pendant Yes Yes Yes Yes Yes Yes Yes Yes
Modes of
CMV Yes, Assist Yes, Assist Yes, Assist Yes Yes Yes Yes Yes
On/Off, PLV On/Off, PLV On/Off, PLV
PCV (Bilevel) (Bilevel) Yes Yes Yes Yes Yes
PCV-VG No (BiLevel-VG) Yes Yes (Autoflow) Yes (Autoflow) Yes (Autoflow) Yes (Autoflow) Yes
SIMV-VC Yes, PLV Yes, PLV Yes, PLV Yes Yes Yes Yes Yes
SIMV-PC Yes = BiLevel Yes = BiLevel Yes Yes Yes Yes No Yes
BiLevel Yes Yes Yes Yes Yes Yes Optional Yes (BiVent)
BiLevel-VG No Yes No Yes (Autoflow) Yes (Autoflow) Yes (Autoflow) Yes (Autoflow) yes
APRV Yes (through Yes (through Yes (through Yes Yes Yes No No
CPAP/PSV Yes Yes Yes Yes Yes Yes Yes Yes
Apnea backup Yes in Yes in Yes selectable Yes Yes Yes ? Yes
NIV Yes Yes Yes Yes Yes Yes Yes ?
Other MMV, PPS, ILV MMV, PPS, ILV MMV, PPS, ILV VS, Automode
Controls
Manufacturer GE Healtchcare GE Healtchcare GE Healthcare Dräger Medical Dräger Medical Dräger Medical Dräger Medical Maquet
Model Centiva/5 Centiva/5 Plus Engström Evita-XL Evita-4 Evita-2 Dura Savina Servo-I
Monitored
Tidal Volume Yes Yes Yes Yes Yes Yes Yes Yes
Minute Yes Yes Yes Yes Yes Yes Yes Yes
Breath Rate Yes Yes Yes Yes Yes Yes Yes Yes
Ppeak Yes Yes Yes Yes Yes Yes Yes Yes
PPlateau Yes Yes Yes Yes Yes Yes Yes Yes
Pmean Yes Yes Yes Yes Yes Yes Yes Yes
Manufacturer GE Healtchcare GE Healtchcare GE Healthcare Dräger Medical Dräger Medical Dräger Medical Dräger Medical Maquet
Model Centiva/5 Centiva/5 Plus Engström Evita-XL Evita-4 Evita-2 Dura Savina Servo-I
Alarms
Minute High and Low High and Low High and Low High and Low High and Low High and Low High and Low High and Low
Tidal Volume No No High and Low High and Low High and Low High and Low High and Low No
Peak High and Low High and Low High and Low High and Low High and Low High and Low High and Low High and Low
FiO2 High and Low High and Low High and Low High and Low High and Low High and Low High and Low High and Low
Breath Rate High and Low High and Low High and Low High High High High Yes
Apnea Yes, 10 to 60 sec Yes, 10 to 60 sec Yes, 5 to 60 sec Yes Yes Yes Yes Yes
Leak Yes Yes Yes ? ? ? ? ?
Other etCO2, Vti, etCO2, Vti, etCO2, Vti,
Manufacturer GE Healtchcare GE Healtchcare GE Healthcare Dräger Medical Dräger Medical Dräger Medical Dräger Medical Maquet
Model Centiva/5 Centiva/5 Plus Engström Evita-XL Evita-4 Evita-2 Dura Savina Servo-I
Features
ARC Yes Yes Yes Yes Yes Yes No No
ASR Yes Yes Yes Yes Yes Yes No No
APD Yes Yes Yes No No No No No
Nebulizer Yes, Pneumatic Yes, Pneumatic Yes, Integrated Yes Yes Yes Yes
fully fully control on
compensated compensated ventilator,
electronic
micropump,
Procedures Insp and Exp d
Insp andt Exp
dd
Pause, Manual Pause
Breath Manual breath
SBT procedure
P0.1
t
Miscellaneou
Flow Pneumotachogra Pneumotachogra Hot wire Hot wire Hot Wire Hot wire Hot wire Ultrasonic
Internal Yes Yes Yes (typical 2 No No Yes Yes
External Yes Yes No Yes Yes Yes ?
Comments
Please Accessories for Accessories for Turbine driven Accessories for
consult Intra-hospital Intra-hospital Intra-hospital
specification transport transport transport
sheets as
several of the
biliti
Manufacturer Maquet Puritan Bennett Puritan Bennett Puritan Bennett Hamilton Hamilton Viasys Viasys
Model Servo-S 840 760 740 Galileo Gold Rafael Silver Avea Vela
User
Display size 12" 2x 10.5" No graphics No graphics 12" 6.5" 12" 10"
Physical Touch Touch screen LED, Keys LED, Keys 2 adjustment Adjustment wheel Touch screen Touch
Characteristic Screen Wheel and quick wheels Wheel and quick screen
s Direct keys keys Wheel and
adjustment quick keys
Detachable Yes No No graphics No graphics No No No No
Pendant Yes No No No No No No Yes
Modes of
CMV Yes Yes Yes Yes Yes Yes Yes Yes
Controls
Manufacturer Maquet Puritan Bennett Puritan Bennett Puritan Bennett Hamilton Hamilton Viasys Viasys
Model Servo-S 840 760 740 Galileo Gold Rafael Silver Avea Vela
Tidal Volume 100 to 2000 25 to 2500 ml 40 to 2000 ml 40 to 2000 ml 10 to 2000 ml 50 to 1800 ml 2 to 2500 ml 50 to
ml 2000ml
Breath Rate 1 to 100 bpm 1 to 100 bpm 3 to 70 bpm 3 to 70 bpm 1 to 120 bpm 1 to 80 bpm 1 to 150 bpm 2 to 80 bpm
I:E Ratio 4:1 to 1:10 4:1 to 1:299 4:1 to 1:99 4:1 to 1:99 4:1 to 1:9 4:1 to 1:9
Set NA 3 to 150 LPM 3 to 150 LPM 3 to 150 LPM 1 to 180 LPM 0 to 120 LPM 0.4 to 150 LPM 10 to 140
P-Limit NA N/A N/A N/A N/A N/A N/A N/A
P-Insp 0 to 120 5 to 90 cmH2O 5 to 80 cmH2O N/A 5 to 100 5 to 50 cmH2O 0 to 90 cmH2O 1 to 100
PS 0 to 120 0 to 70 cmH2O 0 to 70 cmH2O 0 to 70 cmH2O 0 to 100 5 to 50 cmH2O 0 to 90 cmH2O 1 to 60
PEEP 0 to 50 0 to 45 cmH2O 0 to 35 cmH2O 0 to 35 cmH2O 0 to 50 cmH2O 0 to 20 cmH2O 0 to 50 cmH2O 0 to 30
P-Ramp 0 - 0.4 s 1 to 100% (?) 5 to 100 (factor) No 25 to 200 ms N/A 1 to 9
PS-Ramp N/A N/A N/A N/A N/A N/A
FiO2 21 to 100% 21% to 100% 21% to 100% 21 to 100% 21 to 100% 21% to 100% 21% to
Flow Trigger 0.5 to 20 LPM 1 to 20 LPM 1 to 20 LPM 0.5 to 15 LPM 2 to 10 LPM 0.1 to 20 LPM 1 to 8 LPM
Pressure 0.1 to 20 LPM
ByFlow Yes (?) Yes (?) Yes (?) 4 to 30 LPM Fixed (?) 0.4 to 5 LPM 10 to 20
Trigger ? ? ? No No
PS Endflow 1 to 45% Yes No 10 to 40% No 5 to 45% 5 to 30%
Insp Plateau 0 to 2 sec 0 to 2 sec 0 to 2 sec 0 to 70% No 0 to 2 sec 0 to 2 sec
Other
Monitored
Tidal Volume Yes Yes Yes Yes Yes Yes Yes Yes
Minute Yes Yes Yes Yes Yes Yes Yes Yes
Breath Rate Yes Yes Yes Yes Yes Yes Yes Yes
Ppeak Yes Yes Yes Yes Yes Yes Yes Yes
PPlateau Yes Yes Yes Yes Yes No
Pmean Yes Yes Yes Yes Yes Yes
Manufacturer Maquet Puritan Bennett Puritan Bennett Puritan Bennett Hamilton Hamilton Viasys Viasys
Model Servo-S 840 760 740 Galileo Gold Rafael Silver Avea Vela
Alarms
Minute High and High and Low High and Low High and Low Low and High Low and high
Tidal Volume No
Peak High and High and Low High and Low High and Low Low and High High and Low
FiO2 High and ? High and Low High and Low Low and High High and Low
Breath Rate Yes High High High Low and High High
Apnea Yes Yes Yes (fixed) Yes (fixed) Yes Yes (fixed)
Leak ? ? No (?) No (?) ? No
Other
Manufacturer Maquet Puritan Bennett Puritan Bennett Puritan Bennett Hamilton Hamilton Viasys Viasys
Model Servo-S 840 760 740 Galileo Gold Rafael Silver Avea Vela
Features
ARC No No No Yes No
ASR only 100% only 100% only 100% 100 % only No
APD ? No No No No
Nebulizer External or Yes No (?) No (?) Yes Yes
Ultrasonic
Procedures
Miscellaneou
Flow Ultrasonic Hot wire ? ? Variable orifice Variable orifice Variable oriffice Variable
Internal Yes Yes Yes Yes Yes Yes
External Yes ? Yes Yes No
Comments
Please Piston driven Piston driven Optional bulit in Turbine
consult compressor (scroll driven
specification pump)
sheets as Heliox delivery
several of the capability
biliti