Professional Documents
Culture Documents
CA-1500 SM #2 Hydraulic&Mech
CA-1500 SM #2 Hydraulic&Mech
2
Reagent syringe
Reagent pipette
MV4 (with heating 2
M
function)
Rinse cup
MV5
MV6
MV1
M
Rinse cup
MV2
MV3
Cooler
r c unit No. 8
MV11
Vacuum (Source
Distilled water pressure)
MV8
Vacuum (Source pressure) 1.0 kg/cm2
2
1.0 kg/cm MV7
MV10
Distilled water
(Rinse) Waste
(1) The instrument does not become READY until the temperature comes to be in the
specified range. (It becomes READY in the Service Mode.)
(2) Check Sample Plate, remained reagent volume (only when set), and reaction tube trash
box.
(3) Monitor the float SW of external detergent container and waste container. (only when set)
(4) Check Light Shield Cover and STAT Cover switches.
NOTE: If it cannot be reset, “Lamp Unit Move Error” occurs, and the operation
immediately stops.
• Error Detection Condition
1) Although the motor is activated, Optical Filter Unit cannot move
from the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Optical Filter Unit cannot move physically.
3) Home position sensor malfunctions.
4) Failed motor drive circuit
NOTE: If they cannot be moved, “Sample Arm Move Error (Z)” or “Reagent Arm
Move Error (Z)” occurs, and the operation immediately stops.
• Error Detection Condition
1) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Failed home position sensor
3) Failed motor drive circuit
(b) Catcher Unit Z ascends to the move position (which is 28 mm above the current
position).
(c) Reset Catcher Unit XY to the home position.
NOTE: If it cannot be reset, “Catcher Move Error (X) or (Y)” occurs, and the
operation immediately stops.
• Error Detection Condition
1) Although the motor is activated, Catcher Unit XY cannot move from
the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Catcher Unit XY cannot move physically.
3) Failed home position sensor
4) Failed motor drive circuit
NOTE: If it cannot be reset, “Catcher Move Error (Z)” occurs, and the operation
immediately stops.
• Error Detection Condition
1) Although the motor is activated, Catcher Unit Z cannot move from
the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Catcher Unit Z cannot move physically.
3) Failed home position sensor
4) Failed motor drive circuit
(e) Catcher Unit moves to Trash Unit and discards reaction tubes. Usually, there are
no tubes on Catcher Unit, however, when the operation stops due to an error,
there may be any reaction tubes on Catcher Unit.
NOTE: If Catcher Unit cannot correctly discard the reaction tubes, “Catcher
Move Error (Trash Unit)” occurs and the operation immediately stops.
• Error Detection Condition
1) Catcher Unit Z cannot move from the home position sensor 0.7
seconds after starting descending.
• Possible Cause
1) Catcher Unit Z cannot descend because Trash Unit is full of
reaction tubes.
2) Catcher Unit Z cannot descend due to the incorrect XY position.
(f) Catcher Unit XY moves to the home position and Catcher Unit Z ascends to the
move position.
NOTE: If it cannot be reset, “Sample Arm Move Error (Z)” occurs, and the
operation immediately stops.
• Error Detection Condition
1) Although the motor is activated, Sample Arm Z cannot move from
the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Failed pipette crash sensor (The sensor does not turn OFF.)
3) Failed motor drive circuit
4) Sample Arm Z cannot move physically.
NOTE: If it cannot be reset, “Sample Arm Move Error (X) or (Y)” occurs, and the
operation immediately stops.
• Error Detection Condition
1 Although the motor is activated, Sample Arm XY cannot move from
the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Failed home position sensor
3) Failed motor drive circuit
4) Sample Arm XY cannot move physically.
NOTE: If it cannot be reset, “Reagent Arm Move Error (Z)” occurs, and the
operation immediately stops.
• Error Detection Condition
1) Although the motor is activated, Reagent Arm Z cannot move from
the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Failed pipette crash sensor (The sensor does not turn OFF.)
3) Failed motor drive circuit
4) Reagent Arm Z cannot move physically.
NOTE: If it cannot be reset, “Reagent Arm Move Error (X) or (Y)” occurs, and
the operation immediately stops.
• Error Detection Condition
1) Although the motor is activated, Reagent Arm XY cannot move
from the home position sensor.
2) The motor does not stop at the home position sensor.
• Possible Cause
1) Failed motor
2) Failed home position sensor
3) Failed motor drive circuit
4) Reagent Arm XY cannot move physically.
NOTE: If the supplying operation is not correctly completed, the message “No
Detergent” is displayed and the analysis operation does not start.
• Error Detection Condition
1) After 30 seconds from starting supplying water (Solenoid Valve No. 9
turns ON), Detergent Chamber cannot be filled with detergent and
the float switch of Detergent Chamber is OFF (open).
• Possible Cause
1) Pneumatic Unit malfunctions or incorrect pressure
2) Clogging in the supplying line
3) Failed float switch
NOTE: If the draining operation is not correctly completed, the message “Drain
Waste Error” is displayed and the analysis operation does not start.
• Error Detection Condition
1) After 2 seconds from starting draining water (Solenoid Valve No. 7
turns ON), Waste Chamber cannot be drained and the float switch of
Waste Chamber is OFF (open).
• Possible Cause
1) Pneumatic Unit malfunctions or incorrect pressure
2) Clogging in the draining line
3) Failed float switch
1) Catcher XY moves to Detector Unit, and Catcher Z descends and catches the
reaction tube.
2) Catcher Z ascends to the move position.
3) Catcher XY moves to Trash Unit and Catcher Z descends. If Catcher Z cannot
correctly descend, “Catcher Move Error (Trash Unit)” occurs and the operation
immediately stops. (This error occurs when Trash Unit is full of reaction tubes or the
Catcher cannot transfer the reaction tube to the correct position.)
Required detergent volume for rinsing inner Sample Pipette: approx. 1700 µL
Required detergent volume for rinsing outer Sample Pipette: approx. 2200 µL
In the FDP, D-Dimmer Analysis Modes, the rinsing sequence using CA CLEAN I is
performed after aspirating sample. This sequence is operated for 30 seconds. After
aspirating CA CLEAN I, the above Sample Pipette rinsing is performed twice. The
required aspiration volume of CA CLEAN I is: Sample Aspiration Volume + Diluent
Volume + a
This sequence can be set to ON or OFF in the Analysis Procedure Settings, also the
detergent can be changed.
The required detergent aspiration volume on the default settings is described in the
following table.
(µL)
CA CLEAN I CA CLEAN II
AT III R1 100 –
AT III R2 100 –
a 2PI R1 100 –
a 2PI R2 – –
Plg R1 100 –
Plg R2 – –
PC R1 100 –
PC R2 – –
Required detergent volume for rinsing inner Reagent Pipette: approx. 1200 µL
Required detergent volume for rinsing outer Reagent Pipette: approx. 2200 µL
Required detergent volume for rinsing inner Reagent Pipette: approx. 300 µL
Required detergent volume for rinsing outer Reagent Pipette: approx. 900 µL
Required detergent volume for rinsing inner Reagent Pipette: approx. 1200 µL
Required detergent volume for rinsing outer Reagent Pipette: approx. 2200 µL
2.3.1 Outline
The CA-1500 detecting unit provides four detecting blocks, each of which provides four
detecting wells and seven incubator wells.
The temperature of each detecting unit is maintained at 37°C, with the heater and
thermistor.
For coagulation detection, light emitted from the red LED is directed to a sample. The light
is scattered by 90° and is received by the photo-diode, then A/D conversion is performed.
For transmitted-light detection, light emitted from the halogen lamp is directed to a sample
in the detecting port. The transmitted light is received by the photo-diode, then A/D
conversion is performed.
2.3.2 Specifications/Functions
(1) Coagulation Detection Unit
Photometer : Four wells
Incubator : Seven wells
Light source : 660 nm LED
Heater : 30 W (Film heater)
Protecting function : 76°C thermal fuse
(2) Transmitted-light Detection Unit
Photometer : Four wells
Light source : Halogen lamp
Heater : 100 V, 30 W (Film heater)
Protecting function : 76°C thermal fuse
2.4.1 Outline
(1) This unit cools coagulation reagents, chromogenic reagents, factor deficit plasma, and
control plasma to prevent these materials from being deactivated.
(2) Precipitable coagulative reagents are stirred.
(3) The pressure air supplied from the pneumatic unit is cooled to saturate the air with
moisture. (Drain function)
NOTE: The moisture of the saturated air is removed by the No. 2 drain filer assembly.
2.4.2 Specifications/Functions
(1) Cooling function
Cooling element: Three Peltier elements
Target temperature : 15 ± 2°C (when room temperature is 15°C to 30°C)
Time required to reach target temperature
: Within 30 minutes (In the range of the specified operating
temperatures)
Radiator : One set of radiator fin and DC fan
(2) Stirring function
Drive : AC synchronous motor
Stirring capacity : The stirrer rotates at approx. 50 rpm.
Stirring position : Four positions at the back of the 31 mm diameter reagent
bottle holder
2.5.1 Outline
The Sampler Plate Holdter holds sample plates for primary sampling sample. The of the
sample plate is detected and indicated by the 2-color LEDs. These LEDs also indicate if the
sample plate can be dismounted or not.
2.5.2 Specifications/Functions
Number of mountable plates: Five plates max.
Plate detection: Photo-interrupter TLP1204 (C3)
Indicator: 2-color LEDs
[Reference]
TLP1204(C1) 228-4611-2
310497102
LED : Light emitter No. 18
973-4981-9
* Five sets of the No. 14 light emitters are linked with each other.
2.6.1 Outline
The No. 2 rotary feeder is a part of the automatic sample tube feeding function. After
sample tubes are aligned in the feeder, they are picked up one by one on the rotary feeder,
and fed into the catcher. At the feed position, the micro-switch detects the presence or
absence of a sample tube. At the same position, the sample is dispensed.
2.6.2 Specifications/Functions
Feeding method : Fed by the swiveling rotor
Drive : DC motor with gear head (5 V)
Rotation speed : 9 rpm
Feeding capacity : Approx. 2.1 seconds/tube
2.7.1 Outline
The sampler feeder automatically feeds sample tubes with the rotary feeder. The sample
tubes irregularly placed in the hopper are aligned in the rail of the feeder.
The reflective sensor at the top of the rail detects the presence of a sample tube so that the
sample tube feeder can be activated only when required.
2.7.2 Specifications/Functions
Aligning method : Sample tubes are picked up through vertical motion of the rail
between two arc plates.
Number of sample tubes that can be loaded
: 300 tubes max.
Feeding capacity : Approx. 5 seconds/tube (average)
Drive : DC motor with gear head (12 VDC)
Rotation speed : 120 rpm
Reflective Sensor
2.8.1 Specifications/Functions
Box detector : One reflective photo-interrupter
Capacity of discharge box : Approx. 280 tubes
Box fixing method : Leaf spring
2.9.2 Specifications/Functions
Location of each sensor
S NS 1 0 S NS 5 S NS 4 S NS 1 S NS 2
S NS 8
S NS 9
S NS 1 1
S NS 6
S NS 3
Micro -switch se n so r
Th ro u g h b e am se n so r
Re fle ctive se n so r
P h o to -in te rru p te r se n so r