Foam Air Permeability Tester
Operation Manual
Model: HD‑W810
1
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Preface
Thanks for you choose our company products, we not only provide high quality, but also reliable
after‑sales service.
In order to ensure the safe of the user and the integrity of the machine, please read the operate
instruction carefully before use the instrument, pay more attention to the using notes. This
operating manual will introduce the design principles, design criteria, structure, operating norms,
calibration, maintenance, some failure case and solutions, electrical diagram so on. If this
operating manual in reference to various “test requirement”, “standards”, all of them just for your
reference. Please review relevant standards or data by yourself if you feel there is disagreement.
Special statement
‧ The operating manual can t be basis as serve as any request to our company.
‧ The interpretation of this manual shall be in our company.
Haida International Equipment Co., Ltd Phone: +86
0769 89055588
Website www.haidatestequipment.com
1
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Directory
I.Attention of security‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑3
II.Instruction‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑4‑6
III. Instrument description‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑7‑8
IV. Operation specifications‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑9‑13
V. Common faults and solutions‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑14
VI. Maintance‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑14
VII. Quality assurance‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑15
VIII. Notes‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑16
VIIII.Installation manual‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑17‑21
2
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Safety Attention
1. Safety labels
These following labels are very important, please pay attention to them carefully when you using
the tester.
Dangerous:
This shows that if don t follow the guideline, operator may get hurt.
!
Warning:
! This shows that if don t follow the guideline, equipment may get damaged.
Attention:
This shows that the item may affect testing result and accuracy.
Note:
This shows auxiliary statements during the machine operation.
2. Warning Labels:
Warning
! This sign warns where need refer to operation manual.
Dangerous voltage mark
The sign represents for dangerous high voltage.
3
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
II. Introduction
1. Usage:
This instrument is used to test the permeability of materials such as special industrial fabrics,
general fabrics, knitted fabrics, coated fabrics, nonwovens and industrial filter paper, and its
performance meets or exceeds the standard requirements of GB / T5453‑1997 Determination of
fabric permeability for test equipment.
2. Main features:
1.2.1. The instrument adopts high‑precision sensor instead of the traditional water column
pressure difference test
1.2.2. It can use the computer to calculate the test data, eliminating the need for manual
conversion table
4
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
1.2.3. The nozzle can be replaced according to each nozzle measurement range. It can
automatically fix the sample to make the operation simple, and the accuracy and efficiency.is
higher.
1.2.4. It can be connected with the computer on the machine operation, automatic rendering
of real‑time test data curve, easy to observe.
1.2.5. High‑definition color touch screen display, convenient and quick.
1.2.6. The test parameters can be adjusted directly by setting the number. The parameter
units can be set as required.
1.2.7. It is equipped with micro thermal printer.
1.2.8. The stent of the instrument is made of cold‑rolled steel sheet by directly stamping,
surface spray treatment, simple and beautiful, light and stable; and no corrosion, easy to clean and
so on.1.2.9. The flow tube is made of seamless steel pipe which is durable, stable and sealed. And it
sets the door and the door lock device not only to ensure the replacement of the nozzle is
convenient, but also to ensure a good air tightness.
1.2.10. The specimen tightening device is controlled by the cylinder. When starts test, the
specimen is clamped automatically under the pressure, and the pressing force is enough to ensure
the close fit of the specimen fixture.
3. Main specifications and technical indicators:
1.3.1. Test pressure:1~4000Pa (nozzles) 1‑1000 Pa (sample)
1.3.2. Measuring air permeability:1~9999mm/s
5
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
1.3.3. Test error:≤±2%
1.3.4. Sample thickness:≤12mm
1.3.5. Air intake adjustment:Dynamic adjustment by data feedback
1.3.6. Test areas:5cm2,10cm2,50cm2,100cm2
1.3.7. Sample diameter:Ф50mm(≈19.6cm2)
Ф70mm(≈38.5cm2)
1.3.8. Nozzle:11 Nos.
No. 00 01 02 03 04 05 06 07 08 09 10
1.3.9. Data processing capacity:≤20‑30times test
1.3.10. Printer: Micro thermal printer
1.3.11. Data port:RS232
1.3.12. Power:AC 220V±10V 50Hz
1.3.13. Power consumption:1200W
1.3.14.Weight:80Kg
1.3.15. Outside dimension:1080×670×1080mm (L×W×H)
4. Work environment and conditions:
1.4.1. Room temperature control range:10℃~35℃
1.4.2. Relative humidity:≤80%
1.4.3. There is no vibration, no corrosive media and no strong electromagnetic interference in
the surrounding environment.
1.4.4. Grounding requirements: grounding resistance is less than 5Ω。
6
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
III. Instrument Description
1. Instrument structure:
1. Power outlet 2. Switch 3. Serial interface 4. Micro printer 5. Display and operation panel 6. Sample fixture 7. Fixture handle 8.
Sample pressure ring 9. Sample table 10. Flow device door 11. Suction fan 12. Feedback adjustment device 13. Emergency stop
switch 14. Air interface
2. Main components:
The external structure of the instrument consists of a bracket, a sample tightening device, a flow device, a display panel and other
parts. The internal structure of the instrument is composed of pressure sensor, CPU data processor, suction fan, feedback
adjustment device and so on.
3. Test Principle
According to the prescribed method and test parameters, the sample is held in the air inlet of
the fabric air permeability tester, and then adjust the fan speed to achieve the required pressure
on both sides of the fabric; according to the nozzle diameter and pressure difference between the
two sides measure the air permeability of fabric which refers to the amount of air flow per unit
7
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
area (mm / s) per unit time, the flow rate of the fabric on both sides at a predetermined pressure
difference, that is air permeability.
4. Key function description:
【Power switch H】
The total power switch of the instrument. Toggling button when lights is on
which means turning on, toggling button when lights is off which means turning off,
【Emergency stop switch】
It can effectively interrupt the power supply of equipment quickly.
Press to power off, turn right to resume normal power.
8
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
IV. Operation Specification
1. Check before starting
1.1. Check if the power plug is secure.
1.2. Check if the air supply is secure.
1.3. Check all nozzles and sample area plates are complete.
1.4. Check if the pump pressure is between 0.4 ~ 0.7MPa.
2. Check after starting
2.1. Turn on the host power.
2.2. Check if the specimen holder is in the raised position, otherwise check the air supply.
2.3. Check the color touch screen display normally, otherwise check the line is loose.
2.4. Check if the specimen on the current fixture is consistent with the sample as required,
otherwise replace.
2.5. According to the needs to do sample area selection and installation.
2.6. After selecting the nozzle for installation, close tightly the feedback adjustment door.
2.5. If the above is normal, you can start the test.
3. Test operation
The display panel is a color touch screen, and the keys and the display function operation is
as bellow:
9
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
3.1. Welcome interface, display company LOGO, instrument name and model.
【Log】 【Set】 【Work】 【Cali】 【Repo】
five buttons were used to switch the various interfaces.
3.2. Setting interface
In this interface, please set the required sample area, the result unit of air permeability, the
pressure of sample, nozzle number and test times. Press the enter key to switch to the work
interface automatically.
10
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
【Sample area】:There are 6 kinds of work area optional, directly click on the corresponding area
button.
【Result unit】:There are seven units optional, directly click on the corresponding unit button.
【Sample pressure】:Click on the sample pressure input box to enter the desired setting.
【Nozzle】:Click the nozzle number input box, you can enter the required nozzle number, 0‑10.
【Times】:According to the test requirements to enter the test times, and you can enter 1‑20
times.
Click on the bottom right corne【r 】
to save the settings parameters, and automatically enter
the work interface.
3.3.Work interface
Nozzles Sample area plate
11
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Feedback device door Sample table base
First, according to the set parameters, open the feedback device door, load the suitable nozzle,
and then close the feedback device door. Followed by loading the corresponding set of test area
into the sample table base.
Key function:
【Run】:start the test
【Stop】:stop the test
【Rst】:The reset value is cleared when the pressure value is large。
【Del】:Delete the data of test results, it can be used when test results vary widely
【 】:Used to create a new test group.
3.4. Calibration interface
The instrument shall be calibrated well when leaving the factory and generally the user has no
need to operate this function.
12
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
3.5. Report interface
The keys of up and down are used to query the data of the previous group and the next group
respectively, and the keys of left and right are used to query each data corresponding to the group.
Press the delete key, for the lack of memory, delete all the data to continue to save the
experimental data. Press the Print key to print all the data and statistics corresponding to the
group of the current query.
13
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
V. Common Faults And Solutions
1. The value of nozzle pressure or pressure difference of sample do not beat or value is too large.
Check if the sensor is loose, if loose, please plug it in.
Check if the instrument is not cleared zero when turning on, if not, please click the Clear
button on the work interface.
2. There was no obvious upward trend in the pressure during the test
If the trachea connecting the sensor and the feedback device is inserted.
Check if the sensor is loose, if loose, please plug it in.
Check if the feedback device door is closed tightly.
3. Test starts, the fixture is not under pressure or under pressure is not enough.
Check if the air supply port is accessible.
Check if the air pressure is between 0.4 ~ 0.7MPa.
VI. Maintenance
1. The machine should be placed on the solid table.
2. Please do not move the machine when using it.
3. Choose the correct power.
4. When the machine has problem, please contact us.
5. There is good ventilation.
6. Lube: fill lube (common one) at the rotating part of the machine (screw) every month
regularly.
7. Please keep the machine clean after using it.
8. For controller, please clean it by dry cloth not wet cloth.
9. Please do well grounding.
14
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
VI. Quality Assurance
ⅰ、Quality assurance items
The guarantee period is one year (consumer goods and travel cost are not included)
ⅱ、the main quality warranty maintenance certificate
1.please keep this guarantee letter, if you lost it, please connect with us in a month.
2. If this guarantee letter has been altered or it has no our stamp, it is useless.
ⅲ、The following conditions need to pay reasonably even in the assurance period:
1. Natural reasons
2. Operating mistakes
3. Voltage is not fit for our operation instruction
4. Repack it without our guides
5. Damaged for borrowing to others
6. Damaged for authorized machine modification
7. Damaged for authorized calibration
ⅳ、Attentions
1. Where the original in Dongguan, Shenzhen installation of our products, if the product moved to
outside Dongguan, Shenzhen‑area use, whether or not within the effective warranty period, the
transportation and travel expenses of the service personnel should be paid by the client.
2.All customers outside of Guangdong province, whether or not in the warranty period, the
service of transportation and travel expenses paid by the customer.
3.The apparatus try not to be used in the following situations:
3.1、Vibration, rocking the occasion.。
3.2、Direct sunlight.
3.3、Hot, dusty, damp places.
3.4、To ensure safe, AC supply of the machine should be well grounded.
3.5、Do not use strong solvents (such as: benzene, nitro oil) washing machine.
3.6、 Do not inject water and debris into the machine, prevent damage to electrical
components and electrical shock.
3.7、Instrument displays the disassembly and debugging can only be measured by the State
Department approved the units and the company, other people not allowed to overhaul.
15
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Notes
Any omissions in this instruction and any add instructions owe to the updated of the
products are written here.
This user is for your operating reference, the notes in it are very useful, please keep it.
Our products are developing, this is user just fit for this kind of tester. Thank you!
COPY IS FORBIDEN UNLESS OUR CONSENT
16
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
Installation Manual
1. The installation space requirements: The installation
space requirement,which is a basic and necessary requirement for the instrument,are as follows:
2.Installation conditions:
2.1 The installation environment requirements.
The environment requirements of the instrument: The temperature
requires 10℃ ~ 35 ℃, the humidity requirement requires below 85% RH, no vibration in the surrounding environment, no
corrosive medium, no strong electromagnetic interference.
2.2 The installation power and capacity requirements.
Power requires AC220V±10% 50Hz, maximum power requires 1200W, no direct sunlight, no vibration interference,
grounding requirements grounding resistance is less than 5Ω.
17
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
3.Receiving device.
When carrying the instrument,you`d better not to recline or inverted or upside down,or will produce large vibrations,and
easily broke the glassware in the instrument.
After the equipment is unpacked, the equipment and accessories should be inspected and inspected according to the
packing list. If there is a shortage or damage, please inform the company in the shortest time so that we can deal with it
in time.
4. Host installation
4.1 Fixing host After
the host is in place, the ground support force needs 80KG, and the brake pressure of the four casters of the instrument is pressed
downwards. As shown in the figure, the fixed position of the foot is fixed to the ground and the instrument cannot be moved.
4.2 Installing sample base The
sample base is composed of a sample stage, a sample net, a silicone pad, and a sample retainer ring. The sample net, the
silica gel pad, and the sample base is installed after the sample collapse ring are sequentially placed in the sample stage.
18
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
4.3 Fixing of the fixture on the sample The clamp
on the sample consists of the sample area seat and the sample pressure ring. Four four screws are mounted on the four protruding
screws on the sample area seat, and then the sample pressure ring is placed on the sample area seat.
The four springs should be aligned with the four holes of the sample pressure ring, and the left and right screws of the sample
pressure ring should be aligned with the left and right copper grooves of the sample area seat.
19
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
4.4 Aligning of sample center position The center
of the sample holder is aligned and controlled by four positions A, E, F, and H. The A clamp adjust the front and rear positions,
the H screw fixing fixture, and the two screws E and F adjust the left and right positions of the sample holder.
4.5 Unfolding of sample table The two
table tops can be directly opened,which are supported by two gas push rods, and can be folded by light pressing.
20
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
5.Electrical installation
5.1 Power connection
The power supply of this instrument requires AC220V±10% 50Hz, and the A port is the power socket. Before plugging in, you
can see if there is any ground wire. If there is no ground wire on the power socket, you should make a regular ground wire and
put it into the power socket. The test can be started after the check is correct.
Tip: The connection of the ground wire is very important. If there is no qualified ground wire connection, it will have a greater
impact on the electrical control system of the device. Therefore, the ground wire must be connected before the device is installed.
5.2 Instrument computer and software installation Install the
computer desk next to the instrument and assemble the randomly configured computers. The operating system is
installed, and the computer requirements are in the software operation manual. Copy the software installation package to the
computer hard disk through the randomly configured program CD or USB flash drive, install the host computer system software,
21
www.haidatestequipment.com
Machine Translated by Google
Operation manual
________________________________________________________________________________
and key software operation manual. Connect the connection line between the computer and the host computer. The E port is the
communication port between the instrument and the computer.
5.3 Gas source installation H is the
air source interface, please ensure that the air pump pressure is between 0.4 and 0.7 MPa.
6.Installation problem In order
to ensure the normal operation of the instrument and the safety of the customer, if you have any installation problems, please
contact our after‑sales service personnel.