Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

AL HOTY CALIBRATION SERVICES

Verification Report
Troxler Gauge Model: 3430 Page 1/2
Serial Number: 26837
Daily Density Standard Count: 1636
Daily Moisture Standard Count: 337
Validator Serial Number: 08021260
Validation Date: 12/27/2022
Failure Cutoff: 32.0. Failure cutoff for Moisture is always ±1 pcf (±16 kg/m3)

Print Date: 12/27/2022

Depth Gauge Count ValiDator Density Calculated Density Error (kg/m3)


_________ ____________ __________________ ___________________ ______________
Moisture 193 540.0 534.8 -5.21
0 488.5 2131.2 2131.4 0.18
50 1831 2145.3 2145.3 0.02
100 2085 2006.1 2006.3 0.21
150 1615 2017.5 2017.8 0.30
200 619 2359.7 2359.4 -0.28
250 451 2214.6 2214.0 -0.61
300 374 2019.0 2017.0 -2.03

Calibration Standard Count: 1636


Calibration Date: 12/27/2022

Depth A B x 1000 C
______ _________ ________ ________
0 4.69134 1.45755 -0.09667
50 10.67050 0.99196 0.11806
100 11.54958 0.99566 0.25120
150 11.54923 1.12646 0.16717
200 26.07753 1.90576 -0.10209
250 16.18545 1.84565 -0.01682
300 11.48538 1.95361 -0.01668

Moisture Parameters:

E Fx1000
_________ _________
0.04491 0.98692
Verification Uncertainty
Troxler Gauge Model: 3430 Page 2/2
Serial Number: 26837
Validator Serial Number: 08021260
Validation Date: 12/27/2022

Errors listed on the Verification Report indicate the difference between the values read by the gauge and the actual values
assigned to the ValiDator block at each depth. These values are a measure of the accuracy of the gauge with the calibration
constants listed in the Verification Report and meet the requirements for verification of ASTM D7759, ASTM D6938 and
AASHTO T310.

Maximum expanded uncertainty using the ValiDator for verification at k=2 are presented below for all depths according to
the requirements of ASTM D6938, D2950 and D7759. The values below are in percent (%) and indicate the maximum
expanded uncertainty for the range of densities normally encountered in construction material testing, low [109.4 lb/ft3 (1752
kg/m3)], medium [135.6 lb/ft3 2172 kg/m3)], high [162.1 lb/f3 (2597 kg/m3)] and moisture (34 lb/ft3 (545 kg/m3).

Max Uncertainty Density Moisture


_______________ _______ ________
BS 1.434 2.880
50 0.940
75 0.876
100 0.704
125 0.662
150 0.623
175 0.643
200 0.671
225 0.746
250 0.752
275 0.866
300 0.984

Note: Expanded uncertainty calculated using NIST Technical Note 1297 in percent for density and moisture verification by
the ValiDator block at k=2