Jurnal Mutiara Elektromedik, 27-31

ANALISIS KALIBRASI CENTRIFUGES MENGGUNAKAN METODE


ECRI 456-0595

Zainal Abidin Siregar1, Yulizham2


1,2
Fakultas Sain dan Teknologi, Universitas Sari Mutiara Indonesia
email:zainalsiregar11@gmail.com

ABSTRACT

Health equipment is a supporting tool for medical services. Health services that are well supported by
equipment that is good, safe and in accordance with its function so that there is no error in the
patient's diagnosis. Calibration is a series of activities that form a relationship between the value
indicated by a measuring instrument or a measurement system, or a value represented by a measuring
material, with known values that are related to the magnitude measured in certain conditions. In other
words. Calibration is an activity to determine the conventional truth of the value of measuring and
measuring instruments by comparing traceable standards to national and international standards for
measurement and / or international units and certified reference materials. Calibration is carried out
by trained technicians so no calibration procedure errors occur.
This research was conducted at the Medan Health Facilities Security Center with a type of approach
quantitatively by analyzing the speed (RPM) of centrifuges using a tachometer. Repeated
measurements or more will get better results for data analysis in this study.
The results of the research conducted at the Medan Health Facilities Security Center were obtained
by u_exp data in the range (1.32-1.92RPM) and centrifuges were declared suitable for use.

Keywords: Maintenance, Speed, Testing and Calibration.

1. PENDAHULUAN yang mampu tertelusur (traceable) ke


Menurut ISO/IEC Guide 17025:2005 standar nasional maupun internasional
dan Vocabulary of International untuk satuan ukuran dan/atau
Metrology (VIM) Kalibrasi adalah internasional dan bahan-bahan acuan
serangkaian kegiatan yang membentuk tersertifikasi.( KIM LIPI., 2016,)
hubungan antara nilai yang ditunjukkan Tujuan kalibrasi peralatan kesehatan
oleh instrumen ukur atau sistem adalah untuk mencapai ketertelusuran
pengukuran, atau nilai yang diwakili pengukuran. Hasil pengukuran dapat
oleh bahan ukur, dengan nilai-nilai yang dikaitkan/ditelusur sampai ke standar
sudah diketahui yang berkaitan dari yang lebih tinggi/teliti (standar primer
besaran yang diukur dalam kondisi nasional dan / internasional), melalui
tertentu. rangkaian perbandingan yang tak
Dengan kata lain. Kalibrasi adalah terputus dan Menentukan deviasi
kegiatan untuk menentukan kebenaran (penyimpangan) kebenaran nilai
konvensional nilai penunjukkan alat konvensional penunjukan suatu
ukur dan bahan ukur dengan cara instrument ukur. Agar pengguna (user)
membandingkan terhadap standar ukur tidak salah mendiagnosa pasien melalui

27
Jurnal Mutiara Elektromedik 201 8,Vol. 2 No. 1
Jurnal Mutiara Elektromedik, 27-31

cairan sample yang telah di ambil dari 2. Untuk mengetahui cara menentukan
tubuh pasien. layak atau tidak layak pakai
Adapun manfaat kalibrasi untuk menjaga
kondisi instrumen ukur dan bahan ukur
2. METODE PENELITIAN
agar tetap sesuai dengan spesefikasinya.
Untuk mendukung sistem mutu yang Pendekatan yang digunakan dalam
diterapkan di berbagai rumah sakit pada penelitian ini adalah kuantitatif dan
peralatan laboratorium dan Bisa Deskriftif yaitu analisis alat centrifuges
mengetahui perbedaan (penyimpangan) menggunakan tachometer yang ada di
antara harga benar dengan harga yang Laboratorium Kelistrikan Balai
ditunjukkan oleh alat ukur.( KIM LIPI., Pengamanan Fasilitas Kesehatan Medan
2016,) (BPFK). Metode yang digunakan adalah
Peralatan kesehatan merupakan sarana Data primer di peroleh dari hasil
penunjang untuk penyelenggaraan observasi yang di lakukan terhadap
pelayanan kesehatan baik secara centrifuge yang ada Alat dan Peralatan
langsung maupun tidak langsung. Penelitian
Peralatan kesehatan semakin banyak
jenis dan pengelompokannya sesuai Adapun alat yang digunakan:
dengan kebutuhan dan fungsi alat 1. Centrifuges General
tersebut. Secara khusus peralatan Merk : Hanshin
kesehatan di bidang laboratorium seperti Tipe : HC-16D
alat pemisah cairan sangat perlu di S/N : ZH01018
bidang laboratorium alat pemisah darah Kecepatan Putar : 4000 RPM
ini lebih di kenal dengan nama 2. Photo Tachometer
Centrifuge. Dimana alat ini dapat Merk : Lutron
menghasilkan kecepatan putaran 0 - Tipe : DT-2236
3000 RPM, 0 - 20000 RPM, dan 0 – S/N 18004071
50000 RPM. Maka untuk mengetahui Resolusi penngukuran : 0,5-20.000
keakuratan data hasil pemisahaan cairan Rpm
maka sangat perlu di lakukan pengujian
dan kalibrasi yang di lakukan teknisi
kesehatan di bidang kalibrasi.
Prosodure Kalibrasi Centrifuges Sesuai
Standar Operasional Kerja Laboratorium
. BPFK :
1. Siapkan Lembar kerja
2. Siapkan peralatan kalibrasi yang
akan digunakan
3. Lakukan pendataan administrasi
meliputi merk, type, nomor seri,
Tujuan dilakukannya penelitian ini tanggal pengujian, lokasi
adalah: pengujian, dan ruangan alat
1. Untuk mengetahui teknik kalibrasi 4. Lakukan pengukuran suhu dan
centrifuge. kelembaban, ruangan sebelum
dan sesudah melaksanakan
kegiatan kalibrasi
28
Jurnal Mutiara Elektromedik 201 8,Vol. 2 No. 1
Jurnal Mutiara Elektromedik, 27-31

5. Pengamatan n : banyaknya data pengukuran


A. Pengamatan fisik dan Fungsi n1 : hasil data pengukuran
alat
i. Badan dan permukaan :
Periksa bagian luar dan
kondisi kebersihan fisik 4. Hitung nilai ketidakpastian daya
secara menyeluruh. Pastikan ulang pembacaan terhadap titik ukur
meja tempat centrifuges yang sama (repeatability)/
dalam keadaan datar dan ketidakpastian tipe A. Karena hanya
tidak miring dilakukan 5 kali pengambilan data
ii. Pelabelan : periksa bahwa atau lebih < (lebih kecil) dari 10 data
semua keberadaan plakat, untuk tiap tiap titik pengukuran maka
label dan penunjuk masih untuk ketidakpastian daya ulang
sesuai dan terbaca pembacaan menggunakan rumus 2.2
a. Catat kondisi tersebut ( baik ketidakpastian pada pengukuran
atau tidak baik) pada lembar berulang
kerja 5. Hitung ketidakpastian daya
baca/resolusi
3. PEMBAHASAN 6. Hitung ketidakpastian
Analisa Data 7. Ketidakpastian standar
8. Ketidakpastiann gabungan
Adapun langkah-langkah kerja 9. Ketidakpastian yang diperluas/
penganalisaan adalah sebagai berikut : bentangan
1. Model matematik untuk keselahan
4. HASIL
Dari hasil penelitian terhadap kalibrasi
Dimana : Centrifuges yang dilaksanakan Di
K : Nilai kesalahan laboratorium Kelistrikan BPFK medan
P test : Penunjukan Centrifuges diperoleh hasil pada tabel dibawah ini.
P standar : Penunjukan Standar Merk :Centrifuges general
Model/type :HC-16D
2. Adapun sumber-sumber Nomor Seri :ZHO101S
Ketidakpastian Terdiri dari : Lokasi alat : Laboraturium BPFK
1. Ketidakpastian daya ulang
Pembacaan (repeatability)
2. Ketidakpastian daya baca
(resolusi)
3. Ketidakpastian standar dari
Sertifikat Kalibrasi
4. Ketidakpastian standar atau
acuan
3. Hitung nilai rata rata Hasil
pengukuran:
Rumus
Dimana :
x: Jumlah data pengukuran

29
Jurnal Mutiara Elektromedik 201 8,Vol. 2 No. 1
Jurnal Mutiara Elektromedik, 27-31

5. SIMPULAN 7. Alat dinyatakan layak pakai dengan


Dari hasil pemeriksaan yang dilakukan toleransi 10% untuk tiap titik
terhadap kalibrasi Centrifuges general yang pengukuran
ada di laboratorium kelistrikan BPFK Medan 8. Stamp kalibrasi yang di tempelkan
Diperoleh hasil dari enam data dinyatakan pada hali berlaku 1 tahun sejak stamp
layak pakai dan telah memenuhi syarat dan di tempelkan dan telah memenuhi
ketentuan berlaku. syarat dan ketetntuan berlaku
1. Pengukuran dari titik 2000-4000
RPM di peroleh hasil Pengukuran
rata- rata (0,50-3,83 RPM)
2. Pengukuran dari titik 2000-4000
RPM di peroleh hasil Pengukuran
Standar deviasi (1,64-4,00 RPM)
3. Pengukuran dari titik 2000-4000
RPM di peroleh hasil Pengukuran
Ketidakpastian Tipe A (0,25-0,66
RPM)
4. Pengukuran dari titik 2000-4000
RPM di peroleh hasil Pengukuran
Ketidakpastian Tachometer (0,55-
0,85 RPM)
5. Pengukuran dari titik 2000-4000
RPM di peroleh hasil Pengukuran
Ketidakpastian gabungan (0,67-0,94
RPM)
6. Dari hasil analia data yang di Bab IV
dinyatakan alat ini masih dibawah
toleransi 10% untuk setiap
pengukuran

30
Jurnal Mutiara Elektromedik 201 8,Vol. 2 No. 1
Jurnal Mutiara Elektromedik, 27-31

6. Spesifikasi tachometer
6. REFERENSI https;//lutron.id/shop/tachometer/t
1. KIM LIPI., 2016, Defenisi achometers-lutron-dt-2236/
kalibrasi, pusat penelitian
metrologi, Tangerang.
2. Ryan Andrian Pratama 2017, alat
kalibrasi centriduges non contact
menggunakan microcontroller
atmega 8, skripsi, Universitas
Muhamadiyah Yogyakarta
3. Djonosaputro, B Darmawan.,
1984, Teori Ketidakpastian
Menggunakan SI , Penerbit ITB
Bandung
4. Burhanudin, M 2011. Dasar
ketidakpastian pengukuran,
jakarta : Salemba Medika.
5. LIPI., 2016, ketidakpastian
pengukuran, Pusat Penelitian
Metrologi, Tangerang.

31
Jurnal Mutiara Elektromedik 201 8,Vol. 2 No. 1